Metrohm inline sample preparation
Techniques for analytical success
Metrohm Ion Chromatography
Bart Cleeren, Metrohm België
ic_sampl_prep_26062003_e.ppt 2
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 3
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 4
IC IC
IC IC
IC
Metrohm Ion Chromatography
Ultrafiltration
Inline sample preparation
Ultrafiltration cell
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
S Sample
P2 Peristaltic pump
A position fill
C
P1
E D
2
ic_sampl_prep_26062003_e.ppt 5
IC IC
IC IC
IC
Metrohm Ion Chromatography
Ultrafiltration Cell
Inline sample preparation
No particles on the acceptor side
Big particles are flushed away
and do not clog the membrane
High flow on the sample side
Low flow on the acceptor side
ic_sampl_prep_26062003_e.ppt 6
IC IC
IC IC
IC
Metrohm Ion Chromatography
Ultrafiltration
Inline sample preparation
# 1 Sample filtration
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
S Sample
P2 Peristaltic pump
A position fill
C
P1
E D
2
ic_sampl_prep_26062003_e.ppt 7
IC IC
IC IC
IC
Metrohm Ion Chromatography
Ultrafiltration
Inline sample preparation
# 2 Sample injection
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
S Sample
P2 Peristaltic pump
A position inject
C
P1
E D
2
ic_sampl_prep_26062003_e.ppt 8
IC IC
IC IC
IC
Metrohm Ion Chromatography
Ultrafiltration: vegetable juice
Inline sample preparation
A Supp 4 – 250
1 Organics n.q.
2 Organics n.q.
3 Cl– 4141
4 NO3
– 102
5 HPO4
2– 577
6 SO4
2– 842
7 Oxalate n.q.
ppt
NaHCO3/Na2CO3; 1.7/1.8 mmol/L;
dilution 1:100; 0.15 µm rec. cellul.
ic_sampl_prep_26062003_e.ppt 9
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 10
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Dialysis cell, acceptor solution
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
P3 Peristaltic pump (2 Ch.)
Dialysis cell
A Acceptor solution
Dialysis
C
P1
E D
2
A E
P1
P2
P3
A D
C
X
X
ic_sampl_prep_26062003_e.ppt 11
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 2 Stopped flow dialysis
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
Injection Valve «B»
Dialysis cell
A Acceptor solution
Dialysis
C
P1
E D
2
A
P3
P2
A
P1
C
D
E
ic_sampl_prep_26062003_e.ppt 12
IC IC
IC IC
IC
Metrohm Ion Chromatography
Dialysis cell
Inline sample preparation
Continuous flow dialysis
Exclusion diameter 0.2 µm
Ions can pass
Oil droplets, fat micelles, particles
are excluded
Strong concentration gradient
No equilibrium concentration
Cellulose acetate membrane
ic_sampl_prep_26062003_e.ppt 13
IC IC
IC IC
IC
Metrohm Ion Chromatography
Dialysis cell
Inline sample preparation
Stopped flow dialysis
Dialysis time 2 ... 8 min
Sample volume 10 mL
No loss of concentration on the
sample side
Equilibrium concentration
Dialysis rate > 99 %
No filtration, no pre concentration
Unique technique, that had
significantly improves the field of
application for IC
ic_sampl_prep_26062003_e.ppt 14
IC IC
IC IC
IC
Metrohm Ion Chromatography
Dialysis: milk
Inline sample preparation
Metrosep Dual 1 – 70
1 Chloride 938
2 Nitrate n.q.
3 Phosphate n.q
4 Sufate 103
ppm
NaHCO3/Na2CO3; 2.4/2.0 mmol/L;
dilution 1:5; 0.2 µm cellul. acetat
ic_sampl_prep_26062003_e.ppt 15
IC IC
IC IC
IC
Metrohm Ion Chromatography
Dialysis: cutting oil emulsion
Inline sample preparation
Metrosep Dual 2
1 Chloride 74.7
2 Nitrite 20.5.
3 Nitrate 23.9
4 n. det. -
5 Phosphate 25.2
6 Sulfate 192.7
7 n. det. -
NaHCO3/Na2CO3; 2.0/1.3 mmol/L;
dilution 1:5; 0.2 µm cellul. acetat
ic_sampl_prep_26062003_e.ppt 16
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 17
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 1 Sample preconcentration
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
C
P1
D
P3 2
S
Inline pre concentration
E
C
P1
P 2
D
ic_sampl_prep_26062003_e.ppt 18
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 2 Sample analysis
E Mobile phase: eluent
P 1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P 2 Peristaltic pump
C
P1
E D
P3 2
S
Inline preconcentration
E
C
D
P1
P 2
ic_sampl_prep_26062003_e.ppt 19
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline preconcentration: ultra pure water
Inline sample preparation
Metrosep A Supp 5 – 150
1 Acetate 2.3
2 Formate 22.0
3 Chloride 13.0
4 Carbonate n.q.
5 Nitrate 4.5
6 Phosphate 16.0
7 Sulfate 8.5
8 Oxalate 12.0
ppt
NaHCO3/Na2CO3; 1.0/3.2 mmol/L;
20 mL preconc,
ic_sampl_prep_26062003_e.ppt 20
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline preconcentration: ultra pure water
Inline sample preparation
Metrosep Dual 2
1 Fluoride 45 – 150
2 Chloride 100 – 210
3 Bromide 0 – 100
4 Nitrate 10 – 112
5 Sulfate 40 – 145
ppt
NaHCO3/Na2CO3; 2.0/3.2 mmol/L;
10 mL preconc, spiked 100 ppt
ic_sampl_prep_26062003_e.ppt 21
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 22
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
Transfer solution «T»
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
Injection Valve «B»
P3 Peristaltic pump
T Transfer solution
A position fill, B position fill
C
P1
E D
2
T
Matrix elimination
ic_sampl_prep_26062003_e.ppt 23
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
# 1 Preconcentrator column inline
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
Injection Valve «B»
P3 Peristaltic pump
T Transfer solution
A position fill, B position fill
C
P1
E D
2
T
Matrix elimination
ic_sampl_prep_26062003_e.ppt 24
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
# 2 Sample transfer, matrix elim.
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
Injection Valve «B»
P3 Peristaltic pump
T Transfer solution
A position fill, B position inject
C
P1
E D
2
T
Matrix elimination
ic_sampl_prep_26062003_e.ppt 25
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
# 3 Sample inject
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2 Peristaltic pump
Injection Valve «B»
P3 Peristaltic pump
T Transfer solution
A position inject, B position inject
C
P1
E D
2
T
Matrix elimination
ic_sampl_prep_26062003_e.ppt 26
IC IC
IC IC
IC
Metrohm Ion Chromatography
Matrix elimination: organic solvents
Inline sample preparation
50 µL iPrOH
B with matrix elimination
A without matrix elimination
ppt
F–
Ac–
443 Cl–
30 NO3
– 28 SO4
2–
425 Cl–
24 NO3
– 17 SO4
2–
ic_sampl_prep_26062003_e.ppt 27
IC IC
IC IC
IC
Metrohm Ion Chromatography
Matrix elimination: 35 % H2O2
Inline sample preparation
left:
without m.e.
Loop 20 µL
right:
with m.e.
Loop 20 µL
transfer vol . 2 mL
ic_sampl_prep_26062003_e.ppt 28
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 29
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
833.0030 IC Sample Prep. Unit
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2, 3 Peristaltic pumps
Injection Valve «B»
T Transfer solution
833 IC Sample Prep
A position inject, B position fill
C
P1
E D
2
T
Matrix neutralization
ic_sampl_prep_26062003_e.ppt 30
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
# 1 Preconcentrator column inline
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2, 3 Peristaltic pumps
Injection Valve «B»
T Transfer solution
833 IC Sample Prep
A position fill, B position fill
C
P1
E D
2
T
Matrix neutralization
ic_sampl_prep_26062003_e.ppt 31
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
# 2 Sample transfer/neutralization
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2, 3 Peristaltic pumps
Injection Valve «B»
T Transfer solution
833 IC Sample Prep
A position fill, B position inject
C
P1
E D
2
T
Matrix neutralization
ic_sampl_prep_26062003_e.ppt 32
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
# 4 Sample inject
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2, 3 Peristaltic pumps
Injection Valve «B»
T Transfer solution
833 IC Sample Prep
A position inject, B position inject
C
P1
E D
2
T
Matrix neutralization
ic_sampl_prep_26062003_e.ppt 33
IC IC
IC IC
IC
0.74 F–
3.6 Cl–
0.97 NO3
–
0.98 HPO4
2–
1.1 SO4
2–
Metrohm Ion Chromatography
Matrix neutralization: caustic soda
Inline sample preparation
30 % NaOH, 20 µL loop
B with matrix elimination
A without matrix elimination
ppm
ic_sampl_prep_26062003_e.ppt 34
IC IC
IC IC
IC
Metrohm Ion Chromatography
Matrix neutralization: caustic soda
Inline sample preparation
Metrosep Dual 2
1 Chloride 25.08
2 Carbonate n.q.
3 Chlorate 4.76
4 Sulfate 10.43
ppm
NaHCO3/Na2CO3; 1.7/1.8 mmol/L;
50 % NaOH, 20 µL loop
ic_sampl_prep_26062003_e.ppt 35
IC IC
IC IC
IC
1
Metrohm Ion Chromatography
Matrix neutralization: ammonia
Inline sample preparation
Metrosep A Supp 4 – 250
1 Formate n.q.
2 Chloride 0.03
3 Nitrite n.q.
4 Bromide 0.95
5 Nitrat 0.04
6 Carbonate n.q.
7 Sulfate 0.13
ppm
NaHCO3/Na2CO3; 1.7/1.8 mmol/L;
26 % NH3, 20 µL loop
2
3
4
5 6
7
ic_sampl_prep_26062003_e.ppt 36
IC IC
IC IC
IC
P3
Metrohm Ion Chromatography
Inline sample preparation
OH cartridge
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P2, 3 Peristaltic pumps
Injection Valve «B»
T Transfer solution
Cartridge
A position fill, B position fill
C
P1
E D
2
T
Matrix neutralization
Ctr
ic_sampl_prep_26062003_e.ppt 37
IC IC
IC IC
IC
Metrohm Ion Chromatography
Matrix neutralization: HCl
Inline sample preparation
10 µL
HCl direct injection
ppm
Metrosep C2 – 250
ic_sampl_prep_26062003_e.ppt 38
IC IC
IC IC
IC
Metrohm Ion Chromatography
Matrix neutralization: HCl
Inline sample preparation
Metrosep C2 – 250
1 Nickel n.q.
2 Sodium 5.4
3 Ammonium 1.3
4 Potassium 0.8
5 Magnesium 1.5
6 Calcium 0.8
ppm
Tartaric-/dipicolinic acid; 4.0/0.75
mmol/L; 7 mol HCl, 10 µL loop
1
2
3
4 5 6
ic_sampl_prep_26062003_e.ppt 39
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Ultra filtration
Stop flow dialysis
Pre concentration technique
Matrix elimination
Matrix neutralization
In- line calibration
ic_sampl_prep_26062003_e.ppt 40
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Standard S, flow path
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P1,2,3 Peristaltic pumps
Injection Valve «B»
S Standard
A position inject, B position fill
Inline calibration
C
P1
E D
P3 2
S
ic_sampl_prep_26062003_e.ppt 41
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 1 Measurement of standard
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P1,2,3 Peristaltic pumps
Injection Valve «B»
S Standard
A position inject, B position fill
C
P1
E D
P3 2
S
Inline calibration
ic_sampl_prep_26062003_e.ppt 42
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 2 Preconcentrator inline
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P1,2,3 Peristaltic pumps
Injection Valve «B»
S Standard
A position fill, B position fill
C
P1
E D
P3 2
S
Inline calibration
ic_sampl_prep_26062003_e.ppt 43
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 3 Transfer of standard
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P1,2,3 Peristaltic pumps
Injection Valve «B»
S Standard
A position fill, B position inject
C
P1
E D
P3 2
S
Inline calibration
ic_sampl_prep_26062003_e.ppt 44
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 4 Inject of standard
E Mobile phase: eluent
P1 High pressure pump
Injection valve «A»
C Analytical column
»MSM» Suppressor
D Conductivity detector
P1,2,3 Peristaltic pumps
Injection Valve «B»
S Standard
A position inject, B position inject
C
P1
E D
P3 2
S
Inline calibration
ic_sampl_prep_26062003_e.ppt 45
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
# 1: Transfer of 2 mL ultrapure
water to the preconcentrator
Blank value
# 2: Transfer of 2 mL ultrapure
water to the preconcentrator.
Valve B is thrown from fill to inject
and back once
Calibration level 1
# 3: Transfer of 2 mL ultrapure
water to the preconcentrator.
Valve B is thrown from fill to inject
and back twice
Calibration level 2
# 4: Transfer of 2 mL ultrapure
water to the preconcentrator.
Valve B is thrown from fill to inject
and back three times
Calibration level 3, etc
Inline calibration
ic_sampl_prep_26062003_e.ppt 46
IC IC
IC IC
IC
Metrohm Ion Chromatography
Inline sample preparation
Inline calibration
100 ppt 100 ... 400 ppt
ic_sampl_prep_26062003_e.ppt 47
IC IC
IC IC
IC
inline sample preparation ?
Metrohm can do that
Metrohm Ion Chromatography

More Related Content

Similar to inline sample preparation.ppt

Vitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiterVitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiter
Rick Youngblood
 
04352464
0435246404352464
04352464
iiui
 

Similar to inline sample preparation.ppt (17)

NMDC Scheme Tracing By Yogesh Yadav.pptx
NMDC Scheme Tracing  By Yogesh Yadav.pptxNMDC Scheme Tracing  By Yogesh Yadav.pptx
NMDC Scheme Tracing By Yogesh Yadav.pptx
 
Vitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiterVitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiter
 
Biodegradation of CAHs in River Sediments
Biodegradation of CAHs in River SedimentsBiodegradation of CAHs in River Sediments
Biodegradation of CAHs in River Sediments
 
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 cod endress+hauser datasheet-cod analyzerStamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
 
MALDI tof
MALDI tofMALDI tof
MALDI tof
 
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
Stamolys ca71 fe endress+hauser datasheet-iron analyzerStamolys ca71 fe endress+hauser datasheet-iron analyzer
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
 
04352464
0435246404352464
04352464
 
PLL (Tsmc 0.18 process)
PLL (Tsmc 0.18 process)PLL (Tsmc 0.18 process)
PLL (Tsmc 0.18 process)
 
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
 
Q46H79S-Total-Chlorine-1.pdf
Q46H79S-Total-Chlorine-1.pdfQ46H79S-Total-Chlorine-1.pdf
Q46H79S-Total-Chlorine-1.pdf
 
Lc training basic_hplc_20_a modifi
Lc training basic_hplc_20_a  modifiLc training basic_hplc_20_a  modifi
Lc training basic_hplc_20_a modifi
 
Cro (cathod ray oscilloscope)
Cro (cathod ray oscilloscope)Cro (cathod ray oscilloscope)
Cro (cathod ray oscilloscope)
 
ATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdfATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdf
 
lec05.pdf
lec05.pdflec05.pdf
lec05.pdf
 
PSPICE seminar
PSPICE seminarPSPICE seminar
PSPICE seminar
 
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 mn endress+hauser datasheet-manganese analyserStamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
 
Split core ct, rogowski coil (2016)
Split core ct, rogowski coil (2016)Split core ct, rogowski coil (2016)
Split core ct, rogowski coil (2016)
 

More from waleedElazab (10)

Instrumentation of LC (HPLC-IC)18-22.pdf
Instrumentation of LC (HPLC-IC)18-22.pdfInstrumentation of LC (HPLC-IC)18-22.pdf
Instrumentation of LC (HPLC-IC)18-22.pdf
 
02The Art and Science of Chemical Analysis.pptx
02The Art and Science of  Chemical Analysis.pptx02The Art and Science of  Chemical Analysis.pptx
02The Art and Science of Chemical Analysis.pptx
 
01Form and Style for ASTM Standards.pptx
01Form and Style for ASTM Standards.pptx01Form and Style for ASTM Standards.pptx
01Form and Style for ASTM Standards.pptx
 
AAS Applications for heavy metal determination
AAS Applications for heavy metal determinationAAS Applications for heavy metal determination
AAS Applications for heavy metal determination
 
methods for continues emission gas monitoring
methods for continues emission gas  monitoringmethods for continues emission gas  monitoring
methods for continues emission gas monitoring
 
CV welazab.pptx
CV welazab.pptxCV welazab.pptx
CV welazab.pptx
 
Reliable year round source Mus.pdf
Reliable year round source Mus.pdfReliable year round source Mus.pdf
Reliable year round source Mus.pdf
 
Oil content.pdf
Oil content.pdfOil content.pdf
Oil content.pdf
 
Injection water quality problem.pdf
Injection water quality problem.pdfInjection water quality problem.pdf
Injection water quality problem.pdf
 
Injection water quality issues.pdf
Injection water quality issues.pdfInjection water quality issues.pdf
Injection water quality issues.pdf
 

Recently uploaded

Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...
Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...
Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...
The Hebrew University of Jerusalem
 
Heat Index related presentation ppt in india
Heat Index related presentation ppt in indiaHeat Index related presentation ppt in india
Heat Index related presentation ppt in india
SriSravani2
 

Recently uploaded (20)

Christmas Palm Trees in Florida The Ultimate Guide to Festive Landscaping wit...
Christmas Palm Trees in Florida The Ultimate Guide to Festive Landscaping wit...Christmas Palm Trees in Florida The Ultimate Guide to Festive Landscaping wit...
Christmas Palm Trees in Florida The Ultimate Guide to Festive Landscaping wit...
 
Global warming, Types, Causes and Effects.
Global warming, Types, Causes and Effects.Global warming, Types, Causes and Effects.
Global warming, Types, Causes and Effects.
 
CAUSES,EFFECTS,CONTROL OF DEFORESTATION.pptx
CAUSES,EFFECTS,CONTROL OF DEFORESTATION.pptxCAUSES,EFFECTS,CONTROL OF DEFORESTATION.pptx
CAUSES,EFFECTS,CONTROL OF DEFORESTATION.pptx
 
Hertwich_EnvironmentalImpacts_BuildingsGRO.pptx
Hertwich_EnvironmentalImpacts_BuildingsGRO.pptxHertwich_EnvironmentalImpacts_BuildingsGRO.pptx
Hertwich_EnvironmentalImpacts_BuildingsGRO.pptx
 
Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...
Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...
Determination of Total Iodine using ICP-MS in Israeli Bottled and Tap Water: ...
 
Role of nanotechnology in management of stored grain pests of cereals and pulses
Role of nanotechnology in management of stored grain pests of cereals and pulsesRole of nanotechnology in management of stored grain pests of cereals and pulses
Role of nanotechnology in management of stored grain pests of cereals and pulses
 
Book ℂall Girls Navi Mumbai Hire Me Neha 9910780858 Top Class ℂall Girl Servi...
Book ℂall Girls Navi Mumbai Hire Me Neha 9910780858 Top Class ℂall Girl Servi...Book ℂall Girls Navi Mumbai Hire Me Neha 9910780858 Top Class ℂall Girl Servi...
Book ℂall Girls Navi Mumbai Hire Me Neha 9910780858 Top Class ℂall Girl Servi...
 
ADBB 5cladba Precursor JWH018 +85244677121
ADBB 5cladba Precursor JWH018 +85244677121ADBB 5cladba Precursor JWH018 +85244677121
ADBB 5cladba Precursor JWH018 +85244677121
 
Smart Watering Solutions for Your Garden
Smart Watering Solutions for Your GardenSmart Watering Solutions for Your Garden
Smart Watering Solutions for Your Garden
 
Heat Index related presentation ppt in india
Heat Index related presentation ppt in indiaHeat Index related presentation ppt in india
Heat Index related presentation ppt in india
 
My Museum presentation by Jamilyn Gonzalez
My Museum presentation by Jamilyn GonzalezMy Museum presentation by Jamilyn Gonzalez
My Museum presentation by Jamilyn Gonzalez
 
Mastering Sustainable Living: Your Comprehensive Guide to Greener Choices in ...
Mastering Sustainable Living: Your Comprehensive Guide to Greener Choices in ...Mastering Sustainable Living: Your Comprehensive Guide to Greener Choices in ...
Mastering Sustainable Living: Your Comprehensive Guide to Greener Choices in ...
 
A Complete Guide to Understanding Air Quality Monitoring.pptx
A Complete Guide to Understanding Air Quality Monitoring.pptxA Complete Guide to Understanding Air Quality Monitoring.pptx
A Complete Guide to Understanding Air Quality Monitoring.pptx
 
Palynology: History, branches, basic principles and application, collection o...
Palynology: History, branches, basic principles and application, collection o...Palynology: History, branches, basic principles and application, collection o...
Palynology: History, branches, basic principles and application, collection o...
 
Rising temperatures also mean that more plant pests are appearing earlier and...
Rising temperatures also mean that more plant pests are appearing earlier and...Rising temperatures also mean that more plant pests are appearing earlier and...
Rising temperatures also mean that more plant pests are appearing earlier and...
 
A Review on Integrated River Basin Management and Development Master Plan of ...
A Review on Integrated River Basin Management and Development Master Plan of ...A Review on Integrated River Basin Management and Development Master Plan of ...
A Review on Integrated River Basin Management and Development Master Plan of ...
 
Town and Country Planning-he term 'town planning' first appeared in 1906 and ...
Town and Country Planning-he term 'town planning' first appeared in 1906 and ...Town and Country Planning-he term 'town planning' first appeared in 1906 and ...
Town and Country Planning-he term 'town planning' first appeared in 1906 and ...
 
Yil Me Hu Summer 2023 Edition - Nisqually Salmon Recovery Newsletter
Yil Me Hu Summer 2023 Edition - Nisqually Salmon Recovery NewsletterYil Me Hu Summer 2023 Edition - Nisqually Salmon Recovery Newsletter
Yil Me Hu Summer 2023 Edition - Nisqually Salmon Recovery Newsletter
 
Production, dispersal, sedimentation and taphonomy of spores/pollen
Production, dispersal, sedimentation and taphonomy of spores/pollenProduction, dispersal, sedimentation and taphonomy of spores/pollen
Production, dispersal, sedimentation and taphonomy of spores/pollen
 
Urban Farming: 3 Benefits, Challenges & The Rise of Green Cities | CIO Women ...
Urban Farming: 3 Benefits, Challenges & The Rise of Green Cities | CIO Women ...Urban Farming: 3 Benefits, Challenges & The Rise of Green Cities | CIO Women ...
Urban Farming: 3 Benefits, Challenges & The Rise of Green Cities | CIO Women ...
 

inline sample preparation.ppt

  • 1. ic_sampl_prep_26062003_e.ppt 1 IC IC IC IC IC Metrohm inline sample preparation Techniques for analytical success Metrohm Ion Chromatography Bart Cleeren, Metrohm België
  • 2. ic_sampl_prep_26062003_e.ppt 2 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 3. ic_sampl_prep_26062003_e.ppt 3 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 4. ic_sampl_prep_26062003_e.ppt 4 IC IC IC IC IC Metrohm Ion Chromatography Ultrafiltration Inline sample preparation Ultrafiltration cell E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector S Sample P2 Peristaltic pump A position fill C P1 E D 2
  • 5. ic_sampl_prep_26062003_e.ppt 5 IC IC IC IC IC Metrohm Ion Chromatography Ultrafiltration Cell Inline sample preparation No particles on the acceptor side Big particles are flushed away and do not clog the membrane High flow on the sample side Low flow on the acceptor side
  • 6. ic_sampl_prep_26062003_e.ppt 6 IC IC IC IC IC Metrohm Ion Chromatography Ultrafiltration Inline sample preparation # 1 Sample filtration E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector S Sample P2 Peristaltic pump A position fill C P1 E D 2
  • 7. ic_sampl_prep_26062003_e.ppt 7 IC IC IC IC IC Metrohm Ion Chromatography Ultrafiltration Inline sample preparation # 2 Sample injection E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector S Sample P2 Peristaltic pump A position inject C P1 E D 2
  • 8. ic_sampl_prep_26062003_e.ppt 8 IC IC IC IC IC Metrohm Ion Chromatography Ultrafiltration: vegetable juice Inline sample preparation A Supp 4 – 250 1 Organics n.q. 2 Organics n.q. 3 Cl– 4141 4 NO3 – 102 5 HPO4 2– 577 6 SO4 2– 842 7 Oxalate n.q. ppt NaHCO3/Na2CO3; 1.7/1.8 mmol/L; dilution 1:100; 0.15 µm rec. cellul.
  • 9. ic_sampl_prep_26062003_e.ppt 9 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 10. ic_sampl_prep_26062003_e.ppt 10 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Dialysis cell, acceptor solution E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump P3 Peristaltic pump (2 Ch.) Dialysis cell A Acceptor solution Dialysis C P1 E D 2 A E P1 P2 P3 A D C X X
  • 11. ic_sampl_prep_26062003_e.ppt 11 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 2 Stopped flow dialysis E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump Injection Valve «B» Dialysis cell A Acceptor solution Dialysis C P1 E D 2 A P3 P2 A P1 C D E
  • 12. ic_sampl_prep_26062003_e.ppt 12 IC IC IC IC IC Metrohm Ion Chromatography Dialysis cell Inline sample preparation Continuous flow dialysis Exclusion diameter 0.2 µm Ions can pass Oil droplets, fat micelles, particles are excluded Strong concentration gradient No equilibrium concentration Cellulose acetate membrane
  • 13. ic_sampl_prep_26062003_e.ppt 13 IC IC IC IC IC Metrohm Ion Chromatography Dialysis cell Inline sample preparation Stopped flow dialysis Dialysis time 2 ... 8 min Sample volume 10 mL No loss of concentration on the sample side Equilibrium concentration Dialysis rate > 99 % No filtration, no pre concentration Unique technique, that had significantly improves the field of application for IC
  • 14. ic_sampl_prep_26062003_e.ppt 14 IC IC IC IC IC Metrohm Ion Chromatography Dialysis: milk Inline sample preparation Metrosep Dual 1 – 70 1 Chloride 938 2 Nitrate n.q. 3 Phosphate n.q 4 Sufate 103 ppm NaHCO3/Na2CO3; 2.4/2.0 mmol/L; dilution 1:5; 0.2 µm cellul. acetat
  • 15. ic_sampl_prep_26062003_e.ppt 15 IC IC IC IC IC Metrohm Ion Chromatography Dialysis: cutting oil emulsion Inline sample preparation Metrosep Dual 2 1 Chloride 74.7 2 Nitrite 20.5. 3 Nitrate 23.9 4 n. det. - 5 Phosphate 25.2 6 Sulfate 192.7 7 n. det. - NaHCO3/Na2CO3; 2.0/1.3 mmol/L; dilution 1:5; 0.2 µm cellul. acetat
  • 16. ic_sampl_prep_26062003_e.ppt 16 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 17. ic_sampl_prep_26062003_e.ppt 17 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 1 Sample preconcentration E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump C P1 D P3 2 S Inline pre concentration E C P1 P 2 D
  • 18. ic_sampl_prep_26062003_e.ppt 18 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 2 Sample analysis E Mobile phase: eluent P 1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P 2 Peristaltic pump C P1 E D P3 2 S Inline preconcentration E C D P1 P 2
  • 19. ic_sampl_prep_26062003_e.ppt 19 IC IC IC IC IC Metrohm Ion Chromatography Inline preconcentration: ultra pure water Inline sample preparation Metrosep A Supp 5 – 150 1 Acetate 2.3 2 Formate 22.0 3 Chloride 13.0 4 Carbonate n.q. 5 Nitrate 4.5 6 Phosphate 16.0 7 Sulfate 8.5 8 Oxalate 12.0 ppt NaHCO3/Na2CO3; 1.0/3.2 mmol/L; 20 mL preconc,
  • 20. ic_sampl_prep_26062003_e.ppt 20 IC IC IC IC IC Metrohm Ion Chromatography Inline preconcentration: ultra pure water Inline sample preparation Metrosep Dual 2 1 Fluoride 45 – 150 2 Chloride 100 – 210 3 Bromide 0 – 100 4 Nitrate 10 – 112 5 Sulfate 40 – 145 ppt NaHCO3/Na2CO3; 2.0/3.2 mmol/L; 10 mL preconc, spiked 100 ppt
  • 21. ic_sampl_prep_26062003_e.ppt 21 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 22. ic_sampl_prep_26062003_e.ppt 22 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation Transfer solution «T» E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump Injection Valve «B» P3 Peristaltic pump T Transfer solution A position fill, B position fill C P1 E D 2 T Matrix elimination
  • 23. ic_sampl_prep_26062003_e.ppt 23 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation # 1 Preconcentrator column inline E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump Injection Valve «B» P3 Peristaltic pump T Transfer solution A position fill, B position fill C P1 E D 2 T Matrix elimination
  • 24. ic_sampl_prep_26062003_e.ppt 24 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation # 2 Sample transfer, matrix elim. E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump Injection Valve «B» P3 Peristaltic pump T Transfer solution A position fill, B position inject C P1 E D 2 T Matrix elimination
  • 25. ic_sampl_prep_26062003_e.ppt 25 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation # 3 Sample inject E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2 Peristaltic pump Injection Valve «B» P3 Peristaltic pump T Transfer solution A position inject, B position inject C P1 E D 2 T Matrix elimination
  • 26. ic_sampl_prep_26062003_e.ppt 26 IC IC IC IC IC Metrohm Ion Chromatography Matrix elimination: organic solvents Inline sample preparation 50 µL iPrOH B with matrix elimination A without matrix elimination ppt F– Ac– 443 Cl– 30 NO3 – 28 SO4 2– 425 Cl– 24 NO3 – 17 SO4 2–
  • 27. ic_sampl_prep_26062003_e.ppt 27 IC IC IC IC IC Metrohm Ion Chromatography Matrix elimination: 35 % H2O2 Inline sample preparation left: without m.e. Loop 20 µL right: with m.e. Loop 20 µL transfer vol . 2 mL
  • 28. ic_sampl_prep_26062003_e.ppt 28 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 29. ic_sampl_prep_26062003_e.ppt 29 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation 833.0030 IC Sample Prep. Unit E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2, 3 Peristaltic pumps Injection Valve «B» T Transfer solution 833 IC Sample Prep A position inject, B position fill C P1 E D 2 T Matrix neutralization
  • 30. ic_sampl_prep_26062003_e.ppt 30 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation # 1 Preconcentrator column inline E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2, 3 Peristaltic pumps Injection Valve «B» T Transfer solution 833 IC Sample Prep A position fill, B position fill C P1 E D 2 T Matrix neutralization
  • 31. ic_sampl_prep_26062003_e.ppt 31 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation # 2 Sample transfer/neutralization E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2, 3 Peristaltic pumps Injection Valve «B» T Transfer solution 833 IC Sample Prep A position fill, B position inject C P1 E D 2 T Matrix neutralization
  • 32. ic_sampl_prep_26062003_e.ppt 32 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation # 4 Sample inject E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2, 3 Peristaltic pumps Injection Valve «B» T Transfer solution 833 IC Sample Prep A position inject, B position inject C P1 E D 2 T Matrix neutralization
  • 33. ic_sampl_prep_26062003_e.ppt 33 IC IC IC IC IC 0.74 F– 3.6 Cl– 0.97 NO3 – 0.98 HPO4 2– 1.1 SO4 2– Metrohm Ion Chromatography Matrix neutralization: caustic soda Inline sample preparation 30 % NaOH, 20 µL loop B with matrix elimination A without matrix elimination ppm
  • 34. ic_sampl_prep_26062003_e.ppt 34 IC IC IC IC IC Metrohm Ion Chromatography Matrix neutralization: caustic soda Inline sample preparation Metrosep Dual 2 1 Chloride 25.08 2 Carbonate n.q. 3 Chlorate 4.76 4 Sulfate 10.43 ppm NaHCO3/Na2CO3; 1.7/1.8 mmol/L; 50 % NaOH, 20 µL loop
  • 35. ic_sampl_prep_26062003_e.ppt 35 IC IC IC IC IC 1 Metrohm Ion Chromatography Matrix neutralization: ammonia Inline sample preparation Metrosep A Supp 4 – 250 1 Formate n.q. 2 Chloride 0.03 3 Nitrite n.q. 4 Bromide 0.95 5 Nitrat 0.04 6 Carbonate n.q. 7 Sulfate 0.13 ppm NaHCO3/Na2CO3; 1.7/1.8 mmol/L; 26 % NH3, 20 µL loop 2 3 4 5 6 7
  • 36. ic_sampl_prep_26062003_e.ppt 36 IC IC IC IC IC P3 Metrohm Ion Chromatography Inline sample preparation OH cartridge E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P2, 3 Peristaltic pumps Injection Valve «B» T Transfer solution Cartridge A position fill, B position fill C P1 E D 2 T Matrix neutralization Ctr
  • 37. ic_sampl_prep_26062003_e.ppt 37 IC IC IC IC IC Metrohm Ion Chromatography Matrix neutralization: HCl Inline sample preparation 10 µL HCl direct injection ppm Metrosep C2 – 250
  • 38. ic_sampl_prep_26062003_e.ppt 38 IC IC IC IC IC Metrohm Ion Chromatography Matrix neutralization: HCl Inline sample preparation Metrosep C2 – 250 1 Nickel n.q. 2 Sodium 5.4 3 Ammonium 1.3 4 Potassium 0.8 5 Magnesium 1.5 6 Calcium 0.8 ppm Tartaric-/dipicolinic acid; 4.0/0.75 mmol/L; 7 mol HCl, 10 µL loop 1 2 3 4 5 6
  • 39. ic_sampl_prep_26062003_e.ppt 39 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Ultra filtration Stop flow dialysis Pre concentration technique Matrix elimination Matrix neutralization In- line calibration
  • 40. ic_sampl_prep_26062003_e.ppt 40 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Standard S, flow path E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P1,2,3 Peristaltic pumps Injection Valve «B» S Standard A position inject, B position fill Inline calibration C P1 E D P3 2 S
  • 41. ic_sampl_prep_26062003_e.ppt 41 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 1 Measurement of standard E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P1,2,3 Peristaltic pumps Injection Valve «B» S Standard A position inject, B position fill C P1 E D P3 2 S Inline calibration
  • 42. ic_sampl_prep_26062003_e.ppt 42 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 2 Preconcentrator inline E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P1,2,3 Peristaltic pumps Injection Valve «B» S Standard A position fill, B position fill C P1 E D P3 2 S Inline calibration
  • 43. ic_sampl_prep_26062003_e.ppt 43 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 3 Transfer of standard E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P1,2,3 Peristaltic pumps Injection Valve «B» S Standard A position fill, B position inject C P1 E D P3 2 S Inline calibration
  • 44. ic_sampl_prep_26062003_e.ppt 44 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 4 Inject of standard E Mobile phase: eluent P1 High pressure pump Injection valve «A» C Analytical column »MSM» Suppressor D Conductivity detector P1,2,3 Peristaltic pumps Injection Valve «B» S Standard A position inject, B position inject C P1 E D P3 2 S Inline calibration
  • 45. ic_sampl_prep_26062003_e.ppt 45 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation # 1: Transfer of 2 mL ultrapure water to the preconcentrator Blank value # 2: Transfer of 2 mL ultrapure water to the preconcentrator. Valve B is thrown from fill to inject and back once Calibration level 1 # 3: Transfer of 2 mL ultrapure water to the preconcentrator. Valve B is thrown from fill to inject and back twice Calibration level 2 # 4: Transfer of 2 mL ultrapure water to the preconcentrator. Valve B is thrown from fill to inject and back three times Calibration level 3, etc Inline calibration
  • 46. ic_sampl_prep_26062003_e.ppt 46 IC IC IC IC IC Metrohm Ion Chromatography Inline sample preparation Inline calibration 100 ppt 100 ... 400 ppt
  • 47. ic_sampl_prep_26062003_e.ppt 47 IC IC IC IC IC inline sample preparation ? Metrohm can do that Metrohm Ion Chromatography