Radiation Safety Training Module: Diagnostic Radiology
Medical Imaging Techniques
Radiological Safety Division
Atomic Energy Regulatory Board
Contents
• Objective
• Expected questions to know after
studying this lecture
• Introduction
• Modalities in Radiology
• Radiography
• Detectors used in X-ray Imaging
• Fluoroscopy
• Mammography
• Dental Radiography
• Bone Mineral Densitometry
• Computed Tomography
• Image Properties
• Typical Technical Parameters of
Diagnostic X-ray Equipment
• Summary/ Learning Outcomes
• Expected Questions
• References and sources for
additional information
Objective
• To become familiar with basic knowledge of the components of the radiographic
chain.
• To become familiar with the component of the fluoroscopy system (design, technical
parameters that affect the fluoroscopic image quality and Quality Control).
• To understand the principles of interventional radiology system including equipment
design, operational consideration and Quality Control.
• To understand the principles of dental radiology system and mammography system
including design and Quality Control.
• To understand the principles and the technology of CT equipment.
Topics to know after studying this lecture
• Different modalities of X-ray imaging
• Different Detectors used in X-ray Imaging
• How X-ray image is formed in radiography?
• Difference between mammography system and diagnostic X-ray system
• Difference between Dental (intra-oral) and Dental (OPG) equipment
• Difference between radiography & CT images
Introduction
• X-rays were first discovered by German Physicist, W. C. Roentgen in 1895.
• In India, medical x-rays were first used in1898 within three years of the discovery of X-
rays.
• In 1972, Computed Tomography was first introduced in medical field.
• In 1913, Albert Solomon, German Surgeon, was the first scientist who used X-ray
techniques in breast cancer radiology which came to be known as mammography.
• In 1960, John R. Cameron invented the bone densitometer.
5
Introduction
• Medical imaging of the human body requires some form of energy (radiation). In imaging techniques used
in radiology, the energy used to produce the image must be capable of penetrating tissues.
• In diagnostic X-ray imaging, images are formed by the interaction of the X-ray beam with the patient.
• As the X-ray beam passes through the patient, the photons interact with the body tissues and are
absorbed/scattered by the patient. The degree of absorption is related to the density of the material that is in
the beam’s path.
• Dense objects (such as bone and metal) have a high degree of photon absorption, while less dense objects
(such as fat and water) absorbs less photons.
• The differential absorption of photons by different materials in the photons' path results in the beam exiting
the patient with different intensities. This is known as transmitted beam.
• A detector is used to measure the intensity variation, thus providing information on the different densities in
the beam’s path.
Imaging Modalities in Radiology
The different modes of making images are referred to as imaging modalities.
Each Imaging Modality has its own applications in medicine.
Various Modalities of Radiology are:
• Radiography
• Fluoroscopy
• Mammography
• Dental Radiography
• Bone Mineral Densitometry
• Computed Tomography
Origin of X-ray Source remains same for
all x-ray imaging devices
i.e X-ray tube
Evacuated
glass tube
Target
Filament
Radiography
• Radiography X-ray equipment has an x-ray tube on one side and an x-ray detector on the other side
of the patient. A short duration pulse of x-rays is emitted by the x-ray tube, a large fraction of the x-
rays interacts in the patient, some of the x-rays pass through the patient (transmitted x-rays) and
reach the detector. These transmitted x-rays form the radiographic image on the film/detector.
• In radiography, the image is formed with screen-film system, CR cassette or with digital detectors.
• In screen-film radiography, areas of high intensity (thus low material absorption) within transmitted
beam result in more blackening of the film, while areas of low intensity (thus high material absorption)
will result in less blackening of the film. The film will remain white in areas with no photons.
• Human body is made up of tissues with varying densities, in the film black corresponds to tissues with
little attenuation (such as air) and white corresponds to tissue with a high degree of attenuation (such
as bone).
Radiography....
• Radiography is a transmission imaging
modality.
• Radiography is also referred as projection
imaging modality.
• Projection imaging refers to the acquisition
of a two–dimensional image of the patient’s
three- dimensional anatomy.
• Typical tube voltages range from 50 -150
kVp.
Typical layout of a Radiography
room
Bone
Air
Soft
tissue
X
Primary
collimation
Film, fluorescent screen
or image intensifier
Beam intensity
at detector level
« Latent »
radiological
Image
formed by
transmitted
beam
Antiscatter Grid
Scattered
radiation
X-ray Image Formation
Image acquisition-Detectors used in Radiography
1) Screen film based
2) Photostimulable phosphor plates (PSP).
–Called CR (computed radiography)
3)Direct digital radiography (flat panel detectors)
-------Direct conversion (selenium)
-------Indirect conversion (scintillation)
12
Detectors used in X-ray Imaging
Screen- Film System
• The X-ray film is sandwiched between two intensifying screens. Radiographic film consists of one or two layers of
film emulsion {grains of silver halide, silver bromide (AgBr )and silver iodide (AgI) bounded in a gelatin base} coated
on a flexible sheet of Mylar (a type of plastic).
• Intensifying screens are made of a scintillating material, called phosphor. Earlier screens use calcium tungstate
phosphors, while now rare earth screens use gadolinium or lanthanum phosphors. Gadolinium Oxy Bromide and
Lanthanum Oxy Sulphide are some commonly used intensifying screen phosphors.
• Intensifying Screens are used to reduce the radiation dose to the patient in diagnostic radiography.
• As X-ray films are more sensitive to light than X-rays, the Intensifying screen are used to convert X-rays falling on it
into light. Each X-ray photon falling on the screen releases many light photons, which reach the film and form the
image.
• Image formation is done by transmitted radiation. However, scatter radiation also reaches the film, thereby reducing
the image quality. The antiscatter grid (bucky) is used to cut off the scatter radiation reaching the film. It is placed
above the screen-film during imaging.
Film-Screen Imaging
Traditionally, all X-ray image capture has been through X-ray film
Film base
Emulsion
Emulsion
Adhesive
layer
Protective
layer
Intensifying screens
• Film is relatively insensitive to X-rays directly
• Only about 2% of the X-rays would interact with the emulsion
• Requires unacceptably high doses to give a diagnostic image
• An intensifying screen is a phosphor sheet the same size as the film, which
converts the X-rays to a pattern of light photons
• The intensity of the light is proportional to the intensity of X-rays
• The pattern of light is then captured by the film
• One exception is intraoral dental radiography, where screens are not
practical
The film-cassette
• Flat, light tight box with pressure pads to
ensure film in good contact with the
screen(s) mounted on the front (and back)
• The tube side of the cassette is low atomic
number material (Z~6) to minimise
attenuation
• Rear of cassette often lead backed to
minimise back scatter (not in
mammography)
Film Processing
• The invisible latent image is made visible
by processing
• There are three stages to this process;
• Development
• Fixing
• Washing
Computed Radiography (CR)
Computed Radiography
• Computed radiography (CR) is a marketing term for photostimulable phosphor
(PSP) detector systems.
• When X-rays are absorbed by photostimulable
• Phosphors, some light is promptly emitted much of absorbed energy is trapped
in the PSP screen and can be read out later
• CR imaging plates are made of BaFBr and BaFI
• The imaging plate is exposed in a procedure identical to screen-film radiography
and the CR cassette is then brought to a CR reader unit
• The cassette is moved into the reader unit and the imaging plate is
mechanically removed
Rota ing Mirror
Tr.ans
a1lion
R.orUer'
S
PSP digitizer
Cassette and PSP
Workstation
Digital Radiography
Digital Radiography
Digital Radiography (DR)
• Receptor provides direct digital output
• No processor / reader required
• Images available in < 15 seconds
• Much less work for technologist
Digital Radiography System
• Digital radiography is
performed with a system of
the following functional
components:
• A digital image receptor
• A digital image processing unit
• An image management system
Risk of increased doses
• The wide dynamic range of digital detectors
results in good image quality while using high
dose technique at the entrance of the detector
and at the entrance of the patient
• This is not possible with conventional screen-film
systems since high dose techniques always result
in an image which is too dark
Detectors used in X-ray Imaging…
Flat Panel Detectors
Indirect Flat panel Detectors: Indirect flat panel detectors are sensitive to visible light, and an X-ray intensifying
screen is used to convert incident X-rays to light, which is then detected by the flat panel detector. Silicon is
commonly used flat panel detector.
Direct Flat panel Detectors: Direct flat panel detectors are made from a layer of photoconductor material on top
of a Thin-Film-Transistor (TFT) array. Selenium is commonly used as the photoconductor.
Indirect Flat Panel Detector Direct Flat Panel Detector
Fluoroscopy
• Fluoroscopy refers to the continuous acquisition of a
sequence of x-ray images over time, essentially a real-time
x-ray movie of the patient. It is called dynamic imaging.
• Most general-purpose fluoroscopy systems use television
technology, which provides images at the rate of 30 frames
per second.
• Newer fluoroscopy systems allow the acquisition of a real-
time digital sequence of images (digital video), that can be
played back as a movie loop.
• Fluoroscopy is used for positioning catheters in arteries, for
visualizing contrast agents in the gastrointestinal (GI) tract,
and for other medical applications such as invasive
therapeutic procedures where real-time image feedback is
necessary.
27
Fluoroscopy….
Fluoroscopy mode of Operation
Continuous Fluoroscopy
• Continuous fluoroscopy is the basic form of fluoroscopy. It uses a continuously on x-ray beam
using typically 0.5 and 4 mA (depending upon patient thicknesses).
• A video camera displays the image at 30 frames/sec, so that each fluoroscopic frame requires 33
milliseconds.
Pulsed Fluoroscopy
• In pulsed fluoroscopy, the x-ray generator produces a series of short x-ray pulses. Pulsed
fluoroscopy offers better image quality in fluoroscopic procedures where body motion is high.
• Pulsed fluoroscopy at variable frame rates (typically 30,15 and 7.5 frames/sec) allows the
fluoroscopist to reduce temporal resolution when it is not needed sparing dose in return.
Last –Frame –Hold
• When the fluoroscopist takes his or her foot off of the fluoroscopy pedal, rather than seeing a
blank monitor, last-frame-hold enables the last live image to be shown continuously, until the next
image acquisition.
• Last-frame-hold is very convenient and can reduce the dose to the patient.
Detectors used in Fluoroscopy…
29
Image Intensifier
The function of the x-ray Image Intensifier is to convert an x-ray image
into a minified light image.
Input screen (CsI): conversion of incident X-rays into light photons
1 X –ray photon creates  3,000 light photons
Photocathode: conversion of light photons into electrons
only 10 to 20% of light photons are converted into photoelectrons
Electrodes : focalization of electrons onto the output screen.
electrodes provide the electronic magnification
Output screen: conversion of accelerated electrons into light photons.
In fluoroscopy equipment following detectors are used:
• Intensifying Screen: Used in conventional fluoroscopy
• Image Intensifier: (discussed below)
• Digital detectors: As discussed in the earlier slides of Radiography
Fluoroscopy….
30
C-Arm equipment
• Ability to maneuver c-arm around patient
without moving the patient
• Often used as “mobile” unit
• Critical care units
• Operation Theatre
• Modern units allow fluro + radiography
Fluoroscopy….
31
Angiography
• Angiography is a specialized fluoroscopic
examination in which a contrast agent is used to
highlight vasculature in the patient.
• Contrast is a radiographic (high density) material
injected into the blood vessels of the patient.
• Vessels containing contrast show up dark on the
image, while areas without contrast show up bright.
• Advanced techniques, such as Digital Subtraction
Angiography can be utilized to improve vessel
visualization and also guide percutaneous tools.
Fluoroscopy…..
Interventional radiology
•Interventional radiology uses the high capacity C-arm
equipment with additional features to guide minimally
invasive surgical procedures.
•Interventional radiology comprises fluoroscopically
guided therapeutic and diagnostic techniques.
•Interventional techniques are used during diagnostic
interventions or for therapeutic purposes as well as
during surgical procedures to guide or monitor the
surgeon’s actions.
•A good knowledge of equipment specification and
characteristics is essential for an effective optimization of
radiation protection
Typical layout of an IR facility
Mammography
• X-ray mammography is the most reliable method of detecting breast
cancer.
• It is employed both as a screening tool and for diagnosis.
• Mammography also uses x-rays for breast imaging; however, there
are fundamental differences between a mammography system and a
diagnostic x-ray system. Due to the tissue characteristics of the
breast and pathology of interest, mammography systems utilize lower
tube potential (15-35 kVp ).
• Molybdenum (Mo) and Rhodium (Rh) are the commonly used
target/filter combinations in mammography. Characteristic X-ray
produced by these target materials is used for breast imaging.
• In addition, two compression plates are used to decrease breast
thickness and minimize motion, thus resulting in the less scatter
radiation and better overall image quality.
Typical layout of Mammo room
A) Dental (Intra-oral x-ray examination)
• Dedicated equipment used for radiography of tooth with low power
stationary anode tube
• Tube is deployed on an extendable arm for easier positioning
• Tube potential : 60-70 kVp
• Tube current : 6-7 mA
• Exposure is varied by altering the exposure time
• Exposure time varies from 30 ms to 2.5 s
• Focus-skin distance 10 cm for 60 kV,
20 cm for 60-75 kVp
• Diameter of lead cone (collimator) diameter shall be less than 7.5 cm
at the end of cone.
It uses x-ray film (without film) or sensor for imaging purpose as detector.
Dental Radiography
B) Dental (OPG)
• Both film and Tube head rotate around the patient:
The X-ray tube rotates around the patient head in one direction and the film
rotates in the opposite direction, while the patient sits (or stands) in a
stationary position.
• The collimator used in the panoramic x-ray machine is a lead plate with a
narrow vertical slit.
• The Tube head always rotates behind the patient head as the film rotates in
front of the patient.
• Tube potential 85-90 kV
• Tube current: 4-15 mA
• Exposure time : 12 s (typically standard OPG scan)
0.16-3.2 s (cephalometric projection)
• Flat panoramic cassette 15 cm x 30cm
Dental Radiography…
C) Dental (CBCT)
• Dental Cone Beam CT scanner uses a cone shaped x-ray beam
rather than a conventional linear fan beam, as in the case of
common CT, to provide images of the skull bony structures.
• Tube potential: 85-90 kV
• Tube current: 4-15 mA
• Leakage from tube housing: less than 1.0 mGy in one hour at 1m
from focus of Dental Cone Beam CT scanner.
Dental Radiography…
Bone Mineral Densitometry Equipment
• Bone density or bone mineral density (BMD) is the
amount of bone mineral in bone tissue.
• Bone density scanning, also called dual-energy x-ray
absorptiometry (DEXA) or bone densitometry.
• Two x-ray beams with different energies are aimed at
the patient’s bones. When soft tissue absorption is
subtracted out, the BMD can be determined.
• Tube potential : 80-140 kV
• Tube current : 1.5-7 mA
• Bone density testing is used to assess the strength of
the bones and the probability of fracture in persons at
risk for osteoporosis.
37
Computed Tomography
• Computed Tomography (CT) was introduced into clinical
practice in 1972 and revolutionized X-ray imaging by
providing high quality images which reproduced
transverse cross sections of the body.
• The technique offered in particular improved low
contrast resolution for better visualization of soft tissue,
but with relatively high absorbed radiation dose.
• CT uses a rotating X-ray tube, with the beam in the form
of a thin slice (about 0.5 - 10 mm).
• The “image” is a simple array of X-ray intensity, and
many hundreds of these are used to make the CT
image, which is a “slice” through the patient.
Helical Spiral (CT)
• If the X –ray tube can rotate constantly, the patient can then be moved continuously through the beam, making the
examination much faster.
• For helical scanners to work, the X –ray tube must rotate continuously.
• Helical CT imaging allows for the continuous movement of the CT Couch during imaging.
• Scanning Geometry
X-ray beam
Table movement
• Continuous Data Acquisition and Table Feed
Computed Tomography
Generator
• High frequency, 30 - 70 kW
X- ray tube
• Rotating anode, high thermal capacity: 3-7 MHU
• Dual focal spot sizes: about 0.8 and 1.4 mm
Gantry
• Aperture: > 70 cm of diameter
• Detectors: solid state; > 600 detectors
• Scanning time: <1 s, 1 - 4 s
• Slice thickness: 0.5 - 10 mm
• Spiral scanning: up to 1400 mm
Computed Tomography
• Computed Tomography (CT) images are produced by passing
x-rays through the body, at a large number of angles, by
rotating the x-ray tube around the body.
• Multiple linear array detectors, opposite the x-ray source,
collect the transmission projection data. The numerous data
points collected in this manner are synthesized by a computer
into a tomographic image of the patient.
• The term tomography refers to a picture (graph) of a slice
(tomo).
• CT is transmission technique that results in images of individual
slices of tissue in the patient.
• The advantage of a tomographic image over projection image
is its ability to display the anatomy in a slice of tissue in the
absence of over or underlying structures .
Typical layout of CT room
Computed Tomography
Sr. No. Type of diagnostic x-ray
equipment
Maximum operating
potential (kVp)
Maximum operating
current (mA)
1 Radiography/Radiography &
fluoroscopy
150 1100
2 C-Arm 110-120 150
3 Dental (intra-oral) 60-70 6-7 (up to 10 mA)
4 Dental (extra-oral) [OPG/CBCT] 85-90 10-15
5 Computed Tomography 140 800
6 Interventional Radiology 150 1250
7 Mammography 49 (Mainly 35 kVp) 200
8 Bone Densitometer 140 5-7
Typical Technical Parameters of Diagnostic X-ray Equipment
Summary/Learning Outcomes
• This Presentation gives a brief overview of all diagnostic x-ray imaging
modalities.
• This presentation elaborates the technical aspects of various X-ray
imaging modalities, image detection systems, operating parameters and
image properties of medical diagnostic X-ray equipment.
• This presentation will be helpful for the medical professionals associated
with use of diagnostic X-ray equipment for understanding the various x-
ray based imaging modalities and their applications in the field of
medicine.
Expected Questions
Q.1 How radiography examination is performed of a patient?
Ans. Radiography is performed with an x-ray source on one side and an x-ray detector on the other
side of the patient. A short duration pulse of x-rays is emitted by the x-ray tube, a large fraction of
the x-rays interacts in the patient, some of the x-rays pass through the patient (transmitted x-rays)
and reach the detector. These transmitted x-rays form the radiographic image on the film/detector.
Q.2 Why intensifying screens are used in radiographic examinations?
Ans. Intensifying Screens are used to reduce the radiation dose to the patient in diagnostic radiography.
The main function of the Intensifying screen is to convert x-rays falling on it into light.
Q.3 What is the function of image intensifier?
Ans. The function of the x-ray Image Intensifier is to convert an x- ray image into a minified light image.
Q.4 What is fluoroscopy?
Ans. Fluoroscopy refers to the continuous acquisition of a sequence of x-ray images over time, essentially a real-time x-ray
movie of the patient. It is called dynamic imaging.
Q.5 How occupational exposure can be minimized in interventional radiological procedures?
Ans. Using protective lead aprons
Avoiding direct beam
Maintaining distance
Limiting the total beam–on time
Avoiding oblique lateral projections
Q.6 Which type of x-ray radiation is used in mammography examinations?
Ans. Characteristic x-ray radiation is used in mammography examinations.
Q.7 What is the limit of tube housing leakage for dental (intra oral) x-ray equipment?
Ans. Leakage from tube housing shall be leas than 0.25 mGy in one hour at 1m from focus of dental (intra-oral) equipment.
Expected Questions
Q.8 How Dental (OPG) scans of patient is performed?
Ans. In Dental (OPG) scans, the X-ray tube rotates around the patient head in one direction and the film
rotates in the opposite direction, while the patient sits (or stands) in a stationary position.
Q.9 What is the advantage of tomographic image over projection image?
Ans. The advantage of a tomographic image over projection image is its ability to display the anatomy in a
slab (slice) of tissue in the absence of over or underlying structures. It improves its image contrast
Q.10 What do you mean about the contrast of the image?
Ans. Contrast in an image is the difference in the gray scale of the image. It is the difference in darkness of
the image and its surroundings.
Expected Questions
References and sources for additional information
• The Essential Physics of Medical Imaging (J. T. Bushberg, J.A. Seibert, E.M.
Leidholdt, J M Boone)
• The Physics of Radiology (H.E. Johns, J.R. Cunnighnam)
• IAEA Presentations on Diagnostic Radiology
List of presentations in the training Module
Basics of Diagnostic X-ray Equipment
Biological effects of Radiations
Medical X-ray imaging techniques
Planning of Diagnostic X-ray facilities
Quality Assurance of X-ray equipment
Quality Assurance of Computed Tomography equipment
Radiation Protection in Diagnostic Radiology Practice
Causes, prevention and investigation of excessive exposures in
diagnostic radiology
Regulatory Requirements for Diagnostic Radiology Practice
3-MEDICAL-IMAGING………….     ……………………………….

More Related Content

Similar to 3-MEDICAL-IMAGING…………. ……………………………….

Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
mr_koky
 
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptxLecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
faiz3334
 

Similar to 3-MEDICAL-IMAGING…………. ………………………………. (20)

Radiology_Equipment_Lec-3_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-3_Dr. Emad Taleb.pdfRadiology_Equipment_Lec-3_Dr. Emad Taleb.pdf
Radiology_Equipment_Lec-3_Dr. Emad Taleb.pdf
 
radiation protection.pptx
radiation protection.pptxradiation protection.pptx
radiation protection.pptx
 
IMAGING TECHNIQUES
IMAGING TECHNIQUESIMAGING TECHNIQUES
IMAGING TECHNIQUES
 
Radiographic Considerations in Implants
Radiographic Considerations in ImplantsRadiographic Considerations in Implants
Radiographic Considerations in Implants
 
CR & DR
CR & DRCR & DR
CR & DR
 
digital imaging ppt.ppt
digital imaging ppt.pptdigital imaging ppt.ppt
digital imaging ppt.ppt
 
Technique 1 inroduction
Technique 1 inroductionTechnique 1 inroduction
Technique 1 inroduction
 
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
Floroscopy new microsoft office powerpoint 97 2003 presentation (2)
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
file004736.ppt
file004736.pptfile004736.ppt
file004736.ppt
 
CR DR system.pptx
CR DR system.pptxCR DR system.pptx
CR DR system.pptx
 
Radiographic image formation
Radiographic image formationRadiographic image formation
Radiographic image formation
 
Radiography Testing.ppt
Radiography Testing.pptRadiography Testing.ppt
Radiography Testing.ppt
 
Radiography Testing for Btech metallurgical and materirials science engineering
Radiography Testing for Btech metallurgical and materirials science engineeringRadiography Testing for Btech metallurgical and materirials science engineering
Radiography Testing for Btech metallurgical and materirials science engineering
 
intensifying screens and grids.pptx
intensifying screens and grids.pptxintensifying screens and grids.pptx
intensifying screens and grids.pptx
 
Cbct
CbctCbct
Cbct
 
X ray lecture 3
X ray lecture 3X ray lecture 3
X ray lecture 3
 
X ray machine l
X ray machine lX ray machine l
X ray machine l
 
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptxLecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
Lecture 3 & 4 anam sanam chick ldkfdlsfldfjdlsjfdlks .pptx
 

Recently uploaded

Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...
Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...
Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...
Levi Shapiro
 
Obat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di Makassar
Obat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di MakassarObat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di Makassar
Obat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di Makassar
clarintahafafa
 
Cara menggugurkan kandungan paling ampuh 08561234742
Cara menggugurkan kandungan paling ampuh 08561234742Cara menggugurkan kandungan paling ampuh 08561234742
Cara menggugurkan kandungan paling ampuh 08561234742
Jual obat penggugur 08561234742 Cara menggugurkan kandungan 08561234742
 
Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...
Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...
Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...
icha27638
 
Cytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi Arabia
Cytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi ArabiaCytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi Arabia
Cytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi Arabia
jaanualu31
 
Spauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCESpauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCE
DR.PRINCE C P
 

Recently uploaded (20)

Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...
Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...
Healthcare Market Overview, May 2024: Funding, Financing and M&A, from Oppenh...
 
Unlock the Secrets to Optimizing Ambulatory Operations Efficiency and Change ...
Unlock the Secrets to Optimizing Ambulatory Operations Efficiency and Change ...Unlock the Secrets to Optimizing Ambulatory Operations Efficiency and Change ...
Unlock the Secrets to Optimizing Ambulatory Operations Efficiency and Change ...
 
Obat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di Makassar
Obat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di MakassarObat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di Makassar
Obat Aborsi Makassar WA 085226114443 Jual Obat Aborsi Cytotec Asli Di Makassar
 
Cara menggugurkan kandungan paling ampuh 08561234742
Cara menggugurkan kandungan paling ampuh 08561234742Cara menggugurkan kandungan paling ampuh 08561234742
Cara menggugurkan kandungan paling ampuh 08561234742
 
Anthony Edwards We Want Dallas T-shirtsAnthony Edwards We Want Dallas T-shirts
Anthony Edwards We Want Dallas T-shirtsAnthony Edwards We Want Dallas T-shirtsAnthony Edwards We Want Dallas T-shirtsAnthony Edwards We Want Dallas T-shirts
Anthony Edwards We Want Dallas T-shirtsAnthony Edwards We Want Dallas T-shirts
 
Making change happen: learning from "positive deviancts"
Making change happen: learning from "positive deviancts"Making change happen: learning from "positive deviancts"
Making change happen: learning from "positive deviancts"
 
Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...
Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...
Obat aborsi Jakarta Timur Wa 081225888346 Jual Obat aborsi Cytotec asli Di Ja...
 
mHealth Israel_Healthcare Finance and M&A- What Comes Next
mHealth Israel_Healthcare Finance and M&A- What Comes NextmHealth Israel_Healthcare Finance and M&A- What Comes Next
mHealth Israel_Healthcare Finance and M&A- What Comes Next
 
Navigating Conflict in PE Using Strengths-Based Approaches
Navigating Conflict in PE Using Strengths-Based ApproachesNavigating Conflict in PE Using Strengths-Based Approaches
Navigating Conflict in PE Using Strengths-Based Approaches
 
Session-1-MBFHI-A-part-of-the-Global-Strategy.ppt
Session-1-MBFHI-A-part-of-the-Global-Strategy.pptSession-1-MBFHI-A-part-of-the-Global-Strategy.ppt
Session-1-MBFHI-A-part-of-the-Global-Strategy.ppt
 
Coach Dan Quinn Commanders Feather T Shirts
Coach Dan Quinn Commanders Feather T ShirtsCoach Dan Quinn Commanders Feather T Shirts
Coach Dan Quinn Commanders Feather T Shirts
 
I urgently need a love spell caster to bring back my ex. +27834335081 How can...
I urgently need a love spell caster to bring back my ex. +27834335081 How can...I urgently need a love spell caster to bring back my ex. +27834335081 How can...
I urgently need a love spell caster to bring back my ex. +27834335081 How can...
 
Cytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi Arabia
Cytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi ArabiaCytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi Arabia
Cytotec 200mcg tab in Riyadh (+919101817206// Get Abortion Pills in Saudi Arabia
 
Mike Lowe’s cancer fight lowe strong shirt
Mike Lowe’s cancer fight lowe strong shirtMike Lowe’s cancer fight lowe strong shirt
Mike Lowe’s cancer fight lowe strong shirt
 
Leadership Style - Code and Rapid Response Workshop
Leadership Style - Code and Rapid Response WorkshopLeadership Style - Code and Rapid Response Workshop
Leadership Style - Code and Rapid Response Workshop
 
Technology transfer documentation and strategies
Technology transfer documentation and strategiesTechnology transfer documentation and strategies
Technology transfer documentation and strategies
 
Top^Clinic ^%[+27785538335__Safe*Abortion Pills For Sale In Soweto
Top^Clinic ^%[+27785538335__Safe*Abortion Pills For Sale In SowetoTop^Clinic ^%[+27785538335__Safe*Abortion Pills For Sale In Soweto
Top^Clinic ^%[+27785538335__Safe*Abortion Pills For Sale In Soweto
 
Personnel and Equipment - Code and Rapid Response Workshop
Personnel and Equipment - Code and Rapid Response WorkshopPersonnel and Equipment - Code and Rapid Response Workshop
Personnel and Equipment - Code and Rapid Response Workshop
 
Session-3-Promoting-Breastfeeding-During-Pregnancy.ppt
Session-3-Promoting-Breastfeeding-During-Pregnancy.pptSession-3-Promoting-Breastfeeding-During-Pregnancy.ppt
Session-3-Promoting-Breastfeeding-During-Pregnancy.ppt
 
Spauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCESpauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCE
 

3-MEDICAL-IMAGING…………. ……………………………….

  • 1. Radiation Safety Training Module: Diagnostic Radiology Medical Imaging Techniques Radiological Safety Division Atomic Energy Regulatory Board
  • 2. Contents • Objective • Expected questions to know after studying this lecture • Introduction • Modalities in Radiology • Radiography • Detectors used in X-ray Imaging • Fluoroscopy • Mammography • Dental Radiography • Bone Mineral Densitometry • Computed Tomography • Image Properties • Typical Technical Parameters of Diagnostic X-ray Equipment • Summary/ Learning Outcomes • Expected Questions • References and sources for additional information
  • 3. Objective • To become familiar with basic knowledge of the components of the radiographic chain. • To become familiar with the component of the fluoroscopy system (design, technical parameters that affect the fluoroscopic image quality and Quality Control). • To understand the principles of interventional radiology system including equipment design, operational consideration and Quality Control. • To understand the principles of dental radiology system and mammography system including design and Quality Control. • To understand the principles and the technology of CT equipment.
  • 4. Topics to know after studying this lecture • Different modalities of X-ray imaging • Different Detectors used in X-ray Imaging • How X-ray image is formed in radiography? • Difference between mammography system and diagnostic X-ray system • Difference between Dental (intra-oral) and Dental (OPG) equipment • Difference between radiography & CT images
  • 5. Introduction • X-rays were first discovered by German Physicist, W. C. Roentgen in 1895. • In India, medical x-rays were first used in1898 within three years of the discovery of X- rays. • In 1972, Computed Tomography was first introduced in medical field. • In 1913, Albert Solomon, German Surgeon, was the first scientist who used X-ray techniques in breast cancer radiology which came to be known as mammography. • In 1960, John R. Cameron invented the bone densitometer. 5
  • 6. Introduction • Medical imaging of the human body requires some form of energy (radiation). In imaging techniques used in radiology, the energy used to produce the image must be capable of penetrating tissues. • In diagnostic X-ray imaging, images are formed by the interaction of the X-ray beam with the patient. • As the X-ray beam passes through the patient, the photons interact with the body tissues and are absorbed/scattered by the patient. The degree of absorption is related to the density of the material that is in the beam’s path. • Dense objects (such as bone and metal) have a high degree of photon absorption, while less dense objects (such as fat and water) absorbs less photons. • The differential absorption of photons by different materials in the photons' path results in the beam exiting the patient with different intensities. This is known as transmitted beam. • A detector is used to measure the intensity variation, thus providing information on the different densities in the beam’s path.
  • 7. Imaging Modalities in Radiology The different modes of making images are referred to as imaging modalities. Each Imaging Modality has its own applications in medicine. Various Modalities of Radiology are: • Radiography • Fluoroscopy • Mammography • Dental Radiography • Bone Mineral Densitometry • Computed Tomography
  • 8. Origin of X-ray Source remains same for all x-ray imaging devices i.e X-ray tube Evacuated glass tube Target Filament
  • 9. Radiography • Radiography X-ray equipment has an x-ray tube on one side and an x-ray detector on the other side of the patient. A short duration pulse of x-rays is emitted by the x-ray tube, a large fraction of the x- rays interacts in the patient, some of the x-rays pass through the patient (transmitted x-rays) and reach the detector. These transmitted x-rays form the radiographic image on the film/detector. • In radiography, the image is formed with screen-film system, CR cassette or with digital detectors. • In screen-film radiography, areas of high intensity (thus low material absorption) within transmitted beam result in more blackening of the film, while areas of low intensity (thus high material absorption) will result in less blackening of the film. The film will remain white in areas with no photons. • Human body is made up of tissues with varying densities, in the film black corresponds to tissues with little attenuation (such as air) and white corresponds to tissue with a high degree of attenuation (such as bone).
  • 10. Radiography.... • Radiography is a transmission imaging modality. • Radiography is also referred as projection imaging modality. • Projection imaging refers to the acquisition of a two–dimensional image of the patient’s three- dimensional anatomy. • Typical tube voltages range from 50 -150 kVp. Typical layout of a Radiography room
  • 11. Bone Air Soft tissue X Primary collimation Film, fluorescent screen or image intensifier Beam intensity at detector level « Latent » radiological Image formed by transmitted beam Antiscatter Grid Scattered radiation X-ray Image Formation
  • 12. Image acquisition-Detectors used in Radiography 1) Screen film based 2) Photostimulable phosphor plates (PSP). –Called CR (computed radiography) 3)Direct digital radiography (flat panel detectors) -------Direct conversion (selenium) -------Indirect conversion (scintillation) 12
  • 13. Detectors used in X-ray Imaging Screen- Film System • The X-ray film is sandwiched between two intensifying screens. Radiographic film consists of one or two layers of film emulsion {grains of silver halide, silver bromide (AgBr )and silver iodide (AgI) bounded in a gelatin base} coated on a flexible sheet of Mylar (a type of plastic). • Intensifying screens are made of a scintillating material, called phosphor. Earlier screens use calcium tungstate phosphors, while now rare earth screens use gadolinium or lanthanum phosphors. Gadolinium Oxy Bromide and Lanthanum Oxy Sulphide are some commonly used intensifying screen phosphors. • Intensifying Screens are used to reduce the radiation dose to the patient in diagnostic radiography. • As X-ray films are more sensitive to light than X-rays, the Intensifying screen are used to convert X-rays falling on it into light. Each X-ray photon falling on the screen releases many light photons, which reach the film and form the image. • Image formation is done by transmitted radiation. However, scatter radiation also reaches the film, thereby reducing the image quality. The antiscatter grid (bucky) is used to cut off the scatter radiation reaching the film. It is placed above the screen-film during imaging.
  • 14. Film-Screen Imaging Traditionally, all X-ray image capture has been through X-ray film Film base Emulsion Emulsion Adhesive layer Protective layer
  • 15. Intensifying screens • Film is relatively insensitive to X-rays directly • Only about 2% of the X-rays would interact with the emulsion • Requires unacceptably high doses to give a diagnostic image • An intensifying screen is a phosphor sheet the same size as the film, which converts the X-rays to a pattern of light photons • The intensity of the light is proportional to the intensity of X-rays • The pattern of light is then captured by the film • One exception is intraoral dental radiography, where screens are not practical
  • 16. The film-cassette • Flat, light tight box with pressure pads to ensure film in good contact with the screen(s) mounted on the front (and back) • The tube side of the cassette is low atomic number material (Z~6) to minimise attenuation • Rear of cassette often lead backed to minimise back scatter (not in mammography)
  • 17. Film Processing • The invisible latent image is made visible by processing • There are three stages to this process; • Development • Fixing • Washing
  • 19. Computed Radiography • Computed radiography (CR) is a marketing term for photostimulable phosphor (PSP) detector systems. • When X-rays are absorbed by photostimulable • Phosphors, some light is promptly emitted much of absorbed energy is trapped in the PSP screen and can be read out later • CR imaging plates are made of BaFBr and BaFI • The imaging plate is exposed in a procedure identical to screen-film radiography and the CR cassette is then brought to a CR reader unit • The cassette is moved into the reader unit and the imaging plate is mechanically removed
  • 21. PSP digitizer Cassette and PSP Workstation Digital Radiography
  • 23. Digital Radiography (DR) • Receptor provides direct digital output • No processor / reader required • Images available in < 15 seconds • Much less work for technologist
  • 24. Digital Radiography System • Digital radiography is performed with a system of the following functional components: • A digital image receptor • A digital image processing unit • An image management system
  • 25. Risk of increased doses • The wide dynamic range of digital detectors results in good image quality while using high dose technique at the entrance of the detector and at the entrance of the patient • This is not possible with conventional screen-film systems since high dose techniques always result in an image which is too dark
  • 26. Detectors used in X-ray Imaging… Flat Panel Detectors Indirect Flat panel Detectors: Indirect flat panel detectors are sensitive to visible light, and an X-ray intensifying screen is used to convert incident X-rays to light, which is then detected by the flat panel detector. Silicon is commonly used flat panel detector. Direct Flat panel Detectors: Direct flat panel detectors are made from a layer of photoconductor material on top of a Thin-Film-Transistor (TFT) array. Selenium is commonly used as the photoconductor. Indirect Flat Panel Detector Direct Flat Panel Detector
  • 27. Fluoroscopy • Fluoroscopy refers to the continuous acquisition of a sequence of x-ray images over time, essentially a real-time x-ray movie of the patient. It is called dynamic imaging. • Most general-purpose fluoroscopy systems use television technology, which provides images at the rate of 30 frames per second. • Newer fluoroscopy systems allow the acquisition of a real- time digital sequence of images (digital video), that can be played back as a movie loop. • Fluoroscopy is used for positioning catheters in arteries, for visualizing contrast agents in the gastrointestinal (GI) tract, and for other medical applications such as invasive therapeutic procedures where real-time image feedback is necessary. 27
  • 28. Fluoroscopy…. Fluoroscopy mode of Operation Continuous Fluoroscopy • Continuous fluoroscopy is the basic form of fluoroscopy. It uses a continuously on x-ray beam using typically 0.5 and 4 mA (depending upon patient thicknesses). • A video camera displays the image at 30 frames/sec, so that each fluoroscopic frame requires 33 milliseconds. Pulsed Fluoroscopy • In pulsed fluoroscopy, the x-ray generator produces a series of short x-ray pulses. Pulsed fluoroscopy offers better image quality in fluoroscopic procedures where body motion is high. • Pulsed fluoroscopy at variable frame rates (typically 30,15 and 7.5 frames/sec) allows the fluoroscopist to reduce temporal resolution when it is not needed sparing dose in return. Last –Frame –Hold • When the fluoroscopist takes his or her foot off of the fluoroscopy pedal, rather than seeing a blank monitor, last-frame-hold enables the last live image to be shown continuously, until the next image acquisition. • Last-frame-hold is very convenient and can reduce the dose to the patient.
  • 29. Detectors used in Fluoroscopy… 29 Image Intensifier The function of the x-ray Image Intensifier is to convert an x-ray image into a minified light image. Input screen (CsI): conversion of incident X-rays into light photons 1 X –ray photon creates  3,000 light photons Photocathode: conversion of light photons into electrons only 10 to 20% of light photons are converted into photoelectrons Electrodes : focalization of electrons onto the output screen. electrodes provide the electronic magnification Output screen: conversion of accelerated electrons into light photons. In fluoroscopy equipment following detectors are used: • Intensifying Screen: Used in conventional fluoroscopy • Image Intensifier: (discussed below) • Digital detectors: As discussed in the earlier slides of Radiography
  • 30. Fluoroscopy…. 30 C-Arm equipment • Ability to maneuver c-arm around patient without moving the patient • Often used as “mobile” unit • Critical care units • Operation Theatre • Modern units allow fluro + radiography
  • 31. Fluoroscopy…. 31 Angiography • Angiography is a specialized fluoroscopic examination in which a contrast agent is used to highlight vasculature in the patient. • Contrast is a radiographic (high density) material injected into the blood vessels of the patient. • Vessels containing contrast show up dark on the image, while areas without contrast show up bright. • Advanced techniques, such as Digital Subtraction Angiography can be utilized to improve vessel visualization and also guide percutaneous tools.
  • 32. Fluoroscopy….. Interventional radiology •Interventional radiology uses the high capacity C-arm equipment with additional features to guide minimally invasive surgical procedures. •Interventional radiology comprises fluoroscopically guided therapeutic and diagnostic techniques. •Interventional techniques are used during diagnostic interventions or for therapeutic purposes as well as during surgical procedures to guide or monitor the surgeon’s actions. •A good knowledge of equipment specification and characteristics is essential for an effective optimization of radiation protection Typical layout of an IR facility
  • 33. Mammography • X-ray mammography is the most reliable method of detecting breast cancer. • It is employed both as a screening tool and for diagnosis. • Mammography also uses x-rays for breast imaging; however, there are fundamental differences between a mammography system and a diagnostic x-ray system. Due to the tissue characteristics of the breast and pathology of interest, mammography systems utilize lower tube potential (15-35 kVp ). • Molybdenum (Mo) and Rhodium (Rh) are the commonly used target/filter combinations in mammography. Characteristic X-ray produced by these target materials is used for breast imaging. • In addition, two compression plates are used to decrease breast thickness and minimize motion, thus resulting in the less scatter radiation and better overall image quality. Typical layout of Mammo room
  • 34. A) Dental (Intra-oral x-ray examination) • Dedicated equipment used for radiography of tooth with low power stationary anode tube • Tube is deployed on an extendable arm for easier positioning • Tube potential : 60-70 kVp • Tube current : 6-7 mA • Exposure is varied by altering the exposure time • Exposure time varies from 30 ms to 2.5 s • Focus-skin distance 10 cm for 60 kV, 20 cm for 60-75 kVp • Diameter of lead cone (collimator) diameter shall be less than 7.5 cm at the end of cone. It uses x-ray film (without film) or sensor for imaging purpose as detector. Dental Radiography
  • 35. B) Dental (OPG) • Both film and Tube head rotate around the patient: The X-ray tube rotates around the patient head in one direction and the film rotates in the opposite direction, while the patient sits (or stands) in a stationary position. • The collimator used in the panoramic x-ray machine is a lead plate with a narrow vertical slit. • The Tube head always rotates behind the patient head as the film rotates in front of the patient. • Tube potential 85-90 kV • Tube current: 4-15 mA • Exposure time : 12 s (typically standard OPG scan) 0.16-3.2 s (cephalometric projection) • Flat panoramic cassette 15 cm x 30cm Dental Radiography…
  • 36. C) Dental (CBCT) • Dental Cone Beam CT scanner uses a cone shaped x-ray beam rather than a conventional linear fan beam, as in the case of common CT, to provide images of the skull bony structures. • Tube potential: 85-90 kV • Tube current: 4-15 mA • Leakage from tube housing: less than 1.0 mGy in one hour at 1m from focus of Dental Cone Beam CT scanner. Dental Radiography…
  • 37. Bone Mineral Densitometry Equipment • Bone density or bone mineral density (BMD) is the amount of bone mineral in bone tissue. • Bone density scanning, also called dual-energy x-ray absorptiometry (DEXA) or bone densitometry. • Two x-ray beams with different energies are aimed at the patient’s bones. When soft tissue absorption is subtracted out, the BMD can be determined. • Tube potential : 80-140 kV • Tube current : 1.5-7 mA • Bone density testing is used to assess the strength of the bones and the probability of fracture in persons at risk for osteoporosis. 37
  • 38. Computed Tomography • Computed Tomography (CT) was introduced into clinical practice in 1972 and revolutionized X-ray imaging by providing high quality images which reproduced transverse cross sections of the body. • The technique offered in particular improved low contrast resolution for better visualization of soft tissue, but with relatively high absorbed radiation dose. • CT uses a rotating X-ray tube, with the beam in the form of a thin slice (about 0.5 - 10 mm). • The “image” is a simple array of X-ray intensity, and many hundreds of these are used to make the CT image, which is a “slice” through the patient.
  • 39. Helical Spiral (CT) • If the X –ray tube can rotate constantly, the patient can then be moved continuously through the beam, making the examination much faster. • For helical scanners to work, the X –ray tube must rotate continuously. • Helical CT imaging allows for the continuous movement of the CT Couch during imaging. • Scanning Geometry X-ray beam Table movement • Continuous Data Acquisition and Table Feed Computed Tomography
  • 40. Generator • High frequency, 30 - 70 kW X- ray tube • Rotating anode, high thermal capacity: 3-7 MHU • Dual focal spot sizes: about 0.8 and 1.4 mm Gantry • Aperture: > 70 cm of diameter • Detectors: solid state; > 600 detectors • Scanning time: <1 s, 1 - 4 s • Slice thickness: 0.5 - 10 mm • Spiral scanning: up to 1400 mm Computed Tomography
  • 41. • Computed Tomography (CT) images are produced by passing x-rays through the body, at a large number of angles, by rotating the x-ray tube around the body. • Multiple linear array detectors, opposite the x-ray source, collect the transmission projection data. The numerous data points collected in this manner are synthesized by a computer into a tomographic image of the patient. • The term tomography refers to a picture (graph) of a slice (tomo). • CT is transmission technique that results in images of individual slices of tissue in the patient. • The advantage of a tomographic image over projection image is its ability to display the anatomy in a slice of tissue in the absence of over or underlying structures . Typical layout of CT room Computed Tomography
  • 42. Sr. No. Type of diagnostic x-ray equipment Maximum operating potential (kVp) Maximum operating current (mA) 1 Radiography/Radiography & fluoroscopy 150 1100 2 C-Arm 110-120 150 3 Dental (intra-oral) 60-70 6-7 (up to 10 mA) 4 Dental (extra-oral) [OPG/CBCT] 85-90 10-15 5 Computed Tomography 140 800 6 Interventional Radiology 150 1250 7 Mammography 49 (Mainly 35 kVp) 200 8 Bone Densitometer 140 5-7 Typical Technical Parameters of Diagnostic X-ray Equipment
  • 43. Summary/Learning Outcomes • This Presentation gives a brief overview of all diagnostic x-ray imaging modalities. • This presentation elaborates the technical aspects of various X-ray imaging modalities, image detection systems, operating parameters and image properties of medical diagnostic X-ray equipment. • This presentation will be helpful for the medical professionals associated with use of diagnostic X-ray equipment for understanding the various x- ray based imaging modalities and their applications in the field of medicine.
  • 44. Expected Questions Q.1 How radiography examination is performed of a patient? Ans. Radiography is performed with an x-ray source on one side and an x-ray detector on the other side of the patient. A short duration pulse of x-rays is emitted by the x-ray tube, a large fraction of the x-rays interacts in the patient, some of the x-rays pass through the patient (transmitted x-rays) and reach the detector. These transmitted x-rays form the radiographic image on the film/detector. Q.2 Why intensifying screens are used in radiographic examinations? Ans. Intensifying Screens are used to reduce the radiation dose to the patient in diagnostic radiography. The main function of the Intensifying screen is to convert x-rays falling on it into light. Q.3 What is the function of image intensifier? Ans. The function of the x-ray Image Intensifier is to convert an x- ray image into a minified light image.
  • 45. Q.4 What is fluoroscopy? Ans. Fluoroscopy refers to the continuous acquisition of a sequence of x-ray images over time, essentially a real-time x-ray movie of the patient. It is called dynamic imaging. Q.5 How occupational exposure can be minimized in interventional radiological procedures? Ans. Using protective lead aprons Avoiding direct beam Maintaining distance Limiting the total beam–on time Avoiding oblique lateral projections Q.6 Which type of x-ray radiation is used in mammography examinations? Ans. Characteristic x-ray radiation is used in mammography examinations. Q.7 What is the limit of tube housing leakage for dental (intra oral) x-ray equipment? Ans. Leakage from tube housing shall be leas than 0.25 mGy in one hour at 1m from focus of dental (intra-oral) equipment. Expected Questions
  • 46. Q.8 How Dental (OPG) scans of patient is performed? Ans. In Dental (OPG) scans, the X-ray tube rotates around the patient head in one direction and the film rotates in the opposite direction, while the patient sits (or stands) in a stationary position. Q.9 What is the advantage of tomographic image over projection image? Ans. The advantage of a tomographic image over projection image is its ability to display the anatomy in a slab (slice) of tissue in the absence of over or underlying structures. It improves its image contrast Q.10 What do you mean about the contrast of the image? Ans. Contrast in an image is the difference in the gray scale of the image. It is the difference in darkness of the image and its surroundings. Expected Questions
  • 47. References and sources for additional information • The Essential Physics of Medical Imaging (J. T. Bushberg, J.A. Seibert, E.M. Leidholdt, J M Boone) • The Physics of Radiology (H.E. Johns, J.R. Cunnighnam) • IAEA Presentations on Diagnostic Radiology
  • 48. List of presentations in the training Module Basics of Diagnostic X-ray Equipment Biological effects of Radiations Medical X-ray imaging techniques Planning of Diagnostic X-ray facilities Quality Assurance of X-ray equipment Quality Assurance of Computed Tomography equipment Radiation Protection in Diagnostic Radiology Practice Causes, prevention and investigation of excessive exposures in diagnostic radiology Regulatory Requirements for Diagnostic Radiology Practice