3. INTRODUCTION
Intra Ocular Pressure : IOP is the pressure within the
eyeball
Intra Ocular Tension : IOT is the pressure exerted by the
intraocular contents on the outer coats of eye
Normal IOP : It is that pressure that does not lead to
glaucomatous damage of the optic nerve head.
Normal mean IOP = 15.5 +/- 2.57 mmHg & 2 SD above
mean is 20.5 mmHg
4. TECHNIQUES FOR MEASURING IOP
Techniques
Manometry
Indentation
1. Von Graefe
2. Schiotz
Instrumental Digital
Direct Indirect (Tonometry)
Applanation
Non Contact
Contact
1. Grolman NCT
2. Ocular response
analyzer
Variable force Variable area
Contour
Pascal’s DCT
Maklakov
1. Goldmann
2. Perkins
6. MANOMETRY
Catheter is inserted into AC of eye
Other end is connected to a manometric device
MOST accurate method
NOT feasible in humans (invasive procedure)
7. DIGITAL TONOMETRY
Response of eyeball to pressure applied by pulp of finger
METHOD: rest both hands on patient’s forehead &
alternately apply just enough pressure on the globe (above
the superior tarsal plate) to indent it slightly with 1 index
finger while feeling the compliance with the other
Indents easily/ firm to touch
NOT accurate (subjective)
8. DIGITAL TONOMETRY
Advantages
• Easiest to perform
• No equipments
• No anesthesia
• No stain
• Estimation of IOP with irregular corneas, where
applanation tonometry is not possible.
Disadvantages
• Reading inaccurate
• Subjective
• Over-estimation or under-estimation of IOP
9. AN IDEAL TONOMETER
Accurate & reasonable IOP measurement
Convenient to use
Simple to calibrate
Stable from day to day
Easier to standardize
Free of maintenance problems
10. INDENTATION TONOMETRY
Shape of deformation -
TRUNCATED CONE
Precise shape - variable &
Unpredictable
Displace large intraocular
Volume
Conversion tables based
on empirical data used
Prototype - Schiøtz
tonometer
APPLANATION TONOMETRY
Shape of deformation -
FLATTENING
Precise shape – constant
Displace small intraocular
Volume
Mathematical calculations
for IOP
Differentiated on the basis
of variable measured
11. SCHIØTZ TONOMETER
Handle : to hold the instrument
Foot plate : rests on the cornea
Plunger : moves freely within a
shaft in the foot plate
Bent lever : short arm and long
arm acting as a pointer needle
The degree to which the plunger
indents the cornea is indicated by
the movement of this needle on a
scale
Weights : 5.5 g weight is
permanently fixed to the plunger,
which can be increased to 7.5 and
10gm.
12. PRINCIPLE
The weight of tonometer on the eye increases the
actual IOP (Po) to a higher level (Pt)
The change in pressure from Po to Pt is an expression
of the resistance of the eye (scleral rigidity) to the
displacement of fluid
P(t) = P(o) + E
IOP with Tonometer in position
Pt = Actual IOP Po + Scleral Rigidity E
Because the tonometer actually measures Pt, it is
necessary to estimate Po for each scale reading & weight.
13.
14. On the basis of,
Friedenwald formula, a set of
conversion tables for IOP was
made
The indicated scale reading
and the plunger weight are
converted to an IOP
measurement
More the plunger indents
the cornea, higher the scale
reading and lower the IOP
Each scale unit represents
0.05 mm protrusion of the
plunger
15. GOLDMANN APPLANATION
TONOMETER (1954)
Goldmann based his concept of tonometry on the
Modified Imbert- Fick Law
W + S = PA1 + B
When A1 = 7.35 mm2 S balances B & W equals P
This internal area of applanation is obtained when diam.
of external area of applanation is 3.06 mm
Volume of displacement produced by applanating this
area is approx. 0.50 mm3
So P is very close to actual IOP & ocular rigidity does
not significantly influence measurements
16. GOLDMANN APPLANATION
TONOMETER (1954)
Applanation tonometry is based on the Imbert-Fick
principle, which states that the pressure (p) inside an ideal
dry, thin walled sphere equals the force (F) necessary to
flatten its surface divided by the area of the flattening (A).
Cornea being aspherical, wet, and slightly inflexible fails
to follow the law.
Moisture creates surface tension (S) or capillary attraction
of tear film for tonometry head.
Lack of flexibility requires force to bend the cornea (B)
which is independent of internal pressure.
The central thickness of cornea is about 0.55 mm and the
outer area of corneal flattening differs from the inner area of
flattening (A1). It is this inner area which is of importance.
17. The pressure (P) of a body of fluid encapsulated within
a sphere is directly proportional to the force (W) required
to applanate an area (A) of the sphere:
W = PA
IMBERT – FICK PRINCIPLE (1885)
18. W + S = PA1 + B
W = tonometer force
S = surface tension of pre-corneal tear film
P = intra-ocular pressure
A1 = inner corneal area of applanation
B = corneal rigidity
MODIFIED IMBERT-FICK’S LAW
20. The two beam-splitting prism within the
applanating unit optically convert the circular area of
corneal contact into 2 semicircles
BIPRISM PROBE
21. The fluorescent semicircles
are viewed through the biprism
and the force against the cornea
is adjusted until the inner edges
overlap.
22. FALSELY LOW IOP
Inadequate flourescein
thin cornea
corneal edema
with the rule astigmatism
(1mm Hg per 4 D)
prolonged contact
Repeated tonometry
FALSELY HIGH IOP
Excess flourescein
thick cornea
steep cornea
against the rule astigmatism
(1mm Hg per 3D)
wider meniscus
Widening the lid fissure
excessively
Elevating the eyes more
than 15°
23. POTENTIAL ERRORS
Thin cornea
Thick cornea
Astigmatism > 3D
Irregular cornea
Inadequate fluorescein
Too much fluorescein
Tonometer out of calibration
Repeated tonometry
Elevating eyes > 15°
Pressing on the eyelids or globe
Squeezing of the eyelids
Observer bias (expectations and even Numbers)
24. PERKINS TONOMETER (1965)
Developed by ES Perkins
Hand-held applanation tonometer
Uses same prism tips as GAT
The prism is illuminated by battery powered bulbs
Advantages Over GAT:
• Portable & can be used in any position of pt.
• Infants/ children
• in the O.T.
• at the bedside for non-ambulatory pt.
25. GROLMAN NCT (1972)
Introduced by Grolman in 1972
NCT has 3 subsystems:
1. Alignment system: It aligns patient’s eye in 3
dimensions (axial/ vertical/ lateral)
2. Optoelectronic applanation monitoring system:
a. Transmitter directs a collimated beam of light at
corneal apex
b. Receiver & detector accept only parallel coaxial
rays of light reflected from cornea
c. Timer measures from an internal reference to
the point of peak light intensity
3. Pneumatic system: It generates a puff of room air
directed against cornea
27. GROLMAN NCT (1972)
Advantages
• Screening procedure
• Can be operated by non-medical personnel
• No anesthesia required
• No contamination
• No chance of corneal abrasion
Limitations
• IOP is near normal, accuracy decreases with increase
in IOP & in eyes with abnormal cornea or poor fixation
• Sub-epithelial air bubbles after repeated use of NCT
(rare)
28. PULSAIR TONOMETER (1986)
New Non Contact Tonometer
Keeler Pulsair Intellipuff is a portable hand-held tonometer
Launched in European market in April 2007
Based on the same principle as the Grolman NCT
29. TONOPEN
The most commonly used Mackay-Marg type tonometer
today is the Tono-Pen
FDA approved in March 2006
Portable handheld instrument with a strain gauge that
creates an electrical signal as the footplate flattens the
cornea
Micro-strain gauge technology
Computerised battery - operated pocket tonometer
Instrument is 18 cm in length and weighs 60 g
Converts IOP into electric waves
Wave form is internally analyzed by a microprocessor
Average of 3 to 6 readings of IOP
30. As the area of applanation of the Tonopen is smaller
than GAT (2.36 mm2 Vs 7.35mm2) therefore, theoretically
the difference between applanating pressure & IOP is
reduced due to reduced corneal resistance of a smaller
contact area
It is particularly useful in community eye camps, on
ward rounds ,children, irregular surfaces, measuring
through an amniotic membrane patch graft, to read from
the sclera
A disposable latex cover which is discarded after each
use provides infection control
31. PASCAL’S DYNAMIC CONTOUR
TONOMETER (2003)
First totally new concept in tonometry was described by
Kanngiesser et al. in 2005
Based on principle of contour matching
Principle: By surrounding and matching the contour of
a sphere (or a portion thereof ), the pressure on the outside
equals the pressure on the inside
Designed to reduce the influence of biomechanical
properties of cornea on measurement of IOP
32.
33. Cup-like plastic device with contour matched tip
Concave surface of radius 10.5 mm, which approx. to
the shape of a normal cornea when pressure on both sides
is equal
Probe is placed in contact with cornea with constant
force of 1 g
The integrated piezo-electric pressure sensor
automatically begins to acquire data
Measures IOP 100 times per second
A complete measurement cycle requires about 8 sec of
contact time
The device also measures the variation in pressure that
occurs with cardiac cycle (Ocular Pulse Amplitude)
34. DCT is more accurate than Goldmann tonometry &
pneumo-tonometer
Not affected by corneal thickness
IOP is not altered by corneal refractive surgery that
thins the cornea
DCT tells us ocular pulse amplitude
OPA may be indicative of the status of ocular blood flow
(low OPA = low ocular blood flow)
OPA is increased over normal ( 1.2 – 4 mmHg) in most
forms of glaucoma & may be related to the level of IOP
35. CONCLUSION
Elevated IOP has been known to be associated with
glaucoma for over a millenium
All our instruments give us an indirect measure of IOP
Most of the newer devices can measure higher IOPs
accurately but we are often interested in lower IOPs as
well (which is best measured by GAT)
Despite all the limitations, GAT remains the GOLD
STANDARD for IOP measurement.