SlideShare a Scribd company logo
Analytical Balance Calibration 
Learn how to calibrate the Analytical Balance as Drift check, Performance check and 
Measurement uncertainty check. 
CALIBRATION OF TOP PAN BALANCE 
Calibration procedure: 
Internal calibration 
It shall be performed as per the manufacturer’s instructions. 
Drift check 
For the calculation of the drift 10 mg weight shall be used. 
Note down the 10 measurements of the 10 mg weight in the performance check log. ( 
Attachment no. 01) 
Variation in the observed weight from the mean value does not exceeds ± 0.2 mg. 
The 10 mg weight shall meet the performance check criteria of the mass value i.e. 0.1% of 
actual mass value. 
E.g. For all the 10 measurements of 10 mg weight, variation in the weighings cannot exceed 
0.01mg. 
Performnce check 
After the Auto calibration put 1 mg, 2 mg, 5 mg, 10 mg and 20 mg weights individually. The 
measurement shall be within the 0.1% of actual mass value of the individual weight as given in 
the performance check log.( Attachment no. 02) 
Measurement uncertainty check 
The measurement of uncertainty shall be carried out by using 10 mg weight. 
Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03) 
Calculate the measurement of uncertainty as follows. 
Measurement uncertainty = Standard Deviation X 3 
Actual Mass Value 
Measurement of uncertainty shall be Not More Than 0.001. 
Calibration frequency: It shall be performed daily, after any maintenance, after the relocation 
and after the power failure of the balance. 
Each manufacturer supplies the maximum loading condition of the balance. The calibration of 
the balance shall be designed in such a way that the performance check weights cover the entire 
loading range of the balance. Lower and higher load limits shall be checked for the performance 
check. Before calibration make sure that the level bubble is in the center of the indicator and if
internal calibration facility is available in the balance then it shall also be performed daily before 
performance check of the balance. 
Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight 
might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the 
maximum capacity of the balance to select the correct check weight. 
Attachment-1: Performance Check Log of Microbalance 
Instrument:_____________________ Code No :______________ 
Make :_____________________ Model :_______________ 
Date 
Level indicator 
check 
1 mg 2 mg 5 mg 10 mg 20 mg 
Weight no. and upto 
Std. Wt. Cal No. Þ 
Limit 
Þ 
Bubble should 
be in Center 
+ 0.1% of 
Std. Weight 
( to ) 
+ 0.1% of 
Std. Weight 
( to ) 
+ 0.1% of 
Std. Weight 
( to ) 
+ 0.1% of 
Std. Weight 
( to ) 
+ 0.1% of 
Std. Weight 
( to ) 
Attachment-2: Auto Performance Check Log of Microbalance 
Instrument:_____________________ Code No 
:______________ 
Make :_____________________ Model :___________ 
____ 
Check level indicator (Bubble should be in center) : _________________________________ 
(Paste Performance check chart here) 
Calibrated by : _______________ Date : ____________________ 
Checked by : _______________ Date : ____________________
Attachment-3: Measurement Uncertainity Check of Microbalance 
Instrument:_____________________ Code No :______________ 
Make :_____________________ Model :_______________ 
Date Std. 
Wt. 
Std. 
Wt. 
Cal. 
No. 
Observation Observation of 
Uncertainty Less 
than or equal to 
0.001 
Weight 
box no. 
and 
valid 
upto 
Done 
by 
Checked by Remark 
10 
mg 
Calculation 
Measurement uncertainty = Standard Deviation X 3 
Actual Mass Value 
Drift and Its Importance in Analytical Balances 
Know the causes, importance and factors effecting the drift in analytical balances. 
In the pharmaceutical industry and bioscience research field, many laboratories make 
use of analytical balances. The analytical balances used in the bioscience research and 
pharmaceutical industry are very sensitive. These balances can be heavily affected by the 
way the measuring personnel handle them and by the environment in which they are 
installed. The environment in the pharmaceutical laboratory needs to be assessed by 
running assessment tests. Based on the results of these assessments should be 
proposed concrete measures for improving the lab environment. 
The weighing instruments in a pharmaceutical lab should always operate is quick and 
accurate. It should be avoided any events in the lab that would affect the analytical 
balance measurements results, such as opening and closing a breeze break door, for 
example. This is not an optimal way of conducting analytical balance weight 
measurements. When the breeze break door is open, the weighing area’s temperature 
will change because the air within the breeze break changes. 
Analytical balances used in a pharmaceutical lab are used for extremely precise 
measurements, such as the microbalance capable of measuring 1 millionth of a gram, 
for instance. These highly accurate weighing instruments are used for quality control of 
the processes involved in the pharmaceutical industry manufacturing production.
Related: Calculation for Determination of Weighing Range of Balances 
Sometimes a phenomenon called “drift” is experienced in the weighing instruments, 
including high precision analytical balances. This undesired phenomenon consists in 
measurements changing in one direction or displays becoming unstable with the 
passage of time. Unstable readings may occur with no weight applied or the weight 
readings do not stabilize due to the drift effect. This can be explained by the static 
electricity accumulated due to the dry environment and friction from insulated material. 
The pharmaceutical production lines are having clean environments, controlled with 24 
hour air conditioning. In such an area the humidity levels is often below 20 percent. By 
moving around objects in such a dry environment the friction causes building up of 
static electricity. Research has proved that people working in such a research lab or 
pharmaceutical production line can build up themselves around 10,000 volts of 
electricity. The effects of the static electricity become even greater under these 
circumstances and they can lead to errors of dozen milligrams. 
In order to avoid the static electricity build up and the apparition of the drift effect on 
the analytical balances, it would be necessary that the humidity level in the weighing 
instrument's installation environment to be increased over 40 percent. In case that the 
static electrical build up occurs faster than the electrical discharge, it is necessary to 
conduct weighing operations only after removing electrical charges from the weighing 
sample and to introduce a static eliminator. It should be avoided the use of plastic 
containers for the weighed items and the operators of the analytical balances should 
always stay on an anti-static floor covering. 
Another environmental factor that can dramatically affect analytical balances stability is 
temperature. Temperature control is therefore imperative in avoiding the drift 
phenomenon. This includes both the maintaining the constant temperature of the 
weighing instrument as well as controlling the room temperature. For the best 
temperature stability, the room should be constantly maintained whiting two 
temperature degrees variation, day and night. The weighing instrument should be 
always plugged in and turned on, in order to maintain its temperature constant as well. 
Generally balances have an error of 2 ppm/°C. 
Calculation for Determination of Weighing Range of Balances 
know how to determine the operating range of the balances used for pharmaceuticals including 
the analytical balances.
Balances are used to weigh different things as raw material, dispensed material and 
samples. Sometimes very low quantity is weighed for analysis. An error in the weighed 
material can cause a big variation in analytical results. 
Following circumstances may cause the error in the weighing: 
 Balance for analytical use must be closed; an open balance may cause 
inaccurate weighing. 
 Hygroscopic material gains weight when it comes in contact of the air 
causing unstable weighing. 
 Air flow or current in the balance room may cause disturbance in accurate 
weighing. 
 Unleveled surface or balance may lead to inaccurate weighing. 
 Vibrating surface can also cause inaccurate weighing. 
Assay, related substances and other quantitative analytical tests require accurate 
weighing for better results. Near the lower and upper capacity of the balance, weighing 
may be inaccurate. Therefore, a weighing range must be defined for all balances. 
Related: Calibration of Analytical Balance 
Weighing Range for Balances: 
Some manufacturers direct in their operating manual to weigh minimum 1.0 mg on 
balance having 0.01 mg least count i.e. least count X 100. But the thumb rule followed in 
pharmaceutical industries for lower limit is least count X 50 and upper limit is 80% of 
capacity of the balance. If any balance has operating range from 0.01 mg to 110 gm 
then it should be used for weighing the material from 0.01X50 = 0.50 mg to 110X80% = 
88.00 gm. Detailed guidance on the balances can be learned through OIMLR76-1 Non-automatic 
Weighing Instruments. 
But according to USP <41> Weights and Balances, a simple theoretical calculation can 
be used to calculate the minimum weight for any balance. To weigh 10 mg sample you 
have to use 10X0.1% = 0.01 mg least count balance. It shows that if you have a balance 
with least count 0.01 mg, you can weigh minimum 10 mg sample on it. Other weights 
can be calculated accordingly.

More Related Content

What's hot

Analytical balance
Analytical balanceAnalytical balance
Analytical balance
Chandra Prakash Singh
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & usp
GANESH NIGADE
 
IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS
prakash64742
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
ArpitSuralkar
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
Solairajan A
 
SITE ACCEPTANCE TEST
SITE ACCEPTANCE TESTSITE ACCEPTANCE TEST
Analysis of raw materials
Analysis of raw materialsAnalysis of raw materials
Analysis of raw materials
KULDEEPKUMAR498
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
Sanathoiba Singha
 
Manufacturing operation and control
Manufacturing operation and controlManufacturing operation and control
Manufacturing operation and control
Malay Pandya
 
IMPURITIES AND STABILITY STUDIES
IMPURITIES AND STABILITY STUDIESIMPURITIES AND STABILITY STUDIES
IMPURITIES AND STABILITY STUDIES
prakash64742
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
PUNEET NIRMAL
 
Impurities in residual solvents
Impurities in residual solventsImpurities in residual solvents
Impurities in residual solvents
MehulJain143
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method Validation
Naila Kanwal
 
Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation' Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation'
Sanathoiba Singha
 
Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETER
Dr.K.Venkateswara raju
 
Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR
KUNDLAJAYALAKSHMI
 
Validation of equipments
Validation of equipmentsValidation of equipments
Validation of equipments
suhasini
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationGaurav Kr
 
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
begalavenu
 
Microbiological assay of vitamins B series
 Microbiological assay of vitamins B series Microbiological assay of vitamins B series
Microbiological assay of vitamins B series
PHARMA WORLD
 

What's hot (20)

Analytical balance
Analytical balanceAnalytical balance
Analytical balance
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & usp
 
IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
SITE ACCEPTANCE TEST
SITE ACCEPTANCE TESTSITE ACCEPTANCE TEST
SITE ACCEPTANCE TEST
 
Analysis of raw materials
Analysis of raw materialsAnalysis of raw materials
Analysis of raw materials
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Manufacturing operation and control
Manufacturing operation and controlManufacturing operation and control
Manufacturing operation and control
 
IMPURITIES AND STABILITY STUDIES
IMPURITIES AND STABILITY STUDIESIMPURITIES AND STABILITY STUDIES
IMPURITIES AND STABILITY STUDIES
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Impurities in residual solvents
Impurities in residual solventsImpurities in residual solvents
Impurities in residual solvents
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method Validation
 
Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation' Notes for the subject 'Pharmaceutical Validation'
Notes for the subject 'Pharmaceutical Validation'
 
Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETER
 
Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR
 
Validation of equipments
Validation of equipmentsValidation of equipments
Validation of equipments
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
 
Microbiological assay of vitamins B series
 Microbiological assay of vitamins B series Microbiological assay of vitamins B series
Microbiological assay of vitamins B series
 

Similar to Calibration of analytical balance

CALIBRATIONS.pptx
CALIBRATIONS.pptxCALIBRATIONS.pptx
CALIBRATIONS.pptx
SanjayKumarUchibagle
 
SANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxSANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptx
sanjayudps2016
 
Qualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniQualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri Soni
Mayuri Soni
 
Techniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptxTechniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptx
AREEBA SHAFIQ
 
Calibrating the Measuring Chain
Calibrating the Measuring ChainCalibrating the Measuring Chain
Calibrating the Measuring Chain
Tacuna Systems
 
Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)
Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)
Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)
Muhammad Shahzad
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
SohailPattan
 
Qualification of instruments
Qualification of instrumentsQualification of instruments
Qualification of instruments
Navdha Soni
 
How to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization EquipmentHow to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization Equipment
Setra Systems
 
Balances
BalancesBalances
Pipette Calibration Worksheet & Guidelines -M. Asif
Pipette Calibration Worksheet & Guidelines -M. AsifPipette Calibration Worksheet & Guidelines -M. Asif
Pipette Calibration Worksheet & Guidelines -M. AsifRana Muhammad Asif
 
Errors in pharmaceutical analysis
Errors  in  pharmaceutical  analysis Errors  in  pharmaceutical  analysis
Errors in pharmaceutical analysis
Bindu Kshtriya
 
Introduction to analysis- Pharmaceutical Analysis
Introduction to analysis- Pharmaceutical AnalysisIntroduction to analysis- Pharmaceutical Analysis
Introduction to analysis- Pharmaceutical Analysis
Sanchit Dhankhar
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
MamunReja1
 
21 hand out on waste quantification -samantha
21  hand out on waste quantification -samantha21  hand out on waste quantification -samantha
21 hand out on waste quantification -samantha
zubeditufail
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Malarvili14
 
Instrumentation & Control Lab Manual.pdf
Instrumentation & Control Lab Manual.pdfInstrumentation & Control Lab Manual.pdf
Instrumentation & Control Lab Manual.pdf
NTU Faisalabad
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stroundhydrologywebsite1
 

Similar to Calibration of analytical balance (20)

CALIBRATIONS.pptx
CALIBRATIONS.pptxCALIBRATIONS.pptx
CALIBRATIONS.pptx
 
SANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptxSANJAY CALIBRATIONS.pptx
SANJAY CALIBRATIONS.pptx
 
Qualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniQualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri Soni
 
Techniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptxTechniques Protocols for Instrumental Analysis final.pptx
Techniques Protocols for Instrumental Analysis final.pptx
 
Calibrating the Measuring Chain
Calibrating the Measuring ChainCalibrating the Measuring Chain
Calibrating the Measuring Chain
 
Calibration Project
Calibration ProjectCalibration Project
Calibration Project
 
Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)
Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)
Validation & calibration in pharma (M shahzad (PAKISTAN) 03325637049)
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
Qualification of instruments
Qualification of instrumentsQualification of instruments
Qualification of instruments
 
How to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization EquipmentHow to Select a Pressure Transducer for Medical Sterilization Equipment
How to Select a Pressure Transducer for Medical Sterilization Equipment
 
Balances
BalancesBalances
Balances
 
Pipette Calibration Worksheet & Guidelines -M. Asif
Pipette Calibration Worksheet & Guidelines -M. AsifPipette Calibration Worksheet & Guidelines -M. Asif
Pipette Calibration Worksheet & Guidelines -M. Asif
 
Errors in pharmaceutical analysis
Errors  in  pharmaceutical  analysis Errors  in  pharmaceutical  analysis
Errors in pharmaceutical analysis
 
Introduction to analysis- Pharmaceutical Analysis
Introduction to analysis- Pharmaceutical AnalysisIntroduction to analysis- Pharmaceutical Analysis
Introduction to analysis- Pharmaceutical Analysis
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
 
21 hand out on waste quantification -samantha
21  hand out on waste quantification -samantha21  hand out on waste quantification -samantha
21 hand out on waste quantification -samantha
 
Presentation for APPON
Presentation for APPONPresentation for APPON
Presentation for APPON
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
 
Instrumentation & Control Lab Manual.pdf
Instrumentation & Control Lab Manual.pdfInstrumentation & Control Lab Manual.pdf
Instrumentation & Control Lab Manual.pdf
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 

Recently uploaded

263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,
sisternakatoto
 
24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all
DrSathishMS1
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
Anujkumaranit
 
Cervical & Brachial Plexus By Dr. RIG.pptx
Cervical & Brachial Plexus By Dr. RIG.pptxCervical & Brachial Plexus By Dr. RIG.pptx
Cervical & Brachial Plexus By Dr. RIG.pptx
Dr. Rabia Inam Gandapore
 
Superficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptxSuperficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptx
Dr. Rabia Inam Gandapore
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
Little Cross Family Clinic
 
Physiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdfPhysiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdf
MedicoseAcademics
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
SumeraAhmad5
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
Krishan Murari
 
Ophthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE examOphthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE exam
KafrELShiekh University
 
Non-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdfNon-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdf
MedicoseAcademics
 
New Drug Discovery and Development .....
New Drug Discovery and Development .....New Drug Discovery and Development .....
New Drug Discovery and Development .....
NEHA GUPTA
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
Levi Shapiro
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
pal078100
 
Knee anatomy and clinical tests 2024.pdf
Knee anatomy and clinical tests 2024.pdfKnee anatomy and clinical tests 2024.pdf
Knee anatomy and clinical tests 2024.pdf
vimalpl1234
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Dr KHALID B.M
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
LanceCatedral
 
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Savita Shen $i11
 
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #GirlsFor Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
Savita Shen $i11
 
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
Swetaba Besh
 

Recently uploaded (20)

263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,
 
24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
 
Cervical & Brachial Plexus By Dr. RIG.pptx
Cervical & Brachial Plexus By Dr. RIG.pptxCervical & Brachial Plexus By Dr. RIG.pptx
Cervical & Brachial Plexus By Dr. RIG.pptx
 
Superficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptxSuperficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptx
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
 
Physiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdfPhysiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdf
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
 
Ophthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE examOphthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE exam
 
Non-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdfNon-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdf
 
New Drug Discovery and Development .....
New Drug Discovery and Development .....New Drug Discovery and Development .....
New Drug Discovery and Development .....
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
 
Knee anatomy and clinical tests 2024.pdf
Knee anatomy and clinical tests 2024.pdfKnee anatomy and clinical tests 2024.pdf
Knee anatomy and clinical tests 2024.pdf
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
 
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
 
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #GirlsFor Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
 
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
 

Calibration of analytical balance

  • 1. Analytical Balance Calibration Learn how to calibrate the Analytical Balance as Drift check, Performance check and Measurement uncertainty check. CALIBRATION OF TOP PAN BALANCE Calibration procedure: Internal calibration It shall be performed as per the manufacturer’s instructions. Drift check For the calculation of the drift 10 mg weight shall be used. Note down the 10 measurements of the 10 mg weight in the performance check log. ( Attachment no. 01) Variation in the observed weight from the mean value does not exceeds ± 0.2 mg. The 10 mg weight shall meet the performance check criteria of the mass value i.e. 0.1% of actual mass value. E.g. For all the 10 measurements of 10 mg weight, variation in the weighings cannot exceed 0.01mg. Performnce check After the Auto calibration put 1 mg, 2 mg, 5 mg, 10 mg and 20 mg weights individually. The measurement shall be within the 0.1% of actual mass value of the individual weight as given in the performance check log.( Attachment no. 02) Measurement uncertainty check The measurement of uncertainty shall be carried out by using 10 mg weight. Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03) Calculate the measurement of uncertainty as follows. Measurement uncertainty = Standard Deviation X 3 Actual Mass Value Measurement of uncertainty shall be Not More Than 0.001. Calibration frequency: It shall be performed daily, after any maintenance, after the relocation and after the power failure of the balance. Each manufacturer supplies the maximum loading condition of the balance. The calibration of the balance shall be designed in such a way that the performance check weights cover the entire loading range of the balance. Lower and higher load limits shall be checked for the performance check. Before calibration make sure that the level bubble is in the center of the indicator and if
  • 2. internal calibration facility is available in the balance then it shall also be performed daily before performance check of the balance. Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight. Attachment-1: Performance Check Log of Microbalance Instrument:_____________________ Code No :______________ Make :_____________________ Model :_______________ Date Level indicator check 1 mg 2 mg 5 mg 10 mg 20 mg Weight no. and upto Std. Wt. Cal No. Þ Limit Þ Bubble should be in Center + 0.1% of Std. Weight ( to ) + 0.1% of Std. Weight ( to ) + 0.1% of Std. Weight ( to ) + 0.1% of Std. Weight ( to ) + 0.1% of Std. Weight ( to ) Attachment-2: Auto Performance Check Log of Microbalance Instrument:_____________________ Code No :______________ Make :_____________________ Model :___________ ____ Check level indicator (Bubble should be in center) : _________________________________ (Paste Performance check chart here) Calibrated by : _______________ Date : ____________________ Checked by : _______________ Date : ____________________
  • 3. Attachment-3: Measurement Uncertainity Check of Microbalance Instrument:_____________________ Code No :______________ Make :_____________________ Model :_______________ Date Std. Wt. Std. Wt. Cal. No. Observation Observation of Uncertainty Less than or equal to 0.001 Weight box no. and valid upto Done by Checked by Remark 10 mg Calculation Measurement uncertainty = Standard Deviation X 3 Actual Mass Value Drift and Its Importance in Analytical Balances Know the causes, importance and factors effecting the drift in analytical balances. In the pharmaceutical industry and bioscience research field, many laboratories make use of analytical balances. The analytical balances used in the bioscience research and pharmaceutical industry are very sensitive. These balances can be heavily affected by the way the measuring personnel handle them and by the environment in which they are installed. The environment in the pharmaceutical laboratory needs to be assessed by running assessment tests. Based on the results of these assessments should be proposed concrete measures for improving the lab environment. The weighing instruments in a pharmaceutical lab should always operate is quick and accurate. It should be avoided any events in the lab that would affect the analytical balance measurements results, such as opening and closing a breeze break door, for example. This is not an optimal way of conducting analytical balance weight measurements. When the breeze break door is open, the weighing area’s temperature will change because the air within the breeze break changes. Analytical balances used in a pharmaceutical lab are used for extremely precise measurements, such as the microbalance capable of measuring 1 millionth of a gram, for instance. These highly accurate weighing instruments are used for quality control of the processes involved in the pharmaceutical industry manufacturing production.
  • 4. Related: Calculation for Determination of Weighing Range of Balances Sometimes a phenomenon called “drift” is experienced in the weighing instruments, including high precision analytical balances. This undesired phenomenon consists in measurements changing in one direction or displays becoming unstable with the passage of time. Unstable readings may occur with no weight applied or the weight readings do not stabilize due to the drift effect. This can be explained by the static electricity accumulated due to the dry environment and friction from insulated material. The pharmaceutical production lines are having clean environments, controlled with 24 hour air conditioning. In such an area the humidity levels is often below 20 percent. By moving around objects in such a dry environment the friction causes building up of static electricity. Research has proved that people working in such a research lab or pharmaceutical production line can build up themselves around 10,000 volts of electricity. The effects of the static electricity become even greater under these circumstances and they can lead to errors of dozen milligrams. In order to avoid the static electricity build up and the apparition of the drift effect on the analytical balances, it would be necessary that the humidity level in the weighing instrument's installation environment to be increased over 40 percent. In case that the static electrical build up occurs faster than the electrical discharge, it is necessary to conduct weighing operations only after removing electrical charges from the weighing sample and to introduce a static eliminator. It should be avoided the use of plastic containers for the weighed items and the operators of the analytical balances should always stay on an anti-static floor covering. Another environmental factor that can dramatically affect analytical balances stability is temperature. Temperature control is therefore imperative in avoiding the drift phenomenon. This includes both the maintaining the constant temperature of the weighing instrument as well as controlling the room temperature. For the best temperature stability, the room should be constantly maintained whiting two temperature degrees variation, day and night. The weighing instrument should be always plugged in and turned on, in order to maintain its temperature constant as well. Generally balances have an error of 2 ppm/°C. Calculation for Determination of Weighing Range of Balances know how to determine the operating range of the balances used for pharmaceuticals including the analytical balances.
  • 5. Balances are used to weigh different things as raw material, dispensed material and samples. Sometimes very low quantity is weighed for analysis. An error in the weighed material can cause a big variation in analytical results. Following circumstances may cause the error in the weighing:  Balance for analytical use must be closed; an open balance may cause inaccurate weighing.  Hygroscopic material gains weight when it comes in contact of the air causing unstable weighing.  Air flow or current in the balance room may cause disturbance in accurate weighing.  Unleveled surface or balance may lead to inaccurate weighing.  Vibrating surface can also cause inaccurate weighing. Assay, related substances and other quantitative analytical tests require accurate weighing for better results. Near the lower and upper capacity of the balance, weighing may be inaccurate. Therefore, a weighing range must be defined for all balances. Related: Calibration of Analytical Balance Weighing Range for Balances: Some manufacturers direct in their operating manual to weigh minimum 1.0 mg on balance having 0.01 mg least count i.e. least count X 100. But the thumb rule followed in pharmaceutical industries for lower limit is least count X 50 and upper limit is 80% of capacity of the balance. If any balance has operating range from 0.01 mg to 110 gm then it should be used for weighing the material from 0.01X50 = 0.50 mg to 110X80% = 88.00 gm. Detailed guidance on the balances can be learned through OIMLR76-1 Non-automatic Weighing Instruments. But according to USP <41> Weights and Balances, a simple theoretical calculation can be used to calculate the minimum weight for any balance. To weigh 10 mg sample you have to use 10X0.1% = 0.01 mg least count balance. It shows that if you have a balance with least count 0.01 mg, you can weigh minimum 10 mg sample on it. Other weights can be calculated accordingly.