Key factors in the choice and
development of a cleaning strategy
Part II
M. L. Bernuzzi
Manager, R&D, Fedegari Autoclavi SpA
S. Ferretti
Ph.D, R&D, Fedegari Autoclavi SpA
This article discusses the removal of bacterial endotoxins in a washing process and the need for a
continuous monitoring activity according to common cleaning validation practices. Results of washing
tests performed with different load materials following an effective cleaning strategy to remove
microbiological soil are presented. In addition, this paper compares two different online assay methods:
conductivity analysis and Total Organic Carbon analysis.
Keywords: Cleaning, Microbiological Soil, Bacterial Endotoxins, Conductivity Analysis, Total Organic
Carbon Analysis.
Introduction
The purpose of a washing treatment consists in
removing the soil present on the load. Some aspects
of soil removal, strictly of a chemical type, were
previously discussed [1]. However, a new cleaning
requirement is becoming increasingly widespread:
treating a different soil, which is not chemical but
microbiological. A very frequent requirement is
indeed to remove bacterial endotoxins.
Once the washing process has been validated, the
chemical and/or biological soil has been removed,
and its absence on the load has been demonstrated,
the process must remain in a continuous
“monitoring mode”. Periodic checks or online
monitoring of the rinse water contribute to:
•	 Increasing the safety margin that makes a
validated washing process, reliable and always
reproducible in its results;
•	 Protecting the end user of the product from
an unexpected result that does not conform to
expectations.
Removal of bacterial
endotoxins
The U.S. FDA Cleaning validation guide states
that “…control of the bioburden through adequate
cleaning and storage of equipment is important to
ensure that subsequent sterilization or sanitization
procedures achieve the necessary assurance of
sterility” [2]. Failure to remove/destroy endotoxins
is certainly a limitation of steam sterilization;
therefore, the “endotoxin content” of a generic
load, intended for subsequent sterilization, is often
monitored and controlled during its handling,
production and washing.
Endotoxin removal can be investigated during the
validation cycles of the cleaning process. In fact,
three endotoxin log reductions, which is a standard
requirement for the depyrogenation process, might
be the additional target of a washing procedure.
Obviously, the required reduction would not be
achieved due to the high temperatures but due
to the mechanical removal effect dictated by the
cleaning medium, such as water, steam or aqueous
solution with the addition of detergents.
DOC. 386143.2
1/7
�
Copyright © 2015 Fedegari Group
fedegari.com
Case study
The test, performed at the Research and
Development department of Fedegari, removed
bacterial endotoxins from different materials by
a washing cycle and constituted a preliminary
assessment of the effectiveness of this process
in terms of endotoxin load reduction. The goal
of this study was to verify the 3 log reduction of
the inoculated endotoxin load on the surface of
coupons measuring 100 mm x 100 mm.
The use of coupons in the initial setup of a washing
process is suggested by PDATechnical Report No.
29: Points to consider for Cleaning Validation [3].
Stainless steal, 316 L, polypropylene and tempered
glass coupons were investigated. The choice
of materials to be inoculated was of a practical
nature: materials usually found in pharmaceutical
practices.
TEST PROCEDURES
The materials were depyrogenated through a
thermal treatment performed at 250°C. Due to the
high temperatures, it was not possible to expose
the polypropylene coupons to the same process.
This way, the plastic coupons were disinfected
with a common biocidal agent representing a
worst case in this study.
Every coupon was inoculated with a predefined
quantity of endotoxins (Endotoxin Indicator
ENDOSAFE, Charles River - Source: Escherichia
coli 055:B5) and the inoculum was dried at 40°C
by means of a laboratory heater. The drying time
depended on the treated material. Appropriate
supports for the coupons were studied to ensure
that the washing medium reached the inoculated
coupon surface (see Figure 1).
In addition, an appropriate cleaning equipment,
composed of rotating arms with nozzles, was
designed and built. Since the effectiveness of
the process in terms of the intended goal was
not known, a standard cycle already applied at a
Figure 1 - Stainless steel coupons placed on
special supports inside the washing machine
customer facility was used. This cycle was mainly
composed by various washing phases with water
recirculation and a washing phase with water
disposal, injection and simultaneous drain of
water.
The washing machine located in the Research and
Development laboratory of Fedegari is supplied
with deionized water originating from a reverse
osmosis plant. In order to remove from the results
any endotoxin content due to the contribution of
the water entering the system, an analysis of the
water was carried out.
The endotoxin content was evaluated by applying
a “Bacterial Endotoxin Test – BET”, as described
in the various international Pharmacopoeias,
harmonized in this regard [4-5]. This test is
capable of identifying or quantifying endotoxins
originating from gram-negative bacteria by using
Limulus Polyphemus amebocyte lysate.
2/7
�
Copyright © 2015 Fedegari Group
fedegari.com
There are three techniques for this type of test:
• gel clot;
• turbidimetric;
• chromogenic.
The European Pharmacopoeia [5] indicates the
following analysis methods:
• Method A. Gel-clot method: limit test
• Method B. Gel-clot method: semi-quantitative
test
• Method C. Turbidimetric kinetic method
• Method D. Chromogenic kinetic method
• Method E. Chromogenic end-point method
• Method F. Turbidimetric end-point method
Thetechniqueappliedwasthechromogenickinetic
one (method D of the European Pharmacopeia).
The instrument used (see Figure 2), Endosafe
PTS (Charles River), is a portable cartridge
spectrophotometer, which measures the endotoxin
content as a function of the time required to
develop a given quantity of color. The higher is the
endotoxin content; the lower is the response time.
Figure2 –EndosafePTSinstrument,manufactured
by Charles River, used for the evaluation of the
endotoxin load of the samples being considered.
Figure 3 - Coupons immersed in water for
injection; recovery of endotoxin content.
At the end of the washing treatment, the sam-
ples were extracted from the machine, placed in
a non-pyrogenic container and immersed in a de-
fined volume of water for injection (see Figure 3).
After approximately 60 minutes of immersion and
continuous manual agitation, a series of dilutions
of the solution was performed. It allowed the in-
strument to operate in its correct reading range.
The same procedure was applied for positive con-
trols: coupons of various materials that were ino-
culated and not subjected to the washing process.
Comparing the positive control with the after tre-
atment coupon results revealed a logarithmic re-
duction of approximately 3 log for the different
materials tested. See Table 1 for more details.
Managing the unexpected:
online control
Even for cleaning processes established and
validated according to common Cleaning
Validation practices, there are online controls
that can help to “manage the unexpected”. These
procedures allow continuous monitoring of the
parameters indicating the performance of the
3/7
�
Copyright © 2015 Fedegari Group
fedegari.com
Table 1 - Endotoxin load of the coupons before
and after treatment and corresponding logarithmic
reduction.
process and are useful to identify any anomaly in
the treatment.
The use of rinsing water samples can keep the
process in a “state of control” during routine
production and is prescribed in the guidelines for
validation processes written by the U.S. FDA [6]
in 2011: “...Ongoing assurance is gained during
routine production that the process remains in a
state of control”
Appropriate analytical methods can be applied
to meet the previously described requirements:
analysis of conductivity and of total organic
carbon on the waste water of the process are a
typical example.
ONLINE CONDUCTIVITY ANALYSIS METHOD
The most widespread method for online analysis
of waste water quality in a washing system is
the measurement of conductivity. Conductivity
analysis is a non-specific method that allows
to detect ionic substances dissociated in water
samples but is unable to distinguish among
different ions.
Conductivity analysis can be used to measure
residues of API (Active Pharmaceutical
Ingredients)ordetergents,iftheyareofionicnature.
The conductivity value of the final rinse water is
an additional requirement commonly included in
the performance parameters of a cleaning process.
E.g.: ‘If the readouts of the conductivity at the end
of the rinse phase is higher than the preset value,
the system must continue rinsing of all washing
lines (repetition of maximum number of phases)
until the required value is obtained’. URS of a
Pharma company.
Application of online conductivity analysis on
Fedegari steam washers
In washing machines, the conductivity sensor is
integrated in the hydraulic system and connected
to the process controller (see Figure 4).
Analysis of conductivity
•	 Measurement of water conductivity is
performed during final rinsing. For this
purpose, a conductivity meter is installed in
the drain line.
•	 Rinses are repeated automatically until the
conductivity reading is lower than a preset
value or until the maximum preset number of
rinses has been reached. If the conductivity set
point is not reached within the set number of
rinses, the process controller reports an alarm.
In this manner, the operator can apply the
procedures required to solve the anomaly in
progress.
•	 The conductivity meter is fully managed
by ‘Thema4’ process controller installed in
Fedegari steam washers.
4/7
1 Samples Endotoxin load Logarithmic re-
ductions
1a Positive
control: 316L
stainless steel
coupon
968 EU/ml
3.01
1b 316L stainless
steel coupon
after treatment
0.952 EU/ml
2 Samples Endotoxin load Logarithmic re-
ductions
2a Positive con-
trol: tempered
glass coupon
1001 EU/ml
3.02
2b Tempered
glass coupon
after treatment
0.965 EU/ml
3 Samples Endotoxin load Logarithmic re-
ductions
3a Positive control:
polypropylene
coupon
897 EU/ml
3.32
3b Polypropylene
coupon after
treatment
0.434 EU/ml
�
Copyright © 2015 Fedegari Group
fedegari.com
TOC analyzers manufactured by various suppliers
and based on different measurement methods are
commercially available. However, the following
characteristics are common to all TOC analysis
devices, regardless of the supplier:
•	 the end result of the TOC measurement is
derived from conductivity readings,
•	 the concentration of the carbon content derived
from organic material present in the water is
measured in parts per billion (ppb).
The TOC analysis is a highly sensitive method, but
it is clearly nonspecific. In fact, the instrument is
capable of detecting organic residues of products
that have degraded during washing processes but
is unable to identify the analyte in the sample.
Application of online analysis of the TOC sensor
on Fedegari washing machines
This case study regards the online application of
the conductivity and Total Organic Carbon sensors
in Fedegari steam washers. In these machines, both
the conductivity meter and the TOC analyzer are
integrated in the hydraulic system and connected
to the process controller (see Figure 5).
Figure 5 - Schematic view of the online application
of conductivity and Total Organic Carbon sensors
in Fedegari washing machines.
Figure 4 - Schematic view of the online application
of conductivity sensors in Fedegari steam washers.
METHOD FOR ONLINE ANALYSIS OF TOTAL
ORGANIC CARBON
Until recently, conductivity analysis was the only
online method used to monitor the quality of
the water of the last rinse. Recent technological
developments related to the measurement of Total
Organic Carbon (TOC) in water samples have
allowed to integrate it in a washing system.
Why is this parameter monitored as well?
Most of the products involved in biotechnological
processes are characterized by its organic nature;
so simple conductivity analysis cannot detect them.
A TOC analyzer is used to determine residues
of detergents and API (Active Pharmaceutical
Ingredients) composed of organic molecules when
the identity of the analyte is not required.
TOC analysis can be applied to every sample
containing organic carbon. The TOC analyzer
oxidizes the organic carbon present in the
molecule that constitutes the sample and measures
the resulting carbon dioxide obtained from the
oxidized organic molecule. The instrument is
capable of converting organic compounds into
inorganic compounds, which then can be analyzed
by conductivity measurements.
5/7
�
Copyright © 2015 Fedegari Group
fedegari.com
Analysis of Total Organic Carbon (TOC)
•	 The conductivity of the rinse water is measured
during final rinsing through a conductivity
sensor (which is external to the TOC sensor)
located on the drain line. The rinses are
repeated automatically until the conductivity
value is lower than a predetermined value
included within the operating limits of the
TOC analyzer.
•	 At the end of this phase, the system for
monitoring Total Organic Carbon is activated
and a controlled volume of water from the final
rinse is conveyed to the TOC sensor, opening
an isolation valve.
•	 The TOC sensor uses the oxidation induced
by an ultraviolet lamp to determine the
concentration of Total Organic Carbon that is
contained in the sample of water from the final
rinse. Integrated conductivity sensors provide
real-time conductivity measurements before
and after sample oxidation; the final value
of the Total Organic Carbon derives from
conductivity readings.
•	 Once sampling has ended, the measured TOC
value is transmitted to the process controller
of the washing machine. The controller can be
programmed to repeat the washing phase until
a predefined TOC value or a maximum number
of rinses is reached. If the preset TOC value is
not reached within the maximum number of
rinses, then the Process Controller reports an
alarm. In this manner, the operator can apply
the procedure required to solve the anomaly in
progress.
•	 As highlighted in the preceding application,
the conductivity meter and the TOC sensor are
fully managed by ‘Thema4’ process controller
installed on Fedegari washing machines.
While the conductivity measurement is
instantaneous, the introduction of an online TOC
analysis entails a longer cycle time. The washing
system is in fact in a standby state during sample
analysis in order to assess the need to perform
further rinsing.
Nevertheless, comparison between off-line and
online measurements clearly shows that the online
application of conductivity and TOC allows
to eliminate manual sampling, the transport of
samplestolaboratories,thedelayinlaboratorytime
analysis, the "human variable" in the assessment
of the results, and the risk of cross contamination.
Online application allows continuous monitoring
of process parameters in the reporting of any errors
during a previously validated process.
A schematic summary of the advantages/
disadvantages of the online application of
conductivity and TOC is provided hereafter.
Advantages of the online conductivity and TOC
analysis methods
•	 Faster analyses: elimination of manual
sampling
•	 Reduced risk of cross contamination:
elimination of manual sampling
•	 Reduced costs: no manual analysis in a
laboratory
•	 Reduction of inactivity periods: waiting for
laboratory analysis
•	 Analysis reliability : continuous monitoring of
process parameters
•	 Real-time "alarm system"
•	 High sensitivity of the measurement
Disadvantages of the online conductivity and TOC
analysis methods
•	 Nonspecific methods
•	 Only for the TOC analyzer: initial cost of
instrument
FINAL CONSIDERATIONS
Removal of bacterial endotoxins - The washing
tests performed at Fedegari R&D department
have proven to be effective in removing not only
chemical but also microbiological soil. A more
intense washing process is expected to yield a
reduction of endotoxin load distinctly higher
6/7
�
Copyright © 2015 Fedegari Group
fedegari.com
than 3 log. A supplementary test, starting from
a “chemically dirty” load, might constitute an
additional challenge in the removal of inoculated
bacterial endotoxins.
Online analyses of Total Organic Carbon and
conductivity are complementary for continuous
monitoring of the washing process performances.
The conductivity meter and the TOC sensor can
be fully integrated in the hydraulic system of a
washing machine and connected to the process
controller, as in the case of Fedegari steam washers.
The complete integration with the controller
allows regulatory compliance to the requirements
of the FDA 21 CFR part 11: the data related to
each deviation from the standard required in terms
of washing performance are recorded and stored in
the system history. Through these applications, the
process remains in a state of control during routine
production as indicated by the FDA itself [6].
References
1.	 M.L. Bernuzzi. Fattori chiave nella scelta/
sviluppo di una strategia di lavaggio; 1a parte.
ASCCA News, 2014; 4: 6-9.
2.	 Guide to Inspections of Validation of Cleaning
Practices; U.S. Food and DrugAdministration:
1993. Available at: www.fda.gov/ICECI/
Inspections/InspectionGuides/ucm074922.
htm (accessed February 24, 2015)
3.	 Technical Report No. 29: Points to Consider
for Cleaning Validation; Parenteral Drug
Association: Bethesda, MD, 1998.
4.	 USP 37 General Chapter <85> BACTERIAL
ENDOTOXINS TEST, 2014.
5.	 European Pharmacopoeia 8.0. Bacterial
Endotoxins, 01/2014:20614.
6.	 Guidance for Industry. Process Validation:
General Principals and Practice; U.S. Food
and Drug Administration: [on-line] 2011.
Available at: www.fda.gov/downloads/
Drugs/Guidances/UCM070336.pdf‎ (accessed
February 24, 2015)
1.	
7/7

More Related Content

What's hot

USP &lt;1116> and its impact on Microbiology
USP &lt;1116> and its impact on MicrobiologyUSP &lt;1116> and its impact on Microbiology
USP &lt;1116> and its impact on MicrobiologyTim Sandle, Ph.D.
 
Update on-pda-tr22-revisions-for-aseptic-process-simulations
Update on-pda-tr22-revisions-for-aseptic-process-simulationsUpdate on-pda-tr22-revisions-for-aseptic-process-simulations
Update on-pda-tr22-revisions-for-aseptic-process-simulationsHainer Martínez Cabrera
 
Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Tim Sandle, Ph.D.
 
Sterile Facilities
Sterile FacilitiesSterile Facilities
Sterile FacilitiesEric Sipe
 
Microbial conamination control
Microbial conamination controlMicrobial conamination control
Microbial conamination controlpragatk
 
Usp 36 Chapter 1116 environment monitoring
Usp 36 Chapter 1116 environment monitoringUsp 36 Chapter 1116 environment monitoring
Usp 36 Chapter 1116 environment monitoringdelli_intralab
 
Fundamentals Of Contamination Control
Fundamentals Of Contamination ControlFundamentals Of Contamination Control
Fundamentals Of Contamination Controlrwwlws
 
Cleaning and Disinfecting iv Hoods and Rooms
Cleaning and Disinfecting iv Hoods and RoomsCleaning and Disinfecting iv Hoods and Rooms
Cleaning and Disinfecting iv Hoods and RoomsJerry Fahrni
 
FACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIES
FACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIESFACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIES
FACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIESNEHA SINGH
 
Environmental Characterization of Controlled Rooms
Environmental Characterization of Controlled RoomsEnvironmental Characterization of Controlled Rooms
Environmental Characterization of Controlled Roomsangelsalaman
 
Aseptic process tech & advanced sterile product mfg rashmi nasare
Aseptic process tech & advanced sterile product mfg  rashmi nasareAseptic process tech & advanced sterile product mfg  rashmi nasare
Aseptic process tech & advanced sterile product mfg rashmi nasareRASHMINasare
 
Environmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingEnvironmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingJerry Fahrni
 
Contamination control in pharmaceutical industry
Contamination control in pharmaceutical industryContamination control in pharmaceutical industry
Contamination control in pharmaceutical industryclientscomp
 
Aseptic processing
Aseptic processingAseptic processing
Aseptic processingShivaram
 
Environmental monitoring - surface sampling
Environmental monitoring - surface samplingEnvironmental monitoring - surface sampling
Environmental monitoring - surface samplingJerry Fahrni
 
Parenteral production and aseptic area
Parenteral production and aseptic areaParenteral production and aseptic area
Parenteral production and aseptic areaShaik Sana
 

What's hot (20)

USP &lt;1116> and its impact on Microbiology
USP &lt;1116> and its impact on MicrobiologyUSP &lt;1116> and its impact on Microbiology
USP &lt;1116> and its impact on Microbiology
 
Environmental monitoring
Environmental monitoringEnvironmental monitoring
Environmental monitoring
 
Update on-pda-tr22-revisions-for-aseptic-process-simulations
Update on-pda-tr22-revisions-for-aseptic-process-simulationsUpdate on-pda-tr22-revisions-for-aseptic-process-simulations
Update on-pda-tr22-revisions-for-aseptic-process-simulations
 
Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges
 
Manufacturing facility of parentarals as per schedule m
Manufacturing facility of parentarals as per schedule mManufacturing facility of parentarals as per schedule m
Manufacturing facility of parentarals as per schedule m
 
Sterile Facilities
Sterile FacilitiesSterile Facilities
Sterile Facilities
 
Microbial conamination control
Microbial conamination controlMicrobial conamination control
Microbial conamination control
 
Usp 36 Chapter 1116 environment monitoring
Usp 36 Chapter 1116 environment monitoringUsp 36 Chapter 1116 environment monitoring
Usp 36 Chapter 1116 environment monitoring
 
Fundamentals Of Contamination Control
Fundamentals Of Contamination ControlFundamentals Of Contamination Control
Fundamentals Of Contamination Control
 
Cleaning and Disinfecting iv Hoods and Rooms
Cleaning and Disinfecting iv Hoods and RoomsCleaning and Disinfecting iv Hoods and Rooms
Cleaning and Disinfecting iv Hoods and Rooms
 
FACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIES
FACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIESFACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIES
FACTORS IN THE DESIGN OF PARENTERAL PRODUCTION FACILITIES
 
Environmental Characterization of Controlled Rooms
Environmental Characterization of Controlled RoomsEnvironmental Characterization of Controlled Rooms
Environmental Characterization of Controlled Rooms
 
Aseptic process tech & advanced sterile product mfg rashmi nasare
Aseptic process tech & advanced sterile product mfg  rashmi nasareAseptic process tech & advanced sterile product mfg  rashmi nasare
Aseptic process tech & advanced sterile product mfg rashmi nasare
 
Asepting procesing
Asepting procesingAsepting procesing
Asepting procesing
 
Environmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingEnvironmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testing
 
Ne28. agalloco tr22
Ne28. agalloco tr22Ne28. agalloco tr22
Ne28. agalloco tr22
 
Contamination control in pharmaceutical industry
Contamination control in pharmaceutical industryContamination control in pharmaceutical industry
Contamination control in pharmaceutical industry
 
Aseptic processing
Aseptic processingAseptic processing
Aseptic processing
 
Environmental monitoring - surface sampling
Environmental monitoring - surface samplingEnvironmental monitoring - surface sampling
Environmental monitoring - surface sampling
 
Parenteral production and aseptic area
Parenteral production and aseptic areaParenteral production and aseptic area
Parenteral production and aseptic area
 

Viewers also liked

Integrated solutions for contamination control
Integrated solutions for contamination control Integrated solutions for contamination control
Integrated solutions for contamination control Fedegari Group
 
Tara Versteeg - Resume
Tara Versteeg - ResumeTara Versteeg - Resume
Tara Versteeg - ResumeTara Versteeg
 
Marco 2014 iib90_overview_port
Marco 2014 iib90_overview_portMarco 2014 iib90_overview_port
Marco 2014 iib90_overview_portJuan Garay
 
Integrating mRDTs into the health system in Uganda: preparing health workers ...
Integrating mRDTs into the health system in Uganda: preparing health workers ...Integrating mRDTs into the health system in Uganda: preparing health workers ...
Integrating mRDTs into the health system in Uganda: preparing health workers ...Malaria Consortium
 
Steps To Synchronize G-Mail Account With Microsoft Outlook
Steps To Synchronize G-Mail Account With Microsoft OutlookSteps To Synchronize G-Mail Account With Microsoft Outlook
Steps To Synchronize G-Mail Account With Microsoft OutlookClark445Kent
 
Innovation in Chemical Bio-decontamination: from the process to the machine
Innovation in Chemical Bio-decontamination: from the process to the machineInnovation in Chemical Bio-decontamination: from the process to the machine
Innovation in Chemical Bio-decontamination: from the process to the machineFedegari Group
 
It’s all in the detail developing effective health-related job aids
It’s all in the detail  developing effective health-related job aidsIt’s all in the detail  developing effective health-related job aids
It’s all in the detail developing effective health-related job aidsMalaria Consortium
 
Fedegari Group - Company profile - Passion that drives performances.
Fedegari Group - Company profile - Passion that drives performances.Fedegari Group - Company profile - Passion that drives performances.
Fedegari Group - Company profile - Passion that drives performances.Fedegari Group
 
Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]
Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]
Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]Energy Digital Summit
 
Invest in Real Estate What you should think about first how to get started
Invest in Real Estate What you should think about first how to get startedInvest in Real Estate What you should think about first how to get started
Invest in Real Estate What you should think about first how to get startedcvergaraamigo
 
Implementing integrated community case management: stakeholder experiences an...
Implementing integrated community case management: stakeholder experiences an...Implementing integrated community case management: stakeholder experiences an...
Implementing integrated community case management: stakeholder experiences an...Malaria Consortium
 
เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1
เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1
เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1Jee Czjee
 
How to cook a basic stir fry
How to cook a basic stir fryHow to cook a basic stir fry
How to cook a basic stir fryrddunn0720
 
1a budaya&pembelajaran
1a budaya&pembelajaran1a budaya&pembelajaran
1a budaya&pembelajaranSang Kojol
 
Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...
Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...
Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...Energy Digital Summit
 

Viewers also liked (20)

Integrated solutions for contamination control
Integrated solutions for contamination control Integrated solutions for contamination control
Integrated solutions for contamination control
 
Tara Versteeg - Resume
Tara Versteeg - ResumeTara Versteeg - Resume
Tara Versteeg - Resume
 
Marco 2014 iib90_overview_port
Marco 2014 iib90_overview_portMarco 2014 iib90_overview_port
Marco 2014 iib90_overview_port
 
Photo story
Photo storyPhoto story
Photo story
 
Integrating mRDTs into the health system in Uganda: preparing health workers ...
Integrating mRDTs into the health system in Uganda: preparing health workers ...Integrating mRDTs into the health system in Uganda: preparing health workers ...
Integrating mRDTs into the health system in Uganda: preparing health workers ...
 
Steps To Synchronize G-Mail Account With Microsoft Outlook
Steps To Synchronize G-Mail Account With Microsoft OutlookSteps To Synchronize G-Mail Account With Microsoft Outlook
Steps To Synchronize G-Mail Account With Microsoft Outlook
 
Ontologi Peripatetik
Ontologi PeripatetikOntologi Peripatetik
Ontologi Peripatetik
 
Mulla Sadra's philosophy
Mulla Sadra's philosophyMulla Sadra's philosophy
Mulla Sadra's philosophy
 
Innovation in Chemical Bio-decontamination: from the process to the machine
Innovation in Chemical Bio-decontamination: from the process to the machineInnovation in Chemical Bio-decontamination: from the process to the machine
Innovation in Chemical Bio-decontamination: from the process to the machine
 
It’s all in the detail developing effective health-related job aids
It’s all in the detail  developing effective health-related job aidsIt’s all in the detail  developing effective health-related job aids
It’s all in the detail developing effective health-related job aids
 
Fedegari Group - Company profile - Passion that drives performances.
Fedegari Group - Company profile - Passion that drives performances.Fedegari Group - Company profile - Passion that drives performances.
Fedegari Group - Company profile - Passion that drives performances.
 
Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]
Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]
Communicating in Constellations - Steve Rubel [Energy Digital Summit 2015]
 
How women to be a leader?
How women to be a leader? How women to be a leader?
How women to be a leader?
 
Invest in Real Estate What you should think about first how to get started
Invest in Real Estate What you should think about first how to get startedInvest in Real Estate What you should think about first how to get started
Invest in Real Estate What you should think about first how to get started
 
Fasakhova
FasakhovaFasakhova
Fasakhova
 
Implementing integrated community case management: stakeholder experiences an...
Implementing integrated community case management: stakeholder experiences an...Implementing integrated community case management: stakeholder experiences an...
Implementing integrated community case management: stakeholder experiences an...
 
เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1
เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1
เธ‡เธฒเธ™เธ™เธณเน€เธชเธ™เธญ1
 
How to cook a basic stir fry
How to cook a basic stir fryHow to cook a basic stir fry
How to cook a basic stir fry
 
1a budaya&pembelajaran
1a budaya&pembelajaran1a budaya&pembelajaran
1a budaya&pembelajaran
 
Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...
Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...
Brands and Identity: The Power of StoryTelling - Scott Stone [Energy Digital ...
 

Similar to Key factors for developing a pharmaceutical cleaning strategy - Part 2

White paper - Continued process verification methods in cleaning validation
White paper - Continued process verification methods in cleaning validationWhite paper - Continued process verification methods in cleaning validation
White paper - Continued process verification methods in cleaning validationFedegari Group
 
IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...
IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...
IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...IRJET Journal
 
Performance status of common effluent treatment
Performance status of common effluent treatmentPerformance status of common effluent treatment
Performance status of common effluent treatmenteSAT Publishing House
 
IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...
IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...
IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...IRJET Journal
 
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANINGCONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANINGShankar Maind Patil
 
Operation of Sterility Testing Isolators and validation issues
Operation of Sterility Testing Isolators and validation issuesOperation of Sterility Testing Isolators and validation issues
Operation of Sterility Testing Isolators and validation issuesTim Sandle, Ph.D.
 
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...IRJET Journal
 
Non-invasive moisture content measurement system based on the ESP8266 microco...
Non-invasive moisture content measurement system based on the ESP8266 microco...Non-invasive moisture content measurement system based on the ESP8266 microco...
Non-invasive moisture content measurement system based on the ESP8266 microco...journalBEEI
 
Aseptic Media Fill Validation.pptx
Aseptic Media Fill Validation.pptxAseptic Media Fill Validation.pptx
Aseptic Media Fill Validation.pptxMansi792999
 
Auditing of microbiological lab
Auditing of microbiological lab Auditing of microbiological lab
Auditing of microbiological lab KhushbooKunkulol
 
Cleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfCleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfLdiaSabino
 
Uasb wsatewater treatment
Uasb wsatewater treatmentUasb wsatewater treatment
Uasb wsatewater treatmentSakil Ahmed
 

Similar to Key factors for developing a pharmaceutical cleaning strategy - Part 2 (20)

White paper - Continued process verification methods in cleaning validation
White paper - Continued process verification methods in cleaning validationWhite paper - Continued process verification methods in cleaning validation
White paper - Continued process verification methods in cleaning validation
 
TEXTILE MANUFACTURING PROCESS
TEXTILE MANUFACTURING PROCESSTEXTILE MANUFACTURING PROCESS
TEXTILE MANUFACTURING PROCESS
 
IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...
IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...
IRJET- Macrophytic Treatment of Textile Waste Water Together with Root Zone T...
 
Unit 2 ppt new.pptx
Unit 2 ppt new.pptxUnit 2 ppt new.pptx
Unit 2 ppt new.pptx
 
xyz
xyzxyz
xyz
 
Performance status of common effluent treatment
Performance status of common effluent treatmentPerformance status of common effluent treatment
Performance status of common effluent treatment
 
IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...
IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...
IRJET- Use of Natural and Artificial Multimedia Filter as an Adsorbent for Fi...
 
F04915660
F04915660F04915660
F04915660
 
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANINGCONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
CONCEPT OF CIP (Clean In Place ) AND FACILITY CLEANING
 
Endoscope cleaning
Endoscope cleaningEndoscope cleaning
Endoscope cleaning
 
Operation of Sterility Testing Isolators and validation issues
Operation of Sterility Testing Isolators and validation issuesOperation of Sterility Testing Isolators and validation issues
Operation of Sterility Testing Isolators and validation issues
 
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...IRJET-  	  Characterisation of Grey Water and Treatment using Moving Bed Biof...
IRJET- Characterisation of Grey Water and Treatment using Moving Bed Biof...
 
Non-invasive moisture content measurement system based on the ESP8266 microco...
Non-invasive moisture content measurement system based on the ESP8266 microco...Non-invasive moisture content measurement system based on the ESP8266 microco...
Non-invasive moisture content measurement system based on the ESP8266 microco...
 
Aseptic Media Fill Validation.pptx
Aseptic Media Fill Validation.pptxAseptic Media Fill Validation.pptx
Aseptic Media Fill Validation.pptx
 
Auditing of microbiological lab
Auditing of microbiological lab Auditing of microbiological lab
Auditing of microbiological lab
 
Cleaning Validation.pptx
Cleaning Validation.pptxCleaning Validation.pptx
Cleaning Validation.pptx
 
Moisture measurement
Moisture measurementMoisture measurement
Moisture measurement
 
Cleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdfCleaning Validation - A lifecycle approach.pdf
Cleaning Validation - A lifecycle approach.pdf
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Uasb wsatewater treatment
Uasb wsatewater treatmentUasb wsatewater treatment
Uasb wsatewater treatment
 

More from Fedegari Group

CAMPIONI HI TECH - Corriere della Sera - Giulia Cimpanelli
CAMPIONI HI TECH - Corriere della Sera - Giulia CimpanelliCAMPIONI HI TECH - Corriere della Sera - Giulia Cimpanelli
CAMPIONI HI TECH - Corriere della Sera - Giulia CimpanelliFedegari Group
 
Un modello 4.0 - Giovanni Felice
Un modello 4.0 - Giovanni Felice Un modello 4.0 - Giovanni Felice
Un modello 4.0 - Giovanni Felice Fedegari Group
 
Integrated Solution for Filling Al- kegs with Sterile Antibiotics
Integrated Solution for Filling Al- kegs with Sterile AntibioticsIntegrated Solution for Filling Al- kegs with Sterile Antibiotics
Integrated Solution for Filling Al- kegs with Sterile AntibioticsFedegari Group
 
Factores clave en la elecciòn y desarrollo de una strategia de limpieza
Factores clave en la elecciòn y desarrollo de una strategia de limpiezaFactores clave en la elecciòn y desarrollo de una strategia de limpieza
Factores clave en la elecciòn y desarrollo de una strategia de limpiezaFedegari Group
 
Métodos de verificaciòn continua del proceso en validaciòn de limpieza
Métodos de verificaciòn continua del proceso en validaciòn de limpiezaMétodos de verificaciòn continua del proceso en validaciòn de limpieza
Métodos de verificaciòn continua del proceso en validaciòn de limpiezaFedegari Group
 
Sistemi integrati per l'efficienza produttiva
Sistemi integrati per l'efficienza produttivaSistemi integrati per l'efficienza produttiva
Sistemi integrati per l'efficienza produttivaFedegari Group
 
Pavia e i due uomini d'acciaio: Fedegari e Mo
Pavia e i due uomini d'acciaio: Fedegari e MoPavia e i due uomini d'acciaio: Fedegari e Mo
Pavia e i due uomini d'acciaio: Fedegari e MoFedegari Group
 
Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...
Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...
Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...Fedegari Group
 
Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017
Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017
Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017Fedegari Group
 
Fedegari Training Calendar 2017
Fedegari Training Calendar 2017Fedegari Training Calendar 2017
Fedegari Training Calendar 2017Fedegari Group
 
Articolo Fedegari - Il Sole 24 ore - 09/12/16
Articolo Fedegari - Il Sole 24 ore - 09/12/16Articolo Fedegari - Il Sole 24 ore - 09/12/16
Articolo Fedegari - Il Sole 24 ore - 09/12/16Fedegari Group
 
White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...
White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...
White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...Fedegari Group
 
Fedegari Warehousing Solutions
Fedegari  Warehousing SolutionsFedegari  Warehousing Solutions
Fedegari Warehousing SolutionsFedegari Group
 
Почему изменения в асептическом производстве являются неизбежными?
Почему изменения в асептическом производстве являются неизбежными?Почему изменения в асептическом производстве являются неизбежными?
Почему изменения в асептическом производстве являются неизбежными?Fedegari Group
 
FEDEGARI: from raw material to integrated systems
FEDEGARI: from raw material to integrated systemsFEDEGARI: from raw material to integrated systems
FEDEGARI: from raw material to integrated systemsFedegari Group
 
Fedegari After Sales Services
Fedegari After Sales ServicesFedegari After Sales Services
Fedegari After Sales ServicesFedegari Group
 
L'eccellenza è un percorso senza fine...
L'eccellenza è un percorso senza fine...L'eccellenza è un percorso senza fine...
L'eccellenza è un percorso senza fine...Fedegari Group
 
INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...
INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...
INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...Fedegari Group
 
Fedegari Training & Education Calendar 2016
Fedegari Training & Education Calendar 2016Fedegari Training & Education Calendar 2016
Fedegari Training & Education Calendar 2016Fedegari Group
 

More from Fedegari Group (20)

CAMPIONI HI TECH - Corriere della Sera - Giulia Cimpanelli
CAMPIONI HI TECH - Corriere della Sera - Giulia CimpanelliCAMPIONI HI TECH - Corriere della Sera - Giulia Cimpanelli
CAMPIONI HI TECH - Corriere della Sera - Giulia Cimpanelli
 
Un modello 4.0 - Giovanni Felice
Un modello 4.0 - Giovanni Felice Un modello 4.0 - Giovanni Felice
Un modello 4.0 - Giovanni Felice
 
Integrated Solution for Filling Al- kegs with Sterile Antibiotics
Integrated Solution for Filling Al- kegs with Sterile AntibioticsIntegrated Solution for Filling Al- kegs with Sterile Antibiotics
Integrated Solution for Filling Al- kegs with Sterile Antibiotics
 
Factores clave en la elecciòn y desarrollo de una strategia de limpieza
Factores clave en la elecciòn y desarrollo de una strategia de limpiezaFactores clave en la elecciòn y desarrollo de una strategia de limpieza
Factores clave en la elecciòn y desarrollo de una strategia de limpieza
 
Métodos de verificaciòn continua del proceso en validaciòn de limpieza
Métodos de verificaciòn continua del proceso en validaciòn de limpiezaMétodos de verificaciòn continua del proceso en validaciòn de limpieza
Métodos de verificaciòn continua del proceso en validaciòn de limpieza
 
Sistemi integrati per l'efficienza produttiva
Sistemi integrati per l'efficienza produttivaSistemi integrati per l'efficienza produttiva
Sistemi integrati per l'efficienza produttiva
 
Pavia e i due uomini d'acciaio: Fedegari e Mo
Pavia e i due uomini d'acciaio: Fedegari e MoPavia e i due uomini d'acciaio: Fedegari e Mo
Pavia e i due uomini d'acciaio: Fedegari e Mo
 
Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...
Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...
Un cuore tecnologico al servizio dell'industria farmaceutica - Articolo Fedeg...
 
Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017
Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017
Svolta Tecnologica - Articolo Fedegari Tecn'è Gennaio 2017
 
Fedegari Training Calendar 2017
Fedegari Training Calendar 2017Fedegari Training Calendar 2017
Fedegari Training Calendar 2017
 
Articolo Fedegari - Il Sole 24 ore - 09/12/16
Articolo Fedegari - Il Sole 24 ore - 09/12/16Articolo Fedegari - Il Sole 24 ore - 09/12/16
Articolo Fedegari - Il Sole 24 ore - 09/12/16
 
White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...
White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...
White Paper - Thermal sterilization of Pre-filled Syringes (PFS) with hyaluro...
 
Fedegari Warehousing Solutions
Fedegari  Warehousing SolutionsFedegari  Warehousing Solutions
Fedegari Warehousing Solutions
 
Почему изменения в асептическом производстве являются неизбежными?
Почему изменения в асептическом производстве являются неизбежными?Почему изменения в асептическом производстве являются неизбежными?
Почему изменения в асептическом производстве являются неизбежными?
 
FEDEGARI: from raw material to integrated systems
FEDEGARI: from raw material to integrated systemsFEDEGARI: from raw material to integrated systems
FEDEGARI: from raw material to integrated systems
 
Excellence is a path.
Excellence is a path.Excellence is a path.
Excellence is a path.
 
Fedegari After Sales Services
Fedegari After Sales ServicesFedegari After Sales Services
Fedegari After Sales Services
 
L'eccellenza è un percorso senza fine...
L'eccellenza è un percorso senza fine...L'eccellenza è un percorso senza fine...
L'eccellenza è un percorso senza fine...
 
INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...
INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...
INSIGHTS IN CHEMICAL BIO-DECONTAMINATION: design, applications and validation...
 
Fedegari Training & Education Calendar 2016
Fedegari Training & Education Calendar 2016Fedegari Training & Education Calendar 2016
Fedegari Training & Education Calendar 2016
 

Recently uploaded

Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Memriyagarg453
 
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...chandigarhentertainm
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...indiancallgirl4rent
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...Ahmedabad Call Girls
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMuzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012adityaroy0215
 
Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...
Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...
Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...Sheetaleventcompany
 
Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Vipesco
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.ktanvi103
 
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetTirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...
Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...russian goa call girl and escorts service
 
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In RaipurCall Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipurgragmanisha42
 
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...indiancallgirl4rent
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...Gfnyt.com
 

Recently uploaded (20)

Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Patiala Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMuzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
VIP Call Girl DLF Phase 2 Gurgaon (Noida) Just Meet Me@ 9711199012
 
Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...
Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...
Punjab❤️Call girls in Mohali ☎️7435815124☎️ Call Girl service in Mohali☎️ Moh...
 
Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
 
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetTirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...
Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...Call Girls Service In Goa  💋 9316020077💋 Goa Call Girls  By Russian Call Girl...
Call Girls Service In Goa 💋 9316020077💋 Goa Call Girls By Russian Call Girl...
 
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In RaipurCall Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
Call Girl Raipur 📲 9999965857 ヅ10k NiGhT Call Girls In Raipur
 
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
(Ajay) Call Girls in Dehradun- 8854095900 Escorts Service 50% Off with Cash O...
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
 

Key factors for developing a pharmaceutical cleaning strategy - Part 2

  • 1. � Copyright © 2015 Fedegari Group fedegari.com Key factors in the choice and development of a cleaning strategy Part II M. L. Bernuzzi Manager, R&D, Fedegari Autoclavi SpA S. Ferretti Ph.D, R&D, Fedegari Autoclavi SpA This article discusses the removal of bacterial endotoxins in a washing process and the need for a continuous monitoring activity according to common cleaning validation practices. Results of washing tests performed with different load materials following an effective cleaning strategy to remove microbiological soil are presented. In addition, this paper compares two different online assay methods: conductivity analysis and Total Organic Carbon analysis. Keywords: Cleaning, Microbiological Soil, Bacterial Endotoxins, Conductivity Analysis, Total Organic Carbon Analysis. Introduction The purpose of a washing treatment consists in removing the soil present on the load. Some aspects of soil removal, strictly of a chemical type, were previously discussed [1]. However, a new cleaning requirement is becoming increasingly widespread: treating a different soil, which is not chemical but microbiological. A very frequent requirement is indeed to remove bacterial endotoxins. Once the washing process has been validated, the chemical and/or biological soil has been removed, and its absence on the load has been demonstrated, the process must remain in a continuous “monitoring mode”. Periodic checks or online monitoring of the rinse water contribute to: • Increasing the safety margin that makes a validated washing process, reliable and always reproducible in its results; • Protecting the end user of the product from an unexpected result that does not conform to expectations. Removal of bacterial endotoxins The U.S. FDA Cleaning validation guide states that “…control of the bioburden through adequate cleaning and storage of equipment is important to ensure that subsequent sterilization or sanitization procedures achieve the necessary assurance of sterility” [2]. Failure to remove/destroy endotoxins is certainly a limitation of steam sterilization; therefore, the “endotoxin content” of a generic load, intended for subsequent sterilization, is often monitored and controlled during its handling, production and washing. Endotoxin removal can be investigated during the validation cycles of the cleaning process. In fact, three endotoxin log reductions, which is a standard requirement for the depyrogenation process, might be the additional target of a washing procedure. Obviously, the required reduction would not be achieved due to the high temperatures but due to the mechanical removal effect dictated by the cleaning medium, such as water, steam or aqueous solution with the addition of detergents. DOC. 386143.2 1/7
  • 2. � Copyright © 2015 Fedegari Group fedegari.com Case study The test, performed at the Research and Development department of Fedegari, removed bacterial endotoxins from different materials by a washing cycle and constituted a preliminary assessment of the effectiveness of this process in terms of endotoxin load reduction. The goal of this study was to verify the 3 log reduction of the inoculated endotoxin load on the surface of coupons measuring 100 mm x 100 mm. The use of coupons in the initial setup of a washing process is suggested by PDATechnical Report No. 29: Points to consider for Cleaning Validation [3]. Stainless steal, 316 L, polypropylene and tempered glass coupons were investigated. The choice of materials to be inoculated was of a practical nature: materials usually found in pharmaceutical practices. TEST PROCEDURES The materials were depyrogenated through a thermal treatment performed at 250°C. Due to the high temperatures, it was not possible to expose the polypropylene coupons to the same process. This way, the plastic coupons were disinfected with a common biocidal agent representing a worst case in this study. Every coupon was inoculated with a predefined quantity of endotoxins (Endotoxin Indicator ENDOSAFE, Charles River - Source: Escherichia coli 055:B5) and the inoculum was dried at 40°C by means of a laboratory heater. The drying time depended on the treated material. Appropriate supports for the coupons were studied to ensure that the washing medium reached the inoculated coupon surface (see Figure 1). In addition, an appropriate cleaning equipment, composed of rotating arms with nozzles, was designed and built. Since the effectiveness of the process in terms of the intended goal was not known, a standard cycle already applied at a Figure 1 - Stainless steel coupons placed on special supports inside the washing machine customer facility was used. This cycle was mainly composed by various washing phases with water recirculation and a washing phase with water disposal, injection and simultaneous drain of water. The washing machine located in the Research and Development laboratory of Fedegari is supplied with deionized water originating from a reverse osmosis plant. In order to remove from the results any endotoxin content due to the contribution of the water entering the system, an analysis of the water was carried out. The endotoxin content was evaluated by applying a “Bacterial Endotoxin Test – BET”, as described in the various international Pharmacopoeias, harmonized in this regard [4-5]. This test is capable of identifying or quantifying endotoxins originating from gram-negative bacteria by using Limulus Polyphemus amebocyte lysate. 2/7
  • 3. � Copyright © 2015 Fedegari Group fedegari.com There are three techniques for this type of test: • gel clot; • turbidimetric; • chromogenic. The European Pharmacopoeia [5] indicates the following analysis methods: • Method A. Gel-clot method: limit test • Method B. Gel-clot method: semi-quantitative test • Method C. Turbidimetric kinetic method • Method D. Chromogenic kinetic method • Method E. Chromogenic end-point method • Method F. Turbidimetric end-point method Thetechniqueappliedwasthechromogenickinetic one (method D of the European Pharmacopeia). The instrument used (see Figure 2), Endosafe PTS (Charles River), is a portable cartridge spectrophotometer, which measures the endotoxin content as a function of the time required to develop a given quantity of color. The higher is the endotoxin content; the lower is the response time. Figure2 –EndosafePTSinstrument,manufactured by Charles River, used for the evaluation of the endotoxin load of the samples being considered. Figure 3 - Coupons immersed in water for injection; recovery of endotoxin content. At the end of the washing treatment, the sam- ples were extracted from the machine, placed in a non-pyrogenic container and immersed in a de- fined volume of water for injection (see Figure 3). After approximately 60 minutes of immersion and continuous manual agitation, a series of dilutions of the solution was performed. It allowed the in- strument to operate in its correct reading range. The same procedure was applied for positive con- trols: coupons of various materials that were ino- culated and not subjected to the washing process. Comparing the positive control with the after tre- atment coupon results revealed a logarithmic re- duction of approximately 3 log for the different materials tested. See Table 1 for more details. Managing the unexpected: online control Even for cleaning processes established and validated according to common Cleaning Validation practices, there are online controls that can help to “manage the unexpected”. These procedures allow continuous monitoring of the parameters indicating the performance of the 3/7
  • 4. � Copyright © 2015 Fedegari Group fedegari.com Table 1 - Endotoxin load of the coupons before and after treatment and corresponding logarithmic reduction. process and are useful to identify any anomaly in the treatment. The use of rinsing water samples can keep the process in a “state of control” during routine production and is prescribed in the guidelines for validation processes written by the U.S. FDA [6] in 2011: “...Ongoing assurance is gained during routine production that the process remains in a state of control” Appropriate analytical methods can be applied to meet the previously described requirements: analysis of conductivity and of total organic carbon on the waste water of the process are a typical example. ONLINE CONDUCTIVITY ANALYSIS METHOD The most widespread method for online analysis of waste water quality in a washing system is the measurement of conductivity. Conductivity analysis is a non-specific method that allows to detect ionic substances dissociated in water samples but is unable to distinguish among different ions. Conductivity analysis can be used to measure residues of API (Active Pharmaceutical Ingredients)ordetergents,iftheyareofionicnature. The conductivity value of the final rinse water is an additional requirement commonly included in the performance parameters of a cleaning process. E.g.: ‘If the readouts of the conductivity at the end of the rinse phase is higher than the preset value, the system must continue rinsing of all washing lines (repetition of maximum number of phases) until the required value is obtained’. URS of a Pharma company. Application of online conductivity analysis on Fedegari steam washers In washing machines, the conductivity sensor is integrated in the hydraulic system and connected to the process controller (see Figure 4). Analysis of conductivity • Measurement of water conductivity is performed during final rinsing. For this purpose, a conductivity meter is installed in the drain line. • Rinses are repeated automatically until the conductivity reading is lower than a preset value or until the maximum preset number of rinses has been reached. If the conductivity set point is not reached within the set number of rinses, the process controller reports an alarm. In this manner, the operator can apply the procedures required to solve the anomaly in progress. • The conductivity meter is fully managed by ‘Thema4’ process controller installed in Fedegari steam washers. 4/7 1 Samples Endotoxin load Logarithmic re- ductions 1a Positive control: 316L stainless steel coupon 968 EU/ml 3.01 1b 316L stainless steel coupon after treatment 0.952 EU/ml 2 Samples Endotoxin load Logarithmic re- ductions 2a Positive con- trol: tempered glass coupon 1001 EU/ml 3.02 2b Tempered glass coupon after treatment 0.965 EU/ml 3 Samples Endotoxin load Logarithmic re- ductions 3a Positive control: polypropylene coupon 897 EU/ml 3.32 3b Polypropylene coupon after treatment 0.434 EU/ml
  • 5. � Copyright © 2015 Fedegari Group fedegari.com TOC analyzers manufactured by various suppliers and based on different measurement methods are commercially available. However, the following characteristics are common to all TOC analysis devices, regardless of the supplier: • the end result of the TOC measurement is derived from conductivity readings, • the concentration of the carbon content derived from organic material present in the water is measured in parts per billion (ppb). The TOC analysis is a highly sensitive method, but it is clearly nonspecific. In fact, the instrument is capable of detecting organic residues of products that have degraded during washing processes but is unable to identify the analyte in the sample. Application of online analysis of the TOC sensor on Fedegari washing machines This case study regards the online application of the conductivity and Total Organic Carbon sensors in Fedegari steam washers. In these machines, both the conductivity meter and the TOC analyzer are integrated in the hydraulic system and connected to the process controller (see Figure 5). Figure 5 - Schematic view of the online application of conductivity and Total Organic Carbon sensors in Fedegari washing machines. Figure 4 - Schematic view of the online application of conductivity sensors in Fedegari steam washers. METHOD FOR ONLINE ANALYSIS OF TOTAL ORGANIC CARBON Until recently, conductivity analysis was the only online method used to monitor the quality of the water of the last rinse. Recent technological developments related to the measurement of Total Organic Carbon (TOC) in water samples have allowed to integrate it in a washing system. Why is this parameter monitored as well? Most of the products involved in biotechnological processes are characterized by its organic nature; so simple conductivity analysis cannot detect them. A TOC analyzer is used to determine residues of detergents and API (Active Pharmaceutical Ingredients) composed of organic molecules when the identity of the analyte is not required. TOC analysis can be applied to every sample containing organic carbon. The TOC analyzer oxidizes the organic carbon present in the molecule that constitutes the sample and measures the resulting carbon dioxide obtained from the oxidized organic molecule. The instrument is capable of converting organic compounds into inorganic compounds, which then can be analyzed by conductivity measurements. 5/7
  • 6. � Copyright © 2015 Fedegari Group fedegari.com Analysis of Total Organic Carbon (TOC) • The conductivity of the rinse water is measured during final rinsing through a conductivity sensor (which is external to the TOC sensor) located on the drain line. The rinses are repeated automatically until the conductivity value is lower than a predetermined value included within the operating limits of the TOC analyzer. • At the end of this phase, the system for monitoring Total Organic Carbon is activated and a controlled volume of water from the final rinse is conveyed to the TOC sensor, opening an isolation valve. • The TOC sensor uses the oxidation induced by an ultraviolet lamp to determine the concentration of Total Organic Carbon that is contained in the sample of water from the final rinse. Integrated conductivity sensors provide real-time conductivity measurements before and after sample oxidation; the final value of the Total Organic Carbon derives from conductivity readings. • Once sampling has ended, the measured TOC value is transmitted to the process controller of the washing machine. The controller can be programmed to repeat the washing phase until a predefined TOC value or a maximum number of rinses is reached. If the preset TOC value is not reached within the maximum number of rinses, then the Process Controller reports an alarm. In this manner, the operator can apply the procedure required to solve the anomaly in progress. • As highlighted in the preceding application, the conductivity meter and the TOC sensor are fully managed by ‘Thema4’ process controller installed on Fedegari washing machines. While the conductivity measurement is instantaneous, the introduction of an online TOC analysis entails a longer cycle time. The washing system is in fact in a standby state during sample analysis in order to assess the need to perform further rinsing. Nevertheless, comparison between off-line and online measurements clearly shows that the online application of conductivity and TOC allows to eliminate manual sampling, the transport of samplestolaboratories,thedelayinlaboratorytime analysis, the "human variable" in the assessment of the results, and the risk of cross contamination. Online application allows continuous monitoring of process parameters in the reporting of any errors during a previously validated process. A schematic summary of the advantages/ disadvantages of the online application of conductivity and TOC is provided hereafter. Advantages of the online conductivity and TOC analysis methods • Faster analyses: elimination of manual sampling • Reduced risk of cross contamination: elimination of manual sampling • Reduced costs: no manual analysis in a laboratory • Reduction of inactivity periods: waiting for laboratory analysis • Analysis reliability : continuous monitoring of process parameters • Real-time "alarm system" • High sensitivity of the measurement Disadvantages of the online conductivity and TOC analysis methods • Nonspecific methods • Only for the TOC analyzer: initial cost of instrument FINAL CONSIDERATIONS Removal of bacterial endotoxins - The washing tests performed at Fedegari R&D department have proven to be effective in removing not only chemical but also microbiological soil. A more intense washing process is expected to yield a reduction of endotoxin load distinctly higher 6/7
  • 7. � Copyright © 2015 Fedegari Group fedegari.com than 3 log. A supplementary test, starting from a “chemically dirty” load, might constitute an additional challenge in the removal of inoculated bacterial endotoxins. Online analyses of Total Organic Carbon and conductivity are complementary for continuous monitoring of the washing process performances. The conductivity meter and the TOC sensor can be fully integrated in the hydraulic system of a washing machine and connected to the process controller, as in the case of Fedegari steam washers. The complete integration with the controller allows regulatory compliance to the requirements of the FDA 21 CFR part 11: the data related to each deviation from the standard required in terms of washing performance are recorded and stored in the system history. Through these applications, the process remains in a state of control during routine production as indicated by the FDA itself [6]. References 1. M.L. Bernuzzi. Fattori chiave nella scelta/ sviluppo di una strategia di lavaggio; 1a parte. ASCCA News, 2014; 4: 6-9. 2. Guide to Inspections of Validation of Cleaning Practices; U.S. Food and DrugAdministration: 1993. Available at: www.fda.gov/ICECI/ Inspections/InspectionGuides/ucm074922. htm (accessed February 24, 2015) 3. Technical Report No. 29: Points to Consider for Cleaning Validation; Parenteral Drug Association: Bethesda, MD, 1998. 4. USP 37 General Chapter <85> BACTERIAL ENDOTOXINS TEST, 2014. 5. European Pharmacopoeia 8.0. Bacterial Endotoxins, 01/2014:20614. 6. Guidance for Industry. Process Validation: General Principals and Practice; U.S. Food and Drug Administration: [on-line] 2011. Available at: www.fda.gov/downloads/ Drugs/Guidances/UCM070336.pdf‎ (accessed February 24, 2015) 1. 7/7