Design of the Pulse Oximetry Measurement Circuit and Its
Sensing System Based On CMOS
Yang Wang1
, Liying Chen2
1
(Tianjin Poly-technic University, School of Electronic and Information Engineering, 300387, China)
2
(Tianjin Poly-technic University, School of Electronic and Information Engineering, 300387, China)
Abstract: The pulse oximetry circuit and its sensing system is designed based on the standard CMOS technology
of 0.18um. The reflection oxygen sensor is used to collect the pulse oximeter signal of human body, then the
collected physiological signals are processed by the data processing circuit The data processing circuit is
composed of two parts: the amplifying circuit and the band-pass filter circuit, and the pulse oximeter data
processed by the data processing circuit is written into the tag through the SPI communication The RFID reader
read the data in the RFID tag through wireless communication, and display the data . The experimental results
show that the maximum error is ±1%. The maximum error of the pulse is ±1.9%. The stability and feasibility of
pulse blood oxygen sensing system is demonstrated in this paper and it will have a good application prospect in
the direction of wearable medical wisdom research.
Keywords: measurement circuit, pulse oximetry, wireless sensor networks, reflection type, active RFID tag
I. Introduction
With the aging of the society is becoming more and more serious, the cost of medical care for the elderly
and the proportion of medical resources is increasing. In order to alleviate the pressure on the health care system,
the traditional medical center model needs to be transformed into a family centered model. Wearable medical
device is a device that integrates signal acquisition, processing and data transmission. The physiological
parameters such as pulse, oxygen saturation and temperature were collected, processed and analyzed by the
biomedical sensors on the human body. The data can be displayed timely feedback to the user in the device, and
can also be used for real-time monitoring of users. It does effectively carry out the prevention and treatment of
disease.
At present, the most mature technology in the detection of blood oxygen saturation [1] in the human body
is the transmission detection structure. However, the reflection type instrument has been paid more and more
attention, because of the limitation of the current type of testing instruments in the acquisition of the human body.
The light source and the light detector of the reflective pulse oxygen sensor is placed on the same side of the
detection part. The detector receives the light reflected back to human tissue, and convert optical signals into
electrical signal as the output of the sensor. The reflective optical detection mode is less limited by the detection
position, so that the detection position is more flexible. The measurement can be achieved in despite of the limit of
position.
With the development of medical sensor technology, the transmission type detection method of pulse
oximetry has been gradually replaced by reflective detection mode. It has become the main measurement method
for detecting pulse oximetry, and received great attention and research in the field of medicine. In this paper, the
reflective probe is used to collect the signal, and the measurement circuit is designed with UMC 180nm CMOS
standard technology. After the amplification and filtering of the collected signal, the signal processed by the
circuit will be transmitted further to other modules.
II. Related Works
The components of the absorption of light in the blood mainly include oxygenated hemoglobin (HbO2)
and reduced hemoglobin (Hb). Blood oxygen saturation is defined as the percentage of HbO2 in the blood, which
is the sum of the concentration of HbO2 and the concentration of Hb in the blood. The concentration of HbO2 is an
important physiological parameter for the metabolism of human tissues. The definition of blood oxygen
saturation:
2
2
2 100%
HbO
HbO Hb
C
SpO
C C
 

(1)
In the formula, Spo2 represents the value of oxygen saturation, CHbo2 represents the concentration of
HbO2 in the blood, and CHb indicates the concentration of Hb in the blood. Bill Lambert's law states that light has
certain attenuation characteristics in non-homogeneous scattering medium, the absorption degree of the light in
human issues has certain difference in different medium. The measurement of oxygen saturation [2,3] is based on
Design of the pulse oximetry measurement circuit and its sensing system based on CMOS
DOI: 10.9790/1676-1201046470 www.iosrjournals.org 65 | Page
Lambert Bill's law as the theoretical basis, the light passes through the medium of interference without
considering the absorption of skin and muscle and other tissue to light condition. The blood HbO2 and Hb have
different absorption rates of different wavelengths. Different wavelengths of light can emit the device as a light
source. We can calculate the content of hemoglobin in blood components through the measurement of the
attenuation degree after the light passing through the blood.
The measurement of the pulse and blood oxygen saturation is based on the principle that the light
absorption quantity will change with the expansion or contraction of the artery, while the light absorption of other
organizations are almost constant. The amount of light absorbed by the arterial blood changes when the arterial
pulse. The amount of light absorbed by the skin, muscles, bones, venous blood, and non-pulsating arteries is
constant. Heart contraction or expansion will affect the arterial volume, according to Lambert Bill's law. Change
of the optical path will affect the absorbance, ripple component of artery is PPG (pulse wave) signal. According to
the photon diffusion equation and time resolved spectroscopy technique, the formula of the reflected light
intensity can be obtained:
2 2
3 5
[(1 ) ]
exp( )exp( )
4( , )
(1 ) (4 )
s
a
s
g
ct
DcR t
g Dc t
 


 

 
 


(2)
When the time is long enough, the formula (2) is approximately equal to ac ,W that represents the change of
the reflected light intensity, is directly proportional to the absorption coefficient s .
aW c
(3)
According to the Lambert Bill law, when a beam of light intensity of 0I is perpendicular to a sample with a
concentration of C , the transmitted light intensity I and 0I exist.
( )
0
CL
I I e  

(4)
In the formula, ( )  is the absorption coefficient of the sample molecule to the wavelength  ; L is the path
of light. In order to carry out the transformation, the absorption coefficient a is obtained.
0
1
ln( ) a
I
C
L I
   
(5)
Assuming that Hb and HbC are the absorption coefficient and concentration of Hb in blood, 2HbO
and 2HbOC are respectively the absorption coefficient and concentration of HbO2 in blood. When the light
irradiates on the human body with two kinds of wavelength of red and infrared light as light source, the absorption
coefficient of HbO2 and Hb on the two wavelengths of light can type (5) respectively, only considering the effect
of absorption of HbO2 in blood and Hb.
2 2
R R R
a Hb Hb Hbo HboC C    (6)
2 2
IR IR IR
a Hb Hb Hbo HboC C    (7)
When light passes through human tissue and blood vessels, it can be divided into two parts: DC component and
AC component. The change rate of light intensity in the tissue is /AC DCW I I binding (2), (6) and (7)
According to the double beam method, the following formula will be derived.
2 2
2 2
R R
Hb Hb Hbo HboR
IR IR
IR Hb Hb Hbo Hbo
C CW
W C C
 
 



(8)
/
/
R R
AC DCR
IR IR
IR AC DC
I IW
A
W I I
  (9)
The formula (9) is brought into (1),
Design of the pulse oximetry measurement circuit and its sensing system based on CMOS
DOI: 10.9790/1676-1201046470 www.iosrjournals.org 66 | Page
2 2
2
R IR
Hb Hb R
R R R R
Hb Hbo Hb Hbo IR
W
spo
W
 
   
 
 
(10)
In the formula,
IR
Hb R
Hb and 2
R
Hbo are all constant, so we can get the formula of blood oxygen saturation:
2 1 2spo k k A  (11)
The reflective pulse oximetry can be calculated by the AC and DC information of the pulse wave signal.
III. The Pulse Oximeter Measurement Circuit
In this paper, the pulse oximetry sensor (DCM03) is used to detect the pulse oxygen saturation, which has
the advantages of small volume, low power consumption and high precision. The key to measure the blood
oxygen saturation is to get the pulse waveform. The pulse signal of human body is a physiological signal of low
frequency and low amplitude, which is about 0.5mV. In theory, the whole measurement system needs to be
enlarged more than 200 times to achieve the requirement of ADC conversion voltage. The pulse frequency is
concentrated in 0.5Hz ~ 5Hz, and the maximum is less than 40Hz. In the process of signal acquisition, the
movement of the body and mental stress will produce baseline drift and noise. At the same time, there will
inevitably be power frequency interference and electromagnetic interference. This requires the suppression of the
base drift circuit and the two filter circuit to filter the interference. The structure of the pulse oximeter circuit based
on CMOS is shown in Fig.1.
Preamplifiercircuit
Suppressorcircuit
LPF
HPF
twostage
amplification
Power management module(bandgap)
input
DC
AC AC
Fig.1 structure of the pulse oximeter circuit
3.1 amplifier circuit
The pulse signal collected by the pulse sensor is very weak, the signal amplitude is 20mV ~ 50mV. In
order to achieve the conversion voltage of ADC, the input signal of the circuit module should be increased by
50~80 times. In order to improve the anti-interference performance of the circuit (to avoid the distortion of the
pulse signal caused by a large gain), the pulse oximeter measurement circuit is composed of preamplifier circuit
and two-stage amplifier circuit. The preamplifier circuit is shown in Fig.2, the two-stage amplifier circuit is shown
in Fig. 3. Among them, the preamplifier circuit uses differential amplifier. The differential amplifier circuit has a
capability of strong common mode rejection (82.5dB), which can suppress the temperature drift, voltage drift and
effectively offset the random noise.
VCC
V1
V2
Vb
GND
A1
A2
A3
C0
C1
R4
R5
R0
R1
R2
R3 R6
R7
Vout
VCC
Vb
ENA
Vin
GND
Vout
A2R0
R1
R2
Fig.2 Preamplifier circuit Fig.3 two-stage amplifier circuit
Design of the pulse oximetry measurement circuit and its sensing system based on CMOS
DOI: 10.9790/1676-1201046470 www.iosrjournals.org 67 | Page
3.2 band pass filter
After the signal is amplified, it needs to filter the interference. The band-pass filter circuit is designed to
eliminate the interference and noise, including low-pass filter and high-pass filter. Low pass filter circuit uses
two-order Butterworth low-pass filter circuit. This type of active circuit can not only remove the interference
effectively, but also reduce the signal attenuation. At the same time, the signal can be amplified to some extent.
The pulse signal after the low-pass filter circuit is composed of DC signal and AC signal. Among them, the DC
signal is the volt level, while the AC signal is millivolt level. Therefore, it is necessary to amplify the AC signal
individually. In this way, the AC and DC signals should be isolated, and then the AC signal can be amplified,
some low-frequency interference can also be filtered out. Finally, the AC signal is fed into the two stage amplifier
to obtain the value of ACI 、 DCI .
IV. System Design
The pulse oximetry sensor system based on CMOS is composed of four modules: the reflective pulse
oximetry module, data processing module, RFID transceiver module and the PC interface. In the process of data
transmission, this system involves two kinds of communication modes including wired communication and
wireless communication [4]. The communication between sensor and MCU, MCU and the tag, the reader and PC
are all through the wired communication mode. The transmission of sensor data between RFID tag and RFID
reader is accomplished by wireless communication. The data processing module will not only process the pulse
and blood oxygen information [5, 6] collected by the sensor, but also establish a connection with the RFID tag. At
the same time, the pulse and blood oxygen data collected and processed are transferred to the register of the RFID
tag. The data transmission between the sensor and the data processing module is through the serial port
connection, and the data transmission between the data processing module and the RFID tag is communicated
through the SPI interface.
RFID transceiver module includes the RFID active tags and the RFID reader. The RFID tag [7, 8] and the
reader achieve the data transmission based on GEN IS0/IEC 18000-6C protocol. Based on the communication
between the data processing modules, the whole communication process: MCU transmits the data in the register to
the RFID tag through the SPI interface. The reader communicates with the tag in accordance with the radio
frequency communication protocol and reads the data in the internal register of the RFID tag. Finally the RFID
reader read the data in the RFID tag through wireless communication [9], and display the data on PC interface.
The overall design structure diagram of pulse oximetry system is shown in Fig.4.
sensor MCU tag reader PC
UART SPI
Pulse oximetry
detection module
Data
processing
module
Data will be
written into the
tag through SPI
The data will be
transmitted through
wireless
communication
Display
interface
light Detecor
Fig.4 the structure diagram of pulse oximetry system
V. Process of The Data Transmission
After the reader and tag are connected, the pulse oximeter system will be tested. The pulse oximeter
sensor is connected with the host computer through the USB interface, which can directly display the red light and
infrared light PPG signal measured by the sensor. According to the principle of dual wavelength detection, the
PPG signal can be transformed into digital signal, which can be used for further transmission of the data. The PPG
signal waveform of the sensor is shown in Fig.5.
Design of the pulse oximetry measurement circuit and its sensing system based on CMOS
DOI: 10.9790/1676-1201046470 www.iosrjournals.org 68 | Page
Fig. 5 waveform of PPG signal collected by sensor
In the pulse oximetry system, the pulse oximeter is connected with the data processing module through
the UART interface. The data processing module is connected with the RFID tag through the SPI interface. Then
the data in the register of data processing module is transferred to the internal register of the tag. The RFID reader
and RFID tags constitute the RFID transceiver module, it works in the 860-960MHz frequency range. The data
transmission is based on the EPC C1 GEN2/ISO 18000-6C protocol. The reader interface is shown in Fig.6, and
the reader starts working after clicking on the connection button. Click on the label and open a single label cycle
recognition mode, the reader will identify the tag within the working range. The data window displays the label
number, storage area and storage address of the tag identified by the reader.
Fig. 6 the reader interface
The data read by the reader is shown in Fig.6, the contents of the data window in the graph are the tag
information identified by the reader and the data in each storage area. The reader is connected with the host
computer through the USB interface, which can display the value and waveform of the pulse and oxygen
saturation in the computer interface. The pulse oximeter information displayed on the computer interface is shown
in Fig.7.
Design of the pulse oximetry measurement circuit and its sensing system based on CMOS
DOI: 10.9790/1676-1201046470 www.iosrjournals.org 69 | Page
Fig. 7 the information displayed in the computer interface Fig. 8 Pulse oximetry data
Click on the record, the pulse and oxygen saturation data received by the host computer can be saved in
the form of text. The real time data detected by the pulse oximeter system is shown in Fig.8. Users can check their
information on a regular basis. By analyzing and processing the real time data of pulse and blood oxygen
saturation, we can know the health status and the development trend. It can play a good role in monitoring the
health of the user.
VI. Experimental Validation
After the system hardware and software debugging, the finger probe BSJ09001C of Shanghai Berry
Company was chosen in order to further verify the feasibility and accuracy of the system. Then the measured data
are compared with the data obtained by the system. The measurement range of oxygen saturation is 35%-100%,
and the measurement error is ±1%. The pulse rate measurement range is 25-250bpm, the measurement error is
± 2%. 50 people of different ages were selected for the test, and each person had no strenuous exercise before the
test. The subjects were given at least five minutes before the signal extraction. During the measurement process,
the room needs to be quiet, and the results are recorded after the measurement data is stable. Then there will be a
comparison and analysis of the data between the system and the oximeter. The results of oxygen saturation and
pulse rate are shown in Fig.9 (a), (b). Among them, series 1 represents the data measured by the oxygen sensor,
and series 2 represents the data measured from the system.
(a) (b)
Fig.9 Results of blood oxygen saturation and pulse rate.
The error analysis of the results of blood oxygen saturation and pulse measurement is shown in Fig.10, in
which the series 1 and 2 represent the percentage error of pulse rate and oxygen saturation respectively. It can be
concluded that the maximum error is ±1%, and the maximum error of pulse rate is ±1.9%.The experimental
objects are selected scientifically, and the test results are representative. Although there are individual differences
in the oxygen saturation and pulse rate of the test object, the transmission stability of the reflective pulse oximetry
system can be guaranteed. The accuracy of oxygen saturation and pulse rate is: the error range of oxygen
saturation is ±1%, and the error range of pulse rate is ±2%.It can be seen that the reflective pulse oximetry
Design of the pulse oximetry measurement circuit and its sensing system based on CMOS
DOI: 10.9790/1676-1201046470 www.iosrjournals.org 70 | Page
sensing system has good stability and accuracy. The experimental results show that the accuracy and reliability of
the system meet the design requirements, and it is proved that the system can be used to monitor the pulse and
blood oxygen saturation.
Fig.10 the results of error analysis of the system
VII.Conclusion
The pulse oximetry circuit and its sensing system is designed based on the CMOS technology. The
principle of pulse oximetry circuit is analyzed, and the corresponding circuit is designed to realize the acquisition
and amplification of pulse oximeter signal. It can realize the function of real-time monitoring of pulse oximetry in
the middle and far distance with the combination of microprocessor and wireless sensor technology. The
experimental results show the feasibility of the proposed scheme, and the system has the advantages of stable
operation, high precision and low power consumption. With the development of science and technology and the
continuous improvement of the medical level, the characteristics of miniaturization and wireless monitoring have
a good prospect in the field of wearable medical research.
Acknowledgments
This research is funded by Tianjin Research Program of Application Foundation and Advanced Technology (No.
15JCYBJC16300), and Tianjin science and Technology Commissioner Project (No. 16JCTPJC45500).
References
[1] Y. Mendelson, Noninvasive pulse oximetry utilizing skin reflectance photo plethysmography. IEEE Transactions on Biomedical
Engineering, 35 (10), 1988, 798-805
[2] V. Kirk, S. Bohn, W. Flemons, and J. Rammers, Comparison of home oximetry monitoring with laboratory polysomnography in
children. Amer. College Chest Phys, 2003, 1702-1708
[3] S. Ley, D. Laqua, P. Husar, Simulation of Photon Propagation in Multi-layered Tissue for Non-invasive Fetal Pulse Oximetry, The
15th International Conference on Biomedical Engineering, 2013, 356-359
[4] S. Warren, J Lebak, J Yao, at al, Interoperability and security in wireless body area network infrastructures [C], Conf Proc IEEE Eng
Med Biol Soc, 2005, 837 -840
[5] T. Polk, W. Walker, A Hande, Wireless telemetry for oxygen saturation measurements [A] .IEEE Biomedical Circuits and Systems
(BiOCAS) Conference Proceedings [C],London , UK ,November , 2006 .
[6] S. Arms, C. Townsend, C. Churchill, ,Power management for energy h harvesting wireless sensors [A] .SPIE International
Symposium on Smart Structures and Smart Materials Proceedings [C], 2005.
[7] S.LEE, S.JOO, C.LEE, An enhanced dynamic framed slotted ALOHA algorithm for RFID tag[C].Proceedings of The Second Annual
International Conference on Mobile and Ubiquitous Systems: Networking and Services, Washington, DC: IEEE Computer Society,
2005,166-172
[8] Dirk Hihnel, Wolfram Burgard, Dieter Fox, et al, Mapping and Localization with RFID Technology[C].IEEE international
Conference on robotics on & automation, 2004, 1015-1020
[9] W. Choi and D I Oh. R-LIM, an Affordable Library Search System Based on RFID[C]. International Conference on Hybrid
Information Technology, Cheju Island, 2006, 103-108

More Related Content

Similar to Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS

Pulse_Oximeter_SoC
Pulse_Oximeter_SoCPulse_Oximeter_SoC
Pulse_Oximeter_SoC
Tianhao Li
 
PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1
Ahmed Ghouri
 
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGProblems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
IJSRD
 
Heart rate detection using hilbert transform
Heart rate detection using hilbert transformHeart rate detection using hilbert transform
Heart rate detection using hilbert transform
eSAT Journals
 

Similar to Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS (20)

Pulseoximetry
PulseoximetryPulseoximetry
Pulseoximetry
 
Pulse oximetry ppt
Pulse oximetry pptPulse oximetry ppt
Pulse oximetry ppt
 
Pulse_Oximeter_SoC
Pulse_Oximeter_SoCPulse_Oximeter_SoC
Pulse_Oximeter_SoC
 
PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1
 
FinalReport
FinalReportFinalReport
FinalReport
 
WorldComp2016_Jenny
WorldComp2016_JennyWorldComp2016_Jenny
WorldComp2016_Jenny
 
3a-Wearable sensors for physiological signal measurement.pdf
3a-Wearable sensors for physiological signal measurement.pdf3a-Wearable sensors for physiological signal measurement.pdf
3a-Wearable sensors for physiological signal measurement.pdf
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
 
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGProblems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
 
pulse oximetry final
pulse oximetry finalpulse oximetry final
pulse oximetry final
 
Wireless Body Area Networks for Healthcare: A Survey
Wireless Body Area Networks for Healthcare: A SurveyWireless Body Area Networks for Healthcare: A Survey
Wireless Body Area Networks for Healthcare: A Survey
 
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
 
Non invasive
Non invasiveNon invasive
Non invasive
 
Medical dissection lab pulse oximeter
Medical dissection lab pulse oximeterMedical dissection lab pulse oximeter
Medical dissection lab pulse oximeter
 
Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Heart rate detection using hilbert transform
Heart rate detection using hilbert transformHeart rate detection using hilbert transform
Heart rate detection using hilbert transform
 
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEWMeasurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
 
IJET-V3I1P26
IJET-V3I1P26IJET-V3I1P26
IJET-V3I1P26
 
Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...
 
PsterPresentation2016
PsterPresentation2016PsterPresentation2016
PsterPresentation2016
 

More from IOSRJEEE

Investigation of Power Quality Indices in Jordanian Distribution Grid
Investigation of Power Quality Indices in Jordanian Distribution GridInvestigation of Power Quality Indices in Jordanian Distribution Grid
Investigation of Power Quality Indices in Jordanian Distribution Grid
IOSRJEEE
 
Improving Distribution System Performance in Deregulated Electricity Industry...
Improving Distribution System Performance in Deregulated Electricity Industry...Improving Distribution System Performance in Deregulated Electricity Industry...
Improving Distribution System Performance in Deregulated Electricity Industry...
IOSRJEEE
 
Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...
Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...
Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...
IOSRJEEE
 
Magneto Optic Current Transformer Technology (MOCT)
Magneto Optic Current Transformer Technology (MOCT)Magneto Optic Current Transformer Technology (MOCT)
Magneto Optic Current Transformer Technology (MOCT)
IOSRJEEE
 
Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...
Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...
Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...
IOSRJEEE
 

More from IOSRJEEE (20)

GM Distributed Generation Inverters in a Micro grid by Controlling Energy Man...
GM Distributed Generation Inverters in a Micro grid by Controlling Energy Man...GM Distributed Generation Inverters in a Micro grid by Controlling Energy Man...
GM Distributed Generation Inverters in a Micro grid by Controlling Energy Man...
 
Load Performances of a PIC Based Speed Governor for Micro Hydro Power Plant
Load Performances of a PIC Based Speed Governor for Micro Hydro Power PlantLoad Performances of a PIC Based Speed Governor for Micro Hydro Power Plant
Load Performances of a PIC Based Speed Governor for Micro Hydro Power Plant
 
A New Design in Parabolic Trough Using Heat Retention
A New Design in Parabolic Trough Using Heat RetentionA New Design in Parabolic Trough Using Heat Retention
A New Design in Parabolic Trough Using Heat Retention
 
Investigation of Power Quality Indices in Jordanian Distribution Grid
Investigation of Power Quality Indices in Jordanian Distribution GridInvestigation of Power Quality Indices in Jordanian Distribution Grid
Investigation of Power Quality Indices in Jordanian Distribution Grid
 
Improving Distribution System Performance in Deregulated Electricity Industry...
Improving Distribution System Performance in Deregulated Electricity Industry...Improving Distribution System Performance in Deregulated Electricity Industry...
Improving Distribution System Performance in Deregulated Electricity Industry...
 
High Efficiency Dc-Dc Converter for Renewable Energy Applications and High Vo...
High Efficiency Dc-Dc Converter for Renewable Energy Applications and High Vo...High Efficiency Dc-Dc Converter for Renewable Energy Applications and High Vo...
High Efficiency Dc-Dc Converter for Renewable Energy Applications and High Vo...
 
Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...
Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...
Performance Indicators For Grid-Connected PV Systems: A Case Study In Casabla...
 
A Review of Gas Sensors Based on Semiconducting Metal Oxide
A Review of Gas Sensors Based on Semiconducting Metal OxideA Review of Gas Sensors Based on Semiconducting Metal Oxide
A Review of Gas Sensors Based on Semiconducting Metal Oxide
 
Stability Control System for a Two-Wheeler
Stability Control System for a Two-WheelerStability Control System for a Two-Wheeler
Stability Control System for a Two-Wheeler
 
Voltage Sag and Swell Identification Using FFT Analysis and Mitigation with DVR
Voltage Sag and Swell Identification Using FFT Analysis and Mitigation with DVRVoltage Sag and Swell Identification Using FFT Analysis and Mitigation with DVR
Voltage Sag and Swell Identification Using FFT Analysis and Mitigation with DVR
 
Design of Intrusion Tolerance System based on Service Redundancy Level
Design of Intrusion Tolerance System based on Service Redundancy LevelDesign of Intrusion Tolerance System based on Service Redundancy Level
Design of Intrusion Tolerance System based on Service Redundancy Level
 
Compensation of Harmonics of Fully Controlled Loads by Using SAHF
Compensation of Harmonics of Fully Controlled Loads by Using SAHFCompensation of Harmonics of Fully Controlled Loads by Using SAHF
Compensation of Harmonics of Fully Controlled Loads by Using SAHF
 
Design of Half Bridge LLC Resonant Converter for Low Voltage Dc Applications
Design of Half Bridge LLC Resonant Converter for Low Voltage Dc ApplicationsDesign of Half Bridge LLC Resonant Converter for Low Voltage Dc Applications
Design of Half Bridge LLC Resonant Converter for Low Voltage Dc Applications
 
Design of an Integrated Power Factor Converter with PI Controller for Low Pow...
Design of an Integrated Power Factor Converter with PI Controller for Low Pow...Design of an Integrated Power Factor Converter with PI Controller for Low Pow...
Design of an Integrated Power Factor Converter with PI Controller for Low Pow...
 
Magneto Optic Current Transformer Technology (MOCT)
Magneto Optic Current Transformer Technology (MOCT)Magneto Optic Current Transformer Technology (MOCT)
Magneto Optic Current Transformer Technology (MOCT)
 
On Essentiality of RoHS in EE R&D Ecosystem
On Essentiality of RoHS in EE R&D EcosystemOn Essentiality of RoHS in EE R&D Ecosystem
On Essentiality of RoHS in EE R&D Ecosystem
 
A high stable on-chip CMOS temperature sensor
A high stable on-chip CMOS temperature sensorA high stable on-chip CMOS temperature sensor
A high stable on-chip CMOS temperature sensor
 
Improvement of Load Power Factor by Using Capacitor
Improvement of Load Power Factor by Using CapacitorImprovement of Load Power Factor by Using Capacitor
Improvement of Load Power Factor by Using Capacitor
 
Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...
Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...
Modeling and Simulation of Three Phase Induction Machine Using Written Pole T...
 
Design of LED Driver Power Based on LNK417EG
Design of LED Driver Power Based on LNK417EGDesign of LED Driver Power Based on LNK417EG
Design of LED Driver Power Based on LNK417EG
 

Recently uploaded

Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
jaanualu31
 
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
HenryBriggs2
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdf
Kamal Acharya
 
Introduction to Robotics in Mechanical Engineering.pptx
Introduction to Robotics in Mechanical Engineering.pptxIntroduction to Robotics in Mechanical Engineering.pptx
Introduction to Robotics in Mechanical Engineering.pptx
hublikarsn
 

Recently uploaded (20)

Worksharing and 3D Modeling with Revit.pptx
Worksharing and 3D Modeling with Revit.pptxWorksharing and 3D Modeling with Revit.pptx
Worksharing and 3D Modeling with Revit.pptx
 
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
 
Memory Interfacing of 8086 with DMA 8257
Memory Interfacing of 8086 with DMA 8257Memory Interfacing of 8086 with DMA 8257
Memory Interfacing of 8086 with DMA 8257
 
Online food ordering system project report.pdf
Online food ordering system project report.pdfOnline food ordering system project report.pdf
Online food ordering system project report.pdf
 
PE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and propertiesPE 459 LECTURE 2- natural gas basic concepts and properties
PE 459 LECTURE 2- natural gas basic concepts and properties
 
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptxS1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
S1S2 B.Arch MGU - HOA1&2 Module 3 -Temple Architecture of Kerala.pptx
 
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
scipt v1.pptxcxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx...
 
Theory of Time 2024 (Universal Theory for Everything)
Theory of Time 2024 (Universal Theory for Everything)Theory of Time 2024 (Universal Theory for Everything)
Theory of Time 2024 (Universal Theory for Everything)
 
Introduction to Geographic Information Systems
Introduction to Geographic Information SystemsIntroduction to Geographic Information Systems
Introduction to Geographic Information Systems
 
Electromagnetic relays used for power system .pptx
Electromagnetic relays used for power system .pptxElectromagnetic relays used for power system .pptx
Electromagnetic relays used for power system .pptx
 
Signal Processing and Linear System Analysis
Signal Processing and Linear System AnalysisSignal Processing and Linear System Analysis
Signal Processing and Linear System Analysis
 
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...Max. shear stress theory-Maximum Shear Stress Theory ​  Maximum Distortional ...
Max. shear stress theory-Maximum Shear Stress Theory ​ Maximum Distortional ...
 
8th International Conference on Soft Computing, Mathematics and Control (SMC ...
8th International Conference on Soft Computing, Mathematics and Control (SMC ...8th International Conference on Soft Computing, Mathematics and Control (SMC ...
8th International Conference on Soft Computing, Mathematics and Control (SMC ...
 
fitting shop and tools used in fitting shop .ppt
fitting shop and tools used in fitting shop .pptfitting shop and tools used in fitting shop .ppt
fitting shop and tools used in fitting shop .ppt
 
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
 
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
Convergence of Robotics and Gen AI offers excellent opportunities for Entrepr...
 
Hostel management system project report..pdf
Hostel management system project report..pdfHostel management system project report..pdf
Hostel management system project report..pdf
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdf
 
Introduction to Robotics in Mechanical Engineering.pptx
Introduction to Robotics in Mechanical Engineering.pptxIntroduction to Robotics in Mechanical Engineering.pptx
Introduction to Robotics in Mechanical Engineering.pptx
 

Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS

  • 1. IOSR Journal of Electrical and Electronics Engineering (IOSR-JEEE) e-ISSN: 2278-1676,p-ISSN: 2320-3331, Volume 12, Issue 1 Ver. IV (Jan. – Feb. 2017), PP 64-70 www.iosrjournals.org DOI: 10.9790/1676-1201046470 www.iosrjournals.org 64 | Page Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS Yang Wang1 , Liying Chen2 1 (Tianjin Poly-technic University, School of Electronic and Information Engineering, 300387, China) 2 (Tianjin Poly-technic University, School of Electronic and Information Engineering, 300387, China) Abstract: The pulse oximetry circuit and its sensing system is designed based on the standard CMOS technology of 0.18um. The reflection oxygen sensor is used to collect the pulse oximeter signal of human body, then the collected physiological signals are processed by the data processing circuit The data processing circuit is composed of two parts: the amplifying circuit and the band-pass filter circuit, and the pulse oximeter data processed by the data processing circuit is written into the tag through the SPI communication The RFID reader read the data in the RFID tag through wireless communication, and display the data . The experimental results show that the maximum error is ±1%. The maximum error of the pulse is ±1.9%. The stability and feasibility of pulse blood oxygen sensing system is demonstrated in this paper and it will have a good application prospect in the direction of wearable medical wisdom research. Keywords: measurement circuit, pulse oximetry, wireless sensor networks, reflection type, active RFID tag I. Introduction With the aging of the society is becoming more and more serious, the cost of medical care for the elderly and the proportion of medical resources is increasing. In order to alleviate the pressure on the health care system, the traditional medical center model needs to be transformed into a family centered model. Wearable medical device is a device that integrates signal acquisition, processing and data transmission. The physiological parameters such as pulse, oxygen saturation and temperature were collected, processed and analyzed by the biomedical sensors on the human body. The data can be displayed timely feedback to the user in the device, and can also be used for real-time monitoring of users. It does effectively carry out the prevention and treatment of disease. At present, the most mature technology in the detection of blood oxygen saturation [1] in the human body is the transmission detection structure. However, the reflection type instrument has been paid more and more attention, because of the limitation of the current type of testing instruments in the acquisition of the human body. The light source and the light detector of the reflective pulse oxygen sensor is placed on the same side of the detection part. The detector receives the light reflected back to human tissue, and convert optical signals into electrical signal as the output of the sensor. The reflective optical detection mode is less limited by the detection position, so that the detection position is more flexible. The measurement can be achieved in despite of the limit of position. With the development of medical sensor technology, the transmission type detection method of pulse oximetry has been gradually replaced by reflective detection mode. It has become the main measurement method for detecting pulse oximetry, and received great attention and research in the field of medicine. In this paper, the reflective probe is used to collect the signal, and the measurement circuit is designed with UMC 180nm CMOS standard technology. After the amplification and filtering of the collected signal, the signal processed by the circuit will be transmitted further to other modules. II. Related Works The components of the absorption of light in the blood mainly include oxygenated hemoglobin (HbO2) and reduced hemoglobin (Hb). Blood oxygen saturation is defined as the percentage of HbO2 in the blood, which is the sum of the concentration of HbO2 and the concentration of Hb in the blood. The concentration of HbO2 is an important physiological parameter for the metabolism of human tissues. The definition of blood oxygen saturation: 2 2 2 100% HbO HbO Hb C SpO C C    (1) In the formula, Spo2 represents the value of oxygen saturation, CHbo2 represents the concentration of HbO2 in the blood, and CHb indicates the concentration of Hb in the blood. Bill Lambert's law states that light has certain attenuation characteristics in non-homogeneous scattering medium, the absorption degree of the light in human issues has certain difference in different medium. The measurement of oxygen saturation [2,3] is based on
  • 2. Design of the pulse oximetry measurement circuit and its sensing system based on CMOS DOI: 10.9790/1676-1201046470 www.iosrjournals.org 65 | Page Lambert Bill's law as the theoretical basis, the light passes through the medium of interference without considering the absorption of skin and muscle and other tissue to light condition. The blood HbO2 and Hb have different absorption rates of different wavelengths. Different wavelengths of light can emit the device as a light source. We can calculate the content of hemoglobin in blood components through the measurement of the attenuation degree after the light passing through the blood. The measurement of the pulse and blood oxygen saturation is based on the principle that the light absorption quantity will change with the expansion or contraction of the artery, while the light absorption of other organizations are almost constant. The amount of light absorbed by the arterial blood changes when the arterial pulse. The amount of light absorbed by the skin, muscles, bones, venous blood, and non-pulsating arteries is constant. Heart contraction or expansion will affect the arterial volume, according to Lambert Bill's law. Change of the optical path will affect the absorbance, ripple component of artery is PPG (pulse wave) signal. According to the photon diffusion equation and time resolved spectroscopy technique, the formula of the reflected light intensity can be obtained: 2 2 3 5 [(1 ) ] exp( )exp( ) 4( , ) (1 ) (4 ) s a s g ct DcR t g Dc t              (2) When the time is long enough, the formula (2) is approximately equal to ac ,W that represents the change of the reflected light intensity, is directly proportional to the absorption coefficient s . aW c (3) According to the Lambert Bill law, when a beam of light intensity of 0I is perpendicular to a sample with a concentration of C , the transmitted light intensity I and 0I exist. ( ) 0 CL I I e    (4) In the formula, ( )  is the absorption coefficient of the sample molecule to the wavelength  ; L is the path of light. In order to carry out the transformation, the absorption coefficient a is obtained. 0 1 ln( ) a I C L I     (5) Assuming that Hb and HbC are the absorption coefficient and concentration of Hb in blood, 2HbO and 2HbOC are respectively the absorption coefficient and concentration of HbO2 in blood. When the light irradiates on the human body with two kinds of wavelength of red and infrared light as light source, the absorption coefficient of HbO2 and Hb on the two wavelengths of light can type (5) respectively, only considering the effect of absorption of HbO2 in blood and Hb. 2 2 R R R a Hb Hb Hbo HboC C    (6) 2 2 IR IR IR a Hb Hb Hbo HboC C    (7) When light passes through human tissue and blood vessels, it can be divided into two parts: DC component and AC component. The change rate of light intensity in the tissue is /AC DCW I I binding (2), (6) and (7) According to the double beam method, the following formula will be derived. 2 2 2 2 R R Hb Hb Hbo HboR IR IR IR Hb Hb Hbo Hbo C CW W C C        (8) / / R R AC DCR IR IR IR AC DC I IW A W I I   (9) The formula (9) is brought into (1),
  • 3. Design of the pulse oximetry measurement circuit and its sensing system based on CMOS DOI: 10.9790/1676-1201046470 www.iosrjournals.org 66 | Page 2 2 2 R IR Hb Hb R R R R R Hb Hbo Hb Hbo IR W spo W           (10) In the formula, IR Hb R Hb and 2 R Hbo are all constant, so we can get the formula of blood oxygen saturation: 2 1 2spo k k A  (11) The reflective pulse oximetry can be calculated by the AC and DC information of the pulse wave signal. III. The Pulse Oximeter Measurement Circuit In this paper, the pulse oximetry sensor (DCM03) is used to detect the pulse oxygen saturation, which has the advantages of small volume, low power consumption and high precision. The key to measure the blood oxygen saturation is to get the pulse waveform. The pulse signal of human body is a physiological signal of low frequency and low amplitude, which is about 0.5mV. In theory, the whole measurement system needs to be enlarged more than 200 times to achieve the requirement of ADC conversion voltage. The pulse frequency is concentrated in 0.5Hz ~ 5Hz, and the maximum is less than 40Hz. In the process of signal acquisition, the movement of the body and mental stress will produce baseline drift and noise. At the same time, there will inevitably be power frequency interference and electromagnetic interference. This requires the suppression of the base drift circuit and the two filter circuit to filter the interference. The structure of the pulse oximeter circuit based on CMOS is shown in Fig.1. Preamplifiercircuit Suppressorcircuit LPF HPF twostage amplification Power management module(bandgap) input DC AC AC Fig.1 structure of the pulse oximeter circuit 3.1 amplifier circuit The pulse signal collected by the pulse sensor is very weak, the signal amplitude is 20mV ~ 50mV. In order to achieve the conversion voltage of ADC, the input signal of the circuit module should be increased by 50~80 times. In order to improve the anti-interference performance of the circuit (to avoid the distortion of the pulse signal caused by a large gain), the pulse oximeter measurement circuit is composed of preamplifier circuit and two-stage amplifier circuit. The preamplifier circuit is shown in Fig.2, the two-stage amplifier circuit is shown in Fig. 3. Among them, the preamplifier circuit uses differential amplifier. The differential amplifier circuit has a capability of strong common mode rejection (82.5dB), which can suppress the temperature drift, voltage drift and effectively offset the random noise. VCC V1 V2 Vb GND A1 A2 A3 C0 C1 R4 R5 R0 R1 R2 R3 R6 R7 Vout VCC Vb ENA Vin GND Vout A2R0 R1 R2 Fig.2 Preamplifier circuit Fig.3 two-stage amplifier circuit
  • 4. Design of the pulse oximetry measurement circuit and its sensing system based on CMOS DOI: 10.9790/1676-1201046470 www.iosrjournals.org 67 | Page 3.2 band pass filter After the signal is amplified, it needs to filter the interference. The band-pass filter circuit is designed to eliminate the interference and noise, including low-pass filter and high-pass filter. Low pass filter circuit uses two-order Butterworth low-pass filter circuit. This type of active circuit can not only remove the interference effectively, but also reduce the signal attenuation. At the same time, the signal can be amplified to some extent. The pulse signal after the low-pass filter circuit is composed of DC signal and AC signal. Among them, the DC signal is the volt level, while the AC signal is millivolt level. Therefore, it is necessary to amplify the AC signal individually. In this way, the AC and DC signals should be isolated, and then the AC signal can be amplified, some low-frequency interference can also be filtered out. Finally, the AC signal is fed into the two stage amplifier to obtain the value of ACI 、 DCI . IV. System Design The pulse oximetry sensor system based on CMOS is composed of four modules: the reflective pulse oximetry module, data processing module, RFID transceiver module and the PC interface. In the process of data transmission, this system involves two kinds of communication modes including wired communication and wireless communication [4]. The communication between sensor and MCU, MCU and the tag, the reader and PC are all through the wired communication mode. The transmission of sensor data between RFID tag and RFID reader is accomplished by wireless communication. The data processing module will not only process the pulse and blood oxygen information [5, 6] collected by the sensor, but also establish a connection with the RFID tag. At the same time, the pulse and blood oxygen data collected and processed are transferred to the register of the RFID tag. The data transmission between the sensor and the data processing module is through the serial port connection, and the data transmission between the data processing module and the RFID tag is communicated through the SPI interface. RFID transceiver module includes the RFID active tags and the RFID reader. The RFID tag [7, 8] and the reader achieve the data transmission based on GEN IS0/IEC 18000-6C protocol. Based on the communication between the data processing modules, the whole communication process: MCU transmits the data in the register to the RFID tag through the SPI interface. The reader communicates with the tag in accordance with the radio frequency communication protocol and reads the data in the internal register of the RFID tag. Finally the RFID reader read the data in the RFID tag through wireless communication [9], and display the data on PC interface. The overall design structure diagram of pulse oximetry system is shown in Fig.4. sensor MCU tag reader PC UART SPI Pulse oximetry detection module Data processing module Data will be written into the tag through SPI The data will be transmitted through wireless communication Display interface light Detecor Fig.4 the structure diagram of pulse oximetry system V. Process of The Data Transmission After the reader and tag are connected, the pulse oximeter system will be tested. The pulse oximeter sensor is connected with the host computer through the USB interface, which can directly display the red light and infrared light PPG signal measured by the sensor. According to the principle of dual wavelength detection, the PPG signal can be transformed into digital signal, which can be used for further transmission of the data. The PPG signal waveform of the sensor is shown in Fig.5.
  • 5. Design of the pulse oximetry measurement circuit and its sensing system based on CMOS DOI: 10.9790/1676-1201046470 www.iosrjournals.org 68 | Page Fig. 5 waveform of PPG signal collected by sensor In the pulse oximetry system, the pulse oximeter is connected with the data processing module through the UART interface. The data processing module is connected with the RFID tag through the SPI interface. Then the data in the register of data processing module is transferred to the internal register of the tag. The RFID reader and RFID tags constitute the RFID transceiver module, it works in the 860-960MHz frequency range. The data transmission is based on the EPC C1 GEN2/ISO 18000-6C protocol. The reader interface is shown in Fig.6, and the reader starts working after clicking on the connection button. Click on the label and open a single label cycle recognition mode, the reader will identify the tag within the working range. The data window displays the label number, storage area and storage address of the tag identified by the reader. Fig. 6 the reader interface The data read by the reader is shown in Fig.6, the contents of the data window in the graph are the tag information identified by the reader and the data in each storage area. The reader is connected with the host computer through the USB interface, which can display the value and waveform of the pulse and oxygen saturation in the computer interface. The pulse oximeter information displayed on the computer interface is shown in Fig.7.
  • 6. Design of the pulse oximetry measurement circuit and its sensing system based on CMOS DOI: 10.9790/1676-1201046470 www.iosrjournals.org 69 | Page Fig. 7 the information displayed in the computer interface Fig. 8 Pulse oximetry data Click on the record, the pulse and oxygen saturation data received by the host computer can be saved in the form of text. The real time data detected by the pulse oximeter system is shown in Fig.8. Users can check their information on a regular basis. By analyzing and processing the real time data of pulse and blood oxygen saturation, we can know the health status and the development trend. It can play a good role in monitoring the health of the user. VI. Experimental Validation After the system hardware and software debugging, the finger probe BSJ09001C of Shanghai Berry Company was chosen in order to further verify the feasibility and accuracy of the system. Then the measured data are compared with the data obtained by the system. The measurement range of oxygen saturation is 35%-100%, and the measurement error is ±1%. The pulse rate measurement range is 25-250bpm, the measurement error is ± 2%. 50 people of different ages were selected for the test, and each person had no strenuous exercise before the test. The subjects were given at least five minutes before the signal extraction. During the measurement process, the room needs to be quiet, and the results are recorded after the measurement data is stable. Then there will be a comparison and analysis of the data between the system and the oximeter. The results of oxygen saturation and pulse rate are shown in Fig.9 (a), (b). Among them, series 1 represents the data measured by the oxygen sensor, and series 2 represents the data measured from the system. (a) (b) Fig.9 Results of blood oxygen saturation and pulse rate. The error analysis of the results of blood oxygen saturation and pulse measurement is shown in Fig.10, in which the series 1 and 2 represent the percentage error of pulse rate and oxygen saturation respectively. It can be concluded that the maximum error is ±1%, and the maximum error of pulse rate is ±1.9%.The experimental objects are selected scientifically, and the test results are representative. Although there are individual differences in the oxygen saturation and pulse rate of the test object, the transmission stability of the reflective pulse oximetry system can be guaranteed. The accuracy of oxygen saturation and pulse rate is: the error range of oxygen saturation is ±1%, and the error range of pulse rate is ±2%.It can be seen that the reflective pulse oximetry
  • 7. Design of the pulse oximetry measurement circuit and its sensing system based on CMOS DOI: 10.9790/1676-1201046470 www.iosrjournals.org 70 | Page sensing system has good stability and accuracy. The experimental results show that the accuracy and reliability of the system meet the design requirements, and it is proved that the system can be used to monitor the pulse and blood oxygen saturation. Fig.10 the results of error analysis of the system VII.Conclusion The pulse oximetry circuit and its sensing system is designed based on the CMOS technology. The principle of pulse oximetry circuit is analyzed, and the corresponding circuit is designed to realize the acquisition and amplification of pulse oximeter signal. It can realize the function of real-time monitoring of pulse oximetry in the middle and far distance with the combination of microprocessor and wireless sensor technology. The experimental results show the feasibility of the proposed scheme, and the system has the advantages of stable operation, high precision and low power consumption. With the development of science and technology and the continuous improvement of the medical level, the characteristics of miniaturization and wireless monitoring have a good prospect in the field of wearable medical research. Acknowledgments This research is funded by Tianjin Research Program of Application Foundation and Advanced Technology (No. 15JCYBJC16300), and Tianjin science and Technology Commissioner Project (No. 16JCTPJC45500). References [1] Y. Mendelson, Noninvasive pulse oximetry utilizing skin reflectance photo plethysmography. IEEE Transactions on Biomedical Engineering, 35 (10), 1988, 798-805 [2] V. Kirk, S. Bohn, W. Flemons, and J. Rammers, Comparison of home oximetry monitoring with laboratory polysomnography in children. Amer. College Chest Phys, 2003, 1702-1708 [3] S. Ley, D. Laqua, P. Husar, Simulation of Photon Propagation in Multi-layered Tissue for Non-invasive Fetal Pulse Oximetry, The 15th International Conference on Biomedical Engineering, 2013, 356-359 [4] S. Warren, J Lebak, J Yao, at al, Interoperability and security in wireless body area network infrastructures [C], Conf Proc IEEE Eng Med Biol Soc, 2005, 837 -840 [5] T. Polk, W. Walker, A Hande, Wireless telemetry for oxygen saturation measurements [A] .IEEE Biomedical Circuits and Systems (BiOCAS) Conference Proceedings [C],London , UK ,November , 2006 . [6] S. Arms, C. Townsend, C. Churchill, ,Power management for energy h harvesting wireless sensors [A] .SPIE International Symposium on Smart Structures and Smart Materials Proceedings [C], 2005. [7] S.LEE, S.JOO, C.LEE, An enhanced dynamic framed slotted ALOHA algorithm for RFID tag[C].Proceedings of The Second Annual International Conference on Mobile and Ubiquitous Systems: Networking and Services, Washington, DC: IEEE Computer Society, 2005,166-172 [8] Dirk Hihnel, Wolfram Burgard, Dieter Fox, et al, Mapping and Localization with RFID Technology[C].IEEE international Conference on robotics on & automation, 2004, 1015-1020 [9] W. Choi and D I Oh. R-LIM, an Affordable Library Search System Based on RFID[C]. International Conference on Hybrid Information Technology, Cheju Island, 2006, 103-108