This document discusses key concepts in radiation dosimetry for medical imaging, including:
1) It defines common radiation protection terms like absorbed dose, equivalent dose, effective dose and explains their appropriate and limited uses for evaluating patient risk from medical imaging.
2) It notes that effective dose should not be used to assess risk to individuals from medical exposures and that organ doses are more relevant.
3) It discusses how uncertainty increases at each step moving from measurement of organ doses to estimation of effective dose and risk, limiting the usefulness of effective dose for medical applications.
4) For medical imaging, more detailed knowledge of organ doses, dose distributions and patient factors is needed than just effective dose alone to properly assess radiation
The document discusses the International Commission on Radiological Protection (ICRP), which sets standards for radiation protection. The ICRP relies on the linear no-threshold model to establish dose limits for workers and the public. This model assumes that any amount of radiation exposure increases cancer risk proportionally. The ICRP cites data from studies of atomic bomb survivors and other exposed groups to determine that radiation carries a 5% increased risk of cancer per sievert of lifetime dose. Using this risk factor, the ICRP calculates annual dose limits of 20 millisieverts for occupational workers and 1 millisievert for members of the public. Though other models question the linear no-threshold model, the ICRP maintains it is a
Radiation protection course for radiologists L7Amin Amin
1) The document summarizes key aspects of radiation monitoring and protection for radiologists, including recommended dose limits, monitoring of exposed personnel, and design of radiation facilities.
2) It discusses wearing personal dosimeters under or over lead aprons to accurately measure radiation exposure to covered and uncovered body parts.
3) The document also outlines what should be included in radiation protection surveys and reports, such as equipment details, radiation measurements, and recommendations.
The document discusses recommendations from ICRP 60 & 103 regarding radiation protection. It begins with background on natural and artificial radiation sources and their effects. It then summarizes the evolution of ICRP recommendations over time, from early annual dose limits of 1000 mSv reduced gradually to current limits. Key concepts discussed include justification of practices, optimization of protection, and application of dose limits. Occupational, public, and medical exposure dose limits are provided. ICRP 103 introduced changes like new tissue weighting factors and computational phantoms.
Radiation is energy that is given off by particular materials and devices.
Radiation protection, also known as radiological protection, is defined by the International Atomic Energy Agency (IAEA) as "The protection of people from harmful effects of exposure to ionizing radiation, and the means for achieving this". Exposure can be from a source of radiation external to the human body or due to internal irradiation caused by the ingestion of radioactive contamination
This document discusses central axis depth doses in water for both SSD and SAD techniques. For SSD technique:
- Percentage depth dose (PDD) curves measure attenuation at different depths and are affected by beam quality, field size, and SSD.
- Buildup region occurs as secondary electrons deposit energy downstream, increasing dose with depth until maximum.
- Depth dose maximum (zmax) depends on beam energy and field size.
- PDD increases with larger field sizes due to increased scatter radiation.
- PDD increases with longer SSD due to the inverse square law of radiation intensity.
Radiation protection involves protecting people from harmful effects of ionized radiation. Sources of radiation exposure include occupational, non-occupational natural sources like radon and cosmic radiation, and man-made sources like medical radiation. The biological effects of radiation include prompt personal effects from high doses occurring within days, delayed personal effects from chronic low doses like cancer, and racial effects from hereditary changes. Permissible radiation doses are regulated and the quantities used to measure radiation include activity, exposure, absorbed dose, and dose equivalent. Effective dose equivalent considers tissue sensitivity and radiation type to measure biological damage risk from radiation exposure.
The document discusses the International Commission on Radiological Protection (ICRP), which sets standards for radiation protection. The ICRP relies on the linear no-threshold model to establish dose limits for workers and the public. This model assumes that any amount of radiation exposure increases cancer risk proportionally. The ICRP cites data from studies of atomic bomb survivors and other exposed groups to determine that radiation carries a 5% increased risk of cancer per sievert of lifetime dose. Using this risk factor, the ICRP calculates annual dose limits of 20 millisieverts for occupational workers and 1 millisievert for members of the public. Though other models question the linear no-threshold model, the ICRP maintains it is a
Radiation protection course for radiologists L7Amin Amin
1) The document summarizes key aspects of radiation monitoring and protection for radiologists, including recommended dose limits, monitoring of exposed personnel, and design of radiation facilities.
2) It discusses wearing personal dosimeters under or over lead aprons to accurately measure radiation exposure to covered and uncovered body parts.
3) The document also outlines what should be included in radiation protection surveys and reports, such as equipment details, radiation measurements, and recommendations.
The document discusses recommendations from ICRP 60 & 103 regarding radiation protection. It begins with background on natural and artificial radiation sources and their effects. It then summarizes the evolution of ICRP recommendations over time, from early annual dose limits of 1000 mSv reduced gradually to current limits. Key concepts discussed include justification of practices, optimization of protection, and application of dose limits. Occupational, public, and medical exposure dose limits are provided. ICRP 103 introduced changes like new tissue weighting factors and computational phantoms.
Radiation is energy that is given off by particular materials and devices.
Radiation protection, also known as radiological protection, is defined by the International Atomic Energy Agency (IAEA) as "The protection of people from harmful effects of exposure to ionizing radiation, and the means for achieving this". Exposure can be from a source of radiation external to the human body or due to internal irradiation caused by the ingestion of radioactive contamination
This document discusses central axis depth doses in water for both SSD and SAD techniques. For SSD technique:
- Percentage depth dose (PDD) curves measure attenuation at different depths and are affected by beam quality, field size, and SSD.
- Buildup region occurs as secondary electrons deposit energy downstream, increasing dose with depth until maximum.
- Depth dose maximum (zmax) depends on beam energy and field size.
- PDD increases with larger field sizes due to increased scatter radiation.
- PDD increases with longer SSD due to the inverse square law of radiation intensity.
Radiation protection involves protecting people from harmful effects of ionized radiation. Sources of radiation exposure include occupational, non-occupational natural sources like radon and cosmic radiation, and man-made sources like medical radiation. The biological effects of radiation include prompt personal effects from high doses occurring within days, delayed personal effects from chronic low doses like cancer, and racial effects from hereditary changes. Permissible radiation doses are regulated and the quantities used to measure radiation include activity, exposure, absorbed dose, and dose equivalent. Effective dose equivalent considers tissue sensitivity and radiation type to measure biological damage risk from radiation exposure.
This document provides a history of radiation and radiotherapy from its discoveries in the late 19th century to modern applications. It describes how radiation was discovered in 1895 by Roentgen and how Becquerel discovered radioactive elements. It then outlines the early uses of radiation in medicine by Grubbe and how fractionation was developed. It discusses the advancements of technologies like cobalt-60 units, linear accelerators, 3D conformal radiotherapy, IMRT and IGRT that have allowed more precise targeting of radiation doses to tumors.
The document discusses the history and development of artificial intelligence over the past 70 years. It outlines some of the key milestones in AI research from the early work in the 1950s to modern advances in deep learning. While progress has been made, fully general artificial intelligence that can match or exceed human levels of intelligence remains an ongoing challenge that researchers continue working to achieve.
Electron beam therapy uses accelerated electrons to treat superficial tumors. Electrons interact with matter through inelastic collisions that cause ionization and excitation, and elastic collisions that scatter the electrons. This gives electron beams a characteristically sharp dose drop-off beyond the tumor depth. Key applications of electron beams include treatment of skin cancers, chest wall irradiation for breast cancer, and boost doses to lymph nodes.
This document summarizes the expectations and key learnings from a linear accelerator acceptance, commissioning, and annual QA training that occurred from September to November 2008. The training covered:
1. Fundamental concepts of linear accelerators, beam production, safety features, and the acceptance testing process.
2. Techniques for collecting beam data needed for commissioning, including measurements and data definitions.
3. Procedures for linear accelerator QA and other treatment machine QA on an annual basis.
Key topics included the beamline components that produce photon and electron beams, characteristics of linear accelerator beams, the importance of acceptance testing and commissioning the machine properly, and techniques for annual QA tests.
Radiation Dose Units and Dose Limits- Avinesh ShresthaAvinesh Shrestha
Describes different units of radiation dose and the dose limits in diagnostic radiology imaging. Discuses different radiation units described by ICRU. Describes different radiation dose limits given by different organizations like ICRP, NCRP, AERB.
This document provides information on medical radiation safety. It discusses natural and man-made sources of radiation exposure, units used to measure radiation doses, and key principles of radiation protection including minimizing time, distance, and shielding. The document also covers radiation risks and perceptions, dose limits for occupational exposure, and requirements for radioactive waste management programs.
In 2000 IAEA published another International Code of Practice.
“Absorbed Dose Determination in External Beam Radiotherapy” (Technical Report Series No. 398)
Recommending procedures to obtain the absorbed dose in water from measurements made with an ionisation chamber in external beam radiotherapy (EBRT).
This document discusses optically stimulated luminescence (OSL) dosimetry and its applications in radiotherapy. It provides an overview of OSL principles, readers, and stimulation methods. Aluminum oxide (Al2O3:C) OSL dosimeters are commonly used and have good dosimetric characteristics including dose linearity, minimal energy dependence, and suitability for clinical measurements. The document reviews dosimeter characteristics, advantages, clinical applications in phantom and in vivo measurements, and concludes that OSL dosimeters can provide accurate dosimetric data for a variety of radiotherapy uses.
Cavity theory.. Radiotherapy..
I explained about Bragg-gray, Spencer attix and Burlin theory..
In future I'll try to explain this with some more points. So wait for the updation.
I referred Radiation oncology (IAEA) book and
Introduction to Radiological Physics and Radiation Dosimetry by Frank Herbert Attix book
Electron beam therapy uses electrically charged particles called electrons that are generated by a linear accelerator to treat superficial cancers. It deposits dose uniformly from the surface to a specific depth before rapidly falling off, sparing deeper tissues. Electron energies up to 20 MeV can treat disease within 6 cm of the surface. Accessories like applicators, cutouts, bolus and internal shields are used to shape the beam for treatment fields and protect healthy tissues. Precise dose specification and reporting is important for electron therapy due to the rapid dose fall-off and higher skin doses compared to prescription depth.
This document discusses radiation protection and provides definitions, types of radiation effects, sources of radiation exposure, units of measurement, dose limits, and techniques to reduce radiation exposure in medical imaging. It defines radiation protection as protecting people from harmful effects of ionizing radiation. It describes stochastic and deterministic effects and lists examples of radiation anomalies. It also outlines regulatory bodies, dose limits for occupational workers and the public, and principles of radiation safety including time, distance, shielding and reducing exposure.
This document provides an overview of key concepts in radiobiology for radiotherapy. It discusses the biological effects of ionizing radiation, including deterministic effects which have a dose threshold and include tissue injuries, and stochastic effects which have no threshold and include cancer induction. Fractionation of radiation doses is explained, along with the 5 R's that influence radiation response: repair, repopulation, reoxygenation, redistribution, and radiosensitivity. Direct and indirect radiation actions on DNA are also summarized.
This document discusses concepts and instruments used in dosimetry. It defines key terms like absorbed dose, exposure, and kerma. It explains dosimetry protocols like TG-51 and TRS-398 which provide standards for calibrating dosimeters. Common dosimeters discussed include ionization chambers like thimble chambers and parallel-plate chambers, as well as Geiger-Muller counters. Calibration of dosimeters involves various correction factors to account for influences like temperature, pressure and polarity.
Computed Tomography Dose Index, Includes various CTDI parameters and the way of calculating effective dose from various Computed Tomography procedures along with their conversion factor.
This document provides an overview of key concepts in radiation protection for diagnostic radiology, including:
- Medical exposure involves exposing patients for diagnosis or treatment, following principles of justification and optimization.
- Justification involves assessing if a procedure does more good than harm at the individual, generic, and general levels.
- Optimization aims to keep patient doses as low as reasonably achievable given image quality needs.
- Guidance levels indicate typical dose levels and help identify unusually high exposures requiring review. They are not dose limits.
The document discusses absolute dosimetry in radiation therapy. It defines absolute dosimetry as a direct measurement of ionization or absorbed dose under standard conditions, such as calorimetry or measuring electrons released, without requiring calibration. Absolute dosimetry provides a known dose value that other measurements can be compared to. The document also outlines concepts and factors involved in radiation dosimetry, such as temperature and pressure corrections, and discusses the use of phantoms made of materials like water, solid water and PMMA.
This document discusses different types of radiation dosimeters, including their properties, uses, and limitations. It focuses on ionization chamber dosimetry systems, describing cylindrical thimble chambers, parallel plate chambers, spherical chambers, well chambers, and pencil chambers. It also discusses thermoluminescent dosimeters (TLD), noting that TLD crystals absorb and trap radiation energy then emit light when heated, allowing dose measurement. The document provides an overview of key factors such as linearity, dose rate dependence, energy dependence, and directional dependence for dosimeters.
This document discusses radiation protection for medical workers. It explains that radiologists, radiographers and other medical staff are subject to some radiation exposure through their work, so radiation protection aims to keep doses as low as possible to prevent health risks. It describes different types of radiation effects and dose limits for workers. Various methods of radiation monitoring are outlined, including use of dosimeters like film badges, pocket ionization chambers and thermoluminescent dosimeters to measure individual radiation exposures.
1. Electronic Portal Imaging Devices (EPIDs) are imaging devices mounted on linear accelerators opposite the MV x-ray source.
2. EPIDs have a wide variety of applications including real-time patient setup verification during treatment and determining beam blocking shapes and leaf positions.
3. Commercially available EPIDs include scanning liquid-filled ion chamber devices, camera-based devices, and active matrix flat panel detectors. They provide localization quality images with doses less than 3 cGy.
1) Radiation protection standards aim to restrict radiation risks while allowing beneficial uses of radiation. Exposure standards are set to limit stochastic effects probabilistically and prevent deterministic effects.
2) Radiation can cause both stochastic (random) effects like cancer that have no threshold and increase in risk with increasing dose, as well as non-stochastic (deterministic) effects like burns that have thresholds below which they will not occur.
3) Various quantities like absorbed dose, equivalent dose, and effective dose are used to quantify radiation exposure and the associated biological harm from different radiation types. Operational quantities allow direct measurement and monitoring.
This document provides a history of radiation and radiotherapy from its discoveries in the late 19th century to modern applications. It describes how radiation was discovered in 1895 by Roentgen and how Becquerel discovered radioactive elements. It then outlines the early uses of radiation in medicine by Grubbe and how fractionation was developed. It discusses the advancements of technologies like cobalt-60 units, linear accelerators, 3D conformal radiotherapy, IMRT and IGRT that have allowed more precise targeting of radiation doses to tumors.
The document discusses the history and development of artificial intelligence over the past 70 years. It outlines some of the key milestones in AI research from the early work in the 1950s to modern advances in deep learning. While progress has been made, fully general artificial intelligence that can match or exceed human levels of intelligence remains an ongoing challenge that researchers continue working to achieve.
Electron beam therapy uses accelerated electrons to treat superficial tumors. Electrons interact with matter through inelastic collisions that cause ionization and excitation, and elastic collisions that scatter the electrons. This gives electron beams a characteristically sharp dose drop-off beyond the tumor depth. Key applications of electron beams include treatment of skin cancers, chest wall irradiation for breast cancer, and boost doses to lymph nodes.
This document summarizes the expectations and key learnings from a linear accelerator acceptance, commissioning, and annual QA training that occurred from September to November 2008. The training covered:
1. Fundamental concepts of linear accelerators, beam production, safety features, and the acceptance testing process.
2. Techniques for collecting beam data needed for commissioning, including measurements and data definitions.
3. Procedures for linear accelerator QA and other treatment machine QA on an annual basis.
Key topics included the beamline components that produce photon and electron beams, characteristics of linear accelerator beams, the importance of acceptance testing and commissioning the machine properly, and techniques for annual QA tests.
Radiation Dose Units and Dose Limits- Avinesh ShresthaAvinesh Shrestha
Describes different units of radiation dose and the dose limits in diagnostic radiology imaging. Discuses different radiation units described by ICRU. Describes different radiation dose limits given by different organizations like ICRP, NCRP, AERB.
This document provides information on medical radiation safety. It discusses natural and man-made sources of radiation exposure, units used to measure radiation doses, and key principles of radiation protection including minimizing time, distance, and shielding. The document also covers radiation risks and perceptions, dose limits for occupational exposure, and requirements for radioactive waste management programs.
In 2000 IAEA published another International Code of Practice.
“Absorbed Dose Determination in External Beam Radiotherapy” (Technical Report Series No. 398)
Recommending procedures to obtain the absorbed dose in water from measurements made with an ionisation chamber in external beam radiotherapy (EBRT).
This document discusses optically stimulated luminescence (OSL) dosimetry and its applications in radiotherapy. It provides an overview of OSL principles, readers, and stimulation methods. Aluminum oxide (Al2O3:C) OSL dosimeters are commonly used and have good dosimetric characteristics including dose linearity, minimal energy dependence, and suitability for clinical measurements. The document reviews dosimeter characteristics, advantages, clinical applications in phantom and in vivo measurements, and concludes that OSL dosimeters can provide accurate dosimetric data for a variety of radiotherapy uses.
Cavity theory.. Radiotherapy..
I explained about Bragg-gray, Spencer attix and Burlin theory..
In future I'll try to explain this with some more points. So wait for the updation.
I referred Radiation oncology (IAEA) book and
Introduction to Radiological Physics and Radiation Dosimetry by Frank Herbert Attix book
Electron beam therapy uses electrically charged particles called electrons that are generated by a linear accelerator to treat superficial cancers. It deposits dose uniformly from the surface to a specific depth before rapidly falling off, sparing deeper tissues. Electron energies up to 20 MeV can treat disease within 6 cm of the surface. Accessories like applicators, cutouts, bolus and internal shields are used to shape the beam for treatment fields and protect healthy tissues. Precise dose specification and reporting is important for electron therapy due to the rapid dose fall-off and higher skin doses compared to prescription depth.
This document discusses radiation protection and provides definitions, types of radiation effects, sources of radiation exposure, units of measurement, dose limits, and techniques to reduce radiation exposure in medical imaging. It defines radiation protection as protecting people from harmful effects of ionizing radiation. It describes stochastic and deterministic effects and lists examples of radiation anomalies. It also outlines regulatory bodies, dose limits for occupational workers and the public, and principles of radiation safety including time, distance, shielding and reducing exposure.
This document provides an overview of key concepts in radiobiology for radiotherapy. It discusses the biological effects of ionizing radiation, including deterministic effects which have a dose threshold and include tissue injuries, and stochastic effects which have no threshold and include cancer induction. Fractionation of radiation doses is explained, along with the 5 R's that influence radiation response: repair, repopulation, reoxygenation, redistribution, and radiosensitivity. Direct and indirect radiation actions on DNA are also summarized.
This document discusses concepts and instruments used in dosimetry. It defines key terms like absorbed dose, exposure, and kerma. It explains dosimetry protocols like TG-51 and TRS-398 which provide standards for calibrating dosimeters. Common dosimeters discussed include ionization chambers like thimble chambers and parallel-plate chambers, as well as Geiger-Muller counters. Calibration of dosimeters involves various correction factors to account for influences like temperature, pressure and polarity.
Computed Tomography Dose Index, Includes various CTDI parameters and the way of calculating effective dose from various Computed Tomography procedures along with their conversion factor.
This document provides an overview of key concepts in radiation protection for diagnostic radiology, including:
- Medical exposure involves exposing patients for diagnosis or treatment, following principles of justification and optimization.
- Justification involves assessing if a procedure does more good than harm at the individual, generic, and general levels.
- Optimization aims to keep patient doses as low as reasonably achievable given image quality needs.
- Guidance levels indicate typical dose levels and help identify unusually high exposures requiring review. They are not dose limits.
The document discusses absolute dosimetry in radiation therapy. It defines absolute dosimetry as a direct measurement of ionization or absorbed dose under standard conditions, such as calorimetry or measuring electrons released, without requiring calibration. Absolute dosimetry provides a known dose value that other measurements can be compared to. The document also outlines concepts and factors involved in radiation dosimetry, such as temperature and pressure corrections, and discusses the use of phantoms made of materials like water, solid water and PMMA.
This document discusses different types of radiation dosimeters, including their properties, uses, and limitations. It focuses on ionization chamber dosimetry systems, describing cylindrical thimble chambers, parallel plate chambers, spherical chambers, well chambers, and pencil chambers. It also discusses thermoluminescent dosimeters (TLD), noting that TLD crystals absorb and trap radiation energy then emit light when heated, allowing dose measurement. The document provides an overview of key factors such as linearity, dose rate dependence, energy dependence, and directional dependence for dosimeters.
This document discusses radiation protection for medical workers. It explains that radiologists, radiographers and other medical staff are subject to some radiation exposure through their work, so radiation protection aims to keep doses as low as possible to prevent health risks. It describes different types of radiation effects and dose limits for workers. Various methods of radiation monitoring are outlined, including use of dosimeters like film badges, pocket ionization chambers and thermoluminescent dosimeters to measure individual radiation exposures.
1. Electronic Portal Imaging Devices (EPIDs) are imaging devices mounted on linear accelerators opposite the MV x-ray source.
2. EPIDs have a wide variety of applications including real-time patient setup verification during treatment and determining beam blocking shapes and leaf positions.
3. Commercially available EPIDs include scanning liquid-filled ion chamber devices, camera-based devices, and active matrix flat panel detectors. They provide localization quality images with doses less than 3 cGy.
1) Radiation protection standards aim to restrict radiation risks while allowing beneficial uses of radiation. Exposure standards are set to limit stochastic effects probabilistically and prevent deterministic effects.
2) Radiation can cause both stochastic (random) effects like cancer that have no threshold and increase in risk with increasing dose, as well as non-stochastic (deterministic) effects like burns that have thresholds below which they will not occur.
3) Various quantities like absorbed dose, equivalent dose, and effective dose are used to quantify radiation exposure and the associated biological harm from different radiation types. Operational quantities allow direct measurement and monitoring.
Stereotactic body radiotherapy (SBRT) delivers high-dose radiation to tumors in a small number of fractions using high precision. For prostate SBRT, the target and organs at risk are contoured on planning CT. A dose of 35-38Gy in 5 fractions is used as primary treatment for low risk prostate cancer. Rigid image guidance and intrafraction monitoring are important to minimize setup errors. ExacTrac X-ray positioning co-registers X-rays with digitally reconstructed radiographs and corrects for rotational and translational deviations, achieving sub-millimeter accuracy. This allows safe dose escalation for prostate SBRT.
The document discusses the International Commission on Radiation Units and Measurements (ICRU). It summarizes that ICRU defines physical quantities and units related to ionizing radiation. It has published over 85 reports on topics like dose specifications, clinical terms, and recommendations for intensity modulated radiation therapy. ICRU works to establish international standards for radiation quantities, units, and nomenclature used in radiation oncology and other applications of ionizing radiation.
Radiation safety in diagnostic nuclear medicineSGPGIMS
1. Radiation is a form of energy emitted by atoms in the form of electromagnetic waves or particles. Ionizing radiation can eject electrons from atoms and produce ions, while non-ionizing radiation excites electrons.
2. People are exposed to ionizing radiation from natural and man-made sources. Naturally occurring sources include terrestrial radiation, cosmic radiation, and internal radiation. Medical procedures such as CT scans, nuclear medicine exams, and fluoroscopy account for over 90% of man-made radiation exposure.
3. Radiation protection aims to take advantage of the benefits of radiation use while preventing deterministic effects and limiting stochastic effects to acceptable levels. Occupational dose limits are higher than public limits, and some populations like
Radiation safety in diagnostic nuclear medicineSGPGIMS
1. Radiation is a form of energy emitted by atoms in the form of electromagnetic waves or particles. Ionizing radiation can eject electrons from atoms and produce ions, while non-ionizing radiation excites electrons.
2. People are exposed to ionizing radiation from natural and man-made sources. Naturally occurring sources include terrestrial radiation, cosmic radiation, and internal radiation. Medical procedures such as CT scans, nuclear medicine exams, and fluoroscopy account for over 90% of man-made radiation exposure.
3. Radiation protection aims to take advantage of the benefits of radiation use while preventing deterministic effects and limiting stochastic effects to acceptable levels. Occupational dose limits are higher than public limits, and some populations like
CT Dose Issues.pptx on the factors to be considered on radiation protectionsanyengere
summary, mobile radiography allows for the diagnostic imaging of patients who are unable to be seen in the X-ray examination room. Therefore, mobile X-ray equipment is useful for patients who have difficulty with movement. However, staff are exposed to scattered radiation from the patient, and can receive potentially harmful radiation doses during radiography. The protection of staff is of utmost importance; therefore, we investigated the occupational radiation doses received by RTs, particularly eye doses, using phantom measurements. RTs can be located close to a patient (i.e., the source of scattered radiation) during mobile radiography. As eye doses can be significant, protective measures are essential for RTs. Protective aprons are important for protecting RTs, as is increasing the distance from the radiation source (i.e., the patient). Lead glasses may also be necessary for protecting the eyes of RTs. To reduce RT radiation exposure, RTs should remain distant from the patient if possible. However, because this distance may hinder verification of the patient’s condition, RTs sometimes work in close proximity to patients. This is a patient phantom study. In future, the data may need validation by comparison with personal RT dosimeter records. It is important to evaluate the radiation doses delivered to RTs during mobile radiography, as well as the scattered radiation distribution, to ensure adequate protection. Further comparison studies may be needed using the Monte Carlo method.
radiographers and nurses have a responsibility to ensure that no one is within the radiation field during the X-ray exposure of the patient. This is achieved by informing all persons in the immediate area that an X-ray exposure is about to be made and asking them to stand a safe distance from the radiation field area.
Shielding
Placing a barrier of lead or concrete between the radiation source and an individual provides protection from X-radiation (Jones and Taylor, 2006; Ehrlich and Coakes, 2017). During mobile radiography, anyone assisting in an examination and staying in the radiation field should wear a lead-rubber apron or stand behind a mobile lead screen. Generally, walls in special care units where ionising radiation is used are designed to contain the radiation produced by the mobile X-ray tube within a set of criteria and limits determined by relevant legislation (Hart et al, 2002).
Radiation protection during mobile radiography
Nurses' understanding and adherence to radiation protection control measures during mobile radiography is of paramount importance in protecting patients, themselves and members of the public visiting the ward/unit. However, some research studies have found limited awareness and non-adherence to radiation protection control measures among nurses during mobile radiography (Anim-Sampong et al, 2015; Luntsi et al, 2016; Azimi et al, 2018). This can be attributed to a lack of radiation protection awareness programmes for nurses working
The document discusses intensity-modulated radiation therapy (IMRT) for head and neck cancers. It describes how IMRT improves target coverage and sparing of organs-at-risk like the parotid glands compared to conventional radiation therapy. Studies show IMRT reduces the risk of xerostomia and improves quality of life outcomes for patients.
The document discusses various aspects of radiation oncology and radiotherapy clinical trial design. It provides an overview of the evolution of radiation therapy techniques from the 1960s to present. It also covers important considerations for radiation oncology trials, including target volume delineation, dose schedules, quality assurance measures, and assessing toxicity. Multidisciplinary collaboration and factors influencing radiation sensitivity are also briefly discussed.
This study analyzed the cumulative radiation exposure from imaging of 315 children with cancer between 2019-2021 at a single institution. The majority (50%) received cumulative effective doses less than 20 mSv, while 5.4% received doses between 20-50 mSv and 2 patients received over 50 mSv. The risk of radiation-induced cancer remains low for most patients, but more caution is needed as survival rates increase and patients face cumulative risks from treatment burdens like radiation exposure. The authors recommend strengthening radiation protection practices, ensuring appropriateness of procedures, considering alternative imaging modalities when possible, adopting new scanners with lower radiation doses, and monitoring patients' radiation exposure histories.
Essentials of radiation therapy and cancer immunotherapy by Dr. Basil TumainiBasil Tumaini
Radiation therapy uses high-energy radiation to treat cancer and control its symptoms. It works by damaging DNA in cancer cells and preventing them from reproducing. The radiation oncology team includes radiation oncologists, medical physicists, dosimetrists, radiation therapists and nurses. Treatment involves simulation, planning and delivery, with the goal of maximizing dose to the tumor while minimizing it to healthy tissues. External beam radiation uses linear accelerators to deliver photon or particle beams in precise fractions over weeks, while brachytherapy places radioactive sources directly in or near tumors. New techniques like IMRT and SBRT further improve targeting accuracy.
NTCP MODELLING OF ACUTE TOXICITY IN CARCINOMA CERVIX TREATED WITH CONCURRENT ...Dr. Rituparna Biswas
1. The study aimed to develop a predictive nomogram and dose constraints for hematological toxicity in cervical cancer patients treated with chemoradiation including IMRT.
2. Thirty-seven patients were treated with IMRT and cisplatin, and bone marrow was re-delineated to include the entire marrow volume.
3. Dose-volume histograms were combined with toxicity data to create a nomogram from which hematological toxicity probabilities can be estimated based on bone marrow dosimetry.
This document summarizes key aspects of the International Commission on Radiation Units and Measurements (ICRU) Report 83 from 2010 on prescribing, recording, and reporting photon beam intensity-modulated radiation therapy (IMRT). The ICRU Report 50 from 1993 and Report 62 from 1999 established guidelines for defining target volumes like gross tumor volume, clinical target volume, and planning target volume. ICRU Report 83 aimed to update these guidelines for IMRT, which uses non-uniform fluence and dose distributions compared to earlier conformal radiation techniques. Key changes included separating the planning target volume into internal and setup margins, classifying organs at risk, and defining new metrics like the planning organ at risk volume and conformity index for evaluating IM
Radiation Protection Better titles and descriptions lead to more readerssudheendrapv
This document discusses radiation protection and provides recommendations for dose limiting. It begins with an introduction to radiation biology and how radiation can damage living tissue. It then defines common radiation units like exposure, absorbed dose, absorbed dose equivalent, effective dose and others. The document outlines the biological effects of radiation like deterministic and stochastic effects. It provides threshold doses for different deterministic effects. Finally, it presents dose limiting recommendations from ICRP for occupational exposures, public exposures, and embryo-fetus exposure.
This document discusses the history and techniques of stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT). It begins by outlining the early development of SRS by Lars Leksell in the 1950s. It then defines key terms like SRS, SBRT, and fractionated stereotactic radiosurgery. The document goes on to discuss the rationale and advantages of SRS/SBRT, including its ability to deliver high radiation doses with steep dose gradients using multiple beams and image guidance. It also covers topics like tumor oxygenation, cell kill mechanisms, and recent technological advances in the field like VMAT, flattening filter free beams, and 4D
EFFECTS OF X-RAY RADIATION EXPOSURE TOWARD LYMPHOCYTES OF RADIOGRAPHERS IN AB...irjes
X-ray radiation sources in ABCD Hospitalisused as one of the healthfacilities and the role of management, itis not maximized as well in providing protection against the radiographer. The use of personal protective equipmentisstilla rare thingdone right by the radiologist. This studyaimed to analyze the influence of X-ray radiation to the lymphocytes of radiographer in the ABCD Hospital. This studyis a quantitative studyconductedin 4 Hospitals in Mataram, West NusaTenggara (NTB) in June and October 2014. The populations in thisstudywere all radiographerswhoworking in ABCD Hospital as many as 30 people. The sampling technique usedis simple randomsamplingwhere a sample size of 28 people. Data wasanalyzed by usingregressionanalysis. Theseresultsindicatethat the radiologistcharacteristics affect the lymphocytes wereage (p = 0.028), radiation protection training (p = 0.046), use of APD (p = 0.026) and radiation dose (p = 0.046). Radiation protection efforts at A hospital are still not good and B, C and D hospitals are good.
Adaptive radiotherapy (ART) can improve treatment for head and neck cancer patients. ART involves modifying the treatment plan based on anatomical changes observed during radiation therapy delivery. For head and neck cancer, target volumes and organs at risk often change significantly over the course of treatment due to factors like weight loss or tumor shrinkage. Studies have shown ART can improve dose distribution by reducing dose to organs at risk while maintaining or improving tumor dose coverage. Clinical benefits of ART include improved local tumor control and fewer treatment toxicities. ART is most beneficial for patients experiencing greater anatomical changes, such as those with more advanced tumors or significant weight loss.
The document summarizes highlights from breakout sessions at a conference on predictive models for decision making in prostate cancer. It discusses how the radiation oncologist and technique affect outcomes, and factors that predict toxicity after external beam radiotherapy or brachytherapy for sexual, urinary, and GI dysfunction. Future predictive models need to incorporate dosimetric, clinical, genetic, and molecular factors using standardized toxicity reporting and integrating patient-reported outcomes with dose-volume data.
1) Stereotactic radiosurgery (SRS), stereotactic radiotherapy (SRT), and stereotactic body radiotherapy (SBRT) use focused radiation beams and precise targeting to deliver high doses of radiation to small, well-defined tumors with minimal damage to surrounding healthy tissue.
2) Studies have found stereo treatments improve survival rates for certain brain tumors compared to traditional treatments and offer an effective non-invasive option for inoperable lung tumors.
3) The advantages of stereo treatments include targeting small tumors with high radiation doses using fewer treatment sessions, resulting in reduced side effects and improved quality of life for patients.
Similar to Improving Patient Radiation Protection or Evaluating Risks in Medical Imaging? What matters most? (20)
El documento resume la historia de 29 años de la Federación de Radioprotección de América Latina y El Caribe (FRALC), desde su creación en 1993 hasta la actualidad. La FRALC fue fundada para promover la cooperación entre las sociedades de protección radiológica de la región a través de congresos, publicaciones y otras actividades. Actualmente agrupa a 17 sociedades con más de 1200 miembros de países de América Latina y el Caribe. La FRALC ha jugado un papel importante en el desarrollo de la protección radiológica
Este documento trata sobre la protección radiológica en la radiología intervencionista en niños. Explica que los procedimientos intervencionistas en pediatría son tratamientos mínimamente invasivos que ofrecen menores riesgos y estancias hospitalarias más cortas. Sin embargo, los niños son más sensibles a la radiación que los adultos, por lo que es importante optimizar los procedimientos para minimizar la dosis de radiación. El documento proporciona varias estrategias para lograr esto, como el uso de protocolos de dosis reducida, col
Aportes de la Sociedad Peruana de Radioprotección para mejorar la protección ...Eduardo Medina Gironzini
La Sociedad Peruana de Radioprotección (SPR) fue fundada en 1987 con el objetivo de promover estudios e investigaciones para mejorar la protección radiológica en el país. Está integrada por más de 70 profesionales de diferentes campos. A lo largo de sus 31 años de existencia, la SPR ha organizado congresos nacionales e internacionales, simposios, cursos y otras actividades para difundir conocimientos sobre protección radiológica.
Este documento describe la Red Latinoamericana de Protección Radiológica en Medicina (Red LAPRAM), incluyendo su fundador y presidente Eduardo Medina Gironzini, sus objetivos de fortalecer la protección radiológica en aplicaciones médicas en América Latina, y sus actividades desde 2018 como seminarios web, grupos de trabajo y la creación de un canal en YouTube.
Seminario Web
"Herramientas y técnicas para la Gestión del Conocimiento Nuclear"
Claudio Henrique dos Santos Grecco, PostDoc
Organizado por la Red LAPRAM
2 de octubre 2020
Este documento habla sobre los campos electromagnéticos (CEM) y sus propiedades físicas. Explica que todo movimiento de cargas eléctricas genera campos eléctricos y magnéticos. Si el movimiento es cíclico, como en una antena de transmisión, se generan perturbaciones que se propagan y pueden transmitir energía a distancia. También describe la velocidad de propagación de la radiación electromagnética y la relación entre los campos eléctrico y magnético.
These lecture slides, by Dr Sidra Arshad, offer a simplified look into the mechanisms involved in the regulation of respiration:
Learning objectives:
1. Describe the organisation of respiratory center
2. Describe the nervous control of inspiration and respiratory rhythm
3. Describe the functions of the dorsal and respiratory groups of neurons
4. Describe the influences of the Pneumotaxic and Apneustic centers
5. Explain the role of Hering-Breur inflation reflex in regulation of inspiration
6. Explain the role of central chemoreceptors in regulation of respiration
7. Explain the role of peripheral chemoreceptors in regulation of respiration
8. Explain the regulation of respiration during exercise
9. Integrate the respiratory regulatory mechanisms
10. Describe the Cheyne-Stokes breathing
Study Resources:
1. Chapter 42, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 36, Ganong’s Review of Medical Physiology, 26th edition
3. Chapter 13, Human Physiology by Lauralee Sherwood, 9th edition
Promoting Wellbeing - Applied Social Psychology - Psychology SuperNotesPsychoTech Services
A proprietary approach developed by bringing together the best of learning theories from Psychology, design principles from the world of visualization, and pedagogical methods from over a decade of training experience, that enables you to: Learn better, faster!
Histololgy of Female Reproductive System.pptxAyeshaZaid1
Dive into an in-depth exploration of the histological structure of female reproductive system with this comprehensive lecture. Presented by Dr. Ayesha Irfan, Assistant Professor of Anatomy, this presentation covers the Gross anatomy and functional histology of the female reproductive organs. Ideal for students, educators, and anyone interested in medical science, this lecture provides clear explanations, detailed diagrams, and valuable insights into female reproductive system. Enhance your knowledge and understanding of this essential aspect of human biology.
Does Over-Masturbation Contribute to Chronic Prostatitis.pptxwalterHu5
In some case, your chronic prostatitis may be related to over-masturbation. Generally, natural medicine Diuretic and Anti-inflammatory Pill can help mee get a cure.
Osteoporosis - Definition , Evaluation and Management .pdfJim Jacob Roy
Osteoporosis is an increasing cause of morbidity among the elderly.
In this document , a brief outline of osteoporosis is given , including the risk factors of osteoporosis fractures , the indications for testing bone mineral density and the management of osteoporosis
The skin is the largest organ and its health plays a vital role among the other sense organs. The skin concerns like acne breakout, psoriasis, or anything similar along the lines, finding a qualified and experienced dermatologist becomes paramount.
Improving Patient Radiation Protection or Evaluating Risks in Medical Imaging? What matters most?
1. 1
Adjunct Assistant Professor (Radiology)
The George Washington University School of Medicine and Health Sciences
Cari Borrás, D.Sc., FACR, FAAPM, FIOMP
Washington DC, USA
Chair, AAPM International Educational Activities Committee
Improving Patient Radiation
Protection or Evaluating Risks in
Medical Imaging?
What matters most?
2. Lecture Outline
▲ Radiation Protection Dosimetry Terms
▲ Radiology Dosimetry Terms
• Machine related
• Patient related
▲ Diagnostic Reference Levels
▲ Patient Organ Doses
• Measurements
• Look up Tables
• Monte Carlo Simulations
• Dicom Standards: RDSR & P-RDSR
2
3. 3
Radiation Protection Dosimetric
Quantities and Units (ICRP / ICRU)
▲ Incorporate Exposures
• External and Internal
▲ Are Different For:
• Low doses of (low-LET) radiation < 0.5 Gy
o Stochastic (probabilistic) effects, LNT is applicable
• Cancer induction
• Hereditary diseases
• High doses of radiation
o Tissue reactions (deterministic effects), threshold exists
4. 4
RP Dosimetric Quantities and Units
Absorbed Dose, D
Where d ε is the mean energy imparted by
ionizing radiation in a volume element and
d m is the mass of the matter in that volume
The SI unit is J kg-1 and the special name is
gray (Gy)
D = d ε / d m
5. 5
RP Dosimetric Quantities and Units
Tissue Reactions
RBE : Relative Biological Effectiveness
differs for
• different biological endpoints and
• different tissues or organs
Dose to Tissue = Absorbed Dose * RBE
The SI unit is J kg-1 and the special name is gray (Gy)
6. 6
RP Dosimetric Quantities and Units
Stochastic Effects
ICRP 26 (1977) ICRP 60 (1991) ICRP 103 (2007)
* Equivalent Dose Equivalent Dose#
Effective Dose
Equivalent
Effective Dose Effective Dose
* No specific term
# Radiation Weighted Dose proposed but not accepted
The SI unit is J kg-1 and the special name is sievert (Sv)
Evolution of Terminology
7. 7
RP Dosimetric Quantities and Units
Stochastic Effects (Sv)
Equivalent Dose, HT, in a tissue T:
HT = ΣR wR D T,R
wR is the radiation weighting factor, which accounts for
the detriment caused by different types of radiation
relative to photon irradiation
D T,R is the absorbed dose averaged over the tissue
T due to radiation R
wR values are derived from in vivo and in vitro RBE studies
They are independent of dose and dose rate in the low dose region
8. 8
Radiation Weighting Factors in
1991 (ICRP 60) and in 2007 (ICRP 103)
Radiation type and energy range 1991 2007
Photons 1 1
Electrons and muons 1 1
Protons (1991, 2007), pions (2007) 5 2
Alpha particles, fission fragments, heavy ions 20 20
Neutrons, energy < 10 keV 5
Continuous
Function
10 keV to 100 keV 10
> 100 keV to 2 MeV 20
> 2 MeV to 20 MeV 10
> 20 MeV 5
10. 10
RP Dosimetric Quantities and Units
Stochastic Effects (Sv)
E = ΣT wT H T = ΣT ΣR wT wR D R,T
Effective Dose, E
ΣT wT = 1
wT represents the relative contribution of that
tissue or organ to the total detriment resulting
from uniform irradiation of the body
A uniform dose distribution in the whole body gives
an effective dose numerically equal to the radiation-
weighted dose in each organ and tissue of the body
11. 11
Tissue Weighting Factors in 1977 and 1991
Tissue ICRP 26 ICRP 60
Bone surface 0.03 0.01
Bladder 0.05
Breast 0.15 0.05
Colon 0.12
Gonads 0.25 0.20
Liver 0.05
Lungs 0.12 0.12
Esophagus 0.05
Red bone marrow 0.12 0.12
Skin 0.01
Stomach 0.12
Thyroid 0.03 0.05
Remainder 0.30 0.05
TOTAL 1.0 1.0
13. 13
RP Dosimetric Quantities and Units
▲ Incorporate Exposures
• External
• Internal
In this presentation we will not discuss the
magnitudes for exposures from radionuclides:
Committed Equivalent Dose
Committed Effective Dose
14. 14
RP Dosimetric Quantities and Units
Stochastic Effects
Collective Effective Dose, S
(due to Individual Effective Doses E1 and E2)
• d N / d E : number of individuals who experience
an effective dose between E and E + d E
• ΔT specifies the time period within which the
effective doses are summed
15. 15
Limitations of
Equivalent and Effective Doses
▲ Are not directly measurable
▲ Point quantities needed for area monitoring (in
a non-isotropic radiation field, effective dose
depends on the body’s orientation in that field)
▲ Instruments for radiation monitoring need to
be calibrated in terms of a measurable
quantity for which calibration standards exist
Operational protection quantities are needed!
16. 16
RP Operational Quantities - ICRU
Dose Equivalent, H
H = Q * D (Sv)
Where: D = Absorbed Dose
Q = Quality Factor, function of L∞ (LET)
Where: DL is the distribution
of D in L for the charged
particles contributing to D
At a point in tissue:
17. 17
H*(10) and HP (10) – photons > 12 keV and neutrons
HP (0.07) – α and β particles and doses to extremities
Ω in RP usually not specified. Instead,
Maximum H’(0.07, Ω) is obtained
by rotating meter seeking maximum reading
Hp (3) – lens of the eyes
19. 19
System of Quantities for Radiological Protection
Absorbed dose, D
Equivalent dose, HT, in an
organ or tissue T
Effective dose, E
Committed doses,
HT (τ) and E(τ)
Collective effective dose, S
For external exposure
Dose quantities for
area monitoring and
individual monitoring
For internal exposure
Activity quantities in
combination with
biokinetic models and
computations
Operational
Quantities
Dose Quantities
defined in the body
21. Are the radiation protection (RP)
dosimetry terms applicable to
patient exposures in diagnostic
medical imaging and
interventional radiology
procedures?
21
22. 22
RP Dosimetric Quantities and Units
▲ Retrospective dose assessments
▲ Epidemiological studies without careful
consideration of the uncertainties and
limitations of the models and values used
▲ Estimation of specific individual human
exposures and risk
E is calculated averaging
gender, age and individual sensitivity
Caveats
Effective Dose should not be used for
23. 23
RP Dosimetric Quantities and Units
Caveats
Dose to Individuals
Absorbed doses to organs or tissues should be
used with the most appropriate biokinetic
parameters, biological effectiveness of the
ionizing radiation and risk factor data, taking
into consideration the associated uncertainties.
Medical exposures fall in this category!
28. Effective Dose vs Organ Doses
in Medical Exposures
Effective Dose is an adequate parameter to
intercompare doses from different radiological
techniques in order to optimize protection
However, to assess risks it is necessary to
determine organ doses
28
35. 35
Conclusion
For an assessment of the risk due to
induction of stochastic and deterministic
effects by medical x-ray imaging detailed
knowledge is required of organ doses,
absorbed-dose distribution, and the age
and gender of the group of patients
concerned, rather than effective dose.
ICRU 74
38. Dosimetric and Geometric Quantities for
Determination of Patient Dose (ICRU 74, 2005)
38
PKA represents the
integral of air kerma
across the entire x-ray
beam emitted from the
x-ray tube.
Its units are Gy cm2
Tolerance ± 35 % for
> 2.5 Gy cm2
The accuracy of the
display can be checked
directly or indirectly
40. Ka,r (mGy)
“Interventional reference point”, “Cumulative reference point air
kerma”, “Cumulative dose”, “Patient entrance reference point”
40IEC 60601-2-43, 2000 & NCRP 168, 2010
The accuracy of the Ka,r display is
checked with an ion chamber
Tolerance ± 35 % for ≻ 100
mGy
Ka,r (mGy)
Ka,r approximates Ka,e for
adult patients undergoing
cardiac interventions, but
overestimates it for patients in
cerebrovascular interventions.
41. 41
Protection Dosimetry
▲ Kai, Kae, PKA
▲ Maximum (Peak) Organ Dose
• Skin
• Eye Lens
▲ Stochastic Effects
▲ Deterministic Effects
Patient Follow-up may be needed in Interventional Radiology
These terms are the ones mostly used in Diagnostic Reference Levels
45. 45
PKA
Ka,i
HJ Khoury, 2009
Determination of PKA
DIRECT INDIRECT
X Ray tube output
Imaging parameters
Rad area
Removable
KAP meter
on collimator
exit
KAP (DAP)
Display on
control
48. 48
High pixel electronic noise can be reduced by
incorporating a solid state avalanche layer of
a-Se over the TFT array, which amplifies the
signal (HARP)
Summary of Digital Detectors
for Radiography / Fluoroscopy
JR Scheuermann, 2018JA Seibert, 2018
49. CR and DR systems assess the recorded signal through
histogram analysis
Tests with defined beam conditions are used to verify
that correct indicators are being reported
Recommended exposure indicator ranges are used by
technologists to check each radiographic exposure
JA Seibert 2018
Exposure Indicators
50. Region to assess signal indicator
Systems vary in the
region used to assess the
signal for an image.
Full Image
Regular regions
Corresponding
histograms JA Seibert 2018
51. Region to assess signal indicator
IEC 62494-1
Gray histogram for the entire image
Black histogram for the anatomic region
(relevant region)
JA Seibert 2018
52. Computation of an exposure indicator
Median value of the signal values determined in the histogram
of the relevant image region
Manufacturers have proprietary methods
Algorithms, values, and calibration methods are widely
different, leading to confusion amongst users
Inappropriate image segmentation can produce inaccuracies
and incorrect feedback values
JA Seibert 2018
53. Manufacturer Symbol 5 Gy 10 Gy 20 Gy
Canon REX 50 100 200
IDC (ST = 200) F# -1 0 1
Philips (CR-Fuji) EI 200 100 50
Philips (DR) EI 200 400 800
Fuji S 400 200 100
Carestream EI 1700 2000 2300
Siemens EI 500 1000 2000
Approximate EI Values vs. Receptor Exposure
….. The need for a standard is clearly evident
Estimated receptor exposure
JA Seibert 2018
IEC 62494-1
54. Calibration of Radiography EI value
Fuji – CR & DR
Follow manufacturer documentation
Measure incident AK
Compare to indicated value / 100
JA Seibert 2018
55. Deviation Index (DI)
Exposure Indices
𝐷𝐼 = 10 × 𝑙𝑜𝑔10
𝐸𝐼
𝐸𝐼 𝑇(𝑏. 𝑣)
EIT is a target index value that is to be determined for each body
part b, view , procedure type, and clinical site
When EI equals EIT, DI = 0
DI = +3.0 for 2x target exposure
DI = -3.0 for ½x target exposure
± 1 is one step on a standard generator mAs
control or AEC compensation (ISO R5 scale)
JA Seibert 2018
56. Target Exposure Index - EIT
EIT depends on detector type, examination type,
diagnostic question and other parameters
Establishing EIT values requires feedback from
technologists and radiologists working with MP
EIT values are (must be) provided as a data base in the
digital imaging system
Be aware that systems have default values that might
be inappropriate – Must review all values & protocols!
JA Seibert 2018
57. Target exposure index, EIT
Examples:
CR adult PA chest, desired S#: EIT = 700
CR pediatric chest EIT = 500
Noting the efficiency of CR is ~½ that of DR, the EIT for DR devices
are adjusted accordingly
DR adult chest EIT = 350
DR pediatric chest EIT = 250
Extremities: higher EIT
Large patients: higher EIT ??
JA Seibert 2018
59. Caveats
The EI does not describe patient dose
EI is derived from detector signal (dose at the detector)
Best indicator for patient dose is PKA (mGy-cm2)
The EI is not a dose measurement tool
Dose calibration only valid at one radiation quality
Same EI obtained on different digital systems might not
have similar image quality
Influence of detector DQE, scattered radiation, beam quality
JA Seibert 2018
60. Why is incident detector exposure index
(EI) important?
• Is proportional to the image SNR (for given DQE)
Signal to Noise Ratio “image quality”
• Is indirectly related to patient exposure
• Is not linked with image appearance as with screen-
film receptors
• Assists the technologist in identifying appropriate
“equivalent speed” and therefore SNR
JA Seibert 2010 60
68. ▲ Kerma Rate vs Phantom Thickness
• Variables
o Tube Potential (kV)
o Tube Current (mA)
o Pulse Width (ms)
o Cu filter (new angio systems)
▲ Maximum Patient Surface Kerma Rate
68
Automatic Brightness Control
Check: Fluoro and Cine
70. 70
Variation of air kerma rate, tube potential, tube current and
Cu filtration vs water-equivalent phantom - FLUOROSCOPY
N. Lunelli, 2012
Flat Panel
71. 71
IAEA NAHU No. 24
DOSIMETRY IN
DIAGNOSTIC
RADIOLOGY FOR
PAEDIATRIC
PATIENTS
2013
75. 75
CT Dosimetry
Measurements can be done with the ion chamber in air at the
isocenter or in a CT (FDA) phantom using appropriate corrections
76. BSS
MSAD
I
D z dz
N I
N I
1
2
2
( )
U .S. C D R H
MSAD
I
D z dzN I
I
I
1
2
2
, ( ) CTDI
N T
D z dz
T
T
1
7
7
( )
EC
CTDI
T
D z dz
1
( ) DLP CTDI T NW
i
CTDI CTDI CTDIW cm c cm p
1
3
2
310 10, ,
CT Dosimetry Before Helical CT
78. 78
Volume CTDIW (CTDIvol)
CTDIvol = CTDIW • N • T / I
where:
I = the table increment per axial scan, or the table increment
per rotation of the x-ray tube in a helical scan. In helical CT,
the term pitch (P) is defined as the ratio of the table
increment per tube rotation to the nominal (total) width of the
radiation beam. Hence,
Pitch = I / (N • T) and CTDIvol = CTDIW /pitch
CTDIvol is the parameter that best estimates the average dose
at a point with the scan volume for a particular scan protocol.
Dose Length Product (DLP)
DLP = CTDIvol (mGy) • scan length (cm)
CT Dosimetry After Helical CT
80. Current dose reporting methods
▲ Computed Tomography Dose Index, CTDIvol (mGy)
• Provides dose comparison for scan protocols or scanners
• Useful for obtaining “benchmark” data
• Not good for estimating patient dose
▲ Dose Length Product (DLP): CTDIvol × scan length
• Volume dose delivered to the patient (mGy-cm)
• In limited scan range, DLP is less useful, e.g., density-time
studies such as brain perfusion
▲ Effective Dose: a crude measure of whole body dose
• Conversion factors are generated from Monte Carlo transport
methods in standardized phantoms
• Not intended for individual patient dose metrics
• Estimated from DLP
Adapted from JA Seibert 2011
84. Family of physical phantoms
Cynthia McCollough, Mayo Clinic
standard phantoms
Tom Toth & Keith Strauss
Monte Carlo phantoms (1 – 50 cm)
John M. Boone, UC Davis
Anthropomorphic Monte Carlo phantoms
Mike McNitt-Gray, UCLA
AAPM Task Group 204 – Size-Specific CT Dose
JA Seibert 2011
100. 100
Protection Dosimetry
▲ Kai, Kae, PKA
▲ Maximum (Peak) Organ Dose
• Skin
• Eye Lens
▲ Stochastic Effects
▲ Deterministic Effects
Patient Follow-up may be needed in Interventional Radiology
These terms are used for Diagnostic Refence Levels
101. 101
To Optimize Radiation Protection
The best way is to establish
Diagnostic Reference Levels (DRLs)
… derived from
the data from wide
scale quality
surveys … for the
most frequent
examinations in
diagnostic
radiology...
UK 2000
75% Percentile
105. 105
Achievable dose
A dose which serves as a
goal for optimization
efforts. This dose is
achievable by standard
techniques and
technologies in
widespread use, while
maintaining clinical image
quality adequate for the
diagnostic purpose. The
achievable dose is typically
set at the median value of
the dose distribution.
2012
111. DRL Requirements – BSS
A review is conducted to determine whether the
optimization of protection and safety for patients
is adequate, or whether corrective action is
required if, for a given radiological procedure:
i. typical doses or activities exceed the relevant
diagnostic reference level; or
ii. typical doses or activities fall substantially
below the relevant diagnostic reference level
and the exposures do not provide useful
diagnostic information or do not yield the
expected medical benefit to the patient. 111
112. Patient Organ Doses
▲ Medical Imaging
• Radiography
• Mammography
• Fluoroscopy
• CT
▲ Interventional Radiology
112
113. Organ Dose Determination
▲ Direct Radiation Measurements
▲ Table Look Up
▲ Monte Carlo Simulations using Patient and Radiation
Transport Modeling
• Mathematical phantoms
• Special features of the active bone marrow
• Voxel phantoms
• Anthropometric phantoms
▲ Calculations from Imaging System Dose Metrics
• and/or „Dose Report‟
▲ DICOM Standards
113
114. Threshold doses for approximately 1% morbidity incidence
114
ICRP 118 , 2012
115. 115
Trends in average effective doses resulting from selected
diagnostic medical examinations (UNSCEAR 2008)
Examination
Average effective dose per examination (mSv)
Health care level I
1970–1979 1980–1990 1991–1996 1997–2007
Chest
radiography 0.25 0.14 0.14 0.07
Abdomen X-ray 1.9 1.1 0.53 0.82
Mammography 1.8 1 0.51 0.26
CT scan 1.3 4.4 8.8 7.4
Angiography 9.2 6.8 12 9.3
116. Effective Dose vs Organ Doses
in Medical Exposures (ICRU, ICRP)
Effective Dose is an adequate parameter to
intercompare doses from different radiological
techniques in order to optimize protection
However, to assess risks
it is necessary to determine organ doses
116
121. 121
Patient Dose - Mammography
Kerma in Air
Reproducibility and Linearity
Average Glandular Dose
Half Value Layer
Thickness of Compressed Breast
Estimation of Breast Tissue Composition
(Dg)av = (DgN)av * Ka,i
125. 125
Skin Dose Measurements - TLD
SILVA et al., 2009
TLDs placed directly
on a patient’s back
Matrix of TLDs placed on a sheet
under a patient undergoing a
cardiac interventional procedure
134. 134
Skin Dose (DT) Calculations
Ka,e = Ka,r ftable (SRD/SSD)2 BSF
DT = Ka,e ( ) ≃ Ka,e 1.06
Ka,e: Entrance surface Air Kerma
Ka,r: Air Kerma at Reference Point
ftable: Table and pad attenuation
factor
SRD: Source-reference distance
SSD: Source-entrance surface
µen,a / ρ: Air
mean energy
absorption
coefficient
µen,T / ρ: Skin
mean energy
absorption
coefficient
µen,T
ρ
µen,a
ρ
135. 135
Coronary Angioplasty Results
Loc Dose (Gy)
A 0.43
B 2.27
C 3.92
D 0.57
E 3.56
F 3.97
G 2.28
H 1.39
I 0.41
J 0.9
L 1.39
M 1.22
N 0.28
SILVA et al., 2009
136. 136
Maximum Skin Dose - Results
Neuroradiology Interventions
N. Lunelli, 2012
Reaction in the patient –
MSD=8030 mGy
139. CT Dose Reporting…… HOW?
▲ Scanner dose measurement indicators: CTDIvol & DLP
▲ How to get the CT provided data?
• Dose summary page and Optical Character Recognition
• Open-source or commercial “dose gathering” products
Image
DOSE
Summary
page
Adapted from JA Seibert 2011
142. Radimetrics eXposure: Dose calculation engine
• Receives CT study
• Extracts patient dose
metric information
• Pushes dose metrics
to radiology report
• Maintains database
• Provides dashboards
142
JA Seibert 2013
143. Summary
▲ Neither CTDIvol nor DLP should be used to
estimate effective dose or potential cancer risk
for any individual patient
▲ Estimation of organ dose and use of age- and
gender-specific risk coefficients are necessary to
determine individual risk
▲ Investigations using Monte Carlo photon
transport within CT scan data, identification /
segmentation of organs, and tabulating organ
doses are a start to individual, customized dose
measurements
JA Seibert 2011
144. Radiation Dose to Pediatric Patients
of Different Body Stature from CT
Exams Using Deformable Realistic
Phantoms
HPS, 59th Annual Meeting, Baltimore, MD, 2014
Stabin, M. et al. Vanderbilt University Nashville, TN, USA
• Using the Geant4 Monte Carlo toolkit, created a
radiation transport code to simulate patients
undergoing exams on a CT scanner similar to that
at Vanderbilt University Children's Hospital.
• Used measured values of dose in a physical
phantom to calibrate the simulated output from the
Geant CT source.
149. Conclusions
• As with internal dosimetry, using deformable
NURBS phantoms greatly facilitates the
development of phantom-based dosimetry.
• We have used a Geant4-based CT dosimetry
program to calculate doses to a wide range of
patient and phantom models.
• These data can be used to produce patient-
individualized organ and effective doses that
are far better than CT scanner reported doses.
Stabin et al., 2014
150. So, what should be used for
patient risk?
▲ Using Monte Carlo photon transport on organ-segmented CT scan data of
patients
▲ Estimation of specific individual’s organ doses
▲ Accumulating organ dose for each instance
▲ Applying age- and sex-specific risk coefficients
. . . . . .
▲ This is a large undertaking, and will take time for implementation
JA Seibert 2011
151. organ dosesCT scan & patient
parameters
Monte Carlo modeling should be
the basis for patient CT dosimetry
Monte
Carlo
JA Seibert 2011
157. DICOM RDSR from Philips (rooms 3-4 San Carlos Hospital)
157
Radiation Dose Structured Report 67: pages, 30 runs of fluoro + 10 runs of cine; all
technical, dose and geometry details included (part 3 Fluoro example)
Irradiation Event X-Ray Data
Acquisition Plane : Single Plane
DateTime Started : 2013-08-12, 12:51:07.958
Irradiation Event Type : Fluoroscopy
Reference Point Definition : 15cm below BeamIsocenter
Irradiation Event UID :
1.3.46.670589.28.3711502481496.20130812125107317.
1
Dose Area Product = 1.8E-06 Gy.m2
Dose (RP) = 0.00017185537709Gy
Positioner Primary Angle = 1.6 °
Positioner Secondary Angle = -0.1°
X-Ray Filters
X-Ray Filter Type : Strip filter
X-Ray Filter Material : Copper or Copper compound
X-Ray Filter Thickness Minimum = 0.9 mm
X-Ray Filter Thickness Maximum = 0.9 mm
X-Ray Filters
X-Ray Filter Type : Strip filter
X-Ray Filter Material : Aluminum or Aluminum
compound
X-Ray Filter Thickness Minimum = 1 mm
X-Ray Filter Thickness Maximum = 1 mm
Fluoro Mode : Pulsed
Pulse Rate = 7.5 pulse/s
Number of Pulses = 11no units
X-Ray Tube Current = 120 mA
Distance Source to Isocenter = 765 mm
KVP = 97.97 kV
Pulse Width = 9.2 ms
Irradiation Duration = 1.466 s
Patient Table Relationship : headfirst
Patient Orientation : recumbent
Patient Orientation Modifier : supine
Target Region : Chest
Number of Frames = 11no units
SubImages per Frame = 1 no units
Wedges and Shutters
Bottom Shutter = 82.5mm
Left Shutter = 82.5mm
Right Shutter = 82.5mm
Top Shutter = 82.5mm
Beam Position
Longitudinal Beam Position = 1562mm
Beam Angle = 0 °
Table Height Position = 920 mm
E. Vañó, 2018
158. DICOM RDSR from Philips (rooms 3-4 San Carlos Hospital)
158
Radiation Dose Structured Report 67: pages, 30 runs of fluoro + 10 runs of cine; all technical,
dose and geometry details included (part 4 cine example)
Irradiation Event X-Ray Data (series 1 manual
note)
Acquisition Plane : Single Plane
DateTime Started : 2013-08-12, 12:52:31.098
Irradiation Event Type : Stationary Acquisition
Reference Point Definition : 15cm below
BeamIsocenter
Acquired Image : Image X-Ray Angiographic
Image Storage (SOP Instance UID: )
Irradiation Event UID :
1.3.46.670589.28.3711502481496.201308121252
30407.1
Dose Area Product = 0.0004013 Gy.m2
Dose (RP) = 0.06913345231013Gy
Positioner Primary Angle = 11.8°
Positioner Secondary Angle = -0.1°
X-Ray Filters
X-Ray Filter Type : No Filter
X-Ray Tube Current = 865.6 mA
Distance Source to Isocenter = 765 mm
KVP = 82.57 kV
Pulse Width = 7.3 ms
Distance Source to Detector = 1068mm
Irradiation Duration = 5.2 s
Patient Table Relationship : headfirst
Patient Orientation : recumbent
Patient Orientation Modifier : supine
Target Region : Chest
Number of Frames = 78no units
SubImages per Frame = 1 no units
Wedges and Shutters
Bottom Shutter = 67.5mm
Left Shutter = 67mm
Right Shutter = 67mm
Top Shutter = 67.5mm
Beam Position
Longitudinal Beam Position = 1562mm
Beam Angle = 0 °
Table Height Position = 920 mm
E. Vañó, 2018
159. Study level – summary of a procedure II
Sample of the graphical representation of
DICOM RDSR data of a neuro-
interventional study carried out in a bi-
plane cath-lab, that allows to see the
different fluoroscopy and acquisition
modes used during the procedure.
J.M. Fernandez-Soto et al., 2016
160. Sample of a radiochromic film image
placed at the patient back in an
interventional cardiology procedure
(right) and two types of dose maps
obtained from the DICOM RDSR for the
same procedure. This also allows the
estimation of the maximum dose at the
skin entrance.
Skin Dose Maps
(JM Fernandez-Soto et al., 2016)
163. Supplement 191 Patient Dose SR
• Current Radiation Dose SR contains only information about the x-ray system or information
the x-ray system can determine, e.g.:
• radiation output, geometry, x-ray source, detector system, etc.
• Estimation of patient/organ dose requires knowledge of:
• Radiation beam characteristics that interact with patient
• Models of the patient/organs
• Models of radiation interaction within the patient
• Methods to do patient dose estimations are being developed and improved continuously
• storage of these estimations in a different object would allow more versatile utilization of
the data
163D. Pelc, 2016
164. 164
RDSR
IOD
- X-Ray exposure techniques
- Table, Gantry Angle, Beam
Geometry, collimation, …
- Dose measures: CTDI, DAP, ...
X-Ray
Equipment
Modality
IOD
Patient Dose Determination: Data Flow Requirements
Organ Dose
Reporter
System
Modality Frame of Reference
(FOR)
Signifies part of Supplement 191 Patient RDSR
D. Pelc, 2016
165. 165Signifies part of Supplement 191 Patient RDSR
RDSR
IOD
- X-Ray exposure techniques
- Table, Gantry Angle, Beam
Geometry, collimation, …
- Dose measures: CTDI, DAP, ...
X-Ray
Equipment
Modality
IOD
Organ Dose
Reporter
System
Patient Model
Registration
IOD
Equipment Information
- Table dimension
- Attenuating material, …
Modality Frame of Reference
(FOR)
Patient Model
FOR
Patient Dose Determination: Data Flow Requirements
D. Pelc, 2016
166. 166Signifies part of Supplement 191 Patient RDSR
RDSR
IOD
- X-Ray exposure techniques
- Table, Gantry Angle, Beam
Geometry, collimation, …
- Dose measure: CTDI, DAP, ...
Registration
IOD
X-Ray
Equipment
Modality
IOD
Patient Model
Organ Dose
Reporter
System
Patient
Dose SR
Calculated data for
documentation and reference
to images and RDSR’s
Patient Dose Determination: Data Flow Requirements
Equipment Information
- Table dimension
- Attenuating material, …
Patient Dose 2D view/map
(e.g. iso-dose map)
Patient Dose Surface
3D map/view
Modality Frame of Reference
(FOR)
Patient Model
FOR
D. Pelc, 2016
167. Final Considerations: What Matters Most?
▲ Patient organ doses will be useful for refining dose response
curves and carry out epidemiological studies
▲ Will determining patient doses allow us to estimate patient risk?
▲ Risk is population-based, it cannot be applied to a single
individual! Risk is a probability!!
▲ Do we need individualized dosimetry for radiation protection?
▲ If the goal is to protect the patient, can we just use diagnostic
reference levels, which are usually machine parameters,
something much simpler to determine than organ doses?
▲ Can we report such parameters in the patient chart? Some
countries / states require it….
168. Agency
Regulatory Requirement
(ID: Individual Dose)
FDA &
NEMA
Kai limit for fluoroscopy. Dg limit for
mammography.
All new CT scanners to comply with the NEMA
XR-25 Dose Check Standard
State of
California
ID: CTDIvol and DLP included in each patient
record and transferred to PACS. Effective dose
limits for repeated exams unless approved by a
physician. All CT facilities accredited.
State of Texas
Kai limits for some radiography exams. Maintain
records for output of fluoroscopically-guided
interventions to include cumulative Ka,r or PKA.
ID for CT: Dose recording and reporting plus
skin dose determination. Recommends
establishing a “CT reference level” and patient
follow up when value is exceeded.
.
169. Agency Accreditation Requirements
The Joint
Commission
CT: CTDIvol and DLP or SSDE specific for
each exam, summarized by series or anatomic
area and retrievable. Review and analyze
values if they exceed expected dose index
ranges. Compare with external benchmarks.
Fluoro: Establish radiation exposure and skin
dose threshold levels. If exceeded, review
procedure and evaluate patient (>15 Gy =
Sentinel event). Ka,r or PKA documented in a
retrievable format.
Both: Evaluation of equipment performance
and imaging protocols.
American
College of
Radiology
Dg limit, CTDIvol limit for adult and pediatric
head and abdomen CT scans. DRLs
recommended.
170. Example from a California Patient CT Chart
This patient received a total of 1 exposure event(s) during this CT
examination. The CTDIvol and DLP radiation dose values for each exposure
are: Exposure: 1; Series: 2; Anatomy: Neck; Phantom: 32 cm; CTDIvol: 19;
DLP: 498 The dose indicators for CT are the volume Computed
Tomography (CT) Dose Index (CTDIvol) and the Dose Length Product
(DLP), and are measured in units of mGy and mGy-cm, respectively. These
indicators are not patient dose, but values generated from the CT scanner
acquisition factors. The report includes radiation exposure data for
exposures received during this examination. If multiple reports are
produced from this examination, the exposure data is duplicated in each
report. The exposure data reported is indicative, but not determinative, of
the radiation dose received by this patient.