International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2302
DIAGNOSIS OF DIABETES MELLITUS USING ACETONE AS BIOMARKER IN
HUMAN BREATH
Naveen Kumar T1, Vijayakumar J2, Sruthi K3
1PG Student, Department of Electronics and Instrumentation, Bharathiar University, Coimbatore, Tamilnadu,
India
2Associate Professor, Department of Electronics and Instrumentation, Bharathiar University, Coimbatore,
Tamilnadu, India
3Ph.D Research Scholar, Department of Electronics and Instrumentation, Bharathiar University, Coimbatore,
Tamilnadu, India
-----------------------------------------------------------------------***-------------------------------------------------------------------
ABSTRACT
This project is to design and develop a non-invasive method to diagnose diabetes mellitus which is commonly known
as Diabetes. It is a group of metabolic disorders characterized by a high blood sugar level over a prolonged period of
time. It is essential to monitor the blood glucose levels of a diabetic patient frequently. As of 2019, an estimated 463
million people had diabetes worldwide and it causes approximately 4.2 million deaths. It is the 7th leading causation of
death globally. Hence it is necessary to diagnose the disease earlier, so that we can prevent the patient from suffering.
It can be done using a technique called breath analysis. Breath analysis is a technique in which the diseases in human
beings can be diagnosed using human breath. Human breath naturally consists of over 3500 Volatile Organic
Compounds (VOC) other than of major gases like oxygen, nitrogen and carbon dioxide. Each VOC is a biomarker to
various diseases and metabolism of our body. Likewise, acetone is one of the 3500 VOCs and it is a biomarker to
diabetes mellitus. Hence the measurement of acetone gas in our breath leads to the detection of diabetes mellitus. The
normal range of acetone gas in our breath is less than 0.8 ppm. If the value exceeds the normal range, the person is
affected with diabetes. The acetone level in human breath is detected using a Grove HCHO – VOC gas sensor. The
sensor is connected to the microcontroller which will control the entire operation. The data collected from the sensor
will be analyzed for the disease. If an abnormal change is detected in the level of acetone, it will alert the person such
that the person requires immediate medication. The main advantage of this project is the non-invasive way of
detecting diabetes which does not involve any pain staking invasive laboratory methods.
Keywords: Diabetes Mellitus, Biomarker, Diagnose, Acetone, Breath Analysis.
1. INTRODUCTION
Diabetes mellitus (DM) is a disease of imperfect control of blood levels of glucose. In diabetes mellitus, human body
has trouble moving glucose, which is a type of sugar, from blood into cells. This leads to the glucose level high in blood
and not enough of it in cells, and the cells need glucose as a source of energy. In general, the human body controls how
much glucose is in the blood relative to how much gets into the cells with the help of two hormones: insulin and
glucagon. Insulin is used to reduce blood glucose levels, and glucagon is used to increase blood glucose.
Insulin is secreted by beta cells in the center of the islets, and glucagon is secreted by alpha cells in the periphery of
the islets. Insulin decreases the amount of glucose in the blood and glucagon does exactly the opposite, it increases the
blood glucose level in blood. Diabetes mellitus is diagnosed when the blood glucose levels get too high, and this is seen
among 10% of the world population. There are two dominant types of diabetes - Type 1 and Type 2 and has
subclassifications – gestational diabetes, drug-induced diabetes and diabetic ketoacidosis (DKA). About 10% of people
with diabetes have Type 1, and the remaining 90% of people with diabetes have Type 2.
The Type 1 diabetes results from failure of the pancreas to produce enough insulin due to loss of beta cells. It appears
as childhood or adolescence; it can also develop in adults. In diabetes Type 1, destruction of beta cells usually starts
early in life, but sometimes up to 90% of the beta cells are destroyed before symptoms crop up. Four clinical
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2303
symptoms of uncontrolled diabetes, that all sound similar, are polyphagia (poly - a lot; phaga - eating), glycosuria
(glucose - glucose; uria – urine), polyuria (poly – a lot; uria – urine), and polydipsia (poly – a lot; dipsia – thirst).
Type 2 diabetes begins with insulin resistance, a condition in which cells fail to respond to insulin properly. It appears
commonly among adults.
Gestational diabetes occurs when pregnant women without a previous history of diabetes develop high blood sugar
levels. It returns to normal soon after delivery.
One really serious complication with diabetes type 1 is called diabetic ketoacidosis (DKA). Diabetic ketoacidosis (DKA)
is a life-threatening problem of diabetes mellitus. Hence it is essential to detect the diabetes level and to undertake the
necessary treatment in order to maintain it. Otherwise, it causes serious impact on patient’s health and finally it leads
to death. DKA impact from a shortage of insulin; in response, the body switches to burn fatty acids, which produces
ketone bodies in acidic nature. That ketone bodies break down into acetone and escape as a gas while exhaled from
the lungs, which gives a sweet fruity smell to a person’s breath. DKA is diagnosed when testing finds low blood pH,
high blood sugar, and ketoacids in either the urine or blood. In these type of diabetes produces ketone bodies in blood
urine and breath (acetone – which produces fruity smell).
Currently, diabetes mellitus is diagnosed by both invasive (finger pricking -blood test) and non-invasive (urine test)
methods. Here this is one of the non-invasive methods of diagnosing using breath. Human breath analysis is one of the
earlier methods to diagnose disease. There are about over 3500 Volatile Organic Compounds present in our exhaled
breath, one of such VOC is acetone. The VOC’s contained in exhaled human breath are detected with the help of the
sensors. It is found that acetone acts as the biomarker for diabetes mellitus. By detecting and mapping the
concentration of acetone present in the exhaled human breath we can observe the prior stages of the disease. This can
be done with the help of breath analysis technique. By mapping the concentration of acetone and diagnosing with the
standard values we can ensure that the patient is affected with diabetes or not. Detection of acetone can be done with
the help of GROVE – HCHO VOC gas sensor. The normal acetone level in exhaled breath of healthy person is less than
(0.1 PPM – 0.8 PPM). The patients who are affected from diabetes will have acetone concentration more than (1.8
PPM – 5.0 PPM). The detected concentration is displayed in the display so that they know their state of health.
Through this method, we can ensure a person’s safety by continuously monitoring the concentration of ammonia in
exhaled breath at a particular time interval. It is one of the cost-efficient ways to detect the disease such that the poor
people can afford it easily. One of the main advantages of this method that it does not involve risk factors like finger
pricking and pain stacking methods involved in laboratory methods.
2. LITERATURE REVIEW
Fahad Usman., et al. [1] identified acetone as a non-invasive biomarker for diagnosing diabetes from exhaled human
breath. They studied and discussed about diabetes, death rate, biomarker and biosensors, and concentration of
acetone for normal (0.1 ppm - 0.8 ppm) and diabetes (1.8 ppm - 5.0 ppm) persons. They gathered standard
concentration from WHO.
Claire Turner., [2] investigates the potential for monitoring compounds in breath and emitted through skin for
inferring blood glucose concentration. Potential markers and an assessment of their suitability for non-invasive
monitoring are discussed. The varying technologies developed for monitoring VOCs (volatile organic compounds) in
breath and from skin of diabetics using acetone as a biomarker. They discussed about breath analysis and
concentration of acetone in diabetics and non-diabetics person.
Valentine Saasa., et al. [3] describe the methods used in the field of breath analysis to monitor and diagnose diabetes
mellitus. Currently the diagnosis and monitoring of blood glucose and ketone bodies that are used in clinical studies
involve the use of blood tests. This method entails pricking fingers for a drop of blood and placing a drop on a
sensitive area of a strip which is pre-inserted into an electronic reading instrument. Furthermore, it is painful,
invasive and expensive, and can be unsafe if proper handling is not undertaken. Human breath analysis provides a
non-invasive and rapid method for detecting various VOCs (volatile organic compounds) that are indicators for
different diseases. In patients with diabetes mellitus, the body produces excess amounts of ketones such as beta-
hydroxybutyrate, acetoacetate, and acetone. Acetone is exhaled during respiration.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2304
Dixit K., et al. [4] research on non-invasive blood glucose monitoring has led to the development of various devices,
easing diabetes management through comfortable, minimally invasive/non-invasive continuous blood-glucose
monitoring. Nevertheless, they still suffer from issues such as lag time and the requirement of frequent calibration.
Breath analysis is newly budding domain for diagnosing and monitoring the disease using non-invasive method. It is
painless, safe, and allows repetitive sampling. However, the intricacies of breath analysis need to be well-studied to
enable accurate, reliable, and reproducible monitoring.
Righettoni M., et al. [5] proposed about Si-doped WO3 chemo-resistive gas sensors have been fabricated by flame
spray pyrolysis. These nanoparticles had excellent acetone sensing properties with a great potential for application in
non-invasive medical diagnostic by breath analysis. In fact, theWO3 ε-phase thermal stability was enhanced greatly by
Si-doping (up to 500 °C) allowing detection in the ppb range of acetone in ideal (dry air) and realistic (90%) breath
conditions. However, gas sensing performances such as the sensor response and the response time are also dependent
on the measurement system and conditions.
Maureen I. Harris., et al. [6] proposed that most of the US people do not check their blood glucose level because of cost
and income. So, they proposed about the importance of blood glucose control in the prevention of diabetes
complications and the role of self-monitoring in achieving blood glucose control, it may be prudent for physicians and
their patients to make greater use of this technique.
Chen, X., et al. [7] proposed an innovative method of lung cancer pathologic analysis and early diagnosis at the cellular
level was introduced. An optimum HS-SPME-GC method was used to detect the odors of some cells in human breath,
including squamous cell carcinoma cells, adenocarcinoma cells, bronchioloalveolar carcinoma cells, non-small cell
carcinoma cells, bronchial epithelial cells, tastebud cells, osteogenic cells, and lipocytes, and we were able to obtain an
odor chromatogram of these cells.
3. METHODOLOGY
In this chapter we are going to discuss about the proposed methodology or working of our project. In this project we
proposed Human Breath Analysis technique to identify the diabetes mellitus in a human. It is a technique in which
various diseases in humans can be identified easily through exhaled human breath. In this technique, human breath is
analyzed for the normal concentration of ammonia in order to identify the liver cirrhosis. One of the biggest
advantages of this technique is the early detection of diseases in human beings in a non- invasive way.
Figure 2.1 HUMAN BREATH ANALYSIS TECHNIQUE
It is done with the help of gas sensing technology. For sensing the acetone gas presented in our human breath as one
of the 3500 VOCs, we have used a type of acetone gas sensor called Grove HCHO VOC gas sensor. It is a general-
purpose sensor that it can be used for various purposes. The sensor is powered with the microcontroller which has a
power source pin of 3.3V. The microcontroller used here is PIC16F877A which is the brain of this entire system.
The sensor will sense the input acetone gas as soon as it is powered. At the same time, the microcontroller was
programmed in such a way to read the input signal. Here the analog output pin of the sensor is used to sense the gas
since it gives the raw values continuously. The analog output from the sensor is not in a readable form. Hence it is
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2305
converted into readable digital form with the help of 10-bit Analog to Digital Converter which is already present in the
microcontroller.
The output which is sensed by the sensor is analyzed and compared with the normal standard concentration of the
acetone gas which is also programmed in the microcontroller. The normal standard concentration of acetone in
exhaled human breath is found out by performing the literature survey. The corresponding output will be displayed in
the displaying unit. Hence the person can be verified with their corresponding present health status. If the output
value is confirmed with diabetes mellitus, person have to get medication at right time. If the output is normal, the
entire process can be repeated in case of any critical situation occurs in future.
It will greatly pave the way for detecting diseases using this breath analysis technique in these modern days as it
possesses many advantages over other invasive laboratory methods. The test can also be done in a short period of
time.
Figure 2.2 WORK FLOW PROCESS
2.1 SENSOR CALIBRATION
The calibration of sensor is very important in doing any project work. It is very important before using any sensors in
project. It is a process of rectifying the undesired outputs to desired outputs. Sensors are made up of electronic and
electrical components. Since those components are sensitive to various factors like temperature, humidity, external
shocks and vibration etc., it will lead to the undesired output results. The sudden change in the working environment
also leads to the deviation of output results such that the measured outputs differ from the expected outputs. Hence
calibration plays a significant role in increases the chance of getting desired output results. It will improve the
performance and accuracy of the sensor by removing errors in measurements.
In our project, we use GROVE HCHO – VOC gas sensor for detecting acetone gas in human breath. Hence it is necessary
to calibrate the sensor for accurate output. The operation of this sensor is based on the changing resistance of
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2306
the heated coil thus measuring the concentration of the particular gas. Two parameters are involved while calibrating
the sensor. They are as follows:
 R0 – Sensor resistance of instantaneous concentration of measuring gas
 Rs – Internal resistance of the sensor changing in clean air or environment
The concentration measurement of a particular gas is also depending on the ratio of Rs and R0. Hence it is found out
as a primary stage of calibrating the sensor. The sensor will show some random values as output when it is connected
to microcontroller. It is necessary to calibrate it to zero. For that purpose, we are going to take average of that random
values as a secondary step of calibration. Hence, we coded the microcontroller in such a way to calculate the average
value using looping controls. The loop will run for 500 times and the average of the values was taken.
The calculated average was converted into equivalent sensor voltage since this conversion calibrate the sensor to zero
in an effective manner. The Rs value of the sensor is also calculated every time by converting the values into
equivalent sensor voltage. The obtained R0 value will be used to find the ratio for changing Rs value at every step.
Figure 2.3 SENSOR OUTPUT BEFORE AND AFTER CALIBRATION
4. MODELING AND ANALYSIS
In this project we are using a GROVE HCHO VOC gas sensor to analyze the concentration of acetone in human breath.
Hence this experiment plays a very important role to analyze the characteristics of the sensor and to determine the
applicability of this particular type sensor in real time monitoring. It is conducted to verify that the above used sensor
meets the above characteristics for the efficient measurement of acetone in human breath in real time application.
3.1 ANALYSIS I
The analysis is carried out with the detection of acetone gas which is to be measured in human breath. But human
breath has minute level of acetone and it is volatile. So, acetone solution from laboratory is taken as sample for this
analysis. The general calibration of the sensor is carried out and made the sensor ready to detect the acetone. To
generate the acetone odor or gas, we have used acetone solution in same concentrations and in different environment.
The acetone solution (10ml) was taken in a beaker to conduct the experiment. Before commencing of the experiment,
the sensor was preheated by supplying the power for about thirty minutes for obtaining the best results. This analysis
was carried out in an open environment such that the gas sensor is placed near to the beaker containing acetone
solution. Then the sensor detected the acetone gas which is evolved from the solution. The data obtained from the
sensor is viewed in the monitor and it is recorded for further analysis. The data was collected for every second and the
collected data was tabulated for further analysis.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2307
TABLE 3.1 READINGS OF ACETONE DATA 1
TIME (IN S) CONCENTRATION OF ACETONE (IN PPM)
1 0.30
5 0.39
10 0.60
15 0.83
20 1.11
25 1.43
30 1.80
35 2.21
40 2.48
45 2.62
55 2.82
60 2.92
The data which is tabulated are used to plot the graph against time in seconds for analyzing the response of sensor in
detecting the concentration of acetone gas which is released from the acetone solution. From the graph, we inferred
the characteristics of sensor in determining the acetone concentration in solution clearly.
Figure 3.1 GRAPHICAL REPRESENTATION OF ACETONE DATA 1
3.2 ANALYSIS II
Another experiment analysis using the different concentration of acetone solution (10ml) was carried out to analyze
the performance of sensor in detecting the gas. In this analysis, the experiment was done in a closed environment.
Then the sensor detected the acetone gas which is evolved from the solution. The data obtained from the sensor is
viewed in the monitor and it is recorded for further analysis. The data was collected for every second and the collected
data was tabulated for further analysis.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2308
Table 3.2 READINGS OF ACETONE DATA 2
TIME (IN S) CONCENTRATION OF ACETONE (IN PPM)
1 0.40
5 0.53
10 0.70
15 0.95
20 1.20
25 1.45
30 1.70
35 1.95
40 2.20
45 2.45
50 2.70
55 2.95
60 3.30
The data which is tabulated are used to plot the graph against time in seconds for analyzing the response of sensor in
detecting the concentration of acetone gas which is released from the acetone solution. From the graph, we inferred
the characteristics of sensor in determining the acetone concentration in solution clearly.
Figure 3.2 GRAPHICAL REPRESENTATION OF ACETONE DATA 2
5. RESULTS AND DISCUSSION
Experiment analysis was carried out regarding the performance characteristics of the sensor and its behavior. The
analysis was done in different concentrations of acetone solution. In this chapter we will discuss about the results
obtained from each concentration.
In experiment analysis I, acetone solution (10ml) is used for generating acetone odor and the gas sensor is made to
detect the gas in open environment. The data was collected and it is observed that the response of the sensor towards
acetone gas is very high. The results obtained from the sensor was also identified as very high in detecting the acetone
gas in open environment.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2309
In the experiment analysis II, the same concentration of acetone solution (10 ml) is used and the data was collected
from the sensor in closed environment. It is also observed from the results that the sensor is very sensitive in
detecting the acetone gas. The output obtained from the sensor proved its high sensitivity again showing in its crystal-
clear data and the same is verified in plotting the data in graph. Hence from both the analysis I and II, it is found that
the sensor has high sensitivity towards acetone gas and proved that the usage of Grove HCHO VOC gas sensor in this
project is correct. In this experiment, PIC16F877A and NodeMCU is used. PIC is user friendly and cost efficient. So, PIC
is used for diagnose and analyze of acetone and the output is displayed in LCD (16X2). Here NodeMCU is used for the
same purpose of analysis and the output is displayed in mobile application (Blynk application).
Figure 7.1 LCD OUTPUT
All the necessary characteristics of sensor was studied verified by conducting different experimental analysis.
Hence the selection of Grove HCHO VOC gas sensor for this project is proved as a rightful choice to detect the
abnormal detection of acetone which will be used in detecting diabetes mellitus. The sensor has a range of 10PPM –
1000PPM in detecting acetone but the concentration of acetone in breath actually falls under 0.8 – 5.0 PPM. Hence it
requires a low range of sensors for detecting acetone in exhaled breath for that level. The sensor has to be specifically
designed using MOS technologies to apply this project in real time. So, I have done this project as a prototype model
suggesting that this can be done in future appropriately.
6. CONCLUSION
Diagnosis of diabetes mellitus using human breath will play a major role in future. Research found that human breath
analysis technique has a great potential to diagnose various other diseases like diabetes mellitus. The simple and cost-
efficient means of diagnosing diabetes mellitus and is one of the future techniques. The time consumption of
diagnosing the disease is also very less comparing to traditional laboratory diagnosing methods. An important
highlight of this diagnosing technique is the non-invasive way such that it only uses the exhaled human breath for
diagnosis. It does not involve any finger pricking method of diagnosis which leads to pain staking. It is very safe to
undergo the diagnosis and it does not involve any hazardous procedures unlike in other X-rays or MRI scanning
methods. This technique has many complexities like requirement of standard procedures, designing of appropriate
gas sensors with low range and high accuracy but if it is are currently going for standards and procedures in
incorporating this technique in today’s world. Further enhancements can be done in future like
Figure 7.2 BLYNK APPLICATION OUTPUT.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072
© 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2310
 Diagnosing through Image Processing Techniques
 Perform Machine Learning (ML) or AI algorithms
 Application of IoT techniques
 Data Collection using Cloud Technologies
It will be an additional advantage for this project if the above-mentioned technologies get added to this project. Then
it will become as a very suitable method of diagnosing diseases in this modern world. It will also be very useful for the
patients also for periodical analysis of their state of health.
7. REFERENCES
[1] Usman, F., Dennis, J.O., Ahmed, A.Y., Meriaudeau, F., Ayodele, O.B. and Rabih, A.A., 2018. A review of biosensors for
non-invasive diabetes monitoring and screening in human exhaled breath. Ieee Access, 7, pp.5963-5974.
[2] Turner, C., 2011. Potential of breath and skin analysis for monitoring blood glucose concentration in
diabetes. Expert review of molecular diagnostics, 11(5), pp.497-503.
[3] Saasa, V., Malwela, T., Beukes, M., Mokgotho, M., Liu, C.P. and Mwakikunga, B., 2018. Sensing technologies for
detection of acetone in human breath for diabetes diagnosis and monitoring. Diagnostics, 8(1), p.12.
[4] Dixit, K., Fardindoost, S., Ravishankara, A., Tasnim, N. and Hoorfar, M., 2021. Exhaled Breath Analysis for Diabetes
Diagnosis and Monitoring: Relevance, Challenges and Possibilities. Biosensors, 11(12), p.476.
[5] Righettoni, M. and Tricoli, A., 2011. Toward portable breath acetone analysis for diabetes detection. Journal of
breath research, 5(3), p.037109.
[6] Harris, M.I., Cowie, C.C. and Howie, L.J., 1993. Self-monitoring of blood glucose by adults with diabetes in the United
States population. Diabetes care, 16(8), pp.1116-1123.
[7] Chen, X., Xu, F., Wang, Y., Pan, Y., Lu, D., Wang, P., Ying, K., Chen, E. and Zhang, W., 2007. A study of the volatile
organic compounds exhaled by lung cancer cells in vitro for breath diagnosis. Cancer: Interdisciplinary International
Journal of the American Cancer Society, 110(4), pp.835-844.

More Related Content

Similar to DIAGNOSIS OF DIABETES MELLITUS USING ACETONE AS BIOMARKER IN HUMAN BREATH

Nano Robots for Continuous Blood Glucose Diagnosis
Nano Robots for Continuous Blood Glucose DiagnosisNano Robots for Continuous Blood Glucose Diagnosis
Nano Robots for Continuous Blood Glucose Diagnosisijtsrd
 
IJBET-4-Narasimham et al
IJBET-4-Narasimham et alIJBET-4-Narasimham et al
IJBET-4-Narasimham et alGaurav Kaila
 
F inal presentation vernon ems glucometer smr 2015
F inal presentation vernon ems glucometer smr 2015F inal presentation vernon ems glucometer smr 2015
F inal presentation vernon ems glucometer smr 2015cmarth
 
SMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEM
SMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEMSMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEM
SMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEMijics
 
Smart Phone for Personal Health Care Monitoring System
Smart Phone for Personal Health Care Monitoring SystemSmart Phone for Personal Health Care Monitoring System
Smart Phone for Personal Health Care Monitoring Systemijcisjournal
 
Future Of Nanomedicine In Diabetes Treatment.pdf
Future Of Nanomedicine In Diabetes Treatment.pdfFuture Of Nanomedicine In Diabetes Treatment.pdf
Future Of Nanomedicine In Diabetes Treatment.pdfPharmacy 7065
 
IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...
IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...
IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...IRJET Journal
 
Kidney Diseases Prediction Using Hybrid Ensemble Learning
Kidney Diseases Prediction Using Hybrid Ensemble LearningKidney Diseases Prediction Using Hybrid Ensemble Learning
Kidney Diseases Prediction Using Hybrid Ensemble LearningIRJET Journal
 
Impact of patient counseling on diabetes mellitus patients in the territory c...
Impact of patient counseling on diabetes mellitus patients in the territory c...Impact of patient counseling on diabetes mellitus patients in the territory c...
Impact of patient counseling on diabetes mellitus patients in the territory c...SriramNagarajan17
 
Diagnosis of diabetes using classification mining techniques [
Diagnosis of diabetes using classification mining techniques [Diagnosis of diabetes using classification mining techniques [
Diagnosis of diabetes using classification mining techniques [IJDKP
 
Updates on management of Adolescent PCOS An evidence based approach
Updates on management of Adolescent PCOS An evidence based approachUpdates on management of Adolescent PCOS An evidence based approach
Updates on management of Adolescent PCOS An evidence based approachAboubakr Elnashar
 
The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...
The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...
The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...YogeshIJTSRD
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightIRJET Journal
 
Blood Glucose (Sugar)
Blood Glucose (Sugar)Blood Glucose (Sugar)
Blood Glucose (Sugar)Tayyab Minhas
 
The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...
The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...
The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...Randox Reagents
 

Similar to DIAGNOSIS OF DIABETES MELLITUS USING ACETONE AS BIOMARKER IN HUMAN BREATH (20)

Nano Robots for Continuous Blood Glucose Diagnosis
Nano Robots for Continuous Blood Glucose DiagnosisNano Robots for Continuous Blood Glucose Diagnosis
Nano Robots for Continuous Blood Glucose Diagnosis
 
IJBET-4-Narasimham et al
IJBET-4-Narasimham et alIJBET-4-Narasimham et al
IJBET-4-Narasimham et al
 
F inal presentation vernon ems glucometer smr 2015
F inal presentation vernon ems glucometer smr 2015F inal presentation vernon ems glucometer smr 2015
F inal presentation vernon ems glucometer smr 2015
 
SMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEM
SMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEMSMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEM
SMART PHONE FOR PERSONAL HEALTH CARE MONITORING SYSTEM
 
6416ijics01
6416ijics016416ijics01
6416ijics01
 
Smart Phone for Personal Health Care Monitoring System
Smart Phone for Personal Health Care Monitoring SystemSmart Phone for Personal Health Care Monitoring System
Smart Phone for Personal Health Care Monitoring System
 
Future Of Nanomedicine In Diabetes Treatment.pdf
Future Of Nanomedicine In Diabetes Treatment.pdfFuture Of Nanomedicine In Diabetes Treatment.pdf
Future Of Nanomedicine In Diabetes Treatment.pdf
 
IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...
IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...
IRJET - Complications of Diabetes with Cardiovascular Diseases Diagnostic usi...
 
Research Paper
Research PaperResearch Paper
Research Paper
 
Kidney Diseases Prediction Using Hybrid Ensemble Learning
Kidney Diseases Prediction Using Hybrid Ensemble LearningKidney Diseases Prediction Using Hybrid Ensemble Learning
Kidney Diseases Prediction Using Hybrid Ensemble Learning
 
Impact of patient counseling on diabetes mellitus patients in the territory c...
Impact of patient counseling on diabetes mellitus patients in the territory c...Impact of patient counseling on diabetes mellitus patients in the territory c...
Impact of patient counseling on diabetes mellitus patients in the territory c...
 
Diagnosis of diabetes using classification mining techniques [
Diagnosis of diabetes using classification mining techniques [Diagnosis of diabetes using classification mining techniques [
Diagnosis of diabetes using classification mining techniques [
 
Diabetes
DiabetesDiabetes
Diabetes
 
Updates on management of Adolescent PCOS An evidence based approach
Updates on management of Adolescent PCOS An evidence based approachUpdates on management of Adolescent PCOS An evidence based approach
Updates on management of Adolescent PCOS An evidence based approach
 
The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...
The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...
The Role of Gedunin Against Aquaporin 2 for the Possible Treatment of Diabete...
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
 
Biosensor lab report
Biosensor lab reportBiosensor lab report
Biosensor lab report
 
Blood Glucose (Sugar)
Blood Glucose (Sugar)Blood Glucose (Sugar)
Blood Glucose (Sugar)
 
Journal of Research in Diabetes & Metabolism
Journal of Research in Diabetes & MetabolismJournal of Research in Diabetes & Metabolism
Journal of Research in Diabetes & Metabolism
 
The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...
The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...
The Role of Adiponectin in Obesity and its Clinical Utility in Obesity-Associ...
 

More from IRJET Journal

TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...IRJET Journal
 
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURESTUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTUREIRJET Journal
 
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...IRJET Journal
 
Effect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil CharacteristicsEffect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil CharacteristicsIRJET Journal
 
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...IRJET Journal
 
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...IRJET Journal
 
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...IRJET Journal
 
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...IRJET Journal
 
A REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADASA REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADASIRJET Journal
 
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...IRJET Journal
 
P.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD ProP.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD ProIRJET Journal
 
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...IRJET Journal
 
Survey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare SystemSurvey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare SystemIRJET Journal
 
Review on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridgesReview on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridgesIRJET Journal
 
React based fullstack edtech web application
React based fullstack edtech web applicationReact based fullstack edtech web application
React based fullstack edtech web applicationIRJET Journal
 
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...IRJET Journal
 
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.IRJET Journal
 
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...IRJET Journal
 
Multistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic DesignMultistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic DesignIRJET Journal
 
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...IRJET Journal
 

More from IRJET Journal (20)

TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
 
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURESTUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
 
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
 
Effect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil CharacteristicsEffect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil Characteristics
 
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
 
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
 
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
 
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
 
A REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADASA REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADAS
 
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
 
P.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD ProP.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD Pro
 
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
 
Survey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare SystemSurvey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare System
 
Review on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridgesReview on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridges
 
React based fullstack edtech web application
React based fullstack edtech web applicationReact based fullstack edtech web application
React based fullstack edtech web application
 
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
 
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
 
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
 
Multistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic DesignMultistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic Design
 
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
 

Recently uploaded

Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)Suman Mia
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVRajaP95
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerAnamika Sarkar
 
Current Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLCurrent Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLDeelipZope
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...RajaP95
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxDeepakSakkari2
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxupamatechverse
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...Call Girls in Nagpur High Profile
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxpranjaldaimarysona
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZTE
 

Recently uploaded (20)

Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)Software Development Life Cycle By  Team Orange (Dept. of Pharmacy)
Software Development Life Cycle By Team Orange (Dept. of Pharmacy)
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
 
Current Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCLCurrent Transformer Drawing and GTP for MSETCL
Current Transformer Drawing and GTP for MSETCL
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
IMPLICATIONS OF THE ABOVE HOLISTIC UNDERSTANDING OF HARMONY ON PROFESSIONAL E...
 
Biology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptxBiology for Computer Engineers Course Handout.pptx
Biology for Computer Engineers Course Handout.pptx
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptx
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
High Profile Call Girls Nashik Megha 7001305949 Independent Escort Service Na...
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptx
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
ZXCTN 5804 / ZTE PTN / ZTE POTN / ZTE 5804 PTN / ZTE POTN 5804 ( 100/200 GE Z...
 

DIAGNOSIS OF DIABETES MELLITUS USING ACETONE AS BIOMARKER IN HUMAN BREATH

  • 1. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2302 DIAGNOSIS OF DIABETES MELLITUS USING ACETONE AS BIOMARKER IN HUMAN BREATH Naveen Kumar T1, Vijayakumar J2, Sruthi K3 1PG Student, Department of Electronics and Instrumentation, Bharathiar University, Coimbatore, Tamilnadu, India 2Associate Professor, Department of Electronics and Instrumentation, Bharathiar University, Coimbatore, Tamilnadu, India 3Ph.D Research Scholar, Department of Electronics and Instrumentation, Bharathiar University, Coimbatore, Tamilnadu, India -----------------------------------------------------------------------***------------------------------------------------------------------- ABSTRACT This project is to design and develop a non-invasive method to diagnose diabetes mellitus which is commonly known as Diabetes. It is a group of metabolic disorders characterized by a high blood sugar level over a prolonged period of time. It is essential to monitor the blood glucose levels of a diabetic patient frequently. As of 2019, an estimated 463 million people had diabetes worldwide and it causes approximately 4.2 million deaths. It is the 7th leading causation of death globally. Hence it is necessary to diagnose the disease earlier, so that we can prevent the patient from suffering. It can be done using a technique called breath analysis. Breath analysis is a technique in which the diseases in human beings can be diagnosed using human breath. Human breath naturally consists of over 3500 Volatile Organic Compounds (VOC) other than of major gases like oxygen, nitrogen and carbon dioxide. Each VOC is a biomarker to various diseases and metabolism of our body. Likewise, acetone is one of the 3500 VOCs and it is a biomarker to diabetes mellitus. Hence the measurement of acetone gas in our breath leads to the detection of diabetes mellitus. The normal range of acetone gas in our breath is less than 0.8 ppm. If the value exceeds the normal range, the person is affected with diabetes. The acetone level in human breath is detected using a Grove HCHO – VOC gas sensor. The sensor is connected to the microcontroller which will control the entire operation. The data collected from the sensor will be analyzed for the disease. If an abnormal change is detected in the level of acetone, it will alert the person such that the person requires immediate medication. The main advantage of this project is the non-invasive way of detecting diabetes which does not involve any pain staking invasive laboratory methods. Keywords: Diabetes Mellitus, Biomarker, Diagnose, Acetone, Breath Analysis. 1. INTRODUCTION Diabetes mellitus (DM) is a disease of imperfect control of blood levels of glucose. In diabetes mellitus, human body has trouble moving glucose, which is a type of sugar, from blood into cells. This leads to the glucose level high in blood and not enough of it in cells, and the cells need glucose as a source of energy. In general, the human body controls how much glucose is in the blood relative to how much gets into the cells with the help of two hormones: insulin and glucagon. Insulin is used to reduce blood glucose levels, and glucagon is used to increase blood glucose. Insulin is secreted by beta cells in the center of the islets, and glucagon is secreted by alpha cells in the periphery of the islets. Insulin decreases the amount of glucose in the blood and glucagon does exactly the opposite, it increases the blood glucose level in blood. Diabetes mellitus is diagnosed when the blood glucose levels get too high, and this is seen among 10% of the world population. There are two dominant types of diabetes - Type 1 and Type 2 and has subclassifications – gestational diabetes, drug-induced diabetes and diabetic ketoacidosis (DKA). About 10% of people with diabetes have Type 1, and the remaining 90% of people with diabetes have Type 2. The Type 1 diabetes results from failure of the pancreas to produce enough insulin due to loss of beta cells. It appears as childhood or adolescence; it can also develop in adults. In diabetes Type 1, destruction of beta cells usually starts early in life, but sometimes up to 90% of the beta cells are destroyed before symptoms crop up. Four clinical
  • 2. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2303 symptoms of uncontrolled diabetes, that all sound similar, are polyphagia (poly - a lot; phaga - eating), glycosuria (glucose - glucose; uria – urine), polyuria (poly – a lot; uria – urine), and polydipsia (poly – a lot; dipsia – thirst). Type 2 diabetes begins with insulin resistance, a condition in which cells fail to respond to insulin properly. It appears commonly among adults. Gestational diabetes occurs when pregnant women without a previous history of diabetes develop high blood sugar levels. It returns to normal soon after delivery. One really serious complication with diabetes type 1 is called diabetic ketoacidosis (DKA). Diabetic ketoacidosis (DKA) is a life-threatening problem of diabetes mellitus. Hence it is essential to detect the diabetes level and to undertake the necessary treatment in order to maintain it. Otherwise, it causes serious impact on patient’s health and finally it leads to death. DKA impact from a shortage of insulin; in response, the body switches to burn fatty acids, which produces ketone bodies in acidic nature. That ketone bodies break down into acetone and escape as a gas while exhaled from the lungs, which gives a sweet fruity smell to a person’s breath. DKA is diagnosed when testing finds low blood pH, high blood sugar, and ketoacids in either the urine or blood. In these type of diabetes produces ketone bodies in blood urine and breath (acetone – which produces fruity smell). Currently, diabetes mellitus is diagnosed by both invasive (finger pricking -blood test) and non-invasive (urine test) methods. Here this is one of the non-invasive methods of diagnosing using breath. Human breath analysis is one of the earlier methods to diagnose disease. There are about over 3500 Volatile Organic Compounds present in our exhaled breath, one of such VOC is acetone. The VOC’s contained in exhaled human breath are detected with the help of the sensors. It is found that acetone acts as the biomarker for diabetes mellitus. By detecting and mapping the concentration of acetone present in the exhaled human breath we can observe the prior stages of the disease. This can be done with the help of breath analysis technique. By mapping the concentration of acetone and diagnosing with the standard values we can ensure that the patient is affected with diabetes or not. Detection of acetone can be done with the help of GROVE – HCHO VOC gas sensor. The normal acetone level in exhaled breath of healthy person is less than (0.1 PPM – 0.8 PPM). The patients who are affected from diabetes will have acetone concentration more than (1.8 PPM – 5.0 PPM). The detected concentration is displayed in the display so that they know their state of health. Through this method, we can ensure a person’s safety by continuously monitoring the concentration of ammonia in exhaled breath at a particular time interval. It is one of the cost-efficient ways to detect the disease such that the poor people can afford it easily. One of the main advantages of this method that it does not involve risk factors like finger pricking and pain stacking methods involved in laboratory methods. 2. LITERATURE REVIEW Fahad Usman., et al. [1] identified acetone as a non-invasive biomarker for diagnosing diabetes from exhaled human breath. They studied and discussed about diabetes, death rate, biomarker and biosensors, and concentration of acetone for normal (0.1 ppm - 0.8 ppm) and diabetes (1.8 ppm - 5.0 ppm) persons. They gathered standard concentration from WHO. Claire Turner., [2] investigates the potential for monitoring compounds in breath and emitted through skin for inferring blood glucose concentration. Potential markers and an assessment of their suitability for non-invasive monitoring are discussed. The varying technologies developed for monitoring VOCs (volatile organic compounds) in breath and from skin of diabetics using acetone as a biomarker. They discussed about breath analysis and concentration of acetone in diabetics and non-diabetics person. Valentine Saasa., et al. [3] describe the methods used in the field of breath analysis to monitor and diagnose diabetes mellitus. Currently the diagnosis and monitoring of blood glucose and ketone bodies that are used in clinical studies involve the use of blood tests. This method entails pricking fingers for a drop of blood and placing a drop on a sensitive area of a strip which is pre-inserted into an electronic reading instrument. Furthermore, it is painful, invasive and expensive, and can be unsafe if proper handling is not undertaken. Human breath analysis provides a non-invasive and rapid method for detecting various VOCs (volatile organic compounds) that are indicators for different diseases. In patients with diabetes mellitus, the body produces excess amounts of ketones such as beta- hydroxybutyrate, acetoacetate, and acetone. Acetone is exhaled during respiration.
  • 3. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2304 Dixit K., et al. [4] research on non-invasive blood glucose monitoring has led to the development of various devices, easing diabetes management through comfortable, minimally invasive/non-invasive continuous blood-glucose monitoring. Nevertheless, they still suffer from issues such as lag time and the requirement of frequent calibration. Breath analysis is newly budding domain for diagnosing and monitoring the disease using non-invasive method. It is painless, safe, and allows repetitive sampling. However, the intricacies of breath analysis need to be well-studied to enable accurate, reliable, and reproducible monitoring. Righettoni M., et al. [5] proposed about Si-doped WO3 chemo-resistive gas sensors have been fabricated by flame spray pyrolysis. These nanoparticles had excellent acetone sensing properties with a great potential for application in non-invasive medical diagnostic by breath analysis. In fact, theWO3 ε-phase thermal stability was enhanced greatly by Si-doping (up to 500 °C) allowing detection in the ppb range of acetone in ideal (dry air) and realistic (90%) breath conditions. However, gas sensing performances such as the sensor response and the response time are also dependent on the measurement system and conditions. Maureen I. Harris., et al. [6] proposed that most of the US people do not check their blood glucose level because of cost and income. So, they proposed about the importance of blood glucose control in the prevention of diabetes complications and the role of self-monitoring in achieving blood glucose control, it may be prudent for physicians and their patients to make greater use of this technique. Chen, X., et al. [7] proposed an innovative method of lung cancer pathologic analysis and early diagnosis at the cellular level was introduced. An optimum HS-SPME-GC method was used to detect the odors of some cells in human breath, including squamous cell carcinoma cells, adenocarcinoma cells, bronchioloalveolar carcinoma cells, non-small cell carcinoma cells, bronchial epithelial cells, tastebud cells, osteogenic cells, and lipocytes, and we were able to obtain an odor chromatogram of these cells. 3. METHODOLOGY In this chapter we are going to discuss about the proposed methodology or working of our project. In this project we proposed Human Breath Analysis technique to identify the diabetes mellitus in a human. It is a technique in which various diseases in humans can be identified easily through exhaled human breath. In this technique, human breath is analyzed for the normal concentration of ammonia in order to identify the liver cirrhosis. One of the biggest advantages of this technique is the early detection of diseases in human beings in a non- invasive way. Figure 2.1 HUMAN BREATH ANALYSIS TECHNIQUE It is done with the help of gas sensing technology. For sensing the acetone gas presented in our human breath as one of the 3500 VOCs, we have used a type of acetone gas sensor called Grove HCHO VOC gas sensor. It is a general- purpose sensor that it can be used for various purposes. The sensor is powered with the microcontroller which has a power source pin of 3.3V. The microcontroller used here is PIC16F877A which is the brain of this entire system. The sensor will sense the input acetone gas as soon as it is powered. At the same time, the microcontroller was programmed in such a way to read the input signal. Here the analog output pin of the sensor is used to sense the gas since it gives the raw values continuously. The analog output from the sensor is not in a readable form. Hence it is
  • 4. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2305 converted into readable digital form with the help of 10-bit Analog to Digital Converter which is already present in the microcontroller. The output which is sensed by the sensor is analyzed and compared with the normal standard concentration of the acetone gas which is also programmed in the microcontroller. The normal standard concentration of acetone in exhaled human breath is found out by performing the literature survey. The corresponding output will be displayed in the displaying unit. Hence the person can be verified with their corresponding present health status. If the output value is confirmed with diabetes mellitus, person have to get medication at right time. If the output is normal, the entire process can be repeated in case of any critical situation occurs in future. It will greatly pave the way for detecting diseases using this breath analysis technique in these modern days as it possesses many advantages over other invasive laboratory methods. The test can also be done in a short period of time. Figure 2.2 WORK FLOW PROCESS 2.1 SENSOR CALIBRATION The calibration of sensor is very important in doing any project work. It is very important before using any sensors in project. It is a process of rectifying the undesired outputs to desired outputs. Sensors are made up of electronic and electrical components. Since those components are sensitive to various factors like temperature, humidity, external shocks and vibration etc., it will lead to the undesired output results. The sudden change in the working environment also leads to the deviation of output results such that the measured outputs differ from the expected outputs. Hence calibration plays a significant role in increases the chance of getting desired output results. It will improve the performance and accuracy of the sensor by removing errors in measurements. In our project, we use GROVE HCHO – VOC gas sensor for detecting acetone gas in human breath. Hence it is necessary to calibrate the sensor for accurate output. The operation of this sensor is based on the changing resistance of
  • 5. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2306 the heated coil thus measuring the concentration of the particular gas. Two parameters are involved while calibrating the sensor. They are as follows:  R0 – Sensor resistance of instantaneous concentration of measuring gas  Rs – Internal resistance of the sensor changing in clean air or environment The concentration measurement of a particular gas is also depending on the ratio of Rs and R0. Hence it is found out as a primary stage of calibrating the sensor. The sensor will show some random values as output when it is connected to microcontroller. It is necessary to calibrate it to zero. For that purpose, we are going to take average of that random values as a secondary step of calibration. Hence, we coded the microcontroller in such a way to calculate the average value using looping controls. The loop will run for 500 times and the average of the values was taken. The calculated average was converted into equivalent sensor voltage since this conversion calibrate the sensor to zero in an effective manner. The Rs value of the sensor is also calculated every time by converting the values into equivalent sensor voltage. The obtained R0 value will be used to find the ratio for changing Rs value at every step. Figure 2.3 SENSOR OUTPUT BEFORE AND AFTER CALIBRATION 4. MODELING AND ANALYSIS In this project we are using a GROVE HCHO VOC gas sensor to analyze the concentration of acetone in human breath. Hence this experiment plays a very important role to analyze the characteristics of the sensor and to determine the applicability of this particular type sensor in real time monitoring. It is conducted to verify that the above used sensor meets the above characteristics for the efficient measurement of acetone in human breath in real time application. 3.1 ANALYSIS I The analysis is carried out with the detection of acetone gas which is to be measured in human breath. But human breath has minute level of acetone and it is volatile. So, acetone solution from laboratory is taken as sample for this analysis. The general calibration of the sensor is carried out and made the sensor ready to detect the acetone. To generate the acetone odor or gas, we have used acetone solution in same concentrations and in different environment. The acetone solution (10ml) was taken in a beaker to conduct the experiment. Before commencing of the experiment, the sensor was preheated by supplying the power for about thirty minutes for obtaining the best results. This analysis was carried out in an open environment such that the gas sensor is placed near to the beaker containing acetone solution. Then the sensor detected the acetone gas which is evolved from the solution. The data obtained from the sensor is viewed in the monitor and it is recorded for further analysis. The data was collected for every second and the collected data was tabulated for further analysis.
  • 6. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2307 TABLE 3.1 READINGS OF ACETONE DATA 1 TIME (IN S) CONCENTRATION OF ACETONE (IN PPM) 1 0.30 5 0.39 10 0.60 15 0.83 20 1.11 25 1.43 30 1.80 35 2.21 40 2.48 45 2.62 55 2.82 60 2.92 The data which is tabulated are used to plot the graph against time in seconds for analyzing the response of sensor in detecting the concentration of acetone gas which is released from the acetone solution. From the graph, we inferred the characteristics of sensor in determining the acetone concentration in solution clearly. Figure 3.1 GRAPHICAL REPRESENTATION OF ACETONE DATA 1 3.2 ANALYSIS II Another experiment analysis using the different concentration of acetone solution (10ml) was carried out to analyze the performance of sensor in detecting the gas. In this analysis, the experiment was done in a closed environment. Then the sensor detected the acetone gas which is evolved from the solution. The data obtained from the sensor is viewed in the monitor and it is recorded for further analysis. The data was collected for every second and the collected data was tabulated for further analysis.
  • 7. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2308 Table 3.2 READINGS OF ACETONE DATA 2 TIME (IN S) CONCENTRATION OF ACETONE (IN PPM) 1 0.40 5 0.53 10 0.70 15 0.95 20 1.20 25 1.45 30 1.70 35 1.95 40 2.20 45 2.45 50 2.70 55 2.95 60 3.30 The data which is tabulated are used to plot the graph against time in seconds for analyzing the response of sensor in detecting the concentration of acetone gas which is released from the acetone solution. From the graph, we inferred the characteristics of sensor in determining the acetone concentration in solution clearly. Figure 3.2 GRAPHICAL REPRESENTATION OF ACETONE DATA 2 5. RESULTS AND DISCUSSION Experiment analysis was carried out regarding the performance characteristics of the sensor and its behavior. The analysis was done in different concentrations of acetone solution. In this chapter we will discuss about the results obtained from each concentration. In experiment analysis I, acetone solution (10ml) is used for generating acetone odor and the gas sensor is made to detect the gas in open environment. The data was collected and it is observed that the response of the sensor towards acetone gas is very high. The results obtained from the sensor was also identified as very high in detecting the acetone gas in open environment.
  • 8. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2309 In the experiment analysis II, the same concentration of acetone solution (10 ml) is used and the data was collected from the sensor in closed environment. It is also observed from the results that the sensor is very sensitive in detecting the acetone gas. The output obtained from the sensor proved its high sensitivity again showing in its crystal- clear data and the same is verified in plotting the data in graph. Hence from both the analysis I and II, it is found that the sensor has high sensitivity towards acetone gas and proved that the usage of Grove HCHO VOC gas sensor in this project is correct. In this experiment, PIC16F877A and NodeMCU is used. PIC is user friendly and cost efficient. So, PIC is used for diagnose and analyze of acetone and the output is displayed in LCD (16X2). Here NodeMCU is used for the same purpose of analysis and the output is displayed in mobile application (Blynk application). Figure 7.1 LCD OUTPUT All the necessary characteristics of sensor was studied verified by conducting different experimental analysis. Hence the selection of Grove HCHO VOC gas sensor for this project is proved as a rightful choice to detect the abnormal detection of acetone which will be used in detecting diabetes mellitus. The sensor has a range of 10PPM – 1000PPM in detecting acetone but the concentration of acetone in breath actually falls under 0.8 – 5.0 PPM. Hence it requires a low range of sensors for detecting acetone in exhaled breath for that level. The sensor has to be specifically designed using MOS technologies to apply this project in real time. So, I have done this project as a prototype model suggesting that this can be done in future appropriately. 6. CONCLUSION Diagnosis of diabetes mellitus using human breath will play a major role in future. Research found that human breath analysis technique has a great potential to diagnose various other diseases like diabetes mellitus. The simple and cost- efficient means of diagnosing diabetes mellitus and is one of the future techniques. The time consumption of diagnosing the disease is also very less comparing to traditional laboratory diagnosing methods. An important highlight of this diagnosing technique is the non-invasive way such that it only uses the exhaled human breath for diagnosis. It does not involve any finger pricking method of diagnosis which leads to pain staking. It is very safe to undergo the diagnosis and it does not involve any hazardous procedures unlike in other X-rays or MRI scanning methods. This technique has many complexities like requirement of standard procedures, designing of appropriate gas sensors with low range and high accuracy but if it is are currently going for standards and procedures in incorporating this technique in today’s world. Further enhancements can be done in future like Figure 7.2 BLYNK APPLICATION OUTPUT.
  • 9. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 09 Issue: 05 | May 2022 www.irjet.net p-ISSN: 2395-0072 © 2022, IRJET | Impact Factor value: 7.529 | ISO 9001:2008 Certified Journal | Page 2310  Diagnosing through Image Processing Techniques  Perform Machine Learning (ML) or AI algorithms  Application of IoT techniques  Data Collection using Cloud Technologies It will be an additional advantage for this project if the above-mentioned technologies get added to this project. Then it will become as a very suitable method of diagnosing diseases in this modern world. It will also be very useful for the patients also for periodical analysis of their state of health. 7. REFERENCES [1] Usman, F., Dennis, J.O., Ahmed, A.Y., Meriaudeau, F., Ayodele, O.B. and Rabih, A.A., 2018. A review of biosensors for non-invasive diabetes monitoring and screening in human exhaled breath. Ieee Access, 7, pp.5963-5974. [2] Turner, C., 2011. Potential of breath and skin analysis for monitoring blood glucose concentration in diabetes. Expert review of molecular diagnostics, 11(5), pp.497-503. [3] Saasa, V., Malwela, T., Beukes, M., Mokgotho, M., Liu, C.P. and Mwakikunga, B., 2018. Sensing technologies for detection of acetone in human breath for diabetes diagnosis and monitoring. Diagnostics, 8(1), p.12. [4] Dixit, K., Fardindoost, S., Ravishankara, A., Tasnim, N. and Hoorfar, M., 2021. Exhaled Breath Analysis for Diabetes Diagnosis and Monitoring: Relevance, Challenges and Possibilities. Biosensors, 11(12), p.476. [5] Righettoni, M. and Tricoli, A., 2011. Toward portable breath acetone analysis for diabetes detection. Journal of breath research, 5(3), p.037109. [6] Harris, M.I., Cowie, C.C. and Howie, L.J., 1993. Self-monitoring of blood glucose by adults with diabetes in the United States population. Diabetes care, 16(8), pp.1116-1123. [7] Chen, X., Xu, F., Wang, Y., Pan, Y., Lu, D., Wang, P., Ying, K., Chen, E. and Zhang, W., 2007. A study of the volatile organic compounds exhaled by lung cancer cells in vitro for breath diagnosis. Cancer: Interdisciplinary International Journal of the American Cancer Society, 110(4), pp.835-844.