This document summarizes a capstone project to design a temperature logger for blood analyzer cartridges. The project aimed to modify existing cartridge PCBs to log temperatures during shipment and storage using a microcontroller, thermistor, and battery. The design was tested and found to accurately measure temperature within ±2°C while extending battery life to over 3 months. Testing confirmed the design could interface with existing blood analyzer instruments and read stored temperature measurements.