Introduction
The efficiency of a column is measured by theoretical plates, Nth, and can be normalized
with the length of the column to give the height equivalent theoretical plate, called HETP
or H. The Van Deemter equation describes the various factors influencing H, and is
divided into eddy diffusion, longitudinal diffusion, and mass transfer terms. The relative
importance of these factors varies with mobile phase velocity. Particle size and
morphology contribute to H, along with a variety of other factors. Understanding the van
Deemter equation allows the determination of the optimum mobile phase velocity.
The theoretical plate number Nth shows the relation between retention time and peak
width and describes column quality and separation power.
Factors affecting column efficiency (plate number)
 Column length
 Particle size
 Packing quality
 Linear velocity (flow)
 Instrument quality (dead volume)
 Retention factor
Stationary phase particle size is one of the most important factors in the van Deemter
equation. For a given column length, the plate number (Nth) is inversely related to the
particle size of the column packing. The smaller the particles, the higher the plate number
and the separation power.
The plate number is also dependent on the flow rate (F) of the mobile phase. There is a
certain velocity, the so-called optimum flow, at which the plate number is highest (and H
is lowest). A lower or a higher flow rate provides less plates (higher H). In routine HPLC,
columns are always operated at velocities above the optimum. The reduced column
efficiency is less significant than the shorter analysis time at the higher than optimal flow
rates.
To describe the contribution of the above factors, several so-called plate height equations
have been developed. A plate height equation expresses the correlation between plate
height and mobile phase velocity. Best known is the van Deemter equation, which
describes the various contributions to plate height (H). In this equation the parameters
that influence the overall peak width are expressed in three terms:
H=A+Bu+C×u(1)(1)H=A+Bu+C×u
H = HETP (plate height)
A = eddy diffusion term
B = longitudinal diffusion term
u = linear velocity
C = Resistance to mass transfer coefficient
Peak height and peak broadening are governed by kinetic processes in the column such
as molecular dispersion, diffusion and slow mass transfer. Identical molecules travel
differently in the column due to probability processes. The three processes that contribute
to peak broadening described in the van Deemter equation are:
 A-term: eddy diffusion: The column packing consists of particles with flow
channels in between. Due to the difference in packing and particle shape, the
speed of the mobile phase in the various flow channels differs and analyte
molecules travel along different flow paths through the channnels.
 B-term: longitudinal diffusion: Molecules traverse the column under influence of
the flowing mobile phase. Due to molecular diffusion, slight dispersions of the
mean flow rate will be the result.
 C-term: resistance against mass transfer. A chromatographic system is in
dynamic equilibrium. As the mobile phase is moving continuously, the system has
to restore this equilibrium continuously. Since it takes some time to restore
equilibrium (resistance to mass transfer), the concentration profiles of sample
components between mobile and stationary phase are always slightly shifted. This
results in additional peak broadening.
Van Deemter H-u curve
The van Deemter equation is graphically expressed in the H-u curve, which is a plot of
the plate height as a function of the mobile phase velocity.
The H-u curve shows that:
 The A-term is independent of u and does not contribute to the shape of the H-u
curve..
 The contribution of the B-term is negligible at normal operating conditions. This
is due to the fact that the molecular diffusion coefficient in a liquid medium is very
small.
 The C- term increases linearily with mobile phase velocity and its
contribution to the H-u curve is therefore considerable. A small C-term leads to a
fairly flat ascending portion of the H-u curve at higher mobile phase velocities.
This means that the separation can be carried out at higher mobile phase
velocities without sacrificing separation quality.
Van Deemter H/u curve
The H-u curve is very useful to determine the optimum mobile phase velocity uopt at
which the highest column efficiency wil be attained. Below this velocity the column
efficiency will decrease rapidly, whereas above the optimum velocity there is only a slight
decrease in efficiency.
General rules of thumb:
 The smaller the plate height H, the more efficient the column. However, at the
optimum flow rate (with the lowest H), the analysis time will in most cases be
unacceptably long.
 Below the optimum flow rate the analysis time is too long and the quality of the
separation suffers because of longitudinal diffusion (contribution of the B-term).
 At extremely high flows, both the separation quality and the pressure drop
across the column will become unacceptable.
 The practical flow is often two or three times the optimum velocity. At these
values, the minor loss of efficiency is still acceptable.
 Small particles (< 5 µm) allow an increase in the flow to reduce the analysis
time without significantly lowering resolution between the peaks of interest.
Separation and (optimum) velocity
Eddy Diffusion (A-term)
The A-term in the Van Deemter equation describes the peak broadening due the
presence of stationary phase particles in the column. In the literature it is called ' eddy
diffusion'.
Eddy diffusion occurs as a result of multiple flow paths through a packed column bed.
Analyte molecules cannot flow along a straight path through the column. Each analyte
molecule follows a different flow path, which causes dispersion (differential path lengths).
Some molecules will travel more slowly and will lag behind the zone (peak) centre, others
will follow shorter paths and are slightly ahead. This effect results in slight differences in
retention.
The homogeneity of the packed bed strongly affects the A-term. The probability of
differential pathlengths due to radial movements is greater in a column having a poorly
packed bed. Eddy diffusion is important for columns with internal diameters of 2 - 5 mm.
However, when a column of this size is packed regularly with small, spherical particles
with small dispersion in particle size, the contribution from eddy diffusion in the van
Deemter equation will be a minimal.
CHROMEDIA PROGRAM
Eddy diffusion in the van Deemter equation is independent of the flow velocity of the
mobile phase. In reality it appears that this is not entirely true, particularly at low flow
rates. This can be seen in the above H-u curve.
Factors influencing A-term ('eddy' diffusion)
 Particle size dp
 Particle shape (regular or irregular?)
 Particle pore structure / shape
 Quality of the column packing
 Wall effects (material, roughness, column diameter)
Eddy diffusion is affected by the packing material in the column, specifically:
 Particle size: The larger the particles, the stronger the dispersion effect.
 Particle shape: For regularly shaped particles (spherical) the path length between
the particles is smaller than for irregular particles. This is due to the fact that
spherical particles can more easily form a regularly packed column bed.
Conversely, an irregularly packed bed consists of flow channels of different shapes
and diameters, resulting in different mobile phase velocities.
Longitudinal Diffusion (B-term)
The B-term in the van Deemter equation, also known as longitudinal diffusion, refers to
the diffusion of individual analyte molecules in the mobile phase along the longitudinal
direction of a column. Longitudinal diffusion contributes to peak broadening only at very
low flow rates below the minimum (optimum) plate heigth.
Molecular diffusion takes place independent of the longitudinal (= axial) flow direction.
Longitudinal diffusion is the result of concentration differences in the mobile phase. In the
center of the peak zone the concentration is at its maximum. The concentration before
and after the peak zone is lower. This results in diffusion, both in the direction of the
mobile phase flow as well as in the opposite direction.
Some molecules move faster and others will move slower relative to the average velocity,
resulting in peak broadening. This effect will be relatively great at long residence times in
the column, which is the case at low flow rates. As the flow rate increases, this effect will
contribute less to the total peak broadening. In practice, it is best to select flow rates that
minimise the effect of longitudinal diffusion on column efficiency.
CHROMEDIA PROGRAM
Longitudinal diffusion is proportional to the molecular diffusion coefficient. It obeys Fick’s
law and is therefore affected by viscosity, temperature and molecular size (~ weight).
Increased temperature and diffusion coefficient increase the B term, while increased
viscosity decreases it. Diffusion coefficients for components in liquids are generally very
small. Because the mobile phase velocity is much higher than the diffusion coefficient of
the component in solution, the B term is hardly significant in liquid chromatography and
is typically neglected.
Influences on B-term (longitudnial diffusion):
 Linear velocity of the mobile phase
 Diffusion coefficient of analyte in the mobile phase Dm
 Mobile phase viscosity γ
 Temperature
 Type of analyte (molecular mass)
Mass Transfer (C-term)
The C-term in the van Deemter equation relates to the mass transfer of sample
components between the stationary phase and the mobile phase during separation. The
overall C-term is divided into two separate mass transfer terms:
 Cm-term, describing the contributions to peak broadening in the mobile
phase. Because the linear velocity of the mobile phase is lower closer to the
column wall (or the stationary phase particles) than in the center (or further away
from the particles), the analyte molecules experience different velocities. This
results in peak broadening in the mobile phase. This phenomenon is described
by the Cm term.
 Cs-term, describing the contributions to peak broadening in the stationary
phase. The Cs-term is determined by the amount of stationary phase (low is
advantageous for the efficiency) and the extent of interaction of the sample on
the phase (represented by the retention factor) and the distances the sample
molecules have to travel.
Sample components (analytes) are retained in a chromatographic column due to their
interaction with the stationary phase. Analyte molecules present in the flowing mobile
phase diffuse towards the mobile/stationary pase interface, and enter into the stationary
phase. To maintain partition equilibrium, some molecules will return to the mobile phase.
After some time there will be a reverse process as the analytes move from the stationary
phase into the mobile phase. This results in a continuous mass transfer taking place
between the flowing mobile phase and the stagnant stationary phase during the course
of the separation.
Resistance to mass transfer is dependent on the speed with which the partition
equilibrium between mobile and stationary phase is obtained. Since the resistance to
mass transfer in the mobile phase is not the same for all molecules of one type of
analyte (it depends on the location in the column at a particular time and the distance
they have to travel to partition between phases) this will also result in peak broadening
of that analyte in the column.
Resistance against mass transfer in mobile phase
C" is a constant factor, related to the retention factor k of the peak of interest. Omega is
related to the solvent volume of the column between and in the particles (pores),
accessible for the mobile phase. Another important parameter affecting the Cm-term is
particle diameter. Smaller particles will result in narrower flow channels and less
contribution to mass transfer in the mobile phase.
Influences on Cm-term:
 Particle size dp
 Linear velocity u of the mobile phase
 Diffusion coefficient in the mobile phase
 Porosity of the packing particles
 Viscosity of the mobile phase
 Retention factor k
 Temperature
When analyte molecules can move quickly back and forth between mobile and
stationary phase, the column efficiency is greatly improved. The speed with which
analyte molecules move back and forth between phases is related to the diffusion
coefficient of the analyte molecule in a particular mobile phase. Analyte molecules with
large diffusion coefficients in the mobile phase will need less time to move to the
mobile/stationary phase interface. The same applies for the movement into and out of
the stationary phase. The viscosity and temperature also have an effect on diffusion.
Resistance Against Mass Transfer in Stationary Phase
Equilibrium is reached faster with a larger molecular diffusion coefficient for the mobile
phase and with a smaller particle diameter. This means that the contribution of the Cm
and Cs to plate height decreases and the separation is more efficient. The contribution of
the two C-terms to the overall peak broadening appears to increase with the linear velocity
of the mobile phase. As mobile phase velocity increases, the exchange of molecules
between mobile phase and stationary phase must be fast enough to keep up if the
equilibrium between the phases is to be maintained. If equilibrium is not maintained, the
peaks broaden and the efficiency and separation quality decrease.
Influences on Cs-term:
 Quality of stationary phase
 Diffusion coefficient in stationary phase
 Retention factor k
 Temperature
 Particle size
 Mobile phase velocity
The k value influences both C terms. The relative band broadening and thus the plate
number appears to be dependent on the extent of interaction of a component with the
stationary phase. Therefore, it becomes important to consider which peak is used to
calculate a column's plate number. This dependence is strongest at the beginning of the
chromatogram and decreases at higher retention factors. Thus Nth is the best calculated
from a peak somewhere at the end of the chromatogram (k>5).

More Related Content

What's hot

Factors affecting uv visible spectroscopy
Factors affecting uv visible spectroscopyFactors affecting uv visible spectroscopy
Factors affecting uv visible spectroscopy
wadhava gurumeet
 
Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
Vharsha Haran
 

What's hot (20)

Factors influencing chemical shift
Factors influencing chemical shiftFactors influencing chemical shift
Factors influencing chemical shift
 
Plate theory of Chromatography
Plate theory of ChromatographyPlate theory of Chromatography
Plate theory of Chromatography
 
Super crtical fluid chromatography ppt
Super crtical fluid chromatography pptSuper crtical fluid chromatography ppt
Super crtical fluid chromatography ppt
 
Column efficiency parameters
Column efficiency parametersColumn efficiency parameters
Column efficiency parameters
 
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
 
Theory of high performance liquid chromatography ppt
Theory of high performance liquid chromatography pptTheory of high performance liquid chromatography ppt
Theory of high performance liquid chromatography ppt
 
Factors affecting uv visible spectroscopy
Factors affecting uv visible spectroscopyFactors affecting uv visible spectroscopy
Factors affecting uv visible spectroscopy
 
Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
 
Capillary Electrophoresis
Capillary ElectrophoresisCapillary Electrophoresis
Capillary Electrophoresis
 
High Performance Liquid Chromatography (HPLC)
High Performance Liquid Chromatography (HPLC)High Performance Liquid Chromatography (HPLC)
High Performance Liquid Chromatography (HPLC)
 
I R spectroscopy & its application
I R spectroscopy & its application I R spectroscopy & its application
I R spectroscopy & its application
 
column chromatography ppt
column chromatography pptcolumn chromatography ppt
column chromatography ppt
 
Colum chromatography
Colum chromatographyColum chromatography
Colum chromatography
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Chromatography
ChromatographyChromatography
Chromatography
 
Ion pair chromatography final
Ion pair chromatography finalIon pair chromatography final
Ion pair chromatography final
 
Chromatography Lec 1
Chromatography Lec 1 Chromatography Lec 1
Chromatography Lec 1
 
HPLC
HPLCHPLC
HPLC
 
Ion exchange chromatography
Ion exchange chromatographyIon exchange chromatography
Ion exchange chromatography
 

Similar to Van meeter equaiton

Liquid chromatography still striving for high efficiency2
Liquid chromatography still striving for high efficiency2Liquid chromatography still striving for high efficiency2
Liquid chromatography still striving for high efficiency2
guest63ff7d
 
Liquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyLiquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiency
guest63ff7d
 
Chromatography introduction ppt by Akshay patel
Chromatography introduction ppt by Akshay patelChromatography introduction ppt by Akshay patel
Chromatography introduction ppt by Akshay patel
akshay patel
 
Presentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq AliPresentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq Ali
Shaiq Ali
 
Two Phase Flow Research
Two Phase Flow ResearchTwo Phase Flow Research
Two Phase Flow Research
Rex Liu
 

Similar to Van meeter equaiton (20)

Column chromatography.pdf
Column chromatography.pdfColumn chromatography.pdf
Column chromatography.pdf
 
Fundamentals of Rate theory in Chromatography
Fundamentals of Rate theory in ChromatographyFundamentals of Rate theory in Chromatography
Fundamentals of Rate theory in Chromatography
 
RATE TH,PLATE SYS SUTIABILITY.pptx
RATE TH,PLATE SYS SUTIABILITY.pptxRATE TH,PLATE SYS SUTIABILITY.pptx
RATE TH,PLATE SYS SUTIABILITY.pptx
 
Liquid chromatography still striving for high efficiency2
Liquid chromatography still striving for high efficiency2Liquid chromatography still striving for high efficiency2
Liquid chromatography still striving for high efficiency2
 
Liquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyLiquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiency
 
Chromatography introduction ppt by Akshay patel
Chromatography introduction ppt by Akshay patelChromatography introduction ppt by Akshay patel
Chromatography introduction ppt by Akshay patel
 
9553317.ppt
9553317.ppt9553317.ppt
9553317.ppt
 
Chromatography HPLC.ppt
Chromatography HPLC.pptChromatography HPLC.ppt
Chromatography HPLC.ppt
 
Chromatography Basics
Chromatography BasicsChromatography Basics
Chromatography Basics
 
Lecture 4 analytical chemistry 4-second stage
Lecture 4 analytical chemistry 4-second stageLecture 4 analytical chemistry 4-second stage
Lecture 4 analytical chemistry 4-second stage
 
THEORIES OF GAS CHROMATOGRAPHY: RATE THEORY AND PLATE THEORY
THEORIES OF GAS CHROMATOGRAPHY: RATE THEORY AND PLATE THEORYTHEORIES OF GAS CHROMATOGRAPHY: RATE THEORY AND PLATE THEORY
THEORIES OF GAS CHROMATOGRAPHY: RATE THEORY AND PLATE THEORY
 
Rheology
RheologyRheology
Rheology
 
Chromatographic techniques
Chromatographic techniquesChromatographic techniques
Chromatographic techniques
 
Chromatographic techniques
Chromatographic techniquesChromatographic techniques
Chromatographic techniques
 
Presentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq AliPresentation of chromatography by Shaiq Ali
Presentation of chromatography by Shaiq Ali
 
Theory and application of High Performance Liquid Chromatography (HPLC)
Theory and application of High Performance Liquid Chromatography (HPLC)Theory and application of High Performance Liquid Chromatography (HPLC)
Theory and application of High Performance Liquid Chromatography (HPLC)
 
Two Phase Flow Research
Two Phase Flow ResearchTwo Phase Flow Research
Two Phase Flow Research
 
Gas Chromatography
Gas Chromatography Gas Chromatography
Gas Chromatography
 
Chromatography
ChromatographyChromatography
Chromatography
 
Unit41.pptx
Unit41.pptxUnit41.pptx
Unit41.pptx
 

More from Sattar kayani (17)

Q3 (a)
Q3 (a)Q3 (a)
Q3 (a)
 
314 supp 6_isom_form
314 supp 6_isom_form314 supp 6_isom_form
314 supp 6_isom_form
 
314 supp 6_isom_form
314 supp 6_isom_form314 supp 6_isom_form
314 supp 6_isom_form
 
Reagent types chem
Reagent types chemReagent types chem
Reagent types chem
 
4 converted (1)
4 converted (1)4 converted (1)
4 converted (1)
 
Water soluble
Water solubleWater soluble
Water soluble
 
Water soluble vs Fat Soluable
Water soluble vs Fat SoluableWater soluble vs Fat Soluable
Water soluble vs Fat Soluable
 
6 defects
6 defects6 defects
6 defects
 
Create wireless ad hoc internet connection on windows 10
Create wireless ad hoc internet connection on windows 10Create wireless ad hoc internet connection on windows 10
Create wireless ad hoc internet connection on windows 10
 
Dua
DuaDua
Dua
 
Kinds of sentence
Kinds of sentenceKinds of sentence
Kinds of sentence
 
Lecture 8
Lecture 8Lecture 8
Lecture 8
 
Artificial intelligence cs607 handouts lecture 11 - 45
Artificial intelligence   cs607 handouts lecture 11 - 45Artificial intelligence   cs607 handouts lecture 11 - 45
Artificial intelligence cs607 handouts lecture 11 - 45
 
Dsys guide37
Dsys guide37Dsys guide37
Dsys guide37
 
A dhoc networks
A dhoc networksA dhoc networks
A dhoc networks
 
03 basic computer_network
03 basic computer_network03 basic computer_network
03 basic computer_network
 
Class01
Class01Class01
Class01
 

Recently uploaded

LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.
Cherry
 
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptxTHE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
ANSARKHAN96
 
Reboulia: features, anatomy, morphology etc.
Reboulia: features, anatomy, morphology etc.Reboulia: features, anatomy, morphology etc.
Reboulia: features, anatomy, morphology etc.
Cherry
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Sérgio Sacani
 
Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.
Cherry
 
development of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virusdevelopment of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virus
NazaninKarimi6
 
Human genetics..........................pptx
Human genetics..........................pptxHuman genetics..........................pptx
Human genetics..........................pptx
Cherry
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
levieagacer
 

Recently uploaded (20)

Site Acceptance Test .
Site Acceptance Test                    .Site Acceptance Test                    .
Site Acceptance Test .
 
Role of AI in seed science Predictive modelling and Beyond.pptx
Role of AI in seed science  Predictive modelling and  Beyond.pptxRole of AI in seed science  Predictive modelling and  Beyond.pptx
Role of AI in seed science Predictive modelling and Beyond.pptx
 
Cot curve, melting temperature, unique and repetitive DNA
Cot curve, melting temperature, unique and repetitive DNACot curve, melting temperature, unique and repetitive DNA
Cot curve, melting temperature, unique and repetitive DNA
 
FS P2 COMBO MSTA LAST PUSH past exam papers.
FS P2 COMBO MSTA LAST PUSH past exam papers.FS P2 COMBO MSTA LAST PUSH past exam papers.
FS P2 COMBO MSTA LAST PUSH past exam papers.
 
LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.
 
module for grade 9 for distance learning
module for grade 9 for distance learningmodule for grade 9 for distance learning
module for grade 9 for distance learning
 
Gwalior ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Gwalior ESCORT SERVICE❤CALL GIRL
Gwalior ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Gwalior ESCORT SERVICE❤CALL GIRLGwalior ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Gwalior ESCORT SERVICE❤CALL GIRL
Gwalior ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Gwalior ESCORT SERVICE❤CALL GIRL
 
Genome sequencing,shotgun sequencing.pptx
Genome sequencing,shotgun sequencing.pptxGenome sequencing,shotgun sequencing.pptx
Genome sequencing,shotgun sequencing.pptx
 
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptxTHE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
 
Selaginella: features, morphology ,anatomy and reproduction.
Selaginella: features, morphology ,anatomy and reproduction.Selaginella: features, morphology ,anatomy and reproduction.
Selaginella: features, morphology ,anatomy and reproduction.
 
Cyanide resistant respiration pathway.pptx
Cyanide resistant respiration pathway.pptxCyanide resistant respiration pathway.pptx
Cyanide resistant respiration pathway.pptx
 
Reboulia: features, anatomy, morphology etc.
Reboulia: features, anatomy, morphology etc.Reboulia: features, anatomy, morphology etc.
Reboulia: features, anatomy, morphology etc.
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
 
Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.
 
Kanchipuram Escorts 🥰 8617370543 Call Girls Offer VIP Hot Girls
Kanchipuram Escorts 🥰 8617370543 Call Girls Offer VIP Hot GirlsKanchipuram Escorts 🥰 8617370543 Call Girls Offer VIP Hot Girls
Kanchipuram Escorts 🥰 8617370543 Call Girls Offer VIP Hot Girls
 
development of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virusdevelopment of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virus
 
Site specific recombination and transposition.........pdf
Site specific recombination and transposition.........pdfSite specific recombination and transposition.........pdf
Site specific recombination and transposition.........pdf
 
Human genetics..........................pptx
Human genetics..........................pptxHuman genetics..........................pptx
Human genetics..........................pptx
 
Genome organization in virus,bacteria and eukaryotes.pptx
Genome organization in virus,bacteria and eukaryotes.pptxGenome organization in virus,bacteria and eukaryotes.pptx
Genome organization in virus,bacteria and eukaryotes.pptx
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
 

Van meeter equaiton

  • 1. Introduction The efficiency of a column is measured by theoretical plates, Nth, and can be normalized with the length of the column to give the height equivalent theoretical plate, called HETP or H. The Van Deemter equation describes the various factors influencing H, and is divided into eddy diffusion, longitudinal diffusion, and mass transfer terms. The relative importance of these factors varies with mobile phase velocity. Particle size and morphology contribute to H, along with a variety of other factors. Understanding the van Deemter equation allows the determination of the optimum mobile phase velocity. The theoretical plate number Nth shows the relation between retention time and peak width and describes column quality and separation power. Factors affecting column efficiency (plate number)  Column length  Particle size  Packing quality  Linear velocity (flow)  Instrument quality (dead volume)  Retention factor Stationary phase particle size is one of the most important factors in the van Deemter equation. For a given column length, the plate number (Nth) is inversely related to the particle size of the column packing. The smaller the particles, the higher the plate number and the separation power. The plate number is also dependent on the flow rate (F) of the mobile phase. There is a certain velocity, the so-called optimum flow, at which the plate number is highest (and H is lowest). A lower or a higher flow rate provides less plates (higher H). In routine HPLC, columns are always operated at velocities above the optimum. The reduced column efficiency is less significant than the shorter analysis time at the higher than optimal flow rates. To describe the contribution of the above factors, several so-called plate height equations have been developed. A plate height equation expresses the correlation between plate height and mobile phase velocity. Best known is the van Deemter equation, which describes the various contributions to plate height (H). In this equation the parameters that influence the overall peak width are expressed in three terms: H=A+Bu+C×u(1)(1)H=A+Bu+C×u H = HETP (plate height) A = eddy diffusion term B = longitudinal diffusion term
  • 2. u = linear velocity C = Resistance to mass transfer coefficient Peak height and peak broadening are governed by kinetic processes in the column such as molecular dispersion, diffusion and slow mass transfer. Identical molecules travel differently in the column due to probability processes. The three processes that contribute to peak broadening described in the van Deemter equation are:  A-term: eddy diffusion: The column packing consists of particles with flow channels in between. Due to the difference in packing and particle shape, the speed of the mobile phase in the various flow channels differs and analyte molecules travel along different flow paths through the channnels.  B-term: longitudinal diffusion: Molecules traverse the column under influence of the flowing mobile phase. Due to molecular diffusion, slight dispersions of the mean flow rate will be the result.  C-term: resistance against mass transfer. A chromatographic system is in dynamic equilibrium. As the mobile phase is moving continuously, the system has to restore this equilibrium continuously. Since it takes some time to restore equilibrium (resistance to mass transfer), the concentration profiles of sample components between mobile and stationary phase are always slightly shifted. This results in additional peak broadening. Van Deemter H-u curve The van Deemter equation is graphically expressed in the H-u curve, which is a plot of the plate height as a function of the mobile phase velocity. The H-u curve shows that:  The A-term is independent of u and does not contribute to the shape of the H-u curve..  The contribution of the B-term is negligible at normal operating conditions. This is due to the fact that the molecular diffusion coefficient in a liquid medium is very small.  The C- term increases linearily with mobile phase velocity and its contribution to the H-u curve is therefore considerable. A small C-term leads to a fairly flat ascending portion of the H-u curve at higher mobile phase velocities. This means that the separation can be carried out at higher mobile phase velocities without sacrificing separation quality.
  • 3. Van Deemter H/u curve The H-u curve is very useful to determine the optimum mobile phase velocity uopt at which the highest column efficiency wil be attained. Below this velocity the column efficiency will decrease rapidly, whereas above the optimum velocity there is only a slight decrease in efficiency. General rules of thumb:  The smaller the plate height H, the more efficient the column. However, at the optimum flow rate (with the lowest H), the analysis time will in most cases be unacceptably long.  Below the optimum flow rate the analysis time is too long and the quality of the separation suffers because of longitudinal diffusion (contribution of the B-term).  At extremely high flows, both the separation quality and the pressure drop across the column will become unacceptable.  The practical flow is often two or three times the optimum velocity. At these values, the minor loss of efficiency is still acceptable.  Small particles (< 5 µm) allow an increase in the flow to reduce the analysis time without significantly lowering resolution between the peaks of interest. Separation and (optimum) velocity
  • 4. Eddy Diffusion (A-term) The A-term in the Van Deemter equation describes the peak broadening due the presence of stationary phase particles in the column. In the literature it is called ' eddy diffusion'. Eddy diffusion occurs as a result of multiple flow paths through a packed column bed. Analyte molecules cannot flow along a straight path through the column. Each analyte molecule follows a different flow path, which causes dispersion (differential path lengths). Some molecules will travel more slowly and will lag behind the zone (peak) centre, others will follow shorter paths and are slightly ahead. This effect results in slight differences in retention. The homogeneity of the packed bed strongly affects the A-term. The probability of differential pathlengths due to radial movements is greater in a column having a poorly packed bed. Eddy diffusion is important for columns with internal diameters of 2 - 5 mm. However, when a column of this size is packed regularly with small, spherical particles with small dispersion in particle size, the contribution from eddy diffusion in the van Deemter equation will be a minimal. CHROMEDIA PROGRAM Eddy diffusion in the van Deemter equation is independent of the flow velocity of the mobile phase. In reality it appears that this is not entirely true, particularly at low flow rates. This can be seen in the above H-u curve. Factors influencing A-term ('eddy' diffusion)  Particle size dp  Particle shape (regular or irregular?)  Particle pore structure / shape  Quality of the column packing  Wall effects (material, roughness, column diameter) Eddy diffusion is affected by the packing material in the column, specifically:  Particle size: The larger the particles, the stronger the dispersion effect.  Particle shape: For regularly shaped particles (spherical) the path length between the particles is smaller than for irregular particles. This is due to the fact that spherical particles can more easily form a regularly packed column bed. Conversely, an irregularly packed bed consists of flow channels of different shapes and diameters, resulting in different mobile phase velocities.
  • 5. Longitudinal Diffusion (B-term) The B-term in the van Deemter equation, also known as longitudinal diffusion, refers to the diffusion of individual analyte molecules in the mobile phase along the longitudinal direction of a column. Longitudinal diffusion contributes to peak broadening only at very low flow rates below the minimum (optimum) plate heigth. Molecular diffusion takes place independent of the longitudinal (= axial) flow direction. Longitudinal diffusion is the result of concentration differences in the mobile phase. In the center of the peak zone the concentration is at its maximum. The concentration before and after the peak zone is lower. This results in diffusion, both in the direction of the mobile phase flow as well as in the opposite direction. Some molecules move faster and others will move slower relative to the average velocity, resulting in peak broadening. This effect will be relatively great at long residence times in the column, which is the case at low flow rates. As the flow rate increases, this effect will contribute less to the total peak broadening. In practice, it is best to select flow rates that minimise the effect of longitudinal diffusion on column efficiency. CHROMEDIA PROGRAM Longitudinal diffusion is proportional to the molecular diffusion coefficient. It obeys Fick’s law and is therefore affected by viscosity, temperature and molecular size (~ weight). Increased temperature and diffusion coefficient increase the B term, while increased viscosity decreases it. Diffusion coefficients for components in liquids are generally very small. Because the mobile phase velocity is much higher than the diffusion coefficient of the component in solution, the B term is hardly significant in liquid chromatography and is typically neglected. Influences on B-term (longitudnial diffusion):  Linear velocity of the mobile phase  Diffusion coefficient of analyte in the mobile phase Dm  Mobile phase viscosity γ  Temperature  Type of analyte (molecular mass)
  • 6. Mass Transfer (C-term) The C-term in the van Deemter equation relates to the mass transfer of sample components between the stationary phase and the mobile phase during separation. The overall C-term is divided into two separate mass transfer terms:  Cm-term, describing the contributions to peak broadening in the mobile phase. Because the linear velocity of the mobile phase is lower closer to the column wall (or the stationary phase particles) than in the center (or further away from the particles), the analyte molecules experience different velocities. This results in peak broadening in the mobile phase. This phenomenon is described by the Cm term.  Cs-term, describing the contributions to peak broadening in the stationary phase. The Cs-term is determined by the amount of stationary phase (low is advantageous for the efficiency) and the extent of interaction of the sample on the phase (represented by the retention factor) and the distances the sample molecules have to travel. Sample components (analytes) are retained in a chromatographic column due to their interaction with the stationary phase. Analyte molecules present in the flowing mobile phase diffuse towards the mobile/stationary pase interface, and enter into the stationary phase. To maintain partition equilibrium, some molecules will return to the mobile phase. After some time there will be a reverse process as the analytes move from the stationary phase into the mobile phase. This results in a continuous mass transfer taking place between the flowing mobile phase and the stagnant stationary phase during the course of the separation. Resistance to mass transfer is dependent on the speed with which the partition equilibrium between mobile and stationary phase is obtained. Since the resistance to mass transfer in the mobile phase is not the same for all molecules of one type of analyte (it depends on the location in the column at a particular time and the distance they have to travel to partition between phases) this will also result in peak broadening of that analyte in the column. Resistance against mass transfer in mobile phase
  • 7. C" is a constant factor, related to the retention factor k of the peak of interest. Omega is related to the solvent volume of the column between and in the particles (pores), accessible for the mobile phase. Another important parameter affecting the Cm-term is particle diameter. Smaller particles will result in narrower flow channels and less contribution to mass transfer in the mobile phase. Influences on Cm-term:  Particle size dp  Linear velocity u of the mobile phase  Diffusion coefficient in the mobile phase  Porosity of the packing particles  Viscosity of the mobile phase  Retention factor k  Temperature When analyte molecules can move quickly back and forth between mobile and stationary phase, the column efficiency is greatly improved. The speed with which analyte molecules move back and forth between phases is related to the diffusion coefficient of the analyte molecule in a particular mobile phase. Analyte molecules with large diffusion coefficients in the mobile phase will need less time to move to the mobile/stationary phase interface. The same applies for the movement into and out of the stationary phase. The viscosity and temperature also have an effect on diffusion. Resistance Against Mass Transfer in Stationary Phase Equilibrium is reached faster with a larger molecular diffusion coefficient for the mobile phase and with a smaller particle diameter. This means that the contribution of the Cm and Cs to plate height decreases and the separation is more efficient. The contribution of the two C-terms to the overall peak broadening appears to increase with the linear velocity
  • 8. of the mobile phase. As mobile phase velocity increases, the exchange of molecules between mobile phase and stationary phase must be fast enough to keep up if the equilibrium between the phases is to be maintained. If equilibrium is not maintained, the peaks broaden and the efficiency and separation quality decrease. Influences on Cs-term:  Quality of stationary phase  Diffusion coefficient in stationary phase  Retention factor k  Temperature  Particle size  Mobile phase velocity The k value influences both C terms. The relative band broadening and thus the plate number appears to be dependent on the extent of interaction of a component with the stationary phase. Therefore, it becomes important to consider which peak is used to calculate a column's plate number. This dependence is strongest at the beginning of the chromatogram and decreases at higher retention factors. Thus Nth is the best calculated from a peak somewhere at the end of the chromatogram (k>5).