SlideShare a Scribd company logo
KAVITA SHASTRI
OPTOMETRIST(TMH)
CONTENTS :
Introduction
Corneal Anatomy
Techniques
Clinical utility
INTRODUCTION :
 Pachymetry (Greek words: Pachos= thick+metry) is term used for
measurement of corneal thickness.
 It is important indicator of health status of the cornea especially
of corneal endothelial function.
 The thickness of cornea was first reported in ancient textbooks
on physiological optics (Helmholtz and Gullstrand).
CORNEAL ANATOMY :
CORNEAL THICKNESS IN NORMAL EYES
 It ranges from 0.7-0.9 mm at the limbus and varies
between 0.49-0.56 mm at the center.
 The CCT reading of 0.7 mm or more is indicative of
endothelial decompensation.
 Peripheral corneal thickness is asymmetric so that
temporal cornea is thinnest followed by the inferior
cornea.
CORNEAL THICKNESS IN NEWBORNS AND
CHILDREN :
 Is similar to that of adult cornea.
 It has been found that cornea on day one is significantly thicker and
decreases in thickness as child grows older.
 The average corneal thickness in infants is 585±52 microns.
 The superior peripheral cornea is thinnest in newborn.
Normal Values
Normal corneal thickness – 500-575 microns.
 Abnormal results
Abnormally thick or thin measurements may indicate –
 corneal thinning.
 Corneal edema
 Lower or higher than suspected IOP.
FACTORS AFFECTING CCT :
 Higher in young patients, male patients and diabetic
patients.
 The mean CCT in black children is thinner than that
of white children.
 CCT does not correlate with refraction or systemic
hypertension.
TECHNIQUES :
ULTRASONIC OPTICAL ALTERNATIVE
MEASUREMENTS
• Conventional ultrasonic
pachymetry
• Manual • Pentacam
• Ultrasound
biomicroscpy (UBM).
• Specular microscopy • Pachycam
• Optical coherence
Tomography(OCT)
• Ocular response
analyzer (ORA).
• Optical coherence
interferometry
• Confocal microscopy
• Laser Doppler
interferometry
• Scanning Slit
technology
1.ULTRASONIC PACHYMETRY :
 Broadly accepted as method of choice, regarded as the gold
standard.
 PRINCIPLE :
 Instruments functions by measuring the amount of time
(transmit time) needed for ultrasound pulse pass from one end
of transducer to descemet’s membrane and back to transducer.
 Corneal thickness=transmit time × propagation velocity)/2
 Speed of sound in cornea : current standard is 1640 m/sec
COMPONENTS :
 PROBE HANDLE - It has piezoelectric crystal that emits
an ultrasonic beam of 20 MHz
 TRANSDUCER – It sends ultrasound rays through the
probe to the cornea and receives echoes from the cornea.
 TIP - The diameter of the tip should not be more than
2mm.
ADVANTAGES DISADVANTAGES
• Faster ,simpler and easy to use. • Contact method.
• Consistent and eliminates
interoobserver variation.
• Accuracy is dependent on the
perpendicularity of the probe’s
application to the cornea.
• Portable • Reproducibility relies on the precise
probe placement on the center of the
cornea.
• Dry(no coupling medium
required)
• Low resolution.
• Can be used intraoperatively. • Not accurate in edematous cornea.
 High resolution ultrasound machine which images anterior segment of
the eye.
 Using a 50-MHz imaging probe or high-frequency UBM, reproducible
images of the cross-sectional anterior chamber anatomy with high
resolution are achieved.
 12.5–50 MHz probe , the depth of penetration is lesser (4mm)than
conventional ultrasound.
2. ULTRASOUND
BIOMICROSCOPY:
ADVANTAGES DISADVANTAGES
• AS+ Corneal thickness • The main limitation of UBM is that it
requires a water-bath coupling media
and a very experienced examiner.
• Useful in cases where cornea is opaque • Contact method and patient to lie
supine
• Various layers of cornea can be
identified.
• Cannot be used intraoperatively.
3.MANUAL OPTICAL PACHYMETER:
 This was the original method to measure
corneal thickness.
 The instrument contains two Plano glass
plates that splits the image of the corneal parallelepiped.
 There are two methods to measure corneal thickness.
 Just touch method.
 Overlap method.
 The corneal thickness is then directly read from the scale.
 Range is 0-1.2mm, with a least gradation of 0.02 mm.
ADVANTAGE DISADVANTAGE
• Non contact method. • Lack of accuracy in measurements,
the usual range of error with an
optical pachymeter is ±2%
• Lack of repeatability.
• Requires slit lamp and therefore has
poor portability and cannot be used
in operating room.
4. SPECULAR PACHYMETRY :
 This is the oldest method to measure corneal thickness.
 PRINCIPLE : This measures the distance between the
anterior and the posterior surfaces of cornea and depends
on the focusing of light rays through front back cornea
unlike sound waves in ultrasound pachymeter.
 2 types 1.contact
2.Non contact
ADVANTAGES DISADVANTAGES
• Operator independent. • The exact point where reading is taken is
not known.
• Non invasive. • Time consuming.
• Simultaneous measurement of
cell count.
• Less reproducible.
• Impractical to use in operation room.
• Clinical use is limited to corneas free of
edema, scarring, deposits or opacities
that may distort light transmission.
5.SLIT LAMP PACHYMETER:
 Elevation based system which uses scanning slit
technology.
 Capable of assessment of multiple functions thickness
profile, anterior and posterior topography, elevation,
and anterior chamber depth.
 PRINCIPLE : It measures anterior and posterior
elevations by comparing it to a best fit sphere.
Demonstration of slit-lamp adapted ultrasonic pachymetry. The ultrasound probe is
inserted into the Goldmann tonometry mount, and under joystick control, it can be
softly and precisely used to touch the surface of the cornea.
ADVANTAGES DISADVANTAGES
• It gives wide field pachymetry
measures across the entire cornea.
• It overestimates corneal thickness by
5%
• It also identifies the thinnest part of
cornea.
• Tendency to underestimate corneal
thickness in keratoconic, post PRK,
and post-Lasik eyes.
• Corneal alignment is not required • The measurements are adjusted for
normal prolate shape of cornea.
Change of shape may interfere with
the reconstruction algorithms
• Can be used to calculate optical
zones in corneal refractive surgeries.
• Decreased accuracy in measuring
corneal thickness when clinically
significant haze is present.
6. ANTERIOR SEGMENT OPTICAL
COHERANCE TOMOGRAPHY :
 High resolution, non-contact OCT customized
anterior segment.
 Gives color coded map of corneal thickness.
ADVANTAGES :
 Non contact
 Rapid acquisition during the pachymetry scan ensures an
accurate and repeatable pachymetry map.
 High resolution
 It measures and documents both corneal flap thickness and
residual stromal thickness immediately following LASIK
surgery.
 Measures through corneal opacity.
7. OPTICAL LOW COHERENCE REFLECTOMETRY:
 The instrument is attached
to a slit lamp.
 Measure corneal thickness
to a precision of 1micron.
 PRINCIPLE:
Based on Michelson interferometer. It uses diode laser
beam.
ADVANTAGES DISADVANTAGES
• Precise 1 micron measurement. • Measures only central corneal
thickness.
• Automatic alignment.
• Non-contact.
• Real-time data acquisition and
display.
• Convenient and easy.
• Variability of measurement is
significantly lower.
• Intraoperative measurements
possible.
8. CONFOCAL MICROSCOPY:
 This unique method offers the ability to examine objects at
high magnification ranging from 20 X to 500 X.
 It also measures thickness of each layer by using
computerized scanning system providing the total corneal
thickness in studied area.
 Beside endothelium examination also measures endothelial
cell count (density) which is comparable to specular
microscopy.
 It offers the possibility to visualize structures posterior to
haze, scars or edema in the cornea.
ADVANTAGES DISADVANTAGES
• Moderate to good repeatability,
particularly for measurements of thin
layers such as epithelial and bowman’s
layer thickness.
• Poor agreement with ultrasound
pachymetry, it apparently
overestimates corneal thickness.
• Following LASIK surgery, flap
thickness can also be obtained.
• The precision of measurements with
this technique will vary with contact
lens hydration, post-lens tear film
thickness and observation angle.
• The Z-scan curve can be used to assess
the level and location of corneal haze
associated with the various corneal
dystrophies.
• Slower data acquisition.
• Poor penetration of corneal opacity.
9. LASER DOPPLER INTERFEROMETRY :
 It is a noncontact technique that uses a dual-beam
infrared laser Doppler interferometry to measure
corneal thickness.
 There is limited data available in the literature
regarding this technique.
10. PENTACAM :
 It is3-dimensional (3D) rotating scheimpflug.
 It can perform five functions in 2 sec.
1. Scheimpflug image of anterior segment.
2. Pachymetry
3. 3-D anterior chamber analyzer
4. Corneal topography.
5. Cataract analyzer.
ADVANTAGES DISADVANTAGES
• Non invasiveness, non contact. • It underestimates the corneal
thickness in comparison to
ultrasonic pachymetry.
• Even minute eye movement are
captured and corrected
simultaneously.
• Gives precise representation and
repeatability.
APPLICATIONS:
1. Preoperative planning for corneal refractive surgery
2. Glaucoma screening
3. IOP modification with regard to corneal thickness
4. Keratoconus detection and quantification.
11. PACHYCAM :
 Compact and portable non-contact pachymeter with built in keratometer. It
can be mounted on slit lamp. It automatically corrects IOP.
 Image acquisition is done with help of a 3-D alignment screen.
ADVATAGES :
1. Non contact.
2. Compact, portable and light weight.
12. OCULAR RESPONSE
ANALYZER :
 Newer modality for measuring biomechanical properties of cornea.
 Utilizes a rapid air impulse and measures delays in inward and outward
applanation of cornea resulting in two different pressure values
 The difference in two pressure values gives corneal thickness.
CLINICAL UTILITY
 LASIK
 GLAUCOMA
 CORNEAL TRANSPLANT
 KERATOCONOUS
USE IN LASIK
Essential prior to a Lasik procedure for ensuring
sufficient corneal thickness to prevent abnormal
bulging of cornea ( ectasia). Differences in CCT
between pre op and intra op readings.
Differences in central corneal thickness between
pre op and intra op reading. Intra op readings are
always used for setting blasé length and depth of
incision.
USE IN GLAUCOMA
 CCT statistically significant predictor of development of glaucoma.
 In congenital glaucoma to assess amount of corneal edema.
 For applying correction factor in actual IOP determination.
 Results suggested that IOP measurements need to be adjusted for
abnormally thick or thin corneas.
 The target IOP is lower for a thin cornea and higher for thick
cornea.
• Eyes with thick corneas have a true IOP that is lower than the
measured IOP.
• Thus individuals with thicker cornea may be mis- classified as having
ocular hypertension.
FACTS ABOUT CCT IN GLAUCOMA :
 Bechmann in the yr. 2000 found following association of CCT with
different forms of glaucoma.
 Increased CCT measurement are found in patients with ocular
hypertension, which can lead to falsely elevated IOP readings.
 Decreased CCT is found in patients with low tension glaucoma, resulting
in falsely reduced IOP measurements.
 CCT is found to be lower in patients with Pseudoexfoliation syndrome
(PXS) and in POAG.
 There is no difference in corneal thickness in individuals with Pigmentary
glaucoma and PACG.
CORRECTION FACTOR FOR IOP BASED CCT
MEASUREMENT:
 The adjustment factor is applied to the measured
IOP.
CCT
(micro
meters)
IOP
adjustment
(mmHg
CCT
(micro
meters)
IOP
adjustment
(mmHg)
CCT
(micro
meters)
IOP
adjustment
(mmHg)
445 +7 515 +2 585 -3
455 +6 522 +1 595 -4
465 +6 535 +1 605 -4
475 +5 545 0 615 -5
485 +4 555 -1 625 -6
495 +4 565 -1 635 -6
505 +3 575 -2 645 -7
CORNEAL TRANSPLANTS GRAFTS
Assessing candidates for penetrating
keratoplasty(corneal transplant)
Post operative follow up of keratoplasty patients to
determine endothelial cell function and its
recovery and to become alert to early graft
decompensation.
KERATOCONOUS
Keratoconous is associated with corneal thinning.
Available evidence indicates that ultrasonic
corneal pachymetry is not as accurate as
videokeratography in diagnosing keratoconus.
False +ve and false –ve rates are unacceptably
higher than those obtained by videokeratography.
CLINICAL UTILITY
ASSESSING CORNEA THINNESS:
as in corneal disorders like Terriens and Pellucid
marginal degenerations , keratoconus,keratoglobus ,
post LASIK ectasia.
Other cases of corneal decompensation as in herpetic
endothelitis.
CONTACT LENS CARE:
To access corneal edema.
TAKE HOME MESSAGE :
 The methodologies used in these techniques are based on either ultrasonic or
optical principles.
 While each of the methods have a peculiarity of their own, all have been described
as reliable.
 Keep in mind that systematic differences exist between the different techniques
and result in different values.
 As a result, the measurements cannot simply be substituted between the different
modalities.
 corneal Pachymetry

More Related Content

What's hot

Keratometry
KeratometryKeratometry
Keratometry
ankita mahapatra
 
Binocular balancing
Binocular balancing Binocular balancing
Binocular balancing
Tahseen Jawaid
 
AC/A
AC/AAC/A
AC/A
zarin45
 
Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)
Azizul Islam
 
Trial set
Trial setTrial set
Bandage Contact Lens
Bandage Contact LensBandage Contact Lens
Bandage Contact Lens
Loknath Goswami
 
Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing Mohammad Arman Bin Aziz
 
Testing for npa
Testing for npaTesting for npa
Testing for npa
RAJU RATHORE ™️
 
Pantoscopic tilt
Pantoscopic tiltPantoscopic tilt
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
Azizul Islam
 
Vertex distance and power
Vertex distance and powerVertex distance and power
Dynamic retinoscopy
Dynamic retinoscopyDynamic retinoscopy
Dynamic retinoscopy
Jinal chauhan
 
Soft Contact Lenses: Material, Fitting, and Evaluation
Soft Contact Lenses: Material, Fitting, and EvaluationSoft Contact Lenses: Material, Fitting, and Evaluation
Soft Contact Lenses: Material, Fitting, and Evaluation
Zahra Heidari
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
Indra Prasad Sharma
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
Satish Jeria
 
Scleral contact lens in Ophthalmology
Scleral contact lens in OphthalmologyScleral contact lens in Ophthalmology
Scleral contact lens in Ophthalmology
DrArvindMorya
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondence
Rajeshwori
 
Maddox rod n wing
Maddox rod n wingMaddox rod n wing
Maddox rod n wing
Rizwan Sajjad Chachar
 

What's hot (20)

Keratometry
KeratometryKeratometry
Keratometry
 
Binocular balancing
Binocular balancing Binocular balancing
Binocular balancing
 
AC/A
AC/AAC/A
AC/A
 
Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)
 
Trial set
Trial setTrial set
Trial set
 
Bandage Contact Lens
Bandage Contact LensBandage Contact Lens
Bandage Contact Lens
 
Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing Binocular refraction techniques, binocular balancing
Binocular refraction techniques, binocular balancing
 
Testing for npa
Testing for npaTesting for npa
Testing for npa
 
Pantoscopic tilt
Pantoscopic tiltPantoscopic tilt
Pantoscopic tilt
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Vertex distance and power
Vertex distance and powerVertex distance and power
Vertex distance and power
 
Dynamic retinoscopy
Dynamic retinoscopyDynamic retinoscopy
Dynamic retinoscopy
 
Soft Contact Lenses: Material, Fitting, and Evaluation
Soft Contact Lenses: Material, Fitting, and EvaluationSoft Contact Lenses: Material, Fitting, and Evaluation
Soft Contact Lenses: Material, Fitting, and Evaluation
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Scleral contact lens in Ophthalmology
Scleral contact lens in OphthalmologyScleral contact lens in Ophthalmology
Scleral contact lens in Ophthalmology
 
anomalous retinal correspondence
anomalous retinal correspondenceanomalous retinal correspondence
anomalous retinal correspondence
 
keratometry
 keratometry keratometry
keratometry
 
Maddox rod n wing
Maddox rod n wingMaddox rod n wing
Maddox rod n wing
 
Amsler grid
Amsler gridAmsler grid
Amsler grid
 

Similar to corneal Pachymetry

Corneal pachymetry.ppt
Corneal pachymetry.pptCorneal pachymetry.ppt
Corneal pachymetry.ppt
AhmedAli191572
 
pachy.pptx
pachy.pptxpachy.pptx
pachy.pptx
Divya785180
 
pachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnosticspachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnosticsPaavan Kalra
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipment
Devdutta Nayak
 
Optical Coherence Tomography
Optical Coherence TomographyOptical Coherence Tomography
Optical Coherence Tomography
Tushya Parkash
 
A SCAN, B SCAN & BIOMETRY---19.pptx
A SCAN, B SCAN & BIOMETRY---19.pptxA SCAN, B SCAN & BIOMETRY---19.pptx
A SCAN, B SCAN & BIOMETRY---19.pptx
LavanyaMadabushi
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdf
ssuser0f453c
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculations
rakesh jaiswal
 
AS OCT & UBM - Dr Shylesh B Dabke
AS OCT & UBM - Dr Shylesh B DabkeAS OCT & UBM - Dr Shylesh B Dabke
AS OCT & UBM - Dr Shylesh B Dabke
Shylesh Dabke
 
Oct introduction
Oct  introductionOct  introduction
Oct introduction
Eranda Wannigama
 
OCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , VitrectomyOCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , Vitrectomy
Al Amin
 
Oct
OctOct
OCT
OCTOCT
Newer tests for glaucoma
Newer tests for glaucomaNewer tests for glaucoma
Newer tests for glaucoma
vaishusmail
 
Oct in post seg disorders
Oct in post seg disordersOct in post seg disorders
Oct in post seg disorders
Dinesh Madduri
 
OPTICAL COHERENCE DEMYSTIFIED
OPTICAL COHERENCE DEMYSTIFIED OPTICAL COHERENCE DEMYSTIFIED
OPTICAL COHERENCE DEMYSTIFIED
DINESH and SONALEE
 
biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.ppt
mikaelgirum
 
Refractive surgeries.pptx arjun
Refractive surgeries.pptx arjunRefractive surgeries.pptx arjun
Refractive surgeries.pptx arjun
arjun sapkota
 
LASIK.pptx
LASIK.pptxLASIK.pptx

Similar to corneal Pachymetry (20)

Corneal pachymetry.ppt
Corneal pachymetry.pptCorneal pachymetry.ppt
Corneal pachymetry.ppt
 
pachy.pptx
pachy.pptxpachy.pptx
pachy.pptx
 
pachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnosticspachymetry confocal microscopy cornea ophthalmology diagnostics
pachymetry confocal microscopy cornea ophthalmology diagnostics
 
Pachymetry sivateja
Pachymetry sivatejaPachymetry sivateja
Pachymetry sivateja
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipment
 
Optical Coherence Tomography
Optical Coherence TomographyOptical Coherence Tomography
Optical Coherence Tomography
 
A SCAN, B SCAN & BIOMETRY---19.pptx
A SCAN, B SCAN & BIOMETRY---19.pptxA SCAN, B SCAN & BIOMETRY---19.pptx
A SCAN, B SCAN & BIOMETRY---19.pptx
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdf
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculations
 
AS OCT & UBM - Dr Shylesh B Dabke
AS OCT & UBM - Dr Shylesh B DabkeAS OCT & UBM - Dr Shylesh B Dabke
AS OCT & UBM - Dr Shylesh B Dabke
 
Oct introduction
Oct  introductionOct  introduction
Oct introduction
 
OCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , VitrectomyOCT , Laser therapy for DR , Vitrectomy
OCT , Laser therapy for DR , Vitrectomy
 
Oct
OctOct
Oct
 
OCT
OCTOCT
OCT
 
Newer tests for glaucoma
Newer tests for glaucomaNewer tests for glaucoma
Newer tests for glaucoma
 
Oct in post seg disorders
Oct in post seg disordersOct in post seg disorders
Oct in post seg disorders
 
OPTICAL COHERENCE DEMYSTIFIED
OPTICAL COHERENCE DEMYSTIFIED OPTICAL COHERENCE DEMYSTIFIED
OPTICAL COHERENCE DEMYSTIFIED
 
biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.ppt
 
Refractive surgeries.pptx arjun
Refractive surgeries.pptx arjunRefractive surgeries.pptx arjun
Refractive surgeries.pptx arjun
 
LASIK.pptx
LASIK.pptxLASIK.pptx
LASIK.pptx
 

Recently uploaded

Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Savita Shen $i11
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
Little Cross Family Clinic
 
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
i3 Health
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
pal078100
 
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
Catherine Liao
 
Flu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore KarnatakaFlu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore Karnataka
addon Scans
 
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.GawadHemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
NephroTube - Dr.Gawad
 
Non-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdfNon-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdf
MedicoseAcademics
 
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in StockFactory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
rebeccabio
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
greendigital
 
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #GirlsFor Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
Savita Shen $i11
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
Anujkumaranit
 
Antiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptxAntiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptx
Rohit chaurpagar
 
24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all
DrSathishMS1
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Dr KHALID B.M
 
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists  Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists
Saeid Safari
 
Physiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of TastePhysiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of Taste
MedicoseAcademics
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
Levi Shapiro
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
LanceCatedral
 
THOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation ActTHOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation Act
DrSathishMS1
 

Recently uploaded (20)

Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
 
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
 
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...The hemodynamic and autonomic determinants of elevated blood pressure in obes...
The hemodynamic and autonomic determinants of elevated blood pressure in obes...
 
Flu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore KarnatakaFlu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore Karnataka
 
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.GawadHemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
 
Non-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdfNon-respiratory Functions of the Lungs.pdf
Non-respiratory Functions of the Lungs.pdf
 
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in StockFactory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
Factory Supply Best Quality Pmk Oil CAS 28578–16–7 PMK Powder in Stock
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
 
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #GirlsFor Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
 
Antiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptxAntiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptx
 
24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
 
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists  Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists
 
Physiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of TastePhysiology of Special Chemical Sensation of Taste
Physiology of Special Chemical Sensation of Taste
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
 
THOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation ActTHOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation Act
 

corneal Pachymetry

  • 3. INTRODUCTION :  Pachymetry (Greek words: Pachos= thick+metry) is term used for measurement of corneal thickness.  It is important indicator of health status of the cornea especially of corneal endothelial function.  The thickness of cornea was first reported in ancient textbooks on physiological optics (Helmholtz and Gullstrand).
  • 5. CORNEAL THICKNESS IN NORMAL EYES  It ranges from 0.7-0.9 mm at the limbus and varies between 0.49-0.56 mm at the center.  The CCT reading of 0.7 mm or more is indicative of endothelial decompensation.  Peripheral corneal thickness is asymmetric so that temporal cornea is thinnest followed by the inferior cornea.
  • 6. CORNEAL THICKNESS IN NEWBORNS AND CHILDREN :  Is similar to that of adult cornea.  It has been found that cornea on day one is significantly thicker and decreases in thickness as child grows older.  The average corneal thickness in infants is 585±52 microns.  The superior peripheral cornea is thinnest in newborn.
  • 7. Normal Values Normal corneal thickness – 500-575 microns.  Abnormal results Abnormally thick or thin measurements may indicate –  corneal thinning.  Corneal edema  Lower or higher than suspected IOP.
  • 8. FACTORS AFFECTING CCT :  Higher in young patients, male patients and diabetic patients.  The mean CCT in black children is thinner than that of white children.  CCT does not correlate with refraction or systemic hypertension.
  • 9. TECHNIQUES : ULTRASONIC OPTICAL ALTERNATIVE MEASUREMENTS • Conventional ultrasonic pachymetry • Manual • Pentacam • Ultrasound biomicroscpy (UBM). • Specular microscopy • Pachycam • Optical coherence Tomography(OCT) • Ocular response analyzer (ORA). • Optical coherence interferometry • Confocal microscopy • Laser Doppler interferometry • Scanning Slit technology
  • 10. 1.ULTRASONIC PACHYMETRY :  Broadly accepted as method of choice, regarded as the gold standard.  PRINCIPLE :  Instruments functions by measuring the amount of time (transmit time) needed for ultrasound pulse pass from one end of transducer to descemet’s membrane and back to transducer.  Corneal thickness=transmit time × propagation velocity)/2  Speed of sound in cornea : current standard is 1640 m/sec
  • 11. COMPONENTS :  PROBE HANDLE - It has piezoelectric crystal that emits an ultrasonic beam of 20 MHz  TRANSDUCER – It sends ultrasound rays through the probe to the cornea and receives echoes from the cornea.  TIP - The diameter of the tip should not be more than 2mm.
  • 12.
  • 13. ADVANTAGES DISADVANTAGES • Faster ,simpler and easy to use. • Contact method. • Consistent and eliminates interoobserver variation. • Accuracy is dependent on the perpendicularity of the probe’s application to the cornea. • Portable • Reproducibility relies on the precise probe placement on the center of the cornea. • Dry(no coupling medium required) • Low resolution. • Can be used intraoperatively. • Not accurate in edematous cornea.
  • 14.  High resolution ultrasound machine which images anterior segment of the eye.  Using a 50-MHz imaging probe or high-frequency UBM, reproducible images of the cross-sectional anterior chamber anatomy with high resolution are achieved.  12.5–50 MHz probe , the depth of penetration is lesser (4mm)than conventional ultrasound. 2. ULTRASOUND BIOMICROSCOPY:
  • 15. ADVANTAGES DISADVANTAGES • AS+ Corneal thickness • The main limitation of UBM is that it requires a water-bath coupling media and a very experienced examiner. • Useful in cases where cornea is opaque • Contact method and patient to lie supine • Various layers of cornea can be identified. • Cannot be used intraoperatively.
  • 16. 3.MANUAL OPTICAL PACHYMETER:  This was the original method to measure corneal thickness.  The instrument contains two Plano glass plates that splits the image of the corneal parallelepiped.  There are two methods to measure corneal thickness.  Just touch method.  Overlap method.  The corneal thickness is then directly read from the scale.  Range is 0-1.2mm, with a least gradation of 0.02 mm.
  • 17. ADVANTAGE DISADVANTAGE • Non contact method. • Lack of accuracy in measurements, the usual range of error with an optical pachymeter is ±2% • Lack of repeatability. • Requires slit lamp and therefore has poor portability and cannot be used in operating room.
  • 18. 4. SPECULAR PACHYMETRY :  This is the oldest method to measure corneal thickness.  PRINCIPLE : This measures the distance between the anterior and the posterior surfaces of cornea and depends on the focusing of light rays through front back cornea unlike sound waves in ultrasound pachymeter.  2 types 1.contact 2.Non contact
  • 19.
  • 20. ADVANTAGES DISADVANTAGES • Operator independent. • The exact point where reading is taken is not known. • Non invasive. • Time consuming. • Simultaneous measurement of cell count. • Less reproducible. • Impractical to use in operation room. • Clinical use is limited to corneas free of edema, scarring, deposits or opacities that may distort light transmission.
  • 21. 5.SLIT LAMP PACHYMETER:  Elevation based system which uses scanning slit technology.  Capable of assessment of multiple functions thickness profile, anterior and posterior topography, elevation, and anterior chamber depth.  PRINCIPLE : It measures anterior and posterior elevations by comparing it to a best fit sphere.
  • 22. Demonstration of slit-lamp adapted ultrasonic pachymetry. The ultrasound probe is inserted into the Goldmann tonometry mount, and under joystick control, it can be softly and precisely used to touch the surface of the cornea.
  • 23. ADVANTAGES DISADVANTAGES • It gives wide field pachymetry measures across the entire cornea. • It overestimates corneal thickness by 5% • It also identifies the thinnest part of cornea. • Tendency to underestimate corneal thickness in keratoconic, post PRK, and post-Lasik eyes. • Corneal alignment is not required • The measurements are adjusted for normal prolate shape of cornea. Change of shape may interfere with the reconstruction algorithms • Can be used to calculate optical zones in corneal refractive surgeries. • Decreased accuracy in measuring corneal thickness when clinically significant haze is present.
  • 24. 6. ANTERIOR SEGMENT OPTICAL COHERANCE TOMOGRAPHY :  High resolution, non-contact OCT customized anterior segment.  Gives color coded map of corneal thickness.
  • 25.
  • 26. ADVANTAGES :  Non contact  Rapid acquisition during the pachymetry scan ensures an accurate and repeatable pachymetry map.  High resolution  It measures and documents both corneal flap thickness and residual stromal thickness immediately following LASIK surgery.  Measures through corneal opacity.
  • 27. 7. OPTICAL LOW COHERENCE REFLECTOMETRY:  The instrument is attached to a slit lamp.  Measure corneal thickness to a precision of 1micron.  PRINCIPLE: Based on Michelson interferometer. It uses diode laser beam.
  • 28. ADVANTAGES DISADVANTAGES • Precise 1 micron measurement. • Measures only central corneal thickness. • Automatic alignment. • Non-contact. • Real-time data acquisition and display. • Convenient and easy. • Variability of measurement is significantly lower. • Intraoperative measurements possible.
  • 29. 8. CONFOCAL MICROSCOPY:  This unique method offers the ability to examine objects at high magnification ranging from 20 X to 500 X.  It also measures thickness of each layer by using computerized scanning system providing the total corneal thickness in studied area.  Beside endothelium examination also measures endothelial cell count (density) which is comparable to specular microscopy.  It offers the possibility to visualize structures posterior to haze, scars or edema in the cornea.
  • 30.
  • 31. ADVANTAGES DISADVANTAGES • Moderate to good repeatability, particularly for measurements of thin layers such as epithelial and bowman’s layer thickness. • Poor agreement with ultrasound pachymetry, it apparently overestimates corneal thickness. • Following LASIK surgery, flap thickness can also be obtained. • The precision of measurements with this technique will vary with contact lens hydration, post-lens tear film thickness and observation angle. • The Z-scan curve can be used to assess the level and location of corneal haze associated with the various corneal dystrophies. • Slower data acquisition. • Poor penetration of corneal opacity.
  • 32. 9. LASER DOPPLER INTERFEROMETRY :  It is a noncontact technique that uses a dual-beam infrared laser Doppler interferometry to measure corneal thickness.  There is limited data available in the literature regarding this technique.
  • 33. 10. PENTACAM :  It is3-dimensional (3D) rotating scheimpflug.  It can perform five functions in 2 sec. 1. Scheimpflug image of anterior segment. 2. Pachymetry 3. 3-D anterior chamber analyzer 4. Corneal topography. 5. Cataract analyzer.
  • 34. ADVANTAGES DISADVANTAGES • Non invasiveness, non contact. • It underestimates the corneal thickness in comparison to ultrasonic pachymetry. • Even minute eye movement are captured and corrected simultaneously. • Gives precise representation and repeatability. APPLICATIONS: 1. Preoperative planning for corneal refractive surgery 2. Glaucoma screening 3. IOP modification with regard to corneal thickness 4. Keratoconus detection and quantification.
  • 35. 11. PACHYCAM :  Compact and portable non-contact pachymeter with built in keratometer. It can be mounted on slit lamp. It automatically corrects IOP.  Image acquisition is done with help of a 3-D alignment screen. ADVATAGES : 1. Non contact. 2. Compact, portable and light weight.
  • 36. 12. OCULAR RESPONSE ANALYZER :  Newer modality for measuring biomechanical properties of cornea.  Utilizes a rapid air impulse and measures delays in inward and outward applanation of cornea resulting in two different pressure values  The difference in two pressure values gives corneal thickness.
  • 37. CLINICAL UTILITY  LASIK  GLAUCOMA  CORNEAL TRANSPLANT  KERATOCONOUS
  • 38. USE IN LASIK Essential prior to a Lasik procedure for ensuring sufficient corneal thickness to prevent abnormal bulging of cornea ( ectasia). Differences in CCT between pre op and intra op readings. Differences in central corneal thickness between pre op and intra op reading. Intra op readings are always used for setting blasé length and depth of incision.
  • 39. USE IN GLAUCOMA  CCT statistically significant predictor of development of glaucoma.  In congenital glaucoma to assess amount of corneal edema.  For applying correction factor in actual IOP determination.  Results suggested that IOP measurements need to be adjusted for abnormally thick or thin corneas.  The target IOP is lower for a thin cornea and higher for thick cornea. • Eyes with thick corneas have a true IOP that is lower than the measured IOP. • Thus individuals with thicker cornea may be mis- classified as having ocular hypertension.
  • 40. FACTS ABOUT CCT IN GLAUCOMA :  Bechmann in the yr. 2000 found following association of CCT with different forms of glaucoma.  Increased CCT measurement are found in patients with ocular hypertension, which can lead to falsely elevated IOP readings.  Decreased CCT is found in patients with low tension glaucoma, resulting in falsely reduced IOP measurements.  CCT is found to be lower in patients with Pseudoexfoliation syndrome (PXS) and in POAG.  There is no difference in corneal thickness in individuals with Pigmentary glaucoma and PACG.
  • 41. CORRECTION FACTOR FOR IOP BASED CCT MEASUREMENT:  The adjustment factor is applied to the measured IOP. CCT (micro meters) IOP adjustment (mmHg CCT (micro meters) IOP adjustment (mmHg) CCT (micro meters) IOP adjustment (mmHg) 445 +7 515 +2 585 -3 455 +6 522 +1 595 -4 465 +6 535 +1 605 -4 475 +5 545 0 615 -5 485 +4 555 -1 625 -6 495 +4 565 -1 635 -6 505 +3 575 -2 645 -7
  • 42. CORNEAL TRANSPLANTS GRAFTS Assessing candidates for penetrating keratoplasty(corneal transplant) Post operative follow up of keratoplasty patients to determine endothelial cell function and its recovery and to become alert to early graft decompensation.
  • 43. KERATOCONOUS Keratoconous is associated with corneal thinning. Available evidence indicates that ultrasonic corneal pachymetry is not as accurate as videokeratography in diagnosing keratoconus. False +ve and false –ve rates are unacceptably higher than those obtained by videokeratography.
  • 44. CLINICAL UTILITY ASSESSING CORNEA THINNESS: as in corneal disorders like Terriens and Pellucid marginal degenerations , keratoconus,keratoglobus , post LASIK ectasia. Other cases of corneal decompensation as in herpetic endothelitis. CONTACT LENS CARE: To access corneal edema.
  • 45. TAKE HOME MESSAGE :  The methodologies used in these techniques are based on either ultrasonic or optical principles.  While each of the methods have a peculiarity of their own, all have been described as reliable.  Keep in mind that systematic differences exist between the different techniques and result in different values.  As a result, the measurements cannot simply be substituted between the different modalities.