SlideShare a Scribd company logo
X-Ray Safety
It is recommended that after the completion of each
section that you test your understanding by doing the
quiz related to that section.
Slide Show 1:
Some Background Concepts
This slide show will examine:
1. The formation of X-rays
2. X-ray spectra
X-rays
X-rays and gamma rays are both forms of ionising
radiation. Both are forms of electromagnetic radiation
But they differ in their source of origin.
• X-rays are produced through interactions in
electron shells.
• Gamma rays are produced in the nucleus.
• X-rays are sometimes defined as having wavelengths
between 10-10 and 10-12 m. A more robust definition of
X-rays, however, is their mode of production.
• X-rays are produced through interactions in electron
shells
The Formation of X-rays
To produce x-rays projectile electrons are accelerated from
the negative cathode to the positive anode.
cathode
Anode
Electron beam
Xray
-ve
+ve
When the electrons from the cathode are
accelerated at high voltage to the anode:
• 99% of the energy is dissipated as heat
(anode materials are selected to withstand the
high temperatures they are able to withstand)
• 1% is given off as x-rays.
The x-ray radiation is emitted as:
•bremsstrahlung x-ray radiation (about 80%)
•characteristic x-ray radiation.
and/or
• The energy of the x-rays (keV) is
determined by the voltage applied (kVp).
• The amount of x-rays is determined by the
current (mA).
X-ray Spectra
X-ray spectra are composed of:
1. Continuous bremsstrahlung spectra
2. In most cases, discrete spectra peaks known
as characteristic x-rays.
keV
Bremsstrahlung radiation makes up approximately 80% of the x-ray
beam
Bremsstrahlung Radiation
Bremsstrahlung radiation
X-ray
Projectile electrons originating from the cathode filament
impinge on atoms in the anode and will often pass close by the
nucleus of these atoms.
As the electrons pass through the target atom they slow down,
with a loss in kinetic energy. This energy is emitted as x-rays.
The process is known as bremsstrahlung or “braking energy”.
Bremsstrahlung xrays form a continuous energy
spectra. The frequency distribution is continuous and
shows that the Bremsstrahlung process produces more
low energy that higher energy x-rays. The average
energy is approximately 1/3 of the Emax.
E max
The Emax or the maximum energy of the x-rays measured as
(keV) is equal to voltage applied to the Xray tube (kilovolt peak
or kVp).
E max
For example:
An applied voltage of 70 kVp produces an x-ray spectra with Emax of
70 KeV and average energy of about 23 keV.
70KeV
23keV
Characteristic X-ray Radiation
E max
Characteristic X rays
To produce characteristic x-rays the projectile electrons must
have sufficient energy to displace orbital electrons
• An outer shell electron (usually from the L or M shells) fills the
vacancy in the inner orbital and sheds energy as an x-ray of
characteristic energy.
• If the projectile electron has sufficient energy, it
may cause the ejection of an orbital electron (usually
in the K shell) from an atom in the anode.
• The most common transition is from L to K shell.
• For tungsten the characteristic Xray spectra are represented by
peaks at 58 and 69 keV representing L-to-K and M-to-K shell
transitions respectively.
•Each shell transition has a characteristic energy and this
energy is dependent on the atomic number of the atom.
•M-to-K transitions are less common and are of higher energy.
X-ray
K Shell
L shell
M shell
Production of Characteristic X-rays
Note that the impinging electron from the cathode must have
sufficient energy to displace the K shell electron of the anode.
For a tungsten anode the electrons from the cathode must have at least
69.5 keV to dislodge a K shell electron.
Consequently no lines will appear if the x-ray tube with a
tungsten anode is operated at 20 kVp or 40 kVp
75keV 40keV
This energy is the excitation energy of the electron shells
and the energy is characteristic for each and each element
Note that characteristic X ray spectra are
independent of voltage once the threshold
values have been reached
Window
rotor
anode
Focusing cup
filament
cathode
focal spot
Glass envelope
A Schematic X ray Tube
electron beam
X-ray beam
Window
rotor
anode
Focusing cup
filament
cathode
focal spot
Glass envelope
electron beam
X-ray beam
•The filament is heated to boil off electrons which are
then accelerated to the anode
•The filament is contained within the cathode which is cup shaped to
focus the electrons onto the focus spot on the anode
•Tube currents of 50-800 milliamperes are used whereas filament currents
are in the range of 2-5 amperes
•The anode is usually composed of tungsten or molybdenum as it
must withstand very high temperatures (>3000 degrees C)
•The anode is bevelled at an angle of 12 to 17 degrees in order to
maximise the contact area while focussing the resultant beam
•Correct warm up and stand by procedures are essential
to maximise tube and filament life
rotor
anode
Focusing cup
filament
cathode
Glass envelope
electron beam
End of Section 1
Why not test your understanding using the quiz booklet
Slide Show 2:
Factors affecting x-ray beam
quality and quantity
This slide show will examine:
The factors that affect the quantity of x-rays and
their characteristics
• The energy of the x-rays is
determined by the voltage applied.
• The amount of x-rays is determined
by the current.
Factors affecting x-ray beam
quality and quantity
• Anode material
• Voltage applied (kVp)
• Tube Current (mA)
• Filters used
1. Anode material
Different anode materials will produce different
characteristic x-ray spectra and different amounts of
bremsstrahlung radiation.
2. Voltage (kVp)
Note that increasing the applied voltage or kVp will
increase the maximal energy, the average energy and
the intensity of the x-rays. Characteristic x rays do not
change with a change in kVp
40keV 75keV
3. Tube current (mA)
Increasing the current (ie mA) will not change energy of the
beam only the intensity (i.e. the amount) of x-rays. The
quantity of x-rays is directly proportional to the tube current.
100 mA
200 mA
75 keV
4. Filtration
Often filters of thin aluminium or other metals are used to
filter out low energy x-rays.
Average energy
2mm filter
4mm filter
Filters will increase the average x-ray energy but decreases
the intensity. The maximum energy and the characteristic
x-rays remain unchanged
End of Section 2
Why not test your understanding using the quiz booklet
Slide Show 3.
Interaction of X-ray
Radiation with Matter
This slide show will:
1. Examine the energy transfer that accompanies
interaction of radiation with matter
2. Examine how x-rays interact with matter and the
types of interaction
Interaction of Radiation with Matter
(Transfer of Energy)
•Ionization occurs when the energy transferred is
sufficient to eject electron from the incident atom
•If energy of particle or photon is absorbed - radiation
will appear to have stopped
•If energy not completely deposited in the matter -
remaining energy will pass through.
•If energy is absorbed – ionisation is more likely
The Process of Ionisation
An Ion Pair is created
Slide Show 3:
How X-rays Interact with Matter
This slide show will:
Examine the 5 types of interactions between x-rays
and matter
Interaction of X-rays with Matter
The penetration of x-rays (or conversely the
amount of attenuation) is a function of:
• energy of the photon
• atomic number of irradiated matter
• thickness of irradiated material
• density of irradiated material
X-rays are classified as penetrative radiation
Attenuation of X-rays
X-rays are attenuated as they pass through matter
The degree of that any given material is able to
attenuate x-rays is a function of its atomic number
and its density
Half Value Thickness
It is conventional to refer to measure attenuation in
terms of half value thickness
ie the thickness of material required to reduce an x-
ray to half its original intensity
• Photoelectric effect
Interaction of X-rays with matter
• Compton scattering
• Pair production
There are five types of interactions:
• Coherent Scattering
• Photodisintegration
With the exception of coherent scattering, all can result in
ionisation of tissue
• Photoelectric effect
Interaction of X-rays with matter
• Compton scattering
• Pair production
Dominant in biological materials
X-ray is scattered at angle depending on amount of energy
transferred
X-rays of low energy
X-ray transfers energy to an electron which then ejected
X-rays > 1.02MeV
Four interactions resulting in ionisation:
• Photodisintegration
X-rays> 10MeV
Coherent scattering
• Also known as Classical or Thompson Scattering
• Change in x-ray direction with no ionisation
•Occurs at energies <10 keV
Photoelectric effect
Photoelectron
X-ray
• X-ray transfers energy to an inner shell electron which then ejected.
• Filling the inner shell electron results in a characteristic x-ray.
• Characteristic x-rays from nitrogen, carbon and oxygen have very low
energies.
• The final result is absorption of the x-ray (i.e. there is no exit radiation)
Compton Scattering
Compton electron
•The incident x-ray is scattered by an outer shell electron which is also
ejected (Compton electron)
•The X-ray is scattered at angle depending on amount of energy transferred
•The energy of the incident x-ray is shared between the scattered x-ray and
the Compton electron
•The scattered X ray has lower energy and longer wavelength
The photoelectric effect is responsible for most x-ray attenuation in tissue
Photoelectric attenuation increases with increasing atomic number.
Bone absorbs 4x the x-ray than tissue at lower x-ray energies
Photoelectric attenuation also decreases with increasing energy of the x-ray
Above 26 keV Compton Scattering becomes more dominant
The Medical Application of the Photoelectric Effect
100%
50%
25keV 50keV
Compton
X-ray Energy
Photoelectric
Relative
importance
0.51MeV positron
0.51MeV electron
Pair Production
Occurs with high energy x-ray (> 1.02MeV)
Photonuclear Disintegration
Only occurs with very high energy x-ray (> 10 MeV)
Nuclear fragment
Linear Energy Transfer (LET)
Rate of energy transfer - ionisations per mm
or keV/mm
High LET radiation is not penetrative (ie
energy is deposited in a small distance)
Low LET radiation is penetrative (ie much
less chance energy is deposited in a small
distance)
X-rays are lower LET radiation
Linear Energy Transfer (LET)
Rate of energy transfer - ionisations per mm
or keV/mm
Alpha Not penetrative
Higher LET than beta
X-rays More penetrative
Slightly lower LET than beta
Beta
20
1
1
End of Section 4
Why not test your understanding using the quiz booklet
Slide Show 4:
The Effect of Radiation on
Living Organisms
This slide show will:
1. Examine the effect of ionising radiation on living
organisms
2. Describe units of exposure and dose
Effect of Radiation on Living Organisms
Molecular Effect of Ionising Radiation
•Disruption of bonds - reduced molecular weight
•Alteration of the tertiary and quaternary structure
•Cross-linking
Molecular Effect of Ionising Radiation
Direct Effect
Radiolysis of DNA
Primary feature of high LET radiation
Indirect Effect
Free radicals by radiolysis of water.
2H20 H2O+ + H20-
H2O+ OH. + H+
Hydroxyl radicals react with other molecules (such as DNA)
damaging them.
Effect of Radiation on Cells
•Relationship to dose rate (cell survival curve)
At lower doses cells are able to repair damage without cell death
(shoulder region)
Dose
Cell
Survival
Higher doses cell death occurs directly proportional to
dose
Higher LET there is no shoulder region (i.e. cell repair
mechanism overwhelmed by radiation
•Tissue Type (Law of Bergonne and Tribondeau)
1. Rapidly dividing tissue is more radiosensitive
2. Rapidly growing cells are more radiosensitive
3. Younger and more immature cells are more radiosensitive
4. Mature cells are less radiosensitive
(cf. tissue weighting factor)
NB: Dividing cells are more sensitive in G2 and G1 parts of the cell cycle
•Organ toxicity
Eye lens is particularly sensitive
Effect of Radiation on Tissues and Cells
Stochastic effects
Threshold after which there is an all or nothing effect
e.g. Cancer or genetic effects
Deterministic Effects
Vary with Dose
e.g. lens opacification, blood changes
Total body irradiation
Highly unlikely that an individual would survive a total
exposure of more than 3 Gray without intensive medical
treatment
Partial body irradiation
Cataracts are formed if eyes are exposed to more than 2 Gray
Hair loss occurs at exposures over 3 Gray
Effect of Radiation on Humans
Units of Radiation Exposure and Dose
Units of Radiation Exposure and Dose
• Exposure (Roentgens)
• Dose Equivalence (Sievert)
Relative biological effectiveness of
different types of ionising radiation
• The Effective Dose Rate (Sievert)
• Absorbed dose (Gray)
• Unit is Roentgen
• Amount of x-rays that will cause 1 gram of air to
absorb 86.9 ergs
• Useful for gamma and x-rays only
Exposure
Absorbed dose
• SI Unit is Gray (Gy); old unit is rad
• Dose absorbed by the irradiated material accompanied
by 1 joule (100 ergs) of energy.
• The quantity of energy absorbed per gram per Roentgen
is dependent on the material
• Therefore the absorbed dose is a useful measure and is
applicable to any type or energy of ionising radiation
Dose Equivalence
•Unit is Sievert (Sv); old unit is rem
•Dose is multiplied by a radiation weighting factor
(WR) similar to LET
•Dose Equivalence = D x WR
Radiation Weighting factors of emissions are
approximately:
Alpha particles = 20
Protons, neutrons = 10
Beta particles = 1
Gamma rays and x-rays = 1
The Sievert takes into account the Biological Effectiveness
of the radiation
It can be thought of the absorbed dose of any radiation
that produces the same biological effect as 1 Gray of
therapeutic x-rays
For example:
If 2.5 Sieverts of radiation are required for a given
biological effect – then this could be delivered by 2.5 Gray
of therapeutic x-rays or 0.25 Gray of neutrons
Explanation:
neutrons are 10 x more effective at producing the same
biological effect (ie: have a Quality Factor of 10) and hence
1/10 effective dose of neutrons is required for the same
biological effect.
The Effective Dose
•Unit = Sievert (Sv)
•Takes into account how different parts of the body react
to ionising radiation
Effective Dose Rate = D x Q x wT
wT is a tissue weighting factor for organs and tissues
e.g. wT gonads = 0.2 while wT Skin is 0.01
Prescribed Limits for Dose
Dose
ICRP Prescribed Limits per annum
• Members of public
• Radiation workers
20 mSv per annum above background
150 mSv to eye
500 mSv to hands
1 mSv per annum above background
5 mSv to eye
20 mSv to hands
•Pregnant women must receive no more than 2mSv
per annum
Note that:
•Exposure limits are set for all ‘members of the public’
including pregnant women, babies etc.
•University staff members and students should
consider themselves members of the public for the
purposes of setting exposure limits
Background Dose in NZ
•Background depends on activity (e.g. number of
medical x-rays received in a year)
•Is approx 1.8 mSv per annum
Note- that airline crew on international flights are the most
occupationally exposed group in NZ - 6-8 mSv per annum
received as a result of increased cosmic radiation received at
higher altitudes
Measurement of dose
Geiger Muller meters
Geiger Muller meters only measure ionisation events impinging on the
tube. These meters are useful for detecting point leakage from x-ray
apparatus.
Any readings obtained should be used with care as the high x-ray
intensity can give alarming results Always verify any reading with a
dosimeter
Dosimeters
Measurement of dose can only really be obtained with
dosimeters. These range from film badges, to
thermiluminescent detectors to hand-held monitors.
Hand held dosimeters are available in the University
End of Section 4
Why not test your understanding using the quiz booklet
Slide Show 5:
Safety with X-ray Apparatus
• While most X-ray generated in the x-ray units are of low
energy and the x-ray beam is very narrow, their intensity
is very high. If an some one was able to put his hand in
the path of the xray beam he would sustain an x-ray burn.
• In order to ensure safe operation of such machines the
manufacturers have installed interlocks and designed the
operation of the machine so that it is extremely difficult for the
radiographer to come in contact with the x-ray beam.
General safety regarding the use of X-ray
Equipment
The cornerstone of safe operation of x-ray
units is to ensure:
1. Only trained radiographers who are aware of the safe
operation use these machines.
2. The safe operation of the interlocks are checked regularly
3. X-ray leakage and scatter are monitored on a regular basis
4. The equipment is secured against unauthorised and
untrained use
Safety with X-ray Apparatus
Four Principles of Radiation Safety:
1. Minimise Exposure Time
2. Maximise Distance from Source
3. Use Correct Shielding
4. Follow Manufacturers Instructions
5. Follow the ALARA principle -
keep dose As Low As Reasonably Achievable
1. Time
•Ensure all personnel minimise time spent close
to machine
2. Distance
•Exposure falls at the square of distance from
source
•Therefore use distance to your advantage
3. Shielding
•It is important to periodically verify that shielding
is functioning properly
•Most X-ray machines have adequate
shielding provided
•Never tamper with or alter shielding
4. Follow Manufacturers Instructions
•Never over-ride safety interlocks
•Only authorised operators of X-ray apparatus who are
familiar with safe operation of machine use the
apparatus.
•Document all repairs
•Ensure there are procedures to prevent unauthorised
use of or access to X-ray equipment (ie only authorised
key-holders can use apparatus)
Legal Obligations
•All x-ray apparatus must have a current licence-holder
in charge of the apparatus.
•All x-ray apparatus must be operated in accordance
with the current NNRA Safe Code of Practice
•All x-ray equipment must be secured against unauthorised use
•NNRA must be notified of any disposal of x-ray equipment
NNRA Requirements for safe operation of x-
ray equipment
•There must be a current licence holder in charge
•All users are authorised by licence holder
•Prominent warning signs on doors and equipment
•Access to the machine is restricted by locking
access to room and restricting access to switch
•Log of authorised users and log of use and repairs
•All users are properly trained (training is documented)
•Code of Practice readily available to all users
•Interlocks are tested periodically
•Repairs are undertaken by authorised persons
•Equipment is periodically monitored for x-ray scatter and
leakage
•Emergency procedures are in place to ensure machine is
powered down in an emergency
TIDINESS AND CLEANLINESS OF THE
WORKING ENVIRONMENT
• Always dust the equipment: dust can
prevent the free movt. of trolley
wheels.
• Dust may cause tube columns to jump
the truck.
• Failure to remove dust from the TV
monitor screen will reduce image
quality as will dirt on intensifying screen
STORAGE
• Equipment must be stored carefully and correctly to avoid
damage to equipment or injury to personnel.
• Unit should be switched off after use, mechanical breaks applied
and cable stored correctly in the case of mobile equipment
• Grid requires special arrangement to ensure that they will not
be damaged
• TV monitor should be protected from bright light as this with
time reduce the phosphor efficiency
• X-ray tube should be switched off after use to conserve the
filament life
• Radiographer should “prepare” for short possible time to
conserve the rotor bearing
GOOD PRACTICE
• Radiographer should avoid the use of
force when fitting the parts together,
and ensure they are properly coupled
• X-ray tube should be run up before
heavy loads are applied to a cold anode
• Ensure QC tests are carried out
regularly
LOG BOOK
• Keep log book for each x-ray unit to record
faults:
• Apart from being a catalogue of emergency
failure, it ensures that minor faults and
suggestive symptoms are brought to the
attention of the service engineer
• It provide firms evidence that a particular
problem has received attention
• It provides information about an x-ray unit
End of Section 5
Why not test your understanding using the quiz booklet
TROUBLESHOOTING IN
DIAGNOSTIC RADIOLOGY
• Trouble-shooting is a procedure of
observation, making suitable tests to
either eliminate, or confirm, a suspect
section of equipment.
TYPICAL PROBLEMS
• Operator error.
• Equipment incorrectly calibrated.
• Faulty connecting plugs, sockets, or cables.
• A safety interlock is preventing equipment
operation.
• Electrical or electronic failure.
• High-tension cable or X-ray tube failure.
• Mechanical problems.
• Alignment adjustments
OBSERVATION OF A PROBLEM
• It is good to observe problem when it
occurs such as:
• Is there an operator error?
• A burning smell? Where does it come from?
• Is there an increase in temperature? For
example: The X-ray tube housing has become
very hot to touch.
OBSERVATION OF A
PROBLEM
• —A lock coil, in the area where a burning
smell is observed, is very hot to touch.
• Unusual sound. What sort of sound? Where
is it from?
• Absence of sound. For example:
• —No anode rotation noise from the X-ray
tube.
• —Ventilation fans are quiet
OBSERVATION OF A
PROBLEM
• Wrong mechanical operation. Look for obstructions,
or loose sections. For example:
—An indicator knob on a control panel has slipped into
a wrong position.
—A film is jammed in the processor.
• Visual observation. For example:
—Appearance of the film immediately as it leaves
the processor.
—Smoke rising from equipment, or a HT cable end.
EQUIPMENT MANUAL
• Referred to equipment manual when
ever there is a problem
• The spare parts illustrations can help
find the physical positions of parts, such
as locating a fuse in equipment.
• In the event of missing manual, order
for replacement immediately
• Service engineer may request for this
manual most of the time
REQUEST FOR ASSISTANCE
• When requesting for help from the
service department the following should
be considered:
• Hospital name, address, fax, and phone number.
• Who to contact at the hospital when discussing the problem.
Include the department and phone number.
• Department and room number for equipment
• location.
• Make and model number of equipment.
REQUEST FOR ASSISTANCE
• A description of the problem. Include any symptoms.
• What tests have been made, and the results.
• Are you asking for advice, or is this a direct request
for a service call?
• If a request for a service call, is an order number
available?
• Any conclusions that were made regarding the cause
of the problem, or what will be needed to correct the
problem.
THANK YOU

More Related Content

What's hot

Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiography
Durga Singh
 
X ray tube
X ray tubeX ray tube
X ray tube
Bipin Joseph
 
Chapter 4 image display
Chapter 4 image displayChapter 4 image display
Chapter 4 image display
Muntaser S.Ahmad
 
Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tube
Santosh Ojha
 
Interactions of X-ray & matter & Attenuation - Dr. Sayak Datta
Interactions of X-ray & matter & Attenuation - Dr. Sayak DattaInteractions of X-ray & matter & Attenuation - Dr. Sayak Datta
Interactions of X-ray & matter & Attenuation - Dr. Sayak Datta
SayakDatta
 
Sensitometry &amp; characteristic curve
Sensitometry &amp; characteristic curveSensitometry &amp; characteristic curve
Sensitometry &amp; characteristic curve
Rahul Midha
 
Radiographic image formation
Radiographic image formationRadiographic image formation
Radiographic image formation
MutahirRehman
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
Rakesh Ca
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
Priyanka Parimala
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
Anupam Niraula
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
Dr pradeep Kumar
 
radiation protection in ct
radiation protection in ctradiation protection in ct
radiation protection in ct
shokoofeh mousavi
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
dypradio
 
Generations of CT scanner
Generations of  CT scannerGenerations of  CT scanner
Generations of CT scanner
sadhamhussain52
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
Ramzee Small
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
Anjan Dangal
 
Image quality
Image qualityImage quality
Image quality
airwave12
 
The x ray tube
The x ray tubeThe x ray tube
The x ray tubeairwave12
 

What's hot (20)

Computed Radiography and digital radiography
Computed Radiography and  digital radiographyComputed Radiography and  digital radiography
Computed Radiography and digital radiography
 
X ray generators
X ray generatorsX ray generators
X ray generators
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Chapter 4 image display
Chapter 4 image displayChapter 4 image display
Chapter 4 image display
 
Recent advancements in modern x ray tube
Recent advancements in modern x ray tubeRecent advancements in modern x ray tube
Recent advancements in modern x ray tube
 
Interactions of X-ray & matter & Attenuation - Dr. Sayak Datta
Interactions of X-ray & matter & Attenuation - Dr. Sayak DattaInteractions of X-ray & matter & Attenuation - Dr. Sayak Datta
Interactions of X-ray & matter & Attenuation - Dr. Sayak Datta
 
Sensitometry &amp; characteristic curve
Sensitometry &amp; characteristic curveSensitometry &amp; characteristic curve
Sensitometry &amp; characteristic curve
 
Radiographic image formation
Radiographic image formationRadiographic image formation
Radiographic image formation
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Radiographic grids
Radiographic gridsRadiographic grids
Radiographic grids
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
radiation protection in ct
radiation protection in ctradiation protection in ct
radiation protection in ct
 
CT ARTIFACTS.pptx
CT ARTIFACTS.pptxCT ARTIFACTS.pptx
CT ARTIFACTS.pptx
 
Generations of CT scanner
Generations of  CT scannerGenerations of  CT scanner
Generations of CT scanner
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Image quality
Image qualityImage quality
Image quality
 
The x ray tube
The x ray tubeThe x ray tube
The x ray tube
 

Similar to x-ray_safety_presentation__2_.ppt

Production of x rays
Production of x rays Production of x rays
Production of x rays
VIneeth C
 
X ray physics
X ray physicsX ray physics
X ray physics
Akshay Joshi
 
X-Ray Production.pptx
X-Ray Production.pptxX-Ray Production.pptx
X-Ray Production.pptx
Sachin Sharma
 
X ray production (Emission and Filtration)
X ray production (Emission and Filtration)X ray production (Emission and Filtration)
X ray production (Emission and Filtration)
Edonna Jim
 
1-Radiation physics had.pptx
1-Radiation physics had.pptx1-Radiation physics had.pptx
1-Radiation physics had.pptx
Mohammed700Almkhlafi
 
production and characteristics of x-rays
production and characteristics of x-raysproduction and characteristics of x-rays
production and characteristics of x-rays
Amal Al-Yasiri
 
Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022
dypradio
 
xray production.pptx
xray production.pptxxray production.pptx
xray production.pptx
Atul Verma
 
x-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptxx-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptx
LokendrasinghChoudha
 
B.Tech sem I Engineering Physics U-IV Chapter 2-X-Rays
B.Tech sem I Engineering Physics U-IV Chapter 2-X-RaysB.Tech sem I Engineering Physics U-IV Chapter 2-X-Rays
B.Tech sem I Engineering Physics U-IV Chapter 2-X-Rays
Abhi Hirpara
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
Manisha Saxena
 
Diagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipmentDiagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipment
Vijaykumar Garasiya
 
X-ray Production
X-ray ProductionX-ray Production
X-ray Production
RiyazAhamed80
 
X ray spectroscopy. ppt
X ray spectroscopy. ppt X ray spectroscopy. ppt
X ray spectroscopy. ppt
AkankshaBehl3
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
DrKajalLimbad
 
Interaction of radiation with matter
Interaction of radiation with matterInteraction of radiation with matter
Interaction of radiation with matter
Abhishek Soni
 
02 x ray production and its interaction with matter
02 x ray production and its interaction with matter02 x ray production and its interaction with matter
02 x ray production and its interaction with matter
Akram Kothwal Syed
 
Atomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptxAtomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptx
Dr. Dheeraj Kumar
 
The x ray beam
The x ray beam The x ray beam
The x ray beam
airwave12
 
B.tech sem i engineering physics u iv chapter 2-x-rays
B.tech sem i engineering physics u iv chapter 2-x-raysB.tech sem i engineering physics u iv chapter 2-x-rays
B.tech sem i engineering physics u iv chapter 2-x-rays
Rai University
 

Similar to x-ray_safety_presentation__2_.ppt (20)

Production of x rays
Production of x rays Production of x rays
Production of x rays
 
X ray physics
X ray physicsX ray physics
X ray physics
 
X-Ray Production.pptx
X-Ray Production.pptxX-Ray Production.pptx
X-Ray Production.pptx
 
X ray production (Emission and Filtration)
X ray production (Emission and Filtration)X ray production (Emission and Filtration)
X ray production (Emission and Filtration)
 
1-Radiation physics had.pptx
1-Radiation physics had.pptx1-Radiation physics had.pptx
1-Radiation physics had.pptx
 
production and characteristics of x-rays
production and characteristics of x-raysproduction and characteristics of x-rays
production and characteristics of x-rays
 
Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022
 
xray production.pptx
xray production.pptxxray production.pptx
xray production.pptx
 
x-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptxx-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptx
 
B.Tech sem I Engineering Physics U-IV Chapter 2-X-Rays
B.Tech sem I Engineering Physics U-IV Chapter 2-X-RaysB.Tech sem I Engineering Physics U-IV Chapter 2-X-Rays
B.Tech sem I Engineering Physics U-IV Chapter 2-X-Rays
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
Diagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipmentDiagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipment
 
X-ray Production
X-ray ProductionX-ray Production
X-ray Production
 
X ray spectroscopy. ppt
X ray spectroscopy. ppt X ray spectroscopy. ppt
X ray spectroscopy. ppt
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 
Interaction of radiation with matter
Interaction of radiation with matterInteraction of radiation with matter
Interaction of radiation with matter
 
02 x ray production and its interaction with matter
02 x ray production and its interaction with matter02 x ray production and its interaction with matter
02 x ray production and its interaction with matter
 
Atomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptxAtomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptx
 
The x ray beam
The x ray beam The x ray beam
The x ray beam
 
B.tech sem i engineering physics u iv chapter 2-x-rays
B.tech sem i engineering physics u iv chapter 2-x-raysB.tech sem i engineering physics u iv chapter 2-x-rays
B.tech sem i engineering physics u iv chapter 2-x-rays
 

More from EmmanuelOluseyi1

Drugs used for the CNS.pptx
Drugs used for the CNS.pptxDrugs used for the CNS.pptx
Drugs used for the CNS.pptx
EmmanuelOluseyi1
 
Pharmacokinetics-4_033758.pptx
Pharmacokinetics-4_033758.pptxPharmacokinetics-4_033758.pptx
Pharmacokinetics-4_033758.pptx
EmmanuelOluseyi1
 
lecture_6_20.ppt
lecture_6_20.pptlecture_6_20.ppt
lecture_6_20.ppt
EmmanuelOluseyi1
 
Radiographic Anatomy(chest abdomen and skeletal).pptx
Radiographic Anatomy(chest abdomen and skeletal).pptxRadiographic Anatomy(chest abdomen and skeletal).pptx
Radiographic Anatomy(chest abdomen and skeletal).pptx
EmmanuelOluseyi1
 
PRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptx
PRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptxPRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptx
PRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptx
EmmanuelOluseyi1
 
UG3 RDG304, 1st SEM. CHEST PROCEDURE.pptx
UG3 RDG304, 1st SEM. CHEST PROCEDURE.pptxUG3 RDG304, 1st SEM. CHEST PROCEDURE.pptx
UG3 RDG304, 1st SEM. CHEST PROCEDURE.pptx
EmmanuelOluseyi1
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
EmmanuelOluseyi1
 
DAY LIGHT PROCESSING ASSIGNMENT.pptx
DAY LIGHT PROCESSING ASSIGNMENT.pptxDAY LIGHT PROCESSING ASSIGNMENT.pptx
DAY LIGHT PROCESSING ASSIGNMENT.pptx
EmmanuelOluseyi1
 
BARIUM PROCEEDURES 1.pptx
BARIUM PROCEEDURES 1.pptxBARIUM PROCEEDURES 1.pptx
BARIUM PROCEEDURES 1.pptx
EmmanuelOluseyi1
 
DARKROOM GROUP ASS-2.pptx
DARKROOM GROUP ASS-2.pptxDARKROOM GROUP ASS-2.pptx
DARKROOM GROUP ASS-2.pptx
EmmanuelOluseyi1
 
cardiomegaly-201230142934.pptx
cardiomegaly-201230142934.pptxcardiomegaly-201230142934.pptx
cardiomegaly-201230142934.pptx
EmmanuelOluseyi1
 
arch-k-signs-in-chest-xray-160210003601.pptx
arch-k-signs-in-chest-xray-160210003601.pptxarch-k-signs-in-chest-xray-160210003601.pptx
arch-k-signs-in-chest-xray-160210003601.pptx
EmmanuelOluseyi1
 
congenitaldiaphragmatichernia-140526233701-phpapp01.pdf
congenitaldiaphragmatichernia-140526233701-phpapp01.pdfcongenitaldiaphragmatichernia-140526233701-phpapp01.pdf
congenitaldiaphragmatichernia-140526233701-phpapp01.pdf
EmmanuelOluseyi1
 
radiationtherapy-100510231849-phpapp02.pdf
radiationtherapy-100510231849-phpapp02.pdfradiationtherapy-100510231849-phpapp02.pdf
radiationtherapy-100510231849-phpapp02.pdf
EmmanuelOluseyi1
 
adrenal sex hormone.pptx
adrenal sex hormone.pptxadrenal sex hormone.pptx
adrenal sex hormone.pptx
EmmanuelOluseyi1
 
ramzeesmall-emphysema-161108164159.pdf
ramzeesmall-emphysema-161108164159.pdframzeesmall-emphysema-161108164159.pdf
ramzeesmall-emphysema-161108164159.pdf
EmmanuelOluseyi1
 
pulmonary-fibrosis-presentation-1228416374524935-8.pdf
pulmonary-fibrosis-presentation-1228416374524935-8.pdfpulmonary-fibrosis-presentation-1228416374524935-8.pdf
pulmonary-fibrosis-presentation-1228416374524935-8.pdf
EmmanuelOluseyi1
 
bariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdf
bariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdfbariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdf
bariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdf
EmmanuelOluseyi1
 
pulmonaryedema-170104150504.pdf
pulmonaryedema-170104150504.pdfpulmonaryedema-170104150504.pdf
pulmonaryedema-170104150504.pdf
EmmanuelOluseyi1
 
positioningofskull-130812100608-phpapp01.pdf
positioningofskull-130812100608-phpapp01.pdfpositioningofskull-130812100608-phpapp01.pdf
positioningofskull-130812100608-phpapp01.pdf
EmmanuelOluseyi1
 

More from EmmanuelOluseyi1 (20)

Drugs used for the CNS.pptx
Drugs used for the CNS.pptxDrugs used for the CNS.pptx
Drugs used for the CNS.pptx
 
Pharmacokinetics-4_033758.pptx
Pharmacokinetics-4_033758.pptxPharmacokinetics-4_033758.pptx
Pharmacokinetics-4_033758.pptx
 
lecture_6_20.ppt
lecture_6_20.pptlecture_6_20.ppt
lecture_6_20.ppt
 
Radiographic Anatomy(chest abdomen and skeletal).pptx
Radiographic Anatomy(chest abdomen and skeletal).pptxRadiographic Anatomy(chest abdomen and skeletal).pptx
Radiographic Anatomy(chest abdomen and skeletal).pptx
 
PRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptx
PRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptxPRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptx
PRINCIPLES OF PHARMOCODYNAMICS 2 [Autosaved].pptx
 
UG3 RDG304, 1st SEM. CHEST PROCEDURE.pptx
UG3 RDG304, 1st SEM. CHEST PROCEDURE.pptxUG3 RDG304, 1st SEM. CHEST PROCEDURE.pptx
UG3 RDG304, 1st SEM. CHEST PROCEDURE.pptx
 
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptxLCU RDG 402  PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
 
DAY LIGHT PROCESSING ASSIGNMENT.pptx
DAY LIGHT PROCESSING ASSIGNMENT.pptxDAY LIGHT PROCESSING ASSIGNMENT.pptx
DAY LIGHT PROCESSING ASSIGNMENT.pptx
 
BARIUM PROCEEDURES 1.pptx
BARIUM PROCEEDURES 1.pptxBARIUM PROCEEDURES 1.pptx
BARIUM PROCEEDURES 1.pptx
 
DARKROOM GROUP ASS-2.pptx
DARKROOM GROUP ASS-2.pptxDARKROOM GROUP ASS-2.pptx
DARKROOM GROUP ASS-2.pptx
 
cardiomegaly-201230142934.pptx
cardiomegaly-201230142934.pptxcardiomegaly-201230142934.pptx
cardiomegaly-201230142934.pptx
 
arch-k-signs-in-chest-xray-160210003601.pptx
arch-k-signs-in-chest-xray-160210003601.pptxarch-k-signs-in-chest-xray-160210003601.pptx
arch-k-signs-in-chest-xray-160210003601.pptx
 
congenitaldiaphragmatichernia-140526233701-phpapp01.pdf
congenitaldiaphragmatichernia-140526233701-phpapp01.pdfcongenitaldiaphragmatichernia-140526233701-phpapp01.pdf
congenitaldiaphragmatichernia-140526233701-phpapp01.pdf
 
radiationtherapy-100510231849-phpapp02.pdf
radiationtherapy-100510231849-phpapp02.pdfradiationtherapy-100510231849-phpapp02.pdf
radiationtherapy-100510231849-phpapp02.pdf
 
adrenal sex hormone.pptx
adrenal sex hormone.pptxadrenal sex hormone.pptx
adrenal sex hormone.pptx
 
ramzeesmall-emphysema-161108164159.pdf
ramzeesmall-emphysema-161108164159.pdframzeesmall-emphysema-161108164159.pdf
ramzeesmall-emphysema-161108164159.pdf
 
pulmonary-fibrosis-presentation-1228416374524935-8.pdf
pulmonary-fibrosis-presentation-1228416374524935-8.pdfpulmonary-fibrosis-presentation-1228416374524935-8.pdf
pulmonary-fibrosis-presentation-1228416374524935-8.pdf
 
bariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdf
bariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdfbariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdf
bariumswallowunisonradiologydeptkhorfakkanhospital-181115040224.pdf
 
pulmonaryedema-170104150504.pdf
pulmonaryedema-170104150504.pdfpulmonaryedema-170104150504.pdf
pulmonaryedema-170104150504.pdf
 
positioningofskull-130812100608-phpapp01.pdf
positioningofskull-130812100608-phpapp01.pdfpositioningofskull-130812100608-phpapp01.pdf
positioningofskull-130812100608-phpapp01.pdf
 

Recently uploaded

NVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control programNVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control program
Sapna Thakur
 
Maxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptx
Maxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptxMaxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptx
Maxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptx
Dr. Rabia Inam Gandapore
 
Physiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of TastePhysiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of Taste
MedicoseAcademics
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
Sujoy Dasgupta
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
greendigital
 
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIONDACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
DR SETH JOTHAM
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
pal078100
 
BENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdf
BENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdfBENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdf
BENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdf
DR SETH JOTHAM
 
Alcohol_Dr. Jeenal Mistry MD Pharmacology.pdf
Alcohol_Dr. Jeenal Mistry MD Pharmacology.pdfAlcohol_Dr. Jeenal Mistry MD Pharmacology.pdf
Alcohol_Dr. Jeenal Mistry MD Pharmacology.pdf
Dr Jeenal Mistry
 
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptxPharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Dr. Rabia Inam Gandapore
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
SumeraAhmad5
 
Flu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore KarnatakaFlu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore Karnataka
addon Scans
 
Ophthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE examOphthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE exam
KafrELShiekh University
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
LanceCatedral
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
Krishan Murari
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
Little Cross Family Clinic
 
Triangles of Neck and Clinical Correlation by Dr. RIG.pptx
Triangles of Neck and Clinical Correlation by Dr. RIG.pptxTriangles of Neck and Clinical Correlation by Dr. RIG.pptx
Triangles of Neck and Clinical Correlation by Dr. RIG.pptx
Dr. Rabia Inam Gandapore
 
New Drug Discovery and Development .....
New Drug Discovery and Development .....New Drug Discovery and Development .....
New Drug Discovery and Development .....
NEHA GUPTA
 
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Savita Shen $i11
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
Levi Shapiro
 

Recently uploaded (20)

NVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control programNVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control program
 
Maxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptx
Maxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptxMaxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptx
Maxilla, Mandible & Hyoid Bone & Clinical Correlations by Dr. RIG.pptx
 
Physiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of TastePhysiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of Taste
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
 
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIONDACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
 
BENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdf
BENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdfBENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdf
BENIGN PROSTATIC HYPERPLASIA.BPH. BPHpdf
 
Alcohol_Dr. Jeenal Mistry MD Pharmacology.pdf
Alcohol_Dr. Jeenal Mistry MD Pharmacology.pdfAlcohol_Dr. Jeenal Mistry MD Pharmacology.pdf
Alcohol_Dr. Jeenal Mistry MD Pharmacology.pdf
 
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptxPharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
 
Flu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore KarnatakaFlu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore Karnataka
 
Ophthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE examOphthalmology Clinical Tests for OSCE exam
Ophthalmology Clinical Tests for OSCE exam
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
 
Triangles of Neck and Clinical Correlation by Dr. RIG.pptx
Triangles of Neck and Clinical Correlation by Dr. RIG.pptxTriangles of Neck and Clinical Correlation by Dr. RIG.pptx
Triangles of Neck and Clinical Correlation by Dr. RIG.pptx
 
New Drug Discovery and Development .....
New Drug Discovery and Development .....New Drug Discovery and Development .....
New Drug Discovery and Development .....
 
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
 

x-ray_safety_presentation__2_.ppt

  • 2. It is recommended that after the completion of each section that you test your understanding by doing the quiz related to that section.
  • 3. Slide Show 1: Some Background Concepts
  • 4. This slide show will examine: 1. The formation of X-rays 2. X-ray spectra
  • 5. X-rays X-rays and gamma rays are both forms of ionising radiation. Both are forms of electromagnetic radiation But they differ in their source of origin. • X-rays are produced through interactions in electron shells. • Gamma rays are produced in the nucleus.
  • 6. • X-rays are sometimes defined as having wavelengths between 10-10 and 10-12 m. A more robust definition of X-rays, however, is their mode of production. • X-rays are produced through interactions in electron shells
  • 8. To produce x-rays projectile electrons are accelerated from the negative cathode to the positive anode. cathode Anode Electron beam Xray -ve +ve
  • 9. When the electrons from the cathode are accelerated at high voltage to the anode: • 99% of the energy is dissipated as heat (anode materials are selected to withstand the high temperatures they are able to withstand) • 1% is given off as x-rays.
  • 10. The x-ray radiation is emitted as: •bremsstrahlung x-ray radiation (about 80%) •characteristic x-ray radiation. and/or
  • 11. • The energy of the x-rays (keV) is determined by the voltage applied (kVp). • The amount of x-rays is determined by the current (mA).
  • 13. X-ray spectra are composed of: 1. Continuous bremsstrahlung spectra 2. In most cases, discrete spectra peaks known as characteristic x-rays. keV Bremsstrahlung radiation makes up approximately 80% of the x-ray beam
  • 15. Bremsstrahlung radiation X-ray Projectile electrons originating from the cathode filament impinge on atoms in the anode and will often pass close by the nucleus of these atoms. As the electrons pass through the target atom they slow down, with a loss in kinetic energy. This energy is emitted as x-rays. The process is known as bremsstrahlung or “braking energy”.
  • 16. Bremsstrahlung xrays form a continuous energy spectra. The frequency distribution is continuous and shows that the Bremsstrahlung process produces more low energy that higher energy x-rays. The average energy is approximately 1/3 of the Emax. E max
  • 17. The Emax or the maximum energy of the x-rays measured as (keV) is equal to voltage applied to the Xray tube (kilovolt peak or kVp). E max For example: An applied voltage of 70 kVp produces an x-ray spectra with Emax of 70 KeV and average energy of about 23 keV. 70KeV 23keV
  • 19. E max Characteristic X rays To produce characteristic x-rays the projectile electrons must have sufficient energy to displace orbital electrons • An outer shell electron (usually from the L or M shells) fills the vacancy in the inner orbital and sheds energy as an x-ray of characteristic energy. • If the projectile electron has sufficient energy, it may cause the ejection of an orbital electron (usually in the K shell) from an atom in the anode.
  • 20. • The most common transition is from L to K shell. • For tungsten the characteristic Xray spectra are represented by peaks at 58 and 69 keV representing L-to-K and M-to-K shell transitions respectively. •Each shell transition has a characteristic energy and this energy is dependent on the atomic number of the atom. •M-to-K transitions are less common and are of higher energy.
  • 21. X-ray K Shell L shell M shell Production of Characteristic X-rays
  • 22. Note that the impinging electron from the cathode must have sufficient energy to displace the K shell electron of the anode. For a tungsten anode the electrons from the cathode must have at least 69.5 keV to dislodge a K shell electron. Consequently no lines will appear if the x-ray tube with a tungsten anode is operated at 20 kVp or 40 kVp 75keV 40keV This energy is the excitation energy of the electron shells and the energy is characteristic for each and each element
  • 23. Note that characteristic X ray spectra are independent of voltage once the threshold values have been reached
  • 24. Window rotor anode Focusing cup filament cathode focal spot Glass envelope A Schematic X ray Tube electron beam X-ray beam
  • 25. Window rotor anode Focusing cup filament cathode focal spot Glass envelope electron beam X-ray beam •The filament is heated to boil off electrons which are then accelerated to the anode •The filament is contained within the cathode which is cup shaped to focus the electrons onto the focus spot on the anode •Tube currents of 50-800 milliamperes are used whereas filament currents are in the range of 2-5 amperes
  • 26. •The anode is usually composed of tungsten or molybdenum as it must withstand very high temperatures (>3000 degrees C) •The anode is bevelled at an angle of 12 to 17 degrees in order to maximise the contact area while focussing the resultant beam •Correct warm up and stand by procedures are essential to maximise tube and filament life rotor anode Focusing cup filament cathode Glass envelope electron beam
  • 27. End of Section 1 Why not test your understanding using the quiz booklet
  • 28. Slide Show 2: Factors affecting x-ray beam quality and quantity
  • 29. This slide show will examine: The factors that affect the quantity of x-rays and their characteristics
  • 30. • The energy of the x-rays is determined by the voltage applied. • The amount of x-rays is determined by the current.
  • 31. Factors affecting x-ray beam quality and quantity • Anode material • Voltage applied (kVp) • Tube Current (mA) • Filters used
  • 32. 1. Anode material Different anode materials will produce different characteristic x-ray spectra and different amounts of bremsstrahlung radiation.
  • 33. 2. Voltage (kVp) Note that increasing the applied voltage or kVp will increase the maximal energy, the average energy and the intensity of the x-rays. Characteristic x rays do not change with a change in kVp 40keV 75keV
  • 34. 3. Tube current (mA) Increasing the current (ie mA) will not change energy of the beam only the intensity (i.e. the amount) of x-rays. The quantity of x-rays is directly proportional to the tube current. 100 mA 200 mA 75 keV
  • 35. 4. Filtration Often filters of thin aluminium or other metals are used to filter out low energy x-rays. Average energy 2mm filter 4mm filter Filters will increase the average x-ray energy but decreases the intensity. The maximum energy and the characteristic x-rays remain unchanged
  • 36. End of Section 2 Why not test your understanding using the quiz booklet
  • 37. Slide Show 3. Interaction of X-ray Radiation with Matter
  • 38. This slide show will: 1. Examine the energy transfer that accompanies interaction of radiation with matter 2. Examine how x-rays interact with matter and the types of interaction
  • 39. Interaction of Radiation with Matter (Transfer of Energy) •Ionization occurs when the energy transferred is sufficient to eject electron from the incident atom •If energy of particle or photon is absorbed - radiation will appear to have stopped •If energy not completely deposited in the matter - remaining energy will pass through. •If energy is absorbed – ionisation is more likely
  • 40. The Process of Ionisation An Ion Pair is created
  • 41. Slide Show 3: How X-rays Interact with Matter
  • 42. This slide show will: Examine the 5 types of interactions between x-rays and matter
  • 43. Interaction of X-rays with Matter The penetration of x-rays (or conversely the amount of attenuation) is a function of: • energy of the photon • atomic number of irradiated matter • thickness of irradiated material • density of irradiated material X-rays are classified as penetrative radiation
  • 44. Attenuation of X-rays X-rays are attenuated as they pass through matter The degree of that any given material is able to attenuate x-rays is a function of its atomic number and its density Half Value Thickness It is conventional to refer to measure attenuation in terms of half value thickness ie the thickness of material required to reduce an x- ray to half its original intensity
  • 45. • Photoelectric effect Interaction of X-rays with matter • Compton scattering • Pair production There are five types of interactions: • Coherent Scattering • Photodisintegration With the exception of coherent scattering, all can result in ionisation of tissue
  • 46. • Photoelectric effect Interaction of X-rays with matter • Compton scattering • Pair production Dominant in biological materials X-ray is scattered at angle depending on amount of energy transferred X-rays of low energy X-ray transfers energy to an electron which then ejected X-rays > 1.02MeV Four interactions resulting in ionisation: • Photodisintegration X-rays> 10MeV
  • 47. Coherent scattering • Also known as Classical or Thompson Scattering • Change in x-ray direction with no ionisation •Occurs at energies <10 keV
  • 48. Photoelectric effect Photoelectron X-ray • X-ray transfers energy to an inner shell electron which then ejected. • Filling the inner shell electron results in a characteristic x-ray. • Characteristic x-rays from nitrogen, carbon and oxygen have very low energies. • The final result is absorption of the x-ray (i.e. there is no exit radiation)
  • 49. Compton Scattering Compton electron •The incident x-ray is scattered by an outer shell electron which is also ejected (Compton electron) •The X-ray is scattered at angle depending on amount of energy transferred •The energy of the incident x-ray is shared between the scattered x-ray and the Compton electron •The scattered X ray has lower energy and longer wavelength
  • 50. The photoelectric effect is responsible for most x-ray attenuation in tissue Photoelectric attenuation increases with increasing atomic number. Bone absorbs 4x the x-ray than tissue at lower x-ray energies Photoelectric attenuation also decreases with increasing energy of the x-ray Above 26 keV Compton Scattering becomes more dominant The Medical Application of the Photoelectric Effect 100% 50% 25keV 50keV Compton X-ray Energy Photoelectric Relative importance
  • 51. 0.51MeV positron 0.51MeV electron Pair Production Occurs with high energy x-ray (> 1.02MeV)
  • 52. Photonuclear Disintegration Only occurs with very high energy x-ray (> 10 MeV) Nuclear fragment
  • 53. Linear Energy Transfer (LET) Rate of energy transfer - ionisations per mm or keV/mm High LET radiation is not penetrative (ie energy is deposited in a small distance) Low LET radiation is penetrative (ie much less chance energy is deposited in a small distance) X-rays are lower LET radiation
  • 54. Linear Energy Transfer (LET) Rate of energy transfer - ionisations per mm or keV/mm Alpha Not penetrative Higher LET than beta X-rays More penetrative Slightly lower LET than beta Beta 20 1 1
  • 55. End of Section 4 Why not test your understanding using the quiz booklet
  • 56. Slide Show 4: The Effect of Radiation on Living Organisms
  • 57. This slide show will: 1. Examine the effect of ionising radiation on living organisms 2. Describe units of exposure and dose
  • 58. Effect of Radiation on Living Organisms Molecular Effect of Ionising Radiation •Disruption of bonds - reduced molecular weight •Alteration of the tertiary and quaternary structure •Cross-linking
  • 59. Molecular Effect of Ionising Radiation Direct Effect Radiolysis of DNA Primary feature of high LET radiation Indirect Effect Free radicals by radiolysis of water. 2H20 H2O+ + H20- H2O+ OH. + H+ Hydroxyl radicals react with other molecules (such as DNA) damaging them.
  • 60. Effect of Radiation on Cells •Relationship to dose rate (cell survival curve) At lower doses cells are able to repair damage without cell death (shoulder region) Dose Cell Survival Higher doses cell death occurs directly proportional to dose Higher LET there is no shoulder region (i.e. cell repair mechanism overwhelmed by radiation
  • 61. •Tissue Type (Law of Bergonne and Tribondeau) 1. Rapidly dividing tissue is more radiosensitive 2. Rapidly growing cells are more radiosensitive 3. Younger and more immature cells are more radiosensitive 4. Mature cells are less radiosensitive (cf. tissue weighting factor) NB: Dividing cells are more sensitive in G2 and G1 parts of the cell cycle •Organ toxicity Eye lens is particularly sensitive Effect of Radiation on Tissues and Cells
  • 62. Stochastic effects Threshold after which there is an all or nothing effect e.g. Cancer or genetic effects Deterministic Effects Vary with Dose e.g. lens opacification, blood changes Total body irradiation Highly unlikely that an individual would survive a total exposure of more than 3 Gray without intensive medical treatment Partial body irradiation Cataracts are formed if eyes are exposed to more than 2 Gray Hair loss occurs at exposures over 3 Gray Effect of Radiation on Humans
  • 63. Units of Radiation Exposure and Dose
  • 64. Units of Radiation Exposure and Dose • Exposure (Roentgens) • Dose Equivalence (Sievert) Relative biological effectiveness of different types of ionising radiation • The Effective Dose Rate (Sievert) • Absorbed dose (Gray)
  • 65. • Unit is Roentgen • Amount of x-rays that will cause 1 gram of air to absorb 86.9 ergs • Useful for gamma and x-rays only Exposure
  • 66. Absorbed dose • SI Unit is Gray (Gy); old unit is rad • Dose absorbed by the irradiated material accompanied by 1 joule (100 ergs) of energy. • The quantity of energy absorbed per gram per Roentgen is dependent on the material • Therefore the absorbed dose is a useful measure and is applicable to any type or energy of ionising radiation
  • 67. Dose Equivalence •Unit is Sievert (Sv); old unit is rem •Dose is multiplied by a radiation weighting factor (WR) similar to LET •Dose Equivalence = D x WR Radiation Weighting factors of emissions are approximately: Alpha particles = 20 Protons, neutrons = 10 Beta particles = 1 Gamma rays and x-rays = 1
  • 68. The Sievert takes into account the Biological Effectiveness of the radiation It can be thought of the absorbed dose of any radiation that produces the same biological effect as 1 Gray of therapeutic x-rays For example: If 2.5 Sieverts of radiation are required for a given biological effect – then this could be delivered by 2.5 Gray of therapeutic x-rays or 0.25 Gray of neutrons Explanation: neutrons are 10 x more effective at producing the same biological effect (ie: have a Quality Factor of 10) and hence 1/10 effective dose of neutrons is required for the same biological effect.
  • 69. The Effective Dose •Unit = Sievert (Sv) •Takes into account how different parts of the body react to ionising radiation Effective Dose Rate = D x Q x wT wT is a tissue weighting factor for organs and tissues e.g. wT gonads = 0.2 while wT Skin is 0.01
  • 71. Dose ICRP Prescribed Limits per annum • Members of public • Radiation workers 20 mSv per annum above background 150 mSv to eye 500 mSv to hands 1 mSv per annum above background 5 mSv to eye 20 mSv to hands •Pregnant women must receive no more than 2mSv per annum
  • 72. Note that: •Exposure limits are set for all ‘members of the public’ including pregnant women, babies etc. •University staff members and students should consider themselves members of the public for the purposes of setting exposure limits
  • 73. Background Dose in NZ •Background depends on activity (e.g. number of medical x-rays received in a year) •Is approx 1.8 mSv per annum Note- that airline crew on international flights are the most occupationally exposed group in NZ - 6-8 mSv per annum received as a result of increased cosmic radiation received at higher altitudes
  • 74. Measurement of dose Geiger Muller meters Geiger Muller meters only measure ionisation events impinging on the tube. These meters are useful for detecting point leakage from x-ray apparatus. Any readings obtained should be used with care as the high x-ray intensity can give alarming results Always verify any reading with a dosimeter Dosimeters Measurement of dose can only really be obtained with dosimeters. These range from film badges, to thermiluminescent detectors to hand-held monitors. Hand held dosimeters are available in the University
  • 75. End of Section 4 Why not test your understanding using the quiz booklet
  • 76. Slide Show 5: Safety with X-ray Apparatus
  • 77. • While most X-ray generated in the x-ray units are of low energy and the x-ray beam is very narrow, their intensity is very high. If an some one was able to put his hand in the path of the xray beam he would sustain an x-ray burn. • In order to ensure safe operation of such machines the manufacturers have installed interlocks and designed the operation of the machine so that it is extremely difficult for the radiographer to come in contact with the x-ray beam. General safety regarding the use of X-ray Equipment
  • 78. The cornerstone of safe operation of x-ray units is to ensure: 1. Only trained radiographers who are aware of the safe operation use these machines. 2. The safe operation of the interlocks are checked regularly 3. X-ray leakage and scatter are monitored on a regular basis 4. The equipment is secured against unauthorised and untrained use
  • 79. Safety with X-ray Apparatus Four Principles of Radiation Safety: 1. Minimise Exposure Time 2. Maximise Distance from Source 3. Use Correct Shielding 4. Follow Manufacturers Instructions 5. Follow the ALARA principle - keep dose As Low As Reasonably Achievable
  • 80. 1. Time •Ensure all personnel minimise time spent close to machine 2. Distance •Exposure falls at the square of distance from source •Therefore use distance to your advantage
  • 81. 3. Shielding •It is important to periodically verify that shielding is functioning properly •Most X-ray machines have adequate shielding provided •Never tamper with or alter shielding
  • 82. 4. Follow Manufacturers Instructions •Never over-ride safety interlocks •Only authorised operators of X-ray apparatus who are familiar with safe operation of machine use the apparatus. •Document all repairs •Ensure there are procedures to prevent unauthorised use of or access to X-ray equipment (ie only authorised key-holders can use apparatus)
  • 83. Legal Obligations •All x-ray apparatus must have a current licence-holder in charge of the apparatus. •All x-ray apparatus must be operated in accordance with the current NNRA Safe Code of Practice •All x-ray equipment must be secured against unauthorised use •NNRA must be notified of any disposal of x-ray equipment
  • 84. NNRA Requirements for safe operation of x- ray equipment •There must be a current licence holder in charge •All users are authorised by licence holder •Prominent warning signs on doors and equipment •Access to the machine is restricted by locking access to room and restricting access to switch •Log of authorised users and log of use and repairs •All users are properly trained (training is documented)
  • 85. •Code of Practice readily available to all users •Interlocks are tested periodically •Repairs are undertaken by authorised persons •Equipment is periodically monitored for x-ray scatter and leakage •Emergency procedures are in place to ensure machine is powered down in an emergency
  • 86. TIDINESS AND CLEANLINESS OF THE WORKING ENVIRONMENT • Always dust the equipment: dust can prevent the free movt. of trolley wheels. • Dust may cause tube columns to jump the truck. • Failure to remove dust from the TV monitor screen will reduce image quality as will dirt on intensifying screen
  • 87. STORAGE • Equipment must be stored carefully and correctly to avoid damage to equipment or injury to personnel. • Unit should be switched off after use, mechanical breaks applied and cable stored correctly in the case of mobile equipment • Grid requires special arrangement to ensure that they will not be damaged • TV monitor should be protected from bright light as this with time reduce the phosphor efficiency • X-ray tube should be switched off after use to conserve the filament life • Radiographer should “prepare” for short possible time to conserve the rotor bearing
  • 88. GOOD PRACTICE • Radiographer should avoid the use of force when fitting the parts together, and ensure they are properly coupled • X-ray tube should be run up before heavy loads are applied to a cold anode • Ensure QC tests are carried out regularly
  • 89. LOG BOOK • Keep log book for each x-ray unit to record faults: • Apart from being a catalogue of emergency failure, it ensures that minor faults and suggestive symptoms are brought to the attention of the service engineer • It provide firms evidence that a particular problem has received attention • It provides information about an x-ray unit
  • 90. End of Section 5 Why not test your understanding using the quiz booklet
  • 91. TROUBLESHOOTING IN DIAGNOSTIC RADIOLOGY • Trouble-shooting is a procedure of observation, making suitable tests to either eliminate, or confirm, a suspect section of equipment.
  • 92. TYPICAL PROBLEMS • Operator error. • Equipment incorrectly calibrated. • Faulty connecting plugs, sockets, or cables. • A safety interlock is preventing equipment operation. • Electrical or electronic failure. • High-tension cable or X-ray tube failure. • Mechanical problems. • Alignment adjustments
  • 93. OBSERVATION OF A PROBLEM • It is good to observe problem when it occurs such as: • Is there an operator error? • A burning smell? Where does it come from? • Is there an increase in temperature? For example: The X-ray tube housing has become very hot to touch.
  • 94. OBSERVATION OF A PROBLEM • —A lock coil, in the area where a burning smell is observed, is very hot to touch. • Unusual sound. What sort of sound? Where is it from? • Absence of sound. For example: • —No anode rotation noise from the X-ray tube. • —Ventilation fans are quiet
  • 95. OBSERVATION OF A PROBLEM • Wrong mechanical operation. Look for obstructions, or loose sections. For example: —An indicator knob on a control panel has slipped into a wrong position. —A film is jammed in the processor. • Visual observation. For example: —Appearance of the film immediately as it leaves the processor. —Smoke rising from equipment, or a HT cable end.
  • 96. EQUIPMENT MANUAL • Referred to equipment manual when ever there is a problem • The spare parts illustrations can help find the physical positions of parts, such as locating a fuse in equipment. • In the event of missing manual, order for replacement immediately • Service engineer may request for this manual most of the time
  • 97. REQUEST FOR ASSISTANCE • When requesting for help from the service department the following should be considered: • Hospital name, address, fax, and phone number. • Who to contact at the hospital when discussing the problem. Include the department and phone number. • Department and room number for equipment • location. • Make and model number of equipment.
  • 98. REQUEST FOR ASSISTANCE • A description of the problem. Include any symptoms. • What tests have been made, and the results. • Are you asking for advice, or is this a direct request for a service call? • If a request for a service call, is an order number available? • Any conclusions that were made regarding the cause of the problem, or what will be needed to correct the problem.