Study of related factors about positive biological monitoring of steam
sterilization and emergent treatment.
Huang Hao, Zhou Xiao Li*, Zeng Ai Ying, He Jia, Zhang Pu Yue, Ruan Hong Mei, Qin Nian
Central Sterile Supply Center of West China Hospital at Sichuan University, No 37, Guoxue Road, Chengdu City,
Sichuan Province, PR China
Abstract
Objective: This study aimed to gain relevant experience from an emergent treatment about steam
sterilization rapid biological monitoring (BI).
Methods: To analyse the positive result of BI of steam sterilization and take effective measures, including
stopping distributing, recalling all items processed back to the last negative BI, analyzing the physical
and chemical monitoring back to the last negative BI, and repairing sterilizer.
Results: Recall 36 surgical instrument packages, use other sterilization method, and distribute the items
after BI. No surgery was disturbed. The reason that Belimed MST9-6-18HS2 prevacuum steam sterilizer
failed was aging of rubber gasket inside pressure switch, which led to leak. After replacing the sterilizer
pressure switches and accessories, BI and Bowie-Dick tests were passed for three times and sterilizer
was reused.
Conclusion: Except for the conventional maintenance, the sterilizer needed regular and deep inspection
and timely replacement of wearing or related accessories. Effective emergent treatment should be taken
to control the medical risks produced by positive BI.
Keywords: Biological monitoring, Positive reports, Emergent treatment, Related research.
Accepted on October 24, 2017
Introduction
Biological monitoring (BI) is the gold standard using non-
pathogenic bacteria spores to determine and evaluate the effect
of steam sterilization [1]. The use of spores based standard BI
has been accepted practice for establishing that a sterilization
process has been effective in sterilizing a given batch of
surgical instruments in acute care facilities and medical device
manufacturing for over 30 years [2]. If these spores are
destroyed, it is assumed that any contaminating organisms in
the load have also been killed, as these organisms have lower
resistance than the spores, and are present in lower numbers
[3]. The traditional biological indicators were inoculated paper
strips inside envelopes, which were transferred to sterile
culture medium following processing, and incubated for 7 d
[4]. Recent years, rapid BI is widely used to get the result
quickly and accurately, which shorten the detection time
greatly [5]. From Zhang et al. they have compared the accuracy
of the sterilization BI effect of three kinds of pressure steam
sterilization integrating indicators by real surveillance [6].
Another study also reported the monitor of sterilization effect
of different kinds of chemical indicator for steam sterilization
[7]. All the surgical instruments will not be distributed until BI
was negative, which makes the sterilization process safe. In the
Central Sterile Supply Department of West China Hospital at
Sichuan University, one positive case of BI in steam
sterilization was reported in January 1, 2014. It was timely
analysed and effective measures were taken to make sure the
safe use.
Materials and Methods
Materials
Belimed MST9-6-18HS2 prevacuum steam sterilizer was used.
The sterilization temperature was set to 134°C, sterilization
time was set to 6 min. 3 M 41382 integrating Process
Challenge Device (PCD) containing 3 M 1292 biological
indicator and class 5 integrating indicator was placed in the
area of the sterilizer chamber when the sterilization cycle
began. The PCD was placed on the bottom section of the
sterilizer drain near the gate. The item name, the end user of
the items and the sterilizer number was written on record sheet
outside the PCD. The BI was performed by 3 M 290 auto
reader after physical monitoring and chemical monitoring of
the sterilization cycle was passed.
Biological monitoring operating
The 3 M 41382 integrated PCD was used in accordance with
introduction for users. Each information of biological indicator
in the reading hole, such as pot number and type, was recorded.
Positive control from same lot number as processed biological
indicators at the same day was put in incubation well for
ISSN 0970-938X
www.biomedres.info
Biomed Res 2018 Volume 29 Issue 1 142
Biomedical Research 2018; 29 (1): 142-145
incubating in 1 h. The BI result was obtained after 3 h. The
yellow light indicating the incubation was ongoing. The green
light indicating the negative result was detected, and the red
light indicating the positive result was detected (Figure 1). No
bacteria will grow if the media in the processed BI tube
remained purple, which provided negative result. The culture
media in positive control tube turned yellow and cloudy that
indicated a growth of bacteria and positive. Processed BI
negative result with positive control indicated acceptable
sterilization process (Figure 2) [8].
Figure 1. Processed positive results.
Figure 2. Processed BI negative results.
Positive results: In the morning, the nurse reported a positive
case. Two red (+) symbol was displayed on the LCD panel of
auto reader. One biological indicator was processed, another
was positive control. The other lights displayed green-symbol,
which indicated negative result (Figure 3). The culture media
in the processed indicator and positive control were yellow and
cloudy, which illustrated the growth of bacteria (Figure 4). The
media color of rest processed indicators were purple, which
meant no bacterial growth. The color of chemical indicator
outside the tube was brown, which meant the BIs were exposed
to the sterilization process. The color of chemical indicator
outside the tube was pink, which meant the BIs were not
exposed to the sterilization process. In conclusion, positive BI
case was reported.
Figure 3. Green-symbol indicating negative results.
Figure 4. Illustrating the growth of bacteria.
Emergency measures-stop distributing
All processed items distributing were stopped immediately.
Nurse on duty reported to supervisor followed recall
procedure. This batch items were reprocessed by other
sterilizer or sterilization method in order to ensure the supply
of surgical instruments and clinical items.
Recall and reprocess
According to “Central sterile supply department (CSSD) Part
III: surveillance standard for cleaning, disinfection and
sterilization”, all the unused sterile items should be recalled as
soon as possible back to the last negative BI. The items should
be reprocessed and the causes should be analysed for
improvement. The sterilizer can be reused after three
consecutive BI [9]. According to this protocol, the instruments
were recalled and reprocessed. Other autoclave sterilizers were
used to process the items.
Hao/Li/Ying/Jia/Yue/Mei/Nian
143 Biomed Res 2018 Volume 29 Issue 1
Suspend of sterilize
The number of the dysfunctional sterilizer was recorded and
suspended. The logo read "fault, suspended".
Rapid biological monitoring equipment performance
detecting
The biological indicator processed by functional sterilizer was
put in the well with positive control. No positive alarm after
three hours as normal operation. Thus, the cause of BI
equipment malfunction was excluded.
Sterilizer leak test
The engineer arrived within two hours. Empty chamber leak
test of sterilizer was conducted after analysis. It was found that
the leak was for the error of pressure switch, not the error of
sterilizer door seal after leak test.
Maintenance of sterilizer
Engineers replaced the pressure switch of MST9-6-18HS2
prevacuum steam sterilizer. According to “Central sterile
supply department (CSSD) Part III: surveillance standard for
cleaning, disinfection and sterilization” [9], three consecutive
Bowie-Dick tests were run after sterilizer maintenance service.
Then physical monitoring, chemical monitoring and BI were
completed with three consecutive sterilization programs. In the
end, the hollow PCD tests were run for three times to make
sterilizer properly qualified before put into use. The details and
data of incident were saved for future reference according to
relevant provisions.
Results
In the steam sterilization positive BI report, 36 surgical
instrument packages were recalled wherein one implant
package, which were all repackaged to sterile by using other
steam sterilizers. Surgical instruments were delayed 1.5 h to
use. Implants were distributed after BI passed in 4.5 h. No
surgery and clinical use were affected. The rubber gasket of the
pressure switch of MST9-6-18HS2 prevacuum steam sterilizer
was aging to leak. The leak rate is greater than 3 mpa/min.
Sterilizer was normally enabled to use after replacement of
pressure switch and success of leak test with no leak.
Discussion
Principle of biological monitoring
Steam sterilization was widely used in hospital because of
good effect, short time and low cost. Sterilization process
conditions should be evaluated by physical, biological, and
chemical monitoring. Physical monitoring reflected if the
sterilization cycle was complete. Chemical monitoring had
many impact factors. Therefore, BI was the most accurate and
reliable method in quality control of steam sterilization [10].
Biological indicators monitored high-pressure steam
sterilization under 132°C, 4min condition with resistant
bacterial spores to make Sterility Assurance Level (SAL)
expressed as 10-6. Biological indicator was applied to ensure
the entire sterilization process can achieve at the level of SAL.
It reflected a variety of factors led to sterilization failure
comprehensively. 3 M rapid biological indicator contained a
standardized amount of Bacillus stearothermophilus spores.
The culture media of biological indicator involved water, pH
indicator, nutritious substances and fluorescent substrate.
Rapid BI provided a strong guarantee for the safety of surgical
instruments [11].
Analysis of biological monitoring
Positive result of rapid BI could not prove sterilization failure
[12]. Some factors also led to false positive results, such as no
proper six-step hand-washing before the operating, chemical
indicator outside the self-contained BI touched by fingers,
indicator cap untimely pressed and other improper operation of
biological indicators. All operations may be polluted that could
provide false positive results. In this case, the validity of
biological indicators, the production lot to ensure the
comparability between processed and unprocessed BI (Figure
2), incubation temperature within the range of 56 ± 1°C, the
cleanliness of the reading hole, the pollution in the incubation
well, fail working condition caused by the unrepaired lamps
which may cause the temperature detector error, foreign
substance barrier leading to interpretation errors, etc. In this
case, operations were followed the recommendation, biological
indicators were effective and the detector had no faults after
investigating. Above all, concluded no false positive results.
Analysis of sterilizer
It should be checked that the class 5 integrating indicator in
integrated PCD, the chemical indicators inside and outside the
package of sterilization items. The record of chemical
monitoring especially Bowie-Dick test in the past week should
be checked to find the reason for sterilizer malfunction. No
Bowie-Dick test failure and chemical indicators meant
acceptable sterilization process. The physical monitoring
records should be checked to ensure the consistency between
sterilization cycle program and the sterile items, the parameters
of sterilization program and the alarm reminder. After that the
physical monitoring should be compared with other normal
qualified physical monitoring records. The evacuation times,
evacuation duration, final pressure, sterilization temperature
and sterilization duration should be checked in order to look
for physical problems. The manufacturer engineer was
contacted to discover sterilizer hardware problems by a leak
test. Maximum leak rate of Belimed MST9-6-18HS2
prevacuum steam sterilizer was less than 3 mpa/min. The
detection value was greater than 3 mpa/min proved the
sterilizer leakage. Engineers found that pressure switch
problem caused sterilizer leakage in an exact testing method.
The leakage existed but the Bowie-Dick test still passed,
because the leak rate was less than predetermined value of
sterilizer. After Bowie-Dick test, the rubber gasket of pressure
switch inside prevacuum steam sterilizer aged and leaked
Study of related factors about positive biological monitoring of steam sterilization and emergent treatment
Biomed Res 2018 Volume 29 Issue 1 144
worse until leakage rate was greater than 3 mpa/min, which
resulted in the sterilization failure. Some experience was
gained through this case that steam sterilizer should be
regularly checked by special and deep inspection. The timely
replacement of accessories was necessary to ensure the
sterilization quality.
Conclusion
The low probability of positive BI and the less experienced
nurses to deal with the positive results of rapid BI led to
incorrect or untimely treatment, which brought out unsafety of
processed items. Except for the daily maintenance of steam
sterilizer, there should be regular special and deep inspection,
and the timely replacement of accessories. When the positive
BI occurred, the surveillance people should pay great attention
to report according to process. Effective measures should be
taken to control medical risks and ensure medical safety as
positive BI appeared.
References
1. Qin G, Li L. Impact factors and effection of biological
monitoring in steam sterilization. Chinese J Disinfect 2012;
29: 621.
2. Gordon PE. P389: A rapid biological indicator for sterility
assurance. Antimicrob Resist Infect Contr 2013; 2: 389.
3. Pflug IJ, Odlaug TE. Biological indicators in the
pharmaceutical and the medical device industry. J Parent
Sci Technol Publ Parent Drug Assoc 1986; 40: 242.
4. Belkin NL. Comparison of a rapid readout biological
indicator for steam sterilization with four conventional
biological indicators and five chemical indicators. Infect
Off J Soc Hosp Epidemiolog Am 1996; 17: 773.
5. Chen X. Defect analysis of rapid biological monitoring. J
Med Theor Pract 2012; 25: 1905-1906.
6. Zhang J, Xin-Wu LI, Zhang W. Study on the monitor of
sterilization effect of differene kinds of chemical indicator
for pressure steam sterilization. Chinese J Disinfect 2015;
9: 233.
7. Belkin NL. Comparison of a rapid readout biological
indicator for steam sterilization with four conventional
biological indicators and five chemical indicators. Infect
Contr Off J Soc Hosp Epidemiolog Am 1996; 17: 773.
8. Yu L, Fu C. Analysis of positive results in steam
sterilization biological monitoring. Chinese J Disinfect
2009; 26: 439.
9. Chinese Ministry of Health. Central sterile supply
department (CSSD) Part III: surveillance standard for
cleaning, disinfection and sterilization. Chinese Stand
Books 2009.
10. Shen J, Zhang L. Progress of steam sterilizers. Chinese J
Disinfect 2007; 24: 271.
11. Xiao Q. Fitting of chemical monitoring and biological
monitoring in steam sterilization. Chinese J Nosocomiol
2012; 22: 141.
12. Chen J, Tie J, Wang C. Research of rapid biological
indicators culture method consistency in steam sterilization.
Chinese Prev Med 2013; 14: 831.
*Correspondence to
Zhou Xiao Li
Central Sterile Supply Center of West China Hospital at
Sichuan University
Sichuan Province
PR China
Hao/Li/Ying/Jia/Yue/Mei/Nian
145 Biomed Res 2018 Volume 29 Issue 1

More Related Content

What's hot

377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY
GOVIND YuvRaju
 
Ipqc for parenterals
Ipqc for parenteralsIpqc for parenterals
Ipqc for parenterals
ceutics1315
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
RavichandraNadagouda
 

What's hot (20)

Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Validation of sterilization methods
Validation of sterilization methodsValidation of sterilization methods
Validation of sterilization methods
 
Cleaning validation presentation
Cleaning validation presentationCleaning validation presentation
Cleaning validation presentation
 
Sterilization Validation for Medical Devices
Sterilization Validation for Medical DevicesSterilization Validation for Medical Devices
Sterilization Validation for Medical Devices
 
Bioburden Validation Strategy for Cleaning Validation
Bioburden Validation Strategy for Cleaning ValidationBioburden Validation Strategy for Cleaning Validation
Bioburden Validation Strategy for Cleaning Validation
 
Managing risk in cleaning validation
Managing risk in cleaning validationManaging risk in cleaning validation
Managing risk in cleaning validation
 
377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY377218-GMP-TRAINING-STERILE-FACILITY
377218-GMP-TRAINING-STERILE-FACILITY
 
Filter validation
Filter validationFilter validation
Filter validation
 
Bioburden method validation
Bioburden method validationBioburden method validation
Bioburden method validation
 
Ipqc for parenterals
Ipqc for parenteralsIpqc for parenterals
Ipqc for parenterals
 
Risk Management in Sterile Environments
Risk Management in Sterile Environments Risk Management in Sterile Environments
Risk Management in Sterile Environments
 
Basic Concepts of Cleaning validation
Basic Concepts of Cleaning validationBasic Concepts of Cleaning validation
Basic Concepts of Cleaning validation
 
Lifecycle Approach to Cleaning Validation
Lifecycle Approach to Cleaning ValidationLifecycle Approach to Cleaning Validation
Lifecycle Approach to Cleaning Validation
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
1. cleaning validation basic concept
1. cleaning validation basic concept1. cleaning validation basic concept
1. cleaning validation basic concept
 
Microbiological Environmental Monitoring in Pharmaceutical Facility
Microbiological Environmental Monitoring in Pharmaceutical FacilityMicrobiological Environmental Monitoring in Pharmaceutical Facility
Microbiological Environmental Monitoring in Pharmaceutical Facility
 
Aseptic processing operation
Aseptic processing operationAseptic processing operation
Aseptic processing operation
 
Good containment practices
Good containment practicesGood containment practices
Good containment practices
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 

Similar to Study of-related-factors-about-positive-biological-monitoring-of-steam-sterilization-and-emergent-treatment

Similar to Study of-related-factors-about-positive-biological-monitoring-of-steam-sterilization-and-emergent-treatment (20)

SporeCheck
SporeCheckSporeCheck
SporeCheck
 
Unit 2 ppt new.pptx
Unit 2 ppt new.pptxUnit 2 ppt new.pptx
Unit 2 ppt new.pptx
 
Quality improvement and cost reduction (project by project appoach)
Quality improvement and cost reduction (project by project appoach)Quality improvement and cost reduction (project by project appoach)
Quality improvement and cost reduction (project by project appoach)
 
Good laboratory practises
Good laboratory practisesGood laboratory practises
Good laboratory practises
 
Biological indicators for /certified fixed orthodontic courses by Indian dent...
Biological indicators for /certified fixed orthodontic courses by Indian dent...Biological indicators for /certified fixed orthodontic courses by Indian dent...
Biological indicators for /certified fixed orthodontic courses by Indian dent...
 
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationEU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
 
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter ValidationEU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
EU GMP Annex 1 Draft: Implications on Sterilizing Grade Filter Validation
 
Evaluation of the efficiency of sterilization methods. Sterility indicators
Evaluation of the efficiency of sterilization methods.Sterility indicatorsEvaluation of the efficiency of sterilization methods.Sterility indicators
Evaluation of the efficiency of sterilization methods. Sterility indicators
 
Bioburden
Bioburden  Bioburden
Bioburden
 
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
 
Cleaning validation-volume-iii
Cleaning validation-volume-iiiCleaning validation-volume-iii
Cleaning validation-volume-iii
 
BIJGIM-20232112.pdf
BIJGIM-20232112.pdfBIJGIM-20232112.pdf
BIJGIM-20232112.pdf
 
cleaning validation
cleaning validationcleaning validation
cleaning validation
 
Change control oos oot
Change control oos ootChange control oos oot
Change control oos oot
 
cleaning validation_٠٧١٥٣٠.pptx
cleaning validation_٠٧١٥٣٠.pptxcleaning validation_٠٧١٥٣٠.pptx
cleaning validation_٠٧١٥٣٠.pptx
 
Usp 1116
Usp 1116Usp 1116
Usp 1116
 
D7 Laundry Sanitizer and Disinfectant Test
D7 Laundry Sanitizer and Disinfectant Test D7 Laundry Sanitizer and Disinfectant Test
D7 Laundry Sanitizer and Disinfectant Test
 
Cleaning Validation in Pharma Industry.pdf
Cleaning Validation in Pharma Industry.pdfCleaning Validation in Pharma Industry.pdf
Cleaning Validation in Pharma Industry.pdf
 
Key factors for developing a pharmaceutical cleaning strategy - Part 2
Key factors for developing a pharmaceutical cleaning strategy - Part 2Key factors for developing a pharmaceutical cleaning strategy - Part 2
Key factors for developing a pharmaceutical cleaning strategy - Part 2
 
An Over View of Sterile Filtration Validation: A Key Elements for Sterile Dru...
An Over View of Sterile Filtration Validation: A Key Elements for Sterile Dru...An Over View of Sterile Filtration Validation: A Key Elements for Sterile Dru...
An Over View of Sterile Filtration Validation: A Key Elements for Sterile Dru...
 

More from eko_apt (14)

2. Sosialisasi untuk Industri Pengolahan Daging.pdf
2. Sosialisasi untuk Industri Pengolahan Daging.pdf2. Sosialisasi untuk Industri Pengolahan Daging.pdf
2. Sosialisasi untuk Industri Pengolahan Daging.pdf
 
Pedoman dasar-teknik-aseptis
Pedoman dasar-teknik-aseptisPedoman dasar-teknik-aseptis
Pedoman dasar-teknik-aseptis
 
Manajemen risiko ristekdikti
Manajemen risiko ristekdiktiManajemen risiko ristekdikti
Manajemen risiko ristekdikti
 
M df-01-manual-risk-management-rev.-00-copy-controlled
M df-01-manual-risk-management-rev.-00-copy-controlledM df-01-manual-risk-management-rev.-00-copy-controlled
M df-01-manual-risk-management-rev.-00-copy-controlled
 
Pedoman jr
Pedoman jrPedoman jr
Pedoman jr
 
Bahan workshop mr
Bahan workshop mrBahan workshop mr
Bahan workshop mr
 
M df-01-manual-risk-management-rev.-00-copy-controlled
M df-01-manual-risk-management-rev.-00-copy-controlledM df-01-manual-risk-management-rev.-00-copy-controlled
M df-01-manual-risk-management-rev.-00-copy-controlled
 
Bahan workshop mr
Bahan workshop mrBahan workshop mr
Bahan workshop mr
 
Guide to risk management 2
Guide to risk management 2Guide to risk management 2
Guide to risk management 2
 
Usability specification-document-template
Usability specification-document-templateUsability specification-document-template
Usability specification-document-template
 
Usability engineering-file
Usability engineering-fileUsability engineering-file
Usability engineering-file
 
Rcdso 4884 potential causes of a positive bi v.2
Rcdso 4884 potential causes of a positive bi v.2Rcdso 4884 potential causes of a positive bi v.2
Rcdso 4884 potential causes of a positive bi v.2
 
Digital 20361525 pr-agatha dwi setiastuti-pt molex
Digital 20361525 pr-agatha dwi setiastuti-pt molexDigital 20361525 pr-agatha dwi setiastuti-pt molex
Digital 20361525 pr-agatha dwi setiastuti-pt molex
 
4113230009 bab i
4113230009 bab i4113230009 bab i
4113230009 bab i
 

Recently uploaded

Physiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdfPhysiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdf
MedicoseAcademics
 
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Halo Docter
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan 087776558899
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
MedicoseAcademics
 

Recently uploaded (20)

Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024
 
TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...
TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...
TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...
 
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
 
Physiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdfPhysiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdf
 
SEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdf
SEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdfSEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdf
SEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdf
 
Test bank for critical care nursing a holistic approach 11th edition morton f...
Test bank for critical care nursing a holistic approach 11th edition morton f...Test bank for critical care nursing a holistic approach 11th edition morton f...
Test bank for critical care nursing a holistic approach 11th edition morton f...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
 
VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Physicochemical properties (descriptors) in QSAR.pdf
Physicochemical properties (descriptors) in QSAR.pdfPhysicochemical properties (descriptors) in QSAR.pdf
Physicochemical properties (descriptors) in QSAR.pdf
 
Intro to disinformation and public health
Intro to disinformation and public healthIntro to disinformation and public health
Intro to disinformation and public health
 
VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...
 
spinal cord disorders and paraplegia .
spinal cord disorders  and  paraplegia .spinal cord disorders  and  paraplegia .
spinal cord disorders and paraplegia .
 
ABO Blood grouping in-compatibility in pregnancy
ABO Blood grouping in-compatibility in pregnancyABO Blood grouping in-compatibility in pregnancy
ABO Blood grouping in-compatibility in pregnancy
 
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptxCreeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
 
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happenedPart I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
 
Top 10 Most Beautiful Chinese Pornstars List 2024
Top 10 Most Beautiful Chinese Pornstars List 2024Top 10 Most Beautiful Chinese Pornstars List 2024
Top 10 Most Beautiful Chinese Pornstars List 2024
 
Face and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptxFace and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptx
 

Study of-related-factors-about-positive-biological-monitoring-of-steam-sterilization-and-emergent-treatment

  • 1. Study of related factors about positive biological monitoring of steam sterilization and emergent treatment. Huang Hao, Zhou Xiao Li*, Zeng Ai Ying, He Jia, Zhang Pu Yue, Ruan Hong Mei, Qin Nian Central Sterile Supply Center of West China Hospital at Sichuan University, No 37, Guoxue Road, Chengdu City, Sichuan Province, PR China Abstract Objective: This study aimed to gain relevant experience from an emergent treatment about steam sterilization rapid biological monitoring (BI). Methods: To analyse the positive result of BI of steam sterilization and take effective measures, including stopping distributing, recalling all items processed back to the last negative BI, analyzing the physical and chemical monitoring back to the last negative BI, and repairing sterilizer. Results: Recall 36 surgical instrument packages, use other sterilization method, and distribute the items after BI. No surgery was disturbed. The reason that Belimed MST9-6-18HS2 prevacuum steam sterilizer failed was aging of rubber gasket inside pressure switch, which led to leak. After replacing the sterilizer pressure switches and accessories, BI and Bowie-Dick tests were passed for three times and sterilizer was reused. Conclusion: Except for the conventional maintenance, the sterilizer needed regular and deep inspection and timely replacement of wearing or related accessories. Effective emergent treatment should be taken to control the medical risks produced by positive BI. Keywords: Biological monitoring, Positive reports, Emergent treatment, Related research. Accepted on October 24, 2017 Introduction Biological monitoring (BI) is the gold standard using non- pathogenic bacteria spores to determine and evaluate the effect of steam sterilization [1]. The use of spores based standard BI has been accepted practice for establishing that a sterilization process has been effective in sterilizing a given batch of surgical instruments in acute care facilities and medical device manufacturing for over 30 years [2]. If these spores are destroyed, it is assumed that any contaminating organisms in the load have also been killed, as these organisms have lower resistance than the spores, and are present in lower numbers [3]. The traditional biological indicators were inoculated paper strips inside envelopes, which were transferred to sterile culture medium following processing, and incubated for 7 d [4]. Recent years, rapid BI is widely used to get the result quickly and accurately, which shorten the detection time greatly [5]. From Zhang et al. they have compared the accuracy of the sterilization BI effect of three kinds of pressure steam sterilization integrating indicators by real surveillance [6]. Another study also reported the monitor of sterilization effect of different kinds of chemical indicator for steam sterilization [7]. All the surgical instruments will not be distributed until BI was negative, which makes the sterilization process safe. In the Central Sterile Supply Department of West China Hospital at Sichuan University, one positive case of BI in steam sterilization was reported in January 1, 2014. It was timely analysed and effective measures were taken to make sure the safe use. Materials and Methods Materials Belimed MST9-6-18HS2 prevacuum steam sterilizer was used. The sterilization temperature was set to 134°C, sterilization time was set to 6 min. 3 M 41382 integrating Process Challenge Device (PCD) containing 3 M 1292 biological indicator and class 5 integrating indicator was placed in the area of the sterilizer chamber when the sterilization cycle began. The PCD was placed on the bottom section of the sterilizer drain near the gate. The item name, the end user of the items and the sterilizer number was written on record sheet outside the PCD. The BI was performed by 3 M 290 auto reader after physical monitoring and chemical monitoring of the sterilization cycle was passed. Biological monitoring operating The 3 M 41382 integrated PCD was used in accordance with introduction for users. Each information of biological indicator in the reading hole, such as pot number and type, was recorded. Positive control from same lot number as processed biological indicators at the same day was put in incubation well for ISSN 0970-938X www.biomedres.info Biomed Res 2018 Volume 29 Issue 1 142 Biomedical Research 2018; 29 (1): 142-145
  • 2. incubating in 1 h. The BI result was obtained after 3 h. The yellow light indicating the incubation was ongoing. The green light indicating the negative result was detected, and the red light indicating the positive result was detected (Figure 1). No bacteria will grow if the media in the processed BI tube remained purple, which provided negative result. The culture media in positive control tube turned yellow and cloudy that indicated a growth of bacteria and positive. Processed BI negative result with positive control indicated acceptable sterilization process (Figure 2) [8]. Figure 1. Processed positive results. Figure 2. Processed BI negative results. Positive results: In the morning, the nurse reported a positive case. Two red (+) symbol was displayed on the LCD panel of auto reader. One biological indicator was processed, another was positive control. The other lights displayed green-symbol, which indicated negative result (Figure 3). The culture media in the processed indicator and positive control were yellow and cloudy, which illustrated the growth of bacteria (Figure 4). The media color of rest processed indicators were purple, which meant no bacterial growth. The color of chemical indicator outside the tube was brown, which meant the BIs were exposed to the sterilization process. The color of chemical indicator outside the tube was pink, which meant the BIs were not exposed to the sterilization process. In conclusion, positive BI case was reported. Figure 3. Green-symbol indicating negative results. Figure 4. Illustrating the growth of bacteria. Emergency measures-stop distributing All processed items distributing were stopped immediately. Nurse on duty reported to supervisor followed recall procedure. This batch items were reprocessed by other sterilizer or sterilization method in order to ensure the supply of surgical instruments and clinical items. Recall and reprocess According to “Central sterile supply department (CSSD) Part III: surveillance standard for cleaning, disinfection and sterilization”, all the unused sterile items should be recalled as soon as possible back to the last negative BI. The items should be reprocessed and the causes should be analysed for improvement. The sterilizer can be reused after three consecutive BI [9]. According to this protocol, the instruments were recalled and reprocessed. Other autoclave sterilizers were used to process the items. Hao/Li/Ying/Jia/Yue/Mei/Nian 143 Biomed Res 2018 Volume 29 Issue 1
  • 3. Suspend of sterilize The number of the dysfunctional sterilizer was recorded and suspended. The logo read "fault, suspended". Rapid biological monitoring equipment performance detecting The biological indicator processed by functional sterilizer was put in the well with positive control. No positive alarm after three hours as normal operation. Thus, the cause of BI equipment malfunction was excluded. Sterilizer leak test The engineer arrived within two hours. Empty chamber leak test of sterilizer was conducted after analysis. It was found that the leak was for the error of pressure switch, not the error of sterilizer door seal after leak test. Maintenance of sterilizer Engineers replaced the pressure switch of MST9-6-18HS2 prevacuum steam sterilizer. According to “Central sterile supply department (CSSD) Part III: surveillance standard for cleaning, disinfection and sterilization” [9], three consecutive Bowie-Dick tests were run after sterilizer maintenance service. Then physical monitoring, chemical monitoring and BI were completed with three consecutive sterilization programs. In the end, the hollow PCD tests were run for three times to make sterilizer properly qualified before put into use. The details and data of incident were saved for future reference according to relevant provisions. Results In the steam sterilization positive BI report, 36 surgical instrument packages were recalled wherein one implant package, which were all repackaged to sterile by using other steam sterilizers. Surgical instruments were delayed 1.5 h to use. Implants were distributed after BI passed in 4.5 h. No surgery and clinical use were affected. The rubber gasket of the pressure switch of MST9-6-18HS2 prevacuum steam sterilizer was aging to leak. The leak rate is greater than 3 mpa/min. Sterilizer was normally enabled to use after replacement of pressure switch and success of leak test with no leak. Discussion Principle of biological monitoring Steam sterilization was widely used in hospital because of good effect, short time and low cost. Sterilization process conditions should be evaluated by physical, biological, and chemical monitoring. Physical monitoring reflected if the sterilization cycle was complete. Chemical monitoring had many impact factors. Therefore, BI was the most accurate and reliable method in quality control of steam sterilization [10]. Biological indicators monitored high-pressure steam sterilization under 132°C, 4min condition with resistant bacterial spores to make Sterility Assurance Level (SAL) expressed as 10-6. Biological indicator was applied to ensure the entire sterilization process can achieve at the level of SAL. It reflected a variety of factors led to sterilization failure comprehensively. 3 M rapid biological indicator contained a standardized amount of Bacillus stearothermophilus spores. The culture media of biological indicator involved water, pH indicator, nutritious substances and fluorescent substrate. Rapid BI provided a strong guarantee for the safety of surgical instruments [11]. Analysis of biological monitoring Positive result of rapid BI could not prove sterilization failure [12]. Some factors also led to false positive results, such as no proper six-step hand-washing before the operating, chemical indicator outside the self-contained BI touched by fingers, indicator cap untimely pressed and other improper operation of biological indicators. All operations may be polluted that could provide false positive results. In this case, the validity of biological indicators, the production lot to ensure the comparability between processed and unprocessed BI (Figure 2), incubation temperature within the range of 56 ± 1°C, the cleanliness of the reading hole, the pollution in the incubation well, fail working condition caused by the unrepaired lamps which may cause the temperature detector error, foreign substance barrier leading to interpretation errors, etc. In this case, operations were followed the recommendation, biological indicators were effective and the detector had no faults after investigating. Above all, concluded no false positive results. Analysis of sterilizer It should be checked that the class 5 integrating indicator in integrated PCD, the chemical indicators inside and outside the package of sterilization items. The record of chemical monitoring especially Bowie-Dick test in the past week should be checked to find the reason for sterilizer malfunction. No Bowie-Dick test failure and chemical indicators meant acceptable sterilization process. The physical monitoring records should be checked to ensure the consistency between sterilization cycle program and the sterile items, the parameters of sterilization program and the alarm reminder. After that the physical monitoring should be compared with other normal qualified physical monitoring records. The evacuation times, evacuation duration, final pressure, sterilization temperature and sterilization duration should be checked in order to look for physical problems. The manufacturer engineer was contacted to discover sterilizer hardware problems by a leak test. Maximum leak rate of Belimed MST9-6-18HS2 prevacuum steam sterilizer was less than 3 mpa/min. The detection value was greater than 3 mpa/min proved the sterilizer leakage. Engineers found that pressure switch problem caused sterilizer leakage in an exact testing method. The leakage existed but the Bowie-Dick test still passed, because the leak rate was less than predetermined value of sterilizer. After Bowie-Dick test, the rubber gasket of pressure switch inside prevacuum steam sterilizer aged and leaked Study of related factors about positive biological monitoring of steam sterilization and emergent treatment Biomed Res 2018 Volume 29 Issue 1 144
  • 4. worse until leakage rate was greater than 3 mpa/min, which resulted in the sterilization failure. Some experience was gained through this case that steam sterilizer should be regularly checked by special and deep inspection. The timely replacement of accessories was necessary to ensure the sterilization quality. Conclusion The low probability of positive BI and the less experienced nurses to deal with the positive results of rapid BI led to incorrect or untimely treatment, which brought out unsafety of processed items. Except for the daily maintenance of steam sterilizer, there should be regular special and deep inspection, and the timely replacement of accessories. When the positive BI occurred, the surveillance people should pay great attention to report according to process. Effective measures should be taken to control medical risks and ensure medical safety as positive BI appeared. References 1. Qin G, Li L. Impact factors and effection of biological monitoring in steam sterilization. Chinese J Disinfect 2012; 29: 621. 2. Gordon PE. P389: A rapid biological indicator for sterility assurance. Antimicrob Resist Infect Contr 2013; 2: 389. 3. Pflug IJ, Odlaug TE. Biological indicators in the pharmaceutical and the medical device industry. J Parent Sci Technol Publ Parent Drug Assoc 1986; 40: 242. 4. Belkin NL. Comparison of a rapid readout biological indicator for steam sterilization with four conventional biological indicators and five chemical indicators. Infect Off J Soc Hosp Epidemiolog Am 1996; 17: 773. 5. Chen X. Defect analysis of rapid biological monitoring. J Med Theor Pract 2012; 25: 1905-1906. 6. Zhang J, Xin-Wu LI, Zhang W. Study on the monitor of sterilization effect of differene kinds of chemical indicator for pressure steam sterilization. Chinese J Disinfect 2015; 9: 233. 7. Belkin NL. Comparison of a rapid readout biological indicator for steam sterilization with four conventional biological indicators and five chemical indicators. Infect Contr Off J Soc Hosp Epidemiolog Am 1996; 17: 773. 8. Yu L, Fu C. Analysis of positive results in steam sterilization biological monitoring. Chinese J Disinfect 2009; 26: 439. 9. Chinese Ministry of Health. Central sterile supply department (CSSD) Part III: surveillance standard for cleaning, disinfection and sterilization. Chinese Stand Books 2009. 10. Shen J, Zhang L. Progress of steam sterilizers. Chinese J Disinfect 2007; 24: 271. 11. Xiao Q. Fitting of chemical monitoring and biological monitoring in steam sterilization. Chinese J Nosocomiol 2012; 22: 141. 12. Chen J, Tie J, Wang C. Research of rapid biological indicators culture method consistency in steam sterilization. Chinese Prev Med 2013; 14: 831. *Correspondence to Zhou Xiao Li Central Sterile Supply Center of West China Hospital at Sichuan University Sichuan Province PR China Hao/Li/Ying/Jia/Yue/Mei/Nian 145 Biomed Res 2018 Volume 29 Issue 1