This document discusses the principles and performance of nebulizers. It describes how pneumatic nebulizers work by using compressed gas to break liquid medications into small respirable droplets. Key factors that influence nebulizer performance include the gas flow, fill volume, solution characteristics, and patient breathing pattern. The performance of different nebulizers can vary significantly in terms of respirable dose delivered, droplet size, and nebulization time. Both technical and patient factors need to be considered to optimize drug delivery from nebulizers.
in this topic the technique of chest physiotherapy, indications, contradications of chest physiotherapy are explained. different positions used in postural drainage are briefed.
In medicine, a nebuliser or nebulizer (see spelling differences) is a drug delivery device used to administer medication in the form of a mist inhaled into the lungs. Nebulizers are commonly used for the treatment of cystic fibrosis, asthma, COPD and other respiratory diseases.
An incentive Spirometer is a device that increases pulmonary function, and also clears Secretions of the air pathway.
It reduces Postoperative Pulmonary Complications.
It also stimulates cough.
It will give Visual Feedback to the Patient and encourages the Patient.
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...Team Consulting Ltd
In this article, David Harris, Head of Respiratory Drug Delivery, Team Consulting, taps into a powerful combination of detailed anatomical and functional understanding of the human respiratory system, pulmonary drug delivery technology and formulation expertise, and mathematical modelling techniques, in order to put forward the case for high-resistance swirl chambers in dry-powder inhalers, and a rational strategy for optimising the design and thus maximising therapeutic efficacy.
in this topic the technique of chest physiotherapy, indications, contradications of chest physiotherapy are explained. different positions used in postural drainage are briefed.
In medicine, a nebuliser or nebulizer (see spelling differences) is a drug delivery device used to administer medication in the form of a mist inhaled into the lungs. Nebulizers are commonly used for the treatment of cystic fibrosis, asthma, COPD and other respiratory diseases.
An incentive Spirometer is a device that increases pulmonary function, and also clears Secretions of the air pathway.
It reduces Postoperative Pulmonary Complications.
It also stimulates cough.
It will give Visual Feedback to the Patient and encourages the Patient.
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...Team Consulting Ltd
In this article, David Harris, Head of Respiratory Drug Delivery, Team Consulting, taps into a powerful combination of detailed anatomical and functional understanding of the human respiratory system, pulmonary drug delivery technology and formulation expertise, and mathematical modelling techniques, in order to put forward the case for high-resistance swirl chambers in dry-powder inhalers, and a rational strategy for optimising the design and thus maximising therapeutic efficacy.
Handout for Inhalational Anaesthesia CME held in 2013 by the Department of Anaesthesiology, JNMC, Belagavi. Authors have been credited in each chapter.
International Journal of Engineering Research and Applications (IJERA) is an open access online peer reviewed international journal that publishes research and review articles in the fields of Computer Science, Neural Networks, Electrical Engineering, Software Engineering, Information Technology, Mechanical Engineering, Chemical Engineering, Plastic Engineering, Food Technology, Textile Engineering, Nano Technology & science, Power Electronics, Electronics & Communication Engineering, Computational mathematics, Image processing, Civil Engineering, Structural Engineering, Environmental Engineering, VLSI Testing & Low Power VLSI Design etc.
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journeygreendigital
Tom Selleck, an enduring figure in Hollywood. has captivated audiences for decades with his rugged charm, iconic moustache. and memorable roles in television and film. From his breakout role as Thomas Magnum in Magnum P.I. to his current portrayal of Frank Reagan in Blue Bloods. Selleck's career has spanned over 50 years. But beyond his professional achievements. fans have often been curious about Tom Selleck Health. especially as he has aged in the public eye.
Follow us on: Pinterest
Introduction
Many have been interested in Tom Selleck health. not only because of his enduring presence on screen but also because of the challenges. and lifestyle choices he has faced and made over the years. This article delves into the various aspects of Tom Selleck health. exploring his fitness regimen, diet, mental health. and the challenges he has encountered as he ages. We'll look at how he maintains his well-being. the health issues he has faced, and his approach to ageing .
Early Life and Career
Childhood and Athletic Beginnings
Tom Selleck was born on January 29, 1945, in Detroit, Michigan, and grew up in Sherman Oaks, California. From an early age, he was involved in sports, particularly basketball. which played a significant role in his physical development. His athletic pursuits continued into college. where he attended the University of Southern California (USC) on a basketball scholarship. This early involvement in sports laid a strong foundation for his physical health and disciplined lifestyle.
Transition to Acting
Selleck's transition from an athlete to an actor came with its physical demands. His first significant role in "Magnum P.I." required him to perform various stunts and maintain a fit appearance. This role, which he played from 1980 to 1988. necessitated a rigorous fitness routine to meet the show's demands. setting the stage for his long-term commitment to health and wellness.
Fitness Regimen
Workout Routine
Tom Selleck health and fitness regimen has evolved. adapting to his changing roles and age. During his "Magnum, P.I." days. Selleck's workouts were intense and focused on building and maintaining muscle mass. His routine included weightlifting, cardiovascular exercises. and specific training for the stunts he performed on the show.
Selleck adjusted his fitness routine as he aged to suit his body's needs. Today, his workouts focus on maintaining flexibility, strength, and cardiovascular health. He incorporates low-impact exercises such as swimming, walking, and light weightlifting. This balanced approach helps him stay fit without putting undue strain on his joints and muscles.
Importance of Flexibility and Mobility
In recent years, Selleck has emphasized the importance of flexibility and mobility in his fitness regimen. Understanding the natural decline in muscle mass and joint flexibility with age. he includes stretching and yoga in his routine. These practices help prevent injuries, improve posture, and maintain mobilit
Ethanol (CH3CH2OH), or beverage alcohol, is a two-carbon alcohol
that is rapidly distributed in the body and brain. Ethanol alters many
neurochemical systems and has rewarding and addictive properties. It
is the oldest recreational drug and likely contributes to more morbidity,
mortality, and public health costs than all illicit drugs combined. The
5th edition of the Diagnostic and Statistical Manual of Mental Disorders
(DSM-5) integrates alcohol abuse and alcohol dependence into a single
disorder called alcohol use disorder (AUD), with mild, moderate,
and severe subclassifications (American Psychiatric Association, 2013).
In the DSM-5, all types of substance abuse and dependence have been
combined into a single substance use disorder (SUD) on a continuum
from mild to severe. A diagnosis of AUD requires that at least two of
the 11 DSM-5 behaviors be present within a 12-month period (mild
AUD: 2–3 criteria; moderate AUD: 4–5 criteria; severe AUD: 6–11 criteria).
The four main behavioral effects of AUD are impaired control over
drinking, negative social consequences, risky use, and altered physiological
effects (tolerance, withdrawal). This chapter presents an overview
of the prevalence and harmful consequences of AUD in the U.S.,
the systemic nature of the disease, neurocircuitry and stages of AUD,
comorbidities, fetal alcohol spectrum disorders, genetic risk factors, and
pharmacotherapies for AUD.
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...i3 Health
i3 Health is pleased to make the speaker slides from this activity available for use as a non-accredited self-study or teaching resource.
This slide deck presented by Dr. Kami Maddocks, Professor-Clinical in the Division of Hematology and
Associate Division Director for Ambulatory Operations
The Ohio State University Comprehensive Cancer Center, will provide insight into new directions in targeted therapeutic approaches for older adults with mantle cell lymphoma.
STATEMENT OF NEED
Mantle cell lymphoma (MCL) is a rare, aggressive B-cell non-Hodgkin lymphoma (NHL) accounting for 5% to 7% of all lymphomas. Its prognosis ranges from indolent disease that does not require treatment for years to very aggressive disease, which is associated with poor survival (Silkenstedt et al, 2021). Typically, MCL is diagnosed at advanced stage and in older patients who cannot tolerate intensive therapy (NCCN, 2022). Although recent advances have slightly increased remission rates, recurrence and relapse remain very common, leading to a median overall survival between 3 and 6 years (LLS, 2021). Though there are several effective options, progress is still needed towards establishing an accepted frontline approach for MCL (Castellino et al, 2022). Treatment selection and management of MCL are complicated by the heterogeneity of prognosis, advanced age and comorbidities of patients, and lack of an established standard approach for treatment, making it vital that clinicians be familiar with the latest research and advances in this area. In this activity chaired by Michael Wang, MD, Professor in the Department of Lymphoma & Myeloma at MD Anderson Cancer Center, expert faculty will discuss prognostic factors informing treatment, the promising results of recent trials in new therapeutic approaches, and the implications of treatment resistance in therapeutic selection for MCL.
Target Audience
Hematology/oncology fellows, attending faculty, and other health care professionals involved in the treatment of patients with mantle cell lymphoma (MCL).
Learning Objectives
1.) Identify clinical and biological prognostic factors that can guide treatment decision making for older adults with MCL
2.) Evaluate emerging data on targeted therapeutic approaches for treatment-naive and relapsed/refractory MCL and their applicability to older adults
3.) Assess mechanisms of resistance to targeted therapies for MCL and their implications for treatment selection
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stockrebeccabio
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
Telegram: bmksupplier
signal: +85264872720
threema: TUD4A6YC
You can contact me on Telegram or Threema
Communicate promptly and reply
Free of customs clearance, Double Clearance 100% pass delivery to USA, Canada, Spain, Germany, Netherland, Poland, Italy, Sweden, UK, Czech Republic, Australia, Mexico, Russia, Ukraine, Kazakhstan.Door to door service
Hot Selling Organic intermediates
The prostate is an exocrine gland of the male mammalian reproductive system
It is a walnut-sized gland that forms part of the male reproductive system and is located in front of the rectum and just below the urinary bladder
Function is to store and secrete a clear, slightly alkaline fluid that constitutes 10-30% of the volume of the seminal fluid that along with the spermatozoa, constitutes semen
A healthy human prostate measures (4cm-vertical, by 3cm-horizontal, 2cm ant-post ).
It surrounds the urethra just below the urinary bladder. It has anterior, median, posterior and two lateral lobes
It’s work is regulated by androgens which are responsible for male sex characteristics
Generalised disease of the prostate due to hormonal derangement which leads to non malignant enlargement of the gland (increase in the number of epithelial cells and stromal tissue)to cause compression of the urethra leading to symptoms (LUTS
Explore natural remedies for syphilis treatment in Singapore. Discover alternative therapies, herbal remedies, and lifestyle changes that may complement conventional treatments. Learn about holistic approaches to managing syphilis symptoms and supporting overall health.
Title: Sense of Smell
Presenter: Dr. Faiza, Assistant Professor of Physiology
Qualifications:
MBBS (Best Graduate, AIMC Lahore)
FCPS Physiology
ICMT, CHPE, DHPE (STMU)
MPH (GC University, Faisalabad)
MBA (Virtual University of Pakistan)
Learning Objectives:
Describe the primary categories of smells and the concept of odor blindness.
Explain the structure and location of the olfactory membrane and mucosa, including the types and roles of cells involved in olfaction.
Describe the pathway and mechanisms of olfactory signal transmission from the olfactory receptors to the brain.
Illustrate the biochemical cascade triggered by odorant binding to olfactory receptors, including the role of G-proteins and second messengers in generating an action potential.
Identify different types of olfactory disorders such as anosmia, hyposmia, hyperosmia, and dysosmia, including their potential causes.
Key Topics:
Olfactory Genes:
3% of the human genome accounts for olfactory genes.
400 genes for odorant receptors.
Olfactory Membrane:
Located in the superior part of the nasal cavity.
Medially: Folds downward along the superior septum.
Laterally: Folds over the superior turbinate and upper surface of the middle turbinate.
Total surface area: 5-10 square centimeters.
Olfactory Mucosa:
Olfactory Cells: Bipolar nerve cells derived from the CNS (100 million), with 4-25 olfactory cilia per cell.
Sustentacular Cells: Produce mucus and maintain ionic and molecular environment.
Basal Cells: Replace worn-out olfactory cells with an average lifespan of 1-2 months.
Bowman’s Gland: Secretes mucus.
Stimulation of Olfactory Cells:
Odorant dissolves in mucus and attaches to receptors on olfactory cilia.
Involves a cascade effect through G-proteins and second messengers, leading to depolarization and action potential generation in the olfactory nerve.
Quality of a Good Odorant:
Small (3-20 Carbon atoms), volatile, water-soluble, and lipid-soluble.
Facilitated by odorant-binding proteins in mucus.
Membrane Potential and Action Potential:
Resting membrane potential: -55mV.
Action potential frequency in the olfactory nerve increases with odorant strength.
Adaptation Towards the Sense of Smell:
Rapid adaptation within the first second, with further slow adaptation.
Psychological adaptation greater than receptor adaptation, involving feedback inhibition from the central nervous system.
Primary Sensations of Smell:
Camphoraceous, Musky, Floral, Pepperminty, Ethereal, Pungent, Putrid.
Odor Detection Threshold:
Examples: Hydrogen sulfide (0.0005 ppm), Methyl-mercaptan (0.002 ppm).
Some toxic substances are odorless at lethal concentrations.
Characteristics of Smell:
Odor blindness for single substances due to lack of appropriate receptor protein.
Behavioral and emotional influences of smell.
Transmission of Olfactory Signals:
From olfactory cells to glomeruli in the olfactory bulb, involving lateral inhibition.
Primitive, less old, and new olfactory systems with different path
These simplified slides by Dr. Sidra Arshad present an overview of the non-respiratory functions of the respiratory tract.
Learning objectives:
1. Enlist the non-respiratory functions of the respiratory tract
2. Briefly explain how these functions are carried out
3. Discuss the significance of dead space
4. Differentiate between minute ventilation and alveolar ventilation
5. Describe the cough and sneeze reflexes
Study Resources:
1. Chapter 39, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 34, Ganong’s Review of Medical Physiology, 26th edition
3. Chapter 17, Human Physiology by Lauralee Sherwood, 9th edition
4. Non-respiratory functions of the lungs https://academic.oup.com/bjaed/article/13/3/98/278874
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
Principles for nebulization
1. Nebulizers: Principles and Performance
Dean R Hess PhD RRT FAARC
Introduction
Pneumatic Nebulizers
Principle of Operation
Clinically Important Characteristics of Nebulizer Performance
Technical Factors Affecting Nebulizer Performance
Patient Factors Affecting Nebulizer Performance
Designs to Enhance Nebulizer Performance
Use of Reservoir Bags to Collect Aerosol During the Expiratory Phase
Breath-Enhanced Nebulizers
Breath-Actuated Nebulizers
Continuous Nebulization
Nebulizers for Specific Applications
Ultrasonic Nebulizers
Summary
[Respir Care 2000;45(6):609–622] Key words: nebulizer, pneumatic nebu-
lizer, ultrasonic nebulizer, continuous nebulization, aerosol therapy.
Introduction
Nebulizers are used to convert liquids into aerosols of a
size that can be inhaled into the lower respiratory tract.
The process of pneumatically converting a bulk liquid into
small droplets is called atomization. Pneumatic nebulizers
have baffles incorporated into their design so that most of
the droplets delivered to the patient are within the respi-
rable size range of 1–5 m. Ultrasonic nebulizers use
electricity to convert a liquid into respirable droplets.
Although the first choice of aerosol generator for the
delivery of bronchodilators and steroids is the metered-
dose inhaler,1,2 nebulizers remain useful for several rea-
sons. First, some drugs for inhalation are available only in
solution form. Second, some patients cannot master the
correct use of metered-dose inhalers or dry powder inhal-
ers. Third, some patients prefer the nebulizer over other
aerosol generating devices. The physiologic benefits of
metered-dose inhalers and nebulizers are virtually equiv-
alent,3,4 and the choice of device is often based on clini-
cian or patient preference rather than clear superiority of
one approach over the other. Although cost savings have
been suggested with the use of metered-dose inhalers com-
pared to nebulizers, these benefits may be overestimated.3
The purpose of this paper is to review the performance
characteristics of nebulizers. Both pneumatic and ultra-
sonic nebulizer designs will be considered.
Pneumatic Nebulizers
Nebulizers are the oldest form of aerosol generation.
Although they have been commonly used for many years,
their basic design and performance has changed little over
the past 25 years. Nebulizers are most commonly used for
bronchodilator administration, and it is well established
that nebulized bronchodilators produce a physiologic re-
sponse. Because bronchodilators are relatively inexpen-
sive, there is little market pressure to improve nebulizer
performance. In fact, the market generally prefers an in-
expensive nebulizer rather than a high-performance neb-
Dean R Hess PhD RRT FAARC is affiliated with Massachusetts General
Hospital and Harvard Medical School, Boston, Massachusetts.
A version of this paper was presented by Dr Hess during the RESPIRA-
TORY CARE Journal Consensus Conference, “Aerosols and Delivery De-
vices,” held September 24–26, 1999 in Bermuda.
Correspondence: Dean R Hess PhD RRT FAARC, Respiratory Care,
Ellison 401, Massachusetts General Hospital, 55 Fruit Street, Boston MA
02114. E-mail: dhess@partners.org
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 609
2. ulizer for bronchodilator administration. However, there
are newer drugs available for inhalation that are expensive
and for which precise dosing may be important. These
include dornase alfa, tobramycin, and pentamidine. Neb-
ulizer performance is affected by both technical and pa-
tient-related factors (Table 1).
Principle of Operation
The operation of a pneumatic nebulizer requires a pres-
surized gas supply as the driving force for liquid atomi-
zation (Fig. 1).5–10 Compressed gas is delivered through a
jet, causing a region of negative pressure. The solution to
be aerosolized is entrained into the gas stream and is sheared
into a liquid film. This film is unstable and breaks into
droplets because of surface tension forces. A baffle is
placed in the aerosol stream, producing smaller particles
and causing larger particles to return to the liquid reser-
voir. More than 99% of the particles may be returned to
the liquid reservoir.8 The aerosol is delivered into the in-
spiratory gas stream of the patient. Before delivery into the
patient’s respiratory tract, the aerosol can be further con-
ditioned by environmental factors such as the relative hu-
midity of the carrier gas.11–14
Nebulizer nozzles are of two types (Fig. 2).15 With the
internal mixing design, gas flow interacts with the solution
prior to leaving the exit port. With external mixing, gas
and the solution interact after both leave the nozzle. Mod-
ifications on these designs are used by nebulizer manufac-
turers, without clear superiority of one approach over the
other.
Determinants of droplet size produced by nebulizers
include the characteristics of the solution (density, viscos-
ity, surface tension), the velocities of the gas and solution,
and the flow rates for the gas and the solution.5,15 The most
important factors are gas velocity and the ratio of liquid to
gas flow.5 An increase in gas velocity decreases droplet
size, whereas an increase in the ratio of liquid to gas flow
increases particle size. It is interesting to note that gas
velocity affects the flow rates for both the gas and the
solution. Thus, it is impossible to separately control the
primary factors affecting droplet size from nebulizers.
An important consideration in the use of nebulizers is
the dead volume of the device. Dead volume refers to the
amount of solution that is trapped inside the nebulizer and
is thus not made available for inhalation. The dead volume
is typically in the range of 1 to 3 mL. Dead volume is
minimized by using a conical shape of the nebulizer, by
decreasing the surface area of the internal surface of the
nebulizer, and by improving the wetness of the plastic
surface of the nebulizer.5,15 To reduce medication loss due
to dead volume, clinicians and patients may tap the neb-
ulizer periodically during therapy, which has been shown
to increase nebulizer output.16 Therapy may also be con-
tinued past the point of inconsistent nebulization (sputter-
ing) in an attempt to deliver medication from the dead
volume, but this has been reported to be unproductive.17
Fig. 1. Basic components of the design of pneumatic nebulizers.
(Adapted from Reference 6.)
Table 1. Factors Affecting Penetration and Deposition of
Therapeutic Aerosols Delivered via Jet Nebulizer
Technical Factors
Manufacturer of nebulizer
Gas flow used to power nebulizer
Fill volume of nebulizer
Solution characteristics
Composition of the driving gas
Designs to enhance nebulizer output
Continuous versus breath-actuated
Patient Factors
Breathing pattern
Nose versus mouth breathing
Composition of inspired gas
Airway obstruction
Positive pressure delivery
Artificial airway and mechanical ventilation Fig.2.Internalmixingandexternalmixingnebulizerdesigns.(Adapt-
ed from Reference 5.)
NEBULIZERS: PRINCIPLES AND PERFORMANCE
610 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6
3. Due to evaporative water loss, the solution in the neb-
ulizer becomes increasingly concentrated during the neb-
ulization time.18–20 For this reason, gravimetric methods
underestimate nebulizer output.21,22 Due to the evaporative
effects, the nebulizer solution cools.18 Solution tempera-
ture affects nebulizer output, with output varying directly
with temperature.15,23 The droplet size produced by the
nebulizer also varies directly with temperature.18
Clinically Important Characteristics of Nebulizer
Performance
The most important characteristic of nebulizer perfor-
mance is the respirable dose provided for the patient. The
respirable dose is determined by the mass output of the
nebulizer and the size of the droplets that are produced.
The droplet size should be 2–5 m for airway deposition
and 1–2 m for parenchymal deposition.24 Droplet size is
usually reported as mass median aerodynamic diameter
(MMAD), which is the diameter around which the mass of
the aerosol is equally divided. Note that MMAD is used to
characterize the population of droplets produced; it does
not refer to the size of individual droplets. Because the
volume, and hence the mass, of the droplet is determined
by the cube of the radius (volume ϭ 4/3 r3
), most of the
particles will be smaller than the MMAD. The respirable
dose is sometimes reported as respirable mass, which is
the output of droplets from a nebulizer in the respirable
range of 1–5 m.25,26
Other important characteristics of nebulizer performance
include nebulization time, cost, ease of use, and require-
ments for cleaning and sterilization. Nebulization time is
important for patient compliance in the outpatient setting
and clinician supervision for hospitalized patients. A short
nebulization time that delivers an effective dose is desir-
able. Many nebulizers are low-cost, mass-produced, sin-
gle-patient-use devices. This results in variability in per-
formanceamongdevices,25–28 whichmightnotbeimportant
for bronchodilator delivery, but which could be important
for delivery of other inhaled medications.
Technical Factors Affecting Nebulizer Performance
Several studies have reported performance differences
between nebulizers from different manufacturers.23,25–
27,29–36 Performance differences among nebulizers from the
same manufacturer have been reported.37–39 Hess et al25
evaluated the performance of 17 nebulizers, using a model
of spontaneous breathing. They reported a respirable mass
available to the patient that was severalfold greater from
some nebulizers than from others (Fig. 3). Performance
differences between nebulizers may have clinical implica-
tions.37,40–43 In healthy subjects, Hardy et al42 reported
aerosol deposition from some pneumatic nebulizers that
was twice that of others. In subjects with chronic stable
asthma, Johnson et al43 reported differences in bronchodi-
lation between nebulizers from different manufacturers.
Because of cost considerations, disposable single-pa-
tient-use nebulizers are typically used for many treatments.
The effects of repetitive use and cleaning were evaluated
by Standaert et al.44 In that study, nebulizers were found to
function correctly for 100 repeated uses, provided they
were properly maintained. Proper maintenance consisted
of washing with soapy water, rinsing, and air drying after
each use. Each nebulizer was also subjected to a daily
30-minute soak in 2.5% acetic acid. In the same study,44 it
Fig. 3. Comparison of the output of 17 commercially available pneumatic nebulizers. Respirable mass ϭ particles 1–5 m delivered to
mouthpiece with simulated spontaneous breathing; 2.5 mg albuterol placed into nebulizer cup. (Adapted from Reference 25.)
NEBULIZERS: PRINCIPLES AND PERFORMANCE
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 611
4. was reported that the nebulizers started to fail after 40 uses
if they were not cleaned after each use.
Several studies have reported greater output from pneu-
matic nebulizers when the fill volume is in-
creased.23,25,33,34,45 This is probably because nebulizers
have a fixed dead volume, and thus an increase in fill
volume reduces the proportion of dead volume within the
nebulizer. Although nebulizer output increases with a
greater fill volume, there is also an increase in nebulization
time.25 The nebulization time can be reduced when a larger
fill volume is used by increasing the flow to power the
nebulizer.23,25 A nebulizer fill volume of 4–5 mL is rec-
ommended.
Output increases with an increased flow to power the
nebulizer.23,25,33 An increase in flow also decreases the
droplet size produced by nebulizers.25,39,46–48 A flow of 8
L/min is recommended. Flows lower than this result in
decreased nebulizer performance. A flow greater than this
may result in increased drug loss during the expiratory
phase, which offsets the effect of greater flow on nebulizer
output.25
It is not commonly appreciated that the drug formula-
tion can affect nebulizer performance. MacNeish et al38
reported differences in nebulizer output with two formu-
lations of albuterol. Nebulizer output was significantly
greater with the formulation containing the preservative
benzalkonium chloride, probably because of its surface
activity (Fig. 4). Large droplets were seen to adhere to the
walls of the nebulizer with the preservative-free formula-
tion, whereas foaming was seen to occur with the preser-
vative-containing formulation. Others have also reported
effects of drug formulation on nebulizer output.39,49 It is
interesting to note that metered-dose inhalers have always
been tested and approved as a drug-delivery-system com-
bination. Newer drug solutions have also been approved
for a specific nebulizer (eg, pentamidine, ribavirin, dor-
nase alpha, tobramycin).
The density of the gas powering the nebulizer affects
nebulizer performance. Hess et al50 reported the effect of
heliox (80% helium, 20% oxygen) on nebulizer function.
The inhaled mass of albuterol was significantly reduced
when the nebulizer was powered with heliox, and there
was a greater than twofold increase in nebulization time
with heliox. An increased flow with heliox produced a
respirable mass output similar to that produced when the
nebulizer was powered with air. These results are explained
by Bernoulli’s principle, which predicts that the decreased
density of heliox increases the velocity at which the gas
leaves the jet orifice and produces less negative pressure to
entrain drug solution.50
Patient Factors Affecting Nebulizer Performance
The breathing pattern of the patient affects the amount
of aerosol deposited in the lower respiratory tract. This
partially explains differences in aerosol deposition between
children and adults. To improve aerosol penetration and
deposition in the lungs, the patient should be encouraged
to use a slow and deep breathing pattern.51 Because of the
effect of breathing pattern on drug delivery from nebuliz-
ers, in vitro evaluations of nebulizer performance should
be conducted in a manner that simulates the breathing
pattern of a patient.52
Inhaled aerosols can be administered using a mouth-
piece or a face mask.53–57 Bronchodilator responses occur
with both techniques, and some have argued that the se-
lection of interface should be based on patient preference.53
However, it should be appreciated that the nasal passages
effectively filter droplets delivered from the nebulizer.
Everard et al54 reported a nearly 50% reduction in aerosol
Fig. 4. Comparison of albuterol output of a pneumatic nebulizer, using two formulations of albuterol. (Drawn from data in Reference 38.)
NEBULIZERS: PRINCIPLES AND PERFORMANCE
612 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6
5. delivery to the lungs with nasal inhalation. Whether a
mouthpiece or a face mask is used, it is important to in-
struct the patient to inhale through the mouth. Use of a
mouthpiece may encourage oral breathing.54 It is interest-
ing to note that asthmatic patients switch their breathing
route from the nasal route to the oronasal route during
acute exacerbations and, even when not acutely broncho-
constricted, switch to oronasal breathing when wearing a
face mask.58
Airway caliber affects lung delivery of nebulized bron-
chodilators.59–61 Lipworth et al61 reported lower plasma
albuterol concentrations and attenuated bronchodilator re-
sponses in patients with severe asthma than in normal
subjects or those with mild asthma. It is ironic that the air
flow obstruction that produces the need for inhaled bron-
chodilator therapy also decreases the effectiveness of that
therapy.
Several studies have reported greater pulmonary pene-
tration of aerosols in patients with stable asthma and with
acute airway constriction during heliox breathing.62–66 Be-
cause of the lower density and greater viscosity of heliox,
gas flow becomes less turbulent, which theoretically im-
proves the transport of aerosols through constricted air-
ways to more peripheral lung regions. Henderson et al67
reported no significant advantage of heliox-driven nebu-
lizer therapy over oxygen-driven nebulizer therapy. How-
ever, they67 did not account for the effect of heliox on
nebulizer function,50 and this may have contributed to their
negative findings.
Nebulizer therapy has been used in combination with
noninvasive ventilation. Pollack et al68 randomized pa-
tients with acute asthma to receive either bronchodilator
therapy with a nebulizer and face mask or with a nebulizer,
nasal mask, and BiPAP (bi-level positive airway pressure
system made by Respironics). The BiPAP settings were:
inspiratory pressure 10 cm H2O, expiratory pressure 5 cm
H2O. The patients who received bronchodilator therapy
with BiPAP had a greater improvement in peak flow (Fig.
5). Although these results are intriguing, further positive
reports are needed before widespread acceptance of this
practice. Interestingly, this approach to nebulizer therapy
is reminiscent of intermittent positive-pressure breathing,
which was abandoned many years ago as a method for
delivery of inhaled bronchodilators to patients with asthma.
Nebulizer therapy is commonly used in mechanically
ventilated patients, and this topic has been reviewed in
detail elsewhere.69–73 A number of factors are known to
affect aerosol delivery from nebulizers during mechanical
ventilation (Table 2).12 There are disadvantages of nebu-
lizer use during mechanical ventilation, such as circuit
contamination,74 decreased ability of the patient to trigger
the ventilator75 (if the nebulizer is not powered by the
ventilator), and increases in the delivered tidal volume and
airway pressure76 (if the nebulizer is not powered by the
ventilator). The nebulizer is less efficient than the me-
tered-dose inhaler during mechanical ventilation, but the
nebulizer delivers a greater dose to the lower respiratory
tract.77
Designs to Enhance Nebulizer Performance
In recent years, several nebulizer designs have become
available to decrease the amount of aerosol lost during the
expiratory phase.78 These include reservoir bags to collect
aerosol during the expiratory phase, the use of a vented
design to increase the nebulizer output during the inspira-
tory phase (breath-enhanced nebulizers), and nebulizers
that only generate aerosol during the inspiratory phase
(breath-actuated nebulizers). Because these designs im-
prove drug delivery to the patient, they have the potential
to reduce treatment time, which should improve patient
compliance with nebulizer therapy.
Use of Reservoir Bags to Collect Aerosol
During the Expiratory Phase
For many years, it has been a common practice to use a
T-piece and corrugated tubing as a reservoir for small-
Fig. 5. Peak expiratory flow (PEF) responses with albuterol delivery
by nebulizer with bilevel positive airway pressure (BiPAP) or by
nebulizer alone. (Drawn from data in Reference 68.)
Table 2. Factors Affecting Aerosol Delivery from Nebulizers
During Mechanical Ventilation
Endotracheal tube size
Position of nebulizer placement in the circuit
Type of nebulizer and fill volume
Humidification of the inspired gas
Treatment time
Duty cycle (I:E ratio)
Ventilator brand
I:E ratio ϭ ratio of inspiratory time to expiratory time.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 613
6. volume nebulizers.79 In the late 1980s and early 1990s,
there were reports of increased aerosol delivery to the
lower respiratory tract when a plastic chamber was used
with the nebulizer to capture aerosol during the expiratory
phase, and provide that to the patient during the subse-
quent inspiration.80,81 In the United States, a similar con-
cept was incorporated into the Circulaire and AeroTee
designs (Fig. 6). Both of these designs use a 750 mL bag
to store aerosol during exhalation, but differ in how they
prevent rebreathing. The Circulaire uses a one-way valve
to prevent exhaled gas from entering the reservoir bag,
whereas the AeroTee allows some exhaled gas to enter the
bag.82 These designs also decrease environmental contam-
ination with the aerosol that is generated. The Circulaire
incorporates a variable inspiratory/expiratory resister that
is set to maximize inspiration from the reservoir bag, and
to provide a positive expiratory pressure effect.
Mason et al83 reported an MMAD of 0.51 m with the
Circulaire. Compared with a conventional nebulizer, they
also reported better lung deposition, less gastrointestinal
deposition, and less drug loss to the environment (Fig. 7).
However, there are several important observations about
these results. First, of 9 normal subjects, 2 actually had
decreased pulmonary deposition with the Circulaire. This
illustrates why caution must be exercised when applying
group data to individual patients. Second, the conventional
nebulizer used by Mason et al83 does not perform as well
as the nebulizer incorporated into the Circulaire. Thus, it is
unclear whether the results were the effect of the reservoir
bag or the nebulizer. The MMAD reported by Mason et
al83 for the Circulaire is not ideal. For maximal pulmonary
deposition of bronchodilators, an MMAD of 1–5 m is
more desirable.
In another study by Mason et al,84 the Circulaire was
compared to a conventional nebulizer for bronchodilator
delivery in patients with chronic obstructive pulmonary
disease. In that study, the pulmonary deposition and ther-
apeutic effect were similar for the Circulaire and the con-
ventional nebulizer. Hoffman et al85 compared the Circu-
laire to a conventional nebulizer for bronchodilator delivery
in patients with acute bronchospasm presenting to an emer-
gency department. They reported a greater improvement in
bronchospasm (measured by peak flow) in the Circulaire
group. In this study,85 like those by Mason et al,83,84 the
nebulizer used with the Circulaire may have been superior
to the conventional nebulizer that was used and, thus, the
study may have compared the performance of nebulizers
rather than the effect of the reservoir bag.
Breath-Enhanced Nebulizers
The traditional nebulizer design incorporates the nebu-
lizer sidestream to the air flow of the patient. Some newer
nebulizers use a mainstream design with valves. In this
valved open-vent design, the patient breathes through the
nebulizer during inspiration, which enhances the nebu-
lizer output. During the expiratory phase, a one-way
valve directs patient flow away from the nebulizer cham-
ber (Fig. 8).
This design has been evaluated in several studies, which
have reported greater pulmonary deposition with this de-
sign than with a conventional nebulizer.86–88 A potential
advantage of the open-vent nebulizer design is an im-
provement in nebulizer output with an increase in inspira-
tory flow. Coates et al31 reported a greater output of to-
bramycin with increased inspiratory flow, using an open-
vent nebulizer, whereas changes in inspiratory flow did
not affect the output of the conventional nebulizer (Fig. 9).
As with conventional nebulizers, performance differences
between breath-enhanced nebulizers have been report-
ed.88,89
Breath-Actuated Nebulizers
Aerosol waste during the expiratory phase can be elim-
inated if the nebulizer is only active during the inspiratory
phase. Methods to manually actuate the nebulizer during
the inspiratory phase have been available for many
years.45,90 It is also of interest to note that this design is
commonly used in mechanical ventilator-actuated de-
signs.91,92 Both pneumatically and electronically controlled
breath-actuated nebulizers have recently become commer-
cially available. Their role in clinical application is yet to
be determined.
Continuous Nebulization
Since the late 1980s, there has been considerable clin-
ical and academic interest in the use of continuous aero-
solized bronchodilators for the treatment of acute asth-
ma93–106 (Table 3). These studies suggest that this therapy
is safe, at least as effective as intermittent nebulization,
and may be superior to intermittent nebulization in pa-
tients with the most severe pulmonary function.
Several configurations have been described for contin-
uous nebulization.107 These include frequent refilling of
Fig. 6. Two designs of nebulizer device that use a reservoir bag.
Circulaire (left) and AeroTee (right).
NEBULIZERS: PRINCIPLES AND PERFORMANCE
614 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6
7. the nebulizer,93,101,104,106 use of a nebulizer and infusion
pump (Fig. 10),94,96,99,102,103,108 and use of a large-volume
nebulizer.95,97,98,106,109 Berlinski et al110 reported a consis-
tent and adequate aerosol production by a large-volume
nebulizer over a 4-hour period of operation. Reisner et
al,111 however, reported a more consistent aerosol delivery
with a small-volume nebulizer attached to an infusion pump
than with a large-volume nebulizer. A commonly used
large-volume nebulizer for this therapy is the High-output
Extended Aerosol Respiratory Therapy (HEART) nebu-
lizer. Raabe et al112 reported a detailed evaluation of the
performance of the HEART nebulizer. At a flow of 10–15
L/min, the aerosol output was 38–50 L of aerosolized
drug per liter of gas flow, and the solution output was
30–56 mL/hr.
McPeck et al113 compared the HEART nebulizer to a
conventional small-volume nebulizer in a model of adult
and pediatric breathing. With the adult breathing pattern
they reported similar aerosol delivery from the HEART
nebulizer and small-volume nebulizer. For the pediatric
breathing pattern the aerosol delivery from the small-vol-
ume nebulizer was greater than from the HEART. Both
Raabe et al112 and McPeck et al113 reported an MMAD of
about 2 m with the HEART nebulizer. An important
finding of McPeck et al113 was that the albuterol delivery
from the HEART nebulizer was significantly less than the
target dose from the manufacturer’s recommended setup.
Nebulizers for Specific Applications
Specially constructed small-volume nebulizers should
be used when contamination of the ambient environment
with the aerosolized drug needs to be avoided.35,114,115 The
most common example is aerosolized pentamidine.116 The
nebulizer is fitted with one-way valves and filters to pre-
vent gross contamination of the environment. Examples of
these devices include the Cadema Aero-Tech II and the
Respirgard II. These devices produce a very small particle
size, with an MMAD of about 1–2 m, which is necessary
to improve alveolar deposition of the drug.
The Small-Particle Aerosol Generator is used specifi-
cally to aerosolize ribavirin (Virazole).117–119 The device
consists of a nebulizer and a drying chamber. The drying
chamber reduces the MMAD of particles to about 1.3 m.
There are concerns about the potential adverse effects of
this drug on health care workers when ribavirin is used.
For this reason, a scavenging system should be used when
ribavirin is administered.120–122 This is a double-enclosure
system, with a ribavirin administration hood or mask in-
side a tent. Two high-flow vacuum scavenging systems
Fig. 7. Percent of drug delivery to the lungs, gastrointestinal (GI) tract, and room using a Circulaire nebulizer system. (Drawn from data in
Reference 83.)
Fig. 8. Schematic representation of the function of a breath-en-
hanced nebulizer. Courtesy of Pari Respiratory Equipment.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 615
8. aspirate ribavirin from the system through high-efficiency
particulate air filters.
Ultrasonic Nebulizers
Ultrasonic nebulizers have been clinically available since
the 1960s.123,124 Small-volume ultrasonic nebulizers are
commercially available for delivery of inhaled bronchodi-
lators.125–131 Althoughseveralstudiesreportedgreaterbron-
chodilator response with ultrasonic nebulizers than with
other aerosol generators,127,128 this has not been confirmed
in other studies.129–131 Large-volume ultrasonic nebulizers
are used to deliver inhaled antibiotics in patients with cys-
Fig. 10. System for continuous bronchodilator administration using a conventional nebulizer and an infusion pump. (From Reference 103,
with permission.)
Fig. 9. Tobramycin output from a conventional nebulizer (Hudson) and a breath-enhanced nebulizer (Pari) with changes in inspiratory flow.
RF ϭ respirable fraction. (From Reference 31, with permission.)
NEBULIZERS: PRINCIPLES AND PERFORMANCE
616 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6
9. Table 3. Summary of Studies Reporting Use of Continuous Nebulization
Author Study Population Drug
Research
Design
Continuous Nebulizer
Design
Major Finding
Portnoy93 Children with
severe asthma
Terbutaline;
1–12 mg/h
Case series Frequent refilling of
conventional
nebulizer
Continuously nebulized terbutaline safe and effective
for treatment of severe asthma.
Moler103 Children with
severe asthma
Terbutaline; 4
mg/h
Case series Conventional nebulizer
with infusion pump
Continuously nebulized terbutaline is an effective
therapy for severe asthma.
Calcacone95 Adults with
acute asthma
Albuterol; 10
mg over 2 h
Randomized
controlled
trial
Large volume nebulizer Continuous nebulization was as effective as
intermittent nebulization.
Chipps109 Children with
bronchospasm
Terbutaline; 4
mg/h
Case series HEART nebulizer 18 of 23 cases showed significant improvement with
continuous nebulization.
Papo96
Children with
status asthma
Albuterol; 0.3
mg/kg/h
Randomized
controlled
trial
Conventional nebulizer
with infusion pump
Continuous nebulization was safe and resulted in more
rapid improvement than intermittent nebulization.
Lin97 Adults with
acute asthma
Albuterol; 30
mg over 110
min
Randomized
controlled
trial
HEART nebulizer Continuous nebulization most beneficial in patients
with FEV1 Ͻ 50% predicted.
Rudnitsky98 Adults with
acute asthma
Albuterol; 10
mg over 70
min
Randomized
controlled
trial
HEART nebulizer Continuous nebulization may be of benefit for patients
with peak flow Ͻ 200 L/min.
Lin99
Adults with
acute asthma
Albuterol; 0.4
mg/kg/h for
4 h
Case series Conventional nebulizer
with infusion pump
High dose continuous nebulization can result in
markedly elevated serum albuterol levels and
potential cardiac stimulation.
Katz100 Infants and
children with
bronchospasm
Albuterol; 3
mg/kg/h
Case series Not reported Continuous albuterol safe and without significant
evidence of cardiotoxicity.
Olshaker101 Adults with
acute asthma
Albuterol; 7.5
mg over 1 h
Case series Frequent refilling of
conventional
nebulizer
Continuous nebulization was safe and effective.
Baker105 Adults with
acute asthma
Albuterol; 10
mg/h
Retrospective
case control
Conventional nebulizer
with infusion pump
Continuous and intermittent nebulization were similar
in terms of safety, morbidity, and mortality.
Reisner111 Adults with
acute asthma
Albuterol; 7.5
mg/h
Randomized
controlled
trial
Conventional nebulizer
with infusion pump
Continuous nebulization was as safe and effective as
intermittent nebulization.
Moler103 Children with
acute asthma
Terbutaline; 16
mg over 8 h
Randomized
controlled
trial
Conventional nebulizer
with infusion pump
Continuous nebulization produced similar plasma
terbutaline levels and cardiovascular effects as
intermittent nebulization.
Shrestha104 Adults with
acute asthma
Albuterol; 2.5
mg or 7.5 mg
over 2 h
Randomized
controlled
trial
Frequent refilling of
conventional
nebulizer
The standard dose continuous nebulization group had
the greatest improvement with the fewest side
effects.
Weber106 Adults with
acute asthma
Albuterol at 10
mg/h;
ipratropium at
1 mg/h
Randomized
controlled
trial
HEART nebulizer There was no significant difference in outcomes for
patients receiving continuous albuterol alone or
continuous albuterol with ipratropium.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 617
10. tic fibrosis (eg, tobramycin).132–135 Ultrasonic nebulizers
have also been used during mechanical ventilation,136–139
where they have an advantage in that they do not augment
tidal volume, as occurs with pneumatic nebulizers.
Table 4 lists advantages and disadvantages of ultrasonic
nebulizers for medication delivery. Table 5 lists factors
affecting output from ultrasonic nebulizers.140 A potential
issue with the use of ultrasonic nebulizers is the possibility
of drug inactivation by the ultrasonic waves,141 although
this has not been shown to occur with commonly-used
nebulized medications.
The ultrasonic nebulizer uses a piezoelectric transducer
to produce ultrasonic waves that pass through the solution
and aerosolize it at the surface of the solution. The ultra-
sonic nebulizer creates particle sizes of about 1–6 m
MMAD, depending on the manufacturer of the device.140
The volume output of the ultrasonic nebulizer is about 1–6
mL/min, depending on the manufacturer of the device.140
An ultrasonic nebulizer has 3 components: the power
unit, the transducer, and a fan. The power unit converts
electrical energy to high-frequency ultrasonic waves at a
frequency of 1.3–2.3 megahertz.140 The frequency of the
ultrasonic waves determines the size of the particles, with
an inverse relationship between frequency and particle size.
The frequency is not user adjustable. The power unit also
controls the amplitude of the ultrasonic waves. This is user
adjustable, with an increase in amplitude resulting in an
increase in output from the ultrasonic nebulizer. The trans-
ducer vibrates at the frequency of the ultrasonic waves
applied to it (piezoelectric effect). The transducer is found
in two shapes, concave (focused) and flat (unfocused).140
Concave transducers produce a higher output but require a
constant level of solution for proper operation. The con-
version of ultrasonic energy to mechanical energy by the
transducer produces heat, which is absorbed by the solu-
tion over the transducer.
In some ultrasonic nebulizers, the solution to be nebu-
lized is placed directly over the transducer. In others, the
solution to be nebulized is placed into a nebulization cham-
ber and a water couplant chamber is placed between the
transducer and the medication chamber. A fan is used to
deliver the aerosol produced by the ultrasonic nebulizer to
the patient, or the aerosol is evacuated from the nebuliza-
tion chamber by the inspiratory flow of the patient.
Summary
Nebulizers have been used clinically for many years.
Despite the increasing use of metered-dose inhalers and
dry powder inhalers, it is likely that nebulizers will con-
tinue to be used in selected patients. A number of factors
affect nebulizer performance, and these should be appre-
ciated by clinicians who use these devices. Several new
designs have recently become available that improve the
performance of the nebulizer, but their cost-effectiveness
remains to be determined.
REFERENCES
1. AARC Clinical Practice Guideline. Selection of aerosol delivery
device. American Association for Respiratory Care. Respir Care
1992;37(8):891–897.
2. O’Donohue WJ Jr. Guidelines for the use of nebulizers in the home
and at domiciliary sites: report of a consensus conference. NAM-
DRC Consensus Group. Chest 1996;109(3):814–820.
3. Turner MO, Gafni A, Swan D, FitzGerald JM. A review and eco-
nomic evaluation of bronchodilator delivery methods in hospital-
ized patients. Arch Intern Med 1996;156(18):2113–2118.
4. Cates CJ. Holding chambers versus nebulizers for beta-agonist treat-
ment of acute asthma. The Cochrane Database or Systematic Re-
views, 1999.
5. Dalby RN, Tiano SL, Hickey AJ. Medical devices for the delivery
of therapeutic aerosols to the lungs. In: Hickey AJ, editor. Inhala-
tion aerosols: physical and biological basis for therapy. New York:
Marcel Decker; 1996: 441–473.
6. Newman SP. Aerosol generators and delivery systems. Respir Care
1991;36(9):939–951.
7. Dennis JH, Hendrick DJ. Design characteristics for drug nebulizers.
J Med Eng Technol 1992;16(2):63–68.
8. Smye SW, Jollie MI, Littlewood JM. A mathematical model of
some aspects of jet nebuliser performance. Clin Phys Physiol Meas
1991;12(3):289–300.
9. Niven RW, Brain JD. Some functional characteristics of air-jet
nebulizers. Int J Pharm 1994;104:73–85.
10. Nerbrink O, Dahlback M, Hansson HC. Why do medical nebulizers
difffer in their output and particle size characteristics? J Aerosol
Med 1994;7:259–276.
Table 4. Advantages and Disadvantages of Ultrasonic Nebulizers
Advantages
Little patient coordination required
Small dead volume
Quiet
Aerosol accumulates during exhalation
High doses possible
No chlorofluorocarbon release
Fast drug delivery
Disadvantages
Expensive
Contamination possible
Prone to electrical and mechanical breakdown
Not all drug formulations available
Drug preparation required
Table 5. Factors Affecting Output from Ultrasonic Nebulizers
Fluid characteristics: density, viscosity, surface tension, vapor pressure
Piezoelectric transducer: frequency of vibration, amplitude of
vibration, configuration (focused or flat)
Coupling of medication chamber to transducer
Medication chamber: size, baffles
Flow from fan
NEBULIZERS: PRINCIPLES AND PERFORMANCE
618 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6
11. 11. Fuller HD, Dolovich MB, Chambers C, Newhouse MT. Aerosol
delivery during mechanical ventilation: a predictive in vitro lung
model. J Aerosol Med 1992;5:251–259.
12. AARC Clinical Practice Guideline. Selection of device, adminis-
tration of bronchodilator, and evaluation of response to therapy in
mechanically ventilated patients. American Association for Respi-
ratory Care. Respir Care 1999;44(1):105–113.
13. Diot P, Morra L, Smaldone GC. Albuterol delivery in a model of
mechanical ventilation: comparison of metered-dose inhaler and
nebulizer efficiency. Am J Respir Crit Care Med 1995;152(4 Pt
1):1391–1394.
14. O’Riordan TG, Greco MJ, Perry RJ, Smaldone GC. Nebulizer func-
tion during mechanical ventilation. Am Rev Respir Dis 1992;145(5):
1117–1122.
15. Niven RW. Atomization and nebulizers. In: Hickey AJ, editor.
Inhalation aerosols: physical and biological basis for therapy. New
York: Marcel Decker; 1996: 273–312.
16. Everard ML, Evans M, Milner AD. Is tapping jet nebulisers worth-
while? Arch Dis Child 1994;70(6):538–539.
17. Malone RA, Hollie MC, Glynn-Barnhart A, Nelson HS. Optimal
duration of nebulized albuterol therapy. Chest 1993;104(4):1114–
1118.
18. Phipps PR, Gonda I. Droplets produced by medication nebulizers:
some factors affecting their size and solute concentration. Chest
1990;97(6):1327–1332.
19. Wood JA, Wilson RSE, Bray C. Changes in salbutamol concentra-
tion in the reservoir solution of a jet nebulizer. Br J Dis Chest
1986;80(2):164–169.
20. Stapleton KW, Finlay WH. Determining solution concentration
within aerosol droplets output by jet nebulizers. J Aerosol Sci 1995;
26:137–145.
21. O’Callaghan C, Clarke AR, Milner AD. Inaccurate calculation of
drug output from nebulisers. Eur J Pediatr 1989;148(5):473–474.
22. Tandon R, McPeck M, Smaldone GC. Measuring nebulizer output:
aerosolproductionvsgravimetricanalysis.Chest1997;111(5):1361–
1365.
23. Clay MM, Pavia D, Newman SP, Lennard-Jones T, Clarke SW.
Assessment of jet nebulisers for lung aerosol therapy. Lancet 1983;
2(8350):592–594.
24. Aerosol consensus statement—1991. American Association for Re-
spiratory Care. Respir Care 1991;36(9):916–921.
25. Hess D, Fisher D, Williams P, Pooler S, Kacmarek RM. Medication
nebulizer performance: effects of diluent volume, nebulizer flow,
and nebulizer brand. Chest 1996;110(2):498–505.
26. Loffert DT, Ikle D, Nelson HS. A comparison of commercial jet
nebulizers. Chest 1994;106(6):1788–1792.
27. Alvine GF, Rodgers P, Fitzsimmons KM, Ahrens RC. Disposable
jet nebulizers. How reliable are they? Chest 1992;101(2):316–319.
28. Hollie MC, Malone RA, Skufca RM, Nelson HS. Extreme variabil-
ity in aerosol output of the DeVilbiss 646 jet nebulizer. Chest
1991;100(5):1339–1344.
29. Langford SA, Allen MB. Salbutamol output from two jet nebuliz-
ers. Respir Med 1993;87(2):99–103.
30. Ho SL, Coates AL. Effect of dead volume on the efficiency and the
cost to deliver medications in cystic fibrosis with four disposable
nebulizers. Can Respir J 1999;6(3):253–260.
31. Coates AL, MacNeish CF, Lands LC, Meisner D, Kelemen S,
Vadas EB. A comparison of the availability of tobramycin for
inhalation from vented vs unvented nebulizers. Chest 1998;113(4):
951–956.
32. Arossa W, Quagliotti F, Sala M, Spinaci S. De Candussio G. Dif-
ferent performance of two commercial nebulizers. Respiration 1984;
46(1):128–132.
33. Hess D, Horney D, Snyder T. Medication-delivery performance of
eight small-volume, hand-held nebulizers: effects of diluent vol-
ume, gas flowrate, and nebulizer model. Respir Care 1989;34(8):
717–723.
34. Hurley PK, Smye SW, Cunliffe H. Assessment of antibiotic aerosol
generation using commercial jet nebulizers. J Aerosol Med 1994;
7(3):217–228.
35. Smaldone GC, Perry RJ, Deutsch DG. Characteristics of nebulizers
used in the treatment of AIDS-related pneumocystis carinii pneu-
monia. J Aerosol Med 1988;1:113–126.
36. Waldrep JC, Keyhani K, Black M, Knight V. Operating character-
istics of 18 different continuous-flow jet nebulizers with beclometha-
sone dipropionate liposome aerosol. Chest 1994;105(1):106–110.
37. Hartley-Sharpe CJ, Booth H, Johns DP, Walters EH. Differences in
aerosol output and airways responsiveness between the DeVilbiss
40 and 45 hand held nebulisers. Thorax 1995;50(6):635–638.
38. MacNeish CF, Meisner D, Thibert R, Kelemen S, Vadas EB, Coates
AL. A comparison of pulmonary availability between Ventolin (al-
buterol) nebules and Ventolin (albuterol) Respirator Solution. Chest
1997;111(1):204–208.
39. Coates AL, MacNeish CF, Meisner D, Kelemen S, Thibert R, Mac-
Donald J, Vadas E. The choice of jet nebulizer, nebulizing flow,
and addition of albuterol affects the output of tobramycin aerosols.
Chest 1997;111(5):1206–1212.
40. Newman SP, Woodman G, Clarke SW. Deposition of carbenicillin
aerosols in cystic fibrosis: effects of nebuliser system and breathing
pattern. Thorax 1988;43(4):318–322.
41. O’Doherty M, Thomas S, Page C, Bradbeer C, Nunan T, Bateman
N. Pulmonary deposition of nebulised pentamidine isethionate: ef-
fect of nebuliser type, dose, and volume of fill. Thorax 1990;45(6):
460–464.
42. Hardy JG, Newman SP, Knoch M. Lung deposition from four
nebulizers. Respir Med 1993;87(6):461–465.
43. Johnson MA, Newman SP, Bloom R, Talaee N, Clarke SW. De-
livery of albuterol and ipratropium bromide from two nebulizer
systems in chronic stable asthma: efficacy and pulmonary deposi-
tion. Chest 1989;96(1):6–10.
44. Standaert TA, Morlin GL, Williams-Warren J, Joy P, Pepe MS,We-
ber A, Ramsey BW. Effects of repetitive use and cleaning tech-
niques of disposable jet nebulizers on aerosol generation. Chest
1998;114(2):577–586.
45. Kradjan WA, Lakshminarayan S. Efficiency of air compressor-
driven nebulizers. Chest 1985;87(4):512–516.
46. Hadfield JW, Windebank WJ, Bateman JRM. Is driving gas flow
rate clinically important for nebulizer therapy? Br J Dis Chest 1986;
80(1):50–54.
47. Clay MM, Pavia D, Newman SP, Clarke SW. Factors influencing
the size distribution of aerosols from jet nebulisers. Thorax 1983;
38(10):755–759.
48. Everard ML, Clark AR, Milner AD. Drug delivery from jet nebu-
lisers. Arch Dis Child 1992;67(5):586–591.
49. Flament MP, Leterme P, Burnouf T, Gayot A. Influence of formu-
lation on jet nebulisation quality of alpha 1 protease inhibitor. Int
J Pharm 1999;178(1):101–109.
50. Hess DR, Acosta FL, Ritz RH, Kacmarek RM, Camargo CA Jr. The
effect of heliox on nebulizer function using a beta-agonist bron-
chodilator. Chest 1999;115(1):184–189.
51. Martonen T, Yang Y. Deposition mechanics of pharmaceutical par-
ticles in human airways. In: Hickey AJ, editor. Inhalation aerosols:
physical and biological basis for therapy New York: Marcel Deck-
er; 1996: 3–27.
52. Smaldone GC. Drug delivery via aerosol systems: concept of “aero-
sol inhaled”. J Aerosol Med 1991;4(3):229–235.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 619
12. 53. Steventon RD, Wilson RSE. Facemask or mouthpiece for delivery
of nebulized bronchodilator aerosols? Br J Dis Chest 1981;75(1):
88–90.
54. Everard ML, Hardy JG, Milner AD. Comparison of nebulised aero-
sol deposition in the lungs of healthy adults following oral and
nasal inhalation. Thorax 1993;48(10):1045–1046.
55. Lowenthal D, Kattan M. Facemasks versus mouthpieces for aerosol
treatment of asthmatic children. Pediatr Pulmonol 1992;14(3):192–
196.
56. Smedsaas-Lofvenberg A, Nilsson K, Moa G, Axelsson I. Nebuli-
zation of drugs in a nasal CPAP system. Acta Paediatr 1999;88(1):
89–92.
57. Parkes SN, Bersten AD. Aerosol kinetics and bronchodilator effi-
cacy during continuous positive airway pressure delivered by face
mask. Thorax 1997;52(2):171–175.
58. Kairaitis K, Garlick SR, Wheatley JR, Amis TC. Route of breathing
in patients with asthma. Chest 1999;116(6):1646–1652.
59. Melchor R, Biddiscombe MF, Mak VHF, Short MD, Spiro SG.
Lung deposition patterns of directly labelled salbutamol in normal
subjects and in patients with reversible airflow obstruction. Thorax
1993;48(5):506–511.
60. Pavia D, Thompson ML, Clarke SW, Shannon HS. Effect of lung
function and mode of inhalation on penetration of aerosol into the
human lung. Thorax 1977;32(2):194–197.
61. Lipworth BJ, Clarke DJ. Effects of airway calibre on lung delivery
of nebulised salbutamol. Thorax 1997;52(12):1036–1039.
62. Swift DL, Carpin JC, Mitzner W. Pulmonary penetration and dep-
osition of aerosols in different gases: fluid flow effects. Ann Occup
Hyg 1982;26(1–4):109–117.
63. Svartengren M, Anderson M, Philipson K, Camner P. Human lung
deposition of particles suspended in air or in helium/oxygen mix-
ture. Exp Lung Res 1989;15(4):575–585.
64. Esch JL, Specktor DM, Lippmann M. Effect of lung airway branch-
ing pattern and gas composition on particle deposition: II. Experi-
mental studies in human and canine lungs. Exp Lung Res 1988;
14(3):321–348.
65. Anderson M, Svartengren M, Bylin G, Philipson K, Camner P.
Deposition in asthmatics of particles inhaled in air or in helium-
oxygen. Am Rev Respir Dis 1993;147(3):524–528.
66. Anderson M, Svartengren M, Philioson K, Camner P. Deposition in
man of particles suspended in air or in helium-oxygen mixture at
different flow rates. J Aerosol Med 1990;3:209–216.
67. Henderson SO, Acharya P, Kilaghbian T, Perez J, Korn CS, Chan
LS. Use of heliox-driven nebulizer therapy in the treatment of acute
asthma. Ann Emerg Med 1999;33(2):141–146.
68. Pollack CV Jr, Fleisch KB, Dowsey K. Treatment of acute bron-
chospasm with beta-adrenergic agonist aerosols delivered by a na-
sal bilevel positive airway pressure circuit. Ann Emerg Med 1995;
2(65):552–557.
69. Dhand R, Tobin MJ. Bronchodilator delivery with metered-dose
inhalers in mechancically ventilated patients. Eur Respir J 1996;9:
585–595.
70. Dhand R, Tobin MJ. Inhaled bronchodilator therapy in mechani-
cally-ventilated patients (review). Am J Respir Crit Care Med 1997;
156(1):3–10.
71. Mantous CA, Hall JB. Administration of therapeutic aerosols to
mechanically ventilated patients (review). Chest 1994;106(2):560–
571.
72. Hess D. Inhaled bronchodilators during mechanical ventilation: de-
livery techniques, evaluation of response, and cost-effectiveness.
Respir Care 1994;39(2):105–122.
73. Fink JB, Tobin MJ, Dhand R. Bronchodilator therapy in mechan-
ically ventilated patients. Respir Care 1999;44(1):53–69.
74. Craven DE, Lichtenberg DA, Goularte TA, Make BJ, McCabe WR.
Contaminated medication nebulizers in mechanical ventilator cir-
cuits: source of bacterial aerosols. Am J Med 1984;77(5):834–838.
75. Beaty CD, Ritz RH, Benson MS. Continuous in-line nebulizers
complicate pressure support ventilation. Chest 1989;96(6):1360–
1363.
76. Hanhan U, Kissoon N, Payne M, Taylor C, Murphy S, DeNicola
LK. Effects of in-line nebulization on preset ventilatory variables.
Respir Care 1993;38(5):474–478.
77. Marik P, Hogan J, Krikorian J. A comparison of bronchodilator
therapy delivered by nebulization and metered-dose inhaler in me-
chanically ventilated patients. Chest 1999;115(6):1653–1657.
78. Dennis JH. A review of issues relating to nebulizer standards. J
Aerosol Med 1998;11 Suppl 1:S73–S77.
79. Pisut FM. Comparison of medication delivery by T-nebulizer with
inspiratory and expiratory reservoir. Respir Care 1989;34(11):985–
988.
80. Marshall LM, Francis PW, Khafagi FA. Aerosol deposition in cys-
tic fibrosis using an aerosol conservation device and a conventional
jet nebulizer. J Paediatr Child Health 1994;30(1):65–67.
81. Thomas SH, Langford JA, George RDG, Geddes DM. Improving
the efficacy of drug administration with jet nebulizers (letter). Lan-
cet 1988;1(8577):126.
82. Piper SD. In vitro comparison of the Circulaire and AeroTee to a
traditional nebulizer T-Piece with corrugated tubing. Respir Care
2000;45(3):313–319.
83. Mason JW, Miller WC, Small S. Comparison of aerosol delivery
via Circulaire system vs conventional small volume nebulizer. Re-
spir Care 1994;39(12):1157–1161.
84. Mason JW, Miller WC. Comparison of aerosol delivery via circu-
laire nebulizer system versus a disposable nebulizer in COPD pa-
tients. Respir Care 1996;41(11):1006–1008.
85. Hoffman L, Smithline H. Comparison of Circulaire to conventional
small volume nebulizer for the treatment of bronchospasm in the
emergency department. Respir Care 1997;42(12):1170–1174.
86. Newnham DM, Lipworth BJ. Nebuliser performance, pharmacoki-
netics, airways and systemic effects of salbutamol given via a novel
nebuliser system (“Venstream”). Thorax 1994;49(8):762–770.
87. Newman SP, Pitcairn GR, Hooper G, Knoch M. Efficient drug
delivery to the lungs from a continuously operated open-vent neb-
ulizer and low pressure compressor system. Eur Respir J 1994;7(6):
1177–1181.
88. Devadason SG, Everard ML, Linto JM. Le Souef PN. Comparison
of drug delivery from conventional versus “Venturi” nebulizers.
Eur Respir J 1997;10(11):2479–2483.
89. Barry PW, O’Callaghan C. The output of budesonide from nebu-
lizers. J Allergy Clin Immunol 1998;102(2):321–322.
90. Suez DS, Chai H. A standard method of intermittent inhaled ther-
apy via a jet nebulizer. Ann Allergy 1986;57(4):245–248.
91. McPeck M, O’Riordan TG, Smaldone GC. Choice of mechanical
ventilator: influence on nebulizer performance. Respir Care 1993;
38(8):887–895.
92. Hughes JM, Saez J. Effects of nebulizer mode and position in a
mechanical ventilator circuit on dose efficiency. Respir Care 1987;
32(12):1131–1135.
93. Portnoy J, Aggarwal J. Continuous terbutaline nebulization for the
treatment of severe exacerbations of asthma in children. Ann Al-
lergy 1988;60(4):368–371.
94. Moler FW, Hurwitz ME, Custer JR. Improvement in clinical asthma
score and PaCO2 in children with severe asthma treated with con-
tinuously nebulized terbutaline. J Allergy Clin Immunol 1988;81(6):
1101–1109.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
620 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6
13. 95. Calcacone A, Wolkove N, Stern E, Afilalo M, Rosental TM, Kre-
isman H. Continuous nebulization of albuterol (salbutamol) in acute
asthma. Chest 1990;97(3):693–697.
96. Papo MC, Frank J, Thompson AE. A prospective, randomized study
of continuous versus intermittent nebulized albuterol for severe
status asthmaticus in children. Crit Care Med 1993;21(10):1479–
1486.
97. Lin RY, Sauter D, Newman T, Sirleaf J, Walters J, Tavakol M.
Continuous versus intermittent albuterol nebulization in the treat-
ment of acute asthma. Ann Emerg Med 1993;22(12):1847–1853.
98. Rudnitsky GS, Eberlein RS, Schoffstall JM, Mazur JE, Spivey WH.
Comparison of intermittent and continuosly nebulized albuterol for
treatment of asthma in an urban emergency department. Ann Emerg
Med 1993;22(12):1842–1846.
99. Lin RY, Smith AJ, Hergenroeder P. High serum albuterol levels
and tachycardia in adult asthmatics treated with high-dose contin-
uously aerosolized albuterol. Chest 1993;103(1):221–225.
100. Katz RW, Kelly HW, Crowley MR, Grad R, McWilliams BC,
Murphy SJ. Safety of continuously nebulized albuterol for bron-
chospasm in infants and children. Pediatrics 1993;92(5):666–669.
101. Olshaker J, Jerrard D, Barish RA, Brandt G, Hooper F. The efficacy
and safety of a continuous albuterol protocol for the treatment of
acute adult asthma attacks. Am J Emerg Med 1993;11(2):131–133.
102. Reisner C, Kotch A, Dworkin G. Continuous versus frequent in-
termittent nebulization of albuterol in acute asthma: a randomized,
prospective study. Ann Allergy Asthma Immunol 1995;75(1):41–
47.
103. Moler FW, Johnson CE, Van Laanen C, Palmisano JM, Nasr SZ,
Akingbola O. Continuous versus intermittent nebulized terbutaline:
plasma levels and effects. Am J Respir Crit Care Med 1995;151(3
Pt 1):602–606.
104. Shrestha M, Bidadi K, Gourlay S, Hayes J. Continuous vs inter-
mittent albuterol, at high and low doses, in the treatment of severe
acute asthma in adults. Chest 1996;110(1):42–47.
105. Baker EK, Willsie SK, Marinac JS, Salzman GA. Continuously
nebulized albuterol in severe exacerbations of asthma in adults: a
case controlled study. J Asthma 1997;34:521–530.
106. Weber EJ, Levitt MA, Covington JK, Gambrioli E. Effect of con-
tinuously nebulized ipratropium bromide plus albuterol on emer-
gency department length of stay and hospital admission rates in
patients with acute bronchospasm: a randomized, controlled trial.
Chest 1999;115(4):937–944.
107. Portnoy J, Nadel G, Amado M, Willsie-Ediger S. Continuous neb-
ulization for status asthmaticus (review). Ann Allergy 1992;69(1):
71–79.
108. Voss KR, Willsie-Ediger SK, Pyszczynske DR, Nelson KA. De-
scription of a delivery method for continuously aerosolized albu-
terol in status asthmaticus. J Asthma 1990;27(1):37–39.
109. Chipps BE, Blackney DA, Black LE, Moody RR, Marino JT, Ridell
RC, Wong GA. Vortran high output extended aerosol respiratory
therapy (HEART) for delivery of continuously nebulized terbutal-
ine for the treatment of acute bronchospasm. Pediatr Asthma Al-
lergy Immunol 1990;471–277.
110. Berlinski A, Waldrep JC. Four hours of continuous albuterol neb-
ulization. Chest 1998;114(3):847–853.
111. Reisner C, Lee J, Kotch A, Dworkin G. Comparison of volume
output from two different continuous nebulizer systems. Ann Al-
lergy Asthma Immunol 1996;76(2):209–213.
112. Raabe OG, Wong TM, Wong GB, Roxburgh JW, Piper SD, Lee JI.
Continuous nebulization therapy for asthma with aerosols of beta2
agonists. Ann Allergy Asthma Immunol 1998;80(6):499–508.
113. McPeck M, Tandon R, Hughes K, Smaldone GC. Aerosol delivery
during continuous nebulization. Chest 1997;111(5):1200–1205.
114. Corkery KJ, Luce JM, Montgomery AB. Aerosolized pentamidine
for treatment and prophylaxis of Pneumocystis carinii pneumonia:
an update. Respir Care 1988;33(8):676–685.
115. Vinciguerra C, Smaldone G. Treatment time and patient tolerance
for pentamidine delivery by Respirgard II and AeroTech II. Respir
Care 1990;35(11):1037–1041.
116. Matthys H, Herceg R. Dosing strategies for aerosol delivery to the
lung parenchyma, with specific recommendations for pentamidine.
Respir Care 1991;36(9):989–993.
117. Demers RR, Parker J, Frankel LR, Smith DW. Administration of
ribavirin to neonatal and pediatric patients during mechanical ven-
tilation. Respir Care 1986;31(12):1188–1195.
118. American Academy of Pediatrics Committee on Infectious Dis-
eases. Use of ribavirin in the treatment of respiratory syncytial virus
infection. Pediatrics 1993;92(3):501–504.
119. Smith DW, Frankel LR, Mathers LH, Tang AT, Ariagno RL, Prober
CG. A controlled trial of aerosolized ribavirin in infants receiving
mechanical ventilation for severe respiratory syncytial virus infec-
tion. N Engl J Med 1991;325(1):24–29.
120. Kacmarek RM, Kratohvil J. Evaluation of a double-enclosure dou-
ble-vacuum unit scavenging system for ribavirin administration.
Respir Care 1992;37(1):37–45.
121. Charney W, Corkery KJ, Kraemer R, Wugofski L. Engineering and
administrative controls to contain aerosolized ribavirin: results of
simulation and application to one patient. Respir Care 1990;35(11):
1042–1048.
122. Wahlin B, Malmstrom B, Soop M, Hellstrom LG. A pediatric can-
opy system for aerosol administration and minimized environmen-
tal pollution. Acta Anaesthesiol Scand 1996;40(8 Pt 1):932–939.
123. Stevens HR, Albregt HB. Assessment of ultrasonic nebulization.
Anesthesiology 1966;27(5):648–653.
124. Modell JH, Giammona ST, Davis JH. Effect of chronic exposure to
ultrasonic aerosols on the lung. Anesthesiology 1967;28(4):680–
688.
125. Dennis JH, Stenton SC, Beach JR, Avery AJ, Walters EH, Hendrick
DJ. Jet and ultrasonic nebuliser output: use of a new method for
direct measurement of aerosol output. Thorax 1990;45(10):728–
732.
126. Newman SP, Pellow GD, Clarke SW. In vitro comparison of
DeVilbiss jet and ultrasonic nebulizers. Chest 1987;92(6):991–994.
127. Fok TF, Lam K, Ng PC, So HK, Cheung KL, Wong W, So KW.
Randomised crossover trial of salbutamol aerosol delivered by me-
tered-dose inhaler, jet nebuliser, and ultrasonic nebuliser in chronic
lung disease. Arch Dis Child Fetal Neonatal Ed 1998;79(2):F100–
F104.
128. Ballard RD, Bogin RM, Pak J. Assessment of bronchodilator re-
sponse to a beta-adrenergic delivered from an ultrasonic nebulizer.
Chest 1991;100(2):410–415.
129. Nakanishi AK, Lamb BM, Foster C, Rubin BK. Ultrasonic nebu-
lization of albuterol is no more effective than jet nebulization for
the treatment of acute asthma in children. Chest 1997;111(6):1505–
1508.
130. Larsen KR, Svendsen UG, Molgaard F, Petersen BN. Comparabil-
ity of albuterol delivered by a piezoelectric device versus metered-
dose inhaler in patients with chronic obstructive airways disease. J
Aerosol Med 1998;11(2):81–88.
131. Olivenstein R, Wolkove N, Cohen C, Frank H, Kreisman H. A
comparison of responses to albuterol delivered by two aerosol de-
vices. Chest 1986;90(3):392–395.
132. Eisenberg J, Pepe M, Williams-Warren J, Vasiliev M, Montgomery
AB, Smith AL, Ramsey BW. A comparison of peak sputum tobra-
mycin concentration in patients with cystic fibrosis using jet and
ultrasonic nebulizer systems. Aerosolized Tobramycin Study Group.
Chest 1997;111(4):955–962.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6 621
14. 133. O’Riordan TG, Amram JC. Effect of nebulizer configuration on de-
livery of aerosolized tobramycin. J Aerosol Med 1997;10(1):13–23.
134. Weber A, Morlin G, Cohen M, Williams-Warren J, Ramsey B,
Smith A. Effect of nebulizer type and antibiotic concentration on
device performance. Pediatr Pulmonol 1997;23(4):249–260.
135. Weber A, Smith A, Williams-Warren J, Ramsey B, Covert DS.
Nebulizer delivery of tobramycin to the lower respiratory tract.
Pediatr Pulmonol 1994;17(5):331–339.
136. Harvey CJ, O’Doherty MJ, Page CJ, Thomas SHL, Nunan TO,
Treacher DF. Comparison of jet and ultrasonic nebulizer pulmonary
aerosol deposition during mechanical ventilation. Eur Respir J 1997;
10(4):905–909.
137. Thomas SH, O’Doherty MJ, Page CJ, Treacher DF, Nunan TO.
Delivery of ultrasonic nebulized aerosols to a lung model during me-
chanical ventilation. Am Rev Respir Dis 1993;148(4 Pt 1):872–877.
138. Kemming GI, Kreyling W, Habler O, Merkel M, Kleen M, Welte
M, et al. Aerosol production and aerosol droplet size distribution
during mechanical ventilation (IPPV) with a new ultrasonic nebu-
lizer. Eur J Med Res 1996;1(7):321–327.
139. Williams L, Fletcher GC, Daniel M, Kinsella J. A simple in vitro
method for the evaluatin of an ultrasonic nebulizer for drug delivery
to intubated, ventilated patients and the effect of nebulizer and
ventilator settings on the uptake of fluid from the nebulizer cham-
ber. Eur J Anaesthesiol 1999;16(7):479–484.
140. Greenspan BJ. Ultrasonic and electrohydrodynamic methods for
aerosol generation. In: Hickey AJ, editor. Inhalation aerosols: phys-
ical and biological basis for therapy. New York: Marcel Decker;
1996: 313–335.
141. Gale AE. Drug degeneration during ultrasonic nebulization. J Aero-
sol Sci 1985;16:265.
NEBULIZERS: PRINCIPLES AND PERFORMANCE
622 RESPIRATORY CARE • JUNE 2000 VOL 45 NO 6