Ct radiation dose & safety lecture 1

  1. 1. ‫تعلم‬ ‫تكن‬ ‫مالم‬ ‫وعلمك‬‫تعلم‬ ‫تكن‬ ‫مالم‬ ‫وعلمك‬ ‫ال‬ ‫فضل‬ ‫وكان‬‫ال‬ ‫فضل‬ ‫وكان‬ ‫عظيما‬ ‫عليك‬‫عظيما‬ ‫عليك‬
  2. 2. Basics Of Computed Tomography (CT) Lecture Ten GGamalamal FFathallaathalla MM.. MMahdalyahdaly gamal_mahdaly@hotmail.comgamal_mahdaly@hotmail.com
  3. 3. Radiation Safety and ProtectionRadiation Safety and Protection The potentially harmful effects of ionizingThe potentially harmful effects of ionizing radiation are either:radiation are either:  Stochastic effectsStochastic effects OrOr  Deterministic effectsDeterministic effects
  4. 4. Stochastic effectsStochastic effects  Effects where theEffects where the probabilityprobability of the occurrenceof the occurrence increases with radiation exposure.increases with radiation exposure. e.g.e.g. carcinogenesis and genetic effects.carcinogenesis and genetic effects. i.e.i.e. ((The((The probabilityprobability, but, but not severitynot severity,, of the end point condition, isof the end point condition, is dose-dependentdose-dependent))))
  5. 5. Deterministic effectsDeterministic effects  Effects related to aEffects related to a threshold dosethreshold dose,, belowbelow which the effect is not detected butwhich the effect is not detected but aboveabove thisthis threshold dose , the probability that the effectthreshold dose , the probability that the effect will occur is virtuallywill occur is virtually 100%.100%.  TheThe severityseverity increases with increased dose.increases with increased dose. e.g.e.g. Erythema, epilation, desquamation, cataract,Erythema, epilation, desquamation, cataract, fibrosis and hematopoietic damage.fibrosis and hematopoietic damage.
  6. 6. CT DosimetryCT Dosimetry  To evaluate theTo evaluate the potential risk/benefitpotential risk/benefit of CTof CT scanning, dose should be estimated.scanning, dose should be estimated.  2 main questions are to be answered:2 main questions are to be answered: 1.1. How muchHow much radiation dose is my CT scannerradiation dose is my CT scanner delivering to the patient?delivering to the patient? 2.2. How to compareHow to compare my scanner dose to that ofmy scanner dose to that of other CT scanners?other CT scanners?
  7. 7. Dose MeasurementsDose Measurements  Erythema doseErythema dose  Slight film fogSlight film fog  Roentgen (R)Roentgen (R)  Exposure doseExposure dose  Absorbed doseAbsorbed dose  Equivalent doseEquivalent dose
  8. 8. Classic and SI units of radiationClassic and SI units of radiation dosedose Dose type Unit Abbr Unit Abbr Exposure (air kerma) For x-ray or gamma ray ionization in air only Roentgen R Coulomb per kilogram C/kg Absorption Energy deposited by any type of radiation in any material Radiation absorbed dose Rad Gray Gy Equivalent Biologic effect caused by any radiation in a living organism Roentgen equivalent mammal Rem Sievert Sv Classic units SI units
  9. 9. ExposureExposure  Ionizing radiation ionizes the gas molecules intoIonizing radiation ionizes the gas molecules into electrically charged ions . This has traditionallyelectrically charged ions . This has traditionally been measured in terms of exposure Ebeen measured in terms of exposure E Where Q is total electric charge produced and MWhere Q is total electric charge produced and M is the mass of air.is the mass of air.  The old exposure unit of 1 Roentgen isThe old exposure unit of 1 Roentgen is equivalent to 2.58x10equivalent to 2.58x10-4-4 C.kgC.kg-1-1 .. Since 1 Gy =3x10Since 1 Gy =3x10-2-2 C.kgC.kg-1-1 then 1R = 8.6 Gy.then 1R = 8.6 Gy. E= Q/M (C.Kg-1 )
  10. 10. Absorbed dose (Gy)Absorbed dose (Gy) Is the energy absorbed by exposed material (air or tissue) in joules/kg E absorbed by a mass M of tissue: E/M is the the gray (Gy) 1 Gy = 1x10-3 J.g-1 Absorbed Dose = E/M (Gy)
  11. 11. Dose equivalent (Sievert)Dose equivalent (Sievert)  Tissue damage due to different type ofTissue damage due to different type of ionizing radiation ( Gamma, X, Beta, andionizing radiation ( Gamma, X, Beta, and Alfa) varies considerably.Alfa) varies considerably.  Dose equivalent allows for this byDose equivalent allows for this by multiplying the absorbed dose (grays) bymultiplying the absorbed dose (grays) by a weighting factor which depends on thea weighting factor which depends on the type of radiation.type of radiation. Grays x Q = Sieverts (Sv)
  12. 12. CT Radiation Dose CT Radiation Dose has 3 unique features: Axial CT image is very much collimated with minimal scattered radiation. Dose is evenly distributed ( in modern CTs) due to rotational acquistion. To achieve high contrast resolution, CT needs high SNR which necessiates high Dose/ Volume i.e. high technique. PA Chest XR needs only 120 kVp, 2 mAs Chest CT may need up to120 kVp, 200 mAs
  13. 13. X-ray beam geometryX-ray beam geometry Most modern scanners employ a fan-Shaped X-rayMost modern scanners employ a fan-Shaped X-ray beambeam Factors affecting Radiation Dose in CT
  14. 14. The width of x-ray beam is viewed form sideThe width of x-ray beam is viewed form side CollimatorCollimator determines the width:determines the width: An ideal dose distributionAn ideal dose distribution along z axis is shown (B).along z axis is shown (B). Actual bell-shaped doseActual bell-shaped dose distribution curve (C).distribution curve (C). A B C A Radiation Dose in CT
  15. 15. 1616
  16. 16. 1717 Radiation Dose The main X-ray interaction mechanism in CT is Compton scattering. CT slice acquisition delivers a considerable dose from scatter to adjacent tissues outside the primary beam path. As slice number increases, scattering also increases.
  17. 17. Methods of measuring patient doseMethods of measuring patient dose Of the many dose measurement methods, weOf the many dose measurement methods, we are going to consider only:are going to consider only:  TheThe pencil ionizationpencil ionization chamberchamber method.method.  CT dose indexCT dose index (CTDI) method.(CTDI) method.  Multiple scan average doseMultiple scan average dose (MSAD)(MSAD) method.method.
  18. 18. Radiation Dose Measurement - CTDI CTDI is the dose to any point in the patient including scatter from 7 CT slices in both directions (H/F & F/H) ( a total of 14 slices). The multiple scan average dose MSAD can be estimated using a single scan by measuring the CT Dose Index (CTDI). CTDI can be measured using a pencil ionization chamber in phantoms simulating head (16 cm diameter acrylic) & bodies (32 cm diameter acrylic. Doses at the patient surface may be higher than in the patient center.
  19. 19. In head scans, the surface-to-center ratio is approximately 1:1. In body scans, the surface-to-center ratio is approximately 2:1. CTDI measurements are done at both the surface CTDI (Periphery) & center CTDI (Center) of the phantom & then combined to give CTDIw CTDIw = (2/3 CTDIperipheral + 1/3 CTDIcenter)
  20. 20. 2121 Dose Measurement - CTDI
  21. 21. Dose indexDose index  The CT dose index (CTDI) is mathematically defined as :The CT dose index (CTDI) is mathematically defined as :  Where n is the number of distinct planes of the dataWhere n is the number of distinct planes of the data collected during one revolution,collected during one revolution, sw is the slice width (insw is the slice width (in mm), D(z) is the dose distribution, z is the dimensionmm), D(z) is the dose distribution, z is the dimension along the patient’s axis. For axial CT scanners and spiralalong the patient’s axis. For axial CT scanners and spiral CT scanners with single array of detectors, n = 1. ForCT scanners with single array of detectors, n = 1. For multi slices CT scanner n is the number of activemulti slices CT scanner n is the number of active detector rows.detector rows. *The integral sign merely instructs the user to determine the area single curve (D(z). *
  22. 22. Measuring the CTDIMeasuring the CTDI  CTDI is measured using long cylindricalCTDI is measured using long cylindrical ionization chamber and radiation doseionization chamber and radiation dose from single slice.from single slice.  The ionization chamber receives radiationThe ionization chamber receives radiation from all parts of dose distribution D(z)from all parts of dose distribution D(z) because its length is bigger than the widthbecause its length is bigger than the width of the X-ray beam.of the X-ray beam.
  23. 23.  The total charge from the ionization chamber isThe total charge from the ionization chamber is proportional to the integral in the CTDI definition.proportional to the integral in the CTDI definition. •Where Q is the total charge collected during single scan and Cf is calibration factor of ionization chamber. •Because the ionization chamber measures the exposure and not the dose we, convert Roentgen to cGy.
  24. 24. The integral in Equation is numerically equal to the area ( shaded region ) of the dose distribution Note that the CTDI can be increased by increasing the area under the curve. The area can be increased by either : 1.increasing the intensity of radiation, which Raises top the of the curve. Or 2. widening the beam , usually by opening the x-ray collimator.
  25. 25. Multiple Scan Average DoseMultiple Scan Average Dose  To measure radiation dose received byTo measure radiation dose received by patient from a series of CT scans,patient from a series of CT scans, between each two scans the patient isbetween each two scans the patient is moved by a bed index distance, each slicemoved by a bed index distance, each slice delivers its characteristic bell-shapeddelivers its characteristic bell-shaped dose.dose.  Finally, we have multiple consecutive bell-Finally, we have multiple consecutive bell- shaped doses.shaped doses.
  26. 26. A series of 7spaced scans at a constant bed index along the Z- axis is acquired to produce 7 bell- shaped dose distribution curves (TOP). Summation of these doses results in the (BOTTOM) curve. The total dose ( BOTTOM) curve has peaks where the bell-shaped curves overlap. The dotted line through the total dose curve is the multiple scan average dose. The MSAD is defined as the average dose (at a particular depth from the surface) resulting from a large number of successive slices. The MSAD refers to all the dose delivered to the tissues including dose due to scatter from all the successive slices.
  27. 27. Multi-scan average doseMulti-scan average dose (MSAD)(MSAD) CTDI can be related to the MSAD by this equation Where BI is the bed index or slice spacing (in mm), Sw is slice width (in mm), and the number of active arrays of detectors.
  28. 28. CTDI-Dose Measurement   Decreasing kVp reduces dose while other factors are constantDecreasing kVp reduces dose while other factors are constant CTDI values for body scans are lower than those for head scans due to greater attenuation of X-rays in the body. These values DO NOT quantify patient risk because they DO NOT consider the number of slices NOR the radiosensitivity of the irradiated organs. CTDI values increase with kVp, so decreasing kVp while other factors remain constant reduces the CTDI values.
  29. 29. Effect o kVp On Radiation Dose kVp not only controls the image contrast but also controls the amount of penetration that the x-ray beam will have as it traverses the patient Parameter 80 kV 120 kV 140 kV Image Contrast Best Intermediate Poor Noise Most Average Least Penetration Least Average Most Patient Dose per mAs Lowest Intermediate Highest