MAGNIFICATION
(MACRORADIOGRAPHY/
MICRORADIOGRAPHY)
DISTORTION
PARTHA JYOTI DAS
MSc MIT
Contents
 Macroradiography
 Principles
 Magnification with fixed FFD
 Magnification with fixed FOD
 Unsharpness with macroradiography
 Applications
 Microradiography
 Mass miniature radiography
 Distortion
 References
The technique of producing an image by direct
magnification is called macroradiography.
Enables fine anatomical details than the original
and accurate diagnosis.
Macroradiograp
hy
Principles
 Magnified image can be produced
by increasing the object to film
distance in which the x rays
diverging from the point source
will produce a directly magnified
image.
Relation:
 The
relationship
between
size of
object and
it’s image.
FFD
FOD
OFD
Image Size
X-ray Focus
Film
Plane
Object
Plane
Principles
 Magnification (M) can be calculated by the
following formula:
focus to film distance (FFD)
focus to object distance(FOD)
. The focus to object distance (or object to film
distance) is taken from the mid level or part.
M =
MAGNIFICATION WITH FIXED
FFD
Magnification is increased by bringing the object
nearer to the x ray tube
FOD = FFD/Magnification.
For eg : With a fixed FFD of 100cm, if a
magnification factor of 1.6 is required then;
FOD = 100/1.6 = 62.5 cm
MAGNIFICATION WITH FIXED
FOD
 The required magnification is obtained by
moving the film away from the object.
 The required OFD for a given magnification is
then calculated from
OFD = FOD(M – 1)
Eg. FOD is fixed at 100cm , given magnification
factor 1.6
OFD = 100(1.6 – 1) = 60cm
UNSHARPNESS WITH
MACRORADIOGRAPHY
Movement
Unsharpness
Geometric
Unsharpness
MOVEMENT UNSHARPNESS
 Direct magnification by increased object to
film distance can be carried out if there is
complete immobilization of the patient
 Any movement due to lack of immobilization or
involuntary movement will be magnified on the
radiograph due to increased OFD
Movement
unsharpness
Some steps
 Use of immobilization devices, e.g. supports,
binders,sandbags and pads.
 Instructions should be given to the patient to
remain still.
GEOMETRIC UNSHARPNESS
 Occurs because the source of x rays is not a
point source and any distance between the
object and film will cause an image penumbra
in addition to magnifying the image
 The geometrical unsharpness for given focal
spot size calculated by
Focal Spot Size x Object-Film
Distance
Ug =
Focus-Object Distance
Geometric unsharpness
How to minimize Ug:-
 Focal Spot size:- By reducing Focal Spot
Size.
 Object image Distance:- By reducing object
image distence.
 Focus object Distance:-By increasing the
Focus object distence.
Relationship between focal spot
size and magnification
 For a given focal spot size there is a limit to
magnify the image without significant loss of
details.
APPLICATIONS
Limited use of macroradiography
 Dacrocystography of Skull (Lacrimal ducts after injection
of contrast)
 Imaging the carpal bones in cases of suspected fracture
of scaphoid
 Localised areas of the lung
 Localised areas during angiography eg: cerebral
angiography
Macroradiograph of wrist with a magnification
factor of 2
Microradiography
 An imaging technique used to examine
very small objects or minute details, often
with the use of high-voltages, very small
focal spot size, and an ultrafine film
emulsion.
Ultrafine Emulsion film
 Photo print made by these films performs rich
sharpness, high definition, strong scratch
resistance, as well as long lasting by pigment
Microradiography
 It is also called as chest fluoroscopy.
 The image on a fluorescent screen is
photographed on a small film. Screen and film
are separated and the image is focused on the
film by a camera lens
 Miniature radiographs (70mmx70mm or
100mmx100mm roll or cut films were used).
 A permanent record is possible.
Mass Miniature Radiography(MMR)
 In order to control tuberculosis, medical research
council, London appointed their Committee on
Tuberculosis in 1941.
 The detection of symptomless or latent pulmonary
tuberculosis by mass radiography had been the
subject of much careful investigation.
 An excellent instrument was designed by the
council for this purpose and in 1943 the Ministry of
Health announced that a limited number of mobile
miniature radiography sets.
The equipment consist of
 Fluorescent screen(emits blue or green light)
 A mirror optics system on recording exposure
on film
Survey in progress
Control Panel
Apparatus for MMR
1.Power unit:- High tension unit
 Four valve rectification
 An output of from 100 to 400mA
 A maximum kilovoltage of 100 kVp and 91 kVp
respectively
2. X-ray Tube
 The rotating anode X-ray tube
 Dual focus, Focal-spot sizes are 1mm and
2mm square.
 An optical centring device.
 Adjustable diaphragm
3.Fluorescent Screen and Grid:
 16 by 16 inches fluorescent screen
 Covered by a sheet of protective lead glass.
 Yellow-blue or green type of screen
 A stationary grid is also present which
reduces scatter
4.The Camera.:
 An electrically operated camera is linked with the
X-ray exposure switch.
 Lead glass protection is present on the camera side of
screen
5.The lens: a number of different lenses
Developing MMR films
 Special tanks meant for roll films on a spiral frame
Development time was 7min at 68 degree F
Viewing of MMR FILMS
 Projector for viewing miniature radiographs
Magnifying glasses
DISTORTION
DISTORTION
 Misrepresentation of the true size and
shape of an object is called distortion.
 Distortion results from unequal
magnification of different parts of the
same object.
Foca
l
Spot
Film
•The object AB is
tilted with respect to
the plane of the film.
•The size and shape
of image vary with
the amount of tilting.
Distortion
A
B
A B
Distortion of the image of an object will be
different in different parts of the x-ray beam.
 Distortion of thick object occurs if they are
not directly in the central part of the x-ray
beam.
 When the part to be imaged does not lay parallel
with the IR (cassette).
 If the Central Ray is not perpendicular to the part.
 Caused by improper alignment object with relation
to tube focus and film .
 Distortion of an object will be different in different
part of the x-ray beam.
Elongated Foreshortened Normal
 GOOD RADIOGRAPHS
 MAKES EVERYONE HAPPY
REFERENCES
 CHESNEY’S RADIOGRAPHIC IMAGING, 6TH EDITION
 JOHN BALL, TONY PRICE HDCR
 CLARK’S POSITIONING IN RADIOGRAPHY, 12TH
EDITION
 A. Stewart Whitley Charles Sloane Graham Hoadley Adrian D. Moore
Chrissie W. Alsop
.
Thank You

More Related Content

What's hot

Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
Sahith Reddy
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
Rad Tech
 

What's hot (20)

Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
B.Sc RADIOLOGY QUESTION AND ANSWER BANK - POSITIONING
 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiography
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancement
 
Radiographic cassettes
Radiographic cassettesRadiographic cassettes
Radiographic cassettes
 
Production and control of scatter radiation (beam
Production and control of scatter radiation (beamProduction and control of scatter radiation (beam
Production and control of scatter radiation (beam
 
Viewing and storage of x ray film
Viewing and storage of x ray filmViewing and storage of x ray film
Viewing and storage of x ray film
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968
 
Soft tissue radiography.pptx
Soft tissue radiography.pptxSoft tissue radiography.pptx
Soft tissue radiography.pptx
 

Similar to Magnification(macro and micro radiography), distortion

Distortion
DistortionDistortion
Distortion
mr_koky
 
Microscopy fin
Microscopy finMicroscopy fin
Microscopy fin
MUBOSScz
 
Measurement of corneal curvature
Measurement of corneal curvatureMeasurement of corneal curvature
Measurement of corneal curvature
GauriSShrestha
 

Similar to Magnification(macro and micro radiography), distortion (20)

Photorefraction
PhotorefractionPhotorefraction
Photorefraction
 
radiographic Image quality by Astuti Mishra
 radiographic Image quality by Astuti Mishra radiographic Image quality by Astuti Mishra
radiographic Image quality by Astuti Mishra
 
factor affecting quality of radiograph
factor affecting quality of radiographfactor affecting quality of radiograph
factor affecting quality of radiograph
 
02 principle of photography and imaging
02 principle of photography and imaging02 principle of photography and imaging
02 principle of photography and imaging
 
Stereoscopy in Dentistry
Stereoscopy in DentistryStereoscopy in Dentistry
Stereoscopy in Dentistry
 
Cone beam computed tomography (2)
Cone beam computed tomography (2)Cone beam computed tomography (2)
Cone beam computed tomography (2)
 
Distortion
DistortionDistortion
Distortion
 
projection geometry
projection geometryprojection geometry
projection geometry
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
 
Coneal topography instrumentation, techniques, procedures, limitations, advan...
Coneal topography instrumentation, techniques, procedures, limitations, advan...Coneal topography instrumentation, techniques, procedures, limitations, advan...
Coneal topography instrumentation, techniques, procedures, limitations, advan...
 
Microscopy fin
Microscopy finMicroscopy fin
Microscopy fin
 
Characterization of materials lec2 3
Characterization of materials  lec2 3Characterization of materials  lec2 3
Characterization of materials lec2 3
 
Orthopantomogram
OrthopantomogramOrthopantomogram
Orthopantomogram
 
PROJECTION GEOMETRY/ dental implant courses
PROJECTION GEOMETRY/ dental implant coursesPROJECTION GEOMETRY/ dental implant courses
PROJECTION GEOMETRY/ dental implant courses
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
 
Image characteristics,latent image,film processing.
Image characteristics,latent image,film processing.Image characteristics,latent image,film processing.
Image characteristics,latent image,film processing.
 
Measurement of corneal curvature
Measurement of corneal curvatureMeasurement of corneal curvature
Measurement of corneal curvature
 
Bino presentation2015
Bino presentation2015Bino presentation2015
Bino presentation2015
 
Panoramic imagining
Panoramic imagining Panoramic imagining
Panoramic imagining
 
spect .pptx
spect .pptxspect .pptx
spect .pptx
 

Recently uploaded

raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMalda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Sheetaleventcompany
 
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetdhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Punjab Call Girls Contact Number +919053,900,678 Punjab Call Girls
Punjab Call Girls Contact Number +919053,900,678 Punjab Call GirlsPunjab Call Girls Contact Number +919053,900,678 Punjab Call Girls
Punjab Call Girls Contact Number +919053,900,678 Punjab Call Girls
@Chandigarh #call #Girls 9053900678 @Call #Girls in @Punjab 9053900678
 
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Patna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Patna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetPatna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Patna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetkozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
Sheetaleventcompany
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
mahaiklolahd
 
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetneemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
Sheetaleventcompany
 

Recently uploaded (20)

raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMalda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Sexy Call Girl Dharmapuri Arshi 💚9058824046💚 Dharmapuri Escort Service
Sexy Call Girl Dharmapuri Arshi 💚9058824046💚 Dharmapuri Escort ServiceSexy Call Girl Dharmapuri Arshi 💚9058824046💚 Dharmapuri Escort Service
Sexy Call Girl Dharmapuri Arshi 💚9058824046💚 Dharmapuri Escort Service
 
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
 
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetdhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
dhanbad Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Dehradun Call Girls 8854095900 Call Girl in Dehradun Uttrakhand
Dehradun Call Girls 8854095900 Call Girl in Dehradun  UttrakhandDehradun Call Girls 8854095900 Call Girl in Dehradun  Uttrakhand
Dehradun Call Girls 8854095900 Call Girl in Dehradun Uttrakhand
 
Punjab Call Girls Contact Number +919053,900,678 Punjab Call Girls
Punjab Call Girls Contact Number +919053,900,678 Punjab Call GirlsPunjab Call Girls Contact Number +919053,900,678 Punjab Call Girls
Punjab Call Girls Contact Number +919053,900,678 Punjab Call Girls
 
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Patna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Patna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetPatna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Patna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetkozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
kozhikode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
💚 Punjabi Call Girls In Chandigarh 💯Lucky 🔝8868886958🔝Call Girl In Chandigarh
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
 
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...Call Girl in Indore 8827247818 {Low Price}👉   Nitya Indore Call Girls  * ITRG...
Call Girl in Indore 8827247818 {Low Price}👉 Nitya Indore Call Girls * ITRG...
 
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetneemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
 

Magnification(macro and micro radiography), distortion

  • 2. Contents  Macroradiography  Principles  Magnification with fixed FFD  Magnification with fixed FOD  Unsharpness with macroradiography  Applications  Microradiography  Mass miniature radiography  Distortion  References
  • 3. The technique of producing an image by direct magnification is called macroradiography. Enables fine anatomical details than the original and accurate diagnosis. Macroradiograp hy
  • 4. Principles  Magnified image can be produced by increasing the object to film distance in which the x rays diverging from the point source will produce a directly magnified image.
  • 5. Relation:  The relationship between size of object and it’s image. FFD FOD OFD Image Size X-ray Focus Film Plane Object Plane
  • 6. Principles  Magnification (M) can be calculated by the following formula: focus to film distance (FFD) focus to object distance(FOD) . The focus to object distance (or object to film distance) is taken from the mid level or part. M =
  • 7. MAGNIFICATION WITH FIXED FFD Magnification is increased by bringing the object nearer to the x ray tube FOD = FFD/Magnification. For eg : With a fixed FFD of 100cm, if a magnification factor of 1.6 is required then; FOD = 100/1.6 = 62.5 cm
  • 8.
  • 9. MAGNIFICATION WITH FIXED FOD  The required magnification is obtained by moving the film away from the object.  The required OFD for a given magnification is then calculated from OFD = FOD(M – 1) Eg. FOD is fixed at 100cm , given magnification factor 1.6 OFD = 100(1.6 – 1) = 60cm
  • 10.
  • 12. MOVEMENT UNSHARPNESS  Direct magnification by increased object to film distance can be carried out if there is complete immobilization of the patient  Any movement due to lack of immobilization or involuntary movement will be magnified on the radiograph due to increased OFD
  • 14. Some steps  Use of immobilization devices, e.g. supports, binders,sandbags and pads.  Instructions should be given to the patient to remain still.
  • 15. GEOMETRIC UNSHARPNESS  Occurs because the source of x rays is not a point source and any distance between the object and film will cause an image penumbra in addition to magnifying the image
  • 16.  The geometrical unsharpness for given focal spot size calculated by Focal Spot Size x Object-Film Distance Ug = Focus-Object Distance
  • 18. How to minimize Ug:-  Focal Spot size:- By reducing Focal Spot Size.  Object image Distance:- By reducing object image distence.  Focus object Distance:-By increasing the Focus object distence.
  • 19. Relationship between focal spot size and magnification  For a given focal spot size there is a limit to magnify the image without significant loss of details.
  • 20. APPLICATIONS Limited use of macroradiography  Dacrocystography of Skull (Lacrimal ducts after injection of contrast)  Imaging the carpal bones in cases of suspected fracture of scaphoid  Localised areas of the lung  Localised areas during angiography eg: cerebral angiography
  • 21. Macroradiograph of wrist with a magnification factor of 2
  • 22. Microradiography  An imaging technique used to examine very small objects or minute details, often with the use of high-voltages, very small focal spot size, and an ultrafine film emulsion.
  • 23. Ultrafine Emulsion film  Photo print made by these films performs rich sharpness, high definition, strong scratch resistance, as well as long lasting by pigment
  • 24. Microradiography  It is also called as chest fluoroscopy.  The image on a fluorescent screen is photographed on a small film. Screen and film are separated and the image is focused on the film by a camera lens  Miniature radiographs (70mmx70mm or 100mmx100mm roll or cut films were used).  A permanent record is possible.
  • 25. Mass Miniature Radiography(MMR)  In order to control tuberculosis, medical research council, London appointed their Committee on Tuberculosis in 1941.  The detection of symptomless or latent pulmonary tuberculosis by mass radiography had been the subject of much careful investigation.  An excellent instrument was designed by the council for this purpose and in 1943 the Ministry of Health announced that a limited number of mobile miniature radiography sets.
  • 26. The equipment consist of  Fluorescent screen(emits blue or green light)  A mirror optics system on recording exposure on film
  • 28. Apparatus for MMR 1.Power unit:- High tension unit  Four valve rectification  An output of from 100 to 400mA  A maximum kilovoltage of 100 kVp and 91 kVp respectively
  • 29. 2. X-ray Tube  The rotating anode X-ray tube  Dual focus, Focal-spot sizes are 1mm and 2mm square.  An optical centring device.  Adjustable diaphragm
  • 30. 3.Fluorescent Screen and Grid:  16 by 16 inches fluorescent screen  Covered by a sheet of protective lead glass.  Yellow-blue or green type of screen  A stationary grid is also present which reduces scatter 4.The Camera.:  An electrically operated camera is linked with the X-ray exposure switch.  Lead glass protection is present on the camera side of screen 5.The lens: a number of different lenses
  • 31. Developing MMR films  Special tanks meant for roll films on a spiral frame Development time was 7min at 68 degree F
  • 32.
  • 33. Viewing of MMR FILMS  Projector for viewing miniature radiographs Magnifying glasses
  • 35. DISTORTION  Misrepresentation of the true size and shape of an object is called distortion.  Distortion results from unequal magnification of different parts of the same object.
  • 36. Foca l Spot Film •The object AB is tilted with respect to the plane of the film. •The size and shape of image vary with the amount of tilting. Distortion A B A B
  • 37. Distortion of the image of an object will be different in different parts of the x-ray beam.
  • 38.  Distortion of thick object occurs if they are not directly in the central part of the x-ray beam.
  • 39.  When the part to be imaged does not lay parallel with the IR (cassette).  If the Central Ray is not perpendicular to the part.
  • 40.  Caused by improper alignment object with relation to tube focus and film .  Distortion of an object will be different in different part of the x-ray beam.
  • 42.  GOOD RADIOGRAPHS  MAKES EVERYONE HAPPY
  • 43. REFERENCES  CHESNEY’S RADIOGRAPHIC IMAGING, 6TH EDITION  JOHN BALL, TONY PRICE HDCR  CLARK’S POSITIONING IN RADIOGRAPHY, 12TH EDITION  A. Stewart Whitley Charles Sloane Graham Hoadley Adrian D. Moore Chrissie W. Alsop .