Continuous stirred tank reactor (CSTR):
Conductivity measurement
Name of student: Shwan Sarwan Sadiq
Group: B
Date of Exp. OCT 28th
2015
Submission date: NOV 11th
2015
Supervisor: ms.lameha
Chemical Engineering Department
Chemical reactors lab
3rd stage
(
(
Table of content:
Aim of the experiment 3
Theory 4 &5
Apparatus 6&7
Methodology
8,9,10,11,12,13,14,15,16,17,18
Data sheet 19
Graphs 20
Discussion 21,22&23
References 24
Aim of the experiment:
Measurement of the conductivity for a base – water
Mixture along time at high temperature using different
Concentrations.
Theory:
CSTR:
_it’s called ‘mixed reactor ‘, ‘back mix reactor’, ‘Continuous stirred
Tank reactor CSTR ‘. It’s a reactor in which contents are
well mixed And uniform throughout. Thus, the exit stream
from the reactor has the same composition as the fluid
within the reactor. The pattern of flow is mixed flow and is
used primarily for liquid phase reactions.
More specifically, continuous stirred tanks are used for
relatively slow reactions of liquids and slurries. .
_Its normal operation is at steady state, where the conditions
in the
Reactor don't change with time and it’s assumed to be
perfectly mixed,
So the contents have relatively uniform properties such as
temperature, Density, etc.
Conductivity:
Conductivity is a measure of how well a solution conducts
electricity. To carry a current a solution must contain charged
particles, or ions. Most Conductivity measurements are made
Continue..
in aqueous solutions, and the ions responsible for the
conductivity come from electrolytes dissolved in the Water.
Salts (like sodium chloride and magnesium sulfate), acids (like
Hydrochloric acid and acetic acid), and bases (like sodium
hydroxide and Ammonia) are all electrolytes. Although water
itself is not an electrolyte, it does have a very small
conductivity, implying that at least some ions are Present.
The ions are hydrogen and hydroxide, and they originate
from the Dissociation of molecular water.
_In conductivity works, it is necessary to work with as pure
As possible solvents because any impurity affects to
Conductivity therefore, the distilled water has to be used
During the experiments rather than tap water.
Apparatus:
To study CSTR reactor, it is necessary to use a service
Module and Interface called QUSC. This unit provides the
reagents and the thermostatized water to the reactor under
study.
Base module and interface, QUSC:
QUSC. Service Unit:
_This unit is common for the Chemical Reactors, and can
work with one or several reactors.
_ Accommodation and exchange system of the reactors,
quick and easy to handle.
_It supplies all the services for the operation of each reactor
_ Anodized aluminum structure and panels of painted steel.
_ Main metallic elements in stainless steel.
_ Diagram in the front panel with similar distribution to the
elements in the real unit.
Continue..
_2 Peristaltic dosing pumps with variable speed. Flow rate
up to 3 l. /h.
(Unit standard disposition). With another disposition, they
could reach a flow rate up to 10 l./h.
_ Thermostatic bath of 6 l. capacity. Temperature control of
hot water in order to maintain the reactor temperature.
_Pump of 3 l. /min., to impel the thermostatization water
from the bath to the reactor.
_2 Tanks for the reagents, of 1 l. capacity each one, made in
Pyrex glass.
_The control of the reaction is carried out by a conductivity
sensor,
Which allows the reaction evolution parameterization in real
time.
_ Three “J” type temperature sensors, one to know the
thermostatic
bath temperature in a continuous way and two sensors
to know the
Continue..
Water temperature at the thermostatic bath water inlet and
outlet.
_ Quick connectors with shutoff valve that enable an easy
coupling of
The Service Unit to the chosen reactor.
_All elements of this unit are chemically resistant.
Continue..
Electronic Console:
Metallic box.
_ Temperature sensors connectors.
_ Digital display for temperature sensors.
_ Selector for temperature sensors.
_ Peristaltic pumps controllers and switches.
_ Water pump switch.
_ Stirrer switches.
_heating element controller.
The reagents supply circuit is constituted by a PTFE pipe of 6
mm. The reagents are introduced into two Pyrex vessels
of 1 liter each placed at the rear part. There are two
Continue..
Peristaltic pumps to move the reagents. The following valves
are located at the Module’s base:
_The 6 mm valves (1) are mounted at the peristaltic
pumps outlet and they must be connected to the reagent
inlet of the Reactor.
_The ball valves with the mark (2) are of water outlet and
inlet.
Continue..
They have a tube of 8 mm which must be connected to
the water inlet and outlet of the reactors. The
connections are Interchangeable between them.
_The temperature control system consists of a thermostatic
Bath, whose temperature is controlled by means of a PID
Control on this bath temperature. The thermostated water
System is also composed of a thermostatization water
impeller pump, with variable flow by means of the valves (2).
_The data acquisition and process control system is
Centralized in the electronic interface connected to the
different elements which constitute both the base
module and the reactor module.
Continuous Stirred Tank Reactor Module:
The body of the reactor, made in Pyrex glass, has a capacity
of 2 liters and it is specially designed to work in
continuous, although it also allows to work in discontinuous
if the reagents are introduced to the desired volume and
Continue..
the pump are stopped. There is a manual discharge valve at
the lower part for this purpose.
_During the operation in continuous, the final product outlet
is carried out by the stainless steel tube (4) which is
regulable in height, loosening the brass nut.
This allows to work with different reaction volumes.
_The stirring system (1) with speed control and indication
allows The study of the influence of both continuous and
discontinuous Agitation in the reaction kinetics.
Continue..
_ System to heat the reaction formed by a stainless steel coil
(3) through which the water from the Service Module bath
circulates. The water inlets to the coil from the service
module will be done through the fittings of 8 mm placed
at the lower part of the reactor.
_The monitoring of the reaction is carried out using a
conductivity cell with conductivity meter, which allows
measuring the evolution of the reaction in real time.
_The thermostatic bath has been designed to keep constant
the temperature around the reactor. A volume of 9 liters has
been taken, because this volume allows the control of the
temperature of water every 0.1ºC.
Continue..
The conductivity meter or conductivity sensor:
Applies a temperature compensation factor by default to
Compensate for the change of the measurement due to
Temperature. Such factor has a factory-set value of 2%,
which
implies a correction of 2% per degree increased with respect
to the reference temperature. Depending on the practical
exercise to be performed, we may want this value be 0%, so
that it does not compensate for the temperature influence
(when we want to study ionic conductivities) or have a
specific value between 0 and 5% (when we want to
Continue..
associate conductivity and conversion of the reaction directly).
The default factor is 2%, but it can be optimized by
studying the conductivity relation of each species with
the temperature and then adjusting the compensation
factor (between 0 and 5%) that better fits each reaction
1) To compensate for the compensation factor of the
Conductivity meter, press “TEMP”. The value adjusted
At that moment will appear in the screen. Press “TEMP”
Again to exit the screen.
2) To adjust the % of the compensation factor press“TEMP”
and then the key “FACTOR ADJ” successively until obtaining
the desired value (Between 0 and 5%). Press “TEMP” again
to exit.
Continue.. (CSTR)
Methodology:
1-dissolve 4 grams of NaOH in some distilled water .
2-fill a bottle of 800 ml of the NaOH mixture and
additional distilled water. And another bottle of water.
3-fix the bottles in their places and tide the pipes and
valves.
4-power on the Electronic Console and the heater to
increase the heat of the water in the bath and read the
first temperature reading and its conductivity.
5-open the pump and read the readings of temperature
and the conductivity for every 30 seconds.
6-after finishing readings for several times turn of the
pump and the Electronic Console then pour all the liquid
in the reactor.
Graphs:
Discussion:
Conductivity is a measure of how well a solution conducts electricity. To
carry a current a solution must contain charged particles, or ions. Most
conductivity measurements are made in aqueous solutions, and the ions
responsible for the conductivity come from electrolytes dissolved in the
water. Salts (like sodium chloride and magnesium sulfate), acids (like
hydrochloric acid and acetic acid), and bases (like sodium hydroxide and
ammonia) are all electrolytes. Although water itself is not an electrolyte, it
does have a very small conductivity, implying that at least some ions are
present. The ions are hydrogen and hydroxide, and they originate from the
dissociation of molecular water.
Salts, acids, and bases are electrolytes. They dissolve in water to form ions.
Although water is not an electrolyte, a very small concentration of
hydrogen and hydroxide ions are always present in pure water.
Increasing the temperature of an electrolyte solution always increases the
conductivity. The increase is significant, between 1.5 and 5.0% per °C. To
compensate for temperature changes, conductivity readings are commonly
corrected to the value at a reference temperature, typically 25°C. All
process conductivity sensors have integral temperature sensors that allow
Continue..
the analyzer to measure the process temperature and correct the raw
conductivity. Three temperature correction algorithms are in common use.
When we have to repeated conductivity measurement it means that there
is a mistake in increasing temperature so that and it proves that the ions
that conduct in the same temperature are equal in their conductivity
motion.
Position of conductivity cell
Make sure that all the poles of the conductivity cell are completely covered
by the sample. Always position a 2-pole cell in the center of the measuring
vessel.
Low conductivity measurements:
• use a flow-through cell, to avoid atmospheric contamination from carbon
dioxide,
• use cells with low cell constant, 1 cm-1 or lower,
• use non-platinized cells for easier cleaning and faster response,
• make sure that your instrument is able to apply an appropriate
measuring frequency
High conductivity measurements
• Use platinised cells to avoid polarisation errors, preferably 4-pole cells.
• Use cells with a high cell constant (1 cm-1) or higher if possible.
Continue..
• Do not dilute samples in order to bring them into measuring range.
Conductivity is not proportional to concentration at high levels.
• Make sure that your instrument is able to apply an appropriate measuring
frequency.
Since the accuracy of any measurement depends on proper calibration, a
fresh standard must always be used. Ideally, sample beakers and sensor
should be rinsed two to three times with the sample as the presence of
contaminants can lead to additional errors in conductivity results
When setting up a conductivity measurement system, the lab manager
must make appropriate choices relating to TC(temperature compensation),
such as o whether TC should be used, o setting the reference temperature,
o choosing a linear or nonlinear TC, and o choosing the appropriate TC
factor.
References:
1.http://www2.emersonprocess.com/siteadmincenter/
PM%20Rosemount%20Analytical%20Documents/Liq
_ADS_43-018.pdf
page 1
2.http://us.mt.com/dam/MT-
NA/pHCareCenter/Conductivity_Reduce_Common_E
rrors_WP.pdf
page 3
3.http://igz.ch/de/produkte/uebersicht.asp?action=do
wnload&fileid=8205
page 21
4.http://www.thermofisher.co.nz/Uploads/file/Environ
mental-Industrial/Environmental-Monitoring-
Safety/Water-Monitoring-Treatment/Tech-Tip-Top-
Ten-Conductivity-Mistakes-NZ.pdf
page 1

More Related Content

What's hot

Plate Heat Exchanger Lab Report Group B4
Plate Heat Exchanger Lab Report Group B4Plate Heat Exchanger Lab Report Group B4
Plate Heat Exchanger Lab Report Group B4
Janet Mok
 
Full design for acetone production
Full design for acetone productionFull design for acetone production
Full design for acetone production
dinaelgayar83
 
Packed columns
Packed columnsPacked columns
Packed columns
Doaa Shaheen
 
Design and Operation of a Distillation Column for the Binary Mixture
Design and Operation of a Distillation Column for the Binary MixtureDesign and Operation of a Distillation Column for the Binary Mixture
Design and Operation of a Distillation Column for the Binary Mixture
Jonathan Sherwin
 

What's hot (20)

Reactive distillation
Reactive distillationReactive distillation
Reactive distillation
 
Plate Heat Exchanger Lab Report Group B4
Plate Heat Exchanger Lab Report Group B4Plate Heat Exchanger Lab Report Group B4
Plate Heat Exchanger Lab Report Group B4
 
Project Report
Project ReportProject Report
Project Report
 
Multiple reactors
Multiple reactorsMultiple reactors
Multiple reactors
 
Gas absorbtion
Gas absorbtionGas absorbtion
Gas absorbtion
 
Cooling tower full report
Cooling tower full reportCooling tower full report
Cooling tower full report
 
Full design for acetone production
Full design for acetone productionFull design for acetone production
Full design for acetone production
 
Sedimentation
SedimentationSedimentation
Sedimentation
 
Packed columns
Packed columnsPacked columns
Packed columns
 
conversion and reactor sizing
conversion and reactor sizingconversion and reactor sizing
conversion and reactor sizing
 
Chemical reaction engineering
Chemical reaction engineeringChemical reaction engineering
Chemical reaction engineering
 
Trays in distillation column
Trays in distillation columnTrays in distillation column
Trays in distillation column
 
Plate heat exchangers
Plate heat exchangersPlate heat exchangers
Plate heat exchangers
 
Azeotropic Distillation
Azeotropic DistillationAzeotropic Distillation
Azeotropic Distillation
 
Evaporators
EvaporatorsEvaporators
Evaporators
 
Thermosyphon Reboiler & its type with operational parameter.
Thermosyphon  Reboiler & its type with operational parameter.Thermosyphon  Reboiler & its type with operational parameter.
Thermosyphon Reboiler & its type with operational parameter.
 
Heat exchanger design
Heat exchanger designHeat exchanger design
Heat exchanger design
 
Design and Operation of a Distillation Column for the Binary Mixture
Design and Operation of a Distillation Column for the Binary MixtureDesign and Operation of a Distillation Column for the Binary Mixture
Design and Operation of a Distillation Column for the Binary Mixture
 
Differential method of analysis of data
Differential method of analysis of dataDifferential method of analysis of data
Differential method of analysis of data
 
Evaporator
EvaporatorEvaporator
Evaporator
 

Similar to Conductivity measurement in CSTR water-base(according to temperature)

reactor lab single CSTR reactor.pdf
reactor lab single CSTR reactor.pdfreactor lab single CSTR reactor.pdf
reactor lab single CSTR reactor.pdf
DimaJawhar
 
reactor design lab continuous stirred tank reactor
reactor design lab continuous stirred tank reactorreactor design lab continuous stirred tank reactor
reactor design lab continuous stirred tank reactor
DimaJawhar
 
Meeting w13 chapter 4 part 3
Meeting w13   chapter 4 part 3Meeting w13   chapter 4 part 3
Meeting w13 chapter 4 part 3
Hattori Sidek
 
reactor design batch reactor
reactor design batch reactorreactor design batch reactor
reactor design batch reactor
DimaJawhar
 
Electrical conductivity
Electrical conductivityElectrical conductivity
Electrical conductivity
ECRD IN
 

Similar to Conductivity measurement in CSTR water-base(according to temperature) (20)

Conductivity measurement water-base(temperature is not included)
Conductivity measurement water-base(temperature is not included)Conductivity measurement water-base(temperature is not included)
Conductivity measurement water-base(temperature is not included)
 
reactor lab single CSTR reactor.pdf
reactor lab single CSTR reactor.pdfreactor lab single CSTR reactor.pdf
reactor lab single CSTR reactor.pdf
 
reactor design lab continuous stirred tank reactor
reactor design lab continuous stirred tank reactorreactor design lab continuous stirred tank reactor
reactor design lab continuous stirred tank reactor
 
Meeting w13 chapter 4 part 3
Meeting w13   chapter 4 part 3Meeting w13   chapter 4 part 3
Meeting w13 chapter 4 part 3
 
Isothermal semi batch_reactor
Isothermal semi batch_reactorIsothermal semi batch_reactor
Isothermal semi batch_reactor
 
reactor design batch reactor
reactor design batch reactorreactor design batch reactor
reactor design batch reactor
 
RD Lab - Exp-03 - G-A1.pdf
RD Lab - Exp-03 - G-A1.pdfRD Lab - Exp-03 - G-A1.pdf
RD Lab - Exp-03 - G-A1.pdf
 
Lab 2.docx
Lab 2.docxLab 2.docx
Lab 2.docx
 
Iisrt z abhijit devaraj
Iisrt z abhijit devarajIisrt z abhijit devaraj
Iisrt z abhijit devaraj
 
IRJET- Experimental Analysis on Performance of Closed Loop Pulsating Heat Pip...
IRJET- Experimental Analysis on Performance of Closed Loop Pulsating Heat Pip...IRJET- Experimental Analysis on Performance of Closed Loop Pulsating Heat Pip...
IRJET- Experimental Analysis on Performance of Closed Loop Pulsating Heat Pip...
 
RD Lab - Exp-12 - G-A1.pdf
RD Lab - Exp-12 - G-A1.pdfRD Lab - Exp-12 - G-A1.pdf
RD Lab - Exp-12 - G-A1.pdf
 
Exp 1 (B)
Exp 1 (B)Exp 1 (B)
Exp 1 (B)
 
IRJET- Experimental Research of Heat Transfer using Nano – Fluid in Radiator
IRJET- Experimental Research of Heat Transfer using Nano – Fluid in RadiatorIRJET- Experimental Research of Heat Transfer using Nano – Fluid in Radiator
IRJET- Experimental Research of Heat Transfer using Nano – Fluid in Radiator
 
Reactor Design 1
Reactor Design 1Reactor Design 1
Reactor Design 1
 
Automated process control systems
Automated process control systemsAutomated process control systems
Automated process control systems
 
Electrical conductivity
Electrical conductivityElectrical conductivity
Electrical conductivity
 
Electrical conductivity
Electrical conductivityElectrical conductivity
Electrical conductivity
 
RD Lab - Exp-08 - G-A1.pdf
RD Lab - Exp-08 - G-A1.pdfRD Lab - Exp-08 - G-A1.pdf
RD Lab - Exp-08 - G-A1.pdf
 
RD Lab - Exp-04 - G-A1.pdf
RD Lab - Exp-04 - G-A1.pdfRD Lab - Exp-04 - G-A1.pdf
RD Lab - Exp-04 - G-A1.pdf
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhi
 

More from sarkawtn

Water pollution
Water pollutionWater pollution
Water pollution
sarkawtn
 

More from sarkawtn (15)

Linear heat conduction(Brass 25)
Linear heat conduction(Brass 25)Linear heat conduction(Brass 25)
Linear heat conduction(Brass 25)
 
Temperature measurement
Temperature measurementTemperature measurement
Temperature measurement
 
Global warming
Global warming Global warming
Global warming
 
Water pollution
Water pollutionWater pollution
Water pollution
 
Water pollution
Water pollutionWater pollution
Water pollution
 
Tablet and Smartphone
Tablet and SmartphoneTablet and Smartphone
Tablet and Smartphone
 
Air purification
Air purificationAir purification
Air purification
 
Nano-materials using in chemical engineering (Nano Catalyst)
Nano-materials using in chemical engineering (Nano Catalyst)Nano-materials using in chemical engineering (Nano Catalyst)
Nano-materials using in chemical engineering (Nano Catalyst)
 
sodium (natrium)Na
sodium (natrium)Nasodium (natrium)Na
sodium (natrium)Na
 
sodium Na (natrium)
sodium Na (natrium)sodium Na (natrium)
sodium Na (natrium)
 
Dead weight piston gauge
Dead weight piston gaugeDead weight piston gauge
Dead weight piston gauge
 
Boyles law (expansion)
Boyles law (expansion)Boyles law (expansion)
Boyles law (expansion)
 
Marcet boiler
Marcet boiler Marcet boiler
Marcet boiler
 
Boyles law (compresion) (relation between pressure & volume)
Boyles law (compresion) (relation between pressure & volume)Boyles law (compresion) (relation between pressure & volume)
Boyles law (compresion) (relation between pressure & volume)
 
enviromental pollution & air pollution
enviromental pollution & air pollutionenviromental pollution & air pollution
enviromental pollution & air pollution
 

Recently uploaded

Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power Play
Epec Engineered Technologies
 

Recently uploaded (20)

COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
 
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced LoadsFEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and properties
 
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
NO1 Top No1 Amil Baba In Azad Kashmir, Kashmir Black Magic Specialist Expert ...
 
UNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptxUNIT 4 PTRP final Convergence in probability.pptx
UNIT 4 PTRP final Convergence in probability.pptx
 
AIRCANVAS[1].pdf mini project for btech students
AIRCANVAS[1].pdf mini project for btech studentsAIRCANVAS[1].pdf mini project for btech students
AIRCANVAS[1].pdf mini project for btech students
 
fitting shop and tools used in fitting shop .ppt
fitting shop and tools used in fitting shop .pptfitting shop and tools used in fitting shop .ppt
fitting shop and tools used in fitting shop .ppt
 
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
💚Trustworthy Call Girls Pune Call Girls Service Just Call 🍑👄6378878445 🍑👄 Top...
 
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
 
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxS1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
 
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptxOrlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
Orlando’s Arnold Palmer Hospital Layout Strategy-1.pptx
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdf
 
DC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationDC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equation
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power Play
 
Linux Systems Programming: Inter Process Communication (IPC) using Pipes
Linux Systems Programming: Inter Process Communication (IPC) using PipesLinux Systems Programming: Inter Process Communication (IPC) using Pipes
Linux Systems Programming: Inter Process Communication (IPC) using Pipes
 
Electromagnetic relays used for power system .pptx
Electromagnetic relays used for power system .pptxElectromagnetic relays used for power system .pptx
Electromagnetic relays used for power system .pptx
 
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
 
Introduction to Data Visualization,Matplotlib.pdf
Introduction to Data Visualization,Matplotlib.pdfIntroduction to Data Visualization,Matplotlib.pdf
Introduction to Data Visualization,Matplotlib.pdf
 

Conductivity measurement in CSTR water-base(according to temperature)

  • 1. Continuous stirred tank reactor (CSTR): Conductivity measurement Name of student: Shwan Sarwan Sadiq Group: B Date of Exp. OCT 28th 2015 Submission date: NOV 11th 2015 Supervisor: ms.lameha Chemical Engineering Department Chemical reactors lab 3rd stage ( (
  • 2. Table of content: Aim of the experiment 3 Theory 4 &5 Apparatus 6&7 Methodology 8,9,10,11,12,13,14,15,16,17,18 Data sheet 19 Graphs 20 Discussion 21,22&23 References 24
  • 3. Aim of the experiment: Measurement of the conductivity for a base – water Mixture along time at high temperature using different Concentrations.
  • 4. Theory: CSTR: _it’s called ‘mixed reactor ‘, ‘back mix reactor’, ‘Continuous stirred Tank reactor CSTR ‘. It’s a reactor in which contents are well mixed And uniform throughout. Thus, the exit stream from the reactor has the same composition as the fluid within the reactor. The pattern of flow is mixed flow and is used primarily for liquid phase reactions. More specifically, continuous stirred tanks are used for relatively slow reactions of liquids and slurries. . _Its normal operation is at steady state, where the conditions in the Reactor don't change with time and it’s assumed to be perfectly mixed, So the contents have relatively uniform properties such as temperature, Density, etc. Conductivity: Conductivity is a measure of how well a solution conducts electricity. To carry a current a solution must contain charged particles, or ions. Most Conductivity measurements are made
  • 5. Continue.. in aqueous solutions, and the ions responsible for the conductivity come from electrolytes dissolved in the Water. Salts (like sodium chloride and magnesium sulfate), acids (like Hydrochloric acid and acetic acid), and bases (like sodium hydroxide and Ammonia) are all electrolytes. Although water itself is not an electrolyte, it does have a very small conductivity, implying that at least some ions are Present. The ions are hydrogen and hydroxide, and they originate from the Dissociation of molecular water. _In conductivity works, it is necessary to work with as pure As possible solvents because any impurity affects to Conductivity therefore, the distilled water has to be used During the experiments rather than tap water.
  • 6. Apparatus: To study CSTR reactor, it is necessary to use a service Module and Interface called QUSC. This unit provides the reagents and the thermostatized water to the reactor under study. Base module and interface, QUSC: QUSC. Service Unit: _This unit is common for the Chemical Reactors, and can work with one or several reactors. _ Accommodation and exchange system of the reactors, quick and easy to handle. _It supplies all the services for the operation of each reactor _ Anodized aluminum structure and panels of painted steel. _ Main metallic elements in stainless steel. _ Diagram in the front panel with similar distribution to the elements in the real unit.
  • 7. Continue.. _2 Peristaltic dosing pumps with variable speed. Flow rate up to 3 l. /h. (Unit standard disposition). With another disposition, they could reach a flow rate up to 10 l./h. _ Thermostatic bath of 6 l. capacity. Temperature control of hot water in order to maintain the reactor temperature. _Pump of 3 l. /min., to impel the thermostatization water from the bath to the reactor. _2 Tanks for the reagents, of 1 l. capacity each one, made in Pyrex glass. _The control of the reaction is carried out by a conductivity sensor, Which allows the reaction evolution parameterization in real time. _ Three “J” type temperature sensors, one to know the thermostatic bath temperature in a continuous way and two sensors to know the
  • 8. Continue.. Water temperature at the thermostatic bath water inlet and outlet. _ Quick connectors with shutoff valve that enable an easy coupling of The Service Unit to the chosen reactor. _All elements of this unit are chemically resistant.
  • 9. Continue.. Electronic Console: Metallic box. _ Temperature sensors connectors. _ Digital display for temperature sensors. _ Selector for temperature sensors. _ Peristaltic pumps controllers and switches. _ Water pump switch. _ Stirrer switches. _heating element controller. The reagents supply circuit is constituted by a PTFE pipe of 6 mm. The reagents are introduced into two Pyrex vessels of 1 liter each placed at the rear part. There are two
  • 10. Continue.. Peristaltic pumps to move the reagents. The following valves are located at the Module’s base: _The 6 mm valves (1) are mounted at the peristaltic pumps outlet and they must be connected to the reagent inlet of the Reactor. _The ball valves with the mark (2) are of water outlet and inlet.
  • 11. Continue.. They have a tube of 8 mm which must be connected to the water inlet and outlet of the reactors. The connections are Interchangeable between them. _The temperature control system consists of a thermostatic Bath, whose temperature is controlled by means of a PID Control on this bath temperature. The thermostated water System is also composed of a thermostatization water impeller pump, with variable flow by means of the valves (2). _The data acquisition and process control system is Centralized in the electronic interface connected to the different elements which constitute both the base module and the reactor module. Continuous Stirred Tank Reactor Module: The body of the reactor, made in Pyrex glass, has a capacity of 2 liters and it is specially designed to work in continuous, although it also allows to work in discontinuous if the reagents are introduced to the desired volume and
  • 12. Continue.. the pump are stopped. There is a manual discharge valve at the lower part for this purpose. _During the operation in continuous, the final product outlet is carried out by the stainless steel tube (4) which is regulable in height, loosening the brass nut. This allows to work with different reaction volumes. _The stirring system (1) with speed control and indication allows The study of the influence of both continuous and discontinuous Agitation in the reaction kinetics.
  • 13. Continue.. _ System to heat the reaction formed by a stainless steel coil (3) through which the water from the Service Module bath circulates. The water inlets to the coil from the service module will be done through the fittings of 8 mm placed at the lower part of the reactor. _The monitoring of the reaction is carried out using a conductivity cell with conductivity meter, which allows measuring the evolution of the reaction in real time. _The thermostatic bath has been designed to keep constant the temperature around the reactor. A volume of 9 liters has been taken, because this volume allows the control of the temperature of water every 0.1ºC.
  • 14. Continue.. The conductivity meter or conductivity sensor: Applies a temperature compensation factor by default to Compensate for the change of the measurement due to Temperature. Such factor has a factory-set value of 2%, which implies a correction of 2% per degree increased with respect to the reference temperature. Depending on the practical exercise to be performed, we may want this value be 0%, so that it does not compensate for the temperature influence (when we want to study ionic conductivities) or have a specific value between 0 and 5% (when we want to
  • 15. Continue.. associate conductivity and conversion of the reaction directly). The default factor is 2%, but it can be optimized by studying the conductivity relation of each species with the temperature and then adjusting the compensation factor (between 0 and 5%) that better fits each reaction 1) To compensate for the compensation factor of the Conductivity meter, press “TEMP”. The value adjusted At that moment will appear in the screen. Press “TEMP” Again to exit the screen. 2) To adjust the % of the compensation factor press“TEMP” and then the key “FACTOR ADJ” successively until obtaining the desired value (Between 0 and 5%). Press “TEMP” again to exit.
  • 17. Methodology: 1-dissolve 4 grams of NaOH in some distilled water . 2-fill a bottle of 800 ml of the NaOH mixture and additional distilled water. And another bottle of water. 3-fix the bottles in their places and tide the pipes and valves. 4-power on the Electronic Console and the heater to increase the heat of the water in the bath and read the first temperature reading and its conductivity. 5-open the pump and read the readings of temperature and the conductivity for every 30 seconds. 6-after finishing readings for several times turn of the pump and the Electronic Console then pour all the liquid in the reactor.
  • 19. Discussion: Conductivity is a measure of how well a solution conducts electricity. To carry a current a solution must contain charged particles, or ions. Most conductivity measurements are made in aqueous solutions, and the ions responsible for the conductivity come from electrolytes dissolved in the water. Salts (like sodium chloride and magnesium sulfate), acids (like hydrochloric acid and acetic acid), and bases (like sodium hydroxide and ammonia) are all electrolytes. Although water itself is not an electrolyte, it does have a very small conductivity, implying that at least some ions are present. The ions are hydrogen and hydroxide, and they originate from the dissociation of molecular water. Salts, acids, and bases are electrolytes. They dissolve in water to form ions. Although water is not an electrolyte, a very small concentration of hydrogen and hydroxide ions are always present in pure water. Increasing the temperature of an electrolyte solution always increases the conductivity. The increase is significant, between 1.5 and 5.0% per °C. To compensate for temperature changes, conductivity readings are commonly corrected to the value at a reference temperature, typically 25°C. All process conductivity sensors have integral temperature sensors that allow
  • 20. Continue.. the analyzer to measure the process temperature and correct the raw conductivity. Three temperature correction algorithms are in common use. When we have to repeated conductivity measurement it means that there is a mistake in increasing temperature so that and it proves that the ions that conduct in the same temperature are equal in their conductivity motion. Position of conductivity cell Make sure that all the poles of the conductivity cell are completely covered by the sample. Always position a 2-pole cell in the center of the measuring vessel. Low conductivity measurements: • use a flow-through cell, to avoid atmospheric contamination from carbon dioxide, • use cells with low cell constant, 1 cm-1 or lower, • use non-platinized cells for easier cleaning and faster response, • make sure that your instrument is able to apply an appropriate measuring frequency High conductivity measurements • Use platinised cells to avoid polarisation errors, preferably 4-pole cells. • Use cells with a high cell constant (1 cm-1) or higher if possible.
  • 21. Continue.. • Do not dilute samples in order to bring them into measuring range. Conductivity is not proportional to concentration at high levels. • Make sure that your instrument is able to apply an appropriate measuring frequency. Since the accuracy of any measurement depends on proper calibration, a fresh standard must always be used. Ideally, sample beakers and sensor should be rinsed two to three times with the sample as the presence of contaminants can lead to additional errors in conductivity results When setting up a conductivity measurement system, the lab manager must make appropriate choices relating to TC(temperature compensation), such as o whether TC should be used, o setting the reference temperature, o choosing a linear or nonlinear TC, and o choosing the appropriate TC factor.