Introduction To Tissue Processor
• Purpose of Device
-A device intended for use in the histology laboratory to
process the tissues taken from the body for diagnosis of
disease processes and to produce microscopic slides that can
be viewed under the microscope.
-A tissue processor is used to prepare tissue samples for
analysis by fixing, staining, dehydrating or decalcifying them.
-The processors are mostly single unit devices that can
accommodate a variety of processing techniques to suit the
different needs of the laboratory, therefore improving the
efficiency of tissue processing
Figurative description about the device
Rear side of the device
Introduction…
• Device’s fundamental functional principle
-The basic principle of tissue processor is to transfer the tissue in
different fluid for a specified time in a desired environment. So far,
there are two types of tissue processor.
1.Tissue transfer processor
• Here, the bucket of tissue is transferred from a carousel to other after a
specified time. There are also several containers of reagents.
Cont…
• The tissue remains in a basket which contains 30-100 cassettes.
• The basket containing the tissue is then submerged in the specific
containers for a particular time which is automatically transferred to the
next container.
• A gentle agitation is created by rotatory movement or vertical oscillation
of the tissue in the basket which causes a continuous removal of fluid
from the surface by fresh medium.
Cont...
• A microprocessor determines the time schedule and transfer of tissue in
each container.
• A gentle agitation is created by vertical oscillation or by rotatory
movement of the tissue basket.
Cont...
2. Fluid transfer processor
-This is a closed system processor where tissues are kept in the container.
-The container is periodically filled with particular fluid.
-The fluid is pumped out from the container containing the tissue after a
certain period.
-The container is again filled with the fluid required for the next step. The
fluid transfer processor has some advantages such as the vacuum pressure
that makes the system faster and efficient and there is no chance for tissue
drying.
Device’s fundamental functional principle
• Tissue processing
-It is concerned with the diffusion of various substances into
and out of porous tissues.
-Diffusion results from the tendency of processing reagents to
equalize concentrations both inside and outside blocks of
tissue.
-The reagent molecules diffuse down a concentration gradient
and move from where they are at a high concentration to
where they are at a lower concentration.
Cont…
-This movement requires no energy since it always
progresses down the concentration gradient.
Cont…
• The rate of diffusion is measured by Fick’s law which states that
Introduction…
• General Set-up and Requirements
-Activating the vacuum function(instruments with the vacuum function
only)
- Press ARROW RIGHT to move to the next entry position, the cursor
blinks at the two positions next to the ‘V’ for vacuum function
- To activate the vacuum, press PLUS
Cont…
-To deactivate the vacuum press MINUS
-Selecting infiltration times per station
-Press ARROW RIGHT to move to the next entry position ,the cursor blinks
at the entry digits
-
Cont…
-Press PLUS or MINUS respectively to enter the hours.
- To move the cursor to the next digit for data entry, push the ARROW
RIGHT key, the cursor flashes on both digits destined for entering the
minutes
- Press the PLUS or MINUS to enter the minutes.
Cont…
-Displaying total program duration.
-Once a program is set up, total run time can be displayed.
- Press CLOCK to display the total run time of the program.
Cont…
-The total duration displayed here is 1 day , 0hours and 12 minutes.
-To quit the total run time indication , press CLOCK.
-To end programming you have to leave the programming mode.
-Press PROG to quit programming.
Servicing Technician related knowledge
Iteration of functional Principles and Device set-up
- The principle of tissue processing is designed to remove all extractable
water from the tissue, replacing it with a support medium that provides
sufficient rigidity to enable sectioning of the tissue without damage or
distortion.
-The principle is based on the use of chemicals dehydrate and clear the
tissue so that it can be embedded into molten paraffin wax which stands
as a support medium for the tissue in order to easy sectioning
Iteration of functional Principles and
Device set-up…
DEHYDRATION
- The first stage in tissue processing is dehydration (the removal of water).
In tissues, water is present in both free and bound forms and needs to be
removed before processing can continue.
-Dehydration is usually carried out using alcohols (such as ethanol) but
these can dissolve certain cellular components such as lipids.
-Although dehydration can also cause tissue shrinkage, the stage is
necessary in all infiltration methods, except where tissues are supported
by an aqueous embedding medium (such as water-soluble waxes).
Iteration of functional Principles and Device
set-up…
2. CLEARING
-Clearing is the transition step between dehydration and infiltration with
the embedding medium.
-The term clearing arises because some solvents have a high refractive
index.
-When dehydrated tissues are placed into these reagents, they are
rendered transparent. This property is used to determine the endpoint and
duration of the clearing step since the presence of opaque areas indicates
incomplete dehydration
3. Infiltration and embedding
3.1 Infiltration
-This is the saturation of tissue cavities and cells by a supporting
substance which is generally the medium in which they are finally
embedded.
-The most common agent of choice is paraffin wax which is molten
when hot and solid when cold.
-An infiltrating and embedding medium should ideally be molten
between 30°C and 60°C and suitable for sectioning.
-Additionally, the properties of the medium should be similar to those
of the tissues to be sectioned with regard to density and elasticity.
Infiltration and embedding…
-Additionally, the properties of the medium should be similar to those of
the tissues to be sectioned with regard to density and elasticity
-The use of vacuum infiltration is often used to help complete
impregnation of tissues with wax.
-This is carried using a molten wax or other medium under reduced
pressure. Vacuum assistance helps to not only reduce the time tissues are
subjected to heat but it also assists in the complete removal of any
remaining solvent.
Infiltration and embedding…
3.2. Embedding
-Paraffin embedding is the standard method used in histology laboratories
to produce blocks of tissue for section cutting (microtomy).
-This process is usually carried out using an embedding center and involves
surrounding the tissues by a medium such as paraffin wax which when
cooled and solidified will provide sufficient support for section cutting or
microtomy.
Tissue Processing Cycle
https://slideplayer.com/slide/9288238/
Identification of functional assemblies
and components
-These processors are characterized by the transfer of tissues, contained
within a basket, through a series of stationary reagents arranged in-line or
in a circular carousel plan.
-The rotary or carousel is the most common model of automatic tissue
processor, and is provided with 9-10 reagent and 2-3 wax positions, with a
capacity of 30-110 cassettes depending upon the model.
-Fluid agitation is achieved by vertical oscillation or rotary motion of the
tissue basket. Processing schedules are card-notched, pin or touch pad
programmed
Identification of functional assemblies and
components…
-These processors allow maximum flexibility in the choice of reagents and
schedules that can be run on them.
-These machines have a rapid turn-around time for day or night
processing.
-In more recent models the tissue basket is enclosed within an integrated
fume hood during agitation and transfer cycles thus overcoming the
disadvantages of earlier styles.
Maintenance and Troubleshooting
Warnings and precautions
All the maintenance work should be performed after cut-off the machines
power.
-Everyone should realize that some components of the tissue processor
may be in danger of infection when dismantled and cleaned.
-Do not try to touch the machine while it is processing.
Maintenance and Troubleshooting …
• General precautions:
-Take care when using to avoid hot surface areas as they may cause
burns
-Avoid contact with hot wax as it may cause burns.
-Always validate that a sufficient amount of wax and stock chemicals are
within the stock tank and bottles.
Maintenance and Troubleshooting …
• Safety
• Personal Protective Equipment
(PPE):
-Lab coat or gown
-Fully closed shoes
-Gloves
-Safety glasses as required
https://www.clinicallab.com/proper-use-of-personal-
protective-equipment-in-the-lab-256
Maintenance and Troubleshooting …
• Cleaning preparation
-Lift the carousel in the manual processing mode.
-Remove all station containers from the platform.
-Disconnect the wax bath plugs, slightly lift the wax
bath lids and remove the paraffin containers.
Maintenance and Troubleshooting …
• Validate the wax is fully melted before usage
• Clean all tissues from the machine upon completion of its usage.
• Take care when obtaining new metal rack from within the path centre as
they may contain melted wax, even when in the storage area.
Cleaning …
-Remove residual paraffin from the station holders, lid seals and lid
holders.
-Use a soft plastic spatula for removing the paraffin.
-Use a moist cleaning cloth to clean platform, lacquered instrument
surfaces and control panel.
-Clean the rims of the glass or aluminum reagent containers and of the
wax baths.
Cont…
-The glass containers are dishwasher-safe, never clean the aluminum
containers in an automatic dishwasher.
-Cleaning the Plexiglas fume containment shields of instruments with
fume control system.
-When cleaning the instrument, no liquid may enter in contact with any of
the electrical connections or the interior of the instrument.
Maintenance Schedule
Activity Maintenance Frequency Type of Maintenance
Replacing Mop up reagents Immediately after their expired date Replacing
Replacing Fuse When the old ones are burned. Replacing
Clean the rims of the glass or
aluminum reagent containers and of
the wax baths
Daily Cleaning
Clean the aluminum containers by
hand with mild household
detergents.
Daily Cleaning
Replace activated carbon filters. Every four months. Replacing
clean the
carousel axle with a cleaning cloth
and subsequently apply a thin coat
of machine oil.
Monthly Cleaning and greasing
Parts of Circuit board and their functions
• Horpos Board
-The light barriers located in this component transmit signals to the control
electronics.
-It enable the instrument to localize the horizontal movement of the
containers.
-The code disk is the corresponding sensing device.
Parts of Circuit board and their functions …
Verpos Board
- The light barriers located in this component transmit signals to the
control electronics.
-It enable the instrument to localize the vertical position of the
container at the given moment.
-Loading position or the two vertical end points of the continuous up
and down movement
• Code Disk
-The movement of the code disk is congruous to the container movement
.
-The corresponding position signals are generated on the Horpos
components ,and from there transmitted to the control electronics
Parts of Circuit board and their functions …
• Control Electronics board
-It contains power supply +15v which supply for the electronics, with -5v
for the control panel contrast. With motor electronics which switch the
drive motor in to different modes operation e.g. Lift ,break, stop …
-With the motor voltage and with the distribution voltage for fume
extraction ,magnetic valves and vacuum pumps.
-Sensing the wax bath function, If a connected wax is heated
corresponding signal is transmitted to the control board, and once the
paraffin is molten another signal is transmitted.
-Transmit signals back and forth for verpos and Horpos boards.
Parts of Circuit board and their functions …
Troubleshooting and Approaching Technique
S/No Warning code Couse Solution
1 Warning code
W:01, W:02, W:03
in manual pro-
cessing mode
Paraffin in wax bath stations
10/11/12
still solid.
-To quit the warning code, press any key.
-Check if the paraffin is really molten.
-If you find that the paraffin is molten, pause the
automatic processing cycle and - in the manual
processing mode - by pressing KEY and the
corresponding ARROW button
simultaneously, lower the basket into the wax bath or
lift it out of the wax bath.
2 Warning code
"W:04" "W:05"
"W:06"
Total duration of program from
start until the basket reaches the
first wax bath is less than 8 hours.
Paraffin in stations 10/11/12
possibly still solid upon arrival
of tissue basket.
Check if there is sufficient time for the paraffin to liquefy
completely until the basket arrives at the wax bath.
-If this is not ensured, fill the wax bath with molten
paraffin.
- To override / quit the warning code, press KEY and
START simultaneously.
Cont...
S/No Error Code Root Couse Solution
1 Error code "E:01" RAM battery Defective or Empty Exchange batteries
2 Error code
"E:03"
-A movement is not carried out as scheduled.
-Mechanical stillness/Jammed
-Optosensers Dirty
-Defective Motor
-Horpos Board Defective
-Control electronics board defective
-Clean optosensers
-Change defective motor .
-Exchange rotation sensing device.
-Exchange control electronics board
3 Error code "E:04" -Error in vertical sensor mechanism (Drive)
-Dirty Optosensor
-Verpos bored defective
-Control electronics board defective
-Clean Optosensers
-Exchange Verpos board
-Exchange control electronics board.
4 Error Code “E:05” -Error in rotation sensing mechanism
-Dirty Optosensor
-Horpos board defective
-Verpos bored defective
-Control electronics board defective
-Clean optosensers
-Exchange Horpos boaard
-Exchange Verpos board
-Exchange control electronics board.
Cont…
S/No Error Code Couse Solution
5 Error code "E:7" -Battery in timer component is empty
-Timer component empty or defective
component type :DS12887
- Exchange defective Timer component.
6 Error code "E:08" -Wax bath one not connected or
defective or connected with wrong
socket.
-Excess temperature cut-off activated
-Control electronics board defective
-Connect wax bath 1, exchange if necessary, check
socket.
-Disconnect from mains and let it cool.
-Exchange defective electronic bored.
7 Error code "E:09" -Wax bath number 2 is not connected
or defective.
-Wax number 2 cut-off temperature
activated.
-Control electronics board defective
-Connect wax bath 2 ,exchange if necessary, check
socket.
-Disconnect from mains and let it cool.
-Exchange defective electronic bored.
Cont…
S/No Error Code Couse Solution
8 Error code "E:10" - Wax bath number 3 is not connected
or defective.
-Wax number 3 cut-off temperature
activated.
-Control electronics board defective
-Connect wax bath3,exchange if necessary, check
socket.
-Disconnect from mains and let it cool.
-Exchange defective electronic bored.
9 Error code "E:11" - Processer data lost.
-Battery-buffered RAM defective
-EMV Problem
-Exchange battery
battery type :M48Z30
For further specification contact manufacturer
10 Error code "E:13" - Implausible data entry - Check data entry, alter settings if necessary .
Troubleshooting(Common Problems)
• Power Failure
-It is displayed when power is restored after a power failure, but only if the
processing cycle in progress has been impaired by the power failure.
-In case of a prolonged power failure the tissue can be removed from the
instrument manually and also be re- located to the next station by entirely
manual operation.
-Once the power is restored, processing will be resumed as programmed.
Possible impairment of the processing cycle through the interruption is
displayed.
Cont…
• Exchanging Fuse
-First unplug the equipment from source.
-Pull out locking devices and fuse holder at the rear side of the device
-Exchange fuses
-Make sure the right voltage is selected.
-Make sure the fuse current is compatible.
Thank you!

More Related Content

What's hot

Gross Examination, Selection, Collection and Fixation of Specimen
Gross Examination, Selection, Collection and Fixation of Specimen Gross Examination, Selection, Collection and Fixation of Specimen
Gross Examination, Selection, Collection and Fixation of Specimen
ghulam abbas
 
CYTOLOGY OF CSF
CYTOLOGY OF CSFCYTOLOGY OF CSF
CYTOLOGY OF CSF
Musa Khan
 
Technical faults in histopathology lab
Technical faults in histopathology labTechnical faults in histopathology lab
Technical faults in histopathology lab
magdy abdelghany
 

What's hot (20)

Histopatholgy staining by suchit kumar
Histopatholgy staining by suchit kumarHistopatholgy staining by suchit kumar
Histopatholgy staining by suchit kumar
 
Fixatives
FixativesFixatives
Fixatives
 
2. fixatives and fixation seminar
2. fixatives and fixation seminar2. fixatives and fixation seminar
2. fixatives and fixation seminar
 
DECALCIFICATION
DECALCIFICATIONDECALCIFICATION
DECALCIFICATION
 
Decalcificaion
DecalcificaionDecalcificaion
Decalcificaion
 
Tissue Embedding Procedure
Tissue Embedding ProcedureTissue Embedding Procedure
Tissue Embedding Procedure
 
Histotechniques
HistotechniquesHistotechniques
Histotechniques
 
AUTOMATED STAINING DEVICE.pdf
AUTOMATED STAINING DEVICE.pdfAUTOMATED STAINING DEVICE.pdf
AUTOMATED STAINING DEVICE.pdf
 
Special stains by sowmya
Special stains by sowmyaSpecial stains by sowmya
Special stains by sowmya
 
Cryostat and frozen section
Cryostat and frozen sectionCryostat and frozen section
Cryostat and frozen section
 
Section Cutting
Section CuttingSection Cutting
Section Cutting
 
Gross Examination, Selection, Collection and Fixation of Specimen
Gross Examination, Selection, Collection and Fixation of Specimen Gross Examination, Selection, Collection and Fixation of Specimen
Gross Examination, Selection, Collection and Fixation of Specimen
 
Tissue Processing
Tissue ProcessingTissue Processing
Tissue Processing
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cells
 
Decalcification
Decalcification  Decalcification
Decalcification
 
CYTOLOGY OF CSF
CYTOLOGY OF CSFCYTOLOGY OF CSF
CYTOLOGY OF CSF
 
Clearing
ClearingClearing
Clearing
 
Special stains in cytology
Special stains in cytologySpecial stains in cytology
Special stains in cytology
 
Technical faults in histopathology lab
Technical faults in histopathology labTechnical faults in histopathology lab
Technical faults in histopathology lab
 
Special stains in hematology
Special stains in hematologySpecial stains in hematology
Special stains in hematology
 

Similar to Tissue processing ppt.pptx

Chapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptx
Chapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptxChapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptx
Chapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptx
ErmiyasBeletew
 

Similar to Tissue processing ppt.pptx (20)

18.Histopathology Section.pptx
18.Histopathology Section.pptx18.Histopathology Section.pptx
18.Histopathology Section.pptx
 
Tissue processing by dr manzoor
Tissue processing by dr manzoorTissue processing by dr manzoor
Tissue processing by dr manzoor
 
Automation histopathology
Automation histopathologyAutomation histopathology
Automation histopathology
 
Chapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptx
Chapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptxChapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptx
Chapter III.jjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjjpptx
 
Basic Histopathological Techniques.pptx
Basic Histopathological Techniques.pptxBasic Histopathological Techniques.pptx
Basic Histopathological Techniques.pptx
 
Tissue Processing of Histopathology.pptx
Tissue Processing  of Histopathology.pptxTissue Processing  of Histopathology.pptx
Tissue Processing of Histopathology.pptx
 
Basic Histopathological Techniques (1).pptx
Basic Histopathological Techniques (1).pptxBasic Histopathological Techniques (1).pptx
Basic Histopathological Techniques (1).pptx
 
Tissue processing DMLT II 2023.pptx
Tissue processing DMLT II 2023.pptxTissue processing DMLT II 2023.pptx
Tissue processing DMLT II 2023.pptx
 
Downstream processing
Downstream processingDownstream processing
Downstream processing
 
Cssd (central sterile supply department)
Cssd (central sterile supply department)Cssd (central sterile supply department)
Cssd (central sterile supply department)
 
CSSD PPT.pptx
CSSD PPT.pptxCSSD PPT.pptx
CSSD PPT.pptx
 
62120 tissue processing
62120 tissue processing62120 tissue processing
62120 tissue processing
 
Tissue processing المعدل.ppt
Tissue processing المعدل.pptTissue processing المعدل.ppt
Tissue processing المعدل.ppt
 
HISTOTECHNIQUES MICROTOMY.ppt · version 1.pdf
HISTOTECHNIQUES  MICROTOMY.ppt · version 1.pdfHISTOTECHNIQUES  MICROTOMY.ppt · version 1.pdf
HISTOTECHNIQUES MICROTOMY.ppt · version 1.pdf
 
5 lecture preparation histological specimens.pdf
5  lecture preparation histological specimens.pdf5  lecture preparation histological specimens.pdf
5 lecture preparation histological specimens.pdf
 
Tissue processing
Tissue processingTissue processing
Tissue processing
 
Central sterile supply department(cssd)
Central sterile supply department(cssd)Central sterile supply department(cssd)
Central sterile supply department(cssd)
 
CSSD
CSSDCSSD
CSSD
 
4457652.ppt
4457652.ppt4457652.ppt
4457652.ppt
 
Single-Use Tangential Flow Filtration for Closed Processing
Single-Use Tangential Flow Filtration for Closed ProcessingSingle-Use Tangential Flow Filtration for Closed Processing
Single-Use Tangential Flow Filtration for Closed Processing
 

Recently uploaded

Spauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCESpauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCE
DR.PRINCE C P
 
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
rajveerescorts2022
 
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North CarolinaTIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
Mebane Rash
 
Cash Payment 😋 +9316020077 Goa Call Girl No Advance *Full Service
Cash Payment 😋  +9316020077 Goa Call Girl No Advance *Full ServiceCash Payment 😋  +9316020077 Goa Call Girl No Advance *Full Service
Cash Payment 😋 +9316020077 Goa Call Girl No Advance *Full Service
Real Sex Provide In Goa
 
Goa Call Girls Service +9316020077 Call GirlsGoa By Russian Call Girlsin Goa
Goa Call Girls Service  +9316020077 Call GirlsGoa By Russian Call Girlsin GoaGoa Call Girls Service  +9316020077 Call GirlsGoa By Russian Call Girlsin Goa
Goa Call Girls Service +9316020077 Call GirlsGoa By Russian Call Girlsin Goa
Real Sex Provide In Goa
 
No Advance 931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...
No Advance  931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...No Advance  931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...
No Advance 931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...
Real Sex Provide In Goa
 
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
rajveerescorts2022
 
❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...
❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...
❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...
rajveerescorts2022
 
ITM HOSPITAL The hospital has also been recognised as the best emerging hosp...
ITM  HOSPITAL The hospital has also been recognised as the best emerging hosp...ITM  HOSPITAL The hospital has also been recognised as the best emerging hosp...
ITM HOSPITAL The hospital has also been recognised as the best emerging hosp...
jvomprakash
 

Recently uploaded (20)

clostridiumbotulinum- BY Muzammil Ahmed Siddiqui.pptx
clostridiumbotulinum- BY Muzammil Ahmed Siddiqui.pptxclostridiumbotulinum- BY Muzammil Ahmed Siddiqui.pptx
clostridiumbotulinum- BY Muzammil Ahmed Siddiqui.pptx
 
ISO 15189 2022 standards for laboratory quality and competence
ISO 15189 2022 standards for laboratory quality and competenceISO 15189 2022 standards for laboratory quality and competence
ISO 15189 2022 standards for laboratory quality and competence
 
The Events of Cardiac Cycle - Wigger's Diagram
The Events of Cardiac Cycle - Wigger's DiagramThe Events of Cardiac Cycle - Wigger's Diagram
The Events of Cardiac Cycle - Wigger's Diagram
 
Spauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCESpauldings classification ppt by Dr C P PRINCE
Spauldings classification ppt by Dr C P PRINCE
 
The 2024 Outlook for Older Adults: Healthcare Consumer Survey
The 2024 Outlook for Older Adults: Healthcare Consumer SurveyThe 2024 Outlook for Older Adults: Healthcare Consumer Survey
The 2024 Outlook for Older Adults: Healthcare Consumer Survey
 
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
 
Leading large scale change: a life at the interface between theory and practice
Leading large scale change: a life at the interface between theory and practiceLeading large scale change: a life at the interface between theory and practice
Leading large scale change: a life at the interface between theory and practice
 
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North CarolinaTIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
TIME FOR ACTION: MAY 2024 Securing A Strong Nursing Workforce for North Carolina
 
Post marketing surveillance in Japan, legislation and.pptx
Post marketing surveillance in Japan, legislation and.pptxPost marketing surveillance in Japan, legislation and.pptx
Post marketing surveillance in Japan, legislation and.pptx
 
Cash Payment 😋 +9316020077 Goa Call Girl No Advance *Full Service
Cash Payment 😋  +9316020077 Goa Call Girl No Advance *Full ServiceCash Payment 😋  +9316020077 Goa Call Girl No Advance *Full Service
Cash Payment 😋 +9316020077 Goa Call Girl No Advance *Full Service
 
Goa Call Girls Service +9316020077 Call GirlsGoa By Russian Call Girlsin Goa
Goa Call Girls Service  +9316020077 Call GirlsGoa By Russian Call Girlsin GoaGoa Call Girls Service  +9316020077 Call GirlsGoa By Russian Call Girlsin Goa
Goa Call Girls Service +9316020077 Call GirlsGoa By Russian Call Girlsin Goa
 
TEST BANK For Little and Falace's Dental Management of the Medically Compromi...
TEST BANK For Little and Falace's Dental Management of the Medically Compromi...TEST BANK For Little and Falace's Dental Management of the Medically Compromi...
TEST BANK For Little and Falace's Dental Management of the Medically Compromi...
 
VIP ℂall Girls Madhavaram Chennai 7877925207 WhatsApp: Me All Time Serviℂe Av...
VIP ℂall Girls Madhavaram Chennai 7877925207 WhatsApp: Me All Time Serviℂe Av...VIP ℂall Girls Madhavaram Chennai 7877925207 WhatsApp: Me All Time Serviℂe Av...
VIP ℂall Girls Madhavaram Chennai 7877925207 WhatsApp: Me All Time Serviℂe Av...
 
ACNE VULGARIS , ALLERGIES, ECZEMA, PEMPHIGUS.pdf
ACNE VULGARIS , ALLERGIES, ECZEMA, PEMPHIGUS.pdfACNE VULGARIS , ALLERGIES, ECZEMA, PEMPHIGUS.pdf
ACNE VULGARIS , ALLERGIES, ECZEMA, PEMPHIGUS.pdf
 
VIP ℂall Girls Bodakdev Ahmedabad 7427069034 WhatsApp: Me All Time Serviℂe Av...
VIP ℂall Girls Bodakdev Ahmedabad 7427069034 WhatsApp: Me All Time Serviℂe Av...VIP ℂall Girls Bodakdev Ahmedabad 7427069034 WhatsApp: Me All Time Serviℂe Av...
VIP ℂall Girls Bodakdev Ahmedabad 7427069034 WhatsApp: Me All Time Serviℂe Av...
 
No Advance 931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...
No Advance  931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...No Advance  931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...
No Advance 931~602~0077 Goa ✂️ Call Girl , Indian Call Girl Goa For Full nig...
 
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
❤️ Chandigarh Call Girls Service ☎️99158-51334☎️ Escort service in Chandigarh...
 
VIP Just Call 9548273370 Lucknow Top Class Call Girls Number | 8630512678 Esc...
VIP Just Call 9548273370 Lucknow Top Class Call Girls Number | 8630512678 Esc...VIP Just Call 9548273370 Lucknow Top Class Call Girls Number | 8630512678 Esc...
VIP Just Call 9548273370 Lucknow Top Class Call Girls Number | 8630512678 Esc...
 
❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...
❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...
❤️ Chandigarh Call Girls ☎️99158-51334☎️ Escort service in Chandigarh ☎️ Chan...
 
ITM HOSPITAL The hospital has also been recognised as the best emerging hosp...
ITM  HOSPITAL The hospital has also been recognised as the best emerging hosp...ITM  HOSPITAL The hospital has also been recognised as the best emerging hosp...
ITM HOSPITAL The hospital has also been recognised as the best emerging hosp...
 

Tissue processing ppt.pptx

  • 1. Medical Device presentation on Course Title:-Tissue processing Machine Model:- Leica TP1020 Venue:-Hawassa University Comprehensive Specialized Hospital Hawasaa,Ethiopia Date:-2/22/2023
  • 2. Introduction To Tissue Processor • Purpose of Device -A device intended for use in the histology laboratory to process the tissues taken from the body for diagnosis of disease processes and to produce microscopic slides that can be viewed under the microscope. -A tissue processor is used to prepare tissue samples for analysis by fixing, staining, dehydrating or decalcifying them. -The processors are mostly single unit devices that can accommodate a variety of processing techniques to suit the different needs of the laboratory, therefore improving the efficiency of tissue processing
  • 4. Rear side of the device
  • 5. Introduction… • Device’s fundamental functional principle -The basic principle of tissue processor is to transfer the tissue in different fluid for a specified time in a desired environment. So far, there are two types of tissue processor. 1.Tissue transfer processor • Here, the bucket of tissue is transferred from a carousel to other after a specified time. There are also several containers of reagents.
  • 6. Cont… • The tissue remains in a basket which contains 30-100 cassettes. • The basket containing the tissue is then submerged in the specific containers for a particular time which is automatically transferred to the next container. • A gentle agitation is created by rotatory movement or vertical oscillation of the tissue in the basket which causes a continuous removal of fluid from the surface by fresh medium.
  • 7. Cont... • A microprocessor determines the time schedule and transfer of tissue in each container. • A gentle agitation is created by vertical oscillation or by rotatory movement of the tissue basket.
  • 8. Cont... 2. Fluid transfer processor -This is a closed system processor where tissues are kept in the container. -The container is periodically filled with particular fluid. -The fluid is pumped out from the container containing the tissue after a certain period. -The container is again filled with the fluid required for the next step. The fluid transfer processor has some advantages such as the vacuum pressure that makes the system faster and efficient and there is no chance for tissue drying.
  • 9. Device’s fundamental functional principle • Tissue processing -It is concerned with the diffusion of various substances into and out of porous tissues. -Diffusion results from the tendency of processing reagents to equalize concentrations both inside and outside blocks of tissue. -The reagent molecules diffuse down a concentration gradient and move from where they are at a high concentration to where they are at a lower concentration.
  • 10. Cont… -This movement requires no energy since it always progresses down the concentration gradient.
  • 11. Cont… • The rate of diffusion is measured by Fick’s law which states that
  • 12. Introduction… • General Set-up and Requirements -Activating the vacuum function(instruments with the vacuum function only) - Press ARROW RIGHT to move to the next entry position, the cursor blinks at the two positions next to the ‘V’ for vacuum function - To activate the vacuum, press PLUS
  • 13. Cont… -To deactivate the vacuum press MINUS -Selecting infiltration times per station -Press ARROW RIGHT to move to the next entry position ,the cursor blinks at the entry digits -
  • 14. Cont… -Press PLUS or MINUS respectively to enter the hours. - To move the cursor to the next digit for data entry, push the ARROW RIGHT key, the cursor flashes on both digits destined for entering the minutes - Press the PLUS or MINUS to enter the minutes.
  • 15. Cont… -Displaying total program duration. -Once a program is set up, total run time can be displayed. - Press CLOCK to display the total run time of the program.
  • 16. Cont… -The total duration displayed here is 1 day , 0hours and 12 minutes. -To quit the total run time indication , press CLOCK. -To end programming you have to leave the programming mode. -Press PROG to quit programming.
  • 17. Servicing Technician related knowledge Iteration of functional Principles and Device set-up - The principle of tissue processing is designed to remove all extractable water from the tissue, replacing it with a support medium that provides sufficient rigidity to enable sectioning of the tissue without damage or distortion. -The principle is based on the use of chemicals dehydrate and clear the tissue so that it can be embedded into molten paraffin wax which stands as a support medium for the tissue in order to easy sectioning
  • 18. Iteration of functional Principles and Device set-up… DEHYDRATION - The first stage in tissue processing is dehydration (the removal of water). In tissues, water is present in both free and bound forms and needs to be removed before processing can continue. -Dehydration is usually carried out using alcohols (such as ethanol) but these can dissolve certain cellular components such as lipids. -Although dehydration can also cause tissue shrinkage, the stage is necessary in all infiltration methods, except where tissues are supported by an aqueous embedding medium (such as water-soluble waxes).
  • 19. Iteration of functional Principles and Device set-up… 2. CLEARING -Clearing is the transition step between dehydration and infiltration with the embedding medium. -The term clearing arises because some solvents have a high refractive index. -When dehydrated tissues are placed into these reagents, they are rendered transparent. This property is used to determine the endpoint and duration of the clearing step since the presence of opaque areas indicates incomplete dehydration
  • 20. 3. Infiltration and embedding 3.1 Infiltration -This is the saturation of tissue cavities and cells by a supporting substance which is generally the medium in which they are finally embedded. -The most common agent of choice is paraffin wax which is molten when hot and solid when cold. -An infiltrating and embedding medium should ideally be molten between 30°C and 60°C and suitable for sectioning. -Additionally, the properties of the medium should be similar to those of the tissues to be sectioned with regard to density and elasticity.
  • 21. Infiltration and embedding… -Additionally, the properties of the medium should be similar to those of the tissues to be sectioned with regard to density and elasticity -The use of vacuum infiltration is often used to help complete impregnation of tissues with wax. -This is carried using a molten wax or other medium under reduced pressure. Vacuum assistance helps to not only reduce the time tissues are subjected to heat but it also assists in the complete removal of any remaining solvent.
  • 22. Infiltration and embedding… 3.2. Embedding -Paraffin embedding is the standard method used in histology laboratories to produce blocks of tissue for section cutting (microtomy). -This process is usually carried out using an embedding center and involves surrounding the tissues by a medium such as paraffin wax which when cooled and solidified will provide sufficient support for section cutting or microtomy.
  • 24. Identification of functional assemblies and components -These processors are characterized by the transfer of tissues, contained within a basket, through a series of stationary reagents arranged in-line or in a circular carousel plan. -The rotary or carousel is the most common model of automatic tissue processor, and is provided with 9-10 reagent and 2-3 wax positions, with a capacity of 30-110 cassettes depending upon the model. -Fluid agitation is achieved by vertical oscillation or rotary motion of the tissue basket. Processing schedules are card-notched, pin or touch pad programmed
  • 25. Identification of functional assemblies and components… -These processors allow maximum flexibility in the choice of reagents and schedules that can be run on them. -These machines have a rapid turn-around time for day or night processing. -In more recent models the tissue basket is enclosed within an integrated fume hood during agitation and transfer cycles thus overcoming the disadvantages of earlier styles.
  • 26. Maintenance and Troubleshooting Warnings and precautions All the maintenance work should be performed after cut-off the machines power. -Everyone should realize that some components of the tissue processor may be in danger of infection when dismantled and cleaned. -Do not try to touch the machine while it is processing.
  • 27. Maintenance and Troubleshooting … • General precautions: -Take care when using to avoid hot surface areas as they may cause burns -Avoid contact with hot wax as it may cause burns. -Always validate that a sufficient amount of wax and stock chemicals are within the stock tank and bottles.
  • 28. Maintenance and Troubleshooting … • Safety • Personal Protective Equipment (PPE): -Lab coat or gown -Fully closed shoes -Gloves -Safety glasses as required https://www.clinicallab.com/proper-use-of-personal- protective-equipment-in-the-lab-256
  • 29. Maintenance and Troubleshooting … • Cleaning preparation -Lift the carousel in the manual processing mode. -Remove all station containers from the platform. -Disconnect the wax bath plugs, slightly lift the wax bath lids and remove the paraffin containers.
  • 30. Maintenance and Troubleshooting … • Validate the wax is fully melted before usage • Clean all tissues from the machine upon completion of its usage. • Take care when obtaining new metal rack from within the path centre as they may contain melted wax, even when in the storage area.
  • 31. Cleaning … -Remove residual paraffin from the station holders, lid seals and lid holders. -Use a soft plastic spatula for removing the paraffin. -Use a moist cleaning cloth to clean platform, lacquered instrument surfaces and control panel. -Clean the rims of the glass or aluminum reagent containers and of the wax baths.
  • 32. Cont… -The glass containers are dishwasher-safe, never clean the aluminum containers in an automatic dishwasher. -Cleaning the Plexiglas fume containment shields of instruments with fume control system. -When cleaning the instrument, no liquid may enter in contact with any of the electrical connections or the interior of the instrument.
  • 33. Maintenance Schedule Activity Maintenance Frequency Type of Maintenance Replacing Mop up reagents Immediately after their expired date Replacing Replacing Fuse When the old ones are burned. Replacing Clean the rims of the glass or aluminum reagent containers and of the wax baths Daily Cleaning Clean the aluminum containers by hand with mild household detergents. Daily Cleaning Replace activated carbon filters. Every four months. Replacing clean the carousel axle with a cleaning cloth and subsequently apply a thin coat of machine oil. Monthly Cleaning and greasing
  • 34. Parts of Circuit board and their functions • Horpos Board -The light barriers located in this component transmit signals to the control electronics. -It enable the instrument to localize the horizontal movement of the containers. -The code disk is the corresponding sensing device.
  • 35. Parts of Circuit board and their functions … Verpos Board - The light barriers located in this component transmit signals to the control electronics. -It enable the instrument to localize the vertical position of the container at the given moment. -Loading position or the two vertical end points of the continuous up and down movement
  • 36. • Code Disk -The movement of the code disk is congruous to the container movement . -The corresponding position signals are generated on the Horpos components ,and from there transmitted to the control electronics Parts of Circuit board and their functions …
  • 37. • Control Electronics board -It contains power supply +15v which supply for the electronics, with -5v for the control panel contrast. With motor electronics which switch the drive motor in to different modes operation e.g. Lift ,break, stop … -With the motor voltage and with the distribution voltage for fume extraction ,magnetic valves and vacuum pumps. -Sensing the wax bath function, If a connected wax is heated corresponding signal is transmitted to the control board, and once the paraffin is molten another signal is transmitted. -Transmit signals back and forth for verpos and Horpos boards. Parts of Circuit board and their functions …
  • 38. Troubleshooting and Approaching Technique S/No Warning code Couse Solution 1 Warning code W:01, W:02, W:03 in manual pro- cessing mode Paraffin in wax bath stations 10/11/12 still solid. -To quit the warning code, press any key. -Check if the paraffin is really molten. -If you find that the paraffin is molten, pause the automatic processing cycle and - in the manual processing mode - by pressing KEY and the corresponding ARROW button simultaneously, lower the basket into the wax bath or lift it out of the wax bath. 2 Warning code "W:04" "W:05" "W:06" Total duration of program from start until the basket reaches the first wax bath is less than 8 hours. Paraffin in stations 10/11/12 possibly still solid upon arrival of tissue basket. Check if there is sufficient time for the paraffin to liquefy completely until the basket arrives at the wax bath. -If this is not ensured, fill the wax bath with molten paraffin. - To override / quit the warning code, press KEY and START simultaneously.
  • 39. Cont... S/No Error Code Root Couse Solution 1 Error code "E:01" RAM battery Defective or Empty Exchange batteries 2 Error code "E:03" -A movement is not carried out as scheduled. -Mechanical stillness/Jammed -Optosensers Dirty -Defective Motor -Horpos Board Defective -Control electronics board defective -Clean optosensers -Change defective motor . -Exchange rotation sensing device. -Exchange control electronics board 3 Error code "E:04" -Error in vertical sensor mechanism (Drive) -Dirty Optosensor -Verpos bored defective -Control electronics board defective -Clean Optosensers -Exchange Verpos board -Exchange control electronics board. 4 Error Code “E:05” -Error in rotation sensing mechanism -Dirty Optosensor -Horpos board defective -Verpos bored defective -Control electronics board defective -Clean optosensers -Exchange Horpos boaard -Exchange Verpos board -Exchange control electronics board.
  • 40. Cont… S/No Error Code Couse Solution 5 Error code "E:7" -Battery in timer component is empty -Timer component empty or defective component type :DS12887 - Exchange defective Timer component. 6 Error code "E:08" -Wax bath one not connected or defective or connected with wrong socket. -Excess temperature cut-off activated -Control electronics board defective -Connect wax bath 1, exchange if necessary, check socket. -Disconnect from mains and let it cool. -Exchange defective electronic bored. 7 Error code "E:09" -Wax bath number 2 is not connected or defective. -Wax number 2 cut-off temperature activated. -Control electronics board defective -Connect wax bath 2 ,exchange if necessary, check socket. -Disconnect from mains and let it cool. -Exchange defective electronic bored.
  • 41. Cont… S/No Error Code Couse Solution 8 Error code "E:10" - Wax bath number 3 is not connected or defective. -Wax number 3 cut-off temperature activated. -Control electronics board defective -Connect wax bath3,exchange if necessary, check socket. -Disconnect from mains and let it cool. -Exchange defective electronic bored. 9 Error code "E:11" - Processer data lost. -Battery-buffered RAM defective -EMV Problem -Exchange battery battery type :M48Z30 For further specification contact manufacturer 10 Error code "E:13" - Implausible data entry - Check data entry, alter settings if necessary .
  • 42. Troubleshooting(Common Problems) • Power Failure -It is displayed when power is restored after a power failure, but only if the processing cycle in progress has been impaired by the power failure. -In case of a prolonged power failure the tissue can be removed from the instrument manually and also be re- located to the next station by entirely manual operation. -Once the power is restored, processing will be resumed as programmed. Possible impairment of the processing cycle through the interruption is displayed.
  • 43. Cont… • Exchanging Fuse -First unplug the equipment from source. -Pull out locking devices and fuse holder at the rear side of the device -Exchange fuses -Make sure the right voltage is selected. -Make sure the fuse current is compatible.