M A D E B Y: L A L I T S I N G H
S T U D E N T O F M - P H A R M
( P H A R M A C E U T I C A L A N A LY S I S )
I S F C O L L E G E O F P H A R M A C Y, M O G A
Factor affecting UV-Vis
Absorption
 There are a range of factors that can affect the ultraviolet
and visible absorption characteristics of an organic
compound.
Concentration
of the sample
Temperature
of the sample
pH of the
sample
Solvent
Factor
Effect of Solvent
 The choice of solvent can shift peaks to shorter or longer
wavelengths.
 This will depend on the nature of the interaction of the
particular solvent with the environment of the
chromophore in the excited state of the molecule.
 Depending on the chromophore in the particular analyte,
changes in the polarity of the solvent can influence shifts
to longer or shorter wavelengths.
 For instance, it is usually seen that ethanol solutions give
longer wavelength maxima than hexane solutions.
Effect of Sample Concentration
 As you might expect, sample concentration is proportional
to the intensity of the absorption.
 At high concentrations however, molecular interactions
(for example, polymerisation) can take place causing
changes to the position and shape of absorption bands.
 Such an outcome can affect the linearity of the
relationship between sample concentration and
absorbance (remember Beer’s Law).
Conti…
 Sample concentration ∞ Absorbance
 Higher the concentration of the analyte show higher abs.
they don’t follow Lambert beer law.
 Optimum concentration of analyte will be use , the abs.
will below then 0.8 .
Effect of Sample pH
 The pH of the sample solution can have a significant
impact on absorption spectra.
 If pH is known to be a factor, a remedy is to prepare the
sample in a suitable buffer solution so as to maintain the
pH at a steady value.
 The buffer though needs to be transparent over the
wavelength range of the measurements – if the buffer
absorbs radiation, absorbance readings attributed to the
analyte may be higher than they should because the buffer
and analyte absorptions will add together at each
wavelength.
Conti…
 If the optimum pH
buffer solution is
suitable with
analyte
 The abs. spectra
will show clear
peak of the analyte
• If the optimum pH
buffer solution is
not suitable with
analyte
• The abs. spectra
will not show
clear peak of the
analyte, they
merge with buffer
peak.
Effect of Sample Temperature
 Temperature impacts absorption measurements by various
means:
 Expansion or contraction of the solvent - leading to
lower/higher concentrations and absorbances, a particular
issue with some organic solvents.
 Shifts in equilibria between the chemical forms of an
analyte – the nature of the absorbing species may be
changed.
 Changes to reaction rates - enzymatic reactions are
particularly sensitive to temperature.
Conti…
 Sample solution
Heat
 Expansion or contraction of the solvent
 lower/higher concentrations
 Absorbance will alter(May be higher abs. or lower abs.)
Factor affecting uv vis absorption lalit

Factor affecting uv vis absorption lalit

  • 1.
    M A DE B Y: L A L I T S I N G H S T U D E N T O F M - P H A R M ( P H A R M A C E U T I C A L A N A LY S I S ) I S F C O L L E G E O F P H A R M A C Y, M O G A Factor affecting UV-Vis Absorption
  • 2.
     There area range of factors that can affect the ultraviolet and visible absorption characteristics of an organic compound. Concentration of the sample Temperature of the sample pH of the sample Solvent Factor
  • 3.
    Effect of Solvent The choice of solvent can shift peaks to shorter or longer wavelengths.  This will depend on the nature of the interaction of the particular solvent with the environment of the chromophore in the excited state of the molecule.  Depending on the chromophore in the particular analyte, changes in the polarity of the solvent can influence shifts to longer or shorter wavelengths.  For instance, it is usually seen that ethanol solutions give longer wavelength maxima than hexane solutions.
  • 4.
    Effect of SampleConcentration  As you might expect, sample concentration is proportional to the intensity of the absorption.  At high concentrations however, molecular interactions (for example, polymerisation) can take place causing changes to the position and shape of absorption bands.  Such an outcome can affect the linearity of the relationship between sample concentration and absorbance (remember Beer’s Law).
  • 5.
    Conti…  Sample concentration∞ Absorbance  Higher the concentration of the analyte show higher abs. they don’t follow Lambert beer law.  Optimum concentration of analyte will be use , the abs. will below then 0.8 .
  • 6.
    Effect of SamplepH  The pH of the sample solution can have a significant impact on absorption spectra.  If pH is known to be a factor, a remedy is to prepare the sample in a suitable buffer solution so as to maintain the pH at a steady value.  The buffer though needs to be transparent over the wavelength range of the measurements – if the buffer absorbs radiation, absorbance readings attributed to the analyte may be higher than they should because the buffer and analyte absorptions will add together at each wavelength.
  • 7.
    Conti…  If theoptimum pH buffer solution is suitable with analyte  The abs. spectra will show clear peak of the analyte • If the optimum pH buffer solution is not suitable with analyte • The abs. spectra will not show clear peak of the analyte, they merge with buffer peak.
  • 8.
    Effect of SampleTemperature  Temperature impacts absorption measurements by various means:  Expansion or contraction of the solvent - leading to lower/higher concentrations and absorbances, a particular issue with some organic solvents.  Shifts in equilibria between the chemical forms of an analyte – the nature of the absorbing species may be changed.  Changes to reaction rates - enzymatic reactions are particularly sensitive to temperature.
  • 9.
    Conti…  Sample solution Heat Expansion or contraction of the solvent  lower/higher concentrations  Absorbance will alter(May be higher abs. or lower abs.)