UV-visible spectroscopy involves using light in the UV-visible spectral region to analyze chemical substances. It works on the principle of Beer-Lambert's law, where absorbance is directly proportional to concentration and path length. Different functional groups and conjugated systems can absorb light at characteristic wavelengths. The technique is used for quantitative and qualitative analysis of samples through measurement of absorption spectra. It provides information about electronic transitions and molecular structure of compounds.
3. Quantitative analysis of sample by using UV and Visible light is called UV Spectroscopy.
It is also known as electronic spectroscopy.
UV-Visible Spectroscopy (or Spectrophotometry) is a quantitative technique used to measure
how much a chemical substance absorbs light.
Different types of EMR & their Wavelength range:
Type Wavelength Range
Gamma rays <0.001nm
X-rays 10-2 to 102 Å
Ultra Violet (UV)
1. Far
2. Near
3. Visible
10 – 200 nm
200 – 400 nm
400 – 750 nm
Infra Red (IR)
1. Near
2. Mid
3. Far
0.75 – 2.5 mµ
2.5 – 50 mµ
50 – 1000 mµ
Microwave 0.1 – 100 cm
Radiowave 1 – 1000 m
Energy Wavelength
INTRODUCTION
3
4. Color Produced at Wavelength
VIBGYOR
EMR Spectrum
4
Colour Wavelength
Violet 400-420 nm
Indigo 420-440 nm
Blue 440-490 nm
Green 490-570 nm
Yellow 570-585 nm
Orange 585-620 nm
Red 620-780 nm
5. UV Visible spectroscopy based on the principle of Beers and Lamberts law.
Beers law: Absorption is directly proportional to concentration.
Lamberts law: Absorption is directly proportional to pathlength.
Principle: Any molecule has either n, π, σ or a combination of these electrons. Molecules
containing these bonding and bonding electrons absorb the energy in the form of UV light
and undergo a transition from the ground state to an excited state. By the characteristics
absorption peak, the nature of electron present and hence the molecular structure can be
determined.
I0 It
Incident light Transmitted light
PRINCIPLE
5
6. Molecules contains different types of electrons.
Every molecule absorb energy in different amount.
Absorption of EMR causes electrons to be excited which results in promotion of bonding
or non-bonding to anti-bonding orbitals.
Molecules contains 3 types of electrons:
Transition that takes place when electrons in a molecule are excited from one energy
level to higher energy level.
ELECTRONIC TRANSITIONS
6
7. Types of electronic transition
1. σ – σ*
2. σ – π*
3. n - σ*
4. n - π*
5. π - σ*
6. π - π* σ – π* and π - σ* are only theoretically possible. 7
8. 1. σ – σ*
This type of transition takes place between lower occupied molecular orbitals to
higher occupied molecular orbitals.
Such transition requires more energy.
Shows absorbance at 125 nm.
e.g. Alkane – Methane
2. n - σ*
Saturated compound containing 1 hetero atom with unshared pair of electrons.
Requires less energy.
Show absorbance at 150-250 nm.
Sensitive to H-bonding.
e.g. Aldehyde, ketones, ether, alcohol, etc.
3. π - π*
Takes place in aromatic compound or unsaturated compound.
Transition occurs at longer wavelength.
e.g. alkenes, alkynes, cyanides, azo compounds
4. n - π*
In this, an electrons of unshared pair of electrons on hetero atom get excited to π*anti-
bonding orbitals.
It requires less energy.
e.g. Saturated aldehyde
8
9. Molecules or parts of molecules that absorb light in UV visible region and
responsible for imparting color to the compounds.
Covalently unsaturated group responsible for electronic transition.
e.g. C=C, C=O
Chromogen: Comp. containing chromophores
Types of Chromophores:
1. Independent chromophore:
Single chromophore impart color to compounds.
e.g. Azo group, Nitro group (-NO)
2. Dependent chromophore
e.g. Acetone: 1 Ketonic group – Colorless
2 Ketonic group – Yellow
3 Ketonic group - Orange
CHROMOPHORE
9
10. AUXOCHROMES
Saturated group with non-bonding electrons when attached to chromophore, it will alter
wavelength as well as absorption of substance.
Group of atoms that get attached to chromophore and increases their color intensity.
Benzene Aniline
255 nm 280 nm
10
11. SPECTRAL SHIFTS
The position of λmax and absorption intensity can be modified in different ways by some
structural or solvent changes.
11
12. Bathochromic shift
It is an effect in which absorption maximum is shifted towards longer wavelength due to
presence of auxochromes or by change of solvent.
Also called as Red shift.
Hypsochromic Shift
Absorption maximum is shifted toward shorter wavelength.
Also called as Blue shift.
Hyperchromic Shift
It is an effect due to which intensity of absorption maximum increases.
Hypochromic shift
Intensity of absorption maximum decreases.
12
13. SOLVENT EFFECT ON ABSORPTION MAXIMA
Choice of solvent used in UV is important because of solvent can change λmax .
Most suitable solvent is that it does not absorb radiation in the region under
investigation.
The position as well as intensity of absorption maximum get shifted for a particular
chromophore by changing the polarity of solvent.
By increasing the polarity of solvent compound such as dienes and conjugated
hydrocarbons do not experience any appreciable shift.
In general absorption maximum for non-polar compound is usually shifted with the
change in polarity of solvent.
Polarity Lambda max
e.g. Lambda max of Acetone in water and hexane:
Water: 264 nm
Hexane: 279 nm
Water > Methanol > Ethanol > Benzene > Hexane
Polarity
13
14. Solvent Wavelength of absorption
Water 205
Cyclohexane 210
Ethanol 210
Methanol 210
Chloroform 245
Benzene 280
Alpha, beta unsaturated compounds show different shift:
1. n - π*
Absorption band moves to shorter wavelength by increasing the polarity of solvent.
Less intense
2. π - π*
Absorption band moves to longer wavelength by increasing the polarity of solvent.
More intense
Group is more polar in ground state than excited: Shorter wavelength & Vive versa.
Increase in polarity of solvent shift n - π* and n - σ* band to shorter wavelength and
π - π* to longer wavelength 14
15. BEER AND LAMBERT’S LAW
Beer’s Law
When monochromatic light passes through an absorbing medium its intensity decreases
exponentially as the concentration of the absorbing medium increases.
Absorbance is directly proportional to the Concentration.
A ∝ C
Lambert’s Law
When monochromatic light passes through an absorbing medium its intensity decreases
exponentially as the length of the absorbing medium increases.
Absorbance is directly proportional to the Pathlength.
A ∝ l
Beer’s Lambert’s law
When a monochromatic light is passed through a solution containing absorbing species,
decrease in intensity of light with Pathlength is proportional to Concentration of solution
and intensity of light.
15
17. I0 It
A > 0, T < 100%
T =
𝐼
𝐼0
Now, A = log10 (
1
𝑇
) ---------- 2), As A ∝
1
𝑇
Put the value of T into equation 2)
A = log10 (
1
𝐼/𝐼0
)
A = log10 (
𝐼0
𝐼
) ----------- 3)
According to equation 1) and 3)
A = ε Cl
17
18. DEVIATION
As per law A ∝ C
Positive deviation: Less concentration, Increase in absorbance
Negative deviation: Increase in concentration, Decrease in absorbance
18
19. Types of Deviation
1. Real deviation
Occurs due to higher concentration of molecules in solution and also due to RI of
absorbing species.
Only dilute solution follows the Beer’s law very closely.
2. Instrumental deviation
It happens mainly due to:
Polychromatic light
Stray radiation reaching the detector
Sensitivity changes in detector employed
Fluctuation in radiation source
Defect in detector amplification system
3. Chemical deviation
Due to chemical changes such as pH, association, dissociation, etc.
If absorbing species react with solvent, they may produce more species in solution with
varying absorptivity value. 19
21. 1. RADIATION SOURCE
It is a device that generates a beam of radiation with sufficient power of energy.
Best source of light is that which is more stable, intense and gives range of spectrum
from 200-800 nm. (UV visible range).
Types of radiation sources:
A) Line sources: emit limited number of radiation
B) Continuous sources: Intensity of radiation changes with wavelength.
Commonly used radiation sources in UV visible spectrophotometer are:
Hydrogen discharge lamp
Duterium lamp
Xenon discharge lamp
Mercury arc lamp
21
22. Hydrogen discharge lamp
Hydrogen gas is stored under high pressure.
When electric discharge is passed through lamp, excited hydrogen
molecules will produced which emits UV radiations.
Deuterium lamp
Similar to hydrogen lamp, but Deuterium is used.
Offers 3-5 times more intensity than hydrogen lamp.
Xenon discharge lamp
Xenon is filled under 10-30 atm pressure.
Has 2 tungsten electrodes.
Greater intensity than hydrogen lamp.
22
23. 2. WAVELENGTH SELECTOR
It consist of Slit, dispersive element.
For the selection of particular wavelength, as light source produce light of different
wavelength.
Used for selection of desired wavelength.
Monochromator
Convert Polychromatic light to monochromatic
Consist of 3 parts:
Entrance slit
Prism or Grating (Dispersive element)
23
24. Types:
1. Prism monochromator
a) Single pass monochromator
b) Double pass monochromator
2. Grating monochromator
Highly polished surface of glass, quartz or fused silica.
Device which get the light from light source & passes the light of desired wavelength
which is required for sample analysis.
24
25. 3. SAMPLE CELL
These are the sample holder made from glass, quartz or fused silica.
Cells, cuvettes or test tube are commonly used for holding the sample.
Reflection Absorption Transmittance
4. DETECTOR (PHOTOMETRIC DETECTOR)
Device that converts light signal into electric signal.
Light or intensity of transmitted radiation by a sample is collected on detector device.
This is to measure the amount of transmitted radiation.
Ideal requirements:
High sensitivity
Low noise level
Short response time
Give quantitative response.
25
27. 1. Phototube detector
Composed of evacuated glass tube which consist of a photocathode (light signal) and
a collector anode (electric signal).
Photocathode is coated with layer of Ca, K which liberate electron that goes to
anode.
This flow of electrons towards anode produce a current which is proportional to
intensity of light radiation.
It is more sensitive because of high degree of amplification is used.
27
28. 2. Photomultiplier tube
Principle: Multiplication of photoelectrons by secondary emission of electrons.
It consist of photocathode and series of anode (dynode chain).
Multiplication of secondary electrons: 4-5 times.
It can detect very low signal.
Finally large number of electrons arrive at anode.
Number of electrons falling on the collector measures the intensity of light incident on
cathode surface.
Response time: 10-9 Second.
28
29. 3. Barrier layer cell
Operates without use of battery.
Made from base plate of metals (Iron or aluminium) that act as electrode on this thin
layer of semiconductor material.
When light radiation falls on selenium layer, electrons are generated at selenium silver
interface.
These electrons are collected by silver, which creates a electric voltage difference
between 2 electrodes i.e. silver and base cell and causes flow of current.
This flow is directly proportional to incident light.
29
30. 5. Signal processor & Read out system
Used for amplification of electric signal & alteration of DC to AC.
Generated electric signal is amplified & finally recorded in computerized system.
30
31. APPLICATIONS
To find extent of conjugation (Conjugation can shift λmax to longer wavelength).
To distinguish between conjugated and non-conjugated compound (Wavelength
comparison).
Identification of isomer.
Identification of unknown compound (Comparing spectrum with known).
Detection of functional group (Presence or absence of chromophore).
Examination of polynuclear hydrocarbons (Multiple aromatic rings).
Structural elucidation.
Identification of compound in different solvents.
Detection of impurities.
31
32. 32
Quantitative analysis
1. Spectrophotometric titrations
Process of determining quantity of sample by adding increment of titrant until end
point.
After titration, plot the graph Absorbance vs Volume of titrant
Titration follows Beer’s law.
2. Single component analysis
If only drug molecule in a sample absorb radiations at λmax of drug.
Method: Standard absorptivity value
By using Beer’s curve
3. Multicomponent analysis
If sample has multiple drugs & all of them absorb radiations at same wavelength.
Excipient also absorb radiations.
Method: Simultaneous equation method
Differential spectrophotometry.
33. 33
References
1. Ashutosh Kar (2005). Pharmaceutical Drug Analysis, New Age International
Publishers, Page no. 293-311.
2. Y. R. Sharma (2013). Elementary Organic Spectroscopy, S. Chand & Company
Pvt. Ltd., Page no. 11-72.
3. Gurdeep Chatwal (2019). Instrumental methods of chemical analysis,
Himalaya Publishing House, Page no. 156-175.
34.
35. 35
Explain principle, instrumentation and applications of UV-Visible spectroscopy.
State & derive Beer’s Lambert’s law along with its limitations.
Describe spectral shift & solvent effect on absorption spectra in UV spectroscopy.
Define: Chromophores, Auxochromes,
Absorbance, Transmittance
Explain the detectors used in UV-Visible spectroscopy.
Discuss the concept of EMR & explain electronic transitions in UV-Visible spectroscopy.
????????????????????????????????