Lecture 2
External Beam Therapy (EBT)
Most people think of radiation therapy as coming from a
machine outside of the body, but radiation therapy can be
given in a number of ways. Sometimes radiation is given
more than one way at the same time, or different types of
radiation may be given one after the other. Some ways
radiation can be given include:
1- Eternal radiation therapy
2- Internal radiation therapy
Lecture 2
External radiation therapy
Definition of External beam radiation
(EBRT)
External beam therapy (EBT), also called external
radiation therapy,
External beam radiation (EBRT) is the most widely
used type of radiation therapy, and it most often
uses photon beams.
The radiation comes from a machine outside the
body and is focused on the cancer. (EBRT) is a
major treatment modality in radiation therapy.
Definition of External beam radiation
(EBRT)
• It utilizes ionizing radiation sources outside a patient’s body to treat
(mostly) malignant conditions in the patient.
• Ionizing radiations used in EBRT include photons, electrons, protons,
neutrons, and heavy ions.
• External photon beam radiotherapy is usually carried out with more
than one radiation beam in order to achieve a uniform dose
distribution inside the target volume and an as low as possible a
dose in healthy tissues surrounding the target. ICRU International
Commission on Radiation unit Report No. 50 recommends a target
dose uniformity within +7% and –5% of the dose delivered to a well
defined prescription point within the target
Treatment machines for external beam
radiotherapy
introduction
Introduction
The study and use of ionizing radiation in medicine
started with three important discoveries:
• X rays by Wilhelm Roentgen in 1895.
• Natural radioactivity by Henri Becquerel in 1896.
• Radium-226 by Pierre and Marie Curie in 1898.
Introduction
• The invention of the 60Co
teletherapy unit by H.E. Johns
in Canada in the early 1950s
provided a tremendous boost in
the quest for higher photon
energies and placed the cobalt
unit at the forefront of
radiotherapy for a number of
years.
• Most modern cobalt therapy
machines are arranged on a
gantry so that the source may
rotate about a horizontal axis
referred to as the machine
isocentre axis.
Introduction
• Development of medical accelerator
eclipsed cobalt units and became the most
widely used as a radiation source (photons
& electrons)in modern radiotherapy, as it
cover a wide range of energies.
• More exotic particles, such as protons,
neutrons, heavy ions and negative, all
produced by special accelerators, are also
sometimes used for radiotherapy.
X RAY BEAMS AND X RAY UNITS
X RAY BEAMS AND X RAY UNITS
Clinical X ray beams typically range in energy between 10
kV and 50 MV and are produced when electrons with
kinetic energies between 10 keV and 50 MeV are
decelerated in special metallic targets.
In the target most of the electron’s kinetic energy is
transformed into heat, and a small fraction of the kinetic
energy is emitted in the form of x ray photons which are
divided into two categories:
• Characteristic x rays
• Bremsstrahlung photons
• Result from Coulomb interactions between the incident
electrons and atomic orbital electrons of the target
material (collision loss).
• In a given Coulomb interaction between the incident
electron and an orbital electron, the orbital electron is
ejected from its shell and an electron from a higher level
shell fills the resulting orbital vacancy.
• The energy difference between the two shells may either
be
A. Emitted from the atom in the form of a characteristic
photon (characteristic X ray) or
B. Transferred to an orbital electron that is ejected from
the atom as an Auger electron.
1- Characteristic X rays
2. Bremsstrahlung (continuous) X rays
• Bremsstrahlung X rays result from Coulomb
interactions between the incident electron and the
nuclei of the target material.
• During the Coulomb interaction between the
incident electron and the nucleus, the incident
electron is decelerated and loses part of its kinetic
energy in the form of bremsstrahlung photons .
• The bremsstrahlung spectrum produced in a given
X ray target depends on: the kinetic energy of the
incident electron as well as on the thickness and
atomic number Z of the target.
X rays are used in diagnostic radiology for diagnosis
of disease and in radiation oncology (radiotherapy)
for treatment of disease.
Superficial X rays,
X rays produced by electrons with kinetic energies
between 10 keV and 100 keV .
Orthovoltage
X rays produced by electrons with kinetic energies
between 100 keV and500 keV .
Megavoltage X rays
X rays produced by electrons with kinetic energies
above 1 MeV.
Clinical X ray beams
Superficial and orthovoltage X rays are produced
with X ray tubes(machines),
while megavoltage X rays are most commonly
produced with linacs and sometimes with betatrons
and microtrons.
X ray machines for radiotherapy
Superficial and orthovoltage X rays used in radiotherapy are
produced with X ray machines.
The main components of a radiotherapeutic X ray machine
are
 X-ray tube
 Target cooling system
 Control console
 X-ray power generator
Medical Linear Accelerators in
Radiation Therapy
introduction
What is a Linear Accelerator?
• A linear accelerator is a device that uses high
Radio-Frequency (RF)electromagnetic waves to
accelerate charged particle (i.e. electrons) to high
energies in a linear path, inside a tube like
structure called the accelerator waveguide.
• This is the most common device to treat cancer
with external beam radiation.
How does it work?
• The linear accelerator (Linac), uses microwave
technology to accelerate electrons in a part of the linac
called waveguide, then allows these electrons to
collide with a heavy metal target. As a result of these
collisions, high energy X-Rays (Photons) are produced
from the target.
• These high energy photons will be directed to the
patient's tumor and shaped as they exit the linac to
conform to the shape of the tumor. •
• Radiation can be delivered to the tumor from any angle
by rotating the gantry and moving the treatment
couch.
Linac generations:
• During the past 40 years, medical linacs have gone through five
distinct generations, each one increasingly more sophisticated:
1. Low energy x rays (4-6 MV).
2. Medium energy x rays (10-15 MV) and electrons.
3. High energy x rays (18-25 MV) and electrons.
4. High energy photons and electrons: computer controlled
Operation; dynamic wedge; electronic portal imaging
device (EPID); multileaf collimator (MLC).
5. High energy photons and electrons: photon beam intensity
modulation with MLC; full dynamic conformal dose delivery
with intensity modulated beams produced with an MLC.
Safety of linac installations
The complexity of modern linacs raises concerns as to safety of
operation from the point of view of patients and operators. The
International Electro technical Commission (IEC) document addresses
three categories of safety issues
1. Mechanical
2. Electrical
3. Radiation
And establishes specific requirements mainly for the manufacturers
of linacs in the design and construction of linacsfor use in
radiotherapy. It also covers some radiation safety aspects of linac
installation in customer’s treatment rooms.
Components of modern linacs
Linacs are usually mounted isocentrically and the
operational systems are distributed over five
major and distinct sections of the machine, the:
● Gantry;
● Gantry stand or support;
● Modulator and power supply
● Patient support assembly (i.e. treatment table);
● Control console.
Gantry:
Helps direct the X-ray (photons) or electron beams to the
tumor.
• It rotates 360 degrees around a line/point, called the
Isocenter.
• The gantry has the following components:
1. Electron Gun
2. Accelerator guide.
3. Treatment Head.
Gantry:
Schematic of the treatment head of a modern linear accelerator
operating in photon-production mode (A) or electron-production
mode (B).
Q1 : Choose the correct answer
The machine used for radiotherapy is called as _________
a) Therapeutic X – Ray machine
b) Linear Accelerator
c) Radioactive Ultrasound Machine
-------- result from Coulomb interactions between the incident electron and the nuclei of
the target material
a) Characteristic x-ray
b) Gamma Rays
c) Bremsstrahlung Radiation
The radiation comes from a machine outside the body and is focused on the cancer called
a) External radiotherapy
b) Internal radiotherapy
c) Thermotherapy
Result from Coulomb interactions between the incident electrons and atomic orbital
electrons of the target material
a) Characteristic x-ray
b) Gamma Rays
c) Bremsstrahlung Radiation
Question 2: Write the suitable expression for the following definition:
• A. The main components of a radiotherapeutic X ray machine are
1-………………
2-…………………..
3-…………………..
4-………………….
• X ray target depends on ………………….., ……………….,……………..
• Ionizing radiations used in EBRT include …………., ………., ……………., ……….., and ……………….
• The radiation comes from a machine outside the body and is focused on the cancer called
………………………..
……………………………… result from Coulomb interactions between the incident electrons
and atomic orbital electrons of the target material.
• The kinetic energy is emitted in the form of x ray photons which are divided into two
categories: …………………………………….. ,…………………………………………
• ICRU International Commission on Radiation unit Report No. 50 recommends a target dose
uniformity within ……..% and ………% of the dose delivered to a well defined prescription
point within the target.
• Transferred to an orbital electron that is ejected from the atom as an ………………..
• ………………..result from Coulomb interactions between the incident electron and the nuclei
of the target material.
• A. International Electro technical Commission (IEC) document addresses three
categories of safety issues
……………..
………………
……………..
• ................Helps direct the X-ray (photons) or electron beams to the tumor.
• The main components of Gantry in LINAC are
………………….
…………………..
…………………..
• D. ………………It is responsible for producing electrons and injecting them into the
accelerator structure.
• E. ……………….. Changes the direction of the electron beam, downwards toward the
patient .
• F. ………………. is required to create a flattened beam of sufficient area, uniformity and
symmetry
Question 3: Explain The Following
1. Superficial X rays,
2. Orthovoltage
3. Megavoltage X rays
4. EBRT
5. Linear Accelerator and How does it work?
6. Linear generation
7. Components of modern linacs
8. Treatment head of linacs and component it
9. beam flatting filter and scattering foil
Question4 : Arrange the given words on the suitable space of the following figure?
(Gantry, electron gun, , stand , Isocenter, treatment table, Accelerating wave , X-ray
target, electron beam transport and RF Power generation)
Use some videos to explain the linear
accelerator
• https://www.youtube.com/watch?v=jSgnWfb
Ex1A
• https://www.youtube.com/watch?v=lZ9cGVax
Oes
• https://www.youtube.com/watch?v=hy9atKA
qAf4&list=PLk59CaRJX5aZWsX5hyLmQQNGfKE
H5iY-J&index=3
Thank you for your attention

Lecture (2)

  • 1.
  • 2.
    Most people thinkof radiation therapy as coming from a machine outside of the body, but radiation therapy can be given in a number of ways. Sometimes radiation is given more than one way at the same time, or different types of radiation may be given one after the other. Some ways radiation can be given include: 1- Eternal radiation therapy 2- Internal radiation therapy
  • 3.
  • 4.
  • 5.
    Definition of Externalbeam radiation (EBRT) External beam therapy (EBT), also called external radiation therapy, External beam radiation (EBRT) is the most widely used type of radiation therapy, and it most often uses photon beams. The radiation comes from a machine outside the body and is focused on the cancer. (EBRT) is a major treatment modality in radiation therapy.
  • 6.
    Definition of Externalbeam radiation (EBRT) • It utilizes ionizing radiation sources outside a patient’s body to treat (mostly) malignant conditions in the patient. • Ionizing radiations used in EBRT include photons, electrons, protons, neutrons, and heavy ions. • External photon beam radiotherapy is usually carried out with more than one radiation beam in order to achieve a uniform dose distribution inside the target volume and an as low as possible a dose in healthy tissues surrounding the target. ICRU International Commission on Radiation unit Report No. 50 recommends a target dose uniformity within +7% and –5% of the dose delivered to a well defined prescription point within the target
  • 7.
    Treatment machines forexternal beam radiotherapy
  • 8.
  • 9.
    Introduction The study anduse of ionizing radiation in medicine started with three important discoveries: • X rays by Wilhelm Roentgen in 1895. • Natural radioactivity by Henri Becquerel in 1896. • Radium-226 by Pierre and Marie Curie in 1898.
  • 10.
    Introduction • The inventionof the 60Co teletherapy unit by H.E. Johns in Canada in the early 1950s provided a tremendous boost in the quest for higher photon energies and placed the cobalt unit at the forefront of radiotherapy for a number of years. • Most modern cobalt therapy machines are arranged on a gantry so that the source may rotate about a horizontal axis referred to as the machine isocentre axis.
  • 11.
    Introduction • Development ofmedical accelerator eclipsed cobalt units and became the most widely used as a radiation source (photons & electrons)in modern radiotherapy, as it cover a wide range of energies. • More exotic particles, such as protons, neutrons, heavy ions and negative, all produced by special accelerators, are also sometimes used for radiotherapy.
  • 12.
    X RAY BEAMSAND X RAY UNITS
  • 13.
    X RAY BEAMSAND X RAY UNITS Clinical X ray beams typically range in energy between 10 kV and 50 MV and are produced when electrons with kinetic energies between 10 keV and 50 MeV are decelerated in special metallic targets. In the target most of the electron’s kinetic energy is transformed into heat, and a small fraction of the kinetic energy is emitted in the form of x ray photons which are divided into two categories: • Characteristic x rays • Bremsstrahlung photons
  • 14.
    • Result fromCoulomb interactions between the incident electrons and atomic orbital electrons of the target material (collision loss). • In a given Coulomb interaction between the incident electron and an orbital electron, the orbital electron is ejected from its shell and an electron from a higher level shell fills the resulting orbital vacancy. • The energy difference between the two shells may either be A. Emitted from the atom in the form of a characteristic photon (characteristic X ray) or B. Transferred to an orbital electron that is ejected from the atom as an Auger electron. 1- Characteristic X rays
  • 15.
    2. Bremsstrahlung (continuous)X rays • Bremsstrahlung X rays result from Coulomb interactions between the incident electron and the nuclei of the target material. • During the Coulomb interaction between the incident electron and the nucleus, the incident electron is decelerated and loses part of its kinetic energy in the form of bremsstrahlung photons . • The bremsstrahlung spectrum produced in a given X ray target depends on: the kinetic energy of the incident electron as well as on the thickness and atomic number Z of the target.
  • 16.
    X rays areused in diagnostic radiology for diagnosis of disease and in radiation oncology (radiotherapy) for treatment of disease. Superficial X rays, X rays produced by electrons with kinetic energies between 10 keV and 100 keV . Orthovoltage X rays produced by electrons with kinetic energies between 100 keV and500 keV . Megavoltage X rays X rays produced by electrons with kinetic energies above 1 MeV. Clinical X ray beams
  • 17.
    Superficial and orthovoltageX rays are produced with X ray tubes(machines), while megavoltage X rays are most commonly produced with linacs and sometimes with betatrons and microtrons.
  • 18.
    X ray machinesfor radiotherapy Superficial and orthovoltage X rays used in radiotherapy are produced with X ray machines. The main components of a radiotherapeutic X ray machine are  X-ray tube  Target cooling system  Control console  X-ray power generator
  • 20.
    Medical Linear Acceleratorsin Radiation Therapy
  • 21.
  • 24.
    What is aLinear Accelerator? • A linear accelerator is a device that uses high Radio-Frequency (RF)electromagnetic waves to accelerate charged particle (i.e. electrons) to high energies in a linear path, inside a tube like structure called the accelerator waveguide. • This is the most common device to treat cancer with external beam radiation.
  • 25.
    How does itwork? • The linear accelerator (Linac), uses microwave technology to accelerate electrons in a part of the linac called waveguide, then allows these electrons to collide with a heavy metal target. As a result of these collisions, high energy X-Rays (Photons) are produced from the target. • These high energy photons will be directed to the patient's tumor and shaped as they exit the linac to conform to the shape of the tumor. • • Radiation can be delivered to the tumor from any angle by rotating the gantry and moving the treatment couch.
  • 29.
    Linac generations: • Duringthe past 40 years, medical linacs have gone through five distinct generations, each one increasingly more sophisticated: 1. Low energy x rays (4-6 MV). 2. Medium energy x rays (10-15 MV) and electrons. 3. High energy x rays (18-25 MV) and electrons. 4. High energy photons and electrons: computer controlled Operation; dynamic wedge; electronic portal imaging device (EPID); multileaf collimator (MLC). 5. High energy photons and electrons: photon beam intensity modulation with MLC; full dynamic conformal dose delivery with intensity modulated beams produced with an MLC.
  • 30.
    Safety of linacinstallations The complexity of modern linacs raises concerns as to safety of operation from the point of view of patients and operators. The International Electro technical Commission (IEC) document addresses three categories of safety issues 1. Mechanical 2. Electrical 3. Radiation And establishes specific requirements mainly for the manufacturers of linacs in the design and construction of linacsfor use in radiotherapy. It also covers some radiation safety aspects of linac installation in customer’s treatment rooms.
  • 31.
    Components of modernlinacs Linacs are usually mounted isocentrically and the operational systems are distributed over five major and distinct sections of the machine, the: ● Gantry; ● Gantry stand or support; ● Modulator and power supply ● Patient support assembly (i.e. treatment table); ● Control console.
  • 33.
  • 34.
    Helps direct theX-ray (photons) or electron beams to the tumor. • It rotates 360 degrees around a line/point, called the Isocenter. • The gantry has the following components: 1. Electron Gun 2. Accelerator guide. 3. Treatment Head. Gantry:
  • 50.
    Schematic of thetreatment head of a modern linear accelerator operating in photon-production mode (A) or electron-production mode (B).
  • 51.
    Q1 : Choosethe correct answer The machine used for radiotherapy is called as _________ a) Therapeutic X – Ray machine b) Linear Accelerator c) Radioactive Ultrasound Machine -------- result from Coulomb interactions between the incident electron and the nuclei of the target material a) Characteristic x-ray b) Gamma Rays c) Bremsstrahlung Radiation The radiation comes from a machine outside the body and is focused on the cancer called a) External radiotherapy b) Internal radiotherapy c) Thermotherapy Result from Coulomb interactions between the incident electrons and atomic orbital electrons of the target material a) Characteristic x-ray b) Gamma Rays c) Bremsstrahlung Radiation
  • 52.
    Question 2: Writethe suitable expression for the following definition: • A. The main components of a radiotherapeutic X ray machine are 1-……………… 2-………………….. 3-………………….. 4-…………………. • X ray target depends on ………………….., ……………….,…………….. • Ionizing radiations used in EBRT include …………., ………., ……………., ……….., and ………………. • The radiation comes from a machine outside the body and is focused on the cancer called ……………………….. ……………………………… result from Coulomb interactions between the incident electrons and atomic orbital electrons of the target material. • The kinetic energy is emitted in the form of x ray photons which are divided into two categories: …………………………………….. ,………………………………………… • ICRU International Commission on Radiation unit Report No. 50 recommends a target dose uniformity within ……..% and ………% of the dose delivered to a well defined prescription point within the target. • Transferred to an orbital electron that is ejected from the atom as an ……………….. • ………………..result from Coulomb interactions between the incident electron and the nuclei of the target material.
  • 53.
    • A. InternationalElectro technical Commission (IEC) document addresses three categories of safety issues …………….. ……………… …………….. • ................Helps direct the X-ray (photons) or electron beams to the tumor. • The main components of Gantry in LINAC are …………………. ………………….. ………………….. • D. ………………It is responsible for producing electrons and injecting them into the accelerator structure. • E. ……………….. Changes the direction of the electron beam, downwards toward the patient . • F. ………………. is required to create a flattened beam of sufficient area, uniformity and symmetry
  • 54.
    Question 3: ExplainThe Following 1. Superficial X rays, 2. Orthovoltage 3. Megavoltage X rays 4. EBRT 5. Linear Accelerator and How does it work? 6. Linear generation 7. Components of modern linacs 8. Treatment head of linacs and component it 9. beam flatting filter and scattering foil
  • 55.
    Question4 : Arrangethe given words on the suitable space of the following figure? (Gantry, electron gun, , stand , Isocenter, treatment table, Accelerating wave , X-ray target, electron beam transport and RF Power generation)
  • 56.
    Use some videosto explain the linear accelerator • https://www.youtube.com/watch?v=jSgnWfb Ex1A • https://www.youtube.com/watch?v=lZ9cGVax Oes • https://www.youtube.com/watch?v=hy9atKA qAf4&list=PLk59CaRJX5aZWsX5hyLmQQNGfKE H5iY-J&index=3
  • 57.
    Thank you foryour attention