Break Down voltage insulating oil
1. Test break down voltage
comparison between IEC 60156
and step voltage application method
Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
Statistical measurements: mean, standard deviation.....
2. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
Increase voltage: 2 kV/s ± 0,2 kV/s
oil stirring by
magnetic anchor
Waiting time
between tests: 30 s
INSTRUMENT
OPERATING CONDITIONS
Temperature Test: 20°C ± 5°C
Distance between
electrodes: 2.5 mm
3. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
OIL TYPE: MINERAL
METHOD OF INTERVENTION
Execution of 10 tests, replicated 10 times on the same sample
Calculation of the mean and standard deviation on each replica
Calculation of the mean and standard deviation on a pair of replicates (20
tests), eliminating the 4 HIGHEST and 4 lower measures of each replica
Calculation of the mean and standard deviation, the median, the mode and
the frequency of the total 100 tests
Water content: 9 mg/kg
4. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
1^ 2^ 3^ 4^ 5^ 6^ 7^ 8^ 9^ 10^
Mean 58 60 51 61 58 64 57 58 63 61
St. Dev. ± 8 ± 9 ± 9 ± 4 ± 6 ± 5 ± 6 ± 8 ± 5 ± 4
St. Dev. % 13 15 17 7 11 7 11 8 7 4
Sequences of 10 determinations each
mean St. Dev. St.Dev. %
5. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
1^+2^ 3^+4^ 5^+6^ 7^+8^ 9^+10^
mean 58 58 62 56 62
St. Dev. ± 3 ± 5 ± 9 ± 2 ± 2
St. Dev. % 5,8 8,4 2,8 3,3 3
Sequences of 20 determinations each, eliminating the 4 HIGHEST and 4 lowest measures
mean St. Dev. St.Dev. %
6. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
100 test
Average St. Dev. St. Dev.
%
Mode Median min.
value
Max
value
59,1 ± 4,3 7,3 56 59,5 39 74
classi frequenza
1 38 42 2
2 43 48 7
3 49 54 14
4 55 60 34
5 61 66 29
6 67 72 12
7 73 78 2
range
Kurtosis= - 0,158
«platicurtica»
Class Range Frequence
BDV FREQUENCE
9. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
Increase voltage: 2 kV/s ± 0,2 kV/s
Stirring: manual
1^phase: voltage application with a to 10 kV
for 10 seconds
OPERATIVE CONDITION
Temperature Test: 20°C ± 5°C
Waiting time
between tests: 30 s
2nd phase: application of the voltage with
a increase voltage of 2 kV/s ± 0,2 kV/s
until the cancellation of the applied electric
field
electrode distance: 2,5 mm
10. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
OIL TYPE: MINERAL
METHODS OF INTERVENTION
Execution of 10 tests, replicated 10 times on the same sample
Calculation of the mean each replica
Calculation of the mean on each pair of replicas (20 tests), eliminating the 4
highest and 4 lowest measures of each replica
Water content: 17 mg/kg
Calculation of the mean on each replica (10 tests), eliminating the 2 highest
and 2 lowest measurements of each replica
electrode distance: 2,5 mm
14. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
1^+2^ 3^+4^ 5^+6^ 7^+8^ 9^+10^
MEAN 56,7 53,3 53,2 57,6 56,1
St. Dev ± 1,2 ± 1,7 ± 1,2 ± 2,7 ± 1,4
St. Dev.% 2,1 2,1 2,3 4,7 2,1
Sequences of 20 determinations each, eliminating the 4 highest and 4 lowest measurements
TEST BDV FOR STEP
MEAN St. Dev. St.Dev. %
15. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
60 tests ( pairs of "truncated" replicas)
Mean Dev.St. Dev.st.
%
Mode Median min.
Value
Max
value
55,4 ± 1,6 2,7 55 55 50 62
Kurtosis= - 0,67
CLASSE RANGE FREQUENZA
1 49-51 7
2 52-54 15
3 55-57 23
4 58-60 13
5 61-63 2
BDV FREQUENCE
Class Range Frequence
17. Mean Dev.St.
55,4 ± 1,6
Dev.st.
Mean
0,028
Dev.st.
Mean
0,072
Mean Dev.St.
59,1 ± 4,3
Graphical representation of coefficient of variation
(standard deviation/mean ratio) versus mean breakdown voltage)
ref. IEC 60156
TEST FOR STEP IEC 60156
18. Pierluigi De Horatiis
Laboratory Quality Manager L.M.P. srl
Electrode cleaning
electrode distance control oil test cell filling
BEFORE
AFTER
20. THE TESTS WERE CARRIED OUT AT THE LABORATORY OF:
Team:
PIERLUIGI DE HORATIIS
FABRIZIO PANZAVOLTA
ROBERTO PETROCCHI
FLAVIO CALVANI