SlideShare a Scribd company logo
1 of 125
Lecture One:
Ultrasound Machine
History
Uses of ultrasonic energy in the 1940s. Left, in gastric ulcers. Right, in arthritis
Ultrasonic
therapy
generator, the
"Medi-Sonar" in
the 1950s.
A British
ultrasonic
apparatus for the
treatment of
Meniere's disease
in the late 1950s
History
Denier's Ultrasonoscopic apparatus
with ultrasound generator, emitter
transducer and oscilloscope. This
can be adapted for both therapeutic
and diagnostic purposes
The first hand-held imaging instrument
was developed by John Wild and John
Reid in the early 1950's
Physics of Ultrasound
Sound
• Mechanical &
Longitudinal wave
• Travels in a straight Line
• Human ear range is 20Hz
- 20kHz
• Cannot travel through
Vacuum
Ultrasound
• Mechanical & Longitudinal
wave
• Exceeding the upper limit
of human hearing
• > 20,000H or 20kHz.
4
Characteristics of Ultrasounds
• Velocity
• Frequency
• Wavelength
• Amplitude
• The Relationship b/w v, f and λ
is:
v = f λ
• Average speed of ultrasound in
body is 1540m/sec
• Imaging range is 2 to 20MHz
5
Power and Intensity
• A sound wave transports Energy through a medium from a source. Energy is
measured in joules (J)
• The Power, P,produce by a source of sound is the rate at which it produces energy.
Power is measured in watts (W) where 1 W = 1 J/s
• The Intensity, I, associated with a sound wave is the power per unit area. Intensity is
measured in W/m2
• The power and intensity associated with a wave increase with the pressure amplitude,
p
Intensity, I  p2
Power, P  p
Wavelength, Frequency and Speed
• Waves are characterised by their
wavelength, frequency and speed
• The Wavelength,  , is the distance
between consecutive peaks or other
similar points on the wave.
• The Frequency, f, is the number of
oscillations per second
• Frequency is measured in Hertz (Hz)
where 1 Hz is one oscillation per second.
Characteristics of Ultrasounds
Velocity
8
Advanatges
• Real-time
• Portable
• No ionizing radiation
• Inexpensive?
• Side-side comparison
• Patients love it!
• Guide procedures
Disadvanatges
• Operator dependent
• Expensive?
• Limited penetration through bone and air
• Imaging deep structures
Ultrasound machine
Basic Ultrasound Physics
Amplitude
oscillations/sec = frequency - expressed in Hertz (Hz)
Bats: Navigating with ultrasound
• Bats make high-pitched chirps which are too high for
humans to hear. This is called ultrasound
• Like normal sound, ultrasound echoes off objects
• The bat hears the echoes and works out what caused
them
• Dolphins also navigate with ultrasound
• Submarines use a similar method called sonar
• We can also use ultrasound to look inside the body…
• An ultrasound element acts like a bat.
• Emit ultrasound and detect echoes
• Map out boundary of object
• Medical Ultrasound 2MHz to 16MHz
• Electric field applied to
piezoelectric crystals located
on transducer surface
• Mechanical vibration of
crystals creates sound waves
Each crystal produces an US
wave
• Summation of all waves
forms the US beam
• Wave reflects as echo that
vibrates transducer
• Vibrations produce electrical
pulses
• Scanner processes and
transforms to image
Ultrasound Production
• Transducer produces ultrasound pulses
• These elements convert electrical energy into a
mechanical ultrasound wave
• Reflected echoes return to the scan head which
converts the ultrasound wave into an electrical signal
Frequency vs. Resolution
• The frequency also affects the QUALITY of the ultrasound image
– The HIGHER the frequency, the BETTER the resolution
• The LOWER the frequency, the LESS the resolution
• A 12 MHz transducer has very good resolution, but cannot
penetrate very deep into the body
• A 3 MHz transducer can penetrate deep into the body, but the
resolution is not as good as the 12 MHz
Image Formation
Electrical signal produces ‘dots’ on the screen
• Brightness of the dots is proportional to the
strength of the returning echoes
• Location of the dots is determined by travel
time. The velocity in tissue is assumed constant
at 1540m/sec
Distance = Velocity
Time
16
17
18
Activity 1
• If an echo is detected from a reflector in human soft tissuein 0.065 ms
. How far is the reflector from the transducer.
• Acoustic impedance (AI) is dependent on the density of
the material in which sound is propagated
- the greater the impedance the denser the material.
• Reflections comes from the interface of different AI’s
• greater  of the AI = more signal reflected
• works both ways (send and receive directions)
Medium 1 Medium 2 Medium 3
Transducer
Interactions of Ultrasound with
Tissue
Interactions of Ultrasound with
Tissue
• Reflection
• Refraction
• Transmission
• Attenuation
Interactions of Ultrasound with Tissue
• Reflection
– The ultrasound reflects off tissue and returns to
the transducer, the amount of reflection depends on
differences in acoustic impedance
– The ultrasound image is formed from reflected echoes
transducer
Reflection
• Reflection (specular reflection) occurs at
tissue boundaries where there is a difference
in the acoustic impedance, Z, of the two
tissues
• When the incident ultrasound wave is
perpendicular to the boundary, a fraction of
it’s energy is reflected (an echo) directly
back towards the source
• The remaining energy is transmitted into the
second tissue and continues in the initial
direction
Z1
Z2
Incident
Reflectio
n (echo)
Transmission
31
Reflection
• The fraction of ultrasound intensity reflected
at an interface is given by the intensity
reflection coefficient, R
• The fraction of ultrasound energy reflected
depends on the difference between the Z
values of the two materials
• R increases rapidly as the difference in Z
increases
Z1
Z2
Ii
Ir
It
R =
Z2 –Z1
Z1 +Z2
i
Ir
I
= ( )
2
Reflection
• The fraction of ultrasound intensity
transmitted at an interface is given by the
intensity transmission coefficient, T
• Ultrasound imaging is only possible when
the wave propagates through materials with
similar acoustic impedances – only a small
fraction of energy is reflected and the rest is
transmitted
Z1
Z2
Ii
Ir
It
T =
It
Ii
= 1 - R
Reflection
TissueInterface R T
Liver – Fat 0.01 0.99
Fat– Muscle 0.02 0.98
Muscle - Bone 0.41 0.59
Muscle -Air 0.99 0.01
At soft tissue – soft tissue
interfaces 1-2% of the
ultrasound intensity is
reflected
At soft tissue – air
interfaces 99% of the
incident intensity is
reflected
• At soft tissue – air or soft tissue – bone interfaces, a large proportion of the
incident intensity is reflected, making anatomy beyond such interfaces
unobservable
• Acoustic coupling gel is used between the face of the ultrasound transducer
and skin to eliminate air pockets
Reflection
• When the wave is not incident perpendicular
to the interface, non-normal incidence, the
reflected angle is equal to the incident angle
(i.e. θr = θi)
• Echoes are directed away from the source of
ultrasound and may be undetected
• The transmitted wave does not continue in the
incident direction (i.e. θt ≠ θi)
• The change in direction is described by
Refraction
Z1
Z2
Incident Reflection
(echo)
Transmission
(refraction)
θi θr
θt
Activity
2
Approximately what fraction of an ultrasound beam is transmitted from an
interface between two media with Z values of 1.65 and 1.55?
41
ACTIVITY 3
Calculate the remaining intensity of a 100 W/m^2 ultrasound pulse that
loses 30 dB while traveling through tissue?
56
57
Activity 4
•
59
Transmission
– Some of the ultrasound waves continue deeper into
the body
– These waves will reflect from deeper tissue
structures
transducer
Interactions of Ultrasound with Tissue
Attenuation
– Defined - the deeper the wave travels in the
body, the weaker it becomes -3 processes:
reflection, absorption, refraction
– Air (lung)> bone > muscle > soft tissue >blood >
water
Interactions of Ultrasound with Tissue
Reflected Echo’s
• Strong Reflections = White dots
Diaphragm, tendons, bone
‘Hyperechoic’
Weaker Reflections =
Grey dots
– Most solid organs,
– thick fluid – ‘isoechoic’
Reflected Echo’s
Reflected Echo’s
• No Reflections = Black dots
– Fluid within a cyst, urine, blood
‘Hypoechoic’ or echofree
What determines how far ultrasound waves can
travel?
• The FREQUENCY of the transducer
– The HIGHER the frequency, the LESS it can
penetrate
– The LOWER the frequency, the DEEPER it can
penetrate
– Attenuation is directly related to frequency
7.5 MHz
3.5 MHz
Thinner crystal produces
smaller sound waves.
The Transducer
Thicker crystal
produces bigger sound
waves.
The Transducer
The LOWER the
frequency the better
the penetration
Bigger, Stronger
The HIGHER frequency
the less the penetration
Smaller, weaker
3.5 MHz 7.5 MHz
The Transducer
Short pulses of sound are sent (transmits) into the body
and then the transducer listens for the returning signals
(receives).
The ultrasound system processes the returning signals
into images that are displayed on the ultrasound monitor
Transmits
Waits
Receives
The strength of the returning echoes also
depends on the differences in the acoustic
impedance between various structures.
Acoustic impedance relates to tissue density.
The greater the difference in density between
two structures, the stronger the returning echo
Examples:
different: aorta and liver
same: kidney and liver
Grayscale Imaging
• Attenuation:
A decrease in the strength of the sound wave as it
passes through tissue and further into the body.
• Acoustic Impedance:
The resistance of the sound wave traveling through
tissue .Each tissue has its own acoustic impedance
due to the density of the tissue.
• Through Transmission:
There is no attenuation of the sound wave traveling
through the tissue.
Grayscale Imaging
WHITE DOTS = STRONG = e.g., bone
BLACK DOTS = NO reflections = e.g., fluid
GRAY (different shades) = WEAKER
reflections
Grayscale Imaging
The strength of the returning echo is
directly related to the angle at which the
ultrasound beam strikes an interface.
Grayscale Imaging
The more perpendicular the
ultrasound beam, the stronger the
returning echo.
Resolution
• Clarity of picture
• The ability to identify structures very close
together./
• Ability of equipment to detect
2 separate reflectors in tissue and to display
them as 2 separate reflectors on the monitor
without merging them.
Image Resolution
Types of Resolution
Axial Ability to identify structures that are one in
front of the other
Lateral Ability to identify structures that are side by
side
Temporal Ability to accurately locate a moving structure
Spatial Ability to display very small structures in their
correct anatomic location.
Next class
• Principles of A-Mode, B-Mode, M-Mode
• Application area of ultrasound
• parts of ultrasound machineineParts of
Ultrasound Machine
Summary of previous class
• SOUND to IMAGE
• producing a sound wave,
• receiving echoes, and
• interpreting those echoes
A-mode
• A-mode (amplitude mode) is the simplest type
of ultrasound.
• A single transducer scans a line through the
body with the echoes plotted on screen as a
function of depth.
• One use of the A scan is to measure length. For
an example, ophthalmologists can use it to
measure the diameter of the eye ball.
B-mode or 2D mode
• B-mode (brightness mode) ultrasound: a linear
array of transducers simultaneously scans a
plane through the body that can be viewed as a
two-dimensional image on screen.
• B-flow is a mode that digitally highlights moving
reflectors (mainly red blood cells) while
suppressing the signals from the surrounding
stationary tissue.
• It can visualize flowing blood and surrounding
stationary tissues simultaneously
M-mode
• In M-mode (motion mode) ultrasound, pulses are
emitted in quick succession – each time, either an A-
mode or B-mode image is taken. Over time, this is
analogous to recording a video in ultrasound. As the
organ boundaries that produce reflections move
relative to the probe, this can be used to determine
the velocity of specific organ structures.
• Used to study rhythmically moving structures
Application
• In the field of obstetrics and gynecology, where
ionizing radiation is to be avoided whenever possible.
• To evaluate fetal size and maturity and fetal
and placental position.
• It is a fast, relatively safe, and reliable techni
que for diagnosing multiple pregnancies. Uteri
ne tumors and other pelvic
masses, including abscesses, can be identified
by ultrasonography.
3.5MHz
7.5 MHz
Axial Resolution
The shorter the pulse,
the better the axial
resolution
Increasing the
frequency increases
axial resolution
Characteristics of Sound
Frequency
Sound
Frequency
Sound
Penetration
Axial
Resolution
High
Low
A transducer
with a large
surface area
will resolve
better in the
lateral
dimension
• Very important for ultrasound guidance with
needles/probes
Lateral Resolution
Parts of Ultrasound Machine
Transducer
Probe
Pulse
control
CPU
(Central
Processing
Unit)
Key Board Display
Storage
device
Printer
62
Parts of Ultrasound Machine
63
Components/ parts of Ultrasound machine
• Transducer probe - probe that sends and receives the sound waves
• Central processing unit (CPU) - computer that does all of the
calculations and contains the electrical power supplies for itself and
the transducer probe
• Transducer pulse controls - changes the amplitude, frequency and
duration of the pulses emitted from the transducer probe
• Display - displays the image from the ultrasound data processed by
the CPU
• Keyboard/cursor - inputs data and takes measurements from the
display
• Disk storage device (hard, floppy, CD) - stores the acquired images
• Printer - prints the image from the displayed data
Transducer
Transducer Probe
• Mouth and ears of the machine.
• It makes the sound waves and receives the echoes
• Piezoelectric crystals
• Electric current is applied and change shape rapidly
• When sound waves hit the crystals, they emit electrical
currents
66
Transducer
Transducer (pulse controls)
• The operator, called the ultra sonographer, changes the
amplitude, frequency and duration of the pulses
emitted from the transducer probe
67
Types of Ultrasound Transducer
68
Central Processing Unit (CPU)
• Brain of an ultrasound machine
• Contains the microprocessor, memory and amplifiers
• The transducer receives electrical currents from the CPU
and sends electrical pulses that are created by returning
echoes
69
Keyboard/Cursor
It allows the operator to add notes and to take measurements
of the image
70
Display
• Displays the image from the ultrasound data processed by
the CPU
• This image can be either in black-and-white or color,
depending upon the model of the ultrasound machine
71
How is an image formed on the monitor?
• Strong Reflections = White dots
Pericardium, calcified structures, diaphragm
• Weaker Reflections = Grey dots
Myocardium, valve tissue, vessel walls, liver
• No Reflections = Black dots
Intra-cardiac cavities, gall bladder
72
Disk Storage & Printers
The processed data and images can be stored on disks
• Hard disks, floppy disks, compact disks (CDs), or digital
video disks (DVDs)
• Most of the time, ultrasound scans are filled on floppy disks
and stored with the patient's medical records
• Most ultrasound machines have printers which are thermal
73
Block Diagram
74
Main Parts of Block Diagram
• User interface
• Controller
• Front End
•Scanner(beamforming and signal processing)
•Back End
75
User interface
• Desired mode of operation
• Information in and out of the system through connectors to the system
• VCRs & memory storage devices such as read/write CD-ROMs and DAT
drives
76
Controller
• System have one or more microprocessors
• Senses the settings of the controls and input devices
• Control the hardware to function in the desired mode.
• Estimate the level of acoustic output in real time
77
Front End
• This grouping within the scanner is the gateway of signals
going in and out of the selected transducer
• Pulses are sent to the transducer from the transmitter
circuitry
• Pulse-echo signals from the body are received by array
elements These signals then pass on to the receive beam
former.
78
Front End
• Organizing the many signals of the elements
• The transmit beamformer sends pulses to the elements
• Analog to Digital Converter
• Filters
79
Scanner(beamforming and signal processing)
• This grouping of functions is associated with image
formation, display and image metrics
• Image formation is achieved by organizing the lines and
putting them through a digital scan converter that
transforms them into a scan format for display on a video or
PC monitor.
• Image overlays containing alpha-numeric characters and
other information are added in image planes
80
30-Nov-18 Medical imaging modalities 60
Ultrasoun
d
Transducer
Ultrasound Transducer
Major components of transducers include:
the piezoelectric material
matching layer
backing block
acoustic absorber
insulating cover
Sensor electrodes and
transducer housing
piezoelectric
material
• A piezoelectric material (often a crystal or ceramic) is the
functional component of the transducer.
• It converts electrical energy into mechanical (sound) energy
by
• physical deformation of the crystal structure.
• Conversely, mechanical pressure applied to its surface creates
electrical energy.
• Piezoelectric materials are characterized by: a well-defined
molecular arrangement of electrical dipoles
Cont’d…
• An electrical dipole is a molecular entity containing
• positive and negative electric charges that has no net charge.
• When mechanically compressed by an externally applied
pressure,
• the alignment of the dipoles is disturbed from the equilibrium position to
cause an imbalance of the charge distribution.
• A potential difference (voltage) is created across the element
• with one surface maintaining a net positive charge and one surface a net
negative charge.
• Surface electrodes measure the voltage,
• which is proportional to the incident mechanical pressure amplitude.
• Conversely, application of an external voltage through
conductors attached to the surface electrodes induces
• the mechanical expansion and contraction of the transducer element
Piezo-electric
materials
Cont’d
…
• There are natural and synthetic piezoelectric materials.
• An example of a natural piezoelectric material is quartz crystal
• Ultrasound transducers for medical imaging applications employ:
• a synthetic piezoelectric ceramic, most often lead-zirconate-titanate (PZT).
30-Nov-18 Medical imaging modalities 78
Cont’d ….
• The thickness of layer is equal to one-fourth the wavelength of center
operating frequency
• determined from the center operating frequency of the transducer and speed of
sound in the matching layer.
• In addition to the matching layers, acoustic coupling gel (with acoustic
impedance similar to soft tissue)
• is used between the transducer and the skin of the patient
• to eliminate air pockets that could attenuate and reflect the ultrasoundbeam.
Activity 5
What is the optimal thickness of the matching layer used for PZT
transducer of 1 milli- meter thickness (assume the speed of sound in PZT
is 4000 m/s)?
Image Data
Acquisition
• Images are created using a pulse echo method of ultrasound production
and detection.
• Each pulse transmits directionally into the patient, and then experiences
partial reflections from tissue interfaces
• that create echoes, which return to the transducer.
Cont’d ….
Image formation using the pulse echo approach requires a number of
hardware components:
the beam former,
pulser,
receiver,
amplifier,
scan converter/image memory, and
display system
The transmit/receive
switch
The transmit/receive switch, synchronized with the pulser,
isolates the high voltage used for pulsing (~150 V) from the
sensitive amplification stages during receive mode with
induced voltages ranging from ~1 V to ~2 ~V from the
returning echoes.
After the ring-down time, when vibration of the piezoelectric
material has stopped, the transducer electronics are switched to
sensing small voltages caused by the returning echoes, over a
period up to about 1000 micro sec (1milli sec).
91
92
93
94
95
96
Basic Principals of Image Formation
Pulse Echo Principal
• A short ultrasound pulse is delivered to the tissues, and where there are changes in the
acoustic properties of the tissue, a fraction of the pulse is reflected (an echo) and
returns to the source (pulse-echo principal)
• Collection of the echoes and analysis of their amplitudes provides information about
the tissues along the path of travel
Tissue 1 Tissue 2 Tissue 3
Transducer
US Pulse
Reflected Echoes
Pulse Echo Principal
Tissue 1 Tissue 2 Tissue 3
Transducer
US Pulse
Reflected Echoes
The ultrasound pulse will travel at the speed of sound and the time between the pulse
emission and echo return will be known.
Therefore, the depth, d, at which the echo was generated can be determined and
spatially encoded in the depth direction.
Distance (D) = speed (c) x time (t)  2d = c t
Tomographic Imaging
Repeating this process many times with incremental changes in pulse direction allow
a volume to be sampled and a tomographic image to be formed.
i.e. A tomographic image is formed from a large number of image lines, where each
line in the image is produced by a pulse echo sequence
Transducer
101
Modes of
ultrasound
102
B-Mode Image
 A B-mode image is a cross-sectional image representing tissues and organ
boundaries within the body
 Constructed from echoes which are generated by reflection of US waves at tissue
boundaries, and scattering from small irregularities within tissues
 Each echo is displayed at a point in the image which corresponds to the relative
position of its origin within the body
 The brightness of the image at each point is related to the strength (amplitude) of
the echo
 B-mode = Brightness mode
B-Mode Image – How Long Does it Take?
1. Minimum time for one line = (2 x depth) / speed of sound = 2D / c seconds
2. Each frame of image contains N lines
3. Time for one frame = 2ND / c seconds
E.g. D = 12 cm, c = 1540 m/s, Frame rate = 20 frames per second
Frame rate = c / 2ND
N = c / (2D x Frame rate) = 320 lines (poor - approx half of standard TV)
Additional interpolated lines are inserted between image lines to boost image quality to
the human eye
4. Time is very important!!!
Time Gain Compensation (TGC)
• The deeper the source of echo Smaller signal intensity
• Due signal attenuation in tissue and reduction of the initial US beam intensity by
reflections
• Operator can TGC use to artificially ‘boost’ the signals from deeper tissues to
compensate for this (like a graphic equaliser)
M-Mode Image
• Can be used to observe the motion of tissues (e.g.Echocardiography)
• Image the same position (one image line) repeatly.
• One direction of display is used to represent time rather thanspace
Transducer at fixed point Time
Depth
More on ultrasound
Mod
es
• A-mode: A-mode (amplitude mode) is the simplest type of
ultrasound. A single transducer scans a line through the body with
the echoes plotted on screen as a function of depth. Therapeutic
ultrasound aimed at a specific tumor or calculus is also A-mode,
to allow for pinpoint accurate focus of the destructive wave
energy.
• B-mode or 2D mode: In B-mode (brightness mode) ultrasound, a
linear array of transducers simultaneously scans a plane through
the body that can be viewed as a two-dimensional image on screen.
More commonly known as 2D mode now.
• M-mode: In M-mode (motion mode) ultrasound, pulses are
emitted in quick succession – each time, either an A-mode or B-
mode image is taken. Over time, this is analogous to recording a
video in ultrasound. As the organ boundaries that produce
reflections move relative to the probe, this can be used to
determine the velocity of specific organ structures.
More on Imaging
Modes
• Doppler mode: This mode makes use of the Doppler effect in
measuring and visualizing blood flow
• Color Doppler: Velocity information is presented as a color coded overlay on
top of a B-mode image
• Continuous Doppler: Doppler information is sampled along a line through
the body, and all velocities detected at each time point is presented (on a time
line)
• Pulsed wave (PW) Doppler: Doppler information is sampled from only a
small sample volume (defined in 2D image), and presented on a timeline
• Duplex: a common name for the simultaneous presentation of 2D and
(usually) PW Doppler information. (Using modern ultrasound machines
color Doppler is almost always also used, hence the alternative name
Triplex.)
• Harmonic mode: In this mode a deep penetrating fundamental frequency is
emitted into the body and a harmonic overtone is detected. In this way depth
penetration can be gained with improved lateral resolution.
Ultrasound maintenance
Lab activity???
109
The end …..
Any ?

More Related Content

What's hot

TRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptx
TRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptxTRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptx
TRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptxVANI PUSHPA MUDAVATH
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Prajwith Rai
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducersRaghu Kishore Galla
 
Qaulity Control of CT
Qaulity Control of CTQaulity Control of CT
Qaulity Control of CTAsim Gilani
 
Computed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationComputed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationAnjan Dangal
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifactsansaripv
 
Difference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT ScannerDifference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT ScannerAbdul Rehman (R)(CT)(MR)
 
Mri coils
Mri coilsMri coils
Mri coilsSelf
 
Future Of Radiology
Future Of RadiologyFuture Of Radiology
Future Of RadiologyRMLIMS
 
Encoding and image formation
Encoding and image formationEncoding and image formation
Encoding and image formationAnkit Mishra
 
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...ANAND GURMAITA
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct pptAnand Rk
 
Ultrasound imaging
Ultrasound imagingUltrasound imaging
Ultrasound imagingAnjan Dangal
 
Computed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DRComputed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DRAshim Budhathoki
 
ULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESINDIA ULTRASOUND
 
Ultrasound physcics, techniques & knobology
Ultrasound physcics, techniques & knobologyUltrasound physcics, techniques & knobology
Ultrasound physcics, techniques & knobologySuzanneCain2
 

What's hot (20)

TRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptx
TRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptxTRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptx
TRIPLE PHASE CT AND ITS TECHNIQUE BY VANI PUSHPA (2).pptx
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)
 
CR & DR
CR & DRCR & DR
CR & DR
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducers
 
Qaulity Control of CT
Qaulity Control of CTQaulity Control of CT
Qaulity Control of CT
 
Ultrasound Normal Anatomy of Major Organs
Ultrasound Normal Anatomy of Major OrgansUltrasound Normal Anatomy of Major Organs
Ultrasound Normal Anatomy of Major Organs
 
Computed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationComputed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector Configuration
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
Difference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT ScannerDifference Between Single Slice and Multi Slice CT Scanner
Difference Between Single Slice and Multi Slice CT Scanner
 
Mri coils
Mri coilsMri coils
Mri coils
 
Future Of Radiology
Future Of RadiologyFuture Of Radiology
Future Of Radiology
 
Doppler physics ppt
Doppler physics pptDoppler physics ppt
Doppler physics ppt
 
Encoding and image formation
Encoding and image formationEncoding and image formation
Encoding and image formation
 
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
INTRODUCTION TO ULTRASOUND IMAGING ( ULTRASONGRAPHY ) - ANAND GURMAITA (BSc. ...
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct ppt
 
Ultrasound imaging
Ultrasound imagingUltrasound imaging
Ultrasound imaging
 
Computed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DRComputed radiography and digital radiography- CR/DR
Computed radiography and digital radiography- CR/DR
 
ULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLES
 
Ultrasound physcics, techniques & knobology
Ultrasound physcics, techniques & knobologyUltrasound physcics, techniques & knobology
Ultrasound physcics, techniques & knobology
 
CR and DR ppt
CR and DR pptCR and DR ppt
CR and DR ppt
 

Similar to BIS CHAPTER ONE.pptx

Ultrasonography final
Ultrasonography finalUltrasonography final
Ultrasonography finalravikumarbme
 
4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.pptRohit Bansal
 
Physic Of Ultrasound
Physic Of UltrasoundPhysic Of Ultrasound
Physic Of UltrasoundKhalis Karim
 
Principles of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxPrinciples of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxGemechisAsefa1
 
Ultrasound & properties
Ultrasound &  propertiesUltrasound &  properties
Ultrasound & propertiesMahesh Kumar
 
9 ultrasound part i (2)
9 ultrasound part i (2)9 ultrasound part i (2)
9 ultrasound part i (2)Abeer Texon
 
BASICS OF US & DUPLEX (NEW).ppt
BASICS OF US & DUPLEX (NEW).pptBASICS OF US & DUPLEX (NEW).ppt
BASICS OF US & DUPLEX (NEW).pptmohamedabushanab1
 
Applications of Ultrasound in Medicine
Applications of Ultrasound in MedicineApplications of Ultrasound in Medicine
Applications of Ultrasound in MedicinePranay Dutta
 
production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-Lushinga Mourice
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfghadaElbanby1
 
Physics of ultrasound and echocardiography
Physics of ultrasound and echocardiographyPhysics of ultrasound and echocardiography
Physics of ultrasound and echocardiographyjeetshitole
 
8159968.ppt
8159968.ppt8159968.ppt
8159968.pptYtchechy
 
Presentation1, basic principle of ultrasound.
Presentation1, basic principle of ultrasound.Presentation1, basic principle of ultrasound.
Presentation1, basic principle of ultrasound.Abdellah Nazeer
 
Emergency us physics and instrumentation
Emergency us physics  and instrumentationEmergency us physics  and instrumentation
Emergency us physics and instrumentationdrnikahmad
 
Basics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallBasics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallGamal Agmy
 

Similar to BIS CHAPTER ONE.pptx (20)

Ultrasonography final
Ultrasonography finalUltrasonography final
Ultrasonography final
 
4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt4.Physics of Ultrasound.ppt
4.Physics of Ultrasound.ppt
 
Physic Of Ultrasound
Physic Of UltrasoundPhysic Of Ultrasound
Physic Of Ultrasound
 
Principles of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptxPrinciples of Ultrasound Physics.pptx
Principles of Ultrasound Physics.pptx
 
Ultrasound & properties
Ultrasound &  propertiesUltrasound &  properties
Ultrasound & properties
 
9 ultrasound part i (2)
9 ultrasound part i (2)9 ultrasound part i (2)
9 ultrasound part i (2)
 
BASICS OF US & DUPLEX (NEW).ppt
BASICS OF US & DUPLEX (NEW).pptBASICS OF US & DUPLEX (NEW).ppt
BASICS OF US & DUPLEX (NEW).ppt
 
Applications of Ultrasound in Medicine
Applications of Ultrasound in MedicineApplications of Ultrasound in Medicine
Applications of Ultrasound in Medicine
 
Us physics (4)
Us physics (4)Us physics (4)
Us physics (4)
 
production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-production of ultrasound and physical characteristics-
production of ultrasound and physical characteristics-
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdf
 
PRINCIPLES OF ULTRASONOGRAPHY
PRINCIPLES OF ULTRASONOGRAPHYPRINCIPLES OF ULTRASONOGRAPHY
PRINCIPLES OF ULTRASONOGRAPHY
 
Physics of ultrasound and echocardiography
Physics of ultrasound and echocardiographyPhysics of ultrasound and echocardiography
Physics of ultrasound and echocardiography
 
PHYSICS REPORT.pptx
PHYSICS REPORT.pptxPHYSICS REPORT.pptx
PHYSICS REPORT.pptx
 
8159968.ppt
8159968.ppt8159968.ppt
8159968.ppt
 
Us physics (6)
Us physics (6)Us physics (6)
Us physics (6)
 
Presentation1, basic principle of ultrasound.
Presentation1, basic principle of ultrasound.Presentation1, basic principle of ultrasound.
Presentation1, basic principle of ultrasound.
 
Ultrasound Physics Jr.pptx
Ultrasound Physics  Jr.pptxUltrasound Physics  Jr.pptx
Ultrasound Physics Jr.pptx
 
Emergency us physics and instrumentation
Emergency us physics  and instrumentationEmergency us physics  and instrumentation
Emergency us physics and instrumentation
 
Basics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallBasics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wall
 

More from Tirusew1

DNA PCR2 minituarizatinon altemechanismration
DNA PCR2 minituarizatinon altemechanismrationDNA PCR2 minituarizatinon altemechanismration
DNA PCR2 minituarizatinon altemechanismrationTirusew1
 
Digital image processing fundamental explanation
Digital image processing fundamental explanationDigital image processing fundamental explanation
Digital image processing fundamental explanationTirusew1
 
biofluid mechanics full chapter presentation
biofluid mechanics full chapter presentationbiofluid mechanics full chapter presentation
biofluid mechanics full chapter presentationTirusew1
 
chapter-two ppt for digital image enhancement
chapter-two ppt for digital image enhancementchapter-two ppt for digital image enhancement
chapter-two ppt for digital image enhancementTirusew1
 
biomedical image processing analysis ppt
biomedical image processing analysis pptbiomedical image processing analysis ppt
biomedical image processing analysis pptTirusew1
 
intermiate equipments(1).ppt
intermiate equipments(1).pptintermiate equipments(1).ppt
intermiate equipments(1).pptTirusew1
 
basic ELECtronics(2).ppt
basic ELECtronics(2).pptbasic ELECtronics(2).ppt
basic ELECtronics(2).pptTirusew1
 
Chapter 6 Fluorescence Microscopy.ppt
Chapter 6 Fluorescence Microscopy.pptChapter 6 Fluorescence Microscopy.ppt
Chapter 6 Fluorescence Microscopy.pptTirusew1
 
Engineering and medical ethics.ppt
Engineering and medical ethics.pptEngineering and medical ethics.ppt
Engineering and medical ethics.pptTirusew1
 
Chapter 2S1.pptx
Chapter 2S1.pptxChapter 2S1.pptx
Chapter 2S1.pptxTirusew1
 
Chapter 2S2.pptx
Chapter 2S2.pptxChapter 2S2.pptx
Chapter 2S2.pptxTirusew1
 
UOG_BME_Curriculum_(Autosaved).pdf
UOG_BME_Curriculum_(Autosaved).pdfUOG_BME_Curriculum_(Autosaved).pdf
UOG_BME_Curriculum_(Autosaved).pdfTirusew1
 
2015.09 ATTF 1186C Incubator (Paulo).pdf
2015.09 ATTF 1186C Incubator (Paulo).pdf2015.09 ATTF 1186C Incubator (Paulo).pdf
2015.09 ATTF 1186C Incubator (Paulo).pdfTirusew1
 
All document for BIO-II course.pptx
All document for BIO-II course.pptxAll document for BIO-II course.pptx
All document for BIO-II course.pptxTirusew1
 
anesthesia machine.pptx
anesthesia machine.pptxanesthesia machine.pptx
anesthesia machine.pptxTirusew1
 
introduction to mechanical ventilator.pptx
introduction to mechanical ventilator.pptxintroduction to mechanical ventilator.pptx
introduction to mechanical ventilator.pptxTirusew1
 
dialysis tetorial ppt.pptx
dialysis tetorial ppt.pptxdialysis tetorial ppt.pptx
dialysis tetorial ppt.pptxTirusew1
 
tiru ppt.ppt
tiru ppt.ppttiru ppt.ppt
tiru ppt.pptTirusew1
 

More from Tirusew1 (18)

DNA PCR2 minituarizatinon altemechanismration
DNA PCR2 minituarizatinon altemechanismrationDNA PCR2 minituarizatinon altemechanismration
DNA PCR2 minituarizatinon altemechanismration
 
Digital image processing fundamental explanation
Digital image processing fundamental explanationDigital image processing fundamental explanation
Digital image processing fundamental explanation
 
biofluid mechanics full chapter presentation
biofluid mechanics full chapter presentationbiofluid mechanics full chapter presentation
biofluid mechanics full chapter presentation
 
chapter-two ppt for digital image enhancement
chapter-two ppt for digital image enhancementchapter-two ppt for digital image enhancement
chapter-two ppt for digital image enhancement
 
biomedical image processing analysis ppt
biomedical image processing analysis pptbiomedical image processing analysis ppt
biomedical image processing analysis ppt
 
intermiate equipments(1).ppt
intermiate equipments(1).pptintermiate equipments(1).ppt
intermiate equipments(1).ppt
 
basic ELECtronics(2).ppt
basic ELECtronics(2).pptbasic ELECtronics(2).ppt
basic ELECtronics(2).ppt
 
Chapter 6 Fluorescence Microscopy.ppt
Chapter 6 Fluorescence Microscopy.pptChapter 6 Fluorescence Microscopy.ppt
Chapter 6 Fluorescence Microscopy.ppt
 
Engineering and medical ethics.ppt
Engineering and medical ethics.pptEngineering and medical ethics.ppt
Engineering and medical ethics.ppt
 
Chapter 2S1.pptx
Chapter 2S1.pptxChapter 2S1.pptx
Chapter 2S1.pptx
 
Chapter 2S2.pptx
Chapter 2S2.pptxChapter 2S2.pptx
Chapter 2S2.pptx
 
UOG_BME_Curriculum_(Autosaved).pdf
UOG_BME_Curriculum_(Autosaved).pdfUOG_BME_Curriculum_(Autosaved).pdf
UOG_BME_Curriculum_(Autosaved).pdf
 
2015.09 ATTF 1186C Incubator (Paulo).pdf
2015.09 ATTF 1186C Incubator (Paulo).pdf2015.09 ATTF 1186C Incubator (Paulo).pdf
2015.09 ATTF 1186C Incubator (Paulo).pdf
 
All document for BIO-II course.pptx
All document for BIO-II course.pptxAll document for BIO-II course.pptx
All document for BIO-II course.pptx
 
anesthesia machine.pptx
anesthesia machine.pptxanesthesia machine.pptx
anesthesia machine.pptx
 
introduction to mechanical ventilator.pptx
introduction to mechanical ventilator.pptxintroduction to mechanical ventilator.pptx
introduction to mechanical ventilator.pptx
 
dialysis tetorial ppt.pptx
dialysis tetorial ppt.pptxdialysis tetorial ppt.pptx
dialysis tetorial ppt.pptx
 
tiru ppt.ppt
tiru ppt.ppttiru ppt.ppt
tiru ppt.ppt
 

Recently uploaded

Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerAnamika Sarkar
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineeringmalavadedarshan25
 
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptxDecoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptxJoão Esperancinha
 
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2RajaP95
 
chaitra-1.pptx fake news detection using machine learning
chaitra-1.pptx  fake news detection using machine learningchaitra-1.pptx  fake news detection using machine learning
chaitra-1.pptx fake news detection using machine learningmisbanausheenparvam
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girlsssuser7cb4ff
 
CCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdf
CCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdfCCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdf
CCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdfAsst.prof M.Gokilavani
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSCAESB
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort servicejennyeacort
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx959SahilShah
 
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)dollysharma2066
 
Sachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective IntroductionSachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective IntroductionDr.Costas Sachpazis
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile servicerehmti665
 
complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...asadnawaz62
 
power system scada applications and uses
power system scada applications and usespower system scada applications and uses
power system scada applications and usesDevarapalliHaritha
 

Recently uploaded (20)

Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
 
Internship report on mechanical engineering
Internship report on mechanical engineeringInternship report on mechanical engineering
Internship report on mechanical engineering
 
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptxDecoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
Decoding Kotlin - Your guide to solving the mysterious in Kotlin.pptx
 
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2HARMONY IN THE HUMAN BEING - Unit-II UHV-2
HARMONY IN THE HUMAN BEING - Unit-II UHV-2
 
chaitra-1.pptx fake news detection using machine learning
chaitra-1.pptx  fake news detection using machine learningchaitra-1.pptx  fake news detection using machine learning
chaitra-1.pptx fake news detection using machine learning
 
Call Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call GirlsCall Girls Narol 7397865700 Independent Call Girls
Call Girls Narol 7397865700 Independent Call Girls
 
CCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdf
CCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdfCCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdf
CCS355 Neural Network & Deep Learning Unit II Notes with Question bank .pdf
 
GDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentationGDSC ASEB Gen AI study jams presentation
GDSC ASEB Gen AI study jams presentation
 
young call girls in Green Park🔝 9953056974 🔝 escort Service
young call girls in Green Park🔝 9953056974 🔝 escort Serviceyoung call girls in Green Park🔝 9953056974 🔝 escort Service
young call girls in Green Park🔝 9953056974 🔝 escort Service
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
 
Application of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptxApplication of Residue Theorem to evaluate real integrations.pptx
Application of Residue Theorem to evaluate real integrations.pptx
 
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
 
Sachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective IntroductionSachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
🔝9953056974🔝!!-YOUNG call girls in Rajendra Nagar Escort rvice Shot 2000 nigh...
 
Call Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile serviceCall Girls Delhi {Jodhpur} 9711199012 high profile service
Call Girls Delhi {Jodhpur} 9711199012 high profile service
 
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCRCall Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
Call Us -/9953056974- Call Girls In Vikaspuri-/- Delhi NCR
 
complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...
 
power system scada applications and uses
power system scada applications and usespower system scada applications and uses
power system scada applications and uses
 

BIS CHAPTER ONE.pptx

  • 2. History Uses of ultrasonic energy in the 1940s. Left, in gastric ulcers. Right, in arthritis Ultrasonic therapy generator, the "Medi-Sonar" in the 1950s. A British ultrasonic apparatus for the treatment of Meniere's disease in the late 1950s
  • 3. History Denier's Ultrasonoscopic apparatus with ultrasound generator, emitter transducer and oscilloscope. This can be adapted for both therapeutic and diagnostic purposes The first hand-held imaging instrument was developed by John Wild and John Reid in the early 1950's
  • 4. Physics of Ultrasound Sound • Mechanical & Longitudinal wave • Travels in a straight Line • Human ear range is 20Hz - 20kHz • Cannot travel through Vacuum Ultrasound • Mechanical & Longitudinal wave • Exceeding the upper limit of human hearing • > 20,000H or 20kHz. 4
  • 5. Characteristics of Ultrasounds • Velocity • Frequency • Wavelength • Amplitude • The Relationship b/w v, f and λ is: v = f λ • Average speed of ultrasound in body is 1540m/sec • Imaging range is 2 to 20MHz 5
  • 6. Power and Intensity • A sound wave transports Energy through a medium from a source. Energy is measured in joules (J) • The Power, P,produce by a source of sound is the rate at which it produces energy. Power is measured in watts (W) where 1 W = 1 J/s • The Intensity, I, associated with a sound wave is the power per unit area. Intensity is measured in W/m2 • The power and intensity associated with a wave increase with the pressure amplitude, p Intensity, I  p2 Power, P  p
  • 7. Wavelength, Frequency and Speed • Waves are characterised by their wavelength, frequency and speed • The Wavelength,  , is the distance between consecutive peaks or other similar points on the wave. • The Frequency, f, is the number of oscillations per second • Frequency is measured in Hertz (Hz) where 1 Hz is one oscillation per second.
  • 9. Advanatges • Real-time • Portable • No ionizing radiation • Inexpensive? • Side-side comparison • Patients love it! • Guide procedures
  • 10. Disadvanatges • Operator dependent • Expensive? • Limited penetration through bone and air • Imaging deep structures
  • 12. Basic Ultrasound Physics Amplitude oscillations/sec = frequency - expressed in Hertz (Hz)
  • 13. Bats: Navigating with ultrasound • Bats make high-pitched chirps which are too high for humans to hear. This is called ultrasound • Like normal sound, ultrasound echoes off objects • The bat hears the echoes and works out what caused them • Dolphins also navigate with ultrasound • Submarines use a similar method called sonar • We can also use ultrasound to look inside the body…
  • 14. • An ultrasound element acts like a bat. • Emit ultrasound and detect echoes • Map out boundary of object • Medical Ultrasound 2MHz to 16MHz
  • 15. • Electric field applied to piezoelectric crystals located on transducer surface • Mechanical vibration of crystals creates sound waves Each crystal produces an US wave • Summation of all waves forms the US beam • Wave reflects as echo that vibrates transducer • Vibrations produce electrical pulses • Scanner processes and transforms to image
  • 16. Ultrasound Production • Transducer produces ultrasound pulses • These elements convert electrical energy into a mechanical ultrasound wave • Reflected echoes return to the scan head which converts the ultrasound wave into an electrical signal
  • 17. Frequency vs. Resolution • The frequency also affects the QUALITY of the ultrasound image – The HIGHER the frequency, the BETTER the resolution • The LOWER the frequency, the LESS the resolution • A 12 MHz transducer has very good resolution, but cannot penetrate very deep into the body • A 3 MHz transducer can penetrate deep into the body, but the resolution is not as good as the 12 MHz
  • 18. Image Formation Electrical signal produces ‘dots’ on the screen • Brightness of the dots is proportional to the strength of the returning echoes • Location of the dots is determined by travel time. The velocity in tissue is assumed constant at 1540m/sec Distance = Velocity Time
  • 19. 16
  • 20. 17
  • 21. 18
  • 22. Activity 1 • If an echo is detected from a reflector in human soft tissuein 0.065 ms . How far is the reflector from the transducer.
  • 23. • Acoustic impedance (AI) is dependent on the density of the material in which sound is propagated - the greater the impedance the denser the material. • Reflections comes from the interface of different AI’s • greater  of the AI = more signal reflected • works both ways (send and receive directions) Medium 1 Medium 2 Medium 3 Transducer Interactions of Ultrasound with Tissue
  • 24. Interactions of Ultrasound with Tissue • Reflection • Refraction • Transmission • Attenuation
  • 25. Interactions of Ultrasound with Tissue • Reflection – The ultrasound reflects off tissue and returns to the transducer, the amount of reflection depends on differences in acoustic impedance – The ultrasound image is formed from reflected echoes transducer
  • 26. Reflection • Reflection (specular reflection) occurs at tissue boundaries where there is a difference in the acoustic impedance, Z, of the two tissues • When the incident ultrasound wave is perpendicular to the boundary, a fraction of it’s energy is reflected (an echo) directly back towards the source • The remaining energy is transmitted into the second tissue and continues in the initial direction Z1 Z2 Incident Reflectio n (echo) Transmission
  • 27. 31
  • 28. Reflection • The fraction of ultrasound intensity reflected at an interface is given by the intensity reflection coefficient, R • The fraction of ultrasound energy reflected depends on the difference between the Z values of the two materials • R increases rapidly as the difference in Z increases Z1 Z2 Ii Ir It R = Z2 –Z1 Z1 +Z2 i Ir I = ( ) 2
  • 29. Reflection • The fraction of ultrasound intensity transmitted at an interface is given by the intensity transmission coefficient, T • Ultrasound imaging is only possible when the wave propagates through materials with similar acoustic impedances – only a small fraction of energy is reflected and the rest is transmitted Z1 Z2 Ii Ir It T = It Ii = 1 - R
  • 30. Reflection TissueInterface R T Liver – Fat 0.01 0.99 Fat– Muscle 0.02 0.98 Muscle - Bone 0.41 0.59 Muscle -Air 0.99 0.01 At soft tissue – soft tissue interfaces 1-2% of the ultrasound intensity is reflected At soft tissue – air interfaces 99% of the incident intensity is reflected • At soft tissue – air or soft tissue – bone interfaces, a large proportion of the incident intensity is reflected, making anatomy beyond such interfaces unobservable • Acoustic coupling gel is used between the face of the ultrasound transducer and skin to eliminate air pockets
  • 31. Reflection • When the wave is not incident perpendicular to the interface, non-normal incidence, the reflected angle is equal to the incident angle (i.e. θr = θi) • Echoes are directed away from the source of ultrasound and may be undetected • The transmitted wave does not continue in the incident direction (i.e. θt ≠ θi) • The change in direction is described by Refraction Z1 Z2 Incident Reflection (echo) Transmission (refraction) θi θr θt
  • 32. Activity 2 Approximately what fraction of an ultrasound beam is transmitted from an interface between two media with Z values of 1.65 and 1.55? 41
  • 33. ACTIVITY 3 Calculate the remaining intensity of a 100 W/m^2 ultrasound pulse that loses 30 dB while traveling through tissue? 56
  • 34. 57
  • 36. 59
  • 37.
  • 38. Transmission – Some of the ultrasound waves continue deeper into the body – These waves will reflect from deeper tissue structures transducer Interactions of Ultrasound with Tissue
  • 39. Attenuation – Defined - the deeper the wave travels in the body, the weaker it becomes -3 processes: reflection, absorption, refraction – Air (lung)> bone > muscle > soft tissue >blood > water Interactions of Ultrasound with Tissue
  • 40. Reflected Echo’s • Strong Reflections = White dots Diaphragm, tendons, bone ‘Hyperechoic’
  • 41. Weaker Reflections = Grey dots – Most solid organs, – thick fluid – ‘isoechoic’ Reflected Echo’s
  • 42. Reflected Echo’s • No Reflections = Black dots – Fluid within a cyst, urine, blood ‘Hypoechoic’ or echofree
  • 43. What determines how far ultrasound waves can travel? • The FREQUENCY of the transducer – The HIGHER the frequency, the LESS it can penetrate – The LOWER the frequency, the DEEPER it can penetrate – Attenuation is directly related to frequency
  • 44. 7.5 MHz 3.5 MHz Thinner crystal produces smaller sound waves. The Transducer Thicker crystal produces bigger sound waves.
  • 45. The Transducer The LOWER the frequency the better the penetration Bigger, Stronger The HIGHER frequency the less the penetration Smaller, weaker 3.5 MHz 7.5 MHz
  • 46. The Transducer Short pulses of sound are sent (transmits) into the body and then the transducer listens for the returning signals (receives). The ultrasound system processes the returning signals into images that are displayed on the ultrasound monitor Transmits Waits Receives
  • 47. The strength of the returning echoes also depends on the differences in the acoustic impedance between various structures. Acoustic impedance relates to tissue density. The greater the difference in density between two structures, the stronger the returning echo Examples: different: aorta and liver same: kidney and liver Grayscale Imaging
  • 48. • Attenuation: A decrease in the strength of the sound wave as it passes through tissue and further into the body. • Acoustic Impedance: The resistance of the sound wave traveling through tissue .Each tissue has its own acoustic impedance due to the density of the tissue. • Through Transmission: There is no attenuation of the sound wave traveling through the tissue. Grayscale Imaging
  • 49. WHITE DOTS = STRONG = e.g., bone BLACK DOTS = NO reflections = e.g., fluid GRAY (different shades) = WEAKER reflections Grayscale Imaging
  • 50. The strength of the returning echo is directly related to the angle at which the ultrasound beam strikes an interface. Grayscale Imaging The more perpendicular the ultrasound beam, the stronger the returning echo.
  • 51. Resolution • Clarity of picture • The ability to identify structures very close together./ • Ability of equipment to detect 2 separate reflectors in tissue and to display them as 2 separate reflectors on the monitor without merging them.
  • 52. Image Resolution Types of Resolution Axial Ability to identify structures that are one in front of the other Lateral Ability to identify structures that are side by side Temporal Ability to accurately locate a moving structure Spatial Ability to display very small structures in their correct anatomic location.
  • 53. Next class • Principles of A-Mode, B-Mode, M-Mode • Application area of ultrasound • parts of ultrasound machineineParts of Ultrasound Machine
  • 54. Summary of previous class • SOUND to IMAGE • producing a sound wave, • receiving echoes, and • interpreting those echoes
  • 55. A-mode • A-mode (amplitude mode) is the simplest type of ultrasound. • A single transducer scans a line through the body with the echoes plotted on screen as a function of depth. • One use of the A scan is to measure length. For an example, ophthalmologists can use it to measure the diameter of the eye ball.
  • 56. B-mode or 2D mode • B-mode (brightness mode) ultrasound: a linear array of transducers simultaneously scans a plane through the body that can be viewed as a two-dimensional image on screen. • B-flow is a mode that digitally highlights moving reflectors (mainly red blood cells) while suppressing the signals from the surrounding stationary tissue. • It can visualize flowing blood and surrounding stationary tissues simultaneously
  • 57. M-mode • In M-mode (motion mode) ultrasound, pulses are emitted in quick succession – each time, either an A- mode or B-mode image is taken. Over time, this is analogous to recording a video in ultrasound. As the organ boundaries that produce reflections move relative to the probe, this can be used to determine the velocity of specific organ structures. • Used to study rhythmically moving structures
  • 58. Application • In the field of obstetrics and gynecology, where ionizing radiation is to be avoided whenever possible. • To evaluate fetal size and maturity and fetal and placental position. • It is a fast, relatively safe, and reliable techni que for diagnosing multiple pregnancies. Uteri ne tumors and other pelvic masses, including abscesses, can be identified by ultrasonography.
  • 59. 3.5MHz 7.5 MHz Axial Resolution The shorter the pulse, the better the axial resolution Increasing the frequency increases axial resolution
  • 61. A transducer with a large surface area will resolve better in the lateral dimension • Very important for ultrasound guidance with needles/probes Lateral Resolution
  • 62. Parts of Ultrasound Machine Transducer Probe Pulse control CPU (Central Processing Unit) Key Board Display Storage device Printer 62
  • 63. Parts of Ultrasound Machine 63
  • 64. Components/ parts of Ultrasound machine
  • 65. • Transducer probe - probe that sends and receives the sound waves • Central processing unit (CPU) - computer that does all of the calculations and contains the electrical power supplies for itself and the transducer probe • Transducer pulse controls - changes the amplitude, frequency and duration of the pulses emitted from the transducer probe • Display - displays the image from the ultrasound data processed by the CPU • Keyboard/cursor - inputs data and takes measurements from the display • Disk storage device (hard, floppy, CD) - stores the acquired images • Printer - prints the image from the displayed data
  • 66. Transducer Transducer Probe • Mouth and ears of the machine. • It makes the sound waves and receives the echoes • Piezoelectric crystals • Electric current is applied and change shape rapidly • When sound waves hit the crystals, they emit electrical currents 66
  • 67. Transducer Transducer (pulse controls) • The operator, called the ultra sonographer, changes the amplitude, frequency and duration of the pulses emitted from the transducer probe 67
  • 68. Types of Ultrasound Transducer 68
  • 69. Central Processing Unit (CPU) • Brain of an ultrasound machine • Contains the microprocessor, memory and amplifiers • The transducer receives electrical currents from the CPU and sends electrical pulses that are created by returning echoes 69
  • 70. Keyboard/Cursor It allows the operator to add notes and to take measurements of the image 70
  • 71. Display • Displays the image from the ultrasound data processed by the CPU • This image can be either in black-and-white or color, depending upon the model of the ultrasound machine 71
  • 72. How is an image formed on the monitor? • Strong Reflections = White dots Pericardium, calcified structures, diaphragm • Weaker Reflections = Grey dots Myocardium, valve tissue, vessel walls, liver • No Reflections = Black dots Intra-cardiac cavities, gall bladder 72
  • 73. Disk Storage & Printers The processed data and images can be stored on disks • Hard disks, floppy disks, compact disks (CDs), or digital video disks (DVDs) • Most of the time, ultrasound scans are filled on floppy disks and stored with the patient's medical records • Most ultrasound machines have printers which are thermal 73
  • 75. Main Parts of Block Diagram • User interface • Controller • Front End •Scanner(beamforming and signal processing) •Back End 75
  • 76. User interface • Desired mode of operation • Information in and out of the system through connectors to the system • VCRs & memory storage devices such as read/write CD-ROMs and DAT drives 76
  • 77. Controller • System have one or more microprocessors • Senses the settings of the controls and input devices • Control the hardware to function in the desired mode. • Estimate the level of acoustic output in real time 77
  • 78. Front End • This grouping within the scanner is the gateway of signals going in and out of the selected transducer • Pulses are sent to the transducer from the transmitter circuitry • Pulse-echo signals from the body are received by array elements These signals then pass on to the receive beam former. 78
  • 79. Front End • Organizing the many signals of the elements • The transmit beamformer sends pulses to the elements • Analog to Digital Converter • Filters 79
  • 80. Scanner(beamforming and signal processing) • This grouping of functions is associated with image formation, display and image metrics • Image formation is achieved by organizing the lines and putting them through a digital scan converter that transforms them into a scan format for display on a video or PC monitor. • Image overlays containing alpha-numeric characters and other information are added in image planes 80
  • 81. 30-Nov-18 Medical imaging modalities 60
  • 82.
  • 84. Ultrasound Transducer Major components of transducers include: the piezoelectric material matching layer backing block acoustic absorber insulating cover Sensor electrodes and transducer housing
  • 85.
  • 86. piezoelectric material • A piezoelectric material (often a crystal or ceramic) is the functional component of the transducer. • It converts electrical energy into mechanical (sound) energy by • physical deformation of the crystal structure. • Conversely, mechanical pressure applied to its surface creates electrical energy. • Piezoelectric materials are characterized by: a well-defined molecular arrangement of electrical dipoles
  • 87. Cont’d… • An electrical dipole is a molecular entity containing • positive and negative electric charges that has no net charge. • When mechanically compressed by an externally applied pressure, • the alignment of the dipoles is disturbed from the equilibrium position to cause an imbalance of the charge distribution. • A potential difference (voltage) is created across the element • with one surface maintaining a net positive charge and one surface a net negative charge. • Surface electrodes measure the voltage, • which is proportional to the incident mechanical pressure amplitude. • Conversely, application of an external voltage through conductors attached to the surface electrodes induces • the mechanical expansion and contraction of the transducer element
  • 89.
  • 90.
  • 91.
  • 92.
  • 93.
  • 94.
  • 95.
  • 96. Cont’d … • There are natural and synthetic piezoelectric materials. • An example of a natural piezoelectric material is quartz crystal • Ultrasound transducers for medical imaging applications employ: • a synthetic piezoelectric ceramic, most often lead-zirconate-titanate (PZT).
  • 97. 30-Nov-18 Medical imaging modalities 78
  • 98. Cont’d …. • The thickness of layer is equal to one-fourth the wavelength of center operating frequency • determined from the center operating frequency of the transducer and speed of sound in the matching layer. • In addition to the matching layers, acoustic coupling gel (with acoustic impedance similar to soft tissue) • is used between the transducer and the skin of the patient • to eliminate air pockets that could attenuate and reflect the ultrasoundbeam.
  • 99. Activity 5 What is the optimal thickness of the matching layer used for PZT transducer of 1 milli- meter thickness (assume the speed of sound in PZT is 4000 m/s)?
  • 100. Image Data Acquisition • Images are created using a pulse echo method of ultrasound production and detection. • Each pulse transmits directionally into the patient, and then experiences partial reflections from tissue interfaces • that create echoes, which return to the transducer.
  • 101. Cont’d …. Image formation using the pulse echo approach requires a number of hardware components: the beam former, pulser, receiver, amplifier, scan converter/image memory, and display system
  • 102.
  • 103.
  • 104.
  • 105. The transmit/receive switch The transmit/receive switch, synchronized with the pulser, isolates the high voltage used for pulsing (~150 V) from the sensitive amplification stages during receive mode with induced voltages ranging from ~1 V to ~2 ~V from the returning echoes. After the ring-down time, when vibration of the piezoelectric material has stopped, the transducer electronics are switched to sensing small voltages caused by the returning echoes, over a period up to about 1000 micro sec (1milli sec).
  • 106. 91
  • 107. 92
  • 108. 93
  • 109. 94
  • 110. 95
  • 111. 96
  • 112. Basic Principals of Image Formation
  • 113. Pulse Echo Principal • A short ultrasound pulse is delivered to the tissues, and where there are changes in the acoustic properties of the tissue, a fraction of the pulse is reflected (an echo) and returns to the source (pulse-echo principal) • Collection of the echoes and analysis of their amplitudes provides information about the tissues along the path of travel Tissue 1 Tissue 2 Tissue 3 Transducer US Pulse Reflected Echoes
  • 114. Pulse Echo Principal Tissue 1 Tissue 2 Tissue 3 Transducer US Pulse Reflected Echoes The ultrasound pulse will travel at the speed of sound and the time between the pulse emission and echo return will be known. Therefore, the depth, d, at which the echo was generated can be determined and spatially encoded in the depth direction. Distance (D) = speed (c) x time (t)  2d = c t
  • 115. Tomographic Imaging Repeating this process many times with incremental changes in pulse direction allow a volume to be sampled and a tomographic image to be formed. i.e. A tomographic image is formed from a large number of image lines, where each line in the image is produced by a pulse echo sequence Transducer
  • 117. 102
  • 118. B-Mode Image  A B-mode image is a cross-sectional image representing tissues and organ boundaries within the body  Constructed from echoes which are generated by reflection of US waves at tissue boundaries, and scattering from small irregularities within tissues  Each echo is displayed at a point in the image which corresponds to the relative position of its origin within the body  The brightness of the image at each point is related to the strength (amplitude) of the echo  B-mode = Brightness mode
  • 119. B-Mode Image – How Long Does it Take? 1. Minimum time for one line = (2 x depth) / speed of sound = 2D / c seconds 2. Each frame of image contains N lines 3. Time for one frame = 2ND / c seconds E.g. D = 12 cm, c = 1540 m/s, Frame rate = 20 frames per second Frame rate = c / 2ND N = c / (2D x Frame rate) = 320 lines (poor - approx half of standard TV) Additional interpolated lines are inserted between image lines to boost image quality to the human eye 4. Time is very important!!!
  • 120. Time Gain Compensation (TGC) • The deeper the source of echo Smaller signal intensity • Due signal attenuation in tissue and reduction of the initial US beam intensity by reflections • Operator can TGC use to artificially ‘boost’ the signals from deeper tissues to compensate for this (like a graphic equaliser)
  • 121. M-Mode Image • Can be used to observe the motion of tissues (e.g.Echocardiography) • Image the same position (one image line) repeatly. • One direction of display is used to represent time rather thanspace Transducer at fixed point Time Depth
  • 122. More on ultrasound Mod es • A-mode: A-mode (amplitude mode) is the simplest type of ultrasound. A single transducer scans a line through the body with the echoes plotted on screen as a function of depth. Therapeutic ultrasound aimed at a specific tumor or calculus is also A-mode, to allow for pinpoint accurate focus of the destructive wave energy. • B-mode or 2D mode: In B-mode (brightness mode) ultrasound, a linear array of transducers simultaneously scans a plane through the body that can be viewed as a two-dimensional image on screen. More commonly known as 2D mode now. • M-mode: In M-mode (motion mode) ultrasound, pulses are emitted in quick succession – each time, either an A-mode or B- mode image is taken. Over time, this is analogous to recording a video in ultrasound. As the organ boundaries that produce reflections move relative to the probe, this can be used to determine the velocity of specific organ structures.
  • 123. More on Imaging Modes • Doppler mode: This mode makes use of the Doppler effect in measuring and visualizing blood flow • Color Doppler: Velocity information is presented as a color coded overlay on top of a B-mode image • Continuous Doppler: Doppler information is sampled along a line through the body, and all velocities detected at each time point is presented (on a time line) • Pulsed wave (PW) Doppler: Doppler information is sampled from only a small sample volume (defined in 2D image), and presented on a timeline • Duplex: a common name for the simultaneous presentation of 2D and (usually) PW Doppler information. (Using modern ultrasound machines color Doppler is almost always also used, hence the alternative name Triplex.) • Harmonic mode: In this mode a deep penetrating fundamental frequency is emitted into the body and a harmonic overtone is detected. In this way depth penetration can be gained with improved lateral resolution.