SlideShare a Scribd company logo
1 of 67
CT Image Quality Parameters
Spatial
Resolution
Image
Noise
Contrast
Resolution
Artifacts
Spatial Resolution
• Quantifies image blurring
• “Ability to discriminate objects of varying
density a small distance apart against a
uniform background”
• Minimum separation required between
two high contrast objects for them to be
resolved as two objects
Resolvable Object Size &
Limiting Resolution
• Smallest resolvable high contrast object
• Often expressed as line pairs / cm
• “Pair” is one object + one space
One
Pair
Geometric Factors affecting
Spatial Resolution
• Focal spot size
• detector aperture width
• slice thickness or collimation
–Less variation likely for
thinner slices
–attenuation variations within
a voxel are averaged
• partial volume effect
Geometric Unsharpness & CT
• Object details
distributed (blurred)
over several detectors
decrease spatial
resolution
• detector aperture size
–must be smaller than
object for object to be
resolved
Focal Spot
Detectors
Small Object
to be Imaged
Contrast Resolution
• Ability of an imaging system to
demonstrate small changes in tissue
contrast
• The difference in contrast necessary to
resolve 2 large areas in image as separate
structures
CT Image Noise
• Standard deviation of pixel values
S(xi - xmean)2
Noise (s) = -------------------
(n-1)
Xi = individual pixel value
Xmean = average of all pixel values in ROI
n =total # pixels in ROI
Noise Level
• Units
– CT numbers (HU’s)
or
– % contrast
• Example
– CT # range: 1000 HU’s
– Standard deviation: 3 HU’s
– Noise level is 3 or 3 / 1000 X 100 = 0.3%
CT Image Noise
• Fluctuation of CT
#’s in an image of
uniform material
(water)
• Usually described
as standard
deviation of pixel
values
Noise Measurement in CT
• Scan water phantom
• Select regions of interest
• Take mean & standard
deviation in each region
• Standard deviation is
noise in ROI
CT Spatial vs. Contrast Resolution
• Spatial & contrast resolution interact
– High contrast objects are easier to resolve
• One can improve one at the expense of the
other
• Can only improve both by increasing dose
Increasing
object size
Increasing contrast
Contrast & Detail
• Larger objects easy to see even at low contrast
Increasing
object size
Increasing contrast
Contrast & Detail
• Small objects only visible at high contrast
Increasing
object size
Increasing contrast
CT Number
• Calculated from reconstructed pixel
attenuation coefficient
(mt - mW)
CT # = 1000 X ------------
mW
Where:
ut = linear attenuation coefficient for tissue in pixel
uW = linear attenuation coefficient for water
Linearity
• Linear relationship of CT #’s to linear
attenuation coefficients of objects
• Checked with phantom of several
known materials
– average CT # of each material obtained
from ROI analysis
– Compare CT #’s with known coefficients
17
Acceptance Test:
• Performed upon newly installed equipment
to verify purchase specification and to
establish performance base line.
 Frequency:
• To be performed once
only when new
machine is installed
ACCEPTANCE TEST IN CT
• spatial
resolution
• contrast
resolution
• noise
• slice width
• kVp waveform
• average &
standard
deviation of
water phantom
CT #
• scatter & leakage
CT test Equipment
- water cylinder phantom
- System evaluation phantom
• high contrast resolution ,slice thickness ,laser
accuracy
• low contrast detectibility
• noise and field uniformity
- Head and Body phantom
Jun-20
CRHC 2013 Jaipur
20zzzz
Testing of CT number for water
Procedure
• water filled plastic cylinder ( 20 cm diameter)
• scan and reconstruct the image ,place ROI
200-300 pixels in centre of field
Tolerance limits
• +/- 3HU at the centre
• +/- 5 HU at peripheral location
• outside the image -1000HU +/- 5 HU
Testing of CT number accuracy for
water
Measurement of field uniformity
Testing of image Noise
Jun-20
CRHC 2013 Jaipur
24
At 100% contrast difference the size
of the bar/hole pattern that could be
resolvable should be 1.6 mm ( 3.12
lp/cm).
High contrast sensitivity
Testing of High contrast sensitivity
High contrast (spatial ) Resolution
High contrast (spatial ) Resolution
Low contrast resolution/Detectability
Jun-20
CRHC 2013 Jaipur
29
Low contrast sensitivity
5.0 mm at 1% contrast
difference (minimum)
Slice Thickness
Alignment of Laser light
CT Dose Measurement
(CTDI)
 20% of the
quoted value
(Expected)
 40% of the
quoted value
(maximum)


T
T
sngle dzzDnTCTDI
7
7
)()/1(
CTDI100
• Computed tomography dose index 100
(CTDI100)


cm
cm
sngle dzzDNTCTDI
5
5
100 )()/1(
where N is the number of sections per scan, T is the
nominal width of each acquired section, f is the air
kerma to dose conversion factor, E is measured air kerma
and L is the pencil chamber length (100 mm)
LEfNTCTDI  )/1(100
CTDIvol
pitch/wvol CTDICTDI 
where pitch is the table distance traveled in a full
rotation / total collimated width of the beam
Dose-length Product
• Dose-length product (DLP)
• Units of DLP generally mGy-cm
• Analogous to dose-area-product (DAP) for
projection imaging
lengthscan volCTDIDLP
Jun-20 CRHC 2013 Jaipur 36
Jun-20
CRHC 2013 Jaipur
38
Periodic QA tests for Computed
Tomography (CT) unit
Daily
- indicators, mechanical and safety checks
Weekly
-Image noise , CT number of water
Monthly
CT Number uniformity
Contrast resolution
Spatial resolution
Jun-20
CRHC 2013 Jaipur
39
QA tests for Computed Tomography
(CT) unit Contd---
Semi-annual
• laser alignment
• slice thickness
• position accuracy
Annual
• patient dose measurement
• Beam HVL for 140 kVp and 120 kVp
• Area survey
ACR CT Accreditation Phantom
High contrast
resolution
4
Uniformity
& noise
Distance
Accuracy
& SSP
3
Low
contrast
resolution
Alignment
CT #
Slice width
2 1
Head Foot
Module 1
• Alignment, CT number accuracy, slice thickness
Module 1
• CT number accuracy
– Polyethylene: between -107 and -87 HU
– Water: between -7 and +7 HU (-5 and +5 HU
preferred)
– Acrylic: between +110 and +130 HU
– Bone: between +850 and +970 HU
– Air: between -1005 and -970 HU
• Slice thickness
– Measure at 7, 5, 3 and 1 mm (HRC)
– Must be accurate to within 1.5 mm
Slice Thickness
• Each wire is 0.5 mm slice thickness
• Count number of wires >50% as bright as the
center wire divide by 2
Module 2
• Low contrast resolution
0.6% difference
from background
Module 2
• All four 6-mm rods must be visible
• Record contrast
(no control limit)
Module 3
• Uniformity,
noise, distance,
section
sensitivity profile
Uniformity Control Limit
• Mean in all
edge ROIs
within 5 HU of
mean in center
ROI
• Center ROI
between -7 HU
and +7 HU (-5
HU and +5 HU
preferred)
• No artifacts
Module 4
• High contrast resolution (bar patterns)
– 4-12 lp/cm
High Contrast Resolution
• Control limit:
– Adult abdomen
protocol
• at least 5 lp/mm
– High resolution
chest protocol
• at least 6 lp/mm
Catphan
The Phantom Laboratory, http://www.phantomlab.com/catphan.html
Catphan 600 Modules
• CTP404 Slice width, CT #, pixel size
• CTP591 Bead geometry
• CTP528 Line pair resolution, point source
• CTP515 Low contrast resolution
• CTP486 Uniformity
CPT404
• Slice width, CT number accuracy, pixel size
Slice Width
• Slice width = FWHM(wire ramp) x tan (23)
CT Number Accuracy
Material CT Number
Air -1000
PMP -200
LDPE -100
Water 0
Polystyrene -35
Acrylic 120
Delrin 340
Teflon 990
CPT591
• Bead geometry
– Calculation of slice width and MTF
CPT528
• High contrast spatial resolution
– 1-21 lp/cm bar patterns
• 2 point sources
– 0.28 mm (subpixel)
– MTF measurement
CPT515
• Low contrast resolution targets 3 mm
5 mm
7 mm
Subslice
Supra-slice
CPT486
• Uniformity
– Profiles in horizontal and vertical directions
indicate
possible
cupping
or capping
Uniformity
Mammography Machine
Jun-20 61
Acceptance tests for
Mammography Unit
Focal Spot test
for f < 0.8 mm + 0.5 f
Accuracy of tube potential (kVp) ± 1 Kv
Accuracy of exposure timer ± 10 %
Linearity of mA loading station
Coefficient of mA linearity < 0.1
Linearity of timer
Coefficient of timer linearity < 0.1
Jun-20
CRHC 2013 Jaipur
62
Recommended values of HVL for 28 kVp
0.3 mm Al  HVL  0.37 mm Al
( for Mo target & Mo filter )
Recommended values of HVL for 49 kVp
0.8 mm Al  HVL  1.0 mm Al
(for Mo target & Rh filter )
1.1 mm Al  HVL  1.6 mm Al
(for Mo target & Al filter)
Al equivalence of the  0.1 mm of Al
compression device  2.0 mm of
plexi glass
Imaging Setup
Compression Paddle
Compressed Breast
Table
Grid
Image Detector
Chest Wall
X-ray Source
X-ray Beam
ACR Accreditation Phantom
(RMI 156 )
RMI 156 Phantom
ACR minimum score is 4 visible fibers, 3 speck groups and 3 masses
SETUP For Entrance Skin Exposure
Measurement
Ionization
Chamber Breast
Equivalent
Phantom
X-Ray Beam
Table Film
Film
Compression
Paddle
Entrance Skin Exposure Measurement
Dosimeter
Mammography
Phantom

More Related Content

What's hot

AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxZubairUlIslam5
 
Generations of CT scanner
Generations of  CT scannerGenerations of  CT scanner
Generations of CT scannersadhamhussain52
 
K Space in MRI
K Space in MRIK Space in MRI
K Space in MRIKajal Jha
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptxdypradio
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxThejaTej6
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomographyBeuniquewithNehaSing
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexVivek Elangovan
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose IndexAnjan Dangal
 
Mri coils
Mri coilsMri coils
Mri coilsSelf
 
Ct quality control
Ct quality controlCt quality control
Ct quality controlSAJIDEJAZ1
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyTarun Goyal
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2mr_koky
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scanMohd Aiman Azmardi
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)Upakar Paudel
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct pptAnand Rk
 
Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)globalsoin
 

What's hot (20)

AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptxAREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
AREA MONITORING DEVICES BY ZUBAIRUL ISLAM.pptx
 
Generations of CT scanner
Generations of  CT scannerGenerations of  CT scanner
Generations of CT scanner
 
Quality control
Quality  control Quality  control
Quality control
 
K Space in MRI
K Space in MRIK Space in MRI
K Space in MRI
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose Index
 
Mri coils
Mri coilsMri coils
Mri coils
 
Ct quality control
Ct quality controlCt quality control
Ct quality control
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and Fluoroscopy
 
Filters
FiltersFilters
Filters
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
 
MRI artifacts
MRI artifactsMRI artifacts
MRI artifacts
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct ppt
 
Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)
 

Similar to CT Image Quality Parameters Explained

Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedyYashawant Yadav
 
Image Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptxImage Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptxivanKeshari
 
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSPERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSBhuvi palaniswamy
 
Computed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalComputed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalSambasivaselli R
 
Physics_of_CT , CT machine and it’s parts, ct generations
Physics_of_CT , CT machine and it’s parts, ct generationsPhysics_of_CT , CT machine and it’s parts, ct generations
Physics_of_CT , CT machine and it’s parts, ct generationsNeetuMeena56
 
QC of CT using Catphan 600
QC of CT using Catphan 600QC of CT using Catphan 600
QC of CT using Catphan 600Asim Gilani
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.pptMITS
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.pptNDShaikh
 
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
PARAMETERS AND ASSOCIATED  Trade-Offs, MRIPARAMETERS AND ASSOCIATED  Trade-Offs, MRI
PARAMETERS AND ASSOCIATED Trade-Offs, MRIBeuniquewithNehaSing
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfAngelina Protik
 

Similar to CT Image Quality Parameters Explained (20)

Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedy
 
Image Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptxImage Quality And Artifacts in Computed Tomography.pptx
Image Quality And Artifacts in Computed Tomography.pptx
 
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERSPERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
PERFORMANCE EVALUATION OF COMPUTED TOMOGRAPHY (CT) SCANNERS
 
Ayaz_portfolio
Ayaz_portfolioAyaz_portfolio
Ayaz_portfolio
 
Computed tomography - Physics and Clinical
Computed tomography - Physics and ClinicalComputed tomography - Physics and Clinical
Computed tomography - Physics and Clinical
 
Sensors
SensorsSensors
Sensors
 
Physics_of_CT , CT machine and it’s parts, ct generations
Physics_of_CT , CT machine and it’s parts, ct generationsPhysics_of_CT , CT machine and it’s parts, ct generations
Physics_of_CT , CT machine and it’s parts, ct generations
 
QC of CT using Catphan 600
QC of CT using Catphan 600QC of CT using Catphan 600
QC of CT using Catphan 600
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
PPT Sensors.ppt
PPT Sensors.pptPPT Sensors.ppt
PPT Sensors.ppt
 
Sensors.ppt
Sensors.pptSensors.ppt
Sensors.ppt
 
Sensors
SensorsSensors
Sensors
 
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
PARAMETERS AND ASSOCIATED  Trade-Offs, MRIPARAMETERS AND ASSOCIATED  Trade-Offs, MRI
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 

More from Self

Image formation of Radiology from Radiation
Image formation of Radiology from RadiationImage formation of Radiology from Radiation
Image formation of Radiology from RadiationSelf
 
Tomographic equipment
Tomographic equipmentTomographic equipment
Tomographic equipmentSelf
 
Multisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographyMultisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographySelf
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopySelf
 
Aerb guidelines for Room installation
Aerb guidelines for Room installationAerb guidelines for Room installation
Aerb guidelines for Room installationSelf
 
Barium swallow examination
Barium swallow examinationBarium swallow examination
Barium swallow examinationSelf
 
Roles and responsibility of Radiology professional
Roles and responsibility of Radiology professionalRoles and responsibility of Radiology professional
Roles and responsibility of Radiology professionalSelf
 
Dexa and its aplication
Dexa and its aplicationDexa and its aplication
Dexa and its aplicationSelf
 
Various Pelvic lines
Various Pelvic linesVarious Pelvic lines
Various Pelvic linesSelf
 
Chest decubitus
Chest decubitusChest decubitus
Chest decubitusSelf
 
Pelvimetry
PelvimetryPelvimetry
PelvimetrySelf
 
Planning and organization of Radiology Department
Planning and organization of Radiology DepartmentPlanning and organization of Radiology Department
Planning and organization of Radiology DepartmentSelf
 
Aims of rp
Aims of rpAims of rp
Aims of rpSelf
 

More from Self (13)

Image formation of Radiology from Radiation
Image formation of Radiology from RadiationImage formation of Radiology from Radiation
Image formation of Radiology from Radiation
 
Tomographic equipment
Tomographic equipmentTomographic equipment
Tomographic equipment
 
Multisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographyMultisection tomography and Transaxial Tomography
Multisection tomography and Transaxial Tomography
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopy
 
Aerb guidelines for Room installation
Aerb guidelines for Room installationAerb guidelines for Room installation
Aerb guidelines for Room installation
 
Barium swallow examination
Barium swallow examinationBarium swallow examination
Barium swallow examination
 
Roles and responsibility of Radiology professional
Roles and responsibility of Radiology professionalRoles and responsibility of Radiology professional
Roles and responsibility of Radiology professional
 
Dexa and its aplication
Dexa and its aplicationDexa and its aplication
Dexa and its aplication
 
Various Pelvic lines
Various Pelvic linesVarious Pelvic lines
Various Pelvic lines
 
Chest decubitus
Chest decubitusChest decubitus
Chest decubitus
 
Pelvimetry
PelvimetryPelvimetry
Pelvimetry
 
Planning and organization of Radiology Department
Planning and organization of Radiology DepartmentPlanning and organization of Radiology Department
Planning and organization of Radiology Department
 
Aims of rp
Aims of rpAims of rp
Aims of rp
 

Recently uploaded

APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon AUnboundStockton
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991RKavithamani
 
_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting Data_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting DataJhengPantaleon
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxNirmalaLoungPoorunde1
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13Steve Thomason
 
PSYCHIATRIC History collection FORMAT.pptx
PSYCHIATRIC   History collection FORMAT.pptxPSYCHIATRIC   History collection FORMAT.pptx
PSYCHIATRIC History collection FORMAT.pptxPoojaSen20
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxpboyjonauth
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
Introduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher EducationIntroduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher Educationpboyjonauth
 
Separation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesSeparation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesFatimaKhan178732
 
How to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxHow to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxmanuelaromero2013
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 

Recently uploaded (20)

APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
 
Crayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon ACrayon Activity Handout For the Crayon A
Crayon Activity Handout For the Crayon A
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
 
_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting Data_Math 4-Q4 Week 5.pptx Steps in Collecting Data
_Math 4-Q4 Week 5.pptx Steps in Collecting Data
 
Employee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptxEmployee wellbeing at the workplace.pptx
Employee wellbeing at the workplace.pptx
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13
 
PSYCHIATRIC History collection FORMAT.pptx
PSYCHIATRIC   History collection FORMAT.pptxPSYCHIATRIC   History collection FORMAT.pptx
PSYCHIATRIC History collection FORMAT.pptx
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
Introduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptxIntroduction to AI in Higher Education_draft.pptx
Introduction to AI in Higher Education_draft.pptx
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Introduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher EducationIntroduction to ArtificiaI Intelligence in Higher Education
Introduction to ArtificiaI Intelligence in Higher Education
 
Separation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and ActinidesSeparation of Lanthanides/ Lanthanides and Actinides
Separation of Lanthanides/ Lanthanides and Actinides
 
How to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptxHow to Make a Pirate ship Primary Education.pptx
How to Make a Pirate ship Primary Education.pptx
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 

CT Image Quality Parameters Explained

  • 1.
  • 2. CT Image Quality Parameters Spatial Resolution Image Noise Contrast Resolution Artifacts
  • 3. Spatial Resolution • Quantifies image blurring • “Ability to discriminate objects of varying density a small distance apart against a uniform background” • Minimum separation required between two high contrast objects for them to be resolved as two objects
  • 4. Resolvable Object Size & Limiting Resolution • Smallest resolvable high contrast object • Often expressed as line pairs / cm • “Pair” is one object + one space One Pair
  • 5. Geometric Factors affecting Spatial Resolution • Focal spot size • detector aperture width • slice thickness or collimation –Less variation likely for thinner slices –attenuation variations within a voxel are averaged • partial volume effect
  • 6. Geometric Unsharpness & CT • Object details distributed (blurred) over several detectors decrease spatial resolution • detector aperture size –must be smaller than object for object to be resolved Focal Spot Detectors Small Object to be Imaged
  • 7. Contrast Resolution • Ability of an imaging system to demonstrate small changes in tissue contrast • The difference in contrast necessary to resolve 2 large areas in image as separate structures
  • 8. CT Image Noise • Standard deviation of pixel values S(xi - xmean)2 Noise (s) = ------------------- (n-1) Xi = individual pixel value Xmean = average of all pixel values in ROI n =total # pixels in ROI
  • 9. Noise Level • Units – CT numbers (HU’s) or – % contrast • Example – CT # range: 1000 HU’s – Standard deviation: 3 HU’s – Noise level is 3 or 3 / 1000 X 100 = 0.3%
  • 10. CT Image Noise • Fluctuation of CT #’s in an image of uniform material (water) • Usually described as standard deviation of pixel values
  • 11. Noise Measurement in CT • Scan water phantom • Select regions of interest • Take mean & standard deviation in each region • Standard deviation is noise in ROI
  • 12. CT Spatial vs. Contrast Resolution • Spatial & contrast resolution interact – High contrast objects are easier to resolve • One can improve one at the expense of the other • Can only improve both by increasing dose Increasing object size Increasing contrast
  • 13. Contrast & Detail • Larger objects easy to see even at low contrast Increasing object size Increasing contrast
  • 14. Contrast & Detail • Small objects only visible at high contrast Increasing object size Increasing contrast
  • 15. CT Number • Calculated from reconstructed pixel attenuation coefficient (mt - mW) CT # = 1000 X ------------ mW Where: ut = linear attenuation coefficient for tissue in pixel uW = linear attenuation coefficient for water
  • 16. Linearity • Linear relationship of CT #’s to linear attenuation coefficients of objects • Checked with phantom of several known materials – average CT # of each material obtained from ROI analysis – Compare CT #’s with known coefficients
  • 17. 17 Acceptance Test: • Performed upon newly installed equipment to verify purchase specification and to establish performance base line.  Frequency: • To be performed once only when new machine is installed
  • 18. ACCEPTANCE TEST IN CT • spatial resolution • contrast resolution • noise • slice width • kVp waveform • average & standard deviation of water phantom CT # • scatter & leakage
  • 19. CT test Equipment - water cylinder phantom - System evaluation phantom • high contrast resolution ,slice thickness ,laser accuracy • low contrast detectibility • noise and field uniformity - Head and Body phantom
  • 20. Jun-20 CRHC 2013 Jaipur 20zzzz Testing of CT number for water Procedure • water filled plastic cylinder ( 20 cm diameter) • scan and reconstruct the image ,place ROI 200-300 pixels in centre of field Tolerance limits • +/- 3HU at the centre • +/- 5 HU at peripheral location • outside the image -1000HU +/- 5 HU
  • 21. Testing of CT number accuracy for water
  • 22. Measurement of field uniformity
  • 24. Jun-20 CRHC 2013 Jaipur 24 At 100% contrast difference the size of the bar/hole pattern that could be resolvable should be 1.6 mm ( 3.12 lp/cm). High contrast sensitivity
  • 25. Testing of High contrast sensitivity
  • 26. High contrast (spatial ) Resolution
  • 27. High contrast (spatial ) Resolution
  • 29. Jun-20 CRHC 2013 Jaipur 29 Low contrast sensitivity 5.0 mm at 1% contrast difference (minimum)
  • 32. CT Dose Measurement (CTDI)  20% of the quoted value (Expected)  40% of the quoted value (maximum)   T T sngle dzzDnTCTDI 7 7 )()/1(
  • 33. CTDI100 • Computed tomography dose index 100 (CTDI100)   cm cm sngle dzzDNTCTDI 5 5 100 )()/1( where N is the number of sections per scan, T is the nominal width of each acquired section, f is the air kerma to dose conversion factor, E is measured air kerma and L is the pencil chamber length (100 mm) LEfNTCTDI  )/1(100
  • 34. CTDIvol pitch/wvol CTDICTDI  where pitch is the table distance traveled in a full rotation / total collimated width of the beam
  • 35. Dose-length Product • Dose-length product (DLP) • Units of DLP generally mGy-cm • Analogous to dose-area-product (DAP) for projection imaging lengthscan volCTDIDLP
  • 36. Jun-20 CRHC 2013 Jaipur 36
  • 37.
  • 38. Jun-20 CRHC 2013 Jaipur 38 Periodic QA tests for Computed Tomography (CT) unit Daily - indicators, mechanical and safety checks Weekly -Image noise , CT number of water Monthly CT Number uniformity Contrast resolution Spatial resolution
  • 39. Jun-20 CRHC 2013 Jaipur 39 QA tests for Computed Tomography (CT) unit Contd--- Semi-annual • laser alignment • slice thickness • position accuracy Annual • patient dose measurement • Beam HVL for 140 kVp and 120 kVp • Area survey
  • 40. ACR CT Accreditation Phantom High contrast resolution 4 Uniformity & noise Distance Accuracy & SSP 3 Low contrast resolution Alignment CT # Slice width 2 1 Head Foot
  • 41. Module 1 • Alignment, CT number accuracy, slice thickness
  • 42. Module 1 • CT number accuracy – Polyethylene: between -107 and -87 HU – Water: between -7 and +7 HU (-5 and +5 HU preferred) – Acrylic: between +110 and +130 HU – Bone: between +850 and +970 HU – Air: between -1005 and -970 HU • Slice thickness – Measure at 7, 5, 3 and 1 mm (HRC) – Must be accurate to within 1.5 mm
  • 43. Slice Thickness • Each wire is 0.5 mm slice thickness • Count number of wires >50% as bright as the center wire divide by 2
  • 44. Module 2 • Low contrast resolution 0.6% difference from background
  • 45. Module 2 • All four 6-mm rods must be visible • Record contrast (no control limit)
  • 46. Module 3 • Uniformity, noise, distance, section sensitivity profile
  • 47. Uniformity Control Limit • Mean in all edge ROIs within 5 HU of mean in center ROI • Center ROI between -7 HU and +7 HU (-5 HU and +5 HU preferred) • No artifacts
  • 48. Module 4 • High contrast resolution (bar patterns) – 4-12 lp/cm
  • 49. High Contrast Resolution • Control limit: – Adult abdomen protocol • at least 5 lp/mm – High resolution chest protocol • at least 6 lp/mm
  • 50. Catphan The Phantom Laboratory, http://www.phantomlab.com/catphan.html
  • 51. Catphan 600 Modules • CTP404 Slice width, CT #, pixel size • CTP591 Bead geometry • CTP528 Line pair resolution, point source • CTP515 Low contrast resolution • CTP486 Uniformity
  • 52. CPT404 • Slice width, CT number accuracy, pixel size
  • 53. Slice Width • Slice width = FWHM(wire ramp) x tan (23)
  • 54. CT Number Accuracy Material CT Number Air -1000 PMP -200 LDPE -100 Water 0 Polystyrene -35 Acrylic 120 Delrin 340 Teflon 990
  • 55. CPT591 • Bead geometry – Calculation of slice width and MTF
  • 56. CPT528 • High contrast spatial resolution – 1-21 lp/cm bar patterns • 2 point sources – 0.28 mm (subpixel) – MTF measurement
  • 57. CPT515 • Low contrast resolution targets 3 mm 5 mm 7 mm Subslice Supra-slice
  • 58. CPT486 • Uniformity – Profiles in horizontal and vertical directions indicate possible cupping or capping
  • 61. Jun-20 61 Acceptance tests for Mammography Unit Focal Spot test for f < 0.8 mm + 0.5 f Accuracy of tube potential (kVp) ± 1 Kv Accuracy of exposure timer ± 10 % Linearity of mA loading station Coefficient of mA linearity < 0.1 Linearity of timer Coefficient of timer linearity < 0.1
  • 62. Jun-20 CRHC 2013 Jaipur 62 Recommended values of HVL for 28 kVp 0.3 mm Al  HVL  0.37 mm Al ( for Mo target & Mo filter ) Recommended values of HVL for 49 kVp 0.8 mm Al  HVL  1.0 mm Al (for Mo target & Rh filter ) 1.1 mm Al  HVL  1.6 mm Al (for Mo target & Al filter) Al equivalence of the  0.1 mm of Al compression device  2.0 mm of plexi glass
  • 63. Imaging Setup Compression Paddle Compressed Breast Table Grid Image Detector Chest Wall X-ray Source X-ray Beam
  • 65. RMI 156 Phantom ACR minimum score is 4 visible fibers, 3 speck groups and 3 masses
  • 66. SETUP For Entrance Skin Exposure Measurement Ionization Chamber Breast Equivalent Phantom X-Ray Beam Table Film Film Compression Paddle
  • 67. Entrance Skin Exposure Measurement Dosimeter Mammography Phantom