ANAESTHESIA VENTILATORS
:CLASSIFICATION, PNEUMATICALLY
DRIVEN BELLOWS VENTILATORS,
MECHANICALLY DRIVEN PISTON
VENTILATORS
PRESENTOR- DR SNIGDHA
MODERATOR – DR BHAVNA GUPTA
DEFINITION
 A ventilator is an automatic device designed to provide
or augment patient ventilation
MODERN HISTORY
 John Hunter developed double bellows for resuscitation in 1775
- one for blowing air in and the other for drawing bad air out.
Draeger Medical designed an artificial breathing
device“Draeger Pulmoter” in 1911 that was used by fire and
police units
NEGATIVE PRESSURE VENTILATORS
 From mid 1800-1900s, devices were invented that
applied negative pressure around body or thoracic cavity.
Two successful designs became popular; one is -body of
patient was enclosed in an iron box or cylinder and
patient’s head protruded out of end 'iron lungs’.
 Second design was a box or shell that fitted over
thoracic area only (chest cuirass).
SCANDINAVIAN POLIO EPIDEMIC - 1952
 Between July-December of 1952, in Copenhagen, 2722
patients with poliomyelitis were treated in Community
Disease Hospital of which 315 patients required ventilatory
support. Many principles of IPPV were defined during that
time –including use of cuffed tubes, periodic breaths and
weaning by reduction of assisted breaths.
ERA OF RESPIRATORY INTENSIVE CARE
 Two types of ventilators and two modes of mechanical
ventilation evolved during this period-
1) Ventilator with pressure cycled (PCV).Two ventilators were
commonly used for PCV in 1960’s and 1970’s; the Bird Mark 7
and the Bennet PR2.
2) Volume cycled ventilator –VCV.
 The term ‘weaning’ was used to explain various techniques to
test the quality of patient’s spontaneous ventilation before
extubation
OLD ANAESTHESIA VENTILATORS
 Offers only volume control ventilation
 Separate models or different bellows assemblies were required
for adult and pediatric patients.
 DeliveredVt was affected by fresh gas flow and breathing
system compliance.
 User had to manually enable low pressure alarm when the
ventilator was turned ON.
 APL valve should have to be close and turn the bag/ventilator
switch when turning on ventilator.
OLD ANAESTHESIA VENTILATORS
 Could not provide as high inspiratory pressures or flows
as their ICU counterparts (eg. If positive end-expiratory
pressure (PEEP) were needed
 The anesthesia provider had to add a PEEP valve to the
anesthesia breathing system
NEW ANAESTHESIA VENTILATOR
 An integral PEEP valve is present , and many have several
ventilatory modes.
 High inspiratory pressures and flows can be delivered
 Ventilator can deliver volumes for a wide range of patients
from smallest child to the largest adult.
 Overcomes effect of fresh gas, breathing system
compliance and gas compression on tidal volume.
 Turning ventilator ON involves fewer steps and
automatically enables the low airway pressure alarm
CLASSIFICATION
A. POWER :
a) Low powered - generate only modest gas pressures required
to deliver Vt with normal and near-normal compliances and
resistances.
b) High powered - generate pressures sufficient to overcome the
increase in airways resistance and/or reduction in lung
compliance that are seen in diseased lungs;
Require addition of certain safety features to protect patients
with both normal and abnormal lungs from excessive pressures
B. CYCLING MECHANISM
 A ventilator cycle between two phases -inspiratory and
expiratory.
lnspiratory cycling
 During inspiratory phase, ventilator delivers- (a)a volume of
gas, over (b) a given period of time, producing (c) an increase in
airways pressure, also a change in the pattern of (d) flow
(inspiratory waveform) .
VOLUME CYCLING
 ventilator terminate inspiratory cycling -point at which pre-
determined volume of gas has left the ventilator and
switches its internal mechanism to allow exhalation to
occur.
TIME CYCLING
 Incorporated in most new ventilators.
 Mechanical, pneumatic or electronic timers used to control
operation of inspiratory and expiratory valves that govern
cycling of ventilator ; it functions independently of delivered
tidal volume
PRESSURE CYCLING
 ventilators sense predetermined airway pressure in order to
terminate inspiratory phase.
FLOW CYCLING
 Recognition of flow pattern changes has been used to cycle
ventilators.
 rarely employed nowadays.
Expiratory Cycling
expiratory volume-cycled ventilator have mechanism-
1. for terminating expiratory phase when reservoir bellows has filled
to desired tidal volume required for the next inspiration
2. to identify a selected airways pressure at the end of exhalation that
would trigger the next inspiratory phase
3. to switch to the inspiratory phase when desired flow rate at end of
exhalation was reached, or
4. of terminating expiratory phase after a predetermined time.
VENTILATION MODES IN ANAESTHESIA
VENTILATORS
 Volume control ventilation
 Pressure-controlled ventilation
 PCV volume guaranteed
 Synchronised intermittent mandatory ventilation
 Pressure-support ventilation
VOLUME CONTROL VENTILATION
 Preset volume is delivered with a constant flow
 Peak inflation pressure varies with patient’s compliance and
airway resistance.
 Modern anaesthesia ventilators can deliver Vt : 20-1500 ml.
 Typical ventilator settings inVCV:
•Tidal volume: 6-10 ml/kg body weight.
• Rate: 8-12 breaths/min.
• PEEP: 0-5 cm H2O to start with and titrated.
PRESSURE-CONTROLLED VENTILATION
Inspiratory pressure is maintained constant andVt varies.
Inspired volume varies according to changes in compliance and
airway resistance.
Target pressure is adjusted to produce a reasonableVt to avoid
extremes of atelectasis and volutrauma.
 PCV mode is useful in neonatal surgeries, in pregnancy , in
laproscopic surgery and patients with ARDS.
PCV VOLUME GUARANTEED
 ventilator operates as in PC- mode, but a tidal volume target is
also set.
 It ensure that patient receives uniformVt regardless of
compliance changes caused by packs, retractors, position,
surgical exposure or relaxation
 ventilator delivers presetVt with low pressure using a
decelerating flow.
 PCV-VG breaths are characterised by a decelerating flow and a
square pressure waveform.
INTERMITTENT MANDATORY VENTILATION
 used for weaning patients from mechanical ventilation
 IMV-ventilator delivers mechanical (mandatory, automatic)
breaths at a preset rate and permits spontaneous, unassisted
breaths of a controllable inspiratory gas mixture between
mechanical breaths.
 spontaneous breaths utilizes either continuous gas flow within
circuit or a demand valve that opens to allow gas to flow from
a reservoir.
 Continuous gas flow at a rate greater than peak inspiratory
flow involves no additional work of breathing but requires a
large volume of fresh gas.
SYNCHRONISED INTERMITTENT MANDATORY
VENTILATION
 Synchronizes ventilator-delivered breaths with patient's
spontaneous breaths.
 Time between end of each mandatory breath and beginning of
next is subdivided into spontaneous breathing time and trigger
time.
 Reduces incidence of patient-ventilator disharmony and need
for sedation or narcosis for patient to tolerate mechanical
ventilation.
MANDATORY MINUTE VENTILATION
 Similar to IMV mode except that minimum minute volume is
set rather than R/R.
 Ventilator measures spontaneous minute volume, if found less
than preset mandatory minute volume, the difference b/w two
is delivered as mandatory breaths by ventilator at preset flow
&Vt.
 Suited for patients with variable respiratory drive
PRESSURE SUPPORT
 designed to augment patient's spontaneous breathing by
applying positive pressure to airway in response to patient-
initiated breaths.
 Disadvantage - if patient fails to make any respiratory effort, no
pressure-supported breaths will be initiated.
 A supported breath may be pressure or flow initiated
 When selected flow or sub-baseline pressure caused by a
spontaneous breath is reached, flow from ventilator begins and
set pressure is quickly reached and modulates flow to maintain
that pressure.
 Desired tidal volume should be calculated and pressure support
level adjusted so that the desired volume is delivered.
 If exhaled volume is inadequate, inspiratory pressure should be
increased or inspiratory rise time decreased (if adjustable). PEEP
may cause an increase in tidal volume.
 As the patient's effort increases, inspiratory pressure can be
reduced.
CLASSIFICATION(according to
application)
1. MECHANICALTHUMBS
2. MINUTEVOLUME DIVIDERS
3. BAG SQEEZERS
4. INTERMITTENT BLOWERS
MECHANICAL THUMBS
MINUTE VOLUME DIVIDERS
 uses a continuous source pressurised gas fed into a ventilator
system -collected by a reservoir-continually pressurised by a
spring, a weight or its own elastic recoil.
 It has one inspiratory valve and another expiratory valve, which
are linked together and operated by a “bistable” mechanism.
 Examples of minute volume dividers are East-Freeman
automatic vent, the Flomasta and Manley MP3
BAG SQUEEZERS
 employed in conjunction with a circle or Mapleson D system.
 Various type of bag (bellow squeezers)-
A. rising bellows arrangement B. descending bellows
arrangement C. pneumatic piston with mechanical linkage
D. pneumatic piston E. cam driven linkage from an electric motor
F. screw threaded piston (worm drive) powered by electric
motor.
Manley servovent®, Penlon Nuffield 400 series ventilator,
Ohmeda 7800, Servo 900 series
INTERMITTENT BLOWERS
 Ventilators are driven by a source of gas or air, at a pressure of
45-60 psi.The driving gas is normally delivered to patient
undiluted, but it may be passed through a venturi device so that
air, oxygen or anaesthetic gases may be added to it.
 Major component is - electronically timed and activated
proportional flow valve or a pneumatically timed oscillator that
divides driving gas into tidal volumes; size and rate of which can
be adjusted.
 Pneupac and Penlon Nuffield 200 series ventilator
CLASSIFICATION OF INTERMITTENT BLOWERS:
A. Basic resuscitator B. Sophisticated resuscitator
C. Ventilator for intensive care D. Anaesthetic
ventilator for Mapleson D system
WORKING PRINCIPLE
PNEUMATICALLY DRIVEN BELLOWSVENTILATORS
bellows are analogous to reservoir bag in breathing circuit
 Act as interface between breathing system gas and
ventilator driving gas.
 driving gas circuit is located outside bellows and patient gas
circuit inside bellows.
FUNCTIONING OF BELLOWS-IN–BOX VENTILATOR
INSPIRATION
INSPIRATORY PAUSE
EXHALATION
EXPIRATORY PAUSE
WORKING PRINCIPLE OF MECHANICALLY
DRIVEN PISTON VENTILATORS
 ventilators use electric motor to compress gas in breathing
circuit; Motor’s force compresses gas within piston, raising
pressure within it- causing gas to flow into patient’s lungs
 area of piston is fixed, so volume delivered by piston directly
related to linear movement of piston.
SPONTANEOUS INHALATION
SPONTANEOUS EXHALATION
MANUAL
MECHANICAL VENTILATION (INSPIRATION)
MID EXHALATION
EXHALATION ( LATER PART)
ADVANTAGE AND DISADVANTAGE OF
PNEUMATICALLY DRIVEN BELLOWS
VENTILATORS OVER MECHANICALLY
DRIVEN PISTON VENTILATORS
Advantages of piston ventilators:
 Greater precision of tidal volume delivery due to rigid piston
design, decreased compliance losses
 Greater precision of pressure control with the use of pressure
sensors
 Electrical control instead of pneumatic control
 Economical use of wall oxygen (wall oxygen not used to compress
a bellows, only used for patient delivery)
 A perforation in the bellows can allow the driving gas to be
delivered to the patient.This can potentially cause barotrauma or
unpredictable inspired gas concentrations, including oxygen and
volatile anesthetic.
 No intrinsic PEEP (compared to ascending bellows (2-3 cm water
are mandatory due to the design of ventilator spill valve).)
 Quiet
Disadvantages of piston ventilators
 Leak in piston diaphragm can lead to hypoventilation
 Possible entrainment of room air as piston returns to
filled position
 Loss of the familiar visible behaviour of a standing
bellows during disconnections.
 Quiet and less easy to hear regular cycling.
Anaesthesia ventilators

Anaesthesia ventilators

  • 1.
    ANAESTHESIA VENTILATORS :CLASSIFICATION, PNEUMATICALLY DRIVENBELLOWS VENTILATORS, MECHANICALLY DRIVEN PISTON VENTILATORS PRESENTOR- DR SNIGDHA MODERATOR – DR BHAVNA GUPTA
  • 2.
    DEFINITION  A ventilatoris an automatic device designed to provide or augment patient ventilation
  • 3.
    MODERN HISTORY  JohnHunter developed double bellows for resuscitation in 1775 - one for blowing air in and the other for drawing bad air out. Draeger Medical designed an artificial breathing device“Draeger Pulmoter” in 1911 that was used by fire and police units
  • 5.
    NEGATIVE PRESSURE VENTILATORS From mid 1800-1900s, devices were invented that applied negative pressure around body or thoracic cavity. Two successful designs became popular; one is -body of patient was enclosed in an iron box or cylinder and patient’s head protruded out of end 'iron lungs’.  Second design was a box or shell that fitted over thoracic area only (chest cuirass).
  • 7.
    SCANDINAVIAN POLIO EPIDEMIC- 1952  Between July-December of 1952, in Copenhagen, 2722 patients with poliomyelitis were treated in Community Disease Hospital of which 315 patients required ventilatory support. Many principles of IPPV were defined during that time –including use of cuffed tubes, periodic breaths and weaning by reduction of assisted breaths.
  • 8.
    ERA OF RESPIRATORYINTENSIVE CARE  Two types of ventilators and two modes of mechanical ventilation evolved during this period- 1) Ventilator with pressure cycled (PCV).Two ventilators were commonly used for PCV in 1960’s and 1970’s; the Bird Mark 7 and the Bennet PR2. 2) Volume cycled ventilator –VCV.  The term ‘weaning’ was used to explain various techniques to test the quality of patient’s spontaneous ventilation before extubation
  • 9.
    OLD ANAESTHESIA VENTILATORS Offers only volume control ventilation  Separate models or different bellows assemblies were required for adult and pediatric patients.  DeliveredVt was affected by fresh gas flow and breathing system compliance.  User had to manually enable low pressure alarm when the ventilator was turned ON.  APL valve should have to be close and turn the bag/ventilator switch when turning on ventilator.
  • 10.
    OLD ANAESTHESIA VENTILATORS Could not provide as high inspiratory pressures or flows as their ICU counterparts (eg. If positive end-expiratory pressure (PEEP) were needed  The anesthesia provider had to add a PEEP valve to the anesthesia breathing system
  • 11.
    NEW ANAESTHESIA VENTILATOR An integral PEEP valve is present , and many have several ventilatory modes.  High inspiratory pressures and flows can be delivered  Ventilator can deliver volumes for a wide range of patients from smallest child to the largest adult.  Overcomes effect of fresh gas, breathing system compliance and gas compression on tidal volume.  Turning ventilator ON involves fewer steps and automatically enables the low airway pressure alarm
  • 12.
    CLASSIFICATION A. POWER : a)Low powered - generate only modest gas pressures required to deliver Vt with normal and near-normal compliances and resistances. b) High powered - generate pressures sufficient to overcome the increase in airways resistance and/or reduction in lung compliance that are seen in diseased lungs; Require addition of certain safety features to protect patients with both normal and abnormal lungs from excessive pressures
  • 13.
    B. CYCLING MECHANISM A ventilator cycle between two phases -inspiratory and expiratory. lnspiratory cycling  During inspiratory phase, ventilator delivers- (a)a volume of gas, over (b) a given period of time, producing (c) an increase in airways pressure, also a change in the pattern of (d) flow (inspiratory waveform) .
  • 14.
    VOLUME CYCLING  ventilatorterminate inspiratory cycling -point at which pre- determined volume of gas has left the ventilator and switches its internal mechanism to allow exhalation to occur. TIME CYCLING  Incorporated in most new ventilators.  Mechanical, pneumatic or electronic timers used to control operation of inspiratory and expiratory valves that govern cycling of ventilator ; it functions independently of delivered tidal volume
  • 15.
    PRESSURE CYCLING  ventilatorssense predetermined airway pressure in order to terminate inspiratory phase. FLOW CYCLING  Recognition of flow pattern changes has been used to cycle ventilators.  rarely employed nowadays.
  • 16.
    Expiratory Cycling expiratory volume-cycledventilator have mechanism- 1. for terminating expiratory phase when reservoir bellows has filled to desired tidal volume required for the next inspiration 2. to identify a selected airways pressure at the end of exhalation that would trigger the next inspiratory phase 3. to switch to the inspiratory phase when desired flow rate at end of exhalation was reached, or 4. of terminating expiratory phase after a predetermined time.
  • 17.
    VENTILATION MODES INANAESTHESIA VENTILATORS  Volume control ventilation  Pressure-controlled ventilation  PCV volume guaranteed  Synchronised intermittent mandatory ventilation  Pressure-support ventilation
  • 18.
    VOLUME CONTROL VENTILATION Preset volume is delivered with a constant flow  Peak inflation pressure varies with patient’s compliance and airway resistance.  Modern anaesthesia ventilators can deliver Vt : 20-1500 ml.  Typical ventilator settings inVCV: •Tidal volume: 6-10 ml/kg body weight. • Rate: 8-12 breaths/min. • PEEP: 0-5 cm H2O to start with and titrated.
  • 20.
    PRESSURE-CONTROLLED VENTILATION Inspiratory pressureis maintained constant andVt varies. Inspired volume varies according to changes in compliance and airway resistance. Target pressure is adjusted to produce a reasonableVt to avoid extremes of atelectasis and volutrauma.  PCV mode is useful in neonatal surgeries, in pregnancy , in laproscopic surgery and patients with ARDS.
  • 23.
    PCV VOLUME GUARANTEED ventilator operates as in PC- mode, but a tidal volume target is also set.  It ensure that patient receives uniformVt regardless of compliance changes caused by packs, retractors, position, surgical exposure or relaxation  ventilator delivers presetVt with low pressure using a decelerating flow.  PCV-VG breaths are characterised by a decelerating flow and a square pressure waveform.
  • 24.
    INTERMITTENT MANDATORY VENTILATION used for weaning patients from mechanical ventilation  IMV-ventilator delivers mechanical (mandatory, automatic) breaths at a preset rate and permits spontaneous, unassisted breaths of a controllable inspiratory gas mixture between mechanical breaths.  spontaneous breaths utilizes either continuous gas flow within circuit or a demand valve that opens to allow gas to flow from a reservoir.  Continuous gas flow at a rate greater than peak inspiratory flow involves no additional work of breathing but requires a large volume of fresh gas.
  • 25.
    SYNCHRONISED INTERMITTENT MANDATORY VENTILATION Synchronizes ventilator-delivered breaths with patient's spontaneous breaths.  Time between end of each mandatory breath and beginning of next is subdivided into spontaneous breathing time and trigger time.  Reduces incidence of patient-ventilator disharmony and need for sedation or narcosis for patient to tolerate mechanical ventilation.
  • 28.
    MANDATORY MINUTE VENTILATION Similar to IMV mode except that minimum minute volume is set rather than R/R.  Ventilator measures spontaneous minute volume, if found less than preset mandatory minute volume, the difference b/w two is delivered as mandatory breaths by ventilator at preset flow &Vt.  Suited for patients with variable respiratory drive
  • 29.
    PRESSURE SUPPORT  designedto augment patient's spontaneous breathing by applying positive pressure to airway in response to patient- initiated breaths.  Disadvantage - if patient fails to make any respiratory effort, no pressure-supported breaths will be initiated.  A supported breath may be pressure or flow initiated  When selected flow or sub-baseline pressure caused by a spontaneous breath is reached, flow from ventilator begins and set pressure is quickly reached and modulates flow to maintain that pressure.
  • 30.
     Desired tidalvolume should be calculated and pressure support level adjusted so that the desired volume is delivered.  If exhaled volume is inadequate, inspiratory pressure should be increased or inspiratory rise time decreased (if adjustable). PEEP may cause an increase in tidal volume.  As the patient's effort increases, inspiratory pressure can be reduced.
  • 31.
    CLASSIFICATION(according to application) 1. MECHANICALTHUMBS 2.MINUTEVOLUME DIVIDERS 3. BAG SQEEZERS 4. INTERMITTENT BLOWERS
  • 32.
  • 33.
    MINUTE VOLUME DIVIDERS uses a continuous source pressurised gas fed into a ventilator system -collected by a reservoir-continually pressurised by a spring, a weight or its own elastic recoil.  It has one inspiratory valve and another expiratory valve, which are linked together and operated by a “bistable” mechanism.  Examples of minute volume dividers are East-Freeman automatic vent, the Flomasta and Manley MP3
  • 35.
    BAG SQUEEZERS  employedin conjunction with a circle or Mapleson D system.  Various type of bag (bellow squeezers)- A. rising bellows arrangement B. descending bellows arrangement C. pneumatic piston with mechanical linkage D. pneumatic piston E. cam driven linkage from an electric motor F. screw threaded piston (worm drive) powered by electric motor. Manley servovent®, Penlon Nuffield 400 series ventilator, Ohmeda 7800, Servo 900 series
  • 37.
    INTERMITTENT BLOWERS  Ventilatorsare driven by a source of gas or air, at a pressure of 45-60 psi.The driving gas is normally delivered to patient undiluted, but it may be passed through a venturi device so that air, oxygen or anaesthetic gases may be added to it.  Major component is - electronically timed and activated proportional flow valve or a pneumatically timed oscillator that divides driving gas into tidal volumes; size and rate of which can be adjusted.  Pneupac and Penlon Nuffield 200 series ventilator
  • 39.
    CLASSIFICATION OF INTERMITTENTBLOWERS: A. Basic resuscitator B. Sophisticated resuscitator C. Ventilator for intensive care D. Anaesthetic ventilator for Mapleson D system
  • 40.
    WORKING PRINCIPLE PNEUMATICALLY DRIVENBELLOWSVENTILATORS bellows are analogous to reservoir bag in breathing circuit  Act as interface between breathing system gas and ventilator driving gas.  driving gas circuit is located outside bellows and patient gas circuit inside bellows.
  • 41.
  • 44.
  • 45.
  • 46.
  • 47.
  • 48.
    WORKING PRINCIPLE OFMECHANICALLY DRIVEN PISTON VENTILATORS  ventilators use electric motor to compress gas in breathing circuit; Motor’s force compresses gas within piston, raising pressure within it- causing gas to flow into patient’s lungs  area of piston is fixed, so volume delivered by piston directly related to linear movement of piston.
  • 51.
  • 52.
  • 53.
  • 54.
  • 55.
  • 56.
  • 57.
    ADVANTAGE AND DISADVANTAGEOF PNEUMATICALLY DRIVEN BELLOWS VENTILATORS OVER MECHANICALLY DRIVEN PISTON VENTILATORS Advantages of piston ventilators:  Greater precision of tidal volume delivery due to rigid piston design, decreased compliance losses  Greater precision of pressure control with the use of pressure sensors  Electrical control instead of pneumatic control
  • 58.
     Economical useof wall oxygen (wall oxygen not used to compress a bellows, only used for patient delivery)  A perforation in the bellows can allow the driving gas to be delivered to the patient.This can potentially cause barotrauma or unpredictable inspired gas concentrations, including oxygen and volatile anesthetic.  No intrinsic PEEP (compared to ascending bellows (2-3 cm water are mandatory due to the design of ventilator spill valve).)  Quiet
  • 59.
    Disadvantages of pistonventilators  Leak in piston diaphragm can lead to hypoventilation  Possible entrainment of room air as piston returns to filled position  Loss of the familiar visible behaviour of a standing bellows during disconnections.  Quiet and less easy to hear regular cycling.

Editor's Notes

  • #36 Bellows may be squeezed pneumatically by placing it in a canister and feeding a driver gas in the space between canister and bellows or squeezed mechanically by means of a motor and suitable gears and levers or by a spring or a weight.