The document discusses various life support and life saving equipment. It describes patient monitors and their key components like ECG, pulse oximetry, blood pressure, temperature and respiration monitors. It explains the basic theory, functions, operations, maintenance and troubleshooting of equipment like patient monitors, infusion pumps, syringe pumps, ventilators, bubble CPAP, defibrillator monitors and baby incubators. It focuses on the working and testing of ECG, pulse oximetry, blood pressure and other vital components of patient monitoring systems.
Laporan ini berisi tentang alat-alat kesehatan berupa : Infuse Pump, Automatic Processing Film (APF), Meja Operasi, Blood Bank, Microwave Diathermy, dan Doppler.
Disini disertakan spesifikasi, blok diagram serta cara kerja alat.
Selamat menikmati :)
Jangan lupa tinggalkan komentar positifnya ya sob :*
Laporan ini berisi tentang alat-alat kesehatan berupa : Infuse Pump, Automatic Processing Film (APF), Meja Operasi, Blood Bank, Microwave Diathermy, dan Doppler.
Disini disertakan spesifikasi, blok diagram serta cara kerja alat.
Selamat menikmati :)
Jangan lupa tinggalkan komentar positifnya ya sob :*
Dental Equipments - We are manufacturer and suppliers of Dental Equipment, Dental Instruments Dental Products & Tools including Dental Chair, Dental Air Compressor, Ultra Sonic Dental Scaler, Dental Light Cure Unit, Jaw & Teeth Set and Dental Head Simulator Unit
DT Technologies specialize in outsourcing & providing dental laboratories equipment and materials . Our aim is to provide excellent products that are supported by exceptional and easy-to-access services. We offer a variety of dental milling machines, CAD/CAM Equipment for dentists and labs.
A Noninvasive Alternative to +dP/dtmax: Peak Aortic Blood AccelerationInsideScientific
During this webinar, sponsored by Indus Instruments, Dr. Anilkumar Reddy of the Baylor College of Medicine shows how peak acceleration of blood flow velocity in the ascending aorta can be used as a surrogate, noninvasive measurement to evaluate cardiac contractility. Importantly, this technique enables serial measurements in the same animal, leading to reduced animal-to-animal variability, fewer subjects and shorter data collection times.
Specifically, Dr. Reddy presents preliminary findings from his validation studies in mice, where peak aortic acceleration and peak +dP/dt were compared. He highlights the utility of the Doppler method showing how measurements of mitral inflow and myocardial perfusion capacity through coronary flow reserve can be used to noninvasively assess diastolic function. Furthermore, he presents specific examples of noninvasive cardiac measurements and discuss how they scale in translational research from mice to mammals.
Key topics covered during this webinar included:
- Evaluating cardiac contractility using peak aortic acceleration
- Investigating cardiac relaxation using mitral peak early velocity to peak atrial velocity ratio
- Interpreting myocardial perfusion capacity through coronary flow reserve at baseline and with disease or other conditions
- How Doppler Flow Velocity measurements can be used in translational research from mice to mammals
Utilizing Noninvasive Blood Flow Velocity Measurements for Cardiovascular Phe...InsideScientific
Dr. Anilkumar Reddy of the Baylor College of Medicine presents data from his research outlining the importance of blood flow velocity measurement and shows examples of translational data. He provides an overview of Doppler flow velocity measurement technology and compares data obtained from complimentary devices such as 3D echo ultrasound and transit-time flow systems. Several models are presented showing how many selected measurements scale up in translational research from mice to mammals.
During this presentation the audience learned how Flow Velocity measurements can reliably assess the following parameters in rodents:
Systolic and diastolic cardiac function
Myocardial perfusion & coronary reserve
Pressure overload
Aortic stiffness
Peripheral perfusion
Swine animal model in pressure volume loop (pvl) researchFilip Konecny
Slide presentation is for researcher/ clinician who is interested in pressure-volume loop (PVL) research using swine model. Slides are presented in logical order as you would be getting ready for hemodynamic experiment.
Heamodynamic Monitoring System Using Photoplethysmographyabhijith prabha
A proposed system for combining pulse oximeter and blood pressure measuring apparatus to a single novel device. System will be using optical method, photoplethysmography (PPG) for taking the measurement, thus foolproof and hazard free method introduced.
The Roman Empire A Historical Colossus.pdfkaushalkr1407
The Roman Empire, a vast and enduring power, stands as one of history's most remarkable civilizations, leaving an indelible imprint on the world. It emerged from the Roman Republic, transitioning into an imperial powerhouse under the leadership of Augustus Caesar in 27 BCE. This transformation marked the beginning of an era defined by unprecedented territorial expansion, architectural marvels, and profound cultural influence.
The empire's roots lie in the city of Rome, founded, according to legend, by Romulus in 753 BCE. Over centuries, Rome evolved from a small settlement to a formidable republic, characterized by a complex political system with elected officials and checks on power. However, internal strife, class conflicts, and military ambitions paved the way for the end of the Republic. Julius Caesar’s dictatorship and subsequent assassination in 44 BCE created a power vacuum, leading to a civil war. Octavian, later Augustus, emerged victorious, heralding the Roman Empire’s birth.
Under Augustus, the empire experienced the Pax Romana, a 200-year period of relative peace and stability. Augustus reformed the military, established efficient administrative systems, and initiated grand construction projects. The empire's borders expanded, encompassing territories from Britain to Egypt and from Spain to the Euphrates. Roman legions, renowned for their discipline and engineering prowess, secured and maintained these vast territories, building roads, fortifications, and cities that facilitated control and integration.
The Roman Empire’s society was hierarchical, with a rigid class system. At the top were the patricians, wealthy elites who held significant political power. Below them were the plebeians, free citizens with limited political influence, and the vast numbers of slaves who formed the backbone of the economy. The family unit was central, governed by the paterfamilias, the male head who held absolute authority.
Culturally, the Romans were eclectic, absorbing and adapting elements from the civilizations they encountered, particularly the Greeks. Roman art, literature, and philosophy reflected this synthesis, creating a rich cultural tapestry. Latin, the Roman language, became the lingua franca of the Western world, influencing numerous modern languages.
Roman architecture and engineering achievements were monumental. They perfected the arch, vault, and dome, constructing enduring structures like the Colosseum, Pantheon, and aqueducts. These engineering marvels not only showcased Roman ingenuity but also served practical purposes, from public entertainment to water supply.
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptxEduSkills OECD
Andreas Schleicher presents at the OECD webinar ‘Digital devices in schools: detrimental distraction or secret to success?’ on 27 May 2024. The presentation was based on findings from PISA 2022 results and the webinar helped launch the PISA in Focus ‘Managing screen time: How to protect and equip students against distraction’ https://www.oecd-ilibrary.org/education/managing-screen-time_7c225af4-en and the OECD Education Policy Perspective ‘Students, digital devices and success’ can be found here - https://oe.cd/il/5yV
How to Make a Field invisible in Odoo 17Celine George
It is possible to hide or invisible some fields in odoo. Commonly using “invisible” attribute in the field definition to invisible the fields. This slide will show how to make a field invisible in odoo 17.
The Indian economy is classified into different sectors to simplify the analysis and understanding of economic activities. For Class 10, it's essential to grasp the sectors of the Indian economy, understand their characteristics, and recognize their importance. This guide will provide detailed notes on the Sectors of the Indian Economy Class 10, using specific long-tail keywords to enhance comprehension.
For more information, visit-www.vavaclasses.com
The Art Pastor's Guide to Sabbath | Steve ThomasonSteve Thomason
What is the purpose of the Sabbath Law in the Torah. It is interesting to compare how the context of the law shifts from Exodus to Deuteronomy. Who gets to rest, and why?
Model Attribute Check Company Auto PropertyCeline George
In Odoo, the multi-company feature allows you to manage multiple companies within a single Odoo database instance. Each company can have its own configurations while still sharing common resources such as products, customers, and suppliers.
2. Garis-Garis Besar Mata Kuliah
• Konsep – konsep dasar peralatan Lifesupport dan Lifesaving.
• Patient Monitor : Dasar teori dan Fungsi, Blok diagram fungsional,
pengoperasian pemeliharaan, trouble shooting dan pengujian EKG, SpO2,
NIBP, IBP dan temperatur.
• Bedside-Central monitor Dasar teori dan fungsii, pengoperasian,
pemeliharaan, troubleshooting dan pengujian alat.
• Infussion Pump : Dasar teori dan fungsii, pengoperasian, pemeliharaan,
troubleshooting dan pengujian Infussion Pump.
• Syringe Pump : Dasar teori dan fungsii, pengoperasian, pemeliharaan,
troubleshooting dan pengujian Syrnge pump.
• Ventilator: Dasar teori dan fungsi, pengoperasian, pemeliharaan,
troubleshooting dan pengujian ventilator.
• Bubble CPAP: Dasar teori dan fungsii, pengoperasian, pemeliharaan,
troubleshooting dan pengujian ventilator.
• Defibrillator Monitor : Dasar teori dan fungsi, cara pengoperasian,
pemeliharaan dan troubleshooting dan pengujian defibrillator.
• Baby Incubator: Dasar teori dan fungsi, cara pengoperasian, pemeliharaan
dan troubleshooting.
3. Konsep Dasar
Peralatan Life Support dan Life Saving merupakan alat
alat yang dapat mendukung hidup manusia yang
mengalami kegagalan pada fungsi organ vitalnya misal
jantung, paru-paru.
Life support refers to the treatments and techniques
performed in an emergency in order to support life after
the failure of one or more vital organs. Healthcare
providers and emergency medical tecnicians are
generally certified to perform basic and advanced life
support procedures; however, basic life support is
sometimes provided at the scene of an emergency by
family members or bystanders before emergency
services arrive
4. Patient Monitor
A typical high end patient monitor system has 5 basic
subsystems: ECG, pulse oximetry, blood pressure, body
temperature and respiration. Typically the most critical
components in each system are the sensor circuits.
Each module uses a different sensor and signal
conditioning circuit. For example, the ECG uses electrodes
to measure the electric pulse from the heart. The pulse
oximetry (SpO2) uses a light-emitting diode and light
sensor to measure oxygen content. Blood pressure is
typically measured using a piezo-resistive pressure
transducer.
5. Oksimetri
Oksimeter adalah alat yang sangat diperlukan
oleh klinisi yang melakukan tindakan/
intervensi bedah untuk mengetahui saturasi
oksigen dalam darah. Komponen oksimeter ini
biasanya diintegrasikan dalam peralatan
monitor yang tampil dan menjadi panduan
dokter ahli bedah dan anestesi di ruang
operasi untuk memantau kondisi pasien.
Oksimetri termasuk alat kategori non invasiv
Fungsinya mengamati titik jenuh oksigen di
pembuluh darah arteri.
6. Oksimetri menggunakan light emitting diode
berwarna merah dan infra red yang tidak
nampak,alat ini menampilkan frekuensi
denyut jantung dan saturasi
oksigen,parameter yang menjadi andalan dan
sangat berguna untuk mengetahui kondisi
pasien saat pemeriksaan dan atau
intervensi/pembedahan.
18. BLOOD PRESSURE
In blood pressure biometric modules, the most
critical function is the pressure sensor circuit. Here
precision amplifiers are used to detect very small
signals from the transducer and amplify them to a
level suitable for ADC processing. This is typically
followed by an active filter to limited unwanted
noise at higher frequencies. Amplifiers with low
noise, low drift and high gain are necessary to
minimize measurement errors and ensure accurate
readings.
31. • Durasi
• Polaritas
• Magnitude
• Heart Rate : bpm (beat per minute)
RR
60
BPM
DATA ECG YANG PENTING :
32.
33.
34. Pola Grafik EKG
• Gelombang P: depolarisasi atrium
• Gelombang Q: depolarisasi di berkas his
• Gelombang R: depolarisasi menyebar dr
bgn dalam ke bgn luar dasar ventrikel
• Segmen PR: waktu yg dibutuhkan oleh
impuls dari SA node ke AV node; terjadi
perlambatan AV node
• Gelombang S: depolarisasi menyebar
naik dr bgn dasar ventrikel
• Kompleks QRS: depolarisasi ventrikel
• Segmen ST: waktu sejak akhir
depolarisasi ventrikel sebelum terjadi
repolarisasi (fase plateau); saat tjd
kontraksi & pengosongan ventrikel
• Gelombang T: repolarisasi atrium
• Interval TP: waktu saat terjadinya
relaksasi & pengisian ventrikel
35. SPESIFIKASI ECG
• Impedansi input : > 5 MW
• Response frekuensi : + 0,5 dB -3 dB (0,14 Hz –
25 Hz hingga 100 Hz)
AMPLITUDA GELOMBANG ECG
38. Troubleshooting ECG
No Fault Possible Cause Solution
1
ECG traces
have artifacts
or base line
drift
Improper
grounding
Try with battery power only. If the
recording improves than problem is with
grounding.
Check the grounding.
Power the machine from another outlet
with proper electrical ground
2
ECG traces
have artifacts in
one or more
traces but no
on all traces
Improper
electrode
connection with
patient or
problem with
the ECG cable
Check the patient cable continuity with
continuity tester . Replace cable if found
faulty
Check the electrode expiration date.
Check patient skin preparation .
Check limb electrodes end chest
electrode for damage replace if
necessary.
3
Paper feed not
advancing
Incorrect paper
loading
Use instructions to reload paper
39. 4
Printing not
clear or not
uniform
Printing
head
problem
Adjust the printing head
temperature or position.
Clean the printing head with
head cleaner. If no
improvement , replace the
printing head.
Check the paper roller and
replace if not smooth
5
The machine
shuts down
after a few
minutes
while on
battery
power
Problem
with battery
or charging
circuit
Recharge the unit overnight
If there is no improvement
then replace the battery
If still no improvement refer
to technician