Introduction
to Flow
measurement
VPInstruments
2
Normal cubic meter (m3
n)
Gases do have a certain mass, but the density of gases is strongly dependent on pressure and temperature
Normal conditions: defines the mass of a gas at a fixed temperature and pressure.
3
Normal cubic meter (m3
n)
• Volume: 1m3
• Temperature: 0°C
• Pressure: 1013,25 mbar
• Relative humidity: 0%
• Mass: 1293 gram
m3
n
4
The ideal gas law
Pressure * Volume / Temperature = Constant
P∗V
T
=nm∗R
V
½ V
P
2P
5
Normal conditions may vary
Other common normal conditions:
• 20˚C reference (FAD): See ISO 1217
• 15˚ reference: See DIN 1533
• CAGI: 68 F, 14.5 PSIA, 0% water vapor pressure
Nm3/hr = V ∗ (TN / T) ∗ (P / PN)
6
From FAD to normal cubic meter
100 m3/hr FAD ∗ 273,16/(273,16+20) ∗ 1000/1013,25 = 91,95 Nm3/hr
FAD/Nm3 factor = 1,0874 → 8,7% difference when comparing numbers
20 deg
1000
20 deg
1000
20 deg
1013,25
20 deg
1000 0 deg
1013,25
FAD Nm3
7
Thermabridge™ technology
Core technology of the VPFlowScope®
8
ThermabridgeTM technology
Vout = k ∗ λ ∗ ρ ∗ v ∗ (Ts−Tg)
• Vout = output voltage
• k = sensor (geometrical) constant
• λ = thermal conductivity of the gas
• ρ = density of the gas
• v = actual velocity in in m / sec
• Ts = sensor temperature
• Tg = gas temperature
9
Normal meter per second
Thermal mass flow: velocity in Normal meter per second
Rule of thumb:
“Actual velocity times absolute pressure” (forget temperature)
Example: 8 m/sec∗ 7 bar abs ~ 56 mn/sec
10
Principles compared
Thermal Vortex
DP –
Orifice plate
DP –
Cone meter
Coriolis
Turbine/
rotary
displacement
Clamp on
ultrasonic
Mass flow Yes Optional Optional Optional Yes Optional Optional
Meter run 20D 15D 15D 5D 0D 10D 20D
Pressure loss Low Medium/high high high Low Low Low
Dirty air Fouling OK Clogging Clogging Internal fouling Faillure OK
Wet Air Spikes OK, spikes OK OK, orientation Yes, but affects
reading
Faillure Spikes
Range 1:250 1:10 1:10 1:10 1:100 1:100 1:100
Accuracy 2% 2% 2% 2% 0.5 .. 1% 0.5…1 % 1%
Purchase price $ $ $ $ $$$$ $$ $$$
Maintenance Medium Low Medium Medium Low High Low
11
Value of mass flow and 3-in-1
We believe in: Pressure ∗ Flow = Voltage ∗ Current – Energy!
• Flow, Pressure, Temperature combined
• ISO 11011 (e.g. must measure pressure on multiple locations)
• Difference between supply and demand
• How to detect cause of a pressure drop if you dont measure?
• Internal accounting: 7 bar is more expensive than 6 bar
• Flow & temperature to identify issues with compressed
air driers and (water) cooled compressors
12
Velocity to mass flow
True diameter Measured error Measured diameter % of error
50 3 53 12.36
100 3 103 6.09
200 3 203 3.02
300 3 303 2.01
Check the inner pipe diameter
Area A = π∗ 0.25 ∗ D2
Wrong entry of D → big deviation
Example:
V = 60mn/sec
D: 50 mm → Q = 0,117 m3
n / sec
D: 54 mm → Q = 0,137 m3
n / sec
Difference: 17%
13
Sensing direction
14
Thermal mass flow = pressure independent
Why? The sensor counts molecules:
More pressure = denser air* = more molecules = more heat loss*
* : linear with pressure
2 x P
1 x P
1Q 2Q
15
Differential pressure technology
Dynamic
pressure port
Positive side
Dynamic pressure port
Negative side
Temperature
FLOW
VPFlowScope
Transmitter housing
- Absolute pressure sensor
- Differential pressure sensor
Locking ring
16
Flow range is critical
Thermal vs DP technology: 1:300 range vs 1:10 (1:5) range
Thermal:
• From leakages to high flow
• Dry air only
• Temp range up to ~60°C
DP:
• Only medium to high flow rates
• Not suitable for leakage/low flows
• Wet air and dry air
• Temp range up to ~150°C (or higher)
17
Bi-directional flow: air can go both ways!
18
Importance of bi-directional flow measurement
Demand side (reverse flow)
Tank
Production line
+1002 mn
3
+941 mn
3
- 61 mn
3
____________
+880 mn
3
Flow
Flow
POU tank
Production line
Production line
19
Bi-directional flow examples
• Ring networks
• Multiple compressor rooms
• Complex (old) compressed air networks
• Large receiver tanks
• Non return valves in compressors
• Leaking drains and seals
Extremely useful information for audits and permanent installations.
20
Turbulent vs laminar flow
Reynolds number
ρ∗v∗D
η
=
density∗velocity∗tubediameter
dynamic viscosity
Air dynamic viscosity = 17,1 ∗ 10 −6, which is very
small compared to water or syrup
• Compressed air lines: Flow is always turbulent!
• Turbulent: Velocity profile is nearly flat:
Less sensitive for insertion depth
21
Flow meter
selection
22
Know the basics: Process data
Key to correct measurements and to prevent damage to your flow meter
• Type of gas
• Flow range
• Humidity (dry/saturated)
• (Inner) Diameter
• Pressure
• Temperature range
23
Model selection
In-line meters: 0.5, 1, 2 inch
Flow range leads direct to model,
Check the diameter to ensure quick mechanical installation.
Insertion probes: 2 inch and up
The (normalized) velocity range needs to be calculated from the flow
range to determine the right model.
Normalized velocity = mass flow [m3
n/hr] / [3600] / tuber area
24
½” 1” 2” > 2”
Insertion probes
In-line meters
Size
(inch)
Min flow
m3
n/hr
Max flow
m3
n/hr
Min flow
SCFM
Max flow SCFM
½ inch 0,32 80 0.19 50
1 inch 0,88 250 0,52 150
2 inch 3,53 1000 2,06 600
Flow ranges
25
Flow ranges
Thermal: Schedule 40 Standard Seamless Carbon Steel Pipe
Size
(inch)
DN
ID
(inch)
ID
(mm)
Min flow
SCFM
Max flow
SCFM
Min flow
m3
n/hr
Max flow
m3
n/hr
2 50 2,1 52,5 2,3 688 3,9 1,169
4 100 4,0 102,3 8,7 2,610 15 4,435
6 150 6,1 154,1 20 5,924 34 10,065
12 300 11,9 303,2 77 22,953 130 38,995
20 500 18,8 477,8 190 56,996 323 96,832
½” 1” 2” > 2”
Insertion probes
In-line meters
26
Not sure? Use VPCalculator
Use our calculator to determine which flow
sensor you need.
See online Velocity Calculator
https://www.vpinstruments.com/selection-
guide/velocity/
27
When creating a project quote
Gather enough information to be prepared
• P&ID (Process and Instrumentation Diagram)
• Plant map, to scale
• Pictures of installation points
• Use the on-line demand side table or Excel form
• Use bi-directional as default option
28
Flow meter
installation
29
Read the manual…
• User manual
• VP Academy
• Instruction video
• This training
• Helpdesk “how to” articles
30
Check the direct environment
• Avoiding excessive external heat
• Protect against external water damage (IP)
• Eliminating excessive system condensate
• When having condensation in gas
(100% condensation → VPS DP)
• Avoid corrosive atmosphere where possible
• Use the right cables and cable locations
• Eliminate mechanical vibration and potential
danger
31
Check air quality
• Water
• Oil
• Drains
• Particles
• (Cone strainer)
32
Check temperature
• Thermal mass: limited range: high temperature will cause sensor signal → 0
• Temperature fluctuations: lagging temperature compensation
• Coming from cold outside → inside (audits) : allow 1 hour acclimatization for
best results.
• DP: high range, and reasonably fast response
33
Installation location: Avoid excess water
Piping not in scale
34
2. Find a straight pipe
General minimum rule:
• 20*D upstream length (even 40*D preferred)
• 5*D downstream length (10*D preferred)
The longer the better
Without conditioners:
All other claimed shorter lengths are B.S.
It’s based on physics, not technology
35
3. Use the best orientation
Take into account:
• Ease of wiring, maintenance, read out of and
access to display
• Install at least at an angle of 15° upwards. ALWAYS if possible
• Piping table! Minimum up- and downstream length.
Longer = better
• In case of doubt: Communicate with the end user
30o
36
Recap
• Thermal mass : dry air
• Differential pressure : wet air
• In Line : 0,5, 1 and 2 inch
• Probes : > 2 inch
• Always check : Temperature, Pressure
diameter and range
• Ring network : Bi – directional
• General : Use the checklist!
Things to avoid:
• Large and fast temperature swings
• Excessive condensate in the pipe
• Short meter runs
37
www.facebook.com/vpinstruments/
www.linkedin.com/company/vpinstruments
THANK YOU!
VPInstruments
info@vpinstruments.com
www.vpinstruments.com

More Related Content

Similar to Introduction into Flow Measurment Technology_2021_Dec6.pdf

Condumax W CLS19 endress+hauser datasheet-conductivity sensors
Condumax W CLS19 endress+hauser datasheet-conductivity sensorsCondumax W CLS19 endress+hauser datasheet-conductivity sensors
Condumax W CLS19 endress+hauser datasheet-conductivity sensors
ENVIMART
 
1.)(485 pages) C&I Measurements and Conversions.docx
1.)(485 pages) C&I Measurements and Conversions.docx1.)(485 pages) C&I Measurements and Conversions.docx
1.)(485 pages) C&I Measurements and Conversions.docx
ArjunSen24
 
Basics on Cooling Tower & Water.pdf sana parveen
Basics on Cooling Tower & Water.pdf sana parveenBasics on Cooling Tower & Water.pdf sana parveen
Basics on Cooling Tower & Water.pdf sana parveen
Sana Khan
 
Final presentation
Final presentationFinal presentation
Final presentation
Asma-ul Husna
 
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensorsCondumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
ENVIMART
 
CCS 120 endress+hauser datasheet-sensor for total chlorine
CCS 120 endress+hauser datasheet-sensor for total chlorineCCS 120 endress+hauser datasheet-sensor for total chlorine
CCS 120 endress+hauser datasheet-sensor for total chlorine
ENVIMART
 
STACK ANALYSIS MONITORING STACK S AMPLING
STACK ANALYSIS MONITORING STACK S AMPLINGSTACK ANALYSIS MONITORING STACK S AMPLING
STACK ANALYSIS MONITORING STACK S AMPLING
SarangBonde2
 
Hedland Variable Area Flow Meters
Hedland Variable Area Flow MetersHedland Variable Area Flow Meters
Hedland Variable Area Flow Meters
Bell Flow Systems
 
Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13
Edress Hauser: Flow Meter, Level, Pressure, Temperature
 
Vortex vn2000 datasheet badger meter flow meters hot tap insertion meter
Vortex vn2000 datasheet badger meter flow meters hot tap insertion meterVortex vn2000 datasheet badger meter flow meters hot tap insertion meter
Vortex vn2000 datasheet badger meter flow meters hot tap insertion meter
ENVIMART
 
Tp+manual
Tp+manualTp+manual
Tp+manual
Jeferson Gomez
 
Vortex flow meter manual
Vortex flow meter manualVortex flow meter manual
fanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdffanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdf
BlentBulut5
 
FAN MODULE.pdf
FAN MODULE.pdfFAN MODULE.pdf
FAN MODULE.pdf
NITIN ASNANI
 
Vortex Flow Meters- Badger Meter
Vortex Flow Meters- Badger MeterVortex Flow Meters- Badger Meter
Vortex Flow Meters- Badger Meter
Bell Flow Systems
 
KEMET Webinar -C44U_C44P-R Power Can Film Capacitors
KEMET Webinar -C44U_C44P-R Power Can Film CapacitorsKEMET Webinar -C44U_C44P-R Power Can Film Capacitors
KEMET Webinar -C44U_C44P-R Power Can Film Capacitors
Ivana Ivanovska
 
Vapour pressure-model-1354 atac
Vapour pressure-model-1354 atacVapour pressure-model-1354 atac
Vapour pressure-model-1354 atac
European Tech Serv
 
How to select the most suitable flow meter for your application's gas flow me...
How to select the most suitable flow meter for your application's gas flow me...How to select the most suitable flow meter for your application's gas flow me...
How to select the most suitable flow meter for your application's gas flow me...
TIPL5
 
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsCondumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
ENVIMART
 

Similar to Introduction into Flow Measurment Technology_2021_Dec6.pdf (20)

Condumax W CLS19 endress+hauser datasheet-conductivity sensors
Condumax W CLS19 endress+hauser datasheet-conductivity sensorsCondumax W CLS19 endress+hauser datasheet-conductivity sensors
Condumax W CLS19 endress+hauser datasheet-conductivity sensors
 
1.)(485 pages) C&I Measurements and Conversions.docx
1.)(485 pages) C&I Measurements and Conversions.docx1.)(485 pages) C&I Measurements and Conversions.docx
1.)(485 pages) C&I Measurements and Conversions.docx
 
Basics on Cooling Tower & Water.pdf sana parveen
Basics on Cooling Tower & Water.pdf sana parveenBasics on Cooling Tower & Water.pdf sana parveen
Basics on Cooling Tower & Water.pdf sana parveen
 
Final presentation
Final presentationFinal presentation
Final presentation
 
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensorsCondumax CLS15_15D-endress+hauser datasheet-conductivity sensors
Condumax CLS15_15D-endress+hauser datasheet-conductivity sensors
 
CCS 120 endress+hauser datasheet-sensor for total chlorine
CCS 120 endress+hauser datasheet-sensor for total chlorineCCS 120 endress+hauser datasheet-sensor for total chlorine
CCS 120 endress+hauser datasheet-sensor for total chlorine
 
STACK ANALYSIS MONITORING STACK S AMPLING
STACK ANALYSIS MONITORING STACK S AMPLINGSTACK ANALYSIS MONITORING STACK S AMPLING
STACK ANALYSIS MONITORING STACK S AMPLING
 
Hedland Variable Area Flow Meters
Hedland Variable Area Flow MetersHedland Variable Area Flow Meters
Hedland Variable Area Flow Meters
 
Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13Conductivity Sensor for High-Temperature Applications W CLS 13
Conductivity Sensor for High-Temperature Applications W CLS 13
 
Vortex vn2000 datasheet badger meter flow meters hot tap insertion meter
Vortex vn2000 datasheet badger meter flow meters hot tap insertion meterVortex vn2000 datasheet badger meter flow meters hot tap insertion meter
Vortex vn2000 datasheet badger meter flow meters hot tap insertion meter
 
Tp+manual
Tp+manualTp+manual
Tp+manual
 
Vortex flow meter manual
Vortex flow meter manualVortex flow meter manual
Vortex flow meter manual
 
fanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdffanmodule-230428052900-38a217b0.pdf
fanmodule-230428052900-38a217b0.pdf
 
FAN MODULE.pdf
FAN MODULE.pdfFAN MODULE.pdf
FAN MODULE.pdf
 
Vortex Flow Meters- Badger Meter
Vortex Flow Meters- Badger MeterVortex Flow Meters- Badger Meter
Vortex Flow Meters- Badger Meter
 
KEMET Webinar -C44U_C44P-R Power Can Film Capacitors
KEMET Webinar -C44U_C44P-R Power Can Film CapacitorsKEMET Webinar -C44U_C44P-R Power Can Film Capacitors
KEMET Webinar -C44U_C44P-R Power Can Film Capacitors
 
Vapour pressure-model-1354 atac
Vapour pressure-model-1354 atacVapour pressure-model-1354 atac
Vapour pressure-model-1354 atac
 
DTT31
DTT31DTT31
DTT31
 
How to select the most suitable flow meter for your application's gas flow me...
How to select the most suitable flow meter for your application's gas flow me...How to select the most suitable flow meter for your application's gas flow me...
How to select the most suitable flow meter for your application's gas flow me...
 
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensorsCondumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
Condumax w CLS 21 D-endress+hauser datasheet-conductivity sensors
 

Recently uploaded

Standard Reomte Control Interface - Neometrix
Standard Reomte Control Interface - NeometrixStandard Reomte Control Interface - Neometrix
Standard Reomte Control Interface - Neometrix
Neometrix_Engineering_Pvt_Ltd
 
Chat application through client server management system project.pdf
Chat application through client server management system project.pdfChat application through client server management system project.pdf
Chat application through client server management system project.pdf
Kamal Acharya
 
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Dr.Costas Sachpazis
 
Automobile Management System Project Report.pdf
Automobile Management System Project Report.pdfAutomobile Management System Project Report.pdf
Automobile Management System Project Report.pdf
Kamal Acharya
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
AJAYKUMARPUND1
 
Cosmetic shop management system project report.pdf
Cosmetic shop management system project report.pdfCosmetic shop management system project report.pdf
Cosmetic shop management system project report.pdf
Kamal Acharya
 
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdfHybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
fxintegritypublishin
 
Forklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella PartsForklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella Parts
Intella Parts
 
Democratizing Fuzzing at Scale by Abhishek Arya
Democratizing Fuzzing at Scale by Abhishek AryaDemocratizing Fuzzing at Scale by Abhishek Arya
Democratizing Fuzzing at Scale by Abhishek Arya
abh.arya
 
weather web application report.pdf
weather web application report.pdfweather web application report.pdf
weather web application report.pdf
Pratik Pawar
 
Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024
Massimo Talia
 
power quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptxpower quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptx
ViniHema
 
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
ydteq
 
Scaling in conventional MOSFET for constant electric field and constant voltage
Scaling in conventional MOSFET for constant electric field and constant voltageScaling in conventional MOSFET for constant electric field and constant voltage
Scaling in conventional MOSFET for constant electric field and constant voltage
RCC Institute of Information Technology
 
Courier management system project report.pdf
Courier management system project report.pdfCourier management system project report.pdf
Courier management system project report.pdf
Kamal Acharya
 
ASME IX(9) 2007 Full Version .pdf
ASME IX(9)  2007 Full Version       .pdfASME IX(9)  2007 Full Version       .pdf
ASME IX(9) 2007 Full Version .pdf
AhmedHussein950959
 
Top 13 Famous Civil Engineering Scientist
Top 13 Famous Civil Engineering ScientistTop 13 Famous Civil Engineering Scientist
Top 13 Famous Civil Engineering Scientist
gettygaming1
 
road safety engineering r s e unit 3.pdf
road safety engineering  r s e unit 3.pdfroad safety engineering  r s e unit 3.pdf
road safety engineering r s e unit 3.pdf
VENKATESHvenky89705
 
WATER CRISIS and its solutions-pptx 1234
WATER CRISIS and its solutions-pptx 1234WATER CRISIS and its solutions-pptx 1234
WATER CRISIS and its solutions-pptx 1234
AafreenAbuthahir2
 
Toll tax management system project report..pdf
Toll tax management system project report..pdfToll tax management system project report..pdf
Toll tax management system project report..pdf
Kamal Acharya
 

Recently uploaded (20)

Standard Reomte Control Interface - Neometrix
Standard Reomte Control Interface - NeometrixStandard Reomte Control Interface - Neometrix
Standard Reomte Control Interface - Neometrix
 
Chat application through client server management system project.pdf
Chat application through client server management system project.pdfChat application through client server management system project.pdf
Chat application through client server management system project.pdf
 
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
Sachpazis:Terzaghi Bearing Capacity Estimation in simple terms with Calculati...
 
Automobile Management System Project Report.pdf
Automobile Management System Project Report.pdfAutomobile Management System Project Report.pdf
Automobile Management System Project Report.pdf
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
 
Cosmetic shop management system project report.pdf
Cosmetic shop management system project report.pdfCosmetic shop management system project report.pdf
Cosmetic shop management system project report.pdf
 
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdfHybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdf
 
Forklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella PartsForklift Classes Overview by Intella Parts
Forklift Classes Overview by Intella Parts
 
Democratizing Fuzzing at Scale by Abhishek Arya
Democratizing Fuzzing at Scale by Abhishek AryaDemocratizing Fuzzing at Scale by Abhishek Arya
Democratizing Fuzzing at Scale by Abhishek Arya
 
weather web application report.pdf
weather web application report.pdfweather web application report.pdf
weather web application report.pdf
 
Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024
 
power quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptxpower quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptx
 
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
一比一原版(UofT毕业证)多伦多大学毕业证成绩单如何办理
 
Scaling in conventional MOSFET for constant electric field and constant voltage
Scaling in conventional MOSFET for constant electric field and constant voltageScaling in conventional MOSFET for constant electric field and constant voltage
Scaling in conventional MOSFET for constant electric field and constant voltage
 
Courier management system project report.pdf
Courier management system project report.pdfCourier management system project report.pdf
Courier management system project report.pdf
 
ASME IX(9) 2007 Full Version .pdf
ASME IX(9)  2007 Full Version       .pdfASME IX(9)  2007 Full Version       .pdf
ASME IX(9) 2007 Full Version .pdf
 
Top 13 Famous Civil Engineering Scientist
Top 13 Famous Civil Engineering ScientistTop 13 Famous Civil Engineering Scientist
Top 13 Famous Civil Engineering Scientist
 
road safety engineering r s e unit 3.pdf
road safety engineering  r s e unit 3.pdfroad safety engineering  r s e unit 3.pdf
road safety engineering r s e unit 3.pdf
 
WATER CRISIS and its solutions-pptx 1234
WATER CRISIS and its solutions-pptx 1234WATER CRISIS and its solutions-pptx 1234
WATER CRISIS and its solutions-pptx 1234
 
Toll tax management system project report..pdf
Toll tax management system project report..pdfToll tax management system project report..pdf
Toll tax management system project report..pdf
 

Introduction into Flow Measurment Technology_2021_Dec6.pdf

  • 2. 2 Normal cubic meter (m3 n) Gases do have a certain mass, but the density of gases is strongly dependent on pressure and temperature Normal conditions: defines the mass of a gas at a fixed temperature and pressure.
  • 3. 3 Normal cubic meter (m3 n) • Volume: 1m3 • Temperature: 0°C • Pressure: 1013,25 mbar • Relative humidity: 0% • Mass: 1293 gram m3 n
  • 4. 4 The ideal gas law Pressure * Volume / Temperature = Constant P∗V T =nm∗R V ½ V P 2P
  • 5. 5 Normal conditions may vary Other common normal conditions: • 20˚C reference (FAD): See ISO 1217 • 15˚ reference: See DIN 1533 • CAGI: 68 F, 14.5 PSIA, 0% water vapor pressure Nm3/hr = V ∗ (TN / T) ∗ (P / PN)
  • 6. 6 From FAD to normal cubic meter 100 m3/hr FAD ∗ 273,16/(273,16+20) ∗ 1000/1013,25 = 91,95 Nm3/hr FAD/Nm3 factor = 1,0874 → 8,7% difference when comparing numbers 20 deg 1000 20 deg 1000 20 deg 1013,25 20 deg 1000 0 deg 1013,25 FAD Nm3
  • 8. 8 ThermabridgeTM technology Vout = k ∗ λ ∗ ρ ∗ v ∗ (Ts−Tg) • Vout = output voltage • k = sensor (geometrical) constant • λ = thermal conductivity of the gas • ρ = density of the gas • v = actual velocity in in m / sec • Ts = sensor temperature • Tg = gas temperature
  • 9. 9 Normal meter per second Thermal mass flow: velocity in Normal meter per second Rule of thumb: “Actual velocity times absolute pressure” (forget temperature) Example: 8 m/sec∗ 7 bar abs ~ 56 mn/sec
  • 10. 10 Principles compared Thermal Vortex DP – Orifice plate DP – Cone meter Coriolis Turbine/ rotary displacement Clamp on ultrasonic Mass flow Yes Optional Optional Optional Yes Optional Optional Meter run 20D 15D 15D 5D 0D 10D 20D Pressure loss Low Medium/high high high Low Low Low Dirty air Fouling OK Clogging Clogging Internal fouling Faillure OK Wet Air Spikes OK, spikes OK OK, orientation Yes, but affects reading Faillure Spikes Range 1:250 1:10 1:10 1:10 1:100 1:100 1:100 Accuracy 2% 2% 2% 2% 0.5 .. 1% 0.5…1 % 1% Purchase price $ $ $ $ $$$$ $$ $$$ Maintenance Medium Low Medium Medium Low High Low
  • 11. 11 Value of mass flow and 3-in-1 We believe in: Pressure ∗ Flow = Voltage ∗ Current – Energy! • Flow, Pressure, Temperature combined • ISO 11011 (e.g. must measure pressure on multiple locations) • Difference between supply and demand • How to detect cause of a pressure drop if you dont measure? • Internal accounting: 7 bar is more expensive than 6 bar • Flow & temperature to identify issues with compressed air driers and (water) cooled compressors
  • 12. 12 Velocity to mass flow True diameter Measured error Measured diameter % of error 50 3 53 12.36 100 3 103 6.09 200 3 203 3.02 300 3 303 2.01 Check the inner pipe diameter Area A = π∗ 0.25 ∗ D2 Wrong entry of D → big deviation Example: V = 60mn/sec D: 50 mm → Q = 0,117 m3 n / sec D: 54 mm → Q = 0,137 m3 n / sec Difference: 17%
  • 14. 14 Thermal mass flow = pressure independent Why? The sensor counts molecules: More pressure = denser air* = more molecules = more heat loss* * : linear with pressure 2 x P 1 x P 1Q 2Q
  • 15. 15 Differential pressure technology Dynamic pressure port Positive side Dynamic pressure port Negative side Temperature FLOW VPFlowScope Transmitter housing - Absolute pressure sensor - Differential pressure sensor Locking ring
  • 16. 16 Flow range is critical Thermal vs DP technology: 1:300 range vs 1:10 (1:5) range Thermal: • From leakages to high flow • Dry air only • Temp range up to ~60°C DP: • Only medium to high flow rates • Not suitable for leakage/low flows • Wet air and dry air • Temp range up to ~150°C (or higher)
  • 17. 17 Bi-directional flow: air can go both ways!
  • 18. 18 Importance of bi-directional flow measurement Demand side (reverse flow) Tank Production line +1002 mn 3 +941 mn 3 - 61 mn 3 ____________ +880 mn 3 Flow Flow POU tank Production line Production line
  • 19. 19 Bi-directional flow examples • Ring networks • Multiple compressor rooms • Complex (old) compressed air networks • Large receiver tanks • Non return valves in compressors • Leaking drains and seals Extremely useful information for audits and permanent installations.
  • 20. 20 Turbulent vs laminar flow Reynolds number ρ∗v∗D η = density∗velocity∗tubediameter dynamic viscosity Air dynamic viscosity = 17,1 ∗ 10 −6, which is very small compared to water or syrup • Compressed air lines: Flow is always turbulent! • Turbulent: Velocity profile is nearly flat: Less sensitive for insertion depth
  • 22. 22 Know the basics: Process data Key to correct measurements and to prevent damage to your flow meter • Type of gas • Flow range • Humidity (dry/saturated) • (Inner) Diameter • Pressure • Temperature range
  • 23. 23 Model selection In-line meters: 0.5, 1, 2 inch Flow range leads direct to model, Check the diameter to ensure quick mechanical installation. Insertion probes: 2 inch and up The (normalized) velocity range needs to be calculated from the flow range to determine the right model. Normalized velocity = mass flow [m3 n/hr] / [3600] / tuber area
  • 24. 24 ½” 1” 2” > 2” Insertion probes In-line meters Size (inch) Min flow m3 n/hr Max flow m3 n/hr Min flow SCFM Max flow SCFM ½ inch 0,32 80 0.19 50 1 inch 0,88 250 0,52 150 2 inch 3,53 1000 2,06 600 Flow ranges
  • 25. 25 Flow ranges Thermal: Schedule 40 Standard Seamless Carbon Steel Pipe Size (inch) DN ID (inch) ID (mm) Min flow SCFM Max flow SCFM Min flow m3 n/hr Max flow m3 n/hr 2 50 2,1 52,5 2,3 688 3,9 1,169 4 100 4,0 102,3 8,7 2,610 15 4,435 6 150 6,1 154,1 20 5,924 34 10,065 12 300 11,9 303,2 77 22,953 130 38,995 20 500 18,8 477,8 190 56,996 323 96,832 ½” 1” 2” > 2” Insertion probes In-line meters
  • 26. 26 Not sure? Use VPCalculator Use our calculator to determine which flow sensor you need. See online Velocity Calculator https://www.vpinstruments.com/selection- guide/velocity/
  • 27. 27 When creating a project quote Gather enough information to be prepared • P&ID (Process and Instrumentation Diagram) • Plant map, to scale • Pictures of installation points • Use the on-line demand side table or Excel form • Use bi-directional as default option
  • 29. 29 Read the manual… • User manual • VP Academy • Instruction video • This training • Helpdesk “how to” articles
  • 30. 30 Check the direct environment • Avoiding excessive external heat • Protect against external water damage (IP) • Eliminating excessive system condensate • When having condensation in gas (100% condensation → VPS DP) • Avoid corrosive atmosphere where possible • Use the right cables and cable locations • Eliminate mechanical vibration and potential danger
  • 31. 31 Check air quality • Water • Oil • Drains • Particles • (Cone strainer)
  • 32. 32 Check temperature • Thermal mass: limited range: high temperature will cause sensor signal → 0 • Temperature fluctuations: lagging temperature compensation • Coming from cold outside → inside (audits) : allow 1 hour acclimatization for best results. • DP: high range, and reasonably fast response
  • 33. 33 Installation location: Avoid excess water Piping not in scale
  • 34. 34 2. Find a straight pipe General minimum rule: • 20*D upstream length (even 40*D preferred) • 5*D downstream length (10*D preferred) The longer the better Without conditioners: All other claimed shorter lengths are B.S. It’s based on physics, not technology
  • 35. 35 3. Use the best orientation Take into account: • Ease of wiring, maintenance, read out of and access to display • Install at least at an angle of 15° upwards. ALWAYS if possible • Piping table! Minimum up- and downstream length. Longer = better • In case of doubt: Communicate with the end user 30o
  • 36. 36 Recap • Thermal mass : dry air • Differential pressure : wet air • In Line : 0,5, 1 and 2 inch • Probes : > 2 inch • Always check : Temperature, Pressure diameter and range • Ring network : Bi – directional • General : Use the checklist! Things to avoid: • Large and fast temperature swings • Excessive condensate in the pipe • Short meter runs