9953056974 Young Call Girls In Mahavir enclave Indian Quality Escort service
Energy of Radiation and Electromagnetic Spectrum
1. Where E= Energy of
radiation
h=Planck’s
constant
µ=Frequency of
radiation
The energy of an EMR can be
given by the following equation:
E=hµ
1
2. Frequency(µ)=
c/
λ
Where c=velocity of light in
vacuum λ= wavelength
2
:-Hence, E=hµ
E=hc/λ
Therefore,energy of a radiation depends upon
frequency and wavelength of radiation.
3. The arrangement obtained by arranging various
types of electromagnetic waves or radiations in
order of their increasing wavelegth or
decreasing frequencies is called
electromagnetic spectrum.
The electromagnetic spectrum is divided into a
number of regions; these are artificial divisions in
the sense that they have been defined solely as
a result of differences in the instrumentation
required for producing and detecting radiation
of a given frequency range. 3
4. Electromagnetic Radiation
Electromagnetic radiation consist of discrete packages
of energy which are called as photons.
Frequency (ν):
◦ It is defined as the number of times electrical field
radiation oscillates in one second.
◦ The unit for frequency is Hertz (Hz).
1 Hz = 1 cycle per second
Wavelength (λ):
◦ It is the distance between two nearest parts of the
wave in the same phase i.e. distance between two
nearest crest or troughs.
5. The relationship between wavelength & frequency can
be written as:
c = ν λ
As photon is subjected to energy, so
E = h ν = h c / λ
7. Gamma Ray Region:0.02-1 A
X- rays:1 -10A
UV-Visible: Vaccum-1-180
uv-180-400
Visible 400-800nm
IR: IR Near: 0.7-2.5 Micron
IR :2.5 -15 Micron
far IR-15-200 Micron
Microwave:0.1mm-1 cm
Radio waves:10m-1cm
11. Spectroscopy is the measurement
and interpretation of electromagnetic
radiation absorbed or emitted when
the molecules or atoms or ions of a
sample moves from one energy state
to another energy state
SPECTROSCOPY
11
12. 1)Atomic spectroscopy :Here,the changes in energy
takes place at atomic level.
Eg: Atomic absorption spectroscopy,
Flame photometry
2)Molecular spectroscopy:Here,the changes in
energy takes place at molecular level.
Eg: UVspectroscopy,colorimetry,infra red
spectroscopy
12
13. Absorption spectrophotometry can be
defined as the measurement of absorption of
radiant energy by various substances.
It involves the measurement of
absorptive capacity for radiant energy in the
visible,UV and IR regions of the spectrum.
13
14. o λ- 400-800nm
o Coloured substance absorbs light of different λ
in different manner and hence get an
absorption curve
o The λ at which maximum absorption takes
place is called as λmax
o λmax is characteristic for every coloured
substance
14
15. On plotting a graph of concentration v/s
absorbance,we get a calibration curve that is
useful in determining the concentration or
amount of a drug substance in the given sample
solution.
17. UV spectroscopy is concerned with the study of
absorption
17
of uv radiation which ranges from 200-400nm.
Valence electrons absorb the energy thereby
molecules undergoes transition from ground
state to excited state.
This absorption is characteristic and depends
on the nature of electrons present.
:
18. 18
UV Spectroscopy
Observed electronic transitions
From the molecular orbital diagram, there are several possible electronic
transitions that can occur, each of a different relative energy:
Energy
s*
p
s
p*
n
s
s
p
n
n
s*
p*
p*
s*
p*
alkanes
carbonyls
unsaturated cmpds.
O, N, S, halogens
carbonyls
21. σ electron from orbital is excited to
corresponding anti-bonding orbital σ*.
The energy required is large for this transition.
The organic compounds in which all the valence
shell electrons are involved in the formation of
σ bond do not show absorption in normal uv
region (200-400nm)
This transition is observed with saturated
compounds.
1) σ-σ*
21
22. Eg: Methane(CH₄) has C-H bond only and can
undergo σ-
σ* transition and shows absorption maxima at
122nm.
The usual spectroscopic technique cannot be used
below
200 nm.
To study this high energy transition,the entire
region should be evacuated (Vacuum uv
region)
Here,the excitation ocuurs with net retention of
electronic spin
This region is less informative 22
23. π electron in a bonding orbital is
excited to corresponding anti-bonding
orbital π*.
Energy required is less when compared to n-σ*
Compounds containing multiple bonds
like
alkenes,alkynes,carbonyls,nitriles,arom
atic compounds etc undergo π-π*
transition.
Eg:Alkenes generally absorb in the region 170-
205nm.
2) π-π*
23
24. Absorption usually occurs in the
ordinary uv spectrophotometer
Absorption bands in unconjugated alkenes
(170- 190nm)
Absorption bands in carbonyls (180 nm)
Introduction of alkyl group in olefinic
linkage produces bathochromic shift
24
25. Saturated compounds containing one hetero
atom with unshared pair of electrons(n) like
O,N,S and halogens are capable of n-σ*
transition.
These transition require less energy
than σ-σ* transition.
In saturated alkyl halides, the energy required
for transition decrease with increase in the size
of halogen atom (or decrease in
electronegativity)
3) n-σ*
25
26. Eg:Methyl chloride has a λmax of
173nm. Methyl iodide has a λmax
of 258nm.
This type of transition is very sensitive to
hydrogen bonding
Eg: Alcohol &amines
Hydrogen bonding shift the uv
absorptions to shorter wavelength.
26
27. An electron from non-bonding orbital is promoted to
anti-bonding π*orbital.
Compounds containing double bonds
involving hetero atoms(C=O,N=O) undergo
such type of transitions.
This transition require minimum energy out of all
transitions and shows absorption band at
longer wavelength around 300nm.
Eg:Saturated aldehydes shows both type of
transitions (n-π*, π-π*) at {low energy and high
energy} around 290 and 180 nm.
4) n-π*
27
30. Chromophore is defined as the nucleus or
any isolated covalently bonded group
responsible for the absorptionof light radiation.
Any group which exhibits absorption of
electromagnetic radiations in the visible or
ultraviolet region.
C=C , C=O ,NO2 etc
Some of the important chromophores
arecarbonyls,acids,esters,nitrile,ethylenic
groups. 30
32. These are co-ordinatively saturated or un-
saturated groups which themselves do not
absorb radiations,but when present alongwith
a chromophore enhances the absorbing
properties of chromophore.
Also known as colour enhancing group.
All auxochromes have one or more non-
bonding pair of electrons.
-NH2 ,-OH ,-OR,-COOH etc
It extend the conjugation of a chromophore by
sharing the non-bonding electrons. 2
9
35. When the absorption maxima(λmax)of a
compound shifts to longer wavelength,it is
known as bathochromic shift or red shift.
The effect is due to the presence of auxochrome
by change of solvent.
Eg: The n-π* transition for carbonyl
compounds experiencesbathochromic shift
when the polarityof solvent is decreased.
1) Bathochromic shift(red shift)
35
36. When the absorption maxima (λmax) of a
compound shifts to a shorter wavelength,it is
known as hypsochromic shift or blue shift.
The effect is due to the presence of a group
causes removal of conjugation or by
change of solvent.
2) Hypsochromic shift(blue shift)
36
37. Eg
:
Aniline shows blue shift in acidic medium since it
loses conjugation.Aniline(280nm) &Anilinium
ion (- 203nm).
37
41. The spectrum consist of sharp peaks and each
peak will correspond to the promotion of
electron from one electronic level to another.
During promotion,the electron moves from a given
vibrational and rotational level within one electronic
mode to the other within the next electronic mode.
Thus,there will be a large no of possible transitions
Hence,not just one but a large no. of wavelengths
which are close enough will be absorbed
resulting in the formation of bands 3
9
42. 1) K Band
42
K-Bands originate due to π-π* transition
from a compound containing a
conjugated system
Such type of bands arise in
compounds like dienes,polyenes and
enones etc.
Compound Transition λmax(nm) εmax
Acetophenon
e
π-π* 240 13,000
1,3-butadiene π-π* 217 21,000
43. R-Band transition originate due to n-π* transition
of a single chromophoric group and having
atleast one lone pair of electrons on the hetero
atom
These are less intense with εmax value below
100
Compound Transition λmax(nm) εmax
Acetone n-π* 270 15
Acetaldehyd
e
n-π* 293 12
43
44. Such type of bands arise due to π-π*
transition in aromatic or hetero-aromatic
molecules.
Benzene shows absorption peaks between
230- 270nm.when a chromophoric group is
attached to the benzene ring ,the B-Bands
are observed at longer wavelengths than
the more intense K-Bands.
Compound Transition λmax(nm) εmax
Benzene π-π* 255 215
Phenol π- π* 270 1450 44
45. E-Band originate due to the electronic
transitions in the benzenoid systems of three
ethylenic bonds which are in closed cyclic
conjugation.
These are further characterized as E1and E2
bands
E1 band which appear at shorter wavelength is
usually more intense than the E2 band for the
same compound which appears at longer
wavelength.
Compound E1 Band E1 Band E2 Band E2 Band
λmax(nm) εmax λmax(nm) εmax
Benzene 184 50,000 204 79,000
Napthalene 221 133,000 286 9,300
45
47. BEER’S LAW
According to this law,when a beam of monochromatic
radiation is passed through a solution of absorbing species,the
intensity of beam of monochromatic light decreases exponentially
with increase in concentration of absorbing species
LAMBERT’S LAW
-dI/dc α I
Lambert’s law states that the rate of decrease of
intensity of monochromatic light with the thickness of the
medium is directly proportional to the intensity of incident
light.
-dI/dt α I
47
48. I₀ is the radiation coming in
I is the radiation coming
out
49. -dI α I
dc
-dI =KI
dc
-dI =Kdc
dc
-lnI=kc+b-----------1
On integration ,b is constant of integration
When concentration is 0 ,no absorbance hence I=l0
Sustituting in equation 1
-lnI0=b
substituting the value of b in eq 1
-lnI=kc-lnI0
lnI0-lnI=kc
ln I0 =kc (since log A –logB = log A)
I B
50. I0 /I = ekc (removing natural logarithm)
I /I0 =e-kc (making inverse on both sides)
I=I0e-kc -------------------2
Lambert’s law
-dI/dt α I
Equation can be simplified like equation 1
Thus we get
I=I0e-kt-----------------------3
Equation 2 and 3 can be combined to get
I=I0e-kct
I/I0=10-kct(converting natural logarithm to base 10)
I0/I=10kct
Log I0/I=Kct(taking log on both sides)----------------4
Transmittance (T)=I/I0 and absorbance(A)=log1/T
Hence A=log 1/I/I0
A=log I0/I----------------5
51. ɛ =E 1% x molecular weight
1cm 10
E1% means the absorbance of 1% w/v Solution 1cm using a
path length of 1cm
57. .
3.CHEMICAL DEVIATIONS:
Association of molecules
This can be explained by taking the examples
of methylene blue at small concentration(10‾⁵
molar) and at concentration above 10‾⁵molar.
57
58. Dissociation of molecules
This can be explained by the fact that
dichromate ions posses their maximum
absorbance at 450nm which is orange in
colour .But upon dilution,it will be dissociated to
chromate ions having maximum absorbance
at 410nm which is yellow in colour.
This law is not valid in case if the absorbing
material is coagulated into a small number of
large units.
This law shows deviation if the absorbing
material at the required wavelength contains
presence of impurities.
58
61. Absorption spectrophotometry in the ultraviolet and visible
region is considered to be one of the oldest physical method
for quantitative analysis and structural elucidation.
Wavelength
• UV- 200-400nm
• VISIBLE- 400-800nm
61
62. PHOTOMETER
SPECTOPHOTOMETER
COLORIMETER
PHOTOMETER: An instrument for measuring the
intensity of light or the relative intensity of a pair of
lights. Also called an illuminometer. It utilizes filter
to isolate a narrow wavelength region.
62
63. SPECTOPHOTOMETER: An instrument measures the
ratio, or a function of the two, of the radiant power of two
EM beams over a large wavelength region. It utilizes
dispersing element (Prisms/Gratings) instead of filters, to
scan large wavelength region.
63
COLORIMETER: An instrument which is used for
measuring absorption in the visible region is generally
called colorimeter.
64. source of radiant energy.
Collimating system.
monochromator system.
sample holder or container to hold sample.
detector system of collecting transmitted radiation.
suitable amplifier or readout device.
64
66. REQUIREMENTS OF AN IDEALSOURCE
It should be stable and should not allow fluctuations.
It should emit light of continuous spectrum of high and
uniform intensity over the entire wavelength region in which
it’s used.
It should provide incident light of sufficient intensity for the
transmitted energy to be detected at the end of optic path.
It should not show fatigue on continued use.
66
67. TUNGSTEN HALOGEN LAMP
67
Its construction is similar to a house hold lamp.
The bulb contains a filament of Tungsten fixed in evacuated
condition and then filled with inert gas.
The filament can be heated up to 3000 k, beyond this
Tungsten starts sublimating.
It is used when polychromatic light is required. To prevent this
along with inert gas some amount of halogen is introduced
(usually Iodine).
68. Sublimated form of tungsten reacts with Iodine to
form Tungsten –Iodine complex.
Which migrates back to the hot filament where it
decomposes and Tungsten get deposited.
DEMERIT:
It emits the major portion of its radiant energy in
near IR region of the spectrum.
68
69. HYDROGEN DISCHARGE LAMP:
In Hydrogen discharge lamp pair of electrodes is enclosed in a
glass tube (provided with silica or quartz window for UV
radiation to pass trough) filled with hydrogen gas.
When current is passed trough these electrodes maintained at
high voltage, discharge of electrons occurs which excites
hydrogen molecules which in turn cause emission of UV
radiations in near UV region.
They are stable and robust.
69
70. XENON DISCHARGE LAMP:
70
It possesses two tungsten electrodes separated by some distance.
These are enclosed in a glass tube (for visible) with quartz or fused
silica and xenon gas is filled under pressure.
An intense arc is formed between electrodes by applying high
voltage. This is a good source of continuous plus additional intense
radiation. Its intensity is higher than the hydrogen discharge lamp.
DEMERIT:
The lamp since operates at high voltage becomes very hot during
operation and hence needs thermal insulation.
71. In mercury arc lamp, mercury vapor is stored under high
pressure and excitation of mercury atoms is done by electric
discharge.
DEMERIT:
Not suitable for continuous spectral studies,(because it doesn’t
give continuous radiations).
71
72. The radiation emitted by the source is collimated (made
parallel) by lenses, mirrors and slits.
LENSES:
Materials used for the lenses must be transparent to the
radiation being used.
Ordinary silicate glass transmits between 350 to 3000 nm
and is suitable for visible and near IR region.
Quartz or fused silica is used as a material for lenses to work
below 300nm.
72
73. MIRRORS
73
These are used to reflect, focus or collimate light beams in
spectrophotometer.
To minimize the light loss, mirrors are aluminized on their
front surfaces.
74. SLITS:
74
Slit is an important device in resolving polychromatic
radiation into monochromatic radiation.
To achieve this, entrance slit and exit slit are used.
The width of slit plays an important role in resolution of
polychromatic radiation.
75. It is a device used to isolate the radiation of the desired
wavelength from wavelength of the continuous spectra.
Following types of monochromatic devices are used.
1. Filters
2. Prisms
3. Gratings
75
76. Selection of filters is usually done on a compromise between
peak transmittance and band pass width; the former should be as
high as possible and latter as narrow as possible.
1. Absorption filters- works by selective absorption of unwanted
radiation and transmits the radiation which is required.
Examples- Glass and Gelatin filters.
76
77. Selection of absorption
the following procedure:
Draw a filter wheel.
filter is done according to
77
Write the color VIBGYOR in clockwise or anticlockwise
manner, omitting Indigo.
78. If solution to be analyzed is BLUE in color a filter having a
complimentary color ORANGE is used in the analysis.
78
Similarly, we can select the required filter in colorimeter, based
upon the color of the solution.
79. An Absorption glass filter is made of solid sheet of glass
that has been colored by pigments which Is dissolved or
dispersed in the glass.
79
The color in the glass filters are produced by
metal oxides like (V, Cr, Mn, Fe, Ni, Co, Cu etc.).
incorporating
80. Gelatin filter is an example of absorption filter prepared by
adding organic pigments; here instead of solid glass sheets thin
gelatin sheets are used. Gelatin filters are not use now days.
80
It tends to deteriorate with time and gets affected by the heat and
moisture. The color of the dye gets bleached.
81. MERITS:-
Simple in construction
Cheaper
Selection of the filter is easy
DEMERITS:-
Less accurate
Band pass (bandwidth) is more (±20-30nm) i.e. if we have
to measure at 400nm; we get radiation from 370-430nm.
Hence less accurate results are obtained.
81
82. Works on the interference phenomenon, causes rejection of
unwanted wavelength by selective reflection.
It is constructed by using two parallel glass plates, which are
silvered internally and separated by thin film of dielectric
material of different (CaF2, SiO, MgF2) refractive index. These
filters have a band pass of 10-15nm with peak transmittance of
40-60%.
82
83. Merits -
Provide greater transmittance and narrower band pass (10-
15nm) as compare to absorption filter.
Inexpensive
Additional filters can be used to cut off undesired wavelength.
83
84.
Prism is made from glass, Quartz or fused silica.
Quartz or fused silica is the choice of material of UV
spectrum.
When white light is passed through glass prism, dispersion
of polychromatic light in rainbow occurs. Now by rotation
of the prism different wavelengths of the spectrum can be
made to pass through in exit slit on the sample.
The effective wavelength depends on the dispersive power
of prism material and the optical angle of the prism.
84
86. • There are two types of mounting in an instrument one is called
‘Cornu type’(refractive), which has an optical angle of 60o
and its adjusted such that on rotation the emerging light is
allowed to fall on exit slit.
• The other type is called “Littrow type”(reflective), which has
optical angle 30o and its one surface is aluminized with
reflected light back to pass through prism and to emerge on the
same side of the light source i.e. light doesn’t pass through the
prism on other side.
86
87. Are most effective one in converting a polychromatic light to
monochromatic light. As a resolution of +/- 0.1nm could be
achieved by using gratings, they are commonly used in
spectrophotometers.
Gratings are of two types.
1. Diffraction grating.
2. Transmission gratings.
87
88. More refined dispersion of light is obtained by means of
diffraction gratings.
These consist of large number of parallel lines ( grooves)
about 15000-30000/ inch is ruled on highly polished surface of
aluminum.
these gratings are replica made from master gratings by
coating the original master grating with a epoxy resin and are
removed after setting
88
89. To make the surface reflective, a deposit of aluminum
is made on the surface. In order to minimize to
greater amounts of scattered radiation and
appearance of unwanted radiation of other spectral
orders, the gratings are blazed to concentrate the
radiation into a single order.
89
90. It is similar to diffraction grating but refraction takes
place instead of reflection. Refraction produces
reinforcement. this occurs when radiation transmitted
through grating reinforces with the partially refracted
radiation.
90
91. Grating gives higher and linear dispersions compared to
prism monochromator.
like
Can be used over wide wavelength ranges.
Gratings can be constructed with materials
aluminium which is resistant to atmospheric moisture.
Provide light of narrow wavelength.
No loss of energy due to absorption.
91
92. Comparison Prism Grating
Made of Glass-: Visible
Quartz/fused silica-: UV
Alkali halide:- IR
Grooved on highly polished
surface like alumina.
Working Principle Angle of Incident Law of diffraction
nλ= d (sini±sinθ)
Merits/demerits Prisms give non-liner
dispersion hence no
overlap of spectral order.
It can’t be used over
consideration wavelength
ranges.
Prisms are not sturdy and
long lasting.
Grating gives liner dispersion
hence overlap of spectral
order.
It can be used over
considerable wavelength
ranges.
Grating are sturdy and long
lasting
92
93. The cells or cuvettes are used for handling liquid samples.
The cell may either be rectangular or cylindrical in nature.
For study in UV region; the cells are prepared from quartz or
fused silica whereas color corrected fused glass is used for
visible region.
The surfaces of absorption cells must be kept scrupulously
clean. No fingerprints or blotches should be present on cells.
Cleaning is carried out washing with distilled water or with
dilute alcohol, acetone.
93
95. Device which converts light energy into electrical signals, that
are displayed on readout devices.
The transmitted radiation falls on the detector which
determines the intensity of radiation absorbed by sample
The following types of detectors are employed in instrumentation
of absorption spectrophotometer
1. Barrier layer cell/Photovoltaic cell
2. Phototubes/ Photo emissive tube
3. Photomultiplier tube
95
96. Requirements of an ideal detector:-
It should give quantitative response.
It should have high sensitivity and low noise level.
It should have a short response time.
It should provide signal or response quantitative to wide
spectrum of radiation received.
96
97. The detector has a thin film metallic layer coated with silver or
gold and acts as an electrode.
It also has a metal base plate which acts as another electrode.
These two layers are separated by a semiconductor layer of
selenium.
97
98. When light radiation falls on selenium layer, electrons become
mobile and are taken up by transparent metal layer.
This creates a potential difference between two electrodes &
causes the flow of current.
When it is connected to galvanometer, a flow of current
observed which is proportional to the intensity and wavelength
of light falling on it.
98
101. Consists of a evacuated glass tube with a photocathode and a
collector anode.
The surface of photocathode is coated with a layer of elements
like cesium, silver oxide or mixture of them.
When radiant energy falls on photosensitive cathode, electrons
are emitted which are attracted to anode causing current to
flow.
More sensitive compared to barrier layer cell and therefore
widely used.
101
102. The principle employed in this detector is that, multiplication
of photoelectrons by secondary emission of electrons.
In a vacuum tube, a primary photo-cathode is fixed which
receives radiation from the sample.
Some eight to ten dynodes are fixed each with increasing
potential of 75-100V higher than preceding one.
Near the last dynode is fixed an anode or electron collector
electrode.
Photo-multiplier is extremely sensitive to light and is best
suited where weaker or low radiation is received
102
104. Depending upon the monochromators (filters or dispersing
device) used to isolate and transmit a narrow beam of radiant
energy from the incident light determines whether the
instrument is classified as Photometer or a Spectrophotometer.
Spectrophotometers used here detects the percentage
transmittance of light radiation, when light of certain
intensity & frequency range is passed through the sample.
Both can be a single beam or double beam optical system.
104
105. • Light from the source is carried through lens and/or through
aperture to pass through a suitable filter.
• The type of filter to be used is governed by the colour of the
solution.
• The sample solution to be analysed is placed in cuvettes.
105
107. After passing through the solution, the light strikes the surface
of detector (barrier-layer cell or phototube) and produces
electrical current.
The output of current is measured by the deflection of needle
of light-spot galvanometer or micro ammeter. This meter is
calibrated in terms of transmittance as well as optical density.
The readings of solution of both standard and unknown are
recorded in optical density units after adjusting instrument to a
reagent blank.
107
109. Double beam instrument is the one in which two beams are
formed in the space by a U shaped mirror called as beam
splitter or beam chopper .
Chopper is a device consisting of a circular disc. One third of
the disc is opaque and one third is transparent, remaining one
third is mirrored. It splits the monochromatic beam of light
into two beams of equal intensities.
109
112. Advantages of single & double
beam spectrophotometer
112
Single beam-
Simple in construction, Easy to use and economical
Double beam-
It facilitates rapid scanning over wide λ region.
Fluctuations due to radiation source are minimised.
It doesn’t require adjustment of the transmittance at 0% and
100% at each wavelength.
It gives ratio of intensities of sample & reference beams
simultaneously.
113. Single
beam
113
Any fluctuation in the intensity of radiation sources affects the
absorbance.
Continuous spectrum is not obtained.
Double beam
Construction is complicated.
Instrument is expensive.
115. 2 Radiant energy intensity
changes with fluctuation
of voltage.
It permits a large degree
of inherent
compensation for
fluctuations in the
intensity of the radiant
energy.
3 It measure the total
amount of transmitted
light reaching the detector
It measures the
percentage of light
absorbed by the sample.
115
116. 4 In single beam it’s not
possible to compare blank
and sample together.
In double beam it’s
possible to do direct one
step comparison of sample
in one path with a standard
in the other path.
5 In single beam radiant
energy wavelength has to
be adjusted every time.
In this scanning can be
done over a wide
wavelength region
6 Working on single beam is
tedious and time
consuming.
Working on double beam is
fast and non tedious.
58
117. Instrumental Analysis, Skoog, Fifth edition, Page no.312-316
Instrumental methods of chemical analysis, Gurdeep R.
chatwal. Page no2.116-2.122
Elementary organic analysis, Principles and chemical
applications , Y R Shrama, page no12-14
A textbook of pharmaceutical analysis, kasturi A V
,Vol 310th
ed., 169-81
11
7
119. Light Source-
Function of temperature
Continuous radiation
Adequate intensity
stable
Visible:
Tungsten filament lamps
Tungsten halogen
Carbon arc
Uv:
Hydrogen-Deuterium discharge :two electrodes in a deuterium filled silica envelop
Xenon discharge
Mercuric arc
Monochromator- Filters:
Absorption filters.. complementary
Interference filters…glass plates .. Silvered internally… thin film of dielectric material..
Monochromaters
Prisms:
Refractive type
120. 120
Instrumentation – Sample Handling
1. Virtually all UV spectra are recorded solution-phase
2. Cells can be made of plastic, glass or quartz
3. Only quartz is transparent in the full 200-700 nm range; plastic and glass
are only suitable for visible spectra
4. Concentration (we will cover shortly) is empirically determined
A typical sample cell (commonly called a cuvet):
121. 121
5. Solvents must be transparent in the region to be observed; the
wavelength where a solvent is no longer transparent is referred to as the
cutoff
6. Since spectra are only obtained up to 200 nm, solvents typically only
need to lack conjugated p systems or carbonyls
Common solvents and cutoffs:
acetonitrile 190
chloroform 240
cyclohexane 195
1,4-dioxane 215
95% ethanol 205
n-hexane 201
methanol 205
isooctane 195
water 190
122. Detectors:
Barrier layer cell:
No power supply
Current proportional to light intensity
Metallic plate… layer selenium …conducting material layer…
Photo tubes
Evacuated tubes…photo catode … collector anode…
Photo multiplier tubes
Sensitive and expensive
Multiplication of initial photo electons
Several anodes with gradually increasing potential..
Never goes zero
Dark current