B-scan ultrasonography produces real-time images of ocular structures using high frequency sound waves. It is useful for evaluating conditions like retinal detachment, tumors, and vitreous opacities. The technique involves placing a transducer probe on the eye to emit ultrasound and receive echoes. Different probe positions provide transverse, longitudinal, or axial scans of the eye. Normal tissues like vitreous and retina appear echolucent or reflective on scans depending on their structure and composition. Pathologies are identified based on their appearance, location, and movement patterns seen on the images. B-scan ultrasonography is a non-invasive imaging method useful when the ocular media is opaque.
Ultrasonography uses ultrasound to image tissues within the body. A-scan ultrasonography provides a one-dimensional view of the eye by measuring the echoes of ultrasound waves. It can be used to detect and measure tumors, assess eye structures for IOL calculation, and interpret pathology. The ultrasound is reflected at interfaces between tissues, appearing as spikes on the display. Immersion techniques provide more accurate measurements than contact techniques by avoiding compression artifacts. Limitations include artifacts, small lesions, missed foreign bodies, and misalignment issues.
This document provides an overview of B-scan ultrasonography. It begins with an introduction to B-scans and their use in providing qualitative and quantitative assessment of the eye and orbit. It then discusses the physics and principles behind ultrasound, including reflection, absorption, resolution and other key concepts. The document outlines the components and use of B-scan ultrasound machines, including different probe orientations and scanning techniques. It concludes with clinical applications and indications for B-scan ultrasonography in evaluating ocular pathology.
Ultrasonography, also known as B-scan, was first used in ophthalmology in the 1940s. It uses high frequency sound waves to generate images of the inside of the eye. B-scans can be used to evaluate conditions like tumors, retinal detachments, and vitreous opacities. The document discusses the history, physics, principles and various applications of B-scan ultrasonography for examining the eye. Key aspects covered include probe orientation, scan types, interpretation of echogenicity and advantages in providing a non-invasive evaluation of intraocular structures.
A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation Mero Eye
Hello Everyone, Namaste!
We would like to notify you all that Mero Eye Foundation is going to conduct an "EYE TALKS-Webinar", we will be having our session.
Speaker Name: MR AMIT KUMAR SINGH
Topic: "A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation"
DATE – MONDAY, 11th MAY 2020 @ 01.45PM IST, 02.00PM NPT (GTM +5.45)
B-scan ultrasonography uses high frequency sound waves to produce 2D images of ocular structures. It can be used to evaluate the anterior segment, posterior segment, tumors, vitreous pathology, retinal detachments, and more. The probe transmits sound waves which bounce off tissues and return echoes that are amplified and displayed. This allows visualization of the retina, choroid, lens, vitreous humor and other structures. B-scan is useful for diagnosing and monitoring many ocular conditions.
USG B scan is a noninvasive imaging technique used to assess ocular structures. It works by emitting high frequency sound waves into the eye, which are reflected back to a probe and converted into an image. Key principles include sound traveling faster in solids than liquids, stronger reflections occurring at interfaces of different densities, and perpendicular angle of incidence providing best images. Clinical applications include evaluating conditions that prevent normal examination like corneal scarring or dense cataracts. It can differentiate pathologies like vitreous hemorrhage from asteroid hyalosis.
Ultrasonography is a non-invasive imaging technique used to examine the eye. The B-scan was developed in the 1950s and 1960s and allows cross-sectional imaging of the eye. It works by emitting high frequency soundwaves into the eye and receiving echoes to create images. The B-scan examines features such as lesions through their shape, location, texture and mobility. Proper technique is required for high quality images, including centering lesions and using an appropriate probe frequency and gain. B-scans are useful for diagnosing various pathologies by comparing features to normal anatomical structures.
The document discusses B-scan ultrasound, providing a history of its development and describing the technical aspects and clinical applications. It notes that B-scan utilizes high frequency sound waves to produce two-dimensional images, and was first introduced in 1958. The document outlines the physics behind B-scan, describing how sound waves are reflected and the factors that determine resolution. Clinical uses mentioned include evaluating vitreous opacities, retinal detachments, and tumors.
Ultrasonography uses ultrasound to image tissues within the body. A-scan ultrasonography provides a one-dimensional view of the eye by measuring the echoes of ultrasound waves. It can be used to detect and measure tumors, assess eye structures for IOL calculation, and interpret pathology. The ultrasound is reflected at interfaces between tissues, appearing as spikes on the display. Immersion techniques provide more accurate measurements than contact techniques by avoiding compression artifacts. Limitations include artifacts, small lesions, missed foreign bodies, and misalignment issues.
This document provides an overview of B-scan ultrasonography. It begins with an introduction to B-scans and their use in providing qualitative and quantitative assessment of the eye and orbit. It then discusses the physics and principles behind ultrasound, including reflection, absorption, resolution and other key concepts. The document outlines the components and use of B-scan ultrasound machines, including different probe orientations and scanning techniques. It concludes with clinical applications and indications for B-scan ultrasonography in evaluating ocular pathology.
Ultrasonography, also known as B-scan, was first used in ophthalmology in the 1940s. It uses high frequency sound waves to generate images of the inside of the eye. B-scans can be used to evaluate conditions like tumors, retinal detachments, and vitreous opacities. The document discusses the history, physics, principles and various applications of B-scan ultrasonography for examining the eye. Key aspects covered include probe orientation, scan types, interpretation of echogenicity and advantages in providing a non-invasive evaluation of intraocular structures.
A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation Mero Eye
Hello Everyone, Namaste!
We would like to notify you all that Mero Eye Foundation is going to conduct an "EYE TALKS-Webinar", we will be having our session.
Speaker Name: MR AMIT KUMAR SINGH
Topic: "A quick guide to Ophthalmic Ultrasound/ B-Scan interpretation"
DATE – MONDAY, 11th MAY 2020 @ 01.45PM IST, 02.00PM NPT (GTM +5.45)
B-scan ultrasonography uses high frequency sound waves to produce 2D images of ocular structures. It can be used to evaluate the anterior segment, posterior segment, tumors, vitreous pathology, retinal detachments, and more. The probe transmits sound waves which bounce off tissues and return echoes that are amplified and displayed. This allows visualization of the retina, choroid, lens, vitreous humor and other structures. B-scan is useful for diagnosing and monitoring many ocular conditions.
USG B scan is a noninvasive imaging technique used to assess ocular structures. It works by emitting high frequency sound waves into the eye, which are reflected back to a probe and converted into an image. Key principles include sound traveling faster in solids than liquids, stronger reflections occurring at interfaces of different densities, and perpendicular angle of incidence providing best images. Clinical applications include evaluating conditions that prevent normal examination like corneal scarring or dense cataracts. It can differentiate pathologies like vitreous hemorrhage from asteroid hyalosis.
Ultrasonography is a non-invasive imaging technique used to examine the eye. The B-scan was developed in the 1950s and 1960s and allows cross-sectional imaging of the eye. It works by emitting high frequency soundwaves into the eye and receiving echoes to create images. The B-scan examines features such as lesions through their shape, location, texture and mobility. Proper technique is required for high quality images, including centering lesions and using an appropriate probe frequency and gain. B-scans are useful for diagnosing various pathologies by comparing features to normal anatomical structures.
The document discusses B-scan ultrasound, providing a history of its development and describing the technical aspects and clinical applications. It notes that B-scan utilizes high frequency sound waves to produce two-dimensional images, and was first introduced in 1958. The document outlines the physics behind B-scan, describing how sound waves are reflected and the factors that determine resolution. Clinical uses mentioned include evaluating vitreous opacities, retinal detachments, and tumors.
This document discusses visual field testing and perimetry. It defines the visual field and describes common visual field defects. It then covers the indications, methods, and terminology of visual field testing. Specific details are provided on threshold testing strategies, reliability indices, and how to interpret visual field printout maps and global indices. Criteria for diagnosing glaucomatous visual field loss and detecting progression over time are also outlined.
B-scan ultrasonography provides two-dimensional images of the eye that can reveal information about the shape, location, extension, mobility, and thickness of tissues. It uses high frequency sound waves reflected off structures in the eye. The transducer sends pulses and receives echoes to build an image. B-scan is useful when the ocular media is opaque and for evaluating conditions like tumors, detachments, inflammation and measuring the eye's dimensions. Pathological features seen on B-scan include vitreous hemorrhage, asteroid hyalosis, retinoschisis, choroidal detachment, retinal detachment in various configurations, cysticercosis, choroidal melanoma and more.
Optical coherence tomography (OCT) provides high-resolution, cross-sectional imaging of the retina and optic nerve head. It uses light waves instead of sound waves to capture micrometer-scale resolutions. OCT is a non-contact, non-invasive imaging technique that correlates well with retinal histology. The document discusses the principles, advantages, and procedures of OCT, summarizing key aspects of image acquisition and analysis for clinical and research applications.
Ultrasonography uses sound waves to image the eye and orbit. It was first developed in the 1950s and has since become an important tool for ocular imaging. Ultrasound uses high frequency sound pulses that reflect off structures in the eye to produce images. There are two main types: A-scan which produces a one-dimensional image, and B-scan which produces a two-dimensional cross-sectional image. Ultrasound is useful for evaluating the posterior segment in opaque media, measuring tissue thickness, and detecting intraocular and orbital lesions. It is a non-invasive tool commonly used to diagnose and monitor various ocular diseases.
Ophthalmic ultrasonography uses sound waves to evaluate the eye and orbit. It can assess tumors, retinal detachments, and foreign bodies when the eye is opaque. The A-scan provides one-dimensional measurements of internal structures. The B-scan gives a two-dimensional cross-section, displaying reflections as varying shades of gray. Together they characterize lesions by location, size, internal reflectivity, structure, and vascularity. Ultrasound is used preoperatively for cataract surgery planning and to evaluate intraocular tumors, accurately measuring their dimensions to guide treatment. Common indications also include opaque media evaluation and orbital disorders.
This document discusses using optical coherence tomography (OCT) to analyze the macula, retinal nerve fiber layer (RNFL), and optic nerve head in patients with glaucoma or suspected glaucoma. It describes how OCT can measure macular thickness, RNFL thickness, and optic disc parameters. Five case studies are presented showing how structural changes seen on OCT correlate with functional defects on visual field tests or clinical findings. The document concludes by mentioning Doppler OCT may help understand the role of blood flow in glaucoma and other optic neuropathies.
Corneal topography provides a graphic representation of the geometrical properties of the corneal surface. It uses techniques such as Placido disk, photokeratoscopy, videokeratoscopy, and slit imaging to map over 8000 points across the corneal surface. This provides detailed information about the shape and irregularities of the cornea which can then be used to diagnose conditions that degrade vision and guide treatment.
The document describes the use of various Pentacam maps and indices for screening patients for keratoconus, including:
1) The standard 4-map composite report, keratoconus map, Holladay report, and Belin/Ambrosio Enhanced Ectasia Display.
2) Key features to examine on each map include anterior and posterior elevation maps, pachymetry maps, curvature maps, and indices values.
3) The Belin/Ambrosio Enhanced Ectasia Display aims to improve sensitivity by calculating an "enhanced" best fit sphere reference surface that excludes the thinnest corneal region, highlighting differences between normal and ectatic corneas.
1) Biometry is the process of measuring the eye to determine the ideal intraocular lens power for cataract surgery. It involves measuring the corneal power and axial length of the eye.
2) Traditional A-scan ultrasound biometry measures axial length using sound waves, but has limitations like variable corneal compression. Newer devices like the IOL Master use optical interferometry and are non-contact.
3) Proper technique and accounting for factors like intraocular lens material are important for accurate biometry and intraocular lens power calculation. Inaccuracies can result in postoperative refractive surprises.
Ultrasound biomicroscopy (UBM) provides high-resolution imaging of ocular structures in the anterior segment of the eye using 50 MHz ultrasound. UBM allows visualization of tissues like the ciliary body and zonules that are not visible by slit lamp examination. UBM can be used to qualitatively and quantitatively evaluate the anterior segment structures and has applications in diagnosing and monitoring conditions like glaucoma, corneal diseases, tumors, and intraocular lenses. While UBM provides excellent detail, it has limitations including only being able to image about 5mm into the eye and requiring contact with the eye, unlike anterior segment OCT which is non-contact.
The document discusses principles of perimetry, which is the measurement of visual functions across the visual field. It describes the history of automated perimeters beginning in 1970. Static perimetry uses computerized testing to determine contrast sensitivity thresholds at preset locations, while kinetic perimetry manually maps sensitivity points along meridians. Both methods are used to identify decreases in retinal sensitivity indicative of conditions like glaucoma. Automated static perimetry provides quantifiable and reproducible data but is time-consuming, while kinetic perimetry rapidly defines field contours but requires more operator skill.
The document discusses how to interpret visual field tests, specifically the Humphrey Visual Field test. It provides details on:
- The anatomy and physiology of the visual field and hill of vision.
- Types of perimetry tests including static, kinetic, threshold, and supra-threshold tests.
- Components and procedures of Humphrey Visual Field testing including stimuli, test patterns like 24-2 and 10-2, and testing types.
- What the test printout shows including reliability indices, threshold values, deviation maps, and gaze tracking records.
- What abnormalities are looked for in glaucoma, neurological diseases, and retinal diseases and how the test helps in diagnosis and monitoring of these conditions.
This document discusses ocular biometry and ultrasound. It begins with definitions of biometrics and ultrasound terminology. It then describes the different modes of ultrasound - A-scan, B-scan and M-scan. Key components of ultrasound devices like transducers, amplifiers and velocities of sound through ocular tissues are explained. Factors affecting ultrasound reflection and penetration are outlined. The document concludes with an introduction to ocular biometry procedures and a brief history.
This document discusses corneal topography, which refers to studying the shape of the corneal surface. Various techniques for corneal topography are described, including keratometry, keratoscopy, rasterstereography, and interferometry. Key corneal topography systems such as Placido disc topographers, slit imaging topographers, and laser holographic interferometry systems are summarized. The document also reviews display formats for topography data and clinical applications of corneal topography analysis.
Gonioscopy allows visualization of the anterior chamber angle to evaluate for angle closure and diagnose glaucoma. It was pioneered in the early 20th century with the introduction of contact lenses to eliminate total internal reflection at the cornea. Direct gonioscopy uses contact lenses for a straight view, while indirect gonioscopy uses prisms for an inverted image at the slit lamp. Examination of angle structures like the trabecular meshwork and classification systems help diagnose angle closure and glaucoma. Gonioscopy is used for diagnostic and therapeutic purposes like laser and surgery.
This document summarizes key aspects of perimetry testing. It defines the normal visual field and describes how perimetry can be used to detect functional vision loss and monitor disease progression. Two main types of perimetry are discussed: kinetic and static. Details are provided on testing strategies, stimuli brightness, interpreting results like total deviation and reliability indices. The document emphasizes the importance of perimetry in glaucoma and neurological diagnosis and management.
Investigations in a case of corneal ulcer in a clinic setting what to dodrvishuankad
This document discusses investigations for corneal ulcers in a clinic setting. It begins by outlining the importance of microbiological workup for corneal ulcers. The key investigations discussed are smear examination using stains like Gram and KOH to visualize organisms, and culture using media suited to common bacteria, fungi, Acanthamoeba, and other pathogens. Details are provided on collection and handling of corneal scrapings, as well as characteristics and diagnostic features of common organisms causing bacterial, fungal and Acanthamoeba keratitis. The importance of proper sample collection and meticulous laboratory workup is emphasized to arrive at an accurate diagnosis and guide treatment.
This document discusses different types of drusen and their characterization using multimodal imaging. It introduces soft drusen, cuticular drusen, and subretinal drusenoid deposits. Each type has distinct features on color photography, infrared imaging, autofluorescence, and optical coherence tomography. The document also discusses how drusen type relates to choroidal thickness and risk of developing different forms of neovascularization or geographic atrophy. A new classification system is proposed that better accounts for drusen subtypes and progression pathways in age-related macular degeneration.
The document discusses visual field testing in glaucoma. It defines the visual field and perimetry, and describes the major types of clinical perimetry tests including full threshold, SITA standard, and SITA fast on Humphrey and normal, dynamic, and TOP strategies on Octopus. It explains parameters such as test patterns, reliability, age-corrected plots, tests like GHT and Bebie curve, and global indices including MD, PSD, SF, and CPSD. The purpose of visual field testing in glaucoma is to detect and monitor disease by measuring light sensitivity across the retinal field.
Ultrasound is useful for evaluating ocular structures. It uses sound waves above the audible range that are reflected by tissues. The transducer produces pulses and detects echoes to create images. Different scan types provide 1D (A-scan) or 2D (B-scan) views. A-scans show acoustic density as spikes while B-scans show shape, location and extent of structures. Proper technique and interpretation of features like reflectivity, mobility and shadowing allow ultrasound to diagnose conditions like retinal detachment, tumors, inflammation and optic nerve diseases.
UBM provides high resolution imaging of the anterior segment structures in a non-invasive manner. It can image structures like the ciliary body and zonules that
This document discusses visual field testing and perimetry. It defines the visual field and describes common visual field defects. It then covers the indications, methods, and terminology of visual field testing. Specific details are provided on threshold testing strategies, reliability indices, and how to interpret visual field printout maps and global indices. Criteria for diagnosing glaucomatous visual field loss and detecting progression over time are also outlined.
B-scan ultrasonography provides two-dimensional images of the eye that can reveal information about the shape, location, extension, mobility, and thickness of tissues. It uses high frequency sound waves reflected off structures in the eye. The transducer sends pulses and receives echoes to build an image. B-scan is useful when the ocular media is opaque and for evaluating conditions like tumors, detachments, inflammation and measuring the eye's dimensions. Pathological features seen on B-scan include vitreous hemorrhage, asteroid hyalosis, retinoschisis, choroidal detachment, retinal detachment in various configurations, cysticercosis, choroidal melanoma and more.
Optical coherence tomography (OCT) provides high-resolution, cross-sectional imaging of the retina and optic nerve head. It uses light waves instead of sound waves to capture micrometer-scale resolutions. OCT is a non-contact, non-invasive imaging technique that correlates well with retinal histology. The document discusses the principles, advantages, and procedures of OCT, summarizing key aspects of image acquisition and analysis for clinical and research applications.
Ultrasonography uses sound waves to image the eye and orbit. It was first developed in the 1950s and has since become an important tool for ocular imaging. Ultrasound uses high frequency sound pulses that reflect off structures in the eye to produce images. There are two main types: A-scan which produces a one-dimensional image, and B-scan which produces a two-dimensional cross-sectional image. Ultrasound is useful for evaluating the posterior segment in opaque media, measuring tissue thickness, and detecting intraocular and orbital lesions. It is a non-invasive tool commonly used to diagnose and monitor various ocular diseases.
Ophthalmic ultrasonography uses sound waves to evaluate the eye and orbit. It can assess tumors, retinal detachments, and foreign bodies when the eye is opaque. The A-scan provides one-dimensional measurements of internal structures. The B-scan gives a two-dimensional cross-section, displaying reflections as varying shades of gray. Together they characterize lesions by location, size, internal reflectivity, structure, and vascularity. Ultrasound is used preoperatively for cataract surgery planning and to evaluate intraocular tumors, accurately measuring their dimensions to guide treatment. Common indications also include opaque media evaluation and orbital disorders.
This document discusses using optical coherence tomography (OCT) to analyze the macula, retinal nerve fiber layer (RNFL), and optic nerve head in patients with glaucoma or suspected glaucoma. It describes how OCT can measure macular thickness, RNFL thickness, and optic disc parameters. Five case studies are presented showing how structural changes seen on OCT correlate with functional defects on visual field tests or clinical findings. The document concludes by mentioning Doppler OCT may help understand the role of blood flow in glaucoma and other optic neuropathies.
Corneal topography provides a graphic representation of the geometrical properties of the corneal surface. It uses techniques such as Placido disk, photokeratoscopy, videokeratoscopy, and slit imaging to map over 8000 points across the corneal surface. This provides detailed information about the shape and irregularities of the cornea which can then be used to diagnose conditions that degrade vision and guide treatment.
The document describes the use of various Pentacam maps and indices for screening patients for keratoconus, including:
1) The standard 4-map composite report, keratoconus map, Holladay report, and Belin/Ambrosio Enhanced Ectasia Display.
2) Key features to examine on each map include anterior and posterior elevation maps, pachymetry maps, curvature maps, and indices values.
3) The Belin/Ambrosio Enhanced Ectasia Display aims to improve sensitivity by calculating an "enhanced" best fit sphere reference surface that excludes the thinnest corneal region, highlighting differences between normal and ectatic corneas.
1) Biometry is the process of measuring the eye to determine the ideal intraocular lens power for cataract surgery. It involves measuring the corneal power and axial length of the eye.
2) Traditional A-scan ultrasound biometry measures axial length using sound waves, but has limitations like variable corneal compression. Newer devices like the IOL Master use optical interferometry and are non-contact.
3) Proper technique and accounting for factors like intraocular lens material are important for accurate biometry and intraocular lens power calculation. Inaccuracies can result in postoperative refractive surprises.
Ultrasound biomicroscopy (UBM) provides high-resolution imaging of ocular structures in the anterior segment of the eye using 50 MHz ultrasound. UBM allows visualization of tissues like the ciliary body and zonules that are not visible by slit lamp examination. UBM can be used to qualitatively and quantitatively evaluate the anterior segment structures and has applications in diagnosing and monitoring conditions like glaucoma, corneal diseases, tumors, and intraocular lenses. While UBM provides excellent detail, it has limitations including only being able to image about 5mm into the eye and requiring contact with the eye, unlike anterior segment OCT which is non-contact.
The document discusses principles of perimetry, which is the measurement of visual functions across the visual field. It describes the history of automated perimeters beginning in 1970. Static perimetry uses computerized testing to determine contrast sensitivity thresholds at preset locations, while kinetic perimetry manually maps sensitivity points along meridians. Both methods are used to identify decreases in retinal sensitivity indicative of conditions like glaucoma. Automated static perimetry provides quantifiable and reproducible data but is time-consuming, while kinetic perimetry rapidly defines field contours but requires more operator skill.
The document discusses how to interpret visual field tests, specifically the Humphrey Visual Field test. It provides details on:
- The anatomy and physiology of the visual field and hill of vision.
- Types of perimetry tests including static, kinetic, threshold, and supra-threshold tests.
- Components and procedures of Humphrey Visual Field testing including stimuli, test patterns like 24-2 and 10-2, and testing types.
- What the test printout shows including reliability indices, threshold values, deviation maps, and gaze tracking records.
- What abnormalities are looked for in glaucoma, neurological diseases, and retinal diseases and how the test helps in diagnosis and monitoring of these conditions.
This document discusses ocular biometry and ultrasound. It begins with definitions of biometrics and ultrasound terminology. It then describes the different modes of ultrasound - A-scan, B-scan and M-scan. Key components of ultrasound devices like transducers, amplifiers and velocities of sound through ocular tissues are explained. Factors affecting ultrasound reflection and penetration are outlined. The document concludes with an introduction to ocular biometry procedures and a brief history.
This document discusses corneal topography, which refers to studying the shape of the corneal surface. Various techniques for corneal topography are described, including keratometry, keratoscopy, rasterstereography, and interferometry. Key corneal topography systems such as Placido disc topographers, slit imaging topographers, and laser holographic interferometry systems are summarized. The document also reviews display formats for topography data and clinical applications of corneal topography analysis.
Gonioscopy allows visualization of the anterior chamber angle to evaluate for angle closure and diagnose glaucoma. It was pioneered in the early 20th century with the introduction of contact lenses to eliminate total internal reflection at the cornea. Direct gonioscopy uses contact lenses for a straight view, while indirect gonioscopy uses prisms for an inverted image at the slit lamp. Examination of angle structures like the trabecular meshwork and classification systems help diagnose angle closure and glaucoma. Gonioscopy is used for diagnostic and therapeutic purposes like laser and surgery.
This document summarizes key aspects of perimetry testing. It defines the normal visual field and describes how perimetry can be used to detect functional vision loss and monitor disease progression. Two main types of perimetry are discussed: kinetic and static. Details are provided on testing strategies, stimuli brightness, interpreting results like total deviation and reliability indices. The document emphasizes the importance of perimetry in glaucoma and neurological diagnosis and management.
Investigations in a case of corneal ulcer in a clinic setting what to dodrvishuankad
This document discusses investigations for corneal ulcers in a clinic setting. It begins by outlining the importance of microbiological workup for corneal ulcers. The key investigations discussed are smear examination using stains like Gram and KOH to visualize organisms, and culture using media suited to common bacteria, fungi, Acanthamoeba, and other pathogens. Details are provided on collection and handling of corneal scrapings, as well as characteristics and diagnostic features of common organisms causing bacterial, fungal and Acanthamoeba keratitis. The importance of proper sample collection and meticulous laboratory workup is emphasized to arrive at an accurate diagnosis and guide treatment.
This document discusses different types of drusen and their characterization using multimodal imaging. It introduces soft drusen, cuticular drusen, and subretinal drusenoid deposits. Each type has distinct features on color photography, infrared imaging, autofluorescence, and optical coherence tomography. The document also discusses how drusen type relates to choroidal thickness and risk of developing different forms of neovascularization or geographic atrophy. A new classification system is proposed that better accounts for drusen subtypes and progression pathways in age-related macular degeneration.
The document discusses visual field testing in glaucoma. It defines the visual field and perimetry, and describes the major types of clinical perimetry tests including full threshold, SITA standard, and SITA fast on Humphrey and normal, dynamic, and TOP strategies on Octopus. It explains parameters such as test patterns, reliability, age-corrected plots, tests like GHT and Bebie curve, and global indices including MD, PSD, SF, and CPSD. The purpose of visual field testing in glaucoma is to detect and monitor disease by measuring light sensitivity across the retinal field.
Ultrasound is useful for evaluating ocular structures. It uses sound waves above the audible range that are reflected by tissues. The transducer produces pulses and detects echoes to create images. Different scan types provide 1D (A-scan) or 2D (B-scan) views. A-scans show acoustic density as spikes while B-scans show shape, location and extent of structures. Proper technique and interpretation of features like reflectivity, mobility and shadowing allow ultrasound to diagnose conditions like retinal detachment, tumors, inflammation and optic nerve diseases.
UBM provides high resolution imaging of the anterior segment structures in a non-invasive manner. It can image structures like the ciliary body and zonules that
This document discusses ultrasonography in ophthalmology. It defines ultrasonography as using acoustic waves with frequencies greater than 20kHz. The main advantages of ultrasonography in ophthalmology are that it is easy to use, involves no ionizing radiation, allows for excellent tissue differentiation, and is cost-effective. The primary uses in ophthalmology are for evaluating the posterior segment in hazy media, detecting intraocular and orbital lesions or foreign bodies, and performing ocular biometry for intraocular lens power calculation. The document describes the basic components of ultrasonography machines and provides examples of A-scan and B-scan display modes and various scanning positions and techniques.
Optical coherence tomography (OCT) provides high resolution, cross-sectional images of the retina. OCT uses light waves to generate tomographic scans. Early OCT systems had axial resolutions of 10 μm and scan speeds of 400 scans/second. Newer spectral domain OCT systems have higher resolutions of 1-15 μm and faster scan speeds of up to 52,000 scans/second, allowing better visualization of retinal layers and pathology. OCT is used to qualitatively and quantitatively evaluate retinal morphology and thickness.
This document provides tips for performing and interpreting ultrasound B-scans of the posterior eye segment. It discusses the basic physics and principles of ultrasound imaging. It outlines indications for B-scans such as opaque ocular media, retinal conditions, and ocular trauma. Tips are provided on probe parts and orientation, different examination techniques, normal anatomical structures, and common pathologies. Examples of B-scan images from case studies are also presented to demonstrate clinical applications and findings. The goal is to make ultrasound imaging of the eye easier to understand.
OCT allows for high-resolution cross-sectional imaging of the retina. It provides micron-level resolution, enabling visualization of the retinal layers. OCT is a non-contact, non-invasive technique useful for qualitative and quantitative analysis of the retina and monitoring of morphological changes. It can detect and measure retinal thickness, volume, and parameters like RNFL thickness. While it provides advantages over other modalities, OCT also has limitations like difficulty imaging through opaque media. It operates using low-coherence interferometry and is useful for evaluating a variety of posterior segment diseases.
This document discusses various biometry instruments and equipment used to calculate intraocular lens (IOL) power for cataract surgery. It describes how keratometry, A-scan ultrasound biometry, and non-contact devices like the IOLMaster measure important ocular dimensions needed for IOL power calculations, including corneal power, axial length, and anterior chamber depth. It also discusses IOL power calculation formulas from first to fourth generation and factors that influence formula choice, such as eye length, anterior chamber depth, and IOL placement in the eye. Accurate biometry is emphasized as key to achieving the desired postoperative refractive outcome.
This document provides an overview of ultrasonography principles, methods, and interpretation for ophthalmic use. It discusses the history of ultrasonography, describes A-scan and B-scan display methods, and outlines the examination procedure and interpretation of scans. Key points covered include how ultrasound waves are generated and propagated through ocular tissues, factors that affect resolution, and how scans are oriented and labeled to identify anatomical structures.
Optical coherence tomography (OCT) is a non-invasive imaging technique that uses light to generate high-resolution cross-sectional images of the retina and anterior segment of the eye. OCT was invented in 1991 and has since advanced from time domain to spectral domain and swept-source versions with higher speed and resolution. It works by measuring the echo time delay and intensity of backscattered light to create images, and can qualitatively and quantitatively assess retinal thickness, layers, and pathology. OCT has numerous clinical applications in ophthalmology for diseases of the retina, choroid and anterior segment.
Three key points about imaging the orbit:
1. CT scans provide the best view of bony details and calcifications in the orbit, and can detect small fractures and foreign bodies. Slice thickness and tissue windows must be optimized for diagnostic quality.
2. Different x-ray views (like Waters, Caldwell's, and lateral) allow visualization of specific orbital structures and are useful for identifying pathology in different areas.
3. Features seen on imaging like changes in bone density, orbital size and shape, and structures like the optic canal can indicate conditions like tumors, infections, fractures, and vascular abnormalities affecting the orbit. Precise imaging analysis is important for diagnosis.
OCT is used to non-invasively image the retina in cross-section with micrometer-level resolution. It works by measuring the interference of light reflected from retinal structures. OCT was developed in 1991 and uses near-infrared light wavelengths of 840nm and 1310nm. OCT provides high-resolution 2D images of the retina and can integrate data points over depth to form 3D representations. It is useful for diagnosing and monitoring many retinal diseases.
This document discusses the corneal endothelium and techniques for assessing its health and function. The corneal endothelium is a single layer of hexagonal cells that maintains corneal clarity by pumping fluid out of the stroma. Assessment techniques described include specular microscopy, which examines cell density, morphology, and patterns at high magnification; confocal microscopy and anterior segment OCT for in vivo imaging; and ultrasound pachymetry to measure corneal thickness as an indicator of endothelial function. Common endothelial diseases like Fuchs' dystrophy and conditions affecting assessment are also reviewed.
This document discusses the corneal endothelium and techniques for assessing its health and function. The corneal endothelium is a single layer of hexagonal cells that maintains corneal clarity by pumping fluid out of the stroma. Assessment techniques described include specular microscopy, which allows analysis of endothelial cell density, morphology, and patterns under high magnification; confocal microscopy; anterior segment OCT; and ultrasound pachymetry to measure corneal thickness as an indicator of endothelial function. Common indications for assessment include pre- and post-operative evaluation, and evaluation of donor corneas for transplantation.
This document discusses the corneal endothelium and techniques for assessing its health and function. The corneal endothelium is a single layer of hexagonal cells that maintains corneal clarity by pumping fluid out of the stroma. Assessment techniques described include specular microscopy, which examines cell density, morphology, and patterns at high magnification; confocal microscopy and anterior segment OCT for in vivo imaging; and ultrasound pachymetry to measure corneal thickness as an indicator of endothelial function. Common endothelial diseases like Fuchs' dystrophy and conditions affecting assessment are also reviewed.
- Proptosis refers to forward displacement of the eyeball beyond the orbital rim, usually due to an increased mass or volume within the orbit. Common causes include tumors, inflammation, trauma, thyroid eye disease, and vascular lesions.
- Examination involves measuring exophthalmometry, inspecting for signs of mass effect, pulsation or displacement of the globe, and evaluating for neurologic deficits. Imaging with CT or MRI is important to characterize the lesion causing proptosis.
Optical coherence tomography (OCT) provides high-resolution cross-sectional imaging of coronary arteries and atherosclerotic plaque. It uses near-infrared light instead of sound and has 10 times the resolution of intravascular ultrasound. OCT can image the vessel wall and identify clinically relevant features like fibrous caps, calcium, necrotic cores, inflammation, and thrombus. It allows assessment of atherosclerosis and plaque vulnerability in vivo with measurements like fibrous cap thickness. Clinical applications include evaluation of atherosclerosis, characterization of thin-capped fibroatheromas, and assessment of inflammation.
This document describes various methods of illumination used with a slit lamp to examine different parts of the eye. Diffuse illumination allows for a general survey of the eye while optic section, parallelepiped, and retroillumination techniques are used to view specific structures like the cornea, lens, and vitreous in more detail. Different angles of illumination like tangential, conical beam, and oscillatory help observe surface textures, cells in the aqueous humor, and lens opacities. Precise illumination techniques are crucial for comprehensive eye exams.
Indirect ophthalmoscopy and fundus drawingSonali Singh
This document provides an overview of binocular indirect ophthalmoscopy and fundus drawing techniques. It discusses the history and working principles of the indirect ophthalmoscope. Examination techniques are described, including positioning, use of the condensing lens, and scleral indentation. Fundus drawing methods such as color coding retinal findings are also outlined. Proper cleaning and sterilization of equipment is emphasized.
it describes the bony anatomy including the femoral head , acetabulum, labrum . also discusses the capsule , ligaments . muscle that act on the hip joint and the range of motion are outlined. factors affecting hip joint stability and weight transmission through the joint are summarized.
This presentation was provided by Steph Pollock of The American Psychological Association’s Journals Program, and Damita Snow, of The American Society of Civil Engineers (ASCE), for the initial session of NISO's 2024 Training Series "DEIA in the Scholarly Landscape." Session One: 'Setting Expectations: a DEIA Primer,' was held June 6, 2024.
How to Setup Warehouse & Location in Odoo 17 InventoryCeline George
In this slide, we'll explore how to set up warehouses and locations in Odoo 17 Inventory. This will help us manage our stock effectively, track inventory levels, and streamline warehouse operations.
Executive Directors Chat Leveraging AI for Diversity, Equity, and InclusionTechSoup
Let’s explore the intersection of technology and equity in the final session of our DEI series. Discover how AI tools, like ChatGPT, can be used to support and enhance your nonprofit's DEI initiatives. Participants will gain insights into practical AI applications and get tips for leveraging technology to advance their DEI goals.
This presentation includes basic of PCOS their pathology and treatment and also Ayurveda correlation of PCOS and Ayurvedic line of treatment mentioned in classics.
The simplified electron and muon model, Oscillating Spacetime: The Foundation...RitikBhardwaj56
Discover the Simplified Electron and Muon Model: A New Wave-Based Approach to Understanding Particles delves into a groundbreaking theory that presents electrons and muons as rotating soliton waves within oscillating spacetime. Geared towards students, researchers, and science buffs, this book breaks down complex ideas into simple explanations. It covers topics such as electron waves, temporal dynamics, and the implications of this model on particle physics. With clear illustrations and easy-to-follow explanations, readers will gain a new outlook on the universe's fundamental nature.
2. CONTENTS
• History
• Introduction
• Instrumentation
• Technique of Examination
• Indications
• Evaluation of Ocular Structures
• Ultrasound in Intraocular Pathology
3. HISTORY
• In 1880, Curie Brothers first demonstrated Piezoelectric Effect
• 1n 1949, Ludwig used ultrasound to detect gallstones.
• Ophthalmologic B Scan was first introduced by Baum and Greenwood
in 1958.
• Commercially developed by Coleman et al in 1970’s.
• Technique was emphasized by Karl Ossoinig.
4. INTRODUCTION
• B Scan Ultrasonography is an important adjuvant for the clinical
assessment of various ocular and orbital diseases.
• It produces grey scale, real time, two dimensional images of ocular
tissues
• Ultrasound is a:
Longitudinal Wave
Alternating Compressions and Rarefactions
Frequency: Above 20,000 Hz
Similar to Light Waves : Reflected, Refracted and Absorbed
6. Based on principles of Pulse Echo
Technology.
Echoes are generated at adjoining
tissue interfaces. Greater the difference,
stronger the echo.
Greater the Frequency – Greater the
Resolution – Lower the Penetration.
Lower the Frequency – Lower the Resolution
– Greater the Penetration
7. Sound waves from
Transducer
Hits the target
tissue
Echoes are
received by the
Receiver
Amplification of
signals
Display of image
on screen
9. TRANSDUCER
• Device which converts Electrical
Energy into Sound Energy and vice
versa.
• Parts:
• Piezoelectric Plate
• Backing Layer
• Acoustic Matching Layer
• Acoustic Lens
10. • PIEZOELECTRIC ELEMENT:
• Generates Ultrasonic Waves.
• Coated on both sides with electrodes to which
voltage is applied.
• Oscillation of element generates the sound
waves
• Most common: Lead Zirconate Titnate
• BACKING LAYER:
• Located behind the Piezoelectric element
• Dampens excessive vibrations from probe
• Improves image resolution
• ACOUSTIC MATCHING LAYER:
• Located in front of the Piezoelectric element
• Reduces reflections from acoustic impedance
between probe and object.
• Improves transmission.
11. • ACOUSTIC LENS:
• Grey coloured rubber on tip.
• Helps in focusing Ultrasonic Waves as a beam.
• AMPLIFIER
• DISPLAY MONITOR
13. AA-MODE
• Time- Amplitude Mode
• Seen as vertical deflections from a
baseline
• For interpretation of tissue reflectivity
• Uses one beam of ultrasound
• Brightness Mode
• Image recorded as bright and dim
dots
• For anatomical information: provides
cross-sectional images of globe and
orbit
• Uses a parallel beam of ultrasonic
waves
B-MODE
MODES OF ULTRASOUND
15. ULTRASOUND PROBE
• Emits focused sound beam at
frequency of 10MHz
• Mark on probe indicates beam
orientation
• Area towards which mark is directed
appears at the top of the echogram
on display screen
16. PROBE POSITIONING
TRANSVERSE
• Most Common
• Lateral extend
• 6 clock hours are
examined at a time.
• Probe is parallel to limbus
LONGITUDINAL
• Radial
• Anteroposterior extend of
lesion
• One clock hour examined
at a time
• Probe is perpendicular to
limbus
AXIAL
• Eye is held in primary
gaze
• Probe centered on the
cornea incorporating the
Lens
17. TRANSVERSE SCAN
• Produces circumferential slice through
several meridians
• If examining:
Nasal Area: 12-6 clock hours
Temporal Area: 6-12 clock hours
Superior Area: 9-3 clock hours
Inferior Area: 3-9 clock hours
18. LONGITUDINAL SCAN
• Probe marker towards centre of the
Cornea.
• Optic disc and posterior aspect of
globe.
19. AXIAL SCAN
• Probe centered on cornea
• Evaluates macular region
• Documents lesions and membranes in
relation to optic disc
• Decreased resolution of posterior
portion of globe.
20. PROCEDURE
• The patient is either reclining on a
chair or lying on a couch.
• Probe can be placed directly on
conjunctiva, cornea or on the lids
• Lowest possible decibel gain
consistent with the maintenance of
adequate intensity should be used.
22. EVALUATION OF OCULAR STRUCTURES
Amount of reflection of
ultrasound energy
Absorption of ultrasound energy
Angle of incidence of sound
Shape/ Size/ Smoothness of
interface
23. AMOUNT OF REFLECTED ENERGY
GAIN (decibels)
Higher Gain: displays weaker
echoes like Vitreous
Opacities
Better Penetration
Lower Gain: stronger echoes
like Retina and Sclera
Better resolution
NATURE OF
SURFACE
Radiopaqu
e
Radiotranslucen
t
24. ABSORPTION OF SOUND ENERGY
1. Absorption/ Attenuation: Gradually
all sound energy is absorbed as heat
eg Tumours
2. Shadowing: Sound is strongly
reflected, nothing passes through it.
Leaves dark shadows behind eg
Optic Nerve Head Drusen, Air Bubble
3. Reverberation: Collection of
reflected sounds bouncing back and
forth between tissue boundaries eg
Foreign Body
25. ANGLE OF INCIDENCE
• Probe should be held perpendicular
to the area of interest to achieve a
strong echo – Bright Image
• If held at an angle, some amount of
sound is reflected away – Dim Image
26. SHAPE, SIZE AND SMOOTHNESS OF
SURFACE
DOT LIKE LESIONS:
Vitreous Floaters, VH, Vitreous Exudates
MEMBRANOUS LESIONS:
PVD, RD, Vitreous membranes
MASS LESIONS:
Choroidal or Retinal Tumours
32. ASTEROID HYALOSIS
• Formation of Calcium Soaps within
vitreous gel
• Bright , round signal on B Scan
• Each opacity has its own spike on A
Scan
• Crystals are suspended, exhibit
dynamics of vitreous movement
38. ACUTE RETINAL DETACHMENT
• Detached neurosensory retina
appears as a membrane in vitreous
space.
• Highly reflective sheet like tissue
• Mobile , slightly folded
39. CHRONIC RETINAL DETACHMENT
• Detached retina appears thickened
• Decrease in the aftermovement
amplitude of the retina due to
massive periretinal proliferation by
Muller cells and astrocytes.
• Vitreous contracts leading to funnel
shaped detachment. As it further
contracts, it leads to formation of
cyclitic membranes extending from
vitreous base
• Cysts , subretinal opacities
40. RETINAL
DETACHMENT
• Always attached to the optic disc
• 100% spike on A-Scan
• Moderate aftermovements (recent
RD)
• High echogenicity
• Visible on Low Gain
• With or without disc insertion
• <100% spike on A-Scan
• Marked aftermovements
• Low – Medium echogenicity
• Disappears on Low Gain
POSTERIOR VITREOUS
DETACHMENT
REFLECTIVITY OF THE PERIPHERY CAN DIFFERENTIATE BETWEEN THE
TWO IN DIFFICULT SITUATIONS LIKE TRAUMA AND INFLAMMATION
41. RETINOSCHISIS
• Splitting within the neurosensory
retina
• Inferotemporal quadrant
• Moderately elevated, thin smooth
dome shaped structure in the
periphery
45. CHOROIDAL MELANOMA
• Biconvex , Homogeneous lesion
• If tumour breaks through the Bruchs
Membrane: mushroom shaped lesion,
collar button lesion
• Solid tissue, therefore no
aftermovements.
46. CHOROIDAL DETACHMENT
• Smooth
• Dome shaped, thick membrane
• Does not insert into the optic disc
• When severe, detached choroid can
meet at the center of the globe –
retina to retina touch – Kissing
Choroidal
• Serous Choroidal Detachments :
Echolucent
47. • Hemorhaggic Choroidal Detachment:
Reflective
• A Scan:
o Maximally high (100%) spike
o Thick
o Double peaked ( retina & choroid)
52. DISLOCATED LENS
• Lens seen floating in the vitreous or
lying against the retina
• Strands of vitreous may be attached
to it
53. PENETRATING AND PERFORATING
INJURY
• Penetrating injury: hemorrhage lines
up along the traumatic tract
• Perforating injury: posterior exit site is
present.
• Vitreous may be incarcerate both
anteriorly and posteriorly
57. ADVANTAGES
• Non invasive
• Performed in office setting
• No exposure to radiation
• Accurate assessment
• Easy follow up
• Limited penetration
• May require patient cooperation
DISADVANTAGES
PROS V/S CONS