SlideShare a Scribd company logo
1 of 108
Mechanical Ventilation
Dr. Abdul-Monim Batiha
Assistant Professor
Critical Care Nursing
• Mechanical Ventilation is ventilation
of the lungs by artificial means
usually by a ventilator.
• A ventilator delivers gas to the lungs
with either negative or positive
pressure.
Purposes:
• To maintain or improve ventilation, &
tissue oxygenation.
• To decrease the work of breathing &
improve patient’s comfort.
Indications:
1- Acute respiratory failure due to:
• Mechanical failure, includes neuromuscular
diseases as Myasthenia Gravis, Guillain-Barré
Syndrome, and Poliomyelitis (failure of the
normal respiratory neuromuscular system)
• Musculoskeletal abnormalities, such as chest wall
trauma (flail chest)
• Infectious diseases of the lung such as pneumonia,
tuberculosis.
2- Abnormalities of pulmonary gas exchange
as in:
• Obstructive lung disease in the form of
asthma, chronic bronchitis or emphysema.
• Conditions such as pulmonary edema,
atelectasis, pulmonary fibrosis.
• Patients who has received general
anesthesia as well as post cardiac arrest
patients often require ventilatory support
until they have recovered from the effects
of the anesthesia or the insult of an arrest.
Criteria for institution of
ventilatory support:
Normal
range
Ventilation
indicated
Parameters
10-20
5-7
65-75
75-100
> 35
< 5
< 15
<-20
A- Pulmonary function
studies:
• Respiratory rate
(breaths/min).
• Tidal volume (ml/kg
body wt)
• Vital capacity (ml/kg
body wt)
• Maximum Inspiratory
Force (cm HO2)
Criteria for institution of
ventilatory support:
Normal
range
Ventilation
indicated
Parameters
7.35-7.45
75-100
35-45
< 7.25
< 60
> 50
B- Arterial blood
Gases
• PH
• PaO2 (mmHg)
• PaCO2 (mmHg)
Types of Mechanical
ventilators:
• Negative-pressure ventilators
• Positive-pressure ventilators.
Negative-Pressure Ventilators
• Early negative-pressure ventilators
were known as “iron lungs.”
• The patient’s body was encased in an
iron cylinder and negative pressure
was generated .
• The iron lung are still occasionally
used today.
• Intermittent short-term negative-pressure
ventilation is sometimes used in patients
with chronic diseases.
• The use of negative-pressure ventilators is
restricted in clinical practice, however,
because they limit positioning and
movement and they lack adaptability to
large or small body torsos (chests) .
• Our focus will be on the positive-pressure
ventilators.
Positive-pressure ventilators
• Positive-pressure ventilators deliver
gas to the patient under positive-
pressure, during the inspiratory
phase.
Types of Positive-Pressure
Ventilators
1- Volume Ventilators.
2- Pressure Ventilators
3- High-Frequency Ventilators
• The volume ventilator is commonly used in
critical care settings.
• The basic principle of this ventilator is that
a designated volume of air is delivered
with each breath.
• The amount of pressure required to deliver
the set volume depends on :-
- Patient’s lung compliance
- Patient–ventilator resistance factors.
1- Volume Ventilators
• Therefore, peak inspiratory pressure
(PIP ) must be monitored in volume
modes because it varies from breath
to breath.
• With this mode of ventilation, a
respiratory rate, inspiratory time,
and tidal volume are selected for the
mechanical breaths.
• The use of pressure ventilators is
increasing in critical care units.
• A typical pressure mode delivers a
selected gas pressure to the patient early
in inspiration, and sustains the pressure
throughout the inspiratory phase.
• By meeting the patient’s inspiratory flow
demand throughout inspiration, patient
effort is reduced and comfort increased.
2- Pressure Ventilators
• Although pressure is consistent with these
modes, volume is not.
• Volume will change with changes in
resistance or compliance,
• Therefore, exhaled tidal volume is the
variable to monitor closely.
• With pressure modes, the pressure level to
be delivered is selected, and with some
mode options (i.e., pressure controlled
[PC], described later), rate and inspiratory
time are preset as well.
3- High-Frequency Ventilators
• High-frequency ventilators use small
tidal volumes (1 to 3 mL/kg) at
frequencies greater than 100
breaths/minute.
• The high-frequency ventilator
accomplishes oxygenation by the
diffusion of oxygen and carbon
dioxide from high to low gradients of
concentration.
• This diffusion movement is increased
if the kinetic energy of the gas
molecules is increased.
• A high-frequency ventilator would be
used to achieve lower peak ventilator
pressures, thereby lowering the risk
of barotrauma.
Classification of positive-pressure
ventilators:
• Ventilators are classified according to how
the inspiratory phase ends. The factor
which terminates the inspiratory cycle
reflects the machine type.
• They are classified as:
1- Pressure cycled ventilator
2- Volume cycled ventilator
3- Time cycled ventilator
1- Volume-cycled ventilator
• Inspiration is terminated after a
preset tidal volume has been
delivered by the ventilator.
• The ventilator delivers a preset tidal
volume (VT), and inspiration stops
when the preset tidal volume is
achieved.
2- Pressure-cycled ventilator
• In which inspiration is terminated
when a specific airway pressure has
been reached.
• The ventilator delivers a preset
pressure; once this pressure is
achieved, end inspiration occurs.
3- Time-cycled ventilator
• In which inspiration is terminated
when a preset inspiratory time, has
elapsed.
• Time cycled machines are not used in
adult critical care settings. They are
used in pediatric intensive care
areas.
Ventilator mode
• The way the machine ventilates the
patient
• How much the patient will
participate in his own ventilatory
pattern.
• Each mode is different in determining
how much work of breathing the
patient has to do.
Modes of Mechanical Ventilation
A- Volume Modes
B- Pressure Modes
A- Volume Modes
1- Assist-control (A/C)
2- Synchronized intermittent
mandatory ventilation (SIMV)
1- Assist Control Mode A/C
• The ventilator provides the patient with a
pre-set tidal volume at a pre-set rate .
• The patient may initiate a breath on his
own, but the ventilator assists by delivering
a specified tidal volume to the patient.
Client can initiate breaths that are
delivered at the preset tidal volume.
• Client can breathe at a higher rate than the
preset number of breaths/minute
• The total respiratory rate is determined by
the number of spontaneous inspiration
initiated by the patient plus the number of
breaths set on the ventilator.
• In A/C mode, a mandatory (or “control”)
rate is selected.
• If the patient wishes to breathe faster, he
or she can trigger the ventilator and
receive a full-volume breath.
• Often used as initial mode of
ventilation
• When the patient is too weak to
perform the work of breathing (e.g.,
when emerging from anesthesia).
Disadvantages:
• Hyperventilation,
2- Synchronized Intermittent
Mandatory Ventilation (SIMV)
• The ventilator provides the patient with a pre-set
number of breaths/minute at a specified tidal
volume and FiO2.
• In between the ventilator-delivered breaths, the
patient is able to breathe spontaneously at his
own tidal volume and rate with no assistance
from the ventilator.
• However, unlike the A/C mode, any breaths taken
above the set rate are spontaneous breaths taken
through the ventilator circuit.
• The tidal volume of these breaths can vary
drastically from the tidal volume set on the
ventilator, because the tidal volume is
determined by the patient’s spontaneous
effort.
• Adding pressure support during
spontaneous breaths can minimize the risk
of increased work of breathing.
• Ventilators breaths are synchronized with
the patient spontaneous breathe.
( no fighting)
• Used to wean the patient from the
mechanical ventilator.
• Weaning is accomplished by
gradually lowering the set rate and
allowing the patient to assume more
work
B- Pressure Modes
1- Pressure-controlled ventilation (PCV)
2- Pressure-support ventilation (PSV)
3- Continuous positive airway pressure
(CPAP)
4- Positive end expiratory pressure (PEEP)
5- Noninvasive bilevel positive airway
pressure ventilation (BiPAP)
1- Control Mode (CM)
Continuous Mandatory Ventilation
( CMV)
• Ventilation is completely provided by the
mechanical ventilator with a preset tidal volume,
respiratory rate and oxygen concentration
• Ventilator totally controls the patient’s ventilation
i.e. the ventilator initiates and controls both the
volume delivered and the frequency of breath.
• Client does not breathe spontaneously.
• Client can not initiate breathe
2- Pressure-Controlled
Ventilation Mode ( PCV)
• The PCV mode is used
– If compliance is decreased and the risk of
barotrauma is high.
– It is used when the patient has persistent
oxygenation problems despite a high FiO2 and
high levels of PEEP.
• The inspiratory pressure level, respiratory rate,
and inspiratory–expiratory (I:E) ratio must be
selected.
2- Pressure-Controlled
Ventilation Mode ( PCV)
• In pressure controlled ventilation the
breathing gas flows under constant
pressure into the lungs during the selected
inspiratory time.
• The flow is highest at the beginning of
inspiration( i.e when the volume is lowest in
the lungs).
• As the pressure is constant the flow is
initially high and then decreases with
increasing filling of the lungs.
• Like volume controlled ventilation PCV is
time controlled.
Advantages of pressure
limitations are:
• 1- reduction of peak pressure and
therefore the risk of barotruma and
tracheal injury.
• 2- effective ventilation.
• Improve gas exchange
• Tidal volume varies with compliance and
airway resistance and must be closely
monitored.
• Sedation and the use of neuromuscular
blocking agents are frequently indicated,
because any patient–ventilator asynchrony
usually results in profound drops in the
SaO2.
• This is especially true when inverse ratios
are used. The “unnatural” feeling of this
mode often requires muscle relaxants to
ensure patient–ventilator synchrony.
• Inverse ratio ventilation (IRV) mode
reverses this ratio so that inspiratory time
is equal to, or longer than, expiratory time
(1:1 to 4:1).
• Inverse I:E ratios are used in conjunction
with pressure control to improve
oxygenation by expanding stiff alveoli by
using longer distending times, thereby
providing more opportunity for gas
exchange and preventing alveolar
collapse.
• As expiratory time is decreased, one must
monitor for the development of
hyperinflation or auto-PEEP. Regional
alveolar overdistension and
barotrauma may occur owing to excessive
total PEEP.
• When the PCV mode is used, the mean
airway and intrathoracic pressures rise,
potentially resulting in a decrease in
cardiac output and oxygen delivery.
Therefore, the patient’s hemodynamic
status must be monitored closely.
• Used to limit plateau pressures that can
cause barotrauma & Severe ARDS
3- Pressure Support Ventilation
( PSV)
• The patient breathes spontaneously while
the ventilator applies a pre-determined
amount of positive pressure to the airways
upon inspiration.
• Pressure support ventilation augments
patient’s spontaneous breaths with
positive pressure boost during inspiration
i.e. assisting each spontaneous inspiration.
• Helps to overcome airway resistance and
reducing the work of breathing.
• Indicated for patients with small
spontaneous tidal volume and
difficult to wean patients.
• Patient must initiate all pressure
support breaths.
• Pressure support ventilation may be
combined with other modes such as
SIMV or used alone for a
spontaneously breathing patient.
• The patient’s effort determines the
rate, inspiratory flow, and tidal
volume.
• In PSV mode, the inspired tidal
volume and respiratory rate must be
monitored closely to detect changes
in lung compliance.
• It is a mode used primarily for
weaning from mechanical ventilation.
• Constant positive airway pressure during
spontaneous breathing
• CPAP allows the nurse to observe the
ability of the patient to breathe
spontaneously while still on the ventilator.
• CPAP can be used for intubated and
nonintubated patients.
• It may be used as a weaning mode and for
nocturnal ventilation (nasal or mask CPAP)
4- Continuous Positive Airway
Pressure (CPAP)
5- Positive end expiratory
pressure (PEEP)
• Positive pressure applied at the end
of expiration during mandatory 
ventilator breath
• positive end-expiratory pressure with
positive-pressure (machine) breaths.
Uses of CPAP & PEEP
• Prevent atelactasis or collapse of alveoli
• Treat atelactasis or collapse of alveoli
• Improve gas exchange & oxygenation
• Treat hypoxemia refractory to oxygen
therapy.(prevent oxygen toxicity
• Treat pulmonary edema ( pressure help
expulsion of fluids from alveoli
6- Noninvasive Bilateral Positive
Airway Pressure Ventilation (BiPAP)
• BiPAP is a noninvasive form of mechanical
ventilation provided by means of a nasal
mask or nasal prongs, or a full-face mask.
• The system allows the clinician to select
two levels of positive-pressure support:
• An inspiratory pressure support level
(referred to as IPAP)
• An expiratory pressure called EPAP
(PEEP/CPAP level).
Common Ventilator Settings
parameters/ controls
• Fraction of inspired oxygen (FIO2)
• Tidal Volume (VT)
• Peak Flow/ Flow Rate
• Respiratory Rate/ Breath Rate /
Frequency ( F)
• Minute Volume (VE)
• I:E Ratio (Inspiration to Expiration
Ratio)
• Sigh
● Fraction of inspired oxygen
(FIO2)
• The percent of oxygen concentration that
the patient is receiving from the
ventilator. (Between 21% & 100%)
(room air has 21% oxygen content).
• Initially a patient is placed on a high level
of FIO2 (60% or higher).
• Subsequent changes in FIO2 are based on
ABGs and the SaO2.
• In adult patients the initial FiO2 may be set at
100% until arterial blood gases can document
adequate oxygenation.
• An FiO2 of 100% for an extended period of time
can be dangerous ( oxygen toxicity) but it can
protect against hypoxemia
• For infants, and especially in premature infants,
high levels of FiO2 (>60%) should be avoided.
• Usually the FIO2 is adjusted to maintain an SaO2
of greater than 90% (roughly equivalent to a
PaO2 >60 mm Hg).
• Oxygen toxicity is a concern when an FIO2 of
greater than 60% is required for more than 25
hours
Signs and symptoms of oxygen toxicity :-
1- Flushed face
2- Dry cough
3- Dyspnea
4- Chest pain
5- Tightness of chest
6- Sore throat
● Tidal Volume (VT)
• The volume of air delivered to a patient
during a ventilator breath.
• The amount of air inspired and expired
with each breath.
• Usual volume selected is between 5 to 15
ml/ kg body weight)
• In the volume ventilator, Tidal volumes of
10 to 15 mL/kg of body weight were
traditionally used.
• the large tidal volumes may lead to
(volutrauma) aggravate the damage
inflicted on the lungs
• For this reason, lower tidal volume targets
(6 to 8 mL/kg) are now recommended.
● Peak Flow/ Flow Rate
• The speed of delivering air per unit of
time, and is expressed in liters per
minute.
• The higher the flow rate, the faster
peak airway pressure is reached and
the shorter the inspiration;
• The lower the flow rate, the longer
the inspiration.
● Respiratory Rate/ Breath
Rate / Frequency ( F)
• The number of breaths the ventilator will
deliver/minute (10-16 b/m).
• Total respiratory rate equals patient rate
plus ventilator rate.
• The nurse double-checks the functioning
of the ventilator by observing the patient’s
respiratory rate.
For adult patients and older children:-
With COPD
• A reduced tidal volume
• A reduced respiratory rate
For infants and younger children:-
• A small tidal volume
• Higher respiratory rate
● Minute Volume (VE)
• The volume of expired air in one
minute .
• Respiratory rate times tidal volume
equals minute ventilation
VE = (VT x F)
• In special cases, hypoventilation or
hyperventilation is desired
In a patient with head injury,
• Respiratory alkalosis may be required to
promote cerebral vasoconstriction, with a
resultant decrease in ICP.
• In this case, the tidal volume and
respiratory rate are increased
( hyperventilation) to achieve the desired
alkalotic pH by manipulating the PaCO2.
In a patient with COPD
• Baseline ABGs reflect an elevated PaCO2
should not hyperventilated. Instead, the
goal should be restoration of the baseline
PaCO2.
• These patients usually have a large
carbonic acid load, and lowering their
carbon dioxide levels rapidly may result in
seizures.
● I:E Ratio (Inspiration to
Expiration Ratio):-
• The ratio of inspiratory time to
expiratory time during a breath
(Usually = 1:2)
● Sigh
• A deep breath.
• A breath that has a greater volume than the tidal
volume.
• It provides hyperinflation and prevents
atelectasis.
• Sigh volume :------------------Usual volume is 1.5 –
2 times tidal volume.
• Sigh rate/ frequency :---------Usual rate is 4 to 8
times an hour.
● Peak Airway Pressure:-
• In adults if the peak airway pressure
is persistently above 45 cmH2O, the
risk of barotrauma is increased and
efforts should be made to try to
reduce the peak airway pressure.
• In infants and children it is unclear
what level of peak pressure may
cause damage. In general, keeping
peak pressures below 30 is desirable.
● Pressure Limit
• On volume-cycled ventilators, the pressure
limit dial limits the highest pressure
allowed in the ventilator circuit.
• Once the high pressure limit is reached,
inspiration is terminated.
• Therefore, if the pressure limit is being
constantly reached, the designated tidal
volume is not being delivered to the
patient.
● Sensitivity(trigger Sensitivity)
• The sensitivity function controls the
amount of patient effort needed to initiate
an inspiration
• Increasing the sensitivity (requiring less
negative force) decreases the amount of
work the patient must do to initiate a
ventilator breath.
• Decreasing the sensitivity increases the
amount of negative pressure that the
patient needs to initiate inspiration and
increases the work of breathing.
• The most common setting for pressure
sensitivity are -1 to -2 cm H2O
• The more negative the number the harder
it to breath.
Ensuring humidification and
thermoregulation
• All air delivered by the ventilator passes through
the water in the humidifier, where it is warmed
and saturated.
• Humidifier temperatures should be kept close to
body temperature 35 ºC- 37ºC.
• In some rare instances (severe hypothermia), the
air temperatures can be increased.
• The humidifier should be checked for adequate
water levels
• An empty humidifier contributes to drying
the airway, often with resultant dried
secretions, mucus plugging and less ability
to suction out secretions.
• Humidifier should not be overfilled as this
may increase circuit resistance and
interfere with spontaneous breathing.
• As air passes through the ventilator to the
patient, water condenses in the
corrugated tubing. This moisture is
considered contaminated and must be
drained into a receptacle and not back into
the sterile humidifier.
• If the water is allowed to build up,
resistance is developed in the circuit and
PEEP is generated. In addition, if moisture
accumulates near the endotracheal tube,
the patient can aspirate the water.
• The nurse and respiratory therapist jointly
are responsible for preventing this
condensation buildup. The humidifier is an
ideal medium for bacterial growth.
Ventilator alarms:-
• Mechanical ventilators comprise audible
and visual alarm systems, which act as
immediate warning signals to altered
ventilation.
• Alarm systems can be categorized
according to volume and pressure (high
and low).
• High-pressure alarms warn of rising
pressures.
• Low-pressure alarms warn of
disconnection of the patient from the
ventilator or circuit leaks.
Complications
of Mechanical Ventilation:-
I- Airway Complications,
II- Mechanical complications,
III- Physiological Complications,
IV- Artificial Airway Complications.
I- Airway Complications
1- Aspiration
2- Decreased clearance of secretions
3- Nosocomial or ventilator-acquired
pneumonia
II- Mechanical complications
1- Hypoventilation with atelectasis with respiratory
acidosis or hypoxemia.
2- Hyperventilation with hypocapnia and respiratory
alkalosis
3- Barotrauma
a- Closed pneumothorax,
b- Tension pneumothorax,
c- Pneumomediastinum,
d- Subcutaneous emphysema.
4- Alarm “turned off”
5- Failure of alarms or ventilator
6- Inadequate nebulization or humidification
7- Overheated inspired air, resulting in hyperthermia
III- Physiological Complications
1- Fluid overload with humidified air and
sodium chloride (NaCl) retention
2- Depressed cardiac function and
hypotension
3- Stress ulcers
4- Paralytic ileus
5- Gastric distension
6- Starvation
7- Dyssynchronous breathing pattern
IV- Artificial Airway Complications
A- Complications related to
Endotracheal Tube:-
1- Tube kinked or plugged
2- Rupture of piriform sinus
3- Tracheal stenosis or tracheomalacia
4- Mainstem intubation with contralateral
(located on or affecting the opposite side of the
•
lung) lung atelectasis
5- Cuff failure
6- Sinusitis
7- Otitis media
8- Laryngeal edema
B- Complications related to
Tracheostomy tube:-
1- Acute hemorrhage at the site
2- Air embolism
3- Aspiration
4- Tracheal stenosis
5- Erosion into the innominate artery with
exsanguination
6- Failure of the tracheostomy cuff
7- Laryngeal nerve damage
8- Obstruction of tracheostomy tube
9- Pneumothorax
10- Subcutaneous and mediastinal emphysema
11- Swallowing dysfunction
12- Tracheoesophageal fistula
13- Infection
14- Accidental decannulation with loss of airway
Nursing care of patients on
mechanical ventilation
Assessment:
1- Assess the patient
2- Assess the artificial airway (tracheostomy
or endotracheal tube)
3- Assess the ventilator
Nursing Interventions
1-Maintain airway patency &
oxygenation
2- Promote comfort
3- Maintain fluid & electrolytes balance
4- Maintain nutritional state
5- Maintain urinary & bowel
elimination
6- Maintain eye , mouth and
cleanliness and integrity:-
7- Maintain mobility/ musculoskeletal
function:-
Nursing Interventions
8- Maintain safety:-
9- Provide psychological support
10- Facilitate communication
11- Provide psychological support &
information to family
12- Responding to ventilator alarms
/Troublshooting
ventilator alarms
13- Prevent nosocomial infection
14- Documentation
Responding To Alarms
• If an alarm sounds, respond immediately
because the problem could be serious.
• Assess the patient first, while you silence
the alarm.
• If you can not quickly identify the
problem, take the patient off the ventilator
and ventilate him with a resuscitation bag
connected to oxygen source until the
physician arrives.
• A nurse or respiratory therapist must
respond to every ventilator alarm.
•Alarms must never be ignored
or disarmed.
• Ventilator malfunction is a
potentially serious problem.
Nursing or respiratory therapists
perform ventilator checks every
2 to 4 hours, and recurrent
alarms may alert the clinician to
the possibility of an equipment-
related issue.
• When device malfunction is
suspected, a second person
manually ventilates the patient
while the nurse or therapist
looks for the cause.
• If a problem cannot be promptly
corrected by ventilator
adjustment, a different machine
is procured so the ventilator in
question can be taken out of
service for analysis and repair by
technical staff.
Causes of Ventilator Alarms
High pressure alarm
• Increased secretions
• Kinked ventilator tubing or
endotracheal tube (ETT)
• Patient biting the ETT
• Water in the ventilator tubing.
• ETT advanced into right mainstem
bronchus.
Low pressure alarm
• Disconnected tubing
• A cuff leak
• A hole in the tubing (ETT or
ventilator tubing)
• A leak in the humidifier
Oxygen alarm
• The oxygen supply is insufficient or is not
properly connected.
High respiratory rate alarm
• Episodes of tachypnea,
• Anxiety,
• Pain,
• Hypoxia,
• Fever.
-
Apnea alarm
• During weaning, indicates that the
patient has a slow Respiratory rate
and a period of apnea.
Temperature alarm
• Overheating due to too low or no
gas flow.
• Improper water levels
•
Methods of Weaning
1- T-piece trial,
2- Continuous Positive Airway Pressure
(CPAP) weaning,
3- Synchronized Intermittent Mandatory
Ventilation (SIMV) weaning,
4- Pressure Support Ventilation (PSV)
weaning.
1- T-Piece trial
• It consists of removing the patient from
the ventilator and having him / her
breathe spontaneously on a T-tube
connected to oxygen source.
• During T-piece weaning, periods of
ventilator support are alternated with
spontaneous breathing.
• The goal is to progressively increase the
time spent off the ventilator.
2-Synchronized Intermittent
Mandatory Ventilation ( SIMV)
Weaning
• SIMV is the most common method of
weaning.
• It consists of gradually decreasing the
number of breaths delivered by the
ventilator to allow the patient to increase
number of spontaneous breaths
3-Continuous Positive Airway Pressure
( CPAP) Weaning
• When placed on CPAP, the patient does all
the work of breathing without the aid of a
back up rate or tidal volume.
• No mandatory (ventilator-initiated)
breaths are delivered in this mode i.e. all
ventilation is spontaneously initiated by
the patient.
• Weaning by gradual decrease in pressure
value
4- Pressure Support Ventilation (PSV)
Weaning
• The patient must initiate all pressure support
breaths.
• During weaning using the PSV mode the level of
pressure support is gradually decreased based on
the patient maintaining an adequate tidal volume
(8 to 12 mL/kg) and a respiratory rate of less
than 25 breaths/minute.
• PSV weaning is indicated for :-
- Difficult to wean patients
- Small spontaneous tidal volume.
Weaning readiness Criteria
• Awake and alert
• Hemodynamically stable, adequately
resuscitated, and not requiring vasoactive
support
• Arterial blood gases (ABGs) normalized or
at patient’s baseline
- PaCO2 acceptable
- PH of 7.35 – 7.45
- PaO2 > 60 mm Hg ,
- SaO2 >92%
- FIO2 ≤40%
• Positive end-expiratory pressure
(PEEP) ≤5 cm H2O
• F < 25 / minute
• Vt 5 ml / kg
• VE 5- 10 L/m (f x Vt)
• VC > 10- 15 ml / kg
• PEP (positive expiratory pressure) >
- 20 cm H2O ( indicates patient’s
ability to take a deep breath &
cough),
• Chest x-ray reviewed for correctable
factors; treated as indicated,
• Major electrolytes within normal
range,
• Hematocrit >25%,
• Core temperature >36°C and <39°C,
• Adequate management of
pain/anxiety/agitation,
• Adequate analgesia/ sedation
(record scores on flow sheet),
• No residual neuromuscular blockade.
Role of nurse before weaning:-
1- Ensure that indications for the implementation
of Mechanical ventilation have improved
2- Ensure that all factors that may interfere with
successful weaning are corrected:-
- Acid-base abnormalitie
- Fluid imbalance
- Electrolyte abnormalities
- Infection
- Fever
- Anemia
- Hyperglycemia
- Protein
- Sleep deprivation
Role of nurse before weaning:-
3- Assess readiness for weaning
4- Ensure that the weaning criteria /
parameters are met.
5- Explain the process of weaning to the patient and
offer reassurance to the patient.
6- Initiate weaning in the morning
when the patient is rested.
7- Elevate the head of the bed & Place
the patient upright
8- Ensure a patent airway and suction if
necessary before a weaning trial,
9- Provide for rest period on ventilator
for 15 – 20 minutes after suctioning.
10- Ensure patient’s comfort & administer
pharmacological agents for comfort, such as
bronchodilators or sedatives as indicated.
11- Help the patient through some of the
discomfort and apprehension.
12- Support and reassurance help the patient
through the discomfort and apprehension
as remains with the patient after initiation
of the weaning process.
13- Evaluate and document the patient’s
response to weaning.
Role of nurse during weaning:-
1- Wean only during the day.
2- Remain with the patient during
initiation of weaning.
3- Instruct the patient to relax and breathe
normally.
4- Monitor the respiratory rate, vital signs,
ABGs, diaphoresis and use of accessory
muscles frequently.
If signs of fatigue or respiratory distress develop.
• Discontinue weaning trials.
Signs of Weaning Intolerance Criteria
• Diaphoresis
• Dyspnea & Labored respiratory pattern
• Increased anxiety ,Restlessness, Decrease
in level of consciousness
• Dysrhythmia,Increase or decrease in heart
rate of > 20 beats /min. or heart rate >
110b/m,Sustained heart rate >20%
higher or lower than baseline
• Increase or decrease in blood pressure of
> 20 mm Hg
Systolic blood pressure >180 mm Hg or
<90 mm Hg
• Increase in respiratory rate of > 10 above
baseline or > 30
Sustained respiratory rate greater than
35 breaths/minute
• Tidal volume ≤5 mL/kg, Sustained minute
ventilation <200 mL/kg/minute
• SaO2 < 90%, PaO2 < 60 mmHg, decrease
in PH of < 7.35.
Increase in PaCO2
Role of nurse after weaning
1- Ensure that extubation criteria are
met .
2- Decanulate or extubat
2- Documentation
Good Luck

More Related Content

What's hot

Basic Mechanical Ventilation
Basic Mechanical VentilationBasic Mechanical Ventilation
Basic Mechanical Ventilation
Andrew Ferguson
 
Haemodynamic monitoring-Minati
Haemodynamic monitoring-MinatiHaemodynamic monitoring-Minati
Haemodynamic monitoring-Minati
Minati Choudhury
 
Identifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - PilbeamIdentifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - Pilbeam
River City Symposium
 
New modes of mechanical ventilation TRC
New modes of mechanical ventilation TRCNew modes of mechanical ventilation TRC
New modes of mechanical ventilation TRC
chandra talur
 

What's hot (20)

Basic Mechanical Ventilation
Basic Mechanical VentilationBasic Mechanical Ventilation
Basic Mechanical Ventilation
 
Mechanical ventilation wave forms
Mechanical ventilation wave formsMechanical ventilation wave forms
Mechanical ventilation wave forms
 
Anatomy of the ventilator
Anatomy of the ventilatorAnatomy of the ventilator
Anatomy of the ventilator
 
Haemodynamic monitoring-Minati
Haemodynamic monitoring-MinatiHaemodynamic monitoring-Minati
Haemodynamic monitoring-Minati
 
Newer modes of ventilation
Newer modes of ventilationNewer modes of ventilation
Newer modes of ventilation
 
PRVC
PRVCPRVC
PRVC
 
Ventilator graphics
Ventilator graphicsVentilator graphics
Ventilator graphics
 
aprv
aprvaprv
aprv
 
Ventilatory management in obstructive airway diseases
Ventilatory management in obstructive airway diseasesVentilatory management in obstructive airway diseases
Ventilatory management in obstructive airway diseases
 
Non Invasive Ventilation
Non Invasive VentilationNon Invasive Ventilation
Non Invasive Ventilation
 
Advanced modes of Mechanical Ventilation-Do we need them?
Advanced modes of Mechanical Ventilation-Do we need them?Advanced modes of Mechanical Ventilation-Do we need them?
Advanced modes of Mechanical Ventilation-Do we need them?
 
COPD Lecture 10 non invasive and invasive mechanical ventilation
COPD Lecture 10   non invasive and invasive mechanical ventilationCOPD Lecture 10   non invasive and invasive mechanical ventilation
COPD Lecture 10 non invasive and invasive mechanical ventilation
 
Physiology of mechanical ventilation upload
Physiology of mechanical ventilation   uploadPhysiology of mechanical ventilation   upload
Physiology of mechanical ventilation upload
 
Hemodynamic parameters & fluid therapy Asim
Hemodynamic parameters &  fluid therapy AsimHemodynamic parameters &  fluid therapy Asim
Hemodynamic parameters & fluid therapy Asim
 
Identifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - PilbeamIdentifying Asynchrony and Solving the Problem - Pilbeam
Identifying Asynchrony and Solving the Problem - Pilbeam
 
NIV (Non Invasive Mechanical Ventilation)
NIV (Non Invasive Mechanical Ventilation)NIV (Non Invasive Mechanical Ventilation)
NIV (Non Invasive Mechanical Ventilation)
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilation
 
Basic modes of mechanical ventilation
Basic modes of mechanical ventilationBasic modes of mechanical ventilation
Basic modes of mechanical ventilation
 
New modes of mechanical ventilation TRC
New modes of mechanical ventilation TRCNew modes of mechanical ventilation TRC
New modes of mechanical ventilation TRC
 
Artificial airways
Artificial airwaysArtificial airways
Artificial airways
 

Similar to MECHANICAL VENTILATION (1).ppt

basic mecanical ventilation critical nursing.pptx
basic mecanical ventilation critical nursing.pptxbasic mecanical ventilation critical nursing.pptx
basic mecanical ventilation critical nursing.pptx
MosaHasen
 
Mechanical ventilation in neonates by dr naved akhter
Mechanical ventilation in neonates by dr naved akhterMechanical ventilation in neonates by dr naved akhter
Mechanical ventilation in neonates by dr naved akhter
Dr Naved Akhter
 
Mechanical ventilator
Mechanical ventilatorMechanical ventilator
Mechanical ventilator
Suchithra Pv
 

Similar to MECHANICAL VENTILATION (1).ppt (20)

MECHANICAL VENTILATION ventilator final presentation
MECHANICAL VENTILATION ventilator final presentationMECHANICAL VENTILATION ventilator final presentation
MECHANICAL VENTILATION ventilator final presentation
 
Mechanical ventilation (1)
Mechanical ventilation (1)Mechanical ventilation (1)
Mechanical ventilation (1)
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilation
 
basic mecanical ventilation critical nursing.pptx
basic mecanical ventilation critical nursing.pptxbasic mecanical ventilation critical nursing.pptx
basic mecanical ventilation critical nursing.pptx
 
Demonstration on Mechanical Ventilator.pptx
Demonstration on Mechanical Ventilator.pptxDemonstration on Mechanical Ventilator.pptx
Demonstration on Mechanical Ventilator.pptx
 
Mechanical ventilation in neonates by dr naved akhter
Mechanical ventilation in neonates by dr naved akhterMechanical ventilation in neonates by dr naved akhter
Mechanical ventilation in neonates by dr naved akhter
 
Mechanical Ventilation.pdf
Mechanical Ventilation.pdfMechanical Ventilation.pdf
Mechanical Ventilation.pdf
 
Mechanical ventilation
Mechanical ventilationMechanical ventilation
Mechanical ventilation
 
Mechanical Ventilation.pptx
Mechanical Ventilation.pptxMechanical Ventilation.pptx
Mechanical Ventilation.pptx
 
Mechanical ventilation in congenital heart defect
Mechanical ventilation in congenital heart defectMechanical ventilation in congenital heart defect
Mechanical ventilation in congenital heart defect
 
Modes of MECHANICAL VENTILLATION
Modes of  MECHANICAL VENTILLATIONModes of  MECHANICAL VENTILLATION
Modes of MECHANICAL VENTILLATION
 
Mechanical Ventilation by ms preeti Singh
Mechanical Ventilation by ms preeti SinghMechanical Ventilation by ms preeti Singh
Mechanical Ventilation by ms preeti Singh
 
New عرض تقديمي من Microsoft PowerPoint.pptx
New عرض تقديمي من Microsoft PowerPoint.pptxNew عرض تقديمي من Microsoft PowerPoint.pptx
New عرض تقديمي من Microsoft PowerPoint.pptx
 
ventilator mode.pdf
ventilator mode.pdfventilator mode.pdf
ventilator mode.pdf
 
Ventilator mode
Ventilator modeVentilator mode
Ventilator mode
 
Mechanical ventilator
Mechanical ventilatorMechanical ventilator
Mechanical ventilator
 
patient-ventilatorinteractions-181207175855.pdf
patient-ventilatorinteractions-181207175855.pdfpatient-ventilatorinteractions-181207175855.pdf
patient-ventilatorinteractions-181207175855.pdf
 
Patient ventilator interactions
Patient ventilator interactionsPatient ventilator interactions
Patient ventilator interactions
 
Mechanical ventilator basic setting and modes
Mechanical ventilator  basic  setting and modesMechanical ventilator  basic  setting and modes
Mechanical ventilator basic setting and modes
 
Mechancial Ventilation Part One
Mechancial Ventilation Part OneMechancial Ventilation Part One
Mechancial Ventilation Part One
 

More from SharanabasappaDegoan

REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333
REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333
REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333
SharanabasappaDegoan
 
Presentation-Smart-Cities-International-Virtual-Symposium-2021.pptx
Presentation-Smart-Cities-International-Virtual-Symposium-2021.pptxPresentation-Smart-Cities-International-Virtual-Symposium-2021.pptx
Presentation-Smart-Cities-International-Virtual-Symposium-2021.pptx
SharanabasappaDegoan
 

More from SharanabasappaDegoan (20)

REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333
REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333
REAL ESATETE MANAEMENT125213552.3352333333333333333333333333333333333
 
Professional practice and Valuation (1).pptx
Professional practice and Valuation (1).pptxProfessional practice and Valuation (1).pptx
Professional practice and Valuation (1).pptx
 
41261.ppt
41261.ppt41261.ppt
41261.ppt
 
1) Basics on mechanical ventilation (2).ppt
1) Basics on mechanical ventilation (2).ppt1) Basics on mechanical ventilation (2).ppt
1) Basics on mechanical ventilation (2).ppt
 
Project Estimation.ppt
Project Estimation.pptProject Estimation.ppt
Project Estimation.ppt
 
MAINTENANCE.pptx
MAINTENANCE.pptxMAINTENANCE.pptx
MAINTENANCE.pptx
 
homeandbuildingautomationsystems.pptx
homeandbuildingautomationsystems.pptxhomeandbuildingautomationsystems.pptx
homeandbuildingautomationsystems.pptx
 
BT & M Unit3.2.pptx.pptx
BT & M Unit3.2.pptx.pptxBT & M Unit3.2.pptx.pptx
BT & M Unit3.2.pptx.pptx
 
Fundamental of Noise.ppt
Fundamental of Noise.pptFundamental of Noise.ppt
Fundamental of Noise.ppt
 
1338301613.ppt
1338301613.ppt1338301613.ppt
1338301613.ppt
 
Green_Building.pptx
Green_Building.pptxGreen_Building.pptx
Green_Building.pptx
 
hr_om11_ch03.ppt
hr_om11_ch03.ppthr_om11_ch03.ppt
hr_om11_ch03.ppt
 
bms-the-basics-explained.pptx
bms-the-basics-explained.pptxbms-the-basics-explained.pptx
bms-the-basics-explained.pptx
 
MECH3422_1516_01_Intro_BSE.pptx
MECH3422_1516_01_Intro_BSE.pptxMECH3422_1516_01_Intro_BSE.pptx
MECH3422_1516_01_Intro_BSE.pptx
 
bms-the-basics-explained.pptx
bms-the-basics-explained.pptxbms-the-basics-explained.pptx
bms-the-basics-explained.pptx
 
Presentation-Smart-Cities-International-Virtual-Symposium-2021.pptx
Presentation-Smart-Cities-International-Virtual-Symposium-2021.pptxPresentation-Smart-Cities-International-Virtual-Symposium-2021.pptx
Presentation-Smart-Cities-International-Virtual-Symposium-2021.pptx
 
building-services-activities.pptx
building-services-activities.pptxbuilding-services-activities.pptx
building-services-activities.pptx
 
Tech 031 Unit 5pp.ppt
Tech 031 Unit 5pp.pptTech 031 Unit 5pp.ppt
Tech 031 Unit 5pp.ppt
 
admin-TQM.ppt
admin-TQM.pptadmin-TQM.ppt
admin-TQM.ppt
 
BUILDING MANAGMENT.ppt
BUILDING MANAGMENT.pptBUILDING MANAGMENT.ppt
BUILDING MANAGMENT.ppt
 

Recently uploaded

Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoorTop Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
dharasingh5698
 
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
ssuser89054b
 
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Kandungan 087776558899
 
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
dollysharma2066
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdf
ankushspencer015
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
amitlee9823
 

Recently uploaded (20)

University management System project report..pdf
University management System project report..pdfUniversity management System project report..pdf
University management System project report..pdf
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
 
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
 
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoorTop Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
 
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torque
 
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak HamilCara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
Cara Menggugurkan Sperma Yang Masuk Rahim Biyar Tidak Hamil
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.ppt
 
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
 
NFPA 5000 2024 standard .
NFPA 5000 2024 standard                                  .NFPA 5000 2024 standard                                  .
NFPA 5000 2024 standard .
 
Double rodded leveling 1 pdf activity 01
Double rodded leveling 1 pdf activity 01Double rodded leveling 1 pdf activity 01
Double rodded leveling 1 pdf activity 01
 
Call Girls Wakad Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Wakad Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Wakad Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Wakad Call Me 7737669865 Budget Friendly No Advance Booking
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineering
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdf
 
Work-Permit-Receiver-in-Saudi-Aramco.pptx
Work-Permit-Receiver-in-Saudi-Aramco.pptxWork-Permit-Receiver-in-Saudi-Aramco.pptx
Work-Permit-Receiver-in-Saudi-Aramco.pptx
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
 

MECHANICAL VENTILATION (1).ppt

  • 1.
  • 2. Mechanical Ventilation Dr. Abdul-Monim Batiha Assistant Professor Critical Care Nursing
  • 3.
  • 4. • Mechanical Ventilation is ventilation of the lungs by artificial means usually by a ventilator. • A ventilator delivers gas to the lungs with either negative or positive pressure.
  • 5. Purposes: • To maintain or improve ventilation, & tissue oxygenation. • To decrease the work of breathing & improve patient’s comfort.
  • 6. Indications: 1- Acute respiratory failure due to: • Mechanical failure, includes neuromuscular diseases as Myasthenia Gravis, Guillain-Barré Syndrome, and Poliomyelitis (failure of the normal respiratory neuromuscular system) • Musculoskeletal abnormalities, such as chest wall trauma (flail chest) • Infectious diseases of the lung such as pneumonia, tuberculosis.
  • 7. 2- Abnormalities of pulmonary gas exchange as in: • Obstructive lung disease in the form of asthma, chronic bronchitis or emphysema. • Conditions such as pulmonary edema, atelectasis, pulmonary fibrosis. • Patients who has received general anesthesia as well as post cardiac arrest patients often require ventilatory support until they have recovered from the effects of the anesthesia or the insult of an arrest.
  • 8. Criteria for institution of ventilatory support: Normal range Ventilation indicated Parameters 10-20 5-7 65-75 75-100 > 35 < 5 < 15 <-20 A- Pulmonary function studies: • Respiratory rate (breaths/min). • Tidal volume (ml/kg body wt) • Vital capacity (ml/kg body wt) • Maximum Inspiratory Force (cm HO2)
  • 9. Criteria for institution of ventilatory support: Normal range Ventilation indicated Parameters 7.35-7.45 75-100 35-45 < 7.25 < 60 > 50 B- Arterial blood Gases • PH • PaO2 (mmHg) • PaCO2 (mmHg)
  • 10. Types of Mechanical ventilators: • Negative-pressure ventilators • Positive-pressure ventilators.
  • 11. Negative-Pressure Ventilators • Early negative-pressure ventilators were known as “iron lungs.” • The patient’s body was encased in an iron cylinder and negative pressure was generated . • The iron lung are still occasionally used today.
  • 12.
  • 13. • Intermittent short-term negative-pressure ventilation is sometimes used in patients with chronic diseases. • The use of negative-pressure ventilators is restricted in clinical practice, however, because they limit positioning and movement and they lack adaptability to large or small body torsos (chests) . • Our focus will be on the positive-pressure ventilators.
  • 14. Positive-pressure ventilators • Positive-pressure ventilators deliver gas to the patient under positive- pressure, during the inspiratory phase.
  • 15.
  • 16. Types of Positive-Pressure Ventilators 1- Volume Ventilators. 2- Pressure Ventilators 3- High-Frequency Ventilators
  • 17. • The volume ventilator is commonly used in critical care settings. • The basic principle of this ventilator is that a designated volume of air is delivered with each breath. • The amount of pressure required to deliver the set volume depends on :- - Patient’s lung compliance - Patient–ventilator resistance factors. 1- Volume Ventilators
  • 18. • Therefore, peak inspiratory pressure (PIP ) must be monitored in volume modes because it varies from breath to breath. • With this mode of ventilation, a respiratory rate, inspiratory time, and tidal volume are selected for the mechanical breaths.
  • 19. • The use of pressure ventilators is increasing in critical care units. • A typical pressure mode delivers a selected gas pressure to the patient early in inspiration, and sustains the pressure throughout the inspiratory phase. • By meeting the patient’s inspiratory flow demand throughout inspiration, patient effort is reduced and comfort increased. 2- Pressure Ventilators
  • 20. • Although pressure is consistent with these modes, volume is not. • Volume will change with changes in resistance or compliance, • Therefore, exhaled tidal volume is the variable to monitor closely. • With pressure modes, the pressure level to be delivered is selected, and with some mode options (i.e., pressure controlled [PC], described later), rate and inspiratory time are preset as well.
  • 21. 3- High-Frequency Ventilators • High-frequency ventilators use small tidal volumes (1 to 3 mL/kg) at frequencies greater than 100 breaths/minute. • The high-frequency ventilator accomplishes oxygenation by the diffusion of oxygen and carbon dioxide from high to low gradients of concentration.
  • 22. • This diffusion movement is increased if the kinetic energy of the gas molecules is increased. • A high-frequency ventilator would be used to achieve lower peak ventilator pressures, thereby lowering the risk of barotrauma.
  • 23. Classification of positive-pressure ventilators: • Ventilators are classified according to how the inspiratory phase ends. The factor which terminates the inspiratory cycle reflects the machine type. • They are classified as: 1- Pressure cycled ventilator 2- Volume cycled ventilator 3- Time cycled ventilator
  • 24. 1- Volume-cycled ventilator • Inspiration is terminated after a preset tidal volume has been delivered by the ventilator. • The ventilator delivers a preset tidal volume (VT), and inspiration stops when the preset tidal volume is achieved.
  • 25. 2- Pressure-cycled ventilator • In which inspiration is terminated when a specific airway pressure has been reached. • The ventilator delivers a preset pressure; once this pressure is achieved, end inspiration occurs.
  • 26. 3- Time-cycled ventilator • In which inspiration is terminated when a preset inspiratory time, has elapsed. • Time cycled machines are not used in adult critical care settings. They are used in pediatric intensive care areas.
  • 27. Ventilator mode • The way the machine ventilates the patient • How much the patient will participate in his own ventilatory pattern. • Each mode is different in determining how much work of breathing the patient has to do.
  • 28. Modes of Mechanical Ventilation A- Volume Modes B- Pressure Modes
  • 29. A- Volume Modes 1- Assist-control (A/C) 2- Synchronized intermittent mandatory ventilation (SIMV)
  • 30. 1- Assist Control Mode A/C • The ventilator provides the patient with a pre-set tidal volume at a pre-set rate . • The patient may initiate a breath on his own, but the ventilator assists by delivering a specified tidal volume to the patient. Client can initiate breaths that are delivered at the preset tidal volume. • Client can breathe at a higher rate than the preset number of breaths/minute
  • 31. • The total respiratory rate is determined by the number of spontaneous inspiration initiated by the patient plus the number of breaths set on the ventilator. • In A/C mode, a mandatory (or “control”) rate is selected. • If the patient wishes to breathe faster, he or she can trigger the ventilator and receive a full-volume breath.
  • 32. • Often used as initial mode of ventilation • When the patient is too weak to perform the work of breathing (e.g., when emerging from anesthesia). Disadvantages: • Hyperventilation,
  • 33. 2- Synchronized Intermittent Mandatory Ventilation (SIMV) • The ventilator provides the patient with a pre-set number of breaths/minute at a specified tidal volume and FiO2. • In between the ventilator-delivered breaths, the patient is able to breathe spontaneously at his own tidal volume and rate with no assistance from the ventilator. • However, unlike the A/C mode, any breaths taken above the set rate are spontaneous breaths taken through the ventilator circuit.
  • 34. • The tidal volume of these breaths can vary drastically from the tidal volume set on the ventilator, because the tidal volume is determined by the patient’s spontaneous effort. • Adding pressure support during spontaneous breaths can minimize the risk of increased work of breathing. • Ventilators breaths are synchronized with the patient spontaneous breathe. ( no fighting)
  • 35. • Used to wean the patient from the mechanical ventilator. • Weaning is accomplished by gradually lowering the set rate and allowing the patient to assume more work
  • 36. B- Pressure Modes 1- Pressure-controlled ventilation (PCV) 2- Pressure-support ventilation (PSV) 3- Continuous positive airway pressure (CPAP) 4- Positive end expiratory pressure (PEEP) 5- Noninvasive bilevel positive airway pressure ventilation (BiPAP)
  • 37. 1- Control Mode (CM) Continuous Mandatory Ventilation ( CMV) • Ventilation is completely provided by the mechanical ventilator with a preset tidal volume, respiratory rate and oxygen concentration • Ventilator totally controls the patient’s ventilation i.e. the ventilator initiates and controls both the volume delivered and the frequency of breath. • Client does not breathe spontaneously. • Client can not initiate breathe
  • 38. 2- Pressure-Controlled Ventilation Mode ( PCV) • The PCV mode is used – If compliance is decreased and the risk of barotrauma is high. – It is used when the patient has persistent oxygenation problems despite a high FiO2 and high levels of PEEP. • The inspiratory pressure level, respiratory rate, and inspiratory–expiratory (I:E) ratio must be selected.
  • 39. 2- Pressure-Controlled Ventilation Mode ( PCV) • In pressure controlled ventilation the breathing gas flows under constant pressure into the lungs during the selected inspiratory time. • The flow is highest at the beginning of inspiration( i.e when the volume is lowest in the lungs). • As the pressure is constant the flow is initially high and then decreases with increasing filling of the lungs.
  • 40. • Like volume controlled ventilation PCV is time controlled.
  • 41. Advantages of pressure limitations are: • 1- reduction of peak pressure and therefore the risk of barotruma and tracheal injury. • 2- effective ventilation. • Improve gas exchange
  • 42. • Tidal volume varies with compliance and airway resistance and must be closely monitored. • Sedation and the use of neuromuscular blocking agents are frequently indicated, because any patient–ventilator asynchrony usually results in profound drops in the SaO2. • This is especially true when inverse ratios are used. The “unnatural” feeling of this mode often requires muscle relaxants to ensure patient–ventilator synchrony.
  • 43. • Inverse ratio ventilation (IRV) mode reverses this ratio so that inspiratory time is equal to, or longer than, expiratory time (1:1 to 4:1). • Inverse I:E ratios are used in conjunction with pressure control to improve oxygenation by expanding stiff alveoli by using longer distending times, thereby providing more opportunity for gas exchange and preventing alveolar collapse.
  • 44. • As expiratory time is decreased, one must monitor for the development of hyperinflation or auto-PEEP. Regional alveolar overdistension and barotrauma may occur owing to excessive total PEEP. • When the PCV mode is used, the mean airway and intrathoracic pressures rise, potentially resulting in a decrease in cardiac output and oxygen delivery. Therefore, the patient’s hemodynamic status must be monitored closely. • Used to limit plateau pressures that can cause barotrauma & Severe ARDS
  • 45. 3- Pressure Support Ventilation ( PSV) • The patient breathes spontaneously while the ventilator applies a pre-determined amount of positive pressure to the airways upon inspiration. • Pressure support ventilation augments patient’s spontaneous breaths with positive pressure boost during inspiration i.e. assisting each spontaneous inspiration. • Helps to overcome airway resistance and reducing the work of breathing.
  • 46. • Indicated for patients with small spontaneous tidal volume and difficult to wean patients. • Patient must initiate all pressure support breaths. • Pressure support ventilation may be combined with other modes such as SIMV or used alone for a spontaneously breathing patient.
  • 47. • The patient’s effort determines the rate, inspiratory flow, and tidal volume. • In PSV mode, the inspired tidal volume and respiratory rate must be monitored closely to detect changes in lung compliance. • It is a mode used primarily for weaning from mechanical ventilation.
  • 48. • Constant positive airway pressure during spontaneous breathing • CPAP allows the nurse to observe the ability of the patient to breathe spontaneously while still on the ventilator. • CPAP can be used for intubated and nonintubated patients. • It may be used as a weaning mode and for nocturnal ventilation (nasal or mask CPAP) 4- Continuous Positive Airway Pressure (CPAP)
  • 49. 5- Positive end expiratory pressure (PEEP) • Positive pressure applied at the end of expiration during mandatory ventilator breath • positive end-expiratory pressure with positive-pressure (machine) breaths.
  • 50. Uses of CPAP & PEEP • Prevent atelactasis or collapse of alveoli • Treat atelactasis or collapse of alveoli • Improve gas exchange & oxygenation • Treat hypoxemia refractory to oxygen therapy.(prevent oxygen toxicity • Treat pulmonary edema ( pressure help expulsion of fluids from alveoli
  • 51. 6- Noninvasive Bilateral Positive Airway Pressure Ventilation (BiPAP) • BiPAP is a noninvasive form of mechanical ventilation provided by means of a nasal mask or nasal prongs, or a full-face mask. • The system allows the clinician to select two levels of positive-pressure support: • An inspiratory pressure support level (referred to as IPAP) • An expiratory pressure called EPAP (PEEP/CPAP level).
  • 52. Common Ventilator Settings parameters/ controls • Fraction of inspired oxygen (FIO2) • Tidal Volume (VT) • Peak Flow/ Flow Rate • Respiratory Rate/ Breath Rate / Frequency ( F) • Minute Volume (VE) • I:E Ratio (Inspiration to Expiration Ratio) • Sigh
  • 53. ● Fraction of inspired oxygen (FIO2) • The percent of oxygen concentration that the patient is receiving from the ventilator. (Between 21% & 100%) (room air has 21% oxygen content). • Initially a patient is placed on a high level of FIO2 (60% or higher). • Subsequent changes in FIO2 are based on ABGs and the SaO2.
  • 54. • In adult patients the initial FiO2 may be set at 100% until arterial blood gases can document adequate oxygenation. • An FiO2 of 100% for an extended period of time can be dangerous ( oxygen toxicity) but it can protect against hypoxemia • For infants, and especially in premature infants, high levels of FiO2 (>60%) should be avoided. • Usually the FIO2 is adjusted to maintain an SaO2 of greater than 90% (roughly equivalent to a PaO2 >60 mm Hg). • Oxygen toxicity is a concern when an FIO2 of greater than 60% is required for more than 25 hours
  • 55. Signs and symptoms of oxygen toxicity :- 1- Flushed face 2- Dry cough 3- Dyspnea 4- Chest pain 5- Tightness of chest 6- Sore throat
  • 56. ● Tidal Volume (VT) • The volume of air delivered to a patient during a ventilator breath. • The amount of air inspired and expired with each breath. • Usual volume selected is between 5 to 15 ml/ kg body weight)
  • 57. • In the volume ventilator, Tidal volumes of 10 to 15 mL/kg of body weight were traditionally used. • the large tidal volumes may lead to (volutrauma) aggravate the damage inflicted on the lungs • For this reason, lower tidal volume targets (6 to 8 mL/kg) are now recommended.
  • 58. ● Peak Flow/ Flow Rate • The speed of delivering air per unit of time, and is expressed in liters per minute. • The higher the flow rate, the faster peak airway pressure is reached and the shorter the inspiration; • The lower the flow rate, the longer the inspiration.
  • 59. ● Respiratory Rate/ Breath Rate / Frequency ( F) • The number of breaths the ventilator will deliver/minute (10-16 b/m). • Total respiratory rate equals patient rate plus ventilator rate. • The nurse double-checks the functioning of the ventilator by observing the patient’s respiratory rate.
  • 60. For adult patients and older children:- With COPD • A reduced tidal volume • A reduced respiratory rate For infants and younger children:- • A small tidal volume • Higher respiratory rate
  • 61. ● Minute Volume (VE) • The volume of expired air in one minute . • Respiratory rate times tidal volume equals minute ventilation VE = (VT x F) • In special cases, hypoventilation or hyperventilation is desired
  • 62. In a patient with head injury, • Respiratory alkalosis may be required to promote cerebral vasoconstriction, with a resultant decrease in ICP. • In this case, the tidal volume and respiratory rate are increased ( hyperventilation) to achieve the desired alkalotic pH by manipulating the PaCO2.
  • 63. In a patient with COPD • Baseline ABGs reflect an elevated PaCO2 should not hyperventilated. Instead, the goal should be restoration of the baseline PaCO2. • These patients usually have a large carbonic acid load, and lowering their carbon dioxide levels rapidly may result in seizures.
  • 64. ● I:E Ratio (Inspiration to Expiration Ratio):- • The ratio of inspiratory time to expiratory time during a breath (Usually = 1:2)
  • 65. ● Sigh • A deep breath. • A breath that has a greater volume than the tidal volume. • It provides hyperinflation and prevents atelectasis. • Sigh volume :------------------Usual volume is 1.5 – 2 times tidal volume. • Sigh rate/ frequency :---------Usual rate is 4 to 8 times an hour.
  • 66. ● Peak Airway Pressure:- • In adults if the peak airway pressure is persistently above 45 cmH2O, the risk of barotrauma is increased and efforts should be made to try to reduce the peak airway pressure. • In infants and children it is unclear what level of peak pressure may cause damage. In general, keeping peak pressures below 30 is desirable.
  • 67. ● Pressure Limit • On volume-cycled ventilators, the pressure limit dial limits the highest pressure allowed in the ventilator circuit. • Once the high pressure limit is reached, inspiration is terminated. • Therefore, if the pressure limit is being constantly reached, the designated tidal volume is not being delivered to the patient.
  • 68. ● Sensitivity(trigger Sensitivity) • The sensitivity function controls the amount of patient effort needed to initiate an inspiration • Increasing the sensitivity (requiring less negative force) decreases the amount of work the patient must do to initiate a ventilator breath. • Decreasing the sensitivity increases the amount of negative pressure that the patient needs to initiate inspiration and increases the work of breathing.
  • 69. • The most common setting for pressure sensitivity are -1 to -2 cm H2O • The more negative the number the harder it to breath.
  • 70. Ensuring humidification and thermoregulation • All air delivered by the ventilator passes through the water in the humidifier, where it is warmed and saturated. • Humidifier temperatures should be kept close to body temperature 35 ºC- 37ºC. • In some rare instances (severe hypothermia), the air temperatures can be increased. • The humidifier should be checked for adequate water levels
  • 71. • An empty humidifier contributes to drying the airway, often with resultant dried secretions, mucus plugging and less ability to suction out secretions. • Humidifier should not be overfilled as this may increase circuit resistance and interfere with spontaneous breathing. • As air passes through the ventilator to the patient, water condenses in the corrugated tubing. This moisture is considered contaminated and must be drained into a receptacle and not back into the sterile humidifier.
  • 72. • If the water is allowed to build up, resistance is developed in the circuit and PEEP is generated. In addition, if moisture accumulates near the endotracheal tube, the patient can aspirate the water. • The nurse and respiratory therapist jointly are responsible for preventing this condensation buildup. The humidifier is an ideal medium for bacterial growth.
  • 73. Ventilator alarms:- • Mechanical ventilators comprise audible and visual alarm systems, which act as immediate warning signals to altered ventilation. • Alarm systems can be categorized according to volume and pressure (high and low). • High-pressure alarms warn of rising pressures. • Low-pressure alarms warn of disconnection of the patient from the ventilator or circuit leaks.
  • 74. Complications of Mechanical Ventilation:- I- Airway Complications, II- Mechanical complications, III- Physiological Complications, IV- Artificial Airway Complications.
  • 75. I- Airway Complications 1- Aspiration 2- Decreased clearance of secretions 3- Nosocomial or ventilator-acquired pneumonia
  • 76. II- Mechanical complications 1- Hypoventilation with atelectasis with respiratory acidosis or hypoxemia. 2- Hyperventilation with hypocapnia and respiratory alkalosis 3- Barotrauma a- Closed pneumothorax, b- Tension pneumothorax, c- Pneumomediastinum, d- Subcutaneous emphysema. 4- Alarm “turned off” 5- Failure of alarms or ventilator 6- Inadequate nebulization or humidification 7- Overheated inspired air, resulting in hyperthermia
  • 77. III- Physiological Complications 1- Fluid overload with humidified air and sodium chloride (NaCl) retention 2- Depressed cardiac function and hypotension 3- Stress ulcers 4- Paralytic ileus 5- Gastric distension 6- Starvation 7- Dyssynchronous breathing pattern
  • 78. IV- Artificial Airway Complications A- Complications related to Endotracheal Tube:- 1- Tube kinked or plugged 2- Rupture of piriform sinus 3- Tracheal stenosis or tracheomalacia 4- Mainstem intubation with contralateral (located on or affecting the opposite side of the • lung) lung atelectasis 5- Cuff failure 6- Sinusitis 7- Otitis media 8- Laryngeal edema
  • 79. B- Complications related to Tracheostomy tube:- 1- Acute hemorrhage at the site 2- Air embolism 3- Aspiration 4- Tracheal stenosis 5- Erosion into the innominate artery with exsanguination 6- Failure of the tracheostomy cuff 7- Laryngeal nerve damage 8- Obstruction of tracheostomy tube 9- Pneumothorax 10- Subcutaneous and mediastinal emphysema 11- Swallowing dysfunction 12- Tracheoesophageal fistula 13- Infection 14- Accidental decannulation with loss of airway
  • 80. Nursing care of patients on mechanical ventilation Assessment: 1- Assess the patient 2- Assess the artificial airway (tracheostomy or endotracheal tube) 3- Assess the ventilator
  • 81. Nursing Interventions 1-Maintain airway patency & oxygenation 2- Promote comfort 3- Maintain fluid & electrolytes balance 4- Maintain nutritional state 5- Maintain urinary & bowel elimination 6- Maintain eye , mouth and cleanliness and integrity:- 7- Maintain mobility/ musculoskeletal function:-
  • 82. Nursing Interventions 8- Maintain safety:- 9- Provide psychological support 10- Facilitate communication 11- Provide psychological support & information to family 12- Responding to ventilator alarms /Troublshooting ventilator alarms 13- Prevent nosocomial infection 14- Documentation
  • 83. Responding To Alarms • If an alarm sounds, respond immediately because the problem could be serious. • Assess the patient first, while you silence the alarm. • If you can not quickly identify the problem, take the patient off the ventilator and ventilate him with a resuscitation bag connected to oxygen source until the physician arrives. • A nurse or respiratory therapist must respond to every ventilator alarm.
  • 84. •Alarms must never be ignored or disarmed. • Ventilator malfunction is a potentially serious problem. Nursing or respiratory therapists perform ventilator checks every 2 to 4 hours, and recurrent alarms may alert the clinician to the possibility of an equipment- related issue.
  • 85. • When device malfunction is suspected, a second person manually ventilates the patient while the nurse or therapist looks for the cause. • If a problem cannot be promptly corrected by ventilator adjustment, a different machine is procured so the ventilator in question can be taken out of service for analysis and repair by technical staff.
  • 86. Causes of Ventilator Alarms High pressure alarm • Increased secretions • Kinked ventilator tubing or endotracheal tube (ETT) • Patient biting the ETT • Water in the ventilator tubing. • ETT advanced into right mainstem bronchus.
  • 87. Low pressure alarm • Disconnected tubing • A cuff leak • A hole in the tubing (ETT or ventilator tubing) • A leak in the humidifier
  • 88. Oxygen alarm • The oxygen supply is insufficient or is not properly connected.
  • 89. High respiratory rate alarm • Episodes of tachypnea, • Anxiety, • Pain, • Hypoxia, • Fever. -
  • 90. Apnea alarm • During weaning, indicates that the patient has a slow Respiratory rate and a period of apnea.
  • 91. Temperature alarm • Overheating due to too low or no gas flow. • Improper water levels •
  • 92. Methods of Weaning 1- T-piece trial, 2- Continuous Positive Airway Pressure (CPAP) weaning, 3- Synchronized Intermittent Mandatory Ventilation (SIMV) weaning, 4- Pressure Support Ventilation (PSV) weaning.
  • 93. 1- T-Piece trial • It consists of removing the patient from the ventilator and having him / her breathe spontaneously on a T-tube connected to oxygen source. • During T-piece weaning, periods of ventilator support are alternated with spontaneous breathing. • The goal is to progressively increase the time spent off the ventilator.
  • 94. 2-Synchronized Intermittent Mandatory Ventilation ( SIMV) Weaning • SIMV is the most common method of weaning. • It consists of gradually decreasing the number of breaths delivered by the ventilator to allow the patient to increase number of spontaneous breaths
  • 95. 3-Continuous Positive Airway Pressure ( CPAP) Weaning • When placed on CPAP, the patient does all the work of breathing without the aid of a back up rate or tidal volume. • No mandatory (ventilator-initiated) breaths are delivered in this mode i.e. all ventilation is spontaneously initiated by the patient. • Weaning by gradual decrease in pressure value
  • 96. 4- Pressure Support Ventilation (PSV) Weaning • The patient must initiate all pressure support breaths. • During weaning using the PSV mode the level of pressure support is gradually decreased based on the patient maintaining an adequate tidal volume (8 to 12 mL/kg) and a respiratory rate of less than 25 breaths/minute. • PSV weaning is indicated for :- - Difficult to wean patients - Small spontaneous tidal volume.
  • 97. Weaning readiness Criteria • Awake and alert • Hemodynamically stable, adequately resuscitated, and not requiring vasoactive support • Arterial blood gases (ABGs) normalized or at patient’s baseline - PaCO2 acceptable - PH of 7.35 – 7.45 - PaO2 > 60 mm Hg , - SaO2 >92% - FIO2 ≤40%
  • 98. • Positive end-expiratory pressure (PEEP) ≤5 cm H2O • F < 25 / minute • Vt 5 ml / kg • VE 5- 10 L/m (f x Vt) • VC > 10- 15 ml / kg • PEP (positive expiratory pressure) > - 20 cm H2O ( indicates patient’s ability to take a deep breath & cough),
  • 99. • Chest x-ray reviewed for correctable factors; treated as indicated, • Major electrolytes within normal range, • Hematocrit >25%, • Core temperature >36°C and <39°C, • Adequate management of pain/anxiety/agitation, • Adequate analgesia/ sedation (record scores on flow sheet), • No residual neuromuscular blockade.
  • 100. Role of nurse before weaning:- 1- Ensure that indications for the implementation of Mechanical ventilation have improved 2- Ensure that all factors that may interfere with successful weaning are corrected:- - Acid-base abnormalitie - Fluid imbalance - Electrolyte abnormalities - Infection - Fever - Anemia - Hyperglycemia - Protein - Sleep deprivation
  • 101. Role of nurse before weaning:- 3- Assess readiness for weaning 4- Ensure that the weaning criteria / parameters are met. 5- Explain the process of weaning to the patient and offer reassurance to the patient.
  • 102. 6- Initiate weaning in the morning when the patient is rested. 7- Elevate the head of the bed & Place the patient upright 8- Ensure a patent airway and suction if necessary before a weaning trial, 9- Provide for rest period on ventilator for 15 – 20 minutes after suctioning.
  • 103. 10- Ensure patient’s comfort & administer pharmacological agents for comfort, such as bronchodilators or sedatives as indicated. 11- Help the patient through some of the discomfort and apprehension. 12- Support and reassurance help the patient through the discomfort and apprehension as remains with the patient after initiation of the weaning process. 13- Evaluate and document the patient’s response to weaning.
  • 104. Role of nurse during weaning:- 1- Wean only during the day. 2- Remain with the patient during initiation of weaning. 3- Instruct the patient to relax and breathe normally. 4- Monitor the respiratory rate, vital signs, ABGs, diaphoresis and use of accessory muscles frequently. If signs of fatigue or respiratory distress develop. • Discontinue weaning trials.
  • 105. Signs of Weaning Intolerance Criteria • Diaphoresis • Dyspnea & Labored respiratory pattern • Increased anxiety ,Restlessness, Decrease in level of consciousness • Dysrhythmia,Increase or decrease in heart rate of > 20 beats /min. or heart rate > 110b/m,Sustained heart rate >20% higher or lower than baseline
  • 106. • Increase or decrease in blood pressure of > 20 mm Hg Systolic blood pressure >180 mm Hg or <90 mm Hg • Increase in respiratory rate of > 10 above baseline or > 30 Sustained respiratory rate greater than 35 breaths/minute • Tidal volume ≤5 mL/kg, Sustained minute ventilation <200 mL/kg/minute • SaO2 < 90%, PaO2 < 60 mmHg, decrease in PH of < 7.35. Increase in PaCO2
  • 107. Role of nurse after weaning 1- Ensure that extubation criteria are met . 2- Decanulate or extubat 2- Documentation