SlideShare a Scribd company logo
1 of 58
MRI Artifacts and Its
Remedies
Soni Nagarkoti
B.Sc.MIT 3rd year
NAMS Bir Hospital
Overview
• Introduction to artifacts
• Classification of artifacts
• Cause of the artifacts
• Appearance of the artifacts
• Remedies of the artifacts
What is artifact?
• Artifacts defined as either any signal or void in
the images that does not have an anatomic
basis, or as the result of distortion, addition or
deletion of information.
Significances of artifact
• Some artifacts degrade the images and may mask
or even mimic pathology
• Some artifacts are beneficial and we deliberately
create them to demonstrate flow, lesion
characterization and pathology visualization
• Some artifacts are irreversible and are reduced
rather than removed
• As artifacts characterization and identification is
essential, we need to understand them and how to
compensate them
Classification of the artifacts
• Hardware related artifacts
• Software related artifacts
• Artifacts due to physiologic motion
• Artifacts due to tissue heterogeneity and
foreign body
• Artifacts due to fourier transform and nyquist
theorem
Hardware based
1. Zipper artifact
2. Shading artifact
3. Moire artifact
4. Zebra artifact
5. Central point artifact
6. Equipment faults
Software based
1. Truncation artifact
2. Cross excitation/ cross talk
artifact
3. Aliasing artifacts
Motion based
1. Entry slice phenomenon/ in
flow artifact
2. Phase mismapping
Tissue inhomogeneity
based
1. Chemical shift artifacts
2. Susceptibility artifacts
3. Magic angle artifacts
4. Out of phase cancellation
artifact
5. Dielectric artifacts
Fourier transformation and
sampling based
1. Gibbs artifacts
2. Zero fill artifacts
3. Wrap around artifact
4. N/2 artifacts on EPI
Hardware based artifacts
1. Zipper artifact
• Appearance: appears as a dense broken line across
the image perpendicular to the frequency
encoding direction
• cause:
1. extraneous RF entering the room at a frequency
that matches the frequencies expected in the
echo
2. Faulty scan room doors
3. Breach in the RF cage
(The frequency of the interference tends to be of a
much higher amplitude compared to the spin-echo
and therefore appears in the image as a high-
intensity line representing that particular frequency
after FFT)
• Remedy:
1. Scan door should be closed
2. Engineers should locate breach in RF and repair
them
3. Equipments within scan room should be serviced
2.Moire artifacts
• Appearance: appears as a strange looking wavy pattern such
that in
• the fabrics
• Causes:
1.Multiple noise spike
In SE sequences, the noise from the static electric discharge in the scan
room causes spikes in data in the k space. After FFT, these spikes are
reconstructed as parallel lines. If these two spikes occur during same
acquisition, two sets of parallel lines will be reconstructed that interfere
the image and forms a woven pattern in image.
2.Field inhomogenity
This effect is seen in GRE by the combination of aliasing and field
inhomogeneity. aliasing causes anatomy outside FOV produce signals
and field inhomogeneity causes the superimposed signals to be in phase
out phase with the voxels on which it is wrapped. This gives spiky
appearance in the edge of image.
• Remedies:
1. Use of spin echo sequences
2. Use of anti aliasing remedies
3. Humidity in room should be in normal limits
4. Plug on the receive coil should be firmly sited
in the socket
3.Shading artifacts
• Appearance:
Appears as a difference in the signal intrnsity across the
imaging volume
• Causes :
 Uneven excitation of the nuclei within the patientd due
to the RF applied at flip angle less than 90° and 180°
 Abnormal loading of the coil or coupling the coil at one
point
 Patient anatomy touching at one side of the body coil
and coupling at that point
 Overflow of ADC
Remedies
• Loading the coil correctly
• using the proper size coil for patient size and the
examined part
• preventing the patient touching the coil (you can
use foam pads between patient and coil)
• shimming to reduce inhomogeneity of the
magnetic field
• using the proper scanning parameters to set proper
amplitude of applied RF pulses (less amplification
to avoid analog to digital converter over flow)
4.Zebra artifact
• Appearance: appears as the black and white
stripes of zebra on the image obliquely
• Causes:
Data is missing in k space array or set to zero
during scanning
Abrupt change in signal from signal to no
signal
Remedies
• Using SE pulse sequences
• Expanding the FOV
• Using oversampling techniques to prevent
aliasing
• Using surface coils
5.Central point artifact
• Appearance: appears as a focal dot of increased or
decreased signal in the exact centre of the image
often with a surrounding ringing artifact
• This artifact was originally misinterpreted as a
multiple sclerosis plaque or a lacunar infarction
• Causes:
Constant offset of DC voltage in the receiver.
After FFT, the constant offset gives the bright dot
at the centre of image
Remedies
• Scanners have compensation software to avoid this
problem (DC Correction, Baseline Correction
• doubling the acquisition time.
• In case of failure, technical assistance is required.
• Repeating the sequence
• Maintaining constant temperature in scanner
• Software to estimate DC offset and adjust the data in K-
space
• Recalibration of the receiver and software correction
6.Equipment faults
• The loss of a gradient, for example, causes
image distortion, and eddy currents induced in
the gradient coils can cause phase artifacts as
they create additional unwanted phase shifts.
• data-acquisition errors cause a variety of
different artifacts, most of which are
characterized by a geometric appearance in the
image,such as well-defined bands of missing
signal. These
software based artifacts
1.Truncation artifact
• Appearance: appears as a banding artifact at the
interface of high and low signals. It creates the low
intensity band running through high intensity areas.
• Causes:
• results from undersampling of data (too few k-space
lines are filled) so that interfaces of high and low signal
are incorrectly represented on the image.
• is most common when tissue is still producing a high
signal at the end of data collection or when the peak of
the echo is not centered in the middle of the sampling
window.
Remedy
• Undersampling of data should be avoided.
• Phase matrix should be increased. For example,
use a 512 × 512 matrix instead of 512 × 128.
• Techniques that partially fill k spaces should be
avoided
• Fat suppression techniques used in T1-weighted
imaging might reduce this artifact by nulling high
signal from fat at the beginning and end of the
sampling window.
• filters can be used that force the signal amplitude
to zero at the end of the sampling window
2.Cross excitation artifact
• Appearance:
This artifact causes a reduction in SNR in adjacent slices
in the slice stack
• Causes :
• Ideally, the profile of a slice should be square, or rather
rectangular, when viewed from the edge but in practice,
an RF excitation pulse is not able to achieve this.
• The adjacent slices receive energy from the RF
excitation pulse of their neighbors where the overlap
occurs .
• This energy pushes the magnetic moments of nuclei
toward the transverse plane, so that they may become
saturated when they are excited by their intended RF
pulse. This effect is called cross-excitation, and the
resulting saturation reduces SNR.
Remedy
• Cross-excitation can be reduced by ensuring
that there is at least a 30% gap between the
slices.
• This is 30% of the slice thickness itself and
reduces the likelihood of RF exciting adjacent
slices.
3.Aliasing artifact
• Appearance: appears as the wrap where anatomy hat
exist outside the FOV is folded onto the top of anatomy
inside FOV
• Causes:
 Anatomy outside the FOV still experiences the effects
of the gradients and produces a signal if it is within the
receiving volume of the receiver coil.
 The signal from this anatomy has frequencies that are
higher or lower than those within the FOV because
nuclei are positioned on parts of the gradient that
extend beyond the FOV.
 If the frequency exceeds the Nyquist frequency, it is not
accurately digitized and is represented as a lower
frequency
Types of aliasing
1. Frequency wrap
• When the FOV is smaller than the anatomy in the
frequency direction of the image, frequencies
outside the FOV are higher than the Nyquist
frequency and are mapped to a lower frequency.
This is called highfrequency aliasing
2.Phase wrap
• Aliasing along the phase axis of the image is
known as phase wrap. This is caused by
undersampling of data along the phase axis of the
image.
Remedy
• enlarging the FOV to incorporate all signal-
producing anatomy.
• use presaturation bands on areas outside the FOV
that may wrap into the image. These can
sometimes null signal from these areas and reduce
aliasing.
• two antialiasing software methods that
compensate for wrap i.e. antialiasing along
frequency axis and antialiasing along phase axis
motion based artifacts
1.Phase mismapping/ghosting
• Appearance:
Produces replication of moving anatomy across the image
in the phase encoding axis.
• Causes:
 Is produced by anatomy moving along the phase
encoding gradient like anterior abdominal wall during
respiration, eye movement, pulsation of vessels etc
 As anatomy moves during the scan, its reconstructed
signal is misplaced in the phase encoding direction as
the gradient amplitude changes.
 Repetitive motion results in periodic perturbations of
the data collected in k-space.
 After FFT, it is these perturbations that result in
“ghosts” of moving anatomy being mismapped into
incorrect spatial locations across the image
Remedy
• Swapping the phase and frequency encoding direction
• placing presaturation pulses over the source
of motion, causes signal from this motion decreases or is
nulled altogether
• Appropriate breath hold techniques should be instructed
to patients
• In longer sequences, respiratory compensation can be
used
• Respiratory gating and triggering and cardiac gating
and triggering should be used
Inflow artifacts/ entry slice
phenomenon
• Entry slice phenomenon is related to the excitation history
of the nuclei.
• Nuclei that receive repeated RF pulses during the
acquisition are said to be saturated or ‘beaten down’.
• The NMV of these nuclei eventually reach an equilibrium
position, and produce a signal according to the TE,TR, flip
angle and contrast characteristics of the tissue.
• Nuclei that have not received repeated RF pulses are said to
be ‘Fresh’ as their NMV has not been beaten down.
• The signal produced by the ‘fresh’ nuclei and the ‘beaten
down’ nuclei are different
• Stationary nuclei within a slice become saturated after
repeated RF pulses.
Causes
• Entry slice phenomenon occurs when unsaturated spins in
blood first enter into a slice or slices.
• It is characterised by the bright signal in a blood vessel
(artery or vein) at the first slice that the vessel enters.
• Usually, the signal is seen on more than one slice, fading
with distance.
• This mechanism is used in a positive fashion to generate
flight MR angiograms.
• This artifact has been confused with thrombosis with
disastrous results. The characteristic location and if
necessary, the use of gradient echo flow techniques can be
used to differentiate entry slice artifacts from occlusions.
Remedy
• Use of GE techniques are fundamental in the
differentiation between artifact and occlusion.
• Spatial saturation bands place before the first
slice and after the last can be used to eliminate
this artifact
Tissue heterogeneity and foreign
bodies based artifacts
1.Chemical shift artifact
• Appearance:
 Chemical shift artifact causes the misplacement of
signal from fat in the image.
 It can also create both signal voids and signal
superimposition in areas where fat and water interface
 For eg: fluid filled kidney surrounded by the perirenal
fat
 The signal loss represents a boundry where signal from
fat is shifted by certain number of voxels leaving
devoid of signal as a result area on opposite side of
kidney is seen hyperintense
Cause of chemical shift artifact
• The difference in the precessional frequencies between
magnetic moments in fat and water is called chemical shift,
or sometimes fat/water shift.
• The amount of chemical shift is often expressed in arbitrary
units known as parts per million (ppm) of the main
magnetic field strength which value is abt 3.5ppm
• Is caused by different chemical environment between fat
and water
• The self shielding nature of the fat molecule causes
lowering of the larmor frequency of the magnetic moment
of hydrogen nuclei in fat due to this fat and water are placed
in different pixels and they differ in the receive bandwidth
as well as a result they are sampled differently and
displayed independently across FOV
Remedy
• Scanning at lower field strength
• Minimizing FOV
• Use of wider bandwidth
• Use of fat supression techniques
• Higher readout gradients
Magnetic suceptibility artifacts
• Apperance: it produces distortion of image together
with larger signal voids
• Causes:
 the main causes of this artifact are metals within the
imaging volume,
 it can also be seen from naturally occurring
hemorrhage or iron deposition, as they magnetize to a
much greater degree than the surrounding tissue
 is more prominent in gradient-echo sequences, as the
gradient reversal cannot compensate for the phase
difference at the interface
Remedy
• Removal of all metal objects
• Use of spin echo sequences rather than gradient echo
sequence
• Decrease in the TE
• Use of the wide receive bandwidth when scanning
metal implant
Note
Under some circumstances, it aids in diagnosis. Small
hemorrhages are sometimes only seen because they
produce a magnetic susceptibility effect. The use of fruit
juices such as pineapple as negative “contrast agents” in
the bowel also relies on susceptibility-related dephasing.
Magic angle artifacts
• Appearance: produces abnormally high signal intensity
in the tissue that contaims collahen such as tendon
 Is most common in patellar tendon
• Causes:
 Occurs when the collagen structure lie at 55 degree to
the main magnetic field
 The anisotropic shape of the molecules in collagen
causes reduction to zero of spin–spin interactions.
 The T2 decay time increases in collagen structures
when they lie at this angle to B0.
 This causes increased signal intensity in the structure
when a short TE is used
Remedy
• The patient’s anatomy can be altered relative
to the B0 field or increase the TE. If the signal
remains hyperintense, it is likely to be fluid
(Alter the angle of the patient’s anatomy.
Out phase signal cancellation artifacts
• Appearance: produces a ring of dark signal
around certain organs where fat and water
interfaces occur within the same voxel.
 It degrades the image in gradient-echo pulse
sequences because gradient rephasing does not
compensate for this artifact.
• Cause:
• caused by the fact that the magnetic moments of
hydrogen precess at different frequencies, and the
differences in their phase positions are exhibited
by fat and water vectors at discrete points in time.
Out-of-phase signal cancellation show as a black line around the abdominal
organs border at the boundaries between fat and muscle
Remedy
• Selection of TE that matches the periodicity of fat and water
• Using spin echo sequences rather than gradient echo
sequences
• Using in phase TE
Note
• out-of-phase signal cancellation is clinically very useful.
• Any pathology that contains both fat and water exhibits
reduced signal on an out-of-phase image.
• Adrenal tumors are a good example.
• out-of-phase images provide a method of background
suppression in inflow angiography when examining the
cerebral circulation.
• There are computer algorithms that can manipulate data
from in-phase and out-of-phase images to provide fat-only
and water-only images. This is known as the Dixon
Technique.
Dielectric artifacts
• dielectric effect refers to the interaction of
matter with the E component of an
electromagnetic field.
• Appearance: appears as non uniform shading
• Abnormal bright and dark areas due to B1 field
inhomogeneity are frequently noted at very
high fields
Causes
• Caused by shortening of the RF
• At 3 T, the RF wavelength measures 234 cm in air, and
the speed and wavelength of the RF field is shortened
to ~26 cm within the body as a result of dielectric
effects. However, this 26 cm field of view is
approximately the cross-sectional diameter of most
body imaging studies.
• With patient abdominal diameters that exceed the RF
wavelength (e.g. patients with cirrhosis and ascites or
pregnant patients), constructive and destructive
interference patterns may emerge. In body MRI this
may lead to darkening/shading at the center of the
image
Remedy
• Using of the dielectric pads
• Imaging at lower field strengths
• Draining the ascites before imaging
References
• MRI in Practice 5th edison
• www.mriquestion.com
• Radiopaedia.org
• Google.com
Discussion
• Moire artifact
• Wrap around artifact
• motion/ ghost
• Susceptibility artifact
• Ring artifact of CT

More Related Content

What's hot

Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tubeSantosh Ojha
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
Mri coils
Mri coilsMri coils
Mri coilsSelf
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifactsshajitha khan
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedyYashawant Yadav
 
DUAL ENERGY CT-1.pptx
DUAL ENERGY CT-1.pptxDUAL ENERGY CT-1.pptx
DUAL ENERGY CT-1.pptxJyotiDalal22
 
Physics of 3 Tesla MRI & Silent MRI
Physics of 3 Tesla MRI & Silent MRIPhysics of 3 Tesla MRI & Silent MRI
Physics of 3 Tesla MRI & Silent MRIDr Varun Bansal
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pkDr pradeep Kumar
 
K Space in MRI
K Space in MRIK Space in MRI
K Space in MRIKajal Jha
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkDr pradeep Kumar
 
Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.Abdellah Nazeer
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdfmitians
 

What's hot (20)

MRI artifacts
MRI artifactsMRI artifacts
MRI artifacts
 
Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tube
 
CT detectors
CT detectorsCT detectors
CT detectors
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Ct image artifacts
Ct image artifactsCt image artifacts
Ct image artifacts
 
Mri coils
Mri coilsMri coils
Mri coils
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifacts
 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedy
 
DUAL ENERGY CT-1.pptx
DUAL ENERGY CT-1.pptxDUAL ENERGY CT-1.pptx
DUAL ENERGY CT-1.pptx
 
Physics of 3 Tesla MRI & Silent MRI
Physics of 3 Tesla MRI & Silent MRIPhysics of 3 Tesla MRI & Silent MRI
Physics of 3 Tesla MRI & Silent MRI
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pk
 
K Space in MRI
K Space in MRIK Space in MRI
K Space in MRI
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
 
Basics in ct
Basics in ctBasics in ct
Basics in ct
 
Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.
 
MRI Brain
MRI BrainMRI Brain
MRI Brain
 
Image reconstrsuction in ct pdf
Image reconstrsuction in ct pdfImage reconstrsuction in ct pdf
Image reconstrsuction in ct pdf
 

Similar to Artifacts

artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxKeerthan39
 
Artefacts mr & ct
Artefacts   mr & ctArtefacts   mr & ct
Artefacts mr & ctRakesh Ca
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound ArtifactAmer Sohail
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkDr pradeep Kumar
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundArjun Reddy
 
Optical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaOptical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaAkshay Nayak
 
6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptxRohit Bansal
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducersRaghu Kishore Galla
 
Artifacts and Quality Control
Artifacts and Quality ControlArtifacts and Quality Control
Artifacts and Quality ControlChaun Dee
 
Optical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptxOptical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptxSamanZebari1
 
Fingerprint Recognition
Fingerprint RecognitionFingerprint Recognition
Fingerprint RecognitionMirza Baig
 

Similar to Artifacts (20)

ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
Ct artiact
Ct artiactCt artiact
Ct artiact
 
mri artefacts.pptx
mri artefacts.pptxmri artefacts.pptx
mri artefacts.pptx
 
MRI ARTIFACTS.pptx
MRI ARTIFACTS.pptxMRI ARTIFACTS.pptx
MRI ARTIFACTS.pptx
 
Artefacts mr & ct
Artefacts   mr & ctArtefacts   mr & ct
Artefacts mr & ct
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound Artifact
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasound
 
Optical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --maculaOptical coherence tomography(OCT) --macula
Optical coherence tomography(OCT) --macula
 
Usg
UsgUsg
Usg
 
6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx6 CT ARTIFACTS-Rohit.pptx
6 CT ARTIFACTS-Rohit.pptx
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducers
 
Artifacts and Quality Control
Artifacts and Quality ControlArtifacts and Quality Control
Artifacts and Quality Control
 
OCT
OCTOCT
OCT
 
Optical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptxOptical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptx
 
Fingerprint Recognition
Fingerprint RecognitionFingerprint Recognition
Fingerprint Recognition
 
Digital image processing 2
Digital image processing   2Digital image processing   2
Digital image processing 2
 

Recently uploaded

VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiLow Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiSuhani Kapoor
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 

Recently uploaded (20)

VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiLow Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 

Artifacts

  • 1. MRI Artifacts and Its Remedies Soni Nagarkoti B.Sc.MIT 3rd year NAMS Bir Hospital
  • 2. Overview • Introduction to artifacts • Classification of artifacts • Cause of the artifacts • Appearance of the artifacts • Remedies of the artifacts
  • 3. What is artifact? • Artifacts defined as either any signal or void in the images that does not have an anatomic basis, or as the result of distortion, addition or deletion of information.
  • 4. Significances of artifact • Some artifacts degrade the images and may mask or even mimic pathology • Some artifacts are beneficial and we deliberately create them to demonstrate flow, lesion characterization and pathology visualization • Some artifacts are irreversible and are reduced rather than removed • As artifacts characterization and identification is essential, we need to understand them and how to compensate them
  • 5. Classification of the artifacts • Hardware related artifacts • Software related artifacts • Artifacts due to physiologic motion • Artifacts due to tissue heterogeneity and foreign body • Artifacts due to fourier transform and nyquist theorem
  • 6. Hardware based 1. Zipper artifact 2. Shading artifact 3. Moire artifact 4. Zebra artifact 5. Central point artifact 6. Equipment faults Software based 1. Truncation artifact 2. Cross excitation/ cross talk artifact 3. Aliasing artifacts
  • 7. Motion based 1. Entry slice phenomenon/ in flow artifact 2. Phase mismapping Tissue inhomogeneity based 1. Chemical shift artifacts 2. Susceptibility artifacts 3. Magic angle artifacts 4. Out of phase cancellation artifact 5. Dielectric artifacts
  • 8. Fourier transformation and sampling based 1. Gibbs artifacts 2. Zero fill artifacts 3. Wrap around artifact 4. N/2 artifacts on EPI
  • 10. 1. Zipper artifact • Appearance: appears as a dense broken line across the image perpendicular to the frequency encoding direction • cause: 1. extraneous RF entering the room at a frequency that matches the frequencies expected in the echo 2. Faulty scan room doors 3. Breach in the RF cage
  • 11. (The frequency of the interference tends to be of a much higher amplitude compared to the spin-echo and therefore appears in the image as a high- intensity line representing that particular frequency after FFT) • Remedy: 1. Scan door should be closed 2. Engineers should locate breach in RF and repair them 3. Equipments within scan room should be serviced
  • 12. 2.Moire artifacts • Appearance: appears as a strange looking wavy pattern such that in • the fabrics • Causes: 1.Multiple noise spike In SE sequences, the noise from the static electric discharge in the scan room causes spikes in data in the k space. After FFT, these spikes are reconstructed as parallel lines. If these two spikes occur during same acquisition, two sets of parallel lines will be reconstructed that interfere the image and forms a woven pattern in image. 2.Field inhomogenity This effect is seen in GRE by the combination of aliasing and field inhomogeneity. aliasing causes anatomy outside FOV produce signals and field inhomogeneity causes the superimposed signals to be in phase out phase with the voxels on which it is wrapped. This gives spiky appearance in the edge of image.
  • 13.
  • 14. • Remedies: 1. Use of spin echo sequences 2. Use of anti aliasing remedies 3. Humidity in room should be in normal limits 4. Plug on the receive coil should be firmly sited in the socket
  • 15. 3.Shading artifacts • Appearance: Appears as a difference in the signal intrnsity across the imaging volume • Causes :  Uneven excitation of the nuclei within the patientd due to the RF applied at flip angle less than 90° and 180°  Abnormal loading of the coil or coupling the coil at one point  Patient anatomy touching at one side of the body coil and coupling at that point  Overflow of ADC
  • 16. Remedies • Loading the coil correctly • using the proper size coil for patient size and the examined part • preventing the patient touching the coil (you can use foam pads between patient and coil) • shimming to reduce inhomogeneity of the magnetic field • using the proper scanning parameters to set proper amplitude of applied RF pulses (less amplification to avoid analog to digital converter over flow)
  • 17. 4.Zebra artifact • Appearance: appears as the black and white stripes of zebra on the image obliquely • Causes: Data is missing in k space array or set to zero during scanning Abrupt change in signal from signal to no signal
  • 18. Remedies • Using SE pulse sequences • Expanding the FOV • Using oversampling techniques to prevent aliasing • Using surface coils
  • 19. 5.Central point artifact • Appearance: appears as a focal dot of increased or decreased signal in the exact centre of the image often with a surrounding ringing artifact • This artifact was originally misinterpreted as a multiple sclerosis plaque or a lacunar infarction • Causes: Constant offset of DC voltage in the receiver. After FFT, the constant offset gives the bright dot at the centre of image
  • 20. Remedies • Scanners have compensation software to avoid this problem (DC Correction, Baseline Correction • doubling the acquisition time. • In case of failure, technical assistance is required. • Repeating the sequence • Maintaining constant temperature in scanner • Software to estimate DC offset and adjust the data in K- space • Recalibration of the receiver and software correction
  • 21. 6.Equipment faults • The loss of a gradient, for example, causes image distortion, and eddy currents induced in the gradient coils can cause phase artifacts as they create additional unwanted phase shifts. • data-acquisition errors cause a variety of different artifacts, most of which are characterized by a geometric appearance in the image,such as well-defined bands of missing signal. These
  • 23. 1.Truncation artifact • Appearance: appears as a banding artifact at the interface of high and low signals. It creates the low intensity band running through high intensity areas. • Causes: • results from undersampling of data (too few k-space lines are filled) so that interfaces of high and low signal are incorrectly represented on the image. • is most common when tissue is still producing a high signal at the end of data collection or when the peak of the echo is not centered in the middle of the sampling window.
  • 24. Remedy • Undersampling of data should be avoided. • Phase matrix should be increased. For example, use a 512 × 512 matrix instead of 512 × 128. • Techniques that partially fill k spaces should be avoided • Fat suppression techniques used in T1-weighted imaging might reduce this artifact by nulling high signal from fat at the beginning and end of the sampling window. • filters can be used that force the signal amplitude to zero at the end of the sampling window
  • 25. 2.Cross excitation artifact • Appearance: This artifact causes a reduction in SNR in adjacent slices in the slice stack • Causes : • Ideally, the profile of a slice should be square, or rather rectangular, when viewed from the edge but in practice, an RF excitation pulse is not able to achieve this. • The adjacent slices receive energy from the RF excitation pulse of their neighbors where the overlap occurs . • This energy pushes the magnetic moments of nuclei toward the transverse plane, so that they may become saturated when they are excited by their intended RF pulse. This effect is called cross-excitation, and the resulting saturation reduces SNR.
  • 26.
  • 27. Remedy • Cross-excitation can be reduced by ensuring that there is at least a 30% gap between the slices. • This is 30% of the slice thickness itself and reduces the likelihood of RF exciting adjacent slices.
  • 28. 3.Aliasing artifact • Appearance: appears as the wrap where anatomy hat exist outside the FOV is folded onto the top of anatomy inside FOV • Causes:  Anatomy outside the FOV still experiences the effects of the gradients and produces a signal if it is within the receiving volume of the receiver coil.  The signal from this anatomy has frequencies that are higher or lower than those within the FOV because nuclei are positioned on parts of the gradient that extend beyond the FOV.  If the frequency exceeds the Nyquist frequency, it is not accurately digitized and is represented as a lower frequency
  • 29. Types of aliasing 1. Frequency wrap • When the FOV is smaller than the anatomy in the frequency direction of the image, frequencies outside the FOV are higher than the Nyquist frequency and are mapped to a lower frequency. This is called highfrequency aliasing 2.Phase wrap • Aliasing along the phase axis of the image is known as phase wrap. This is caused by undersampling of data along the phase axis of the image.
  • 30. Remedy • enlarging the FOV to incorporate all signal- producing anatomy. • use presaturation bands on areas outside the FOV that may wrap into the image. These can sometimes null signal from these areas and reduce aliasing. • two antialiasing software methods that compensate for wrap i.e. antialiasing along frequency axis and antialiasing along phase axis
  • 32. 1.Phase mismapping/ghosting • Appearance: Produces replication of moving anatomy across the image in the phase encoding axis. • Causes:  Is produced by anatomy moving along the phase encoding gradient like anterior abdominal wall during respiration, eye movement, pulsation of vessels etc  As anatomy moves during the scan, its reconstructed signal is misplaced in the phase encoding direction as the gradient amplitude changes.  Repetitive motion results in periodic perturbations of the data collected in k-space.  After FFT, it is these perturbations that result in “ghosts” of moving anatomy being mismapped into incorrect spatial locations across the image
  • 33. Remedy • Swapping the phase and frequency encoding direction • placing presaturation pulses over the source of motion, causes signal from this motion decreases or is nulled altogether • Appropriate breath hold techniques should be instructed to patients • In longer sequences, respiratory compensation can be used • Respiratory gating and triggering and cardiac gating and triggering should be used
  • 34. Inflow artifacts/ entry slice phenomenon • Entry slice phenomenon is related to the excitation history of the nuclei. • Nuclei that receive repeated RF pulses during the acquisition are said to be saturated or ‘beaten down’. • The NMV of these nuclei eventually reach an equilibrium position, and produce a signal according to the TE,TR, flip angle and contrast characteristics of the tissue. • Nuclei that have not received repeated RF pulses are said to be ‘Fresh’ as their NMV has not been beaten down. • The signal produced by the ‘fresh’ nuclei and the ‘beaten down’ nuclei are different • Stationary nuclei within a slice become saturated after repeated RF pulses.
  • 35. Causes • Entry slice phenomenon occurs when unsaturated spins in blood first enter into a slice or slices. • It is characterised by the bright signal in a blood vessel (artery or vein) at the first slice that the vessel enters. • Usually, the signal is seen on more than one slice, fading with distance. • This mechanism is used in a positive fashion to generate flight MR angiograms. • This artifact has been confused with thrombosis with disastrous results. The characteristic location and if necessary, the use of gradient echo flow techniques can be used to differentiate entry slice artifacts from occlusions.
  • 36. Remedy • Use of GE techniques are fundamental in the differentiation between artifact and occlusion. • Spatial saturation bands place before the first slice and after the last can be used to eliminate this artifact
  • 37. Tissue heterogeneity and foreign bodies based artifacts
  • 38. 1.Chemical shift artifact • Appearance:  Chemical shift artifact causes the misplacement of signal from fat in the image.  It can also create both signal voids and signal superimposition in areas where fat and water interface  For eg: fluid filled kidney surrounded by the perirenal fat  The signal loss represents a boundry where signal from fat is shifted by certain number of voxels leaving devoid of signal as a result area on opposite side of kidney is seen hyperintense
  • 39.
  • 40. Cause of chemical shift artifact • The difference in the precessional frequencies between magnetic moments in fat and water is called chemical shift, or sometimes fat/water shift. • The amount of chemical shift is often expressed in arbitrary units known as parts per million (ppm) of the main magnetic field strength which value is abt 3.5ppm • Is caused by different chemical environment between fat and water • The self shielding nature of the fat molecule causes lowering of the larmor frequency of the magnetic moment of hydrogen nuclei in fat due to this fat and water are placed in different pixels and they differ in the receive bandwidth as well as a result they are sampled differently and displayed independently across FOV
  • 41. Remedy • Scanning at lower field strength • Minimizing FOV • Use of wider bandwidth • Use of fat supression techniques • Higher readout gradients
  • 42. Magnetic suceptibility artifacts • Apperance: it produces distortion of image together with larger signal voids • Causes:  the main causes of this artifact are metals within the imaging volume,  it can also be seen from naturally occurring hemorrhage or iron deposition, as they magnetize to a much greater degree than the surrounding tissue  is more prominent in gradient-echo sequences, as the gradient reversal cannot compensate for the phase difference at the interface
  • 43. Remedy • Removal of all metal objects • Use of spin echo sequences rather than gradient echo sequence • Decrease in the TE • Use of the wide receive bandwidth when scanning metal implant Note Under some circumstances, it aids in diagnosis. Small hemorrhages are sometimes only seen because they produce a magnetic susceptibility effect. The use of fruit juices such as pineapple as negative “contrast agents” in the bowel also relies on susceptibility-related dephasing.
  • 44. Magic angle artifacts • Appearance: produces abnormally high signal intensity in the tissue that contaims collahen such as tendon  Is most common in patellar tendon • Causes:  Occurs when the collagen structure lie at 55 degree to the main magnetic field  The anisotropic shape of the molecules in collagen causes reduction to zero of spin–spin interactions.  The T2 decay time increases in collagen structures when they lie at this angle to B0.  This causes increased signal intensity in the structure when a short TE is used
  • 45. Remedy • The patient’s anatomy can be altered relative to the B0 field or increase the TE. If the signal remains hyperintense, it is likely to be fluid (Alter the angle of the patient’s anatomy.
  • 46. Out phase signal cancellation artifacts • Appearance: produces a ring of dark signal around certain organs where fat and water interfaces occur within the same voxel.  It degrades the image in gradient-echo pulse sequences because gradient rephasing does not compensate for this artifact. • Cause: • caused by the fact that the magnetic moments of hydrogen precess at different frequencies, and the differences in their phase positions are exhibited by fat and water vectors at discrete points in time.
  • 47. Out-of-phase signal cancellation show as a black line around the abdominal organs border at the boundaries between fat and muscle
  • 48. Remedy • Selection of TE that matches the periodicity of fat and water • Using spin echo sequences rather than gradient echo sequences • Using in phase TE Note • out-of-phase signal cancellation is clinically very useful. • Any pathology that contains both fat and water exhibits reduced signal on an out-of-phase image. • Adrenal tumors are a good example. • out-of-phase images provide a method of background suppression in inflow angiography when examining the cerebral circulation. • There are computer algorithms that can manipulate data from in-phase and out-of-phase images to provide fat-only and water-only images. This is known as the Dixon Technique.
  • 49. Dielectric artifacts • dielectric effect refers to the interaction of matter with the E component of an electromagnetic field. • Appearance: appears as non uniform shading • Abnormal bright and dark areas due to B1 field inhomogeneity are frequently noted at very high fields
  • 50. Causes • Caused by shortening of the RF • At 3 T, the RF wavelength measures 234 cm in air, and the speed and wavelength of the RF field is shortened to ~26 cm within the body as a result of dielectric effects. However, this 26 cm field of view is approximately the cross-sectional diameter of most body imaging studies. • With patient abdominal diameters that exceed the RF wavelength (e.g. patients with cirrhosis and ascites or pregnant patients), constructive and destructive interference patterns may emerge. In body MRI this may lead to darkening/shading at the center of the image
  • 51. Remedy • Using of the dielectric pads • Imaging at lower field strengths • Draining the ascites before imaging
  • 52.
  • 53. References • MRI in Practice 5th edison • www.mriquestion.com • Radiopaedia.org • Google.com
  • 55. • Wrap around artifact

Editor's Notes

  1. When electromagnetic waves encounter the human body, several phenomena occur: 1) the wavelength decreases; 2) electrical currents are generated; and 3) wave reflection/refraction may develop at tissue interfaces.