Lecture 7 dental x ray film, processing and processing errors lecture 7
1. Oral Radilogy
Dr. Rawand Samy Mohamed Abu Nahla
Oral Medicine, periodontology& oral Radiology Department.
Dr. Haydar. A. Shafy Faculty Of Dentistry.Dr. Haydar. A. Shafy Faculty Of Dentistry.
El Azhar University.El Azhar University.
3. Introduction
• Film-processing procedures have a direct effect
on the quality of a radiograph.
Latent Image Formation
• When the radiation interacts with the silver halide
crystals in the film emulsion, the image on the film
is produced.
• The image, which is not visible before processing,
is called the latent image.
4. Fig. 1 Scanning electron micrograph of unprocessed emulsion
of Kodak Ultra Speed dental film (5000× magnification).
Note white-appearing unexposed silver bromide grains.
5. Learning Objectives
• Describe the care and maintenance of the processing
solutions, equipment, and equipment accessories
used in manual and automatic film processing.
• Discuss the requirements necessary for the
darkroom.
• List and identify the component parts of an
automatic film processor.
• Describe common time and temperature errors
during film processing.
6. Learning Objectives
• Describe chemical-contamination errors made
during film processing.
• Describe film-handling errors that can occur during
film processing.
• Describe some common lighting errors made during
film processing.
7. Dental X-Ray Film Processing
• Processing is a series of steps that changes the latent
image on the exposed film into a radiograph by
producing a visible image on the film.
• Proper processing is just as important as exposure technique
in producing diagnostic-quality radiographs.
• Radiographs that are non diagnostic because of poor
processing techniques must be retaken, exposing the patient
to unnecessary radiation.
• In many practices intraoral films are processed in an
automatic processor; however, it is still necessary to know
how to process the film manually.
8. The Five Steps in Processing
Dental Radiographs
• Development
• Rinsing
• Fixation
• Washing
• Drying
9. Developing
• Developing is the first step in processing films.
• A chemical solution called the developer is used.
• The purpose of the developer is to chemically
reduce the exposed silver halide crystals to black
metallic silver.
• The developer solution also softens the film
emulsion during this process.
10. Rinsing
• Rinsing of the films is necessary to remove the
developer from the film so that the development
process stops.
• Usually, agitating the film rack for 20 seconds is
sufficient.
• This must be done under safelight conditions.
11. Fixing
• The acidic fixing solution removes the unexposed silver
halide crystals from the film emulsion.
• The fixer also hardens the film emulsion during this process.
• For permanent fixation, the film is kept in the fixer for a
minimum of 10 minutes.
• However, films may be removed from the fixing solution
after 3 minutes for viewing.
• Films that are not properly fixed will fade and turn brown in
a short time.
• Leaving films in the fixer for a long time can remove the
image from the film.
12. Washing
• After fixation, a water bath is used to wash the film.
• The washing step requires about 20 minutes to
thoroughly remove all excess chemicals from the
emulsion.
13. Drying
• The final step in film processing is the drying of the
films.
• Films may be air-dried at room temperature in a
dust-free area or placed in a heated drying cabinet.
• Films must be completely dried before they can be
handled for mounting and viewing.
19. The Darkroom
• The term “light-tight” is often used to describe the
darkroom.
• To be light-tight, no light leaks can be present.
• When you are in the darkroom with the light turned
off, no white light should be visible.
• X-ray film is extremely sensitive to visible white
light.
• Any leakage of white light can cause film fog. A
fogged film appears dull gray, lacks contrast, and is
nondiagnostic.
20. Types of Darkroom Lighting
• Room lighting: An overhead white light provides
adequate lighting for tasks such as cleaning,
restocking of materials, and mixing of chemicals.
• Safelighting: A safelight is a low-intensity light in
the red-orange spectrum. Safelighting provides
enough illumination in the darkroom to process
films safely without exposure of or damage to the
film.
21. Types of Darkroom Lighting
• There must be a a safe distance between the light
and the working area, and the person developing the
film must work quickly to keep the exposure to the
safelight as short as possible.
• Unwrapped films that are left too close to the
safelight or exposed to the safelight for more than 2
to 3 minutes appear fogged.
• A safelight must be placed a minimum of 4 feet
away from the film and working area.
22. Fig. 3 A distance of at least 4 feet must separate the
safelight from the working area.
23. The Processing Tank
• Manual processing is a method that is used to
develop, rinse, fix, and wash dental x-ray films.
• The essential piece of equipment required for
manual processing is a processing tank.
• The processing tank is divided into compartments to
hold the developer solution, water bath, and fixer
solution.
• A processing tank has two insert tanks and one
master tank.
24. Fig. 4 Processing tanks showing developing and fixing
tanks inserts in bath of running water.
25. Requirements for a Darkroom
• Cleanliness at all times
• Infection-control items (e.g., gloves, disinfectant
spray, paper towels)
• Container, labeled with a biohazard label, for
contaminated film packets or barriers
• Recycle container for pieces of lead foil pieces,
which should not be thrown in the trash
• Light-tightness
• Processing tanks for the developer and fixer solution
and a circulating-water bath
26. • Running water with mixing valves to adjust the temperature
• Both a safelight and a source of white (normal) light
• Accurate timer
• Accurate floating thermometer
• Stirring rods or paddles to mix the chemicals and equalize
the temperatures of the solutions
• Safe storage space for chemicals
• Film hangers
• Film-drying rack and film dryer
27. The Automatic Processor
• Automatic film processing is a fast and simple
method used to process dental x-ray films.
• Other than opening the film packet, all steps of film
processing are handled by the automatic processor.
• Automatic film processing requires only 4 to 6
minutes for the development, fixing, washing, and
drying of a film, whereas manual processing
techniques require approximately 1 hour.
28. • The automatic processor maintains the correct
temperature of the solutions and adjusts the
processing time.
• Proper maintenance of the automatic processor
reduces the chance of errors during film processing.
• Many dental offices that have automatic processors
still maintain manual processing equipment as a
standby if the automatic processor malfunctions.
29. Components of the Automatic
Processor
• The processor housing covers all of the component parts.
• The film feed slot is for the unwrapped films to be inserted
into the automatic processor.
• The roller film transporter is a system of rollers that rapidly
moves the film through the compartments.
• The developer and fixer compartments holds the solutions.
The film is transported directly from the developer into the
fixer without a rinsing step.
• The water compartment holds circulating water.
• The drying chamber holds heated air and dries the wet film.
30. Fig. 5 A, An automatic film processor. B, An automatic film
processor equipped with a daylight film loader.
A B
31. Advantages of Automatic Film
Processing
• Less processing time is required.
• Time and temperature are automatically controlled.
• Less equipment is used.
• Less space is required.
32. Processing Errors
• Processing errors may occur for a variety of
reasons, including:
• Time and temperature errors
• Chemical-contamination errors
• Film-handling errors
• Lighting errors
• The dental assistant must be able to recognize the
appearance of common processing errors and know what
to do to prevent such problems from occurring again.
33. The following photographs illustrate common
examples of technical errors that can occur during
film processing.
Such errors result in radiographs that are not
diagnostic, thus requiring retakes and
additional exposure to the patient.
Examples of Common Processing Errors