International Journal of Technical Research and Applications e-ISSN: 2320-8163,
www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287
1 | P a g e
STUDY & PERFORMANCE OF METAL ON
METAL HIP IMPLANTS: A REVIEW
Sumita Chaturvedi1
, Prem Kumar Bharti2
,
Mechanical Engineering Department
Integral University, Dasauli, Kursi Road, Lucknow (U.P),
Lucknow, India
sumita@iul.ac.in
Abstract— Recent joint surgery studies reveal increased
revisions and resurfacing of the metal on metal hip joints. Metal
on metal hip implants were developed more than thirty years ago
and their application has been refined because of availability of
advanced manufacturing techniques and partly by advancements
in material science and engineering. Development of composite
materials may provide greater durability to metal-on-metal hip
implants .This review article is a study of the latest literature of
metal-on-metal hip implants and its various modeling techniques.
Numbers of methods are used for convergence and numerical
solution to investigate the performance of metal-on-metal hip
implant for accurate stable solution. This paper presents analysis
done by various researchers on metal-on-metal hip implants for
wear, lubrication, fatigue, bio-tribo-corrosion, design, toxicity
and resurfacing. After in vivo and in vitro studies, it is found that
all these methods have limitations. There is a need of more
insight for lubrication analysis, geometry of bearings, materials
and input parameters. The information provided in this work is
intended as an aid in the assessment of metal-on-metal hip joints.
Index Terms— Finite element method, Hip joint, MOM hip
implants, Numerical Methods.
I. INTRODUCTION
There are several joints in human body, each have specific
function and motion. Major joints are knee joint, shoulderjoint
& most important hip joint. Hip joint plays an important role to
balance the whole body weight in respect of the requirement of
motion and work. In tribological study, continuous contribution
of researchers improves the performance of MOM hip joint
over a variety of hip implants. After having hip
implant,physically patients may or may not be comfortable
because of various complications that arises due to wear.
This joint is ball-in-socket joint Literature review shows
number of research works on hip joint replacement withvarious
methods. Brown C et al. [1] concluded that in current, there is
no ideal method for hip replacement in view of toxicity,
adverse biological effects andtribological properties depends
upon wear rate. American Academy of Orthopedic Surgeons
advised for Total Hip Replacement (THR) when this joint
damaged due to a fracture ,arthritis like Osteoarthritis ,
Rheumatoid arthritis, Post Traumatic arthritis and avascular
necrosis or any childhood problem. A typical metal-on-metal
hip joint is shown inFigure1. [2]
Figure 1.Typical metal-on-metal hip joint [2]
There are various types of hip implant joint according to
their material and use i.e., ceramic-on-metal, metal-on-metal,
metal-on-polyethylene, ceramic-on-ceramic, ceramic-on-
polyethylene etc. In 2006 D Dowson and Z-M Jin [3] explained
the four type of lubrication pattern in Metal-on-Metal hip joint
tribology like hydrodynamic fluid film lubrication, mixed
lubrication, boundary lubrication,and dry lubrication. They
concluded that the ceramic on polyethylene joint, metal on
metal joints (mild mixed lubrication regime, Co-Cr-Mo alloy
high carbon>0.20%C)are significantly recommended as
compare to metal and ceramic for current hip replacement due
to high strength and low wear rates.
In old practiceship implant, the cobalt (Co), Chromium
(Cr), Molybdenum (Mo) was adopted by DobbsH. S. and M.J.
Minski[4]. They observed that the metal ions are fifty times
stronger than their standard value in urine and conclude that the
metal-on-plastic joint are more relevant. In recent again focus
on the use of metal-on-metal hip implant instead of others
because of its strong capacity for wear and indicated that
second generation of metal-on-metal hip implant are more
efficient for strength to young patients [5][6].
Feng Liu et al. [8] reported numerically work on bearing
geometry and structure support of metal-on-metal hip
implantsin terms of a little clearance for 28 mm head diameter
and analyzed fluid film lubrication in-vivo and in-vitro
conditions. In their study, they adopted spherical mesh through
International Journal of Technical Research and Applications e-ISSN: 2320-8163,
www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287
2 | P a g e
finite element and calculate the minimum film thickness by
integrating the pressure with the help of fast Fourier
transform.Bharma M. S. and C. P. Case[9]mentioned in their
study that the metal-on-metal hip arthroplasty gives more better
result as compared to metal-on-polyethylene
articulation.JagatiaM. and JinZ. M. introduced Reynolds
equation in spherical coordinate and it was solved by Newton-
Raphson Finite Difference Method.This numerical method
applied for a relatively thick actabular cup shows little effect
on contact pressure [10].Leiming G et al. [11] investigated the
effect of loading and motion on lubrication of metal-on-metal
hip replacements. In their analysis, they adopted multigrid
method for three level grid methods for 257 X 257 maximum
nodes.
II. LUBRICATION MODEL & MATERIALS USED
Boundary lubrication is adopted for the study of various
tribological surfaces to enhance the performance, and life of
hip implant and protect their surfaces from wear & friction,
which are in mutual contact. Elastohydrodynamic films are
then adopted for better performance of joint for metal-on-metal
hip implant for this purpose there are two types of models were
generally adopted ball- in-socket andball- on- plane. Hip
implants made up by 316Lstainless steel, cobalt-chromium-
molybdenum and ceramic like alumina and zirconium, titanium
based alloy.D Dowson and Z-M Jin[3] reported that the
ceramic-on-polyethylene or metal-on-polyethylene have some
limitation for strengthand MOM (metal-on-metal) are much
durable when their design is subjected to proper surface
finishFor lubrication analysis, it is assumed that flow of
lubricant is in between two bearing surfaces and in this
Reynolds equation for spherical co-ordinate is solved by
various researchers to find out optimum result.
MengQ. E [12] adoptedtransient elastohydrodynamic
lubrication analysis for novel metal-on-metal of hip prosthesis
with non-spherical femoral bearing surface. Synovial fluidis
used as lubricant in artificial hip joint which a Non-Newtonian
fluid and have a low shear stress and solved Reynolds Equation
by numerical method.
Equation 1
     
       
       
 
 
t
h
R
h
hh
hh
R
p
h
p
h
z
y
x





























































































22
1
2
1
33
sin12
sin
cossinsinsin
coscossinsin
sin6sinsin
Performance & lubrication analysis of metal-on-metal hip
implant investigated by using above Reynolds’s equation. [11-
13] Where, and are spherical coordinates andWx ,Wy and
Wz are the angular velocities presented in figure 2 forflexation-
extension (FE), internal external rotation (IER) and
abduction-adduction(AA) motions [11].These are the types of
relative motion in between femoral head and actabularcup,
under load for demonstration.Most of time this partial
differential equation is widely used for lubrication analysis,
texturing of hip implant for spherical coordinates.
Figure 2.Metal-on-metal hip implant under loading [11]
International Journal of Technical Research and Applications e-ISSN: 2320-8163,
www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287
3 | P a g e
Researchers in this field are applying this equation in
numerical methods. The governing Reynolds equation (2) was
written for all three dimensional motion. [14] [15].
Equation 2
     
       
       
 
 
t
h
R
h
hh
hh
R
p
h
p
h
z
y
x





























































































22
1
2
1
33
sin12
sin
cossinsinsin
coscossinsin
sin6sinsin
III. DISCUSSION
In general there are various numerical methods Multigrid,
fast- Fouriertransform, FDM (Finite difference Method),
Newton-Raphson method were adopted for calculating
convergence and optimum solution for accuracy. According
toGao L. M.and MengQ. E.[13] the four level grids multi
integration method has more effective as compared to fast
Fourier technique.Multigrid method was also adopted by
various researcherswhich is much more fast and
effective.Where FDM method is limited for two elements like
square and rectangle but, there is alsoananother method for
analysis which is now in trend called FEM (Finite Element
Method). FFT (Fast Fourier transform technique with FEM
measures elastic deformation for complicated structure in
artificial hip joint [18].As compared to above methods, the
finite element method will be suggested for lubrication analysis
of metal-on-metal hip implant for finding more realistic results
with fast convergence and calculation. In hip implant expected
time period of stay is in between 10-15 years [16][17].There
are various vivo –vitro studies performed outside and inside
living being .Wear effect on MOM hip implant under vivo
condition and tested in hip simulator , linear wear depthwas
found for geometry of bearing surface which was investigated
for running case[19]. By Clinically metallic wear particles
generated but their volume is less as compared to polyethylene
[20-23] [29].Hip resurfacing and surface texturing improve the
quality of MOM hip implant offer low wear also improve the
quality [19-23.]For analysis following input parameters are put
in under consideration given in table1 [10].
Table 1. Input parameters
S.No. Parameters Value
1. Radial clearance 24-30 µm
2. Cup wall thickness 9.5mm
3. Equivalent support
thickness
2mm
4. Cup inside radius 12-14mm
5. Elastic modulus of metal 210 GPa
The research on two generation of metal-on-metal hip
resurfacing with different cup wall thicknesses and clearances
were investigated and confirmed the importance of the
geometry and input parameters on the performance of MOM
hip implant [24]. Surface texturing found to be more effective
in fluid film lubrication by micro geometrical design on hip
bearing surface [30] [31]
IV. CONCLUSION
Based on the above literaturereview, it is found that there is
a need to develop a new way to compute more realistic results
with appropriate method for analysis of MOM hip implants.
Hip implants depends not only on geometry but also on proper
methods and materials that are adopted for lubrication analysis
, that depends on the film thickness, head and the pressure
distribution around the surface of contact. An advance material
combination with advance simulation methods in MOM hip
implant again increases the interest to adopt this MOM hip
implant in hip replacement joint. So in the field of tribological
study of MOM hip implants more and more realistic studies are
required, to get minimum revised cases because hip
replacement procedure is more severe than previous hip
prosthesis.
V. NOTATIONS
ø Spherical co-
ordinate
p Fluid
pressure
µ Fluid viscosity ρ Density of
fluid
h Fluid film
thickness
ω Angular
velocity
R1 clearance θ Angle of
contact
International Journal of Technical Research and Applications e-ISSN: 2320-8163,
www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287
4 | P a g e
REFERENCES
[1] BrownC,Fisher J and Ingham E.2005 Bilogical effects of
clinically relevant wear particles from metal-on-metal hip
prostheses. University of medical and biological
engineering,University of leeds,UK
,DOI:10.1234/095441105X63291.
[2] https://orthoinfo.aaos.org/en/treatment/total-hip-replacement/
[3] DowsonD and Jin Z. M. 2006Metal-on-metal hip joint
Tribology. Proc. IMechE,Part H:J.Engineering in
Medicine,Vol.220.
[4] DobbsH.S and MinskiM.J 1980Metal ion release after total hip
replacements.Biomaterials. Vol. 1.
[5] H.LucasAnissian,AndreStark,Vivtoria Good, Henrik Dahlstrand,
Ian C. Clarke 2001The Wear Pattern in Metal-on-Metal Hip
Prostheses. John Wiley & Sons, Inc. Biomed Mater Res
(ApplBiomater, 58:673-678.
[6] Oparaugo P, Clarke I ,Malchau H, Herberts P. 2004The
correlation of wear induced osteolysis and revision with
volumetric wear rates.ActaOrthopScand,72:22.
[7] Wang,F.C. ,Jin,Z.M.. 2004 Prediction of elastic deformation of
acetabular cups and femoral heads for lubrication analysis of
artificial hip joints.Proceedings of the Institution of Mechanical
Engineers,Part J ; Journal of Engineering Tribology , 2004, 218
(3),201-208.
[8] Feng Liu, Zhongming. Jin, Paul Robrts, Peter
Grigoris.2007Effect of bearing geometry and structural supports
on transient elastohydrodynamic lubrication of metal-on-metal
hip implants.Journal of Biomechanics, 40; 1340-1349.
[9] M S Bharma and C P Case.2006Biological effects of metal-on-
metal hip replacements. Proc. ImechE,Part H:J. Engineering in
medicine,Vol.220.
[10] M Jagatia and Z M Jin. 2001Elastohydrodynamic lubrication
analysis of metal-on-metal hip prostheses under steady state
entraining motion. Proc. ImechE,Part H:J. Engineering in
medicine , 2001,Vol. 215.
[11] LeimingGao, Fengcai Wang et al.2009Effect of 3D
Physiological loading and motion on elastohydrodynamic
lubrication of MOM total hip replacements.Medical Engineering
& Physics 31; 720-729
[12] QingenMeng,LeimingGao,FengLIU,PeiranYang,John Fisher,
ZhongminJin . 2009 Transient elastohydrodynamic lubrication
analysis of a novel metal-on-metal hip prosthesis with an
asphericalacetabular bearing surface.Journal of Biomechanics,
Vol .24.
[13] L M Gao, QE Meng, F C Wang, P R Yang and Z M Jin.
2007Comparison of numerical methods for elastohydrodynamic
methods for metal-on-metal hip implants:muti-grid verses
Newton–Raphson.Proc.IMechE, Part J: J.Engineering
Tribology, Vol. 221.
[14] GoenkaPK,Booker JF.1980 Spherical Bearings:Static and
dynamic analysis via the finite element method.ASME J
Lubr.Technol, 1980;102(7):308-19
[15] Wang FC,Jin ZM. 2008 Transient elastohydrodynamic
lubrication of hip implants.ASME J Tribol;130(0011007):1-13
[16] Amstutz,H.C.,Campell, P.Kossovsky,N.,and Clark,I.C.1991
Mechanism and clinical significance of weardebris-induced
osteolysis.Clin.Orthop.Relat.Res. ,276,7-17.
[17] Harris,W.H. 1993 The problem is
osteolysisClin.Orthop.Relat.Res., 311,46-53
[18] Z.M Jin.2006Theoretical studies of elastohydrodynamic
lubrication of artificial hip joint.Proc.IMechEPart J:J
Engineering Tribology, Vol.220.
[19] F Liu,ZM Jin, F Hirt, C Rieker,P Roberts and P
Grigoris.2005Effect of wear of bearing surfaces
onelastohydrodynamic lubrication of metal–on-metal hip
implants. Proc. IMechE ,Part H:J.Engineering in Medicine.
Vol.219
[20] A.G Cobb ,T P Schmalzreid. 2006The clinically significance of
metal ion release from cobalt-chromiummetal-on-metal hip joint
arthroplasty. Proc. IMechE Part H:J.EngineeringinMedicine,
Vol.220.
[21] Jacobs,J.J.,Silverton, C.,Hallab, N.J.,Skipor, A.K Pattrson, L.,
Black,J.,andGalante, J.O. 1999 Metal release and excretion from
cementless titanium alloy total knee
replacements.Clin.Orhop.,358,173-180.
[22] Rieker,C.B.,Schon,R.,andKottig, P. 2004 Development and
validation of a second generation metal-on-metal
bearing:laboratory studies and analysis of retrievals.
J.Arthroplasty,19(8),Suppl.3,5 -11.
[23] Mckellop,H.,Park,S.H.,Chiesa,R.,
Doorn,P.,Lu,B.,Normand,P.,Grigoris,P.,and
Amstutz,H.1996Invivo wear three types of metal-on-metal hip
prosthesis during two decades of
useClin.Ortho.,1996,329,Suppl.,S128-S40.
[24] Feng Liu ,Zhongmi ,Jin , P Roberts ,and P Grigoris.
2006Importance of head diameter ,clearance ,andcup wall
thickness in elastohydrodynamic lubrication analysis of metal-
on-metal hipresurfacingprostheses.ProcInstMechEng
H.220(6):695-704
[25] Grigoris, P., Roberts, P., et al. 2005The evolution of hip
resurfacing arthroplasty.Orthop.Clin.NorthAm.,2005,36(2),125-
134
[26] Dowson, D., Hardaker, C., et al. 2004 A hip joint simulator
study of the performance of metal-on-metaljoints.Part I the role
of materials.J.arthroplasty,19(8), Supplement 1, 118-123.
[27] Dowson, D. Hardaker,C., Flett,M.,andIsaac,G.H. 2004 A hip
simulator study of the performance ofmetal-on-metaljoints,Part
II: design .J.Arthroplasty,19(18),124-130.
[28] Rieker,C.B.,Schon,R.,Konard,R.,Liebentritt,G.,Gnepf,.,Shen,M.,
Roberts,andGrigoris,P. 2005 Influence of clearance on in-vitro
tribology of large diameter metal-on-metal articulations
pertraining to resurfacing hip implants.Orthop.,Clin. North
Am.,36(2),135-142.
[29] .Liu,F.,Wang,F.C.,et al. 2004 Steadystateelastohydrodynamic
lubrication analysis of a metal-on-metalcup with an ultrahigh
molecular weight polyethylene backing.Proc.IMechE,Part
H:JEngineering in Medicine, 218(4),261-270.
[30] Liu, F., Z.M., and Dowson,D. 2006 An investigation into the
effect of a sandwich acetabular cup uponlubricant film thickness
in metal-on-metal total hip replacements.Proc. IMechE, Part J: J
Engineering Tribology,220(3),151-158.
[31] L M Gao , Q E Meng Liu, J Fisher, and ZM Jin. 2010 The effect
of aspherical geometry and surfacetexturing on the
elastohydrodynamic lubrication of metal-on-metal hip prosthesis
underphysiological loading and motions.Proc.IMechVol.224
Part C: J. Mechanical EngineeringScience,203-210.
International Journal of Technical Research and Applications e-ISSN: 2320-8163,
www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287
5 | P a g e

More Related Content

What's hot

Failure analysis of pin loaded glass epoxy polystyrene composite plates
Failure analysis of pin loaded glass epoxy polystyrene composite platesFailure analysis of pin loaded glass epoxy polystyrene composite plates
Failure analysis of pin loaded glass epoxy polystyrene composite platesiaemedu
 
Finite Element Analysis of honeycomb using Abaqus
Finite Element Analysis of honeycomb using AbaqusFinite Element Analysis of honeycomb using Abaqus
Finite Element Analysis of honeycomb using AbaqusUdayan Ghosh
 
Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...IAEME Publication
 
Modal Analysis of Fibre Reinforced Composite Beams with a Transverse Crack U...
Modal Analysis of Fibre Reinforced Composite Beams with a  Transverse Crack U...Modal Analysis of Fibre Reinforced Composite Beams with a  Transverse Crack U...
Modal Analysis of Fibre Reinforced Composite Beams with a Transverse Crack U...IJMER
 
Design and Analysis of Total Elbow Replacement Joint
Design and Analysis of Total Elbow Replacement JointDesign and Analysis of Total Elbow Replacement Joint
Design and Analysis of Total Elbow Replacement JointIJERA Editor
 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)IJERD Editor
 
New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...eSAT Publishing House
 
Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...
Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...
Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...Aaron Kadima
 
Dynamic Analysis of Double-Skin Composite Steel Plates
Dynamic Analysis of Double-Skin Composite Steel PlatesDynamic Analysis of Double-Skin Composite Steel Plates
Dynamic Analysis of Double-Skin Composite Steel PlatesIOSR Journals
 
Behavior of thin-walled tubes with combined cross-sectional geometries under ...
Behavior of thin-walled tubes with combined cross-sectional geometries under ...Behavior of thin-walled tubes with combined cross-sectional geometries under ...
Behavior of thin-walled tubes with combined cross-sectional geometries under ...IJAEMSJORNAL
 
Design and fabrication of hybrid composite springs
Design and fabrication of hybrid composite springsDesign and fabrication of hybrid composite springs
Design and fabrication of hybrid composite springsIAEME Publication
 
IRJET- Review of Cold-Formed Built-Up Compression Members
IRJET-  	  Review of Cold-Formed Built-Up Compression MembersIRJET-  	  Review of Cold-Formed Built-Up Compression Members
IRJET- Review of Cold-Formed Built-Up Compression MembersIRJET Journal
 
Sediment; Amphipod; Hyale nigra; Spiked sediment
Sediment; Amphipod; Hyale nigra; Spiked sedimentSediment; Amphipod; Hyale nigra; Spiked sediment
Sediment; Amphipod; Hyale nigra; Spiked sedimentpeertechzpublication
 

What's hot (18)

Failure analysis of pin loaded glass epoxy polystyrene composite plates
Failure analysis of pin loaded glass epoxy polystyrene composite platesFailure analysis of pin loaded glass epoxy polystyrene composite plates
Failure analysis of pin loaded glass epoxy polystyrene composite plates
 
Finite Element Analysis of honeycomb using Abaqus
Finite Element Analysis of honeycomb using AbaqusFinite Element Analysis of honeycomb using Abaqus
Finite Element Analysis of honeycomb using Abaqus
 
Journal Publication
Journal PublicationJournal Publication
Journal Publication
 
Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...Upgradation of non ductile reinforced concrete beamcolumn connections using f...
Upgradation of non ductile reinforced concrete beamcolumn connections using f...
 
Modal Analysis of Fibre Reinforced Composite Beams with a Transverse Crack U...
Modal Analysis of Fibre Reinforced Composite Beams with a  Transverse Crack U...Modal Analysis of Fibre Reinforced Composite Beams with a  Transverse Crack U...
Modal Analysis of Fibre Reinforced Composite Beams with a Transverse Crack U...
 
Design and Analysis of Total Elbow Replacement Joint
Design and Analysis of Total Elbow Replacement JointDesign and Analysis of Total Elbow Replacement Joint
Design and Analysis of Total Elbow Replacement Joint
 
International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)International Journal of Engineering Research and Development (IJERD)
International Journal of Engineering Research and Development (IJERD)
 
4
44
4
 
New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...New study on friction in a mom total hip prosthesis with balls in self direct...
New study on friction in a mom total hip prosthesis with balls in self direct...
 
Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...
Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...
Pull-out tests of ribbed steel reinforcing bars embedded in concrete with ste...
 
Dynamic Analysis of Double-Skin Composite Steel Plates
Dynamic Analysis of Double-Skin Composite Steel PlatesDynamic Analysis of Double-Skin Composite Steel Plates
Dynamic Analysis of Double-Skin Composite Steel Plates
 
Behavior of thin-walled tubes with combined cross-sectional geometries under ...
Behavior of thin-walled tubes with combined cross-sectional geometries under ...Behavior of thin-walled tubes with combined cross-sectional geometries under ...
Behavior of thin-walled tubes with combined cross-sectional geometries under ...
 
Ijetr042313
Ijetr042313Ijetr042313
Ijetr042313
 
115 ieee surya pratap (1)
115 ieee surya pratap (1)115 ieee surya pratap (1)
115 ieee surya pratap (1)
 
Design and fabrication of hybrid composite springs
Design and fabrication of hybrid composite springsDesign and fabrication of hybrid composite springs
Design and fabrication of hybrid composite springs
 
IRJET- Review of Cold-Formed Built-Up Compression Members
IRJET-  	  Review of Cold-Formed Built-Up Compression MembersIRJET-  	  Review of Cold-Formed Built-Up Compression Members
IRJET- Review of Cold-Formed Built-Up Compression Members
 
Cfd analysis of scaffold
Cfd analysis of scaffoldCfd analysis of scaffold
Cfd analysis of scaffold
 
Sediment; Amphipod; Hyale nigra; Spiked sediment
Sediment; Amphipod; Hyale nigra; Spiked sedimentSediment; Amphipod; Hyale nigra; Spiked sediment
Sediment; Amphipod; Hyale nigra; Spiked sediment
 

Similar to STUDY & PERFORMANCE OF METAL ON METAL HIP IMPLANTS: A REVIEW

A Review on Finite Element Analysis of Beam to Column Endplate Bolted Connection
A Review on Finite Element Analysis of Beam to Column Endplate Bolted ConnectionA Review on Finite Element Analysis of Beam to Column Endplate Bolted Connection
A Review on Finite Element Analysis of Beam to Column Endplate Bolted ConnectionIOSR Journals
 
Design a Corrugated Tube for Energy Absorption
Design a Corrugated Tube for Energy AbsorptionDesign a Corrugated Tube for Energy Absorption
Design a Corrugated Tube for Energy AbsorptionIRJET Journal
 
REVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAM
REVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAMREVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAM
REVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAMIRJET Journal
 
IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...
IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...
IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...IRJET Journal
 
Experimental and numerical analysis of expanded pipe using rigid conical shap...
Experimental and numerical analysis of expanded pipe using rigid conical shap...Experimental and numerical analysis of expanded pipe using rigid conical shap...
Experimental and numerical analysis of expanded pipe using rigid conical shap...ahmed Ibrahim
 
Thin-Walled Structures 45 p283-300
Thin-Walled Structures 45 p283-300Thin-Walled Structures 45 p283-300
Thin-Walled Structures 45 p283-300Haipeng Han
 
Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...
Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...
Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...► Victor Yepes
 
Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...
Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...
Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...IRJET Journal
 
Modelling and Analysis of Hybrid Composite Joint Using Fem in Ansys
Modelling and Analysis of Hybrid Composite Joint Using Fem in AnsysModelling and Analysis of Hybrid Composite Joint Using Fem in Ansys
Modelling and Analysis of Hybrid Composite Joint Using Fem in AnsysIOSR Journals
 
Utilization of steel in construction of high performance structures: A Review
Utilization of steel in construction of high performance structures: A ReviewUtilization of steel in construction of high performance structures: A Review
Utilization of steel in construction of high performance structures: A ReviewIRJET Journal
 
Flexural behavior of composite reinforced concrete t beams cast in steel cha...
Flexural behavior of composite reinforced  concrete t beams cast in steel cha...Flexural behavior of composite reinforced  concrete t beams cast in steel cha...
Flexural behavior of composite reinforced concrete t beams cast in steel cha...IAEME Publication
 
Constructive concepts for reducing wear caused by the
Constructive concepts for reducing wear caused by theConstructive concepts for reducing wear caused by the
Constructive concepts for reducing wear caused by theeSAT Publishing House
 
Acta Materialia 55 (2007) 4475–4487 Ab Initio Titan
Acta  Materialia 55 (2007) 4475–4487 Ab Initio  TitanActa  Materialia 55 (2007) 4475–4487 Ab Initio  Titan
Acta Materialia 55 (2007) 4475–4487 Ab Initio TitanDierk Raabe
 
IRJET- A Review - Optimization of Castellated Beam with Stiffener
IRJET-  	  A Review - Optimization of Castellated Beam with StiffenerIRJET-  	  A Review - Optimization of Castellated Beam with Stiffener
IRJET- A Review - Optimization of Castellated Beam with StiffenerIRJET Journal
 
IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...
IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...
IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...IRJET Journal
 
Constructive concepts for reducing wear caused by the friction torque of hip ...
Constructive concepts for reducing wear caused by the friction torque of hip ...Constructive concepts for reducing wear caused by the friction torque of hip ...
Constructive concepts for reducing wear caused by the friction torque of hip ...eSAT Journals
 
LITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESS
LITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESSLITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESS
LITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESSijiert bestjournal
 

Similar to STUDY & PERFORMANCE OF METAL ON METAL HIP IMPLANTS: A REVIEW (20)

A Review on Finite Element Analysis of Beam to Column Endplate Bolted Connection
A Review on Finite Element Analysis of Beam to Column Endplate Bolted ConnectionA Review on Finite Element Analysis of Beam to Column Endplate Bolted Connection
A Review on Finite Element Analysis of Beam to Column Endplate Bolted Connection
 
IJET-V2I6P4
IJET-V2I6P4IJET-V2I6P4
IJET-V2I6P4
 
Design a Corrugated Tube for Energy Absorption
Design a Corrugated Tube for Energy AbsorptionDesign a Corrugated Tube for Energy Absorption
Design a Corrugated Tube for Energy Absorption
 
REVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAM
REVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAMREVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAM
REVIEW STATIC AND DYNAMIC ANALYSIS OF A LAMINATED COMPOSITE BEAM
 
IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...
IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...
IRJET- Fea & Experimental Analysis of Three Point Bending Test of Thin Walled...
 
Experimental and numerical analysis of expanded pipe using rigid conical shap...
Experimental and numerical analysis of expanded pipe using rigid conical shap...Experimental and numerical analysis of expanded pipe using rigid conical shap...
Experimental and numerical analysis of expanded pipe using rigid conical shap...
 
Thin-Walled Structures 45 p283-300
Thin-Walled Structures 45 p283-300Thin-Walled Structures 45 p283-300
Thin-Walled Structures 45 p283-300
 
Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...
Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...
Buckling Analysis and Stability of Compressed Low-Carbon Steel Rods in the El...
 
F012213639
F012213639F012213639
F012213639
 
Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...
Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...
Synthesis Of Various Forms Of Carbon Nanotubes by ARC Discharge Methods – Com...
 
Modelling and Analysis of Hybrid Composite Joint Using Fem in Ansys
Modelling and Analysis of Hybrid Composite Joint Using Fem in AnsysModelling and Analysis of Hybrid Composite Joint Using Fem in Ansys
Modelling and Analysis of Hybrid Composite Joint Using Fem in Ansys
 
Utilization of steel in construction of high performance structures: A Review
Utilization of steel in construction of high performance structures: A ReviewUtilization of steel in construction of high performance structures: A Review
Utilization of steel in construction of high performance structures: A Review
 
Flexural behavior of composite reinforced concrete t beams cast in steel cha...
Flexural behavior of composite reinforced  concrete t beams cast in steel cha...Flexural behavior of composite reinforced  concrete t beams cast in steel cha...
Flexural behavior of composite reinforced concrete t beams cast in steel cha...
 
Constructive concepts for reducing wear caused by the
Constructive concepts for reducing wear caused by theConstructive concepts for reducing wear caused by the
Constructive concepts for reducing wear caused by the
 
Acta Materialia 55 (2007) 4475–4487 Ab Initio Titan
Acta  Materialia 55 (2007) 4475–4487 Ab Initio  TitanActa  Materialia 55 (2007) 4475–4487 Ab Initio  Titan
Acta Materialia 55 (2007) 4475–4487 Ab Initio Titan
 
IRJET- A Review - Optimization of Castellated Beam with Stiffener
IRJET-  	  A Review - Optimization of Castellated Beam with StiffenerIRJET-  	  A Review - Optimization of Castellated Beam with Stiffener
IRJET- A Review - Optimization of Castellated Beam with Stiffener
 
IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...
IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...
IRJET- A Numerical Study on the Effect of Length on Natural Frequency of Sing...
 
Constructive concepts for reducing wear caused by the friction torque of hip ...
Constructive concepts for reducing wear caused by the friction torque of hip ...Constructive concepts for reducing wear caused by the friction torque of hip ...
Constructive concepts for reducing wear caused by the friction torque of hip ...
 
Boneabsorption3
 Boneabsorption3 Boneabsorption3
Boneabsorption3
 
LITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESS
LITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESSLITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESS
LITERATURE REVIEW ON FRACTURE TOUGHNESS AND IMPACT TOUGHNESS
 

More from International Journal of Technical Research & Application

More from International Journal of Technical Research & Application (20)

EXPONENTIAL SMOOTHING OF POSTPONEMENT RATES IN OPERATION THEATRES OF ADVANCED...
EXPONENTIAL SMOOTHING OF POSTPONEMENT RATES IN OPERATION THEATRES OF ADVANCED...EXPONENTIAL SMOOTHING OF POSTPONEMENT RATES IN OPERATION THEATRES OF ADVANCED...
EXPONENTIAL SMOOTHING OF POSTPONEMENT RATES IN OPERATION THEATRES OF ADVANCED...
 
POSTPONEMENT OF SCHEDULED GENERAL SURGERIES IN A TERTIARY CARE HOSPITAL - A T...
POSTPONEMENT OF SCHEDULED GENERAL SURGERIES IN A TERTIARY CARE HOSPITAL - A T...POSTPONEMENT OF SCHEDULED GENERAL SURGERIES IN A TERTIARY CARE HOSPITAL - A T...
POSTPONEMENT OF SCHEDULED GENERAL SURGERIES IN A TERTIARY CARE HOSPITAL - A T...
 
STUDY OF NANO-SYSTEMS FOR COMPUTER SIMULATIONS
STUDY OF NANO-SYSTEMS FOR COMPUTER SIMULATIONSSTUDY OF NANO-SYSTEMS FOR COMPUTER SIMULATIONS
STUDY OF NANO-SYSTEMS FOR COMPUTER SIMULATIONS
 
ENERGY GAP INVESTIGATION AND CHARACTERIZATION OF KESTERITE CU2ZNSNS4 THIN FIL...
ENERGY GAP INVESTIGATION AND CHARACTERIZATION OF KESTERITE CU2ZNSNS4 THIN FIL...ENERGY GAP INVESTIGATION AND CHARACTERIZATION OF KESTERITE CU2ZNSNS4 THIN FIL...
ENERGY GAP INVESTIGATION AND CHARACTERIZATION OF KESTERITE CU2ZNSNS4 THIN FIL...
 
POD-PWM BASED CAPACITOR CLAMPED MULTILEVEL INVERTER
POD-PWM BASED CAPACITOR CLAMPED MULTILEVEL INVERTERPOD-PWM BASED CAPACITOR CLAMPED MULTILEVEL INVERTER
POD-PWM BASED CAPACITOR CLAMPED MULTILEVEL INVERTER
 
DIGITAL COMPRESSING OF A BPCM SIGNAL ACCORDING TO BARKER CODE USING FPGA
DIGITAL COMPRESSING OF A BPCM SIGNAL ACCORDING TO BARKER CODE USING FPGADIGITAL COMPRESSING OF A BPCM SIGNAL ACCORDING TO BARKER CODE USING FPGA
DIGITAL COMPRESSING OF A BPCM SIGNAL ACCORDING TO BARKER CODE USING FPGA
 
MODELLING THE IMPACT OF FLOODING USING GEOGRAPHIC INFORMATION SYSTEM AND REMO...
MODELLING THE IMPACT OF FLOODING USING GEOGRAPHIC INFORMATION SYSTEM AND REMO...MODELLING THE IMPACT OF FLOODING USING GEOGRAPHIC INFORMATION SYSTEM AND REMO...
MODELLING THE IMPACT OF FLOODING USING GEOGRAPHIC INFORMATION SYSTEM AND REMO...
 
AN EXPERIMENTAL STUDY ON SEPARATION OF WATER FROM THE ATMOSPHERIC AIR
AN EXPERIMENTAL STUDY ON SEPARATION OF WATER FROM THE ATMOSPHERIC AIRAN EXPERIMENTAL STUDY ON SEPARATION OF WATER FROM THE ATMOSPHERIC AIR
AN EXPERIMENTAL STUDY ON SEPARATION OF WATER FROM THE ATMOSPHERIC AIR
 
LI-ION BATTERY TESTING FROM MANUFACTURING TO OPERATION PROCESS
LI-ION BATTERY TESTING FROM MANUFACTURING TO OPERATION PROCESSLI-ION BATTERY TESTING FROM MANUFACTURING TO OPERATION PROCESS
LI-ION BATTERY TESTING FROM MANUFACTURING TO OPERATION PROCESS
 
QUALITATIVE RISK ASSESSMENT AND MITIGATION MEASURES FOR REAL ESTATE PROJECTS ...
QUALITATIVE RISK ASSESSMENT AND MITIGATION MEASURES FOR REAL ESTATE PROJECTS ...QUALITATIVE RISK ASSESSMENT AND MITIGATION MEASURES FOR REAL ESTATE PROJECTS ...
QUALITATIVE RISK ASSESSMENT AND MITIGATION MEASURES FOR REAL ESTATE PROJECTS ...
 
SCOPE OF REPLACING FINE AGGREGATE WITH COPPER SLAG IN CONCRETE- A REVIEW
SCOPE OF REPLACING FINE AGGREGATE WITH COPPER SLAG IN CONCRETE- A REVIEWSCOPE OF REPLACING FINE AGGREGATE WITH COPPER SLAG IN CONCRETE- A REVIEW
SCOPE OF REPLACING FINE AGGREGATE WITH COPPER SLAG IN CONCRETE- A REVIEW
 
IMPLEMENTATION OF METHODS FOR TRANSACTION IN SECURE ONLINE BANKING
IMPLEMENTATION OF METHODS FOR TRANSACTION IN SECURE ONLINE BANKINGIMPLEMENTATION OF METHODS FOR TRANSACTION IN SECURE ONLINE BANKING
IMPLEMENTATION OF METHODS FOR TRANSACTION IN SECURE ONLINE BANKING
 
EFFECT OF TRANS-SEPTAL SUTURE TECHNIQUE VERSUS NASAL PACKING AFTER SEPTOPLASTY
EFFECT OF TRANS-SEPTAL SUTURE TECHNIQUE VERSUS NASAL PACKING AFTER SEPTOPLASTYEFFECT OF TRANS-SEPTAL SUTURE TECHNIQUE VERSUS NASAL PACKING AFTER SEPTOPLASTY
EFFECT OF TRANS-SEPTAL SUTURE TECHNIQUE VERSUS NASAL PACKING AFTER SEPTOPLASTY
 
EVALUATION OF DRAINAGE WATER QUALITY FOR IRRIGATION BY INTEGRATION BETWEEN IR...
EVALUATION OF DRAINAGE WATER QUALITY FOR IRRIGATION BY INTEGRATION BETWEEN IR...EVALUATION OF DRAINAGE WATER QUALITY FOR IRRIGATION BY INTEGRATION BETWEEN IR...
EVALUATION OF DRAINAGE WATER QUALITY FOR IRRIGATION BY INTEGRATION BETWEEN IR...
 
THE CONSTRUCTION PROCEDURE AND ADVANTAGE OF THE RAIL CABLE-LIFTING CONSTRUCTI...
THE CONSTRUCTION PROCEDURE AND ADVANTAGE OF THE RAIL CABLE-LIFTING CONSTRUCTI...THE CONSTRUCTION PROCEDURE AND ADVANTAGE OF THE RAIL CABLE-LIFTING CONSTRUCTI...
THE CONSTRUCTION PROCEDURE AND ADVANTAGE OF THE RAIL CABLE-LIFTING CONSTRUCTI...
 
TIME EFFICIENT BAYLIS-HILLMAN REACTION ON STEROIDAL NUCLEUS OF WITHAFERIN-A T...
TIME EFFICIENT BAYLIS-HILLMAN REACTION ON STEROIDAL NUCLEUS OF WITHAFERIN-A T...TIME EFFICIENT BAYLIS-HILLMAN REACTION ON STEROIDAL NUCLEUS OF WITHAFERIN-A T...
TIME EFFICIENT BAYLIS-HILLMAN REACTION ON STEROIDAL NUCLEUS OF WITHAFERIN-A T...
 
A STUDY ON THE FRESH PROPERTIES OF SCC WITH FLY ASH
A STUDY ON THE FRESH PROPERTIES OF SCC WITH FLY ASHA STUDY ON THE FRESH PROPERTIES OF SCC WITH FLY ASH
A STUDY ON THE FRESH PROPERTIES OF SCC WITH FLY ASH
 
AN INSIDE LOOK IN THE ELECTRICAL STRUCTURE OF THE BATTERY MANAGEMENT SYSTEM T...
AN INSIDE LOOK IN THE ELECTRICAL STRUCTURE OF THE BATTERY MANAGEMENT SYSTEM T...AN INSIDE LOOK IN THE ELECTRICAL STRUCTURE OF THE BATTERY MANAGEMENT SYSTEM T...
AN INSIDE LOOK IN THE ELECTRICAL STRUCTURE OF THE BATTERY MANAGEMENT SYSTEM T...
 
OPEN LOOP ANALYSIS OF CASCADED HBRIDGE MULTILEVEL INVERTER USING PDPWM FOR PH...
OPEN LOOP ANALYSIS OF CASCADED HBRIDGE MULTILEVEL INVERTER USING PDPWM FOR PH...OPEN LOOP ANALYSIS OF CASCADED HBRIDGE MULTILEVEL INVERTER USING PDPWM FOR PH...
OPEN LOOP ANALYSIS OF CASCADED HBRIDGE MULTILEVEL INVERTER USING PDPWM FOR PH...
 
PHYSICO-CHEMICAL AND BACTERIOLOGICAL ASSESSMENT OF RIVER MUDZIRA WATER IN MUB...
PHYSICO-CHEMICAL AND BACTERIOLOGICAL ASSESSMENT OF RIVER MUDZIRA WATER IN MUB...PHYSICO-CHEMICAL AND BACTERIOLOGICAL ASSESSMENT OF RIVER MUDZIRA WATER IN MUB...
PHYSICO-CHEMICAL AND BACTERIOLOGICAL ASSESSMENT OF RIVER MUDZIRA WATER IN MUB...
 

Recently uploaded

Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startQuintin Balsdon
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.Kamal Acharya
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdfKamal Acharya
 
AIRCANVAS[1].pdf mini project for btech students
AIRCANVAS[1].pdf mini project for btech studentsAIRCANVAS[1].pdf mini project for btech students
AIRCANVAS[1].pdf mini project for btech studentsvanyagupta248
 
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdfAldoGarca30
 
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...ssuserdfc773
 
Introduction to Geographic Information Systems
Introduction to Geographic Information SystemsIntroduction to Geographic Information Systems
Introduction to Geographic Information SystemsAnge Felix NSANZIYERA
 
Post office management system project ..pdf
Post office management system project ..pdfPost office management system project ..pdf
Post office management system project ..pdfKamal Acharya
 
Introduction to Artificial Intelligence ( AI)
Introduction to Artificial Intelligence ( AI)Introduction to Artificial Intelligence ( AI)
Introduction to Artificial Intelligence ( AI)ChandrakantDivate1
 
Computer Networks Basics of Network Devices
Computer Networks  Basics of Network DevicesComputer Networks  Basics of Network Devices
Computer Networks Basics of Network DevicesChandrakantDivate1
 
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...ronahami
 
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxS1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxSCMS School of Architecture
 
Memory Interfacing of 8086 with DMA 8257
Memory Interfacing of 8086 with DMA 8257Memory Interfacing of 8086 with DMA 8257
Memory Interfacing of 8086 with DMA 8257subhasishdas79
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwaitjaanualu31
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaOmar Fathy
 
UNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxUNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxkalpana413121
 
8086 Microprocessor Architecture: 16-bit microprocessor
8086 Microprocessor Architecture: 16-bit microprocessor8086 Microprocessor Architecture: 16-bit microprocessor
8086 Microprocessor Architecture: 16-bit microprocessorAshwiniTodkar4
 
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...HenryBriggs2
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdfKamal Acharya
 

Recently uploaded (20)

Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdf
 
AIRCANVAS[1].pdf mini project for btech students
AIRCANVAS[1].pdf mini project for btech studentsAIRCANVAS[1].pdf mini project for btech students
AIRCANVAS[1].pdf mini project for btech students
 
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
 
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
 
Introduction to Geographic Information Systems
Introduction to Geographic Information SystemsIntroduction to Geographic Information Systems
Introduction to Geographic Information Systems
 
Post office management system project ..pdf
Post office management system project ..pdfPost office management system project ..pdf
Post office management system project ..pdf
 
Introduction to Artificial Intelligence ( AI)
Introduction to Artificial Intelligence ( AI)Introduction to Artificial Intelligence ( AI)
Introduction to Artificial Intelligence ( AI)
 
Computer Networks Basics of Network Devices
Computer Networks  Basics of Network DevicesComputer Networks  Basics of Network Devices
Computer Networks Basics of Network Devices
 
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
 
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
 
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxS1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
 
Memory Interfacing of 8086 with DMA 8257
Memory Interfacing of 8086 with DMA 8257Memory Interfacing of 8086 with DMA 8257
Memory Interfacing of 8086 with DMA 8257
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS Lambda
 
UNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxUNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptx
 
8086 Microprocessor Architecture: 16-bit microprocessor
8086 Microprocessor Architecture: 16-bit microprocessor8086 Microprocessor Architecture: 16-bit microprocessor
8086 Microprocessor Architecture: 16-bit microprocessor
 
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdf
 

STUDY & PERFORMANCE OF METAL ON METAL HIP IMPLANTS: A REVIEW

  • 1. International Journal of Technical Research and Applications e-ISSN: 2320-8163, www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287 1 | P a g e STUDY & PERFORMANCE OF METAL ON METAL HIP IMPLANTS: A REVIEW Sumita Chaturvedi1 , Prem Kumar Bharti2 , Mechanical Engineering Department Integral University, Dasauli, Kursi Road, Lucknow (U.P), Lucknow, India sumita@iul.ac.in Abstract— Recent joint surgery studies reveal increased revisions and resurfacing of the metal on metal hip joints. Metal on metal hip implants were developed more than thirty years ago and their application has been refined because of availability of advanced manufacturing techniques and partly by advancements in material science and engineering. Development of composite materials may provide greater durability to metal-on-metal hip implants .This review article is a study of the latest literature of metal-on-metal hip implants and its various modeling techniques. Numbers of methods are used for convergence and numerical solution to investigate the performance of metal-on-metal hip implant for accurate stable solution. This paper presents analysis done by various researchers on metal-on-metal hip implants for wear, lubrication, fatigue, bio-tribo-corrosion, design, toxicity and resurfacing. After in vivo and in vitro studies, it is found that all these methods have limitations. There is a need of more insight for lubrication analysis, geometry of bearings, materials and input parameters. The information provided in this work is intended as an aid in the assessment of metal-on-metal hip joints. Index Terms— Finite element method, Hip joint, MOM hip implants, Numerical Methods. I. INTRODUCTION There are several joints in human body, each have specific function and motion. Major joints are knee joint, shoulderjoint & most important hip joint. Hip joint plays an important role to balance the whole body weight in respect of the requirement of motion and work. In tribological study, continuous contribution of researchers improves the performance of MOM hip joint over a variety of hip implants. After having hip implant,physically patients may or may not be comfortable because of various complications that arises due to wear. This joint is ball-in-socket joint Literature review shows number of research works on hip joint replacement withvarious methods. Brown C et al. [1] concluded that in current, there is no ideal method for hip replacement in view of toxicity, adverse biological effects andtribological properties depends upon wear rate. American Academy of Orthopedic Surgeons advised for Total Hip Replacement (THR) when this joint damaged due to a fracture ,arthritis like Osteoarthritis , Rheumatoid arthritis, Post Traumatic arthritis and avascular necrosis or any childhood problem. A typical metal-on-metal hip joint is shown inFigure1. [2] Figure 1.Typical metal-on-metal hip joint [2] There are various types of hip implant joint according to their material and use i.e., ceramic-on-metal, metal-on-metal, metal-on-polyethylene, ceramic-on-ceramic, ceramic-on- polyethylene etc. In 2006 D Dowson and Z-M Jin [3] explained the four type of lubrication pattern in Metal-on-Metal hip joint tribology like hydrodynamic fluid film lubrication, mixed lubrication, boundary lubrication,and dry lubrication. They concluded that the ceramic on polyethylene joint, metal on metal joints (mild mixed lubrication regime, Co-Cr-Mo alloy high carbon>0.20%C)are significantly recommended as compare to metal and ceramic for current hip replacement due to high strength and low wear rates. In old practiceship implant, the cobalt (Co), Chromium (Cr), Molybdenum (Mo) was adopted by DobbsH. S. and M.J. Minski[4]. They observed that the metal ions are fifty times stronger than their standard value in urine and conclude that the metal-on-plastic joint are more relevant. In recent again focus on the use of metal-on-metal hip implant instead of others because of its strong capacity for wear and indicated that second generation of metal-on-metal hip implant are more efficient for strength to young patients [5][6]. Feng Liu et al. [8] reported numerically work on bearing geometry and structure support of metal-on-metal hip implantsin terms of a little clearance for 28 mm head diameter and analyzed fluid film lubrication in-vivo and in-vitro conditions. In their study, they adopted spherical mesh through
  • 2. International Journal of Technical Research and Applications e-ISSN: 2320-8163, www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287 2 | P a g e finite element and calculate the minimum film thickness by integrating the pressure with the help of fast Fourier transform.Bharma M. S. and C. P. Case[9]mentioned in their study that the metal-on-metal hip arthroplasty gives more better result as compared to metal-on-polyethylene articulation.JagatiaM. and JinZ. M. introduced Reynolds equation in spherical coordinate and it was solved by Newton- Raphson Finite Difference Method.This numerical method applied for a relatively thick actabular cup shows little effect on contact pressure [10].Leiming G et al. [11] investigated the effect of loading and motion on lubrication of metal-on-metal hip replacements. In their analysis, they adopted multigrid method for three level grid methods for 257 X 257 maximum nodes. II. LUBRICATION MODEL & MATERIALS USED Boundary lubrication is adopted for the study of various tribological surfaces to enhance the performance, and life of hip implant and protect their surfaces from wear & friction, which are in mutual contact. Elastohydrodynamic films are then adopted for better performance of joint for metal-on-metal hip implant for this purpose there are two types of models were generally adopted ball- in-socket andball- on- plane. Hip implants made up by 316Lstainless steel, cobalt-chromium- molybdenum and ceramic like alumina and zirconium, titanium based alloy.D Dowson and Z-M Jin[3] reported that the ceramic-on-polyethylene or metal-on-polyethylene have some limitation for strengthand MOM (metal-on-metal) are much durable when their design is subjected to proper surface finishFor lubrication analysis, it is assumed that flow of lubricant is in between two bearing surfaces and in this Reynolds equation for spherical co-ordinate is solved by various researchers to find out optimum result. MengQ. E [12] adoptedtransient elastohydrodynamic lubrication analysis for novel metal-on-metal of hip prosthesis with non-spherical femoral bearing surface. Synovial fluidis used as lubricant in artificial hip joint which a Non-Newtonian fluid and have a low shear stress and solved Reynolds Equation by numerical method. Equation 1                           t h R h hh hh R p h p h z y x                                                                                              22 1 2 1 33 sin12 sin cossinsinsin coscossinsin sin6sinsin Performance & lubrication analysis of metal-on-metal hip implant investigated by using above Reynolds’s equation. [11- 13] Where, and are spherical coordinates andWx ,Wy and Wz are the angular velocities presented in figure 2 forflexation- extension (FE), internal external rotation (IER) and abduction-adduction(AA) motions [11].These are the types of relative motion in between femoral head and actabularcup, under load for demonstration.Most of time this partial differential equation is widely used for lubrication analysis, texturing of hip implant for spherical coordinates. Figure 2.Metal-on-metal hip implant under loading [11]
  • 3. International Journal of Technical Research and Applications e-ISSN: 2320-8163, www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287 3 | P a g e Researchers in this field are applying this equation in numerical methods. The governing Reynolds equation (2) was written for all three dimensional motion. [14] [15]. Equation 2                           t h R h hh hh R p h p h z y x                                                                                              22 1 2 1 33 sin12 sin cossinsinsin coscossinsin sin6sinsin III. DISCUSSION In general there are various numerical methods Multigrid, fast- Fouriertransform, FDM (Finite difference Method), Newton-Raphson method were adopted for calculating convergence and optimum solution for accuracy. According toGao L. M.and MengQ. E.[13] the four level grids multi integration method has more effective as compared to fast Fourier technique.Multigrid method was also adopted by various researcherswhich is much more fast and effective.Where FDM method is limited for two elements like square and rectangle but, there is alsoananother method for analysis which is now in trend called FEM (Finite Element Method). FFT (Fast Fourier transform technique with FEM measures elastic deformation for complicated structure in artificial hip joint [18].As compared to above methods, the finite element method will be suggested for lubrication analysis of metal-on-metal hip implant for finding more realistic results with fast convergence and calculation. In hip implant expected time period of stay is in between 10-15 years [16][17].There are various vivo –vitro studies performed outside and inside living being .Wear effect on MOM hip implant under vivo condition and tested in hip simulator , linear wear depthwas found for geometry of bearing surface which was investigated for running case[19]. By Clinically metallic wear particles generated but their volume is less as compared to polyethylene [20-23] [29].Hip resurfacing and surface texturing improve the quality of MOM hip implant offer low wear also improve the quality [19-23.]For analysis following input parameters are put in under consideration given in table1 [10]. Table 1. Input parameters S.No. Parameters Value 1. Radial clearance 24-30 µm 2. Cup wall thickness 9.5mm 3. Equivalent support thickness 2mm 4. Cup inside radius 12-14mm 5. Elastic modulus of metal 210 GPa The research on two generation of metal-on-metal hip resurfacing with different cup wall thicknesses and clearances were investigated and confirmed the importance of the geometry and input parameters on the performance of MOM hip implant [24]. Surface texturing found to be more effective in fluid film lubrication by micro geometrical design on hip bearing surface [30] [31] IV. CONCLUSION Based on the above literaturereview, it is found that there is a need to develop a new way to compute more realistic results with appropriate method for analysis of MOM hip implants. Hip implants depends not only on geometry but also on proper methods and materials that are adopted for lubrication analysis , that depends on the film thickness, head and the pressure distribution around the surface of contact. An advance material combination with advance simulation methods in MOM hip implant again increases the interest to adopt this MOM hip implant in hip replacement joint. So in the field of tribological study of MOM hip implants more and more realistic studies are required, to get minimum revised cases because hip replacement procedure is more severe than previous hip prosthesis. V. NOTATIONS ø Spherical co- ordinate p Fluid pressure µ Fluid viscosity ρ Density of fluid h Fluid film thickness ω Angular velocity R1 clearance θ Angle of contact
  • 4. International Journal of Technical Research and Applications e-ISSN: 2320-8163, www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287 4 | P a g e REFERENCES [1] BrownC,Fisher J and Ingham E.2005 Bilogical effects of clinically relevant wear particles from metal-on-metal hip prostheses. University of medical and biological engineering,University of leeds,UK ,DOI:10.1234/095441105X63291. [2] https://orthoinfo.aaos.org/en/treatment/total-hip-replacement/ [3] DowsonD and Jin Z. M. 2006Metal-on-metal hip joint Tribology. Proc. IMechE,Part H:J.Engineering in Medicine,Vol.220. [4] DobbsH.S and MinskiM.J 1980Metal ion release after total hip replacements.Biomaterials. Vol. 1. [5] H.LucasAnissian,AndreStark,Vivtoria Good, Henrik Dahlstrand, Ian C. Clarke 2001The Wear Pattern in Metal-on-Metal Hip Prostheses. John Wiley & Sons, Inc. Biomed Mater Res (ApplBiomater, 58:673-678. [6] Oparaugo P, Clarke I ,Malchau H, Herberts P. 2004The correlation of wear induced osteolysis and revision with volumetric wear rates.ActaOrthopScand,72:22. [7] Wang,F.C. ,Jin,Z.M.. 2004 Prediction of elastic deformation of acetabular cups and femoral heads for lubrication analysis of artificial hip joints.Proceedings of the Institution of Mechanical Engineers,Part J ; Journal of Engineering Tribology , 2004, 218 (3),201-208. [8] Feng Liu, Zhongming. Jin, Paul Robrts, Peter Grigoris.2007Effect of bearing geometry and structural supports on transient elastohydrodynamic lubrication of metal-on-metal hip implants.Journal of Biomechanics, 40; 1340-1349. [9] M S Bharma and C P Case.2006Biological effects of metal-on- metal hip replacements. Proc. ImechE,Part H:J. Engineering in medicine,Vol.220. [10] M Jagatia and Z M Jin. 2001Elastohydrodynamic lubrication analysis of metal-on-metal hip prostheses under steady state entraining motion. Proc. ImechE,Part H:J. Engineering in medicine , 2001,Vol. 215. [11] LeimingGao, Fengcai Wang et al.2009Effect of 3D Physiological loading and motion on elastohydrodynamic lubrication of MOM total hip replacements.Medical Engineering & Physics 31; 720-729 [12] QingenMeng,LeimingGao,FengLIU,PeiranYang,John Fisher, ZhongminJin . 2009 Transient elastohydrodynamic lubrication analysis of a novel metal-on-metal hip prosthesis with an asphericalacetabular bearing surface.Journal of Biomechanics, Vol .24. [13] L M Gao, QE Meng, F C Wang, P R Yang and Z M Jin. 2007Comparison of numerical methods for elastohydrodynamic methods for metal-on-metal hip implants:muti-grid verses Newton–Raphson.Proc.IMechE, Part J: J.Engineering Tribology, Vol. 221. [14] GoenkaPK,Booker JF.1980 Spherical Bearings:Static and dynamic analysis via the finite element method.ASME J Lubr.Technol, 1980;102(7):308-19 [15] Wang FC,Jin ZM. 2008 Transient elastohydrodynamic lubrication of hip implants.ASME J Tribol;130(0011007):1-13 [16] Amstutz,H.C.,Campell, P.Kossovsky,N.,and Clark,I.C.1991 Mechanism and clinical significance of weardebris-induced osteolysis.Clin.Orthop.Relat.Res. ,276,7-17. [17] Harris,W.H. 1993 The problem is osteolysisClin.Orthop.Relat.Res., 311,46-53 [18] Z.M Jin.2006Theoretical studies of elastohydrodynamic lubrication of artificial hip joint.Proc.IMechEPart J:J Engineering Tribology, Vol.220. [19] F Liu,ZM Jin, F Hirt, C Rieker,P Roberts and P Grigoris.2005Effect of wear of bearing surfaces onelastohydrodynamic lubrication of metal–on-metal hip implants. Proc. IMechE ,Part H:J.Engineering in Medicine. Vol.219 [20] A.G Cobb ,T P Schmalzreid. 2006The clinically significance of metal ion release from cobalt-chromiummetal-on-metal hip joint arthroplasty. Proc. IMechE Part H:J.EngineeringinMedicine, Vol.220. [21] Jacobs,J.J.,Silverton, C.,Hallab, N.J.,Skipor, A.K Pattrson, L., Black,J.,andGalante, J.O. 1999 Metal release and excretion from cementless titanium alloy total knee replacements.Clin.Orhop.,358,173-180. [22] Rieker,C.B.,Schon,R.,andKottig, P. 2004 Development and validation of a second generation metal-on-metal bearing:laboratory studies and analysis of retrievals. J.Arthroplasty,19(8),Suppl.3,5 -11. [23] Mckellop,H.,Park,S.H.,Chiesa,R., Doorn,P.,Lu,B.,Normand,P.,Grigoris,P.,and Amstutz,H.1996Invivo wear three types of metal-on-metal hip prosthesis during two decades of useClin.Ortho.,1996,329,Suppl.,S128-S40. [24] Feng Liu ,Zhongmi ,Jin , P Roberts ,and P Grigoris. 2006Importance of head diameter ,clearance ,andcup wall thickness in elastohydrodynamic lubrication analysis of metal- on-metal hipresurfacingprostheses.ProcInstMechEng H.220(6):695-704 [25] Grigoris, P., Roberts, P., et al. 2005The evolution of hip resurfacing arthroplasty.Orthop.Clin.NorthAm.,2005,36(2),125- 134 [26] Dowson, D., Hardaker, C., et al. 2004 A hip joint simulator study of the performance of metal-on-metaljoints.Part I the role of materials.J.arthroplasty,19(8), Supplement 1, 118-123. [27] Dowson, D. Hardaker,C., Flett,M.,andIsaac,G.H. 2004 A hip simulator study of the performance ofmetal-on-metaljoints,Part II: design .J.Arthroplasty,19(18),124-130. [28] Rieker,C.B.,Schon,R.,Konard,R.,Liebentritt,G.,Gnepf,.,Shen,M., Roberts,andGrigoris,P. 2005 Influence of clearance on in-vitro tribology of large diameter metal-on-metal articulations pertraining to resurfacing hip implants.Orthop.,Clin. North Am.,36(2),135-142. [29] .Liu,F.,Wang,F.C.,et al. 2004 Steadystateelastohydrodynamic lubrication analysis of a metal-on-metalcup with an ultrahigh molecular weight polyethylene backing.Proc.IMechE,Part H:JEngineering in Medicine, 218(4),261-270. [30] Liu, F., Z.M., and Dowson,D. 2006 An investigation into the effect of a sandwich acetabular cup uponlubricant film thickness in metal-on-metal total hip replacements.Proc. IMechE, Part J: J Engineering Tribology,220(3),151-158. [31] L M Gao , Q E Meng Liu, J Fisher, and ZM Jin. 2010 The effect of aspherical geometry and surfacetexturing on the elastohydrodynamic lubrication of metal-on-metal hip prosthesis underphysiological loading and motions.Proc.IMechVol.224 Part C: J. Mechanical EngineeringScience,203-210.
  • 5. International Journal of Technical Research and Applications e-ISSN: 2320-8163, www.ijtra.com Volume 3, Issue 5 (September-October 2015), PP. 284-287 5 | P a g e