This document discusses various types of face masks, airways, and laryngoscopes used in airway management. It describes different face masks including anatomical masks, Rendell-Baker-Soucek pediatric masks, and scented masks. It outlines techniques for using face masks and discusses complications. It also details oropharyngeal and nasopharyngeal airways including specific designs and insertion methods. Finally, it examines rigid and fiberoptic laryngoscopes, describing the components of handles and blades as well as different blade types like the Macintosh laryngoscope.
This document discusses different types of breathing circuits used in anesthesia. It begins by introducing open, semi-closed, and closed breathing circuits. Open circuits are now obsolete and involved pouring anesthetic agents over a mask. Semi-closed circuits include Mapelson circuits A-F, with Type D (Bain) most commonly used for controlled ventilation. Closed circuits involve rebreathing of exhaled gases after carbon dioxide absorption by soda lime, making them very economical. Key components and properties of soda lime and factors affecting its carbon dioxide absorption are described.
This document discusses the laryngeal mask airway (LMA), including its history, design, indications, contraindications, side effects, necessary equipment, proper preparation and placement technique, verification of correct placement, securing, and potential problems. It also describes different types of LMAs such as the flexible, intubating, C-Trach, ProSeal, and classic LMAs.
This document provides information on various types of face masks and oral/nasal airways used in anesthesia. It describes the parts and materials of face masks, including anatomical masks, transparent masks, and scented masks. It discusses techniques for proper face mask placement and complications. It also covers oropharyngeal and nasopharyngeal airways, describing specific types like the Guedel airway and their uses and insertion techniques. Overall, the document is an overview of common airway devices used in anesthesia and their characteristics.
The document discusses various Mapleson breathing systems used for administering anesthesia. It describes the components, functioning, advantages and disadvantages of Mapleson A, B, C, D systems as well as modifications like Mapleson A-Lack system and Bain circuit. It also discusses Mapleson E (T-piece) and F (Jackson Rees) systems used for pediatric patients. The Mapleson A system is best for spontaneous breathing but the expiratory valve is difficult to use. The Bain circuit functions like a Mapleson D system but has lower resistance and better gas mixing properties.
Alternative technique of intubation retromolar, retrograde, submental and oth...Dhritiman Chakrabarti
This document discusses alternative airway techniques such as retrograde intubation, submental intubation, and others. It begins by defining difficult airway situations and providing anatomical details of the larynx. It then describes the technique of retrograde intubation, involving passing a wire through a needle in the cricothyroid membrane and using it to guide an endotracheal tube. Indications for retrograde intubation include facial anomalies limiting mouth opening. The technique of submental intubation is also summarized, involving creating a submental skin incision and tunnel to guide the endotracheal tube. Applications of these alternative techniques include maxillofacial, dental, and plastic surgeries.
The document discusses the history and use of laryngeal mask airways (LMA). It describes how Dr. Brain developed the first LMA prototype in 1981 as a supraglottic device that sits outside the trachea but provides an airway. Over time, different types of LMAs were developed including the classic LMA, ProSeal LMA, reinforced LMA, LMA-Unique, and Supreme LMA. The document outlines the features and proper insertion technique for each LMA and discusses their advantages, such as being less invasive than endotracheal tubes, as well as potential complications if not properly placed.
The document discusses different types of breathing circuits used in anesthesia. It begins by describing the basic components and functions of a breathing circuit, which delivers oxygen and anesthetic gases to patients while removing carbon dioxide. Circuits are classified as open, semi-open, semi-closed, or closed based on how exhaust gases are handled. Several specific circuit types are then outlined in detail, including the Mapleson A, Bain, Ayres T-piece, and Jackson-Rees systems. Key features and uses of each system are provided. Semi-closed circuits are explained as using a carbon dioxide absorber to remove carbon dioxide from exhaled gases so they can be rebreathed, allowing for lower fresh gas flow rates than open systems
Face masks, laryngeal tube, airways yuvarajhavalprit
This document provides information about various airway devices used in anesthesia including face masks, oral and nasal airways, and laryngoscopes. It describes the parts, types, techniques of use, advantages and disadvantages of face masks. It also discusses oropharyngeal airways, nasopharyngeal airways, and different types of laryngoscope blades including Macintosh, Miller, and specialized blades. Complications of airway devices are also mentioned.
This document discusses different types of breathing circuits used in anesthesia. It begins by introducing open, semi-closed, and closed breathing circuits. Open circuits are now obsolete and involved pouring anesthetic agents over a mask. Semi-closed circuits include Mapelson circuits A-F, with Type D (Bain) most commonly used for controlled ventilation. Closed circuits involve rebreathing of exhaled gases after carbon dioxide absorption by soda lime, making them very economical. Key components and properties of soda lime and factors affecting its carbon dioxide absorption are described.
This document discusses the laryngeal mask airway (LMA), including its history, design, indications, contraindications, side effects, necessary equipment, proper preparation and placement technique, verification of correct placement, securing, and potential problems. It also describes different types of LMAs such as the flexible, intubating, C-Trach, ProSeal, and classic LMAs.
This document provides information on various types of face masks and oral/nasal airways used in anesthesia. It describes the parts and materials of face masks, including anatomical masks, transparent masks, and scented masks. It discusses techniques for proper face mask placement and complications. It also covers oropharyngeal and nasopharyngeal airways, describing specific types like the Guedel airway and their uses and insertion techniques. Overall, the document is an overview of common airway devices used in anesthesia and their characteristics.
The document discusses various Mapleson breathing systems used for administering anesthesia. It describes the components, functioning, advantages and disadvantages of Mapleson A, B, C, D systems as well as modifications like Mapleson A-Lack system and Bain circuit. It also discusses Mapleson E (T-piece) and F (Jackson Rees) systems used for pediatric patients. The Mapleson A system is best for spontaneous breathing but the expiratory valve is difficult to use. The Bain circuit functions like a Mapleson D system but has lower resistance and better gas mixing properties.
Alternative technique of intubation retromolar, retrograde, submental and oth...Dhritiman Chakrabarti
This document discusses alternative airway techniques such as retrograde intubation, submental intubation, and others. It begins by defining difficult airway situations and providing anatomical details of the larynx. It then describes the technique of retrograde intubation, involving passing a wire through a needle in the cricothyroid membrane and using it to guide an endotracheal tube. Indications for retrograde intubation include facial anomalies limiting mouth opening. The technique of submental intubation is also summarized, involving creating a submental skin incision and tunnel to guide the endotracheal tube. Applications of these alternative techniques include maxillofacial, dental, and plastic surgeries.
The document discusses the history and use of laryngeal mask airways (LMA). It describes how Dr. Brain developed the first LMA prototype in 1981 as a supraglottic device that sits outside the trachea but provides an airway. Over time, different types of LMAs were developed including the classic LMA, ProSeal LMA, reinforced LMA, LMA-Unique, and Supreme LMA. The document outlines the features and proper insertion technique for each LMA and discusses their advantages, such as being less invasive than endotracheal tubes, as well as potential complications if not properly placed.
The document discusses different types of breathing circuits used in anesthesia. It begins by describing the basic components and functions of a breathing circuit, which delivers oxygen and anesthetic gases to patients while removing carbon dioxide. Circuits are classified as open, semi-open, semi-closed, or closed based on how exhaust gases are handled. Several specific circuit types are then outlined in detail, including the Mapleson A, Bain, Ayres T-piece, and Jackson-Rees systems. Key features and uses of each system are provided. Semi-closed circuits are explained as using a carbon dioxide absorber to remove carbon dioxide from exhaled gases so they can be rebreathed, allowing for lower fresh gas flow rates than open systems
Face masks, laryngeal tube, airways yuvarajhavalprit
This document provides information about various airway devices used in anesthesia including face masks, oral and nasal airways, and laryngoscopes. It describes the parts, types, techniques of use, advantages and disadvantages of face masks. It also discusses oropharyngeal airways, nasopharyngeal airways, and different types of laryngoscope blades including Macintosh, Miller, and specialized blades. Complications of airway devices are also mentioned.
The document provides an overview of the Esophageal-Tracheal Combitube, which is a double-lumen airway device that can be inserted blindly to secure a patient's airway. It has two tubes, one that enters the esophagus and one that positions in the pharynx. Balloons on each tube are inflated to seal the pharynx and esophagus. The device prevents vomiting and can function as an endotracheal tube if inserted into the trachea. Indications for use include injuries, bleeding, difficult intubation, and respiratory arrest. Contraindications include patient height and age restrictions and medical history. Placement and use of the device is described.
Spinal anesthesia involves injecting local anesthetic into the subarachnoid space of the spinal canal. The summary discusses the key points of spinal anesthesia including:
1. The technique involves preparing equipment and positioning the patient before inserting the spinal needle between vertebrae to inject local anesthetic and induce nerve block.
2. Complications include hypotension from sympathetic blockade and post-dural puncture headache from leakage of cerebrospinal fluid through the puncture site in the dura mater.
3. Indications are for lower body and lower abdominal surgeries, with contraindications including infection, coagulopathies, and anatomical abnormalities that prevent safe needle placement.
A nasopharyngeal airway, also known as an NPA, nasal trumpet (because of its flared end), or nose hose, is a type of airway adjunct, a tube that is designed to be inserted into the nasal passageway to secure an open airway
This document provides information about laryngoscopy and different types of laryngoscopes. It begins with an introduction to laryngoscopy and its uses such as endotracheal intubation. It then discusses the history and development of laryngoscopes. The document outlines the different parts of a direct rigid laryngoscope and various blade types including the Macintosh and Miller blades. It also discusses techniques for laryngoscopy and managing a difficult airway. Finally, it covers video laryngoscopy and the use of flexible fiberoptic endoscopes for intubation.
This document provides an overview of supraglottic airway devices. It discusses their history, classifications, indications, contraindications, complications and techniques. It describes some of the major devices including the Classic LMA, LMA Unique, Flexible LMA, LMA Fastrach, Air-Q, and LMA CTrach. Supraglottic devices are used to maintain airway patency and provide ventilation above the vocal cords. They have advantages over face masks and endotracheal tubes in certain situations but also have potential complications if not properly placed.
The breathing system delivers gases from the anesthesia machine to the patient's airway. It starts from the fresh gas inlet and ends where gases escape. The circle system uses a circular pathway where exhaled CO2 is removed by an absorbent in the CO2 absorber. Key components include the absorber, unidirectional valves, inspiratory and expiratory ports, fresh gas inlet, Y-piece, APL valve, and breathing tubes. The absorbent, usually sodalime, neutralizes CO2 through an exothermic reaction in the canister. Placement of components and fresh gas flow influences gas flows and absorbent desiccation. The circle system reduces agent use and waste gas exposure while increasing tracheal warmth and
This document discusses breathing systems used in anesthesia. It defines a breathing system and lists its main components. The key requirements of an effective breathing system are to deliver accurate gas concentrations, eliminate carbon dioxide, minimize dead space, and have low resistance. Various configurations are described, including open, semi-open, semi-closed and closed systems. Popular breathing circuits like Mapleson A, B, C, D, E and F are explained along with the Ayre's T-piece and reservoir bag. The document provides details on how different breathing systems function during spontaneous and controlled ventilation.
This document provides information on non-return valves (unidirectional valves), including their uses, components, classifications, and descriptions of specific valve types. It discusses how unidirectional valves work for both spontaneous and controlled ventilation. Several common valve designs are described, including their functions, evaluations, care requirements, advantages and disadvantages. The key points are:
Unidirectional valves allow gas flow in one direction only and are useful for resuscitation where CO2 absorption is not available. Common valves include the Ambu, E2, Fink and Laerdal designs. Valves are classified by their mechanical characteristics and whether they are for spontaneous, controlled, or both types of ventilation. Proper cleaning and potential hazards are
The document discusses the components and functioning of an anaesthesia work station's high pressure system. It describes the key components including gas cylinders, hanger yokes, cylinder pressure indicators, and pressure regulators. Gas cylinders contain medical gases at high pressure and have valves, handles, pressure relief devices, and markings. Hanger yokes orient and secure cylinders, providing a gas-tight seal. Cylinder pressure indicators display the pressure level in cylinders. Pressure regulators reduce the high cylinder pressure to a lower, constant pressure suitable for use in the anaesthesia machine.
Laryngeal Mask Airway & Igel - An IntroductionHIRANGER
The document provides information about the Laryngeal Mask Airway (LMA) and I-gel devices for airway management:
[1] It discusses the indications, contraindications, equipment, and step-by-step process for inserting the LMA. Proper sizing, lubrication, and positioning are emphasized.
[2] Placement must be verified by ensuring equal breath sounds on both sides and absence of sounds over the epigastrium. Problems like improper deflation or folding can occur.
[3] The I-gel is introduced as a single-use, cuffless airway device with an integral gastric channel and epiglottis blocking ridge to facilitate insertion.
The document describes the Baska Mask, a new supraglottic airway device that aims to address limitations of existing laryngeal masks by having a smaller opening to reduce risk of aspiration, a tab to aid in placement, dual gastric channels for ventilation and suctioning, and a cuffless design that inflates with ventilation rather than requiring separate cuff inflation. Standard placement and removal techniques for the Baska Mask are provided.
The document defines and describes breathing systems used in anesthesia. It discusses the purpose of breathing systems to deliver anesthetic gases and oxygen safely. The key components and requirements of breathing systems are explained, including minimizing resistance to gas flow. Different types of breathing systems are classified and their mechanisms and advantages/disadvantages are summarized.
Airway anatomy its assessment and anaesthetic implicationAPARNA SAHU
The document discusses airway anatomy, including definitions of the airway and its subdivisions. It describes the structures of the upper airway from the oral cavity to the larynx in detail. This includes the muscles, cartilages, and functions of the oral cavity, nose, pharynx, larynx. It discusses the implications of airway anatomy for airway management and anesthesia, such as the need for humidification during intubation. Difficulties that can arise from various anatomical structures are also summarized, such as from deviations of the nasal septum or injuries to the turbinates during nasotracheal intubation.
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewDhritiman Chakrabarti
The document discusses various supraglottic airway devices including the bougie, tracheal light, laryngeal tube, and combitube.
The bougie is an intubation aid that is inserted through the vocal cords to help guide placement of an endotracheal tube. The tracheal light uses transillumination to help visualize placement of an endotracheal tube in difficult airways. The laryngeal tube is a new supraglottic airway device made of silicone that provides an alternative to endotracheal intubation or laryngeal mask airway placement. The combitube is a double lumen tube that can provide ventilation whether placed in the trachea or esoph
This document discusses supraglottic airway devices. It begins by introducing supraglottic airway devices and their purpose of maintaining airway patency above the glottic opening. It then classifies supraglottic devices based on generation, sealing mechanism, number of lumens, and discusses some common devices like the LMA Classic, Unique, Flexible, and Ambu Aura. Indications, contraindications, advantages, disadvantages, proper sizing, insertion technique and signs of correct placement are outlined. Potential problems and methods to reduce aspiration are also reviewed.
This document discusses oropharyngeal airways, which are curved plastic devices inserted into the mouth to prevent the tongue from blocking the airway. They are indicated for unconscious patients without a gag reflex. The proper size is selected by measuring from the corner of the mouth to the angle of the jaw. The airway should be inserted upside down and rotated sideways as it passes the tongue, then positioned with the flange at the teeth to maintain the airway. Failure to manage the airway properly can lead to preventable death, so early detection, rapid intervention and continual reassessment of the airway are important.
The document discusses the components and functioning of the anesthesia machine. It describes the anesthesia machine as integrating components for anesthesia administration. The machine consists of the anesthesia machine itself, ventilator, breathing system, scavenging system, monitors and may include drug delivery systems. The document outlines the history of developments to the anesthesia machine since its original conception in 1917. It also describes the types of machines, standards, and basic schematics including electrical, pneumatic and gas supply components.
Supraglottic airway devices deliver gases above the vocal cords. The Laryngeal Mask Airway (LMA) was invented in 1981 and introduced improvements over prior extraglottic devices. There are now multiple versions of the LMA with variations in material, shape, and features. Proper selection and insertion technique are important for effective use of these airway devices.
The document discusses monitored anesthesia care (MAC), which involves administering drugs to provide anxiolytic, hypnotic, amnestic, and analgesic effects without depressing consciousness below a certain level. It provides guidelines on drug selection and dosing for MAC, including opioids like fentanyl and remifentanil, benzodiazepines like midazolam, propofol, ketamine, and dexmedetomidine. It also discusses factors that can lead to patient agitation during MAC and principles of drug administration via continuous infusion or patient-controlled methods.
This document discusses face masks, airways, laryngoscopes, endotracheal tubes, and extubation. Face masks are used for non-intubated ventilation and come in different sizes. Airways are inserted to prevent the tongue from falling back. Laryngoscopes are used to visualize the glottis and facilitate intubation, and come in different blade shapes like Macintosh and Miller. Endotracheal tubes are inserted into the trachea and come in different types and sizes. Extubation should be performed when respiration is adequate and during inspiration.
The document provides an overview of the Esophageal-Tracheal Combitube, which is a double-lumen airway device that can be inserted blindly to secure a patient's airway. It has two tubes, one that enters the esophagus and one that positions in the pharynx. Balloons on each tube are inflated to seal the pharynx and esophagus. The device prevents vomiting and can function as an endotracheal tube if inserted into the trachea. Indications for use include injuries, bleeding, difficult intubation, and respiratory arrest. Contraindications include patient height and age restrictions and medical history. Placement and use of the device is described.
Spinal anesthesia involves injecting local anesthetic into the subarachnoid space of the spinal canal. The summary discusses the key points of spinal anesthesia including:
1. The technique involves preparing equipment and positioning the patient before inserting the spinal needle between vertebrae to inject local anesthetic and induce nerve block.
2. Complications include hypotension from sympathetic blockade and post-dural puncture headache from leakage of cerebrospinal fluid through the puncture site in the dura mater.
3. Indications are for lower body and lower abdominal surgeries, with contraindications including infection, coagulopathies, and anatomical abnormalities that prevent safe needle placement.
A nasopharyngeal airway, also known as an NPA, nasal trumpet (because of its flared end), or nose hose, is a type of airway adjunct, a tube that is designed to be inserted into the nasal passageway to secure an open airway
This document provides information about laryngoscopy and different types of laryngoscopes. It begins with an introduction to laryngoscopy and its uses such as endotracheal intubation. It then discusses the history and development of laryngoscopes. The document outlines the different parts of a direct rigid laryngoscope and various blade types including the Macintosh and Miller blades. It also discusses techniques for laryngoscopy and managing a difficult airway. Finally, it covers video laryngoscopy and the use of flexible fiberoptic endoscopes for intubation.
This document provides an overview of supraglottic airway devices. It discusses their history, classifications, indications, contraindications, complications and techniques. It describes some of the major devices including the Classic LMA, LMA Unique, Flexible LMA, LMA Fastrach, Air-Q, and LMA CTrach. Supraglottic devices are used to maintain airway patency and provide ventilation above the vocal cords. They have advantages over face masks and endotracheal tubes in certain situations but also have potential complications if not properly placed.
The breathing system delivers gases from the anesthesia machine to the patient's airway. It starts from the fresh gas inlet and ends where gases escape. The circle system uses a circular pathway where exhaled CO2 is removed by an absorbent in the CO2 absorber. Key components include the absorber, unidirectional valves, inspiratory and expiratory ports, fresh gas inlet, Y-piece, APL valve, and breathing tubes. The absorbent, usually sodalime, neutralizes CO2 through an exothermic reaction in the canister. Placement of components and fresh gas flow influences gas flows and absorbent desiccation. The circle system reduces agent use and waste gas exposure while increasing tracheal warmth and
This document discusses breathing systems used in anesthesia. It defines a breathing system and lists its main components. The key requirements of an effective breathing system are to deliver accurate gas concentrations, eliminate carbon dioxide, minimize dead space, and have low resistance. Various configurations are described, including open, semi-open, semi-closed and closed systems. Popular breathing circuits like Mapleson A, B, C, D, E and F are explained along with the Ayre's T-piece and reservoir bag. The document provides details on how different breathing systems function during spontaneous and controlled ventilation.
This document provides information on non-return valves (unidirectional valves), including their uses, components, classifications, and descriptions of specific valve types. It discusses how unidirectional valves work for both spontaneous and controlled ventilation. Several common valve designs are described, including their functions, evaluations, care requirements, advantages and disadvantages. The key points are:
Unidirectional valves allow gas flow in one direction only and are useful for resuscitation where CO2 absorption is not available. Common valves include the Ambu, E2, Fink and Laerdal designs. Valves are classified by their mechanical characteristics and whether they are for spontaneous, controlled, or both types of ventilation. Proper cleaning and potential hazards are
The document discusses the components and functioning of an anaesthesia work station's high pressure system. It describes the key components including gas cylinders, hanger yokes, cylinder pressure indicators, and pressure regulators. Gas cylinders contain medical gases at high pressure and have valves, handles, pressure relief devices, and markings. Hanger yokes orient and secure cylinders, providing a gas-tight seal. Cylinder pressure indicators display the pressure level in cylinders. Pressure regulators reduce the high cylinder pressure to a lower, constant pressure suitable for use in the anaesthesia machine.
Laryngeal Mask Airway & Igel - An IntroductionHIRANGER
The document provides information about the Laryngeal Mask Airway (LMA) and I-gel devices for airway management:
[1] It discusses the indications, contraindications, equipment, and step-by-step process for inserting the LMA. Proper sizing, lubrication, and positioning are emphasized.
[2] Placement must be verified by ensuring equal breath sounds on both sides and absence of sounds over the epigastrium. Problems like improper deflation or folding can occur.
[3] The I-gel is introduced as a single-use, cuffless airway device with an integral gastric channel and epiglottis blocking ridge to facilitate insertion.
The document describes the Baska Mask, a new supraglottic airway device that aims to address limitations of existing laryngeal masks by having a smaller opening to reduce risk of aspiration, a tab to aid in placement, dual gastric channels for ventilation and suctioning, and a cuffless design that inflates with ventilation rather than requiring separate cuff inflation. Standard placement and removal techniques for the Baska Mask are provided.
The document defines and describes breathing systems used in anesthesia. It discusses the purpose of breathing systems to deliver anesthetic gases and oxygen safely. The key components and requirements of breathing systems are explained, including minimizing resistance to gas flow. Different types of breathing systems are classified and their mechanisms and advantages/disadvantages are summarized.
Airway anatomy its assessment and anaesthetic implicationAPARNA SAHU
The document discusses airway anatomy, including definitions of the airway and its subdivisions. It describes the structures of the upper airway from the oral cavity to the larynx in detail. This includes the muscles, cartilages, and functions of the oral cavity, nose, pharynx, larynx. It discusses the implications of airway anatomy for airway management and anesthesia, such as the need for humidification during intubation. Difficulties that can arise from various anatomical structures are also summarized, such as from deviations of the nasal septum or injuries to the turbinates during nasotracheal intubation.
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewDhritiman Chakrabarti
The document discusses various supraglottic airway devices including the bougie, tracheal light, laryngeal tube, and combitube.
The bougie is an intubation aid that is inserted through the vocal cords to help guide placement of an endotracheal tube. The tracheal light uses transillumination to help visualize placement of an endotracheal tube in difficult airways. The laryngeal tube is a new supraglottic airway device made of silicone that provides an alternative to endotracheal intubation or laryngeal mask airway placement. The combitube is a double lumen tube that can provide ventilation whether placed in the trachea or esoph
This document discusses supraglottic airway devices. It begins by introducing supraglottic airway devices and their purpose of maintaining airway patency above the glottic opening. It then classifies supraglottic devices based on generation, sealing mechanism, number of lumens, and discusses some common devices like the LMA Classic, Unique, Flexible, and Ambu Aura. Indications, contraindications, advantages, disadvantages, proper sizing, insertion technique and signs of correct placement are outlined. Potential problems and methods to reduce aspiration are also reviewed.
This document discusses oropharyngeal airways, which are curved plastic devices inserted into the mouth to prevent the tongue from blocking the airway. They are indicated for unconscious patients without a gag reflex. The proper size is selected by measuring from the corner of the mouth to the angle of the jaw. The airway should be inserted upside down and rotated sideways as it passes the tongue, then positioned with the flange at the teeth to maintain the airway. Failure to manage the airway properly can lead to preventable death, so early detection, rapid intervention and continual reassessment of the airway are important.
The document discusses the components and functioning of the anesthesia machine. It describes the anesthesia machine as integrating components for anesthesia administration. The machine consists of the anesthesia machine itself, ventilator, breathing system, scavenging system, monitors and may include drug delivery systems. The document outlines the history of developments to the anesthesia machine since its original conception in 1917. It also describes the types of machines, standards, and basic schematics including electrical, pneumatic and gas supply components.
Supraglottic airway devices deliver gases above the vocal cords. The Laryngeal Mask Airway (LMA) was invented in 1981 and introduced improvements over prior extraglottic devices. There are now multiple versions of the LMA with variations in material, shape, and features. Proper selection and insertion technique are important for effective use of these airway devices.
The document discusses monitored anesthesia care (MAC), which involves administering drugs to provide anxiolytic, hypnotic, amnestic, and analgesic effects without depressing consciousness below a certain level. It provides guidelines on drug selection and dosing for MAC, including opioids like fentanyl and remifentanil, benzodiazepines like midazolam, propofol, ketamine, and dexmedetomidine. It also discusses factors that can lead to patient agitation during MAC and principles of drug administration via continuous infusion or patient-controlled methods.
This document discusses face masks, airways, laryngoscopes, endotracheal tubes, and extubation. Face masks are used for non-intubated ventilation and come in different sizes. Airways are inserted to prevent the tongue from falling back. Laryngoscopes are used to visualize the glottis and facilitate intubation, and come in different blade shapes like Macintosh and Miller. Endotracheal tubes are inserted into the trachea and come in different types and sizes. Extubation should be performed when respiration is adequate and during inspiration.
This document discusses various techniques for airway management. It describes mechanical maneuvers like jaw thrust and head tilt-chin lift to clear obstructions. Common airway adjuncts like oropharyngeal and nasopharyngeal airways are also discussed. Guidelines are provided for sizing and inserting these adjuncts safely. Face masks can be used with one, two, or three hands to maintain a patent airway. Risk factors for difficult mask ventilation and potential complications are outlined.
This document discusses various techniques for airway management. It describes mechanical maneuvers like jaw thrust and head tilt-chin lift to clear obstructions. Common airway adjuncts like oropharyngeal and nasopharyngeal airways are also discussed. Guidelines are provided for sizing and inserting these adjuncts safely. Face masks can be used with one, two, or three hands to maintain a patent airway. Risk factors for difficult mask ventilation and potential complications are also outlined.
The document discusses various types of airway equipment used in anesthesia including face masks, oral and nasopharyngeal airways, laryngoscopes, and endotracheal tubes. It describes the components, uses, sizes and types of each device. Face masks allow administration of gases without an apparatus in the mouth. Oral and nasopharyngeal airways maintain an open airway. Laryngoscopes are used to visualize the larynx for intubation or foreign body removal. Endotracheal tubes provide a conduit for gases into the trachea during anesthesia. The document provides detailed information on the features and proper use of these important airway management devices.
The document discusses various types of airway equipment used in anesthesia including face masks, oral and nasopharyngeal airways, laryngoscopes, and endotracheal tubes. It describes the components, sizes, uses, and complications of each type of equipment. Specific tubes are also outlined such as Cole tubes, spiral embedded tubes, preformed tubes, and tubes designed for laryngectomy and microlaryngeal surgery.
This document provides information on various types of face masks and oral/nasal airways used in anesthesia. It describes the parts and materials of face masks, including anatomical masks, transparent masks, and scented masks. It discusses techniques for proper face mask placement and complications. It also covers oropharyngeal and nasopharyngeal airways, describing specific types like the Guedel airway and their uses and insertion techniques.
This document discusses basic airway equipment used in anesthesia including face masks, airways, laryngeal mask airways, and endotracheal tubes. It describes the parts, materials, sizes, insertion techniques, advantages, disadvantages, and cleaning methods for each device. Common airway devices covered are the face mask, nasal airway, oropharyngeal airway, laryngeal mask airway (including Classic, ProSeal, and other types), and endotracheal tubes. Complications associated with each device are also summarized. The document provides an overview of basic airway management equipment.
The document discusses various airway management devices and techniques used in basic life support. It describes oropharyngeal airways, nasopharyngeal airways, laryngeal mask airways, endotracheal tubes, and difficult airway management devices like the Combitube. Guidelines for sizing and inserting these devices are provided, along with their indications, contraindications, and potential complications. Factors that can make intubation difficult are also outlined.
vital role
Pre-oxygenation: vital
Awake intubation: consider
Alternative airway: have ready
Senior help: call early
Cricothyrotomy: know how to do
Postpone if not urgent
Don't panic, think and act
Document: vitally important
Prepare for worst
Train and practice regularly
The document summarizes different types of supraglottic airway devices including their advantages, indications, and techniques for insertion and removal. It discusses the laryngeal mask airway (LMA) in detail, describing its design, sizes, and proper insertion technique. Other supraglottic devices covered include the LMA Fastrach, Proseal LMA, LMA-C Trach, I-gel, and LMA Supreme. Complications are also outlined. The document provides an overview of supraglottic airways for maintaining the airway.
This document discusses supraglottic airway devices. It begins by defining supraglottic airway devices as those that maintain airway patency by sitting just above the glottic opening without entering the trachea. It then classifies these devices based on generation, sealing mechanism, number of lumens, and other characteristics. Key devices discussed include the LMA Classic, Flexible LMA, ProSeal LMA, and Intubating LMA. Advantages and disadvantages of supraglottic airway devices are provided. Insertion techniques and signs of correct placement are also summarized.
The document discusses difficult airway assessment and management. It defines a difficult airway as situations involving difficult mask ventilation, difficult intubation, difficult placement of a supraglottic airway device, or difficult surgical airway access. It describes predictors of a difficult airway related to patient characteristics and anatomy. It also discusses the importance of assessing the airway and having appropriate equipment and personnel prepared when encountering an anticipated or unanticipated difficult airway.
The document provides an overview of airway anatomy and management techniques. It describes the anatomy starting from the nose down to the trachea. It then discusses factors that can make the airway difficult and techniques for assessing the airway. It explains various airway management techniques including mask ventilation, use of airways, laryngoscopy, intubation, and alternative techniques like LMA and needle cricothyrotomy.
This document discusses supraglottic airway devices. It begins by introducing supraglottic airway devices as those that maintain airway patency by sitting above the glottic opening. It then classifies devices based on generation, sealing mechanism, number of lumens, and discusses indications, contraindications, advantages, and disadvantages of supraglottic airway devices. Specific devices like LMA Classic, Flexible LMA, Ambu Aura, Soft Seal LMA, and Intubating LMA are then described in more detail. Problems associated with devices and techniques to reduce aspiration are also covered.
Endotracheal tubes are used to intubate patients and enable ventilation. They are typically made of PVC or rubber and have features like a Murphy eye, size designations, and a pilot balloon-connected inflation system to create a seal in the trachea. Complications can occur during or after intubation and extubation, like trauma, aspiration, or laryngospasm. Nasotracheal intubation has advantages like patient comfort but risks like trauma or sinusitis. Proper preparation, techniques, and monitoring are important for safe endotracheal intubation.
Endotracheal tubes are used to intubate patients and enable ventilation. They are typically made of PVC or rubber and have features like a Murphy eye, size designations, and a pilot balloon-connected inflation system to create a seal in the trachea. Complications can occur during or after intubation and extubation, like trauma, aspiration, or laryngospasm. Nasotracheal intubation has advantages like patient comfort but risks like trauma or sinusitis. Proper preparation, techniques, and monitoring are important for safe endotracheal intubation.
- The laryngeal mask airway (LMA) is a supraglottic airway device that is placed in the hypopharynx to control the airway during general anesthesia or ventilation. It provides an alternative to endotracheal intubation or use of a face mask. The LMA has advantages like ease of insertion, reduced hemodynamic response, and improved oxygenation during emergence from anesthesia. Potential complications include sore throat, coughing, laryngospasm, and airway obstruction. Proper selection of size, lubrication, and insertion technique are important for successful use of the LMA.
- Supraglottic airway devices such as the laryngeal mask airway (LMA) provide an alternative to endotracheal intubation for maintaining a patient's airway during general anesthesia. The LMA is easy to place, improves cardiovascular stability, and allows for smooth induction and recovery compared to endotracheal intubation. Various types of LMAs have been developed with features like gastric drainage ports to reduce the risk of aspiration. Proper selection of the appropriately sized LMA and correct insertion technique are important for safely managing a patient's airway.
Oropharyngeal airway and nasopharyngeal airway_114337 (1).pptxSuchismita Pal
The document discusses oropharyngeal airways and nasopharyngeal airways. Oropharyngeal airways lift the tongue and epiglottis to prevent airway obstruction. They are made of various materials and sizes and are used to maintain an open airway. Nasopharyngeal airways resemble shortened tracheal tubes and are inserted through the nose to maintain the airway. Various types of each are described along with their uses, insertion techniques, complications, and when one may be preferred over the other.
Similar to #Laryngoscopes face mask-airway & endotracheal tubes (20)
This document discusses airway local blocks and awake intubation. It describes the indications for awake intubation including comorbidities, risk of aspiration, difficult airway assessment, and emergencies. It provides details on the pharmacological agents, equipment, personnel, and techniques used for airway local blocks and awake intubation. Specifically, it outlines common methods for anesthetizing different areas of the airway using lidocaine, including dosage calculations and risks of lidocaine toxicity. The goal is to safely anesthetize the airway to allow for awake intubation.
This document discusses endotracheal tubes and intubation. It covers indications for intubation including airway protection, optimizing gas exchange, decreasing metabolic demand, and reducing work of breathing. Conditions associated with difficult intubation are described such as congenital anomalies, infections, tumors, injuries, and obesity. Proper equipment, tube sizing, intubation technique including positioning and confirmation of placement are outlined. Golden rules of intubation emphasize preparation, oxygenation, skills, confirmation, and monitoring.
Appropriate airway equipment and techniques.Nisar Arain
This document provides an overview of airway anatomy, equipment, and techniques for airway management. It discusses:
1. The importance of airway control and the development of advanced cardiac life support.
2. The objectives of reviewing upper and lower airway anatomy, basic and advanced airway techniques, equipment for difficult airways, and clinical management of the airway.
3. Details of upper airway structures including the nose, oral cavity, pharynx, and larynx. It also reviews lower airway structures like the trachea and lungs.
This document discusses different types of fluid flow and transport mechanisms in cells. It describes laminar and turbulent fluid flow, how they are characterized, and factors that influence each type. It also outlines different transport mechanisms in cells including diffusion, osmosis, facilitated diffusion, active transport, endocytosis, and exocytosis. Active transport uses carrier proteins and cell energy to move substances against a concentration gradient, while passive transport moves substances down a concentration gradient without cell energy.
1. The document discusses theories of anesthesia including the unitary theory and modern theories involving interactions with membrane proteins and specific ion channels.
2. It describes the stages of general anesthesia from analgesia to surgical anesthesia to medullary respiratory paralysis. However, it notes that the excitement stage is rarely seen with modern anesthesia.
3. GABA receptors are identified as an important target for many anesthetic agents. General anesthetics bind to these receptors, causing chloride channel opening and neuronal inhibition, resulting in anesthesia.
Endotracheal intubation and laryngoscopy part 2Nisar Arain
This document discusses various types of endotracheal tubes and laryngoscopy techniques. It describes specialized tubes like armored tubes, RAE tubes, Oxford tubes, and laser-resistant tubes. It covers direct laryngoscopy using curved and straight blades. Optimal conditions for laryngoscopy are outlined, including using the appropriately sized blade and ensuring good muscle relaxation. Reliable signs of correct endotracheal tube placement include capnography, visualization of the tube passing the vocal cords, and fiberoptic bronchoscopy visualization of tracheal rings.
Complications of artificial applications part 5Nisar Arain
This document discusses the complications that can arise from mechanical ventilation through either invasive or non-invasive means. Some key complications mentioned include pneumonia, infections, injuries to the face/lips/pharynx and larynx/trachea, gastrointestinal effects like esophagitis and decreased motility, renal effects from reductions in blood pressure/flow, disrupted sleep, and decubitus ulcers. Proper diagnosis and management of these various complications is important for patients receiving mechanical ventilation support.
This document discusses endotracheal tubes and intubation. It covers indications for intubation including airway protection, optimizing gas exchange, decreasing metabolic demand, and reducing work of breathing. Conditions associated with difficult intubation are described such as congenital anomalies, infections, tumors, and injuries. Airway assessment techniques like mallampati classification, laryngoscopy view, and thyromental distance are explained. Equipment for intubation and sizing endotracheal tubes are outlined. The technique of intubation is described involving positioning the patient in sniffing position and using a laryngoscope to visualize the vocal cords. Confirmation of proper tube placement is emphasized using methods like auscultation and capnography.
This document discusses airway assessment and difficult airways. It outlines various predictors of difficult airways like obesity, short neck, and facial hair. It describes tests to evaluate the airway like thyromental distance, inter-incisor gap, and Mallampati grading. The document emphasizes the importance of a thorough airway assessment prior to intubation to identify potential difficulties and prepare appropriate management strategies for difficult intubations.
The document discusses preoperative airway assessment for anesthesia. It notes that 1-3 out of 100 anesthetized patients have difficult intubation, while 1 out of 1000 have failed intubation and 1 out of 10,000 experience cannot intubate cannot ventilate scenarios. Factors that can increase intubation difficulty include congenital syndromes, anatomical features like teeth and neck structure, and acquired conditions such as decreased jaw or neck mobility. A thorough preoperative assessment including tests of mouth opening, neck movement, thyromental distance, and Mallampati score can help predict and prepare for a potentially difficult airway.
- Imhotep, an ancient Egyptian priest from around 2600 BC, is considered the first physician and treated many diseases. He extracted medicines from plants and had knowledge of anatomy. Ancient Egyptians used opium and hyoscyamus for anesthesia and performed trepanation surgery.
- In ancient Greece and Rome, mandrake juice was used for its narcotic effects before surgeries to ensure insensibility to pain. Arabic translations of Greek medicine advanced Islamic medicine in the Middle Ages. Physicians like Al Zahrawi described many surgeries and instruments.
- The modern history of anesthesia began with William Morton using ether in 1846 and John Snow advancing the field through publications on ether and chlor
This document discusses different types of anesthesia including local, regional, and general anesthesia. It provides details on common regional anesthesia techniques like spinal blocks, epidurals, and caudal blocks. It also describes local anesthesia techniques such as infiltration, nerve blocks, and intravenous regional anesthesia. The document discusses the mechanisms of local anesthetics and some potential complications as well as benefits of local and regional anesthesia compared to general anesthesia.
This document discusses the problem of anesthesia awareness during surgery. It defines anesthesia awareness as a patient becoming conscious during a surgical procedure under general anesthesia and having recall of events. Risk factors include women, younger age, use of total intravenous anesthesia, long surgeries, prior awareness history, and natural red hair. Causes can include light anesthesia, increased anesthetic requirements, or anesthesiologist error. Prevention strategies include pre-operative evaluation, prophylactic benzodiazepines, monitoring anesthetic levels, and post-operative interviews. Methods to monitor consciousness include clinical signs, isolated forearm technique, brain monitoring like BIS, and measurements of lower esophageal sphincter contractions.
- The document discusses various physiological changes that occur with aging and their implications for anesthesia in geriatric patients. Some key points discussed include:
- Cardiovascular changes like decreased cardiac output and increased risk of hypertension. Respiratory changes like reduced lung capacity and cough reflex. Genitourinary changes like reduced kidney function and bladder issues.
- Gastrointestinal changes like decreased motility leading to constipation. Endocrine changes like increased risk of hypothyroidism and bone issues.
- The implications of these changes for anesthesia include risks of hypotension, bradycardia, respiratory complications, slower drug metabolism and clearance, and risks of gastric aspiration and constipation. Careful preoperative evaluation and
The document provides information on general anesthesia including:
1) It discusses the history, goals, and levels of sedation for general anesthesia. Different levels include minimal sedation, moderate sedation, deep sedation, and general anesthesia.
2) The pre-anesthetic evaluation process involves taking a medical history, performing a physical exam including airway assessment, and ordering lab tests.
3) Common anesthetic equipment is described including laryngoscopes, endotracheal tubes, airways, monitors, and intravenous and inhalational drugs used for induction and maintenance of general anesthesia.
The document provides information on the respiratory system, including its structures and functions. It discusses the processes of ventilation, external respiration, transport of gases, and cellular respiration. It describes the structures of the upper respiratory tract such as the nose, pharynx and larynx. It also details the trachea, bronchi, bronchioles, and alveoli. Furthermore, it examines the muscles involved in inspiration and expiration, respiratory volumes and capacities, and the control of breathing.
The document discusses body temperature regulation and abnormalities. It notes that humans maintain a constant core body temperature of around 37°C through heat gain and loss mechanisms controlled by the hypothalamus. When temperature varies by 0.1°C from the set point, the hypothalamus activates heat conservation or dissipation responses. Disorders include hypothermia, where temperature drops below the normal range, and hyperthermia/fever, where the hypothalamus raises the set point in response to pyrogens like bacterial toxins or cytokines. Heat stroke occurs when temperature exceeds the critical threshold of around 105-108°F.
This document discusses triage, which is the process of sorting patients based on the urgency of their condition to prioritize care. It defines triage and outlines its objectives to identify patients, prioritize their needs, track their progress, and identify hazards. It describes different types of triage including simple, advanced, continuous integrated, and reverse triage. It discusses levels of triage from red to green based on factors like vital signs and Glasgow Coma Scale. Finally, it outlines the role and characteristics of EMTs in effectively conducting triage.
The document discusses the design and setup of an operation theatre. It outlines key areas that should be included like an anesthesia room, sterilization room, recovery room, and storeroom. It also describes the daily, weekly, and deep cleaning procedures needed to properly sanitize the operation theatre and prevent infections. Furthermore, it covers the preparation of equipment, instruments, and supplies as well as different sterilization methods like autoclaving, gamma irradiation, and ethylene oxide.
The document discusses the importance of effective communication in healthcare. It introduces SBAR as a standardized communication technique that structures information exchange into 4 steps: Situation, Background, Assessment, and Recommendation. SBAR provides a mechanism to frame conversations, a tool to share concise and focused information, and a standard for communication that can help close hierarchies and indicate effective information transfer between caregivers. The document advocates for SBAR to foster patient safety, develop teamwork, and reduce time spent on patient handovers.
Histololgy of Female Reproductive System.pptxAyeshaZaid1
Dive into an in-depth exploration of the histological structure of female reproductive system with this comprehensive lecture. Presented by Dr. Ayesha Irfan, Assistant Professor of Anatomy, this presentation covers the Gross anatomy and functional histology of the female reproductive organs. Ideal for students, educators, and anyone interested in medical science, this lecture provides clear explanations, detailed diagrams, and valuable insights into female reproductive system. Enhance your knowledge and understanding of this essential aspect of human biology.
- Video recording of this lecture in English language: https://youtu.be/kqbnxVAZs-0
- Video recording of this lecture in Arabic language: https://youtu.be/SINlygW1Mpc
- Link to download the book free: https://nephrotube.blogspot.com/p/nephrotube-nephrology-books.html
- Link to NephroTube website: www.NephroTube.com
- Link to NephroTube social media accounts: https://nephrotube.blogspot.com/p/join-nephrotube-on-social-media.html
Basavarajeeyam is a Sreshta Sangraha grantha (Compiled book ), written by Neelkanta kotturu Basavaraja Virachita. It contains 25 Prakaranas, First 24 Chapters related to Rogas& 25th to Rasadravyas.
Adhd Medication Shortage Uk - trinexpharmacy.comreignlana06
The UK is currently facing a Adhd Medication Shortage Uk, which has left many patients and their families grappling with uncertainty and frustration. ADHD, or Attention Deficit Hyperactivity Disorder, is a chronic condition that requires consistent medication to manage effectively. This shortage has highlighted the critical role these medications play in the daily lives of those affected by ADHD. Contact : +1 (747) 209 – 3649 E-mail : sales@trinexpharmacy.com
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptxHolistified Wellness
We’re talking about Vedic Meditation, a form of meditation that has been around for at least 5,000 years. Back then, the people who lived in the Indus Valley, now known as India and Pakistan, practised meditation as a fundamental part of daily life. This knowledge that has given us yoga and Ayurveda, was known as Veda, hence the name Vedic. And though there are some written records, the practice has been passed down verbally from generation to generation.
3. INTRODUCTION
1-It is the device which allows
administration of gases to
the patient from the breathing
system without introducing
any apparatus to the patients
mouth
4. DESCRIPTION OF MASK
---A Face mask can be made
up of Black rubber, clear
Plastics, Elastomeric
material or combination
of these Parts of the
Face Mask
BODY
-It constitutes the main
part of the mask.
This Transparent
body allows observation
of Moisture, Vomitus,
and secretions etc.
5. SEAL:-
---This comes in contact with the
Face. Two types are available
a-Pad or Cushion:-Inflated with air
b-Flap:-Flexible extension of the body
CONNECTOR:- (Orifice / collar )
a-Opposite to the seal
b-Thickened fitting of 22 mm ID
c-Ring with hooks helps in strapping
the Mask
6. SPECIFIC MASKS
-This can be moulded to conform
to the Anatomy of Face
-It has slightly Malleable rubber
body, a sharp notch for the nose
and a curved chin section
-- 1-Anatomical Mask
7. 2-RANDELL-BAKER-SOUCEK
(RBS) mask
-It is designed for Pediatric
patients < then 10 years
-It has a triangular body
and Low dead space
-It is used in Tracheostomy
and Acromegaly patients
8. Size
Age group Dead space
0 preterm 3 cu.mm
1 infant 4 cu.mm
2 1 to 3 years 8 cu.mm
3 4 to 10 years 12 cu.mm
9. 3-ENDOSCOPY MASK
-Designed to allow mask
ventilation during
Endoscopy
-It has a port or Diaphragm
in the body to allow Fibro
scope insertion
10. 4-SCENTED MASK
-Scent incorporated into the Mask by the
manufacturer or Anesthetic provider
-Added for better acceptability and
pleasant experience during induction
-Its Disadvantage is Ethyl Alcohol in
some flavours may affect accuracy of
Gas monitors -
11. TECHNIQUES OF USE
-The face mask should form tight seal on the
patients face while fitting comfortably in the
user’s hands
-The smallest mask is most desirable because
It will cause least increase in dead space
-It is easy to hold and less likely to result in
pressure on Eyes
-There are several methods of holding Mask
12. ONE HAND METHOD
-The Thumb and the index finger of the Left hand
are placed on mask body on opposite sides of
connecter to push it downward to prevent Leak
-The remaining three fingers are placed on the
mandible such that middle finger is applied to
the mentum, the ring finger on the body of the
mandible and little finger at angle of the
mandible to give Jaw thrust anteriorly.
14. TWO HANDED METHOD
TECHNIQUE
-As it requires both hands, a second person is
necessary for assisted or controlled ventilation
-Here thumbs are placed on the either side of the
body of masks, and index fingers are placed
under the angles of the Jaw, Mandible is lifted
and head is extended
-If a leak is present, downward pressure on mask
can be increased by Anesthiologist’s chin on the
mask elbow
15.
16. TWO HANDED METHOD TECHNIQUE
Two Handed Jaw Thrust
One person stands at head end
of patient and performs jaw
thrust with his left hand at angle
of left mandible while his right
hand compresses the reservoir
bag.
The 2nd person stands at the
patients shoulder facing 1st
person. This persons Rt hand
covers the Lt hand of the 1st
person and the Lt hand
Achieves Rt sided Jaw thrust
and mask seal.
17. Claw Hand Technique
-It is useful for children undergoing short
duration procedures.
-The Anesthesiologists stands at the side of
bed facing the child.
-The face mask is applied to face by using the
Rt hand with the palmar surface facing
upwards. The ring & middle finger are
placed under the angle of jaw & the index
finger & thumb encircle the body of mask.
18. Dead space in face masks
-Face mask and its adaptor increase dead
space and it may contribute 30% or more
of tidal volume in neonates and infants.
- It can be reduced by
a- using smaller size mask
b- increasing pressure on mask
c- blowing a jet of fresh gas into the
mask
d- changing the volume of the seal
cushion
19. Advantages
-Lower incidence of sore throat
- requires less anesthetic depth
than tracheal tube or supraglottic
device
- No need of muscle relaxants
- The face mask may be the most cost
efficient method for short cases
20. Disadvantages
-Anesthesiologist`s hands are tied up
-Higher fresh gas flows are often needed
-During remote anesthesia, airway access
is difficult (CT & MRI)
-Often more episodes of oxygen
desaturation & require more
intraoperative airway manipulations
21. Complications
-Skin problems- dermatitis, pressure necrosis
- Nerve injury
-Eye injury – conjunctival chemosis, eyelid
edema & corneal injuries.
-Gastric Inflation
-Latex allergy
- Lack of co-relation between arterial & end
tidal CO2
-Environmental Pollution with anesthetic
gases
-Jaw Pain & User fatigue.
23. Introduction
-A fundamental responsibility of anesthesia
providers is to maintain a patent airway
-The airway passage has a rigid posterior wall
and a collapsible anterior wall consisting of
the tongue and epiglottis
-Under anesthesia the tongue & epiglottis
falls back into the posterior pharynx
occluding the airway
-The purpose of airway is to lift the tongue
and epiglottis away from the posterior
pharyngeal wall & maintain a patent airway
24.
25. Terminology
Artificial airway
-Any device that aims to maintain oral or
nasal air passages. It may be
-Simple supraglottic device such as
oropharyngeal or nasopharyngeal airways.
These may not be sufficient to maintain the
patency of airway on their own and may
require patients jaw to be supported as well
- Augmented supraglottic device such as LMA
and airway management
-Infraglottic devices such as Endotracheal
tubes, tracheostomy tubes, jet ventilation
catheters.
26. Oropharyngeal Airways
Description
-It extends from lips to pharynx, fitting between
tongue and posterior pharyngeal wall. Made up
of elastomeric or plastic materials. Parts are
-Flange it is at buccal end to prevent it from
moving deeper into mouth & may also serve to
fix airway in place.
-Bite Portion it is straight & fits between
teeth or gums
-Curved portion extends backwards to
correspond the shape of tongue & palate.
27. size colour length(cm)
000 Violet 3.5
00 Blue 4.5
0 Black 5.5
1 White 6.5
2 Green 7.5
3 Orange 8.5
4 Red 9.5
5 Yellow 10.5
SIZES
--The American National Standard specifies that the size of
oral airways be designated by a no. i.e the length in cms.
28. SPECIFIC AIRWAYS
Most frequently used airway & has large flange &
reinforced bite portion with tubular channel for air
exchange & suctioning.
--Guedel Airway
29. Cuffed Oro-Pharyngeal Airway
-It is Guedel`s Airway with an inflatable cuff designed to seal the
oropharynx. It has an integral bite block & a 15mm connector for
attachment of the breathing circuit.
-The cuff is inflated with air to displace the base of tongue and
form a low pressure seal with the pharynx & provide an open
airway.
30. Patil-Syracuse Endoscopic Airway
-It was designed to aid fiber-optic intubation. It is
made of aluminum
- It has lateral channels & a central groove on the
lingual surface to allow passage of fiberscope. A slit
in distal end allows fiberscope to be manipulated in
antero-posterior direction but limits lateral movements.
31. Berman Intubating Airway
--It is tubular along its entire length. It is open on one
side so that it can be split & removed from around
a tracheal tube. It can be used as an oral airway or
as an aid to fiberoptic or blind oro-tracheal intubation
32. Williams Airway Intubator
-It is designed for blind tracheal intubation and can also
be used for fiberoptic intubation or as an oral airway.
-It is available in 2 sizes #9 and #10 which will admit
upto 8 and 8.5 mm tracheal tube respectively.
-The proximal half is cylindrical while distal half is
open on lingual surface
33. Ovassapian Fiberoptic Intubating Airway
-Designed for fiberoptic intubation
-Has a flat lingual surface which gradually widens at
distal end and 2 vertical sidewalls at the buccal end
between the sidewalls are a pair of guide walls that
curve towards each other which are flexible and
permit removal of airways around the tracheal tube
- The proximal half is tubular & acts as bite block. The
distal half is opened posteriorly & it can
accommodate tracheal tube upto 9 mm id
34. .
METHODS OF INSERTION
OF AIRWAYS
- Pharyngeal and laryngeal reflexes should be
depressed before an airway is inserted to
avoid coughing and laryngospasm.
- Selecting the correct size of airway is
important correct size is estimated by
holding the airway next to pts mouth & the
tip should be at the angle of mandible
-The best criteria for proper size & position is
unobstructed gas exchange
35. Oral airway may be inserted
in two ways
--The jaw opened with left
hand
--The airway is inserted
with its concave side
towards the upper
lip and when the junction
of bite portion and curved
section is near the
incisors, the airway is
rotated 180° and slipped
behind the tongue in the
final position
36. A tongue blade is used to
push forward and depress
the tongue & the airway
is inserted with concave
side towards the tongue &
is advanced.
Oral airway may be insertion
37. OTHER USES
-May be used to prevent patient
from biting and occluding ETT
-Protect the tongue from biting
-To facilitate suctioning
-To obtain better mask fit
38. NASOPHARYNGEAL AIRWAY
(nasal trumpet/ nasal airway)
-It extends from nose to pharynx, the pharyngeal
end should be below base of tongue but above
the epiglottis.
-A nasal airway is better tolerated than oral airway
if the pt has intact airway reflexes.
Description
-It resembles a shortened tracheal tube with a
flange at the outer end to prevent it from
completely passing into the naris. It is made
up of plastic or rubber and the size is
designated by the inner diameter in mm
40. Nasopharyngeal airway
specific types
-It is plastic with large flange and flat
distal end and is supplied with an
introducer which has balloon on its
tip
-Before insertion, the introducer is
inserted into the airway until the tip
of balloon is just past the end
-Air is injected until the balloon tip
is inflated to approx outer diameter
of tube
-The complete assembly is lubricated
and inserted through the nostril
after it is in place, the balloon is
deflated & introducer is removed
-- Linder Nasopharyngeal Airway
41. Cuffed Nasopharyngeal Airway
It is similar to a short, cuffed tracheal tube. It is inserted
through the nose into the pharynx; cuff inflated & then is
pulled back until resistance is felt.
Binasal Airway
It consists of 2 nasal airways joined together by an adaptor
for attachment to the breathing system. Can be used to
administer anesthesia
42. INSERTION
The diameter of nasal airway should be the
same as needed to insert the tracheal tube
(0.5-1.0mm smaller than oral tracheal tube).
Before insertion it should be lubricated
thoroughly along its entire length. The airway
is held with the bevel against septum & gently
advanced posteriorly while being rotated back
& forth.
45. --A laryngoscope is an instrument used to
view larynx and adjacent structures, most
commonly for introducing tube into the
trachea
--Laryngoscopes can be considered under two
broad headings
--Retractor type like the Macintosh laryngoscope
--Fibreoptic Laryngoscope
Two types are available
- Rigid fiberoptic laryngoscope
- Flexible fiberoptic laryngoscope
Introduction of Laryngoscopes
46. RIGID LARYNGOSCOPES
-These are manufactured either as single piece or
separate detachable blade and handle.
-For detachable handle & blade the light source is
energized when blade & handle are locked in
operating position.
A single piece laryngoscope has a switch on handle
that controls power to lamp.
Rigid Laryngoscope has 2 parts
1. Handle 2. Blade
47. Handle
-The handle is the part that is held in the hand
during use. It provides the power source for the
light. Most often this is from disposable batteries
-Fiberoptic illuminated laryngoscope may use a
remote electrically operated light source.
-The handle is fitted with a hinge pin that fits a
slot on the base of the blade
49. -Although most blades form right angle with
the handle when ready to use. The angle may
also be acute or obtuse.
- An adapter may be fitted between the handle
and the blade to allow angled to be altered.
Howland lock is such an example.
-The patil-syracuse handle can be positioned
and locked in four different positions
(45°, 90°, 135° and 180°)
51. size Intended use
000 Small premature infant
00 Premature infant
0 Neonate
1 Small child
2 Child
3 Adult
4 Large adult
5 Extra large adult
Blade
-The blade is a component that is inserted into the mouth
-When a blade is available in more than one size, blades
are numbered, with the number increasing with size.
52. Parts of the blade include
1. base
2. spatula/tongue
3. heel
4. flange
5. web
6. tip
7. light source
53. Types of Blades
-There are several types of blades which may be
advantageous in some particular situations.
-- Macintosh Blade
--The Macintosh blade is one of the most popular. The tongue has a
smooth, gentle curve that extends from the base to the tip. In cross-
section, the tongue, web. and flange form a reverse 'Z'. Cervical spine
movement is greater compared to the Miller blade.
54. Left handed Macintosh Blade
-- The left handed Macintosh blade has the flange on the opposite
side from the usual Macintosh blade. This blade may be useful for
the abnormalities of right side of the face or oropharynx, left
handed persons, intubating in the right lateral position, and
positioning a tracheal tube directly on the left side of the mouth.
55. Improved Vision Macintosh Blade
The Improved Vision (IV) Macintosh blade is similar to the
standard version except that the mid portion of the tongue
is concave to allow greater visualization of the larynx
56. Polio Blade
-The blade is offset from the
handle at an obtuse angle to
allow intubation of patients in
iron lung respirators or body
jackets
- Patients with obesity, breast
hypertrophy, kyphosis with
severe barrel chest deformity,
short neck, or restricted neck
mobility.
-Disadvantages of this blade are
that little force can be applied
and control is minimal
58. Tull Macintosh
The Tull (suction) blade is a modified Macintosh that has a
suction port near the tip. The suction channel extends next
to the handle and has a finger controlled valves so that
suction can be controlled by the laryngoscopist.
59. Fink Blade
-The Fink blade is another modification of the
Macintosh
-The tongue is wider and has a sharper curve at
the distal end. The height of the flange is
reduced, especially at the proximal end The
light bulb is placed farther forward than on
the Macintosh
60. Bizarri-Guiffrida Blade
-The Bizarri-Guiffrida blade is a modified Macintosh.
The flange is removed except for a small part that
encases the light bulb.
-This was made to limit damage to the upper teeth
-The blade is useful for patients with a limited mouth
opening, prominent incisors, receding mandible, short
and thick neck or anterior larynx
61. Miller Blade
-The tongue is straight with a
slight upward curve near the
tip
-In cross-section the flange
web, and tongue form a C
with the top fattened
-The lamp may be either on
the right or left side of the
blade.
62. Oxiport Miller Blade
=The Oxiport Miller (also
called Miller/port,
oxyscope) blade has a built-
in tube that allows delivery
of oxygen or other gases
during intubation. The tube
also may be used for
suction.
=Insufflation of oxygen
during intubation using this
blade has been found to
decrease oxygen
desaturation in
spontaneously breathing
anesthetized patients
63. Tull Miller Blade
=Tull (suction) Miller blade is a standard
Miller blade with a suction tube
whose port ends near the tip of the
blade
=Near the handle is a finger-controlled
port that allows control of suction
with a finger.
64. Mathews Blade
-The Mathews blade is a straight blade with a wide and
flattened petaloid configuration tip. It is designed for
difficult nasotracheal intubations
65. Alberts Blade
=It has a cut-away flange to increase the visibility.
There is recess to facilitate tracheal tube insertion.
The blade forms a 67° angle with the handle. It is
used for pediatric pts
66. Double Angle Blade
=The spatula of double angle
blade has 2 angulations 20° and
30°, to improve lifting of the
epiglottis.
=The flange has been
eliminated. The bulb is located
at left edge of the blade
between the 2 curvatures. The
blade may be useful for the
patient with anterior larynx.
67. Flexible tip blade
=It has a hinged tip that is controlled by lever
attached to the proximal end of the blade
=When the lever is pushed towards the handle
the tip of the blade is flexed eg: Mc-Coy, flipper
flex tip etc.
=It may improve the chances of successful
intubation by elevating the epiglottis
particularly in case of difficult intubation.
=Less force and less stress response.
69. Techniques of use
positioning the head
The head should be positioned so that the passage-
way to the larynx is brought into a straight line for
the best possible view of the vocal cords.
The optimal position for most patients is flexion of
35° of the lower cervical spine and extension of the
head of 85-90° at the atlanto-occipital level, the so-
called sniffing position.
In children, it may be unnecessary to flex the lower
cervical vertebrae, and in neonates it may be
necessary to elevate the shoulders because the head
is relatively larger.
70. =The laryngoscope handle is held in the left hand.
The fingers of the right hand are used to open the
mouth and spread the lips apart.
=The blade is inserted at the right side of the mouth
=This reduces the likelihood of incisor teeth damage
and helps push the tongue to the left.
=The blade is advanced on the side of the tongue
toward the right tonsillar fossa, so that the tongue
lies on the left side of the blade
=When the right tonsillar fossa is visualized, the tip of
the blade is moved toward the midline. The blade is
then advanced behind the base of the tongue
elevating it, until the epiglottis comes into view.
71. =There are two methods for elevating the
epiglottis, depending on whether a straight or
curved blade is being used.
Straight Blade Technique
=The blade is made to scoop under the epiglottis
and lift it anteriorly. The vocal cords should be
identified.
=If the blade is advanced too far, it will result in
elevation of the larynx as a whole rather than
exposure of the vocal cords.
=The straight blade can also be inserted into the
vallecula and used in the same manner as a
curved blade.
72. Curved Blade
=After the epiglottis is seen, the blade is
advanced until the tip fits into the
vallecula
=Traction is then applied along the handle
at right angles to the blade to carry the
base of the tongue and the epiglottis
forward.
=The glottis should come into view
74. Bullard laryngoscope
RIGID FIBEROPTIC LARYNGOSCOPES
=Description It has a rigid metal
blade specially shaped to follow the
contour of the oropharynx.
=Fiberoptic bundles for illumination &
operatory viewing are housed in a
sheath on the posterior aspect of
blade
=A viewing arm with eye-piece extends
at a 45°angle from the handle. A
video camera can be attached to
eyepiece for remote viewing.
=A working channel extends from the
scope body to the point where the
light bundles end at the tip. It can
be used for suction, oxygen
insufflations, administration of
L.A/ Saline, or jet ventilation
catheter.
75. a-Pediatric
b-Pediatric long – for babies upto 8-10yrs of age
c-Adult version – for children more than 10yrs &
adults
Advantages
=It causes less cervical spine movement than
conventional laryngoscopy.
=Requires mouth opening of only 7mm
=Rapidity of intubation
= Low risk of failed intubation or trauma to
lips or teeth
-- SIZES
--3 sizes are available
76. Disadvantages
=Requires experience & Maintenance of
skills & is somewhat expensive
=Intubation with ETT >7.5mm may be
more difficult.
=Certain ETT & Double lumen tubes will
not fit the stylet
77. Wu-Scope
It combines a rigid, tubular
blade & a flexible fiberscope.
Flexible Fiberscope portion
The fiberscope has short light &
image transmitting fiberoptic
bundles & tip deflection
control
Blade portion
It has 3 detachable metal parts
handle, main blade & bivalve
element.
Details
78. Handle
=cone shaped tube that receives the fiberscope
at the top & connects to the main blade at
the base. The handle to blade angle is 110°
Main blade & bivalve element
are anatomically shaped
When positioned together, they form 2 passage
ways- a larger 1 for suction catheter or ETT
& a smaller 1 for fiberscope.
=An oxygen channel is alongside the slot for
fiberscope
=Different sizes of blades & bivalve element
can be used with the handle
=The adult blade can accommodate ETT upto
8.5mm
79.
80. Advantages
=Can be used successfully to place both
tracheal & double lumen bronchial
tubes in difficult-to-intubate pts
=No forceps or stylet is needed so risk
of airway injury is low
=Suctioning & oxygen administration
can done simultaneously. Unlike
flexible fiberoptic endoscopy, one can
view ETT passing through glottis into
trachea.
=There is no need for head extension or
tongue lifting.
81. Disadvantages
=Requires experience or
Maintenance of skills & high
initial cost
=Time required to assemble or
disassemble the device is more
=Application of cricoids pressure
decreases the ease of intubation
while using this device.
82. Upsher-Scope
Details
=It consists of C-shaped metal
blade, shaped to approximate
curve of oropharynx.
=The distal part has an
upward curve. To the left are 2
tubes that carry fiberoptic
bundles
Proximally the viewing bundle
terminates in the eye piece
while the light bundle makes
contact with the light source at
the handle.. It is presently
available only in adult size
83. Advantages
=It can be used with all size of adult ETT
=Useful to intubate pts with difficult
airways in a reasonable time
Disadvantages
=Not suitable for naso-tracheal intubation
=In routine intubation this scope has
demonstrated no advantages over the
conventional laryngoscopes
84. VIDEO LARYNGOSCOPE
=It can be created by attaching video system to a
flexible laryngoscope or an optical stylet. It has
many advantages
=The displayed anatomy is magnified & a larger
viewing angle is provided
=In addition to allowing laryngoscopy & intubation to
be visualized, it can be used to observe vocal cord
movements after thyroid surgery, verify tracheal
tube position & aid in tracheal tube exchange
= Results in less neck movement than conventional
laryngoscopy
=By projecting the image it makes a good teaching
tool
=Limitations of these device is the need for a video
endoscopy monitoring system
85. =It has Macintosh blade
attached to the handle
= The image- light bundle
is threaded through a
small guide in the blade
& advanced 2/3 of the
length of the blade
=Provides better view
than traditional
Macintosh Laryngoscope
VIDEO MACINTOSH
INTUBATING LARYNGOSCPE
86. GLIDE SCOPE
Description
=It has a miniature digital camera
underside of a plastic blade.
A light emitting device( LED)
mounted beside the camera
provides illumination
=Blade has a 60° bend at the
midpoint The Glide Scope is
available in adult & pediatric
sizes
=Studies shows this scope yields
a comparable or superior glottis
view compared to direct
laryngoscopy
=It causes less cervical
movements than Macintosh
blade
87. FLEXIBLE FIBEROPTIC
LARYNGOSCOPE
Introduction
=The 1st use of flexible fiberoptic technology in the
airway management can be credited to Dr P Murphy
who in 1967 invented Choledochoscope for intubation
of trachea
=The flexibility of these devices means that they can
be made to follow virtually any anatomical space to
return an image of the objective. The technique of
flexible fibreoptic intubation has become a gold
standard for management of difficult laryngoscopy.
=This has tended to create the impression that the
technique is solution to all difficult airways, but this
is patently not so.
88. Principles & Design
=The pathway through which the illumination &
the image pass consists of thousands of very fine
glass fibers, each typically of 10µm in diameter
Each fiber consists of a central glass core surrounded
by a thin cladding of another type of glass with
different refractive index. As a result of difference
in refractive indices, light entering the glass fiber
undergoes total internal reflection along the length
of the fiber to emerge at the other end
For image transmission, the arrangement of fibers
relative to one another must be identical at either
end of the bundle, as each fiber carries tiny portion
of the overall image. This is called a COHERENT
BUNDLE.
89.
90. Parts
=It is composed of several parts, a light source, handle &
flexible insertion portion.
The handle is the part to be held in hand during use. It
includes the eyepiece, focusing ring, working channel port &
the tip control lever.
The insertion tube carries
2 light bundles
1 image bundle
1 working or biopsy channel
2 angulation wires, which control the more flexible tip of
device
91. = These are held together with a
stainless steel spiral wrap followed
by a stainless steel braid & covered
by waterproof material, to give a
rigid cross section while allowing
overall flexibility. The fiberscope is
axially rigid to twisting forces thus
rotation of the control handle results
in similar rotation of the tip.
=The fiberscope uses a powerful
external cold light source so that the
tissues are not damaged by radiant
heat.
92. Usage
=The intubating fiberscopes are
available with insertion tubes ranging
in size from 2.5mm external diameter
to over 6mm with proportionately
larger working channels
=A device with external diameter of
3.5mm is optimal for use in adults.
The advent of the miniature video
camera has revolutionized fiberoptic
endoscopy
93. Fiberoptic Intubation
=It can be performed via oral or nasal route
=The fiberscope is inserted & advanced behind
the tongue & into the larynx. Once in the
trachea, a tracheal tube which has been
previously loaded onto the fiberscope is
advanced off the scope into the trachea &
then the fiberscope is removed.
=Oral intubation is considered more difficult
than nasal intubation
94. Advantages
=Reliable approach to the difficult airway management.
=Facility to record images for review & documentation.
=Improvements in teaching techniques.
Disadvantages
=More expensive & fragile & difficult to use than rigid
laryngoscope.
=Requires more time than rigid laryngoscopy.
=Requires considerable experience & skill Maintenance
=Laryngeal trauma may occur.
95. Complications of laryngoscopy
1. Dental injury
2. damage to soft tissue and
nerves
3. Injury to cervical spinal cord
4. Circulatory changes
5. Swallowing or aspiration of
foreign body.
6. Shock or Burns
7. Laryngoscope malfunctions
8. Disease transmission
9. TMJ Dislocation.
97. DESCRIPTION
=An endotracheal tube is one
through which anesthetic
gases and respiratory gases
are conveyed into and out of
the trachea.
=It has tracheal & machine
end
=The bevel is defined as the
slanted part of the tube at
the tracheal end.
=When, an opening in the
tube is present on the
opposite side of the bevel
it is designated as Murphy’s
tip endotracheal tube.
98. REQUIREMENTS
OF AN IDEAL ET TUBE
1) Inertness
2) Smoothness of outer surface to avoid
damage to mucosa
3) Inner surface should be smoothed and
non-wettable to prevent building of
secretions.
4) Non-inflammable
5) Transparent
6) Easily sterilized
7) Non kinking
99. 8) Sufficient strength to allow thin wall
construction
9) Thermo plasticity to confirm to anatomic
passage and to be self centering within the
trachea.
10) Non reactive with lubricants or anesthetic
agents
11) Latex free
12) Non injurious catheter tip
Currently used tubes are manufactured from
synthetic rubber, plastic materials and
silicone.
100. GENERAL PRINCIPLES
RESISTANCE AND WORK OF BREATHING
=A tracheal tube places a mechanical burden on
spontaneously breathing patient. It is a source of more
resistance and is an important factor in determining the
work of breathing. The factors determining the
resistance include:
1) Internal diameter: The tube with thick wall
decreases the ID and thereby increases the resistance
and vice-versa.
2) Length: Decreasing the length of the tube decreases the
resistance.
3) Configuration: Abrupt change in the diameter and
direction increases the resistance. Gentle curve
connectors offer less resistance than right-angled ones as
there will be increased resistance because of turbulent
flow of gases.
101. DEAD SPACE
=The volume of tracheal tube and connector is
usually less than that of the natural passage.
=Dead space is normally reduced by intubation.
=In pediatric patients long tubes and
connectors may increase the dead space
considerably.
102. STANDARD MARKING OF ETT
=The markings are situated on the
bevel side above the cuff & are
read from pt end to machine end
=Type of the tube: Oral or nasal or
oral/nasal Size: ID in mm
=External diameter may also be
indicated.
=Manufacturer's name or trade
mark
103. =Tube has Graduated markings, showing the
distance in cms from the patient end.
=Precautions are usually noted: Disposable/Do
not reuse
=Implantation tests (IT) or Z-79 indicating the
tube has been tested for tissue toxicity &
accomplish ANSI standard.
=Opaque lines may also be included at the
patient end or along full length.
104. CUFF SYSTEM
=A Cuff system consists of the cuff & inflation
system.
=The purpose of cuff system is to provide a
seal between tube & tracheal wall to
prevent passage of pharyngeal contents into
the trachea & ensure no gas leaks past the
cuff.
=The cuff also serves to center the tube in
trachea.
105. INFLATION SYSTEM
1)Inflation valve: When a syringe tip is inserted, a
plunger is displaced from its seat & gas can be
injected into the cuff. Upon removal of the syringe
the valve seals and gas cannot escape.
2) Pilot balloon: Its function is to give an indication of
inflation or deflation of the cuff and a rough idea of
the cuff pressure.
106. 3) External inflation tube: The standard
specifies ,
a)The external diameter should not
exceed 2.5mm
b) The inflation tube should be attached
to the ETT at a small angle.
c) The tube should extend at least 3cm
beyond the machine end of the ETT
before a pilot balloon or inflation
valve is incorporated.
4) Inflation lumen: This connects the inflation
tube to the cuff. It is located within the wall
of the tracheal tube.
107. CUFF
The cuff is a inflatable sleeve near
the patient end of ETT.
The cuff material should be strong
and tear resistant but thin, soft
and pliable.
Cuffs are usually made of the
same material as the ETT.
108. CUFF PRESSURE
Intracuff pressure and pressure
on tracheal wall
=It is desirable that cuff seals the airway
without extending so much pressure on the
trachea so that its circulation is not
compromised or trachea is dilated.
=When cuff is inflated in the trachea, there
is a linear decrease in the perfusion. It is
reduced when the lateral pressure on the
tracheal wall attains 30cm of H20 (22mmHg)
and ceases completely at 50cm of H20
(42mmHg).
=So it is recommended that the pressure on the
lateral tracheal wall should be kept between
25-34cm of H20.
109. Intracuff pressure and
use of nitrous oxide
The resting intracuff pressure
and volume of the cuff inflated
with air rise during nitrous oxide
anesthesia, which results in
ischaemia of the tracheal
mucosa or compression of the
tube, and increase in volume
may lead to cuff herniation.
110. to prevent the increasing pressure
includes the following
a) Filling the cuff with gas mixture to be used for
anesthesia or saline
b) Fitting the cuff system with pressure relief
valve or pressure regulating devices.
c) Use of special system e.g. Lanz pressure
regulating valve, sponge cuff and special tubes
like Brandt tubes.
d) Monitoring cuff pressure and deflating the cuff
as needed.
--Steps
111. Types of cuffs
Low volume high pressure cuff
= They have a small diameter at rest and low residual
volume.
=It requires a high intra-cuff pressure to achieve a
seal
with the trachea.
=It has a small area of contact with the trachea and
distends and deforms the trachea to a circular
shape.
=The intra-cuff pressure and lateral pressure on the
tracheal wall increase sharply as increments of air
are added to the cuff. For this reason use of largest
ETT has been advised, so that the cuff will be
minimally inflated when a seal is created
a) Low volume high pressure
b) High volume low pressure
112. Advantages
=Usually reused,
=Less expensive
=Offers better protection against aspiration
=Better visibility during intubation than low
pressure cuffs.
=There is also low incidence of sore throat.
Disadvantages
=The most serious risk associated with
these cuffs is ischemic damage to the
trachea following prolonged use.
113. High volume low pressure cuffs
=They have a high resting volume, large diameter
and a thin competent wall.
=As it is inflated, it first touches the trachea at its
narrow point. As cuff inflation continues the area
of contact become larger and the cuff adopts
itself to irregular tracheal surface.
=If the cuff inflation is continued the area in
contact will be subjected to increasing pressure
and trachea will be distorted to a circular cross
section similar to high pressure cuff.
114.
115. Advantage
=It is relatively easy to pass devices such as esophageal
stethoscope, temperature probes, nasogastric tubes
around a low pressure cuffs
Disadvantage
=These tubes are more difficult to insert,
=Obscure the view of the tube tip and larynx
=The cuff is more friable and thus more likely to
be torn during intubation.
=Incidence of sore throats is greater with these
tubes.
=It may not effectively prevent fluid leakage into
lower airway.
116. THE GUIDELINES TO DETERMINE
THE SIZE OF ETT
Ideal tube in average Adult male = 8.5mm ID
Ideal tube in an average Adult female = 7.5mm ID.
Age is recognized as the most reliable indicator of
appropriate ETT size for children.
3 months & less ------ 3 mm ID
3 - 9 months ------ 3.5 mm ID
Older than 1 year ------ ID in mm
= (16 + age in years)/4
117. Younger than 6 years ---3.5 + age in years/3
= ID in mm
Older than 6years ---- 4.5 + age in years / 4
= ID in mm
Infants below 1kg ----- 2.5mm
Infants 1-2 kg ------ 3.0 mm
Infants 2-3 kg ------ 3.5 mm
Infants 3 kg ---------- 4.0 mm
Choosing a tube whose external diameter is
same width as the patient's distal little finger
118. Patients age Size( mm ID)
Premature neonate 2.5 to 3.0
Full term neonate 3.0 to 3.5
3 months to 1 year 4.0
2 years 4.5
4 years 5.0
6 years 5.5
8 years 6.0
10 years 6.5
12 years 7.0
Other recommendations for size of ET tubes
119. DEPTH OF INSERTION
The tube should be in the middle third of the
trachea with the head in natural position.
The following calculations can be used.
1) Length in cm = age/2 + 12
2) Length in cm = weight in kg/5 + 12
3) Length in cm = height in cms/10 + 5
4) Length in cm = 3 × ID (mm)
120. In adults, the tube should be passed until the
cuff is 2.25 to 2.5cm below the vocal cord
In average size adult patients, securing the
tube at the anterior incisor at 23cm in males
and 21cm in females will usually avoid
endobronchial intubation.
For nasal intubation 5cm should be added to
these length for positioning at the nares.
121. SPECIFIC ET TUBES
Cole tube
=It is uncuffed ETT, designed for
pediatric patients. The patient end is
smaller in diameter than the rest of
the tube
=They are sized according to the
internal diameter of the tracheal
portion. It ranges from 2mm to 5mm..
=It is recommended for neonatal
resuscitation but not for long term
intubation.
=Disadvantage of this tube is that it
cannot be used nasally
122. Spiral embedded tubes
=Also known as Armored tube.
=These tubes have a metal or
nylon spiral woven reinforcing
wire covered both internally
and externally by rubber, PVC
or silicone.
=A stylet is often needed for
intubation.
=These tubes are esp useful in
situations where the tube is
likely to be bent or compressed
as in head & neck surgery
123. Advantages
----Primary advantage of tube is resistance to
kinking and compression.
--The portion of the tube outside the patient
can be easily angled away from the surgical
field without kinking
-- Can be used for patients with Tracheostomies
Disadvantages
----Tube may rotate on the stylet during intubation.
-- Insertion through nose & intubating LMA is difficult.
-- Fixation of these tubes are more difficult.
--If the patient bites the tube it will cause permanent
deformity resulting in obstruction of the tube.
124. =It is preformed to facilitate the
head & neck surgeries.
=The tubes are available in cuffed
uncuffed ,nasal and oral version
=There is a preformed bend in the
tube that may be temporarily
straightened for suctioning.
=Each tube has a rectangular mark
at the center of the bend.
=Distance from this mark to the
distal tip is printed on each tube.
Preformed tubes/Ring-Adair-Elwyn (RAE)
125. =Oral RAE tubes are shorter than the nasal ones. The external
portion is bent at an acute angle so that when in place it rests
on patients chin & the connector over the pts chest.
=Nasal RAE has a curve opposite to the curvature of the oral
tube, so that when in place the outer portion of the tube is
directed towards patient's forehead
126. Advantages
=Easy to secure and reduce the risk of
unintended extubation.
=Breathing system remains away from
surgical field
Disadvantages
=It offers more resistance than conventional
tubes.
=Suctioning is difficult
127. Laryngectomy tube
=Designed for insertion into a
tracheostomy site.
=The tube is preformed in a J
configuration at the pt end
=This allows the part of the
tube external to the patient
to be directed away from
the surgical field.
=The tip may be short and/or
without a bevel to avoid
inadvertent advancement
into a bronchus.
128. Injectoflex Tube
=It is used for laryngeal microsurgery.
=It is a short cuffed silicone tube designed to
be placed below the vocal cords.
=The tube has an embedded wire spiral to
prevent kinking and compression. The cuff
inflation lumen and the inflation tube are
integrally joined in a sheath with a malleable
introducer.
129. Microlaryngeal tracheal surgery tube
It is available with an ID of 4, 5 or 6mm,
each of which has the same length and cuff
diameter as a standard 8mm ID tube.
Designed for microlaryngeal tracheal
surgery.
The small diameter provides better surgical
access
The problems with this tube are incomplete
exhalation & occlusion.
130. Endotrol tubes
A tip control system is incorporated
in this tracheal tube to change the
direction of the tip.
This tube has a ring loop at the
machine end that is connected to
the tip with a cable
Pulling the ring causes the tip of the
tube to move anteriorly.
This is useful when laryngoscopy is
difficult and only epiglottis is seen.
131. Tubes with extra lumens
Tubes are available with one or more separate lumens terminating near
the tip.
They are useful for respiratory gas sampling, airway pressure
monitoring, injection of fluids and drugs & jet ventilation.
132. The LITA (Laryngotracheal
Instillation of Topical anesthesia)
Has additional small bore channel
within the concave surface of the
tube.
10 small holes at the distal 13cm
of the tube allows the injected
medication to be spread both
above & below the cuffs.
This can provide a smooth
emergence from anesthesia
without coughing in most cases.
133. EMG Reinforced Tube
This tube is designed to
monitor recurrent laryngeal
nerve electromyogram
activity during surgery.
The tube is wire-reinforced &
has 4 stainless steel
electrodes above the cuff.
The electrodes are connected
to a monitor.
134. Laser-shield II Tracheal Tube
It is designed for use with CO2 and
KTP lasers.
Made from silicone with an inner
aluminium wrap and a smooth Teflon
outer coating.
The cuff is not laser resistant &
contains methylene blue crystals.
It should be inflated with water or
saline solution.
135. Cottonoids for wrapping around the cuff are supplied
with each tube. These must be moistened and kept
moist during the entire procedure
Disadvantage
Exposure of unprotected parts of the tube proximal &
distal to cuff can result in rapid combustion.
The methylene blue crystals may not fully dissolve &
may obstruct the pilot tube, making it impossible to
deflate.
136. Laser-Flex Tubes
Laser flex tube is a flexible
stainless steel tube with a smooth
surface designed for used with
C02 and KTP
(potassium-titanyl-phosphate)
lasers.
Adult version has two PVC cuffs
and PVC tips with Murphy eye.
The tube cuffs are inflated by two
separate inflation tubes. The
distal cuff can be used if proximal
one is damage by laser.
137. The cuff should be filled with saline. The distal
cuff should be filled first until sealing occurs,
then the proximal cuff is filled with saline
coloured with methylene blue.
Problems
with the laser flex tube include
stiffness, roughness, cannot be trimmed.
The double cuff adds to the time of intubation
and extubation.
138. Laser-tubes
This is made of white rubber & has
a cuff-within-a cuff design.
If the outer cuff is perforated by a
laser beam, the trachea will still be
sealed by inner cuff. The inner cuff
is filled with air & outer with water
or saline.
The shaft above the cuff is covered
by a corrugated silver foil, which is
covered by merocel sponge that
should be moistened with saline
before use.
It is recommended for use with
argon, NdYAG, CO2 lasers.
139. Hi-Lo Evac Tube
It incorporates dedicated
channel which can be used to
clear secretions below the
vocal cords but above the
cuff.
The lumen may be blocked by
secretions.
140. Hi-Lo Jet Tube
It is an uncuffed tube with
additional lumen that can be
used for Jet ventilation,
monitoring airway pressure,
sampling respiratory gases or
administering local