Oscilloscope Lab
Introduction:
The purpose of this lab is to introduce students to the basic tools used by engineers and technicians
in analyzing electronic equipment: the function generator, the analog oscilloscope, and the digital
oscilloscope. The oscilloscope is a tool commonly used by engineers and technicians to analyze and
troubleshoot electronic systems. In addition, real signals, such as those which carry voice or data,
can be very complex and difficult to analyze. The function generator is a device used to generate an
electronic signal with specific known characteristics, thereby enabling an engineer or technician to
test and examine a circuit.
During this lab, the student will use the function generator to generate a number of signals and to
analyze those signals using either of the ā€˜oscilloscopesā€™. The student will become familiar with the
basic waveforms -- sine, square, and triangle waves -- and the components of the waveforms --
amplitude, period, and frequency. At the conclusion of this lab the student should feel comfortable
using the function generator and the oscilloscopes.
Equipment:
Qty Equipment
1 Leader LFG-1300S Function Generator
1 BNC to 2 alligator clips cable
1 Tektronix 2225 Analog Oscilloscope
1 Tektronix P6103 10X probe
1 Hewlett-Packard 54502A Digital Oscilloscope
1 HP 10430A 10X probe
Oscilloscope Lab - Rev 01-05-95 1
Equipment Introduction:
Part I - Introduction to the Function Generator
Overview:
The function generator is used to generate a wide range of alternating-current (AC) signals. A
diagram of the Leader LFG-1300S Function Generator is shown below in Figure 1.
ā€¢ The front panel is divided into six major control groups:
1) Frequency Selection Group;
2) Sweep Group;
3) Amplitude Modulation Group;
4) DC Offset Group;
5) Function, or Waveform Group; and
6) Output Group.
POWER ON
SWEEP AMP MOD DC OFFSET OUTPUT
ATTENUATION
dB
FUNCTION
FREQUENCY
LEADER LFG-1300S FUNCTION GENERATOR
Hz
Figure 1. Front Panel of Function Generator
ā€¢ The power switch on the upper left-hand corner of the unit. The green LED will indicate that
the unit is on.
ā€¢ The three most important groups for this lab are the frequency, function, and output
groups. The remaining three groups, the sweep, amplitude modulation, and the DC offset
groups, will be briefly covered in the lab setup procedures. Should the student desire more
detailed descriptions of these groups, the Leader Function Generator manual is available in
the lab.
Oscilloscope Lab - Rev 01-05-95 2
Frequency Selection Group:
These controls are used select the operating frequency of the function generator. This group
consists of the frequency control knob and the eight frequency multiplier selection buttons.
For example, to set the function generator to an operating frequency of 2000 Hz (2 kHz):
ā€¢ Rotate the frequency control knob to 2.
ā€¢ Select the 1 kHz frequency multiplier button.
With the result that: 2.0 * 1 kHz = 2.0 kHz.
Example two sets the function generator to an operating frequency of 5.5 kHz:
ā€¢ Rotate the frequency control knob to 0.55.
ā€¢ Select the 10 kHz frequency multiplier button.
With the result that: 0.55 * 10 kHz = 5.5 kHz.
Output Group:
1. These controls are used to adjust the amplitude of the generator's output signal. The group
consists of the amplitude-control knob, the three attenuation buttons and the fused 50 ohm BNC
connector.
ā€¢ Although the amplitude knob is not indexed, the amplitude ranges from a few millivolts to
approximately 20 volts.
ā€¢ We will set the amplitude levels by aligning the white line on the amplitude knob to the three
o'clock position (90 degrees right), the nine o'clock position (90 degrees left), or the twelve
o'clock position (straight up).
ā€¢ Notice that rotating the knob fully to the left does not result in a zero amplitude signal.
ā€¢ The attenuation buttons are used to attenuate (decrease) the amplitude of the signal by a
factor measured in decibels. The following relationship will assist in working with the
attenuation buttons:
(dB) = -10 * log10 (Pout/Pin) (if power is the unit of measurement)
or
(dB) = -20 * log10 (Vout/Vin) (if voltage is the unit of measurement)
ā€¢ Example:
Given a 1 volt input signal, what is the change in voltage amplitude if the 10 dB
attenuation button is depressed?
Oscilloscope Lab - Rev 01-05-95 3
Beginning with the equation defined above:
(dB) = -20 * log10 (Vin/Vout)
10dB = -20 * log10 (Vin/Vout)
-0.5 dB = log10 (Vin/Vout)
10
-.5
= Vout/Vin
Vout/Vin = 0.3162
ā€¢ From here, the output voltage can be stated in terms of the input voltage and vice-versa:
Vin = Vout/0.3162
or
Vout = Vin * 0.3162
Since the initial input voltage (Vin) was 1 volt the output voltage (Vout) must be
0.3162V or 316 millivolts.
ā€¢ Note: The attenuation buttons are additive. In other words, if the 10 dB and the 20 dB
buttons are both pressed in, the combined attenuation of the input signal is 30 dB.
2. The fused 50 ohm BNC connector is used for connecting the function generator to other
equipment. The most common connection used to connect the function generator to other
equipment is a cable with a BNC connector on one end and two alligator clips on the other end.
ā€¢ One end of the cable, the base, is used to connect the cable to the BNC connector on the
function generator. The other end, which has two ā€˜alligatorā€™ clips: one positive (red) and one
negative (black), is used to connect the function generator to the device or circuit under
study.
ā€¢ To connect the cable to the function generator:
1. Hold the cable's BNC connector, the base, in one hand.
2. Line up the two grooves on the cable's BNC connector with the two pegs on the
outside of the function generator's BNC connector.
3. Push the cable's BNC connector over the function generator's BNC connector until it
is fully seated.
4. Rotate the cable's BNC connector in a clockwise manner until it is firmly secured to
the function generator's BNC connector.
Oscilloscope Lab - Rev 01-05-95 4
Function/Waveform Selection Group:
This group is used to select the shape of the generated waveform. The group is made up of the six
wave-selector buttons. The six waveforms that the function generator can produce are the sine
wave, the square wave, the triangle wave, two sawtooth waves, and the variable-width pulse
wave.
reference
reference
SINE WAVE TRIANGLE WAVE SQUARE WAVE
SAWTOOTH
PULSE
SAWTOOTH
Figure 2: Available Generated Waveforms
ā€¢ In this lab you will become familiar with the three fundamental signals (waveforms),
the sine wave, the square wave and the triangle wave.
Waveforms:
This section will provide you with some background on waveforms and on measuring waveforms.
It is especially important that you understand how to measure waveforms. This is a skill that is not
only important to this lab, but also to the T-1 Lab and in labs for other Telecommunications
Courses.
ā€¢ There are two types of alternating-current signals, which are also called waveforms:
ā€¢ Periodic
ā€¢ Aperiodic
ā€¢ Periodic signals, such as sine or triangle waves, behave in a uniform manner and repeat
themselves over a given length of time. Each repetition of a repeating signal is called a
period or a cycle.
Oscilloscope Lab - Rev 01-05-95 5
ā€¢ Aperiodic signals, such as analog voice, behave in a non-uniform manner and do not repeat
themselves over any given length of time.
Note: In this lab, we will only work with periodic signals.
ā€¢ When working with AC signals, there are three properties of the signal that we are concerned
with: amplitude, period and frequency.
ā€¢ The amplitude of the wave is defined as the maximum magnitude of the wave. The amplitude
is the vertical component of the signal and is measured in units of volts (V). Since we are
dealing with an AC signal, the voltage will change over a period of time. The maximum
voltage of a signal during its cycle is commonly referred to as the peak voltage (Vp).
ā€¢ The amplitude can be measured from the reference line to the peak (Vp) or from peak-to-peak
(Vpp). See Figure 3 below.
peak-to-peak
amplitude
peak or
maximum
amplitude
reference point
Figure 3: Measuring Amplitude.
ā€¢ With a periodic signal that is symmetrical (equi-distant above and below the reference point),
the peak-to-peak voltage is equal to twice the peak voltage:
Vpp = 2 * Vp
Oscilloscope Lab - Rev 01-05-95 6
ā€¢ The period (T) of the signal is defined as the time it takes for a signal to complete one full
cycle. The period is the horizontal component of the signal, measured in units of seconds (s).
In Figure 4, the period of the signal is measured as 250 milliseconds (250.0 x 10
-3
s).
P E R I O D
o n e s e c o n d
2 5 0 m s e c 2 5 0 m s e c 2 5 0 m s e c 2 5 0 m s e c
Figure 4: Measuring Period.
ā€¢ The frequency (f) of the signal is defined as the rate at which a periodic signal repeats. It is
usually measured in units of Hertz (Hz), where 1 Hz = 1 cycle per second.
ā€¢ In Figure 4, above, you can see four(4) cycles occurring within one second; therefore, the
signal has a frequency of 4 Hz.
ā€¢ The frequency, f, of a wave is inversely related to its period (T):
f = 1/T
ā€¢ Example:
The period of the signal is 250 milliseconds, therefore the frequency of that signal is:
f = 1/T
f = 1/250 milliseconds
f = 4 hertz
Abbreviation Prefix Name Factor
T Tera 1012
G Giga 109
M Mega 106
K Kilo 103
m milli 10-3
Āµ micro 10-6
n nano 10-9
p pico 10-12
Table of Scientific Prefixes
Oscilloscope Lab - Rev 01-05-95 7
PART II - Introduction to the Tektronix 2215 Analog Oscilloscope
Overview:
An oscilloscope is an electronic measuring device which provides a two-dimensional visual
representation of a signal. Because the oscilloscope allows the user to see the signal(s), their
characteristics can be easily measured and observed. The oscilloscope displays a graph of voltage
(on the vertical axis) versus time (on the horizontal axis). Most electrical circuits can be easily
connected to the oscilloscope typically with probes.
ā€¢ There are four major control groups on the Tektronix oscilloscope: (1) the Display group;
(2)the Vertical group; (3) the Horizontal group; and (4) the Trigger group. See Figure 5,
below.
INTENSITY
BEAM FIND
FOCUS
POWER
TRACE SEP
POSITION POSITION
CH1 VOLTS/DIV CH2 VOLTS/DIV
CH! BOTH CH1 ADD ALT CHOP
COURSE FINE
5X 10X 50X
SEC/DIV
NORM INVERT
AC GND DC AC GND DC
SOURCE
COUPLING
1X ALT MAG
CH1 CH2 EXT
LEVEL
MODE
HOLDOFF
VERTICAL HORIZONTAL TRIGGER
TEKTRONIX 2225
Figure 5: Front Panel of Analog Oscilloscope
Display Group
This group is used to display and adjust the signal for optimal viewing. It consists of the display
screen, the intensity-control knob, the beam-find button, the focus-control knob, and the power
switch.
ā€¢ The display screen is laid out in a 8 by 10 centimeter grid. The oscilloscope draws a ā€˜traceā€™
or graph by moving an electron beam across the phosphor coating on the inside of the
cathode-ray tube (CRT). The excited phosphorous glows for a short period of time, thereby
tracing the path of the beam.
ā€¢ The intensity-control knob is used to adjust the brightness of the trace. The level should be
set to the user's preference; however, increasing the intensity beyond a certain point will make
the trace "fuzzy." The intensity level should never be increased past the point where the trace
has sharply defined edges.
Oscilloscope Lab - Rev 01-05-95 8
ā€¢ The beam-find button allows the user to locate the electron beam anytime it's off-screen.
Push the beam find button to temporarily reduce the vertical and horizontal deflection
voltages so that the beam always appears within the 8 by 10 centimeter screen.
ā€¢ The focus-control knob adjusts the electron beam for optimal trace resolution.
Vertical Group:
This group is used to adjust the vertical components (Y-axis) and the vertical position of the signal.
This group consists of the vertical-position knobs, channel-selector switch, volts-per-division-
selector knobs, input-coupling switch, the channel-mode-selector switch, the channel-2-invert
switch, and the BNC connectors. Because the Tektronix 2225 oscilloscope is a two channel
oscilloscope, there is one set of switches for each channel.
ā€¢ The vertical-position controls are used to vertically move the trace of one channel or the
other.
ā€¢ The channel-selector switch, labeled ā€˜CH1/BOTH/CH2ā€™, selects which channels are
displayed on the screen.
ā€¢ The volts-per-division-selector knob sets the vertical scale for each channel's trace.
1. Since a 10X probe is used in this lab, all readings should be made from the ā€˜10Xā€™
box on the knob.
2. Each channel can have a different vertical scale.
3. A division is one "block" on the screen.
4. Each tick mark is one-fifth (0.2) of a division.
5. This knob is marked in both volts and millivolts.
ā€¢ The input-coupling switch, labeled ā€˜AC/GND/DCā€™, selects the coupling mode of that
channel's display. AC means that only the alternating portion of the signal is displayed. DC
will display both the alternating portion of the signal, plus any DC component. GND shows
the 0 V reference level.
ā€¢ The trace-separation knob is used in conjunction with the horizontal-magnification
controls, to be discussed later.
ā€¢ The channel-mode-selector switch, labeled ā€˜ADD/ALT/CHOPā€™, is activated only when
BOTH is selected on the channel-selector switch. ADD graphically adds the Channel 1
signal to the Channel 2 signal. If the channel-2-invert switch (ā€˜NORMAL/INVERTā€™) is set
to INVERT, then ADD will in fact subtract the Channel 2 signal from the Channel 1 signal.
ALT traces one channel, then the next. CHOP works like ALT, but jumps back and forth
between the two channels during a single trace.
Oscilloscope Lab - Rev 01-05-95 9
Horizontal Control Group:
The Horizontal Control Group consists of the coarse and fine position knobs, the horizontal-
magnification switch, the seconds-per-division-selector knob, and the magnification-scale-
selector switch.
ā€¢ The coarse and fine-position-knobs allow the horizontal movement of the traces in a rough
manner and in a precise manner, respectively. These are used to position the traces in a
manner that makes measurement both more convenient and more precise.
The horizontal-magnification switch (ā€˜X1/ALT/MAGā€™) selects the regular (X1) and/or a
horizontally magnified trace.
1. The trace separation knob in the Vertical-Control Group allows for the vertical
separation of the X1 trace from the magnified trace when ALT is selected.
2. If ALT or MAG is selected, then the magnification-scale-selector switch
(ā€˜5X/10X/50Xā€™) is activated.
ā€¢ The seconds-per-division knob sets the time base (horizontal) scale. It is marked in
seconds, milliseconds, and microseconds.
Note: There is only one horizontal scale for both channels.
Trigger Group
In order to display a signal, the oscilloscope must be able to ā€˜lockā€™ onto that signal; the function of
the triggering controls is to do just that. Triggering can be a complicated topic, and is beyond the
scope of this lab. For more details, the student is referred to the Tektronix manual. The trigger
group is made up of the trigger level knob, the rising/falling edge switch, the trigger mode
switch, the holdoff knob, the trigger source switches, and the trigger coupling switch.
ā€¢ The trigger-level knob sets the voltage level at which the oscilloscope will ā€˜triggerā€™. If a
signal is ā€˜runningā€™ -- that is, not stable -- the trigger level may be too high or too low for the
oscilloscope to recognize the signal. Often you can lock in a running signal by rotating this
knob to the left and to the right. This adjustment provides a mechanism for ignoring small
(low voltage) signals that are well below the level of the signal you ar interested in. This
knob is not indexed.
ā€¢ The rising/falling-edge switch selects whether the oscilloscope will trigger on the positive
(rising) or negative (falling) edge of the signal.
ā€¢ The trigger-mode switch will normally be set to AUTO, but sometimes it is necessary to use
NORM. Other settings are not relevant to this lab.
ā€¢ The holdoff knob affects the delay associated with triggering. The student should consult
the manual if more information is desired.
Oscilloscope Lab - Rev 01-05-95 10
ā€¢ The trigger-source switches select which signal the oscilloscope will attempt to lock onto.
Possible choices include CH1, CH2, VERT MODE, or EXTERNAL. Selecting CH1 or
CH2 will make the oscilloscope attempt to trigger on those channels. If no input is available
on that channel, there may be problems when attempting to view both channels
simultaneously. For this reason, it is recommended that this switch be left on VERT MODE,
which provides an automatic trigger on either CH1 or CH2.
ā€¢ The trigger-coupling switch is another more complicated feature which will not be discussed
here. This switch should typically be set to AC.
How To Make Readings on the Oscilloscope:
ā€¢ Peak-to-Peak Voltage:
Use the vertical-position knob to place a peak (positive or negative) on a horizontal
line, keeping the peak on the screen.
Use the horizontal-position knob to set the next (opposite sign) peak on the center
vertical line.
Count the number of divisions between the positive and negative peaks.
Note: The divisions are further subdivided into 1/5 (0.2) by the tick marks.
Multiply the number of divisions from step 3 by the volts/div setting for the channel in
use.
ā€¢ Period:
Use the horizontal-position knobs to align any edge of the signal with a vertical line.
Use the vertical-position knob to place the next identical edge crossing on the X-axis.
Count the number of divisions along the horizontal line to the next crossing in the same
direction.
Multiply the number of divisions from step 3 by the sec/div setting.
Note: The divisions are subdivided into fifths (0.2) by the ā€˜tick marksā€™.
ā€¢ Frequency:
Measure the period of the signal (T).
Calculate frequency (f) using f = 1/T.
Oscilloscope Lab - Rev 01-05-95 11
Figure 6: Measuring a Signal.
ā€¢ Example:
Assuming the Volts/Div knob reads 2 V/div, the above peak-to-peak voltage would be:
Vpp = 2 volts/div * 4.2 div = 8.4 volts
Assuming the Sec/Div control knob read 50ms, the above period would be:
T = 50 milliseconds/div * 5.25 div = 2625.0 milliseconds = 2.625 seconds
Oscilloscope Lab - Rev 01-05-95 12
PART III - Introduction to the HP 5402A Digital Oscilloscope
Overview:
The digital oscilloscope is a test and measuring device, similar to the analog oscilloscope. Instead
of displaying a signal by deflecting an electron beam that is tracing across a CRT, the digital
oscilloscope uses a processor to sample, digitize, and display the incoming analog signal on the
display screen. The digital oscilloscope has a user interface which allows the user to manipulate
and measure the displayed signal. It can be used to store, retrieve, and print waveforms. It can also
be programmed to make measurements automatically or to control other devices. Because the HP
5402A digital oscilloscope is a multi-purpose piece of equipment, it would be very time consuming
to explain out all of its functionality. It is strongly recommended that the student become familiar
with the following reference manuals if more details are needed:
ā€¢ HP 54502A 400 MHz Digitizing Oscilloscope Front-Panel Reference
ā€¢ HP 54502A 400 MHz Digitizing Oscilloscope Programming Reference
These references can be found at the lab tables in the Telecommunications Lab (Room 834).
Front Panel Reference:
The HP 54502A 400 MHz Digitizing Oscilloscope controls are organized into six functional areas
or groups: (1) the Display Group, (2) the System-Control Group, (3) the Menu Group (4) the
Setup Group, (5) the Entry Group, and (6) the Signal-Input Group.
HEWLETT 54502 A 400 MHz
PACKARD DIGITIZING OSCILLOSCOPE
SYSTEM CONTROL SETUP
MENUS
ENTRY
DISPLAY
INPUTS
Figure 7: Front Panel of Analog Oscilloscope
The power switch is a rocker switch located at the back of the machine, immediately above the
power cord.
Oscilloscope Lab - Rev 01-05-95 13
ā€¢ The display group consists of the display screen and the programmable menu selection
buttons immediately to the right of the screen. The function of these buttons is dependent on
the menu selected. An option appearing on the screen may be selected by striking the button
to the right of it.
ā€¢ The system-control group includes the RUN/STOP and the SINGLE buttons, which
control the screen refreshing, and the CLEAR DISPLAY button which clears the screen.
ā€¢ The menu group contains a selection of menus whose function is similar to the front panel
controls on the analog oscilloscope. The procedures below will serve to familiarize you with
the similarities. There is also a āˆ†āˆ†t/āˆ†āˆ†v menu which is helpful in making measurements of
voltage, period, and frequency.
ā€¢ The setup group includes the AUTOSCALE button, which resets the oscilloscope to
default settings and automatically adjusts horizontal and vertical scales for the signal that is
input.
ā€¢ The entry group is used to enter data. The numeric keypad is augmented with four unit
selection buttons, which give a selection of seconds, milliseconds, microseconds,
nanoseconds, volts, and millivolts. Alternatively, the data entry wheel can be used to adjust
settings instead of entering a specific value. The entry group also contains a number of
automatic measurement functions, which are accessed using the blue shift button.
ā€¢ The signal-input group consists of the BNC connectors for each channel, and an auxiliary
connector.
Making Measurements with the Digital Oscilloscope:
There are two ways to make measurements with the HP digital oscilloscope: (1) by using the
āˆ†t/āˆ†v markers, or (2) by using the automatic measurements functions. The procedures for both
these methods are outlined in the lab procedures.
Oscilloscope Lab - Rev 01-05-95 14
Lab Procedures:
Part I - Analog Oscilloscope Procedures:
Overview:
In this part of the lab, you will use the function generator to generate a signal and then use the
analog oscilloscope to make some measurements of the data. If you need any assistance in
identifying or configuring the equipment, please see the on-duty GSA.
Function Generator Setup:
ā€¢ Turn on the function generator.
ā€¢ Make sure that the Sweep, Amplitude Modulation, and DC offset groups are turned OFF.
Note: The Sweep and Amplitude Modulation buttons should be in the OUT position, and the DC
offset knob should be pushed IN.
ā€¢ Select a TRIANGLE wave.
ā€¢ Set frequency to 1.0 kHz.
ā€¢ Set the amplitude knob to 12 o'clock.
ā€¢ Make sure all the attenuation buttons are OUT.
ā€¢ Attach the BNC end of the cable to the BNC socket in the function generator's Output
Group.
ā€¢ Tie back the black lead alligator clip by clipping the black alligator clip to the insulated wire
to avoid short circuits and blown fuses.
Analog Oscilloscope Setup:
ā€¢ Turn on the power to the oscilloscope.
ā€¢ Connect the Tektronix 10X probe to the Channel 1 input on the oscilloscope.
ā€¢ Tie back the black lead on the probe, then expose the hook and clip it to the red alligator clip
on the function generator cable.
ā€¢ Set CH1/BOTH/CH2 switch to CH1.
ā€¢ In the Vertical Control Group, set the Channel 1 AC/GND/DC switch to GND.
Oscilloscope Lab - Rev 01-05-95 15
ā€¢ Use the CH1 vertical position knob to move the ground (0V) reference (horizontal line) to the
center line on the screen.
ā€¢ After referencing the signal to ground, set the AC/GND/DC switch to AC.
ā€¢ Intensity knob: adjust the signal intensity and focus to a comfortable level, by using the
intensity and focus knobs, respectively.
Note: Keep the signal intensity to within a reasonable range to minimize the chances of
burning the phosphor and damaging the display screen.
ā€¢ Set the Volts-per-Division knob to 2 V/div.
Note: you are using a 10X probe, so be sure to take all readings from the 10X position.
ā€¢ Set the NORMAL/INVERT switch to NORMAL.
ā€¢ Set the ADD/ALT/CHOP switch to ALT.
ā€¢ Set X1/ALT/MAG to X1.
ā€¢ Set the seconds-per-division knob to 0.1 ms/div.
ā€¢ Set the rising/falling-edge switch to positive.
ā€¢ Set the trigger-mode switch to AUTO.
ā€¢ Set the trigger-source switch to VERT MODE.
ā€¢ Set the trigger-coupling switch to AC.
Note: If the signal is still running, try adjusting the trigger level or holdoff knobs.
ā€¢ Make sure that the 'cal' knobs on the V/div and S/div knobs are both pushed in and turned all
the way to the right. They will click into position. This is so the scope will be calibrated
properly.
Oscilloscope Lab - Rev 01-05-95 16
Procedures:
1. Using the settings completed above, answer the following questions.
ā€¢ (1.1) What is the setting on the Volts/Div control knob? _______ volts/div
ā€¢ (1.2) How many vertical divisions from peak-to-peak? _______ div
ā€¢ (1.3) What is the peak to peak voltage (Vpp)?
Vpp = _______ volts/div * _______ div = _______ volts
ā€¢ (1.4) What is the setting on the Sec/Div control knob? _______ ___ seconds/div
ā€¢ (1.2) How many horizontal divisions from positive going crossing to positive going crossing?
_______ div
ā€¢ (1.5) What is the period of the signal (T)?
T = _______ seconds/div * _______ div = _______ ___ seconds
ā€¢ (1.6) What is the frequency of the signal (f)? _______ ___ hertz
ā€¢ (1.7) Draw the displayed signal on Graph 1.
Be neat, to scale, and concise.
Be sure to note the (scale) V/div and Sec/div settings.
Graph 1
Oscilloscope Lab - Rev 01-05-95 17
2. Next you will adjust the function generator to output another signal and repeat the measurements
in Step 1.
ā€¢ Generate a square wave between 5 and 10 kHz with an amplitude setting of 3 o'clock.
ā€¢ Adjust the V/div and Sec/div settings to maximize the display of the signal on the CRT. Make
sure you show the signal from peak-to-peak and at least one full cycle (period) of the signal.
ā€¢ (2.1) What is the setting on the Volts/Div control knob? _______ volts/div
ā€¢ (2.2) How many vertical divisions from peak-to-peak? _______ div
ā€¢ (2.3) What is the peak to peak voltage (Vpp)?
Vpp = _______ volts/div * _______ div = _______ volts
ā€¢ (2.4) What is the setting on the Sec/Div control knob? _______ ___ seconds/div
ā€¢ (2.2) How many horizontal divisions from positive going crossing to positive going crossing?
_______ div
ā€¢ (2.5) What is the period of the signal (T)?
T = _______ sec/div * _______ div = _______ ___ seconds
ā€¢ (2.6) What is the frequency of the signal (f)? _______ ___ hertz.
ā€¢ (2.7) Draw the displayed signal on Graph 2.
Be neat, to scale, and concise.
Be sure to note the (scale) V/div and Sec/div settings.
Graph 2
Oscilloscope Lab - Rev 01-05-95 18
3. In this part of the lab, you will experiment with the triggering of the analog oscilloscope. You
will need to make some changes to the controls of the analog oscilloscope and the function
generator, and you will then have to answer a few questions.
ā€¢ Set TRIGGER MODE to NORM.
Set TRIGGER SOURCE to CH1.
ā€¢ Turn the TRIGGER LEVEL knob all the way to the right.
Turn the TRIGGER HOLDOFF knob all the way to the right.
ā€¢ Generate a 1.4 kHz sine wave with the function generator; setting the amplitude knob all
the way to the left. Then display the signal on the oscilloscope.
ā€¢ On the oscilloscope, turn the horizontal position (coarse) knob until the left edge of the
trace is shown.
ā€¢ While watching the display, turn the TRIGGER LEVEL knob to the left until you get a
steady display.
ā€¢ Keep turning the TRIGGER LEVEL knob to the left, and watch the trigger level go down.
Note that the oscilloscope ceases to trigger when the level moves below the signal.
ā€¢ Set the TRIGGER LEVEL as close to the top of the signal as possible while still retaining a
steady picture.
ā€¢ (3.1) What is the peak-to-peak voltage (Vpp)? _______ volts
ā€¢ On the function generator, depress the 10 dB attenuation button.
ā€¢ Use the TRIGGER LEVEL knob to steady the display. It's a delicate adjustment, so watch
the screen carefully while you turn the knob to the left.
ā€¢ (3.2) Why did the signal run or disappear? (not steady or not locked)
ā€¢ (3.3) Without changing the V/div setting, what is the new Vpp? _______ volts
ā€¢ (3.4) Using the formula on page 3, calculate the attenuation in dB? _______dB (note sign!)
ā€¢ Disconnect the analog oscilloscope, and turn off the power.
Oscilloscope Lab - Rev 01-05-95 19
Part II - Digital Oscilloscope Procedures:
Overview:
In this part of the lab, you will use the function generator to generate a signal and then use the
digital oscilloscope to make some measurements of the data. If you need any assistance in
identifying or configuring the equipment, please see the on-duty GSA.
Digital Oscilloscope Setup:
ā€¢ Generate a 6kHz triangle wave with the amplitude turned fully to the right using the
function generator.
ā€¢ Connect the HP probe to the Channel 1 input on the oscilloscope, then connect the probe to
the red lead from the function generator.
ā€¢ Press the AUTOSCALE button to make the oscilloscope automatically adjust its settings.
ā€¢ Select the Channel menu by pressing the CHAN button.
Select Channel 1.
Set v/div to 5.0 v/div, if it is not already set.
Set offset to 0 volts, if not already set.
Select AC.
Select 1Mā„¦ā„¦, if not already set.
Select More by pressing the button to the right of the more label.
Make sure that 10:1 is selected. If not, use the wheel to set it.
Press the More button.
ā€¢ Select the Time base menu by pressing the TIME BASE button.
Set the time base to 50 ĀµĀµs/div, if not already set.
Set delay to 0 seconds.
Select reference: center, if not already set.
Set window: off, if it is on.
Select realtime.
ā€¢ Select the Trigger menu by pressing the TRIG button.
Select AUTO.
Select EDGE.
Select source: channel 1.
ā€¢ Select the Display menu by pressing the DISPLAY button.
Select NORM.
Set persistence: 'minimum' or 'single'.
Set No. of screens: 1.
Select GRID.
Set connect dots: ON.
Oscilloscope Lab - Rev 01-05-95 20
PROCEDURES:
1. Using the settings completed above, answer the following questions.
ā€¢ Select the āˆ†t/āˆ†v menu by pressing the āˆ†āˆ†t/āˆ†āˆ†v button.
ā€¢ Select Vmarkers and Tmarkers.
ā€¢ Select Vmarker 1, then use the data entry wheel to set the marker at the bottom of the signal.
Be sure to adjust the voltage level not the channel number.
ā€¢ Similarly, set Vmarker 2 to the top of the signal.
ā€¢ (1.1) Read āˆ†V at the bottom of the screen.
Vmarker 1: _______ ___ volts
Vmarker 2: _______ ___ volts
āˆ†V: _______ ___ volts
ā€¢ (1.2) Using the same procedure as for the Vmarkers, set the āˆ†T start marker to either a
positive or negative peak.
ā€¢ (1.3) Similarly, set the āˆ†T stop marker to the next positive or negative peak. (Choose the
same polarity as you did in Step 1.2.)
ā€¢ (1.4) Read āˆ†T and 1/āˆ†T (period and frequency) at the bottom of the screen.
Start Marker: _______ ___ seconds
Stop Marker: _______ ___ seconds
āˆ†T (period): _______ ___ seconds
1/āˆ†T (frequency): _______ ___hertz
ā€¢ Turn the markers off.
2. Using the automatic measurement capability of the digital oscilloscope, you will repeat the
measurements you made in Step 1.
ā€¢ (2.1) Press the blue shift button, press button 1 (Vpp), and then press button 1 for channel 1
when 'c#' appears at the bottom of the screen in reverse text.
ā€¢ (2.2) Measure Vpp: _______ ___ volts
Note: If 'm#' or 'f#' appears in reverse text at the bottom of the screen, use the entry
wheel to change this to 'c#'. If this fails, try hitting RECALL and CLEAR
simultaneously to reset the oscilloscope. You would then need to restart with the
pressing the AUTOSCALE in the digital oscilloscope setup section.
Oscilloscope Lab - Rev 01-05-95 21
ā€¢ (2.3) Press the blue shift button, press button 9 (Freq.), and then press button 1 for channel 1
when 'c#' appears at the bottom of the screen in reverse text.
Measure Frequency (f): _______ ___ hertz
ā€¢ (2.4) Press the blue shift button, press button s-V (period) to the right of button 9, and then
press button 1 for channel 1 when 'c#' appears at the bottom of the screen in reverse text.
Measure Period (T): _______ ___ seconds
ā€¢ (2.5) Draw the displayed signal on Graph 3.
Be neat, to scale, and concise.
Be sure to note the (scale) V/div and Sec/div settings.
Graph 3
ā€¢ To clear the measurements from the screen, press the blue shift button and then press the
clear button.
ā€¢ Disconnect the digital oscilloscope from the function generator and turn it off.
ā€¢ Make sure all probes and connectors are disconnected from the equipment and neatly placed
on the shelf above the work area.
ā€¢ Be sure to turn off both oscilloscopes and the function generator.
Oscilloscope Lab - Rev 01-05-95 22

More Related Content

What's hot

Communication Engineering - Chapter 6 - Noise
Communication Engineering - Chapter 6 - NoiseCommunication Engineering - Chapter 6 - Noise
Communication Engineering - Chapter 6 - Noisemkazree
Ā 
Frequency modulation2
Frequency modulation2Frequency modulation2
Frequency modulation2Sarah Krystelle
Ā 
Noise basics and its modelling
Noise basics and its modellingNoise basics and its modelling
Noise basics and its modellingDr Naim R Kidwai
Ā 
Introduction to Oscilloscope and Function generator
Introduction to Oscilloscope and Function generatorIntroduction to Oscilloscope and Function generator
Introduction to Oscilloscope and Function generatorcyberns_
Ā 
Ch4 2 _fm modulator and demodulator15
Ch4 2 _fm modulator and demodulator15Ch4 2 _fm modulator and demodulator15
Ch4 2 _fm modulator and demodulator15Jyothirmaye Suneel
Ā 
System noise temperature @ Satellite Link Design
System noise temperature  @ Satellite Link Design System noise temperature  @ Satellite Link Design
System noise temperature @ Satellite Link Design AJAL A J
Ā 
National college of science and technology
National college of science and technologyNational college of science and technology
National college of science and technologySarah Krystelle
Ā 
communication system Chapter 6
communication system Chapter 6communication system Chapter 6
communication system Chapter 6moeen khan afridi
Ā 
Understanding Noise Figure
Understanding Noise FigureUnderstanding Noise Figure
Understanding Noise FigureMostafa Ali
Ā 
Noise 2.0
Noise 2.0Noise 2.0
Noise 2.0bhavyaw
Ā 
An Introduction to RF Design, Live presentation at EELive 2014
An Introduction to RF Design, Live presentation at EELive 2014An Introduction to RF Design, Live presentation at EELive 2014
An Introduction to RF Design, Live presentation at EELive 2014Rohde & Schwarz North America
Ā 
Introduction
IntroductionIntroduction
IntroductionFyndi Aw
Ā 

What's hot (20)

Communication Engineering - Chapter 6 - Noise
Communication Engineering - Chapter 6 - NoiseCommunication Engineering - Chapter 6 - Noise
Communication Engineering - Chapter 6 - Noise
Ā 
Maala
MaalaMaala
Maala
Ā 
Cauan (2)
Cauan (2)Cauan (2)
Cauan (2)
Ā 
Frequency modulation2
Frequency modulation2Frequency modulation2
Frequency modulation2
Ā 
Noise basics and its modelling
Noise basics and its modellingNoise basics and its modelling
Noise basics and its modelling
Ā 
Introduction to Oscilloscope and Function generator
Introduction to Oscilloscope and Function generatorIntroduction to Oscilloscope and Function generator
Introduction to Oscilloscope and Function generator
Ā 
Ch4 2 _fm modulator and demodulator15
Ch4 2 _fm modulator and demodulator15Ch4 2 _fm modulator and demodulator15
Ch4 2 _fm modulator and demodulator15
Ā 
System noise temperature @ Satellite Link Design
System noise temperature  @ Satellite Link Design System noise temperature  @ Satellite Link Design
System noise temperature @ Satellite Link Design
Ā 
National college of science and technology
National college of science and technologyNational college of science and technology
National college of science and technology
Ā 
Equalisers
EqualisersEqualisers
Equalisers
Ā 
Cauan
CauanCauan
Cauan
Ā 
3 (3)
3 (3)3 (3)
3 (3)
Ā 
3 (2)
3 (2)3 (2)
3 (2)
Ā 
communication system Chapter 6
communication system Chapter 6communication system Chapter 6
communication system Chapter 6
Ā 
Understanding Noise Figure
Understanding Noise FigureUnderstanding Noise Figure
Understanding Noise Figure
Ā 
Noise 2.0
Noise 2.0Noise 2.0
Noise 2.0
Ā 
An Introduction to RF Design, Live presentation at EELive 2014
An Introduction to RF Design, Live presentation at EELive 2014An Introduction to RF Design, Live presentation at EELive 2014
An Introduction to RF Design, Live presentation at EELive 2014
Ā 
Introduction
IntroductionIntroduction
Introduction
Ā 
About li as
About li asAbout li as
About li as
Ā 
FinalProject
FinalProjectFinalProject
FinalProject
Ā 

Similar to Oscilloscope Lab

St
StSt
Stz2x7
Ā 
Ultrasonic Range Finder
Ultrasonic Range FinderUltrasonic Range Finder
Ultrasonic Range FinderHicham Berkouk
Ā 
STSN1132 Radiation Detection #3.pptx
STSN1132 Radiation Detection #3.pptxSTSN1132 Radiation Detection #3.pptx
STSN1132 Radiation Detection #3.pptxNurmaizatulAtikah
Ā 
Design and implementation of analog multipliers with IC's
Design and implementation of analog multipliers with IC'sDesign and implementation of analog multipliers with IC's
Design and implementation of analog multipliers with IC'sheyaci
Ā 
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...AL- AMIN
Ā 
Module-V-Wave-Analyzers-G.Anitha.ppt
Module-V-Wave-Analyzers-G.Anitha.pptModule-V-Wave-Analyzers-G.Anitha.ppt
Module-V-Wave-Analyzers-G.Anitha.pptMANIMEGALAIMECE
Ā 
Lab LCA 1 7
Lab LCA 1 7Lab LCA 1 7
Lab LCA 1 7NUST Stuff
Ā 
cathode ray oscilloscope &function generator
cathode ray oscilloscope &function generatorcathode ray oscilloscope &function generator
cathode ray oscilloscope &function generatormegha agrawal
Ā 
Electronic instrumentation in NMT
Electronic instrumentation in NMTElectronic instrumentation in NMT
Electronic instrumentation in NMTSUMAN GOWNDER
Ā 
Presentation1 comb filter
Presentation1 comb filterPresentation1 comb filter
Presentation1 comb filterPrerna Singh
Ā 
Negitive Feedback in Analog IC Design 02 April 2020
Negitive Feedback in Analog IC Design 02 April 2020  Negitive Feedback in Analog IC Design 02 April 2020
Negitive Feedback in Analog IC Design 02 April 2020 Javed G S, PhD
Ā 
Amplifiers.pdf
Amplifiers.pdfAmplifiers.pdf
Amplifiers.pdfbooksarpita
Ā 

Similar to Oscilloscope Lab (20)

Transistor Amplifire.pptx
Transistor Amplifire.pptxTransistor Amplifire.pptx
Transistor Amplifire.pptx
Ā 
Basics of amplifier
Basics of amplifierBasics of amplifier
Basics of amplifier
Ā 
Oscillators module-01
Oscillators module-01Oscillators module-01
Oscillators module-01
Ā 
St
StSt
St
Ā 
Ultrasonic Range Finder
Ultrasonic Range FinderUltrasonic Range Finder
Ultrasonic Range Finder
Ā 
STSN1132 Radiation Detection #3.pptx
STSN1132 Radiation Detection #3.pptxSTSN1132 Radiation Detection #3.pptx
STSN1132 Radiation Detection #3.pptx
Ā 
Unit 3 betl 305
Unit 3 betl 305Unit 3 betl 305
Unit 3 betl 305
Ā 
Unit 3 betl 305
Unit 3 betl 305Unit 3 betl 305
Unit 3 betl 305
Ā 
Design and implementation of analog multipliers with IC's
Design and implementation of analog multipliers with IC'sDesign and implementation of analog multipliers with IC's
Design and implementation of analog multipliers with IC's
Ā 
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
Ā 
Module-V-Wave-Analyzers-G.Anitha.ppt
Module-V-Wave-Analyzers-G.Anitha.pptModule-V-Wave-Analyzers-G.Anitha.ppt
Module-V-Wave-Analyzers-G.Anitha.ppt
Ā 
Lab LCA 1 7
Lab LCA 1 7Lab LCA 1 7
Lab LCA 1 7
Ā 
EMI Unit II
EMI Unit IIEMI Unit II
EMI Unit II
Ā 
cathode ray oscilloscope &function generator
cathode ray oscilloscope &function generatorcathode ray oscilloscope &function generator
cathode ray oscilloscope &function generator
Ā 
Electronic instrumentation in NMT
Electronic instrumentation in NMTElectronic instrumentation in NMT
Electronic instrumentation in NMT
Ā 
blah
blahblah
blah
Ā 
Presentation1 comb filter
Presentation1 comb filterPresentation1 comb filter
Presentation1 comb filter
Ā 
Negitive Feedback in Analog IC Design 02 April 2020
Negitive Feedback in Analog IC Design 02 April 2020  Negitive Feedback in Analog IC Design 02 April 2020
Negitive Feedback in Analog IC Design 02 April 2020
Ā 
Amplifiers.pdf
Amplifiers.pdfAmplifiers.pdf
Amplifiers.pdf
Ā 
Function generator
Function generatorFunction generator
Function generator
Ā 

More from cyberns_

Keusahawanan Laporan Amali Jualan Oden shalalala
Keusahawanan Laporan Amali Jualan Oden shalalalaKeusahawanan Laporan Amali Jualan Oden shalalala
Keusahawanan Laporan Amali Jualan Oden shalalalacyberns_
Ā 
Kertas kerja Kelab Esporst Dmt 4
Kertas kerja Kelab Esporst Dmt 4Kertas kerja Kelab Esporst Dmt 4
Kertas kerja Kelab Esporst Dmt 4cyberns_
Ā 
Surat rasmi Kelab DMT 4
Surat rasmi Kelab DMT 4Surat rasmi Kelab DMT 4
Surat rasmi Kelab DMT 4cyberns_
Ā 
Minit Mesyuarat Kelab Mobile Technology
Minit Mesyuarat Kelab Mobile TechnologyMinit Mesyuarat Kelab Mobile Technology
Minit Mesyuarat Kelab Mobile Technologycyberns_
Ā 
Implikasi penggunaan telefon bimbit
Implikasi penggunaan telefon bimbitImplikasi penggunaan telefon bimbit
Implikasi penggunaan telefon bimbitcyberns_
Ā 
Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956
Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956
Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956cyberns_
Ā 
Mpu 1181 topic 1 greetings &introductions
Mpu 1181 topic 1 greetings &introductionsMpu 1181 topic 1 greetings &introductions
Mpu 1181 topic 1 greetings &introductionscyberns_
Ā 
Dsm1021sains1 topik3 Gerakanlinear
Dsm1021sains1 topik3 GerakanlinearDsm1021sains1 topik3 Gerakanlinear
Dsm1021sains1 topik3 Gerakanlinearcyberns_
Ā 
Benchmarking
BenchmarkingBenchmarking
Benchmarkingcyberns_
Ā 
Antena proximity sensor
Antena proximity sensorAntena proximity sensor
Antena proximity sensorcyberns_
Ā 
Kelab e sport Kolej Komunit iAmpang
Kelab e sport Kolej Komunit iAmpangKelab e sport Kolej Komunit iAmpang
Kelab e sport Kolej Komunit iAmpangcyberns_
Ā 
Wawasan matlamat konsep
Wawasan matlamat konsepWawasan matlamat konsep
Wawasan matlamat konsepcyberns_
Ā 
Kerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian Tenaga
Kerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian TenagaKerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian Tenaga
Kerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian Tenagacyberns_
Ā 
Dsm1021 sains1 Gerakanlinear
Dsm1021 sains1 GerakanlinearDsm1021 sains1 Gerakanlinear
Dsm1021 sains1 Gerakanlinearcyberns_
Ā 
Proximity sensors
Proximity sensorsProximity sensors
Proximity sensorscyberns_
Ā 
Mobile Device-Architecture
Mobile Device-ArchitectureMobile Device-Architecture
Mobile Device-Architecturecyberns_
Ā 
DSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUM
DSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUMDSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUM
DSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUMcyberns_
Ā 
DSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYADSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYAcyberns_
Ā 
DSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYADSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYAcyberns_
Ā 
UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)
UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)
UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)cyberns_
Ā 

More from cyberns_ (20)

Keusahawanan Laporan Amali Jualan Oden shalalala
Keusahawanan Laporan Amali Jualan Oden shalalalaKeusahawanan Laporan Amali Jualan Oden shalalala
Keusahawanan Laporan Amali Jualan Oden shalalala
Ā 
Kertas kerja Kelab Esporst Dmt 4
Kertas kerja Kelab Esporst Dmt 4Kertas kerja Kelab Esporst Dmt 4
Kertas kerja Kelab Esporst Dmt 4
Ā 
Surat rasmi Kelab DMT 4
Surat rasmi Kelab DMT 4Surat rasmi Kelab DMT 4
Surat rasmi Kelab DMT 4
Ā 
Minit Mesyuarat Kelab Mobile Technology
Minit Mesyuarat Kelab Mobile TechnologyMinit Mesyuarat Kelab Mobile Technology
Minit Mesyuarat Kelab Mobile Technology
Ā 
Implikasi penggunaan telefon bimbit
Implikasi penggunaan telefon bimbitImplikasi penggunaan telefon bimbit
Implikasi penggunaan telefon bimbit
Ā 
Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956
Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956
Sejarah Tingkatan 3 -Kemerdekaan Negara 31 Ogos 1956
Ā 
Mpu 1181 topic 1 greetings &introductions
Mpu 1181 topic 1 greetings &introductionsMpu 1181 topic 1 greetings &introductions
Mpu 1181 topic 1 greetings &introductions
Ā 
Dsm1021sains1 topik3 Gerakanlinear
Dsm1021sains1 topik3 GerakanlinearDsm1021sains1 topik3 Gerakanlinear
Dsm1021sains1 topik3 Gerakanlinear
Ā 
Benchmarking
BenchmarkingBenchmarking
Benchmarking
Ā 
Antena proximity sensor
Antena proximity sensorAntena proximity sensor
Antena proximity sensor
Ā 
Kelab e sport Kolej Komunit iAmpang
Kelab e sport Kolej Komunit iAmpangKelab e sport Kolej Komunit iAmpang
Kelab e sport Kolej Komunit iAmpang
Ā 
Wawasan matlamat konsep
Wawasan matlamat konsepWawasan matlamat konsep
Wawasan matlamat konsep
Ā 
Kerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian Tenaga
Kerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian TenagaKerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian Tenaga
Kerja,Kuasa,Tenaga Keupayaan,Tenaga Kinetik Dan Keabadian Tenaga
Ā 
Dsm1021 sains1 Gerakanlinear
Dsm1021 sains1 GerakanlinearDsm1021 sains1 Gerakanlinear
Dsm1021 sains1 Gerakanlinear
Ā 
Proximity sensors
Proximity sensorsProximity sensors
Proximity sensors
Ā 
Mobile Device-Architecture
Mobile Device-ArchitectureMobile Device-Architecture
Mobile Device-Architecture
Ā 
DSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUM
DSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUMDSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUM
DSM1021-SAINS-1-TOPIK-2-DAYA-MOMENTUM
Ā 
DSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYADSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-LERAIAN-DAYA
Ā 
DSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYADSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYA
DSM1021-SAINS-1-TOPIK-2-DAYA-KESEIMBANGAN-PANDUAN-DAYA
Ā 
UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)
UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)
UI AND UX GRAPHIC DESIGN(SOCIAL NETWORKING)
Ā 

Recently uploaded

How to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxHow to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxmanuelaromero2013
Ā 
Blooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docxBlooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docxUnboundStockton
Ā 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
Ā 
Biting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfBiting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfadityarao40181
Ā 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfSumit Tiwari
Ā 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
Ā 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxpboyjonauth
Ā 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
Ā 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13Steve Thomason
Ā 
Pharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfPharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfMahmoud M. Sallam
Ā 
Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxOH TEIK BIN
Ā 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
Ā 
Hybridoma Technology ( Production , Purification , and Application )
Hybridoma Technology  ( Production , Purification , and Application  ) Hybridoma Technology  ( Production , Purification , and Application  )
Hybridoma Technology ( Production , Purification , and Application ) Sakshi Ghasle
Ā 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
Ā 
ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,
ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,
ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,Virag Sontakke
Ā 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaVirag Sontakke
Ā 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxRaymartEstabillo3
Ā 
_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting Data_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting DataJhengPantaleon
Ā 

Recently uploaded (20)

How to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxHow to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptx
Ā 
Blooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docxBlooming Together_ Growing a Community Garden Worksheet.docx
Blooming Together_ Growing a Community Garden Worksheet.docx
Ā 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
Ā 
Biting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdfBiting mechanism of poisonous snakes.pdf
Biting mechanism of poisonous snakes.pdf
Ā 
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdfEnzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Enzyme, Pharmaceutical Aids, Miscellaneous Last Part of Chapter no 5th.pdf
Ā 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
Ā 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptx
Ā 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Ā 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13
Ā 
Pharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdfPharmacognosy Flower 3. Compositae 2023.pdf
Pharmacognosy Flower 3. Compositae 2023.pdf
Ā 
Model Call Girl in Tilak Nagar Delhi reach out to us at šŸ”9953056974šŸ”
Model Call Girl in Tilak Nagar Delhi reach out to us at šŸ”9953056974šŸ”Model Call Girl in Tilak Nagar Delhi reach out to us at šŸ”9953056974šŸ”
Model Call Girl in Tilak Nagar Delhi reach out to us at šŸ”9953056974šŸ”
Ā 
Solving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptxSolving Puzzles Benefits Everyone (English).pptx
Solving Puzzles Benefits Everyone (English).pptx
Ā 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
Ā 
Staff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSDStaff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSD
Ā 
Hybridoma Technology ( Production , Purification , and Application )
Hybridoma Technology  ( Production , Purification , and Application  ) Hybridoma Technology  ( Production , Purification , and Application  )
Hybridoma Technology ( Production , Purification , and Application )
Ā 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
Ā 
ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,
ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,
ą¤­ą¤¾ą¤°ą¤¤-ą¤°ą„‹ą¤® ą¤µą„ą¤Æą¤¾ą¤Ŗą¤¾ą¤°.pptx, Indo-Roman Trade,
Ā 
Painted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of IndiaPainted Grey Ware.pptx, PGW Culture of India
Painted Grey Ware.pptx, PGW Culture of India
Ā 
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptxEPANDING THE CONTENT OF AN OUTLINE using notes.pptx
EPANDING THE CONTENT OF AN OUTLINE using notes.pptx
Ā 
_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting Data_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting Data
Ā 

Oscilloscope Lab

  • 1. Oscilloscope Lab Introduction: The purpose of this lab is to introduce students to the basic tools used by engineers and technicians in analyzing electronic equipment: the function generator, the analog oscilloscope, and the digital oscilloscope. The oscilloscope is a tool commonly used by engineers and technicians to analyze and troubleshoot electronic systems. In addition, real signals, such as those which carry voice or data, can be very complex and difficult to analyze. The function generator is a device used to generate an electronic signal with specific known characteristics, thereby enabling an engineer or technician to test and examine a circuit. During this lab, the student will use the function generator to generate a number of signals and to analyze those signals using either of the ā€˜oscilloscopesā€™. The student will become familiar with the basic waveforms -- sine, square, and triangle waves -- and the components of the waveforms -- amplitude, period, and frequency. At the conclusion of this lab the student should feel comfortable using the function generator and the oscilloscopes. Equipment: Qty Equipment 1 Leader LFG-1300S Function Generator 1 BNC to 2 alligator clips cable 1 Tektronix 2225 Analog Oscilloscope 1 Tektronix P6103 10X probe 1 Hewlett-Packard 54502A Digital Oscilloscope 1 HP 10430A 10X probe Oscilloscope Lab - Rev 01-05-95 1
  • 2. Equipment Introduction: Part I - Introduction to the Function Generator Overview: The function generator is used to generate a wide range of alternating-current (AC) signals. A diagram of the Leader LFG-1300S Function Generator is shown below in Figure 1. ā€¢ The front panel is divided into six major control groups: 1) Frequency Selection Group; 2) Sweep Group; 3) Amplitude Modulation Group; 4) DC Offset Group; 5) Function, or Waveform Group; and 6) Output Group. POWER ON SWEEP AMP MOD DC OFFSET OUTPUT ATTENUATION dB FUNCTION FREQUENCY LEADER LFG-1300S FUNCTION GENERATOR Hz Figure 1. Front Panel of Function Generator ā€¢ The power switch on the upper left-hand corner of the unit. The green LED will indicate that the unit is on. ā€¢ The three most important groups for this lab are the frequency, function, and output groups. The remaining three groups, the sweep, amplitude modulation, and the DC offset groups, will be briefly covered in the lab setup procedures. Should the student desire more detailed descriptions of these groups, the Leader Function Generator manual is available in the lab. Oscilloscope Lab - Rev 01-05-95 2
  • 3. Frequency Selection Group: These controls are used select the operating frequency of the function generator. This group consists of the frequency control knob and the eight frequency multiplier selection buttons. For example, to set the function generator to an operating frequency of 2000 Hz (2 kHz): ā€¢ Rotate the frequency control knob to 2. ā€¢ Select the 1 kHz frequency multiplier button. With the result that: 2.0 * 1 kHz = 2.0 kHz. Example two sets the function generator to an operating frequency of 5.5 kHz: ā€¢ Rotate the frequency control knob to 0.55. ā€¢ Select the 10 kHz frequency multiplier button. With the result that: 0.55 * 10 kHz = 5.5 kHz. Output Group: 1. These controls are used to adjust the amplitude of the generator's output signal. The group consists of the amplitude-control knob, the three attenuation buttons and the fused 50 ohm BNC connector. ā€¢ Although the amplitude knob is not indexed, the amplitude ranges from a few millivolts to approximately 20 volts. ā€¢ We will set the amplitude levels by aligning the white line on the amplitude knob to the three o'clock position (90 degrees right), the nine o'clock position (90 degrees left), or the twelve o'clock position (straight up). ā€¢ Notice that rotating the knob fully to the left does not result in a zero amplitude signal. ā€¢ The attenuation buttons are used to attenuate (decrease) the amplitude of the signal by a factor measured in decibels. The following relationship will assist in working with the attenuation buttons: (dB) = -10 * log10 (Pout/Pin) (if power is the unit of measurement) or (dB) = -20 * log10 (Vout/Vin) (if voltage is the unit of measurement) ā€¢ Example: Given a 1 volt input signal, what is the change in voltage amplitude if the 10 dB attenuation button is depressed? Oscilloscope Lab - Rev 01-05-95 3
  • 4. Beginning with the equation defined above: (dB) = -20 * log10 (Vin/Vout) 10dB = -20 * log10 (Vin/Vout) -0.5 dB = log10 (Vin/Vout) 10 -.5 = Vout/Vin Vout/Vin = 0.3162 ā€¢ From here, the output voltage can be stated in terms of the input voltage and vice-versa: Vin = Vout/0.3162 or Vout = Vin * 0.3162 Since the initial input voltage (Vin) was 1 volt the output voltage (Vout) must be 0.3162V or 316 millivolts. ā€¢ Note: The attenuation buttons are additive. In other words, if the 10 dB and the 20 dB buttons are both pressed in, the combined attenuation of the input signal is 30 dB. 2. The fused 50 ohm BNC connector is used for connecting the function generator to other equipment. The most common connection used to connect the function generator to other equipment is a cable with a BNC connector on one end and two alligator clips on the other end. ā€¢ One end of the cable, the base, is used to connect the cable to the BNC connector on the function generator. The other end, which has two ā€˜alligatorā€™ clips: one positive (red) and one negative (black), is used to connect the function generator to the device or circuit under study. ā€¢ To connect the cable to the function generator: 1. Hold the cable's BNC connector, the base, in one hand. 2. Line up the two grooves on the cable's BNC connector with the two pegs on the outside of the function generator's BNC connector. 3. Push the cable's BNC connector over the function generator's BNC connector until it is fully seated. 4. Rotate the cable's BNC connector in a clockwise manner until it is firmly secured to the function generator's BNC connector. Oscilloscope Lab - Rev 01-05-95 4
  • 5. Function/Waveform Selection Group: This group is used to select the shape of the generated waveform. The group is made up of the six wave-selector buttons. The six waveforms that the function generator can produce are the sine wave, the square wave, the triangle wave, two sawtooth waves, and the variable-width pulse wave. reference reference SINE WAVE TRIANGLE WAVE SQUARE WAVE SAWTOOTH PULSE SAWTOOTH Figure 2: Available Generated Waveforms ā€¢ In this lab you will become familiar with the three fundamental signals (waveforms), the sine wave, the square wave and the triangle wave. Waveforms: This section will provide you with some background on waveforms and on measuring waveforms. It is especially important that you understand how to measure waveforms. This is a skill that is not only important to this lab, but also to the T-1 Lab and in labs for other Telecommunications Courses. ā€¢ There are two types of alternating-current signals, which are also called waveforms: ā€¢ Periodic ā€¢ Aperiodic ā€¢ Periodic signals, such as sine or triangle waves, behave in a uniform manner and repeat themselves over a given length of time. Each repetition of a repeating signal is called a period or a cycle. Oscilloscope Lab - Rev 01-05-95 5
  • 6. ā€¢ Aperiodic signals, such as analog voice, behave in a non-uniform manner and do not repeat themselves over any given length of time. Note: In this lab, we will only work with periodic signals. ā€¢ When working with AC signals, there are three properties of the signal that we are concerned with: amplitude, period and frequency. ā€¢ The amplitude of the wave is defined as the maximum magnitude of the wave. The amplitude is the vertical component of the signal and is measured in units of volts (V). Since we are dealing with an AC signal, the voltage will change over a period of time. The maximum voltage of a signal during its cycle is commonly referred to as the peak voltage (Vp). ā€¢ The amplitude can be measured from the reference line to the peak (Vp) or from peak-to-peak (Vpp). See Figure 3 below. peak-to-peak amplitude peak or maximum amplitude reference point Figure 3: Measuring Amplitude. ā€¢ With a periodic signal that is symmetrical (equi-distant above and below the reference point), the peak-to-peak voltage is equal to twice the peak voltage: Vpp = 2 * Vp Oscilloscope Lab - Rev 01-05-95 6
  • 7. ā€¢ The period (T) of the signal is defined as the time it takes for a signal to complete one full cycle. The period is the horizontal component of the signal, measured in units of seconds (s). In Figure 4, the period of the signal is measured as 250 milliseconds (250.0 x 10 -3 s). P E R I O D o n e s e c o n d 2 5 0 m s e c 2 5 0 m s e c 2 5 0 m s e c 2 5 0 m s e c Figure 4: Measuring Period. ā€¢ The frequency (f) of the signal is defined as the rate at which a periodic signal repeats. It is usually measured in units of Hertz (Hz), where 1 Hz = 1 cycle per second. ā€¢ In Figure 4, above, you can see four(4) cycles occurring within one second; therefore, the signal has a frequency of 4 Hz. ā€¢ The frequency, f, of a wave is inversely related to its period (T): f = 1/T ā€¢ Example: The period of the signal is 250 milliseconds, therefore the frequency of that signal is: f = 1/T f = 1/250 milliseconds f = 4 hertz Abbreviation Prefix Name Factor T Tera 1012 G Giga 109 M Mega 106 K Kilo 103 m milli 10-3 Āµ micro 10-6 n nano 10-9 p pico 10-12 Table of Scientific Prefixes Oscilloscope Lab - Rev 01-05-95 7
  • 8. PART II - Introduction to the Tektronix 2215 Analog Oscilloscope Overview: An oscilloscope is an electronic measuring device which provides a two-dimensional visual representation of a signal. Because the oscilloscope allows the user to see the signal(s), their characteristics can be easily measured and observed. The oscilloscope displays a graph of voltage (on the vertical axis) versus time (on the horizontal axis). Most electrical circuits can be easily connected to the oscilloscope typically with probes. ā€¢ There are four major control groups on the Tektronix oscilloscope: (1) the Display group; (2)the Vertical group; (3) the Horizontal group; and (4) the Trigger group. See Figure 5, below. INTENSITY BEAM FIND FOCUS POWER TRACE SEP POSITION POSITION CH1 VOLTS/DIV CH2 VOLTS/DIV CH! BOTH CH1 ADD ALT CHOP COURSE FINE 5X 10X 50X SEC/DIV NORM INVERT AC GND DC AC GND DC SOURCE COUPLING 1X ALT MAG CH1 CH2 EXT LEVEL MODE HOLDOFF VERTICAL HORIZONTAL TRIGGER TEKTRONIX 2225 Figure 5: Front Panel of Analog Oscilloscope Display Group This group is used to display and adjust the signal for optimal viewing. It consists of the display screen, the intensity-control knob, the beam-find button, the focus-control knob, and the power switch. ā€¢ The display screen is laid out in a 8 by 10 centimeter grid. The oscilloscope draws a ā€˜traceā€™ or graph by moving an electron beam across the phosphor coating on the inside of the cathode-ray tube (CRT). The excited phosphorous glows for a short period of time, thereby tracing the path of the beam. ā€¢ The intensity-control knob is used to adjust the brightness of the trace. The level should be set to the user's preference; however, increasing the intensity beyond a certain point will make the trace "fuzzy." The intensity level should never be increased past the point where the trace has sharply defined edges. Oscilloscope Lab - Rev 01-05-95 8
  • 9. ā€¢ The beam-find button allows the user to locate the electron beam anytime it's off-screen. Push the beam find button to temporarily reduce the vertical and horizontal deflection voltages so that the beam always appears within the 8 by 10 centimeter screen. ā€¢ The focus-control knob adjusts the electron beam for optimal trace resolution. Vertical Group: This group is used to adjust the vertical components (Y-axis) and the vertical position of the signal. This group consists of the vertical-position knobs, channel-selector switch, volts-per-division- selector knobs, input-coupling switch, the channel-mode-selector switch, the channel-2-invert switch, and the BNC connectors. Because the Tektronix 2225 oscilloscope is a two channel oscilloscope, there is one set of switches for each channel. ā€¢ The vertical-position controls are used to vertically move the trace of one channel or the other. ā€¢ The channel-selector switch, labeled ā€˜CH1/BOTH/CH2ā€™, selects which channels are displayed on the screen. ā€¢ The volts-per-division-selector knob sets the vertical scale for each channel's trace. 1. Since a 10X probe is used in this lab, all readings should be made from the ā€˜10Xā€™ box on the knob. 2. Each channel can have a different vertical scale. 3. A division is one "block" on the screen. 4. Each tick mark is one-fifth (0.2) of a division. 5. This knob is marked in both volts and millivolts. ā€¢ The input-coupling switch, labeled ā€˜AC/GND/DCā€™, selects the coupling mode of that channel's display. AC means that only the alternating portion of the signal is displayed. DC will display both the alternating portion of the signal, plus any DC component. GND shows the 0 V reference level. ā€¢ The trace-separation knob is used in conjunction with the horizontal-magnification controls, to be discussed later. ā€¢ The channel-mode-selector switch, labeled ā€˜ADD/ALT/CHOPā€™, is activated only when BOTH is selected on the channel-selector switch. ADD graphically adds the Channel 1 signal to the Channel 2 signal. If the channel-2-invert switch (ā€˜NORMAL/INVERTā€™) is set to INVERT, then ADD will in fact subtract the Channel 2 signal from the Channel 1 signal. ALT traces one channel, then the next. CHOP works like ALT, but jumps back and forth between the two channels during a single trace. Oscilloscope Lab - Rev 01-05-95 9
  • 10. Horizontal Control Group: The Horizontal Control Group consists of the coarse and fine position knobs, the horizontal- magnification switch, the seconds-per-division-selector knob, and the magnification-scale- selector switch. ā€¢ The coarse and fine-position-knobs allow the horizontal movement of the traces in a rough manner and in a precise manner, respectively. These are used to position the traces in a manner that makes measurement both more convenient and more precise. The horizontal-magnification switch (ā€˜X1/ALT/MAGā€™) selects the regular (X1) and/or a horizontally magnified trace. 1. The trace separation knob in the Vertical-Control Group allows for the vertical separation of the X1 trace from the magnified trace when ALT is selected. 2. If ALT or MAG is selected, then the magnification-scale-selector switch (ā€˜5X/10X/50Xā€™) is activated. ā€¢ The seconds-per-division knob sets the time base (horizontal) scale. It is marked in seconds, milliseconds, and microseconds. Note: There is only one horizontal scale for both channels. Trigger Group In order to display a signal, the oscilloscope must be able to ā€˜lockā€™ onto that signal; the function of the triggering controls is to do just that. Triggering can be a complicated topic, and is beyond the scope of this lab. For more details, the student is referred to the Tektronix manual. The trigger group is made up of the trigger level knob, the rising/falling edge switch, the trigger mode switch, the holdoff knob, the trigger source switches, and the trigger coupling switch. ā€¢ The trigger-level knob sets the voltage level at which the oscilloscope will ā€˜triggerā€™. If a signal is ā€˜runningā€™ -- that is, not stable -- the trigger level may be too high or too low for the oscilloscope to recognize the signal. Often you can lock in a running signal by rotating this knob to the left and to the right. This adjustment provides a mechanism for ignoring small (low voltage) signals that are well below the level of the signal you ar interested in. This knob is not indexed. ā€¢ The rising/falling-edge switch selects whether the oscilloscope will trigger on the positive (rising) or negative (falling) edge of the signal. ā€¢ The trigger-mode switch will normally be set to AUTO, but sometimes it is necessary to use NORM. Other settings are not relevant to this lab. ā€¢ The holdoff knob affects the delay associated with triggering. The student should consult the manual if more information is desired. Oscilloscope Lab - Rev 01-05-95 10
  • 11. ā€¢ The trigger-source switches select which signal the oscilloscope will attempt to lock onto. Possible choices include CH1, CH2, VERT MODE, or EXTERNAL. Selecting CH1 or CH2 will make the oscilloscope attempt to trigger on those channels. If no input is available on that channel, there may be problems when attempting to view both channels simultaneously. For this reason, it is recommended that this switch be left on VERT MODE, which provides an automatic trigger on either CH1 or CH2. ā€¢ The trigger-coupling switch is another more complicated feature which will not be discussed here. This switch should typically be set to AC. How To Make Readings on the Oscilloscope: ā€¢ Peak-to-Peak Voltage: Use the vertical-position knob to place a peak (positive or negative) on a horizontal line, keeping the peak on the screen. Use the horizontal-position knob to set the next (opposite sign) peak on the center vertical line. Count the number of divisions between the positive and negative peaks. Note: The divisions are further subdivided into 1/5 (0.2) by the tick marks. Multiply the number of divisions from step 3 by the volts/div setting for the channel in use. ā€¢ Period: Use the horizontal-position knobs to align any edge of the signal with a vertical line. Use the vertical-position knob to place the next identical edge crossing on the X-axis. Count the number of divisions along the horizontal line to the next crossing in the same direction. Multiply the number of divisions from step 3 by the sec/div setting. Note: The divisions are subdivided into fifths (0.2) by the ā€˜tick marksā€™. ā€¢ Frequency: Measure the period of the signal (T). Calculate frequency (f) using f = 1/T. Oscilloscope Lab - Rev 01-05-95 11
  • 12. Figure 6: Measuring a Signal. ā€¢ Example: Assuming the Volts/Div knob reads 2 V/div, the above peak-to-peak voltage would be: Vpp = 2 volts/div * 4.2 div = 8.4 volts Assuming the Sec/Div control knob read 50ms, the above period would be: T = 50 milliseconds/div * 5.25 div = 2625.0 milliseconds = 2.625 seconds Oscilloscope Lab - Rev 01-05-95 12
  • 13. PART III - Introduction to the HP 5402A Digital Oscilloscope Overview: The digital oscilloscope is a test and measuring device, similar to the analog oscilloscope. Instead of displaying a signal by deflecting an electron beam that is tracing across a CRT, the digital oscilloscope uses a processor to sample, digitize, and display the incoming analog signal on the display screen. The digital oscilloscope has a user interface which allows the user to manipulate and measure the displayed signal. It can be used to store, retrieve, and print waveforms. It can also be programmed to make measurements automatically or to control other devices. Because the HP 5402A digital oscilloscope is a multi-purpose piece of equipment, it would be very time consuming to explain out all of its functionality. It is strongly recommended that the student become familiar with the following reference manuals if more details are needed: ā€¢ HP 54502A 400 MHz Digitizing Oscilloscope Front-Panel Reference ā€¢ HP 54502A 400 MHz Digitizing Oscilloscope Programming Reference These references can be found at the lab tables in the Telecommunications Lab (Room 834). Front Panel Reference: The HP 54502A 400 MHz Digitizing Oscilloscope controls are organized into six functional areas or groups: (1) the Display Group, (2) the System-Control Group, (3) the Menu Group (4) the Setup Group, (5) the Entry Group, and (6) the Signal-Input Group. HEWLETT 54502 A 400 MHz PACKARD DIGITIZING OSCILLOSCOPE SYSTEM CONTROL SETUP MENUS ENTRY DISPLAY INPUTS Figure 7: Front Panel of Analog Oscilloscope The power switch is a rocker switch located at the back of the machine, immediately above the power cord. Oscilloscope Lab - Rev 01-05-95 13
  • 14. ā€¢ The display group consists of the display screen and the programmable menu selection buttons immediately to the right of the screen. The function of these buttons is dependent on the menu selected. An option appearing on the screen may be selected by striking the button to the right of it. ā€¢ The system-control group includes the RUN/STOP and the SINGLE buttons, which control the screen refreshing, and the CLEAR DISPLAY button which clears the screen. ā€¢ The menu group contains a selection of menus whose function is similar to the front panel controls on the analog oscilloscope. The procedures below will serve to familiarize you with the similarities. There is also a āˆ†āˆ†t/āˆ†āˆ†v menu which is helpful in making measurements of voltage, period, and frequency. ā€¢ The setup group includes the AUTOSCALE button, which resets the oscilloscope to default settings and automatically adjusts horizontal and vertical scales for the signal that is input. ā€¢ The entry group is used to enter data. The numeric keypad is augmented with four unit selection buttons, which give a selection of seconds, milliseconds, microseconds, nanoseconds, volts, and millivolts. Alternatively, the data entry wheel can be used to adjust settings instead of entering a specific value. The entry group also contains a number of automatic measurement functions, which are accessed using the blue shift button. ā€¢ The signal-input group consists of the BNC connectors for each channel, and an auxiliary connector. Making Measurements with the Digital Oscilloscope: There are two ways to make measurements with the HP digital oscilloscope: (1) by using the āˆ†t/āˆ†v markers, or (2) by using the automatic measurements functions. The procedures for both these methods are outlined in the lab procedures. Oscilloscope Lab - Rev 01-05-95 14
  • 15. Lab Procedures: Part I - Analog Oscilloscope Procedures: Overview: In this part of the lab, you will use the function generator to generate a signal and then use the analog oscilloscope to make some measurements of the data. If you need any assistance in identifying or configuring the equipment, please see the on-duty GSA. Function Generator Setup: ā€¢ Turn on the function generator. ā€¢ Make sure that the Sweep, Amplitude Modulation, and DC offset groups are turned OFF. Note: The Sweep and Amplitude Modulation buttons should be in the OUT position, and the DC offset knob should be pushed IN. ā€¢ Select a TRIANGLE wave. ā€¢ Set frequency to 1.0 kHz. ā€¢ Set the amplitude knob to 12 o'clock. ā€¢ Make sure all the attenuation buttons are OUT. ā€¢ Attach the BNC end of the cable to the BNC socket in the function generator's Output Group. ā€¢ Tie back the black lead alligator clip by clipping the black alligator clip to the insulated wire to avoid short circuits and blown fuses. Analog Oscilloscope Setup: ā€¢ Turn on the power to the oscilloscope. ā€¢ Connect the Tektronix 10X probe to the Channel 1 input on the oscilloscope. ā€¢ Tie back the black lead on the probe, then expose the hook and clip it to the red alligator clip on the function generator cable. ā€¢ Set CH1/BOTH/CH2 switch to CH1. ā€¢ In the Vertical Control Group, set the Channel 1 AC/GND/DC switch to GND. Oscilloscope Lab - Rev 01-05-95 15
  • 16. ā€¢ Use the CH1 vertical position knob to move the ground (0V) reference (horizontal line) to the center line on the screen. ā€¢ After referencing the signal to ground, set the AC/GND/DC switch to AC. ā€¢ Intensity knob: adjust the signal intensity and focus to a comfortable level, by using the intensity and focus knobs, respectively. Note: Keep the signal intensity to within a reasonable range to minimize the chances of burning the phosphor and damaging the display screen. ā€¢ Set the Volts-per-Division knob to 2 V/div. Note: you are using a 10X probe, so be sure to take all readings from the 10X position. ā€¢ Set the NORMAL/INVERT switch to NORMAL. ā€¢ Set the ADD/ALT/CHOP switch to ALT. ā€¢ Set X1/ALT/MAG to X1. ā€¢ Set the seconds-per-division knob to 0.1 ms/div. ā€¢ Set the rising/falling-edge switch to positive. ā€¢ Set the trigger-mode switch to AUTO. ā€¢ Set the trigger-source switch to VERT MODE. ā€¢ Set the trigger-coupling switch to AC. Note: If the signal is still running, try adjusting the trigger level or holdoff knobs. ā€¢ Make sure that the 'cal' knobs on the V/div and S/div knobs are both pushed in and turned all the way to the right. They will click into position. This is so the scope will be calibrated properly. Oscilloscope Lab - Rev 01-05-95 16
  • 17. Procedures: 1. Using the settings completed above, answer the following questions. ā€¢ (1.1) What is the setting on the Volts/Div control knob? _______ volts/div ā€¢ (1.2) How many vertical divisions from peak-to-peak? _______ div ā€¢ (1.3) What is the peak to peak voltage (Vpp)? Vpp = _______ volts/div * _______ div = _______ volts ā€¢ (1.4) What is the setting on the Sec/Div control knob? _______ ___ seconds/div ā€¢ (1.2) How many horizontal divisions from positive going crossing to positive going crossing? _______ div ā€¢ (1.5) What is the period of the signal (T)? T = _______ seconds/div * _______ div = _______ ___ seconds ā€¢ (1.6) What is the frequency of the signal (f)? _______ ___ hertz ā€¢ (1.7) Draw the displayed signal on Graph 1. Be neat, to scale, and concise. Be sure to note the (scale) V/div and Sec/div settings. Graph 1 Oscilloscope Lab - Rev 01-05-95 17
  • 18. 2. Next you will adjust the function generator to output another signal and repeat the measurements in Step 1. ā€¢ Generate a square wave between 5 and 10 kHz with an amplitude setting of 3 o'clock. ā€¢ Adjust the V/div and Sec/div settings to maximize the display of the signal on the CRT. Make sure you show the signal from peak-to-peak and at least one full cycle (period) of the signal. ā€¢ (2.1) What is the setting on the Volts/Div control knob? _______ volts/div ā€¢ (2.2) How many vertical divisions from peak-to-peak? _______ div ā€¢ (2.3) What is the peak to peak voltage (Vpp)? Vpp = _______ volts/div * _______ div = _______ volts ā€¢ (2.4) What is the setting on the Sec/Div control knob? _______ ___ seconds/div ā€¢ (2.2) How many horizontal divisions from positive going crossing to positive going crossing? _______ div ā€¢ (2.5) What is the period of the signal (T)? T = _______ sec/div * _______ div = _______ ___ seconds ā€¢ (2.6) What is the frequency of the signal (f)? _______ ___ hertz. ā€¢ (2.7) Draw the displayed signal on Graph 2. Be neat, to scale, and concise. Be sure to note the (scale) V/div and Sec/div settings. Graph 2 Oscilloscope Lab - Rev 01-05-95 18
  • 19. 3. In this part of the lab, you will experiment with the triggering of the analog oscilloscope. You will need to make some changes to the controls of the analog oscilloscope and the function generator, and you will then have to answer a few questions. ā€¢ Set TRIGGER MODE to NORM. Set TRIGGER SOURCE to CH1. ā€¢ Turn the TRIGGER LEVEL knob all the way to the right. Turn the TRIGGER HOLDOFF knob all the way to the right. ā€¢ Generate a 1.4 kHz sine wave with the function generator; setting the amplitude knob all the way to the left. Then display the signal on the oscilloscope. ā€¢ On the oscilloscope, turn the horizontal position (coarse) knob until the left edge of the trace is shown. ā€¢ While watching the display, turn the TRIGGER LEVEL knob to the left until you get a steady display. ā€¢ Keep turning the TRIGGER LEVEL knob to the left, and watch the trigger level go down. Note that the oscilloscope ceases to trigger when the level moves below the signal. ā€¢ Set the TRIGGER LEVEL as close to the top of the signal as possible while still retaining a steady picture. ā€¢ (3.1) What is the peak-to-peak voltage (Vpp)? _______ volts ā€¢ On the function generator, depress the 10 dB attenuation button. ā€¢ Use the TRIGGER LEVEL knob to steady the display. It's a delicate adjustment, so watch the screen carefully while you turn the knob to the left. ā€¢ (3.2) Why did the signal run or disappear? (not steady or not locked) ā€¢ (3.3) Without changing the V/div setting, what is the new Vpp? _______ volts ā€¢ (3.4) Using the formula on page 3, calculate the attenuation in dB? _______dB (note sign!) ā€¢ Disconnect the analog oscilloscope, and turn off the power. Oscilloscope Lab - Rev 01-05-95 19
  • 20. Part II - Digital Oscilloscope Procedures: Overview: In this part of the lab, you will use the function generator to generate a signal and then use the digital oscilloscope to make some measurements of the data. If you need any assistance in identifying or configuring the equipment, please see the on-duty GSA. Digital Oscilloscope Setup: ā€¢ Generate a 6kHz triangle wave with the amplitude turned fully to the right using the function generator. ā€¢ Connect the HP probe to the Channel 1 input on the oscilloscope, then connect the probe to the red lead from the function generator. ā€¢ Press the AUTOSCALE button to make the oscilloscope automatically adjust its settings. ā€¢ Select the Channel menu by pressing the CHAN button. Select Channel 1. Set v/div to 5.0 v/div, if it is not already set. Set offset to 0 volts, if not already set. Select AC. Select 1Mā„¦ā„¦, if not already set. Select More by pressing the button to the right of the more label. Make sure that 10:1 is selected. If not, use the wheel to set it. Press the More button. ā€¢ Select the Time base menu by pressing the TIME BASE button. Set the time base to 50 ĀµĀµs/div, if not already set. Set delay to 0 seconds. Select reference: center, if not already set. Set window: off, if it is on. Select realtime. ā€¢ Select the Trigger menu by pressing the TRIG button. Select AUTO. Select EDGE. Select source: channel 1. ā€¢ Select the Display menu by pressing the DISPLAY button. Select NORM. Set persistence: 'minimum' or 'single'. Set No. of screens: 1. Select GRID. Set connect dots: ON. Oscilloscope Lab - Rev 01-05-95 20
  • 21. PROCEDURES: 1. Using the settings completed above, answer the following questions. ā€¢ Select the āˆ†t/āˆ†v menu by pressing the āˆ†āˆ†t/āˆ†āˆ†v button. ā€¢ Select Vmarkers and Tmarkers. ā€¢ Select Vmarker 1, then use the data entry wheel to set the marker at the bottom of the signal. Be sure to adjust the voltage level not the channel number. ā€¢ Similarly, set Vmarker 2 to the top of the signal. ā€¢ (1.1) Read āˆ†V at the bottom of the screen. Vmarker 1: _______ ___ volts Vmarker 2: _______ ___ volts āˆ†V: _______ ___ volts ā€¢ (1.2) Using the same procedure as for the Vmarkers, set the āˆ†T start marker to either a positive or negative peak. ā€¢ (1.3) Similarly, set the āˆ†T stop marker to the next positive or negative peak. (Choose the same polarity as you did in Step 1.2.) ā€¢ (1.4) Read āˆ†T and 1/āˆ†T (period and frequency) at the bottom of the screen. Start Marker: _______ ___ seconds Stop Marker: _______ ___ seconds āˆ†T (period): _______ ___ seconds 1/āˆ†T (frequency): _______ ___hertz ā€¢ Turn the markers off. 2. Using the automatic measurement capability of the digital oscilloscope, you will repeat the measurements you made in Step 1. ā€¢ (2.1) Press the blue shift button, press button 1 (Vpp), and then press button 1 for channel 1 when 'c#' appears at the bottom of the screen in reverse text. ā€¢ (2.2) Measure Vpp: _______ ___ volts Note: If 'm#' or 'f#' appears in reverse text at the bottom of the screen, use the entry wheel to change this to 'c#'. If this fails, try hitting RECALL and CLEAR simultaneously to reset the oscilloscope. You would then need to restart with the pressing the AUTOSCALE in the digital oscilloscope setup section. Oscilloscope Lab - Rev 01-05-95 21
  • 22. ā€¢ (2.3) Press the blue shift button, press button 9 (Freq.), and then press button 1 for channel 1 when 'c#' appears at the bottom of the screen in reverse text. Measure Frequency (f): _______ ___ hertz ā€¢ (2.4) Press the blue shift button, press button s-V (period) to the right of button 9, and then press button 1 for channel 1 when 'c#' appears at the bottom of the screen in reverse text. Measure Period (T): _______ ___ seconds ā€¢ (2.5) Draw the displayed signal on Graph 3. Be neat, to scale, and concise. Be sure to note the (scale) V/div and Sec/div settings. Graph 3 ā€¢ To clear the measurements from the screen, press the blue shift button and then press the clear button. ā€¢ Disconnect the digital oscilloscope from the function generator and turn it off. ā€¢ Make sure all probes and connectors are disconnected from the equipment and neatly placed on the shelf above the work area. ā€¢ Be sure to turn off both oscilloscopes and the function generator. Oscilloscope Lab - Rev 01-05-95 22