TELKOMNIKA Telecommunication, Computing, Electronics and Control
Vol. 18, No. 4, August 2020, pp. 2035~2041
ISSN: 1693-6930, accredited First Grade by Kemenristekdikti, Decree No: 21/E/KPT/2018
DOI: 10.12928/TELKOMNIKA.v18i4.14993  2035
Journal homepage: http://journal.uad.ac.id/index.php/TELKOMNIKA
Uric acid detection in visible spectrum
Afiqah Yaacob1
, Nor Hafizah Ngajikin2
, Nurfatihah Che Abd Rashid3
, Siti Hajar Aminah Ali4
,
Maslina Yaacob5
, Suhaila Isaak6
and Noran Azizan Cholan7
1,2,3,4,5,7
Faculty of Electrical and Electronic Engineering, Universiti Tun Hussein Onn Malaysia, Malaysia
6
School of Electrical Engineering, Faculty of Engineering, Universiti Teknologi Malaysia, Malaysia
Article Info ABSTRACT
Article history:
Received Dec 19, 2019
Revised Mar 19, 2020
Accepted Mar 30, 2020
The measurement of uric acid based on the optical absorption at visible light
spectrum is investigated and tested. Sensing in the visible region was
conducted for determination of suitable wavelength that produces high
sensitivity and accuracy performance based on the Beer-Lambert law
calculation. In this work, the uric acid is detected by detecting sodium urate
as a product of chemical reaction between uric acid with sodium hydroxide
buffer. The setup has been tested for uric acid concentration ranging from
15 mg/dL to 85 mg/dL. Three wavelengths have been analyzed which are
460 nm, 525 nm and 630 nm. Measured data at 460nm wavelength exhibits
the highest sensitivity, which is 0.0012 (mg/dL)-with 86.51% accuracy.
Detection of uric acid at visible light spectrum offers a low-cost sensor based
on visible LEDs and photodiode is possible to be realized.
Keywords:
Beer-Lambert law
Spectrophotometer
Uric acid
This is an open access article under the CC BY-SA license.
Corresponding Author:
Nor Hafizah Ngajikin,
Faculty of Electrical and Electronic Engineering,
Universiti Tun Hussein Onn Malaysia,
86400 Parit Raja, Batu Pahat, Johor, Malaysia.
Email: norhafizah@uthm.edu.my
1. INTRODUCTION
Uric acid sensor has received much attention in recent years due to the increasing number of gout
patients worldwide. Gout is a disease developed from the excess amount of uric acid present in blood
and urine. Uric acid dissolves in blood and travels to the kidney to be eliminated through urine. Excess
amount of uric acid will not only lead to gout, but it can also lead to kidney failure, heart disease, high
cholesterol, diabetes and even hypertension [1, 2]. It is important to regulate the amount uric acid in
maintaining the health of the bloodstream, since low amount of uric acid may lead to atherosclerosis
and stroke [2]. Uric acid can be monitored using two methods; blood test measurement [3-6] and urine test
measurement [1, 7-15]. A healthy range for uric acid levels in blood serum is from 3.5 to 7.2 mg/dL in men
and from 2.6 to 6 mg/dL for women [14]. Meanwhile, the healthy range of uric acid in human urine is ten
times greater than the uric acid levels in blood where it is ranging from 25 to 74 mg/dL [1, 12, 14]. In recent
years, there has been extensive research on uric acid analytical techniques to detect uric acid in human urine
such as electro-analytical, luminescence, chromatography and spectroscopy. As for spectroscopy system
[1, 7-10, 16-21], the system detects uric acid by using chemical reagent or buffer solution to the sample to be
tested by analyzing light absorbance or transmittance value. Spectroscopy shows a promising linear uric acid
detection in a broad range which is from 0.58 to 58.84 mg/dL [14] and 58.84 to 218.56 mg/dL [13]. Thus,
make it suitable for uric acid detection in both blood and urine.
 ISSN: 1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 4, August 2020: 2035 - 2041
2036
As afore mentioned, interference cause by other substance in the sample in the spectroscopy system
is minimize by adding a chemical reagent or buffer solution. In 2007, Yamaguchi has developed a simple and
highly sensitive spectrophotometric method for the determination of uric acid based on fading of
the o-hydroxyhydroquinonephthalein (QP)-palladium (II)-hexadecyltrimethylammonium complex in human
urine. The absorbance measurement has been conducted using a Shimadzu spectrophotometer with deuterium
and tungsten halogen lamp as the light source. This technique however has limitation in linearity range,
where the linearity occurs in between 0.001 mg/dL to 0.02 mg/dL at 635 nm visible operating
wavelength [18]. Later, a simple spectrophotometric method based-uricase enzyme for the detection of uric
acid in normal urine and gout patient’s urine samples has been developed based on the reaction of hydrogen
peroxide (H2O2) with yellow color of 4-Aminodiphenylamine Diazonium Sulfate (variamine blue RT salt) to
yield a pale yellow-green coloured solution at 269 nm wavelength im ultraviolet (UV) spectrum [8].
The calibration plot of different concentrations of uric acid was found to be linear between 9 mg/dL to
234 mg/dL with 20 minutes response time. Although broad sensing range has been achieved, the sensor
requires very long response time. Besides of using chemical reagent, enzymes or buffer, uric acid also can be
detected directly from the solution using spectrophotometer at 294.46 nm wavelength [19]. Although
the method is simple, the effect of interefence by other substance is not studied since there is no chemical
reagent was added. In this work, the uric acid detection at visible region was conducted by using NaOH
buffer solution without any additional chemical reagent. Detection of the uric acid in this work is carried out
by detecting sodium urate as a product of uric acid and NaOH mixing process.
2. THEORY
Analysis of sensitivity and accuracy in this work was calculated using Beer Lambert law.
The Beer-Lambert law defines the attenuation of light to the properties of the sample in terms of
transmittance, T and absorbance, A as stated in (1) and (2-5). In this work, the output light intensity after
passing through cuvette without sample, Iref, and with sample, Io, is measured by using spectrometer.
Configurations of a basic spectrophotometer system with the addition of input and output fiber is illustrated
in Figure 1.
(a) (b)
Figure 1. Spectrophotometer system (a) without sample and (b) with sample
The transmittance value based on (1) can be used to define the absorbance, A, of the sample.
The relationship between transmittance, T and absorbance, A is shown in (2) [20-22].
(1)
(2)
The (2) also relates absorbance, A, concentration of sample, c, sample path length, b, and the absorptivity
of the sample, ɛ. This relationship is known as the Beer-Lambert law. Based on (2), the concentration of
a substance is directly proportional to the amount of light absorbed or inversely proportional to the logarithm
of the transmitted light [23, 24]. The relationship between transmittance, T and absorbance, A is illustrated
in Figure 2.
bc
T
A 
1
log
Iref
Io
T 
TELKOMNIKA Telecommun Comput El Control 
Uric acid detection in visible spectrum (Afiqah Yaacob)
2037
Analytical analysis is done by analyzing the calibration curve which consists of a plot of absorbance
versus concentration series of standard solution [20]. If the curve-fit is linear, the sensitivity of the curve fit is
its slope [25]. Thus, sensitivity is given as in (3).
[(mg/dL)-1
] (3)
is the absorbance difference and is concentration difference. The (3) shows that for a fixed
concentration difference, higher absorbance will contribute to a higher sensitivity of spectrophotometer [26].
The Standard unit for sensitivity of the tested sensor is dependent on the unit used for the sample
concentration. Thus, in this paper, the unit used for sensitivity is (mg/dL)-1
. As for calibration accuracy,
this parameter refers to how close the measured value, Cmeasured with the linear-fit real value, Creal [25].
The (4) shows equation used for accuracy calculation based on the data from absorbance plot in Figure 3.
The following section will describe procedure for sample preparation and experiment that was conducted.
(4)
Figure 2. Graph of absorbance versus
transmittance [22]
Figure 3. Accuracy measurement from absorbance plot
3. EXPERIMENT
Uric acid solution is prepared by mixing uric acid powder into sodium hydroxide (NaOH)
solution [4, 7, 24]. In this research, uric acid and sodium hydroxide solution was purchased from Axon
Scientific. The sample is prepared by mixing uric acid powder in 0.1 M NaOH solution as shown
in Figure 4 (a) and (b). Table 1 shows the amount of uric acid powder that needs to be added to 80 mL of NaOH
solution for producing uric acid sample with 15 mg/dL, 25 mg/dL, 50 mg/dL, 75 mg/dL and 85 mg/dL
concentration. The sample preparation process was carried out at room temperature. Mixture of uric acid
powder and NaOH solution will produce sodium urate and H2O [24]. Higher concentration of uric acid in
NaOH solution will form a solid sodium urate as shown in Figure 4 (c). Therefore, the mixture needs to be
stirred for about 3 minutes before being transferred into the cuvette for fully dilution process as visualized in
Figure 4 (d). In this measurement, 1.5 mL of sample were transferred into each cuvette for the characterization
process as in Figure 4 (e). Figure 4 illustrates the steps taken for standard uric acid sample preparation.
Experiment was carried out by using tungsten-halogen as the light source (1) and spectrometer as
optical detector. In this experiment, variable optical attenuator (2) is used to ensure intensity of light is not
saturated at the spectrometer. The maximum photon count is limit to a maximum peak of 14000 photon count.
Light from optical fibre is collimated into parallel beam using Ocean Optics 74-UV collimating
lens (3) to interact with solution in cuvette (4) as shown in Figure 5. Then, light from cuvette is collected using
the collimating lens and delivered to Ocean Optics HR4000CG-UV-NIR high resolution spectrometer (5).
Ocean Optics OceanView spectrometer operating software is used to obtain and process data from spectrometer.
Uric acid concentration is calculated by comparing intensity of light that passes through sample, Io and does not
pass through sample, Iref. Wavelengths selected to sample uric acid are 460 nm, 525 nm and 630 nm based on
available visible LED in the market.
c
A
ySensitivit



A c
1001 






 

real
measuredreal
C
CC
Accuracy [%]
 ISSN: 1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 4, August 2020: 2035 - 2041
2038
Figure 4. Preparation of standard uric acid sample; (a) uric acid powder (b) NaOH solution
(c) sample after adding NaOH with uric acid producing particles of sodium urate
(d) sample after stirred using glass rod (e) transfer 1.5 ml into cuvette for measurement
Table 1. Amount of uric acid powder added into 80 mL of NaOH solution
Concentration (mg/dL) Uric Acid powder (mg)
15 12
25 20
50 40
75 60
85 68
Figure 5. Experimental setup
4. RESULTS AND DISCUSSION
Figure 6 visualized the intensity spectra observed using HR4000CG-UV-VIS Ocean Optics
spectrometer for SLS201 Stabilized Tungsten Halogen light operated at 400 nm to 700 nm wavelength range.
This figure shows that light intensity will decrease as the concentration of uric acid increases, in which obey
the Beer-Lambert law. The relationship between light intensity with absorbance is associated using (1) and (2).
Calculated absorbance is plotted as in Figure 7 while calculated sensitivity and accuracy is tabulated in Table 2.
Table 2 shows that the highest sensitivity is at 460 nm wavelength which is 0.0012 (mg/dL)-1
.
Although all tested wavelength demonstrate a similar pattern of absorbance due to substance in the sample,
amount of the light absorb is difference, thus exhibit difference sensitivity performance. For accuracy
performance, 525 nm wavelength spectrophotometer system produces accuracy above 90%. The result agrees
well with what has been reported in the previous literature where less error will occur when the system has
absorbance value within the recommended range, which are between 0.2 and 0.8 [22].
TELKOMNIKA Telecommun Comput El Control 
Uric acid detection in visible spectrum (Afiqah Yaacob)
2039
Figure 6. Intensity spectra of different uric acid concentration
Figure 7. Absorbance analytical curve
Table 2. Spectrophotometer performance for selected tested wavelength
Wavelength (nm) Sensitivity (mg/dL)-1
Accuracy (%)
460 0.0012 86.51
525 0.0011 91.17
630 0.0008 84.27
5. CONCLUSION
The research study was carried out to analyze sensitivity and accuracy performance of
spectrophotometer system at different sampled wavelength using absorbance analytical technique.
The absorption was calculated using Beer’s Lambert law formula. Through the analysis, the spectrometer
was able to observe the current concentration of uric acid and its absorption wavelength at visible light
region. In this work, spectrophotometer system operated at 460nm wavelength has the highest sensitivity
while only system at 525 nm wavelength has accuracy higher than 90%. As discussed, the accuracy is
influenced by limitation of Beer-Lambert law where absorbance value beyond 0.2 and 0.8 will have large
error. The spectrophotometer system with visible light source will offer a simple and economical uric acid
detection system and can be applied in biomedical application.
ACKNOWLEDGEMENTS
The authors are grateful to Universiti Tun Hussein Onn Malaysia (UTHM) for supporting this
research work under Postgraduate Research Grant (GPPS), grant no: H408 and Research University Grant
(RUG) TIER 1 Scheme, grant no: H162.
REFERENCES
[1] Azmi N. E., et al., “A simple and sensitive fluorescence based biosensor for the determination of uric acid using H2
O 2-sensitive quantum dots/dual enzymes,” Biosensors and Bioelectronics, vol. 67, pp. 129-133, 2015.
[2] Pormsila W., S. Krähenbühl, P. C. Hauser, “Capillary electrophoresis with contactless conductivity detection for
uric acid determination in biological fluids,” Analytica Chimica Acta, vol. 636, no. 2, pp. 224-228, 2009.
 ISSN: 1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 4, August 2020: 2035 - 2041
2040
[3] Zhuang Q. Q., et al., “Peroxidase-like activity of nanocrystalline cobalt selenide and its application for uric acid
detection,” International Journal of Nanomedicine, vol. 12, pp. 3295-3302, 2017.
[4] Boroumand S., M. A. Chamjangali, G. Bagherian, “Double injection/single detection asymmetric flow injection
manifold for spectrophotometric determination of ascorbic acid and uric acid: Selection the optimal conditions by
MCDM approach based on different criteria weighting methods,” Spectrochimica Acta Part A: Molecular and
Biomolecular Spectroscopy, vol. 174, pp. 203-213, 2017.
[5] Liao L. T., et al., “Evaluation of an electrochemical biosensor for uric acid measurement in human whole blood
samples,” Clin Chim Acta, vol. 436, pp. 72-77, 2014.
[6] Pleskacova A., et al., “Simultaneous Determination of Uric Acid, Xanthine and Hypoxanthine in Human Plasma
and Serum by HPLC–UV: Uric Acid Metabolism Tracking,” Chromatographia, vol. 80, no. 4, pp. 529-536, 2017.
[7] Rocha D. L., Rocha F. R., “A flow-based procedure with solenoid micro-pumps for the spectrophotometric
determination of uric acid in urine,” Microchemical Journal, vol. 94, no. 1, pp. 53-59, 2010.
[8] Hamzah H. H., et al., “Spectrophotometric Determination of Uric Acid in Urine Based-Enzymatic Method Uricase
with 4-Aminodiphenylamine Diazonium Sulfate (Variamine Blue RT Salt),” Journal of Analytical & Bioanalytical
Techniques, vol. S7, no. 011, 2013.
[9] Villa J. E., Poppi R. J., “A portable SERS method for the determination of uric acid using a paper-based substrate
and multivariate curve resolution,” Analyst, vol. 141, no. 6, pp. 1966-1972, 2016.
[10] Jin D., et al., “Quantitative determination of uric acid using CdTe nanoparticles as fluorescence probes,”
Biosensors and Bioelectronics, vol. 77, pp. 359-365, 2016.
[11] Yu J., et al., “A novel chemiluminescence paper microfluidic biosensor based on enzymatic reaction for uric acid
determination,” Biosensors and Bioelectronics, vol. 26, no. 7, pp. 3284-3289, 2011.
[12] Ballesta-Claver J., et al., “Disposable biosensor based on cathodic electrochemiluminescence of tris (2,2-bipyridine)
ruthenium (II) for uric acid determination,” Analytica Chimica Acta, vol. 770, pp. 153-160, 2013.
[13] Pavlíček V., et al., “Very fast electrophoretic determination of creatinine and uric acid in human urine using
a combination of two capillaries with different internal diameters,” Electrophoresis, vol. 35, no. 7, pp. 956-961, 2014.
[14] Fanjul-Bolado P., et al., “Uric Acid Determination by Adsorptive Stripping Voltammetry on Multiwall Carbon
Nanotubes Based Screen-Printed Electrodes,” Electroanalysis, vol. 27, no. 5, pp. 1276-1281, 2015.
[15] Langsi V. K., et al., “Synthesis and characterisation of non-bonded 1.7 μm thin-shell (TS1. 7–100nm) silica
particles for the rapid separation and analysis of uric acid and creatinine in human urine by hydrophilic interaction
chromatography,” Journal of Chromatography A, vol. 1506, pp. 37-44, 2017.
[16] Khajehsharifi H., et al., “The comparison of partial least squares and principal component regression in
simultaneous spectrophotometric determination of ascorbic acid, dopamine and uric acid in real samples,” Arabian
Journal of Chemistry, vol. 4, no. 2, pp. S3451-S3458, 2014.
[17] Martinez-Pérez D., et al., “A reagent less fluorescent sol–gel biosensor for uric acid detection in biological fluids,”
Analytical Biochemistry, vol. 322, no. 2, pp. 238-242, 2003.
[18] Yamaguchi T., et al., “Spectrophotometric determination of uric acid based on fading of o-hydroxyhydroquinonephthalein-
palladium (II)-hexadecyltrimethyl-ammonium complex,” Analytical Sciences, vol. 23, no. 2, pp. 223-226, 2007.
[19] Norazmi N., et al., “Uric acid detection using uv-vis spectrometer,” IOP Conf. Series: Materials Science and
Engineering, 2017.
[20] Hardesty J. H., B. Attili, “Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in
Chemistry,” Collin, 2010.
[21] Rashid N. C. A., et al., “Spectrophotometer with Enhanced Sensitivity for Uric Acid Detection,” Chinese Optics
Letter, vol 17, no. 8, pp. 1-5, 2019.
[22] Marcus T. C., et al., “Alternative wavelength for linearity preservation of Beer–Lambert Law in ozone
concentration measurement,” Microwave and Optical Technology Letters, vol. 57, no. 4, pp. 1013-1016, 2015.
[23] Turgeon M. L., “Linne & Ringsrud's Clinical Laboratory Science - E-Book: The Basics and Routine Techniques,”
Elsevier Health Sciences, 2015.
[24] Teque N., “Uric Acid,” 2018, [Online]. Available: https://www.nacalai.co.jp/ss/ec/ECsrchdetl.cfm?Dum=1&
syohin=3591592&syubetsu=3
[25] Dunn P. F., “Measurement and Data Analysis for Engineering and Science,” Third Edition, CRC Press, 2014.
[26] Welz B., et al., “High-Resolution Continuum Source AAS: The Better Way to Do Atomic Absorption
Spectrometry,” Wiley, 2006.
BIOGRAPHIES OF AUTHORS
Afiqah Yaacob received her Bachelor degree from Universiti Teknologi Mara (UiTM).
Currently, she is pursuing her Masters degree in Faculty of Electrical and Electronic
Engineering, Universiti Tun Hussein Onn Malaysia (UTHM) in the field of
spectrophotometer application.
TELKOMNIKA Telecommun Comput El Control 
Uric acid detection in visible spectrum (Afiqah Yaacob)
2041
Nor Hafizah Ngajikin received her B.E and M.E in Electronic Engineering from Universiti
Teknologi Malaysia (UTM) in 2001 and 2003 respectively. She was awarded a Ph.D from
UTM for her work on MEMS Fabry-Perot optical tunable filter. From 2004-2017,
she served as a senior lecturer and researcher at Lightwave Communication Research Group
(LCRG), UTM. She is currently a lecturer at Universiti Tun Hussein Onn Malaysia.
Her research interest includes an optical devices and sensors for biomedical applications.
Nurfatihah Che Abd Rashid received her B.E and M.E in Electrical Engineering from
Universiti Teknologi Malaysia (UTM) in 2015 and 2019 respectively. She has involved in
research on spectrophotometric detection for uric acid and soil nutrient applications. She is
currently pursuing a PhD degree in Faculty of Electrical and Electronic Engineering,
Universiti Tun Hussein Onn Malaysia (UTHM).
Siti Hajar Aminah Ali is a lecturer at Universiti Tun Hussein Onn Malaysia (UTHM). Her
work focuses on image processing and neural network. She received her Bachelor of
Electrical-Telecommunications Engineering and Master of Electrical-Electronics and
Telecommunications from Universiti Teknologi Malaysia (UTM), Skudai, Malaysia.
Her highest education is Doctor of Philosophy in Engineering from Kobe University, Japan.
Maslina Yaacob was awarded a Ph.D from Universiti Teknologi Malaysia (UTM) for her
work on Wide Range Analysis of Ozone Gas Concentration in Ultraviolet Region. Her B.E
and M.E in Electrical Engineering were also awarded from UTM in 2007 and 2010,
respectively. Her research interest includes an optical devices and optical sensors for gas
sensing applications. She is currently a lecturer and researcher at Universiti Tun Hussein
Onn Malaysia.
Suhaila Isaak received her Bachelor Degree in Electrical and Electronic Engineering from
Universiti Teknologi Malaysia in 1998. In 2001, she completed her Master in Science
(Electronic) from Universiti Putra Malaysia. She was awarded the Ph.D degree in Electrical
and Electronics Engineering from University of Nottingham, England in 2011. Her major
research interests are high-speed Geiger mode avalanche photodiode, analog integrated
CMOS design, VLSI circuit, high speed photon counting on FPGA, and photonic
experimental analysis on biological prototype system in visible light range. Currently,
she has involved on integrated linear array photon camera as spectroscopy system for
macronutrient soil analysis.
Noran Azizan Cholan was born on 31st August 1979 in Segamat, Johor, Malaysia.
He received his bachelor degree in Electronics Engineering from Universiti Tenaga
Nasional Malaysia in 2002. Afterwards in 2004, he obtained his master degree in
Electronics-Telecommunications Engineering form Universiti Teknologi Malaysia. In 2010,
he enrolled as a PhD student in Universiti Putra Malaysia. During his PhD study in 2012,
he went to Swansea University, UK and The Hong Kong Polytechnic University, Hong Kong
for research attachment. He completed his PhD in 2014 before serving as a senior lecturer in
Universiti Tun Hussein Onn Malaysia. As of now, he has been authors/co-authors for 25
journal and 16 conference proceeding papers. His research interests include lasers,
amplifiers, optics modeling, microwave optics and optical sensors.

More Related Content

What's hot

Prediction of ammonia concentration in water based on microwave spectroscopy
Prediction of ammonia concentration in water based on microwave spectroscopyPrediction of ammonia concentration in water based on microwave spectroscopy
Prediction of ammonia concentration in water based on microwave spectroscopyjournalBEEI
 
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...Ninad Phadke
 
Graphene/PSMA composite for gas sensing application
Graphene/PSMA composite for gas sensing applicationGraphene/PSMA composite for gas sensing application
Graphene/PSMA composite for gas sensing applicationjournalBEEI
 
Primary Odor on Consideration of Reducing the Number of Compositions
Primary Odor on Consideration of Reducing the Number of CompositionsPrimary Odor on Consideration of Reducing the Number of Compositions
Primary Odor on Consideration of Reducing the Number of Compositionsinventionjournals
 
An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...
An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...
An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...Fontem Ventures
 
Mdo2014 takeda
Mdo2014 takedaMdo2014 takeda
Mdo2014 takedaEd Tessier
 
Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Arun kumar
 
Co s mos 2014_ruchi
Co s mos 2014_ruchiCo s mos 2014_ruchi
Co s mos 2014_ruchiEd Tessier
 
Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...
Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...
Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...Fontem Ventures
 
Nicotine Retention Following Use of E-Cigarettes
Nicotine Retention Following Use of E-CigarettesNicotine Retention Following Use of E-Cigarettes
Nicotine Retention Following Use of E-CigarettesFontem Ventures
 
Gas chromatography sag2020
Gas chromatography  sag2020Gas chromatography  sag2020
Gas chromatography sag2020SuchetaGaikwad3
 
Application of nanomaterials in lifescience
Application of nanomaterials in lifescienceApplication of nanomaterials in lifescience
Application of nanomaterials in lifescienceArun kumar
 

What's hot (20)

Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...
A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...
A Comparative Analysis of Semiconductor Electroplating Bath Additives by Cali...
 
Prediction of ammonia concentration in water based on microwave spectroscopy
Prediction of ammonia concentration in water based on microwave spectroscopyPrediction of ammonia concentration in water based on microwave spectroscopy
Prediction of ammonia concentration in water based on microwave spectroscopy
 
WorldComp2016_Jenny
WorldComp2016_JennyWorldComp2016_Jenny
WorldComp2016_Jenny
 
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
 
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
Chromatography: Complete Inorganic Elemental Speciation Analysis Solutions fo...
 
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET  By Ninad M...
SPECTROSCOPIC METHOD DEVELOPMENT FOR LOSARTON POTASSIUM IN TABLET By Ninad M...
 
Graphene/PSMA composite for gas sensing application
Graphene/PSMA composite for gas sensing applicationGraphene/PSMA composite for gas sensing application
Graphene/PSMA composite for gas sensing application
 
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
 
Primary Odor on Consideration of Reducing the Number of Compositions
Primary Odor on Consideration of Reducing the Number of CompositionsPrimary Odor on Consideration of Reducing the Number of Compositions
Primary Odor on Consideration of Reducing the Number of Compositions
 
An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...
An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...
An Experimental Method to Determine the Concentration of Nicotine in Exhaled ...
 
Mdo2014 takeda
Mdo2014 takedaMdo2014 takeda
Mdo2014 takeda
 
Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...
 
Ppt spectroscopy
Ppt spectroscopyPpt spectroscopy
Ppt spectroscopy
 
Co s mos 2014_ruchi
Co s mos 2014_ruchiCo s mos 2014_ruchi
Co s mos 2014_ruchi
 
Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...
Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...
Electronic Cigarettes and Indoor Air Quality: A Simple Approach to Modeling P...
 
Nicotine Retention Following Use of E-Cigarettes
Nicotine Retention Following Use of E-CigarettesNicotine Retention Following Use of E-Cigarettes
Nicotine Retention Following Use of E-Cigarettes
 
Gas chromatography sag2020
Gas chromatography  sag2020Gas chromatography  sag2020
Gas chromatography sag2020
 
Application of nanomaterials in lifescience
Application of nanomaterials in lifescienceApplication of nanomaterials in lifescience
Application of nanomaterials in lifescience
 
LCMS
LCMSLCMS
LCMS
 

Similar to Uric acid detection using visible light spectroscopy

Optimization of light source wavelength for ammonia detection in water
Optimization of light source wavelength for ammonia detection in waterOptimization of light source wavelength for ammonia detection in water
Optimization of light source wavelength for ammonia detection in waterTELKOMNIKA JOURNAL
 
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...pharmaindexing
 
Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...IJSIT Editor
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...TELKOMNIKA JOURNAL
 
Simple design flow injection PMMA acrylic sample cell for nitrite determination
Simple design flow injection PMMA acrylic sample cell for nitrite determinationSimple design flow injection PMMA acrylic sample cell for nitrite determination
Simple design flow injection PMMA acrylic sample cell for nitrite determinationHarry Ramza
 
Difference & derivative spectrometry
Difference & derivative spectrometryDifference & derivative spectrometry
Difference & derivative spectrometryaishuanju
 
Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...BRNSS Publication Hub
 
Determination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex ReactionDetermination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex ReactionRatnakaram Venkata Nadh
 
Determination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex ReactionDetermination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex ReactionRatnakaram Venkata Nadh
 
Effects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurementEffects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurementTELKOMNIKA JOURNAL
 
Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...Ratnakaram Venkata Nadh
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsSCIEX
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method IJECEIAES
 
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...IRJET Journal
 
Applicationofu v-spectroscopy-120416145659-phpapp02
Applicationofu v-spectroscopy-120416145659-phpapp02Applicationofu v-spectroscopy-120416145659-phpapp02
Applicationofu v-spectroscopy-120416145659-phpapp02Kirsha K S
 
applicationofu-v-spectroscopy-120416145659-phpapp02
applicationofu-v-spectroscopy-120416145659-phpapp02applicationofu-v-spectroscopy-120416145659-phpapp02
applicationofu-v-spectroscopy-120416145659-phpapp02Kirsha K S
 
Low-cost colorimetric setup for concentration measurement of manganese ions b...
Low-cost colorimetric setup for concentration measurement of manganese ions b...Low-cost colorimetric setup for concentration measurement of manganese ions b...
Low-cost colorimetric setup for concentration measurement of manganese ions b...IJECEIAES
 
FT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturingFT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturingGalaxy Scientific
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightIRJET Journal
 

Similar to Uric acid detection using visible light spectroscopy (20)

Optimization of light source wavelength for ammonia detection in water
Optimization of light source wavelength for ammonia detection in waterOptimization of light source wavelength for ammonia detection in water
Optimization of light source wavelength for ammonia detection in water
 
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
 
Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...Development and validation of hplc method for determination of theophylline a...
Development and validation of hplc method for determination of theophylline a...
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
 
Simple design flow injection PMMA acrylic sample cell for nitrite determination
Simple design flow injection PMMA acrylic sample cell for nitrite determinationSimple design flow injection PMMA acrylic sample cell for nitrite determination
Simple design flow injection PMMA acrylic sample cell for nitrite determination
 
Difference & derivative spectrometry
Difference & derivative spectrometryDifference & derivative spectrometry
Difference & derivative spectrometry
 
Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...Validated High-performance Liquid Chromatography Method for Degradation Study...
Validated High-performance Liquid Chromatography Method for Degradation Study...
 
Determination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex ReactionDetermination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex Reaction
 
Determination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex ReactionDetermination of Satranidazole through Ion-Associative Complex Reaction
Determination of Satranidazole through Ion-Associative Complex Reaction
 
Effects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurementEffects of noises on near infrared sensor for blood glucose level measurement
Effects of noises on near infrared sensor for blood glucose level measurement
 
Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...Quality-by-design-based development and validation of a stability-indicating ...
Quality-by-design-based development and validation of a stability-indicating ...
 
LC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminantsLC-MS/MS analysis of emerging food contaminants
LC-MS/MS analysis of emerging food contaminants
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method
 
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
 
SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...
SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...
SPECTROPHOTOMETRIC MEASUREMENT OF SOLUBILITY TEST TURBIDITY AS AN IMPROVED DI...
 
Applicationofu v-spectroscopy-120416145659-phpapp02
Applicationofu v-spectroscopy-120416145659-phpapp02Applicationofu v-spectroscopy-120416145659-phpapp02
Applicationofu v-spectroscopy-120416145659-phpapp02
 
applicationofu-v-spectroscopy-120416145659-phpapp02
applicationofu-v-spectroscopy-120416145659-phpapp02applicationofu-v-spectroscopy-120416145659-phpapp02
applicationofu-v-spectroscopy-120416145659-phpapp02
 
Low-cost colorimetric setup for concentration measurement of manganese ions b...
Low-cost colorimetric setup for concentration measurement of manganese ions b...Low-cost colorimetric setup for concentration measurement of manganese ions b...
Low-cost colorimetric setup for concentration measurement of manganese ions b...
 
FT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturingFT-NIR as a real-time QC tool for polymer manufacturing
FT-NIR as a real-time QC tool for polymer manufacturing
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
 

More from TELKOMNIKA JOURNAL

Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...TELKOMNIKA JOURNAL
 
Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...TELKOMNIKA JOURNAL
 
Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...TELKOMNIKA JOURNAL
 
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...TELKOMNIKA JOURNAL
 
Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...TELKOMNIKA JOURNAL
 
Efficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antennaEfficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antennaTELKOMNIKA JOURNAL
 
Design and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fireDesign and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fireTELKOMNIKA JOURNAL
 
Wavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio networkWavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio networkTELKOMNIKA JOURNAL
 
A novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bandsA novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bandsTELKOMNIKA JOURNAL
 
Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...TELKOMNIKA JOURNAL
 
Brief note on match and miss-match uncertainties
Brief note on match and miss-match uncertaintiesBrief note on match and miss-match uncertainties
Brief note on match and miss-match uncertaintiesTELKOMNIKA JOURNAL
 
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...TELKOMNIKA JOURNAL
 
Evaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G systemEvaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G systemTELKOMNIKA JOURNAL
 
Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...TELKOMNIKA JOURNAL
 
Reagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensorReagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensorTELKOMNIKA JOURNAL
 
Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...TELKOMNIKA JOURNAL
 
A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...TELKOMNIKA JOURNAL
 
Electroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networksElectroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networksTELKOMNIKA JOURNAL
 
Adaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imagingAdaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imagingTELKOMNIKA JOURNAL
 
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...TELKOMNIKA JOURNAL
 

More from TELKOMNIKA JOURNAL (20)

Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...
 
Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...
 
Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...
 
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
 
Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...
 
Efficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antennaEfficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antenna
 
Design and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fireDesign and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fire
 
Wavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio networkWavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio network
 
A novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bandsA novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bands
 
Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...
 
Brief note on match and miss-match uncertainties
Brief note on match and miss-match uncertaintiesBrief note on match and miss-match uncertainties
Brief note on match and miss-match uncertainties
 
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
 
Evaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G systemEvaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G system
 
Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...
 
Reagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensorReagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensor
 
Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...
 
A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...
 
Electroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networksElectroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networks
 
Adaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imagingAdaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imaging
 
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
 

Recently uploaded

the ladakh protest in leh ladakh 2024 sonam wangchuk.pptx
the ladakh protest in leh ladakh 2024 sonam wangchuk.pptxthe ladakh protest in leh ladakh 2024 sonam wangchuk.pptx
the ladakh protest in leh ladakh 2024 sonam wangchuk.pptxhumanexperienceaaa
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...ranjana rawat
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Call Girls in Nagpur High Profile
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130Suhani Kapoor
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxupamatechverse
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVRajaP95
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSKurinjimalarL3
 
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130Suhani Kapoor
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxpranjaldaimarysona
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSISrknatarajan
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 

Recently uploaded (20)

the ladakh protest in leh ladakh 2024 sonam wangchuk.pptx
the ladakh protest in leh ladakh 2024 sonam wangchuk.pptxthe ladakh protest in leh ladakh 2024 sonam wangchuk.pptx
the ladakh protest in leh ladakh 2024 sonam wangchuk.pptx
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
 
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
Call Girls in Nagpur Suman Call 7001035870 Meet With Nagpur Escorts
 
Introduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptxIntroduction to IEEE STANDARDS and its different types.pptx
Introduction to IEEE STANDARDS and its different types.pptx
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
 
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICSAPPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
APPLICATIONS-AC/DC DRIVES-OPERATING CHARACTERISTICS
 
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
VIP Call Girls Service Hitech City Hyderabad Call +91-8250192130
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptx
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSIS
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 

Uric acid detection using visible light spectroscopy

  • 1. TELKOMNIKA Telecommunication, Computing, Electronics and Control Vol. 18, No. 4, August 2020, pp. 2035~2041 ISSN: 1693-6930, accredited First Grade by Kemenristekdikti, Decree No: 21/E/KPT/2018 DOI: 10.12928/TELKOMNIKA.v18i4.14993  2035 Journal homepage: http://journal.uad.ac.id/index.php/TELKOMNIKA Uric acid detection in visible spectrum Afiqah Yaacob1 , Nor Hafizah Ngajikin2 , Nurfatihah Che Abd Rashid3 , Siti Hajar Aminah Ali4 , Maslina Yaacob5 , Suhaila Isaak6 and Noran Azizan Cholan7 1,2,3,4,5,7 Faculty of Electrical and Electronic Engineering, Universiti Tun Hussein Onn Malaysia, Malaysia 6 School of Electrical Engineering, Faculty of Engineering, Universiti Teknologi Malaysia, Malaysia Article Info ABSTRACT Article history: Received Dec 19, 2019 Revised Mar 19, 2020 Accepted Mar 30, 2020 The measurement of uric acid based on the optical absorption at visible light spectrum is investigated and tested. Sensing in the visible region was conducted for determination of suitable wavelength that produces high sensitivity and accuracy performance based on the Beer-Lambert law calculation. In this work, the uric acid is detected by detecting sodium urate as a product of chemical reaction between uric acid with sodium hydroxide buffer. The setup has been tested for uric acid concentration ranging from 15 mg/dL to 85 mg/dL. Three wavelengths have been analyzed which are 460 nm, 525 nm and 630 nm. Measured data at 460nm wavelength exhibits the highest sensitivity, which is 0.0012 (mg/dL)-with 86.51% accuracy. Detection of uric acid at visible light spectrum offers a low-cost sensor based on visible LEDs and photodiode is possible to be realized. Keywords: Beer-Lambert law Spectrophotometer Uric acid This is an open access article under the CC BY-SA license. Corresponding Author: Nor Hafizah Ngajikin, Faculty of Electrical and Electronic Engineering, Universiti Tun Hussein Onn Malaysia, 86400 Parit Raja, Batu Pahat, Johor, Malaysia. Email: norhafizah@uthm.edu.my 1. INTRODUCTION Uric acid sensor has received much attention in recent years due to the increasing number of gout patients worldwide. Gout is a disease developed from the excess amount of uric acid present in blood and urine. Uric acid dissolves in blood and travels to the kidney to be eliminated through urine. Excess amount of uric acid will not only lead to gout, but it can also lead to kidney failure, heart disease, high cholesterol, diabetes and even hypertension [1, 2]. It is important to regulate the amount uric acid in maintaining the health of the bloodstream, since low amount of uric acid may lead to atherosclerosis and stroke [2]. Uric acid can be monitored using two methods; blood test measurement [3-6] and urine test measurement [1, 7-15]. A healthy range for uric acid levels in blood serum is from 3.5 to 7.2 mg/dL in men and from 2.6 to 6 mg/dL for women [14]. Meanwhile, the healthy range of uric acid in human urine is ten times greater than the uric acid levels in blood where it is ranging from 25 to 74 mg/dL [1, 12, 14]. In recent years, there has been extensive research on uric acid analytical techniques to detect uric acid in human urine such as electro-analytical, luminescence, chromatography and spectroscopy. As for spectroscopy system [1, 7-10, 16-21], the system detects uric acid by using chemical reagent or buffer solution to the sample to be tested by analyzing light absorbance or transmittance value. Spectroscopy shows a promising linear uric acid detection in a broad range which is from 0.58 to 58.84 mg/dL [14] and 58.84 to 218.56 mg/dL [13]. Thus, make it suitable for uric acid detection in both blood and urine.
  • 2.  ISSN: 1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 4, August 2020: 2035 - 2041 2036 As afore mentioned, interference cause by other substance in the sample in the spectroscopy system is minimize by adding a chemical reagent or buffer solution. In 2007, Yamaguchi has developed a simple and highly sensitive spectrophotometric method for the determination of uric acid based on fading of the o-hydroxyhydroquinonephthalein (QP)-palladium (II)-hexadecyltrimethylammonium complex in human urine. The absorbance measurement has been conducted using a Shimadzu spectrophotometer with deuterium and tungsten halogen lamp as the light source. This technique however has limitation in linearity range, where the linearity occurs in between 0.001 mg/dL to 0.02 mg/dL at 635 nm visible operating wavelength [18]. Later, a simple spectrophotometric method based-uricase enzyme for the detection of uric acid in normal urine and gout patient’s urine samples has been developed based on the reaction of hydrogen peroxide (H2O2) with yellow color of 4-Aminodiphenylamine Diazonium Sulfate (variamine blue RT salt) to yield a pale yellow-green coloured solution at 269 nm wavelength im ultraviolet (UV) spectrum [8]. The calibration plot of different concentrations of uric acid was found to be linear between 9 mg/dL to 234 mg/dL with 20 minutes response time. Although broad sensing range has been achieved, the sensor requires very long response time. Besides of using chemical reagent, enzymes or buffer, uric acid also can be detected directly from the solution using spectrophotometer at 294.46 nm wavelength [19]. Although the method is simple, the effect of interefence by other substance is not studied since there is no chemical reagent was added. In this work, the uric acid detection at visible region was conducted by using NaOH buffer solution without any additional chemical reagent. Detection of the uric acid in this work is carried out by detecting sodium urate as a product of uric acid and NaOH mixing process. 2. THEORY Analysis of sensitivity and accuracy in this work was calculated using Beer Lambert law. The Beer-Lambert law defines the attenuation of light to the properties of the sample in terms of transmittance, T and absorbance, A as stated in (1) and (2-5). In this work, the output light intensity after passing through cuvette without sample, Iref, and with sample, Io, is measured by using spectrometer. Configurations of a basic spectrophotometer system with the addition of input and output fiber is illustrated in Figure 1. (a) (b) Figure 1. Spectrophotometer system (a) without sample and (b) with sample The transmittance value based on (1) can be used to define the absorbance, A, of the sample. The relationship between transmittance, T and absorbance, A is shown in (2) [20-22]. (1) (2) The (2) also relates absorbance, A, concentration of sample, c, sample path length, b, and the absorptivity of the sample, ɛ. This relationship is known as the Beer-Lambert law. Based on (2), the concentration of a substance is directly proportional to the amount of light absorbed or inversely proportional to the logarithm of the transmitted light [23, 24]. The relationship between transmittance, T and absorbance, A is illustrated in Figure 2. bc T A  1 log Iref Io T 
  • 3. TELKOMNIKA Telecommun Comput El Control  Uric acid detection in visible spectrum (Afiqah Yaacob) 2037 Analytical analysis is done by analyzing the calibration curve which consists of a plot of absorbance versus concentration series of standard solution [20]. If the curve-fit is linear, the sensitivity of the curve fit is its slope [25]. Thus, sensitivity is given as in (3). [(mg/dL)-1 ] (3) is the absorbance difference and is concentration difference. The (3) shows that for a fixed concentration difference, higher absorbance will contribute to a higher sensitivity of spectrophotometer [26]. The Standard unit for sensitivity of the tested sensor is dependent on the unit used for the sample concentration. Thus, in this paper, the unit used for sensitivity is (mg/dL)-1 . As for calibration accuracy, this parameter refers to how close the measured value, Cmeasured with the linear-fit real value, Creal [25]. The (4) shows equation used for accuracy calculation based on the data from absorbance plot in Figure 3. The following section will describe procedure for sample preparation and experiment that was conducted. (4) Figure 2. Graph of absorbance versus transmittance [22] Figure 3. Accuracy measurement from absorbance plot 3. EXPERIMENT Uric acid solution is prepared by mixing uric acid powder into sodium hydroxide (NaOH) solution [4, 7, 24]. In this research, uric acid and sodium hydroxide solution was purchased from Axon Scientific. The sample is prepared by mixing uric acid powder in 0.1 M NaOH solution as shown in Figure 4 (a) and (b). Table 1 shows the amount of uric acid powder that needs to be added to 80 mL of NaOH solution for producing uric acid sample with 15 mg/dL, 25 mg/dL, 50 mg/dL, 75 mg/dL and 85 mg/dL concentration. The sample preparation process was carried out at room temperature. Mixture of uric acid powder and NaOH solution will produce sodium urate and H2O [24]. Higher concentration of uric acid in NaOH solution will form a solid sodium urate as shown in Figure 4 (c). Therefore, the mixture needs to be stirred for about 3 minutes before being transferred into the cuvette for fully dilution process as visualized in Figure 4 (d). In this measurement, 1.5 mL of sample were transferred into each cuvette for the characterization process as in Figure 4 (e). Figure 4 illustrates the steps taken for standard uric acid sample preparation. Experiment was carried out by using tungsten-halogen as the light source (1) and spectrometer as optical detector. In this experiment, variable optical attenuator (2) is used to ensure intensity of light is not saturated at the spectrometer. The maximum photon count is limit to a maximum peak of 14000 photon count. Light from optical fibre is collimated into parallel beam using Ocean Optics 74-UV collimating lens (3) to interact with solution in cuvette (4) as shown in Figure 5. Then, light from cuvette is collected using the collimating lens and delivered to Ocean Optics HR4000CG-UV-NIR high resolution spectrometer (5). Ocean Optics OceanView spectrometer operating software is used to obtain and process data from spectrometer. Uric acid concentration is calculated by comparing intensity of light that passes through sample, Io and does not pass through sample, Iref. Wavelengths selected to sample uric acid are 460 nm, 525 nm and 630 nm based on available visible LED in the market. c A ySensitivit    A c 1001           real measuredreal C CC Accuracy [%]
  • 4.  ISSN: 1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 4, August 2020: 2035 - 2041 2038 Figure 4. Preparation of standard uric acid sample; (a) uric acid powder (b) NaOH solution (c) sample after adding NaOH with uric acid producing particles of sodium urate (d) sample after stirred using glass rod (e) transfer 1.5 ml into cuvette for measurement Table 1. Amount of uric acid powder added into 80 mL of NaOH solution Concentration (mg/dL) Uric Acid powder (mg) 15 12 25 20 50 40 75 60 85 68 Figure 5. Experimental setup 4. RESULTS AND DISCUSSION Figure 6 visualized the intensity spectra observed using HR4000CG-UV-VIS Ocean Optics spectrometer for SLS201 Stabilized Tungsten Halogen light operated at 400 nm to 700 nm wavelength range. This figure shows that light intensity will decrease as the concentration of uric acid increases, in which obey the Beer-Lambert law. The relationship between light intensity with absorbance is associated using (1) and (2). Calculated absorbance is plotted as in Figure 7 while calculated sensitivity and accuracy is tabulated in Table 2. Table 2 shows that the highest sensitivity is at 460 nm wavelength which is 0.0012 (mg/dL)-1 . Although all tested wavelength demonstrate a similar pattern of absorbance due to substance in the sample, amount of the light absorb is difference, thus exhibit difference sensitivity performance. For accuracy performance, 525 nm wavelength spectrophotometer system produces accuracy above 90%. The result agrees well with what has been reported in the previous literature where less error will occur when the system has absorbance value within the recommended range, which are between 0.2 and 0.8 [22].
  • 5. TELKOMNIKA Telecommun Comput El Control  Uric acid detection in visible spectrum (Afiqah Yaacob) 2039 Figure 6. Intensity spectra of different uric acid concentration Figure 7. Absorbance analytical curve Table 2. Spectrophotometer performance for selected tested wavelength Wavelength (nm) Sensitivity (mg/dL)-1 Accuracy (%) 460 0.0012 86.51 525 0.0011 91.17 630 0.0008 84.27 5. CONCLUSION The research study was carried out to analyze sensitivity and accuracy performance of spectrophotometer system at different sampled wavelength using absorbance analytical technique. The absorption was calculated using Beer’s Lambert law formula. Through the analysis, the spectrometer was able to observe the current concentration of uric acid and its absorption wavelength at visible light region. In this work, spectrophotometer system operated at 460nm wavelength has the highest sensitivity while only system at 525 nm wavelength has accuracy higher than 90%. As discussed, the accuracy is influenced by limitation of Beer-Lambert law where absorbance value beyond 0.2 and 0.8 will have large error. The spectrophotometer system with visible light source will offer a simple and economical uric acid detection system and can be applied in biomedical application. ACKNOWLEDGEMENTS The authors are grateful to Universiti Tun Hussein Onn Malaysia (UTHM) for supporting this research work under Postgraduate Research Grant (GPPS), grant no: H408 and Research University Grant (RUG) TIER 1 Scheme, grant no: H162. REFERENCES [1] Azmi N. E., et al., “A simple and sensitive fluorescence based biosensor for the determination of uric acid using H2 O 2-sensitive quantum dots/dual enzymes,” Biosensors and Bioelectronics, vol. 67, pp. 129-133, 2015. [2] Pormsila W., S. Krähenbühl, P. C. Hauser, “Capillary electrophoresis with contactless conductivity detection for uric acid determination in biological fluids,” Analytica Chimica Acta, vol. 636, no. 2, pp. 224-228, 2009.
  • 6.  ISSN: 1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 4, August 2020: 2035 - 2041 2040 [3] Zhuang Q. Q., et al., “Peroxidase-like activity of nanocrystalline cobalt selenide and its application for uric acid detection,” International Journal of Nanomedicine, vol. 12, pp. 3295-3302, 2017. [4] Boroumand S., M. A. Chamjangali, G. Bagherian, “Double injection/single detection asymmetric flow injection manifold for spectrophotometric determination of ascorbic acid and uric acid: Selection the optimal conditions by MCDM approach based on different criteria weighting methods,” Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, vol. 174, pp. 203-213, 2017. [5] Liao L. T., et al., “Evaluation of an electrochemical biosensor for uric acid measurement in human whole blood samples,” Clin Chim Acta, vol. 436, pp. 72-77, 2014. [6] Pleskacova A., et al., “Simultaneous Determination of Uric Acid, Xanthine and Hypoxanthine in Human Plasma and Serum by HPLC–UV: Uric Acid Metabolism Tracking,” Chromatographia, vol. 80, no. 4, pp. 529-536, 2017. [7] Rocha D. L., Rocha F. R., “A flow-based procedure with solenoid micro-pumps for the spectrophotometric determination of uric acid in urine,” Microchemical Journal, vol. 94, no. 1, pp. 53-59, 2010. [8] Hamzah H. H., et al., “Spectrophotometric Determination of Uric Acid in Urine Based-Enzymatic Method Uricase with 4-Aminodiphenylamine Diazonium Sulfate (Variamine Blue RT Salt),” Journal of Analytical & Bioanalytical Techniques, vol. S7, no. 011, 2013. [9] Villa J. E., Poppi R. J., “A portable SERS method for the determination of uric acid using a paper-based substrate and multivariate curve resolution,” Analyst, vol. 141, no. 6, pp. 1966-1972, 2016. [10] Jin D., et al., “Quantitative determination of uric acid using CdTe nanoparticles as fluorescence probes,” Biosensors and Bioelectronics, vol. 77, pp. 359-365, 2016. [11] Yu J., et al., “A novel chemiluminescence paper microfluidic biosensor based on enzymatic reaction for uric acid determination,” Biosensors and Bioelectronics, vol. 26, no. 7, pp. 3284-3289, 2011. [12] Ballesta-Claver J., et al., “Disposable biosensor based on cathodic electrochemiluminescence of tris (2,2-bipyridine) ruthenium (II) for uric acid determination,” Analytica Chimica Acta, vol. 770, pp. 153-160, 2013. [13] Pavlíček V., et al., “Very fast electrophoretic determination of creatinine and uric acid in human urine using a combination of two capillaries with different internal diameters,” Electrophoresis, vol. 35, no. 7, pp. 956-961, 2014. [14] Fanjul-Bolado P., et al., “Uric Acid Determination by Adsorptive Stripping Voltammetry on Multiwall Carbon Nanotubes Based Screen-Printed Electrodes,” Electroanalysis, vol. 27, no. 5, pp. 1276-1281, 2015. [15] Langsi V. K., et al., “Synthesis and characterisation of non-bonded 1.7 μm thin-shell (TS1. 7–100nm) silica particles for the rapid separation and analysis of uric acid and creatinine in human urine by hydrophilic interaction chromatography,” Journal of Chromatography A, vol. 1506, pp. 37-44, 2017. [16] Khajehsharifi H., et al., “The comparison of partial least squares and principal component regression in simultaneous spectrophotometric determination of ascorbic acid, dopamine and uric acid in real samples,” Arabian Journal of Chemistry, vol. 4, no. 2, pp. S3451-S3458, 2014. [17] Martinez-Pérez D., et al., “A reagent less fluorescent sol–gel biosensor for uric acid detection in biological fluids,” Analytical Biochemistry, vol. 322, no. 2, pp. 238-242, 2003. [18] Yamaguchi T., et al., “Spectrophotometric determination of uric acid based on fading of o-hydroxyhydroquinonephthalein- palladium (II)-hexadecyltrimethyl-ammonium complex,” Analytical Sciences, vol. 23, no. 2, pp. 223-226, 2007. [19] Norazmi N., et al., “Uric acid detection using uv-vis spectrometer,” IOP Conf. Series: Materials Science and Engineering, 2017. [20] Hardesty J. H., B. Attili, “Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry,” Collin, 2010. [21] Rashid N. C. A., et al., “Spectrophotometer with Enhanced Sensitivity for Uric Acid Detection,” Chinese Optics Letter, vol 17, no. 8, pp. 1-5, 2019. [22] Marcus T. C., et al., “Alternative wavelength for linearity preservation of Beer–Lambert Law in ozone concentration measurement,” Microwave and Optical Technology Letters, vol. 57, no. 4, pp. 1013-1016, 2015. [23] Turgeon M. L., “Linne & Ringsrud's Clinical Laboratory Science - E-Book: The Basics and Routine Techniques,” Elsevier Health Sciences, 2015. [24] Teque N., “Uric Acid,” 2018, [Online]. Available: https://www.nacalai.co.jp/ss/ec/ECsrchdetl.cfm?Dum=1& syohin=3591592&syubetsu=3 [25] Dunn P. F., “Measurement and Data Analysis for Engineering and Science,” Third Edition, CRC Press, 2014. [26] Welz B., et al., “High-Resolution Continuum Source AAS: The Better Way to Do Atomic Absorption Spectrometry,” Wiley, 2006. BIOGRAPHIES OF AUTHORS Afiqah Yaacob received her Bachelor degree from Universiti Teknologi Mara (UiTM). Currently, she is pursuing her Masters degree in Faculty of Electrical and Electronic Engineering, Universiti Tun Hussein Onn Malaysia (UTHM) in the field of spectrophotometer application.
  • 7. TELKOMNIKA Telecommun Comput El Control  Uric acid detection in visible spectrum (Afiqah Yaacob) 2041 Nor Hafizah Ngajikin received her B.E and M.E in Electronic Engineering from Universiti Teknologi Malaysia (UTM) in 2001 and 2003 respectively. She was awarded a Ph.D from UTM for her work on MEMS Fabry-Perot optical tunable filter. From 2004-2017, she served as a senior lecturer and researcher at Lightwave Communication Research Group (LCRG), UTM. She is currently a lecturer at Universiti Tun Hussein Onn Malaysia. Her research interest includes an optical devices and sensors for biomedical applications. Nurfatihah Che Abd Rashid received her B.E and M.E in Electrical Engineering from Universiti Teknologi Malaysia (UTM) in 2015 and 2019 respectively. She has involved in research on spectrophotometric detection for uric acid and soil nutrient applications. She is currently pursuing a PhD degree in Faculty of Electrical and Electronic Engineering, Universiti Tun Hussein Onn Malaysia (UTHM). Siti Hajar Aminah Ali is a lecturer at Universiti Tun Hussein Onn Malaysia (UTHM). Her work focuses on image processing and neural network. She received her Bachelor of Electrical-Telecommunications Engineering and Master of Electrical-Electronics and Telecommunications from Universiti Teknologi Malaysia (UTM), Skudai, Malaysia. Her highest education is Doctor of Philosophy in Engineering from Kobe University, Japan. Maslina Yaacob was awarded a Ph.D from Universiti Teknologi Malaysia (UTM) for her work on Wide Range Analysis of Ozone Gas Concentration in Ultraviolet Region. Her B.E and M.E in Electrical Engineering were also awarded from UTM in 2007 and 2010, respectively. Her research interest includes an optical devices and optical sensors for gas sensing applications. She is currently a lecturer and researcher at Universiti Tun Hussein Onn Malaysia. Suhaila Isaak received her Bachelor Degree in Electrical and Electronic Engineering from Universiti Teknologi Malaysia in 1998. In 2001, she completed her Master in Science (Electronic) from Universiti Putra Malaysia. She was awarded the Ph.D degree in Electrical and Electronics Engineering from University of Nottingham, England in 2011. Her major research interests are high-speed Geiger mode avalanche photodiode, analog integrated CMOS design, VLSI circuit, high speed photon counting on FPGA, and photonic experimental analysis on biological prototype system in visible light range. Currently, she has involved on integrated linear array photon camera as spectroscopy system for macronutrient soil analysis. Noran Azizan Cholan was born on 31st August 1979 in Segamat, Johor, Malaysia. He received his bachelor degree in Electronics Engineering from Universiti Tenaga Nasional Malaysia in 2002. Afterwards in 2004, he obtained his master degree in Electronics-Telecommunications Engineering form Universiti Teknologi Malaysia. In 2010, he enrolled as a PhD student in Universiti Putra Malaysia. During his PhD study in 2012, he went to Swansea University, UK and The Hong Kong Polytechnic University, Hong Kong for research attachment. He completed his PhD in 2014 before serving as a senior lecturer in Universiti Tun Hussein Onn Malaysia. As of now, he has been authors/co-authors for 25 journal and 16 conference proceeding papers. His research interests include lasers, amplifiers, optics modeling, microwave optics and optical sensors.