The document discusses the qualification of analytical equipment like electronic balances and pH meters. It explains that qualification includes design, installation, operational and performance qualification to ensure equipment is properly installed and functioning accurately. Specific steps for qualifying balances, such as daily calibration checks with internal weights and yearly checks with external weights, are provided. The two-point calibration method for pH meters using buffer solutions is also described. Acceptance limits and record keeping procedures are outlined to ensure equipment remains calibrated over time.
QUALIFICATION OF MANUFACTURING EQUIPMENTSANKUSH JADHAV
it gives the information about qualification of various manufacturing equipment which is used into the pharmaceutical labs. (only for information purpose)
In this slide contains introduction, qualification, preventive maintenance, requalification method.
Presented by: Malarvannan M (Department of pharmaceutical analysis).RIPER, anantapur
In this slide contains definition and details of Qualification Of HPLC
Presented by: KHALID KUWAITY (Department of pharmaceutical analysis).RIPER, anantapur
University Institute of Pharmaceutical Sciences is a flag bearer of excellence in Pharmaceutical education and research in the country. Here is another initiative to make study material available to everyone worldwide. Based on the new PCI guidelines and syllabus here we have a presentation dealing with qualifications of HPLC which is the " High Performance Liquid Chromatography".
Thank you for reading.
Hope it was of help to you.
UIPS,PU team
It is process of “Establishing documentary evidence that provide a high degree of assurance that a specific process will consistently produce a product meeting its predetermined specifications and quality attributes”.
In the pharmaceutical industry, it is very important that in addition to final testing and compliance of products, it is also assured that the process will consistently produce the expected results.
Validation is action of proving in accordance with the principles of good manufacturing practices, that any procedure, process, equipment, material, activity or system actually leads to expected results.
Cleaning validation is documented evidence with a high degree assurance that one can consistently clean a system or a piece of equipment to predetermined and acceptable limits.
The primary regulatory concern driving the need for cleaning validation is cross contamination of the desired drug substance either by other API from previous batch runs or by residues from the cleaning agents used.
The prime purpose of validating a cleaning process is to ensure compliance with federal and other standard regulations
1. Cross contamination with active ingredients
Contamination of one batch of product with significant levels of residual active ingredients from previous batch cannot be tolerated.
In addition to the obvious problems posed by subjecting consumers or patients to unintended contaminants, potential clinically significant synergistic interactions between pharmacologically active chemicals are a real concern.
2. Contamination with unintended materials or compounds
While inert ingredients used in drug products are generally recognized as safe for human consumption, the routine use, maintenance and cleaning of equipment's provide the potential contamination with such items as equipment parts, lubricants and chemical cleaning agents3. Microbiological contamination
Maintenance , cleaning and storage conditions may provide adventitious microorganisms with the opportunity to proliferate within the processing equipment.
QUALIFICATION OF MANUFACTURING EQUIPMENTSANKUSH JADHAV
it gives the information about qualification of various manufacturing equipment which is used into the pharmaceutical labs. (only for information purpose)
In this slide contains introduction, qualification, preventive maintenance, requalification method.
Presented by: Malarvannan M (Department of pharmaceutical analysis).RIPER, anantapur
In this slide contains definition and details of Qualification Of HPLC
Presented by: KHALID KUWAITY (Department of pharmaceutical analysis).RIPER, anantapur
University Institute of Pharmaceutical Sciences is a flag bearer of excellence in Pharmaceutical education and research in the country. Here is another initiative to make study material available to everyone worldwide. Based on the new PCI guidelines and syllabus here we have a presentation dealing with qualifications of HPLC which is the " High Performance Liquid Chromatography".
Thank you for reading.
Hope it was of help to you.
UIPS,PU team
It is process of “Establishing documentary evidence that provide a high degree of assurance that a specific process will consistently produce a product meeting its predetermined specifications and quality attributes”.
In the pharmaceutical industry, it is very important that in addition to final testing and compliance of products, it is also assured that the process will consistently produce the expected results.
Validation is action of proving in accordance with the principles of good manufacturing practices, that any procedure, process, equipment, material, activity or system actually leads to expected results.
Cleaning validation is documented evidence with a high degree assurance that one can consistently clean a system or a piece of equipment to predetermined and acceptable limits.
The primary regulatory concern driving the need for cleaning validation is cross contamination of the desired drug substance either by other API from previous batch runs or by residues from the cleaning agents used.
The prime purpose of validating a cleaning process is to ensure compliance with federal and other standard regulations
1. Cross contamination with active ingredients
Contamination of one batch of product with significant levels of residual active ingredients from previous batch cannot be tolerated.
In addition to the obvious problems posed by subjecting consumers or patients to unintended contaminants, potential clinically significant synergistic interactions between pharmacologically active chemicals are a real concern.
2. Contamination with unintended materials or compounds
While inert ingredients used in drug products are generally recognized as safe for human consumption, the routine use, maintenance and cleaning of equipment's provide the potential contamination with such items as equipment parts, lubricants and chemical cleaning agents3. Microbiological contamination
Maintenance , cleaning and storage conditions may provide adventitious microorganisms with the opportunity to proliferate within the processing equipment.
Qualification of Dissolution Test Apparatus and Validation of Utility System this presentation will help to enhance your knowledge in validation and qualification area.
This presentation explains about qualifications of HPTLC, types of qualifications, design qualification , installation qualification ,operational qualification, performance qualification ,documentation of qualification .
In this slide contains details about Pharmaceutical validation of water system
Presented by: K VENKATSAI PRASAD (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
Manufacturing Control Systems. J R Controls provides control systems for the manufacturing industry. A typical control system will monitor the progress of parts through the manufacturing and finishing process.
What is IPQC & IPQC Test
Appearance
Drug content determination
pH
Sensitivity test
Spreadability
Rate of absorption
Extrudability
Consistency Test
Rheology & Viscosity
Validation of utility system (water system)ShameerAbid
these slides talked about the validation of utility systems in pharmaceutical industries
with special emphasis on the water system
helpful for pharmaceutical student
In this slide contains Introduction, levels of cleaning, mechanism, sampling method of cleaning validation.
Presented by: P. VENKATESH (Department of pharmaceutical analysis).RIPER, anantapur
In this slide contains definition, validation method of HVAC
Presented by: V NABI RASOOL (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur.
Qualification of Dissolution Test Apparatus and Validation of Utility System this presentation will help to enhance your knowledge in validation and qualification area.
This presentation explains about qualifications of HPTLC, types of qualifications, design qualification , installation qualification ,operational qualification, performance qualification ,documentation of qualification .
In this slide contains details about Pharmaceutical validation of water system
Presented by: K VENKATSAI PRASAD (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
Manufacturing Control Systems. J R Controls provides control systems for the manufacturing industry. A typical control system will monitor the progress of parts through the manufacturing and finishing process.
What is IPQC & IPQC Test
Appearance
Drug content determination
pH
Sensitivity test
Spreadability
Rate of absorption
Extrudability
Consistency Test
Rheology & Viscosity
Validation of utility system (water system)ShameerAbid
these slides talked about the validation of utility systems in pharmaceutical industries
with special emphasis on the water system
helpful for pharmaceutical student
In this slide contains Introduction, levels of cleaning, mechanism, sampling method of cleaning validation.
Presented by: P. VENKATESH (Department of pharmaceutical analysis).RIPER, anantapur
In this slide contains definition, validation method of HVAC
Presented by: V NABI RASOOL (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur.
In this slide contains Calibration vs Qualification and phases of qualification.
Presented by: A.Siddartha Tharun Teja. (Department of industrial pharmacy).
RIPER, anantapur.
In this slide contains definition, validation plan, types of Qualification of Dry Powder Mixture.
Presented by: Ravi Sanker babu .D.V (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
In this slide contains Factors Affecting Resolution In HPLC and its criteria's.
Presented by: M.Sudheeshna. (Department of pharmaceutical analysis).
RIPER,anantpur.
In this slide contains introduction, steps, requirements, principle and quantification methods of HPLC.
Presented by: HIMA BINDHU (Department of pharmaceutical analysis).
RIPER, anantapur
In this slide contains QbD approach in Pharmaceutical development.
Presented by: V NABI RASOOL (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
In this slide contains Investigation, reason, case study of OOS.
Presented by: K Venkatsai Preasad. (Department of pharmaceutical analysis and quality assurance).
RIPER, anantapur.
In this slide contains details about drying and pharmaceutical drying equipment's.
Presented by: K VENKATSAI PRASAD (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
In this slide contains types, working principle, factors affecting, advantage and disadvantage of paper electrophoresis.
Presented by: G.Sai Swetha. (Department of pharmacology),
RIPER, anantapur.
Similar to Qualification of Weights and Measures (20)
JOURNAL CLUB PRESENTATION (20L81S0402-PA & QA)
Presented by: K VENKATSAI PRASAD (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
In this slide contains Study of Quality of Raw Materials and General methods of analysis of Raw materials used in cosmetic manufacture as per BSI
Presented by: P.PAVAN KALYAN (Department of pharmaceutical analysis).RIPER, anantapur
In this slide contains Determination of Acid value, Saponification value and Ester value.
Presented by: P.NARESH (Department of pharmaceutical analysis).RIPER, anantapur
In this slide contains Monographs of Herbal Drugs Study in British Herbal Pharmacopoeia and American Herbal Pharmacopoeia.
Presented by: M.SUDHEESHNA (Department of pharmaceutical analysis ).RIPER, anantapur
More from Raghavendra institute of pharmaceutical education and research . (20)
A brief information about the SCOP protein database used in bioinformatics.
The Structural Classification of Proteins (SCOP) database is a comprehensive and authoritative resource for the structural and evolutionary relationships of proteins. It provides a detailed and curated classification of protein structures, grouping them into families, superfamilies, and folds based on their structural and sequence similarities.
Comparing Evolved Extractive Text Summary Scores of Bidirectional Encoder Rep...University of Maribor
Slides from:
11th International Conference on Electrical, Electronics and Computer Engineering (IcETRAN), Niš, 3-6 June 2024
Track: Artificial Intelligence
https://www.etran.rs/2024/en/home-english/
What is greenhouse gasses and how many gasses are there to affect the Earth.moosaasad1975
What are greenhouse gasses how they affect the earth and its environment what is the future of the environment and earth how the weather and the climate effects.
Slide 1: Title Slide
Extrachromosomal Inheritance
Slide 2: Introduction to Extrachromosomal Inheritance
Definition: Extrachromosomal inheritance refers to the transmission of genetic material that is not found within the nucleus.
Key Components: Involves genes located in mitochondria, chloroplasts, and plasmids.
Slide 3: Mitochondrial Inheritance
Mitochondria: Organelles responsible for energy production.
Mitochondrial DNA (mtDNA): Circular DNA molecule found in mitochondria.
Inheritance Pattern: Maternally inherited, meaning it is passed from mothers to all their offspring.
Diseases: Examples include Leber’s hereditary optic neuropathy (LHON) and mitochondrial myopathy.
Slide 4: Chloroplast Inheritance
Chloroplasts: Organelles responsible for photosynthesis in plants.
Chloroplast DNA (cpDNA): Circular DNA molecule found in chloroplasts.
Inheritance Pattern: Often maternally inherited in most plants, but can vary in some species.
Examples: Variegation in plants, where leaf color patterns are determined by chloroplast DNA.
Slide 5: Plasmid Inheritance
Plasmids: Small, circular DNA molecules found in bacteria and some eukaryotes.
Features: Can carry antibiotic resistance genes and can be transferred between cells through processes like conjugation.
Significance: Important in biotechnology for gene cloning and genetic engineering.
Slide 6: Mechanisms of Extrachromosomal Inheritance
Non-Mendelian Patterns: Do not follow Mendel’s laws of inheritance.
Cytoplasmic Segregation: During cell division, organelles like mitochondria and chloroplasts are randomly distributed to daughter cells.
Heteroplasmy: Presence of more than one type of organellar genome within a cell, leading to variation in expression.
Slide 7: Examples of Extrachromosomal Inheritance
Four O’clock Plant (Mirabilis jalapa): Shows variegated leaves due to different cpDNA in leaf cells.
Petite Mutants in Yeast: Result from mutations in mitochondrial DNA affecting respiration.
Slide 8: Importance of Extrachromosomal Inheritance
Evolution: Provides insight into the evolution of eukaryotic cells.
Medicine: Understanding mitochondrial inheritance helps in diagnosing and treating mitochondrial diseases.
Agriculture: Chloroplast inheritance can be used in plant breeding and genetic modification.
Slide 9: Recent Research and Advances
Gene Editing: Techniques like CRISPR-Cas9 are being used to edit mitochondrial and chloroplast DNA.
Therapies: Development of mitochondrial replacement therapy (MRT) for preventing mitochondrial diseases.
Slide 10: Conclusion
Summary: Extrachromosomal inheritance involves the transmission of genetic material outside the nucleus and plays a crucial role in genetics, medicine, and biotechnology.
Future Directions: Continued research and technological advancements hold promise for new treatments and applications.
Slide 11: Questions and Discussion
Invite Audience: Open the floor for any questions or further discussion on the topic.
Richard's entangled aventures in wonderlandRichard Gill
Since the loophole-free Bell experiments of 2020 and the Nobel prizes in physics of 2022, critics of Bell's work have retreated to the fortress of super-determinism. Now, super-determinism is a derogatory word - it just means "determinism". Palmer, Hance and Hossenfelder argue that quantum mechanics and determinism are not incompatible, using a sophisticated mathematical construction based on a subtle thinning of allowed states and measurements in quantum mechanics, such that what is left appears to make Bell's argument fail, without altering the empirical predictions of quantum mechanics. I think however that it is a smoke screen, and the slogan "lost in math" comes to my mind. I will discuss some other recent disproofs of Bell's theorem using the language of causality based on causal graphs. Causal thinking is also central to law and justice. I will mention surprising connections to my work on serial killer nurse cases, in particular the Dutch case of Lucia de Berk and the current UK case of Lucy Letby.
Seminar of U.V. Spectroscopy by SAMIR PANDASAMIR PANDA
Spectroscopy is a branch of science dealing the study of interaction of electromagnetic radiation with matter.
Ultraviolet-visible spectroscopy refers to absorption spectroscopy or reflect spectroscopy in the UV-VIS spectral region.
Ultraviolet-visible spectroscopy is an analytical method that can measure the amount of light received by the analyte.
Nutraceutical market, scope and growth: Herbal drug technologyLokesh Patil
As consumer awareness of health and wellness rises, the nutraceutical market—which includes goods like functional meals, drinks, and dietary supplements that provide health advantages beyond basic nutrition—is growing significantly. As healthcare expenses rise, the population ages, and people want natural and preventative health solutions more and more, this industry is increasing quickly. Further driving market expansion are product formulation innovations and the use of cutting-edge technology for customized nutrition. With its worldwide reach, the nutraceutical industry is expected to keep growing and provide significant chances for research and investment in a number of categories, including vitamins, minerals, probiotics, and herbal supplements.
1. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 1
QUALIFICATION OF WEIGHTS
AND MEASURES
A Seminar as a part of curricular requirement
for I year M. Pharmacy II semester
Presented by
P. PAVAN KALYAN
(Reg. No.20L81S0705)
Department of Pharmaceutical Analysis
2. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 2
• Introduction to Qualification
• Electronic Balance
• Qualification of Electronic Balance
• pH meter
• Qualification of pH meter
• References
contents
3. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 3
• Qualification is a part of validation. It is action of proving and documenting the
equipment, which are properly installed.
• Qualification is of 4 types:
1. Design Qualification
2. Installation Qualification
3. Operational Qualification
4. Performance Qualification
INTRODUCTION TO QUALIFICATION
4. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 4
ELECTRONIC BALANCE
5. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 5
Many types of balances are used in chemical and biological tests.
These balances must be calibrated to ensure the accuracy of the data
produced by it.
Different types of balances used are as follows :-
6. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
Requirements of the balance :-
It should work under optimal conditions like weighing room, weighing
bench, temperature, light, air etc.
S.NO REQUIREMENTS ACCEPTANCE
1) Weighing bench Should be non-magnetic vibration
proof& dust free
2) Temperature Constant temperature should be
maintained.
Deviations should not exceed 5°c
3) Atmospheric humidity Should be between 40% to 60%
4) Light Should be protected from directsunlight
5) Weighing vessel
Smallest possible weighing vessel
used. Weighing vessel and sample
it contains should have same
temperature. 13
7. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 7
Design Qualification :-
Selection of a manufacturer that can satisfy the needs of the laboratory
Installation Qualification :-
Checking of all requirements set during the selection of instrument such as
electricity, humidity, temperature, etc.
Allowing sufficient shelf space for the equipment, SOPs, operating manuals, etc.
Comparing equipment, as received, with purchase order, including accessories,
spare parts, etc.
QUALIFICATION OF ELECTRONIC
BALANCE
8. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 8
Checking documentation for completeness like operating manuals,
maintenance instructions, health and safety instructions, etc.
Checking equipment for any damage.
Reading the supplier’s safety instructions, if there are any.
Installing hardware following the manufacturer’s recommendation.
Switching on the electronic balance and ensuring that all the modules power up.
Preparing an installation report.
9. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
T
est procedure Acceptance
limits
T
est frequency Remarks
• Measurement of reference
weight by using 10mg,
50mg, 100mg, 500mg, 1g,
5g, 10g and 20g.
• Comparing the actual
results with reference
weights.
0.1% • Daily or when used,
whatever is longer
with internal reference
weights.
• Yearly with traceable
external weights
through instrument
vendor.
External
standard
should
be
traceable
to
national
standard.
Operational Qualification:-
10. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 10
Defining weights and weight classes to be used.
Defining acceptance limits of results.
Defining test intervals.
Defining corrective actions on what to do if the electronic balance does not meet the
criteria, in other words if the results are out of specification.
Performance Qualification:-
11. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 11
Calibration of Electronic Balance
12. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 12
pH METER
13. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
Two Point Calibration Method:-
This is the method in which pH meter is calibrated with 2 different buffers of
known pH and then checking the pH meter against pH 7.0 standard buffer to
confirm the calibration.
Again place the tip of the electrode in the second calibration buffer i.e .
either in pH 4.0 or 9.2 and repeat the procedure.
14. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
PROCEDURE:-
Check the level of the 3M KCL electrode filling solution (electrical
conductor between reference & sample) to ensure that it is within 25mm of the
filling hole (small circular hole on the side of the electrode).
Then rinse the probe with de-ionized water & blot dry with wipes & then
place the tip of the electrode into the pH 7.0 buffer solution and pH 4.0
buffer or pH 9.2 buffer.
The pH meter will automatically read the endpoint when the reading is
stable & appropriate buffer symbol will appear on the display and record the
pH value.
15. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 15
Design Qualification:-
Selection of a manufacturer that can satisfy the needs of the laboratory
Installation Qualification :-
Checking of all requirements set during the selection of instrument such as
electricity, humidity, temperature, etc.
Allowing sufficient shelf space for the equipment, SOPs, operating manuals,
etc.
Comparing equipment, as received, with purchase order, including accessories,
spare parts, etc.
QUALIFICATION OF pH METER
16. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 16
Checking documentation for completeness like operating manuals,
maintenance instructions, health and safety instructions, etc.
Checking equipment for any damage.
Reading the supplier’s safety instructions, if there are any.
Installing hardware following the manufacturer’s recommendation.
Switching on the pH meter and ensuring that all the modules power up.
Preparing an installation report.
17. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
Test
procedure
Acceptance
limits
Test
frequency
Remarks
• Calibration with
buffer solution
at pH 7 and pH 4.
• V
erification with
buffer solution at
pH 5 or pH 6.
<0.005 pH
units
Before
each use
• Buffer solutions are commercially
available in either tablet or
solution form.
• Some pH meters also have a
temperature sensor, which should
be calibrated every six months.
• The electrodes should be
washed between each
measurement and the buffers
should be visually checked for
cleanliness and absence of
microbial growth before each use.
Operational Qualification:-
18. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
Performance Qualification:-
.
Parameter to be
checked
Typical frequency Typical tolerance limit
Slope Each day of use or before each
series of measurements
95.0 to 105.0%
Offset (pH asymmetry) +/-30mV (offset of
manufacturer
+/-
0.5 pH units
at defined
temperature)
Reference buffer
solution
+/-0.05 pH units
A log book is prepared for operators and service technicians to record
all equipment related activities in chronological order.
19. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721
Date & Time
(DD-MM-YY
&
am or pm)
Observed pH
4.01
Buffer
(±0.05)
7.00Buffe
r
(±0.05)
9.2Buffer
(±0.05)
Room
temp
(°C)
Cal. Status
(Pass/Fail)
Calibrated by
(Name &Sign)
Checked By
(Name
&Sign)
Daily pH meterCalibration
20. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 20
Precautions :-
Do not allow the filled solution to run dry and add filled solution when ever the
level falls 25mm below.
Do not leave the electrode in organic solvents as the tip and body may be damaged.
Discard used buffer solutions daily. Do not reuse it.
Interference :-
The electrode can be stored for up to one week in pH 7.0 or 4.0 not in distilled
water.
For longer storage periods remove the wetting cap and fill it with 3M KCL solution
and push it onto the tip of the electrode.
21. RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 21
• Qualification of Equipment Calibration/Qualification of pH Meters PA/PH/OMCL
(13) 86 DEF.
• Qualification of balances to the OMCL network Guidelines “Qualification of
Equipments” PA/PH/OMCL (12) 77 R11.
• P. Bedson, M. Sargent: The development and application of guidance
on equipment qualification of analytical instruments. Accred. Qual.
Assurance (1996) 1, 265–274
Reference: