TELKOMNIKA Telecommunication, Computing, Electronics and Control
Vol.18, No.3, June 2020, pp. 1459~1466
ISSN: 1693-6930, accredited First Grade by Kemenristekdikti, Decree No: 21/E/KPT/2018
DOI: 10.12928/TELKOMNIKA.v18i3.14760  1459
Journal homepage: http://journal.uad.ac.id/index.php/TELKOMNIKA
Effects of noises on near infrared sensor
for blood glucose level measurement
Kiki Prawiroredjo, Engelin Shintadewi Julian
Electrical Engineering Department, Faculty of Industrial Technology, Trisakti University, Indonesia
Article Info ABSTRACT
Article history:
Received Sep 9, 2019
Revised Jan 15, 2020
Accepted Feb 26, 2020
This paper proposed the method of measuring glucose level in solution using
near infrared light (NIR) and photodiode sensor. We studied noises that
occurred on the output signal of NIR sensor in three different room
conditions in order to know the effects on this sensor output voltage stability.
The sensor’s circuit consisted of a 1450 nm NIR light emitting diode,
a photodiode as the receiver, transimpedance amplifier, a notch filter, and
a 4th
order low pass filter. The results indicated that sunlight passing through
windows was the most influencing factor caused the unstable sensor output
voltage. Filters removed the effective voltages and the average sensor output
voltages from the three rooms were 4.6825 V for air media, 2.2809 V for
water media and 2.3368 V for glucose solution media. The output voltages
tended to increase for one-hour measurement about 10 to 40 mV for air
media, 40 to 90 mV for water media and 30 to 80 mV for glucose solution
media. This sensor could only be used in a short time and suitable in a room
without sunlight. Based on the voltage difference of the average sensor
output voltage with water and glucose solution media, the sensor had
the potential to be a blood glucose level meter.
Keywords:
Light emitting diode
Near infrared light
NIR sensor
Photodiode
Sunlight
This is an open access article under the CC BY-SA license.
Corresponding Author:
Kiki Prawiroredjo,
Electrical Engineering Department, Faculty of Industrial Technology,
Trisakti University,
1 Kiai Tapa St., Jakarta 11440, Indonesia.
Email: kiki.prawiroredjo@trisakti.ac.id
1. INTRODUCTION
Diabetes is a chronic, metabolic disease characterized by elevated levels of blood glucose or blood
sugar. Diabetes of all types can lead to complications in many parts of the body and can increase the overall
risk of premature death. The most common is type 2 diabetes, usually in adults, which occurs when the body
becomes resistant to insulin or doesn't make enough insulin [1]. “Some 425 million people worldwide,
or 8.8% of adults 20-79 years, are estimated to have diabetes. About 79% live in low and middle income
countries” [2]. Persons suffering from diabetes may typically monitor their own glucose concentrations with
periodic daily measurements, usually four times each day [3]. Unfortunately many patients do not monitor
their blood glucose level at recommended frequency.It is partly due to the inconvenience, pain and
tediousness of current monitoring techniques and due to the increased risk of infection when the blood skin
barrier is broken, a non-invasive blood glucose monitor would be highly desirable. Many methods have been
proposed for the non-invasive measurements but the main concern in non invasive technique is the difficulty
in achieving high accuracy of results as no blood is directly involved in the process. Since 1980s,
NIR spectroscopy has become a promising technique for in vivo analyses, with advantages such as rapidity,
 ISSN:1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466
1460
non destructivity, and simplicity. Spectroscopic technique is an optical method used to detect blood glucose
by determining the quantity of light which is either absorbed, transmitted or emitted as a function of glucose
concentration [4]. Near-infrared radiation is a low energy electromagnetic wave, it is possible to avoid
the occurrence of radiation damage of the sample [5]. A key finding related to the non-invasive measurement
of glucose is that a major physiological response accompanies changes in glucose and can be detected
non-invasively through the resulting changes in tissue properties [6]. A change of glucose concentration may
modify the optical parameters in tissue, with the result that its glucose concentration can be extracted by
analyzing the received optical signals [7]. A limitation of the near-infrared blood glucose measurement
instruments is that each may be require to be custom calibrated for each individual user [8].
NIR sensor for blood glucose measurement has been investigated to develop non-invasive glucose
monitoring systems and to get a high accuracy result of measurement. Shyqyri et al used 940 nm NIR LED as
transmitter and passed the light beam through human fingertip and received the attenuated signal from
the other side of the finger by a photodiode sensor. Their experimental study confirms a correlation between
the sensor output voltage and glucose concentration levels [9]. Sia used 1450 and 2050 nm NIR LED and
photodiodes as the sensor to measure the voltage of 50, 100, 150 and 200 mg/dL glucose concentration.
As glucose concentration increases, output voltage from photodiode also increases. The relations observed
was linier [10]. Reddy et al developed a smart insulin device for non-invasive blood glucose levelmonitoring
using 1550 nm NIR light on human fingertip. A photodiode received the reflected infrared signal and
converted to an equivalent voltage value. The concentration of glucose level is calculated based on
the body mass index of the patient. Their results showed that there was a strong relationship between glucose
concentration and sensor output voltage. Voltage increased as glucose concentration increased [11].
Unnikrishna et al proposed a method that is based on the principle of absorbance transmittance photometry.
They used 950 nm NIR LED as the transmitter and a phototransistor as the receiver to monitor human blood
glucose [12]. Pande and Joshi in 2015 in their article presented a circuit for noninvasive measurement of
the glucose. They concluded that the transmitter with 1450-nm wavelength is more suitable [13].
A limitation in many photodiode applications is a back ground noise floor which masks the signal
detected by the photodiode. A contributing factor to background noise in a photodiode detector circuit, as in
most electronic circuits, is the parasitic coupling of electromagnetic interference (EMI) into the circuit [14].
Moreover, ambient and/or environment interference (i.e. noise) can adversely affect the measurement
accuracy. Interference is generated by many commonly-used electrical devices. In a typical household for
example, electric power lines and outlets, ambient lights, light dimmers, television or computer displays,
and power supplies or transformers generate electromagnetic interference [15]. In photodiode itself there are
sources of noise, these are flicker (1/f) noise, Johnson noise and shot noise [16]. As a preliminary research in
developing a blood glucose level sensor, we studied whether NIR sensor is susceptible to noises.
We measured the output voltage of the sensor with this proposed design in three different places with
different conditions and with three different media between the NIR light emitting diode and photodiode and
observed noises that occurred. The purpose of this studied was to know the effects of noises to sensor output
voltage stability.
2. RESEARCH METHOD
The block diagram of the proposed system consists of input, process and output section as shown in
Figure 1. The transmitter circuit in the input section is a 1450 nm NIR LED where the glucose absorption is
high [17]. It emits NIR light through media and an infrared photodiode receives the light after passing
through the media. The transmittance setup is recommended for non-invasive glucose monitoring [18].
The LED and InGaAsinfrared photodiode used in this research are from ThorLabs.
Input :
NIR source light
Proccess :
NIR sensor, TIA
filters & amp
Output :
DMM and
oscilloscope
Power supply
Figure 1. System block diagram
TELKOMNIKA Telecommun Comput El Control 
Effects of noises on near infrared sensor for blood glucose level measurement (Kiki Prawiroredjo)
1461
The principle of measuring glucose level with a NIR light source and a photodiode sensor is
the domain of spectrometry. Spectrometry is an analyzing method that studies the interaction between
material substances with electromagnetic wave radiation. This interaction changes the direction of light wave
radiation and decreases the light intensity received by the photo detector. For spectrophotometric
experiments, Beer Lambert’s law is applicable to measure the attenuation of light that is proportional to
the level of glucose solution, thickness of the sample and molar extinction coefficient. The (1) shows
the attenuation A according to Beer Lamberts law [19].
A = log(
I0
I
) = ECl (1)
where Io is transmitted light intensity, I is incident light intensity, E is extinction coefficient, C is level of
glucose solution, and l is length of light path through solution being observed.
In the previous research we measured some glucose concentration in glucose solution with this
proposed design and the results showed that higher glucose concentration yield lower sensor output voltage
but it did not show a good linier trend [20]. This result is concistent with Beer Lambert’s law. In this research
we studied the sources of noise that made the non linier trend of output voltages from the sensor as
the glucose concentration increase. The signal conditioning circuits are a transimpedance amplifier, a notch
filter, a 4th
order low pass filter and a final stage amplifier. The NIR sensor photodiode is fed to a
transimpedance amplifier to convert the current from photodiode sensor to voltage as shown in Figure 2.
The output voltage is filtered by a 50 Hz notch filter and a 5 Hz 4th
order low pass filter to reduce low and
high frequency noises. Figure 3 shows the 50 Hz notch filter. The 50 Hz notch filter is an active filter with a
unity gain operational amplifier at the output stage of the filter to reduce loading effect. The simulation's
result of the notch filter frequency response is shown in Figure 4.
VS
-
RF
IC1
-
+
VS
+
Vo1
CF
Ipd
-
+
IC2
VS
+
VS
-
Vi2
Vo2
R1
R2
R1
C1 C1
C2
Figure 2. Transimpedance amplifier Figure 3. Notch filter
Figure 4. Notch filter frequency response
 ISSN:1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466
1462
The active low pass filter circuit shown in Figure 5 is a two-stages and high gain operational
amplifier which functions are to amplify and filter the appropriate voltage level from input signal [21].
The low pass filter frequency cut off is designed at 5 Hz and the frequency response is Butterworth
approximation. The simulation result of the low pass filter frequency response is shown in Figure 6. In this
experiment we measured the output voltages of the transimpedance amplifier (TIA) and the output voltages
of the final stage amplifier after filters to observe the noises. We measured the voltages with a metrix digital
voltmeter and a digital oscilloscope to observe the noises that occurred and we analyzed the stability of the
sensor output voltages for one-hour measurement.
R4
03
-
+
I
C
3
VS
+
VS
-
V04
C5
C6
R4-
+
IC3
VS
+
VS
-
Vi3
V
R3 R3
C3
C4
Figure 5. 4th
order low pass filter [22]
Figure 6. 4th
order low pass filter frequency response
The media between the LED and photodiode from the NIR sensor were air, water and 200 mg/dL
glucose solution. The water and glucose solution were placed in an acrylic box with 0.5 cm thickness, 5 cm
height and 5 cm width as shown in Figure 7. Aqueous optical paths of 5 mm are required to measure
clinically relevant concentrations of glucose in the first-overtone region. All of the tested sites met this
requirement [23]. The room temperature was constantly maintained at 27° C and the supply voltage was
constantly maintained at 5.08 volts.
Acrylic
box
photo
diode
IR
LED
5 cm
5 cm
LED
circuit
TIA, filters
& amp
Osc and
DMM
0,5 cm
Figure 7. The experiment set up
TELKOMNIKA Telecommun Comput El Control 
Effects of noises on near infrared sensor for blood glucose level measurement (Kiki Prawiroredjo)
1463
The first experiment was conducted on the 6th
floor of building E, in a laboratory room that near a
power transformer for about 100 meters distance from the building (room A). The laboratory's windows were
transparent glass. The second experiment was conducted in a classroom on the 6th
floor of building E lit by
fluorescent lamps and near a power transformer for about 50 meters distance from the building (room B).
This room's windows were covered by multiplex boards. The third experiment was conducted in an office
room on the 12th
floor of building M (room C) with transparent glass windows.
The LED circuit drived the IR LED to emitted NIR light. After passing through media in an acrylic
box, the photodiode sensor received the NIR light and a transimpedance amplifier (TIA) converted
the current from the sensor to voltage. Filters were used to eliminate the low and high frequency noises.
After the final amplifier the output voltage was measured with a digital multimeter. In this experiment we
measured the output voltage with air, water and glucose solution media for one hour to know the stability of
the output voltage. For one-hour measurement we expected a flat linier trend line in voltage versus time graph.
3. RESULTS AND ANALYSIS
The experiment result of the sensor output voltages in three different rooms are shown in Figures 8-10
and Table 1. The output voltages were recorded every five minutes within one hour. We measured the output
voltages with air media between NIR LED and photodiode at 9:35 to 10:35 AM, with water media at 10:40 to
11:40 AM and with glucose solution media at 13:40 to 14:40 PM on different days for each room. Since the
transmitted signal was direct current signal, the received signal should be direct current signal too.
Figure 8. Room A, B and C sensor output voltages with air media
Figure 9. Room A, B and C sensor output voltages with water media
 ISSN:1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466
1464
Figure 10. Room A, B and C sensor output voltages with glucose solution media
Table 1. TIA and sensor average output voltages for room A, B, and C
Room A Room B Room C
TIA Vout [V] Sensor Vout [V] TIA Vout [V] Sensor Vout [V] TIA Vout [V] Sensor Vout [V]
Media VDC Veff VDC Veff VDC Veff VDC Veff VDC Veff VDC Veff
Air 3.7094 0.0000 4.7064 0.0000 3.6895 0.0000 4.6878 0.0000 3.6503 0.0000 4.6533 0.0000
Water 1.5868 0.0369 2.3173 0.0000 1.4984 0.0675 2.3116 0.0000 1.4732 0.0780 2.2140 0.0000
Glucose 1.5179 0.0739 2.3104 0.0000 1.5587 0.0690 2.3687 0.0000 1.5484 0.0775 2.3313 0.0000
As seen in Figure 8 with air media between the NIR LED and photodiode, the sensor output
voltages from room A and B were higher and more stable than the voltages in room C. Room C was the room
which received most sunlight. As seen from room A and B, most of the measured voltages were closed to
linier trend line but in room C most of the measured voltages were far from linier trend line. The red linier
trend line had the most flat slope and it meant that the measured voltages in room B were the most stable
compared to the other rooms. The sensor output voltages tended to increase over time for each room from
10 mV to 40 mV.
As seen in Figure 9, the sensor output voltages from room A and B were higher and more stable than
the voltages in room C, but the voltage stability was worse than with air media. The sensor output voltages
tended to increase over time for each room with water media from 40 mV to 90 mV. As seen from room A,
most of the measured voltages were closed to linier trend line but in room C most of the measured voltages
were far from linier trend line. The blue linier trend line had the most flat slope but most of the measured
voltages were far from the blue linier trend line. The voltage stability of room B was better than the other
rooms. Room B was the room without window, sunlight did not interfere the measurements. As seen from
Figure 10, the sensor output voltages from room B were higher and more stable compared to voltages in
room A and C. The voltages tended to increase over time for each room with glucose solution media from
30 mV to 80 mV.
With air, water and glucose solution media the sensor output voltages tend to increase time by time,
we suspected this was caused by noises in the photodiode components. From thisexperiment it can be seen
that the measurement in room B with covered windows produced the most stable output voltages. The TIA
and sensor average output voltages for room A, B and C is shown in Table 1. The first row of Table 1
showed that with air media, there was no measured effective voltage on the TIA and sensor output voltages.
It meant that noises did not appear with air media for the three rooms. Air media did not absorb the NIR light
from the NIR LED and in these experiments the DC output voltages from room A were higher than room B
and C. The sensor average output voltage for the three rooms with air media was 4.6825 V. The second row
of Table 1 showed that with water media, there were effective voltages on the TIA output voltages; it means
that noises did appear with water media for the three rooms. The effective output voltages were the 50 Hz
interference from the main power supply [24]. High frequency noises also appeared on the oscilloscope with
very small voltage. After passing through the filters, the effective output voltages from the sensor were
TELKOMNIKA Telecommun Comput El Control 
Effects of noises on near infrared sensor for blood glucose level measurement (Kiki Prawiroredjo)
1465
reduced to zero. These sensor DC output voltages from the three rooms were lower almost half of the output
voltages from air media. This result was consistent with our previous result [25]. Water absorbed NIR light
and caused noises [26]. The sensor average output voltage for the three rooms with water media was
2.2809 V. The third row of Table 1 showed that with 200 mg/dL glucose solution media, there were effective
voltages on the TIA output which meant noises did appear with glucose solution media for the three rooms.
After passing through the filter the effective output voltages from the sensor were reduced to zero.
The effective output voltages were a bit higher than the effective voltages with water media for room A and
B; it meant that noises with glucose solution were higher than with water media [25]. The sensor average
output voltage for the three rooms with glucose solution media was 2.3368 V. There was voltage difference
between sensor average output voltage with water and glucose solution media from the three rooms and it
could be concluded that the sensor had the potential as a blood glucose level meter.
4. CONCLUSION
In order to find out the effect of noises on NIR sensor, we studied noise sources in measuring
glucose level with NIR sensor over time and measured the output voltage of the sensor in three different
conditions with three different media between the light source and the sensor. From Table 1 can be seen that
the filters reduced the effective output voltages to zero and it meant that the 50 Hz and high frequency noises
could be removed with filters. The stability of the measurement output voltages from the three different
places depended on the environmental conditions especially sunlight. The sensor output voltages of
the 3-room tended to increase over time and we suspected it was from the noises in the photodiode
component. Hence, this sensor can only be used in a short time and suitable only in a room without sunlight.
In comparison with the other researches generally did not count on the sunlight effect. Based on
the data presented on Table 1, there was voltage difference between average sensor output voltage with water
media and average sensor output voltage with glucose solution media from room A, B and C. It could be
concluded that the sensor had the potential as a blood glucose level meter. In our future research, we target to
put the photodiode on an ear clip to be applied to an earlobe, process the sensor output signal, and develop
software to display the measurement result on a smartphone.
ACKNOWLEDGMENT
We would like to express our gratitude to Trisakti University and DRPM, Ministry of Research,
Technology and Higher Education of the Republic of Indonesia, Grant Number 201/A/LPT/USAKTI/V/2019, for
funding this research.
REFERENCES
[1] WHO, “Global Report on Diabetes,” World Health Organization, 2016.
[2] International Diabetes Federation, “IDF Diabetes Atlas, 8 ed,” International Diabetes Federation, 2017.
[3] K. J. Schlager, “Non-Invasive Near Infrared Measurement of Blood Analyte Concentrations,” US4882492A.
(Patent), 1988.
[4] N. A. Salam, et al., “The Evolution of Non-invasive Blood Glucose Monitoring System for Personal Application,”
Journal of Telecommunication, Electronics and Computer Engineering, vol. 8, no. 1, pp. 59-65, 2016.
[5] K. Maruo and M. Oka, “Method of Determining a Glucose Concentration in a Target by Using an Near Infrared
Spectroscopy,” US5957841A (Patent), 1998.
[6] G. Acosta, et al., “Compact Apparatus for Non-Invasive Measurement of Glucose Through Near-Infrared
Spectroscopy,” CA2476421A1. (Patent), 2003, https://patents.google.com/patent/CA2476421A1/en
[7] Z. Zhao and R. A. Myllyla, “Photoacoustic Blood Glucose and Skin Measurement Based on Optical Scattering
Effect,” Proceedings of Saratov Fall Meeting, vol. 4707, 2001.
[8] R. D. Rosenthal, “Method for Providing Custom Calibration for Near-Infrared Instruments for Measurement of
Blood Glucose,” US5068536A. (Patent), 1991.
[9] S. Haxha and J. Jhoja, “Optical Based Noninvasive Glucose Monitoring Sensor Prototype,” IEEE Photonics
Journal,vol. 8, no. 6, pp. 1-11, Dec 2016.
[10] D. Sia, “Design of A Near Infrared Device for The Study of Glucose Concentration Measurements,”EE4BI6
Electrical Engineering Biomedical Capstones, McMaster University, 2010.
[11] P. S. Reddy and K. Jyostna, “Development of Smart Insulin Device for Non Invasive Blood Glucose Level
Monitoring,” in IEEE 7th International Advance Computing Conference, pp. 516-519, 2017.
[12] K. A. U. Menon, et al., “A Survey on Non-Invasive Blood Glucose Monitoring Using NIR,” in International
Conference on Communication and Signal Processing, India, 2013.
[13] M. Pande and A. Joshi, “Non-Invasive Blood Glucose Measurement,” Proceeding of Int J ComputEng Res,
vol. 5, I. 4 pp. 26-28, 2015, http://www.ijceronline.com/papers/Vol5_issue4/E054026028.pdf
[14] J. P. Coffin, et al., “Photodiode Detector with Integrated Noise Shielding,”US6184521B1. (Patent),1998.
 ISSN:1693-6930
TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466
1466
[15] J. Poeze, et al., “Interference Detector for Patient Monitor,” US 8754776B2. (Patent), 2013.
[16] M. Doğan and A. Tangel, “Shot Noise Measurement in PIN Photodiodes,” International Journalof Computing,
Communication and Instrumentation Engg., vol. 3, no. 2, pp. 297-300, 2016.
[17] C. D. Bobade and M. S. Patil, “Non-invasive Blood Glucose Level Monitoring System for Diabetic Patients using
Near-Infrared Spectroscopy,” American Journal of Computer Science and Information Technology, vol. 4, no. 01,
pp. 1-8, 2016.
[18] Jeon K. J., et al., “Comparison Between Transmittance and Reflectance Measurements in Glucose Determination
Using Near Infrared Spectroscopy,” Journalof Biomedical Optics, vol. 11, no. 1, 2006.
[19] E. S. Julian, et al., “The Model of Near Infrared Sensor Output Voltage as a Function of Glucose Concentration in
Solution,” in The 15th International Conference on Quality in Research (QIR), Bali, pp. 146-149, 2017.
[20] K. Prawiroredjo, et al., “Noise reduction in near infrared-based glucose concentration measurement sensor,”
in IOP Conference Series: Materials Science and Engineering, vol. 532, Jun 2019.
[21] Z. M. Yosuf, et al.,“Design and Fabrication of Cost-Effective Heart-Rate Pulse Monitoring Sensor System,”
TELKOMNIKA Telecommunication Computing Electronics and Control, vol. 17, no. 5, pp. 2497-2504, Oct 2019.
[22] B. G. Irianto, et al., “A low-cost electro-cardiograph machine equipped with sensitivity and paper speed option,”
TELKOMNIKA Telecommunication Computing Electronics and Control, vol. 17, no. 3, pp. 1275-1281, Jun 2019.
[23] J. J. Burmeister and M. A. Arnold, “Evaluation of Measurement Sites for Noninvasive Blood Glucose Sensing with
Near-Infrared Transmission Spectroscopy,” Clinical Chemistry, vol. 45, no. 9, pp. 1621-1627, 1999.
[24] A. C. M. van Rijn, et al., “High-Quality Recording of Bioelectric Events. Part 1 Interference Reduction, Theory and
Practice,” Medical and Biological Engineering and Computer, vol. 28, no. 5, pp. 389-397, Oct 1990.
[25] K. Prawiroredjo and E. S. Julian, “Comparative study of 940 nm and 1450 nm near infrared sensor for glucose
concentration monitoring,” TELKOMNIKA Telecommunication Computing Electronics and Control, vol. 17, no. 2,
pp. 981-985, Apr 2019.
[26] K. Palmer and D. Williams, “Optical properties of water in the near infrared,” Journal of the Optical Societyof
America, vol. 64, no. 8, pp. 1107-1110, Aug 1974.

More Related Content

Similar to Effects of noises on near infrared sensor for blood glucose level measurement

Improving hemoglobin estimation accuracy through standardizing of light-emitt...
Improving hemoglobin estimation accuracy through standardizing of light-emitt...Improving hemoglobin estimation accuracy through standardizing of light-emitt...
Improving hemoglobin estimation accuracy through standardizing of light-emitt...IJECEIAES
 
Non-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose MeasurementNon-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose Measurementijceronline
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...TELKOMNIKA JOURNAL
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method IJECEIAES
 
Characterization of excitation source LEDs and sensors without filters for me...
Characterization of excitation source LEDs and sensors without filters for me...Characterization of excitation source LEDs and sensors without filters for me...
Characterization of excitation source LEDs and sensors without filters for me...TELKOMNIKA JOURNAL
 
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptxRADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptxSrinath Chowdary
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...IRJET Journal
 
upload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptxupload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptxPreethi Natarajan
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightIRJET Journal
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET Journal
 
Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...eSAT Publishing House
 
Non Invasive Glucose Measurement using Raman Spectroscopy
Non Invasive Glucose Measurement using Raman SpectroscopyNon Invasive Glucose Measurement using Raman Spectroscopy
Non Invasive Glucose Measurement using Raman SpectroscopyIRJET Journal
 
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEWMeasurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEWIOSR Journals
 
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...IRJET Journal
 
Microelectronicpill
MicroelectronicpillMicroelectronicpill
MicroelectronicpillRohit Roy
 
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGProblems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGIJSRD
 

Similar to Effects of noises on near infrared sensor for blood glucose level measurement (20)

U34129131
U34129131U34129131
U34129131
 
Improving hemoglobin estimation accuracy through standardizing of light-emitt...
Improving hemoglobin estimation accuracy through standardizing of light-emitt...Improving hemoglobin estimation accuracy through standardizing of light-emitt...
Improving hemoglobin estimation accuracy through standardizing of light-emitt...
 
Non-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose MeasurementNon-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose Measurement
 
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
Comparative study of 940 nm and 1450 nm near infrared sensor for glucose conc...
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method
 
Characterization of excitation source LEDs and sensors without filters for me...
Characterization of excitation source LEDs and sensors without filters for me...Characterization of excitation source LEDs and sensors without filters for me...
Characterization of excitation source LEDs and sensors without filters for me...
 
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptxRADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...IRJET-  	  Development of Portable Device for Measurement of Blood Glucose, T...
IRJET- Development of Portable Device for Measurement of Blood Glucose, T...
 
upload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptxupload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptx
 
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser LightA Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
A Non-Invasive Blood Glucose Monitoring Device using Red Laser Light
 
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
IRJET- Development of Portable Device for Measurement of Blood Glucose, Tempe...
 
Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...
 
Non Invasive Glucose Measurement using Raman Spectroscopy
Non Invasive Glucose Measurement using Raman SpectroscopyNon Invasive Glucose Measurement using Raman Spectroscopy
Non Invasive Glucose Measurement using Raman Spectroscopy
 
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEWMeasurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
 
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
IRJET- Non-Invasive Blood Glucose Measurement Depending on the Blood Dielectr...
 
Microelectronicpill
MicroelectronicpillMicroelectronicpill
Microelectronicpill
 
Conception and Development of a Tele-Medical Interface Dedicated to Tele Mon...
Conception and Development of a Tele-Medical Interface  Dedicated to Tele Mon...Conception and Development of a Tele-Medical Interface  Dedicated to Tele Mon...
Conception and Development of a Tele-Medical Interface Dedicated to Tele Mon...
 
Non invasive
Non invasiveNon invasive
Non invasive
 
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGProblems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
 
40120140507005
4012014050700540120140507005
40120140507005
 

More from TELKOMNIKA JOURNAL

Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...TELKOMNIKA JOURNAL
 
Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...TELKOMNIKA JOURNAL
 
Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...TELKOMNIKA JOURNAL
 
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...TELKOMNIKA JOURNAL
 
Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...TELKOMNIKA JOURNAL
 
Efficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antennaEfficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antennaTELKOMNIKA JOURNAL
 
Design and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fireDesign and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fireTELKOMNIKA JOURNAL
 
Wavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio networkWavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio networkTELKOMNIKA JOURNAL
 
A novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bandsA novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bandsTELKOMNIKA JOURNAL
 
Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...TELKOMNIKA JOURNAL
 
Brief note on match and miss-match uncertainties
Brief note on match and miss-match uncertaintiesBrief note on match and miss-match uncertainties
Brief note on match and miss-match uncertaintiesTELKOMNIKA JOURNAL
 
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...TELKOMNIKA JOURNAL
 
Evaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G systemEvaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G systemTELKOMNIKA JOURNAL
 
Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...TELKOMNIKA JOURNAL
 
Reagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensorReagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensorTELKOMNIKA JOURNAL
 
Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...TELKOMNIKA JOURNAL
 
A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...TELKOMNIKA JOURNAL
 
Electroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networksElectroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networksTELKOMNIKA JOURNAL
 
Adaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imagingAdaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imagingTELKOMNIKA JOURNAL
 
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...TELKOMNIKA JOURNAL
 

More from TELKOMNIKA JOURNAL (20)

Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...Amazon products reviews classification based on machine learning, deep learni...
Amazon products reviews classification based on machine learning, deep learni...
 
Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...Design, simulation, and analysis of microstrip patch antenna for wireless app...
Design, simulation, and analysis of microstrip patch antenna for wireless app...
 
Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...Design and simulation an optimal enhanced PI controller for congestion avoida...
Design and simulation an optimal enhanced PI controller for congestion avoida...
 
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
Improving the detection of intrusion in vehicular ad-hoc networks with modifi...
 
Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...Conceptual model of internet banking adoption with perceived risk and trust f...
Conceptual model of internet banking adoption with perceived risk and trust f...
 
Efficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antennaEfficient combined fuzzy logic and LMS algorithm for smart antenna
Efficient combined fuzzy logic and LMS algorithm for smart antenna
 
Design and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fireDesign and implementation of a LoRa-based system for warning of forest fire
Design and implementation of a LoRa-based system for warning of forest fire
 
Wavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio networkWavelet-based sensing technique in cognitive radio network
Wavelet-based sensing technique in cognitive radio network
 
A novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bandsA novel compact dual-band bandstop filter with enhanced rejection bands
A novel compact dual-band bandstop filter with enhanced rejection bands
 
Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...Deep learning approach to DDoS attack with imbalanced data at the application...
Deep learning approach to DDoS attack with imbalanced data at the application...
 
Brief note on match and miss-match uncertainties
Brief note on match and miss-match uncertaintiesBrief note on match and miss-match uncertainties
Brief note on match and miss-match uncertainties
 
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
Implementation of FinFET technology based low power 4×4 Wallace tree multipli...
 
Evaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G systemEvaluation of the weighted-overlap add model with massive MIMO in a 5G system
Evaluation of the weighted-overlap add model with massive MIMO in a 5G system
 
Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...Reflector antenna design in different frequencies using frequency selective s...
Reflector antenna design in different frequencies using frequency selective s...
 
Reagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensorReagentless iron detection in water based on unclad fiber optical sensor
Reagentless iron detection in water based on unclad fiber optical sensor
 
Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...Impact of CuS counter electrode calcination temperature on quantum dot sensit...
Impact of CuS counter electrode calcination temperature on quantum dot sensit...
 
A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...A progressive learning for structural tolerance online sequential extreme lea...
A progressive learning for structural tolerance online sequential extreme lea...
 
Electroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networksElectroencephalography-based brain-computer interface using neural networks
Electroencephalography-based brain-computer interface using neural networks
 
Adaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imagingAdaptive segmentation algorithm based on level set model in medical imaging
Adaptive segmentation algorithm based on level set model in medical imaging
 
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
Automatic channel selection using shuffled frog leaping algorithm for EEG bas...
 

Recently uploaded

University management System project report..pdf
University management System project report..pdfUniversity management System project report..pdf
University management System project report..pdfKamal Acharya
 
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTINGMANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTINGSIVASHANKAR N
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxupamatechverse
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdfONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdfKamal Acharya
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...ranjana rawat
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Bookingdharasingh5698
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Call Girls in Nagpur High Profile
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdfankushspencer015
 
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxBSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxfenichawla
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordAsst.prof M.Gokilavani
 
Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)simmis5
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 

Recently uploaded (20)

University management System project report..pdf
University management System project report..pdfUniversity management System project report..pdf
University management System project report..pdf
 
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTINGMANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
MANUFACTURING PROCESS-II UNIT-1 THEORY OF METAL CUTTING
 
Introduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptxIntroduction to Multiple Access Protocol.pptx
Introduction to Multiple Access Protocol.pptx
 
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINEDJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Roadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and RoutesRoadmap to Membership of RICS - Pathways and Routes
Roadmap to Membership of RICS - Pathways and Routes
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdfONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
ONLINE FOOD ORDER SYSTEM PROJECT REPORT.pdf
 
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
The Most Attractive Pune Call Girls Budhwar Peth 8250192130 Will You Miss Thi...
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdf
 
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxBSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024
 
Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)Java Programming :Event Handling(Types of Events)
Java Programming :Event Handling(Types of Events)
 
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur EscortsCall Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
Call Girls Service Nagpur Tanvi Call 7001035870 Meet With Nagpur Escorts
 
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 

Effects of noises on near infrared sensor for blood glucose level measurement

  • 1. TELKOMNIKA Telecommunication, Computing, Electronics and Control Vol.18, No.3, June 2020, pp. 1459~1466 ISSN: 1693-6930, accredited First Grade by Kemenristekdikti, Decree No: 21/E/KPT/2018 DOI: 10.12928/TELKOMNIKA.v18i3.14760  1459 Journal homepage: http://journal.uad.ac.id/index.php/TELKOMNIKA Effects of noises on near infrared sensor for blood glucose level measurement Kiki Prawiroredjo, Engelin Shintadewi Julian Electrical Engineering Department, Faculty of Industrial Technology, Trisakti University, Indonesia Article Info ABSTRACT Article history: Received Sep 9, 2019 Revised Jan 15, 2020 Accepted Feb 26, 2020 This paper proposed the method of measuring glucose level in solution using near infrared light (NIR) and photodiode sensor. We studied noises that occurred on the output signal of NIR sensor in three different room conditions in order to know the effects on this sensor output voltage stability. The sensor’s circuit consisted of a 1450 nm NIR light emitting diode, a photodiode as the receiver, transimpedance amplifier, a notch filter, and a 4th order low pass filter. The results indicated that sunlight passing through windows was the most influencing factor caused the unstable sensor output voltage. Filters removed the effective voltages and the average sensor output voltages from the three rooms were 4.6825 V for air media, 2.2809 V for water media and 2.3368 V for glucose solution media. The output voltages tended to increase for one-hour measurement about 10 to 40 mV for air media, 40 to 90 mV for water media and 30 to 80 mV for glucose solution media. This sensor could only be used in a short time and suitable in a room without sunlight. Based on the voltage difference of the average sensor output voltage with water and glucose solution media, the sensor had the potential to be a blood glucose level meter. Keywords: Light emitting diode Near infrared light NIR sensor Photodiode Sunlight This is an open access article under the CC BY-SA license. Corresponding Author: Kiki Prawiroredjo, Electrical Engineering Department, Faculty of Industrial Technology, Trisakti University, 1 Kiai Tapa St., Jakarta 11440, Indonesia. Email: kiki.prawiroredjo@trisakti.ac.id 1. INTRODUCTION Diabetes is a chronic, metabolic disease characterized by elevated levels of blood glucose or blood sugar. Diabetes of all types can lead to complications in many parts of the body and can increase the overall risk of premature death. The most common is type 2 diabetes, usually in adults, which occurs when the body becomes resistant to insulin or doesn't make enough insulin [1]. “Some 425 million people worldwide, or 8.8% of adults 20-79 years, are estimated to have diabetes. About 79% live in low and middle income countries” [2]. Persons suffering from diabetes may typically monitor their own glucose concentrations with periodic daily measurements, usually four times each day [3]. Unfortunately many patients do not monitor their blood glucose level at recommended frequency.It is partly due to the inconvenience, pain and tediousness of current monitoring techniques and due to the increased risk of infection when the blood skin barrier is broken, a non-invasive blood glucose monitor would be highly desirable. Many methods have been proposed for the non-invasive measurements but the main concern in non invasive technique is the difficulty in achieving high accuracy of results as no blood is directly involved in the process. Since 1980s, NIR spectroscopy has become a promising technique for in vivo analyses, with advantages such as rapidity,
  • 2.  ISSN:1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466 1460 non destructivity, and simplicity. Spectroscopic technique is an optical method used to detect blood glucose by determining the quantity of light which is either absorbed, transmitted or emitted as a function of glucose concentration [4]. Near-infrared radiation is a low energy electromagnetic wave, it is possible to avoid the occurrence of radiation damage of the sample [5]. A key finding related to the non-invasive measurement of glucose is that a major physiological response accompanies changes in glucose and can be detected non-invasively through the resulting changes in tissue properties [6]. A change of glucose concentration may modify the optical parameters in tissue, with the result that its glucose concentration can be extracted by analyzing the received optical signals [7]. A limitation of the near-infrared blood glucose measurement instruments is that each may be require to be custom calibrated for each individual user [8]. NIR sensor for blood glucose measurement has been investigated to develop non-invasive glucose monitoring systems and to get a high accuracy result of measurement. Shyqyri et al used 940 nm NIR LED as transmitter and passed the light beam through human fingertip and received the attenuated signal from the other side of the finger by a photodiode sensor. Their experimental study confirms a correlation between the sensor output voltage and glucose concentration levels [9]. Sia used 1450 and 2050 nm NIR LED and photodiodes as the sensor to measure the voltage of 50, 100, 150 and 200 mg/dL glucose concentration. As glucose concentration increases, output voltage from photodiode also increases. The relations observed was linier [10]. Reddy et al developed a smart insulin device for non-invasive blood glucose levelmonitoring using 1550 nm NIR light on human fingertip. A photodiode received the reflected infrared signal and converted to an equivalent voltage value. The concentration of glucose level is calculated based on the body mass index of the patient. Their results showed that there was a strong relationship between glucose concentration and sensor output voltage. Voltage increased as glucose concentration increased [11]. Unnikrishna et al proposed a method that is based on the principle of absorbance transmittance photometry. They used 950 nm NIR LED as the transmitter and a phototransistor as the receiver to monitor human blood glucose [12]. Pande and Joshi in 2015 in their article presented a circuit for noninvasive measurement of the glucose. They concluded that the transmitter with 1450-nm wavelength is more suitable [13]. A limitation in many photodiode applications is a back ground noise floor which masks the signal detected by the photodiode. A contributing factor to background noise in a photodiode detector circuit, as in most electronic circuits, is the parasitic coupling of electromagnetic interference (EMI) into the circuit [14]. Moreover, ambient and/or environment interference (i.e. noise) can adversely affect the measurement accuracy. Interference is generated by many commonly-used electrical devices. In a typical household for example, electric power lines and outlets, ambient lights, light dimmers, television or computer displays, and power supplies or transformers generate electromagnetic interference [15]. In photodiode itself there are sources of noise, these are flicker (1/f) noise, Johnson noise and shot noise [16]. As a preliminary research in developing a blood glucose level sensor, we studied whether NIR sensor is susceptible to noises. We measured the output voltage of the sensor with this proposed design in three different places with different conditions and with three different media between the NIR light emitting diode and photodiode and observed noises that occurred. The purpose of this studied was to know the effects of noises to sensor output voltage stability. 2. RESEARCH METHOD The block diagram of the proposed system consists of input, process and output section as shown in Figure 1. The transmitter circuit in the input section is a 1450 nm NIR LED where the glucose absorption is high [17]. It emits NIR light through media and an infrared photodiode receives the light after passing through the media. The transmittance setup is recommended for non-invasive glucose monitoring [18]. The LED and InGaAsinfrared photodiode used in this research are from ThorLabs. Input : NIR source light Proccess : NIR sensor, TIA filters & amp Output : DMM and oscilloscope Power supply Figure 1. System block diagram
  • 3. TELKOMNIKA Telecommun Comput El Control  Effects of noises on near infrared sensor for blood glucose level measurement (Kiki Prawiroredjo) 1461 The principle of measuring glucose level with a NIR light source and a photodiode sensor is the domain of spectrometry. Spectrometry is an analyzing method that studies the interaction between material substances with electromagnetic wave radiation. This interaction changes the direction of light wave radiation and decreases the light intensity received by the photo detector. For spectrophotometric experiments, Beer Lambert’s law is applicable to measure the attenuation of light that is proportional to the level of glucose solution, thickness of the sample and molar extinction coefficient. The (1) shows the attenuation A according to Beer Lamberts law [19]. A = log( I0 I ) = ECl (1) where Io is transmitted light intensity, I is incident light intensity, E is extinction coefficient, C is level of glucose solution, and l is length of light path through solution being observed. In the previous research we measured some glucose concentration in glucose solution with this proposed design and the results showed that higher glucose concentration yield lower sensor output voltage but it did not show a good linier trend [20]. This result is concistent with Beer Lambert’s law. In this research we studied the sources of noise that made the non linier trend of output voltages from the sensor as the glucose concentration increase. The signal conditioning circuits are a transimpedance amplifier, a notch filter, a 4th order low pass filter and a final stage amplifier. The NIR sensor photodiode is fed to a transimpedance amplifier to convert the current from photodiode sensor to voltage as shown in Figure 2. The output voltage is filtered by a 50 Hz notch filter and a 5 Hz 4th order low pass filter to reduce low and high frequency noises. Figure 3 shows the 50 Hz notch filter. The 50 Hz notch filter is an active filter with a unity gain operational amplifier at the output stage of the filter to reduce loading effect. The simulation's result of the notch filter frequency response is shown in Figure 4. VS - RF IC1 - + VS + Vo1 CF Ipd - + IC2 VS + VS - Vi2 Vo2 R1 R2 R1 C1 C1 C2 Figure 2. Transimpedance amplifier Figure 3. Notch filter Figure 4. Notch filter frequency response
  • 4.  ISSN:1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466 1462 The active low pass filter circuit shown in Figure 5 is a two-stages and high gain operational amplifier which functions are to amplify and filter the appropriate voltage level from input signal [21]. The low pass filter frequency cut off is designed at 5 Hz and the frequency response is Butterworth approximation. The simulation result of the low pass filter frequency response is shown in Figure 6. In this experiment we measured the output voltages of the transimpedance amplifier (TIA) and the output voltages of the final stage amplifier after filters to observe the noises. We measured the voltages with a metrix digital voltmeter and a digital oscilloscope to observe the noises that occurred and we analyzed the stability of the sensor output voltages for one-hour measurement. R4 03 - + I C 3 VS + VS - V04 C5 C6 R4- + IC3 VS + VS - Vi3 V R3 R3 C3 C4 Figure 5. 4th order low pass filter [22] Figure 6. 4th order low pass filter frequency response The media between the LED and photodiode from the NIR sensor were air, water and 200 mg/dL glucose solution. The water and glucose solution were placed in an acrylic box with 0.5 cm thickness, 5 cm height and 5 cm width as shown in Figure 7. Aqueous optical paths of 5 mm are required to measure clinically relevant concentrations of glucose in the first-overtone region. All of the tested sites met this requirement [23]. The room temperature was constantly maintained at 27° C and the supply voltage was constantly maintained at 5.08 volts. Acrylic box photo diode IR LED 5 cm 5 cm LED circuit TIA, filters & amp Osc and DMM 0,5 cm Figure 7. The experiment set up
  • 5. TELKOMNIKA Telecommun Comput El Control  Effects of noises on near infrared sensor for blood glucose level measurement (Kiki Prawiroredjo) 1463 The first experiment was conducted on the 6th floor of building E, in a laboratory room that near a power transformer for about 100 meters distance from the building (room A). The laboratory's windows were transparent glass. The second experiment was conducted in a classroom on the 6th floor of building E lit by fluorescent lamps and near a power transformer for about 50 meters distance from the building (room B). This room's windows were covered by multiplex boards. The third experiment was conducted in an office room on the 12th floor of building M (room C) with transparent glass windows. The LED circuit drived the IR LED to emitted NIR light. After passing through media in an acrylic box, the photodiode sensor received the NIR light and a transimpedance amplifier (TIA) converted the current from the sensor to voltage. Filters were used to eliminate the low and high frequency noises. After the final amplifier the output voltage was measured with a digital multimeter. In this experiment we measured the output voltage with air, water and glucose solution media for one hour to know the stability of the output voltage. For one-hour measurement we expected a flat linier trend line in voltage versus time graph. 3. RESULTS AND ANALYSIS The experiment result of the sensor output voltages in three different rooms are shown in Figures 8-10 and Table 1. The output voltages were recorded every five minutes within one hour. We measured the output voltages with air media between NIR LED and photodiode at 9:35 to 10:35 AM, with water media at 10:40 to 11:40 AM and with glucose solution media at 13:40 to 14:40 PM on different days for each room. Since the transmitted signal was direct current signal, the received signal should be direct current signal too. Figure 8. Room A, B and C sensor output voltages with air media Figure 9. Room A, B and C sensor output voltages with water media
  • 6.  ISSN:1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466 1464 Figure 10. Room A, B and C sensor output voltages with glucose solution media Table 1. TIA and sensor average output voltages for room A, B, and C Room A Room B Room C TIA Vout [V] Sensor Vout [V] TIA Vout [V] Sensor Vout [V] TIA Vout [V] Sensor Vout [V] Media VDC Veff VDC Veff VDC Veff VDC Veff VDC Veff VDC Veff Air 3.7094 0.0000 4.7064 0.0000 3.6895 0.0000 4.6878 0.0000 3.6503 0.0000 4.6533 0.0000 Water 1.5868 0.0369 2.3173 0.0000 1.4984 0.0675 2.3116 0.0000 1.4732 0.0780 2.2140 0.0000 Glucose 1.5179 0.0739 2.3104 0.0000 1.5587 0.0690 2.3687 0.0000 1.5484 0.0775 2.3313 0.0000 As seen in Figure 8 with air media between the NIR LED and photodiode, the sensor output voltages from room A and B were higher and more stable than the voltages in room C. Room C was the room which received most sunlight. As seen from room A and B, most of the measured voltages were closed to linier trend line but in room C most of the measured voltages were far from linier trend line. The red linier trend line had the most flat slope and it meant that the measured voltages in room B were the most stable compared to the other rooms. The sensor output voltages tended to increase over time for each room from 10 mV to 40 mV. As seen in Figure 9, the sensor output voltages from room A and B were higher and more stable than the voltages in room C, but the voltage stability was worse than with air media. The sensor output voltages tended to increase over time for each room with water media from 40 mV to 90 mV. As seen from room A, most of the measured voltages were closed to linier trend line but in room C most of the measured voltages were far from linier trend line. The blue linier trend line had the most flat slope but most of the measured voltages were far from the blue linier trend line. The voltage stability of room B was better than the other rooms. Room B was the room without window, sunlight did not interfere the measurements. As seen from Figure 10, the sensor output voltages from room B were higher and more stable compared to voltages in room A and C. The voltages tended to increase over time for each room with glucose solution media from 30 mV to 80 mV. With air, water and glucose solution media the sensor output voltages tend to increase time by time, we suspected this was caused by noises in the photodiode components. From thisexperiment it can be seen that the measurement in room B with covered windows produced the most stable output voltages. The TIA and sensor average output voltages for room A, B and C is shown in Table 1. The first row of Table 1 showed that with air media, there was no measured effective voltage on the TIA and sensor output voltages. It meant that noises did not appear with air media for the three rooms. Air media did not absorb the NIR light from the NIR LED and in these experiments the DC output voltages from room A were higher than room B and C. The sensor average output voltage for the three rooms with air media was 4.6825 V. The second row of Table 1 showed that with water media, there were effective voltages on the TIA output voltages; it means that noises did appear with water media for the three rooms. The effective output voltages were the 50 Hz interference from the main power supply [24]. High frequency noises also appeared on the oscilloscope with very small voltage. After passing through the filters, the effective output voltages from the sensor were
  • 7. TELKOMNIKA Telecommun Comput El Control  Effects of noises on near infrared sensor for blood glucose level measurement (Kiki Prawiroredjo) 1465 reduced to zero. These sensor DC output voltages from the three rooms were lower almost half of the output voltages from air media. This result was consistent with our previous result [25]. Water absorbed NIR light and caused noises [26]. The sensor average output voltage for the three rooms with water media was 2.2809 V. The third row of Table 1 showed that with 200 mg/dL glucose solution media, there were effective voltages on the TIA output which meant noises did appear with glucose solution media for the three rooms. After passing through the filter the effective output voltages from the sensor were reduced to zero. The effective output voltages were a bit higher than the effective voltages with water media for room A and B; it meant that noises with glucose solution were higher than with water media [25]. The sensor average output voltage for the three rooms with glucose solution media was 2.3368 V. There was voltage difference between sensor average output voltage with water and glucose solution media from the three rooms and it could be concluded that the sensor had the potential as a blood glucose level meter. 4. CONCLUSION In order to find out the effect of noises on NIR sensor, we studied noise sources in measuring glucose level with NIR sensor over time and measured the output voltage of the sensor in three different conditions with three different media between the light source and the sensor. From Table 1 can be seen that the filters reduced the effective output voltages to zero and it meant that the 50 Hz and high frequency noises could be removed with filters. The stability of the measurement output voltages from the three different places depended on the environmental conditions especially sunlight. The sensor output voltages of the 3-room tended to increase over time and we suspected it was from the noises in the photodiode component. Hence, this sensor can only be used in a short time and suitable only in a room without sunlight. In comparison with the other researches generally did not count on the sunlight effect. Based on the data presented on Table 1, there was voltage difference between average sensor output voltage with water media and average sensor output voltage with glucose solution media from room A, B and C. It could be concluded that the sensor had the potential as a blood glucose level meter. In our future research, we target to put the photodiode on an ear clip to be applied to an earlobe, process the sensor output signal, and develop software to display the measurement result on a smartphone. ACKNOWLEDGMENT We would like to express our gratitude to Trisakti University and DRPM, Ministry of Research, Technology and Higher Education of the Republic of Indonesia, Grant Number 201/A/LPT/USAKTI/V/2019, for funding this research. REFERENCES [1] WHO, “Global Report on Diabetes,” World Health Organization, 2016. [2] International Diabetes Federation, “IDF Diabetes Atlas, 8 ed,” International Diabetes Federation, 2017. [3] K. J. Schlager, “Non-Invasive Near Infrared Measurement of Blood Analyte Concentrations,” US4882492A. (Patent), 1988. [4] N. A. Salam, et al., “The Evolution of Non-invasive Blood Glucose Monitoring System for Personal Application,” Journal of Telecommunication, Electronics and Computer Engineering, vol. 8, no. 1, pp. 59-65, 2016. [5] K. Maruo and M. Oka, “Method of Determining a Glucose Concentration in a Target by Using an Near Infrared Spectroscopy,” US5957841A (Patent), 1998. [6] G. Acosta, et al., “Compact Apparatus for Non-Invasive Measurement of Glucose Through Near-Infrared Spectroscopy,” CA2476421A1. (Patent), 2003, https://patents.google.com/patent/CA2476421A1/en [7] Z. Zhao and R. A. Myllyla, “Photoacoustic Blood Glucose and Skin Measurement Based on Optical Scattering Effect,” Proceedings of Saratov Fall Meeting, vol. 4707, 2001. [8] R. D. Rosenthal, “Method for Providing Custom Calibration for Near-Infrared Instruments for Measurement of Blood Glucose,” US5068536A. (Patent), 1991. [9] S. Haxha and J. Jhoja, “Optical Based Noninvasive Glucose Monitoring Sensor Prototype,” IEEE Photonics Journal,vol. 8, no. 6, pp. 1-11, Dec 2016. [10] D. Sia, “Design of A Near Infrared Device for The Study of Glucose Concentration Measurements,”EE4BI6 Electrical Engineering Biomedical Capstones, McMaster University, 2010. [11] P. S. Reddy and K. Jyostna, “Development of Smart Insulin Device for Non Invasive Blood Glucose Level Monitoring,” in IEEE 7th International Advance Computing Conference, pp. 516-519, 2017. [12] K. A. U. Menon, et al., “A Survey on Non-Invasive Blood Glucose Monitoring Using NIR,” in International Conference on Communication and Signal Processing, India, 2013. [13] M. Pande and A. Joshi, “Non-Invasive Blood Glucose Measurement,” Proceeding of Int J ComputEng Res, vol. 5, I. 4 pp. 26-28, 2015, http://www.ijceronline.com/papers/Vol5_issue4/E054026028.pdf [14] J. P. Coffin, et al., “Photodiode Detector with Integrated Noise Shielding,”US6184521B1. (Patent),1998.
  • 8.  ISSN:1693-6930 TELKOMNIKA Telecommun Comput El Control, Vol. 18, No. 3, June 2020: 1459-1466 1466 [15] J. Poeze, et al., “Interference Detector for Patient Monitor,” US 8754776B2. (Patent), 2013. [16] M. Doğan and A. Tangel, “Shot Noise Measurement in PIN Photodiodes,” International Journalof Computing, Communication and Instrumentation Engg., vol. 3, no. 2, pp. 297-300, 2016. [17] C. D. Bobade and M. S. Patil, “Non-invasive Blood Glucose Level Monitoring System for Diabetic Patients using Near-Infrared Spectroscopy,” American Journal of Computer Science and Information Technology, vol. 4, no. 01, pp. 1-8, 2016. [18] Jeon K. J., et al., “Comparison Between Transmittance and Reflectance Measurements in Glucose Determination Using Near Infrared Spectroscopy,” Journalof Biomedical Optics, vol. 11, no. 1, 2006. [19] E. S. Julian, et al., “The Model of Near Infrared Sensor Output Voltage as a Function of Glucose Concentration in Solution,” in The 15th International Conference on Quality in Research (QIR), Bali, pp. 146-149, 2017. [20] K. Prawiroredjo, et al., “Noise reduction in near infrared-based glucose concentration measurement sensor,” in IOP Conference Series: Materials Science and Engineering, vol. 532, Jun 2019. [21] Z. M. Yosuf, et al.,“Design and Fabrication of Cost-Effective Heart-Rate Pulse Monitoring Sensor System,” TELKOMNIKA Telecommunication Computing Electronics and Control, vol. 17, no. 5, pp. 2497-2504, Oct 2019. [22] B. G. Irianto, et al., “A low-cost electro-cardiograph machine equipped with sensitivity and paper speed option,” TELKOMNIKA Telecommunication Computing Electronics and Control, vol. 17, no. 3, pp. 1275-1281, Jun 2019. [23] J. J. Burmeister and M. A. Arnold, “Evaluation of Measurement Sites for Noninvasive Blood Glucose Sensing with Near-Infrared Transmission Spectroscopy,” Clinical Chemistry, vol. 45, no. 9, pp. 1621-1627, 1999. [24] A. C. M. van Rijn, et al., “High-Quality Recording of Bioelectric Events. Part 1 Interference Reduction, Theory and Practice,” Medical and Biological Engineering and Computer, vol. 28, no. 5, pp. 389-397, Oct 1990. [25] K. Prawiroredjo and E. S. Julian, “Comparative study of 940 nm and 1450 nm near infrared sensor for glucose concentration monitoring,” TELKOMNIKA Telecommunication Computing Electronics and Control, vol. 17, no. 2, pp. 981-985, Apr 2019. [26] K. Palmer and D. Williams, “Optical properties of water in the near infrared,” Journal of the Optical Societyof America, vol. 64, no. 8, pp. 1107-1110, Aug 1974.