Radiation Safety Training Module: Diagnostic Radiology
Quality Assurance Criteria for Mammography X-ray Equipment
Radiological Safety Division
Atomic Energy Regulatory Board
CONTENTS
• Expected questions to know after studying this lecture
• Objective/Aim of QA
• Mammography
• Parameters Affecting Image Quality in Mammography Equipment
• Quality Assurance Tests in Mammography Equipment
• Survey of the Mammography Installation
• Summary/Conclusion
• Questions and Answers
• References and sources for additional information
Expected questions to know after studying this lecture
• What do you mean by Quality Assurance in Mammography?
• Why the QA tests are required?
• Name of the QA tests for Mammography equipment ?
• How the leakage radiation of x-ray tube is measured?
• What do you mean of survey of Mammography x-ray installation and how
will you ensure whether Mammography x-ray installation is safe from
radiation safety view point?
Objective / Aim of Quality Assurance (QA)
 Optimum image quality with minimum possible dose
to the patient
Why QA in Mammography?
• Frequent use and aging of the diagnostic X-ray equipment
may cause change in its functional characteristics.
• To avoid retakes which reduces the radiation exposure to patient and
radiological personnel and public.
Parameters Affecting Image Quality
• Operating Potential (kV)
• Operating Tube (mA)
• Exposure Time (msec)
• Effective Focal Spot Size
• Total Filtration
• Detector characteristics
Mammography is a specific type of
imaging unit that uses a low-dose
X-ray system to examine breasts.
What is Mammography
A mammography exam,
called a mammogram, is
used to aid in the early
detection and diagnosis of
breast diseases in women.
Digital
Mammogram
Normal
Mammogram
Mammography
X-ray unit
Development of Mammography
• In 1913, radiographic appearance of breast cancers was first
reported
• 1950’s – Industrial grade X-ray film used
• 1960’s – Xerography introduced – much lower dose
• 1975 – High speed/resolution film introduced
• 1992 – Mammography Quality Standard Test (MQSA)
implemented
Risk Vs. Benefit
• Mortality risk from mammography induced radiation is 5
deaths/1 million patients using screen film mammography. (Pl.
give reference)
• It is more risky to refuse mammography than to allow it.
Breast Cancer Screening
• American Cancer Society and American College of Radiology
suggest that all women over 50 years should undergo mammography
examinations annually.
• Women between 40 and 49 should have one mammography every or
every other year.
• A baseline mammogram should be done at the onset of menopause.
Tissue Equivalent Phantom
for Mammography
Al filter
Multi -parameter
measuring Instrument
Survey meter
Mammography X-ray unit
Parameters Tested are:
• Correspondence between X-ray Field and Image Reception Area
• Effective Focal Spot Size Measurement
• Accelerating Potential (kVp)
• Accuracy of Exposure Time
• Linearity of mA Loading Stations
• Linearity of Timer Loading Stations
• Radiation Out put consistency
• Total Filtration x-ray Tube
Cont…
• Calibration of Compression Device
• Imaging Performance Evaluation
• Constancy of the Automatic Exposure Control (AEC) on Film
Screen Receptors
• Radiation Leakage through Tube Housing and Collimator
1. CORRESPONDENCE BETWEEN X-RAY FIELD
AND IMAGE RECEPTION AREA
• X-ray field to be positioned to cover the region of interest and
the sensitive volumes of the Automatic Exposure Control
• X-ray field is adjusted in normal use for full coverage of the
image reception area
• Result : within ______ mm
CORRESPONDENCE BETWEEN X-RAY FIELD AND IMAGE
RECEPTION AREA (Contd..)
Tolerance: The X-ray field
• shall extend to the edge of the patient support that is designed to be
adjacent to the chest wall of the patient and shall not extend beyond this
edge by more than 5 mm.
• shall not extend by more than 2% of the direct focal distance beyond all
edges of the image receptor area.
(IS13450(Part 2/Sec 45) :2007 Clause 29.203.4, page 22)
2. EFFECTIVE FOCAL SPOT SIZE MEASUREMENT
The ability for resolving the smallest size of the image (i.e detail) in a radiograph depends on the focal
spot size.
Techniques for the measurement of focal spot
• Pin hole
• Slit
• Star pattern or resolution pattern
• Minimum focal spot to skin distance is 20 cm
in mammography with geometric magnification
Tolerance:
Nominal Focal Spot Value Focal Spot Dimension (mm)
f Width Length
0.1 0.1-0.15 0.1-0.15
0.15 0.15-0.23 0.15-0.23
0.2 0.2-0.3 0.2-0.3
0.25 0.25-0.38 0.25-0.38
0.3 0.3-0.45 0.3-0.45
0.4 0.4-0.46 0.4-0.46
Ref: Table 5 IEC 336 :1993 Section 5, page 43
Accelerating Potential (kVp)
The peak potential of the x-ray generator affects quality of the x-ray beam
and exposure to the patient.
Presently solid – state detectors, which employ non-invasive method for
peak tube potential measurement, are quite handy for this test.
Accelerating Potential (kVp)
Tolerance :  1 kV
Accuracy of Exposure Time
If the exposure time of the x-ray unit is not in order, the
radiograph can be under exposed or overexposed.
For this absolute timer method is adopted by measuring set and
measured time with digital timers.
Accuracy of exposure timer ± 10 %
Linearity of mA loading stations
• The tube current (mA) is equal to the number of electrons flowing from the
cathode to the anode per unit time.
• The exposure of the beam for a given kVp and filtration is proportional to the
tube current.
• This test is carried out to check the linearity of radiation output
with respect to change in tube current (mA) stations by keeping
timer station constant at a particular kV station.
Linearity of mA loading stations (contd..)
• Keeping exposure time and kVp constant, radiation output is measured at
different mA stations.
• Measurement for mA loading station is to be repeated for a number of times each
to eliminate statistical errors.
• Each mA loading station readings are averaged and the coefficient of linearity
(CoL) is evaluated from average (mR/mAs) or (mGy/mAs) as follows:
Tolerance: Coefficient of Linearity  0.1
Dose measuring instrument
Coefficient of linearity =
(X max -Xmin )  (Xmax + X min)
Linearity of timer loading stations
• The exposure time is the duration of X-ray production.
• Keeping the kVp and mA constant, radiation output is measured at different timer stations and
Coefficient of Linearity is evaluated.
• This test is carried out to check the linearity of radiation output with respect to change
in timer stations by keeping mA station constant at a particular kV station.
Tolerance: Coefficient of timer linearity < 0.1
• To check the constancy of radiation output.
• Arrange the X-ray source assembly, the diaphragm and the radiation detector for
measurement under narrow beam condition without the compression plate.
• Ensure that the radiation quality of the X-ray beam emerging from the X-ray source
assembly complies with applicable specified conditions for normal use.
RADIATION OUTPUT CONSISTENCY
IS13450(Part 2/Sec 45) :2007 Clause 29.1.105, page 21
• If no such conditions are specified, ensure
that the total filtration in the X-ray source
assembly is such as to comply with the
general standard as applicable
Out put Consistency
• Keeping fixed mA and time, radiation output is measured at various
available kVp stations. average(X) of mR/mAs (mGy/mAs) is
calculated.
• Consistency at each kVp station is checked by Evaluating the
coefficient of variation.
• Coefficient of variation =
Tolerance : The coefficient of variation of measured values of air kerma
shall not be greater than 0.05 for any combination of loading factors
 
1
X
-
Xi
1
2


n
X
Total Filtration of X-ray tube
• To cutoff low energy components from X-ray beam, which do not contribute
to diagnostic image formation but result in unnecessary patient exposure.
• If the filtration is too high, image contrast will be poor and unit will be
overloaded. Therefore it is necessary that the total filtration (inherent + added)
provided for the X-ray tube be as per the recommended value.
• The determination of half value thickness (HVT) of the X-ray beam is the
method of evaluation of total aluminum equivalent filtration of the X-ray tube.
Evaluation of HVL and Total Filtration
0
10
20
30
40
50
60
70
80
90
100
0 1 2 3 4 5
Thickness of added Aluminium filters (mm)
Transmission
(%)
• Total aluminum equivalent filtration of the x-ray tube is evaluated by
determining the half value thickness of the beam.
• Transmission curve of the x-ray beam can be prepared by plotting a graph
between the absorber thickness and corresponding percentage transmission.
• The absorber thickness for 50 % transmission will be the half value
thickness of the x-ray beam.
• Total aluminum filtration could be determined from HVT using calibration
tables.
• Presently, new solid state detectors directly gives the values of HVT and
total filtration used in diagnostic X-ray equipment.
TOTAL FILTRATION AND ALUMINUM EQUIVALENCE OF THE
COMPRESSION DEVICE (contd..)
Operating
Potential
mAs Added filter (mm Al) Compression
Device
HVT mm
Al
30kVp 0.0 0.1 0.2 0.3 0.4 0.5
40kVp
IEC 601-1-3:1994 section 29.201.3, page 33age 29
TOTAL FILTRATION AND ALUMINIUM EQUIVALENCE OF THE
COMPRESSION DEVICE
Tolerance:
Total filtration in mammography equipment with Mo-Mo target filter combination
shall not be less than 0.03 mm of Mo. For any other combination the above values of
first HVL will apply.
The minimum total filtration equivalent to first HVL shall be provided as
First HVL at 30 kVp ≥ 0.3 mm Al
First HVL at 40 kVp ≥ 0.4 mm Al
First HVL at 50 kVp ≥ 0.5 mm Al
CALIBRATION OF COMPRESSION DEVICE
• Test Equipment
• A soft rubber block, 20 mm to 50 mm thick and 100 mm to 120 mm long wide
• Test Procedure:
• Position the X-ray tube-image receptor assembly to acquire a cranio-caudal projection of the breast.
• Place the force balance on the patient support and place the soft rubber block on the sensitive area of the
force balance.
• Operate the compression device, clamping the soft rubber block and record the reading.
• Measure the highest achievable forces for all compression modes.
• Repeat the procedure for at least three other orientations.
IS13450(Part 2/Sec 45) :2007 Clause 22.102.5, page 16
Cranio-caudal (CC)
Projection
The CC view is
particularly valuable in
detecting small valuable
cancer
IS13450(Part 2/Sec 45) :2007 Clause 22.102.5,
page 16
Tolerance :
• No compression device shall be able to apply a force exceeding 300 N.
• For power-driven compression, the compression device shall be able to
apply a force of at least 150 N and it shall be unable to apply a force
exceeding 200 N.
• For power-driven compression, the available operating force shall be
adjustable down to 70 N or less.
• If the value of applied force is displayed, the indication shall be accurate
to ±20 N.
IS13450(Part 2/Sec 45) :2007 Clause 22.102.5, page 16
CALIBRATION OF COMPRESSION DEVICE (Contd..)
IMAGING PERFORMANCE EVALUATION
Tolerance:
To pass the mammography image
quality standards at least
 four fibers,
 three calcification group and
 three masses
must be clearly visible (With no
oblivious artifacts) at an average
glandular dose of less than 3 mGy
CONSTANCY OF THE AUTOMATIC EXPOSURE
CONTROL (AEC) ON FILM SCREEN RECEPTORS
• TEST METHOD
Measure the transmitted doses through the phantoms made of breast tissue
equivalent material, water or polymethyl methacrylate (PMMA), produced with
AEC in operation.
• Tolerance: AEC shall be designed such that the transmitted dose shall not vary by
more than 20% for all phantom thicknesses.
Breast imaging phantom
PMMA Breast Phantom
Radiation Leakage Levels through
Housing and Collimator
Radiation Leakage through Tube Housing
• The radiation leakage measurement is carried with an ionization
radiation survey meter. For checking the leakage radiation, the
collimator of the tube housing is fully closed and the tube is
energized at maximum rated tube potential and current at that
kVp.
• The exposure rate at one meter from the target is measured at
different locations (anode side, cathode side, front back and
top) from the tube housing and collimator.
• From the maximum leakage rate (mR/h) from both tube housing and collimator,
leakage radiation in one hour is computed on the basis of the workload of the
unit.
• Workload used for medical X-ray unit :
40 mA-min in 1 hour
• Max leakage from tube housing
= 40 mAmin X ----Max. leakage mR/hr
60 min X ----- mA used for measurement
Codal Requirements:
The leakage radiation averaged over an area of 10
cm2, with no linear dimension greater than 20 cm
and located at 5 cm from any point on the external
surface of X-ray tube housing does not exceed 0.02
mGy in one hour.
Survey of the X-Ray Installation
• Estimation of Dose Levels from measured dose rate values:
• If x (mR/hr) is the exposure level measured at a point with I (mA)
current and W (mA-min / wk) is the work load of the unit, then the
weekly exposure level at that point can be calculated as follows:
• Exposure level at the point = x mR / hr with I mA
• Weekly exposure = Y * W mR
• Weekly exposures due to all the three beam directions are added to get the
total weekly exposure at any location
• Permissible limit:
• For location of Radiation Worker:
• 20 mSv in a year (40 mR/week)
• For Location of Member of Public: 1 mSv in a year (2mR/week)
y
mA
mR
I
x
hr
mA
mR
I
x







min
60
*
CONCLUSION:
• With all this discussion we can conclude that QA is
• very important in diagnostic radiology facilities to confirm
the real diagnostic information with minimum dose to
patient.
Summary
Parameters affecting image quality
• Operating Potential
• Current/Time (mAs)
• Focal Spot Size
• Filtration
• Detector characteristics
Additional Information
Ready values to remember
Sr No Name of QA test Tolerance
1 kVp  1 kV
2 Accuracy of time  10%
3 mA /mAs / Timer linearity (coeff. of linearity) 0.1
4 Output consistency (Coeff. of variation) 0.05
5 Image Quality  four fibers,
 three calcification group and
 three masses
6 Tolerance for filtration Total filtration for Mo-Mo filter 0.03 mm of
Mo
First HVL at 30 kVp ≥ 0.3 mm Al
First HVL at 40 kVp ≥ 0.4 mm Al
First HVL at 50 kVp ≥ 0.5 mm Al
7 Leakage from X-ray tube 0.02 mGy in 1 hour
Questions and Answers (Min 10)
• Q- What are the x-ray unit operating parameter on which output of x-ray unit depends?
Ans- kVp, mA and time
• Q -What is the tolerance of kV
A -  1 kV
• Q- What is the tolerance for leakage radiation
A – 0.02 mGy in one hour
• Q- What is the unit of work-load
A – mA-min per week
Questions and Answers
• Q- What is the role of filtration in x-ray unit?
A - To cutoff low energy components from X-ray beam, which do not contribute to
diagnostic image formation but result in unnecessary patient exposure.
• Q- What is the role of focal spot in image quality?
A- The ability for resolving the smallest size of the image (i.e detail) in a
radiograph depends on the focal spot size. The effective focal spot size is the length
and width of the focal spot as projected down the central ray in the X-ray field.
• Q. What is the criteria for passing image Quality standard
A. To pass the mammography image quality standards at least four fibers, three calcification group and three
masses must be clearly visible (With no oblivious artifacts) at an average glandular dose of less than 3 mGy.
• Q.Tolerance for total HVL in mammography x-ray tube?
A.- Total filtration in mammography equipment with Mo-Mo target filter combination shall not be less than 0.03
mm of Mo. For any other combination the above values of first HVL will apply. The minimum total filtration
equivalent to first HVL shall be provided as
First HVL at 30 kVp ≥ 0.3 mm Al
First HVL at 40 kVp ≥ 0.4 mm Al
First HVL at 50 kVp ≥ 0.5 mm Al
References and sources for additional information:
• IS 13450 (Part 2/Sec 45) : 2007 / IEC 60601-2-45: 2001 ‘Medical electrical equipment Part 2/45: Particular
requirements for the safety of mammographic X-ray equipment and mammographic stereotactic devices’
 IEC 60601-2-28 : 1993 Medical electrical equipment - Part 2-28: Particular requirements for the safety of X-ray source
assemblies and X-ray tube assemblies for medical diagnosis
• IEC 61223-3-2: 1996 Evaluation and routine testing in medical imaging departments - Part 3-2: Acceptance tests -
Imaging performance of mammographic X-ray equipment
• IEC 60601-1-3 (1994) General requirements for the safety; Section 3 Collateral Standard: General requirements for
radiation protection in diagnostic X-ray equipment
• IEC 60788:1984, Medical radiology - Terminology
• AERB Safety Code for Medical Diagnostic X-ray Equipment and Installations, AERB/SC/MED-2 (Rev.1)
• For additional information please refer to AERB website: www.aerb.gov.in
• http://www.aerb.gov.in/AERBPortal/pages/English/t/XRay/forms/QAMammography.doc
List of presentations in the training Module
Basics of Diagnostic X-ray Equipment
Biological effects of Radiations
Medical X-ray imaging techniques
Planning of Diagnostic X-ray facilities
Quality Assurance of X-ray equipment
Quality Assurance of Computed Tomography equipment
Radiation Protection in Diagnostic Radiology Practice
Causes, prevention and investigation of excessive exposures in
diagnostic radiology
Regulatory Requirements for Diagnostic Radiology Practice
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf

More Related Content

What's hot

Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyRamzee Small
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageSHASHI BHUSHAN
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose IndexAnjan Dangal
 
Emergency radiography
Emergency radiographyEmergency radiography
Emergency radiographyVignesh Viki
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x raySushilPattar
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexVivek Elangovan
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and techniqueArchana Koshy
 
Significance of scatter radiation
Significance of scatter radiationSignificance of scatter radiation
Significance of scatter radiationBikash Subedi
 
Quality Assurance on Diagnostic Equipment/Accessories
Quality Assurance on Diagnostic Equipment/AccessoriesQuality Assurance on Diagnostic Equipment/Accessories
Quality Assurance on Diagnostic Equipment/AccessoriesTheoNtiri1
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionparthajyotidas11
 
Ct quality control
Ct quality controlCt quality control
Ct quality controlSAJIDEJAZ1
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its careSaruGosain
 

What's hot (20)

Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
Quality assurance
Quality assuranceQuality assurance
Quality assurance
 
Viewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic imageViewing and recording the fluoroscopic image
Viewing and recording the fluoroscopic image
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose Index
 
Emergency radiography
Emergency radiographyEmergency radiography
Emergency radiography
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
X ray grids
X ray gridsX ray grids
X ray grids
 
Mammography physics and technique
Mammography  physics and techniqueMammography  physics and technique
Mammography physics and technique
 
Significance of scatter radiation
Significance of scatter radiationSignificance of scatter radiation
Significance of scatter radiation
 
Fluoroscopy
 Fluoroscopy  Fluoroscopy
Fluoroscopy
 
Quality Assurance on Diagnostic Equipment/Accessories
Quality Assurance on Diagnostic Equipment/AccessoriesQuality Assurance on Diagnostic Equipment/Accessories
Quality Assurance on Diagnostic Equipment/Accessories
 
X-Ray Generator
X-Ray GeneratorX-Ray Generator
X-Ray Generator
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Ct quality control
Ct quality controlCt quality control
Ct quality control
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
QUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptxQUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptx
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
 
CT artifact
CT artifact CT artifact
CT artifact
 

Similar to QUALITY-ASSURANCE---MAMMOGRAPHY.pdf

QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.pptBBAdhikari
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentationMumba Chilimboyi
 
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxAALIA ABDULLAH
 
Radiation protection course for radiologists L8
Radiation protection course for radiologists L8Radiation protection course for radiologists L8
Radiation protection course for radiologists L8Amin Amin
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancementYashawant Yadav
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image qualityRad Tech
 
-Lecture 2.ppt
-Lecture 2.ppt-Lecture 2.ppt
-Lecture 2.pptMagde Gad
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptJgdishrathi
 
factors_affecting_xray_production.pdf
factors_affecting_xray_production.pdffactors_affecting_xray_production.pdf
factors_affecting_xray_production.pdfJisni Cheruvathoor
 
Commissioning of Truebeam LINAC
Commissioning of Truebeam LINACCommissioning of Truebeam LINAC
Commissioning of Truebeam LINACVIneeth C
 
Chapter 3 Radiation Characteristics
Chapter 3 Radiation CharacteristicsChapter 3 Radiation Characteristics
Chapter 3 Radiation CharacteristicsKatieHenkel1
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scanMohd Aiman Azmardi
 
Dosimetric evaluation of the MLCs for irregular shaped radiation fields
Dosimetric evaluation of the MLCs for irregular shaped radiation fieldsDosimetric evaluation of the MLCs for irregular shaped radiation fields
Dosimetric evaluation of the MLCs for irregular shaped radiation fieldsIOSR Journals
 

Similar to QUALITY-ASSURANCE---MAMMOGRAPHY.pdf (20)

QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
Quality control
Quality  control Quality  control
Quality control
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.ppt
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
 
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
 
Radiation protection course for radiologists L8
Radiation protection course for radiologists L8Radiation protection course for radiologists L8
Radiation protection course for radiologists L8
 
Mammography.pdf
Mammography.pdfMammography.pdf
Mammography.pdf
 
Mammographic equipment and its advancement
Mammographic equipment and its advancementMammographic equipment and its advancement
Mammographic equipment and its advancement
 
linac QA.pptx
linac QA.pptxlinac QA.pptx
linac QA.pptx
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
-Lecture 2.ppt
-Lecture 2.ppt-Lecture 2.ppt
-Lecture 2.ppt
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
factors_affecting_xray_production.pdf
factors_affecting_xray_production.pdffactors_affecting_xray_production.pdf
factors_affecting_xray_production.pdf
 
Commissioning of Truebeam LINAC
Commissioning of Truebeam LINACCommissioning of Truebeam LINAC
Commissioning of Truebeam LINAC
 
X ray generator basic
X ray generator basicX ray generator basic
X ray generator basic
 
Chapter 3 Radiation Characteristics
Chapter 3 Radiation CharacteristicsChapter 3 Radiation Characteristics
Chapter 3 Radiation Characteristics
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
Tomotherapy
TomotherapyTomotherapy
Tomotherapy
 
Dosimetric evaluation of the MLCs for irregular shaped radiation fields
Dosimetric evaluation of the MLCs for irregular shaped radiation fieldsDosimetric evaluation of the MLCs for irregular shaped radiation fields
Dosimetric evaluation of the MLCs for irregular shaped radiation fields
 
Helical Tomotherapy
Helical TomotherapyHelical Tomotherapy
Helical Tomotherapy
 

Recently uploaded

Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13Steve Thomason
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 
Mastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionMastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionSafetyChain Software
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 

Recently uploaded (20)

Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
Mastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionMastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory Inspection
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 

QUALITY-ASSURANCE---MAMMOGRAPHY.pdf

  • 1. Radiation Safety Training Module: Diagnostic Radiology Quality Assurance Criteria for Mammography X-ray Equipment Radiological Safety Division Atomic Energy Regulatory Board
  • 2. CONTENTS • Expected questions to know after studying this lecture • Objective/Aim of QA • Mammography • Parameters Affecting Image Quality in Mammography Equipment • Quality Assurance Tests in Mammography Equipment • Survey of the Mammography Installation • Summary/Conclusion • Questions and Answers • References and sources for additional information
  • 3. Expected questions to know after studying this lecture • What do you mean by Quality Assurance in Mammography? • Why the QA tests are required? • Name of the QA tests for Mammography equipment ? • How the leakage radiation of x-ray tube is measured? • What do you mean of survey of Mammography x-ray installation and how will you ensure whether Mammography x-ray installation is safe from radiation safety view point?
  • 4. Objective / Aim of Quality Assurance (QA)  Optimum image quality with minimum possible dose to the patient
  • 5. Why QA in Mammography? • Frequent use and aging of the diagnostic X-ray equipment may cause change in its functional characteristics. • To avoid retakes which reduces the radiation exposure to patient and radiological personnel and public.
  • 6. Parameters Affecting Image Quality • Operating Potential (kV) • Operating Tube (mA) • Exposure Time (msec) • Effective Focal Spot Size • Total Filtration • Detector characteristics
  • 7. Mammography is a specific type of imaging unit that uses a low-dose X-ray system to examine breasts. What is Mammography A mammography exam, called a mammogram, is used to aid in the early detection and diagnosis of breast diseases in women. Digital Mammogram Normal Mammogram Mammography X-ray unit
  • 8. Development of Mammography • In 1913, radiographic appearance of breast cancers was first reported • 1950’s – Industrial grade X-ray film used • 1960’s – Xerography introduced – much lower dose • 1975 – High speed/resolution film introduced • 1992 – Mammography Quality Standard Test (MQSA) implemented
  • 9. Risk Vs. Benefit • Mortality risk from mammography induced radiation is 5 deaths/1 million patients using screen film mammography. (Pl. give reference) • It is more risky to refuse mammography than to allow it.
  • 10. Breast Cancer Screening • American Cancer Society and American College of Radiology suggest that all women over 50 years should undergo mammography examinations annually. • Women between 40 and 49 should have one mammography every or every other year. • A baseline mammogram should be done at the onset of menopause.
  • 11. Tissue Equivalent Phantom for Mammography Al filter Multi -parameter measuring Instrument Survey meter Mammography X-ray unit
  • 12. Parameters Tested are: • Correspondence between X-ray Field and Image Reception Area • Effective Focal Spot Size Measurement • Accelerating Potential (kVp) • Accuracy of Exposure Time • Linearity of mA Loading Stations • Linearity of Timer Loading Stations • Radiation Out put consistency • Total Filtration x-ray Tube
  • 13. Cont… • Calibration of Compression Device • Imaging Performance Evaluation • Constancy of the Automatic Exposure Control (AEC) on Film Screen Receptors • Radiation Leakage through Tube Housing and Collimator
  • 14. 1. CORRESPONDENCE BETWEEN X-RAY FIELD AND IMAGE RECEPTION AREA • X-ray field to be positioned to cover the region of interest and the sensitive volumes of the Automatic Exposure Control • X-ray field is adjusted in normal use for full coverage of the image reception area • Result : within ______ mm
  • 15. CORRESPONDENCE BETWEEN X-RAY FIELD AND IMAGE RECEPTION AREA (Contd..) Tolerance: The X-ray field • shall extend to the edge of the patient support that is designed to be adjacent to the chest wall of the patient and shall not extend beyond this edge by more than 5 mm. • shall not extend by more than 2% of the direct focal distance beyond all edges of the image receptor area. (IS13450(Part 2/Sec 45) :2007 Clause 29.203.4, page 22)
  • 16. 2. EFFECTIVE FOCAL SPOT SIZE MEASUREMENT The ability for resolving the smallest size of the image (i.e detail) in a radiograph depends on the focal spot size. Techniques for the measurement of focal spot • Pin hole • Slit • Star pattern or resolution pattern • Minimum focal spot to skin distance is 20 cm in mammography with geometric magnification
  • 17. Tolerance: Nominal Focal Spot Value Focal Spot Dimension (mm) f Width Length 0.1 0.1-0.15 0.1-0.15 0.15 0.15-0.23 0.15-0.23 0.2 0.2-0.3 0.2-0.3 0.25 0.25-0.38 0.25-0.38 0.3 0.3-0.45 0.3-0.45 0.4 0.4-0.46 0.4-0.46 Ref: Table 5 IEC 336 :1993 Section 5, page 43
  • 18. Accelerating Potential (kVp) The peak potential of the x-ray generator affects quality of the x-ray beam and exposure to the patient. Presently solid – state detectors, which employ non-invasive method for peak tube potential measurement, are quite handy for this test.
  • 20. Accuracy of Exposure Time If the exposure time of the x-ray unit is not in order, the radiograph can be under exposed or overexposed. For this absolute timer method is adopted by measuring set and measured time with digital timers. Accuracy of exposure timer ± 10 %
  • 21. Linearity of mA loading stations • The tube current (mA) is equal to the number of electrons flowing from the cathode to the anode per unit time. • The exposure of the beam for a given kVp and filtration is proportional to the tube current. • This test is carried out to check the linearity of radiation output with respect to change in tube current (mA) stations by keeping timer station constant at a particular kV station.
  • 22. Linearity of mA loading stations (contd..) • Keeping exposure time and kVp constant, radiation output is measured at different mA stations. • Measurement for mA loading station is to be repeated for a number of times each to eliminate statistical errors. • Each mA loading station readings are averaged and the coefficient of linearity (CoL) is evaluated from average (mR/mAs) or (mGy/mAs) as follows:
  • 23. Tolerance: Coefficient of Linearity  0.1 Dose measuring instrument Coefficient of linearity = (X max -Xmin )  (Xmax + X min)
  • 24. Linearity of timer loading stations • The exposure time is the duration of X-ray production. • Keeping the kVp and mA constant, radiation output is measured at different timer stations and Coefficient of Linearity is evaluated. • This test is carried out to check the linearity of radiation output with respect to change in timer stations by keeping mA station constant at a particular kV station. Tolerance: Coefficient of timer linearity < 0.1
  • 25. • To check the constancy of radiation output. • Arrange the X-ray source assembly, the diaphragm and the radiation detector for measurement under narrow beam condition without the compression plate. • Ensure that the radiation quality of the X-ray beam emerging from the X-ray source assembly complies with applicable specified conditions for normal use. RADIATION OUTPUT CONSISTENCY IS13450(Part 2/Sec 45) :2007 Clause 29.1.105, page 21 • If no such conditions are specified, ensure that the total filtration in the X-ray source assembly is such as to comply with the general standard as applicable
  • 26. Out put Consistency • Keeping fixed mA and time, radiation output is measured at various available kVp stations. average(X) of mR/mAs (mGy/mAs) is calculated. • Consistency at each kVp station is checked by Evaluating the coefficient of variation. • Coefficient of variation = Tolerance : The coefficient of variation of measured values of air kerma shall not be greater than 0.05 for any combination of loading factors   1 X - Xi 1 2   n X
  • 27. Total Filtration of X-ray tube • To cutoff low energy components from X-ray beam, which do not contribute to diagnostic image formation but result in unnecessary patient exposure. • If the filtration is too high, image contrast will be poor and unit will be overloaded. Therefore it is necessary that the total filtration (inherent + added) provided for the X-ray tube be as per the recommended value. • The determination of half value thickness (HVT) of the X-ray beam is the method of evaluation of total aluminum equivalent filtration of the X-ray tube.
  • 28. Evaluation of HVL and Total Filtration 0 10 20 30 40 50 60 70 80 90 100 0 1 2 3 4 5 Thickness of added Aluminium filters (mm) Transmission (%)
  • 29. • Total aluminum equivalent filtration of the x-ray tube is evaluated by determining the half value thickness of the beam. • Transmission curve of the x-ray beam can be prepared by plotting a graph between the absorber thickness and corresponding percentage transmission. • The absorber thickness for 50 % transmission will be the half value thickness of the x-ray beam. • Total aluminum filtration could be determined from HVT using calibration tables. • Presently, new solid state detectors directly gives the values of HVT and total filtration used in diagnostic X-ray equipment.
  • 30. TOTAL FILTRATION AND ALUMINUM EQUIVALENCE OF THE COMPRESSION DEVICE (contd..) Operating Potential mAs Added filter (mm Al) Compression Device HVT mm Al 30kVp 0.0 0.1 0.2 0.3 0.4 0.5 40kVp IEC 601-1-3:1994 section 29.201.3, page 33age 29
  • 31. TOTAL FILTRATION AND ALUMINIUM EQUIVALENCE OF THE COMPRESSION DEVICE Tolerance: Total filtration in mammography equipment with Mo-Mo target filter combination shall not be less than 0.03 mm of Mo. For any other combination the above values of first HVL will apply. The minimum total filtration equivalent to first HVL shall be provided as First HVL at 30 kVp ≥ 0.3 mm Al First HVL at 40 kVp ≥ 0.4 mm Al First HVL at 50 kVp ≥ 0.5 mm Al
  • 32. CALIBRATION OF COMPRESSION DEVICE • Test Equipment • A soft rubber block, 20 mm to 50 mm thick and 100 mm to 120 mm long wide • Test Procedure: • Position the X-ray tube-image receptor assembly to acquire a cranio-caudal projection of the breast. • Place the force balance on the patient support and place the soft rubber block on the sensitive area of the force balance. • Operate the compression device, clamping the soft rubber block and record the reading. • Measure the highest achievable forces for all compression modes. • Repeat the procedure for at least three other orientations. IS13450(Part 2/Sec 45) :2007 Clause 22.102.5, page 16
  • 33. Cranio-caudal (CC) Projection The CC view is particularly valuable in detecting small valuable cancer IS13450(Part 2/Sec 45) :2007 Clause 22.102.5, page 16
  • 34. Tolerance : • No compression device shall be able to apply a force exceeding 300 N. • For power-driven compression, the compression device shall be able to apply a force of at least 150 N and it shall be unable to apply a force exceeding 200 N. • For power-driven compression, the available operating force shall be adjustable down to 70 N or less. • If the value of applied force is displayed, the indication shall be accurate to ±20 N. IS13450(Part 2/Sec 45) :2007 Clause 22.102.5, page 16 CALIBRATION OF COMPRESSION DEVICE (Contd..)
  • 35. IMAGING PERFORMANCE EVALUATION Tolerance: To pass the mammography image quality standards at least  four fibers,  three calcification group and  three masses must be clearly visible (With no oblivious artifacts) at an average glandular dose of less than 3 mGy
  • 36. CONSTANCY OF THE AUTOMATIC EXPOSURE CONTROL (AEC) ON FILM SCREEN RECEPTORS • TEST METHOD Measure the transmitted doses through the phantoms made of breast tissue equivalent material, water or polymethyl methacrylate (PMMA), produced with AEC in operation. • Tolerance: AEC shall be designed such that the transmitted dose shall not vary by more than 20% for all phantom thicknesses. Breast imaging phantom PMMA Breast Phantom
  • 37. Radiation Leakage Levels through Housing and Collimator
  • 38. Radiation Leakage through Tube Housing • The radiation leakage measurement is carried with an ionization radiation survey meter. For checking the leakage radiation, the collimator of the tube housing is fully closed and the tube is energized at maximum rated tube potential and current at that kVp. • The exposure rate at one meter from the target is measured at different locations (anode side, cathode side, front back and top) from the tube housing and collimator.
  • 39. • From the maximum leakage rate (mR/h) from both tube housing and collimator, leakage radiation in one hour is computed on the basis of the workload of the unit. • Workload used for medical X-ray unit : 40 mA-min in 1 hour • Max leakage from tube housing = 40 mAmin X ----Max. leakage mR/hr 60 min X ----- mA used for measurement
  • 40. Codal Requirements: The leakage radiation averaged over an area of 10 cm2, with no linear dimension greater than 20 cm and located at 5 cm from any point on the external surface of X-ray tube housing does not exceed 0.02 mGy in one hour.
  • 41. Survey of the X-Ray Installation • Estimation of Dose Levels from measured dose rate values: • If x (mR/hr) is the exposure level measured at a point with I (mA) current and W (mA-min / wk) is the work load of the unit, then the weekly exposure level at that point can be calculated as follows: • Exposure level at the point = x mR / hr with I mA
  • 42. • Weekly exposure = Y * W mR • Weekly exposures due to all the three beam directions are added to get the total weekly exposure at any location • Permissible limit: • For location of Radiation Worker: • 20 mSv in a year (40 mR/week) • For Location of Member of Public: 1 mSv in a year (2mR/week) y mA mR I x hr mA mR I x        min 60 *
  • 43. CONCLUSION: • With all this discussion we can conclude that QA is • very important in diagnostic radiology facilities to confirm the real diagnostic information with minimum dose to patient.
  • 44. Summary Parameters affecting image quality • Operating Potential • Current/Time (mAs) • Focal Spot Size • Filtration • Detector characteristics
  • 45. Additional Information Ready values to remember Sr No Name of QA test Tolerance 1 kVp  1 kV 2 Accuracy of time  10% 3 mA /mAs / Timer linearity (coeff. of linearity) 0.1 4 Output consistency (Coeff. of variation) 0.05 5 Image Quality  four fibers,  three calcification group and  three masses 6 Tolerance for filtration Total filtration for Mo-Mo filter 0.03 mm of Mo First HVL at 30 kVp ≥ 0.3 mm Al First HVL at 40 kVp ≥ 0.4 mm Al First HVL at 50 kVp ≥ 0.5 mm Al 7 Leakage from X-ray tube 0.02 mGy in 1 hour
  • 46. Questions and Answers (Min 10) • Q- What are the x-ray unit operating parameter on which output of x-ray unit depends? Ans- kVp, mA and time • Q -What is the tolerance of kV A -  1 kV • Q- What is the tolerance for leakage radiation A – 0.02 mGy in one hour • Q- What is the unit of work-load A – mA-min per week
  • 47. Questions and Answers • Q- What is the role of filtration in x-ray unit? A - To cutoff low energy components from X-ray beam, which do not contribute to diagnostic image formation but result in unnecessary patient exposure. • Q- What is the role of focal spot in image quality? A- The ability for resolving the smallest size of the image (i.e detail) in a radiograph depends on the focal spot size. The effective focal spot size is the length and width of the focal spot as projected down the central ray in the X-ray field.
  • 48. • Q. What is the criteria for passing image Quality standard A. To pass the mammography image quality standards at least four fibers, three calcification group and three masses must be clearly visible (With no oblivious artifacts) at an average glandular dose of less than 3 mGy. • Q.Tolerance for total HVL in mammography x-ray tube? A.- Total filtration in mammography equipment with Mo-Mo target filter combination shall not be less than 0.03 mm of Mo. For any other combination the above values of first HVL will apply. The minimum total filtration equivalent to first HVL shall be provided as First HVL at 30 kVp ≥ 0.3 mm Al First HVL at 40 kVp ≥ 0.4 mm Al First HVL at 50 kVp ≥ 0.5 mm Al
  • 49. References and sources for additional information: • IS 13450 (Part 2/Sec 45) : 2007 / IEC 60601-2-45: 2001 ‘Medical electrical equipment Part 2/45: Particular requirements for the safety of mammographic X-ray equipment and mammographic stereotactic devices’  IEC 60601-2-28 : 1993 Medical electrical equipment - Part 2-28: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis • IEC 61223-3-2: 1996 Evaluation and routine testing in medical imaging departments - Part 3-2: Acceptance tests - Imaging performance of mammographic X-ray equipment • IEC 60601-1-3 (1994) General requirements for the safety; Section 3 Collateral Standard: General requirements for radiation protection in diagnostic X-ray equipment • IEC 60788:1984, Medical radiology - Terminology • AERB Safety Code for Medical Diagnostic X-ray Equipment and Installations, AERB/SC/MED-2 (Rev.1) • For additional information please refer to AERB website: www.aerb.gov.in • http://www.aerb.gov.in/AERBPortal/pages/English/t/XRay/forms/QAMammography.doc
  • 50. List of presentations in the training Module Basics of Diagnostic X-ray Equipment Biological effects of Radiations Medical X-ray imaging techniques Planning of Diagnostic X-ray facilities Quality Assurance of X-ray equipment Quality Assurance of Computed Tomography equipment Radiation Protection in Diagnostic Radiology Practice Causes, prevention and investigation of excessive exposures in diagnostic radiology Regulatory Requirements for Diagnostic Radiology Practice