Introduction to Chromatography
(HPLC)
2
Chromatography began at the turn of 20th century when
Ramsey {Ramsey., Proc.Roy. soc.A76,111 (1905)}
separated mixtures of gases and vapours on adsorbents
like charcoal and Michael Tswett [Tswett, M.,
Ber.deut.botan. Ges., 24, 316 & 386(1906)] separated
plant pigments by liquid chromatography.
William Ramsey Michael Tswett
3
 Tswett is credited as being the "Father of
Chromatography" principally because he coined the term
chromatography (literally colour writing) and
scientifically described the process.
4
MICHAEL TSWETT
 He was a Russian botanist.
 His main interest was to find out the composition of the green
plant pigment, chlorophyll.
 There were two schools of thought on this subject.
 One school thought that chlorophyll was a single solute.
 The other school thought that chlorophyll was a mixture
of different solutes.
5
 A glass column 1-5 cm in diameter and 20-100 cm long was taken.
 The column was packed with the slurry of Calcium Carbonate
made in petroleum ether(60-80).
MICHAEL TSWETT’S EXPERIMENT
6
 The solution of chlorophyll was allowed to move down
slowly and then the column was washed with ethyl alcohol for
few hours.
 A concentrated solution of chlorophyll was prepared in
petroleum ether(60-80) and a small volume (~10cm3
) was
put on the top of the calcium carbonate column in the
glass column as a thin band.
MICHAEL TSWETT’S EXPERIMENT
7
COTTON PLUG
SLURRY OF CaCO3
POWDER IN PETROLEUM
ETHER
COTTON PLUG
CONCENTRATED SOLUTION OF
CHLOROPHYLL IN PETROLEUM
ETHER
WASHING/ ELUTION
WITH ETHANOL
8
β CHLOROPHYLL
α CHLOROPHYLL
XANTHOPHYLL
VIOLA XANTHINE
CAROTENE
COTTON PLUG
SLURRY OF CaCO3
POWDER IN
PETROLEUM ETHER
COTTON PLUG
CONCENTRATED SOLUTION OF
CHLOROPHYLL IN PETROLEUM
ETHER
WASHING/ ELUTION
WITH ETHANOL
9
• Each band was removed from the column and identified.
• It was concluded that chlorophyll is a mixture
of five different solutes. α –chlorophyll, β - chlorophyll,
xanthophyll, viola xanthene, and carotene .
10
α CHLOROPHYLL and β CHLOROPHYLL
11
XANTHOPHYLL
12
VIOLA XANTHENE
13
CAROTENE
14
• Michael Tswett’s main problem was resolved and
in the process he discovered a simple, time saving
and relatively less laborious method of separating
complex solutes from a mixture.
• For example α – chlorophyll (MW 893) differs
from β – chlorophyll by only one methyl group
yet these two solutes are easily separated in
the simplest apparatus.
15
• Today the term chromatography is used for
separations involving colourless solutes also. This
is done to honour the discoverer of the
technique.
• Michael Tswett coined the term “Chromatography for this
new technique of separation.
(Chromatus = colour, graphein = to write)
• He visualized the coloured bands from
chlorophyll as coloured writing.
16
 As a chemist, a researcher or simply as an
inquisitive person, two questions come to our
mind on the results of Tswett’s experiment.
1. Why do the five solutes separate on the column?
2. How do the five solutes separate on the column?
17
Migration velocities increase from
β- chlorophyll to carotene.
“The five solutes are separated on the column
because of the differences in their migration
velocities”
18
“The five solutes have different migration velocities
across the column of CaCO3, because they have
different degrees of adsorption on the surface of
CaCO3”
19
 Carotene has the least degree of adsorption on
CaCO3 surface and hence the highest
migration velocity while β- chlorophyll has the
highest degree of adsorption on CaCO3
surface and hence the lowest migration velocity of
the five solutes.
20
 This is the basis of “Adsorption column
chromatography”.
21
Adsorbents
 The most important agent in chromatography is the adsorbent.
 Therefore, utmost care is to be bestowed in the selection of
proper adsorption material and also the process of building
up a column with it and the maintenance of the column.
22
 It must be insoluble in the solution under analysis and the
solvent that is used for elution
Essential requirements to be satisfied by every adsorbent
 It should not chemically react with the mixture under analysis.
 It should generally be colourless, but this is not a rigid
rule as there are instances where coloured adsorbents
have been used successfully.
Particle-size of the adsorbent material is of primary importance
for successful separation.
23
Too loose , large paricles.
Too tight , small paricles.
24
 Slaked lime (120 mesh) and magnesia are two cheap
adsorbents.
 Fuller's earths are also good adsorbents but unlike the
previous ones they are acidic in reaction. On moistening
they darken considerably.
 Calcium carbonate is a cheap adsorbent of uniform size.
• Sodium sulphate, talc, magnesium sulphate, calcium
phosphate, calcium hydroxide; barium carbonate,
magnesium carbonate, magnesium trisilicate, calcium
sulphate, soda ash, titania, charcoal, sucrose, starch are some
of the other materials.
25
• Adsorbents may be classified as follows according to the
force with which they hold the ions.
Type Examples
Weak Sucrose and starch
Intermediate Calcium carbonate, calcium
phosphate, calcium hydroxide and
magnesium hydroxide
Strong Alumina, charcoal and Fuller's earth.
26
 From Tswett’s experiment if we change slurry
of CaCO3 and replace it with slurry of Al2O3,
will the five solutes form chlorophyll separate
and will the sequence of separation remain the
same?
 In Tswett’s experiment if we replace chlorophyll
extract with a mixture of A+B+C+D+E solutes,
will the solutes separate and what will be the
sequence of their separation?
27
 In Tswett’s experiment if we replace chlorophyll
extract with a mixture of A+B+C+D+E solutes,
will the solutes separate and what will be the
sequence of their separation?
 If we have understood the concept of
adsorption, then it should not be difficult to
answer the above two questions.
28
 In the first instance, we have changed the adsorbent, other
things remaining unchanged.
 The five adsorbates have a new surface of Al2O3.
 If they exhibit differences in the degree of their
adsorption on the surface of Al2O3, they will get
separated.
29
 In the second instance, we have changed the five
solutes, other things remaining unchanged.
 Although the surface of CaCO3 is the same as
that in Tswett’s experiment, the five solutes are
different.
 They will get separated, if they exhibit differences in the degree of
their adsorption on the surface of CaCO3.
30
 Let us consider a situation where in the glass
column we put a liquid which remains stationary
even if the tap is opened.
 If we allow three solutes A, B and C to pass
through the chromatographic column and realize
their separation after some time, we can ask
again the same two questions which we did in
adsorption chromatography.
 Another liquid which is immiscible with the stationary
liquid moves through it.
31
 Why do the three solutes separate?
 How do the three solutes separate?
 The three solutes separate because they have
different migration velocities.
 Solute ‘A’ has the highest migration velocity of the
three and reaches the end of the column first.
 Solute ‘C’ has the lowest migration velocity of the
three and reaches the end of the column last.
32
 The migration velocity increases from solute ‘C’
to solute ‘A’.
 Solute ‘A’ travels the fastest because it is more
soluble in the liquid which moves across the
stationary liquid.
 Solute ‘C’ on the other hand is more
soluble in the stationary phase and
hence lags behind.
33
 The principle of separation here is differences in solubility
or partitioning of solutes between the stationary and mobile
liquid phases.
 It is called as “Partition Column Chromatography”.
34
 If they do not show differences with
reference to any property they cannot be
separated.
 For solutes to get separated from a
mixture, they must exhibit some
differences with reference to some property
like solubility, boiling point, adsorption,
partitioning etc.,
35
Stationary phase: Phase which remains
stationary. It could be a solid or a liquid bonded to an inert
solid support.
Mobile phase: A phase which moves across the
stationary phase. It could be a gas or a liquid.
Elution: The process of washing a chromatographic
column with a liquid or a gas.
Solutes: This term is used in place of compounds or
substances.
36
Normal phase mode of separation: Here the stationary phase
is relatively more polar than the mobile phase.
Reverse phase mode of separation: Here the stationary phase
is relatively less polar than the mobile phase.
37
MOBILE
PHASE
STATIONARY
PHASE
PRINCIPLE OF
SEPARATION
CONTAINER NAME OF THE
CHROMATOGRAPHIC
TECHNIQUE
LIQUID SOLID DIFFERENCES
IN ADSORPTION
COLUMN LIQUID-SOLID ADSORPTION
COLUMN
CHROMATOGRAPHY
LIQUID SOLID DIFFERENCES
IN ADSORPTION
THIN LAYER LIQUID-SOLID ADSORPTION
THIN LAYER
CHROMATOGRAPHY
LIQUID LIQUID DIFFERENCES IN
PARTITIONING
COLUMN LIQUID-LIQUID PARTITION
COLUMN
CHROMATOGRAPHY
LIQUID LIQUID DIFFERENCES IN
PARTITIONING
THIN LAYER LIQUID-LIQUID PARTITION
THIN LAYER
CHROMATOGRAPHY
GAS LIQUID DIFFERENCES IN
PARTITIONING
COLUMN GAS-LIQUID-PARTITION
COLUMN CHROMATIGRAPHY
GAS SOLID DIFFERENCES
IN ADSORPTION
COLUMN GAS-SOLID ADSORPTION
COLUMN
CHROMATOGRAPHY
38
GAS LIQUID/ SOLUTION
CLASSIFICATION OF DIFFERENT CHROMATOGRAPHIC TECHNIQUES
Stationary phase is solid Stationary phase is a liquid coated on the
surface of an inert solid support
gas - solid - adsorption
chromatography
gas - liquid - partition
chromatography
Mobile phase is Gas and all the
solutes being separated are in
gaseous phase under
experimental conditions
Mobile phase is liquid and all the
solutes being separated are in
liquid/solution state
39
Liquid / Solution Chromatography
ADSORPTION : Separation of solutes on the basis of
their degree of adsorption on the
surface of an adsorbent.
PARTITION : Separation of solutes on the basis of
their differences in partition coefficient
of solutes in two mutually immiscible
solvents.ION EXCHANGE : Separation of similarly charged ions
based on the differences in the degree
of their affinity towards the ion exchange
resin.
SIZE EXCLUSION : Separation of solutes based on the
differences in their molecular weights or
sizes.
ELECTROPHORETIC : Separation based on the movement of
charged solutes to opposite poles under
the applied electrical field.
AFFINITY : Separation of solutes based on specific
affinity of solutes on the specific sites of
the stationary phase.
40
• Let us consider a chromatographic column on which
three solutes A, B and C are separated in that order.
• Let there be sensor or a detector fixed at the other
end of the column.
• The detector can sense and count the number of
molecules of solute entering into it.
• The detector does not sense and count the
number of molecules of mobile phase entering
into it.
41
42
• As we can see from the figure, first the mobile phase
will enter into the detector and no response will be
generated.
DETECTOR
A
B
C
COTTON PLUG
COTTON PLUG
43
 Molecules of solute ‘B’ enter the detector next.
They are sensed and counted.
 Mobile phase enters into the detector again and is not sensed and
counted. Molecules of solute ‘C’ enter the detector next.
They are sensed and counted.
 Mobile phase enters into the detector again and is not
sensed and counted.
 Molecules of solute ‘A’ enter the detector next.
They are sensed and counted.
 Mobile phase enters into the detector again and
is not sensed and counted.
44
45
• If we plot a graph of graph of detector response as a
function of time, the above description can be converted
into the following picture.
•
• This is called a chromatogram.
A
B
C
DETECTOR
RESPONSE
TIME
46
A
B
C
DETECTOR
RESPONSE
TIME
tRA
tRB
tRC
DETECTOR
RESPONSE
TIME
X
Y
DETECTOR
RESPONSE
TIME
AUCA
AUCB
AUCC
TIME
DETECTOR
RESPONSE
t wA t wB t wC
47
• A normal chromatogram has certain characteristic features.
All the peaks are bell shaped symmetrical peaks.
 Neighbouring peaks are separated by some distance.
 Each peak has a characteristic retention time
(distance between center of the peak and the
reference point) under given experimental conditions.
This is useful in qualitative Chromatographic analysis.
48
 Each peak has certain area under curve (AUC) and
AUC α Concentration of Solute. This is useful in
quantitative chromatographic analysis. A separate
calibration graph is needed for each peak in a
chromatogram.
 Each peak has certain width at base which
increases with increase in retention time.
49
 Both qualitative and quantitative chromatographic
analysis are relative methods of analysis in which
the standard solutes and the sample solutes are
chromatographed under the same experimental
conditions.
 Volumetric and Gravimetric methods, on the other
hand, are the examples of absolute methods of
quantitative analysis.
50
• It is observed in practical Chromatographic
separations that the spots or bands or peaks are
always broad. They are never very sharp.
 What could be the reason for this?
• If we observe our earlier Chromatogram
carefully, we see that for each solute, solute
molecules start eluting at a particular time and the
elution is complete after a few seconds.
51
• It means that all the molecules of the same solute do
not travel across the stationary phase with the same
speed. Some travel faster than others. Therefore
some reach the detector first and others latter.
• In other words molecules of the same solute spread across
the stationary phase which results in band broadening.
52
• We can visualize four important causes leading to molecular
spreading and band broadening.
• We must find out at least some important causes for molecular
spreading and band broadening.
53
• We should know in the beginning that the particles of
stationary phases are approximately spherical in nature and
the particles are not exactly of the same size.
• When we say that the particle size is 100µ, it really
means that it is the average particle size and that every
particle is not of exact 100µ diameter.
• Secondly, the spherical particles are always
porous in nature. That is, the surface is not
smooth.
54
Let us now imagine a magnified picture of a chromatographic column
• We can see that when the solute molecules travel across the stationary
phase, they have different paths to flow down the column. Each path has
different path length.
COTTON PLUG
DETECTOR
SAMPLE
SOLUTION
WASHING WITH
LIQUID MOBILE
PHASE
55
• The molecular spreading results in band broadening.
• This cause of molecular spreading leading to band broadening
is known as, “Eddy Diffusion” or “Multiple path flows Effect”.
• Solute molecules travelling in shorter paths reach the
detector early.
• Solute molecules travelling in longer paths reach the
detector late.
• As a result of multiple path flows, molecules of the same
solute spread across the stationary phase .
56
1. Eddy diffusion. (difference in the mobile phase stream flow due to non
homogeneity of stationary phase) in packed columns.
57
1. No or minimum Eddy diffusion. (difference in the mobile phase
stream flow due to non homogeneity of stationary phase) in packed
columns.
58
 Can Eddy diffusion be avoided?
• Even if we bring the stationary phase particles closer
together than what they are in the above picture, there
will always be some distance between the
neighbouring particles.
• What is achieved by compact packing of stationary
phase particles in the column probably is that the
numbers of multiple path flows are minimized and
the extent of molecular spreading leading to band
broadening is reduced.
• Eddy diffusion, therefore, cannot be avoided, it can be optimized.
59
• Let us now have a magnified picture of two neighbouring
stationary phase particles.
60
• In the above picture we have three streams in which solute
molecules are moving between two stationary phase
particles.
• Two streams flow closer the surface of the stationary
phase particles.
DETECTOR
61
• Solute molecules in these streams experience resistance
to flow from the surface of the stationary phase particles.
• Their migration velocity will be relatively smaller than those
solute molecules traveling in the middle.
• This obviously results in molecular spreading adding
to band broadening.
• This cause is known as mobile phase mass transfer
and cannot be eliminated.
62
• In closely packed column the contribution of
this cause called “Mobile phase mass transfer” will
be more than in a loosely packed column.
• In a loosely packed column the contribution of
Eddy diffusion will be more.
• In practice we have to strike a balance.
63
Let us now consider two pictures of two magnified porous particles.
• In both the above pictures A and B, while solute molecules travel
across the stationary phase , some of them enter inside the
pores and remain stagnant for sometime before they are flushed
out by the mobile phase which is continuously flowing across the
chromatographic column.
•There is a small difference however.
DETECTOR
A
DETECTOR
B
64
• In picture A the pore is shallow and the solute molecules remain
stagnant along with the mobile phase for short time. This cause is
known as “Stagnant Mobile Phase Mass Transfer”.
• In picture B, solute molecules enter into a deep pore. Inside the deep pore
they have enough time to interact with the liquid stationary phase coated on
the surface of the stationary phase particles.They remain in the deep pore for a
longer period of time as that compared to solute molecules in a shallow pore.
• In either case, the solute molecules which enter into the pores
have relatively smaller migration velocity as when compared to
those which do not enter into the pores.
• This cause is known as “Stationary Phase Mass Transfer”.
• This results in molecular spreading and band broadening.
65
• Eddy diffusion, mobile phase mass transfer, stagnant
mobile phase mass transfer, and stationary phase
mass transfer are the four main causes for molecular
spreading and band broadening.
66
• The causes for molecular spreading leading to
band broadening cannot be eliminated.
• They can be optimized by choosing stationary phase
particles with uniform size having shallow pores and
packing compactly in a chromatographic column.
67
• Let us now consider in quantitative terms what we
have discussed qualitatively about molecular
spreading and band broadening in chromatographic
separations.
68
• In order to do that let us make the following assumptions:
tR - retention time of a solute
L cm - Length of chromatographic column
tO - retention time of the mobile phase
u - migration velocity of the mobile phase
uX - migration velocity of the solute
ns - number of moles of solute in the stationary phase
nm - number of moles of solute in the mobile phase
R - fraction of moles of solute in the mobile phase
69
• Let us imagine that a chromatographic column of
Lcm length is divided into N small sections of equal
length or height.
L cm
• Each small section is called a “Theoretical Plate”
because it is a theoretical concept.
• No such sections or plates exist in reality.
70
• The solute molecules constantly get exchanged
between the stationary phase and the mobile phase
when they travel across the column.
• The solute molecules travel in the forward direction along
with the mobile phase only when they are in the mobile
phase . They remain stagnant when they are in the
stationary phase.
• The length or height of each theoretical plate is such that
when solute molecules travel that distance a new
equilibrium is established.
• There will be, therefore ‘N’ number of such equilibria
across the length of the column.
71
• It is then easy to imagine that more the number of
theoretical plates associated with the given length of
chromatographic column, more will be such equilibria
and solute molecules will have less time and chances
to spread across the stationary phase.
• It means that more the N (number of theoretical
plates), lesser is the molecular spreading and
sharper will be the resulting bands or peaks.
72
• In other words we can say that higher the N associated
with a given length of chromatographic column higher is
“ column separation efficiency.”
• We also know that L / N = H, height equivalent to a
theoretiacal plate. N and H are inversely proportional.
• It can, therefore, be concluded that higher the N or
lower the H associated with a given length of
chromatographic column, higher is the column separation
efficiency.
73
• The following three factors govern the column separation
efficiency.
1. Particle size of stationary phase.
2. Length of chromatographic column.
3. Mobile phase velocity.
74
Particle size of stationary phase
• We normally assume stationary phase particles to be
spherical in shape.
• If we reduce the particle size diameter from 100µ
to 50µ, 25µ, 10µ, 05µ, the surface area will increase as
follows.
75
Particle size in diameter
(microns)
Surface area
(m2
/g)
100 0.01538
50 0.0307
25 0.0615
10 0.153
5 0.307
76
0
20
40
60
80
100
120
0 0.1 0.2 0.3 0.4
Series1
Particlesizeindiameter(microns)
Surfacearea(m2/g)
77
• We can see that as the particle size of spherical particles is
reduced, the surface area increases exponentially.
• When the surface area increases, the amount of the liquid
phase coated also increases proportionately. More the liquid
phase loading more will be the interactions of solute
molecules with the liquid stationery phase, less will be
molecular spreading and band broadening.
• Then we can say that lower the particle size higher will be N
and higher will be the column separation efficiency.
78
• However we have also to remember that as the
stationary phase particles become smaller there is more
resistant to mobile phase flow.
•We have therefore to optimize the particle size without
sacrificing analytical times. Because to achieve good
separation with reasonable analytical time is practically
important and cannot be overlooked.
79
• If with a given length of say 100cm long column,
N = 10,000 then with L = 200cm, N= 20,000, keeping
particle size of stationary phase the same.
• Thus, longer the chromatographic column, higher will be N
and column separation efficiency.
• With longer columns, however, retention time of solutes
will increase and peaks will become broader and analytical time
will increase.
80
• We have to, therefore, optimize the column length just
as we optimized the particle size. Our main aim being
good separation with reasonable analytical time.
81
For example for 10µ diameter porous stationary phase particles
we get a graph as shown below.
• It can be seen that initially as the
mobile phase velocity increases, H
decreases i.e. separation efficiency
increases.
• It reaches at the maximum with
mobile phase velocity ‘x’.
Mobile phase velocity
L/N
X
82
• We may, therefore, be tempted to work at the mobile phase
velocity ‘x’ because at this point we get the maximum
separation efficiency.
• From the above discussion it becomes clear that for getting
good separation efficiency with reasonable separation time
we have to optimize particle size, column length and
mobile phase velocity.
• However at this mobile phase velocity if the separation
times are long, then we will have to work at higher mobile
phase velocity at the cost of lower separation efficiency to
achieve reasonable analytical time.
83
A TYPICAL CHROMATOGRAM
 It is a plot of the detector response in millivolts as a function of
time (usually time is the abscissa and millivolts the ordinate)
Retention Time
S
o
l
v
e
n
t
A
B
C
Inject
Detector
Response
100
0
Time in minutes0
84
• The above Fig. illustrates how resolution can be enhanced by
exploiting the selectivity effect of the mobile phase (i.e, by
switching from 60% ACN to 60% MeOH). It is important to
note that small change of selectivity can have a major effect on
resolution as resolution is proportional to (α -1). Columns of
different bonded phases (i.e,C8, phenyl, CN ,polar- embedded
etc) can also provide different selectivity effects.
85
• Finally, the plate number of the column (N) should be maximized by using a
longer column or a more efficient column. Increasing N is not an efficient way to
achieve resolution since Rs is proportional to √N. Doubling N by doubling the
column length increases analysis time by 2 but only increases resolution by √2 or
by 41%. In contrast, increasing α from 1.05 to 1.1 will almost double the
resolution. For complex samples with many peaks, increasing N is a viable and
the most direct approach.
86
Concept of tailing and fronting
Individual solute molecules act independently of one
another during the chromatographic process.
As, a result they produce a randomized aggregation of retention
times after repeated sorptions and desorptions.
The result of a given solute is a distribution, or peak, whose shape
can be approximated as being normal or Gaussian.
 Nonsymmetrical peaks usually indicate that some undesirable
interaction has taken place during the chromatographic process.
87
•Ideal Guassian Broad Fronting Tailing Doublet
Rate
theory
Faulty
injection.
Degraded
column.
88
The figure below shows some shapes that can occur in actual samples.
89
 Broad peaks like (b) are more common in packed columns and
usually indicate that the kinetics of mass transfer is too slow.
 Asymmetric peaks can be classified as fronting (c) or tailing (d)
depending on the location of asymmetry.
 The extent of asymmetry is defined as the tailing factor (TF).
90
TF = b/a
 Both a and b are measured at 10% of the peak height shown.
 As can be seen from the equation, a tailing peak will have a TF
greater than one.
 The opposite symmetry, fronting, will yield a TF less than one.
91
 The doublet peak like (e) can represent a pair
of solutes that are not adequately separated.
• Repeatability of a doublet peak should be
verified because such peak shape can also
result from faulty injection technique, too
much sample, or degraded columns.
Flow path for method
developmentDefine Goal
Detection
Chromatography
selection (Stationary
phase & Mobile phase)
Sample preparation
Measurement
Matrix Substitution
System suitability
Stability
Validation
Reporting
DETECTION
• Prepare a solution of 10 µg/mL in MeOH:water mixture in
acidic, basic or neutral conditions
•Assess against their corresponding blanks for UV absorbance and
Fluorescence
•Tune the ionisation parameter on LC-MS to assess fragmentation
and potential suitability for LC-MS-MS
Results may indicate likely detection mechanism and detector
The most common HPLC detectors are
– UV/VIS
– Fluorescence
– Refractive index
– Electro chemical
– LCMS
DETECTION
Detection method chosen depending on :
Requirements of assay
UV detection – easiest and most widely available
But, has moderate selectivity and sensitivity
Fluorescence detection- greater sensitivity( 10-1000 times
higher than that of UV) and selective,
But greater care is needed for IS selection
E.C.D detection-comparable sensitivity to fluorescence
and good selectivity (due to response to specific
functional group undergoing oxidation or reduction,
But has serious problem with stability at high
sensitivity work
LCMS highest sensitivity and specificity,
But involves large cost
Separation
Most difficult and time consuming part
-SAMPLE PREPARATION
-Selectively enrich biological sample for quantification at
appropriate level are possible
-CHROMATOGRAPHIC SEPARATION
Resolve and quantify property of analyte/ potential
metabolite and IS in presence of endogenous material
in short time as possible
Chromatographic separation
Fundamental Resolution Equation
Rs = 1 √n k α-1
4 1+ k α
How can resolution be changed ?
Efficiency : √n - Column length, particle size, Column Quality,
Temperature
Retention : k - Eluent Strength, Column stationary phase,
1+ k Temperature
Selectivity : α-1 - Stationary Phase (column), Mobile phase,
α Mobile phase additives, Temperature
In bioanalysis capacity factor more than 4 to 5
- Matrix effect
Sample preparation
• OBJECTIVE
• To carry through as high percentage as possible of
the analyte(s) from original sample to the
determination steps (good recovery)
• To ensure that the analyte remain in their original
state
• To carry through as few of the matrix components
from the original sample to the determination step
• Preparation of sample in solution form, compatible
for introduction in specific instrument (HPLC, LCMS,
GCMS, RIA etc)
Sample preparation
In general there are three different methods of sample
preparation prior to assay: -
– Protein Precipitation
– Liquid - Liquid Extraction
– Solid Phase Extraction
Protein Precipitation
• Denaturation and precipitation of proteins is done by
an addition of an acid, solution containing a heavy
metal ion or water miscible organic solvents
(Methanol, ACN)
• Following admixture of precipitant the sample is
centrifuged to produce a clear supernatant
• Precipitation by addition of organic solvent
• Water miscible organic solvents ACN(x1 vol),
MeOH(x 2 vol), ethanol or acetone
• Use solvent in which analyte is highly soluble
• Mixture of two solvents (MeOH and DMSO)
Protein Precipitation
• Precipitation by addition of inorganic compounds
• Acidic or anionic precipitants trichloroacetic acid
and/or perchloric acid (10-20 % w/v)
• Mobile phase should contain a high molar
concentration of buffer to protect column
• Should not be used for compound which are prone to
acid hydrolysis
• Cationic precipitants such as copper and zinc in
alkaline solution e.g Zinc sulphate in NaOH or barium
hydroxide or copper sulphate in potassium hydroxide
Protein Precipitation
• Disadvantage
Cannot be used for quantifying very low concentrations
due to dilution of sample
Late eluting peak may co-elute with drug peak
distorting peak measurements
Further precipitation may occur if supernatant is
allowed to stand which could clog the
chromatographic column(can be avoided)
Liquid -Liquid Extraction
• Based on extraction of analyte depending on its
partition between an aqueous and an immiscible
organic phase
• The degree of extraction depends on partition
coefficient (KD) which is dependent on
• Organic solvent
• pH
• Ionic strength
• Optimised for maximum recovery and minimum co-
extract
Liquid -Liquid Extraction
• Analyte possessing sufficient lipophilic character gets
extracted in organic layer
• The organic layer is collected, dried and reconstituted
in solvent, mobile phase or a solution of a particular
pH.
• Properties of ideal solvent
• Environmentally acceptable, non toxic and not highly
inflammable, immiscible with water, convenient
specific gravity,suitable volatility, high chemical
stability, not prone to form emulsion
• Desirable compromise is required
Liquid -Liquid Extraction
• Most popular solvents are heptane, hexane, toluene,
diethyl ether , ethyl acetate, dichloromethane and
chloroform
• It is best to use least polar solvent that will effectively
extract drug.
• No advantage in selecting solvent with too much
extracting power it may reduce selectivity of assay
• Acid drugs extracted under acidic condition (pH 2 units
below pKa)
• Basic drugs extracted under basic conditions (pH 2
units above pKa)
Liquid -Liquid Extraction
• Wherever possible high pH is desirable to ensure
cleaner extract
• To limit the carry-over
• Back extraction into aqueous phase with pH adjusted so
that analytes ionize again
• Highly polar ionic drug are extracted by formation of
neutral ion pair complex
• Addition of ion pairing agents tetralkyammonium salts
for acids and alkyl sulphonates (heptane sulphonic acid
for base)
Liquid -Liquid Extraction
Disadvantages
- Emulsion Formation
- Time consuming
- adsorption of analytes on glassware
- Require large volume of organic solvent
- Ionizable compounds require specific pH
adjustment for extraction
Solid Phase ExtractionSolid Phase Extraction
SPE utilizes the principle of selective retention of an
analyte on a chemically modified silica surface/sorbent.
There are four steps involved in SPE
– Conditioning
– Application
– Rinsing
– Elution
Solid Phase ExtractionSolid Phase Extraction
Advantages
– Faster Sample preparation – Average time cut by
2/3
– Less solvent and reagent consumption
– Less hazardous waste
– Greater Recoveries
– Greater Accuracy
– Greater Specificity
– Less sample volume handling
– Useful for both hydro and lipophilic drug
– Easy automation
Disadvantages
– Expensive
Concentration of sample
EVAPORATION OF SAMPLE
Simple but its practical application problematic
-time and attention required
-Loss of sample
-In three ways : degradation, volatilization and
incomplete transfer
-Overcome by use Low Volume Evaporator
-Sample is swept with a inert gas nitrogen.
Standardisation
• External standardization
•Internal standardization
Detector response is recorded as a ratio of the signal of
the analyte to that of internal standard.
Calibration Check
Literature search for methods
•Internet
•Pubmed
•Analytical Abstracts (Royal Society of chemist)
www.rsc.org
•Analytical Profiles
Rules of one and two
• Two basic rules applied while modifying the
method
• Rule of one states
• “In altering any chromatographic parameter and
this include the extraction step only one
parameter is changed at any one time
• Rule of two states
• “A chromatographic observation is not valid until
it has been repeated”
Method ValidationMethod validation is a procedure used to ascertain and
prove that a test method consistently yield what it is
expected to do with adequate accuracy and precision.
The method validation parameters are
Specificity
Ability of an assay to distinguish a drug from its
metabolite, other drugs and endogenous constituents of
biological matrix.
Criteria
Response of interfering peaks (if any) at the retention
time of an analyte should be ≤ 20% of the LLOQ (Low
limit of quantitation) response and should be ≤ 5% of
the internal standard response.
Blank Plasma on LCMS Blank Plasma on HPLC
Sensitivity
• The lowest level of concentration of an analyte that
can be measured in biological matrix
Criteria
• The response of LLOQ should be at least 3 times
greater than any interference peak.
• The between batch accuracy must be ≤ 20%
• The between batch precision must be ≤ 20%
Low Limit of Quantitation
on LCMS
Low Limit of Quantitation
on HPLC
Linearity
A minimum of five non-zero calibration curve
standards including LLOQ and highest concentration
should yield:
• r2
≥ 0.98
∀ ± 20% deviation of LLOQ from nominal
concentration
∀ ± 15% deviation of standards other than LLOQ
Precision & Accuracy (-PA)
• A PA batch is five non-zero calibration curve
standards including LLOQ and highest concentration
with five sets of: -
• LLOQ Quality Control Sample
• LQW Quality Control Sample
• Middle Quality Control Sample
• High Quality Control Sample
Acceptance criteria
• A minimum of five sets of PA batch should run on
different days
• Precision and accuracy must be ≤15% for:
• within PA batch
• Intra day PA batch
• Inter day / Between PA batches
Stability
• The following stability exercise should be conducted
during method validation:
• Standard stock solution
• Bench top
• Autosampler
• Freeze them (I, II& III cycle)
• Long term stability (Storage condition)
Recovery
• This is carried out to ascertain maximum and
consistent extraction of drug from biological fluid.
Ruggedness
• It is carried out by analysing one PA batch processed
using different sets of reagents, different columns,
different analyst and run on different instrument.
Method life cycle
Validation
Development Optimization
As a part of GLP all the exercise carried out
during method validation are well documented
by competing specific pre-designed form, log
books and note books.
The complete method validation should be
audited by an independent quality assurance
person for
• Conduction of activities as per related SOPs
• Pre recording and documentation of data
• Data Analysis
• Accurate reporting
GLP Requirements
Reference
s•Hill, H.M. Analysis of drugs in biological fluid, in :
Townshed, A (ed.) Encyclopedia of Analytical
sciences, Volume 6, London, Academic Press, 1995
•Smyth, M.R.(ed.0 Chemical Analysis in Complex
Matrices, New York, Ellis Harwood Limited, 1992
•Snyder, L.R. Kirkland, J.J. and Glajch, Practical
HPLC Method Development, 2nd
edition, New York,
John Wiley and Sons, Inc., 1997
•Venn, R.F.(ed.) Principle and Practice of bioanalysis,
London, Taylor and Francis
123
References
1.Heftmann, Chromatography Part A: Fundamentals and
Techniques, 6th edition, published by Elsevier, 2004.
2.Skoog, Holler, Nieman, Principles of Instrumental Analysis,
5th edition, published by Thomson Brook/Cole, 2005.
3.Mendham, Denny, Barnes, Thomas, Vogel’s text book of
Quantitative Chemical Analysis, 6th edition, published by
Person Education, 2007.

More Related Content

What's hot

(PART-II)- Development of latent fingerprints using powders method.pptx
(PART-II)- Development of latent fingerprints using powders method.pptx(PART-II)- Development of latent fingerprints using powders method.pptx
(PART-II)- Development of latent fingerprints using powders method.pptx
Applied Forensic Research Sciences
 
Chromatography : A seperation technique
Chromatography : A seperation techniqueChromatography : A seperation technique
Chromatography : A seperation technique
SHIVANEE VYAS
 
Brain electrical oscillation signature profiling(beos)
Brain electrical oscillation signature profiling(beos)Brain electrical oscillation signature profiling(beos)
Brain electrical oscillation signature profiling(beos)
Nitesh Bhatia
 
Dna in criminal justice_complete slides
Dna in criminal justice_complete slidesDna in criminal justice_complete slides
Dna in criminal justice_complete slidesKari Ann Bitgue
 
Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)
karimbscdu
 
ear print.pptx
ear print.pptxear print.pptx
ear print.pptx
MATANGI LAD
 
Thin layer chromatography. (TLC)
Thin layer chromatography. (TLC)Thin layer chromatography. (TLC)
Thin layer chromatography. (TLC)
Manish Rajput
 
conventional methods of fingerprint development
 conventional methods of fingerprint development conventional methods of fingerprint development
conventional methods of fingerprint development
faraharooj
 
Forensic analysis of soil
Forensic analysis of soilForensic analysis of soil
Forensic analysis of soil
Ketan Patil
 
Asphyxia.pptx
Asphyxia.pptxAsphyxia.pptx
Asphyxia.pptx
andy664508
 
Sample preparation for Chromatography
Sample preparation for Chromatography Sample preparation for Chromatography
Sample preparation for Chromatography
latha13061996
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
Vandana Devesh Sharma
 
Comparison microscope
Comparison microscopeComparison microscope
Comparison microscope
Ketan Patil
 
Pattern recognition palm print authentication system
Pattern recognition palm print authentication systemPattern recognition palm print authentication system
Pattern recognition palm print authentication system
Mazin Alwaaly
 
Thin layer chromatography
Thin layer chromatographyThin layer chromatography
Thin layer chromatography
Pratyay Seth
 
Latent fingerprint development
Latent fingerprint developmentLatent fingerprint development
Latent fingerprint development
Tejasvi Bhatia
 
Comparative study of security features of passports of different countries
Comparative study of security features of passports of different countriesComparative study of security features of passports of different countries
Comparative study of security features of passports of different countries
Akashlina Basu
 
Blood grouping from bloodstains.pptx
Blood grouping from bloodstains.pptxBlood grouping from bloodstains.pptx
Blood grouping from bloodstains.pptx
Suchita Rawat
 
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
John Donohue
 
Toolmarks
ToolmarksToolmarks
Toolmarks
Aman Bhargava
 

What's hot (20)

(PART-II)- Development of latent fingerprints using powders method.pptx
(PART-II)- Development of latent fingerprints using powders method.pptx(PART-II)- Development of latent fingerprints using powders method.pptx
(PART-II)- Development of latent fingerprints using powders method.pptx
 
Chromatography : A seperation technique
Chromatography : A seperation techniqueChromatography : A seperation technique
Chromatography : A seperation technique
 
Brain electrical oscillation signature profiling(beos)
Brain electrical oscillation signature profiling(beos)Brain electrical oscillation signature profiling(beos)
Brain electrical oscillation signature profiling(beos)
 
Dna in criminal justice_complete slides
Dna in criminal justice_complete slidesDna in criminal justice_complete slides
Dna in criminal justice_complete slides
 
Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)
 
ear print.pptx
ear print.pptxear print.pptx
ear print.pptx
 
Thin layer chromatography. (TLC)
Thin layer chromatography. (TLC)Thin layer chromatography. (TLC)
Thin layer chromatography. (TLC)
 
conventional methods of fingerprint development
 conventional methods of fingerprint development conventional methods of fingerprint development
conventional methods of fingerprint development
 
Forensic analysis of soil
Forensic analysis of soilForensic analysis of soil
Forensic analysis of soil
 
Asphyxia.pptx
Asphyxia.pptxAsphyxia.pptx
Asphyxia.pptx
 
Sample preparation for Chromatography
Sample preparation for Chromatography Sample preparation for Chromatography
Sample preparation for Chromatography
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
 
Comparison microscope
Comparison microscopeComparison microscope
Comparison microscope
 
Pattern recognition palm print authentication system
Pattern recognition palm print authentication systemPattern recognition palm print authentication system
Pattern recognition palm print authentication system
 
Thin layer chromatography
Thin layer chromatographyThin layer chromatography
Thin layer chromatography
 
Latent fingerprint development
Latent fingerprint developmentLatent fingerprint development
Latent fingerprint development
 
Comparative study of security features of passports of different countries
Comparative study of security features of passports of different countriesComparative study of security features of passports of different countries
Comparative study of security features of passports of different countries
 
Blood grouping from bloodstains.pptx
Blood grouping from bloodstains.pptxBlood grouping from bloodstains.pptx
Blood grouping from bloodstains.pptx
 
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
 
Toolmarks
ToolmarksToolmarks
Toolmarks
 

Viewers also liked

Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysis
patil17
 
Chromatography
ChromatographyChromatography
Chromatography
Kamruzzaman Khan
 
HPLC Osnove SŠFKZ Aleš 18.3.2010
HPLC Osnove SŠFKZ Aleš 18.3.2010HPLC Osnove SŠFKZ Aleš 18.3.2010
HPLC Osnove SŠFKZ Aleš 18.3.2010
Aleš Čerin
 
σύγχρονες θεωρίες για την μάθηση
σύγχρονες θεωρίες για την μάθησησύγχρονες θεωρίες για την μάθηση
σύγχρονες θεωρίες για την μάθησηΓιάννης Παπαδάκης
 
Rheology for the food industry
Rheology for the food industryRheology for the food industry
Rheology for the food industry
Jackie Trịnh
 
Xinobiotic Biodegradation
Xinobiotic BiodegradationXinobiotic Biodegradation
Xinobiotic Biodegradation
Amit Pratush
 
Biotechnological tools
Biotechnological toolsBiotechnological tools
HPLC
HPLCHPLC
basic techniques of biotechnology
basic techniques of biotechnologybasic techniques of biotechnology
basic techniques of biotechnology
shalinikaushik
 
GM crops food security ppt
GM crops food security pptGM crops food security ppt
GM crops food security ppt
Sai Bhaskar Reddy Nakka
 
Gmo
GmoGmo
Gmo
sai abi
 
Genetically Modified Organisms
Genetically Modified OrganismsGenetically Modified Organisms
Genetically Modified Organismsgreengirl825
 
Bioremediation and biodegradation
Bioremediation and biodegradationBioremediation and biodegradation
Bioremediation and biodegradationrharishkmr
 
Genetic Modified Crops
Genetic Modified CropsGenetic Modified Crops
Genetic Modified Crops
pham van chuyen
 
Week 11 - 12 Food And Beverage Service Method 3-2552
Week 11 - 12  Food And Beverage Service Method 3-2552Week 11 - 12  Food And Beverage Service Method 3-2552
Week 11 - 12 Food And Beverage Service Method 3-2552Pavit Tansakul
 
Genetically Modified Organisms (GMO)
Genetically Modified Organisms (GMO)Genetically Modified Organisms (GMO)
Genetically Modified Organisms (GMO)Tanvi Vasani
 
GMO presentation
GMO presentationGMO presentation
GMO presentation
tangeld
 
Genetically modified crops and food Security..scientific facts
Genetically modified crops and food Security..scientific factsGenetically modified crops and food Security..scientific facts
Genetically modified crops and food Security..scientific factsRajdeeep sidhu
 
Food additives
Food additives Food additives
Food additives amarsujith
 
Techniques of electrophoresis
Techniques of electrophoresisTechniques of electrophoresis
Techniques of electrophoresisSayanti Sau
 

Viewers also liked (20)

Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysis
 
Chromatography
ChromatographyChromatography
Chromatography
 
HPLC Osnove SŠFKZ Aleš 18.3.2010
HPLC Osnove SŠFKZ Aleš 18.3.2010HPLC Osnove SŠFKZ Aleš 18.3.2010
HPLC Osnove SŠFKZ Aleš 18.3.2010
 
σύγχρονες θεωρίες για την μάθηση
σύγχρονες θεωρίες για την μάθησησύγχρονες θεωρίες για την μάθηση
σύγχρονες θεωρίες για την μάθηση
 
Rheology for the food industry
Rheology for the food industryRheology for the food industry
Rheology for the food industry
 
Xinobiotic Biodegradation
Xinobiotic BiodegradationXinobiotic Biodegradation
Xinobiotic Biodegradation
 
Biotechnological tools
Biotechnological toolsBiotechnological tools
Biotechnological tools
 
HPLC
HPLCHPLC
HPLC
 
basic techniques of biotechnology
basic techniques of biotechnologybasic techniques of biotechnology
basic techniques of biotechnology
 
GM crops food security ppt
GM crops food security pptGM crops food security ppt
GM crops food security ppt
 
Gmo
GmoGmo
Gmo
 
Genetically Modified Organisms
Genetically Modified OrganismsGenetically Modified Organisms
Genetically Modified Organisms
 
Bioremediation and biodegradation
Bioremediation and biodegradationBioremediation and biodegradation
Bioremediation and biodegradation
 
Genetic Modified Crops
Genetic Modified CropsGenetic Modified Crops
Genetic Modified Crops
 
Week 11 - 12 Food And Beverage Service Method 3-2552
Week 11 - 12  Food And Beverage Service Method 3-2552Week 11 - 12  Food And Beverage Service Method 3-2552
Week 11 - 12 Food And Beverage Service Method 3-2552
 
Genetically Modified Organisms (GMO)
Genetically Modified Organisms (GMO)Genetically Modified Organisms (GMO)
Genetically Modified Organisms (GMO)
 
GMO presentation
GMO presentationGMO presentation
GMO presentation
 
Genetically modified crops and food Security..scientific facts
Genetically modified crops and food Security..scientific factsGenetically modified crops and food Security..scientific facts
Genetically modified crops and food Security..scientific facts
 
Food additives
Food additives Food additives
Food additives
 
Techniques of electrophoresis
Techniques of electrophoresisTechniques of electrophoresis
Techniques of electrophoresis
 

Similar to Chromatogtaphy

Chromatography, types by different approaches, HPLC
Chromatography, types by  different approaches, HPLC Chromatography, types by  different approaches, HPLC
Chromatography, types by different approaches, HPLC
Muhammad Asif Shaheeen
 
Introduction of chromatography
Introduction of chromatographyIntroduction of chromatography
Introduction of chromatography
Deepa Verma
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
Dr Alok Bharti
 
Chromatography and its types
Chromatography and its typesChromatography and its types
Chromatography and its types
SindhBiotech
 
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptxDSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
Mathabhanga College
 
Adsorption Chromatography Assignment.pdf
Adsorption Chromatography Assignment.pdfAdsorption Chromatography Assignment.pdf
Adsorption Chromatography Assignment.pdf
MuhammadFaizan389
 
Lecture 1 analytical chemistry 4-second stage
Lecture 1 analytical chemistry 4-second stageLecture 1 analytical chemistry 4-second stage
Lecture 1 analytical chemistry 4-second stage
Ashraf Saad
 
Column Chromatography
Column ChromatographyColumn Chromatography
Column Chromatography
Mahadev Kammar
 
Chromatography
ChromatographyChromatography
Chromatography
Dr.Prameswari Kasa
 
Extraction chromatography
Extraction chromatographyExtraction chromatography
Extraction chromatography
Shivaji Burungale
 
Column Chromatography
Column ChromatographyColumn Chromatography
Column Chromatography
Sabbir Hoshen
 
Columchromatography rv
Columchromatography rvColumchromatography rv
Columchromatography rv
RAGHAV DOGRA
 
Chromatography
ChromatographyChromatography
ChromatographyDrLeezum
 
Chromatography
ChromatographyChromatography
Chromatography
manzara arshad
 
Introduction to Chromatography (Column chromatography)
Introduction to Chromatography (Column chromatography)Introduction to Chromatography (Column chromatography)
Introduction to Chromatography (Column chromatography)
Ahmed Metwaly
 
Columnchromatography
ColumnchromatographyColumnchromatography
Columnchromatography
Mahendra G S
 
Column chromatography
Column chromatographyColumn chromatography
Column chromatographyRakesh Guptha
 
columnchromatography-131008023940-phpapp02.pdf
columnchromatography-131008023940-phpapp02.pdfcolumnchromatography-131008023940-phpapp02.pdf
columnchromatography-131008023940-phpapp02.pdf
UMAIRASHFAQ20
 
Chromatography.
Chromatography.Chromatography.
Chromatography.
Yasir Mehmood
 
Chromatography.ppt
Chromatography.pptChromatography.ppt
Chromatography.ppt
ZORAIZ HAIDER
 

Similar to Chromatogtaphy (20)

Chromatography, types by different approaches, HPLC
Chromatography, types by  different approaches, HPLC Chromatography, types by  different approaches, HPLC
Chromatography, types by different approaches, HPLC
 
Introduction of chromatography
Introduction of chromatographyIntroduction of chromatography
Introduction of chromatography
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
Chromatography and its types
Chromatography and its typesChromatography and its types
Chromatography and its types
 
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptxDSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
 
Adsorption Chromatography Assignment.pdf
Adsorption Chromatography Assignment.pdfAdsorption Chromatography Assignment.pdf
Adsorption Chromatography Assignment.pdf
 
Lecture 1 analytical chemistry 4-second stage
Lecture 1 analytical chemistry 4-second stageLecture 1 analytical chemistry 4-second stage
Lecture 1 analytical chemistry 4-second stage
 
Column Chromatography
Column ChromatographyColumn Chromatography
Column Chromatography
 
Chromatography
ChromatographyChromatography
Chromatography
 
Extraction chromatography
Extraction chromatographyExtraction chromatography
Extraction chromatography
 
Column Chromatography
Column ChromatographyColumn Chromatography
Column Chromatography
 
Columchromatography rv
Columchromatography rvColumchromatography rv
Columchromatography rv
 
Chromatography
ChromatographyChromatography
Chromatography
 
Chromatography
ChromatographyChromatography
Chromatography
 
Introduction to Chromatography (Column chromatography)
Introduction to Chromatography (Column chromatography)Introduction to Chromatography (Column chromatography)
Introduction to Chromatography (Column chromatography)
 
Columnchromatography
ColumnchromatographyColumnchromatography
Columnchromatography
 
Column chromatography
Column chromatographyColumn chromatography
Column chromatography
 
columnchromatography-131008023940-phpapp02.pdf
columnchromatography-131008023940-phpapp02.pdfcolumnchromatography-131008023940-phpapp02.pdf
columnchromatography-131008023940-phpapp02.pdf
 
Chromatography.
Chromatography.Chromatography.
Chromatography.
 
Chromatography.ppt
Chromatography.pptChromatography.ppt
Chromatography.ppt
 

Recently uploaded

The Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptxThe Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptx
DhatriParmar
 
Pride Month Slides 2024 David Douglas School District
Pride Month Slides 2024 David Douglas School DistrictPride Month Slides 2024 David Douglas School District
Pride Month Slides 2024 David Douglas School District
David Douglas School District
 
Introduction to AI for Nonprofits with Tapp Network
Introduction to AI for Nonprofits with Tapp NetworkIntroduction to AI for Nonprofits with Tapp Network
Introduction to AI for Nonprofits with Tapp Network
TechSoup
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
Jisc
 
The Diamond Necklace by Guy De Maupassant.pptx
The Diamond Necklace by Guy De Maupassant.pptxThe Diamond Necklace by Guy De Maupassant.pptx
The Diamond Necklace by Guy De Maupassant.pptx
DhatriParmar
 
The Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official PublicationThe Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official Publication
Delapenabediema
 
The basics of sentences session 5pptx.pptx
The basics of sentences session 5pptx.pptxThe basics of sentences session 5pptx.pptx
The basics of sentences session 5pptx.pptx
heathfieldcps1
 
Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
DeeptiGupta154
 
Natural birth techniques - Mrs.Akanksha Trivedi Rama University
Natural birth techniques - Mrs.Akanksha Trivedi Rama UniversityNatural birth techniques - Mrs.Akanksha Trivedi Rama University
Natural birth techniques - Mrs.Akanksha Trivedi Rama University
Akanksha trivedi rama nursing college kanpur.
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
Jisc
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
Balvir Singh
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Thiyagu K
 
Chapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptxChapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptx
Mohd Adib Abd Muin, Senior Lecturer at Universiti Utara Malaysia
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
Pavel ( NSTU)
 
TESDA TM1 REVIEWER FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...
TESDA TM1 REVIEWER  FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...TESDA TM1 REVIEWER  FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...
TESDA TM1 REVIEWER FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...
EugeneSaldivar
 
Azure Interview Questions and Answers PDF By ScholarHat
Azure Interview Questions and Answers PDF By ScholarHatAzure Interview Questions and Answers PDF By ScholarHat
Azure Interview Questions and Answers PDF By ScholarHat
Scholarhat
 
Best Digital Marketing Institute In NOIDA
Best Digital Marketing Institute In NOIDABest Digital Marketing Institute In NOIDA
Best Digital Marketing Institute In NOIDA
deeptiverma2406
 
A Strategic Approach: GenAI in Education
A Strategic Approach: GenAI in EducationA Strategic Approach: GenAI in Education
A Strategic Approach: GenAI in Education
Peter Windle
 
BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...
BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...
BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...
Nguyen Thanh Tu Collection
 
Lapbook sobre os Regimes Totalitários.pdf
Lapbook sobre os Regimes Totalitários.pdfLapbook sobre os Regimes Totalitários.pdf
Lapbook sobre os Regimes Totalitários.pdf
Jean Carlos Nunes Paixão
 

Recently uploaded (20)

The Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptxThe Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptx
 
Pride Month Slides 2024 David Douglas School District
Pride Month Slides 2024 David Douglas School DistrictPride Month Slides 2024 David Douglas School District
Pride Month Slides 2024 David Douglas School District
 
Introduction to AI for Nonprofits with Tapp Network
Introduction to AI for Nonprofits with Tapp NetworkIntroduction to AI for Nonprofits with Tapp Network
Introduction to AI for Nonprofits with Tapp Network
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
 
The Diamond Necklace by Guy De Maupassant.pptx
The Diamond Necklace by Guy De Maupassant.pptxThe Diamond Necklace by Guy De Maupassant.pptx
The Diamond Necklace by Guy De Maupassant.pptx
 
The Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official PublicationThe Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official Publication
 
The basics of sentences session 5pptx.pptx
The basics of sentences session 5pptx.pptxThe basics of sentences session 5pptx.pptx
The basics of sentences session 5pptx.pptx
 
Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
 
Natural birth techniques - Mrs.Akanksha Trivedi Rama University
Natural birth techniques - Mrs.Akanksha Trivedi Rama UniversityNatural birth techniques - Mrs.Akanksha Trivedi Rama University
Natural birth techniques - Mrs.Akanksha Trivedi Rama University
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
 
Chapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptxChapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptx
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
 
TESDA TM1 REVIEWER FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...
TESDA TM1 REVIEWER  FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...TESDA TM1 REVIEWER  FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...
TESDA TM1 REVIEWER FOR NATIONAL ASSESSMENT WRITTEN AND ORAL QUESTIONS WITH A...
 
Azure Interview Questions and Answers PDF By ScholarHat
Azure Interview Questions and Answers PDF By ScholarHatAzure Interview Questions and Answers PDF By ScholarHat
Azure Interview Questions and Answers PDF By ScholarHat
 
Best Digital Marketing Institute In NOIDA
Best Digital Marketing Institute In NOIDABest Digital Marketing Institute In NOIDA
Best Digital Marketing Institute In NOIDA
 
A Strategic Approach: GenAI in Education
A Strategic Approach: GenAI in EducationA Strategic Approach: GenAI in Education
A Strategic Approach: GenAI in Education
 
BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...
BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...
BÀI TẬP BỔ TRỢ TIẾNG ANH GLOBAL SUCCESS LỚP 3 - CẢ NĂM (CÓ FILE NGHE VÀ ĐÁP Á...
 
Lapbook sobre os Regimes Totalitários.pdf
Lapbook sobre os Regimes Totalitários.pdfLapbook sobre os Regimes Totalitários.pdf
Lapbook sobre os Regimes Totalitários.pdf
 

Chromatogtaphy

  • 2. 2 Chromatography began at the turn of 20th century when Ramsey {Ramsey., Proc.Roy. soc.A76,111 (1905)} separated mixtures of gases and vapours on adsorbents like charcoal and Michael Tswett [Tswett, M., Ber.deut.botan. Ges., 24, 316 & 386(1906)] separated plant pigments by liquid chromatography. William Ramsey Michael Tswett
  • 3. 3  Tswett is credited as being the "Father of Chromatography" principally because he coined the term chromatography (literally colour writing) and scientifically described the process.
  • 4. 4 MICHAEL TSWETT  He was a Russian botanist.  His main interest was to find out the composition of the green plant pigment, chlorophyll.  There were two schools of thought on this subject.  One school thought that chlorophyll was a single solute.  The other school thought that chlorophyll was a mixture of different solutes.
  • 5. 5  A glass column 1-5 cm in diameter and 20-100 cm long was taken.  The column was packed with the slurry of Calcium Carbonate made in petroleum ether(60-80). MICHAEL TSWETT’S EXPERIMENT
  • 6. 6  The solution of chlorophyll was allowed to move down slowly and then the column was washed with ethyl alcohol for few hours.  A concentrated solution of chlorophyll was prepared in petroleum ether(60-80) and a small volume (~10cm3 ) was put on the top of the calcium carbonate column in the glass column as a thin band. MICHAEL TSWETT’S EXPERIMENT
  • 7. 7 COTTON PLUG SLURRY OF CaCO3 POWDER IN PETROLEUM ETHER COTTON PLUG CONCENTRATED SOLUTION OF CHLOROPHYLL IN PETROLEUM ETHER WASHING/ ELUTION WITH ETHANOL
  • 8. 8 β CHLOROPHYLL α CHLOROPHYLL XANTHOPHYLL VIOLA XANTHINE CAROTENE COTTON PLUG SLURRY OF CaCO3 POWDER IN PETROLEUM ETHER COTTON PLUG CONCENTRATED SOLUTION OF CHLOROPHYLL IN PETROLEUM ETHER WASHING/ ELUTION WITH ETHANOL
  • 9. 9 • Each band was removed from the column and identified. • It was concluded that chlorophyll is a mixture of five different solutes. α –chlorophyll, β - chlorophyll, xanthophyll, viola xanthene, and carotene .
  • 10. 10 α CHLOROPHYLL and β CHLOROPHYLL
  • 14. 14 • Michael Tswett’s main problem was resolved and in the process he discovered a simple, time saving and relatively less laborious method of separating complex solutes from a mixture. • For example α – chlorophyll (MW 893) differs from β – chlorophyll by only one methyl group yet these two solutes are easily separated in the simplest apparatus.
  • 15. 15 • Today the term chromatography is used for separations involving colourless solutes also. This is done to honour the discoverer of the technique. • Michael Tswett coined the term “Chromatography for this new technique of separation. (Chromatus = colour, graphein = to write) • He visualized the coloured bands from chlorophyll as coloured writing.
  • 16. 16  As a chemist, a researcher or simply as an inquisitive person, two questions come to our mind on the results of Tswett’s experiment. 1. Why do the five solutes separate on the column? 2. How do the five solutes separate on the column?
  • 17. 17 Migration velocities increase from β- chlorophyll to carotene. “The five solutes are separated on the column because of the differences in their migration velocities”
  • 18. 18 “The five solutes have different migration velocities across the column of CaCO3, because they have different degrees of adsorption on the surface of CaCO3”
  • 19. 19  Carotene has the least degree of adsorption on CaCO3 surface and hence the highest migration velocity while β- chlorophyll has the highest degree of adsorption on CaCO3 surface and hence the lowest migration velocity of the five solutes.
  • 20. 20  This is the basis of “Adsorption column chromatography”.
  • 21. 21 Adsorbents  The most important agent in chromatography is the adsorbent.  Therefore, utmost care is to be bestowed in the selection of proper adsorption material and also the process of building up a column with it and the maintenance of the column.
  • 22. 22  It must be insoluble in the solution under analysis and the solvent that is used for elution Essential requirements to be satisfied by every adsorbent  It should not chemically react with the mixture under analysis.  It should generally be colourless, but this is not a rigid rule as there are instances where coloured adsorbents have been used successfully. Particle-size of the adsorbent material is of primary importance for successful separation.
  • 23. 23 Too loose , large paricles. Too tight , small paricles.
  • 24. 24  Slaked lime (120 mesh) and magnesia are two cheap adsorbents.  Fuller's earths are also good adsorbents but unlike the previous ones they are acidic in reaction. On moistening they darken considerably.  Calcium carbonate is a cheap adsorbent of uniform size. • Sodium sulphate, talc, magnesium sulphate, calcium phosphate, calcium hydroxide; barium carbonate, magnesium carbonate, magnesium trisilicate, calcium sulphate, soda ash, titania, charcoal, sucrose, starch are some of the other materials.
  • 25. 25 • Adsorbents may be classified as follows according to the force with which they hold the ions. Type Examples Weak Sucrose and starch Intermediate Calcium carbonate, calcium phosphate, calcium hydroxide and magnesium hydroxide Strong Alumina, charcoal and Fuller's earth.
  • 26. 26  From Tswett’s experiment if we change slurry of CaCO3 and replace it with slurry of Al2O3, will the five solutes form chlorophyll separate and will the sequence of separation remain the same?  In Tswett’s experiment if we replace chlorophyll extract with a mixture of A+B+C+D+E solutes, will the solutes separate and what will be the sequence of their separation?
  • 27. 27  In Tswett’s experiment if we replace chlorophyll extract with a mixture of A+B+C+D+E solutes, will the solutes separate and what will be the sequence of their separation?  If we have understood the concept of adsorption, then it should not be difficult to answer the above two questions.
  • 28. 28  In the first instance, we have changed the adsorbent, other things remaining unchanged.  The five adsorbates have a new surface of Al2O3.  If they exhibit differences in the degree of their adsorption on the surface of Al2O3, they will get separated.
  • 29. 29  In the second instance, we have changed the five solutes, other things remaining unchanged.  Although the surface of CaCO3 is the same as that in Tswett’s experiment, the five solutes are different.  They will get separated, if they exhibit differences in the degree of their adsorption on the surface of CaCO3.
  • 30. 30  Let us consider a situation where in the glass column we put a liquid which remains stationary even if the tap is opened.  If we allow three solutes A, B and C to pass through the chromatographic column and realize their separation after some time, we can ask again the same two questions which we did in adsorption chromatography.  Another liquid which is immiscible with the stationary liquid moves through it.
  • 31. 31  Why do the three solutes separate?  How do the three solutes separate?  The three solutes separate because they have different migration velocities.  Solute ‘A’ has the highest migration velocity of the three and reaches the end of the column first.  Solute ‘C’ has the lowest migration velocity of the three and reaches the end of the column last.
  • 32. 32  The migration velocity increases from solute ‘C’ to solute ‘A’.  Solute ‘A’ travels the fastest because it is more soluble in the liquid which moves across the stationary liquid.  Solute ‘C’ on the other hand is more soluble in the stationary phase and hence lags behind.
  • 33. 33  The principle of separation here is differences in solubility or partitioning of solutes between the stationary and mobile liquid phases.  It is called as “Partition Column Chromatography”.
  • 34. 34  If they do not show differences with reference to any property they cannot be separated.  For solutes to get separated from a mixture, they must exhibit some differences with reference to some property like solubility, boiling point, adsorption, partitioning etc.,
  • 35. 35 Stationary phase: Phase which remains stationary. It could be a solid or a liquid bonded to an inert solid support. Mobile phase: A phase which moves across the stationary phase. It could be a gas or a liquid. Elution: The process of washing a chromatographic column with a liquid or a gas. Solutes: This term is used in place of compounds or substances.
  • 36. 36 Normal phase mode of separation: Here the stationary phase is relatively more polar than the mobile phase. Reverse phase mode of separation: Here the stationary phase is relatively less polar than the mobile phase.
  • 37. 37 MOBILE PHASE STATIONARY PHASE PRINCIPLE OF SEPARATION CONTAINER NAME OF THE CHROMATOGRAPHIC TECHNIQUE LIQUID SOLID DIFFERENCES IN ADSORPTION COLUMN LIQUID-SOLID ADSORPTION COLUMN CHROMATOGRAPHY LIQUID SOLID DIFFERENCES IN ADSORPTION THIN LAYER LIQUID-SOLID ADSORPTION THIN LAYER CHROMATOGRAPHY LIQUID LIQUID DIFFERENCES IN PARTITIONING COLUMN LIQUID-LIQUID PARTITION COLUMN CHROMATOGRAPHY LIQUID LIQUID DIFFERENCES IN PARTITIONING THIN LAYER LIQUID-LIQUID PARTITION THIN LAYER CHROMATOGRAPHY GAS LIQUID DIFFERENCES IN PARTITIONING COLUMN GAS-LIQUID-PARTITION COLUMN CHROMATIGRAPHY GAS SOLID DIFFERENCES IN ADSORPTION COLUMN GAS-SOLID ADSORPTION COLUMN CHROMATOGRAPHY
  • 38. 38 GAS LIQUID/ SOLUTION CLASSIFICATION OF DIFFERENT CHROMATOGRAPHIC TECHNIQUES Stationary phase is solid Stationary phase is a liquid coated on the surface of an inert solid support gas - solid - adsorption chromatography gas - liquid - partition chromatography Mobile phase is Gas and all the solutes being separated are in gaseous phase under experimental conditions Mobile phase is liquid and all the solutes being separated are in liquid/solution state
  • 39. 39 Liquid / Solution Chromatography ADSORPTION : Separation of solutes on the basis of their degree of adsorption on the surface of an adsorbent. PARTITION : Separation of solutes on the basis of their differences in partition coefficient of solutes in two mutually immiscible solvents.ION EXCHANGE : Separation of similarly charged ions based on the differences in the degree of their affinity towards the ion exchange resin. SIZE EXCLUSION : Separation of solutes based on the differences in their molecular weights or sizes. ELECTROPHORETIC : Separation based on the movement of charged solutes to opposite poles under the applied electrical field. AFFINITY : Separation of solutes based on specific affinity of solutes on the specific sites of the stationary phase.
  • 40. 40 • Let us consider a chromatographic column on which three solutes A, B and C are separated in that order. • Let there be sensor or a detector fixed at the other end of the column. • The detector can sense and count the number of molecules of solute entering into it. • The detector does not sense and count the number of molecules of mobile phase entering into it.
  • 41. 41
  • 42. 42 • As we can see from the figure, first the mobile phase will enter into the detector and no response will be generated. DETECTOR A B C COTTON PLUG COTTON PLUG
  • 43. 43  Molecules of solute ‘B’ enter the detector next. They are sensed and counted.  Mobile phase enters into the detector again and is not sensed and counted. Molecules of solute ‘C’ enter the detector next. They are sensed and counted.  Mobile phase enters into the detector again and is not sensed and counted.  Molecules of solute ‘A’ enter the detector next. They are sensed and counted.  Mobile phase enters into the detector again and is not sensed and counted.
  • 44. 44
  • 45. 45 • If we plot a graph of graph of detector response as a function of time, the above description can be converted into the following picture. • • This is called a chromatogram. A B C DETECTOR RESPONSE TIME
  • 47. 47 • A normal chromatogram has certain characteristic features. All the peaks are bell shaped symmetrical peaks.  Neighbouring peaks are separated by some distance.  Each peak has a characteristic retention time (distance between center of the peak and the reference point) under given experimental conditions. This is useful in qualitative Chromatographic analysis.
  • 48. 48  Each peak has certain area under curve (AUC) and AUC α Concentration of Solute. This is useful in quantitative chromatographic analysis. A separate calibration graph is needed for each peak in a chromatogram.  Each peak has certain width at base which increases with increase in retention time.
  • 49. 49  Both qualitative and quantitative chromatographic analysis are relative methods of analysis in which the standard solutes and the sample solutes are chromatographed under the same experimental conditions.  Volumetric and Gravimetric methods, on the other hand, are the examples of absolute methods of quantitative analysis.
  • 50. 50 • It is observed in practical Chromatographic separations that the spots or bands or peaks are always broad. They are never very sharp.  What could be the reason for this? • If we observe our earlier Chromatogram carefully, we see that for each solute, solute molecules start eluting at a particular time and the elution is complete after a few seconds.
  • 51. 51 • It means that all the molecules of the same solute do not travel across the stationary phase with the same speed. Some travel faster than others. Therefore some reach the detector first and others latter. • In other words molecules of the same solute spread across the stationary phase which results in band broadening.
  • 52. 52 • We can visualize four important causes leading to molecular spreading and band broadening. • We must find out at least some important causes for molecular spreading and band broadening.
  • 53. 53 • We should know in the beginning that the particles of stationary phases are approximately spherical in nature and the particles are not exactly of the same size. • When we say that the particle size is 100µ, it really means that it is the average particle size and that every particle is not of exact 100µ diameter. • Secondly, the spherical particles are always porous in nature. That is, the surface is not smooth.
  • 54. 54 Let us now imagine a magnified picture of a chromatographic column • We can see that when the solute molecules travel across the stationary phase, they have different paths to flow down the column. Each path has different path length. COTTON PLUG DETECTOR SAMPLE SOLUTION WASHING WITH LIQUID MOBILE PHASE
  • 55. 55 • The molecular spreading results in band broadening. • This cause of molecular spreading leading to band broadening is known as, “Eddy Diffusion” or “Multiple path flows Effect”. • Solute molecules travelling in shorter paths reach the detector early. • Solute molecules travelling in longer paths reach the detector late. • As a result of multiple path flows, molecules of the same solute spread across the stationary phase .
  • 56. 56 1. Eddy diffusion. (difference in the mobile phase stream flow due to non homogeneity of stationary phase) in packed columns.
  • 57. 57 1. No or minimum Eddy diffusion. (difference in the mobile phase stream flow due to non homogeneity of stationary phase) in packed columns.
  • 58. 58  Can Eddy diffusion be avoided? • Even if we bring the stationary phase particles closer together than what they are in the above picture, there will always be some distance between the neighbouring particles. • What is achieved by compact packing of stationary phase particles in the column probably is that the numbers of multiple path flows are minimized and the extent of molecular spreading leading to band broadening is reduced. • Eddy diffusion, therefore, cannot be avoided, it can be optimized.
  • 59. 59 • Let us now have a magnified picture of two neighbouring stationary phase particles.
  • 60. 60 • In the above picture we have three streams in which solute molecules are moving between two stationary phase particles. • Two streams flow closer the surface of the stationary phase particles. DETECTOR
  • 61. 61 • Solute molecules in these streams experience resistance to flow from the surface of the stationary phase particles. • Their migration velocity will be relatively smaller than those solute molecules traveling in the middle. • This obviously results in molecular spreading adding to band broadening. • This cause is known as mobile phase mass transfer and cannot be eliminated.
  • 62. 62 • In closely packed column the contribution of this cause called “Mobile phase mass transfer” will be more than in a loosely packed column. • In a loosely packed column the contribution of Eddy diffusion will be more. • In practice we have to strike a balance.
  • 63. 63 Let us now consider two pictures of two magnified porous particles. • In both the above pictures A and B, while solute molecules travel across the stationary phase , some of them enter inside the pores and remain stagnant for sometime before they are flushed out by the mobile phase which is continuously flowing across the chromatographic column. •There is a small difference however. DETECTOR A DETECTOR B
  • 64. 64 • In picture A the pore is shallow and the solute molecules remain stagnant along with the mobile phase for short time. This cause is known as “Stagnant Mobile Phase Mass Transfer”. • In picture B, solute molecules enter into a deep pore. Inside the deep pore they have enough time to interact with the liquid stationary phase coated on the surface of the stationary phase particles.They remain in the deep pore for a longer period of time as that compared to solute molecules in a shallow pore. • In either case, the solute molecules which enter into the pores have relatively smaller migration velocity as when compared to those which do not enter into the pores. • This cause is known as “Stationary Phase Mass Transfer”. • This results in molecular spreading and band broadening.
  • 65. 65 • Eddy diffusion, mobile phase mass transfer, stagnant mobile phase mass transfer, and stationary phase mass transfer are the four main causes for molecular spreading and band broadening.
  • 66. 66 • The causes for molecular spreading leading to band broadening cannot be eliminated. • They can be optimized by choosing stationary phase particles with uniform size having shallow pores and packing compactly in a chromatographic column.
  • 67. 67 • Let us now consider in quantitative terms what we have discussed qualitatively about molecular spreading and band broadening in chromatographic separations.
  • 68. 68 • In order to do that let us make the following assumptions: tR - retention time of a solute L cm - Length of chromatographic column tO - retention time of the mobile phase u - migration velocity of the mobile phase uX - migration velocity of the solute ns - number of moles of solute in the stationary phase nm - number of moles of solute in the mobile phase R - fraction of moles of solute in the mobile phase
  • 69. 69 • Let us imagine that a chromatographic column of Lcm length is divided into N small sections of equal length or height. L cm • Each small section is called a “Theoretical Plate” because it is a theoretical concept. • No such sections or plates exist in reality.
  • 70. 70 • The solute molecules constantly get exchanged between the stationary phase and the mobile phase when they travel across the column. • The solute molecules travel in the forward direction along with the mobile phase only when they are in the mobile phase . They remain stagnant when they are in the stationary phase. • The length or height of each theoretical plate is such that when solute molecules travel that distance a new equilibrium is established. • There will be, therefore ‘N’ number of such equilibria across the length of the column.
  • 71. 71 • It is then easy to imagine that more the number of theoretical plates associated with the given length of chromatographic column, more will be such equilibria and solute molecules will have less time and chances to spread across the stationary phase. • It means that more the N (number of theoretical plates), lesser is the molecular spreading and sharper will be the resulting bands or peaks.
  • 72. 72 • In other words we can say that higher the N associated with a given length of chromatographic column higher is “ column separation efficiency.” • We also know that L / N = H, height equivalent to a theoretiacal plate. N and H are inversely proportional. • It can, therefore, be concluded that higher the N or lower the H associated with a given length of chromatographic column, higher is the column separation efficiency.
  • 73. 73 • The following three factors govern the column separation efficiency. 1. Particle size of stationary phase. 2. Length of chromatographic column. 3. Mobile phase velocity.
  • 74. 74 Particle size of stationary phase • We normally assume stationary phase particles to be spherical in shape. • If we reduce the particle size diameter from 100µ to 50µ, 25µ, 10µ, 05µ, the surface area will increase as follows.
  • 75. 75 Particle size in diameter (microns) Surface area (m2 /g) 100 0.01538 50 0.0307 25 0.0615 10 0.153 5 0.307
  • 76. 76 0 20 40 60 80 100 120 0 0.1 0.2 0.3 0.4 Series1 Particlesizeindiameter(microns) Surfacearea(m2/g)
  • 77. 77 • We can see that as the particle size of spherical particles is reduced, the surface area increases exponentially. • When the surface area increases, the amount of the liquid phase coated also increases proportionately. More the liquid phase loading more will be the interactions of solute molecules with the liquid stationery phase, less will be molecular spreading and band broadening. • Then we can say that lower the particle size higher will be N and higher will be the column separation efficiency.
  • 78. 78 • However we have also to remember that as the stationary phase particles become smaller there is more resistant to mobile phase flow. •We have therefore to optimize the particle size without sacrificing analytical times. Because to achieve good separation with reasonable analytical time is practically important and cannot be overlooked.
  • 79. 79 • If with a given length of say 100cm long column, N = 10,000 then with L = 200cm, N= 20,000, keeping particle size of stationary phase the same. • Thus, longer the chromatographic column, higher will be N and column separation efficiency. • With longer columns, however, retention time of solutes will increase and peaks will become broader and analytical time will increase.
  • 80. 80 • We have to, therefore, optimize the column length just as we optimized the particle size. Our main aim being good separation with reasonable analytical time.
  • 81. 81 For example for 10µ diameter porous stationary phase particles we get a graph as shown below. • It can be seen that initially as the mobile phase velocity increases, H decreases i.e. separation efficiency increases. • It reaches at the maximum with mobile phase velocity ‘x’. Mobile phase velocity L/N X
  • 82. 82 • We may, therefore, be tempted to work at the mobile phase velocity ‘x’ because at this point we get the maximum separation efficiency. • From the above discussion it becomes clear that for getting good separation efficiency with reasonable separation time we have to optimize particle size, column length and mobile phase velocity. • However at this mobile phase velocity if the separation times are long, then we will have to work at higher mobile phase velocity at the cost of lower separation efficiency to achieve reasonable analytical time.
  • 83. 83 A TYPICAL CHROMATOGRAM  It is a plot of the detector response in millivolts as a function of time (usually time is the abscissa and millivolts the ordinate) Retention Time S o l v e n t A B C Inject Detector Response 100 0 Time in minutes0
  • 84. 84 • The above Fig. illustrates how resolution can be enhanced by exploiting the selectivity effect of the mobile phase (i.e, by switching from 60% ACN to 60% MeOH). It is important to note that small change of selectivity can have a major effect on resolution as resolution is proportional to (α -1). Columns of different bonded phases (i.e,C8, phenyl, CN ,polar- embedded etc) can also provide different selectivity effects.
  • 85. 85 • Finally, the plate number of the column (N) should be maximized by using a longer column or a more efficient column. Increasing N is not an efficient way to achieve resolution since Rs is proportional to √N. Doubling N by doubling the column length increases analysis time by 2 but only increases resolution by √2 or by 41%. In contrast, increasing α from 1.05 to 1.1 will almost double the resolution. For complex samples with many peaks, increasing N is a viable and the most direct approach.
  • 86. 86 Concept of tailing and fronting Individual solute molecules act independently of one another during the chromatographic process. As, a result they produce a randomized aggregation of retention times after repeated sorptions and desorptions. The result of a given solute is a distribution, or peak, whose shape can be approximated as being normal or Gaussian.  Nonsymmetrical peaks usually indicate that some undesirable interaction has taken place during the chromatographic process.
  • 87. 87 •Ideal Guassian Broad Fronting Tailing Doublet Rate theory Faulty injection. Degraded column.
  • 88. 88 The figure below shows some shapes that can occur in actual samples.
  • 89. 89  Broad peaks like (b) are more common in packed columns and usually indicate that the kinetics of mass transfer is too slow.  Asymmetric peaks can be classified as fronting (c) or tailing (d) depending on the location of asymmetry.  The extent of asymmetry is defined as the tailing factor (TF).
  • 90. 90 TF = b/a  Both a and b are measured at 10% of the peak height shown.  As can be seen from the equation, a tailing peak will have a TF greater than one.  The opposite symmetry, fronting, will yield a TF less than one.
  • 91. 91  The doublet peak like (e) can represent a pair of solutes that are not adequately separated. • Repeatability of a doublet peak should be verified because such peak shape can also result from faulty injection technique, too much sample, or degraded columns.
  • 92. Flow path for method developmentDefine Goal Detection Chromatography selection (Stationary phase & Mobile phase) Sample preparation Measurement Matrix Substitution System suitability Stability Validation Reporting
  • 93. DETECTION • Prepare a solution of 10 µg/mL in MeOH:water mixture in acidic, basic or neutral conditions •Assess against their corresponding blanks for UV absorbance and Fluorescence •Tune the ionisation parameter on LC-MS to assess fragmentation and potential suitability for LC-MS-MS Results may indicate likely detection mechanism and detector The most common HPLC detectors are – UV/VIS – Fluorescence – Refractive index – Electro chemical – LCMS
  • 94. DETECTION Detection method chosen depending on : Requirements of assay UV detection – easiest and most widely available But, has moderate selectivity and sensitivity Fluorescence detection- greater sensitivity( 10-1000 times higher than that of UV) and selective, But greater care is needed for IS selection E.C.D detection-comparable sensitivity to fluorescence and good selectivity (due to response to specific functional group undergoing oxidation or reduction, But has serious problem with stability at high sensitivity work LCMS highest sensitivity and specificity, But involves large cost
  • 95. Separation Most difficult and time consuming part -SAMPLE PREPARATION -Selectively enrich biological sample for quantification at appropriate level are possible -CHROMATOGRAPHIC SEPARATION Resolve and quantify property of analyte/ potential metabolite and IS in presence of endogenous material in short time as possible
  • 96. Chromatographic separation Fundamental Resolution Equation Rs = 1 √n k α-1 4 1+ k α How can resolution be changed ? Efficiency : √n - Column length, particle size, Column Quality, Temperature Retention : k - Eluent Strength, Column stationary phase, 1+ k Temperature Selectivity : α-1 - Stationary Phase (column), Mobile phase, α Mobile phase additives, Temperature In bioanalysis capacity factor more than 4 to 5 - Matrix effect
  • 97. Sample preparation • OBJECTIVE • To carry through as high percentage as possible of the analyte(s) from original sample to the determination steps (good recovery) • To ensure that the analyte remain in their original state • To carry through as few of the matrix components from the original sample to the determination step • Preparation of sample in solution form, compatible for introduction in specific instrument (HPLC, LCMS, GCMS, RIA etc)
  • 98. Sample preparation In general there are three different methods of sample preparation prior to assay: - – Protein Precipitation – Liquid - Liquid Extraction – Solid Phase Extraction
  • 99. Protein Precipitation • Denaturation and precipitation of proteins is done by an addition of an acid, solution containing a heavy metal ion or water miscible organic solvents (Methanol, ACN) • Following admixture of precipitant the sample is centrifuged to produce a clear supernatant • Precipitation by addition of organic solvent • Water miscible organic solvents ACN(x1 vol), MeOH(x 2 vol), ethanol or acetone • Use solvent in which analyte is highly soluble • Mixture of two solvents (MeOH and DMSO)
  • 100. Protein Precipitation • Precipitation by addition of inorganic compounds • Acidic or anionic precipitants trichloroacetic acid and/or perchloric acid (10-20 % w/v) • Mobile phase should contain a high molar concentration of buffer to protect column • Should not be used for compound which are prone to acid hydrolysis • Cationic precipitants such as copper and zinc in alkaline solution e.g Zinc sulphate in NaOH or barium hydroxide or copper sulphate in potassium hydroxide
  • 101. Protein Precipitation • Disadvantage Cannot be used for quantifying very low concentrations due to dilution of sample Late eluting peak may co-elute with drug peak distorting peak measurements Further precipitation may occur if supernatant is allowed to stand which could clog the chromatographic column(can be avoided)
  • 102. Liquid -Liquid Extraction • Based on extraction of analyte depending on its partition between an aqueous and an immiscible organic phase • The degree of extraction depends on partition coefficient (KD) which is dependent on • Organic solvent • pH • Ionic strength • Optimised for maximum recovery and minimum co- extract
  • 103. Liquid -Liquid Extraction • Analyte possessing sufficient lipophilic character gets extracted in organic layer • The organic layer is collected, dried and reconstituted in solvent, mobile phase or a solution of a particular pH. • Properties of ideal solvent • Environmentally acceptable, non toxic and not highly inflammable, immiscible with water, convenient specific gravity,suitable volatility, high chemical stability, not prone to form emulsion • Desirable compromise is required
  • 104. Liquid -Liquid Extraction • Most popular solvents are heptane, hexane, toluene, diethyl ether , ethyl acetate, dichloromethane and chloroform • It is best to use least polar solvent that will effectively extract drug. • No advantage in selecting solvent with too much extracting power it may reduce selectivity of assay • Acid drugs extracted under acidic condition (pH 2 units below pKa) • Basic drugs extracted under basic conditions (pH 2 units above pKa)
  • 105. Liquid -Liquid Extraction • Wherever possible high pH is desirable to ensure cleaner extract • To limit the carry-over • Back extraction into aqueous phase with pH adjusted so that analytes ionize again • Highly polar ionic drug are extracted by formation of neutral ion pair complex • Addition of ion pairing agents tetralkyammonium salts for acids and alkyl sulphonates (heptane sulphonic acid for base)
  • 106. Liquid -Liquid Extraction Disadvantages - Emulsion Formation - Time consuming - adsorption of analytes on glassware - Require large volume of organic solvent - Ionizable compounds require specific pH adjustment for extraction
  • 107. Solid Phase ExtractionSolid Phase Extraction SPE utilizes the principle of selective retention of an analyte on a chemically modified silica surface/sorbent. There are four steps involved in SPE – Conditioning – Application – Rinsing – Elution
  • 108. Solid Phase ExtractionSolid Phase Extraction Advantages – Faster Sample preparation – Average time cut by 2/3 – Less solvent and reagent consumption – Less hazardous waste – Greater Recoveries – Greater Accuracy – Greater Specificity – Less sample volume handling – Useful for both hydro and lipophilic drug – Easy automation Disadvantages – Expensive
  • 109. Concentration of sample EVAPORATION OF SAMPLE Simple but its practical application problematic -time and attention required -Loss of sample -In three ways : degradation, volatilization and incomplete transfer -Overcome by use Low Volume Evaporator -Sample is swept with a inert gas nitrogen.
  • 110. Standardisation • External standardization •Internal standardization Detector response is recorded as a ratio of the signal of the analyte to that of internal standard. Calibration Check
  • 111. Literature search for methods •Internet •Pubmed •Analytical Abstracts (Royal Society of chemist) www.rsc.org •Analytical Profiles
  • 112. Rules of one and two • Two basic rules applied while modifying the method • Rule of one states • “In altering any chromatographic parameter and this include the extraction step only one parameter is changed at any one time • Rule of two states • “A chromatographic observation is not valid until it has been repeated”
  • 113. Method ValidationMethod validation is a procedure used to ascertain and prove that a test method consistently yield what it is expected to do with adequate accuracy and precision. The method validation parameters are Specificity Ability of an assay to distinguish a drug from its metabolite, other drugs and endogenous constituents of biological matrix. Criteria Response of interfering peaks (if any) at the retention time of an analyte should be ≤ 20% of the LLOQ (Low limit of quantitation) response and should be ≤ 5% of the internal standard response.
  • 114. Blank Plasma on LCMS Blank Plasma on HPLC
  • 115. Sensitivity • The lowest level of concentration of an analyte that can be measured in biological matrix Criteria • The response of LLOQ should be at least 3 times greater than any interference peak. • The between batch accuracy must be ≤ 20% • The between batch precision must be ≤ 20%
  • 116. Low Limit of Quantitation on LCMS Low Limit of Quantitation on HPLC
  • 117. Linearity A minimum of five non-zero calibration curve standards including LLOQ and highest concentration should yield: • r2 ≥ 0.98 ∀ ± 20% deviation of LLOQ from nominal concentration ∀ ± 15% deviation of standards other than LLOQ Precision & Accuracy (-PA) • A PA batch is five non-zero calibration curve standards including LLOQ and highest concentration with five sets of: - • LLOQ Quality Control Sample • LQW Quality Control Sample • Middle Quality Control Sample • High Quality Control Sample
  • 118. Acceptance criteria • A minimum of five sets of PA batch should run on different days • Precision and accuracy must be ≤15% for: • within PA batch • Intra day PA batch • Inter day / Between PA batches Stability • The following stability exercise should be conducted during method validation: • Standard stock solution • Bench top • Autosampler • Freeze them (I, II& III cycle) • Long term stability (Storage condition)
  • 119. Recovery • This is carried out to ascertain maximum and consistent extraction of drug from biological fluid. Ruggedness • It is carried out by analysing one PA batch processed using different sets of reagents, different columns, different analyst and run on different instrument.
  • 121. As a part of GLP all the exercise carried out during method validation are well documented by competing specific pre-designed form, log books and note books. The complete method validation should be audited by an independent quality assurance person for • Conduction of activities as per related SOPs • Pre recording and documentation of data • Data Analysis • Accurate reporting GLP Requirements
  • 122. Reference s•Hill, H.M. Analysis of drugs in biological fluid, in : Townshed, A (ed.) Encyclopedia of Analytical sciences, Volume 6, London, Academic Press, 1995 •Smyth, M.R.(ed.0 Chemical Analysis in Complex Matrices, New York, Ellis Harwood Limited, 1992 •Snyder, L.R. Kirkland, J.J. and Glajch, Practical HPLC Method Development, 2nd edition, New York, John Wiley and Sons, Inc., 1997 •Venn, R.F.(ed.) Principle and Practice of bioanalysis, London, Taylor and Francis
  • 123. 123 References 1.Heftmann, Chromatography Part A: Fundamentals and Techniques, 6th edition, published by Elsevier, 2004. 2.Skoog, Holler, Nieman, Principles of Instrumental Analysis, 5th edition, published by Thomson Brook/Cole, 2005. 3.Mendham, Denny, Barnes, Thomas, Vogel’s text book of Quantitative Chemical Analysis, 6th edition, published by Person Education, 2007.