SlideShare a Scribd company logo
Techniques Protocols for
Instrumental Analysis
PRESENTED BY:AREEBA SHAFIQ
M.PHILL PHARMACEUTICS SCHOLAR (ISLAMIA UNIVERSITY
BAHAWALPUR,PAKISTAN.
Contents
HPLC
UV-SPECTROPHOTOMETER
PH METER
DIGITAL STIRRER
CENTRIFUGATION Machine
Viscometer
Weighing Balance
Definition of Calibration
Calibration is the comparison of measurement values delivered by a device under
test with those of a calibration standard of known accuracy.
Purpose of Equipment Calibration
• To determine the Precision and Accuracy of the parameters of an equipment in
relation to its use.
• To determine the Uncertainty of the Parameters within a stated confidence level
(usually 95%).
• To maintain traceability of the results provided by the equipment.
Parameters of an Instrument to be
Calibrated
 Any Parameter of an instrument that will affect the quality of the products/test
results has to be calibrated
 (e.g.,a refrigerated centrifuge :speed,time,temperature)
General Requirments of Equipment
Calibration
 All Equipment must be uniquely identified and Labelled.
 Acceptance criteria should be set for each type of equipment.
 ALL equipment should be calibrated prior to being placed into service,recalibrated on
schedule and after major repair.
 Equipment Calibration method must meet the accuracy and precision requirements of
the equipment.
 External Equipment Calibration shall be performed by Organizations nationally
accredited for the Equipment Calibration.
 In-house Equipment Calibration must be performed by trained and Competent staff
using reference standards traceable to national standards.
 Equipment Calibration records should include the acceptance ranges and results
sufficient to support the accuracy and Precision of the equipment within the
acceptance ranges.
 All equipment shall carry calibration status labels with at least the following
information:
• Equipment Identification
• Date of Calibration due
• Signature of Authorized Staff
Equipment Calibration Label
Calibration Method
 Determine the accuracy of a Parameter of an instrument by comparing with a
reference value.
 Determine the precision of a parameter of an instrument by repeat measurments.
Requirments of Reference equipment
used for Calibration
 The reference equipment must have a certified accuracy traceable to
international/National Standards and at least three times(preferably five times)
better than that required for equipment being calibrated.
 The reference equipment should only be used for calibration purpose and stored
securely.
Intervals for Equipment Calibration
 It depends on the frequency of the equipment being used,its condition and
previous calibration results.
 It should ensure detection of malfunction and continued optimal function of the
equipment.
 It should be determined by each institute for each equipment in use based on the
manufacturer’s instruction or national guidelines.
 It varies from one month to ten years.
Weighing Balance
A weighing balance is an instrument that is used to determine the weight or mass of an object.
Calibration of Balance
Requirments:
 Every three years:full calibration comprising accuracy,precision,off-center loading
and hysteresis check
 Every six months:Precision
 Daily or day of use:scale value check
 Each weighing :zero check
Calibration of Balance
Off-center Loading test
• This test is to determine the effect of eccentric loading of the balance
• Use a nominal mass of approximately 1/3-1/2 of the full range of the balance
• Weigh the mass successively at the center on the front,rear ,left and right position
on the lip of the pan by the symmetrical weighing method.(the ‘lip’of the pan is
the area where the flat surface of the pan starts to curve up)
Calibration of Balance
Calibration of Balance
Precision Check
• Take a minimum of 10 successive readings at near zero weight,half range and full
range of the balance .
(Note :the balance should not be zeroed throughout the successive weighting)
• Calculate the σ for these three series of readings
• Take the largest σ as the σ of the balance
• The σ should not be greater than the 0.0005 gm for 200 gmfour-digit balance.
Accuracy Check( 5~10 Point check)
Weigh a single reference mass of value ‘M’ (1/10th of the full range of the balance )
Calculate the correction of each check point (Ci=M-(mi-zi))
Calculate the limit of Performance of the balance F=3 σ + max.of Ci
Hysteresis test
This test checks the variability of a measured mass when its value is reached from the
possible directions of the scale (i.e direction of increasing load and direction of decreasing
load)
Use a Nominal mass,M,~1/2 of the full range of the balance
Zero the balance, Z1
Hysteresis test
Place the mass on the pan and record the scale reading m1
Add extra weights to the Pan until the balance reads full –range.
Remove extra weights (Do not remove the mass) and record scale reading m2
Remove the mass and record zero z2
Repeat the test three times and express the hysteresis as the average of the
differences between (m1- m2) and (z1 - z2).
Daily Scale check
• Check the balance with a mass close to the maximum capacity
the balance or to the weight to be usually weighed.
• It should be within ±3σ.
• Check the mass every six months and re- state the weight of the
mass if it is out of the range.
Acceptance Criteria
 Off-centre loading: not more than σ.
 Hysteresis : not more thanσ.
 F (3σ + max. of Ci): not higher than the resolution of balance (e.g.,
0.0005 gm for a 200 gm 4-digit balance)
 Daily Scale check: within 3σ.
Operation of Weighing Balance
 Check that all connections of the balance are proper & balance is On.
 Ensure that balance is placed on shockproof bench top .
 If the balance is showing some reading on the display then(ensure that nobody
requires that reading press tare key.
 Balance should show 0.000
 Open the door from one side & keep container or sampling bag as required
&close the door & press tare key.
 Then open the door & put required sample.
 Add sample till you get desired weight on display & close the door .
 After getting a stable display ,open the door & take out the sample with
container or sampling bag
 Close the door & press tare key again. (2)
Weighing balance
Centrifuge
A centrifuge is a device that uses centrifugal force to separate various components
of a fluid. This is achieved by spinning the fluid at high speed within a container,
thereby separating fluids of different densities (e.g. cream from milk) or liquids from
solids.
2. Calibration of Centrifugation Machine
Requirements:
 Speed :
Quarterly by Tachometer
 Timer :
Quarterly by calibrated timer
 Temperature :
Semi-annually by calibrated thermometer
Centrifugation machine
Acceptance Criteria:
 Speed :
±10% of the set or displayed speed
 Timer :
±10% of the set time
 Temperature :
±2oC
Viscometer
Viscosity is measure of fluid’s resistance to flow. It is to drive a spindle ( which is
immersed in the test fluid ) through a calibrated spring. The viscous drag of the fluid
against the spindle is measured by the spring deflection. Spring deflection is
measured with a rotary transducer.
Calibration of Viscometer
Procedure of Brookfield Viscometer Calibration
• Ensure that all the connections of Instrument are Proper.
• Turn the Viscometer on
• Takeout the viscosity Standard Solutions in 500 ml beaker,adjust the temperature
to 25ͦC. Attach the required spindle ,adjust require speed %measure the
viscosity.,note down the constant reading.
• Viscosity standards,speeds,spindles required&tolerance limit are as follows:
Acceptance Criteria
• Viscosity should be within the tolerance limit
• Do not transfer the viscosity solution to original container back after use.
Precautions
Take care to avoid any contamination of the standard solutions.
Avoid air bubble entrapment during transfer of Standard solutions.
Frequency
Once in 6 months
Brookfield Viscometer
Operation of hplc
• Check the general cleaning of instrument.
• Switch on the instrument allow displaying the ‘Menu Screen’.
• Keep the mobile phase bottle after filteration and degas.
• Purge the mobile phase line by pressing the ‘PURGE’ Button provided below to
the display on the instrument.
• Connect the HPLC Column whichever required ,as per flow direction mentioned in
column.
• Set the flow rate .Column oven temperature and initial wavelength on system
display.
References
1. https://jbsr.pharmainfo.in/documents/vol2issue2/2010020204.pdf.
2. https://www.pharmaguideline.com/2010/08/normal-0-false-false-
false_13.html#:~:text=Open%20the%20door%20from%20one,with%20container%
20or%20sampling%20bag.
3. http://www.pharmatips.in/Articles/Pharmaceutical-Equipment/Brookfield-
Viscometer-Operation-
Calibration.aspx#:~:text=It%20is%20to%20drive%20a,measured%20with%20a%20
rotary%20transducer.
4. https://www.slideshare.net/solairajananant/calibration-and-validation-of-
analytical-instruments
 http://www.pharmaguidesline.com/sop-for-operation-and-calibration-of-hplc/
 https://www.omega.com/en-us/resources/ph-
meter#:~:text=pH%20meter%20is%20an%20instrument,scale%20of%200%20to%
2014.
 https://www.pharmaguideline.com/2010/01/sop-for-operation-and-calibration-
of-ph.html

More Related Content

Similar to Techniques Protocols for Instrumental Analysis final.pptx

Good Weighing Practices in QC Lab
Good Weighing Practices in QC LabGood Weighing Practices in QC Lab
Good Weighing Practices in QC Lab
Dr. Amsavel A
 
MSA R&R for training in manufacturing industry
MSA R&R for training in manufacturing industryMSA R&R for training in manufacturing industry
MSA R&R for training in manufacturing industry
abhishek558363
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
Ankit Raiyani
 
Calibration of analytical balance
Calibration of analytical balanceCalibration of analytical balance
Calibration of analytical balance
Subrat Acharya
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
Mostafa Mahmoud
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
Gamal Abdel Hamid
 
Qualification of Friability Test Apparatus.pptx
Qualification of Friability Test Apparatus.pptxQualification of Friability Test Apparatus.pptx
Qualification of Friability Test Apparatus.pptx
GNIPST
 
Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical Instruments
ISF COLLEGE OF PHARMACY MOGA
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Pratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Pratik434909
 
Measurement risk and the impact on your processes
Measurement risk and the impact on your processes  Measurement risk and the impact on your processes
Measurement risk and the impact on your processes
Transcat
 
Uncertainty of measurement d dhingra
Uncertainty of measurement d dhingraUncertainty of measurement d dhingra
Uncertainty of measurement d dhingra
Indian Council of Agricultural Research
 
quality control in clinical laboratory
quality control in clinical laboratory quality control in clinical laboratory
quality control in clinical laboratory
DrmanarEmam
 
In-Tolerance Non-Conformance Investigations Webinar Slides
In-Tolerance Non-Conformance Investigations Webinar SlidesIn-Tolerance Non-Conformance Investigations Webinar Slides
In-Tolerance Non-Conformance Investigations Webinar Slides
Transcat
 
Calibration
CalibrationCalibration
Calibration
rakshamhetre
 
QUALITY control in hematology by Dr yogeeta.pptx
QUALITY  control in hematology by Dr yogeeta.pptxQUALITY  control in hematology by Dr yogeeta.pptx
QUALITY control in hematology by Dr yogeeta.pptx
YogeetaTanty1
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptx
AsifSHAIKH50040
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptx
Asif Shaikh
 
Materi Webinar - Verifikasi Timbangan.pdf
Materi Webinar - Verifikasi Timbangan.pdfMateri Webinar - Verifikasi Timbangan.pdf
Materi Webinar - Verifikasi Timbangan.pdf
ProduksiMF
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
Bholakant raut
 

Similar to Techniques Protocols for Instrumental Analysis final.pptx (20)

Good Weighing Practices in QC Lab
Good Weighing Practices in QC LabGood Weighing Practices in QC Lab
Good Weighing Practices in QC Lab
 
MSA R&R for training in manufacturing industry
MSA R&R for training in manufacturing industryMSA R&R for training in manufacturing industry
MSA R&R for training in manufacturing industry
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
Calibration of analytical balance
Calibration of analytical balanceCalibration of analytical balance
Calibration of analytical balance
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Qualification of Friability Test Apparatus.pptx
Qualification of Friability Test Apparatus.pptxQualification of Friability Test Apparatus.pptx
Qualification of Friability Test Apparatus.pptx
 
Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical Instruments
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
Measurement risk and the impact on your processes
Measurement risk and the impact on your processes  Measurement risk and the impact on your processes
Measurement risk and the impact on your processes
 
Uncertainty of measurement d dhingra
Uncertainty of measurement d dhingraUncertainty of measurement d dhingra
Uncertainty of measurement d dhingra
 
quality control in clinical laboratory
quality control in clinical laboratory quality control in clinical laboratory
quality control in clinical laboratory
 
In-Tolerance Non-Conformance Investigations Webinar Slides
In-Tolerance Non-Conformance Investigations Webinar SlidesIn-Tolerance Non-Conformance Investigations Webinar Slides
In-Tolerance Non-Conformance Investigations Webinar Slides
 
Calibration
CalibrationCalibration
Calibration
 
QUALITY control in hematology by Dr yogeeta.pptx
QUALITY  control in hematology by Dr yogeeta.pptxQUALITY  control in hematology by Dr yogeeta.pptx
QUALITY control in hematology by Dr yogeeta.pptx
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptx
 
Calibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptxCalibration of Instruments Asif.pptx
Calibration of Instruments Asif.pptx
 
Materi Webinar - Verifikasi Timbangan.pdf
Materi Webinar - Verifikasi Timbangan.pdfMateri Webinar - Verifikasi Timbangan.pdf
Materi Webinar - Verifikasi Timbangan.pdf
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
 

Recently uploaded

Efficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in AyurvedaEfficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in Ayurveda
Dr. Jyothirmai Paindla
 
Ear and its clinical correlations By Dr. Rabia Inam Gandapore.pptx
Ear and its clinical correlations By Dr. Rabia Inam Gandapore.pptxEar and its clinical correlations By Dr. Rabia Inam Gandapore.pptx
Ear and its clinical correlations By Dr. Rabia Inam Gandapore.pptx
Dr. Rabia Inam Gandapore
 
Clinic ^%[+27633867063*Abortion Pills For Sale In Tembisa Central
Clinic ^%[+27633867063*Abortion Pills For Sale In Tembisa CentralClinic ^%[+27633867063*Abortion Pills For Sale In Tembisa Central
Clinic ^%[+27633867063*Abortion Pills For Sale In Tembisa Central
19various
 
Outbreak management including quarantine, isolation, contact.pptx
Outbreak management including quarantine, isolation, contact.pptxOutbreak management including quarantine, isolation, contact.pptx
Outbreak management including quarantine, isolation, contact.pptx
Pratik328635
 
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx
Holistified Wellness
 
Medical Quiz ( Online Quiz for API Meet 2024 ).pdf
Medical Quiz ( Online Quiz for API Meet 2024 ).pdfMedical Quiz ( Online Quiz for API Meet 2024 ).pdf
Medical Quiz ( Online Quiz for API Meet 2024 ).pdf
Jim Jacob Roy
 
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdfCHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
rishi2789
 
Acute Gout Care & Urate Lowering Therapy .pdf
Acute Gout Care & Urate Lowering Therapy .pdfAcute Gout Care & Urate Lowering Therapy .pdf
Acute Gout Care & Urate Lowering Therapy .pdf
Jim Jacob Roy
 
Post-Menstrual Smell- When to Suspect Vaginitis.pptx
Post-Menstrual Smell- When to Suspect Vaginitis.pptxPost-Menstrual Smell- When to Suspect Vaginitis.pptx
Post-Menstrual Smell- When to Suspect Vaginitis.pptx
FFragrant
 
Complementary feeding in infant IAP PROTOCOLS
Complementary feeding in infant IAP PROTOCOLSComplementary feeding in infant IAP PROTOCOLS
Complementary feeding in infant IAP PROTOCOLS
chiranthgowda16
 
Adhd Medication Shortage Uk - trinexpharmacy.com
Adhd Medication Shortage Uk - trinexpharmacy.comAdhd Medication Shortage Uk - trinexpharmacy.com
Adhd Medication Shortage Uk - trinexpharmacy.com
reignlana06
 
June 2024 Oncology Cartoons By Dr Kanhu Charan Patro
June 2024 Oncology Cartoons By Dr Kanhu Charan PatroJune 2024 Oncology Cartoons By Dr Kanhu Charan Patro
June 2024 Oncology Cartoons By Dr Kanhu Charan Patro
Kanhu Charan
 
The Nervous and Chemical Regulation of Respiration
The Nervous and Chemical Regulation of RespirationThe Nervous and Chemical Regulation of Respiration
The Nervous and Chemical Regulation of Respiration
MedicoseAcademics
 
K CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấu
K CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấuK CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấu
K CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấu
HongBiThi1
 
How to choose the best dermatologists in Indore.
How to choose the best dermatologists in Indore.How to choose the best dermatologists in Indore.
How to choose the best dermatologists in Indore.
Gokuldas Hospital
 
Top Travel Vaccinations in Manchester
Top Travel Vaccinations in ManchesterTop Travel Vaccinations in Manchester
Top Travel Vaccinations in Manchester
NX Healthcare
 
vonoprazan A novel drug for GERD presentation
vonoprazan A novel drug for GERD presentationvonoprazan A novel drug for GERD presentation
vonoprazan A novel drug for GERD presentation
Dr.pavithra Anandan
 
Physical demands in sports - WCSPT Oslo 2024
Physical demands in sports - WCSPT Oslo 2024Physical demands in sports - WCSPT Oslo 2024
Physical demands in sports - WCSPT Oslo 2024
Torstein Dalen-Lorentsen
 
Osteoporosis - Definition , Evaluation and Management .pdf
Osteoporosis - Definition , Evaluation and Management .pdfOsteoporosis - Definition , Evaluation and Management .pdf
Osteoporosis - Definition , Evaluation and Management .pdf
Jim Jacob Roy
 
Tests for analysis of different pharmaceutical.pptx
Tests for analysis of different pharmaceutical.pptxTests for analysis of different pharmaceutical.pptx
Tests for analysis of different pharmaceutical.pptx
taiba qazi
 

Recently uploaded (20)

Efficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in AyurvedaEfficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in Ayurveda
 
Ear and its clinical correlations By Dr. Rabia Inam Gandapore.pptx
Ear and its clinical correlations By Dr. Rabia Inam Gandapore.pptxEar and its clinical correlations By Dr. Rabia Inam Gandapore.pptx
Ear and its clinical correlations By Dr. Rabia Inam Gandapore.pptx
 
Clinic ^%[+27633867063*Abortion Pills For Sale In Tembisa Central
Clinic ^%[+27633867063*Abortion Pills For Sale In Tembisa CentralClinic ^%[+27633867063*Abortion Pills For Sale In Tembisa Central
Clinic ^%[+27633867063*Abortion Pills For Sale In Tembisa Central
 
Outbreak management including quarantine, isolation, contact.pptx
Outbreak management including quarantine, isolation, contact.pptxOutbreak management including quarantine, isolation, contact.pptx
Outbreak management including quarantine, isolation, contact.pptx
 
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx
8 Surprising Reasons To Meditate 40 Minutes A Day That Can Change Your Life.pptx
 
Medical Quiz ( Online Quiz for API Meet 2024 ).pdf
Medical Quiz ( Online Quiz for API Meet 2024 ).pdfMedical Quiz ( Online Quiz for API Meet 2024 ).pdf
Medical Quiz ( Online Quiz for API Meet 2024 ).pdf
 
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdfCHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
 
Acute Gout Care & Urate Lowering Therapy .pdf
Acute Gout Care & Urate Lowering Therapy .pdfAcute Gout Care & Urate Lowering Therapy .pdf
Acute Gout Care & Urate Lowering Therapy .pdf
 
Post-Menstrual Smell- When to Suspect Vaginitis.pptx
Post-Menstrual Smell- When to Suspect Vaginitis.pptxPost-Menstrual Smell- When to Suspect Vaginitis.pptx
Post-Menstrual Smell- When to Suspect Vaginitis.pptx
 
Complementary feeding in infant IAP PROTOCOLS
Complementary feeding in infant IAP PROTOCOLSComplementary feeding in infant IAP PROTOCOLS
Complementary feeding in infant IAP PROTOCOLS
 
Adhd Medication Shortage Uk - trinexpharmacy.com
Adhd Medication Shortage Uk - trinexpharmacy.comAdhd Medication Shortage Uk - trinexpharmacy.com
Adhd Medication Shortage Uk - trinexpharmacy.com
 
June 2024 Oncology Cartoons By Dr Kanhu Charan Patro
June 2024 Oncology Cartoons By Dr Kanhu Charan PatroJune 2024 Oncology Cartoons By Dr Kanhu Charan Patro
June 2024 Oncology Cartoons By Dr Kanhu Charan Patro
 
The Nervous and Chemical Regulation of Respiration
The Nervous and Chemical Regulation of RespirationThe Nervous and Chemical Regulation of Respiration
The Nervous and Chemical Regulation of Respiration
 
K CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấu
K CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấuK CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấu
K CỔ TỬ CUNG.pdf tự ghi chép, chữ hơi xấu
 
How to choose the best dermatologists in Indore.
How to choose the best dermatologists in Indore.How to choose the best dermatologists in Indore.
How to choose the best dermatologists in Indore.
 
Top Travel Vaccinations in Manchester
Top Travel Vaccinations in ManchesterTop Travel Vaccinations in Manchester
Top Travel Vaccinations in Manchester
 
vonoprazan A novel drug for GERD presentation
vonoprazan A novel drug for GERD presentationvonoprazan A novel drug for GERD presentation
vonoprazan A novel drug for GERD presentation
 
Physical demands in sports - WCSPT Oslo 2024
Physical demands in sports - WCSPT Oslo 2024Physical demands in sports - WCSPT Oslo 2024
Physical demands in sports - WCSPT Oslo 2024
 
Osteoporosis - Definition , Evaluation and Management .pdf
Osteoporosis - Definition , Evaluation and Management .pdfOsteoporosis - Definition , Evaluation and Management .pdf
Osteoporosis - Definition , Evaluation and Management .pdf
 
Tests for analysis of different pharmaceutical.pptx
Tests for analysis of different pharmaceutical.pptxTests for analysis of different pharmaceutical.pptx
Tests for analysis of different pharmaceutical.pptx
 

Techniques Protocols for Instrumental Analysis final.pptx

  • 1. Techniques Protocols for Instrumental Analysis PRESENTED BY:AREEBA SHAFIQ M.PHILL PHARMACEUTICS SCHOLAR (ISLAMIA UNIVERSITY BAHAWALPUR,PAKISTAN.
  • 3. Definition of Calibration Calibration is the comparison of measurement values delivered by a device under test with those of a calibration standard of known accuracy.
  • 4. Purpose of Equipment Calibration • To determine the Precision and Accuracy of the parameters of an equipment in relation to its use. • To determine the Uncertainty of the Parameters within a stated confidence level (usually 95%). • To maintain traceability of the results provided by the equipment.
  • 5. Parameters of an Instrument to be Calibrated  Any Parameter of an instrument that will affect the quality of the products/test results has to be calibrated  (e.g.,a refrigerated centrifuge :speed,time,temperature)
  • 6. General Requirments of Equipment Calibration  All Equipment must be uniquely identified and Labelled.  Acceptance criteria should be set for each type of equipment.  ALL equipment should be calibrated prior to being placed into service,recalibrated on schedule and after major repair.  Equipment Calibration method must meet the accuracy and precision requirements of the equipment.  External Equipment Calibration shall be performed by Organizations nationally accredited for the Equipment Calibration.  In-house Equipment Calibration must be performed by trained and Competent staff using reference standards traceable to national standards.
  • 7.  Equipment Calibration records should include the acceptance ranges and results sufficient to support the accuracy and Precision of the equipment within the acceptance ranges.  All equipment shall carry calibration status labels with at least the following information: • Equipment Identification • Date of Calibration due • Signature of Authorized Staff
  • 9. Calibration Method  Determine the accuracy of a Parameter of an instrument by comparing with a reference value.  Determine the precision of a parameter of an instrument by repeat measurments.
  • 10. Requirments of Reference equipment used for Calibration  The reference equipment must have a certified accuracy traceable to international/National Standards and at least three times(preferably five times) better than that required for equipment being calibrated.  The reference equipment should only be used for calibration purpose and stored securely.
  • 11. Intervals for Equipment Calibration  It depends on the frequency of the equipment being used,its condition and previous calibration results.  It should ensure detection of malfunction and continued optimal function of the equipment.  It should be determined by each institute for each equipment in use based on the manufacturer’s instruction or national guidelines.  It varies from one month to ten years.
  • 12. Weighing Balance A weighing balance is an instrument that is used to determine the weight or mass of an object.
  • 13. Calibration of Balance Requirments:  Every three years:full calibration comprising accuracy,precision,off-center loading and hysteresis check  Every six months:Precision  Daily or day of use:scale value check  Each weighing :zero check
  • 14. Calibration of Balance Off-center Loading test • This test is to determine the effect of eccentric loading of the balance • Use a nominal mass of approximately 1/3-1/2 of the full range of the balance • Weigh the mass successively at the center on the front,rear ,left and right position on the lip of the pan by the symmetrical weighing method.(the ‘lip’of the pan is the area where the flat surface of the pan starts to curve up)
  • 16. Calibration of Balance Precision Check • Take a minimum of 10 successive readings at near zero weight,half range and full range of the balance . (Note :the balance should not be zeroed throughout the successive weighting) • Calculate the σ for these three series of readings • Take the largest σ as the σ of the balance • The σ should not be greater than the 0.0005 gm for 200 gmfour-digit balance.
  • 17. Accuracy Check( 5~10 Point check) Weigh a single reference mass of value ‘M’ (1/10th of the full range of the balance ) Calculate the correction of each check point (Ci=M-(mi-zi)) Calculate the limit of Performance of the balance F=3 σ + max.of Ci
  • 18.
  • 19. Hysteresis test This test checks the variability of a measured mass when its value is reached from the possible directions of the scale (i.e direction of increasing load and direction of decreasing load) Use a Nominal mass,M,~1/2 of the full range of the balance Zero the balance, Z1
  • 20. Hysteresis test Place the mass on the pan and record the scale reading m1 Add extra weights to the Pan until the balance reads full –range. Remove extra weights (Do not remove the mass) and record scale reading m2 Remove the mass and record zero z2 Repeat the test three times and express the hysteresis as the average of the differences between (m1- m2) and (z1 - z2).
  • 21.
  • 22. Daily Scale check • Check the balance with a mass close to the maximum capacity the balance or to the weight to be usually weighed. • It should be within ±3σ. • Check the mass every six months and re- state the weight of the mass if it is out of the range.
  • 23. Acceptance Criteria  Off-centre loading: not more than σ.  Hysteresis : not more thanσ.  F (3σ + max. of Ci): not higher than the resolution of balance (e.g., 0.0005 gm for a 200 gm 4-digit balance)  Daily Scale check: within 3σ.
  • 24. Operation of Weighing Balance  Check that all connections of the balance are proper & balance is On.  Ensure that balance is placed on shockproof bench top .  If the balance is showing some reading on the display then(ensure that nobody requires that reading press tare key.  Balance should show 0.000  Open the door from one side & keep container or sampling bag as required &close the door & press tare key.  Then open the door & put required sample.  Add sample till you get desired weight on display & close the door .  After getting a stable display ,open the door & take out the sample with container or sampling bag  Close the door & press tare key again. (2)
  • 26. Centrifuge A centrifuge is a device that uses centrifugal force to separate various components of a fluid. This is achieved by spinning the fluid at high speed within a container, thereby separating fluids of different densities (e.g. cream from milk) or liquids from solids.
  • 27. 2. Calibration of Centrifugation Machine Requirements:  Speed : Quarterly by Tachometer  Timer : Quarterly by calibrated timer  Temperature : Semi-annually by calibrated thermometer
  • 29. Acceptance Criteria:  Speed : ±10% of the set or displayed speed  Timer : ±10% of the set time  Temperature : ±2oC
  • 30. Viscometer Viscosity is measure of fluid’s resistance to flow. It is to drive a spindle ( which is immersed in the test fluid ) through a calibrated spring. The viscous drag of the fluid against the spindle is measured by the spring deflection. Spring deflection is measured with a rotary transducer.
  • 31. Calibration of Viscometer Procedure of Brookfield Viscometer Calibration • Ensure that all the connections of Instrument are Proper. • Turn the Viscometer on • Takeout the viscosity Standard Solutions in 500 ml beaker,adjust the temperature to 25ͦC. Attach the required spindle ,adjust require speed %measure the viscosity.,note down the constant reading. • Viscosity standards,speeds,spindles required&tolerance limit are as follows:
  • 32.
  • 33. Acceptance Criteria • Viscosity should be within the tolerance limit • Do not transfer the viscosity solution to original container back after use. Precautions Take care to avoid any contamination of the standard solutions. Avoid air bubble entrapment during transfer of Standard solutions. Frequency Once in 6 months
  • 35.
  • 36. Operation of hplc • Check the general cleaning of instrument. • Switch on the instrument allow displaying the ‘Menu Screen’. • Keep the mobile phase bottle after filteration and degas. • Purge the mobile phase line by pressing the ‘PURGE’ Button provided below to the display on the instrument. • Connect the HPLC Column whichever required ,as per flow direction mentioned in column. • Set the flow rate .Column oven temperature and initial wavelength on system display.
  • 37. References 1. https://jbsr.pharmainfo.in/documents/vol2issue2/2010020204.pdf. 2. https://www.pharmaguideline.com/2010/08/normal-0-false-false- false_13.html#:~:text=Open%20the%20door%20from%20one,with%20container% 20or%20sampling%20bag. 3. http://www.pharmatips.in/Articles/Pharmaceutical-Equipment/Brookfield- Viscometer-Operation- Calibration.aspx#:~:text=It%20is%20to%20drive%20a,measured%20with%20a%20 rotary%20transducer. 4. https://www.slideshare.net/solairajananant/calibration-and-validation-of- analytical-instruments