SlideShare a Scribd company logo
1 of 42
FUNDAMENTAL
S OF IMAGING
RAD 206
Control of Scatter Radiation
What is focal spot?
What is effective focal spot ?
Which side of the x-ray tube has better resolution ?
Which focal spot size produces better resolution?
Which OID produces better resolution?
Which Object Thickness produces better resolution?
Which Object shape produces better resolution?
OBJECTIVES
Begin discussing factors that influence image detail or
visibility of detail
• Scatter Radiation, factors affecting it, and there effect on
Spatial & Contrast resolution
• Ways to reduce scatter & improve image quality
• Primary beam restriction & Grids
• Technique adjustments when using grids
WHAT IS CONTRAST
What is scatterWhat is scatter
RELATIONSHIP BETWEEN
CONTRAST AND SCATTER
The relationship is INVERSED
IMAGE-FORMING X-RAYS
Two kinds of photons are responsible for
the OD and contrast on an image:
• Photons that pass through
without interacting
and
• Photons that are scattered
though Compton.
EFFECT OF SCATTER
ON THE IMAGE
Which interaction produces scatter radiation ?
FACTORS CONTRIBUTING
TO AN INCREASE IN
SCATTER
• Increased kVpkVp (Beam Energy, Compton)
• Increased partpart thicknessthickness (e.g. Dabbeh)
• Increased x-ray field sizefield size (opening
Collimation to Larger Fields)
Q: what can you do to reduce scatter ?Q: what can you do to reduce scatter ?
1- KVP
At 50 kVp
7979% photoelectric,
21%21% Compton
<1% transmission
At 80 kVp
4646% photoelectric,
52%52% Compton
2% transmission
RELATIONSHIP BETWEEN
KVP AND CONTRAST
How does increasing kVp affect patient dose?
2- PATIENT THICKNESS
Imaging thick parts of the body results in
more scatter radiation than thin parts
IMAGE TEST TOOL
Is patient thickness something the
radiographer can control?
IS PATIENT THICKNESS SOMETHING THEIS PATIENT THICKNESS SOMETHING THE
RADIOGRAPHER CAN CONTROL?RADIOGRAPHER CAN CONTROL?
Normally, No
But Compression Devices can be used in
certain situations which reduces patient
thickness and bringing the object closer
to the IR, and therefore results in better
spatial resolution and contrast resolution
and also lesser exposure factors and
patient dose
COMPRESSION
• Reduces OID
• Improves spatial
resolution
• Improves contrast
resolution
(reducing fog or
noise)
• Reduces patient
dose
COMPRESSION
3- FIELD SIZE
As field size increases, intensity of scatter
radiation also increases rapidly. Especially during
fluoroscopy
Larger radiation field produces more
scatter radiation
COMPARE IMAGES: WHAT DO YOU THINK
ABOUT RADIOGRAPHIC CONTRAST & IMAGE
NOISE?
INCREASE IN SCATTER
• kVpkVp
• PartPart ThicknessThickness
• Field SizeField Size
CONTROL OF
SCATTER RADIATION
Technologists routinely use two types of devices
to reduce the amount of scatter radiation
reaching the IR
Beam Restrictors
Compression
Grids
Before Patient
On patient
After patient
BEAM-RESTRICTING DEVICES
AKA LBD
1. Aperture
Diaphragm
1. Cones or
Cylinders
2. Variable aperture
collimator
APERTURE DIAPHRAGM
The simplest of all beam-restricting
devices
Lead or lead-lined metal diaphragm
attached to the x-ray tube head
The opening in the diaphragm is usually
designed to cover just less than the IR
used
APERTURE DIAPHRAGM
Fixed lead opening
Fixed image receptor
size
Constant SID
Source-to-diaphragm
distance = SDD
CONES & CYLINDERS
Are modifications of the aperture diaphragm
Alignment
is one difficulty
when using
cones
Now mostly
used with spines
teeth & heads
IMPROVED CONTRAST RESOLUTION OF
THE L5 - S1
IMPROVED CONTRAST RESOLUTION OF
THE FRONTAL SINUSES
VARIABLE APERTURE
COLLIMATOR
The most common beam-
restricting device is the
light-localizing variable
aperture collimator
The first part of the
collimator serves to
control off-focus
radiation. What is off-
focus radiation?
OFF - FOCUS RADIATION
ALSO CALLED EXTRA-FOCAL RADIATION
X-ray tubes are designed so that the projectile e-
interacts with the target. However, some of the e-
bounce off the target and land on other areas
This caused x-rays to be produced out side the
focal spot, they are undesirable because it
extends the size of the focal spot, increases
patient skin dose & reduces image contrast
Note: Using a grid does not reduce off-focal radiation
FIRST-STAGE ENTRANCE
SHUTTERING DEVICE
Has multiple
collimator blades
protruding from the
top of the collimator
into the tube housing
THE SECOND-STAGE COLLIMATOR
SHUTTERS
Pb leaves are at least
3 mm thick
They work in pairs and
are independently
controlled
THE COLLIMATOR LAMP & MIRROR
Must be adjusted so
that the projected light
field coincides with
the x-ray beam
Misalignment of the
light field and beam
can result in
collimator cutoff of
anatomic structures
Always keep the collimated area
smaller than the size of the
cassette
What is a PBL
BLD
GRIDS
QUESTIONS …?
Technique adjustments
problems

More Related Content

What's hot

Line focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectLine focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effect
JoaquinLuisRinehart1
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
parthajyotidas11
 

What's hot (20)

Radiographic factor
Radiographic factorRadiographic factor
Radiographic factor
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 
X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
 
Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tube
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolution
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
kV and mA Meters
kV and mA MeterskV and mA Meters
kV and mA Meters
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 
Line focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effectLine focus-principle-and-anode-heel-effect
Line focus-principle-and-anode-heel-effect
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging System
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
computed and digital radiography(2)
computed and digital radiography(2)computed and digital radiography(2)
computed and digital radiography(2)
 
X ray films - mamita
X ray films - mamitaX ray films - mamita
X ray films - mamita
 
Radiographic film ; intensifying screens
Radiographic film ; intensifying screensRadiographic film ; intensifying screens
Radiographic film ; intensifying screens
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Filters.ppt
Filters.pptFilters.ppt
Filters.ppt
 
X ray generator, circuits
X  ray generator, circuitsX  ray generator, circuits
X ray generator, circuits
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 

Similar to Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation

Image quality, digital technology and radiation protection
Image quality, digital technology and radiation protectionImage quality, digital technology and radiation protection
Image quality, digital technology and radiation protection
Rad Tech
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGY
Paul George
 

Similar to Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation (20)

Anode Heal Effect
Anode Heal EffectAnode Heal Effect
Anode Heal Effect
 
Lecture 10 focusing and colimating
Lecture 10 focusing and colimatingLecture 10 focusing and colimating
Lecture 10 focusing and colimating
 
Cbct
CbctCbct
Cbct
 
Aberrometry
AberrometryAberrometry
Aberrometry
 
Image quality
Image qualityImage quality
Image quality
 
Image quality, digital technology and radiation protection
Image quality, digital technology and radiation protectionImage quality, digital technology and radiation protection
Image quality, digital technology and radiation protection
 
Isodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGYIsodose curves RADIATION ONCOLOGY
Isodose curves RADIATION ONCOLOGY
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopy
 
5.1. lithography 1,2.final 2013
5.1. lithography 1,2.final 20135.1. lithography 1,2.final 2013
5.1. lithography 1,2.final 2013
 
Beam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarBeam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattar
 
Production and control of scatter radiation (beam
Production and control of scatter radiation (beamProduction and control of scatter radiation (beam
Production and control of scatter radiation (beam
 
Chapter 5 Lithography _ II.pptx
Chapter 5 Lithography _ II.pptxChapter 5 Lithography _ II.pptx
Chapter 5 Lithography _ II.pptx
 
Laser in ophthalmology
Laser in ophthalmologyLaser in ophthalmology
Laser in ophthalmology
 
Radiographic parameters
Radiographic parametersRadiographic parameters
Radiographic parameters
 
X-ray Machine
X-ray Machine X-ray Machine
X-ray Machine
 
Cone beam ct
Cone beam ctCone beam ct
Cone beam ct
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.
 
Flim screen physics
Flim screen physicsFlim screen physics
Flim screen physics
 
Image characteristics,latent image,film processing.
Image characteristics,latent image,film processing.Image characteristics,latent image,film processing.
Image characteristics,latent image,film processing.
 

More from sehlawi

More from sehlawi (19)

Rad 206 p13
Rad 206 p13Rad 206 p13
Rad 206 p13
 
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p12 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
 
Rad 206 p10 Fundamentals of Imaging -
Rad 206 p10 Fundamentals of Imaging - Rad 206 p10 Fundamentals of Imaging -
Rad 206 p10 Fundamentals of Imaging -
 
Rad 206 p05 Fundamentals of Imaging - Fluoroscopy
Rad 206 p05 Fundamentals of Imaging - FluoroscopyRad 206 p05 Fundamentals of Imaging - Fluoroscopy
Rad 206 p05 Fundamentals of Imaging - Fluoroscopy
 
Rad 206 p03
Rad 206 p03Rad 206 p03
Rad 206 p03
 
Rad 206 p09
Rad 206 p09Rad 206 p09
Rad 206 p09
 
Rad 206 p08
Rad 206 p08Rad 206 p08
Rad 206 p08
 
Rad 206 p04
Rad 206 p04Rad 206 p04
Rad 206 p04
 
Rad 206 introduction
Rad 206 introductionRad 206 introduction
Rad 206 introduction
 
Rad 206 p02
Rad 206 p02Rad 206 p02
Rad 206 p02
 
Rad 206 p01
Rad 206 p01Rad 206 p01
Rad 206 p01
 
Rad 104 hospital practice and care of patients 8 types of catheters 2016
Rad 104 hospital practice and care of patients 8  types of catheters 2016Rad 104 hospital practice and care of patients 8  types of catheters 2016
Rad 104 hospital practice and care of patients 8 types of catheters 2016
 
Rad 104 hospital practice and care of patients 7 safe patient movement and h...
Rad 104 hospital practice and care of patients 7  safe patient movement and h...Rad 104 hospital practice and care of patients 7  safe patient movement and h...
Rad 104 hospital practice and care of patients 7 safe patient movement and h...
 
Rad 104 hospital practice and care of patients 6 infection control 2016
Rad 104 hospital practice and care of patients 6  infection control 2016Rad 104 hospital practice and care of patients 6  infection control 2016
Rad 104 hospital practice and care of patients 6 infection control 2016
 
Rad 104 hospital practice and care of patients 5 anesthesia 2016
Rad 104 hospital practice and care of patients 5  anesthesia 2016Rad 104 hospital practice and care of patients 5  anesthesia 2016
Rad 104 hospital practice and care of patients 5 anesthesia 2016
 
Rad 104 hospital practice and care of patients 3 drugs &amp; contrast 2016
Rad 104 hospital practice and care of patients 3  drugs &amp; contrast 2016Rad 104 hospital practice and care of patients 3  drugs &amp; contrast 2016
Rad 104 hospital practice and care of patients 3 drugs &amp; contrast 2016
 
Rad 104 hospital practice and care of patients 2 terminology p2 - 2016
Rad 104 hospital practice and care of patients 2  terminology p2 - 2016Rad 104 hospital practice and care of patients 2  terminology p2 - 2016
Rad 104 hospital practice and care of patients 2 terminology p2 - 2016
 
Rad 104 hospital practice and care of patients 1 objectives &amp; terminolog...
Rad 104 hospital practice and care of patients 1  objectives &amp; terminolog...Rad 104 hospital practice and care of patients 1  objectives &amp; terminolog...
Rad 104 hospital practice and care of patients 1 objectives &amp; terminolog...
 
Peer observation of teaching ( mahdi) 2016
Peer observation of teaching ( mahdi) 2016Peer observation of teaching ( mahdi) 2016
Peer observation of teaching ( mahdi) 2016
 

Recently uploaded

Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Call Girls in Nagpur High Profile Call Girls
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
Rashmi Entertainment
 

Recently uploaded (20)

Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
Guntur Call Girl Service 📞6297126446📞Just Call Divya📲 Call Girl In Guntur No ...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
Indore Call Girls ❤️🍑7718850664❤️🍑 Call Girl service in Indore ☎️ Indore Call...
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
 
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
💞 Safe And Secure Call Girls Coimbatore🧿 6378878445 🧿 High Class Coimbatore C...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in ChennaiChennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
Chennai Call Girls Service {7857862533 } ❤️VVIP ROCKY Call Girl in Chennai
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICEBhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
Bhopal❤CALL GIRL 9352988975 ❤CALL GIRLS IN Bhopal ESCORT SERVICE
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
Lucknow Call Girls Service { 9984666624 } ❤️VVIP ROCKY Call Girl in Lucknow U...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 

Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation

  • 1. FUNDAMENTAL S OF IMAGING RAD 206 Control of Scatter Radiation
  • 2.
  • 3. What is focal spot? What is effective focal spot ? Which side of the x-ray tube has better resolution ? Which focal spot size produces better resolution? Which OID produces better resolution? Which Object Thickness produces better resolution? Which Object shape produces better resolution?
  • 4. OBJECTIVES Begin discussing factors that influence image detail or visibility of detail • Scatter Radiation, factors affecting it, and there effect on Spatial & Contrast resolution • Ways to reduce scatter & improve image quality • Primary beam restriction & Grids • Technique adjustments when using grids
  • 5. WHAT IS CONTRAST What is scatterWhat is scatter
  • 6. RELATIONSHIP BETWEEN CONTRAST AND SCATTER The relationship is INVERSED
  • 7. IMAGE-FORMING X-RAYS Two kinds of photons are responsible for the OD and contrast on an image: • Photons that pass through without interacting and • Photons that are scattered though Compton.
  • 8. EFFECT OF SCATTER ON THE IMAGE Which interaction produces scatter radiation ?
  • 9.
  • 10. FACTORS CONTRIBUTING TO AN INCREASE IN SCATTER • Increased kVpkVp (Beam Energy, Compton) • Increased partpart thicknessthickness (e.g. Dabbeh) • Increased x-ray field sizefield size (opening Collimation to Larger Fields) Q: what can you do to reduce scatter ?Q: what can you do to reduce scatter ?
  • 11. 1- KVP At 50 kVp 7979% photoelectric, 21%21% Compton <1% transmission At 80 kVp 4646% photoelectric, 52%52% Compton 2% transmission
  • 12. RELATIONSHIP BETWEEN KVP AND CONTRAST How does increasing kVp affect patient dose?
  • 14. Imaging thick parts of the body results in more scatter radiation than thin parts IMAGE TEST TOOL
  • 15. Is patient thickness something the radiographer can control?
  • 16. IS PATIENT THICKNESS SOMETHING THEIS PATIENT THICKNESS SOMETHING THE RADIOGRAPHER CAN CONTROL?RADIOGRAPHER CAN CONTROL? Normally, No But Compression Devices can be used in certain situations which reduces patient thickness and bringing the object closer to the IR, and therefore results in better spatial resolution and contrast resolution and also lesser exposure factors and patient dose
  • 17.
  • 18. COMPRESSION • Reduces OID • Improves spatial resolution • Improves contrast resolution (reducing fog or noise) • Reduces patient dose
  • 20. 3- FIELD SIZE As field size increases, intensity of scatter radiation also increases rapidly. Especially during fluoroscopy
  • 21. Larger radiation field produces more scatter radiation
  • 22. COMPARE IMAGES: WHAT DO YOU THINK ABOUT RADIOGRAPHIC CONTRAST & IMAGE NOISE?
  • 23.
  • 24. INCREASE IN SCATTER • kVpkVp • PartPart ThicknessThickness • Field SizeField Size
  • 25. CONTROL OF SCATTER RADIATION Technologists routinely use two types of devices to reduce the amount of scatter radiation reaching the IR Beam Restrictors Compression Grids Before Patient On patient After patient
  • 26. BEAM-RESTRICTING DEVICES AKA LBD 1. Aperture Diaphragm 1. Cones or Cylinders 2. Variable aperture collimator
  • 27. APERTURE DIAPHRAGM The simplest of all beam-restricting devices Lead or lead-lined metal diaphragm attached to the x-ray tube head The opening in the diaphragm is usually designed to cover just less than the IR used
  • 28. APERTURE DIAPHRAGM Fixed lead opening Fixed image receptor size Constant SID Source-to-diaphragm distance = SDD
  • 29. CONES & CYLINDERS Are modifications of the aperture diaphragm Alignment is one difficulty when using cones Now mostly used with spines teeth & heads
  • 30.
  • 32. IMPROVED CONTRAST RESOLUTION OF THE FRONTAL SINUSES
  • 33. VARIABLE APERTURE COLLIMATOR The most common beam- restricting device is the light-localizing variable aperture collimator The first part of the collimator serves to control off-focus radiation. What is off- focus radiation?
  • 34.
  • 35. OFF - FOCUS RADIATION ALSO CALLED EXTRA-FOCAL RADIATION X-ray tubes are designed so that the projectile e- interacts with the target. However, some of the e- bounce off the target and land on other areas This caused x-rays to be produced out side the focal spot, they are undesirable because it extends the size of the focal spot, increases patient skin dose & reduces image contrast
  • 36. Note: Using a grid does not reduce off-focal radiation
  • 37. FIRST-STAGE ENTRANCE SHUTTERING DEVICE Has multiple collimator blades protruding from the top of the collimator into the tube housing
  • 38. THE SECOND-STAGE COLLIMATOR SHUTTERS Pb leaves are at least 3 mm thick They work in pairs and are independently controlled
  • 39. THE COLLIMATOR LAMP & MIRROR Must be adjusted so that the projected light field coincides with the x-ray beam Misalignment of the light field and beam can result in collimator cutoff of anatomic structures
  • 40. Always keep the collimated area smaller than the size of the cassette What is a PBL BLD
  • 41. GRIDS

Editor's Notes

  1. How many important words can you see Whare are you in this sentence Where am I
  2. As x-ray energy increases Photoelectric and Compton interactions decrease. Explain?
  3. Which part is thicker
  4. BLD
  5. You should always see a 4 sided borders.