This document compares the dosimetric outcomes of forward (F-IMRT) and inverse (I-IMRT) treatment planning for intensity-modulated radiotherapy (IMRT) in treating prostate cancer. Twenty patient plans were re-planned and evaluated using various dosimetric parameters. I-IMRT achieved better target coverage while F-IMRT resulted in more homogeneous dose distributions and was more treatment efficient due to fewer monitor units and shorter treatment time. Both techniques provided comparable organ-at-risk sparing. Further research is needed to evaluate dose escalation between the two methods.
The Advantages of Two Dimensional Techniques (2D) in Pituitary Adenoma TreatmentIOSR Journals
The purpose of the study is to evaluate the two dimensional dose distribution techniques in pituitary adenoma patient treatment in order to provide 2D dose coverage to the target volume while sparing organs at risk (OARs). The CT simulator was used to radiograph 300 patients of pituitary adenomas to conform 2D dose distribution planning inside the tumour bed , and its structures were delineated; including gross target volume (GTV), clinical target volume (CTV), and planning target volume (PTV)], as well as organs at risks (OARs) . Dose distribution analysis was edited to provide 2D dose coverage to the target while sparing organs at risk. The main results of the study were, 2D dose distribution plans increases the unnecessary dose to the critical organs according to their geographical location from the pituitary adenoma site, and the present study , concludes that when the tumour dose increases from 45 to 55 Gy there is a linear proportional increment of dose to the organs at risks, and when the dose is about 60 Gy in 2D, the increment of unnecessary dose to temporal lobe is 0.31 Gy, and to eye is0.34Gy, and to optic chiasm is 0.42 Gy respectively .New techniques, which will lessen the unnecessary dose to OARs, needed to be developed .
The Advantages of Two Dimensional Techniques (2D) in Pituitary Adenoma TreatmentIOSR Journals
The purpose of the study is to evaluate the two dimensional dose distribution techniques in pituitary adenoma patient treatment in order to provide 2D dose coverage to the target volume while sparing organs at risk (OARs). The CT simulator was used to radiograph 300 patients of pituitary adenomas to conform 2D dose distribution planning inside the tumour bed , and its structures were delineated; including gross target volume (GTV), clinical target volume (CTV), and planning target volume (PTV)], as well as organs at risks (OARs) . Dose distribution analysis was edited to provide 2D dose coverage to the target while sparing organs at risk. The main results of the study were, 2D dose distribution plans increases the unnecessary dose to the critical organs according to their geographical location from the pituitary adenoma site, and the present study , concludes that when the tumour dose increases from 45 to 55 Gy there is a linear proportional increment of dose to the organs at risks, and when the dose is about 60 Gy in 2D, the increment of unnecessary dose to temporal lobe is 0.31 Gy, and to eye is0.34Gy, and to optic chiasm is 0.42 Gy respectively .New techniques, which will lessen the unnecessary dose to OARs, needed to be developed .
The Computed Tomography (CT) dose output of some selected hospitals in the Federal capital Territory, Abuja, Nigeria have been determined by calculating the Effective doses of CT Chest and Abdomen-Pelvis of selected hospitals and compared its average with the Mean Reference Dose of CT Chest and Abdomen-Pelvis from four hospitals in the Federal Capital Territory, Abuja, Nigeria. Effective Dose and Scan type were extracted from the CT Chest and Abdomen-Pelvis examinations recorded. The Effective Dose of each patient undergoing the Chest and Abdomen-Pelvis examinations were calculated using the coefficient factor and the DLP values. Patients’ CT dose data from the ages of 18 to 60years from each of the 4 centres for each study type from January, 2013 to December, 2014 was extracted. A total of 112 patients’ CT dose data was extracted. Chest CT Effective Dose ranged from 9.0 to 34.0mSv, while Abdomen-Pelvis CT Effective Dose ranged from 15.9 to 61.0 for all the Centres in Federal Capital Territory, Abuja. This is higher than the recommended Reference Effective Dose range for CT Chest which is from 5 – 7mSv. and for CT Abdomen-Pelvis is from 8 – 14mSv. The mean effective dose from the Chest CT is 21.8mSv and from the Abdomen-Pelvis is 31.9mSv.
Adaptive Radiotherapy at GenesisCare UKGenesisCareUK
A discussion about adaptive radiation therapy, a closed-loop radiation treatment process and how it can improve radiation treatment by systematically monitoring treatment variations and incorporating them to re-optimize the treatment plan early on during the course of treatment.
High Risk disease is defined as “apparent localized cancer that has a high propensity of micro-metastatic disease” (cancer that is not visible on convention radiography, such as bone and CT scans). These cancers, once removed via radiation or surgery, are likely to "return," but in fact, they were never removed in the first place because the cancer cells were outside the treated region.
Therefore, successful eradication of high risk disease requires both aggressive local control and systemic treatment with androgen deprivation therapy and extended field radiation. This lecture will review the most up-to-date data on dose-intensity radiation therapy, pelvic radiation, surgery with adjuvant radiation, and adjuvant hormone therapy. Finally, data on experimental chemotherapy and abiraterone (Zytiga) will be presented.
The Computed Tomography (CT) dose output of some selected hospitals in the Federal capital Territory, Abuja, Nigeria have been determined by calculating the Effective doses of CT Chest and Abdomen-Pelvis of selected hospitals and compared its average with the Mean Reference Dose of CT Chest and Abdomen-Pelvis from four hospitals in the Federal Capital Territory, Abuja, Nigeria. Effective Dose and Scan type were extracted from the CT Chest and Abdomen-Pelvis examinations recorded. The Effective Dose of each patient undergoing the Chest and Abdomen-Pelvis examinations were calculated using the coefficient factor and the DLP values. Patients’ CT dose data from the ages of 18 to 60years from each of the 4 centres for each study type from January, 2013 to December, 2014 was extracted. A total of 112 patients’ CT dose data was extracted. Chest CT Effective Dose ranged from 9.0 to 34.0mSv, while Abdomen-Pelvis CT Effective Dose ranged from 15.9 to 61.0 for all the Centres in Federal Capital Territory, Abuja. This is higher than the recommended Reference Effective Dose range for CT Chest which is from 5 – 7mSv. and for CT Abdomen-Pelvis is from 8 – 14mSv. The mean effective dose from the Chest CT is 21.8mSv and from the Abdomen-Pelvis is 31.9mSv.
Adaptive Radiotherapy at GenesisCare UKGenesisCareUK
A discussion about adaptive radiation therapy, a closed-loop radiation treatment process and how it can improve radiation treatment by systematically monitoring treatment variations and incorporating them to re-optimize the treatment plan early on during the course of treatment.
High Risk disease is defined as “apparent localized cancer that has a high propensity of micro-metastatic disease” (cancer that is not visible on convention radiography, such as bone and CT scans). These cancers, once removed via radiation or surgery, are likely to "return," but in fact, they were never removed in the first place because the cancer cells were outside the treated region.
Therefore, successful eradication of high risk disease requires both aggressive local control and systemic treatment with androgen deprivation therapy and extended field radiation. This lecture will review the most up-to-date data on dose-intensity radiation therapy, pelvic radiation, surgery with adjuvant radiation, and adjuvant hormone therapy. Finally, data on experimental chemotherapy and abiraterone (Zytiga) will be presented.
Dosimetric Consequences of Intrafraction Variation of Tumor Motion in Lung St...semualkaira
The purpose of this study was to investigate the target dose discrepancy caused by intrafraction variation during Stereotactic Body Radiotherapy (SBRT) for lung cancer. Intensity-Modulated Radiation Therapy (IMRT) plans were designed based on Average Computed Tomography (AVG CT) utilizing the Planning Target Volume (PTV) surrounding the 65% and 85% prescription isodoses in both phantom and patient cases
Dosimetric Consequences of Intrafraction Variation of Tumor Motion in Lung St...semualkaira
The purpose of this study was to investigate the target dose discrepancy caused by intrafraction variation during Stereotactic Body Radiotherapy (SBRT) for lung cancer. Intensity-Modulated Radiation Therapy (IMRT) plans were designed based on Average Computed Tomography (AVG CT) utilizing the Planning Target Volume (PTV) surrounding the 65% and 85% prescription isodoses in both phantom and patient cases. Intrafraction variation was simulated by shifting the nominal plan isocenter along six directions from 0.5 mm to 4.5 mm with a 1-mm step size to produce a series of perturbed plans. The dose discrepancy between the initial plan and the perturbed plans was calculated as the percentage of the initial plan
Dosimetric Consequences of Intrafraction Variation of Tumor Motion in Lung St...semualkaira
The purpose of this study was to investigate the target dose discrepancy caused by intrafraction variation during Stereotactic Body Radiotherapy (SBRT) for lung cancer. Intensity-Modulated Radiation Therapy (IMRT) plans were designed based on Average Computed Tomography (AVG CT) utilizing the Planning Target Volume (PTV) surrounding the 65% and 85% prescription isodoses in both phantom and patient cases. Intrafraction variation was simulated by shifting the nominal plan isocenter along six directions from 0.5 mm to 4.5 mm with a 1-mm step size to produce a series of perturbed plans. The dose discrepancy between the initial plan and the perturbed plans was calculated as the percentage of the initial plan. Dose indices, including D99 and D95 for Internal Target Volume (ITV) and Gross Tumor Volume (GTV), were adopted as endpoint samples. The mean dose discrepancy was calculated under the 3-dimensional space distribution. In this study, we found that intrafraction motion can lead to serious dose degradation of the target and ITV in lung SBRT, especially during SBRT with PTV surrounding the lower isodose line. This phenomenon was compromised when 3-dimensional space distribution was considered. This result may provide a prospective reference for target dose degradation due to intrafraction motion during lung SBRT treatment.
The vmat vs other recent radiotherapy techniquesM'dee Phechudi
VMAT is a new type of intensity-modulated radiation therapy (IMRT) treatment technique that uses the same hardware (i.e. a digital linear accelerator) as used for IMRT or conformal treatment, but delivers the radiotherapy treatment using a rotational or arc geometry rather than several static beams.
This technique uses continuous modulation (i.e. moving the collimator leaves) of the multileaf collimator (MLC) fields, continuous change of the fluence rate (the intensity of the X rays) and gantry rotation speed across a single or multiple 360 degree rotations
Dose Evaluation in the Movement Couch of the Total Body Irradiation Technique...iosrjce
IOSR Journal of Applied Physics (IOSR-JAP) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of physics and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications in applied physics. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
The use of high frequency radiation to shrink tumor cells and kill cancer cells is Radiation Oncology. Austin Journal of Radiation Oncology and Cancer is an open access, peer reviewed scholarly journal committed to publication of unique contributions concerned with the cancer and its therapy.
Austin Journal of Radiation Oncology and Cancer accepts original research articles, review articles, case reports, clinical images and rapid communication on all the aspects of radiation therapy and oncology.
A short overview of Image Guided Radiotherapy process in Lung Cancer presented at TMC Kolkata circa 2016. Basic principles and concepts as well as examples are outlined.
Radiation Oncology in 21st Century - Changing the ParadigmsApollo Hospitals
Since its inception radiation therapy has been used as one of
the essential treatment options in the management of malignant and some benign tumors. With better understanding of tumor biology many new molecules have been added to the armamentarium of an oncologist. There is continuous improvement in surgical techniques with more emphasis on minimally invasive, organ- and function-preserving techniques. Neoadjuvant chemotherapy with or without addition of radiation therapy has helped surgeon downsizing the tumor and obtaining clearer margins.
Breast conserving surgery followed by adjuvant radiotherapy is adopted in the early detected cases and mastectomy followed by radiotherapy or chemotherapy in the advanced cases are the general practices.
Advances of Radiation Oncology in CancManagement: Vision for Role of Theranos...CrimsonpublishersCancer
Significant computational and technological advances in radiation therapy have enhanced our ability to more accurately plan and deliver increasing doses of radiation therapy to limited target volumes in many patients with cancer. Recent developments on magnetic resonance on-line imaging and use of implanted markers allow more precise on-time tumor localization with lower doses delivered to surrounding organs at risk leading to less treatment morbidity. Biological markers and molecular imaging (theranostics) will add new dimensions and precision to radiation therapy techniques. Nanoparticles are promising tools in therapeutic programs. Further research in efficacy, safety, cost utility (value) and institution of robust quality assurance programs will be necessary to optimize these contributions in clinical practice.
Intensity-modulated radiotherapy with simultaneous modulated accelerated boos...Enrique Moreno Gonzalez
To present our experience of intensity-modulated radiotherapy (IMRT) with simultaneous modulated accelerated radiotherapy (SMART) boost technique in patients with nasopharyngeal carcinoma (NPC).
An Examination of Effectuation Dimension as Financing Practice of Small and M...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Does Goods and Services Tax (GST) Leads to Indian Economic Development?iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Childhood Factors that influence success in later lifeiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Emotional Intelligence and Work Performance Relationship: A Study on Sales Pe...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Customer’s Acceptance of Internet Banking in Dubaiiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
A Study of Employee Satisfaction relating to Job Security & Working Hours amo...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Consumer Perspectives on Brand Preference: A Choice Based Model Approachiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Student`S Approach towards Social Network Sitesiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Broadcast Management in Nigeria: The systems approach as an imperativeiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
A Study on Retailer’s Perception on Soya Products with Special Reference to T...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
A Study Factors Influence on Organisation Citizenship Behaviour in Corporate ...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Consumers’ Behaviour on Sony Xperia: A Case Study on Bangladeshiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Design of a Balanced Scorecard on Nonprofit Organizations (Study on Yayasan P...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Public Sector Reforms and Outsourcing Services in Nigeria: An Empirical Evalu...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Media Innovations and its Impact on Brand awareness & Considerationiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Customer experience in supermarkets and hypermarkets – A comparative studyiosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Social Media and Small Businesses: A Combinational Strategic Approach under t...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Secretarial Performance and the Gender Question (A Study of Selected Tertiary...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Implementation of Quality Management principles at Zimbabwe Open University (...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Organizational Conflicts Management In Selected Organizaions In Lagos State, ...iosrjce
IOSR Journal of Business and Management (IOSR-JBM) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of business and managemant and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications inbusiness and management. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Richard's entangled aventures in wonderlandRichard Gill
Since the loophole-free Bell experiments of 2020 and the Nobel prizes in physics of 2022, critics of Bell's work have retreated to the fortress of super-determinism. Now, super-determinism is a derogatory word - it just means "determinism". Palmer, Hance and Hossenfelder argue that quantum mechanics and determinism are not incompatible, using a sophisticated mathematical construction based on a subtle thinning of allowed states and measurements in quantum mechanics, such that what is left appears to make Bell's argument fail, without altering the empirical predictions of quantum mechanics. I think however that it is a smoke screen, and the slogan "lost in math" comes to my mind. I will discuss some other recent disproofs of Bell's theorem using the language of causality based on causal graphs. Causal thinking is also central to law and justice. I will mention surprising connections to my work on serial killer nurse cases, in particular the Dutch case of Lucia de Berk and the current UK case of Lucy Letby.
Seminar of U.V. Spectroscopy by SAMIR PANDASAMIR PANDA
Spectroscopy is a branch of science dealing the study of interaction of electromagnetic radiation with matter.
Ultraviolet-visible spectroscopy refers to absorption spectroscopy or reflect spectroscopy in the UV-VIS spectral region.
Ultraviolet-visible spectroscopy is an analytical method that can measure the amount of light received by the analyte.
Professional air quality monitoring systems provide immediate, on-site data for analysis, compliance, and decision-making.
Monitor common gases, weather parameters, particulates.
Cancer cell metabolism: special Reference to Lactate PathwayAADYARAJPANDEY1
Normal Cell Metabolism:
Cellular respiration describes the series of steps that cells use to break down sugar and other chemicals to get the energy we need to function.
Energy is stored in the bonds of glucose and when glucose is broken down, much of that energy is released.
Cell utilize energy in the form of ATP.
The first step of respiration is called glycolysis. In a series of steps, glycolysis breaks glucose into two smaller molecules - a chemical called pyruvate. A small amount of ATP is formed during this process.
Most healthy cells continue the breakdown in a second process, called the Kreb's cycle. The Kreb's cycle allows cells to “burn” the pyruvates made in glycolysis to get more ATP.
The last step in the breakdown of glucose is called oxidative phosphorylation (Ox-Phos).
It takes place in specialized cell structures called mitochondria. This process produces a large amount of ATP. Importantly, cells need oxygen to complete oxidative phosphorylation.
If a cell completes only glycolysis, only 2 molecules of ATP are made per glucose. However, if the cell completes the entire respiration process (glycolysis - Kreb's - oxidative phosphorylation), about 36 molecules of ATP are created, giving it much more energy to use.
IN CANCER CELL:
Unlike healthy cells that "burn" the entire molecule of sugar to capture a large amount of energy as ATP, cancer cells are wasteful.
Cancer cells only partially break down sugar molecules. They overuse the first step of respiration, glycolysis. They frequently do not complete the second step, oxidative phosphorylation.
This results in only 2 molecules of ATP per each glucose molecule instead of the 36 or so ATPs healthy cells gain. As a result, cancer cells need to use a lot more sugar molecules to get enough energy to survive.
Unlike healthy cells that "burn" the entire molecule of sugar to capture a large amount of energy as ATP, cancer cells are wasteful.
Cancer cells only partially break down sugar molecules. They overuse the first step of respiration, glycolysis. They frequently do not complete the second step, oxidative phosphorylation.
This results in only 2 molecules of ATP per each glucose molecule instead of the 36 or so ATPs healthy cells gain. As a result, cancer cells need to use a lot more sugar molecules to get enough energy to survive.
introduction to WARBERG PHENOMENA:
WARBURG EFFECT Usually, cancer cells are highly glycolytic (glucose addiction) and take up more glucose than do normal cells from outside.
Otto Heinrich Warburg (; 8 October 1883 – 1 August 1970) In 1931 was awarded the Nobel Prize in Physiology for his "discovery of the nature and mode of action of the respiratory enzyme.
WARNBURG EFFECT : cancer cells under aerobic (well-oxygenated) conditions to metabolize glucose to lactate (aerobic glycolysis) is known as the Warburg effect. Warburg made the observation that tumor slices consume glucose and secrete lactate at a higher rate than normal tissues.
The increased availability of biomedical data, particularly in the public domain, offers the opportunity to better understand human health and to develop effective therapeutics for a wide range of unmet medical needs. However, data scientists remain stymied by the fact that data remain hard to find and to productively reuse because data and their metadata i) are wholly inaccessible, ii) are in non-standard or incompatible representations, iii) do not conform to community standards, and iv) have unclear or highly restricted terms and conditions that preclude legitimate reuse. These limitations require a rethink on data can be made machine and AI-ready - the key motivation behind the FAIR Guiding Principles. Concurrently, while recent efforts have explored the use of deep learning to fuse disparate data into predictive models for a wide range of biomedical applications, these models often fail even when the correct answer is already known, and fail to explain individual predictions in terms that data scientists can appreciate. These limitations suggest that new methods to produce practical artificial intelligence are still needed.
In this talk, I will discuss our work in (1) building an integrative knowledge infrastructure to prepare FAIR and "AI-ready" data and services along with (2) neurosymbolic AI methods to improve the quality of predictions and to generate plausible explanations. Attention is given to standards, platforms, and methods to wrangle knowledge into simple, but effective semantic and latent representations, and to make these available into standards-compliant and discoverable interfaces that can be used in model building, validation, and explanation. Our work, and those of others in the field, creates a baseline for building trustworthy and easy to deploy AI models in biomedicine.
Bio
Dr. Michel Dumontier is the Distinguished Professor of Data Science at Maastricht University, founder and executive director of the Institute of Data Science, and co-founder of the FAIR (Findable, Accessible, Interoperable and Reusable) data principles. His research explores socio-technological approaches for responsible discovery science, which includes collaborative multi-modal knowledge graphs, privacy-preserving distributed data mining, and AI methods for drug discovery and personalized medicine. His work is supported through the Dutch National Research Agenda, the Netherlands Organisation for Scientific Research, Horizon Europe, the European Open Science Cloud, the US National Institutes of Health, and a Marie-Curie Innovative Training Network. He is the editor-in-chief for the journal Data Science and is internationally recognized for his contributions in bioinformatics, biomedical informatics, and semantic technologies including ontologies and linked data.
Richard's aventures in two entangled wonderlandsRichard Gill
Since the loophole-free Bell experiments of 2020 and the Nobel prizes in physics of 2022, critics of Bell's work have retreated to the fortress of super-determinism. Now, super-determinism is a derogatory word - it just means "determinism". Palmer, Hance and Hossenfelder argue that quantum mechanics and determinism are not incompatible, using a sophisticated mathematical construction based on a subtle thinning of allowed states and measurements in quantum mechanics, such that what is left appears to make Bell's argument fail, without altering the empirical predictions of quantum mechanics. I think however that it is a smoke screen, and the slogan "lost in math" comes to my mind. I will discuss some other recent disproofs of Bell's theorem using the language of causality based on causal graphs. Causal thinking is also central to law and justice. I will mention surprising connections to my work on serial killer nurse cases, in particular the Dutch case of Lucia de Berk and the current UK case of Lucy Letby.
Comparative dosimetry of forward and inverse treatment planning for Intensity- Modulated Radiotherapy of prostate cancer
1. IOSR Journal of Applied Physics (IOSR-JAP)
e-ISSN: 2278-4861.Volume 7, Issue 6 Ver. I (Nov. - Dec. 2015), PP 97-106
www.iosrjournals
DOI: 10.9790/4861-076197106 www.iosrjournals.org 97 | Page
Comparative dosimetry of forward and inverse treatment
planning for Intensity- Modulated Radiotherapy of prostate
cancer
Maha Mokhtar1*
, Ehab M.Attalla1, 2
, Nashaat A. Deiab1, 2
, Ahmed Soltan3
,
H. Abou-Shady3
, Amr Amin 1, 2
1,2
Department of Radiotherapy and nuclear medicine, National Cancer Institute, Cairo University
3
Department of Biophysics, Faculty of Science, Cairo University
Abstract: Purpose: This study compared the dosimetric outcomes and treatment efficiency between forward
(F-IMRT) and inverse intensity modulated radiotherapy treatment technique (I-IMRT) for localized prostate
cancer.
Materials and methods: Twenty patients with localized prostate carcinoma were re-planned on Xio treatment
planning system (TPS) for forward and inverse intensity modulated radiotherapy (F-IMRT and I-IMRT).
Analyses were performed on comparing the dose volume histograms (DVHs) for Planning target volume (PTV)
and the relevant organ at risk (OARs). Target coverage was evaluated with parameters including (Dmean) the
mean target dose ,(Dmax) the maximum target dose, (D50%), (D98%), (D2%),(D5%) and (D95% ), the doses
that covered 50,98,2,5 and 95 % of the volume of the PTV respectively. Also, (V95%) the volume of the target
received 95% of prescribed dose (PD) and (V≤110%) the target volume received less than or equal 110% of
PD. Target dose distribution and conformality was evaluated with the homogeneity indices (HI) , Conformity
Index (CI) and uniformity indices(UI).Treatment efficiency was assessed using total number of monitor units
(MUs), total number of segments, integral dose, planning and treatment time. Normal tissue avoidance of OARs
was evaluated with using (D30) the dose received by 30% volume for rectum and bladder ,(V70), (V65), (V60)
and( V50); volume received at least an absorbed dose 70, 65,60 and 50 Gy respectively. Irradiated body volume
(IBV) at 10 Gy (V10), and 20 Gy (V20) were calculated. Also, (IBV36)and (IBV90) volume received 50and 90 %
of PD were calculated. Finally the integral dose was calculated .
Results:
The mean PTV D95% were 68.76±0.28 Gy, 68.92±0.5Gy for F- IMRT and I- IMRT respectively (p= 0.01),
D98 %was significantly higher for I-IMRT (68.40±7.3 Gy) plans compared to F-IMRT (68.10±7.1 Gy).
Target dose distribution was homogeneous in F-IMRT plans; however, I-IMRT plans had significantly higher CI
(p = 0.02). The mean volume values for rectum and bladder (V70 and V65) were similar and the difference was
not statistically significant in both IMRT plans. The mean volume for both femoral heads V50 was smaller in I-
IMRT than for F-IMRT planning (p = 0.005).The integral dose was statistically significant in I-IMRT compared
to F-IMRT p=0.001.
Conclusion: I-IMRT has achieved better PTV coverage, CI and offered greater degree of OARs sparing while
F-IMRT has better HI and improved the treatment efficiency due to small number of MUs and shorter treatment
time. IMRT planning increases the IBV at 10Gy or less but decreases the IBV at 20 Gy. Further research is
required to evaluate the dose escalation between F-IMRT and I-IMRT.
Keywords: Treatment planning system, Intensity-modulated radiotherapy, forward planning; Inverse planning,
dose volume histogram.
I. Introduction
Radiotherapy is one of cancer treatment modality where ionizing radiation is used to kill the cancer cells [1]
.
It is aimed to deliver maximum dose to the tumor while minimizing the dose to the surrounding unspecified
normal tissues [2]
. Before the time of Intensity modulated radiotherapy (IMRT) the conventional high energy
photon treatment was used. The introduction of the Computerized Tomographic (CT) in the 1980s and
Magnetic Resonance Imaging (MRI) in 1990s has given the radiation oncology a reliable three dimensional
(3D) overview of the patient anatomy where the soft tissues and tumors can be clearly outlined. Thus the need to
deliver conformal treatment fields became obvious and feasible. Three dimensions conformal radiotherapy
(3DCRT) is a change from traditional that uses target and normal tissues identified on serial transverse CT
images, field shape based on beam’s eye view(BEV) projections, volumetric dose calculations, and volumetric
plan evaluation tools such as (DVHs) AAPM report 82, 2003[3]
. DVHs display how much radiation is being
2. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 98 | Page
delivered to a specific volume of tissue and provide a quantitative evaluation of target coverage as well as dose
to normal tissues.
IMRT planning is an extension of 3DCRT that uses ionizing radiation beam intensities and determined by
various TPS based on optimization techniques [4]
. The success of IMRT planning in different tumor sites is
dependent on the accuracy of tumor delineation through different imaging modality [5]
.
Two treatment planning modality are generally applied for step and shoot IMRT, the first method is an
extension of three dimension conformal therapy (3-DCRT) and is defined as forward planning (F-IMRT).
It’s a very simplified form of IMRT which employ a few beams each with few segments S Webb, 2003[6]
. The
planner tried a variety of configurations of beams, wedges and more than one segment (1-3)is used from each
beam direction . These segment shapes are adjusted manually after that, the dose weights of the segments are
optimized in trial and error by the planner using a computer algorithm until a suitable homogenous dose match
is found to the desired dose prescription. The clinical implementation using the F- IMRT planning is relatively
easy, because it is closely related to conventional planning. The clinical implementation of IMRT using forward
planning is relatively easy, because it is closely related to conventional planning.
The other technique is Intensity Modulated Radiation Therapy (IMRT) which we denote as inverse planning .
It’s a fixed delivery technique contain a set of intensity modulated beams enters the patient from multiple
directions, which the multi-leaf collimator(MLC) allowed to move in between the segments when the beam is
off (Step and shoot). The concept of inverse planning IMRT is that, dose distributions are inversely
determined, meaning that the treatment planner must specify the dose distribution that is desired.
The method for calculating the required intensity fluence from each beam segment is done by inverse treatment
planning with high computer system using an iterative process to calculate doses with algorithms, starting with
the desired result( optimization of fluence map) and then working backwards to generate an optimal way to
reach the final goal(desired dose distribution).
Most I-IMRT planning computer systems allow a specification of dose-volume constraints and or dose limits S
Webb, 2003[6]
. Inverse planning is far less related to conventional radiotherapy because the segment shapes are
not defined manually ; However, there are complex clinical situations, which require the use of many beam
directions and segments; though, in these cases, inverse planning may be the more efficient strategy. In I-
IMRT, the number of MUs delivered has increased , reduced the hot spot dose[7]
,increase the integral dose to
normal tissues[ 8-9]
and these has increases the risk of secondary malignancy cancer [ 10]
compared to 3DCRT. In
addition the prolonged treatment time of I-IMRT may affect the treatment accuracy due to increasing the
chances of internal organ movement owing to volume changes ( rectum and bladder) , and patient respiration
over time [11]
. The medical physicist should verify the actual dose delivered to the patient before I-IMRT
treatment. AAPM TG119 [12]
guide line are available on how to verify IMRT plans. The quality assurance (QA),
time involved in planning and delivery are a logical extension of the experience obtained with conformal
radiotherapy. Until The International Commission on Radiation Units and Measurements ( ICRU) 83[13]
Report
was released in 2010 for IMRT only and introducing some different concepts and plan evaluation parameters ,
ICRU 50 [14]
and ICRU 62 [15]
Reports were widely used as an international reference for the prescribing,
recording, and reporting of photon beam radiotherapy, including 3DCRT and IMRT.
The aim of the present study is to compare dosimetric quantitative and qualitative in forward-planning (field-in-
field 3D conformal radiation therapy) and inverse-planning IMRT techniques regarding Physical DVH
treatment parameters (Target +OAR) and dosimetric parameters as target dose distribution, coverage volume ,
normal tissue avoidance, Irradiated body volume (IBV),mean and maximum dose, Homogeneity Index (HI),
Conformal number , Conformity Index (CI),Uniformity index(UI) , Over dose Index (ODI) ,Mu, delivery time
integral dose and treatment efficiency.
II. Materials and Methods
Twenty patients with localized prostate cancer were CT scanned (General Electric (GE) in supine position. All
images obtained with 2.5 mm slice thickness. The PTV and OARs delineation was performed by the radiation
oncologist. Two plans F-IMRT and I-IMRT ( PD =72Gy /36 fraction) made on XIO TPS (CMS; Elekta, version
4.4). All I-IMRT plans were consisted , seven coplanar fields, with 0°, 51°, 100°, 151°, 202°, 253° and 304°
gantry angles ;while, in F-IMRT 5 fields with 0°, 51°, 90°, 270° and 315° gantry angle .
All treatments were delivered by ELEKTA precise (6MV) .
Data collection and plan evaluation
The plans were evaluated qualitatively by comparing, the dose distribution through the patient volume (cut-by-
cut) and quantitatively with the use of DVHs . The maximum dose, mean dose and a set of values ( Dx%) the
percentage dose received by the x% volume of the target volume and (Vx % ) the percentage volume irradiated
by x% of the PD ,were obtained for the OARs. To achieve the target coverage and normal tissue sparing the
following parameters were used.
3. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 99 | Page
HI and CI indexes
Dose homogeneity and dose conformity are independent specifications for evaluating plan quality.
Dose homogeneity characterizes the uniformity of dose distribution within the target volume[16]
and was defined
as the difference between the maximum and minimum dose normalized to the median dose[7]
.
HI = ( D 2% - D 98%) / D50%
Where D2 (maximum), D98 (minimum) and D50 (the median dose to 50%of target volume) represent the doses
received by 2, 98 and 50 %volumes of PTV respectively, and HI = Zero is ideal value.
Dose conformity specifies the degree which the high dose region conforms to the target volume [16]
and it used
for comparison the degree of conformity between plans and was calculated in term of the tumor volume
enclosed by 95% isodose line to the PTV volume (V95% / VPTV), and CI=1 is ideal, Therapy and Oncology
Group RTOG [17-19]
.
Uniformity (UI) and Over dose (OD) Indexes
The dose UI was defined as UI = D5 / D95, Where D5 and D95 are the dose delivered to 5% and 95% of PTV
volume, respectively, [20-21]
.
The ODI was defined as PTV covered by 105% isodose line to PTV volume (V105% / VPTV).
OARs
The degree of organ sparing was compared using DVHs for all patients with F-IMRT and I-IMRT plans.
The guideline for OARs evaluation were described in QUANTEC Group report by Marks, 2010[22]
table (1).
No plan was accepted with hot spot along the rectal wall and bladder to reduce the toxicity risk.
Table (1) Dose constraints to OARs.
Structure Constrains
Bladder V70≤ 25%
V65≤ 50%
Rectum V50<50%
V60<35%
V65<25%
V70<15%
Right and left femoral head V50<10%
Healthy tissues
To find the lower dose to normal unspecified tissue, (IBV10 and IBV20) the normal tissue volume receiving
radiation dose more than 10 Gy and 20 Gy[23-24]
respectively were calculated. To estimate the dose outside PTV,
( IBV36 and IBV90) the normal tissue volume received 50% and 90% of PD respectively were calculated .
To evaluate the dose to the healthy tissues, the concept of integral dose (ID) was used.
Integral dose is defined as the total energy absorbed by the body, multiplied by volume and computed based on
the average organ density, averaged organ dose, and volume.
ID is defined as Integral dose = D .ρ.V (Gy.Kg), Where D, ρ and V are averaged organ dose, averaged organ
density and volume, respectively [23]
. In this study the integral dose is simplicity calculated as:
Integral Dose =Average Dose × Volume (unspecified tissue) (Gy.cc)
ID is defined as Integral dose = D .ρ.V (Gy.Kg), Where D, ρ and V are averaged organ dose, averaged organ
density and volume, respectively [23]
. In this study the integral dose is simplicity calculated as:
Integral Dose=Average Dose × Volume (Gy.cc)
Treatment efficiency
To achieve the treatment efficiency, the mean average segments number, mean MUs, and mean
average treatment time were assessed. Both techniques underwent a “dummy run”, to simulate the treatment
time of a real patient. The treatment time included the time for planning, radiation delivery, in addition to the
time required for QA plan verification.
Statistical analysis
Data were analyzed using SPSS win statistical package version 22. Numerical data were summarized as
means and standard deviations (SD) . The Wilcoxon’s matched-pairs test was used to determine statistical
differences between volumes and doses in F-IMRT vs I-IMRT plans. The dose–volume parameters of target
4. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 100 | Page
volumes and OARs were also measured and compared. Probability (p-value) equal or less than 0.05 is
considered significant.
III. Results
Both I-IMRT and F-IMRT techniques had good result regarding PTV coverage, while I-IMRT had
shown a slightly sharper dose gradient than the F-IMRT plans. Figure (1) showed the isodose distribution of the
F-IMRT and I-IMRT techniques in transverse CT section.
Target coverage, dose Conformity and dose Homogeneity index
Tables (2- 3) summarized the dosimetric results of the PTV. The planning objective in both techniques
were achieved nearly the same result; however ,the differences were statistically not significant in term of
Dmax (75.1±1.53 ,76.3±1.58Gy), D mean (71.48±0.72,71.98±0.78Gy), D50%( 71.82± 0.87,72.2± 0.80 Gy ) ,
D5% (74.33±1.42 ,73.71±1.16Gy) , V≤110 (2.95±2.03.1.99±2.83 %), ODI( 0.1±0.03 , 0.13±0.02 %) and
UI(1.08±0.46, 1.06±0.26 Gy) in F-IMR and I-IMRT respectively .
I-IMRT showed a slightly considerably difference in PTV dose coverage with V95% , D98% , D95, D2 and
the differences were statistically significant compared to F-IMRT table (2).
F-IMRT had better acceptable dose homogeneity (0.097±0.054 Gy) compared to I-IMRT (0.095±0.17 Gy) and
the differences was statistically significant table (3). The isodose distribution shown a bit better conformity to
the PTV in I-IMRT(0.964± 0.011) compared with F-IMRT (0.961±0.010 ) plans and the differences was
statistical significant (p=0.02). I- IMRT reduced the hotspots volume by 4% compared to F-IMRT table (2);
however, the difference was statistically not ssignificant.
Figure 1 the isodose distribution in the transverse CT section (a) F-IMRT and (b) I-IMRT plans.
Organ at risk (OARs)
Table (4 ) and figure 2(b-e) summarize the dosimetric result of OARs.
5. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 101 | Page
Both techniques achieved organ sparing according to QUANTIC [22]
guide constraint ,table (1).
The DVHs analysis for rectum and both femoral heads showed that I-IMRT were better than F-IMRT technique
while in bladder both techniques had similar result; however, these differences were statistically not significant.
The mean dose for rectum (D30% ) was (56.36±9.36, 47.14± 9.96Gy) for F-IMRT and I-IMRT respectively,
and the differences were statistically significant table(4) . D30% for bladder was (10.16± 18.62, 10.90±20.04
Gy) with F-IMRT and I-IMRT respectively; however, the differences were statistically not significant.
Regarding the rectum and bladder volumes received 70, 65, 60, 50 Gy, there was a small difference between
both techniques; however, the differences were statically not significant table (4).
The average mean doses for right femoral head were (30.39 ±6.43, 25.12 ±5.07Gy) for F-IMRT and I-IMRT
respectively ;however, the difference were statistically not significant while, the average mean dose (V50)
were (2.99±4.4, 0.24±0.86%) for F-IMRT and I-IMRT respectively and (p=0.003) .
The average mean dose for left femoral head were (30.71 ±5.19, 24.77 ±6.38Gy) for F-IMRT and I-IMRT
respectively ;however, the differences were statically not significant while, the mean dose (V50) were (3.03 ±
4.25 , 0.14 ± 0.32%) for F-IMRT and I-IMRT respectively,(p=0.06) table (4) .
Table (2) Dmax (Gy), Dmean (Gy), D95% (%), D98% (%),D2%, D5% (%),(%) , V≤110% (%)plans, where Dmax :
maximum dose to, Dmean: mean dose, D min : minimum dose for PTV in F-IMRT and I-IMRT.
Parameter
Dmax (Gy) Dmean (Gy) D50% (Gy) D98% (Gy)
F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT
Mean ± SD 75.1±1.53 76.3±1.58 71.48±0.72 71.98±0.78 71.87± 0.53 72.21±0.8 67.19±0.62 67.75±0.95
P-Value 0.06 0.08 0.12 0.04*
Parameter
D2% (Gy) D5% (Gy) V95% (%) V≤110% (%)
F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT
Mean ± SD 74.23±1.27 74.61±1.46 74.33±1.42 73.71±1.16 97.04±1.002 97.64±1.09 2.95±2.03 1.99± 2.83
P-Value 0.021* 0.19 0.04* 0.1
Parameter Dmin (Gy) D95% (Gy) V90% (%)
F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT
Mean ± SD
58.35±3.9 60.2±2.25 68.76±0.28 68.99±0.54 99.23±4.9 99.49±5.1
P-Value 0.03* 0.01* ≤0.001*
*P-values < 0.05.
Table (3) the homogeneity index (HI), conformity number (CI), monitor units (MU), number of segments,
uniformity index (UI), over dose index, planning time and treatment time in F-IMRT and I-IMRT plans.
Parameter HI CI UI ODI
F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT
Mean ± SD 0.097±0.054 0.095±0.17 0.961±0.010 0.964± 0.011 1.08±0.017 1.06±0.29 0.10±0.03 0.13±0.02
P-Value 0.008* 0.02*
0.232 0. 1
Parameter
MUs Number of segments Planning time Treatment time
F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT F-IMRT I-IMRT
Mean ± SD 665.64±57.44 719.6 ±103.5 5.6±1.71 91.38±9.06 40±9.5 65±13.4 22.5±7.7 38±9
P-Value 0.04*
≤ 0.001*
0.012*
≤ 0.001*
*P-values < 0.05.
6. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 102 | Page
Table (4).The dose–volume parameters for the different OARs for F-IMRT and I-IMRT plans. Where Dmean:
mean dose to organ
OARs DVH parameter F-IMRT I-IMRT P -Value
Rectum
D mean (Gy) 42.58±7.59 40.58±5.61 0.3
D 30%(Gy) 56.36±9.36 47.14± 9.96 0.05*
V50Gy (%) 37.15±12.27 29.74±8.28 0.08
V60Gy (%) 26.7±9.8 17.9±6.47 0.1
V65Gy (%) 20.54±8.59 11.87±5.1 0.1
V70Gy (%) 9.02±7.38 4.82±3.2 0.2
Bladder
D mean (Gy) 40.30±9.5 40.30±6.21 0.5
D 30%(Gy) 10.16±18.62 10.90±20.04 0.2
V65Gy (%) 22.56±8.78 18.37±5.22 0.2
V70Gy (%) 11.01±6.47 10.74±4.5 0.07
RT head of femur
D mean (Gy) 30.39 ±6.43 25.12±5.07 0.06
V50Gy (%) 2.99±4.4 0.24±0.86 0.003*
LT head of femur
D mean (Gy) 30.71±5.19 24.77±6.38 0.06
V50Gy (%) 3.03 ± 4.25 0.14 ± 0.32 0.005*
*P-values < 0.05.
Table (5) Dosimetric parameters for the normal healthy tissue represented as mean and standard deviation (SD).
DVH Parameters F- IMRT (Mean ± SD) I- IMRT( Mean ± SD) P-Value
D mean (cGy) 412.4±139.57 480.25±167.28 ≤0.001*
IrV10 (%) 16.3 ±5.82 19.37 ±6.85 0.04*
IrV20 (%) 10.19 ±3.78 13.48 ±4.97 ≤0.001*
IrV36 (%) 5.65 ±2.4 6.37 ±2.97 0.003*
IrV90 (%) 0.27 ±1.13 0.01 ±0.019 ≤0.001*
Integral dose(cGy) 687.59±425 800.72±608 0.001*
*P-values < 0.05.
Treatment MUs, MLC segment, Integral dose for normal healthy tissue, treatment time and time
efficiency
Table( 3 ) showed that, the average planning time 65 ±13.4 minutes (range 40–75 minutes) was longer
in I-IMRT compared to F-IMRT 40 ±9.5 minutes (range 30–65 minutes) and the difference was statistically
significant. The mean delivery treatment time of I- IMRT (38 ±9 min) was longer than F-IMRT (22.5 ±7.7 min)
and the difference was statistically significant.
The average number of segment was (5.6±1.71, 91.38±7.06) for F-IMRT and I-IMRT respectively, table (2),
and this was statistically significant (p≤ 0.001). I-IMRT increased the number of MUs (719.6 ±103.5 MUs)
compared to F-IMRT (665.64± 57.44MUs) . I-IMRT increased the number of segment by factor15 compared to
F-IMRT .The larger number of MUs in I-IMRT could lead to higher integral, dose table (5) and this was
statistically significant (p=0.04). I-IMRT reduced the hot spot by factor 4% owing to beam modulation during
plan optimization.
Regarding unspecified normal tissue , the mean average volume was smaller with F-IMRT compared to I-
IMRT table (5) in lower radiation dose region (IBV10, IBV20 ) and, the differences were statistically
significant. The mean irradiated volume received 20Gy was smaller than 10 Gy .
In middle radiation dose region (IBV36), F-IMRT was smaller, with bigger normal tissue volume received 50%
of prescribed dose compared to I-IMRT and the difference was statistically significant table(5).
In higher dose region (IBV90), the treated normal tissue volume was higher in F-IMRT compared to I-IMRT
and the difference was statistically significant table (5).
7. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 103 | Page
Figure 2(a-e) Mean DVH for (a) PTV, (b) Rectum, (c) Bladder, (d) RT femur and (e) LT femur.
IV. Discussion
Target coverage, dose conformity and dose homogeneity index
The employment of newer and complex technologies for the treatment of tumors improve target coverage while
keeping normal tissue doses under known thresholds to avoid the complications ;however, conventional forward
technique is used in some cases as pelvis and lungs which have not small PTV margins and dose delivered in
one or two phases according to the plan complexity . Also, it is used in some head and neck cases with certain
beam arrangement and the high dose delivered in phases to save OARs.
it was observed, both techniques rendered similar and nearly equivalent target dose volume distributions that
was able to reach our planning goals and reduce the dose spillage to the healthy tissues ; however, the results
showed that I-IMRT achieved better target dose coverage , higher degree of conformality for the concavity PTV
and reduced high dose region to the surrounding normal tissue and rectal wall.
I-IMRT optimization provided a uniform dose distribution, and produced greater dose inhomogeneity through
steep dose gradients in target volume compared with F-IMRT(P=0.841). The greater dose inhomogeneity was
due to ideal optimized I- IMRT plans which, cannot be executed with the use of existing multileaf collimator
systems. Also, the isocenter (reference point) was placed anywhere inside the treated volume including those
locations that may be near a low-dose region or inside an normal tissue where, it used for positioning the patient
in the treatment machine and this was critical for dose delivery [43]
. our results were met ICRU 83 guidelines .
According to the RTOG and ICRU 83 guidelines, the CI in I-IMRT has able to conform the dose distribution to
the concavity of target volume and achieved improvement with better sparing of critical and uninvolved
surrounding structures. Our results were in contrast to the study reported by Fisher [24]
and consistent with other
studies reported by[25- 26]
.This contradiction may be due to variation in tumor shape and relevant OARs.
I- IMRT reduced hot dose spots in, due to beam modulation during optimization and our result is consistent
with S. Moorthy[7]
.
Treatment MUs, MLC segment and Integral dose for normal healthy tissue
F-IMRT has reduced MUs by approximately 7.53% due to small number of beams and segments compared to F-
IMRT.
a b
c
ed
8. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 104 | Page
F-IMRT has reduced the number of segment and our result was consistent with J. Luc [27]
who reported that,
reducing number of segment could limit the degree of intensity modulation and compromise the dose
distribution.
The mean delivery time of I-IMRT was longer than F-IMRT owing to the larger number of beams, segments,
MUs delivered and the dead time in the gantry rotation from field to field.
F-IMRT has shown to spare more healthy tissue in low dose region at V10Gy and V20Gy compared to I-IMRT.
Similar results were showed that the large percentage of healthy tissue received 10-20Gy of radiation with
3DCRT [25- 26, 28]
. I-IMRT had statistically significant reduction in the volume of healthy tissue irradiated in the
medium dose region (IrV36).I-IMRT has achieved better sparing to normal healthy tissue in high dose region
and showed greater dose spillage outside PTV.
F-IMRT has reduced the integral dose by 14% compared with I-IMRT, our result were consistent with Bland,
et al. and Pirzkall, et al. [8-9]
and in contrast to Shirani, et al. [26]
.
Aoyama, et al. reported that a larger number of MUs would result in a higher integral dose [29]
; however,
Leire, et al. [25]
reported that, the integral dose does not depend on the number of MUs only[29]
, but also, depends
on the target volumes and shapes.
OARs sparing
Both techniques achieved QUANTEC guide lines [22]
and other studies[8]
which recommended that, the relative
volume received 70 Gy (V70) for rectum was less than 20% of total rectal volume and, (V70) for the bladder
[12]
was less than 35%.
I-IMRT decreased the average volume of OARs received high dose by 3.6% (range 0.27-8.8 %).
Regarding the DVHs, figure2(b-e) ,I-IMRT has spared OARs better in high dose region and have a maximum
protection compared to F-IMRT, because in F-IMRT, it should be covered the PTV firstly due to the limitation
dose in OARs.
I-IMRT technique was better in reducing the dose in the high-dose region in bladder; however, F-IMRT reduced
the dose in the low- and medium-dose regions.
I-IMRT reduced the high doses in rectal wall in all patients compared to F-IMRT.
It was reported that the higher radiation dose delivered to small volume of rectum was the primary cause of late
toxicities or late rectal bleeding.
Based on the dosimetric results for the rectum in this study, patients treated with I-IMRT were suffered less
rectal complications; however, the relation between toxicity and rectal dose-volume with I-IMRT would be
investigated with long-term follow up.
Regarding dose-volume relationship, the study has presented the constraints on localization of the OARs during
treatment, as well as the duration of follow-up necessary to determine their toxicity.
Lou, et al. [27]
compared the dosimetry of 3DCRT and IMRT plans based on dose-volume histograms DVHs for
rectum and bladder. The daily variation of rectum and bladder volumes and dose to these structures has been
shown to vary considerably for a given patient during the whole course of the treatment. The DVHs for rectum
and bladder were varied between institutions [20]
because volumes of these structures on planning CT varied
between patients. These variation were independent among institutions , due to the differences in the
institution’s planning criteria and techniques. It was observed that F-IMRT spared rectum and bladder better in
low dose region compared I-IMRT. Although the difficulties in determining the relation between bladder dose-
volume and toxicity, QUANTEC guide lines recommended that no more than 35% and 50% of the bladder
volume (bladder + wallcontents) receive a dose greater than 70 and 65 Gy, and this recommendation was based
on the treatment outcome of patients treated with 3-DCRT. our result were met these constraint.
IMRT were achieved smaller volume (V50) in both femoral heads than F-IMRT due to beam arrangement in
optimization and a choice of treatment technique was a significant predictor for femoral head necrosis.
Treatment planning time and treatment efficiency
The average treatment planning time in I-IMRT ( automated segments) has required 65 min which was work
loaded for the planner due to, computational time for optimization process and planning algorithms .The
planning time would be shortened as the planner had more experience .
The mean treatment time of I-IMRT was longer due to the number of beams, dead time in the gantry rotation
from field to field ,number of segments and the larger number of MUs delivered.
F-IMRT had an advantage in shorting the treatment time, it was usefully not only in reduction the chance of
the intra-fractional motion of patients, but also decreased the internal organ motion and also, reduction the time
that the patients required to maintain a full bladder. Another advantage it was necessary for work loaded
radiotherapy centers and thus decrease the patient waiting list.
9. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 105 | Page
F-IMRT has reduced MUs by approximately 7.53% due to small number of beams and segments compared to
F-IMRT ,and our result was consistent with J. Luc [27]
who reported that, reducing number of segment could
limit the degree of intensity modulation and compromise the dose distribution.
The mean delivery time of I-IMRT was longer than F-IMRT owing to the larger number of beams, segments ,
MUs delivered and the dead time in the gantry rotation from field to field.
F-IMRT has shown to spare more healthy tissue in low dose region at V10Gy and V20Gy compared to I-IMRT.
Similar results showed that the large percentage of healthy tissue received 10-20Gy of radiation with 3DCRT [25-
26, 28]
. I-IMRT had statistically significant reduction in the volume of healthy tissue irradiated in the medium
dose region (IBV36).I-IMRT has achieved better sparing to normal healthy tissue in high dose region and
showed greater dose spillage outside PTV.
F-IMRT has reduced the integral dose by 14% compared with I-IMRT, our result were consistent with Bland,
et al. and Pirzkall, et al. [8-9] and in contrast to Shirani, et al. [26]
.
Aoyama, et al . reported that a larger number of MUs would result in a higher integral dose[ 29 ]; however,
Leire ,et al.[25] reported that, the integral dose does not depend on the number of MUs only[29] , but also,
depends on the target volumes and shapes.
The higher number of MUs in I-IMRT had the effect in increasing the risk of a secondary malignancy due to
an increase in amount of radiation transmitted, scatter and leakage through multi-leaf collimator (MLC). This
result was consistent with other study [27, 29, 30-31]. Kry, et al.[32] reported that I-IMRT may be increased the
10-year incidence of second malignancies from 1% in patients treated with 3DCRT to 1.75% in patients treated
with IMRT.
V. Conclusion
I-IMRT is a better technique compared to F-IMRT but it need some margins to improve and optimize
the dose distributions resulting an improvement PTV coverage
I-IMRT has achieved better CI and PTV coverage. It also has the ability to preserve more OARs at the
expense of decreasing target homogeneity, increases the delivered integral dose resulting in, increasing the risk
of secondary malignancy. Besides, it also achieved reduction in the volume of normal healthy tissue receiving
high dose (IBV90).
On the other hand, F- IMRT has provided better conformality dose distribution, decreased the number
of MUs. It also, irradiated smaller volume of normal healthy tissue in the low to medium dose region and
delivered lower integral doses.
F-IMRT has smaller number of MUs and thus shortens the treatment time was delivered than I-IMRT
which was more comfortable for the patient and useful in decreasing the machine work load in busy
department.
In order to achieve similar results of I-IMRT in term of target coverage and dose homogeneity
distribution, mixed energies ( 6 and 15 MV) are needed.
To choose between the two techniques it is necessary to see the patients’ characteristics case by case.
when the radiation oncologist prescribed a total dose to PTV ≤ 72 Gy, it is useful to choose the conventional
forward technique, while in cases of high dose to PTV ( ≥74 Gy) further dosimetric study is required to
evaluate dose escalation , clinical benefits and treatment efficiency.
References
[1]. Brahme, A. “Optimization of stationary and moving beam radiation therapy techniques.” Radiother. Oncol. 12(2):129–140 ,1985.
[2]. National Cancer Institute. Prostate Cancer Screening. Available at:
http://www.cancer.gov/cancertopics/pdq/screening/prostate/HealthProfessional/page2. Accessed March, 2013.
[3]. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of
the AAPM radiation therapy committee. AAPM Report 82, 2003.
[4]. Wui-Jin Koh and David H.: Moore, Cervical cancer, Gunderson &Tepper, Clinical Radiation Oncology; Chapter 56, 2012.
[5]. Benedick A. Fraass, Shruti Jolly, and Avraham Eisbruch.: Conformal Therapy and Intensity Modulated Radiation Therapy:
Treatment Planning, Treatment Delivery, and Clinical Results, Gunderson & Tepper, Clinical Radiation Oncology; Chapter 15,
2012.
[6]. S Webb the physical basis of IMRT and inverse planning. British Journal of Radiology; 76; 678-689, 2003.
[7]. Suresh Moorthy, M.Phil,. Narayana Murthy, Saroj Kumar Das Majumdar, MD, DNB, Hamdy El Hateer, R. Mohan, V.
Ramanathan, B.Sc, Dosimetric Characteristics of IMRT versus 3DCRT for Intact Breast Irradiation with Simultaneous Integrated
Boost, Austral - Asian Journal of Cancer ;ISSN-0972-2556, Vol. 11, No. 3, July 2012.
[8]. Bland MJ, Altman DJ, Statistical method for assessment agreement between two method of clinical measurements, Lance; 1: 307-
10 , 1986.
[9]. Pirzkall A. Carol M, Comparison of intensity modulated radiotherapy with conventional conformal radiotherapy for complex shape
tumor, Int J radiant, Oncol boil phy; 67: 1135-44,2007.
[10]. Hall EJ and Wuu CS. Radiation-induced second cancers: The impact of 3D-CRT and IMRT. Int J Radiat Oncol Biol Phys; 56(1):
83–88, 2003.
[11]. Huang E, Dong L, Chandra A, Kuban DA, Rosen II, Evans A and Pollack A. Intrafraction prostate motion during IMRT for prostate
cancer. Int J Radiat Oncol Biol Phys; 53(2): 261–268, 2002.
[12]. IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119, 2009.
10. Comparative dosimetry of forward and inverse treatment planning for Intensity
DOI: 10.9790/4861-076197106 www.iosrjournals.org 106 | Page
[13]. Prescribing, Recording, and Reporting Intensity-Modulated Photon-Beam Therapy (IMRT), ICRU Report 83, Journal of the ICRU;
Vol. 10, No. 1, 2010.
[14]. International Commission on Radiation Units and Measurements ICRU report 50: prescribing, recording, and reporting photon
beam therapy. Bethesda: ICRU, 1993.
[15]. International Commission on Radiation Units and measurements ICRU report 62: prescribing, recording, and reporting photon
beam therapy (supplement to ICRU report 50). Bethesda: ICRU,1999
[16]. Shuang Luan, Chao Wang, Danny Z. Chen, A new MLC segmentation algorithm/software for step-and-shoot IMRT delivery,
Medical Physics; Vol. 31, No. 4, April 2004.
[17]. Feuvret L, Noel G, Mazeron JJ and Bey P. Conformity index: a review. Int J Radiat Oncol Biol Phys; 64(2): 333–342, 2006.
[18]. Huchet A, Caudry M, Belkacémi Y, Trouette R, Vendrely V, Causse N, Récaldini L, Atlan D and Maire JP. Volume-effect and
radiotherapy Part II: Volume-effect and normal tissue. Cancer Radiother; 7(5): 353–362, 2003.
[19]. Shaw E, Scott C, Souhami L, Dinapoli R, Kline R, Loeffler J and Farnan N. Single dose radiosurgical treatment of recurrent
previously irradiated primary brain tumors and brain metastases: Final report of RTOG protocol 90-05. Int J Radiat Oncol Biol
Phys; 47(2): 291–298, 2000.
[20]. Wang X, Zhang X, Dong L, et al. Effectiveness of non-coplanar IMRT planning using a parallelized multiresolution beam angle
optimization method for paranasal sinus carcinoma. Int J Radiat Oncol Biol Phys.; 63(2):594-601, 2005.
[21]. Sheng K, Molloy JA, Larner JM, et al. A dosimetric comparison of non-coplanar IMRT versus Helical Tomotherapy for nasal
cavity and paranasal sinus cancer. Radiother Oncol.; 82(2):174-8, 2007.
[22]. Marks LB, “Use Of Normal Tissue Complication Probability Models in the Clinic,” Int. J. Radiation Oncology Biol. Phys.; Vol. 76,
No. 3, Supplement, pp. S10–S19, 2010.
[23]. Chengyu Shi, Jose Penagaricano, Comparison of IMRT treatments plan between linac and helical tomotherapy based on integral
dose and inhomogeneity index, medical dosimetry; vol. 33, No.3, PP. 215-221,2008.
[24]. Fisher J, Scott C, Stevens R,Marconi B, Champion L , Freedman GM, et al. Randomized phase III study comparing best supportive
care to Biafine as a prophylactic agent for radiation induced skin toxicity for women undergoing breast irradiation: Radiation
Therapy Oncology Group (RTO G) 97-13. Int J Radiat Oncol Biol, Phys; 48: 1307–1310, 2000.
[25]. Leire Arbea, Luis Isaac Ramos, Rafael Martínez-Monge, Marta Moreno, Javier AristuLeire. Intensity-modulated radiation therapy
(IMRT) vs.3D conformal radiotherapy (3DCRT) in locally advanced rectal cancer (LARC): dosimetric comparison and clinical
implications Radiation Oncology; 5:17, 2010.
[26]. Shirani K , Nedaie H A , Banaee N 3 , Hassani H , Samiei F , Hajilooei F. Evaluation and comparison of dosimetric parameters in
PTV for prostate cancer via step and shoot IMRT and 3DCRT , journal of advances in physics ;Vol. 6, no.1,2014.
[27]. LucJ.Bos,Marco Schwarz,Werner Bar ,Markus Alber , Ben J Mijnheeer,Joos V. Lebesque and Eugene M.F.Demen "Comparison
between manual and automatic segment generation in step-and-shoot IMRT of prostate cancer “Journal of med. phys.;31 (1)
January 2004.
[28]. A.K. Bhardwaj, T.S. Kehwar, S.K. Chakarvarti, A.S. Oinam, S.C. Sharma '' 3-Dimensional conformal radiotherapy versus intensity
modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic''Iran.j.Radiat.Res. ;( 5(1)7-8, 2007.
[29]. Aoyama H, Westerly DC, Mackie TR, Olivera GH, Bentzen SM, Patel RR, Jaradat H, Tome WA, Ritter MA and Mehta MP.
Integral radiation dose to normal structures with conformal external beam radiation. Int J Radiat Oncol Biol Phys; 64(3): 962–967,
2006.
[30]. Yoo S, Wu QJ, Lee WR and Yin FF. Radiotherapy treatment plans with RapidArc® for prostate cancer involving seminal vesicles
and lymph nodes. Int J Radiat Oncol Biol Phys 2010; 76(3): 935–942.
[31]. EJ and Wuu CS. Radiation-induced second cancers: The impact of 3D-CRT and IMRT. Int J Radiat Oncol Biol Phys; 56(1): 83–88,
2003.
[32]. Kry SF, Salehpour M, Followill DS, Stovall M, Kuban DA, White RA, et al: Outof-field photon and neutron dose equivalents from
step-and-shoot intensity-modulated radiation therapy. Int J Radiat Oncol Biol Phys; 62(4):1204-1216, 2005.