Filters, Collimators and Grids
Christensen’s physics of
diagnostic radiology
Filters
Filtration is the process of shaping the
X-ray beam to increase the ratio of
photons useful for imaging to those
photons that increase patient dose or
decrease image contrast
Diagnostic X-ray beam is
polychromatic, comprising of whole
spectrum of energies. The mean
energy will vary from 1/3rd to one half
of the their peak energy. The first few
centimeters of the tissues receive
much more radiation than the rest of
the body tissues of the patient
Filters
Filters are sheets of metal, attached
at the opening of tube housing, which
will absorb the low energy photons
from the X-ray beam before it reaches
the patient
Types of filtration
The X-ray beam is filtered by absorbers at
three different levels
Inherent Filtration
Added filtration
Patient
Inherent filtration
Glass envelope
The insulating oil surrounding the tube
Window in the tube housing
Added filtration
Aluminum At.No. 13
Copper At. No. 29
Compound Filter copper + Aluminum
Advantage of compound filtration
Copper is used to cut down the thickness
of filter
Copper will absorb high energy photons
and Aluminum will absorb the
characteristic radiation from copper
(8 keV)
Measurement of filtration
Filtration is measured in Aluminum
equivalents which is defined as the
thickness of the Aluminum that would
produce the same degree of
attenuation as the thickness of
material in question
Effects of filtration
Patient exposure and
Exposure factors
60 –kVp beam
Aluminum Exposure dose Decrease in
filtration (mm) to skin (mR) exposure dose (%)
None 2380
0.5 1850 22
1.0 1270 47
3.0 465 80
Inherent filtration at the tube housing is
0.5 to 1.0 Al.Eq.
Below 50 kVp 0.5 mm Aluminum
50-70 kVp 1.5 mm Aluminum
Above 70 kVp 2.5 mm Aluminum
Effect on exposure factors
There will be reduction in the intensity of
X-ray beam as the filters absorb some
photons at all energy levels.
To Compensate the loss of high energy
photons, increase in the exposure
factors (mAs) is required
Types of filters
Single Filter Aluminum
Compund Filter Aluminum +Copper
Wedge Filter Used in the
angiography
Molybdenum Filters used in
Mammography.
K. Edge filter
Molybdenum filters
Used in molybdenum target X-ray tubes
used for mammography
17.5 kev K alpha and 19.6 kev K beta
characterstic radiation of Mo
When operated at 30-40 kVp, Mo will
produce bremsstrahling with energies
higher than 20 kev
Mo filter attenuates these high energy rays
K- Edge filters
These filters make use of K absorption edge
of elements with atomic No. greater than 60.
The purpose of heavy metal filters or K edge
filters is to produce an X-ray beam that has a
high number of photons in the specific energy
range
Enhance contrast for Iodine and barium,
reduce patient dose, and increase tube
loading
X-ray beam restrictors
X-ray beam restrictors
An X-ray beam restrictor is a device
that is attached to the opening in the
X-ray tube housing to regulate the
size and shape of the X-ray beam
Types of restrictors
Aperture diaphragms
Cones & cylinders
Collimators
Aperture diaphragm
It consists of a sheet of lead with a hole in
the center. The size and shape of the hole
determine the size and shape of the X-ray
beam.
Cones
Cones are usually flare shaped
Ideal geometric configuration for an
X-ray beam restrictor.
The flare of the cone is greater than the
flare of the x ray beam.
Cylinders
Beam restriction with cylinder takes
place at the far end of the barrel, so
there is less penumbra.
Aperture diaphragm cone cylinder
P
Disadvantages
Penumbra: Partially exposed periphery of
the X-ray field is called penumbra
Another major disadvantage with cones
and cylinders is severe limitations they
place on the number of available field
sizes
Collimators
Collimator is the best all round X-ray
beam restrictor
These are two types:
Manual collimator
Automatic collimator or PBL (Positive
beam limitation device or automatic light
localised variable aperture collimator)
Advantages
It provides infinite variety of rectangular
or square X-ray fields
The light beam shows the exact center
and configuration of X-ray field
Construction of collimator
X ray
filter
mirror
X ray
&light
beam
bulb
Lower
shutter
Collimator shutters
Testing of X- ray beam and light
beam alignment
Periodic check of alignment of
X-ray beam and light beam is
essential, because the mirror gets
out of adjustment due to frequent
daily use.
Automatic collimators
When the cassette is loaded in the Buckey
tray the sensors in the tray identify the size
and alignment of the cassette and relay
the information to collimator motors, which
positions shutters to the exact size of the
film used.
Functions of X ray beam restrictors
Protects the patient from unnecessary
radiation
It decreases the scatter radiation
The number of scattered radiation depends upon
field size
Small fields generate little scatter, as the field
increases scatter increases
Collimators are only successful in decreasing the
scatter radiation with small fields, so we should
reduce the size of X-ray beams as much as
possible
GRIDS
The radiographic grid consists of a series of
lead foil strips separated by X ray transparent
spacers.
It was invented by DR.GUSTAVE BUCKY in
1913.
Grid is still the most effective way of removing
the scatter radiation from large radiographic
fields
Primary radiation is oriented in the same axis as
the lead strips and passes between them .
Scatter radiation arises from many points within
the patient and most of it is absorbed by the lead
strips
The interspaces of the grids are filled either
with aluminium or some organic compound.
The main purpose of the interspace
material is to support the thin lead foil
strips.
Is defined as the ratio between the height of
the lead strips and the distance between them.
GRID RATIO
The lead strips are approximately 0.05 mm
thick ( lead foil).
The interspaces are much thicker.
Grid ratios are usually expressed as two
numbers, such as 20:1
Ratios usually range from 4:1 to 16:1
the Higher the ratio, the better the grid
functions.
Grid pattern
Is the orientation of the lead strips in their
longitudinal axis.
The two basic grid patterns are :
Linear and
Crossed.
They allow us to angle the x-ray tube along the
length of the grid .
Linear grid
A crossed grid is made up of two superimposed linear
grids that have the same focussing distance
The grid ratio of crossed grids is
equal to the sum of the ratios of
the two linear grids.
A crossed grid made up of two 5:1
linear grids has a ratio of 10:1.
Crossed grids cannot be used with oblique
techniques requiring angulation of the X-ray tube
Crossed grids
Is a grid made up of lead strips that are angled slightly so
that they focus in space.
A focussed grid may be either linear or crossed.
Linear focused grids converge at a line in space called
the convergent line.
Crossed grids converge at a point in space called the
convergent point.
The focal distance is the perpendicular distance
between the grid and the convergent line or point.
Focussed grid
Focussing range
Indicates the distance within which the grid can be used
without significant loss of primary radiation
It is fairly wide for a low-ratio grid and narrow for a high
ratio grid.
A 5:1 grid focused at 40 inches has a focusing range of
approximately 28 to 72 inches.
While a 16:1 grid focused at 40inches has a range of only 38
to 42inches.
Parallel grid
A parallel grid is one in which the lead strips are parallel
They are focused at infinity.
can only be used with either very small Xray fields or long-
target grid distances.
They are frequently used in fluoroscopic spot film
devices.
Lines per inch
Is the number of lead strips per inch of the grid.
Lines per inch = 25.4/D+d
D= thickness of the interspaces
d=thickness of the lead strips(both in millimeters)
Grid cassette
Usually used for portable radiography ,
with a grid built in to the front of the
cassette.
Are focussed and
Have a grid ratio of 4:1 to 8:1
Evaluation of grid performance
The three methods of evaluating grid
performance:
1.Primary transmission(Tp)
2.Bucky factor (B)
3.Contrast improvement factor(K)
Primary transmission
Is the percentage of primary radiation transmitted
through the grid.
Ideally , a grid should transmit 100% of the primary
radiation.
The first measurement is made
with the grid in place
The second measurement is
made after removal of the grid
A ratio of the intensity with the grid to the
intensity without the grid gives the
fractional transmission, which is
multiplied by 100 to give the percentage of
transmission.
intensity with grid Ip
Tp = _____________________ X100
intensity without grid I’p
There is a significant loss of primary
radiation with grids, more with cross grids
Measured primary tranmission
The measured primary transmission is
always less than the calculated primary
transmission:
absorption by the interspace material
manufacturing imperfections
Bucky factor
Is the ratio of the incident radiation falling on the grid to the
transmitted radiation passing through the grid.
It indicates how much we must increase exposure factors
when we change from a non grid to a grid technique.
The Bucky factor indicates the absorption of both primary and
secondary radiation.
It is determined with a
large X-ray field and a
thick phantom
The Bucky factor( B) is a measure of the total
quantity of radiation absorbed from an X-ray
beam by a grid
incident radiation
B = ____________________
transmitted radiation
The transmitted radiation is measured with the
grid in place ,and
The incident radiation is measured after the grid
has been removed
Grid ratio 70 kVp 120 kVp
No grid 1 1
5:1 3 3
8:1 3.5 4
12:1 4 5
16:1 4.5 6
High ratio grids absorb more scatter radiation
and have larger Bucky factors than low-ratio
grids
Higher energy beams generate more scatter
radiation and place a greater demand on a grid’s
performance than lower energy radiation.
The higher the Bucky factor,the
greater the exposure factors and
radiation dosage to the patient.
If the Bucky factor for a particular grid-
energy combination is 5, then exposure
factors and patient exposure both
increase 5 times
Contrast improvement factor (K)
The contrast improvement factor(K) is the ratio of
the contrast with a grid to the contrast without a
grid.
contrast with a grid
K = _________________
contrast without a grid
Is the ultimate test of grid performance.
The contrast improvement factor is dependent on
kVp,field size and phantom thickness.
These three factors determine the amount of
scatter radiation
The larger the amount of scatter radiation,the
poorer the contrast, and the lower the
contrast improvement factor.
It is more closely related to the lead content
of the grid than any other factor.
Generally, the higher the grid ratio,the
higher the contrast improvement factor.
Lead content
The lead content of a grid is expressed in gm/cm2
Imagine cutting a grid up in to 1 cm squares and then
weighing one square.
Its weight in grams is the lead content of the
grid(ignoring the interspace material)
The amount of lead in a grid is a good indicator of
its ability to improve contrast.
There is a definite relationship between the grid
ratio,lead content and the number of lines per inch.
when grids are constructed with many lines per
inch, both the thickness and height of the lead
strips are decreased.
These grids are thinner, and improve contrast
less than grids of comparable ratios with fewer
lines per inch.
Grid cut off
Grid cut off is the loss of primary radiation that occurs when
the images of the lead strips are projected wider than they
would be with ordinary magnification
It is the result of a poor geometric
relation ship between the primary
beam and the lead foil strips .
The resultant radiograph will be
light in the area in which the cutoff
occurs.
With linear grids there may be
uniform lightening of the whole film,
one edge of the film ,or both edges
of the film,depending on how the
cutoff is produced.
The amount of cutt off is always greatest with high
ratio grids and short grid focus distance
There are 4 situations that produce grid cutt
off
1.Focussed grids used upside down
2.Lateral decentering
3.Focussed grid distance decentering
4.Combined lateral and focus grid distance
decentering
Upside down focussed grids
When a focused grid is used upside down, there is
severe peripheral cutoff with a dark band of
exposure in the center of the film and no exposure at
the film’s periphery
Lateral decentering
Results from the X-ray tube being positioned
lateral to the convergent line but at the correct
focal distance
All the lead strips cutoff the same amount of primary
radiation,so there is a uniform loss of radiation over
the entire surface of the grid ,producing a uniformly
light radiograph
This is probably the most common kind of grid
cutoff, but it cannot be recognised by inspection of
the film.
All we see is a light film (that is usually attributed to
incorrect exposure factors.)
The films become progressively lighter as the
amount of lateral decentering increases.
Three factors affect the magnitude of
cutoff :
Grid ratio
Focal distance and
Amount of decentering
The equation for calculating the loss of primary radiation
with lateral decentering is
L=rb/fo x 100
L= loss of primary radiation(%)
r=grid ratio
b=lateral decentering distance (inches)
fo =focal distance of grid (inches)
When exact centering is not
possible , as in portable
radiography, low ratio grids
and long focal distances should
be used whenever possible
Off level grids
When a linear grid is tilted , as it frequently is
in portable radiography, there is a uniform loss
of primary radiation across the entire surface
of the grid
Focus-grid distance decentering
In focus -grid distance decentering , the target of the X-ray
tube is correctly centered to the grid , but it is positioned
above or below the convergent line.
If the target is above the convergent line ,it is called FAR
focus-grid distance decentering
If the target is below the convergent line , it is called NEAR
focus-grid distance decentering.
The cut off is greater with NEAR than with FAR focus-grid
distance decentering.
The cutoff becomes progressively greater with increasing
distance from the film center.
The central portion of the film is not affected, but the
periphery is light.
Parallel grids are focused at infinity
A film taken with a parallel grid has a dark center
and light edges because of near focus-grid distance
decentering
Combined lateral & focus-grid distance decentering
The most commonly recognised kind of grid cutoff is from
combined lateral and focus grid distance decentering
It causes an uneven exposure,resulting in a film that is light
on one side and dark on the other side.
The projected images of the lead strips directly below the
tube target are broader than those on the opposite side, and
the film is light on the near side.
Cutoff is greatest on the side directly under the Xray tube.
The projected images of the lead strips are broader
on the side opposite the tube target than on the
same side, and the film is light on the far side.
Cutoff is least on the side under the Xray tube.
Moving grids
Grids are moved to blur out the shadows cast by
the lead strips.
Most grids are reciprocating,which means they
continuously move 1 to 3 cms back and forth
through out the exposure.
They start moving when the Xray tube anode
begins to rotate.
They eliminate grid lines from the film
Moving grids precautions
The grid must move fast enough to blur its lead
strips
The transverse motion of the grid should be
synchronous with the pulses of the Xray
generator
Disadvantages
They are costly
Subject to failure
May vibrate the Xray table
Put a limit on the minimum exposure
time because they move slowly
increase the patient’s radiation dose
Grid selection
Usually 8:1 grid will give adequate results below 90kVp.
Above 90kVp,12:1 grids are preferred.
There is little decrease in
transmitted scatter beyond an
8:1 ratio grid,
And almost no change
between 12:1 and 16:1
For this reason12:1 grids are
preferable to 16:1 grids for
routine radiography
Air gap technique
Scattered radiation arising in a patient
from comption reactions is dispersed in all
directions.
With an air gap the concentration of
scatterd radiation decreases because
scatters photons fail to reach the film
Used in 2 clinical situations
• Magnification radiology
• Chest radiology
With magnification techniques the object -film
distance is optimised for the screen focal spot
combination and the air gap technique reduces
the scatter radiation
In chest radiography the focal film distance is
usually lengthened from 6-10 ft to restore
sharpness
Exposure factors with air gaps
X-ray tube exposure must be increased for
the air gap technique because of larger
focal film distance
Patients exposures are usually less with
air gap technique
The air gap loses less primary radiation,
so the patient’s exposure is less
THANK YOU

More Related Content

What's hot

X ray beam rest
X ray beam restX ray beam rest
X ray beam restBuandari
 
CT History
CT HistoryCT History
CT Historybartonect
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imgingSahith Reddy
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
Radiographic grid Swastik
Radiographic grid   SwastikRadiographic grid   Swastik
Radiographic grid SwastikBipul Poudel
 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiographyambar hazarika
 
Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiographyDurga Singh
 
Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tubeSantosh Ojha
 
Production and control of scatter radiation (beam
Production and control of scatter radiation (beamProduction and control of scatter radiation (beam
Production and control of scatter radiation (beamSUJAN KARKI
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Sarin Vidya
 
Computed radiography
Computed radiographyComputed radiography
Computed radiographyAnjan Dangal
 
Beam restriction
Beam restrictionBeam restriction
Beam restrictionaslam bs
 
Grids
GridsGrids
Gridsaslam bs
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...DrKajalLimbad
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageSHASHI BHUSHAN
 
Direct digital radiography(1) (1)
Direct digital radiography(1) (1)Direct digital radiography(1) (1)
Direct digital radiography(1) (1)Tarun Kumar
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct Muntaser S.Ahmad
 

What's hot (20)

X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
 
CT History
CT HistoryCT History
CT History
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Radiographic grid Swastik
Radiographic grid   SwastikRadiographic grid   Swastik
Radiographic grid Swastik
 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiography
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiography
 
Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tube
 
Production and control of scatter radiation (beam
Production and control of scatter radiation (beamProduction and control of scatter radiation (beam
Production and control of scatter radiation (beam
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968Xray beam restrictors 1436711280968
Xray beam restrictors 1436711280968
 
Hypotonic duodenography
Hypotonic duodenographyHypotonic duodenography
Hypotonic duodenography
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Beam restriction
Beam restrictionBeam restriction
Beam restriction
 
Grids
GridsGrids
Grids
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic image
 
Direct digital radiography(1) (1)
Direct digital radiography(1) (1)Direct digital radiography(1) (1)
Direct digital radiography(1) (1)
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 

Similar to Filters.ppt

Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptXray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptNagasai Pelala
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Arya Prasad
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiationNitish Rajput
 
Radiology.pptx
Radiology.pptxRadiology.pptx
Radiology.pptxkhaziathar777
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimatorsGeorgina George
 
Analytical instruments introduction
Analytical instruments introductionAnalytical instruments introduction
Analytical instruments introductionSadiq Rahim
 
X ray filters
X ray filtersX ray filters
X ray filtersSaruGosain
 
grids-160225101450 (1).pptx
grids-160225101450 (1).pptxgrids-160225101450 (1).pptx
grids-160225101450 (1).pptxVishnuDutt40
 
Dual Mode Dual Band pass Filter Using Circular Patch Antenna.
Dual Mode Dual Band pass Filter Using Circular Patch Antenna.Dual Mode Dual Band pass Filter Using Circular Patch Antenna.
Dual Mode Dual Band pass Filter Using Circular Patch Antenna.Aishwary Singh
 
Beam restriction technique and devices.pptx
Beam restriction technique and devices.pptxBeam restriction technique and devices.pptx
Beam restriction technique and devices.pptxVANI PUSHPA MUDAVATH
 
Spectrophotometer.ppt (T.Y. Bsc).pptx
Spectrophotometer.ppt (T.Y. Bsc).pptxSpectrophotometer.ppt (T.Y. Bsc).pptx
Spectrophotometer.ppt (T.Y. Bsc).pptxJinalMehta41
 
MicroStrip Antenna
MicroStrip AntennaMicroStrip Antenna
MicroStrip AntennaTarek Nader
 
grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptxrohanjohnjacob
 

Similar to Filters.ppt (20)

Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptXray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.
 
The control of scattered radiation
The control of scattered radiationThe control of scattered radiation
The control of scattered radiation
 
Radiology.pptx
Radiology.pptxRadiology.pptx
Radiology.pptx
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
 
Analytical instruments introduction
Analytical instruments introductionAnalytical instruments introduction
Analytical instruments introduction
 
X ray filters
X ray filtersX ray filters
X ray filters
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
 
Grids JP .ppt
Grids JP .pptGrids JP .ppt
Grids JP .ppt
 
grids-160225101450 (1).pptx
grids-160225101450 (1).pptxgrids-160225101450 (1).pptx
grids-160225101450 (1).pptx
 
Dual Mode Dual Band pass Filter Using Circular Patch Antenna.
Dual Mode Dual Band pass Filter Using Circular Patch Antenna.Dual Mode Dual Band pass Filter Using Circular Patch Antenna.
Dual Mode Dual Band pass Filter Using Circular Patch Antenna.
 
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTSUnit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
 
Grids
GridsGrids
Grids
 
Neelotpal Mitra
Neelotpal MitraNeelotpal Mitra
Neelotpal Mitra
 
Beam restriction technique and devices.pptx
Beam restriction technique and devices.pptxBeam restriction technique and devices.pptx
Beam restriction technique and devices.pptx
 
X ray
X rayX ray
X ray
 
Spectrophotometer.ppt (T.Y. Bsc).pptx
Spectrophotometer.ppt (T.Y. Bsc).pptxSpectrophotometer.ppt (T.Y. Bsc).pptx
Spectrophotometer.ppt (T.Y. Bsc).pptx
 
MicroStrip Antenna
MicroStrip AntennaMicroStrip Antenna
MicroStrip Antenna
 
grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptx
 

More from jayapandiyan Paraman (16)

Heat : Heat is a form of energy produced by natural and artificial sources
Heat : Heat is a form of energy produced by natural and artificial sourcesHeat : Heat is a form of energy produced by natural and artificial sources
Heat : Heat is a form of energy produced by natural and artificial sources
 
Microwave Diathermy JP.ppt
Microwave Diathermy JP.pptMicrowave Diathermy JP.ppt
Microwave Diathermy JP.ppt
 
Therapatic Currents.ppt
Therapatic Currents.pptTherapatic Currents.ppt
Therapatic Currents.ppt
 
LED JP.ppt
LED  JP.pptLED  JP.ppt
LED JP.ppt
 
ultrasound JP.pptx
ultrasound JP.pptxultrasound JP.pptx
ultrasound JP.pptx
 
Sound.ppt
Sound.pptSound.ppt
Sound.ppt
 
Sound waves.ppt
Sound waves.pptSound waves.ppt
Sound waves.ppt
 
SOUND JP .ppt
SOUND JP .pptSOUND JP .ppt
SOUND JP .ppt
 
usg diplay mode.pptx
usg diplay mode.pptxusg diplay mode.pptx
usg diplay mode.pptx
 
ultrasound JP.pptx
ultrasound JP.pptxultrasound JP.pptx
ultrasound JP.pptx
 
X ray generator new.ppt
X ray generator new.pptX ray generator new.ppt
X ray generator new.ppt
 
Radioactivity.pptx
Radioactivity.pptxRadioactivity.pptx
Radioactivity.pptx
 
spect .pptx
spect .pptxspect .pptx
spect .pptx
 
radio pharmaceuticals.ppt
radio pharmaceuticals.pptradio pharmaceuticals.ppt
radio pharmaceuticals.ppt
 
Flouroscopy JP.pptx
Flouroscopy JP.pptxFlouroscopy JP.pptx
Flouroscopy JP.pptx
 
Nuclear medicine Img.pptx
Nuclear medicine Img.pptxNuclear medicine Img.pptx
Nuclear medicine Img.pptx
 

Recently uploaded

Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 

Recently uploaded (20)

Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 

Filters.ppt

  • 1. Filters, Collimators and Grids Christensen’s physics of diagnostic radiology
  • 2. Filters Filtration is the process of shaping the X-ray beam to increase the ratio of photons useful for imaging to those photons that increase patient dose or decrease image contrast
  • 3. Diagnostic X-ray beam is polychromatic, comprising of whole spectrum of energies. The mean energy will vary from 1/3rd to one half of the their peak energy. The first few centimeters of the tissues receive much more radiation than the rest of the body tissues of the patient
  • 4. Filters Filters are sheets of metal, attached at the opening of tube housing, which will absorb the low energy photons from the X-ray beam before it reaches the patient
  • 5. Types of filtration The X-ray beam is filtered by absorbers at three different levels Inherent Filtration Added filtration Patient
  • 6. Inherent filtration Glass envelope The insulating oil surrounding the tube Window in the tube housing
  • 7. Added filtration Aluminum At.No. 13 Copper At. No. 29 Compound Filter copper + Aluminum
  • 8. Advantage of compound filtration Copper is used to cut down the thickness of filter Copper will absorb high energy photons and Aluminum will absorb the characteristic radiation from copper (8 keV)
  • 9. Measurement of filtration Filtration is measured in Aluminum equivalents which is defined as the thickness of the Aluminum that would produce the same degree of attenuation as the thickness of material in question
  • 10. Effects of filtration Patient exposure and Exposure factors
  • 11. 60 –kVp beam Aluminum Exposure dose Decrease in filtration (mm) to skin (mR) exposure dose (%) None 2380 0.5 1850 22 1.0 1270 47 3.0 465 80
  • 12. Inherent filtration at the tube housing is 0.5 to 1.0 Al.Eq. Below 50 kVp 0.5 mm Aluminum 50-70 kVp 1.5 mm Aluminum Above 70 kVp 2.5 mm Aluminum
  • 13. Effect on exposure factors There will be reduction in the intensity of X-ray beam as the filters absorb some photons at all energy levels. To Compensate the loss of high energy photons, increase in the exposure factors (mAs) is required
  • 14. Types of filters Single Filter Aluminum Compund Filter Aluminum +Copper Wedge Filter Used in the angiography Molybdenum Filters used in Mammography. K. Edge filter
  • 15. Molybdenum filters Used in molybdenum target X-ray tubes used for mammography 17.5 kev K alpha and 19.6 kev K beta characterstic radiation of Mo When operated at 30-40 kVp, Mo will produce bremsstrahling with energies higher than 20 kev Mo filter attenuates these high energy rays
  • 16. K- Edge filters These filters make use of K absorption edge of elements with atomic No. greater than 60. The purpose of heavy metal filters or K edge filters is to produce an X-ray beam that has a high number of photons in the specific energy range Enhance contrast for Iodine and barium, reduce patient dose, and increase tube loading
  • 18. X-ray beam restrictors An X-ray beam restrictor is a device that is attached to the opening in the X-ray tube housing to regulate the size and shape of the X-ray beam
  • 19. Types of restrictors Aperture diaphragms Cones & cylinders Collimators
  • 20. Aperture diaphragm It consists of a sheet of lead with a hole in the center. The size and shape of the hole determine the size and shape of the X-ray beam.
  • 21. Cones Cones are usually flare shaped Ideal geometric configuration for an X-ray beam restrictor. The flare of the cone is greater than the flare of the x ray beam.
  • 22. Cylinders Beam restriction with cylinder takes place at the far end of the barrel, so there is less penumbra.
  • 24. Disadvantages Penumbra: Partially exposed periphery of the X-ray field is called penumbra Another major disadvantage with cones and cylinders is severe limitations they place on the number of available field sizes
  • 25. Collimators Collimator is the best all round X-ray beam restrictor These are two types: Manual collimator Automatic collimator or PBL (Positive beam limitation device or automatic light localised variable aperture collimator)
  • 26. Advantages It provides infinite variety of rectangular or square X-ray fields The light beam shows the exact center and configuration of X-ray field
  • 27. Construction of collimator X ray filter mirror X ray &light beam bulb Lower shutter Collimator shutters
  • 28. Testing of X- ray beam and light beam alignment Periodic check of alignment of X-ray beam and light beam is essential, because the mirror gets out of adjustment due to frequent daily use.
  • 29. Automatic collimators When the cassette is loaded in the Buckey tray the sensors in the tray identify the size and alignment of the cassette and relay the information to collimator motors, which positions shutters to the exact size of the film used.
  • 30. Functions of X ray beam restrictors Protects the patient from unnecessary radiation It decreases the scatter radiation
  • 31. The number of scattered radiation depends upon field size Small fields generate little scatter, as the field increases scatter increases Collimators are only successful in decreasing the scatter radiation with small fields, so we should reduce the size of X-ray beams as much as possible
  • 32. GRIDS
  • 33. The radiographic grid consists of a series of lead foil strips separated by X ray transparent spacers. It was invented by DR.GUSTAVE BUCKY in 1913. Grid is still the most effective way of removing the scatter radiation from large radiographic fields
  • 34. Primary radiation is oriented in the same axis as the lead strips and passes between them . Scatter radiation arises from many points within the patient and most of it is absorbed by the lead strips
  • 35. The interspaces of the grids are filled either with aluminium or some organic compound. The main purpose of the interspace material is to support the thin lead foil strips.
  • 36. Is defined as the ratio between the height of the lead strips and the distance between them. GRID RATIO
  • 37. The lead strips are approximately 0.05 mm thick ( lead foil). The interspaces are much thicker. Grid ratios are usually expressed as two numbers, such as 20:1 Ratios usually range from 4:1 to 16:1 the Higher the ratio, the better the grid functions.
  • 38. Grid pattern Is the orientation of the lead strips in their longitudinal axis. The two basic grid patterns are : Linear and Crossed.
  • 39. They allow us to angle the x-ray tube along the length of the grid . Linear grid
  • 40. A crossed grid is made up of two superimposed linear grids that have the same focussing distance The grid ratio of crossed grids is equal to the sum of the ratios of the two linear grids. A crossed grid made up of two 5:1 linear grids has a ratio of 10:1. Crossed grids cannot be used with oblique techniques requiring angulation of the X-ray tube Crossed grids
  • 41. Is a grid made up of lead strips that are angled slightly so that they focus in space. A focussed grid may be either linear or crossed. Linear focused grids converge at a line in space called the convergent line. Crossed grids converge at a point in space called the convergent point. The focal distance is the perpendicular distance between the grid and the convergent line or point. Focussed grid
  • 42. Focussing range Indicates the distance within which the grid can be used without significant loss of primary radiation It is fairly wide for a low-ratio grid and narrow for a high ratio grid. A 5:1 grid focused at 40 inches has a focusing range of approximately 28 to 72 inches. While a 16:1 grid focused at 40inches has a range of only 38 to 42inches.
  • 43. Parallel grid A parallel grid is one in which the lead strips are parallel They are focused at infinity. can only be used with either very small Xray fields or long- target grid distances. They are frequently used in fluoroscopic spot film devices.
  • 44. Lines per inch Is the number of lead strips per inch of the grid. Lines per inch = 25.4/D+d D= thickness of the interspaces d=thickness of the lead strips(both in millimeters)
  • 45. Grid cassette Usually used for portable radiography , with a grid built in to the front of the cassette. Are focussed and Have a grid ratio of 4:1 to 8:1
  • 46. Evaluation of grid performance The three methods of evaluating grid performance: 1.Primary transmission(Tp) 2.Bucky factor (B) 3.Contrast improvement factor(K)
  • 47. Primary transmission Is the percentage of primary radiation transmitted through the grid. Ideally , a grid should transmit 100% of the primary radiation. The first measurement is made with the grid in place The second measurement is made after removal of the grid
  • 48. A ratio of the intensity with the grid to the intensity without the grid gives the fractional transmission, which is multiplied by 100 to give the percentage of transmission. intensity with grid Ip Tp = _____________________ X100 intensity without grid I’p There is a significant loss of primary radiation with grids, more with cross grids
  • 49. Measured primary tranmission The measured primary transmission is always less than the calculated primary transmission: absorption by the interspace material manufacturing imperfections
  • 50. Bucky factor Is the ratio of the incident radiation falling on the grid to the transmitted radiation passing through the grid. It indicates how much we must increase exposure factors when we change from a non grid to a grid technique. The Bucky factor indicates the absorption of both primary and secondary radiation. It is determined with a large X-ray field and a thick phantom
  • 51. The Bucky factor( B) is a measure of the total quantity of radiation absorbed from an X-ray beam by a grid incident radiation B = ____________________ transmitted radiation The transmitted radiation is measured with the grid in place ,and The incident radiation is measured after the grid has been removed
  • 52. Grid ratio 70 kVp 120 kVp No grid 1 1 5:1 3 3 8:1 3.5 4 12:1 4 5 16:1 4.5 6 High ratio grids absorb more scatter radiation and have larger Bucky factors than low-ratio grids Higher energy beams generate more scatter radiation and place a greater demand on a grid’s performance than lower energy radiation.
  • 53. The higher the Bucky factor,the greater the exposure factors and radiation dosage to the patient. If the Bucky factor for a particular grid- energy combination is 5, then exposure factors and patient exposure both increase 5 times
  • 54. Contrast improvement factor (K) The contrast improvement factor(K) is the ratio of the contrast with a grid to the contrast without a grid. contrast with a grid K = _________________ contrast without a grid Is the ultimate test of grid performance. The contrast improvement factor is dependent on kVp,field size and phantom thickness. These three factors determine the amount of scatter radiation
  • 55. The larger the amount of scatter radiation,the poorer the contrast, and the lower the contrast improvement factor. It is more closely related to the lead content of the grid than any other factor. Generally, the higher the grid ratio,the higher the contrast improvement factor.
  • 56. Lead content The lead content of a grid is expressed in gm/cm2 Imagine cutting a grid up in to 1 cm squares and then weighing one square. Its weight in grams is the lead content of the grid(ignoring the interspace material) The amount of lead in a grid is a good indicator of its ability to improve contrast. There is a definite relationship between the grid ratio,lead content and the number of lines per inch.
  • 57. when grids are constructed with many lines per inch, both the thickness and height of the lead strips are decreased. These grids are thinner, and improve contrast less than grids of comparable ratios with fewer lines per inch.
  • 58. Grid cut off Grid cut off is the loss of primary radiation that occurs when the images of the lead strips are projected wider than they would be with ordinary magnification It is the result of a poor geometric relation ship between the primary beam and the lead foil strips . The resultant radiograph will be light in the area in which the cutoff occurs. With linear grids there may be uniform lightening of the whole film, one edge of the film ,or both edges of the film,depending on how the cutoff is produced.
  • 59. The amount of cutt off is always greatest with high ratio grids and short grid focus distance There are 4 situations that produce grid cutt off 1.Focussed grids used upside down 2.Lateral decentering 3.Focussed grid distance decentering 4.Combined lateral and focus grid distance decentering
  • 60. Upside down focussed grids When a focused grid is used upside down, there is severe peripheral cutoff with a dark band of exposure in the center of the film and no exposure at the film’s periphery
  • 61. Lateral decentering Results from the X-ray tube being positioned lateral to the convergent line but at the correct focal distance
  • 62. All the lead strips cutoff the same amount of primary radiation,so there is a uniform loss of radiation over the entire surface of the grid ,producing a uniformly light radiograph This is probably the most common kind of grid cutoff, but it cannot be recognised by inspection of the film. All we see is a light film (that is usually attributed to incorrect exposure factors.) The films become progressively lighter as the amount of lateral decentering increases.
  • 63. Three factors affect the magnitude of cutoff : Grid ratio Focal distance and Amount of decentering The equation for calculating the loss of primary radiation with lateral decentering is L=rb/fo x 100 L= loss of primary radiation(%) r=grid ratio b=lateral decentering distance (inches) fo =focal distance of grid (inches)
  • 64. When exact centering is not possible , as in portable radiography, low ratio grids and long focal distances should be used whenever possible
  • 65. Off level grids When a linear grid is tilted , as it frequently is in portable radiography, there is a uniform loss of primary radiation across the entire surface of the grid
  • 66. Focus-grid distance decentering In focus -grid distance decentering , the target of the X-ray tube is correctly centered to the grid , but it is positioned above or below the convergent line. If the target is above the convergent line ,it is called FAR focus-grid distance decentering If the target is below the convergent line , it is called NEAR focus-grid distance decentering. The cut off is greater with NEAR than with FAR focus-grid distance decentering. The cutoff becomes progressively greater with increasing distance from the film center.
  • 67. The central portion of the film is not affected, but the periphery is light.
  • 68. Parallel grids are focused at infinity A film taken with a parallel grid has a dark center and light edges because of near focus-grid distance decentering
  • 69. Combined lateral & focus-grid distance decentering The most commonly recognised kind of grid cutoff is from combined lateral and focus grid distance decentering It causes an uneven exposure,resulting in a film that is light on one side and dark on the other side. The projected images of the lead strips directly below the tube target are broader than those on the opposite side, and the film is light on the near side. Cutoff is greatest on the side directly under the Xray tube.
  • 70. The projected images of the lead strips are broader on the side opposite the tube target than on the same side, and the film is light on the far side. Cutoff is least on the side under the Xray tube.
  • 71. Moving grids Grids are moved to blur out the shadows cast by the lead strips. Most grids are reciprocating,which means they continuously move 1 to 3 cms back and forth through out the exposure. They start moving when the Xray tube anode begins to rotate. They eliminate grid lines from the film
  • 72. Moving grids precautions The grid must move fast enough to blur its lead strips The transverse motion of the grid should be synchronous with the pulses of the Xray generator
  • 73. Disadvantages They are costly Subject to failure May vibrate the Xray table Put a limit on the minimum exposure time because they move slowly increase the patient’s radiation dose
  • 74. Grid selection Usually 8:1 grid will give adequate results below 90kVp. Above 90kVp,12:1 grids are preferred. There is little decrease in transmitted scatter beyond an 8:1 ratio grid, And almost no change between 12:1 and 16:1 For this reason12:1 grids are preferable to 16:1 grids for routine radiography
  • 75. Air gap technique Scattered radiation arising in a patient from comption reactions is dispersed in all directions. With an air gap the concentration of scatterd radiation decreases because scatters photons fail to reach the film
  • 76. Used in 2 clinical situations • Magnification radiology • Chest radiology With magnification techniques the object -film distance is optimised for the screen focal spot combination and the air gap technique reduces the scatter radiation In chest radiography the focal film distance is usually lengthened from 6-10 ft to restore sharpness
  • 77. Exposure factors with air gaps X-ray tube exposure must be increased for the air gap technique because of larger focal film distance Patients exposures are usually less with air gap technique The air gap loses less primary radiation, so the patient’s exposure is less