Nephlometer
y
Presented by…
Musaddaq Hafeez
(College of allied
health professional
GC university
Faisalabad)
Introduction
 Word derived from Greek word, Nephele
means’ cloud’ + Meter
Definition :
A. In analytical chemistry The measurement of
the concentration of
a solution, suspension or dispersion based
upon its light-scattering properties.
In immunology A technique used to determine
the levels of antibodies or antigens in a
suspension, based on its light-scattering
properties
Introduction
 It is performed by measuring the turbidity in a
water sample by passing light through the
sample being measured.
 In nephelometry the measurement is made by
measuring the light passed through a sample
at an angle
 It is based on the principle that a dilute
suspension of small particles will scatter light
(usually a laser) passed through it rather than
simply absorbing it.
 The amount of scatter is determined by
collecting the light at an angle (usually at 30
and 90 degrees)
Introduction
 .Antibody and the antigen are mixed in
concentrations such that only small
aggregates are formed that do not quickly
settle to the bottom
 . The amount of light scatter is measured
and compared to the amount of scatter
from known mixtures.
 The amount of the unknown is determined
from a standard curve.Nephelometry can
be used to detect either antigen or antibody
Introduction
Purpose :
 Nephelometry is a technique used
in immunology to determine the levels of
several blood plasma proteins.
 For example the total levels of antibodies or classes
 It is important in quantification of free light
chains in diseases such as multiple myeloma.
 Quantification is important for disease classification
and for disease monitoring
 once a patient has been treated (increased skewing
of the ratio between kappa and lambda light
chains after a patient has been treated is an
indication of disease recurrence).
Types
In immunology two types of test run
 End point nephelometry
 tests are run by allowing the
antibody/antigen reaction to run through to
completion (until all of the present reagent
antibodies and the present patient sample
antigens that can aggregate have done so
and no more complexes can form).
 the large particles will fall out of the solution
and cause a false scatter reading, thus
kinetic nephelometry was devised.
Types
 kinetic (rate) nephelometry
 In kinetic nephelometry, the rate of scatter is
measured right after the reagent is added.
 As long as the reagent is constant the rate
of change can be seen as directly related to
the amount of antigen present
Principle
 The principle of nephelometry is based on
the scattering or absorption of light by solid
or colloidal particles suspended in solution.
 When light is passed through the
suspension, part of incident radiant energy
is dissipated by absorption, reflection, and
reaction while remainder is transmitted.
 In nephelometry light is passed through
the sample solution (suspended particles)
directly
 the amount of scattered radiation is
measured generally at 90°C.
principle
 The measurement of intensity of scattered light as
a function of concentration of dispersed phase is
the basis of analysis of nephelometry.
 It is very important to note that in nephelometry
incident and scattered light are of same
wavelength whereas in fluorimeter (in fluorimetry)
scattered light is of longer wavelength than
incident light.
instrumentation
1- light source:
Tungsten its relatively low intensity makes
it less useful for samples with low light
scattering. Alternatives are:
quartz halogen lamp, xenon lamp and
laser which have higher intensities than
tungsten lamp.
2-lens assembly:
Light enter the sample holder through lens
assembly.
Instrumentation
3-there is provision for the insertion of
filter between the sample and source
of light(monochromator).
4- detector (photo –cell):
It is shielded to minimize interference
from stray light.
5-Read out device:
Light intensity is converted to an
electrical signal by the detector .
Factors affecting the scatting
property
1. Particle size & shape
2. Wavelength dependence
3. Distance of observation (Set-up)
4. Polarization of incident light
5. Concentration of particles
6. Molecular weight of particles
Measurement procedure
1-plug the instrument into ground outlet.
2- choose desirable scale from 0-10
starting with the highest conc. (for std
1=scale 10)
3- turn power switch on.
4-selecte desirable range by range
selector at desirable position .
5- select filter required.
Measurement
6- transfer your standards in the cleaned
cell and place them in cell holder.
7- remove the standards.
8- fill the second cell with blank to set zero
.
9- check the reading of the standards
again.
10- measure your unknown.
11- draw calibration curve.
Precautions
-Number and size of the particles
should remain constant if repeated
preparation are made
-clean cell & filter
- avoid air bubbles (high reading).
-dilute sample if there is need.
-prepare the blank ,standards, Sample
at the same time.(to avoid ppt)
Clinical applications
 Widely used to determine concentrations
of unknowns where there is antigen-
antibody reactions such as
 Determination of immunoglobulin's (total,
IgG, IgE, IgM, IgA) in serum and other
biological fluids
 Determination of the concentrations of
individual serum proteins; hemoglobin,
haptoglobin, transferring, c-reactive
protein, !1-antitrypsin, albumin (using
antibodies specific for each protein)
 Determination of the size and number of
particles
Advantage & disadvantage
 Advantage
 Very rapid procedure
 Simplicity in measurement
 High accurate
 Disadvantage
 High cost
 Easily damaged
 Require high power supply
 Used for lower concentration

Nephlometery

  • 1.
    Nephlometer y Presented by… Musaddaq Hafeez (Collegeof allied health professional GC university Faisalabad)
  • 2.
    Introduction  Word derivedfrom Greek word, Nephele means’ cloud’ + Meter Definition : A. In analytical chemistry The measurement of the concentration of a solution, suspension or dispersion based upon its light-scattering properties. In immunology A technique used to determine the levels of antibodies or antigens in a suspension, based on its light-scattering properties
  • 3.
    Introduction  It isperformed by measuring the turbidity in a water sample by passing light through the sample being measured.  In nephelometry the measurement is made by measuring the light passed through a sample at an angle  It is based on the principle that a dilute suspension of small particles will scatter light (usually a laser) passed through it rather than simply absorbing it.  The amount of scatter is determined by collecting the light at an angle (usually at 30 and 90 degrees)
  • 4.
    Introduction  .Antibody andthe antigen are mixed in concentrations such that only small aggregates are formed that do not quickly settle to the bottom  . The amount of light scatter is measured and compared to the amount of scatter from known mixtures.  The amount of the unknown is determined from a standard curve.Nephelometry can be used to detect either antigen or antibody
  • 5.
    Introduction Purpose :  Nephelometryis a technique used in immunology to determine the levels of several blood plasma proteins.  For example the total levels of antibodies or classes  It is important in quantification of free light chains in diseases such as multiple myeloma.  Quantification is important for disease classification and for disease monitoring  once a patient has been treated (increased skewing of the ratio between kappa and lambda light chains after a patient has been treated is an indication of disease recurrence).
  • 6.
    Types In immunology twotypes of test run  End point nephelometry  tests are run by allowing the antibody/antigen reaction to run through to completion (until all of the present reagent antibodies and the present patient sample antigens that can aggregate have done so and no more complexes can form).  the large particles will fall out of the solution and cause a false scatter reading, thus kinetic nephelometry was devised.
  • 7.
    Types  kinetic (rate)nephelometry  In kinetic nephelometry, the rate of scatter is measured right after the reagent is added.  As long as the reagent is constant the rate of change can be seen as directly related to the amount of antigen present
  • 8.
    Principle  The principleof nephelometry is based on the scattering or absorption of light by solid or colloidal particles suspended in solution.  When light is passed through the suspension, part of incident radiant energy is dissipated by absorption, reflection, and reaction while remainder is transmitted.  In nephelometry light is passed through the sample solution (suspended particles) directly  the amount of scattered radiation is measured generally at 90°C.
  • 9.
    principle  The measurementof intensity of scattered light as a function of concentration of dispersed phase is the basis of analysis of nephelometry.  It is very important to note that in nephelometry incident and scattered light are of same wavelength whereas in fluorimeter (in fluorimetry) scattered light is of longer wavelength than incident light.
  • 11.
    instrumentation 1- light source: Tungstenits relatively low intensity makes it less useful for samples with low light scattering. Alternatives are: quartz halogen lamp, xenon lamp and laser which have higher intensities than tungsten lamp. 2-lens assembly: Light enter the sample holder through lens assembly.
  • 12.
    Instrumentation 3-there is provisionfor the insertion of filter between the sample and source of light(monochromator). 4- detector (photo –cell): It is shielded to minimize interference from stray light. 5-Read out device: Light intensity is converted to an electrical signal by the detector .
  • 13.
    Factors affecting thescatting property 1. Particle size & shape 2. Wavelength dependence 3. Distance of observation (Set-up) 4. Polarization of incident light 5. Concentration of particles 6. Molecular weight of particles
  • 14.
    Measurement procedure 1-plug theinstrument into ground outlet. 2- choose desirable scale from 0-10 starting with the highest conc. (for std 1=scale 10) 3- turn power switch on. 4-selecte desirable range by range selector at desirable position . 5- select filter required.
  • 15.
    Measurement 6- transfer yourstandards in the cleaned cell and place them in cell holder. 7- remove the standards. 8- fill the second cell with blank to set zero . 9- check the reading of the standards again. 10- measure your unknown. 11- draw calibration curve.
  • 16.
    Precautions -Number and sizeof the particles should remain constant if repeated preparation are made -clean cell & filter - avoid air bubbles (high reading). -dilute sample if there is need. -prepare the blank ,standards, Sample at the same time.(to avoid ppt)
  • 17.
    Clinical applications  Widelyused to determine concentrations of unknowns where there is antigen- antibody reactions such as  Determination of immunoglobulin's (total, IgG, IgE, IgM, IgA) in serum and other biological fluids  Determination of the concentrations of individual serum proteins; hemoglobin, haptoglobin, transferring, c-reactive protein, !1-antitrypsin, albumin (using antibodies specific for each protein)  Determination of the size and number of particles
  • 18.
    Advantage & disadvantage Advantage  Very rapid procedure  Simplicity in measurement  High accurate  Disadvantage  High cost  Easily damaged  Require high power supply  Used for lower concentration