Characterisation of neutron meriaty
1. Australian Government
CHARACTERISATION OF NEUTRONCHARACTERISATION OF NEUTRON
FIELD IN RIG ROOM AT ANSTOFIELD IN RIG ROOM AT ANSTO
INSTRUMENT CALIBRATION FACILITYINSTRUMENT CALIBRATION FACILITY
Haider MeriatyHaider Meriaty
Safety, Environmental & Radiological AssuranceSafety, Environmental & Radiological Assurance
ANSTO, Locked Bag 2001,ANSTO, Locked Bag 2001, KirraweeKirrawee DC, NSW 2232, AustraliaDC, NSW 2232, Australia
ham@ansto.gov.auham@ansto.gov.au
2. INTRODUCTION
The Instrument Calibration Facility at Australian
Nuclear Science & Technology Organisation
(ANSTO) provides calibration to radiation monitors
for radiation protection applications. The Services
cover neutron, gamma, beta and alpha monitors.
The neutron calibration room was characterised
with 241Am/Be source, dose equivalent neutron
monitor (digipig2222). The room’s size category is
large1.
3. OBJECTIVES
To determine the Fractional Room Return
Scatter at the ICF rig calibration room.
To determine the ambient dose equivalent
response of ICF reference neutron monitor.
To evaluate the appropriate calibration
standards for implementation at ICF.
4. METHODOLOGY
Measurement results were assessed by four different
standard methods, which commonly applied at neutron
calibration facilities. The methods1 included, the Shadow
Shield, the Semi-Empirical, the Polynomial and NCRP-112
shadow shield.
The shadow shield was a truncated cone of two stacked
sections: a solid iron and an aluminum shell filled with
24%w/w LiBr aqueous solution. The properties5 of neutron
absorption and scattering were compared with Li2CO3 wax
shield of NCRP-112 and the former showed superior
properties.
Forsythe and Power Least squared methods3 were used to
fit the polynomial functions.
5. CHARACTERISATION TECHNIQUES
1) Shadow Shield (ISO 10647):
M . FA(l) = k / l 2 (Equation 1)
Monitor readings were taken with and without the shield. The reading difference, corrected for air
attenuation ‘FA(l )’ (Table 1), plotted against the inversed distance squared (Plot 1). The Characteristic
Constance ‘k’ obtained as the slope of the linear function, fitted to data.
2) Semi-Empirical (ISO 10647:
MT / [ . Fl(l) . (1 + A . l )] = R (1 + S . l 2) (Equation 2)
The ratio of Monitor readings to neutron fluence, corrected for air scatter ‘A’ and geometry ‘F1(l)’’, were
taken and plotted versus distance squared (plot 3). Data fitted into linear function. The intercept and slope
of the function provided the monitor response and the fractional room return scatter ‘s’ (Table 5).
3) Polynomial Fitting (ISO 10647):
MT / [ . Fl(l)] = R (1 + B . l + C . l 2) (Equation 3)
The ratio of Monitor readings to neutron fluence, corrected for geometry ‘F1(l)’’ (Table 3), were taken and
plotted versus distance (Plot 4). Data fitted into quadratic function3. The intercept of the function provided
the monitor response, R . Also, linear function fitted to data and compared with the quadratic one.
4) Shadow Shield (NCRP 112):
D. l 2 = Do (1 + S. l 2) (Equation 4)
The product of monitor readings by distance squared (Table 4) were plotted versus distance squared as
well as fitted into linear function. The function’s intercept and slope provided the monitor response and
the fractional room return scatter ‘S’ (Plots 5, 6).
6. TOOLS & INSTRUMENT
Dimensions of Truncated Conical shield:
Section Material Iron Al Shell
Diameter of Section’s Top 20 92.7
Diameter of Section’s Base 92.7 220
Height of section (mm) 200 350
Thickness of Shell (mm) - 1
Floor to Source Height 1640mm
Neutron Monitor Model Digipig 2222A
Neutron Emission 2.08x107s-1 @ 9/12/2009
Shadow Shield
Two truncated conical sections. One of
solid Iron and the other of aluminium
shell (filled with aqueous LiBr).
7. SETUP1: SOURCE, SHADOW SHIELD & MONITOR
The conical shadow shield was placed at half way between the source
rig (red circle) and monitor stand (exception was the measurement at
1m). Steps ladder was removed during measurements.
8. At 1m reference distance, the shadow shield was placed closer to
source guide (vertical tube to left). This arrangement allowed
sufficient scattered neutrons to impinge on monitor.
SETUP2: SOURCE, SHADOW SHIELD & MONITOR
9. FLOOR PLAN OF RIG CALIBRATION ROOM
Neutron
Source
Rail & monitor
stand system
concrete
walls
10. Table 1: Results of Shadow Shield Method by ISO10647 standard, Equation 1. “ l ” represents the
effective source monitor distance. LHS values plotted versus “1/l 2 “ in plot 1. The ratio of (MT-Ms) to
MT was plotted versus l in Plot 2. The straight line illustrated a good test performance of the shadow
shield and direct neutron, obeying the law of inverse distance squared.
Distance,
l
(m)
MT
(uSv/h)
MS
(uSv/h)
1/l 2
(m-2)
Air-Attenuation
Correction FA(l)
Geometry
Correction F1(l)
LHS of
Equation 1
1 230.55 31.44 1.00 1.0089 1.0014 200.56
1.5 111.63 26.30 0.44 1.0134 1.0006 86.40
2 68.58 21.60 0.25 1.0180 1.0004 47.81
2.5 49.25 18.97 0.16 1.0225 1.0002 30.95
3 38.03 17.06 0.11 1.0271 1.0002 21.53
3.5 30.72 14.93 0.08 1.0316 1.0001 16.28
4 25.11 13.77 0.06 1.0362 1.0001 11.74
4.5 21.43 12.28 0.05 1.0409 1.0001 9.53
5 18.55 11.11 0.04 1.0455 1.0001 7.77
RESULTS
11. Plot 1: Direct neutron dose rate versus source-monitor inverse distance squared at ICF.
The “y” function represents the linear fitting of data. The straight line and correlation
value (R2) illustrated a good outcome of the performance test of the shield integrity.
12. Plot 2: Fraction of direct neutron “y” versus source-monitor distance at ICF. Data were
fitted into linear and quadratic functions. The latter function is applied for routine client
calibrations.
14. Plot 3: LHS of Equation 2 plotted versus source-monitor distance squared at rig calibration
room. The “y” function represents the linear fitting of data. R2 represents the fitting correlation
factor.
15. Table 3: Results of Polynomial Fitting Method and Equation 3. LHS values plotted versus distance
(l ) and fitted to quadratic function (RHS) of a, b, c parameters. The data fitted also linearly for
comparison (Plot 4). The percentage difference (%) of RHS to LHS results is listed in column 8.
Distance,
l
(m)
l 2
(m2)
Total
Reading
MT
(uSv/h)
Geometry
Factor,
F1(l)
Fluence
Rate
(n/m2/s)
LHS
(uSv/h per
n/m2/s)
RHS=
R(1+b..l +c..l 2)
1.042e-
4(1+0.2966 l
+0.8671 l 2)
% of
RHS to
LHS
1 1 2.306E+02 1.0014 1.657E+06 1.389E-04 1.360E-04 2.09
1.5 2.25 1.116E+02 1.0006 7.364E+05 1.515E-04 1.526E-04 -0.75
2 4 6.858E+01 1.0004 4.142E+05 1.655E-04 1.697E-04 -2.51
2.5 6.25 4.925E+01 1.0002 2.651E+05 1.857E-04 1.872E-04 -0.77
3 9 3.803E+01 1.0002 1.841E+05 2.065E-04 2.051E-04 0.69
3.5 12.25 3.072E+01 1.0001 1.353E+05 2.271E-04 2.235E-04 1.58
4 16 2.511E+01 1.0001 1.036E+05 2.424E-04 2.424E-04 0.04
4.5 20.25 2.143E+01 1.0001 8.182E+04 2.619E-04 2.617E-04 0.10
5 25 1.855E+01 1.0001 6.627E+04 2.799E-04 2.814E-04 -0.55
RESULTS (continue…)
17. Table 4: Results of Shadow Shield Method of NCRP 112 standard, Equation 4. “SD” and “U”
represent the sample standard deviation and the combined uncertainty “A Type” respectively.
The Readings Difference is the Total minus Shadow Readings. The Fraction of Direct Dose Rate
is the ratio of the Difference to Total Readings.
Distance,
l
(mm)
Total
Reading
(uSv/5min)
Shadow
Reading
(uSv/5min)
Readings
Difference
(uSv/5min)
SD of
Difference
(uSv/5min)
Fraction
of Direct
Dose Rate
U of
Fraction
%U of
Fraction
1000 19.213 2.620 16.593 0.162 0.864 0.011 1.3
1500 9.303 2.192 7.111 0.136 0.764 0.018 2.4
2000 5.715 1.800 3.916 0.055 0.685 0.011 1.6
2500 4.104 1.581 2.523 0.065 0.615 0.018 2.9
3000 3.169 1.422 1.747 0.046 0.551 0.016 3.0
3500 2.560 1.244 1.316 0.042 0.514 0.017 3.4
4000 2.092 1.148 0.945 0.063 0.451 0.032 7.2
4500 1.786 1.023 0.763 0.050 0.427 0.029 6.9
5000 1.546 0.926 0.620 0.040 0.401 0.027 6.6
RESULTS (continue…)
18. Plot 5 : Fraction of direct neutron “y” versus source-monitor distance at ICF. Data were fitted
into linear and quadratic functions and plotted for comparison.
19. Plot 6: Fraction of direct neutron versus source-monitor distance at ICF. Data were fitted into
linear and quadratic functions and plotted for comparison.
20. Table 5: Summary of Characterisation Results at ICF and the Applied Methods. The shadow
shield method of ISO 10647 was addpted in this characterisation for the rig calibration room
at ICF.
Standard Method Numerical
Analysis
Monitor
Reading
@ 1m
[Free
field]
Reading
Unit
Dose
Response
[nominal
uSv per
n/m2]
Fractional
Room
Return
Scatter, S
NCRP 112 Shadow Shield Forsythe 240.12 uSv/h 1.449E-04 0.044
ISO 10647 Shadow Shield Forsythe 200.40 uSv/h 1.210E-04 n/a
ISO 10647 Semi-Empirical
Regression
(Excel 97)
245.24 uSv/n/m2 1.480E-04 0.041
ISO 10647 Semi-Empirical Forsythe 238.44 uSv/n/m2 1.439E-04 -
ISO 10647
Polynomial
(2nd degree)
Forsythe
(9 points)
172.66 uSv/n/m2 1.042E-04 n/a
ISO 10647
Polynomial
(2nd degree)
Power Matrix
(5 points)
178.0 uSv/n/m2 1.073E-04 n/a
RESULTS (continue…)
21. DISCUSSION
The evaluation of LiBr and Li2CO3 properties for neutron demonstrated the advantage of LiBr solution as
shadow shield materials against Li2CO3 waxed cone i.e. 7% higher absorption and 31% less scattering of
neutrons. In addition, the solution did provide a uniformity of absorber distribution at ionic level as well as
simpler and straightforward preparation. The shadow shield assembly passed the performance test well, in
which the direct neutron fluence followed the law of inverse distance squared (plot 1).
The arithmetical mean and sample standard deviation of repeated 5 measurements were used to represent
the average reading and associated statistical dispersion. The uncertainty propagation technique was
applied, to obtain the combine uncertainty (A type) of a given measurement technique’s function4.
The fitted & plotted agreement with the hypothesis of the measurement technique was considered as an
illustration of good accuracy with measurement4. The uncertainty scope given in ISO 10647 is still
applicable to this work and considered as Type B uncertainty.
The values of Characteristic Constant ‘k’, the fraction of free field neutrons and ‘S’ obtained, by the shadow
shield methods, were implemented at ICF rig calibration room i.e. medium to large size of calibration
room1.
The ‘in situ’ calibration of ICF reference monitor was compared and agreed with the monitor calibration
that completed recently by PTB standard laboratory, within 5.6% accuracy.
At 1m reference distance, the shadow shield was placed closer to source i.e. 5cm (photo in setup2). This
arrangement was necessary to allow sufficient air gap between the cone face and monitor, so optimising
the linear relationship1, 9 between air inscatter and net scatters (i.e. inscatter minus outscatter). The ratio of
monitor-air gap to cone length was 73% of the recommended value (for uncertainty < 3%). However, the
shield was placed half way between source and monitor (photo in setup1) in other reference distances,
22. CONCLUSIONS
The Fractional Room Return Scatter (S) at ICF rig calibration room was determined. Its
small value illustrated a good quality feature of the room e.g. minimal scatter (Table 5).
The measured S Values by NCRP112 or ISO10647 techniques (equations 2 & 4) were in
good agreement (93%).
Routine calibrations should be carried out within 3m distance from neutron source in order to
comply with 40% room scatter limit, recommended by ISO10647 standard.
The new type of aqueous LiBr truncated cone was successfully implemented in
measurements of the shadow shield technique.
The fluence and dose rate equivalent responses were determined (Table 5) for ICF
reference monitor (digipig2222A type) and 241Am/Be neutron source. The reference neutron
fluence was taken from the source certificate6 and is now linked to the PTB standard
authority for traceability.
The ‘in situ’ calibration of ICF reference monitor compared and agreed with the monitor
calibration that completed recently by PTB standard laboratory, within 5.6% accuracy
Two extra positions of higher dose rate were determined (extrapolation). They expand the
capability of dose rate range at ICF calibration service that desired by some clients.
The characterisation parameters measured by the different techniques were evaluated and
the Shadow Method1 of ISO10647 was identified as the appropriate one for implementation
at ICF rig calibration room.
Two neutron sources of appropriate activities are recommended, So the full dose rates
range given in ISO10647 can be achieved within the distance limitation for air scatter .
23. REFERENCES
1) ISO 10647, 1996E; ISO 8522:1989E; ISO 8529-3:1998E.
2) NCRP 112 Report, 1991.
3) Southworth, R.W, et al; Digital Computation & Numerical Methods;
1965.
4) Caria, M; Measurement Analysis, 2000.
5) Curtiss, L. F.; Introduction to Neutron Physics, 1959
6) 241Am/Be Certificate of Measurement; NPL Reference N113, 1982.
7) Hunt, J B; The Calibration of Neutron Sensitive Spherical devices,
Radiation Protection & Dosimetry, 1984.
8) Eisenhauer, C M; Review of Scattering Corrections for Calibration of
Neutron Instruments; Radiation Protection & Dosimetry, , 1989.
9) Burger, G, et al.; Guidelines on Calibration of Neutron Measuring
Devices; IAEA; 1988