2. Outline
1. Introduction
2. History
3. Principle
4. Ideal grid
5. Grid construction
• Grid ratio
• Grid frequency
• Grid materials
6. Grid performance
7. Grid types
8. Grid problem
9. Grid selection
10. Patient dose
11. Air gap technique
3. Introduction
The grid is an extremely effective device for reducing the level of
scattered radiation that reaches to the image receptor (IR).
A carefully fabricated series of section of radiopaque material (grid
stripes) alternating with the section of radiolucent materials
(interspace materials).
The grid is positioned between the patient and image receptor.
The grid is designed to transmit only those x-rays whose direction is
on straight line from source to IR.
Scattered radiation is absorbed in the grid materials.
4.
5. History
The grid was first invented in 1913 by Dr Gustav Bucky.
Consisted of wide strips of lead approx. 2cm apart in crisscross pattern.
Despite the crudeness, it removed enough scatter to improve contrast.
6. History
In 1920 Dr Hollis Potter, a Chicago radiologist improved Dr Bucky’s grid design.
He realigned the lead strips so they would run in only one direction.
He made the lead strips thinner and therefore less obvious on the image and design a
device (now known as the potter bucky diaphragm) that allowed the grid to move
during the exposure.
The motion eliminates grid lines.
7. Principle
When the primary beam interact with the patient’s body, the scattered
radiation are produced that may or may not be absorbed depending on
angle of incidence and physical characteristics of the grid.
If the angle of scattered beam is greater enough to interact lead strips, it is
absorbed.
If the angle is slight, the scattered beam will be transmitted as the primary
beam.
Grid removes the scattered radiation before it reaches to the film, therefore
it improves contrast.
8. Ideal grid
Ideal grid;
passes all the primary photons i.e.
photons coming from focal spot.
block all the secondary photons i.e.
photons not coming from focal spot
9. Grid construction
There are three different aspect of grid construction.
1. Grid ratio
2. Grid frequency
3. Grid materials
• Interspace materials
• Lead strips
10. 1:Grid ratio
A grid has a three main dimensions
1. the thickness of the grid stripe(T)
2. the width of the interspace materials(D)
3. the height of the grid(h)
Grid ratio is the height of grid divided by interspace material
thickness.
Grid ratio=h/D
11. Grid ratio
Grid ratio plays a major role in the grid’s ability to improve contrast.
Thus, if the height of the grid is constant and the distance between
the lead strips is decreased, this will result in an increase in the grid
ratio.
It determines how scatter radiation is “cleaned up”.
Grid ratio is expressed as X:1
Typical values; 5:1 to 16:1 – in general (8:1 to 10:1 generally used)
3:1 to 5:1 _ in mammography (4:1 to 5:1 in mammo)
A 5:1 ratio grid will clean up about 85% and 16:1 ratio grid will clean
up about 97% of scattered radiation.
Higher the grid ratio, the more clean up of scatter radiation.
Grid function generally improves with higher ratio.
12. 2: Grid frequency
It is defined as the number of grid lines per inch or
centimeter.
Usually ranges from 60-200 lines /inch (25-80lines /cm)
Most commonly used grids have frequency of 85-103
lines /inch (33-41 lines/cm)
Mammographic grids have frequencies of 80lines /cm
(200lines /inch)
High frequency grids
• More and thinner grid strips
• High radiographic technique
• Higher patient dose
• No significant grid lines on the image
14. 3: Grid materials
A series of radiopaque lead stripes which alternate with radiolucent
materials.
a. Strips are held firmly together then sliced into flat sheets.
b. Lead is the radiopaque material of choice.
1. Interspace materials
• Materials between the grid stripes.
• Width varies from 250-350 micron.
• It maintains the precise separation between the strips.
• Generally constructed from aluminum or plastic fiber.
15. Advantages of Aluminum
Provides selective filtration of scattered x-ray not absorbed
in grid strips.
Non- hydroscopic --does not absorb moisture as plastic
fibers does.
Easy to manufacture as it can be roll into sheets of precise
thickness.
Produce less visible grid lines on the radiograph.
16. Disadvantages
1. Increase the absorption of primary x-ray (low KvP).
2. Higher mAs and higher patient dose.
3. At low KvP, patient radiation dose may be increased by approx;20%.
However, Aluminum has definite advantages over plastic fiber.
Aluminum is used as the cover for the grid to protect it from damage and
moisture.
Fiber interspace grids are preferred where their application can contribute
to lower patient dose, such as in mammography.
17. 2.Grid strips
Grid stripes should be thick enough to absorb the scatter radiation and
thin enough to allow the primary radiation.
The height varies from 2mm-5mm.
Lead is widely used material because;
• It is easy to shape.
• It is relatively inexpensive.
• It has high atomic number and high mass density.
Tungsten, platinum, gold and uranium can also be used but none of them
has overall desirable characteristics that of lead.
18. Grid performance
The principle function of grid is to improve contrast.
There are three methods of evaluating performance;
Contrast improvement factor
Bucky factor
Selectivity
19. Contrast Improvement Factor
Measures improvement in image quality when grid is used.
Grid ratio does not reveal the ability of the grid that how
much image contrast is improved than non grid.
Contrast improvement factor compares contrast
improvement of a radiograph with a grid to that without a
grid.
It is represented by letter “K”.
Contrast improvement factor of 1 indicates no improvement.
20. Contrast Improvement Factor Equation
K = Radiographic contrast with grid
Radiographic contrast without grid
Most grids have a contrast improvement of 1.5 -
2.5
Contrast improvement is higher with higher ratio
grids
K is the complex function of
• X ray emission spectrum
• Patient thickness
• Tissue irradiated
21. Bucky Factor
Also called grid factor.
B=Patient dose with grid
Patient dose without grid
The higher the grid ratio, the higher is the bucky factor.
The amount of radiation hitting the grid will always be greater than
the amount hitting the film.
22. Grid selectivity
Although grids are designed to absorb scatter radiation, they also absorb
some primary radiation.
Grids that absorb a greater percentage of scatter than primary radiation
are described as having a greater degree of selectivity.
The better a grid is at removing scatter, the greater will be the selectivity of
the grid.
This means that a grid with a higher lead content would have a greater
selectivity.
It is represented by Greek sigma ()
=Primary radiation transmitted through grid
Scatter radiation transmitted through grid
23. General Rules Of Grid Characteristics
High ratio grids have high contrast improvement factor
High frequency grids have thin strips of interspace material
Heavy grids have high selectivity and high contrast
improvement factors – the heavier a grid is the more lead it
contains the higher its selectivity and the more efficient is
in cleaning up scatter radiation
25. a) Parallel grid
Simplest type of grid .
Parallel grids are made with the lead and interspace
strips running parallel to one another.
This means that if the grid lines were extended into
space they would never intersect.
More primary radiation cutoff.
Not useful for
Large area image receptor.
Short SID
Parallel grids only reduces scatter in the direction of the
grid lines.
1:Linear grid
26. b) Crossed grid
Two linear grids at right angles to each other
Used for high contrast studies
More efficient in cleanup of scatter radiation than that of parallel grid.
Dr Bucky’s original grid was made using this pattern.
Higher contrast improvement factor than a parallel grid of twice the grid
ratio
Restricted application in clinical radiology
27. Disadvantages of crossed Grid
Grid cut off is the primary disadvantage of a crossed grid.
The Central ray must be perfectly aligned with the center of
the grid.
Tube can not be angled unless x ray tube and table are
properly aligned.
High exposure technique required
28. 2:Focused Grids
Designed to minimize grid cut off.
Lead strips lie on the imaginary radial lines of a
circle centered at the focal spot so they can
coincide with the divergence of the x ray beam.
If properly positioned ,no grid cutoff.
More difficult to manufacture than parallel.
Normally the SID is 100cm in table and 180 cm
in chest radiography.
29. 3) Moving grid
Grids lines are visible if primary x rays are absorbed even grid strips
are very small
Invented by HOLLIS E. POTTER in 1920 with simple idea of moving
the while exposure is made
Focused grids are usually used
Placed in holding mechanism that begins moving just before x-ray
exposure and continues moving after end of exposure
Moves 1-3 inches
Motion blurs out the grid strips
30. Single stroke
Antiquated /Old fashioned.
Grid had to be cocked with a spring mechanism.
Worked in correspond with exposure time.
The mechanism moved once throughout exposure.
Had to be reset for each exposure.
31. Reciprocating gird
With the reciprocating grid, a motor drives the grid back and forth
during the exposure.
Does not have to be reset for each exposure.
The distance of drive is approx. 2cm.
32. Oscillating grid
Moves in a circular motion as opposed to back and forth.
Delicate , Spring like devices located in the four corners hold the grid
centered within the frame.
A powerful electromagnet pulls the grid to one side and releases it at
the beginning of the exposure.
The grid oscillates in a circular fashion around the grid frame,
coming to rest after 20-30 secs.
33. Advantages of moving grid.
• No grid lines are seen.
• Problems occur infrequently.
• Motion blur is undetectable.
Disadvantages
• Mechanical problems may occur.
• Costly.
• Increase the patient radiation dose.
• Very infrequently, the motion is detected in the radiograph
34. Mammographic grid
Moving grid with ratio of 4:1 to 5:1.
Grid frequencies of 30-40 lines /cm.
Bucky factor:2-3
Does not compromise spatial resolution but increases pt. dose.
4:1 ratio grid doubles the pt. dose compared to non grid contact
mammography.
HTC grid is the choice.
35. High transmission cellular grid (HTC)
Reduces scatter radiation in the two directions.
Grid strips are made up of copper.
Interspace material is air.
Physical dimensions-3.8:1
36. Grid problems
Most frequent error – improper positioning
For correct functioning, precisely positioned in relative to the x ray tube target
and central ray of the x ray beam
Problems associated with focused grid
1. off-level
2. off- center
3. off- focus
4. upside down
5. off-center ,off-focus
37. 1. Off level
This occurs when the central ray is
angled across the long axis of the grid
strips/ across the radiographic table.
Improper positioned x-ray tube not the
grid
May occur during the grid tilting during
horizontal beam or in mobile
radiography.
Central rays incident on the grid at angle.
Grid cutoff will occur.
38. 2. Off center
If the central ray is not properly centered
to the centermost interspace of the grid
i.e. lateral shifting of grid
Common problem with mobile x-ray
tables or ceiling suspended tubes.
Arises due to improper tube position
Also called lateral decentering
39. 3. Off focused
Unspecified SIDs selection in case of the focused
grid
Major problem with high ratio grids.
Grid cutoff occurs if distance increases
Grid cut off more severe at the edge than the
center
This problem will not be observed if all chest
radiograph are taken at 180cm SID and all table
radiographs at 100cm SID.
40. 4. Upside down
Major grid problem but will be noticed
easily.
Maximum cutoff on either side of central
ray.
This type of grid error occurs when the
radiographic grid is used upside-down.
It is important that the technologist check
the tube side prior to using a focusing grid.
41. 5. Combined off center and off focused
Most common improper grid position.
Mostly occurs during mobile radiography.
Resultant radiograph with dark on one side and light on the other
i. e. uneven exposure.
Can be easily recognized.
42. Grid selection
Do not use the grid for children
Grid can be used if thickness is >10cm
Moving grid is the best than stationary grid
Moving grid with focused grid strips is common than parallel one
Up to 8:1 ratio is good for <90KVp used and more than 8:1 is good for
>90KVp
Choice of grid is also depends on the size and shape of anatomy that is
being examined
4:1 or 5:1 ratio with 30-40 lines/cm is good for mammography
But the HTC grid is the choice for mammography.
43. Grid selection factor
Patient dose increases with increasing grid ratio.
High ratio grids are used for high kVp examinations.
Patient dose at high kVp is less than that at low kVp.
44. Patient dose
Moving grid increases pt. dose by 15% than the stationary
The use of high KVp and high- ratio grids results in lower patient
doses and equal image quality than the use of low-KVp and low-ratio
grid
Disadvantages of using grid
• Increase patient dose
• Increased radiographic technique required (Usually mAs than kVp)
45. Contd;
Grid ratio mAs increase kVp increase
• No grid 1X 0
• 5:1 2X +8 to 10
• 8:1 4X +13 to 15
• 12:1 6X + 20 to 25
• 16:1 8X + 30 to 40
46. Problem;
• Original: 20mAs with an 8:1 grid then,
Find new mAs with a 12:1 grid
mAs2 = 20 mAs x 6
4
mAs2 = 120
4
mAs2 = 30
47. Air gap technique
The air gap technique is an alternative to use of a
grid.
It has primary application in magnification
radiography to a lesser extent , in chest radiography
and c-spine lateral radiograph.
The technique involves placing the patient at a
greater object image receptor distance (OID).
Image receptor is moved 10 to 15cm from the
patient.
By moving patient away from the image receptor,
the amount of scatter reaching the image receptor
will be reduced.
48. Contd;
The result is improved contrast without the use of grid.
mAs usually increased approx. 10% for every centimeter
of air gap.
The primary disadvantage of air gap technique is the loss
of sharpness that results from increased OID.
49. References
Radiologic Science for Technologist (Physics, Biology and Protection)
by Stewart C. Bushong, 9th Edition.
Chesney equipment
Website articles