Steam sterilization monitoring involves ensuring sterilized items are free from microorganisms. The document discusses several key aspects of steam sterilization monitoring including: (1) sterility assurance levels and D-values which define sterilization effectiveness, (2) common sterilization methods like steam, gamma radiation and hydrogen peroxide, (3) methods for monitoring sterilization like mechanical, chemical and biological indicators, and (4) levels of assurance including equipment, exposure, pack and load control through proper use of indicators and record keeping. Managing a positive biological indicator requires recalling implantable items, retesting the sterilizer, and putting it back in service only after three consecutive negative retests.
2. Sterilization / Sterility
Process used to transform product free from viable
microorganisms (Bacteria, Viruses, Fungi) through
inactivation of microorganisms
Introduction
The absence of microorganisms cannot be proven
A sterility assurance level (SAL) is used to define the sterility
Dr. Mohamed Eleraky 28/2/2019
3. Cont. Introduction
Sterility Assurance Level (SAL)
Probability of a
viable organism
being present
on a product unit
after sterilization
SAL of 10-6 is required
by the FDA for
invasive medical
device
Dr. Mohamed Eleraky 28/2/2019
4. Cont. Introduction
Inactivation of microorganisms
D value:
Time required to
achieve inactivation of
90% of a population of
the microorganism
under specific conditions
Dr. Mohamed Eleraky 28/2/2019
6. Cont. Introduction
Moist heat
Gamma radiation
Hydrogen peroxide gas plasma
Ethylene oxide (ETO)
Common Sterilization methods used in health services:
For moist heat sterilization, There are four parameters of steam sterilization:
steam, pressure, temperature, and time.
Dr. Mohamed Eleraky 28/2/2019
7. Sterilization Quality Control
How do I know my item is safe for the patient???
I. Methods used for Sterilization monitoring
II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
8. Mechanical monitoring
Chemical indicators
Biological indicators
Equipment Control
Exposure control
Pack Control
Load Control
Record Keeping
I. Methods used for monitoring
II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
9. Cycle time.
Temperature.
Pressure
Sterilizers have recording devices that print out these parameters.
Correct readings do not ensure sterilization, but incorrect readings can be the first
indicator for a problem with the sterilization cycle.
I. Methods used for monitoring
A.Mechanical monitoring : through assessing
Dr. Mohamed Eleraky 28/2/2019
10. External chemical indicators (class 1):
They are applied to the outside of a package (e.g., chemical indicator tape and labels).
They change color rapidly when a specific parameter is reached, and they verify that the
package has been exposed to the sterilization process.
Internal chemical indicators and integrators (Class 3-6 ):
They are used inside package to ensure the sterilizing agent has penetrated the packaging
material and actually reached the instruments inside.
Bowie-Dick Sheets inside test packs (class 2) :
Sheets printed with a chemical indicator in the form of diagonal strips which change color to
dark brown or black.
A satisfactory test is indicated by the development of uniform color change indicating rapid
steam penetration, adequate air removal and freedom from significant air.
Chemical strip (Class 2) :
Used inside a hollow load process challenge device (helix test) for testing the efficiency of
rapid steam penetration in pre vacuum sterilizers.
cont. I. Methods used for monitoring
B. Chemical indicators :
Dr. Mohamed Eleraky 28/2/2019
11. Class Description Use
Class 1 Process indicators
Used to indicate that the pack or container has been exposed to the sterilization process and to
differentiate between processed and unprocessed items (tape and labels)
Class 2
Indicators for use in
special tests
e.g. BD sheets and strips used inside process challenge device (PCD) to test for air removal in
vacuum assisted sterilizers
Class 3
Single variable
indicators
React to one of the critical parameter of sterilization and indicate exposure to a sterilization
cycle at stated values (SV) of the chosen parameter.
Critical parameters typically chosen for steam sterilization processes are time or temperature.
Class 4
Multi-variable
indicators
React to two or more of the critical variables (time, temperature and saturated steam) to
indicate exposure to a sterilization cycle as SVs of the chosen variables. They provide
additional monitoring of the sterilization cycle
Class 5 Integrators
React to all critical parameters over a specified range of sterilization cycles. Their
performance has been correlated to the performance of a biological indicator (BI) under its
labeled conditions for use. This class of indicator can be used in place of the BI in many
applications, thereby reducing the overall cost of sterile processing.
Class 6 Emulating indicators
They are cycle verification indicators that are designed to react to all critical variables of
specified sterilization cycles. They emulate the cycle. They have the most stringent tolerance
levels of all chemical indicators.
Different chemical indicators
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
12. Before Processing After Processing
External Chemical Indicator Tape
Class I
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
13. Sterilization pack & pouch
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
14. Tamper evident device—container locks, tape, arrows
Internal Pack Control—chemical indicators placed inside the tray, let us know
that several parameters for sterilization have been met
Tamper evident device—container
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
15. Ex. of Internal Chemical Indicators
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
16. Bowie Dick Test—used to assure the air removal system is working properly in
a prevacuum sterilizer Daily
Bowie Dick Test
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
17. Helix test
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
18. They assess the sterilization process directly by killing known highly resistant
microorganisms.
Contain Geobacillus stearothermophillus- Highly heat resistant, spore-forming
bacteria does not produce toxins and is non-pathogenic.
C. Biological indicators : (Spore Tests)
cont. I. Methods used for monitoring
Dr. Mohamed Eleraky 28/2/2019
19. A. Equipment Control
B. Exposure control
C. Pack Control
D. Load Control
E. Record Keeping
II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
20. A. Equipment Control
Testing of new autoclaves
Testing after repair or maintenance
Periodic testing
Mechanical parameters
Tests for air removal in pre vacuum sterilizers
Spore test can be used for equipment control
cont. II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
It includes:
It is done through:
21. B. Exposure control
They are placed on the external surface of (class 1 external chemical
indicator, e.g. tape and labels) or inside (class 3 internal single parameter
chemical indicator) each sterilized pack.
If the indicator is only inside the pack, it should be visible to the user
examining the package. They are used to identify packs and sets that have
been processed in the sterilization process.
cont. II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
22. C. Pack Control
Chemical indicators and integrators are used to verify the penetration of the sterilization process
inside the pack or set.
Usually, two or more parameters are monitored (multiple parameter internal chemical indicators
and integrators).
The indicator or integrator should be placed in the most difficult to penetrate location within the
pack or tray.
Integrators are helpful in identifying packs or trays with too high a density or mass, incorrect
configuration of the load, or beginning failures of sterilization equipment.
They can help the user to differentiate problems between packs or trays and the sterilizer itself.
cont. II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
23. D. Load Control
1. Biological release
The process by which a load is monitored and released based on the result of a biological
indicator in a test pack.
Test the load with appropriate spore test
Incubate test and control vials for the recommended time
Read the result
Use the result for releasing the load as sterile load
2. Parametric Release
Parametric Release is “declaring the product as sterile based on physical or chemical (or
both) process data rather than on the basis of biological indicators results”.
Sterilize a PCD, double wrapped, with every sterilization cycle and use the test result for
batch release (BMS).
Batch Monitoring System (BMS): Class 2 chemical indicators inside process challenge devices
cont. II. Levels of Sterilization Assurance
Dr. Mohamed Eleraky 28/2/2019
24. E. Record Keeping
The staff should document each sterilizer:
• load number, date and time,
• contents of load,
• outcomes of chemical indicators, integrators, BIs & mechanical indicators ,
• operator’s name or initials.
Any equipment maintenance, repairs, and calibration must be documented on each sterilizer.
Autoclave User Log:
Complete User Log with Every Use
Maintain User Logs for 1-2 Years
Autoclave Maintenance Log:
Complete Maintenance Log for Every Repair
Maintain records of Yearly Calibration
Maintain Maintenance Logs for 3 – 5 Years
Monitoring Results Log:
Complete Results Log with Every Test
Maintain Results Logs for 3 – 5 Years
cont. II. Levels of Sterilization Assurance
28/2/2019
25. Take the sterilizer out of work
Notify the infection prevention and control department
All implantable objects of suspect load (s) should be recalled .
Check that the spore test was done properly.
Check with the hospital maintenance to ensure that the autoclave was used correctly and check for any changes in steam supply
A single positive BI does not indicate a sterilizer malfunction if mechanical and chemical indicators are acceptable and appear to be functioning
properly.
Objects, other than implantable ones do not need to be permanent recalled because of a single positive spore test unless the sterilizer or the
sterilization procedure is defective
A second BI should be immediately processed while holding the suspect items in quarantine.
Repeat the spore test in 3 consecutive sterilizer cycles and if the 3 spore test results were negative put the sterilizer back in service
If the repeated tests showed positive results, the items should be considered non sterile and all objects processed since the last negative test should
be recalled and processed again and the sterilizer should be closed till the manufacturer solve the problem and ensure that it works properly and retest
before use in three consecutive cycles.
The followings are suggested to manage
a positive biological indicator in a steam sterilizer:
Dr. Mohamed Eleraky 28/2/2019