18. COLORIMETER Vs UV-Vis Spectrophotometer
COLORIMETER - Only colored solutions can be
analyzed
UV-Vis spectroscopy - Both colored and
colorless solutions can be analyzed
19.
20.
21. ELECTRONIC TRANSITION
Electronic transitions take place when electrons
in a molecule are excited from one energy level to a
higher energy level.
The energy change provides information on the
structure of a molecule and determines many
molecular properties
35. DEFINITION
The Beer-Lambert law states that the
quantity of light absorbed by a
substance dissolved in a fully
transmitting solvent is directly
proportional to the concentration of the
substance and the path length of the
light through the solution.
37. A = εcl
A = Absorbance
ε = Molar absorptivity or Molar extinction
coefficient or Molar absorption coefficient
C = Concentration
l = Path length (cm)
38. DERIVATION OF LAW
- A spectrophotometer is an apparatus that measures the intensity, energy
carried by the radiation per unit area per unit time, of the light entering a
sample solution and the light going out of a sample solution.
- The two intensities can be expressed as transmittance: the ratio of the
intensity of the exiting light to the entering light or percent transmittance
(%T).
- Different substances absorb different wavelengths of light. Therefore, the
wavelength of maximum absorption by a substance is one of the
characteristic properties of that material.
- A completely transparent substance will have It = I0 and its percent transmittance will
be 100.
- Similarly, a substance which allows no radiation of a particular wavelength to pass
through it will have It = 0, and a corresponding percent transmittance of 0.
39. Deviations to the law
The Beer-Lambert law maintains linearity under specific
conditions only.
The law will make inaccurate measurements at high
concentrations because the molecules of the analyte
exhibit stronger intermolecular and electrostatics
interactions which is due to the lesser amount of space
between molecules.
This can change the molar absorptivity of the analyte.
40. Applications
Generally, it can be used to
determine concentrations of a
particular substance, or determine the
molar absorptivity of a substance.
41. UV-Vis spectrophotometer
1. Single beam UV-Vis spectrophotometer
(only sample can be placed)
2. Double beam UV-Vis spectrophotometer
(Both sample & reference can be placed)
49. Light source
1. Halogen lamp
2. Deuterium lamp
3. Xenon lamp
4. Xenon flash lamp
5. Low pressure mercury lamp
Many spectrophotometers uses a halogen lamp for
the visible range and a deuterium lamp for the
ultraviolet range according to the wavelength
requirement.
50. Collimator
Also called collimating lens
A collimator is a device which narrows a beam of particles or
waves.
To narrow can mean either to cause the directions of motion to
become more aligned in a specific direction (i.e., make collimated
light or parallel rays)
84. Spectrophometric titration -
During a spectrophotometric titration, we
monitor changes in absorbance to
determine the equivalence point.
The end point is where the graph is
discontineous.