POWER LINE
INTERFERENCE
ARTIFACTS IN ECG
RECORDING
ARTIFACTS
■ Artifact on the electrocardiogram can result from a variety of internal and
external causes.
■ Artifact and troubleshooting the problem will be straight forward.
■ When artifact mimics ECG abnormalities that can cause problems for
patient care.
2
BASIC ELECTRONIC RECORDING SYSTEM
3
EFFECTS OF ARTIFACTS ON ECG RECORDINGS (OR) POWER LINE
INTERFERENCE
Power line
interference
Shifting of
the
Baseline
Muscle
Tremor
4
Interference from Power Line
❖ Power line interference is easily recognizable since the interfering voltage in the ECG
❖ This interference may be due to the stray effect of the alternating current on the patient or because
of alternating current fields due to loops in the patient cable
❖ Other causes of interference are loose contacts on the patient cable as well as dirty electrodes.
❖ When the machine or the patient is not properly grounded, power line interference may even
completely obscure the ECG waveform.
❖ The most common cause of 50 Hz interference is the disconnected electrode resulting in a very
strong disturbing signal.
5
❖ Electromagnetic interference from the power lines also results in poor
quality tracings. Electrical equipment such as air-conditioners and X-ray
units draw heavy power-line current, which induce 50 Hz signals in the
input circuits of ECG machines
❖ This noise is very difficult to remove except in those machines which have
very high CMRR.
❖ A practical solution to minimize this problem is physical separation
between the interference causing sources and the patient.
■ The noise can be reduced by partially shielding the patient
6
a) ECG recording with regular spreading of the
curve with super-imposed 50Hz power line
interference signals.
b)Recording with irregular trembling of the ECG
trace with wandering of the base line but
otherwise normal ECG trace.
c) ECG trace without wandering of the baseline.
7
Problem solved by:
■ Use of transient suppressor in the mains lead of the machines helps to
solve this problem.
8
Shifting of the Baseline
❖ A wandering baseline [Fig. (b)] but otherwise normal ECG trace is usually due to the
movement of the patient or electrodes.
❖ The baseline shift can be eliminated by ensuring that the patient lies relaxed and the
electrodes are properly attached.
❖ Baseline wander is usually observed immediately after application of the electrodes. It is due
to a relatively slow establishment of electrochemical equilibrium at the electrode-skin
interface.
❖ This can be minimized by selecting the proper electrode material, which will reach
equilibrium quickly with a good electrode jelly.
9
Muscle Tremor
❖ Irregular trembling of the ECG trace [Fig.(c)], without wandering of the baseline occurs when
the patient is not relaxed or is cold.
❖ It is generally found in the case of older patients. Muscle tremor signals are especially
bothersome on limb leads when a patient moves or the muscles are stretched. Therefore, for
long-term monitoring, the electrodes are applied on the chest and not on the limbs.
❖ For normal routine ECG recordings, the patient must be advised to get warm and to relax so
that muscle tremor from shivering or tension is eliminated.
❖ The most critical component of the ECG recorder is the patient cable. The conventional PVC
insulation gets degraded and becomes rigid and breakable because of the aerification of the
softener.
10
Problem solved by:
❖ Some manufacturers supply a patient-cable made of silicon-rubber, which
provides better elasticity over long periods.
11
12
13
14

More Related Content

What's hot

Components of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardwareComponents of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardwareRaghu Kishore Galla
 
PACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaPACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaDrSravaniVishnubhatl
 
ECG machines -Operation and Maintenance
ECG machines -Operation and MaintenanceECG machines -Operation and Maintenance
ECG machines -Operation and Maintenanceshashi sinha
 
Cardiac Pacemakers
Cardiac Pacemakers Cardiac Pacemakers
Cardiac Pacemakers Anirudhya J
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiologyjmlafroscia
 
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarIImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarPrithvi Puwar
 
various type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removevarious type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removeManish Kumar
 
Electrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentElectrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentGhansyam Rathod
 
Biomedical Signals
Biomedical SignalsBiomedical Signals
Biomedical SignalsKrish599909
 
Implantable Cardiac Devices
Implantable Cardiac Devices Implantable Cardiac Devices
Implantable Cardiac Devices Mohamed Elwakil
 
ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)Anwar Siddiqui
 

What's hot (20)

cardiac output measurment
cardiac output measurmentcardiac output measurment
cardiac output measurment
 
Ecg ppt
Ecg pptEcg ppt
Ecg ppt
 
Holter
HolterHolter
Holter
 
Components of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardwareComponents of Pacemaker and ICDs - understanding the hardware
Components of Pacemaker and ICDs - understanding the hardware
 
PACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaPACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani Vishnubhatla
 
EEG
EEGEEG
EEG
 
PROSTHETIC VALVES
PROSTHETIC VALVESPROSTHETIC VALVES
PROSTHETIC VALVES
 
Ecg Signal Processing
Ecg Signal ProcessingEcg Signal Processing
Ecg Signal Processing
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
ECG machines -Operation and Maintenance
ECG machines -Operation and MaintenanceECG machines -Operation and Maintenance
ECG machines -Operation and Maintenance
 
Holter monitoring
Holter monitoringHolter monitoring
Holter monitoring
 
Cardiac Pacemakers
Cardiac Pacemakers Cardiac Pacemakers
Cardiac Pacemakers
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
 
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarIImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
 
various type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removevarious type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's remove
 
Electrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentElectrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical Equipment
 
Biomedical Signals
Biomedical SignalsBiomedical Signals
Biomedical Signals
 
Implantable Cardiac Devices
Implantable Cardiac Devices Implantable Cardiac Devices
Implantable Cardiac Devices
 
ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)ELECTROCARDIOGRAPHY (ECG)
ELECTROCARDIOGRAPHY (ECG)
 
blood flow measurement
blood flow measurementblood flow measurement
blood flow measurement
 

Similar to ARTIFACTS ON ECG RECORDING

CM ECG Recorder pp.pptx
CM ECG Recorder pp.pptxCM ECG Recorder pp.pptx
CM ECG Recorder pp.pptxMostafaSbaa
 
Final Documentation - Full
Final Documentation - FullFinal Documentation - Full
Final Documentation - FullEmeel Gayed
 
Electroencephalography (EEG)
Electroencephalography (EEG)Electroencephalography (EEG)
Electroencephalography (EEG)aju remesh
 
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition SystemRight-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition SystemJewel Haque
 
BME 311 Lecture Note 4-ECG Systems_11.pptx
BME 311  Lecture Note 4-ECG Systems_11.pptxBME 311  Lecture Note 4-ECG Systems_11.pptx
BME 311 Lecture Note 4-ECG Systems_11.pptxAbhishekAnand286155
 
cardiac monitoring.pptx
cardiac monitoring.pptxcardiac monitoring.pptx
cardiac monitoring.pptxRaiRai72
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) ijceronline
 
She training module electrical safety
She training module electrical safetyShe training module electrical safety
She training module electrical safetysunilkumarkushwaha12
 
ELECTROCARDIOGRAM. pptx
ELECTROCARDIOGRAM. pptxELECTROCARDIOGRAM. pptx
ELECTROCARDIOGRAM. pptxSn Fatima
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Manu Mitra
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slidesmaltiaziz24
 
Telemedicine System For Cardiac Patients
Telemedicine System For Cardiac PatientsTelemedicine System For Cardiac Patients
Telemedicine System For Cardiac PatientsSharad Karwa
 
Ecg basics lecture 7 (The Last Lecture)
Ecg basics lecture 7 (The Last Lecture)Ecg basics lecture 7 (The Last Lecture)
Ecg basics lecture 7 (The Last Lecture)Dr. Abdulkadir Insi
 

Similar to ARTIFACTS ON ECG RECORDING (20)

Factors
FactorsFactors
Factors
 
CM ECG Recorder pp.pptx
CM ECG Recorder pp.pptxCM ECG Recorder pp.pptx
CM ECG Recorder pp.pptx
 
Final Documentation - Full
Final Documentation - FullFinal Documentation - Full
Final Documentation - Full
 
Ecg machine
Ecg machineEcg machine
Ecg machine
 
ECG1.pdf
ECG1.pdfECG1.pdf
ECG1.pdf
 
Electroencephalography (EEG)
Electroencephalography (EEG)Electroencephalography (EEG)
Electroencephalography (EEG)
 
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition SystemRight-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
Right-Leg-Driven Circuit and Instrumentation Amplifier in ECG Acquisition System
 
BME 311 Lecture Note 4-ECG Systems_11.pptx
BME 311  Lecture Note 4-ECG Systems_11.pptxBME 311  Lecture Note 4-ECG Systems_11.pptx
BME 311 Lecture Note 4-ECG Systems_11.pptx
 
Cardiac monitoring.pdf
Cardiac monitoring.pdfCardiac monitoring.pdf
Cardiac monitoring.pdf
 
unit i
 unit i unit i
unit i
 
cardiac monitoring.pptx
cardiac monitoring.pptxcardiac monitoring.pptx
cardiac monitoring.pptx
 
Sec 1 ecg
Sec 1 ecgSec 1 ecg
Sec 1 ecg
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
 
She training module electrical safety
She training module electrical safetyShe training module electrical safety
She training module electrical safety
 
ELECTROCARDIOGRAM. pptx
ELECTROCARDIOGRAM. pptxELECTROCARDIOGRAM. pptx
ELECTROCARDIOGRAM. pptx
 
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
Project Report on ECG Transmitter using Agilent ADS (Advance System Design)
 
Ecg.pptx
Ecg.pptxEcg.pptx
Ecg.pptx
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
 
Telemedicine System For Cardiac Patients
Telemedicine System For Cardiac PatientsTelemedicine System For Cardiac Patients
Telemedicine System For Cardiac Patients
 
Ecg basics lecture 7 (The Last Lecture)
Ecg basics lecture 7 (The Last Lecture)Ecg basics lecture 7 (The Last Lecture)
Ecg basics lecture 7 (The Last Lecture)
 

More from BharathasreejaG

IMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSIMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSBharathasreejaG
 
Blood pressure measurement
Blood pressure measurementBlood pressure measurement
Blood pressure measurementBharathasreejaG
 
Respiration rate measurement
Respiration rate measurement Respiration rate measurement
Respiration rate measurement BharathasreejaG
 
Temperature measurement
Temperature measurement Temperature measurement
Temperature measurement BharathasreejaG
 
Carrier and chopper amplifiers
Carrier and chopper amplifiers Carrier and chopper amplifiers
Carrier and chopper amplifiers BharathasreejaG
 
Types of bio amplifiers
Types of bio amplifiers Types of bio amplifiers
Types of bio amplifiers BharathasreejaG
 
Signal conditioning unit
Signal conditioning unit Signal conditioning unit
Signal conditioning unit BharathasreejaG
 
Bio signal characteristics and recording modes
Bio signal characteristics and recording modesBio signal characteristics and recording modes
Bio signal characteristics and recording modesBharathasreejaG
 
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMSMEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMSBharathasreejaG
 
Electrodes in medical field
Electrodes in medical fieldElectrodes in medical field
Electrodes in medical fieldBharathasreejaG
 
BASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATIONBASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATIONBharathasreejaG
 

More from BharathasreejaG (12)

IMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSIMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERS
 
Pulse rate measurement
Pulse rate measurement Pulse rate measurement
Pulse rate measurement
 
Blood pressure measurement
Blood pressure measurementBlood pressure measurement
Blood pressure measurement
 
Respiration rate measurement
Respiration rate measurement Respiration rate measurement
Respiration rate measurement
 
Temperature measurement
Temperature measurement Temperature measurement
Temperature measurement
 
Carrier and chopper amplifiers
Carrier and chopper amplifiers Carrier and chopper amplifiers
Carrier and chopper amplifiers
 
Types of bio amplifiers
Types of bio amplifiers Types of bio amplifiers
Types of bio amplifiers
 
Signal conditioning unit
Signal conditioning unit Signal conditioning unit
Signal conditioning unit
 
Bio signal characteristics and recording modes
Bio signal characteristics and recording modesBio signal characteristics and recording modes
Bio signal characteristics and recording modes
 
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMSMEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
 
Electrodes in medical field
Electrodes in medical fieldElectrodes in medical field
Electrodes in medical field
 
BASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATIONBASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATION
 

Recently uploaded

9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room servicediscovermytutordmt
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxSayali Powar
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...Sapna Thakur
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
 
The byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptxThe byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptxShobhayan Kirtania
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxiammrhaywood
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 

Recently uploaded (20)

9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room service
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
 
The byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptxThe byproduct of sericulture in different industries.pptx
The byproduct of sericulture in different industries.pptx
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 

ARTIFACTS ON ECG RECORDING

  • 2. ARTIFACTS ■ Artifact on the electrocardiogram can result from a variety of internal and external causes. ■ Artifact and troubleshooting the problem will be straight forward. ■ When artifact mimics ECG abnormalities that can cause problems for patient care. 2
  • 4. EFFECTS OF ARTIFACTS ON ECG RECORDINGS (OR) POWER LINE INTERFERENCE Power line interference Shifting of the Baseline Muscle Tremor 4
  • 5. Interference from Power Line ❖ Power line interference is easily recognizable since the interfering voltage in the ECG ❖ This interference may be due to the stray effect of the alternating current on the patient or because of alternating current fields due to loops in the patient cable ❖ Other causes of interference are loose contacts on the patient cable as well as dirty electrodes. ❖ When the machine or the patient is not properly grounded, power line interference may even completely obscure the ECG waveform. ❖ The most common cause of 50 Hz interference is the disconnected electrode resulting in a very strong disturbing signal. 5
  • 6. ❖ Electromagnetic interference from the power lines also results in poor quality tracings. Electrical equipment such as air-conditioners and X-ray units draw heavy power-line current, which induce 50 Hz signals in the input circuits of ECG machines ❖ This noise is very difficult to remove except in those machines which have very high CMRR. ❖ A practical solution to minimize this problem is physical separation between the interference causing sources and the patient. ■ The noise can be reduced by partially shielding the patient 6
  • 7. a) ECG recording with regular spreading of the curve with super-imposed 50Hz power line interference signals. b)Recording with irregular trembling of the ECG trace with wandering of the base line but otherwise normal ECG trace. c) ECG trace without wandering of the baseline. 7
  • 8. Problem solved by: ■ Use of transient suppressor in the mains lead of the machines helps to solve this problem. 8
  • 9. Shifting of the Baseline ❖ A wandering baseline [Fig. (b)] but otherwise normal ECG trace is usually due to the movement of the patient or electrodes. ❖ The baseline shift can be eliminated by ensuring that the patient lies relaxed and the electrodes are properly attached. ❖ Baseline wander is usually observed immediately after application of the electrodes. It is due to a relatively slow establishment of electrochemical equilibrium at the electrode-skin interface. ❖ This can be minimized by selecting the proper electrode material, which will reach equilibrium quickly with a good electrode jelly. 9
  • 10. Muscle Tremor ❖ Irregular trembling of the ECG trace [Fig.(c)], without wandering of the baseline occurs when the patient is not relaxed or is cold. ❖ It is generally found in the case of older patients. Muscle tremor signals are especially bothersome on limb leads when a patient moves or the muscles are stretched. Therefore, for long-term monitoring, the electrodes are applied on the chest and not on the limbs. ❖ For normal routine ECG recordings, the patient must be advised to get warm and to relax so that muscle tremor from shivering or tension is eliminated. ❖ The most critical component of the ECG recorder is the patient cable. The conventional PVC insulation gets degraded and becomes rigid and breakable because of the aerification of the softener. 10
  • 11. Problem solved by: ❖ Some manufacturers supply a patient-cable made of silicon-rubber, which provides better elasticity over long periods. 11
  • 12. 12
  • 13. 13
  • 14. 14